Service Manual

Ultra-Low Temperature Freezer


MDF-U76VA
Panasonic Healthcare Co., Ltd

Biomedical Business Unit

SM9910280
Effective models

This service manual is effective for following models.

Model name Product code Voltage and Frequency


MDF-U76VA-PA 90302651 115V 60Hz
I
11111111111111 Specifications 111111111111111

.Structural specifications
Item MOF-U76VA
Name Ultra-low Temperature Freezer
External dimensions W1010 x 0870 x H1990 (mm)
Internal dimensions W870 x 0600 x H1400 (mm)
Effective capacity 728 L
Outer door Painted steel
Inner door 2doors, ABS resin panel with stainless frame
Insulation Vacuum insulation panel + rigid polyurethane foamed-in place
Exterior Painted steel
Interior Painted steel
3shelves(adjustable), stainless steel
Shelf
Inner dimensions; W848 x 0533 (mm) Load; 50kg/shelf
Outer door latch 1pc
Outer door lock 1pc
Caster 4pcs (2pcs as leveling fooO
Access port 2 locations, Inner diameter; q>17mm
Back upper side: for back up nozzle
Bottom left back: for recorder sensor
Refrigeration circuit Cascade refrigerating system
Compressor High stage side: Hermetic type, Output: 750W
Low stage side: Hermetic type, Output: 11 DOW
Evaporator High stage side: Cascade condenser
Low stage side: Tube on sheet type
Condenser High stage side: Fin and tube type
Low stage side: Shell and tube type
Refrigerant High stage side: R-404A1n-pentane
Low stage side: R-508B/n-pentane
Refrigerant oil Ze-NIUSL22SA
Power supply 115V 60Hz
Transformer 3kVA booster
Battery Nickel-metal-hydride battery, DC 6 V, 1100 mAh, Auto-recharge (5HR-AAC)
Weight 370 Kg
Optional component Temperature recorder(MTR-G85)
Back-up kit: CVK-UB2/UB2(1)/UBN2
Communication kit: MTR-480, MTR-L03, MTR-5000
Storage lack: IR-224U, IR-220U
Inner door (small) : MOF-7101
• Air intake port is at left side of unit (for release inside pressure and validation).
.Control specifications
Item MDF-U76VA
Temp. controller Micro-processor control system
Temperature setting range: -50°C--90°C (Unit :1°C)
Non-volatile memory
Thermal sensor pt.1000n
Temperature display Blue LED digital display (Unit :1°C)
When a chamber temperature becomes set temperature +5°C-+40°C (Factory
default: +1Q°C), ALARM lamp blinks, audible alarm sounds intermittently after
15minutes past.
High temperature alarm
Remote alarm contact: Normal Open, Normal Close
Contact turns over after 15minutes past.
Allowable contact capacity: Max. 30VDC, 2A
When a chamber temperature becomes set temperature -5°C--40°C (Factory
default: -1Q°C), ALARM lamp blinks, audible alarm sounds intermittently after
15minutes past.
Low temperature alarm
Remote alarm contact; Normal Open, Normal Close
Contact turns over after 15minutes past
Allowable contact capacity: Max. 30VDC, 2A
Door alarm DOOR lamp illuminates when a door is kept opening for 2minutes.
Filter alarm FILTER lamp illuminates and audible alarm sounds intermittently.
ALARM lamp blinks, audible alarm sounds intermittently and remote alarm
Power failure alarm
contact outputs.
Remote alarm terminal 3P : Max. DC30V, 2A N.C.-COM, N.O.-COM
Remote alarm When a temperature alarm or power failure alarm occurs, or when a sensor is
failed, remote alarm contact turns over.
Notice of battery life When battery life expires (approx. 3years), BATTERY lamp illuminates.
Notice of fan motor life When fan motor life expires (approx. 6years), BATTERY lamp blinks.
Status-1: If an AT sensor temperature is lower than O°C or higher than +35°C, a
unit diagnoses that the ambient temperature should be abnormal.
Status-2: If a power supply voltage is poor (15% lower than rated voltage), a unit
STATUS function
diagnoses that the power supply voltage should be abnormal.
Status-3: If a running rate in low stage side compressor is more than 95%, a unit
diagnoses that the unit is operating overloaded.
Lamps: ALARM, BATTERY, STATUS, DOOR, FILTER
Lamps and keys on Buzzer stop key: BUZZER Alarm test key: ALARM TEST
Control panel Status key: STATUS Set key: SET
Digit shift key: ~ Numerical value shift key::it:
Press digit shift key for 5 seconds to display Key Lock mode.
Key Lock
LO: Key Lock is OFF L1: Key Lock is ON
When a cascade sensor temperature is lower than -34°C, low stage side
compressor turns on. When a cascade sensor temperature is higher than -12°C,
Compressor protection low stage side compressor turns off. When a filter sensor temperature is higher
than +56°C, high stage side compressor turns off.
Overload relay
If there are several units in a same site and a power failure is occurred, their start
can be controlled by "Start delay time" to prevent them being active
Start delay time
simultaneously.
Setting range: 3-15 minutes (Unit: 1 minute)
.Performance specifications
Model MDF-U76VA
Maximum cooling performance -86'C at the center of the chamber (AT30'C, no load)
Temperature control range -50'C~-86'C (AT30'C, no load)
Power source 115V, 60Hz
Rated power consumption 1020W
Noise level 49 dB [A] (background noise; 20dB)
Maximum pressure 2600 kPa
Usable conditions AT; +5'C-+30'C Humidity: Less than 80%RH
• Design or specifications will be subject to change without notice.

The recommended wire gage to the outlet from distribution box (breaker box from electrical
supply to receptacle unit is plugged into) is dependent on 'length of wire', the following
information is a good rule of thumb to follow:

• 60ft or less 12ga


• 60ft - 100ft 10ga
• 100ft -150 ft 8ga
• 150ft - 250ft 6ga
Dimensions

(6480m)

'E:::r
- 33,4' (TraY)
-
e
-0

~===t====
I'
II
II
,I
II
,
'I
II
II
II
II
.,
EO
';:
m
~

t:

'"InOJ
"'-
:~
II
~~~~~~
J ----- ------:-3"
-
Top

(1076mm) (33m.) (810mm) (I05mm)


A2,4" 1 3'_ 34 3' 4 I'
(4Smo) (101000) (21 mm) (795m m)
1,8' _ 3S,8' O. S' 31 .3"
(670 •• ) (sao •• )
34.3' 23. S'
5mtn} (Smm
0,2' o. '

1\

~ +
t f
1 J
i _________ ~ E,
E'".
o
E,
e-
o .
f-----------~ "'-
In"'~
e,
0",
"'In

t : e'"

"""
"'.-
1!11 :

1
~lL
="Ii
.~" __________

JI
I
I

E,
eO)
o .
"''''
3-
L

Front Side
Po ..... er supply cord exit posl1lor.
{Main body back}
Cooling unit parts

MDF-U76VA
Specifications
Item
H side L side
Compressor
Type KS240J1NS-7A KS370J1 NS-7A
Code 7FB-0-M101-011-06 7FB-0-M101-001-06
Rating 220V, 60Hz
Refrigerant oil Ze-NIUS L22SA, Charged q'ty 850cc
Cooling system Forced air cooling (partially) and oil cooler
Starting relay AMVL-300TA AMVL-300A
Overload relay MRA999549201
Starting capacitor 1601JF/250VAC x 2 1601JF/250VAC x 2
Running capacitor 151JF/400VAC 251JF/400VAC
Condenser Cascade condenser
Type Fin and tube Coil pipe </l6.35
Condenser 12 columns x 4 lines, P6.35mm
-----
Fin 88pcs
Pre-condenser W350mm -----
Frame pipe </l6.35 -----
Evaporator Cascade condenser Tube on sheet </l9.52
Type Shell and tube </l80 -----
Accumulator </l38 -----
Capillary Ex. capillary
Resistance
78PSI 0.37Mpa 34PSI
PSI· kg/cm 2
Length 3000mm 3000mm 500mm
Outer diameter </l2.4mm <!l2.0mm <!l2.4mm
Inner diameter </l1.2mm </l0.9mm <!l1.2mm
R-404A, Charged q'ty 540g R-508S, Charged q'ty 370g
Refrigerant
n-Pentane (4wt%) 22.5g (36cc) n-Pentane (12.4%wt) 52.4g (83cc)
Dryer 3AXH-9, Char!]ed q'ty 18!] 4AXH-6, Char!]ed q'ty 58!]
Condensing fan Material: ASS, 4 blades, </l230mm -----
Condensing fan SV4-11AS5P
-----
motor runnin!] capacitor: 1.01JF
Thermostat, etc Thermistor, 502AT PT1000n
Heater ----- Capillary heater, 12W x 2P
Oil separator ----- Ze-NiUS L22A: 445cc (384.8g)
1111111111111 Refrigeration circuit 111111111111

H Ish temp, -;p-


Low temp, -»
HIGH STAGE.
SIDE COMPRESSOR

-;p- -;p-
PRE-CONDENSER
<E-
Ii.I
FRAME PIPE

I PRE-CONDENSER
Ii.I YJ Ii. ~/\/\/\.r-----'
I

SUCTION CONDENSER
HEAT
«- EXCHANGER "*"" <E- I

"*"" YJ
~ '"w f CASCADE
0-

"''" CONDENSER
t --7>

~
w
d >:
=>
'-'
ffi
W :i
OIL ~ ~15~Wc~~2~sSOR
"'
'" (D SEPARATOR
~
C>

t ~ --7>
CAPILLARY
TUBE (L EX)
~ EXPANS ION TANK
--7>
CAPILLARY TUBEIL)
HEAT EXCHANGER
nnnPlntc::: on PCB
CN8
CNll #1-#2: Battery
#1-#3: Temp. sensor

CNl
CN7 #1-#3:Switching
#1-#2: Door switch
#5-#6: AT sensor
o power supply

#7-#8: Filter sensor


#9-#10: Cascade sensor
CN2
MTR-480C(option)

CN9
#1-#2: H. Camp relay

CN6
Display PCB
CN4
#1-#2:Temp. control
relay
#3-#4: Heater relay

CN5 CN3
#1-#5: Switch PCB Remote alarm terminal
#l:COM.

