Operating Tables
Operating Tables
Operating Tables
OPERATING TABLES
OM-9T
OM-8P OM-3F
MODEL: OM-8P, OM-7S, OM-3F
ELECTROMECHANICAL
üzümcü
www.uzumcu.com.tr
TA B LE O F C ONT E NT S
1. GENERAL INFORMATION............................................................... 04
2. INTRODUCTION................................................................................ 05
3. FOR YOUR SAFETY.......................................................................... 06
3.1. General Information............................................................................. 06
3.2. Symbols and Explanation..................................................................... 07
3.2.1. Identification Label.............................................................................. 07
4. PACKAGING, TRANSPORTATION AND STORAGE..................... 08-09
5. PRESENTATION................................................................................. 10
5.1. Intended Use and Areas of Use............................................................ 10
5.2. Device and Parts Definition................................................................. 10
5.3. Technical Specifications....................................................................... 11
5.4. Table Movements................................................................................. 12
6. INSTALLATION AND USE OF THE DEVICE................................. 13
6.1. Installation............................................................................................ 13
6.1.1. Leg Rest Assembly............................................................................... 13
6.1.2. Head Rest Assembly............................................................................ 14
6.1.3. Kidney Positioning Rod....................................................................... 14
6.1.4. Power Cable......................................................................................... 15
6.2. Usage.................................................................................................... 16
6.2.1. Hand Control System Movements....................................................... 16
6.2.2. Storage.................................................................................................. 17
6.2.3. Leg Rest Settings.................................................................................. 17
6.2.3.1. Leg Rest Up/Down Movement............................................................ 17
6.2.3.2. Leg Rest Right/Left Movement........................................................... 18
6.2.4. Kidney Surgery Position...................................................................... 19
6.2.5. Head Rest Positioning.......................................................................... 20
6.2.5.1 Up/Down Movement............................................................................ 20
6.2.6. Leg Rest Settings (OM-7S and OM-3F).............................................. 20
6.2.7 Leg Rest Right/Left Movement (OM-7S and OM-3F)........................ 21
6.2.8 Back Rest Movement (OM-7S and OM-3F)........................................ 22
7. ACCESSORIES................................................................................... 23
7.1. Standard Accessories............................................................................ 23
7.2. Standard Accessories Installation......................................................... 24
8. CLEANING AND MAINTENANCE.................................................. 25
8.1. Table Cleaning..................................................................................... 25
8.2. Maintenance......................................................................................... 27
8.2.1. Fuse Replacement................................................................................ 27
8.2.2. Battery Usage....................................................................................... 28
9. PROBABLE FAILURES AND TROUBLESHOOTING.................... 29
10. CIRCUIT DIAGRAMS....................................................................... 30
10.1. Electrical Circuit Diagram................................................................... 30
11. RECYCLING....................................................................................... 31
12. AUTHORIZED SERVICES................................................................ 31
Copyright
This document has been prepared by ÜZÜMCÜ Tıbbi Cihaz ve Medikal
Gaz Sistemleri San. ve Tic. A.Ş.; ÜZÜMCÜ Tıbbi Cihaz ve Medikal Gaz
Sistemleri San. ve Tic. A.Ş. reserves the right to make changes in the
technical specifications of the product. It is forbidden to reproduce or copy
any part of this document without company’s written permission.
CE Declaration of Conformity
1. GENERAL INFORMATION
TR
üzümcü
www.uzumcu.com.tr
Thank you for purchasing ÜZÜMCÜ brand operating table. Our product
has been meticulously designed and produced for offering quality and
performance to you in the best possible way. Please carefully read all
information about the operation and safety of the operating table.
This manual explains the steps of Installation, Use, Cleaning and
Maintenance& Repair for ÜZÜMCÜ brand operating table.
Points to Consider During Handling and Transportation
Please follow the instructions on equipment boxes during handling and
transportation.
Matters Related to Maintenance and Repair
2. INTRODUCTION
Please call the authorized service of the equipment in case of situations that
require Maintenance and Repair.
Information about Connection and Assembly
Please refer to the following parts for connection and assembly information.
TR
üzümcü
www.uzumcu.com.tr
3.1. GENERAL INFORMATION
In order to use and service the operating table by ensuring patient and
user safety;
• Make sure that both this user manual and its warnings and explanations
have been read and understood before servicing the table.
• The buyer is responsible for ensuring that the staff in charge of using or
servicing the table is informed about the content of this user manual.
• Keep the user manual readily available to the staff in charge of using or
servicing the table.
3. FOR YOUR SAFETY
• Use the table only for the purposes explained in this manual.
• Service the table at the periods indicated in Article 8.2 on page 21 of this
manual.
