Brookfield Manual PDF
Brookfield Manual PDF
Brookfield Manual PDF
Operating Instructions
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
I. INTRODUCTION ................................................................................ 3
APPENDIX A -
Cone/Plate Viscometer Set-Up ........................................ 9
APPENDIX B -
Viscosity Ranges ........................................................... 18
APPENDIX C -
Variables in Viscosity Measurements............................ 20
APPENDIX D -
Calibration Procedures .................................................. 22
APPENDIX E -
Models A & A-E Laboratory Stands
with Parts Identification ........................................... 29
APPENDIX F - Fault Diagnosis and Troubleshooting ........................... 31
APPENDIX G - Warranty Repair and Service ........................................ 33
Tear-off Sheet - Viscosity Test Report .................................................... 35
The Viscometer is able to measure over a number of ranges since, for a given spring deflection, the
actual viscosity is proportional to the spindle speed and is related to the spindle's size and shape. For
a material of given viscosity, the resistance will be greater as the spindle size and/or rotational speed
increase. The minimum viscosity range is obtained by using the largest spindle at the highest speed;
the maximum range by using the smallest spindle at the slowest speed.
The higher the torque calibration, the higher the measurement range. The viscosity measurement
range for each torque calibration may be found in Appendix B.
All units of measurement are calculated in units of centipoise (cP) by using a look-up table
known as "the Brookfield Factor Finder" to convert torque reading. The equivalent units of
measurement in the SI system are calculated using the following conversions:
CGS SI
Viscosity: 1 cP = 1 mPa•s mPa•s = milli Pascal - second
Torque: 107 dyne-cm = 1 Newton-m
I.1 Components
1) Dial Viscometer
2) Model A Laboratory Stand
3) Spindle Set with Case (4 spindles for LVF, LVT; 7 Spindles for RVF, RVT, HAT, and HBT).
For Cone/Plate versions: a spindle wrench, one cone spindle and a sample cup (Part No.
CP-44Y) replace the spindle set
4) Guard Leg (LVF, LVT, RVF, and RVT only)
5) Carrying Case
6) Operating Instruction Manual
7) Factor Finder
Please check to be sure that you have received all components, and that there is no damage. If you
are missing any parts, please notify Brookfield Engineering or your local Brookfield agent immedi-
ately. Any shipping damage must be reported to the carrier.
I.3 Specifications
Operating
Temperature: -20°C (-4°F) to 65°C (149°F)
I.4 Set-Up
Note: The position of the laboratory stand clamp assembly is important. Refer to
Appendix E for proper alignment and positioning of the clamp assembly.
Level the viscometer, referring to the bubble level on the instrument. If the
viscometer cannot be leveled, recheck the laboratory stand assembly, as shown on
Sheet #82-0330 (supplied with Model A laboratory stand).
3) Verify that the viscometer's power requirements match your power source before
connecting it to power.
The Brookfield Viscometer is powered by a precision synchronous motor. Exact speeds of rotation
are assured as the motor will turn erratically and spasmodically if synchronization cannot be
maintained.
Speed changes are affected by a gear train having either four or eight speeds. Maximum speed
(RPM) will be set at full clockwise rotation and minimum speed at full counter-clockwise rotation.
Although not absolutely necessary, it is advisable to change speeds while the motor is running.
Four speed Viscometers have a square speed control knob with one number shown on each of four
faces. The instrument’s rotational speed is indicated by the number facing up. Eight speed models
have a square speed control knob with two numbers on each face; by moving the knob through two
complete turns, speeds may be changed in sequence. No trouble will be experienced in differenti-
ating between the two speeds shown on each face since each pair is in the ratio of 20:1. To insure
rotation at the indicated speed, it is important that the face of the knob upon which this speed is shown
be closely parallel to the viscometer’s dial.
LV Viscometers are provided with a set of four spindles and a narrow guard leg; RV Viscometers
come with a set of seven spindles and a wider guard leg; HA and HB Viscometers come with a set
of seven spindles and no guard leg.
The spindles are attached to the viscometer by screwing them to the lower shaft. Note that the
spindles have a left-hand thread. The lower shaft should be held in one hand and lifted up. The
spindle should be screwed to the left. The face of the spindle nut and the matching surface on the
lower shaft should be smooth and clean to prevent eccentric rotation of the spindle. Spindles can
be identified by the number on the side of the spindle nut.
All Brookfield Dial Reading Viscometers are provided with a clutch lever located at the back of the
instrument. Depressing the lever raises the dial against the pointer and “holds” the instrument’s
reading. When the clutch is released the dial is lowered and the pointer will be freed.
Any of the controls on the Viscometer - the motor switch, speed change knob, and clutch - may be
operated independently of the other.
2. Attach the spindle (left-hand thread) to the viscometer lower shaft by lifting the coupling screw
slightly. Hold it firmly with one hand while screwing the spindle on with the other (note left-hand
thread). Avoid putting side thrust on the shaft.