~~~::2=rJ~~!::~::~o1-
~ 1F 0
#2: N.D.
#3:N.C.

o
Connections on PCB

Connections on Temp controller PCB

Connector Connects to Usage

CN1 #1, #3: Switching power supply To supply the power to PCB.

CN2 Network interface To connect to MTR-4S0/L03 (option)

Remote alarm terminal


Remote alarm contact outputs.
#1: COM.
CN3 In normal condition, open for #1-#2 and closed
#2: N.O.
for #1-#3.
#3: N.C.

#1-#2: Temp. control relay To control internal temperature (12VDC)


CN4
#3-#4: Heater relay To supply the power to capitube heater (12VDC)

#1-#5: Switch PCB To connect to each switch


CN5
#6-#7: Buzzer PCB To connect to buzzer PCB

CN6 Display PCB To connect to each LED

To detect door ajar


#1-#2: Door switch
To detect the ambient temperature
#5-#6: AT sensor
CN7 To detect the temperature in condenser outlet
#7-#S: Filter sensor
pipe.
#9-#10: Cascade sensor
To detect the temperature in cascade.

#1-#2: Battery(#1 :6V #2:Battery switch) To supply the power during power failure
CNS
#3-#4:Transformer

CN9 #1-#2: H. Compo relay To control compressor H ON/OFF (12VDC)

CN10 Unused

CN11 #1-#3: Temp. sensor To detect the internal temperature.


III Electrical Parts

MDF-U76VA AC115V,60Hz

Compressor (H) Type KS240J1NS-7A


Code 7FB-0-M101-011-06
Rated voltage (50/60Hz) 220V, 60Hz
Winding resistance C-S(Aux) 1.7811
C-R(Main) 4.3411
Compressor (L) Type KS370J1NS-7A
Code 7FB-0-M101-001-06
Rated voltage (50/60Hz) 220V,60Hz
Winding resistance C-S(Aux) 1.6411
C-R(Main) 3.3511
Starting relay (H) Type AMVL-300TA
Rating AC300V
Starting relay (L) Type AMVL-300A
Rating AC300V
Overload relay (H), (L) Type MRA999549201
Rating 29.5A
Electrolytic capacitor (H) Rating 250VAC, 160.u F
Electrolytic capacitor (L) Rating 250VAC, 160.u F
Running capacitor (H) Rating 400VAC, 15.u F
Running capacitor (L) Rating 400VAC, 25.u F
Condensing fan motor Type SV4-11AA5P
Rating 220-240V
Capitube heater Rating 230V, 11.2W
H Compo relay Type AJM5211F
Contact capacity 20A
Heater relay Type G2R-1A-T
Contact capacity 10A,250VAC
Switching power supply Type ZWS10-12/J
Rated output DC12V, 0.9A (TDK)
Power supply switch Type 1R11AZE201 R
Rating 20A, 250VAC. Breaker SW
Temp. sensor Type THC-663
Rating 100011
AT sensor Type 502AT
Rating 5KI1,25"C
Filter sensor Type 502AT
Rating 5KI1,25"C
Cascade sensor Type 502AT
Rating 5KI1,25"C
Battery switch Type SLE6A2-5
Rating 250VAC,4A
Battery Type 5HR-AAC
Rating 6V,1100mAH
Step-up transformer Type E168-3000
Rating S115V, P225V, 13.3A, 3000VA
Fuse 20A
cations of sensor

The following shows the temperature in thermal sensor (502AT-1) and its resistance value.
Temperature Resistance Temperature Resistance Temperature Resistance Temperature Resistance
ee) (kQ) ee) (kQ) ee) (kQ) ee) (kQ)
-50 154.5 -36 71.80 -22 35.65 0 13.29
-49 145.9 -35 68.15 -21 33.99 5 10.80
-48 137.8 -34 64.71 -20 32.43 10 8.84
-47 130.2 -33 61.48 -19 30.92 15 7.20
-46 123.1 -32 58.43 -18 29.50 20 6.01
-45 116.5 -31 55.55 -17 28.14 25 5.00
-44 110.2 -30 52.84 -16 26.87 30 4.17
-43 104.4 -29 50.23 -15 25.65 35 3.50
-42 98.87 -28 47.77 -14 24.51 40 2.96
-41 93.70 -27 45.45 -13 23.42 45 2.51
-40 88.85 -26 43.26 -12 22.39 50 2.13
-39 84.18 -25 41.19 -11 21.41 55 1.82
-38 79.80 -24 39.24 -10 20.48 60 1.56
-37 75.67 -23 37.39 -5 16.43 65 1.35

The following shows the temperature in thermal sensor (PT1 0000) and its resistance value.
Temperature Resistance Temperature Resistance Temperature Resistance
(Oe) (kQ) ee) (kQ) ee) (kQ)
-140 450.83 -70 729.99 0 1000.0
-130 491.47 -60 769.02 10 1038.0
-120 531.83 -50 807.87 20 1076.0
-110 571.92 -40 846.58 30 1113.8
-100 611.76 -30 885.13 40 1151.4
-90 651.38 -20 923.55 50 1189.0
-80 690.78 -10 961.84 60 1226.4
'-J
..,., --- BATTERY
rltl
~
ttl [OPTION
BK
3:(J\
I MTR-480
MTR-L03 rtl.
l4J BATTERY
3P
TERMINAL
t:J I
..,.,
SWITCH
~ ~
L. ::::l.
jfi'!J

IT
'"0
IN. ::J
[
-I>- (0
C -I>- c:::J~ R BL BK
--,J C-
(J\ >-'
I eN2 eNG eNs al
<
J>
-I>- (0
"'I
>-'
I \Jl al
-0 I
J> :1 l.b.!!L. 1 13
0
0
[
0

~
I---l
10
I---l
Z
Q BL
-;:Jj
o
I---l
BL I -..:;: I
DOOR SWITCH

L.BL
J>
Q L.BL
CAPITUBE HEATER
10 L,BL
J> STARTING, CAP.L' STARTING CAP.L

:s:
/ il:'
r
;,
AT
SENSOR Gl
(';;" y
'II I~
~ FILTER BK

, ~lJ ~
CONDENSING HARNESS PN r
R F'ANMOTOR CAPITUBE , V VI SENSOR S BK ~
S FM HEATER ~r;!
(,1:;::
~
-
~ (t)
BK
'~;:g
!il" Pl
~
-<
~
PK CAPITUBE ~
Cl r
tD
BR
HEATER
''if ~~
n'"
TERMINAL
~ ~
r :t:!;l :J ~
bf=~ ~
@
@ ©-=-1"'1 ~ g
"oioiH-,,""I'L-- BK ","0 GNIY
r C'l '::t'l
r x~
00
~~ ~ <:::
(11""1
" "
READ WIRE ASSY rAN DTDR 1"1~
>n
~3: Z -< ~E
~ ~
("1 Pl
COMPRESSOR H ,..,
5k;
'" <
5t
~Cl
:r"
"
:to
"
"
:t: ~~RNESS
'"
@~
L,"O

...........
BL FUSE
HOLDER BK c~
R
YL4P V

~
S (VI)
OPTION
REMOTE
ALARM,
YL3
('yI L ~8
... D
TERMINAL BATTERY ~
GN/Y
EARTH
~;:gr-
.,r '"
r
IlK IlK ....
~~
~
PO'w'ER
MTR-480

:~ ~ ~S'HITCH I~~ .1ll:..


sM4P(W')
~;o

r=J91 ttf, . _.=_..