• Table should be used with its original antistatic mat and positioned on an
antistatic floor.
• ÜZÜMCÜ R&D Department reserves the right to change and/or abolish
the technical specifications of the device and the information contained in
this manual.
ATTENTION
The metal chassis of the table has electrical grounding. However,
grounding reliability depends on the electrical installation of the
place where the table is used. The table should be used with a TS
3166-compliant electrical installation.
If treatment methods such as cautery, etc. are employed, the patient should
6 be protected by being separated from the metal sections of the table by
means of insulating materials. In case of contact, you may encounter
serious dangers such as electric shock, burn, etc.
Use the table only at 220-240 V, 50 / 60 Hz alternating current. Otherwise,
the device may be damaged or pose dangers such as electric shock.
TR
üzümcü
www.uzumcu.com.tr
3.2. SYMBOLS AND EXPLANATION
Attention! Refer to
accompanying documents. Protective Earthing
50°C
üzümcü
Üzümcü Tıbbi Cihaz ve Medikal Gaz Sistemleri San. ve Tic. A.Ş
Ankara-Konya Devlet Karayolu 29 Km. 06830 Gölbaşı / Ankara - TÜRKİYE
www.uzumcu.com.tr
Novela Senseo
7
REF Novela Class IIa
SN.0900279016 230VAC-50 Hz
2011 160 W+ %15
50°C
0°C
TR
üzümcü
www.uzumcu.com.tr
4.1. PHYSICAL ENVIRONMENT
TR
üzümcü
www.uzumcu.com.tr
4. PACKAGING, TRANSPORTATION AND STORAGE
1
1 1
2
Figure-1
ATTENTION
Never use sharp objects such as knife while opening the nylon package of
the table and the accessories box. Or else, you may damage table mats and
accessories.
While transporting the table, make sure that the table does not turn upside 9
down or fall.
For preventing the occurrence of functional problems in the table, while
transporting the table, get support from the sections no. 2 shown in Figure
1. Never use the sections no. 1.
TR
üzümcü
www.uzumcu.com.tr
5.1. INTENDED USE AND AREAS OF USE
ÜZÜMCÜ brand operating tables are safely used in different surgical
operations with wide accessories options. Table movements are performed by
the electro-motor system without any sound or vibration.
The table is positioned by means of the microprocessor hand control. It is
possible to store 5 different table positions in OM-8P model and 3 different
table positions in OM-3F and OM-7S models.
The table features a built-in battery system that steps in case of power
failures. The table upper tray which transmits x-rays allows for radiological
examinations.
The table upper tray is covered with removable, liquid-tight, seamless,
5. PRESENTATION
special antistatic mats. Head rest and leg rest feature an easily removable
structure.
The table can be easily moved to desired direction thanks to its castors and
can be safely fixed on the ground with its central brake mechanism.
5.2. DEVICE AND PARTS DEFINITION
5
1
10
9
10
OM-8P
8
7
6 13
11
TR
12
OM-9T
üzümcü
www.uzumcu.com.tr
PART NO PART NAME PART NO PART NAME
1 Head Rest 8 Lower Tray Body Carrying Cart
2 Two-Piece Back Rest (OM-8P OM-9T) 9 Hand Control Set
3 Seat 10 Accessories Rail
4 Two-Piece Leg Rest (OM-8P) 11 Lower Tray (OM-9T)
5 Kidney Movement Rod (OM-8P OM-9T) 12 Leg Rest (OM-9T OM-3F)
6 Brake Pedal 13 Viscoelastic Mat
7 Lift Body
Note: Common in all operating table models not indicated in the letterhead.
5. PRESENTATION
Number of Mat Parts 8 5 6
Height Electrical 840-1240 mm Electrical 780-1180 mm Electrical 830-1215 mm
Electrical +70°/-30° Mechanical +70°/-30° Gas Electrical +70°/-25° Gas
Back Rest Springs Springs
Trendelenburg Electrical +25° Electrical +30° Electrical +25°
Reverse Trendelenburg Electrical -25° Electrical -20° Electrical -25°
Lateral Tilt Electrical +25°/-25° Electrical +20°/-20° Electrical +25°/-25°
Sliding 300 mm - 300 mm
Head Rest Gas Springs +35°/-60° Gas Springs +35°/-60° Gas Springs +35°/-60°
1. Part
Mechanical +50°/-75° Gas Springs +30°/-90° and Gas Springs +30°/-90° and
Leg Rest and 90° 80° lateral 80° lateral
2. Part
Mechanical +20°/-90°
Head Rest 220 kg 220 kg 220 kg
Voltage 220-240V AC 50/60Hz 220-240V AC 50/60Hz 220-240V AC 50/60Hz
Kidney Position Convex 295O - 295O
STANDARDS
Category AP Equipment (TS EN 60601-1) IPX4 (TS 3033 EN 60529),
Mode of Operation: Continuous Operation (TS EN 60601-1), Mobile
Operating Table (TS EN 60601-2-46), 11
Conforming to Standards: TS EN 60601-1, TS EN 60601-2-46,
TR
üzümcü
www.uzumcu.com.tr
5.4. TABLE MOVEMENTS
12
Head Rest Movement Leg Rest Movement
TR
Back Rest Movement Kidney Position (OM-9T OM-8P)
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.1. INSTALLATION
Upon unpacking the operating table, take it to the place of installation and fix
it in desired place by stepping on the brake pedal.