To avoid trapping air bubbles under the disc-type spindles, do the following before attaching the
spindle. Begin by immersing the spindle in a diagonal path, slowly drag the spindle across the
3. Lower and center spindle in the test material (600 ml beaker) until the "meniscus" of the fluid is
at the center of the immersion groove on the spindle’s shaft.
4. To make a viscosity measurement, turn the motor switch “ON”. This energizes the viscometer
drive motor. Allow time for the indicated reading to stabilize. The time required for stabilization
will depend on the speed at which the viscometer is running and the characteristics of the sample
fluid.
When making a viscosity measurement, the reading should be noted and multiplied by the factor
appropriate to the viscometer model/spindle/speed combination being used. Refer to the table
below or to the FACTOR FINDER for calculating viscosity. For maximum accuracy, readings
below 10.0% torque (dial reading) should be avoided.
5. Turn the viscometer motor switch “OFF” when changing a spindle, changing samples, etc.
Remove spindle before cleaning.
6. The interpretation of results and the instrument’s use with non-Newtonian and thixotropic
materials is discussed in the booklet, “More Solutions to Sticky Problems” and in Appendix C.
The following tables apply to Brookfield Viscometer Models LV, RV, HA and HB with standard
spindles. They enable the user to convert the percent scale reading into a viscosity value in units of
centipoise.
To convert the viscometer dial reading to a viscosity value in units of centipoise, multiply the reading
noted on dial viscometer by the appropriate factor in the following tables.
LV Series Viscometer
1 2 3 4
.3 200 .3 1K .3 4K .3 20K
6 10 6 50 6 200 6 1K
12 5 12 25 12 100 12 500
30 2 30 10 30 40 30 200
60 1 60 5 60 20 60 100
1 2 3 4 5 6 7
.5 200 .5 800 .5 2K .5 4K .5 8K .5 20K .5 80K
1 100 1 400 1 1K 1 2K 1 4K 1 10K 1 40K
2 50 2 200 2 500 2 1K 2 2K 2 5K 2 20K
2.5 40 2.5 160 2.5 400 2.5 800 2.5 1.6K 2.5 4K 2.5 16K
4 25 4 100 4 250 4 500 4 1K 4 2.5K 4 10K
5 20 5 80 5 200 5 400 5 800 5 2K 5 8K
10 10 10 40 10 100 10 200 10 400 10 1K 10 4K
20 5 20 20 20 50 20 100 20 200 20 500 20 2K
50 2 50 8 50 20 50 40 50 80 50 200 50 800
100 1 100 4 100 10 100 20 100 40 100 100 100 400
HA Series Viscometer
1 2 3 4 5 6 7
.5 400 .5 1.6K .5 4K .5 8K .5 16K .5 40K .5 160K
1 200 1 800 1 2K 1 4K 1 8K 1 20K 1 80K
2 100 2 400 2 1K 2 2K 2 4K 2 10K 2 40K
2.5 80 2.5 320 2.5 800 2.5 1.6K 2.5 3.2K 2.5 8K 2.5 32K
5 40 5 160 5 400 5 800 5 1.6K 5 4K 5 16K
10 20 10 80 10 200 10 400 10 800 10 2K 10 8K
20 10 20 40 20 100 20 200 20 400 20 1K 20 4K
50 4 50 16 50 40 50 80 50 160 50 400 50 1.6K
100 2 100 8 100 20 100 40 100 80 100 200 100 800
HB Series Viscometer
1 2 3 4 5 6 7
.5 1.6K .5 6.4K .5 16K .5 32K .5 64K .5 160K .5 640K
1 800 1 3.2K 1 8K 1 16K 1 32K 1 80K 1 320K
2 400 2 1.6K 2 4K 2 8K 2 16K 2 40K 2 160K
2.5 320 2.5 1.28K 2.5 3.2K 2.5 6.4K 2.5 12.8K 2.5 32K 2.5 128K
5 160 5 640 5 1.6K 5 3.2K 5 6.4K 5 16K 5 64K
10 80 10 320 10 800 10 1.63K 10 3.2K 10 8K 10 32K
20 40 20 160 20 400 20 800 20 1.6K 20 4K 20 16K
50 16 50 64 50 160 50 320 50 640 50 1.6K 50 6.4K
100 8 100 32 100 80 100 160 100 320 100 800 100 3.2K
In taking viscosity measurements with the Dial Viscometer there are two considerations which
pertain to the low viscosity limit of effective measurement.
1) Viscosity measurements should be accepted within the equivalent % Torque Range from
10% to 100% for any combination of spindle/speed rotation.
2) Viscosity measurements should be taken under laminar flow conditions, not under turbulent
flow conditions.