..... f1,;I TRANS ~
HARNESS H it v "'tI

HA H HARNESS P Y BLR BL HARNESS DPHD I ~


: EC

(- HARNESS EC
HARNESS CH
HARNESS P

PK I I TT I
rJI eN
Y R BL
eN
r-1.r.1
I
III
[

I~I~I~~
PK BK BK BK LBL L.BL LBL L,BL
1 2 3 -4 5 6 7 8 9 10 11 ,12
III I~
~
I~TH T]VVV[Vl
PK BR BK BK BR BK W LBL LBL L.BL LBL
GNlYi

I 1III I
III •
'PK

,.
I IVVVVI
I :0
CN12

I:[~
FR'tNE
EARTH LBL

NESS H
L.BLI vi BK ~
1L
eN'

I L- eN· o

~ ,!8 .........
~-~
.... STARTING' BK BK
m
1:1
'-
BK STARTING
Cltdb:lt:d RELAY L RELAY H
:;u;:orr
~~fl~
:t:
ili.cb.
,,""" I. 2D- ...
~

10,,0...(
"0 '"
~ ~ ~
FANMOTrlR
CAP,
2

1
01
3
;Ilj-
<;;
<
0
:0
CN13

~
::O"l]CC'I

1:C~
fTl .... ::O;:o
t::Jz·;r:. ;r,.
7\~-<
Iv N
'"
Y~
~ ~Ni

e1 k IY

3[[: I:NI
r
;,
I....BL Ii" CNS S:W'ITCHING EARTH
r«-< 1 POWER
................. SUPPLY
r<:(..-(
"...
20
......... :::J:,..,
G'lo ... ' "
n" "
nc
'" " lii=l' 1t
..J:I>
:cr-lr"
-If'TI"lC
",'" <:
"'"
"'"
"O~ ~~ ".z
"OZ
n"
""~ .,," ""z" ~:;! nC 1
rz
~
r~
f-* '"
:0"
:t: z
'"
:-o-a:
!J:~ :t::z BATTERY : HS
IBK IR

'"
. 111111111111111 Circuit diagram 111111111111111

~
...
-
,
~
;~ ~ .
i ui ~ ~
l 0~ !~ ~
"
~
< ~

0...
...
",'-'1.-.1
..::.:, ~,
O.Ol"~ L_1'I1.;-z--] !~~~
0g *E *0: 0:~
~
,:::::~,ya
0,01.1" ~
t;: I'JU ."'~
.m.,.L
SET

[)(>
..'-'" t- ;
: h
Ir
""...Lo--
~-u ~
:l'ff'1 DC +12V

i~~ '~ '"'~ CONOI TlON


.Rl.4 :~.

~
. ~. I~ ""~ BUZZER
,---
.~. ,-1.0-
,--1 0- ALAR\,{ TEST

:
I ~~ ~ ,I :

•• m•• """'"
~. I
1'3.8 :, ,,}- ,. .. -- -- -- - --- _______ ,
!S28£~~~~-s : .
P3.5
,
1'3 ••

I::: 1m
,
~i f\ i
0

" . : o
£8Z0!!UC-,V:
L---SOZZe-fn·wif----J
0

.?
~ tl'IO~2UHl •.. ~11
"o,-.!.+.r--h-1r-rrH-+1'-H:'i , 1'2.0
~wP1
1 H 1
~ ~;¥~;~;;~;t If: ~
1111111
1'2.1
P2.2
'"'' ""
'" '"''
"'"
"';;::'
Pl!.7

,
,
1 • JJIJ
to It lli'-~ lli~ I- I" Ii-
0 0

~
2 ~ ~
2 0 W W
~
u
0
oc
<
J
~.

c;
~
~

~
~
0
0
0
..,
~
~

~
J < z
0
u

DISPLA.Y PWB

I I '"
,I
!ila~5
"'0-"'0-

~
0
0
e~
0
0 g
~ ~
IIIIIII Control specifications IIIIIII
1. Key and Switch
BUZZER In alarm condition, audible alarm will silence when this key is pressed, but
remote alarm will be still active and message will not be eliminated.
When a power failure is occurred (during battery back-up), press this key to call
a chamber temperature display for 5 seconds.

ALARM TEST When this key is pressed, unit will step to Alarm Test mode and ALARM lamp
will blink, audible alarm will sound intermittently, digital LED display will go off
and remote alarm will be activate.
A chamber temperature display will appear after approx. 90seconds elapse.
(Auto return)
If Alarm Test is performed when a battery switch is in off position, "E09" will
flash.

SET Press this key to step to setting mode and the 2"' digit in LED display will blink.
Press this key again to memorize a value. Also used to decide and memorize
value in each function and setting mode.

STATUS Press this key during STATUS lamp is on to display occurring status code
("--1", "--2", "--3"). Unused when STATUS lamp is off.
~ In setting mode, press this key to move curser among 1" digit - 3'" digit.
(Digit shift key) If this key is pressed for 5 seconds when a chamber temperature is displayed,
"L_O" will be displayed and go into key lock setting mode.

It] In setting mode, press this key to add numerical value in a displayed digit.
(Numerical value If this key is pressed for 5 seconds when a chamber temperature is displayed,
shift key) "FOO" will be displayed and go into function mode. (use this key and digit shift
key to input function code and press SET key to go into each function mode)

2. Temperature control
Setting range -50°C--90°C
Display range -180-50
How to set Press SET key to step to setting mode and change a value by using It] key
temperature and ~key.
Press SET key to store the value in non-volatile memory.
Unacceptable setting If a value which deviates from setting range is input and SET key is pressed,
value error sound come out and keep setting mode.

3. Key Lock mode


Setting range
How to set Key lock:
° (Release), 1 (Lock)
In a chamber temperature display, press ~ key for 5 seconds to step to
Key Lock mode. ("L_O" or "L_1" is displayed. Factory default display: L_O)
Change a blinking digit to "0" or "1" by using It] key.

4. Function mode
Setting range 00-50
Display range 00-59
00, 16 and 33-43, 44-49, 51-59 are unused.
How to set Function In chamber temperature display, press It] key for 5 seconds to step to
mode: function mode and "FOO" is displayed.
Change a value by using It] key and ~ key.
Press SET key to go into each function mode. If "00" or "16" is input and
press SET key, return to chamber temperature display.

Unacceptable setting f a value which deviates from setting range is input and SET key is pressed,
value: error sound come out and keep setting mode.
5. Error codes
E01: Temp. sensor is open circuited
E02: Temp. sensor is short circuited
E03: Cascade sensor is open circuited
E04: Cascade sensor is short circuited
E05: Filter sensor is open circuited
E06: Filter sensor is short circuited
EO?: AT sensor is open circuited
E08: AT sensor is short circuited
E09: Battery switch is in off position or battery is unconnected
E10: Compressor temperature is abnormal

(1) Temp. sensor


Open circuit (E01): If a temp. sensor temperature is higher than 50°C, E01 and "50" will be
displayed alternately, audible alarm will sound intermittently and remote
alarm contact will be active.
Compressor will be kept running.
Press BUZZER key to silence audible alarm.
Short circuit (E02): If a temp. sensor temperature is lower than -1?0°C, E02 and "-1?0"- "-180"
will be displayed alternately, audible alarm will sound intermittently and
remote alarm contact will be active.
Compressor will be kept running.
Press BUZZER key to silence audible alarm.

