FreeStyle™ Portable
Oxygen Concentrator
DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL. IF YOU ARE
UNABLE TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT YOUR EQUIPMENT PROVIDER BEFORE
ATTEMPTING TO USE THIS EQUIPMENT; OTHERWISE, INJURY OR DAMAGE MAY RESULT.
Smoking while using oxygen is the number one cause of fire injuries and related
deaths. You must follow these safety warnings:
Do not allow smoking, candles, or open flames in the same room with the device or
within 5 feet (1.52 meters) of the oxygen-carrying accessories.
Smoking while wearing an oxygen cannula may cause facial burns and possibly death.
Removing the cannula and putting it on bedding, sofas, or other cushion material will
cause a flash fire when exposed to a cigarette, heat source, or flame.
If you smoke, these 3 steps may save your life:
turn off the oxygen concentrator, take off the cannula, and leave the room where this
device is located.
“No Smoking – Oxygen in Use” signs must be prominently displayed in the home, or where the
oxygen concentrator is in use. Patients and their caregivers must be informed about the dangers of
smoking in the presence of, or while using, medical oxygen.
English: A multilingual version of the manual is available through your Equipment Provider.
Español: Una versión multilingüe del manual está disponible a través de su proveedor de equipo.
Français: Une version multilingue du manuel est disponible par l'intermédiaire de votre fournisseur de matériel.
Deutsche: Eine mehrsprachige Version des Handbuchs ist in Ihrer Geräte-Anbieter.
TABLE OF CONTENTS
Symbols 1-2
Operator Profile 3
Nasal Cannula 15
Filters 24
FreeStyle Accessories 32
Troubleshooting 33-35
Classification 40
Limited Warranty 41
AirSep’s FreeStyle™ Portable Oxygen Concentrator
This Patient Manual will acquaint you with AirSep’s FreeStyle™ Portable Oxygen Concentrator (POC). Make sure you read and
understand all the information contained in this manual before you operate your FreeStyle unit. Should you have any questions, your
Equipment Provider will be happy to answer them for you.
Symbols
Symbols are frequently used on equipment and/or the manual in preference to words with the intention of decreasing the
possibility of misunderstanding caused by language differences. Symbols can also permit easier comprehension of a concept
within a restricted space.
The following table is a list of symbols and definitions used with the FreeStyle Portable Oxygen Concentrator.
Oxygen is a gas that makes up 21% of the room air we breathe. Our bodies depend on a steady supply to function properly. Your
physician prescribed a flow or setting to address your particular respiratory condition.
Although Oxygen is a non-addictive drug, unauthorized oxygen therapy can be dangerous. You must seek medical advice before you
use this oxygen concentrator. The Equipment Provider who supplies your oxygen equipment will demonstrate how to set the
prescribed flow rate.
The FreeStyle Portable Oxygen Concentrator may be used during sleep under the recommendation of a qualified
clinician.
The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gases. In the FreeStyle unit, room air passes
through a regenerative, adsorbent material called “molecular sieve.” This material separates the oxygen from the nitrogen. The result is
a flow of high-concentration oxygen delivered to the patient.
FreeStyle combines advanced oxygen concentrator technology along with oxygen conserving technology for the world’s smallest and
lightest portable oxygen concentrator at just 4.4 lb (2 kg). The FreeStyle efficiently produces its own oxygen, and quickly delivers it as
a pulse of oxygen when you breathe at the very beginning of your inhalation. This eliminates the waste associated with a continuous
flow oxygen device that flows oxygen while you are exhaling. This pulsing of the oxygen is equivalent to continuous flow. FreeStyle
produces the equivalent of up to 3 LPM (liters per minute) in a lightweight package that can be easily carried and used away from the
home.
FreeStyle operates from four different power sources. (Refer to the Power Supplies section of this manual.)
Operator Profile:
AirSep’s Concentrators are intended to supply supplemental Oxygen to users suffering from discomfort due to ailments which
effect the efficiency of ones lungs to transfer the oxygen in air to their bloodstream. POC’s provide the convenience of using a
non-delivery POC system rather than delivery system (O2 tank) which makes the user relatively self-sufficient in terms of in-home use,
ambulation (both within and outside of the home) mobility and overall lifestyle. Oxygen Concentrator use requires a physician’s
prescription, and is not intended for life support use.
Although Oxygen therapy can be prescribed for patients of all ages the typical oxygen therapy patient is older than 65 years of
age and suffers from Chronic Obstructive Pulmonary Disorder (COPD). Patients typically have good cognitive abilities and
MN123-1 rev E 11/12 3
must be able to communicate discomfort. If the user is unable to communicate discomfort, or unable to read and understand the
concentrator labeling and instructions for use, then use is recommended only under the supervision of one who can. If any
discomfort is felt while using the concentrator, patients are advised to contact their healthcare provider. Patients are also
advised to have back-up oxygen available (i.e. cylinder oxygen) in the event of a power outage or concentrator failure. There
are no other unique skills or user abilities required for concentrator use.