The first consideration has to do with the precision of the instrument. All Dial Viscometers have a
full scale range precision of ± 1% for any spindle/speed combination. We discourage taking readings
below 10% of range because the potential viscosity error of ± 1% is a relatively high number
compared to the instrument reading.
The second consideration involves the mechanics of fluid flow. All rheological measurements of
fluid flow properties should be made under laminar flow conditions. Laminar flow is flow wherein
all particle movement is in layers directed by the shearing force. For rotational systems, this means
all fluid movement must be circumferential. When the inertial forces on the fluid become too great,
the fluid can break into turbulent flow wherein the movement of fluid particles becomes random and
the flow can not be analyzed with standard math models. This turbulence creates a falsely high
viscometer reading with the degree of non-linear increase in reading being directly related to the
degree of turbulence in the fluid.
For the following geometries, we have found that an approximate transition point to turbulent flow
occurs as follows:
Turbulent conditions will exist in these situations whenever the RPM/cP ratio exceeds the values
listed above.
Note that the Wells-BrookfieldCone/Plate Viscometer requires the use of a circulating temperature bath
controlling temperature to within +/- 0.1 °C. The bath is connected to the ports on the CP-44Y sample
cup. Brookfield offers a complete line of refrigerated and non-refrigerated (tap water cooling)
circulating water baths.
The following example assumes that the C/P Viscometer is set up on the Model A laboratory stand and
has been leveled. Cone CP-40 is used in the illustrations. The water bath is on, set at the same test
temperature that you will use when measuring the viscosity of your sample and is connected to the sample
cup (Figure A1). Allow enough time for the bath to reach the test temperature. The Viscometer speed
knob should be set at 10 or 12 rpm with motor OFF.
Adjustment
Ring
Cup
Outlet
Bath Bath
Outlet PUMP PUMP
Inlet
OUTLET INLET
Water
Bath Cup Sample
Inlet Cup
Figure A1
These surfaes
must be clean!
Spindle Wrench
Cone
Figure A2
Attach the sample cup and swing the clip under the cup to secure it in place. Take care to avoid hitting
the cone with the cup (Figure A3). Note: Do not add test sample during the gap setting procedure.
Do Not Hit
the CONE
with the CUP!
Figure A3
Start the Viscometer running at 10 or 12 rpm. Readings of % Torque (dial reading) will be utilized for
setting the cone/plate gap. If the display reading jumps to 0.5% of scale (or higher), or will not settle to
zero (indicating that the pins on the cone and cup are hitting), unscrew the adjustment ring by turning
it to the left (clockwise as you look down on the instrument) until the reading settles at 0.0% (Figure A4).
Brookfield Engineering Laboratories, Inc. Page 10 Manual No. M/85-150-P700
Index Mark
Adjustment Ring
Figure A4
Turn the adjustment ring to the right (counterclockwise as you look down on the instrument) in small
increments (one or two divisions of the ring) while watching the Viscometer dial (Figure A5).
Remember that you are trying to raise the cup so that the pin in the cone touches the pin in the cup. Once
you have found this hit point, you can "back off" to create the desired gap between cone and cup. You
must wait at least 6 seconds between movements of the ring.
Turn the adjustment ring until the display reading jumps from 0.0 to 0.5% (HA and HB models) or from
0.0 to 0.5 % (RV models) or 0.0 to 1.0% (LV models). This is the HIT POINT. The cup may have to
be adjusted up and down several times before this deflection is satisfactorily reached. The LV series
Viscometer will be the most sensitive.
Index Mark
Adjustment Ring
Figure A5
Brookfield Engineering Laboratories, Inc. Page 11 Manual No. M/85-150-P700
When you are satisfied the pins are just hitting (by observing acceptable % readings), make a pencil mark
on the adjustment ring directly under the index mark on the pivot housing (Figure A6).
Index Mark
Adjustment Ring
Figure A6
Turn the adjustment ring to the left (clockwise as you look down on the instrument) exactly the width
of one division from the pencil mark you have just made. This will separate the pins by 0.0005" (0.013
mm). This is a very important step because if it is not done, the pins may touch constantly and may cause
wear and damage to the cone, cup and instrument. The Viscometer is now mechanically set and ready
for sample measurement (Figure A7).
Index Mark
Adjustment Ring
with Index Mark
(enlarged)
Adjustment Ring
One Scale
Division
Figure A7
Less than
1 mm
Cone
Sample
Cup
Figure A8
Table A1
1) Ensure that the circulating bath used maintains the stated calibration temperature to within
± 0.1°C.
2) The attachment of the cone spindle and sample cup, and the gap setting between the cone and
cup must be accomplished by following "Cone/Plate Rheometer Set Up" Procedure.
3) Put the proper amount of viscosity standard fluid into the sample cup. The amount varies per
cone spindle (refer to Table A1).
5) Measure the fluid's viscosity and record the viscometer readings (both % torque and cP).
6) See "Interpretation of Test Results" shown below for calculation of total calibration tolerance
(instrument and fluid).