(2) Cascade sensor


Open circuit (E03): If a cascade sensor temperature is lower than -65°C, E03 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Both High and Low side compressors will be forced to turn off.
Press BUZZER key to silence audible alarm.
Short circuit (E04): If a cascade sensor temperature is higher than 60°C, E04 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Both High and Low stage side compressors will be forced to turn off.
Press BUZZER key to silence audible alarm.

(3) Filter sensor


Open circuit (E05): If a filter sensor temperature is lower than -60°C, E05 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
High stage side compressor will be forced to turn off.
Press BUZZER key to silence audible alarm.
Short circuit (E06): If a filter sensor temperature is higher than 130°C, E06 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Press BUZZER key to silence audible alarm.

(4) AT sensor
Open circuit (EO?): If an AT sensor temperature is lower than -60°C, EO? and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Regardless of ambient temperature, operate warm up starting of L side
compressor when AT sensor error is happened.
Press BUZZER key to silence audible alarm.
Short circuit(E08): If an AT sensor temperature is higher than 60°C, E08 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Regardless of ambient temperature, operate warm up starting of L side
compressor when AT sensor error is happened.
Press BUZZER key to silence audible alarm.
(5) Battery SW is off If ALARM TEST key is pressed when battery switch is off position or battery
position or battery is unconnected, E09 will be displayed.
is unconnected(E09):

(6) Compressor If a filter sensor temperature is higher than 54°C, it regards as fan motor
abnormal failure or abnormal compressor temperature.
temperature (E1 0): E10 and chamber temperature will be displayed alternately and high stage
side compressor will be forced to turn off.
Press BUZZER key to silence audible alarm.
If a value which is obtained from filter sensor temperature subtracts an
ambient temperature is equal or lower than 10°C, compressor will turn on.

(7) Error code priority


No.1: Cascade sensor error (E03, E04) ... Compressor is forced to turn off
No.2: Filter sensor error (E05, E06) ... Compressor protection is uncontrollable
No.3: Abnormal compressor temp.(E10) ... Compressor temporary turns off
No.4: Temp. sensor error (E01, E02) ... Compressor is forced to turn on
No.5: AT sensor error (E07, E08) ... Warming-up is forced to be performed in any
ambient temperature.

6. Warning function
High temperature If a chamber temperature is equal or higher than set temperature + high
alarm: temp. alarm set temperature +1°C, ALARM lamp and LED display will blink,
audible alarm will sound intermittently after 10 minutes of delay, and remote
alarm will be active.
If a chamber temperature is equal or lower than set temperature, ALARM
lamp will be off, LED display will blink, audible alarm will silence and remote
alarm will turn off.
Press BUZZER key is to silence audible alarm, but remote alarm will not be
inactive.

Low temp. alarm If a chamber temperature is equal or lower than set temperature - low temp.
alarm set temperature _1°C, ALARM lamp and LED display will blink, audible
alarm will sound intermittently after 10 minutes delay, and remote alarm will
be active.
If a chamber temperature is equal or higher than set temperature, ALARM
lamp will be off, LED display will blink, audible alarm will silence and remote
alarm will turn off.
Press BUZZER key to silence audible alarm, but remote alarm will not be
inactive.

Door alarm If an outer door is left open, DOOR lamp (DP54: red) will turn on. Audible
alarm will sound after 1-15 minutes (default: 2 minutes) of delay. Audible
alarm and remote alarm are not synchronized. Audible alarm will keep
silence by pressing BUZZER key (no ring back function).

Power failure alarm If power supply switch is off or power interrupt during battery switch is in ON
position, it regards as power failure condition.
If a power failure condition keeps for 3 seconds, ALARM lamp will blink,
audible alarm will sound intermittently and remote alarm will be active.
If a power returns within 3 seconds, a unit will start operation at default
settings and remote alarm will be off.
Press BUZZER key to silence audible alarm, but remote alarm will not be
inactive. Remote alarm will keep active until chamber temperature is
stabilized after the power interruption.
Press BUZZER key during power failure condition to display chamber
temperature for about 5 seconds.
7. Status function
When the unit become following 3 conditions, STATUS lamp (DP52: orange) turns on. Press STATUS
key during this condition to display each status code.
Status 1: When AT sensor temperature is lower than O'C or higher than 35'C, display "--1 ".
Status 2: Decrease power supply voltage (between TP7 and TP3: lower than CD2.01V), display
"__2",
Status 3: Running rate is over 95%, display "--3".
Display method
When more than one status is occurred at same time, press STATUS key to display status code in
following order.
" ___ " ---+ "--1" ~ "--2" ---+ "--3" -+ back to "--1"
If there is no status notice, previous status code is displayed (interval is about 3 seconds)
" ___ " ---+ "--1" --+ "--2" --+ "--2" --+ back to "--1"
(when status 1 and 2 are occurred, status 2 is displaying for about 6 seconds)
Return to chamber temperature display after about 90 seconds or all status is disappear.

8. Running rate
Running rate = (ON time I (ON time + OFF time)) x 100%
Measurement of running rate:
It regards as 'cycle start' when a compressor turns on after it turned off once
chamber temperature was lower than set temperature.
Running rate should be measured on and after 2 hours then.

ON time (Min.) = Time until P3.1 in IC5 first reaches from LOW to HIGH
OFF time (Min.) = Time until P3.1 in IC5 reaches from HIGH to LOW

Condition to calculate running rate: 2 hours after cycle start

ON time(Min.) OFF time(Min.) Running rate


-0 -0 Impossible to. obtain
>0 -0 running rate by
-0 >0 calculation (=0%)
>0 >0 0-100%
Note) Running rate cannot be measured when a chamber temperature becomes higher than set
temperature + 4.0'C.
Please wait until a chamber temperature is stabilized.

Obtain for diagnosed value of overload running rate:


Step to F20 and input '000'.

Start to calculate Start to diagnose running rate


running rate

Settemp.

Compressor OFF (1st) 8 hours stabilized Diagnosis is done with every 8 hours
2 hours
Reach to set temp. (Cycle run) until chamber temp is higher than
Set temp + 4.0
Calculation for diagnosed running rate
=(-(Set temp.)XO.9)+((ATXO.9-4.5°C))-((Set temp.+85°C) 110))

Diagnosis of running status


If diagnosed value running rate is more than actual running rate, it diagnosed as 'Normal'
operation (DP52: off).
If diagnosed value running rate is equal or lower than actual running rate, it diagnosed as
'Overloaded' operation (DP52: on).
If it is impossible to obtain actual running rate (=222), it is impossible to diagnose status (DP52:
off).
Diagnosis will be done in every 8 hours.

9. Other functions
Auto Return: If there is no key operations for 90 seconds in setting mode, Key lock mode
and Function mode, return to chamber temperature display.
Note) Auto return does not work in FOg and F10.

Ring Back: In alarming condition, audible alarm will sound again if someone presses
BUZZER key to silence audible alarm. This is to prevent that buzzer sound is
stopped by third person and the operator cannot notice buzzer.
Ring Back time can be set in F25.

Display of sensor F12: Display of temp. sensor temperature


temperature: (Ex. -80.2°C --> Displayed as '80.2')
F13: Display of cascade sensor temperature
(Ex. +6rC --> Displayed as '067')
F14: Display of filter sensor temperature
(Ex. +6rC --> Displayed as '067')
F15: Display of AT sensor temperature
(Ex. +6rC --> Displayed as '067')

Display of battery F03: Display of battery accumulation time


accumulation time: (Ex. 2years and 6months --> Displayed as '02.5')
BATTERY lamp (DP55: orange) will turn on when a display becomes
'02.8'.

<Reset of battery accumulation time>


Step to F06 and input '409'. Press SET key to reset accumulation time.
After reset, BATTERY lamp (DP55: orange) turn off.

Display of condensing F32: Display of condensing fan motor accumulation time


fan motor (Ex. 5years and 8months --> Displayed as '05.8')
accumulation time: BATTERY lamp (DP55: orange) will blink when a display becomes
'05.6'.

<Reset of condensing fan motor accumulation time>


Step to F06 and input '410'. Press SET key to reset accumulation time.
After reset, BATTERY lamp (DP55: orange) stop blinking.

