FT Datalogger - Serie DCX 22 AA - en
FT Datalogger - Serie DCX 22 AA - en
FT Datalogger - Serie DCX 22 AA - en
The data collector DCX-22 AA measures and records ground water levels using KELLER’s two
sensor AA-technology (absolute-absolute). The submersible depth sensor measures the water
level. Barometric pressure variations are measured and compensated with the built-in waterproof
air pressure sensor which is mounted in the electronics housing at the top of the borehole. There
are no ventilation tubes; this means the DCX-22AA data logger is very rugged and suitable for
reliable applications in humid and wet environments. Even short periods of immersion and flooding
will not cause a problem.
The electronics housing is mounted at the top of the borehole to give easy access for data
downloading. The level sensor (diameter 22 mm) is connected via a sealed cable to the bottom
of the electronics housing. Installation is quick and simple, using fixing devices in various sizes,
suitable for cap lock units of different manufacturers and for well access points starting from 1”
(caps starting at 2” include a hole to lower a dip meter). Thus enabling measuring stations to be
set up at considerably lower costs compared to conventional systems.
The electronics employ the latest microprocessor technology, which give high accuracy and re-
solution for the pressure and temperature signals from both the depth sensor and the barometric
sensor. The measured values are mathematically compensated for all linearity and temperature
errors of the pressure sensors.
Level Electronics
The use of a non-volatile memory ensures high data security. Sensor Housing
The modular design of the DCX-22 AA offers the user the two options for collecting the data.
Standard design requires the user to visit the location, connect via data-cable and download
data. The optional ARC-1 unit allows the transmission of data, and instructions (re-program-
ming) to the data-collector from / to a remote location. The data can be sent to any mobile phone
as a short message (SMS).
OPTIONS
ARC-1 Data transmission module (see separate datasheet ARC-1)
Integrated air pressure sensor
with diaphragm protection
Interface Thread for Battery Air Pressure Electronics Mark of the diaphragm position
Plug Fixing Device Housing Sensor Housing on the level sensor
M20x ø 22 ø 22
15,6
216 mm 88
Cap Lock
KELLER AG für Druckmesstechnik KELLER Ges. für Druckmesstechnik mbH Edition 08/2019
CH-8404 Winterthur DE-79798 Jestetten Subject to alterations
+41 52 235 25 25 +49 7745 9214 0 Companies approved to ISO 9001
[email protected] [email protected] www.keller-druck.com
Specifications
Measuring-/ Pressure Ranges 800…1800 mbar abs. for ranges up to 5 mWC (System Length: 5 m)
800…2300 mbar abs. for ranges up to 10 mWC (System Length: 10 m)
Supply Lithium-Battery 3,6 V (Type AA) Temperature Measurement Accuracy typ. ±0,5 °C
Battery Life * 10 years @ 1 measurement/hour Shortest Measuring Range 1x per second
Output RS 485 digital Memory 57’000 measuring values @ storage
Electrical Connection Fischer DEE 103A054 interval ≤ 15 s, otherwise 28’000 measuring
values (always with attributed time)
Pressure Sensor Specifications (for level sensor respectively air pressure sensor) Material Stainless steel 316L (DIN 1.4435)
Linearity typ. 0,02 %FS O-Ring: Viton®
Comp. Temperature Range -10…40 °C (icing not permitted) Weight: Housing / Probe ≈ 335 g / ≈ 200 g (without cable)
Error Band ** typ. 0,05 %FS *** max. 0,1 %FS Tolerance System Length ± 2 cm
Resolution max. 0,0025 %FS
Long Term Stability typ. 0,5 mbar
* exterior influences could reduce battery life ** Linearity + Temperature Error *** optional max. 0,05 %FS
LOGGER 5
The Logger 5 software makes it possible to configure and read autonomous KELLER data loggers. This software assists users during measure-
ments in the field, with processing the data and also with forwarding it to partners or end customers.
Measurement data can be displayed in graphic form, exported, compensated for air pressure or converted into different units. The online function
displays the current device values.
The software is included in the scope of delivery for the interface converter cables, or it can be downloaded free of charge at www.keller-druck.com.
KELLER AG für Druckmesstechnik KELLER Ges. für Druckmesstechnik mbH Edition 08/2019
CH-8404 Winterthur DE-79798 Jestetten Subject to alterations
+41 52 235 25 25 +49 7745 9214 0 Companies approved to ISO 9001
[email protected] [email protected] www.keller-druck.com