Notes: 1) The spindle must rotate at least (5) times before a viscosity reading is taken.
3) Select a viscosity standard fluid that will give viscosity readings between 10% and
100% of full scale range. Refer to Appendix B for viscosity ranges of cone spindles.
Do not use a silicone viscosity standard fluid with a viscosity value greater than 5000
cP with a Cone/Plate Viscometer. Brookfield offers a complete range of mineral oil
viscosity standards suitable for use with Cone/Plate Viscometers as shown in Table
D2 (Appendix D). Consult with Brookfield or an authorized dealer to determine
which fluid is appropriate.
Cone and Plate geometry, as illustrated in Figure A9, is the fixation of a conical vertex perpen-
dicular to and in point contact with a flat plate. When the cone is made very obtuse (θ less than
4°) and rotated at constant speed (ω), precise viscosity measurements are obtained at absolute
and uniform values of shearing rate and stress.
Viscosity (in poise) is the ratio of shear stress to shear rate. Shear stress is related to the summa-
tion of torque ( T ) over the conical surface. Shear rate is related to the cone rotational speed
(ω), and gap width (c) at any radial distance (r) from the center of the rotating cone.
ω Cone
Plate
c
θ
Figure A9
where:
T = % Full Scale Torque (dyne-cm)
r = Cone Radius (cm)
ω = Cone Speed (rad/sec)
θ = Cone Angle (degrees)
Angle Radius
Cone Spindle (deg.) (cm)
LV 673.7
RV 7,187.0
HA 14,374.0
HB 57,496.0
Note: When calibrating 2.4cm diameter cone spindles CP-51 or CP-52 at shear rates greater than
384 sec-1, use Brookfield Viscosity Standards 5,000 cP or lower.
LVF 15 100,000
LVT 15 2,000,000
RVF 100 2,000,000
RVT 100 8,000,000
HAT 200 16,000,000
HBT 800 64,000,000
Viscosity (cP)
SSA & T-Sel Shear Rate
Spindle (1/SEC) LVF LVT
16 0.29N 200 - 20,000 200 - 400,000
18 1.32N 5 - 500 5 - 10,000
25 0.22N 800 - 80,000 800 - 1,600,000
31 0.34N 50 - 5,000 50 - 100,000
34 0.28N 100 - 10,000 100 - 200,000
Viscosity (cP)
SSA & T-Sel Shear Rate
Spindle (1/SEC) RVF RVT
14 0.40N 6,250 - 625,000 1,250 - 2,500,000
15 0.48N 2,500 - 250,000 500 - 1,000,000
21 0.93N 250 - 25,000 50 - 100,000
27 0.34N 1,250 - 125,000 250 - 500,000
28 0.28N 2,500 - 250,000 500 - 1,000,000
29 0.25N 5,000 - 500,000 1,000 - 2,000,000
Viscosity (cP)
SSA & T-Sel Shear Rate
Spindle (1/SEC) HAT HBT
14 0.40N 2,500 - 5,000,000 10,000 - 20,000,000
15 0.48N 1,000 - 2,000,000 4,000 - 8,000,000
21 0.93N 100 - 200,000 400 - 800,000
27 0.34N 500 - 1,000,000 2,000 - 4,000,000
28 0.28N 1,000 - 2,000,000 4,000 - 8,000,000
29 0.25N 2,000 - 4,000,000 8,000 - 16,000,000
N = RPM
Viscosity (cP)
UL Shear Rate
Spindle (1/SEC) LVT RVT HAT HBT
YULA-15 or 15Z 1.22N 1.0 - 2,000 6.4 - 2,000 12.8 - 2,000 51.2 - 2,000
ULA-DIN-Y 1.29N 1.9 - 3,812 12.2 - 5,000 24.4 - 5,000 97.6 - 5,000
N = RPM
Cone/Plate Viscometer
Cone Shear Rate Viscosity (cP)
Spindle (1/SEC) LVTC/P RVTC/P HATC/P HBTC/P
CP-40 7.5N 0.5 - 1,028 3.3 - 6,540 6.6 - 13,080 26.2 - 52,400
CP-41 2.0N 1.9 - 3,840 12.3 - 24,576 24.6 - 49,152 98.4 - 196,600
CP-42 3.84N 1.0 - 2,000 6.4 - 12,800 12.8 - 25,600 51.2 - 102,400
CP-51 3.84N 8.1 - 16,180 51.2 - 102,400 102.4 - 204,800 409.6 - 819,200
CP-52 2.0N 15.5 - 31,065 98.3 - 196,600 196.6 - 393,200 786.4 - 1,572,800
N = RPM
Viscosity (cP)
T-Bar
Spindle LVF LVT
T-A 156 - 3,120 156 - 62,400
T-B 312 - 6,240 312 - 124,800
T-C 780 - 15,600 780 - 312,000
T-D 1,560 - 31,200 1,560 - 624,000
T-E 3,900 - 78,000 3,900 - 1,560,000
T-F 7,800 - 156,000 7,800 - 3,120,000
Viscosity (cP)
T-Bar
Spindle RVF RVT
T-A 2,000 - 100,000 2,000 - 400,000
T-B 4,000 - 200,000 4,000 - 800,000
T-C 10,000 - 500,000 10,000 - 2,000,000
T-D 20,000 - 1,000,000 20,000 - 4,000,000
T-E 50,000 - 2,500,000 50,000 - 10,000,000
T-F 100,000 - 5,000,000 100,000 - 20,000,000
Viscosity (cP)
T-Bar
Spindle HAT HBT
T-A 4,000 - 800,000 16,000 - 3,200,000
T-B 8,000 - 1,600,000 32,000 - 6,400,000
T-C 20,000 - 4,000,000 80,000 - 16,000,000
T-D 40,000 - 8,000,000 160,000 - 32,000,000
T-E 100,000 - 20,000,000 400,000 - 80,000,000
T-F 200,000 - 40,000,000 800,000 - 160,000,000
Rheological Properties
Fluids have different rheological characteristics that can be described by viscometer measurements. We
can then work with these fluids to suit our lab or process conditions.