Forcible ON/OFF F18: If you input 'ODD', compressor will turn off and capillary heater will be
operation in capillary forcibly turned on.
heater: If you input 'O~~' during capillary heater turns on, capillary heater will be
forcibly turned off.
If you input '001', capillary heater will be forcibly turned off.
Compressor will be turned off every 18hours, but capillary heater will
be kept turned off.
10, Function mode
FOO Unused: return to chamber temperature display
F01 Setting of high temperature alarm set temperature
F02 Setting of low temperature alarm set temperature
F03 Display of battery accumulation time
F04 Setting of door alarm delay time
F05 Setting of compressor delay time
F06 Setting of service code (Code: 384), Reset of accumulation time
F07 • Temperature sensor zero calibration
F08 • Cascade sensor zero calibration
F09 • Compressor continuous running mode ... Factory use
F10 • Program running mode ... Factory use
F11 • PCB test mode ... Factory use
F12 • Display of temp. sensor temperature
F13 • Display of cascade sensor temperature
F14 • Display of filter sensor temperature
F15 • Display of AT sensor temperature
F16 Unused: return to chamber temperature display
F17 • Setting/Display of model code (Initialization of non-volatile ROM and memory)
F18 • ON/OFF control of capillary heater
F19 • Setting of capillary heater ON time
F20 • Setting of diagnosed value of overload running rate
F21 Setting of communication 10
F22 Setting of communication mode
F23 • Display of power supply voltage
F24 Linkage between remote alarm and buzzer
F25 Setting of Ring Back time
F26 • Display of actual operation rate
F27 • Display of diagnosed value for overload running rate
F28 • Display of delay time for measuring running rate (2 hrs timer)
F29 • Display of delay time for measuring running rate (8 hrs timer)
F30 • Display of ROM version
F31 • Setting of filter alarm operation ON/OFF
F32 Display of condensing fan motor accumUlation time
F33-F43 Unused
F44 • Adjustment of display of power supply voltage
F45-F49 Unused
F50 Setting of alarm delay time
F51-F59 Unused

• Input service code '384' in F06 prior to use function codes which are marked with "
To cancel service code, input "000" in F06 or turn the power off.

Setting In chamber temperature display, press III key for 5 seconds to display "FOO".
procedure: Input Function code by pressing III key and ~ key.
Press SET key to go into each function mode.

FOO: <Purpose> No function to be set in order for user who enter by mistake to return.
<Operation> Press SET key in "FOO" display to return to chamber temperature
display.

F01: <Purpose> Setting of high temperature alarm set temperature


<Operation> Input F01 and press SET key to display "010" (Factory default).
Setting range is '005-040'.
Change a value by pressing III key and ~key.
Press SET key to store the value and to return to chamber temperature
display.
F02: <Purpose> Setting of low temperature alarm set temperature
<Operation> Input F02 and press SET key to display "-10" (Factory default).
Setting range is "-05"-"-40".
Change a value by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.

F03: <Purpose> Display of battery accumulation time


<Operation> Input F03 and press SET key to display F03 and accumulation time
("00.0" if battery used for 36days or less) alternately.
Press SET key to return to chamber temperature display.

F04: <Purpose> Setting of door alarm delay time


<Operation> Input F04 and press SET key to display '002' (Factory default).
Setting range is '001 '-'015'. (Unit: Minute)
Change a value by pressing ~ key and ~ key.
Press SET key to store the vaJue and to return to chamber temperature
display.

F05: <Purpose> Setting of compressor delay time when a power is supplied (a power
returns from power failure).
<Operation> Input F05 and press SET key to display '003' (Factory default).
Setting range is '003'-'015'. (Unit: Minute)
Change a value by pressing ~ key and ~ key.
Press SET key to store the vaJue and to return to chamber temperature
display.

F06: <Purpose> Setting of service code and reset of accumulation time


<Setting of service code>
Input F06 and press SET key to display '000' (Factory default).
Set service code to "384" by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.

<Reset of battery accumulation time>


Input service code '384' in F06.
Input '409' to reset battery accmulation time and to return to chamber
temperature display.

<Reset of condensing fan motor accumulation time>


Input service code '384' in F06.
Input '410' to reset fan motor accumulation time and to return to
chamber temperature display.

<Cancellation>
Input F06 again and press SET key to display '384'.
Change a value to '000' by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.
A display will be also changed to '000' if you turn a power off and on,
but it will not be stored in non-volatile memory.

Note) Service code '384' is stored in non-volatile memory during battery


back-up.

F07: <Purpose> To match a temperature in temp. sensor with 1/2H air temperature
<Operation> Input service code in F06 prior to use this mode,
Input F07 and press SET key to display '00.0' (Factory default).
Setting range is '-4.9'-'04.9'.
Change a value by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.
F08: <Purpose> To calibrate a temperature in cascade sensor
<Operation> Input service code in F06 prior to use this mode,
Input F08 and press SET key to display '00,0' (Factory default),
Setting range is '-9,9'-'09,9',
Change a value by pressing III key and ~ key,
Press SET key to store the value and to return to chamber temperature
display,

F12: <Purpose> To display a temp, sensor temperature


<Operation> Input service code in F06 prior to use this mode
Input F12 and press SET key to display alternately F12 and "XXX'
(current chamber temperature), Press SET key to return to chamber
temperature display, 3 digits indication, Minus "-" is not indicated,
Ex) "-79,5°C" ---> Indicated as "79,5"

F13: <Purpose> To display a cascade sensor temperature


<Operation> Input service code in F06 prior to use this mode
Input F13 and press SET key to display alternately F13 and "XXX'
(current temperature in cascade sensor), Press SET key to return to
chamber temperature display,

F14: <Purpose> To display a filter sensor temperature


<Operation> Input service code in F06 prior to use this mode,
Input F14 and press SET key to display alternately F14 and "XX.X"
(current temperature in filter sensor), Press SET key to return to
chamber temperature display,

F15: <Purpose> To display an AT sensor temperature


<Operation> Input service code in F06 prior to use this mode
Input F15 and press SET key to display alternately F15 and "XXX'
(current temperature in AT sensor), Press SET key to return to chamber
temperature display,

F17: <Purpose> Change of model code and initialization of non-volatile memory


<Change of model code>
Service code should be input in F06 prior to use this mode,
Input F17 and press SET key to display 'OOX',
Change a value (001-009) by pressing III key and ~ key,
Press SET key to store and return to chamber temperature display,

F18: <Purpose> On/off control for capillary heater


<Operation> Service code should be input in F06 prior to use this mode
Input F18 and Press SET key to display '000' (Factory default),
Change to alternative value '000' or '001' by press III key and ~ key,
Press SET key to store the value and return to chamber temperature
display,
000: Capillary heater will be forcibly turned on when it is turned off,
or Capillary heater will be forcibly turned off when it is turned on
001: Capillary heater will be inactive

F21: <Purpose> Setting of serial communication ID


<Operation> Input F21 and press SET key to display '000' (Factory default),
Setting range is '001' - '255" by pressing III key and ~ key,
Press SET key to return to chamber temperature display,
F22: <Purpose> Setting of serial communication mode
<Operation> Input F22 and press SET key to display '000' (Factory default)
Change a value by pressing ~ key and ~ key.
Press SET key to store the value and return to chamber temperature
display.
Control mode (the 3'd digit)
0: Local (initial)
1: Remote
Baud rate (the 2,d digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps

Note) Setting value will not be changed by operating control panel when
control mode is set in 'Remote'.

F23: <Purpose> Display of power supply voltage (Unit: %)


<Operation> Service code should be input in F06 prior to use this mode.
Input F23 and press SET key to display alternately F23 with 'xxx'
(current power supply voltage).
Press SET key to return to chamber temperature display.

F24: <Purpose> Linkage between remote alarm and buzzer


<Operation> Input F24 and Press SET key to display '000' (Factory default).
Change a value by pressing ~ key and ~ key.
Press SET key to store the value and return to chamber temperature
display.
000: Remote alarm does not link with buzzer
001: Remote alarm links with buzzer

F25: <Purpose> Setting of Ring Back time


<Operation> Input F25 and press SET key to display "030" (Factory default).
Setting range is '000'-'060'.
Change a value by pressing III key and ~key.
Press SET key to store the value and to return to chamber temperature
display.
000: Not Ring Back
010: 10 minutes
020: 20 minutes
030: 30 minutes
040: 40 minutes
050: 50 minutes
060: 60 minutes

F26: <Purpose> Display of running rate (Unit: %)


<Operation> Service code should be input in F06 prior to use this mode.
Input F26 and press SET key to display alternately F26 with "XXX"
(Current running rate).
Press SET key to return to chamber temperature display.

