Kobold KMT

Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

measuring

Thermal •

Mass Flow Meter monitoring



for Compressed Air and Gases analysing

KMT

OO Measuring range: 0.32 ... 63 Nm³/h - 263.4 ... 263 350 Nm³/h


OOAccuracy: ± 1.5% of reading + 0.5% of full scale (≤ DN 50)
or ± 1.5% of reading + 0.8% of full scale (DN 65 ... DN 700)
OO pmax: 16 bar; tmax: 80 °C
OO Connection: R½" ... R 2" ball valve (≤ DN 50) or R½" male
for insertion (DN 65 ... DN 700)
OO Analogue output: 0 - 10 V or 0(4) ... 20 mA
OO Switching /pulse output
OO Option: Digital display
OO Non-return protection device for secure mounting
(DN 65 ... DN 700)

S5
OOMounting and demounting under pressure without
process interruption (with KMT-4)
OO Simple and exact positioning
OO Installation under pressure (with KMT-4)

KOBOLD companies worldwide: KOBOLD Messring GmbH


Nordring 22-24
ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, CHILE, CHINA, COLoMBIA, D-65719 Hofheim/Ts.
CZECHIA, EGYPT, FRANCE, GERMANY, GREAT BRITAIN, HUNGARY, INDIA, INDONESIA, ITALY, Head Office:
MALAYSIA, MEXICO, NETHERLANDS, PERU, POLAND, Republic of Korea, ROMANIA, +49(0)6192 299-0
+49(0)6192 23398
SINGAPORE, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, turkey, USA, VIETNAM [email protected]
2/09- 2014

www.kobold.com

1
Thermal Mass Flow Meter Model KMT

Description •• Non-return protection


The flowmeter of the series KMT, based on the measure- The sensor can only be pushed in one direction during
ment principle of thermal mass flow, is ideally suited for the installation. The sensor cannot return at all, even if it is
measurement of flow of compressed air and gases in released.
pipes with sizes from DN 15 ... DN 700. Measurement of for •• Seal
instance the usage of compressed air, nitrogen, CO2 or other By means of an encapsulated O-ring, no compressed air
non-corrosive gases. can escape under pressure during assembly.
The KMT is setting new standards in terms of measurement •• Precise positioning
accuracy and reproducibility thanks to its application specific The precise positioning with respect to immersion depth
adjustment during production. This flow meter is adjusted and orientation is easy to perform, guaranteeing accurate
under a pressure of 7 bar (abs) (DN 15 ... DN 50) or 9 bar measurement results.
(abs) (DN 65 ... DN 700). Adjusting the device specifically for
its application has the advantage of keeping the actual flow
speed in the pipeline low even with very large flow quantities. For optimum adaptation to different measurement tasks,
Thanks to the more stable flow profile, this low flow speed you can choose between two measuring ranges 0.2 ... 100
facilitates a much better degree of reproducibility and or 0.2 ... 200 Nm/s and three different probe lengths with a
accuracy than if the device were adjusted conventionally un- maximum immersion depth of 165/315/465 mm. The inner
der normal pressure, as flow speeds up to 200 Nm/s can diameter of the distribution pipe which is measured can be
often no longer be controlled under conventional adjustment entered via the USB port and the included configuration soft-
pressures. ware for pipe sizes from DN 65 to DN 300.
The core design of the flow meter is based on the hot An optional tapping sleeve for KMT-4 allows the subsequent
film sensor element, which is produced using the most assembly of the sensor into existing pipelines, and this
modern thin film technology and has already proven itself without interrupting the supply systems.
time and time again in the automotive industry. This flow
sensor features excellent long-term stability, a fast response Areas of Application
time and an extremely high degree of reliability. OO Measurement of consumption of compressed air
Two outputs are available, for further processing of the OO Compressed air counter
measurement data. Depending on the application, these OO Mass flow measurement of industrial gases
outputs can be configured as analogue (current or voltage),
switch output or as pulse output for the measurement of the
Design KMT-1/2/3
consumption.
1 Transmitter
The KMT has an integrated counter for consumption. The
consumed amount is shown on the display and the saved 1 a Enclosure with signal conditioning and optional display
value is not lost even after power outage. The availability of the 1 b Measurement probe with sensor and measurement
consumption amount as a free configurable pulse output is electronics
another helpful feature. 2 Mounting ball valve