FreeStyle has received the US Federal Aviation Administration’s (FAA) acceptance for onboard in-flight use by oxygen
passengers on commercial airlines via a 2006 amendment to SFAR 106.
In addition, as of May 13, 2009, a new Department of Transporation (DOT)/FAA ruling has determined that US-based carriers,
as well as international flights with origination or destination in the US, must allow passengers with FAA-approved portable
oxygen concentrators to use them on board, and in flight, as medically necessary. Check directly with the individual airlines
with which you would like to travel for up-to-date information on their specific POC policies.
This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open flames within
the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe this warning can result in
severe fire, property damage, and/or cause physical injury or death.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort
while using this unit may require additional monitoring. Patients with hearing and/or sight impairment(s) may need
assistance with monitoring alarms
Use no oil, grease, or petroleum-based or other flammable products with the oxygen-carrying accessories or the
FreeStyle unit. Only water based, oxygen compatible lotions or salves should be used. Oxygen accelerates the
combustion of flammable substances.
The incorrect use of the FreeStyle battery can cause the battery to get hot, ignite, and may cause serious injury. Be sure
not to pierce, strike, step on, or drop the battery, or otherwise subject the battery to strong impacts or shocks.
While using the FreeStyle outdoors with the Universal AC/DC power supply, connect the power supply into a Ground
Fault Interrupted (GFI) outlet only.
The FreeStyle should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is unavoidable,
the device should be observed to verify normal operation.
Electrical shock hazard. Disconnect the power cord from the electrical outlet before you clean the unit to prevent
accidental electrical shock hazard. Only your Equipment Provider or a qualified service technician should remove the
covers or service the unit.
Care should be taken to prevent FreeStyle from getting wet or allowing water to enter the unit. This can cause the unit to
malfunction or shut down.
Federal (USA) law restricts this device to sale or rental by order of a physician or other licensed health care provider
In the event of an alarm or you observe that FreeStyle is not working properly; consult the Troubleshooting section in this
manual. If you cannot resolve the problem, consult your Equipment Provider.
The FreeStyle Portable Oxygen Concentrator may be used during sleep under the recommendation of a qualified
clinician.
Operating the FreeStyle unit outside of its normal operating temperature range can affect performance and decrease
battery run time and/or increase battery charge time. (Refer to the Specifications section in this manual.)
Do not operate unit in a restricted or confined space (i.e., a small case or handbag) where ventilation can be limited. This
can cause the FreeStyle unit to overheat and affect performance.
When using FreeStyle in an automobile, boat, or on other DC sources with the Universal AC/DC power supply, make sure
that the vehicle is started and running before connecting the FreeStyle unit. If the Universal AC/DC power supply does not
illuminate and requires resetting, disconnect the power supply from the DC outlet, restart your vehicle, and then reconnect
your power supply into the DC outlet. Failure to follow these instructions can result in the power supply not supplying
power to FreeStyle.
When the automobile in which you are using the FreeStyle unit is turned off, disconnect and remove the unit from the
automobile with you. Do not store FreeStyle in a very hot automobile or in other similar, high-or low-temperature
environments. Operating or storing the unit outside the normal temperature range can affect the performance of the
FreeStyle. (Refer to the Specifications section in this manual.)
If the FreeStyle has been stored for an extended period of time outside its normal operating temperature range, the unit
should be allowed to return to normal operating temperature before being turned on. (Refer to the Specifications section
in this manual.)
It is very important to select only the prescribed level of oxygen. Change the pulse flow selection only under the guidance
of your physician.
Replace the disposable cannula as recommended by the manufacturer or your equipment provider. Additional supplies
are available from your Equipment Provider.
Do not position the Concentrator so that it is difficult to access the power cord.
Do not position the Concentrator so that it is difficult to access the power cord.
Portable and mobile RF communications equipment can effect medical electrical equipment.
Cannula tubing must be non-kinking, which can be used for a total length of up to 25 ft (7.6 m) maximum.
Ensure the cannula is fully inserted and secure. This ensures that the FreeStyle unit can properly detect inspiration for
oxygen delivery.
Storing your FreeStyle for extended periods of time at high temperatures or with a fully charged or completely discharged
battery can degrade its overall battery life.
Depending upon the temperature of the FreeStyle battery, it can take several minutes for the charging cycle to start after
connecting to power. This is a normal condition and is intended for safe charging.
The FreeStyle battery does not need to be fully discharged before recharging. It is recommended to charge the FreeStyle
battery after each use.
Charging may take several minutes after connecting the power to initiate, depending on the battery’s internal operating
temperature. This is a normal condition and is intended for safe battery charging. This circumstance is more likely when
the battery has been fully discharged.
If the internal battery fully depletes and the FreeStyle unit shuts down, the unit cannot be restarted with the AirBelt.
Should this occur, connect your FreeStyle to its Universal AC/DC power supply for a short period of time in order to
provide sufficient internal battery power to start the unit. AirBelt can then be connected to provide additional run time.