Newtonian - These fluids have the same viscosity at different Shear Rates (different
RPM’s) and are called Newtonian over the Shear Rate range they are
measured.
Non-Newtonian - These fluids have different viscosities at different shear rates (different
RPM's). They fall into two groups:
Time Independent
Time Dependent
Thixotropic - A thixotropic material has decreasing viscosity under constant shear rate. If
you set a viscometer at a constant speed, recording cP values over time and
find that the cP values decrease with time, the material is thixotropic.
Brookfield publication, “More Solutions to Sticky Problems” includes a more detailed discussion of
rheological properties and non-Newtonian behavior.
Most fluid viscosities are found to be non-Newtonian. They are Shear Rate dependent on the
measurement conditions. The specifications of the viscometer spindle and chamber geometry will affect
the viscosity readings. If one reading is taken at 2.5 rpm, and a second at 50 rpm, the two cP values
produced will be different because the readings were made at different shear rates. The faster the spindle
speed, the higher the shear rate.
The shear rate of a given measurement is determined by: the rotational speed of the spindle, the size and
shape of the spindle, the size and shape of the container used, and therefore, the distance between the
container wall and the spindle surface.
1) Test temperature
2) Sample container size (or spindle/chamber geometry)
3) Sample volume
4) Viscometer model
5) Spindle used (if using LV (#1-4) or RV (#1-7) attach the guard leg)
6) Test speed or speeds (or the shear rate)
7) Length of time or number of spindle revolutions to record viscosity.
Container size: For Viscosity Standards <30,000 cP, use a 600 ml Low Form Griffin Beaker
having a working volume of 500 ml.
Conditions: The Viscometer should be set according to the operating instructions. The water
bath should be stabilized at test temperature. Viscometers with the letters “LV”
TABLE D1
SILICONE VISCOSITY STANDARD FLUIDS
5 5,000 HT-30,000
10 12,500 HT-60,000
50 30,000 HT-100,000
100 60,000 **25°C, 93.3°C, 149°C
500 100,000
1,000 Refer to Brookfield catalog
for more information.
or “RV” in the model designation should have the guard leg attached.
We recommend that Brookfield Viscosity Standard Fluids be replaced on an annual basis, one year
from date of initial use. These fluids are pure silicone and are not subject to change over time.
However, exposure to outside contaminants through normal use requires replacement on an annual
basis. Contamination may occur by the introduction of solvent, standard of different viscosity or
other foreign material.
Viscosity Standard Fluids may be stored under normal laboratory conditions. Disposal should be in
accordance with state, local and federal regulations as specified on the material safety data sheet.
1) Place the viscosity standard fluid (in the proper container) into the water bath.
2) Lower the Viscometer into measurement position (with guard leg if LV or RV series Viscometer
is used).
3) Attach the spindle to the Viscometer. If you are using a disk-shaped spindle, avoid trapping air
bubbles beneath the disk by first immersing the spindle at an angle, and then connecting it to the
Viscometer.
4) The viscosity standard fluid, together with the spindle, should be immersed in the bath for a
minimum of 1 hour, stirring the fluid periodically, prior to taking measurements.
5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate thermometer.
6) If the fluid is at test temperature (±0.1°C of the specified temperature, normally 25°C), measure
the viscosity and record the Viscometer reading.
Note: The spindle must rotate at least five (5) times before readings are taken.
7) The viscosity reading should equal the cP value on the fluid standard to within the combined
accuracies of the Viscometer and the viscosity standard (as discussed in the section entitled,
Interpretation of Calibration Test Results) which appears later in this section.