Functions
The flow meter KMT consist of the transmitter and the 1a
mounting valve (only for KMT-1/2/3). The transmitter is
modular and consist of the probe and the signal conditio- 1b
ner. The measurement probe contains the sensor element
and the measurement electronics, in which the data of the
factory calibration is stored. The enclosure with the signal
conditioning is mounted either on the measurement probe
2
(compact) or is remote with a sensor cable up to 10 meter
(33 feet). The mounting valve assembly for KMT-1/2/3 allows
for the easy and reliable installation within the pipeline. The
high measurement accuracy is guaranteed by the accurate,
reproducible positioning of the probe within the mounting
valve. (See design KMT-1/2/3). The ball valve assembly allows for the exact alignment of the
sensing head within seconds during instalment and removal,
with only interrupting the process flow for a short moment.
Non-return protection for secure mounting for KMT-4
The ball valve assembly is suitable for pressures up to 16 bar
The non-return protection combines three functions in one (PN  16) and available for pipe diameters DN  15 (½") to
device: DN 50 (2"). During installation in the pipeline, observe the
required inlet and outlet paths as given in the operating
instructions.
 2/09- 2014

2 www.kobold.com No responsibility taken for errors;


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Design KMT-4 Configuration software


The evaluation electronics communicates digitally with the
probe and can be located up to 10 m (32.8 ft) from the probe.

2
1
max. 10m

probe cable
Fühlerkabel
flow sensor
Strömungssensor
non-return protection
Rückschlagschutz

mounting grip
Montagegriff
evaluation unit
Auswerteeinheit

Assembly KMT-4 (DN 65 ... DN 300) The flowmeter can be configured conveniently, to meet


With the right accessories, the KMT-4 flow meter can be the requirements of the application with the standard
easily integrated into any measurement task. configuration software and the integrated USB interface.
An assembly without welding and drilling into the pressurised Functionality
supply line and without flow interruption can be implemented
• Configuration of the output (scale / set point)
very easily with the tapping sleeve.
• Setting the pipe diameter
An optional ½" ball valve on the tapping sleeve enables the
installation and removal of the sensor without interrupting • 2-point user calibration for flow and temperature
the flow in the compressed air line. The ball valve on the • Readout of the counter values
tapping sleeve closes the measuring point pressure-tight after • Reset of min / max values and counter
removing the flow meter. Regular calibration, without taking • Indication of the measurement value
into account the device downtime, is therefore always an
option.

Measurement of consumption (totalizer)


The KMT holds an integrated counter for the usage. The
amount is stored and the data will not be lost due to a power
outage. The availability of the consumption amount as a free
configurable pulse output is another helpful feature.
2/09 - 2014