If the FreeStyle power supply remains connected when the battery is fully charged, the four LEDs will turn off within
approximately 1 hour.
It may be necessary to initially connect the Universal AC/DC FreeStyle power supply to the FreeStyle unit before the unit
will operate for the first time on battery power. Your Equipment Provider may have already performed this step for you.
Use only AirSep Part No. FI144-1 for air intake filter for this unit.
Do not operate FreeStyle without the air intake filter in place. If a second filter is provided, insert the “replacement” filter
before you clean the dirty filter.
Do not attempt any maintenance other than the possible solutions listed within this manual.
The lithium ion rechargeable battery that is used in the FreeStyle unit does not need
to be fully depleted before recharging. It is recommended to charge the battery regardless of the battery’s capacity
level after use.
The battery will charge when the unit is off as well as while the unit is running off the Universal AC/DC power supply.
“No Smoking – Oxygen in Use” signs must be prominently displayed in the home, or where the oxygen concentrator is in
use. Patients and their caregivers must be informed about the dangers of smoking in the presence of, or while using,
medical oxygen.
The incorrect use of AirBelt can cause the battery to get hot, ignite, and can cause serious injury. Be sure not to pierce,
strike, step on, or drop the battery, or otherwise subject the battery to strong impacts or shocks.
Do not attempt to charge the optional AirBelt with the FreeStyle Universal AC/DC power supply, or AirBelt can be
damaged.
Use only the AirBelt power supply provided to charge AirBelt.
Depending upon the temperature of the AirBelt, it can take several minutes for the charging cycle to start after connecting to
power. This is a normal condition and is intended for safe charging.
AirBelt does not need to be fully discharged before recharging. It is recommended to charge AirBelt after each use.
Lithium batteries may permanently lose capacity when exposed to extremely hot temperatures with the batteries fully
charged or completely depleted. For extended storage, it is recommended that batteries be charged 25 to 50% and remain
within a temperature range of 73 °F (23 °C) +/- 2°C.
Optionally, you may also have an AirBelt for extended battery duration. That packaging contains the following, as shown in
Figure 4:
All AirBelt Warnings, Cautions and Notes should be read first before proceeding with your equipment.
See pg 9, "Important Safety Rules for Optional AirBelt".
Other optional accessories include a harness to easily convert the FreeStyle carrying bag to a backpack (part number MI284-1),
as shown in Figure 5. The FreeStyle can also be worn on the waist if desired by feeding the AirBelt or other belt you are
wearing through the loops on the FreeStyle carrying case. (See Figure 6.) The optional accessories bag (part number MI320-1)
enables even more convenient travel when transporting all power accessories and optional AirBelts and/or the harness for use at
your intended destination.
Figure 4: AirBelt and Universal AC/DC Power Supply/Power Cord Figure 5: FreeStyle unit worn backpack style Figure 6: FreeStyle unit worn on the waist
Battery Charging
Check to make sure your unit’s battery is fully charged before venturing out with FreeStyle for the first time or upon subsequent
use. To check the level of charge of the internal battery, press the BATT button on FreeStyle’s keypad. The battery
gauge/indicator(s) LEDs above the BATT button illuminate to indicate the level of internal battery charge (25-100%). Note: The
internal battery is charging whenever the unit is operating on AC or DC power. To charge FreeStyle’s internal battery, simply
connect its Universal AC/DC power supply into the unit’s power connection inlet (as shown in Figure 7). Be certain to first
properly align the power cord to this inlet. To do this, take note of the “D”-shaped plug of both the power cord connector and
the FreeStyle’s inlet connection. These must be properly aligned and when removing the power cord, the release button must be
pressed to remove it from the FreeStyle unit. This ensures that neither the unit nor the power accessories are damaged.
Press to remove
FreeStyle unit’s power cord.
Power Inlet
Optional AirBelt
The optional AirBelt, which can increase FreeStyle’s total battery duration up to 10 hours, requires approximately 3 hours to
completely charge when the external battery is fully depleted. To recharge AirBelt, connect its own AC power supply to the end
of AirBelt’s power cord (as shown in Figure 8), and the other end of the power supply into a proper AC power outlet. A separate
battery gauge/indicator for checking the level of charge on the AirBelt’s battery is located on this accessory’s cord and functions
exactly like the internal battery’s indicator on the keypad.
Depending upon the temperature of the FreeStyle battery, it can take several minutes for the charging cycle to start after
connecting to power. This is a normal condition and is intended for safe charging.
The FreeStyle battery does not need to be fully discharged before recharging. It is recommended to charge the FreeStyle
battery after each use.
AirBelt does not need to be fully discharged before recharging. It is recommended to charge it after each use.
The FreeStyle’s internal battery will completely recharge from its fully depleted state in approximately 3 ½ hours, whether or not the
unit is in use on Universal AC/DC power.