Figure-3 Figure-4
Mount the leg rests and the head rest to the table as explained below.
2
Şekil-5 TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
Figure-6
Figure-7
6.1.3.
KIDNEY POSITIONING
ROD (OM-8P-OM-9T)
Place the kidney positioning
rod into the shaft on the back
rest. You can have kidney
14
positioning movement by
3 1 rotating by means of the clutch
coupling in the shaft (3). Figure
8
Figure-8
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.1.4. POWER CABLE 2
Insert the end no. 2 of the power cable shown in Figure 9 into the table socket
no. 1. After inserting the end no. 3 of the power cable into a grounding plug,
the table becomes ready for use.
2 3
1
Figure-9
ATTENTION
Before using the table for the first time, charge the battery for 20 hours.
For this, just insert the power cable of the table into a grounding plug.
After the battery is charged, unplug the table.
15
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.2. USAGE
6.2.1. HAND CONTROL SYSTEM
1 2
3 4
5 6
7 8
9 10
11 12
Figure-11
NO CONTROL FUNCTION
1 Table Up Press for “Up” function (All Models)
2 Table Down Press for “Down” function (All Models)
3 Back Rest Up Press for “Up” function (OM-8P OM-9T)
16 4 Back Rest Down Press for “Down” function (OM-8P OM-9T)
5 Reverse Trendelenburg Press for “Reverse Trendelenburg” movement (All Models)
6 Trendelenburg Press for “Trendelenburg” movement (All Models)
7 Left Lateral-Tilt Press for “Left Lateral-Tilt” movement (All Models)
8 Right Lateral-Tilt Press for “Right Lateral-Tilt” movement (All Models)
9 Left Sliding Press for “Left Sliding” movement (OM-8P and OM-9T)
10 Right Sliding Press for “Right Sliding” movement (OM-8P and OM-9T)
11 Tray Parallel Press for “Reset” movement (All Models)
12 Storage For storage, press for 3 times (All Models)
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.2.2. STORAGE
1. Bring the table to the position that you want to store (such as Trendelan-
burg, Lateral Tilt, etc.).
2. 7, 8, 9, 10, 11 keys on the hand control are storage keys.
3. Press M key on the hand control set for three times in order to make the
hand control set ready for storage.
4. Of 7, 8, 9, 10, 11 keys, select the one in which you want to store the
position and then press that key for 3 seconds. The hand control set sounds
“beep”.
5. The position that you have selected is stored.
6. In order to call the position you have stored, press M key and the key in
which you have stored the position simultaneously until the table is posi-
tioned.
5. Press 11 key for leveling the table tray.
ATTENTION
The hand control set is manufactured from corrosion resistant materials.
The hand control set is capable of functioning at low voltage. Thus, nec-
essary measures have been taken against electric shock. However, so as to
protect the hand control set from continuous and short-term failures, keep
it away from liquids such as water, etc.
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
sol
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.2.4. KIDNEY SURGICAL
POSITION (for OM-8P OM-9T
model operating tables)
Figure-15 (sadece OM-8P) Rotate the mounted positioning rod
(1) in arrow direction. The operating
table takes kidney position as the
section that is located on the back
rest and comprises of 2 parts (2)
moves up angularly as shown in
Figure 16.
19
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
Figure-17
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.2.7 LEG REST RIGHT/
LEFT MOVEMENT (OM-9T
- OM- 3F)
For positioning the operating
table leg rest outwardly in
horizontal axis, loose the screw
(1) by rotating clockwise. Upon
Figure-19 (0M-9t ve OM-3F) positioning the leg rest as desired
in horizontal axis, fix the leg rest
by rotating the screw in arrow
direction. The leg rest can move
outwardly for 80o as shown in
Figure 20.
21
TR
üzümcü
www.uzumcu.com.tr
6. INSTALLATION AND USE OF THE DEVICE
6.2.8 BACK REST MOVEMENT (OM- 3F)
The back rest of the operating table is moved with the help of the rod (1) that
connects gas springs. In order to position the back rest as desired at certain
angles, move the rod up and down in arrow directions.