F27: <Purpose> Display of diagnosed value for overload running rate


<Operation> Service code should be input in F06 prior to use this mode
Input F27 and press SET key to display alternately F27 with "XXX"
(current diagnosed value for overload running rate).
'000' will be displayed until accumulation time for 8H timer becomes 480
minutes.
Factory default is '095' (fixed value), excepting diagnosed value which
is obtained from calculation in F20.
Press SET key to return to chamber temperature display.
F28: <Purpose> Display of delay time to start measuring running rate
(2hrs timer; 000-120 min)
<Operation> Service code should be input in F06 prior to use this mode
Input F28 and press SET key to display alternately F28 with 'xxx'
(current count value for delay time to start measuring running rate).
Press SET key to return to chamber temperature display.
When a displayed value becomes '120', unit will start measuring
running rate.

F29: <Purpose> Display of delay time to start diagnosing running rate


(8hrs timer; 000-480 min)
<Operation> Service code should be input in F06 prior to use this mode.
Input F29 and press SET key to display alternately F29 with 'xxx'
(present count value for delay time to start diagnosing running rate).
Press SET key to return to chamber temperature display.
8hours timer start counting after 2hours timer expires.
When a delay time becomes '480', unit will start diagnosing running
rate.

F30: <Purpose> ROM version is displayed


<Operation> Service code should be input in F06 prior to yse this mode.
Input F30 and press SET key to display alternately F30 with "XJ<X"
(Ver.1.00 -> "1.00").
Press SET key to return to chamber temperature display.

F31: <Purpose> Setting of audible alarm when filter alarm is occurred


<Operation> Input F31 and press SET key to display "001" (Factory default).
Change to alternative value '000' or '001' by pressing III key and~key.
Press SET key to revert to chamber temperature display.
000: Audible alarm is turned off
001: Audible alarm is turned on

F32: <Purpose> Display of accumulation time of condensing fan motor


<Operation> Input F32 and press SET key to display F32 and accumUlation time
("00.0" if battery used for 36days or less) alternately.
Press SET key to return to chamber temperature display.

F44: <Purpose> Adjustment for difference between display and actual power source
<Operation> Input F44 and press SET key to display '000' (Factory default).
Setting range is '000'-'003'.
Press SET key to return to chamber temperature display.

Setting value Unit status


000 0% (Not changed)
001 3% added to displayed value
002 5% added to displayed value
003 7% added to displayed value

F50: <Purpose> Setting of alarm delay time


<Operation> Input F50 and press SET key to display '015' (Factory default).
Setting range is '000'-'015'.
Change a value by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.
11. L side compressor operating condition
Condition of warm up starting operation
When L side compressor starts in low ambient temperature, unit gives short time voltage
application repeatedly to warm up L side compressor. This is for ensuring startability and operates
required number of times based on ambient temperature condition.
To ensure startability, required number of times is different based on ambient temperature
condition (refer to following list). However, this operation is for areas (used voltage: 100V/115V).
So, this operation is not done for areas (used voltage: 220-240V) and L side compressor start
normally.

Ambient temperature Number of short time Starting


Model code
('C)(F1S) voltage application Operation
Less than S'C 14
S'C-1S'C 9
001/002/004 21 sec ON
1S'C-2S'C 6
/OOS 3 min OFF
More than 2S'C No warm up
Sensor open/short circuit 9
Less than S'C 14
S'C-10'C 9 9 sec ON
010
More than 10'C No warm up 3 min OFF
Sensor open/short circuit 9
003/006/007
No ;yarm up operation
/008/009

Count of L side compressor off period


If L side compressor is stopped for a long time in low ambient temperature condition (by changing
set value, power failure, etc), it incurs low startability. Therefore, it counts its off period and if it
stops more than 120 minutes, operates warm up for L side compressor re-start. Off period counting
starts from L side compressor stop or power failure alarm occurs.

Note) Off period counting operates when the unit power supply is ON or battery switch is ON and
micro computer is operating during power failure. Therefore, counting does not work when the unit
power supply is OFF or power supply returns before power failure alarm is on after micro computer
reset by power failure.

Warm up operation after reset


When chamber temperature is equal or less than set temperature + 10'C, unit regards as instant or
short time power failure and does not operates warm up (regard L side compressor as warm
enough).
However, chamber temperature here is not filtered data for display (TEMPO.PV), use raw data of
PT sensor resistance (TEMPO.PVFIL).
In this case, move to cycle condition immediately and L side compressor start running when it
fulfills one of following conditions after one minute from H side compressor running after delayed
time.
1. Cascade temperature is less thim -34'C
2. Pass 3 minutes (cycle running) or 15 minutes (initial start) after H side compressor running.
12. Timing chart of L side compressor starting operation
Operate L side compressor starting based on following timing.
1. Apply power supply
2. H compo is on after finishing compo delay time.
3. If H side compressor is running more than 3 minutes and cascade sensor temperature become
equal or less than -34°C, H side compressor is OFF. (If H side compressor is running more than
15 minutes, H side compressor is also OFF)
4. L side compressor starts 10 seconds after H side compressor is OFF.
5. L side compressor is OFF after 20 seconds of L side compressor running. After passing 3
minutes, start again.
6. After repeating No.5 operation required times based on ambient temperature, turn H side
compressor ON again, then move to normal operation of L side compressor.

10sec
=;tIC=>
II more than
II 9sec
1min

morethan
3min II
:H ~
I I
: : :<===0>;1 :
ON
Power supply
OFF
H compo operation
ON
OP6: turn on
OFF
L compo operation
ON
DP2
I I OFF

6. L compo
normal operation
1. Apply power supply 2. H compo ON
4. LCOJ.t 6. H compo ON

3. H compo OFF
ON I
5. L compo OFF
4. L compo ON 4. Lcomp. ON
1 min after H compo
is ON and cascade
1 min after H compo 5. L compo OFF sensor temp. is
is ON and cascade less than _34 DC.
sensor temp. is less 5. L compo OFF
than _34 DC. or

or Passed 3 min after


H compo is ON.
Passed 15 min after
H compo is ON.
13. Compressor control (differential) value
Compressor H:
Turns on when a chamber temperature is set temperature -O.4'C.
Compressor H. L:
Turns off when a chamber temperature is set temperature -1.3'C
Compressor L:
1) Turns on when compressor H turns on and a temperature in cascade sensor is -34'C.
2) Turns on after 3 minutes (15 minutes in initial response) since compressor H was turned on
Interval:
It takes 1 minute until one compressor turns on after another one was turned off.

Note) Compressor L does not turn on until 1 minute after compressor H turned on, if a temperature
in cascade sensor is lower than -34'C. Compressor L does not also turn on for 1 minute after
compressor H turn on even if temperature in cascade sensor is lower than -34'C.

Chamber temp. Over I min Chamber temp. Over Imin


=Set temp. -O.4<>C =Set temp. -OAoe
Chamber temp. Comp.delaytime
= Set temp. -I.

Set temp.
Set temp. -OAoC
Settemp. _1.4°C
------------1'~ . . . . .:'f-r_
Cascade Temp.
=-34°C

ON
OFF
ON
OFF

Compressor H protection:
Compressor H turns off to prevent it from being warmed up that is caused by fan motor locked.

Filter sensor tem perature:


Compressor H turns off when a temperature in filter sensor is higher than 56'C. It will turn on again
when the filter sensor temperature is lower than ambient temperature +10'C.

14. Delay time


Compressor delay time (Factory default: 3 minutes)
When a compressor H/L is turned off after applying power or during cycle operation, it has a delay
time to start the compressor again. Delay time can be set in F05.

Temperature alarm delay time (Factory default: 10 minutes)


When high or low temperature alarm is triggered, audible alarm and remote alarm are active after
delay time. (alarm lamp and display are immediately blink) Delay time can be set in F01/F02.

Door alarm delay time (Factory default: 2 minutes)


When outer door is kept opened, audible alarm will be active after delay time. Delay time can be set
in F04.