While charging a fully discharged battery, initially the 25% LED will start blinking rapidly for a few minutes, then start to blink at a
slower rate every (1/2 second).
The LED will continue to blink until 25% capacity is reached. The LED will then turn solid.
Each of the four LEDs, 25% -100%, will blink as stated above, then turn solid when they reach their capacity.
When all LEDs illuminate solid, the battery is fully charged and the LEDs will remain solid for a period of time, then all four LEDs will
turn off.
Lithium batteries may permanently lose capacity when exposed to extremely hot temperatures with the batteries fully
charged or completely depleted. For extended storage, it is recommended that batteries be charged 25 to 50% and remain
within a temperature range of 73° F (23° C) +/- 2°C.
Nasal Cannula
A nasal cannula and tubing are used to deliver oxygen from the FreeStyle unit to the user. The tubing is connected to the unit’s
oxygen outlet (See Figure 9).
AirSep recommends a nasal cannula with 7 ft (2.1 m) of tubing, AirSep Part No. CU002-1. Other lengths of tubing up to 25 ft
(7.6 m) maximum, including nasal cannula, may be used.
Replace the disposable cannula periodically following normal usage. Additional supplies are available from your Equipment
Provider.
Cannula tubing must be non-kinking, which can be used for a total length of up to 25 ft (7.6 m) maximum.
Always follow the cannula manufacturer’s instructions for proper use. Consult your licensed health care provider to determine
how often the cannula should be replaced.
Make sure the cannula is fully inserted and secure. This ensures that the FreeStyle unit can properly detect inspiration for
oxygen delivery.
Alarm/Service
Indicator
Pulse Flow
Indicator(s)
Battery Gauge/
Flow Selection Button Indicator(s)
Handle
(Internal) Battery
Button
Oxygen Outlet
Air Outlet
Air Outlet
Air Intake/Filter
Cover
RTCA DO160
indicates that
FreeStyle unit has
been tested and
approved to Label/Patient
comply with FAA Instructions
standards.
Figure 10: FreeStyle Exterior View - Front
Now that you are familiar with FreeStyle’s components, review the instructions on the following pages to operate the FreeStyle
unit.
Operating Instructions
1. Locate and position the FreeStyle so that the air inlets and air outlets are not obstructed.
2. Power the unit from (a) the internal battery; (b) AirBelt; (c) DC outlet (i.e. automobile or motor boat); or (d) an AC
outlet (i.e. normal household electric). (Refer to the Power Supplies section of this Patient Manual.)
5. Turn the FreeStyle unit on by pressing [1], [2], or [3] button on the unit’s keypad for the liter flow prescribed by your
physician. The green LED above the button selected illuminates. Each time you turn on the FreeStyle unit, a brief alarm
sounds. This indicates that FreeStyle is powered for use.
6. To change the pulse flow setting, press the appropriate button. After the automatic start-up mode, it is normal to hear a
difference in sound as you change the settings.
7. To turn FreeStyle off, press the button corresponding to the pulse flow setting light that is illuminated.
8. When FreeStyle senses inhalation, oxygen is supplied to you through your cannula.
9. The time required to reach maximum oxygen concentration after turning on the FreeStyle unit is approximately two
minutes.
Use no oil, grease, or petroleum-based or other flammable products with the oxygen-carrying accessories or the
FreeStyle unit. Only water based, oxygen compatible lotions or salves should be used. Oxygen accelerates the
combustion of flammable substances.
Each time the FreeStyle unit is turned on; it cycles through an automatic start-up period for approximately seven minutes.
During this time, it maintains a constant speed regardless of setting. After this period, it is normal to hear a change in
speed at the 1 and 2 settings. This variation in sound represents the differences in oxygen production the unit makes at
each setting.
1. Internal Battery: A rechargeable internal battery is located within each FreeStyle unit. When it is fully charged, it
supplies power to the FreeStyle unit for up to 3 ½ hours. An audible alarm sounds when the battery power is getting low.
The alarm is discussed in the Alarm/Light Indicators section of this manual.
Battery Charging: To charge the internal battery, connect FreeStyle to either the Universal AC/DC power supply and a
100-240 volt, 50/60 Hz AC power outlet, or connect it to a DC power outlet in an automobile (boat, motor home, etc.). A
discharged battery requires approximately 3 ½ hours to fully charge. It is recommended to recharge the battery, even if
only partially depleted, as often as possible.
2. Universal AC/DC Power Supply: A Universal AC/DC power supply allows FreeStyle to be connected to a 100-240
volt, 50/60 Hz outlet. The power supply converts 100-240 volt AC to a DC voltage for the FreeStyle unit. When
FreeStyle is operated with the Universal AC/DC power supply, power from the AC outlet powers the unit and recharges
FreeStyle’s battery simultaneously.
3. Optional AirBelt (External Battery Belt): FreeStyle can also be powered by an external battery belt. This belt can be
worn around the waist, and when used in combination with a fully charged internal battery, supplies power to FreeStyle
for up to 10 hours. The belt pack connects to the FreeStyle unit’s power inlet, and it can be recharged by connecting it to
the battery belt’s own AC power supply.