No responsibility taken for errors; www.kobold.com 3


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Technical Data
Measured flow: Volumetric flow at standard Accuracy (in air at 9 bar
conditions acc. DIN 1343 (abs) (and 23 °C for KMT-4
P0 = 1013.25 mbar; i.e. DN 65 ... DN 300)*: ± 1.5 % of reading
t0 = 0 °C (273.15 K) + 0.8 % of full scale
Temperature coefficient: ± 0.1 % of reading  / °C
Measuring range KMT-x1... KMT-x2... Pressure coefficient**: + 0.5 % / bar
Response time t90: < 1 s
DN 15 0.32 ... 63 Nm³/h 0.32 ... 126 Nm³/h
Sample rate: 0.5 s
DN 20 0.57 ... 113 Nm³/h 0.57 ... 226 Nm³/h Measuring range: -20 ... 80 °C
Standardised DN 25 0.90 ... 176 Nm³/h 0.90 ... 352 Nm³/h Accuracy (at 20 °C): ± 0.7 °C
volumetric flow Input: Optional pressure compensation
(air) DN 32 1.45 ... 289 Nm³/h 1.45 ... 578 Nm³/h 4 - 20 mA (2-wire; 14.2 ... 16 VDC)
DN 40 2.26 ... 452 Nm³/h 2.26 ... 904 Nm³/h for pressure sensor
Outputs: Output signal and display
DN 50 3.50 ... 700 Nm³/h 3.50 ... 1400 Nm³/h ranges are freely scalable
Standardised flow Analogue output
(air, nitrogen, CO2, ≤DN 50 0.5 ... 100 Nm/s 0.5 ... 200 Nm/s Voltage: 0 - 10 V max. 1 mA
argon) Current (3-wire): 0 - 20 mA and
4 - 20 mA RL < 500 Ω
Standardised flow
(Helium)
≤DN 50 0.5 ... 100 Nm/s 0.5 ... 120 Nm/s Switching output: Potential-free max. 44 VDC,
500 mA switching capacity
Standardised flow Pulse output: Totaliser,
≤DN 25 0.5 ... 100 Nm/s 0.5 ... 200 Nm/s
(Oxygen) pulse length: 0.02 ... 2 sec.
Digital interface: USB (for configuration)
Flow measuring range in dependence on pipe Electrical connection: Cable entry M 16 x 1,5
diameter (KMT-4) Cable power supply: 18 - 30 VAC/DC
Current consumption: Max. 200 mA (with display)
Inner Measuring range
Pipe Temperature range
Ø KMT-41 KMT-42 Ambient temperature: -20 ... 60 °C (-4 ... 140 °F)
inch mm 0.2 ... 100 Nm/s 0.2 ... 200 Nm/s Medium temperature: -20 ... 80 °C (-4 ... 176 °F)
Storage temperature: -20 ... 60 °C (-4 ... 140 °F)
DN 65 2 ½" 70.3 2.8 ...1397 Nm³/h 2.8 ... 2793 Nm³/h Nominal pressure: PN16 (232 PSI)
DN 80 3" 82.5 3.8 ...1923 Nm³/h 3.8 ... 3847 Nm³/h Humidity: No condensation
Medium: Compressed air or non
DN 100 4" 107.1 6.5 ... 3242 Nm³/h 6.5 ... 6483 Nm³/h
corrosive gases
DN 125 5" 131.7 9.8 ... 4902 Nm³/h 9.8 ... 9803 Nm³/h Display: 2 lines LC-Display, backlighting
Electromagnetic
DN 150 6" 159.3 14.3 ...7171 Nm³/h 14.3 ...14 343 Nm³/h
compatibility: EN61326-1 EN61326-2-3
DN 200 8" 206.5 24.1...12 051 Nm³/h 24.1... 24 101 Nm³/h Industrial Environment
Material
DN 250 10" 260.4 38.3 ...19 163 Nm³/h 38.3 ... 38 325 Nm³/h
Housing: Metal (AlSi3Cu)
DN 300 12" 309.7 54.2 ... 27 105 Nm³/h 57.8 ... 57 840 Nm³/h Probe: Stainless steel
Sensor head: Stainless steel / glass
DN 350 14" 339.6 65.2...32 591 Nm3/h 65.2...65 183 Nm3/h Ball valve
DN 400 16" 388.8 85.4...42 719 Nm3/h 85.4...85 438 Nm3/h (KMT-1/2/3): Brass
Non-return protection
DN 500 20" 486 133.5...66 749 Nm3/h 133.5...133 498 Nm3/h (KMT-4): Brass
Housing protection class: IP 65 / Nema 4
DN 600 24" 585 193.4...96 712 Nm3/h 193.4...193 425 Nm3/h
* The accuracy statement includes the uncertainty of the factory calibration
DN 700 28" 682.6 263.4...131 675 Nm3/h 263.4...263 350 Nm3/h with an enhancement factor k = 2 (2-times standard deviation). The
accuracy was culated in accordance with EA-4/02 and with regard to GUM
(Guide to the Expression of Uncertainty in Measurement).
Accuracy (in air at 7bar
**The pressure dependence is + 0.5 % / bar. The KMT is calibrated at 7 bar
(abs) (and 23 °C for (abs). Thus the error at 7 bar = 0 (e. g. additional error at 10 bar = +1.5 %
KMT-1/2/3 i.e. ≤ DN 50)*: ± 1.5 % of reading of reading. This error can be corrected by entering the actual system
+ 0.5 % of full scale pressure (with the configuration software).
 2/09- 2014

4 www.kobold.com No responsibility taken for errors;


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

The following gases can be measured by the


flowmeter KMT
OO Air
OO Nitrogen
OO Carbon dioxide
OO Helium
OO Argon
OO Oxygen

Use only oil and fat free units for medium oxygen and follow
the corresponding safety regulations.