When a Small Sample Adapter is used, the water jacket is connected to the water bath and the water is
stabilized at the proper temperature:
1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies
with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction
manual).
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging
spindle (or directly attach the solid shaft spindle) to the Viscometer.
4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test
temperature.
Note: The spindle must rotate at least five (5) times before a viscosity reading is taken.
When a Thermosel System is used, the controller stabilizes the Thermo Container at the test temperature.
1) Put the proper amount of HT viscosity standard fluid into the HT-2 sample chamber. The
amount varies with the spindle used. (Refer to the Thermosel instruction manual).
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging
spindle (or directly attach the solid shaft spindle) to the DV-II+.
4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test
temperature.
Note: The spindle must rotate at least five (5) times before a viscosity reading is taken.
When a UL or DIN Adapter is used, the water bath is stabilized at the proper temperature:
1) Put the proper amount of viscosity standard fluid into the UL Tube. (Refer to the UL Adapter
instruction manual).
2) Attach the spindle (with extension link and coupling nut) onto the Viscometer.
4) Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket, connect the
inlet/outlets to the bath external circulating pump.
5) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test
temperature.
Note: The spindle must rotate at least five (5) times before a viscosity reading is taken.
T-Bar spindles should not be used for verifying calibration of the Viscometer.
1) Place the viscosity standard fluid (in the proper container) into the water bath.
2) Attach the spindle to the viscometer. Attach chamber (SA-1Y) and clamp to the viscometer.
3) Lower the viscometer into measurement position. Operate the viscometer at 50 or 60 RPM until
the chamber is fully flooded.
4) The viscosity standard fluid, together with the spindle, should be immersed in the bath for a
minimum of 1 hour, stirring the fluid periodically (operate at 50 or 60 RPM periodically), prior
to taking measurements.
6) If the fluid is at test temperature (+/- 0.1°C of the specified temperature, normally 25°C),
measure the viscosity and record the viscometer reading.
Note: The spindle must rotate at least five (5) times for one minute, whichever is greater
before readings are taken.
7) The viscosity reading should equal the cP value on the viscosity fluid standard to within the
combined accuracies of the viscometer and the standard (as discussed in the section entitled,
Interpretation of Calibration Test Results).
1) Ensure that the circulating bath used maintains the stated calibration temperature to within
± 0.1°C.
2) The attachment of the cone spindle and sample cup, and the gap setting between the cone and
cup must be accomplished by following "Cone/Plate Rheometer Set Up" Procedure (see
Appendix A).
3) Put the proper amount of viscosity standard fluid into the sample cup. The amount varies per
cone spindle (refer to Appendix A - "Cone/Plate Rheometer Set Up" procedure, Table A1).
4) Attach sample cup to viscometer and allow approximately 15 minutes for temperature equilib-
rium.
5) Measure the fluid's viscosity and record the viscometer readings (both % torque and cP).
6) See "Interpretation of Test Results" shown below for calculation of total calibration tolerance
(instrument and fluid).
Notes: 1) The spindle must rotate at least (5) times before a viscosity reading is taken.
3) Select a viscosity standard fluid that will give viscosity readings between 10% and
100% of full scale range. Refer to Appendix B for viscosity ranges of cone spindles.
Do not use a silicone viscosity standard fluid with a viscosity value greater than 5000
cP with a Cone/Plate Viscometer. Brookfield offers a complete range of mineral oil
viscosity standards suitable for use with Cone/Plate Viscometers as shown in TABLE
TABLE D2
OIL VISCOSITY STANDARD FLUIDS
When verifying the calibration of the Viscometer, the instrument and viscosity standard fluid error must
be combined to calculate the total allowable error.
The Dial Viscometer is accurate to ± 1% of any full scale spindle/speed viscosity range.
Brookfield Viscosity Standards Fluids are accurate to ± 1% of their stated value.
Example: Calculate the acceptable range of viscosity using RVF with RV-3 Spindle at 2 RPM;
Brookfield Standard Fluid 12,500 with a viscosity of 12,257 cP at 25°C:
1) Refer to the instructions on the FACTOR FINDER. Calculate the full scale viscosity range by
multiplying the spindle/speed FACTOR by 100.
Note: All spindle/speed factors found on the FACTOR FINDER are equivalent to 1% of
the spindle/speed full scale viscosity range.
2) The viscosity standard fluid is 12,257 cP. Its accuracy is ± 1% of 12,257 or ± 122.57 cP.
4) Therefore, any viscosity reading between 11,634.4 and 12,879.6 cP indicates that the viscom-
eter is operating correctly. Any reading outside these limits may indicate a viscometer problem.
The guard leg was originally designed to protect the spindle during use. The first applications of the
Brookfield Viscometer included hand held operation while measuring fluids in a 55-gallon drum. It is
clear that under those conditions the potential for damage to the spindle was great. Original construction
included a sleeve that protected the spindle from side impact. Early RV guard legs attached to the dial
housing and LV guard legs attached to the bottom of the pivot cup with a twist and lock mechanism.