When using the AirBelt with a fully charged internal battery, the FreeStyle’s internal battery will deplete before the
AirBelt. The AirBelt must be connected to the FreeStyle unit before the FreeStyle’s internal battery is depleted. Observe
and connect the AirBelt to the FreeStyle unit before its internal battery discharges to 25% or less.
If the internal battery fully depletes and the FreeStyle unit shuts down, the unit cannot be restarted with the AirBelt.
Should this occur, connect your FreeStyle to its Universal AC/DC power supply for a short period of time in order
to provide sufficient internal battery power to start the unit. AirBelt can then be connected to provide additional run
time.
Figure 13
AirSep’s new universal power supply for the FreeStyle Portable Oxygen Concentrator (POC) is an all-in-one way to keep you
fully powered, yet traveling even lighter – wherever you go. The universal power supply (Part Number PW022), described in
this insert, enables you to operate your unit while charging its internal battery – all from one component – whether you are
driving to the grocery market, taking a vacation flight, or simply relaxing at a friend’s home. This universal power supply
replaces the two separate AC and DC power supplies that were previously packaged with the FreeStyle unit and covered in your
patient manual.
Your power supply kit is packed complete with three essential accessories, as follows:
• universal power supply (PW017-1), which operates on AC power from 120-240 volts and with multiple DC power
sources, including automobile and airplane power
• removable AC power cord
-CD023-2 US-style
-CD017-2 Euro-style, depending on usage location
-CD025-1 Australian-style, or
-CD017-4 UK-style
• removable DC power cord, which incorporates both the connector for automobile-type outlets and the D-shaped, 4-pin
connector for EmPower®-type outlets commonly used on airlines (CD034-1).
Note: FreeStyle’s internal battery will completely recharge from its fully depleted state in approximately 3½ hours, whether the
unit is in use on AC or DC power.
MN123-1 rev E 11/12 20
Start with the Universal AC/DC Power Supply and FreeStyle Connection
For all procedures below, it is recommended that you first connect the universal power supply to the FreeStyle POC inlet, as
shown in Figure 14. Be sure to properly connect to the unit by ensuring that the carrying case alignment with the FreeStyle unit
inlet allows you to obtain a secure connection so that the cord “locks” in place.
When disconnecting the power supply from the FreeStyle POC, make sure you press down on the tab near the end of the cord to
properly disengage the connection to the unit and release the cord.
Figure 14
1. Make sure your power supply is securely connected to the FreeStyle inlet, as shown in Figure 14.
2. Connect the proper end of the AC electrical cord to the unmarked AC inlet on the power supply, as shown in Figure 15.
3. Connect the opposite end of the AC electrical cord to the wall or other appropriate electric outlet, as shown in Figure 16.
4. To operate the POC, press the FreeStyle [1], [2], or [3] setting on the unit’s top panel, as determined by your prescription.
To turn the unit off, press the setting associated with the LED illuminated above it.
Figure 15 Figure 16
Caution: When using FreeStyle with the universal power supply in an automobile, boat, or other motor vehicle, make sure that
the vehicle is started and running before turning on and operating the FreeStyle unit. If the power supply indicator light does not
illuminate and requires resetting, disconnect the power supply from the motor vehicle’s DC outlet, restart your vehicle, and then
reconnect your power supply into the DC outlet.
1. Make sure your power supply is securely connected to the FreeStyle inlet, as shown in Figure 14.
2. Connect the proper DC power cord end to the inlet marked ACCESSORY DC IN on the power supply, as shown
in Figure 17.
3. Turn on your motor vehicle.
4. Connect the DC power cord to the motor vehicle’s DC outlet, as shown in Figure 18.
5. To operate the POC, press the FreeStyle [1], [2], or [3] setting on the unit’s top panel, as determined by your
prescription. To turn the unit off, press the setting associated with the LED illuminated above it.
Figure 17 Figure 18
1. Make sure your power supply is securely connected to the FreeStyle inlet, as shown in Figure 14.
2. Connect the proper DC power cord end to the ACCESSORY DC IN inlet on the power supply, as shown in Figure 19.
3. Remove the tip of the DC connector by pressing on the tab to expose the 4-pin connector, as shown in Figure 20.
4. Insert the 4-pin connection to the airline passenger DC outlet optionally available at your seat.
5. To operate the POC, press the FreeStyle [1], [2], or [3] setting on the unit’s top panel, as determined by
your prescription. To turn the unit off, press the setting associated with the LED illuminated above it.
Figure 19 Figure 20
For more information on the universal power supply for FreeStyle or any AirSep product accessories, please
contact your local equipment provider.
Figure 21: Removal of Air Intake Figure 22: Removal of Filter from
Filter Cover Filter Cover
Do not operate FreeStyle without the air intake filter in place. If a second filter is provided, insert the “replacement” filter
before you clean the dirty filter.