Power failure alarm delay time (about 3 seconds fixed)


When a power is interrupted, power failure alarm will be occurred after about 3 seconds of delay
time. Delay time cannot be changed.
15. Preventive action for oil logging in capillary
Purpose:
To prevent oil logging in capillary, capillary heater will be active by turning both compressor Hand
L off by regularly period.
Operation:
Both High and Low side compressor will be forcibly turned off and a capillary heater relay (CN4:
3-4) will be active. DP3 (red lamp) will turn on.
Frequency:
8 minutes in every 18 hours (Setting time can be set in F19)
Timing to start operation:
Operation is ordered after 18 hours of unit running (every 18 hours afterwards).
1. 9 seconds after both compressor Hand L are turned off during cycle operation.
2. If the compressor is kept running for more than 60 minutes after operation is ordered,
compressor is forcibly turned off and start operation.
Control of capillarv heater:
Capillary heater will be forcibly turned on or off in F18.

16. Sensor offset


Offset value:
(1) Temperature sensor: +1.3°C (Changeable in FO?)
(2) Cascade sensor: +/-O.O°C (Changeable in F08)
(3) Filter sensor: +/-O.O°C
(4) AT sensor: +/-O.O°C

17. Remote alarm terminal


Operation:
When an alarm is occurred, remote alarm contact (RLY2) switches the position.

CN3
1 - 2 (N.O.) 1 - 3 (N.C.)
Normal Open Close
In alarm Close Open

18. Operation and setting after a power is reset


Settinqs when a power is supplied (Power on resell
Alarms: OFF
Compressors: OFF
Remote alarm: OFF
Ring Back: 30 minutes
Door alarm delay time: 2 minutes
Timers: Reset
2H timer, 8H timer: 0 (Reset)
Warm up starting: ON
Counting of compressor L OFF period: Reset
Setting data: Read by non-volatile memory

Momentarv power failure:


When a chamber temperature is lower than set temperature+1 DoC, unit will determine as
'Momentary power failure' is occurred.

Settings after unit returns from power failure:


Alarms: OFF
Compressors: OFF
Remote alarm: ON
Door alarm delay time: 2 minutes
Timers: Reset
2H timer, 8H timer: 0 (Reset)
Warm up starting: OFF
Counting of compressor L OFF period: Reset
Setting data: Read by non-volatile memory
19. Lamp and display
Lamp operation:
<Display PCB>
DP51: Red lamp: ALARM
In alarm conditions: blink
DP52: Green lamp: STATUS
In STATUS mode: turn on
DP53: Orange lamp: FILTER
In filter alarm: turn on
DP54: Red lamp: DOOR
When door is opening: turn on
DP55: Orange lamp: BATTERY
When battery accumulation time becomes 2.8 years: turn on
When fan motor accumulation time becomes 5.6 years: blink
DP56: Lock (Unused)

<Control PCB>
DP1: Orange lamp
Turn on: no alarm
Turn off: High/low temp. alarm (15min. delay), sensor error,
power failure
DP2: Green lamp
Turn on: L side compressor is ON
Turn off: L side compressor is OFF
DP3: Red lamp
Turn on: Capillary heater is ON
Turn off: Capillary heater is OFF
DP4: Yellow lamp (Unused)
DP6: Green lamp
Turn on: H side compressor is ON
Turn off: H side compressor is OFF

Examples of display:

Chamber temp. -79.5°C Decimal point of chamber temp. -80.0

Set tem p. -80. DoC


CJLJI Sensor offset -5.0

Function F03 :- U ::JI Operation monitoring LCP I: Ipl


Service code 384 -.J I I I_II
-.J CJ I.
Error E01
el::I:1
Set value 004 Accumulation time 8H timer 135 :Hhl
Key Lock
ILl 101
Buzzer tone:
1. Alarms Intermittent tone
2. Key operation Click tone
3. Set value memory Click tone
4. Out of settable range Continuous tone (1 second)
5. Door alarm Intermittent tone (interval is shorter than No.1)
111111111 Parts layout 111111111
<Lower left side>

r.n,ntrnl panel

Step-up transformer

<Lower back side>

/
..J~ Handle
Door latch H/L side compressor

<Lower front side>

Expansion

AT sensor

Filter

Battery switch
Power s,,"tCrL_ _

Temp. control PCB


12P terrnim3l---
W-- - - Power transformer

Switching power supply


Starting ca~lacitor1 U2:!!I""_ _-=::=r-~!!!

Running capatiror H
Starting capaCitor H
'Following data are the reference only, so they do not assure product's performance.

MDF-U76VA Pull-down Temperature AT30°C, 115V60Hz


40

20 ~
\
~

P
~ 0 1\
...
...."::l
...'" -20
\
"E
Co

"
f- -40 1\
-60 ~
~ I'--.
-80

-100
o 2 3! 4 5 6 7 8 9 10 11 12 13 14 15
Time [h]

MDF-U76VA Pull-up Temperature AT30°C


o
-10
V --L.---
~-20
~ /"'
V ----
..." -30
.a
~ -40
/
"Co /
~ -50 ./
f- V
-60 /
V
-70 /
/
-80 /

-90 /
o 1 2 3 4 5 6 7 8 9 10 11 12
Time [h]
MDF-U76VA H/L EVA Temperature AT30°C, 115V60Hz
40
-1/2H Centre

20
f\\ - H EVA OUT
r-
;' '---\ L EVA IN

0 \:'\ - L EVA OUT

\
f? -20 I
'"
~

Q)
\
'-
::l
"til'- -40
It \ \.
I~ \
Q)
c. :
E
Q) -60
~
f-

-80
:
-- --_. '"- . . ",
--- -. ~
--- . -' ..
i~ ---x
-~"-

---"-'-' .. -'",~'-
-'--'., .,.,.. -- - ---.-.-

-100
o 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Time [h]
MDF-U76VA Pressure AT35°C, 115V60Hz
2.8
-HPd
2.4 r---~----~--__-+____~____~____ +-____r-~-HPs
······LPd
-LPs
2.0
~
II)

-g 1.6
~

a.'"
61.2 I---+--+---I---+--===i=-- .... _.... ..... ...- .----1
e
'~" 0.8
II)

a.'-
0.4

0.0
o 2 4 6 8 10 12 14 16 18
Time [h]

MDF-U76VA Current-input AT35°C, 115V60Hz


20 2.0

18 1.8

16 1.6

14 1.4

12 ~ ~
-...... 1.2
'-....-... ~
1.0 ...,
0.8 ..5'c."
-A
6 W 0.6

4 0.4

2 0.2

o 0.0
o 2 4 6 8 10 12 14 16 18
Time [h]
111111111 Temperature uniformity . 17points measuring 1111111111

Back
/'
Front
./
./ i
/~~op ofi~~.=~i~~..
~ ... (
.. ", @
I
1
f--.-c:7..."-....-....-...+i..-.. - - - - -...-....-...-....-..,.-.:1"<: . ..1 70mmH ru"-p-p-e-r-a-re-a-m-ea-s-u-r:-in-g-p-o.,-in-ts------,

(j) ®
(j)-@: 70mm(H) from the top of interior
@ I each
27mm(O) and 42.6mm(W) from

-.:::::: 10mmH corners on Middle shelf.


1
Upper shelf @: 65mm(H) from the top of interior

42.6mmW @: 1Omm(H) from the upper shelf

A-------++--®-9--±-,.i7.:: :::·27mmO
~ CD @
• I [BJ Middle area measuring points
fi&?l.~'~®~!$iM~id?ld~le?ls?iih~e~l1iilf~®!fi!.r4'..::::::.\ Om H ®--®: 10mm(H), 27mm(O), 42.6mm(W)
from each corner on the middle shelf
Mullion
@: 10mm(H) from the middle shelf

® ..::::::: 1o~mH

Lower shelf
Bottom area measuring points
..•......•..•.................... ..!,.......•••••....:". @
."."=-----------+-+--,7. "--=--.;----'- ®-@:70mm(H)fromthebottomofinterior
/// ® • @ / 27mm(O) and 42.6mm(W) from
./ each
....................•...••.••......................................•: ..... .
corner on the middle shelf
@ Bottom of interior ® 70mmH

- -
@ 65mm(H) from the bottom of interior

@: 1Omm(H) from the lower shelf

Left Right
MDF-U76VA Internal Temperature Uniformity (Reference Data)

