Calibration Procedure For BFP-T Turbine Rotor Expansion Sensor
Calibration Procedure For BFP-T Turbine Rotor Expansion Sensor
Calibration Procedure For BFP-T Turbine Rotor Expansion Sensor
Revision
Table of Contents
1. Purpose.................................................................................................................................. 4
2. Scope ..................................................................................................................................... 4
3. Reference Documents ......................................................................................................... 4
4. General Description ............................................................................................................. 4
5. Procedure .............................................................................................................................. 4
5.1) Pre-Checks before maintenance and calibration ............................................................ 4
5.2) Procedure for mounting of BFP turbine rotor expansion sensor:.................................. 4
5.3) Procedure for calibration BFP turbine rotor expansion sensor ..................................... 5
5.4) Procedure for Zero gap voltage setting (No calibration) ................................................ 6
5.5) Procedure for cable lay after GAP adjustment or Calibration ....................................... 6
6. Special Tools/Equipment .................................................................................................... 7
7. General Precautions ............................................................................................................ 7
8. Frequency.............................................................................................................................. 7
9. Responsibility ........................................................................................................................ 7
10. Record/Checklist .................................................................................................................. 7
11. Annexure-I ............................................................................................................................. 7
1. Purpose
The purpose of this document is to make understand the standard calibration procedure of
the BFP turbine rotor expansion Sensors.
2. Scope
There are a total Two No’s of expansion Sensors, one for each turbine.
3. Reference Documents
4. General Description
For the measurement of the BFP turbine rotor expansion 1 no of Eddy Current Sensors is
provided at Turbine Exhaust side. Sensor is mounted on BFP side of the measuring collar
facing towards collar. The Measurement range of sensor is 10 mm. expansion of shaft
towards BFP side is taken as negative from rotor zero position while the measurement
towards BP side is taken as positive. The measurement of the sensor is taken through driver
unit to the monitoring channel of Vibration monitoring panel
5. Procedure
Sensor 1004-527L-012
Driver VW-143PL-3
Extension Cable VK-143-P1
Monitor VM – 5D / 11012
a) Place the calibration kit VZ-30A-3 near the driver unit of HP rotor expansion sensor.
b) Mount the sensor in front of target plate.
c) Adjust the Sensors for Set gap of 3 mm for the zero position by adjusting the
micrometer.
d) Ensure the drivers are connected and communicating to application software.
e) Ensure - 24V DC supply healthiness at Driver Unit.
f) Ensure Sensor is placed at zero position (8 mm from the collar/target plate).
g) Ensure the upper and lower range is set in application software as per annexure –I
for BFP turbine including the safety gap of 3mm.
h) Check for Zero reading in Monitor/DCS and record the voltage reading.
i) if monitor is showing ZERO fix the position and if current reading after adjusting gap
is not ZERO use Zero Shift parameter to Set current position of ‘Rotor Zero’
j) Slide the sensor away from the target plate up to the gap of maximum range in
positive side and check the reading in DCS and the application software in measured
column. If the reading in measured in application software and DCS not matching
adjust the gain and keep the reading to maximum range.
k) Again move the sensor to zero position using the micrometer and matching the
recorded gap voltage for zero position.
l) Slide the sensor away from the target plate at the gap of 1 mm up to +5 mm and
record the reading of gap voltage at driver, DCS reading, and Vibration monitor
system PC reading.
m) Again move the sensor to zero position.
n) Slide the sensor towards the target plate at the gap of 1 mm up to -5 mm and record
the reading of gap voltage at driver, DCS reading, and Vibration monitor system PC
reading..
o) Compare the DCS and field reading.
p) Ensure the gap voltage from deviation from initial calibration report (Annexure-I) is
not very high.
q) Bring the sensors to Zero position
r) Remove the sensor from calibration kit and place safely.
s) Remove the calibration kit from the site.
a) Adjust the Sensors to set gap voltage as mentioned for zero position as per latest
calibration record.
b) Ensure the driver is connected to the sensor and
c) check for Zero reading in Monitor/DCS
d) if monitor is showing ZERO fix the position
e) Lock the sensor position by locking the sliding plate with screw nuts.
f) Tighten the sensor screw with the mounting bracket.
g) Give clearance to mechanical maintenance dept. for placing the bearing pedestal
covers.
6. Special Tools/Equipment
6.1. Digital Multi meter with range microvolt measurement range
6.2. Calibration kit no. VZ-30A
6.3. 3 no of walkie-talkie
7. General Precautions
Ensure healthiness & safety of the entire activities with respect to electrical and
Mechanical system.
Ensure the safety gap of minimum 3mm sensor while rotor is at zero position.
8. Frequency
Calibration to be done yearly once at the time of turbine overhauling or after replacement
of bearing Parts
9. Responsibility
C&I Engineers are responsible for monitoring and reviewing the effectiveness of the system.
The Authorized persons shall ensure that all instruments and instrumentation are regularly
checked and maintained in serviceable condition.
10. Record/Checklist
11. Annexure-I
Reference Calibration Sheet for TDBFP-1A turbine Rotor expansion