Use only AirSep Part No. FI144-1 as the air intake filter for this unit.
Figure 23: Lifting Dust Cover on Keypad Display Figure 24: Pressing Flow Selection Button
It is very important to select only the prescribed level of oxygen. Change the pulse flow selection only under the guidance
of your physician.
Additionally, when the unit is operating on and simultaneously being charged through the Universal AC/DC power supply, the
FreeStyle unit’s battery gauge/indicator(s) display the charge level of the battery and remain on for approximately one hour after
reaching a full charge.
An audible alarm sounds if FreeStyle has a low battery, if the cannula is disconnected, or if performance of the unit is outside
specifications. The light and audible alarm conditions are explained in detail below and summarized on the chart later in this
section of the manual.
In the event of an alarm or you observe FreeStyle is not working properly; consult the Troubleshooting section in this
manual. If you cannot resolve the problem, consult your Equipment Provider.
If you feel discomfort or are experiencing a medical emergency, seek medical assistance immediately.
o Start-Up
A brief alarm sounds at start-up. FreeStyle begins to operate when the alarm stops.
o Low Battery
As the battery power approaches a low level, a brief alarm sounds intermittently, and the yellow 25% Battery gauge /
indicator (Figure 11) light also illuminates intermittently. When this occurs, connect FreeStyle to a DC power outlet or to
an AC power outlet, or change to another source of oxygen within two minutes. When FreeStyle is connected to DC
power outlet or AC power outlet, the unit operates while recharging FreeStyle’s battery simultaneously. The level of
battery charge is indicated by the battery gauge/indicator(s).
o General malfunction
If FreeStyle has a general malfunction, a rapid alarm sounds every ½ second, and the alarm light illuminates red
continuously. When this occurs, the concentration of oxygen that FreeStyle is supplying is below unit specifications.
You should change to another source of oxygen as available, and contact your Equipment Provider.
This unit is not to be used for life support. Geriatric, pediatric, or any other patients unable to communicate discomfort while
using this unit may require additional monitoring. Patients with hearing and/or sight impairment(s) may need assistance with
monitoring alarms.
Indicator Brief, continuous at (Green) pulse; FreeStyle has been You may begin to operate your
start–up continuous light turned on FreeStyle unit.
Alarm Continuous: (Red) alarm; No breath detected by the Check the cannula connection. Ensure
Beep continuous light unit for a predetermined that you are breathing through your
time period. nose. If the alarm persists, contact
your Equipment provider.
Alarm Intermittent: 25% (Yellow) Battery voltage is too low Connect the FreeStyle unit into a DC
Beep, beep, beep… BATT; battery indicator to operate FreeStyle. outlet or an AC outlet immediately.
Alarm Rapid intermittent: (Red) alarm; Breathing rate is Reduce activity, then if necessary use
Beep, beep, beep… intermittent light exceeding the capacity of another source of oxygen as available.
the FreeStyle unit. Contact your Equipment Provider.
Alarm Rapid intermittent: (Red) alarm; General malfunction of the Turn off the unit. Change to another
Beep, beep, beep… continuous light FreeStyle unit has source of oxygen, and contact your
occurred. Equipment Provider.
Cabinet
Turn off the FreeStyle unit, and disconnect the power cord from FreeStyle before you clean the cabinet.
Do not use liquid directly on the FreeStyle unit to clean it. A list of undesirable chemical agents includes but is not
limited to, the following, according to the plastics manufacturer: alcohol and alcohol-based products, concentrated
chlorine-based products (ethylene chloride), and oil-based products (Pine-Sol, Lestoil). These are NOT to be used to
clean the plastic housing on FreeStyle, they can damage the unit’s plastic.
Replace the disposable cannula periodically following normal usage. Additional supplies are available from your
Equipment Provider.
Keep the FreeStyle unit clean and free from moisture and dust. Clean the plastic housing periodically by wiping with a
lint-free cloth or with a mild household cleaner applied with a damp cloth or sponge. Pay special attention to the oxygen
outlet for the cannula connection to make sure it remains free of dust, water, and particles.
At least one time each week, remove the cover on the lower front of the unit, and wash the air intake filter, which is
positioned on the inside of the cover. Your Equipment Provider may advise you to clean it more often, depending upon
your operating conditions. Follow these steps to properly clean the air intake filter:
1. Remove the filter from the air intake filter cover, and wash it in a warm solution of soap and water.
2. Rinse the filter thoroughly, and remove excess water with a soft, adsorbent towel. Ensure that the filter is dry
before replacing it.
Do not operate FreeStyle without the air intake filter in place. If a second filter is provided, insert the “replacement” filter
before you clean the dirty filter.
Dual display
Service Required (yellow)
Alarm/Service Indicator
When the alarm service/indicator illuminates (yellow), it is time for your FreeStyle unit to be inspected/serviced by your
Equipment Provider. After any necessary service and the performance is verified by the Equipment Provider, the alarm
service/indicator light will be reset.