Flow measuring range in dependence on working


pressure (≤ DN 50)
Nm/s

working pressure [bar]

air, nitrogen, oxygen, argon


CO2
helium

Formula to calculate the standard volumetric flow


.
V0 = V0 x id2 x π/4 x 3600
.
V0 = standardised volumetric flow [m³/h]
V0 = standardised flow [m/s]
id = inner pipe diameter [m]
π = 3.1415

Connection Diagram

P +15V 7
Pressuresensor
pressure sensor input
input
I
4...20mA
mA_Input 8
V
mA OUT 1-2 9
analogue* or switching output
Signal output 10
OUT 1-1 Signal-Output
Supply
signal output
voltage
OUT 2-1 11
switching or pulse output
OUT 2-2 12
Vcc 13
18...30V AC/DC ≅ 14
Supplyvoltage
Supply
supply voltage
voltage
GND

* With analogue output OUT 1-1 is connected with GND.


Switching and pulse output are potential-free
2/09 - 2014

No responsibility taken for errors; www.kobold.com 5


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Order Details (Example: KMT-1 14 R 0 0 L 1 N Q 1)

Cable length
Measuring range /
Model Connection Display Sensor /
Installation length (only KMT-4)
Electronic
14 = 0.32 ... 63 Nm³/h for pipe DN 15 (½")
24 = 0.32 ... 126 Nm³/h for pipe DN 15 (½")
KMT-1...
Sensor compact, 15 = 0.57 ... 113 Nm³/h for pipe DN 20 (¾")
direction of flow 25 = 0.57 ... 226 Nm³/h for pipe DN 20 (¾")
from right to left R = thread-ball
16 = 0.90 ... 176 Nm³/h for pipe DN 25 (1")
KMT-2... valve with
Sensor compact, 26 = 0.90 ... 352 Nm³/h for pipe DN 25 (1") G thread
direction of flow 17 = 1.45 ... 289 Nm³/h for pipe DN 32 (1 ¼") N3) = thread-ball
from left to right valve with
27 = 1.45 ... 578 Nm³/h for pipe DN 32 (1 ¼")
KMT-3... NPT thread
remote probe, 18 = 2.26 ... 452 Nm³/h for pipe DN 40 (1 ½")
according to 28 = 2.26 ... 904 Nm³/h for pipe DN 40 (1 ½")
installation (≤DN50)
19 = 3.50... 700 Nm³/h for pipe DN 50 (2")
29 = 3.50 ... 1400 Nm³/h for pipe DN 50 (2")
(see technical details for flow measuring
range in dependence on pipe diameter)
10 = 2.8 ... 1397 Nm³/h for pipe DN 65 (2 ½") / 165 mm
20 = 2.8 ... 2793 Nm³/h for pipe DN 65 (2 ½") / 165 mm 0 = without
1B = 3.8 ... 1923 Nm³/h for pipe DN 80 (3") / 165 mm 21) = 2 m with
2B = 3.8 ... 3847 Nm³/h for pipe DN 80 (3") / 165 mm plug M 12,
1C = 6.5 ... 3242 Nm³/h for pipe DN 100 (4") / 165 mm 4 pin
0 = without
2C = 6.5 ... 6483 Nm³/h for pipe DN 100 (4") / 165 mm Display 51) = 5 m with
plug M 12,
1D = 9.8 ... 4902 Nm³/h for pipe DN 125 (5") / 315 mm 1 = LCD-Display
4 pin
2D = 9.8 ... 9803 Nm³/h for pipe DN 125 (5") / 315 mm
Z1) = 10 m with
1E = 14.3 ... 7171 Nm³/h for pipe DN 150 (6") / 315 mm plug M 12,
2E = 14.3 ... 14 343 Nm³/h for pipe DN 150 (6") / 315 mm 4 pin
1F = 24.1 ... 12 051 Nm³/h for pipe DN 200 (8") / 315 mm
KMT-4 F = R½" male at
2F = 24.1 ... 24 101 Nm³/h for pipe DN 200 (8") / 315 mm
remote probe immersion
(DN65 ... DN700) 1G = 38.3 ... 19 163 Nm³/h for pipe DN 250 (10") / 315 mm
probe
2G = 38.3 ... 38 325 Nm³/h for pipe DN 250 (10") / 315 mm
1H = 54.2 ... 27 105 Nm³/h for pipe DN 300 (12") / 315 mm
2H = 54.2 ... 54 211 Nm³/h for pipe DN 300 (12") / 315 mm
1J = 65.2 ... 32 591 Nm³/h for pipe DN 350 (14") / 465 mm
2J = 65.2 ... 65 183 Nm³/h for pipe DN 350 (14") / 465 mm
1K = 85.4 ... 42 719 Nm³/h for pipe DN 400 (16") / 465 mm
2K = 85.4 ... 85 438 Nm³/h for pipe DN 400 (16") / 465 mm
1L = 133.5 ... 66 749 Nm³/h for pipe DN 500 (20") / 465 mm
2L = 133.5 ... 133 498 Nm³/h for pipe DN 500 (20") / 465 mm
1M = 193.4 ... 96 712 Nm³/h for pipe DN 600 (24") / 465 mm
2M = 193.4 ... 193 425 Nm³/h for pipe DN 600 (24") / 465 mm
1N = 263.4 ... 131 675 Nm³/h for pipe DN 700 (28") / 465 mm
2N = 263.4 ... 263 350 Nm³/h for pipe DN 700 (28") / 465 mm
Order Details (continued) next page
 2/09- 2014