<Conditions>
Ambient temperature: 23/30°C
Source: 115V 60Hz
Load: Unloaded

<Distribution data>
Temperat ure 0 fth e eye I --80°C • al'r t emperat ure
e 'In e ach area (SV-
AT 23 C AT 30 C
Maximum Minimum I M~~([~ of Differential Maximum Minimum
Middle of
Differential

I~ .I!! Left front 77.0 80.3


c ce
-78.7 +1.7 -76.8 80.3
cvcle
78.6 +1.8
®
I@
.
~
Left back
Right front
-77.8
-77.2
-81.5
-80.6
-79.7
-78.9
+1.9
+1.7
-77.6
-77.0
-81.4
-80.5
-79.5
-78.8
+1.9
+1.8
IT
~;g)
""-"- Right back -77.7 -81.1 -79.4 +1.7 -77.5 -81.2 -79.4 +1.9

~- .I!!
::J Center
Left front
77.5
-78.4
-81.0
-80.0
-79.3
-79.2
+1.8
+0.8
-77.4
-78.6
80.9
-80.2
-79.2
-79.4
+1.8
+0.8
-\}2~ . Left back -80.0 -82.5 -81.3 +1.3 -80.2 -82.6 -81.4 +1.2
~ "
'5
Rightfront -78.8 -80.5 -79.7 +0.9 -79.0 -80.7 -79.9 +0.9

-i- ":!! Right back


Center
-80.0
-80.2
-82.5
-81.1
-81.3
80.7·
+1.3
+0.4
-80.0
-80.3
82.7
-81.4
81.4
-80.9
+1.4
+0.6
JJL
~ E
.I!! Left front
Left back
-79.1
-79.6
-82.0
-82.9
-80.6
-81.3
+1.5
+1.7
-79.9
-80.3
-83.8
-84.3
-81.9
-82.3
+2.0
+2.0
~ 0 Right front -78.9 82.4 -80.7 +1.8 -79.7 -83.3 81.5 +1.8
® ==0 Right back -79.1 -82.9 -81.0 +1.9 -80.0 -84.0 -82.0 +2.0
--w
III Center -79.5 -82.5 -81.0 +1.5 -80.3 -84.3 -82.3 +2.0

=I Center of Upper shelf


Center of Lower shelf
-80.4
79.5
-83.0
80.5
-81.7
80.0
+1.3
+0.5
-80.7
-79.3
-
-83.4
-80.3
-82.1
79.8
+1.4
+0.5
Averaae - I 80.2 - -80.6
. °
UnitC
<Distribution data>
Temperature of the cycle in each area (SV--70°C
- air temperature)
AT 23 C AT 30 C
Maximum Middle of
Minimum Differential Maximum Minimum
Middle of
Differential

~(j)-
®
..
Q)
~
Left front
Left back
-68.2
-69.0
-72.6
-74.4
cvcle
-70.4
-71.7
+2.2
+2.7
-67.7
-68.6
-72.3
-74.4
c cle
-70.0
-71.5
+2.3
+2.9
~ Right front -68.2 -72.8 -70.5 +2.3 -67.9 -72.2 -70.1 +2.2
~

@- ::J""-
"- Right back -68.5 -74.1 -71.3 +2.8 -68.2 73.2 70.7 +2.5
@
JID: I!! . Center
Left front
-68.5
-67.5
-73.5
-70.0
-71.0
-68.8
+2.5
+1.3
-68.2
-67.4
-73.2
-70.0
-70.7
-68.7
+2.5
+1.3
I~ .. Left back -70.1 -73.8 -72.0 +1.9 -70.0 -73.8 -71.9 +1.9
~~ '5" Right front -67.7 -70.1 -68.9 +1.2 -67.7 70.2 69.0 +1.3
Right back -70.0 -73.9 -72.0 +2.0 -70.0 . -73.9 -72.0 +2.0

@) ":!! Center -69.0 -71.4 -70.2 +1.2 -69.1 -71.4 -70.3 +1.2

CJ@- .. I!!
E
0
Left front
Left back
Right front
-64.4
-65.7
-64.2
-66.5
-68.2
-66.2
65.5
-67.0
-65.2
+1.1
+1.3
+1.0
64.3
-65.7
-64.2
66.5
68.2
-66.3
-65.4
-67.0
-65.3
+1.1
+1.3
+1.1
(It
-~~ ==0 Right back -64.3 -67.7 -66.0 +1.7 -64.2 -67.7 -66.0 +1.8
@ III Center -64.9 -67.1 -66.0 +1.1 -65.0 -67.2 -66.1 +1.1

=I Center of Upper shelf


Center of Lower shelf
Averaae
-69.8
-70.6
-
-73.7
-72.5
-
71.8
-71.6
-69.1 I
+2.0
+1.0
-
-69.7
-70.3
- I
73.9
72.1
-
71.8
71.2
-69.0 I
+2.1
+0.9
-
. °C
Unit:
Note:This data does not represent a guarantee of product performance.

<Amount of power consumption>


Amount of power consumpt'Ion when d"
nVlng at eye e
AT 23 C AT 30 C
SV 80 C 16.34 18.07
SV -70 C 13.12 14.07
Unlt:kWh/day
Note:This data does not represent a guarantee of product performance.
AT30°C, SV=-80°C, 115V60Hz
-74

(3-76
L
ru
5-78

~
l'~-80
E
~-82
-;;;
E-84
ru
~
c
---86
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]

AT23°C, SV=-80°C, 115V60Hz


-74

0-76
L
ru
5-78

§
~
ru
~-80
"E
~-82
-;;;
E-84
ru
~
c
-- -86
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]

AT30°C, SV=-70°C, 115V60Hz


-62
(3-64
L
~ -66

§
~

~-68
~

~-70
ru
?
~-72
~
c "'(//
a;-74
~
c
---76
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]

AT23°C, SV=-70°C, 115V60Hz


-62
0- 64
L
e-66

§
~

"E-68
~

~-70
~-72
ru
~
~
c
10-74
~
c
---76
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]
Sample load test

<Condition>
500ml water x 240bottles (Total:120L)
Measuring points:a, b, C, d, H EVA OUT, L EVA OUT
Ambient temperature: 30°C

1st shelf: 500ml x 60bottles

2nd shelf: 500ml x 60bottles

3rd shelf: 500ml x 60bottles

4th shelf: 500ml x 60bottles

Pull-down data(SV -80°C)


40

~ ::-....
20
~ ,,~
o
~ l'-
IILoad bl
\
-...... ~oad~1
ILEVA OUTI I'--- -"'\ ~ ILoad1'-
J3'
~-20 \
dl

e \ .'!\
~ -40
Q) HH EVAOU1,
"-
i~,·
"-
,,'" -
~~
0. -~-
E "'-...,
Q) ILoad al
I-
-60
.~

-80

-100
o 6 12 18 24 30 36 42 48 54 60 66 72 78 84
Time [h]

Note:This data does not represent a guarantee of product performance.


Temperatures after unit reaches to -80°C

- H Eva OUT - L Eva OUT ------a --b -c --d


10
o V"---"1
-10 / Ir--
-20 / /
E-30 / \ / /
~
.i3 -40
V 'v ~

V
~
~-50
E
"
f-- -60
.-
-70 / / /
-80 '.
V
,----
j '- ---- --. _. ,-"-, . . l . -- .-"~-

V
• _•• -"# • -

-90
o 10 20 30 40 50 60 70 80
Time [min]

Temperatures after unit reaches to -70°C


--H Eva OUT - L Eva OUT ----a -b -c -d
10
o
-10 / II If

I /
-20
j J
E-30
~
\
V ---- V V
r
-- J
.i3 -40
~
~-50
E
"
f-- -60
/
/ ---- /
/'
--- 1/
./ f-1

-70
-80
-90
o 10 20 30 40 50 60 70 80 90
Time [h]
SV -80°C Pull-up data
40

20
rH EVAOUTI
o v .. ,"

---I---
~

-_.--'-'

l..-- ~ !.---- !.--


E_20 Load cl
---
--- bl
ILEVA OUT ,...--~.-.

e
k k~ ~
j....--- ~
I.---- rLoad ILoad dl
~ -40 :/'
Q)
.' -' ..,..~
/

~~
c. 'iP
E IrLoad ,;j /
~
-60
... / / V
~
t:/'

-80 -.~

-100
o 6 12 18 24 30 36 42 48 54 60 66 72 78 84
Time [h]

Note:This data does not represent a guarantee of product performance.