PW022- Universal AC/DC Power Supply with battery charger including the following power cords as applicable:
1,2,3 or 4
PW022-1 Universal AC/DC Power Supply (4ft / 1.2m) w/ CD023-2 Power Cord 120V (8ft / 2.4m) and
CD034-1DC power cord (33in / 860mm)
PW022-2 Universal AC/DC Power Supply (4ft / 1.2m) w/ CD017-2 Euro Power Cord (8 ft-2 in / 2.5m) and
CD034-1DC power cord (33in / 860mm)
PW022-3 Universal AC/DC Power Supply (4ft / 1.2m) w/ CD025-1 Australian Power Cord 250 VAC (6 ft-6 in / 2.6m) and
CD034-1 DC power cord (33in / 860mm)
PW022-4 Universal AC/DC Power Supply (4ft / 1.2m) w/CD017-4 UK Power Cord (8 ft-2 in / 2.5m) and
CD034-1 DC power cord (33in / 860mm)
Optional AirBelt includes the following:
BT017-1,2,3
or 4 AirBelt with Power Supply including the following power cords as applicable:
BT017-1 AirBelt with Power Supply PW008-1 (BT017 extended cord length 4 ft / 1.2m) w/ CD023-2 Power Cord 120VAC
(8 ft / 2.4m)
BT017-2 AirBelt with Power Supply PW008-2 (BT017 extended cord length 4 ft / 1.2m) w/ CD017-2 Euro Power Cord (8 ft-2 in / 2.5m)
BT017-3 AirBelt with Power Supply PW008-3 (BT017 extended cord length 4 ft / 1.2m) w/ CD025-1 Australian Power Cord 250 VAC
(6 ft-6 in / 2.6m)
BT017-4 AirBelt with Power Supply PW008-4 (BT017 extended cord length 4 ft / 1.2m) w/ CD017-4 UK Power Cord (8 ft-2 in / 2.5m)
Additional or Other Options:
MI240-2 AirBelt Extender
MI078-1 AirBelt Extender New Velcro Version
MI284-1 Shoulder Harness option enables converting the supplied FreeStyle carrying bag to a backpack
FI144-1 Air Intake Filter
MN123-1 rev E 11/12 32
Use of cables and adapters other than those specified, with the exception of cables and adapters sold by the manufacturer
of the medical electrical equipment as replacement parts for internal components, may result in increased emissions of
decreased immunity of the FreeStyle .
Troubleshooting
The FreeStyle product is designed for years of trouble-free use.
If your FreeStyle Portable Oxygen Concentrator fails to operate properly, refer to the chart on the following pages for possible
causes and solutions and, if needed, consult your Equipment Provider.
FreeStyle does not operate when a pulse flow Battery is discharged. Power the unit through the DC outlet, or an AC
selection button is pressed. outlet.
A continuous alarm sounds and the (red) alarm FreeStyle has not detected a breath for 15 Check the cannula connection.
light illuminates continuously. minutes.
Intermittent alarm condition, and the (yellow) Battery requires charging. Connect to a DC or an AC outlet within 2 minutes,
BATT light illuminates intermittently. or connect to the optional AirBelt.
Rapid alarm condition, Battery voltage is too low to operate the Connect to DC or an AC outlet immediately.
and the (yellow) BATT light illuminates FreeStyle unit.
intermittently and FreeStyle shuts down.
Intermittent alarm condition, and the (red) Breathing rate has exceeded the capacity of Reduce activity, and then turn unit off and back on
alarm light illuminates intermittently. the FreeStyle unit. again to reset unit. If necessary, change to another
source of oxygen as available and contact your
Equipment Provider.
Intermittent alarm condition, and the (red) A general malfunction has occurred. Change to another source of oxygen as available, and
alarm light illuminates continuously. contact your Equipment Provider.
Unit does not start on battery power. Unit may be hot or cold if left outdoors such Allow the unit to reach normal operating
as in an automobile. temperature, which may take several minutes if
exposed to temperature extremes. Temporarily
connect your Universal AC/DC power supply to the
unit’s power inlet and power source, as needed, to
reset the unit’s internal battery.
Delay in recharging internal battery. Internal battery exceeds charging temperature. Unit may be operated; however, charging may not
resume until battery temperature is reduced. (See
page 9 notes.)
Unit alarms while in automobile and connected No power to the unit if battery depleted and Disconnect the Universal AC/DC power supply from
to the DC outlet. DC outlet not charging. the automobile outlet, restart the automobile, and
then reconnect the power supply into the automobile
DC outlet to reset the breaker within DC power
supply.
Dimensions: 8.6 in. high x 6.1 in. wide x 3.6 in. deep (21.8 cm high x 15.5 cm wide x 9.1 cm
deep)
Battery duration 3 LPM – 2 hours; 2 LPM – 2.5 hours; 1 LPM – 3.5 hours
(Rechargeable lithium Optional AirBelt when combined with the internal battery:
Battery) 3 LPM – 5 hours; 2 LPM – 6 hours; 1 LPM – 10 hours
Battery cycle life: Approximately 300 cycles, then 80% capacity or below.