6 www.kobold.com No responsibility taken for errors;


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Order Details (continued)

Physical Size Physical Size Output 1 / 


Medium Unit
Output 1 Output 2 Output 2
2 = switching-/counting pulse output
L = air 3 = analogue output 0-10 V/
N = nitrogen N = standard volume Q = consumption [Nm³] counting pulse output
C = CO2 flow [Nm³/h] (Standard setting)
4 = analogue output 4-20 mA /
H = helium 1 = SI units (Standard setting) counting pulse output (Standard)
A = argon 2 = US units (e. g. T = temperature [°C]
N = standard volume 1 = 2 x Switching output
S2) = oxygen (only SCFM, SFPM) M = mass flow [kg/h]
V = standard flow flow [Nm³/h] 7 = analogue output 0-10 V /
up to DN25)
[Nm/s] T = temperature [°C] switching output
Y = other gases
(on request) M = mass flow [kg/h] 8 = analogue output 4-20 mA /
V = standard flow [Nm/s] switching output
1)
Only for KMT-3... and KMT-4 ...
Sensor head and ball valve (wetted parts) are oil-and grease-free. Warning: only oil-and grease-free cleaned devices may be used for oxygen
2)

Not possible with ½" and 1¼


3)

Order Details Replacement Sensor (Example: ERS-KMT-S 1 1 4 K)

Measuring sec-
Measuring
Model Design tion Mounting
range
pipe diameter
4 = DN 15
5 = DN 20
6 = DN 25
K = for ball valve
7 = DN 32
8 = DN 40
9 = DN 50
0 = DN 65
B = DN 80
1 = sensor compact (direction of flow right to left)
C = DN 100
2 = sensor compact (direction of flow left to right) 1 = low
ERS-KMT-S D = DN 125
3 = remote probe (≤ DN 50) 2 = high
E = DN 150
4 = remote probe (DN 65 ... DN 700)
F = DN 200
F = R½" male at
G = DN 250 immersion
probe
H = DN 300
J = DN 350
K = DN 400
L = DN 500
M = DN 600
N = DN 700
2/09 - 2014

No responsibility taken for errors; www.kobold.com 7


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Order Details Replacement Sensor Cable (Version KMT-3 / 4) (Example: ERS-KMT-K 2)

Model

ERS-KMT-K 2 = 2 m with plug M12, 4-pin


ERS-KMT-K 5 = 5 m with plug M12, 4-pin
ERS-KMT-K Z = 10 m with plug M12, 4-pin

Order Details Accessories (for KMT-4)

Model Description Picture / Drawing

ERS-KMT-AS65 tapping sleeve DN 65


ERS-KMT-AS80 tapping sleeve DN 80
ERS-KMT-AS1H tapping sleeve DN 100
ERS-KMT-AS1Z tapping sleeve DN 125
ERS-KMT-AS1F tapping sleeve DN 150
ERS-KMT-AS2H tapping sleeve DN 200
ERS-KMT-AS2F tapping sleeve DN 250
ERS-KMT-AS3H tapping sleeve DN 300