The current guard leg is a band of metal in the shape of the letter U with a bracket at the top that attaches
to the pivot cup of a Brookfield Viscometer/Rheometer. Because it must attach to the pivot cup, the guard
leg cannot be used with a Cone/Plate instrument. A guard leg is supplied with all LV and RV series
instruments, but not with the HA or HB series. It’s shape (shown in Figure 1) is designed to accommodate
the spindles of the appropriate spindle set; therefore, the RV guard leg is wider than the LV due to the
large diameter of the RV #1 spindle. They are not interchangeable.
The calibration of the Brookfield Viscometer/Rheometer is determined using a 600 ml Low Form Griffin
Beaker. The calibration of LV and RV series instruments includes the guard leg. The beaker wall (for
HA/HB instruments) or the guard leg (for LV/RV instruments) define what is called the "outer boundary"
of the measurement. The spindle factors for the LV, RV, and HA/HB spindles were developed with the
above boundary conditions. The spindle factors are used to convert the instrument torque (expressed as
the dial reading or %Torque value) into centipoise. Theoretically, if measurements are made with
different boundary conditions, e.g., without the guard leg or in a container other than 600 ml beaker, then
the spindle factors found on the Factor Finder cannot be used to accurately calculate an absolute
viscosity. Changing the boundary conditions does not change the viscosity of the fluid, but it does change
how the instrument torque is converted to centipoise. Without changing the spindle factor to suit the new
boundary conditions, the calculation from instrument torque to viscosity will be incorrect.
Practically speaking, the guard leg has the greatest effect when used with the #1 & #2 spindles of the LV
and RV spindle sets. Any other LV (#3 & #4) or RV (#3 - #7) spindle can be used in a 600 ml beaker
with or without the guard leg to produce correct results. The HA and HB series Viscometers/Rheometers
are not supplied with guard legs in order to reduce the potential problems when measuring high viscosity
materials. HA/HB spindles #3 through #7 are identical to those spindle numbers in the RV spindle set.
The HA/HB #1 & #2 have slightly different dimensions than the corresponding RV spindles. This
dimensional difference allows the factors between the RV and HA/HB #1 spindles to follow the
same ratios as the instrument torque even though the boundary conditions are different.
The recommended procedures of using a 600 ml beaker and the guard leg are difficult for some customers
to follow. The guard leg is one more item to clean. In some applications the 500 ml of test fluid required
to immerse the spindles in a 600 ml beaker is not available. In practice, a smaller vessel may be used
and the guard leg is removed. The Brookfield Viscometer/Rheometer will produce an accurate and
repeatable torque reading under any measurement circumstance. However, the conversion of this torque
reading to centipoise will only be correct if the factor used was developed for those specific conditions.
Brookfield has outlined a method for recalibrating a Brookfield Viscometer/Rheometer to any measure-
ment circumstance in More Solutions to Sticky Problems, Section 3.3.10. It is important to note that for
many viscometer users the true viscosity is not as important as a repeatable day to day value. This
repeatable value can be obtained without any special effort for any measurement circumstance. But, it
should be known that this type of torque reading will not convert into a correct centipoise value when
using a Brookfield factor if the boundary conditions are not those specified by Brookfield.
The guard leg is a part of the calibration check of the Brookfield LV and RV series Viscometer/
Rheometer. Our customers should be aware of its existence, its purpose and the effect that it may have
on data. With this knowledge, the viscometer user may make modifications to the recommended method
ofBrookfield
operation to suit their
Engineering needs. Inc.
Laboratories, Page 27 Manual No. M/85-150-P700
RV Guardleg LV Guardleg
Figure D1
7 10
1
BROOKFIELD
LABORATORY
VISCOMETER
VS-17SY
UNIVERSAL EX-PROOF
CLAMP ASSEMBLY
16
12
14
13
15
11
4
ITEM PART # DESCRIPTION QTY.
1 VS-20 UPRIGHT ROD 1
2 VS-1 BASE 1
3 VS-3 LEVELING SCREW 3
4 VS-21 JAM NUT 1
5 VS-35 CLAMP 1
6 VS-40Y GEAR SCREW ASSEMBLY 1
7 VS-41Y CLAMP SCREW ASSEMBLY 1
8 VS-29 TENSION INSERT 1
9 VS-29W BELLEVILLE SPRING WASHER 2
10 VS-28 TENSION SCREW 1
11 50S103208S25B SCREW, #10-32 X 1/4 LG. SLOTTED SET 1
3 VSXA-17Y CLAMP ASSEMBLY 1
12
13 50S252024E140 SCREW, 1/4-20 X 3/4 LG. SOC. HD. CAP, 18-8 SS 1
14 VS-46Y EXPLOSION-PROOF GEAR SCREW ASSEMBLY 1
15 VS-45Y EXPLOSION-PROOF CLAMP SCREW ASSEMBLY 1
16 BLM-4E ROD EXTENSION - 4” LONG * OPTIONAL
*for use with Thermosel and Water Baths
Figure E1
Unpacking
Check carefully to see that all the components are received with no concealed damage.