TR
üzümcü
www.uzumcu.com.tr
7.1. STANDARD ACCESSORIES
(1) IV Pole
(2) Arm Support
Figure-22
7. ACCESSORIES
(1) Anesthesia Frame
Figure-23
(1) Belt
23
Figure-24
TR
üzümcü
www.uzumcu.com.tr
7.2. USE OF THE STANDARD ACCESSORIES
Figure-25
7. ACCESSORIES
Figure-26
24
TR
üzümcü
www.uzumcu.com.tr
8.1. TABLE CLEANING
Dilute the solutions that are used for operating rooms cleaning and
disinfection at required rates for using in the stainless steel and polyurethane
mat. Wipe the stainless steel and polyurethane mat with the help of a soft and
üzümcü
www.uzumcu.com.tr
Besides, stainless steel should be cleaned by choosing the method suitable for
the substance that causes contamination.
Remove the dirt on the equipment with mild detergent-added lukewarm water
or with just lukewarm water with the help of a soft cloth or a soft sponge. In
8. CLEANING AND MAINTENANCE
cleaning deep dirt, vinegar in small quantities can also be added to water.
Following cleaning, rinse with hot fresh water again with the help of a clean
cloth and then dry with a clean cloth.
Remove the finger prints and stains on the equipment with chloride-free glass
cleaners or sodium carbonate added to water in small quantities with the
help of a clean cloth. After cleaning, make sure to rinse with lukewarm water
again with the help of a clean cloth and then dry.
For the oil stains on the equipment, first remove the oil that has contaminated
the equipment with a soft cloth or paper towel. Following pre-cleaning
with ammonium solution or detergent (non-scratching) in lukewarm water,
perform routine cleaning operations.
It is possible to use also organic solutions for removing oil stains. However,
make sure that such solutions do not contain chlorine, acetone, methyl
alcohol and mineral spirits.
In case the equipment contacts with acid, immediately deacidify and rub with
ammonium- or sodium bicarbonate-containing solution and then dry.
26
ATTENTION
Before cleaning and servicing the table, make sure that the table is plugged out.
Dyes and solvents (chemicals) may be hazardous for the mat that covers the table.
For this reason, immediately remove the liquids which splash or spill on the mats.
Stainless steel may be scratched in case of contact with hard materials. Prevent
the stainless steel part of the operating table from getting in contact with hard
materials.
TR
üzümcü
www.uzumcu.com.tr
8.2 MAINTENANCE
The table should be periodically controlled for any damage or cut on table’s
power cables and hand control set cables and for table movements (up, back
rest, trendelenburg, lateral-tilt, head rest and leg rest movements). Apart from
Figure-28
27
Figure-29
Figure-27
TR
üzümcü
www.uzumcu.com.tr
8.2.2. BATTERY USE
The battery beeps when it is discharged. Do not charge the battery before it
is discharged (before hearing the beep sound). In order to charge the battery,
8. CLEANING AND MAINTENANCE
insert the power cable of the table into a grounded socket. The battery is
charged in 20 hours.
The rate of self-discharge of the batteries used in operating tables is very low.
These rates are as follows:
The battery of the table which has not been used for 12 months or more
should be charged for 20 hours before use.
The physical life of the battery is 4-5 years. The batteries which have
completed their physical life should be replaced by the authorized service
staff.
ATTENTION
Make sure that the table is plugged out after each use and after 20-hour
charging. Never forget the table as plugged in. Otherwise, the physical life
of the table shortens.
28
TR
üzümcü
www.uzumcu.com.tr
Please get in contact with our authorized technical service for information
about table components and other information that may help in table repair.
ATTENTION
Table’s enclosure should be opened only by trained service staff.
29
TR
üzümcü
www.uzumcu.com.tr
10.1. ELECTRICAL CIRCUIT DIAGRAM
Hand
El Kumanda
Control
Seti
Set
Body
Gövde
Control
Opsiyonel
Optional
10. CIRCUIT DIAGRAMS
KONTROL KUTUSU
Batarya M1 M2 M3 M4 M5 M6
Battery
Trend.T ilt Sliding
Figure-30
30
TR
üzümcü
www.uzumcu.com.tr
11. RECYCLING
If you shall dispose of an operating table or replace any part of it, check the
recyclability of each part. Many of the metals used in the operating table are
stainless steel.
11. RECYCLING
Please implement the recycling procedure related to stainless steel.
For more information about recycling, please get in contact with relevant
institutions and foundations or visit the websites that give information about
recycling.