25

11

welding nipple (stainless steel 1.4301)to be


ERS-KMT-AN
weld at the pipe
Ø 26,67

Ø 18,63
G1/2

G3/4

70

ball valve R½" (brass) mounting and de-


51,5

ERS-KMT-KH mounting under pressure without process


G1/2“

interruption
G3/4“
15

64,5

70

ball valve ½" (brass) for parallel measure-


54

ERS-KMT-KP
G3/4“
G1/2“

ment of pressure or dew point


54,5

Lateral fitting R p1/4“


64,5

16
14

adapter Rp ½" (brass) female to ½" NPT


ERS-KMT-AR15
R p1/2“
NPT1/2“

SW24
Ø 15

male for process connection


20 19
39
 2/09- 2014

8 www.kobold.com No responsibility taken for errors;


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Dimensions [mm] (for KMT-1/2/3 i.e. ≤ DN 50)


Compact KMT-1..., KMT-2...

KMT-1...

v0: 5.43 m/s


T: 23.75 °C

direction of flow is right to left

KMT-2...

v0: 5.43 m/s


T: 23.75 °C

direction of flow is left to right

Remote probe KMT-3...

Ball valve for KMT-1/2/3 (Standard delivery scope)

Ball valve Thread A B


[mm] [mm]
DN 15 Rp ½" 83.7 35
183

183

DN 20 Rp or NPT ¾" 72.7 35


173

173

DN 25 Rp or NPT 1" 88 47,5


DN 32 Rp 1 ¼" 100 120
DN 40 Rp or NPT 1½" 110 150
DN 50 Rp or NPT 2" 131 150
Female thread:
BSP thread acc. EN 10226 (old DIN 2999) or NPT
A A B B
2/09 - 2014

No responsibility taken for errors; www.kobold.com 9


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Dimensions [mm] (for KMT-4 i.e. DN 65 ... DN 300)

Sensor probe Enclosure - signal conditioning unit

115
45

60
38
20

USB-Interface

56
Ø 40
120

285/435/585

SW 36

Assembly - insertion depth


15

variable insertion depth max. 165

R½" thread
marking flow
direction
or 315 (DN 125 ... DN 300)
or 465 (DN 350 ... DN 700)

Ø 13
(DN 65 ... 100)

Ø 14
insertion depth = x + OD
2
12

OD ... outside diameter
x

M12 x 1 connector
for probe cable
Ø
40

0,5 * OD

OD

180 flow direction


 2/09- 2014

10 www.kobold.com No responsibility taken for errors;


subject to change without prior notice.
Thermal Mass Flow Meter Model KMT

Dimensions [mm] (Accessories for KMT-4 i.e. DN 65 ... DN 300)


Ball valve ½" (ERS-KMT-KH) Ball valve ½" for parallel measurement (ERS-KMT-KP)
70 70

51,5

54
G3/4“
G1/2“
G1/2“

G3/4“

Ø15
Ø15

54,5
64,5 Material: brass
lateral fitting for
mounting of
pressure or dew
point sensor 64,5 material: brass

Tapping sleeve (Delivery without ball valve)

25
ball valve ERS-KMT-KH
rubber sealing NBR 70 Shore

G3/4“
∼95

11
22

threaded nipple
185

material: stainless steel 1.4301 125

Pipe Clamping range Max.


[mm (inch)] working pressure
DN 65 (2½") 73 - 93 (2.87 - 3.66) 16 bar (PN16) (232psi)
DN 80 (3") 86 - 106 (3.39 - 4.17) 16 bar (PN16) (232psi)
DN 100 (4") 107 - 127 (4.21 - 5.00) 16 bar (PN16) (232psi)
DN 125 (5") 128 - 148 (5.04 - 5.83) 16 bar (PN16) (232psi)
DN 150 (6") 149 - 171 (5.87 - 6.73) 16 bar (PN16) (232psi)
DN 200 (8") 216 - 236 (8.50 - 9.29) 16 bar (PN16) (232psi)
DN 250 (10") 260 - 280 (10.24 - 11.02) 10 bar (PN10) (145psi)
DN 300 (12") 315 - 335 (12.40 - 13.19) 10 bar (PN10) (145psi)
2/09 - 2014

No responsibility taken for errors; www.kobold.com 11


subject to change without prior notice.