1 base 1 jam nut
3 leveling screws 1 clamp assembly
1 upright rod
Brookfield Engineering Laboratories, Inc. Page 29 Manual No. M/85-150-P700
Remove the three (3) leveling screws from the base and discard the packing material. Remove the jam
nut from the upright rod.
Screw the leveling screws into the base. Insert the threaded end of the upright rod into the hole in the
top of the base and attach the jam nut to the rod on the underside of the base. With the rod gear rack facing
forward (toward the “V” in the base), gently tighten the jam nut.
Viscometer Mounting
Loosen the Viscometer handle retaining nut (if supplied) and slide it down the power cord. Slide the
Viscometer handle (if supplied) toward the cord and remove it from the instrument. Insert the
Viscometer handle core into the hole (with the cut-away slot) in the clamp assembly. Adjust the
instrument level until the bubble is centered from right to left and tighten the Clamp Screw (clockwise).
Note: The small clamp adjusting screw (Figure E1) on the clamp assembly should
be loosened or tightened as necessary to provide smooth height adjustment
and adequate support for the Viscometer.
Remove the hex socket screw from the clamp assembly and separate the clamp. Place the handle of the
Viscometer against the clamp/rod assembly and reinstall the clamp and hex socket screw. Adjust the
instrument level until the bubble is centered from right to left and tighten the Clamp Screw (clockwise).
Caution: Do not tighten the clamp screw unless the handle core is inserted in the clamp
assembly.
Center the Viscometer relative to the stand base and retighten the jam nut as required. Referring to the
Viscometer bubble level, adjust the leveling screws until the instrument is level.
Operation
Listed are some of the more common problems that you may encounter while using your Viscom-
eter. Review these items before you contact Brookfield.
❏ Check the voltage rating on your viscometer (115V, 220V): it must match the wall voltage.
❏ Make sure the speed set knob is set properly and securely at the desired speed.
❏ Inspect viscometer coupling and spindle coupling mating areas and threads for dirt: clean
threads on spindle coupling with a 3/56-inch left-hand tap.
❏ Inspect threads for wear; if the threads are worn, the unit needs service (see Appendix G).
❏ Check to see if spindles rotate eccentrically or wobble. There is an allowable runout for 1/32-
inch in each direction (1/16-inch total) when measured from the bottom of the spindle rotat-
ing in air.
❏ Check to see if the viscometer coupling is bent; if so, the unit is in need of service.
If you are continuing to experience problems with your viscometer, follow this troubleshooting
section to help isolate the potential problem.
❏ Turn the coupling until the red pointer reaches 15-20 on the dial.
If the pointer sticks and/or does not rest at zero, the unit is need of service. See Appendix G for
details on how to return your viscometer.
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix G for
details on how to return your viscometer.
Warranty
a ❏
❏
instrument.
Support pointer shaft with rubber band as shown in Fig-
ure G1.
Pack the instrument in its original case. Cases are avail-
able for immediate shipment from Brookfield. If the
case is not available, take care to wrap the instrument
with enough material to support it. Avoid using foam
peanuts or shredded paper.
DO NOT send the laboratory stand unless there is a prob-
lem with the upright rod, clamp or base. If there is a
problem with the stand, remove the upright rod from the
base and individually wrap each item to avoid contact
with the instrument. Do not put lab stand in viscometer
carrying case.
Fill out the Viscometer Information Sheet (included with
the information packet you received on purchase) with
as much information as possible to help expedite your
service. If you do not have this form, please include a
memo indicating the type of problem you are experi-
encing or the service you need performed. Please also
include a purchase order number for us to bill against.
Mark the outside of the shipping box with handling in-
structions, for example: “Handle with Care” or “Fragile
- Delicate Instrument”.
❏
❏
❏
Remove and return all spindles (properly packed for shipping).
Foam Insert
DIAL
Figure G1
or Tissue Paper
Figure G2
CONE/
PLATE
For cone/plate instruments, please remove the cone spindle and carefully pack in place in the
shipping case. If available, use the original foam insert or roll up one sheet of tissue paper (or
similar) and place between the spindle coupling and cup assembly (see FigureG2). This will
For repair or service outside the United States consult Brookfield Engineering Laboratories, Inc.
or the dealer from whom you purchased the instrument.
SAMPLE MODEL SPINDLE RPM DIAL READING FACTOR VISCOSITY SHEAR TEMP TIME NOTES
% TORQUE cP RATE °C
Page 35
CONCLUSIONS:
This tear-off sheet is a typical example of recorded test data. Please photocopy and retain this tem-