Audible alarms and pulse Start-up – audible and visual (GREEN indicator light);
visual indicators: Pulse flow – visual (GREEN light);
Battery condition – battery level (GREEN indicator lights);
Cannula disconnect- audible and visual (RED alarm light);
System overdraw – audible and visual (RED warning and alarm lights);
High and low pressure – audible and visual (RED alarm light);
Service required: visual (YELLOW alarm light)
Electrostatic ± 6 kV contact ± 6 kV contact Floors should be wood, concrete or ceramic tile. If floors are
discharge (ESD) covered with synthetic material, the relative humidity should be
± 8 kV air ± 8 kV air at least 30 %.
IEC 61000-4-2
Electrical fast ± 2 kV for power ± 2 kV for power Mains power quality should be that of a typical commercial or
transient/burst supply lines supply lines hospital environment.
IEC 61000-4-4 ± 1 kV for input/output Not Applicable
lines
Surge ± 1 kV line to line ± 1 kV line to line Mains power quality should be that of a typical commercial or
hospital environment.
IEC 61000-4-5 ± 2 kV line to earth ± 2 kV line to earth
Voltage dips, short <5 % U T <5 % U T Mains power quality should be that of a typical commercial or
interruptions and (>95 % dip in UT) (>95 % dip in UT ) hospital environment. If the user of the FreeStyle requires
voltage variations for 0,5 cycle for 0,5 cycle continued operation during power mains interruptions, it is
on power. recommended that the FreeStyle be powered from an
40 % U T 40 % U T uninterruptible power supply (UPS) or a battery.
IEC 61000-4-11 (60 % dip in U T) (60 % dip in U T )
for 5 cycles for 5 cycles
70 % U T 70 % U T
(>30 % dip in U T) (30 % dip in U T )
for 25 cycles for 25 cycles
<5 % U T <5 % U T
(>95 % dip in UT) (>95 % dip in UT )
for 5s for 5 s
Power frequency 3 A/m 3 A/m
magnetic field Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
IEC 61000-4-8 hospital environment.
NOTE U T is the a.c. mains voltage prior to application of the test level.
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site surveya should be less than the compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the FreeStyle is used exceeds the applicable RF compliance level above, the FreeStyle should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the FreeStyle .
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RF emissions The Freestyle is suitable for use in all establishments, including domestic establishments
Class B and those directly connected to the public low-voltage power supply network that supplies
CISPR 11 buildings used for domestic purposes.
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions Complies
IEC 61000-3-3
Protection against potential electromagnetic or other interference between the equipment and other devices.
Tested to be in compliance with EN60601-1-2 (EMC).
Tested to be in compliance with RTCA/DO160 Section 21 Category M.
CISPR 11 / EN 55011 Class B Group 1, “Industrial, Scientific, and Medical (ISM) Equipment”
FCC Part 15, Subpart B – Class B Unintentional Radiators
Mode of operation:
Continuous duty.
AirSep Corporation warrants the FreeStyle Oxygen Concentrator to be free from defect in parts for three years (as specified on
the original invoice provided) from the date of delivery to the original purchaser, under normal use and operation. The battery is
warranted for one year. AirSep Corporation’s obligations under this warranty are limited to the repair or replacement of any
such item of equipment (or part thereof) shown to be defective or, at AirSep Corporation’s option, to refund the purchase price
of any such defective item of equipment.
Each item of equipment for which a warranty claim is asserted shall, at the request of AirSep Corporation, be returned on a
prepaid basis with proof of purchase date to the AirSep factory at the expense of the purchaser. The purchaser will be
responsible for return freight charges. Replacement parts shall be warranted as stated above for the unexpired portion of the
original three-year parts warranty (as specified on the original invoice provided). This warranty does not extend to any item or
part subjected to misuse, accident, improper maintenance, or application, or which has been repaired or altered outside of the
AirSep Corporation factory without the express prior written authorization of AirSep Corporation.
THE FOREGOING WARRANTY IS IN LIEU OF ANY OTHER WARRANTY, EXPRESSED OR IMPLIED, IN FACT OR
IN LAW, INCLUDING WITHOUT LIMITATION THE WARRANTY OF MERCHANTABILITY OR THE WARRANTY
OF FITNESS FOR PARTICLUAR PURPOSE. IT IS EXPRESSLY UNDERSTOOD THAT PURCHASER’S SOLE AND
EXCLUSIVE REMEDY FOR DEFECT IN PARTS IS LIMITED TO ENFORCEMENT OF AIRSEP CORPORATION’S
OBLIGATION AS SET FORTH ABOVE, AND AIRSEP CORPORATION SHALL NOT BE LIABLE TO PURCHASER OR
OTHERS FOR LOSS OF USE OF THE EQUIPMENT OR FOR OTHER SPECIAL, INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES.