User Manual For Model MDF-U76VA-PA - 86 C Upright Freezer VIP Series 1440161441 PDF
User Manual For Model MDF-U76VA-PA - 86 C Upright Freezer VIP Series 1440161441 PDF
User Manual For Model MDF-U76VA-PA - 86 C Upright Freezer VIP Series 1440161441 PDF
SM9910280
Effective models
.Structural specifications
Item MOF-U76VA
Name Ultra-low Temperature Freezer
External dimensions W1010 x 0870 x H1990 (mm)
Internal dimensions W870 x 0600 x H1400 (mm)
Effective capacity 728 L
Outer door Painted steel
Inner door 2doors, ABS resin panel with stainless frame
Insulation Vacuum insulation panel + rigid polyurethane foamed-in place
Exterior Painted steel
Interior Painted steel
3shelves(adjustable), stainless steel
Shelf
Inner dimensions; W848 x 0533 (mm) Load; 50kg/shelf
Outer door latch 1pc
Outer door lock 1pc
Caster 4pcs (2pcs as leveling fooO
Access port 2 locations, Inner diameter; q>17mm
Back upper side: for back up nozzle
Bottom left back: for recorder sensor
Refrigeration circuit Cascade refrigerating system
Compressor High stage side: Hermetic type, Output: 750W
Low stage side: Hermetic type, Output: 11 DOW
Evaporator High stage side: Cascade condenser
Low stage side: Tube on sheet type
Condenser High stage side: Fin and tube type
Low stage side: Shell and tube type
Refrigerant High stage side: R-404A1n-pentane
Low stage side: R-508B/n-pentane
Refrigerant oil Ze-NIUSL22SA
Power supply 115V 60Hz
Transformer 3kVA booster
Battery Nickel-metal-hydride battery, DC 6 V, 1100 mAh, Auto-recharge (5HR-AAC)
Weight 370 Kg
Optional component Temperature recorder(MTR-G85)
Back-up kit: CVK-UB2/UB2(1)/UBN2
Communication kit: MTR-480, MTR-L03, MTR-5000
Storage lack: IR-224U, IR-220U
Inner door (small) : MOF-7101
• Air intake port is at left side of unit (for release inside pressure and validation).
.Control specifications
Item MDF-U76VA
Temp. controller Micro-processor control system
Temperature setting range: -50°C--90°C (Unit :1°C)
Non-volatile memory
Thermal sensor pt.1000n
Temperature display Blue LED digital display (Unit :1°C)
When a chamber temperature becomes set temperature +5°C-+40°C (Factory
default: +1Q°C), ALARM lamp blinks, audible alarm sounds intermittently after
15minutes past.
High temperature alarm
Remote alarm contact: Normal Open, Normal Close
Contact turns over after 15minutes past.
Allowable contact capacity: Max. 30VDC, 2A
When a chamber temperature becomes set temperature -5°C--40°C (Factory
default: -1Q°C), ALARM lamp blinks, audible alarm sounds intermittently after
15minutes past.
Low temperature alarm
Remote alarm contact; Normal Open, Normal Close
Contact turns over after 15minutes past
Allowable contact capacity: Max. 30VDC, 2A
Door alarm DOOR lamp illuminates when a door is kept opening for 2minutes.
Filter alarm FILTER lamp illuminates and audible alarm sounds intermittently.
ALARM lamp blinks, audible alarm sounds intermittently and remote alarm
Power failure alarm
contact outputs.
Remote alarm terminal 3P : Max. DC30V, 2A N.C.-COM, N.O.-COM
Remote alarm When a temperature alarm or power failure alarm occurs, or when a sensor is
failed, remote alarm contact turns over.
Notice of battery life When battery life expires (approx. 3years), BATTERY lamp illuminates.
Notice of fan motor life When fan motor life expires (approx. 6years), BATTERY lamp blinks.
Status-1: If an AT sensor temperature is lower than O°C or higher than +35°C, a
unit diagnoses that the ambient temperature should be abnormal.
Status-2: If a power supply voltage is poor (15% lower than rated voltage), a unit
STATUS function
diagnoses that the power supply voltage should be abnormal.
Status-3: If a running rate in low stage side compressor is more than 95%, a unit
diagnoses that the unit is operating overloaded.
Lamps: ALARM, BATTERY, STATUS, DOOR, FILTER
Lamps and keys on Buzzer stop key: BUZZER Alarm test key: ALARM TEST
Control panel Status key: STATUS Set key: SET
Digit shift key: ~ Numerical value shift key::it:
Press digit shift key for 5 seconds to display Key Lock mode.
Key Lock
LO: Key Lock is OFF L1: Key Lock is ON
When a cascade sensor temperature is lower than -34°C, low stage side
compressor turns on. When a cascade sensor temperature is higher than -12°C,
Compressor protection low stage side compressor turns off. When a filter sensor temperature is higher
than +56°C, high stage side compressor turns off.
Overload relay
If there are several units in a same site and a power failure is occurred, their start
can be controlled by "Start delay time" to prevent them being active
Start delay time
simultaneously.
Setting range: 3-15 minutes (Unit: 1 minute)
.Performance specifications
Model MDF-U76VA
Maximum cooling performance -86'C at the center of the chamber (AT30'C, no load)
Temperature control range -50'C~-86'C (AT30'C, no load)
Power source 115V, 60Hz
Rated power consumption 1020W
Noise level 49 dB [A] (background noise; 20dB)
Maximum pressure 2600 kPa
Usable conditions AT; +5'C-+30'C Humidity: Less than 80%RH
• Design or specifications will be subject to change without notice.
The recommended wire gage to the outlet from distribution box (breaker box from electrical
supply to receptacle unit is plugged into) is dependent on 'length of wire', the following
information is a good rule of thumb to follow:
(6480m)
'E:::r
- 33,4' (TraY)
-
e
-0
~===t====
I'
II
II
,I
II
,
'I
II
II
II
II
.,
EO
';:
m
~
t:
'"InOJ
"'-
:~
II
~~~~~~
J ----- ------:-3"
-
Top
1\
~ +
t f
1 J
i _________ ~ E,
E'".
o
E,
e-
o .
f-----------~ "'-
In"'~
e,
0",
"'In
t : e'"
o·
"""
"'.-
1!11 :
1
~lL
="Ii
.~" __________
•
JI
I
I
E,
eO)
o .
"''''
3-
L
Front Side
Po ..... er supply cord exit posl1lor.
{Main body back}
Cooling unit parts
MDF-U76VA
Specifications
Item
H side L side
Compressor
Type KS240J1NS-7A KS370J1 NS-7A
Code 7FB-0-M101-011-06 7FB-0-M101-001-06
Rating 220V, 60Hz
Refrigerant oil Ze-NIUS L22SA, Charged q'ty 850cc
Cooling system Forced air cooling (partially) and oil cooler
Starting relay AMVL-300TA AMVL-300A
Overload relay MRA999549201
Starting capacitor 1601JF/250VAC x 2 1601JF/250VAC x 2
Running capacitor 151JF/400VAC 251JF/400VAC
Condenser Cascade condenser
Type Fin and tube Coil pipe </l6.35
Condenser 12 columns x 4 lines, P6.35mm
-----
Fin 88pcs
Pre-condenser W350mm -----
Frame pipe </l6.35 -----
Evaporator Cascade condenser Tube on sheet </l9.52
Type Shell and tube </l80 -----
Accumulator </l38 -----
Capillary Ex. capillary
Resistance
78PSI 0.37Mpa 34PSI
PSI· kg/cm 2
Length 3000mm 3000mm 500mm
Outer diameter </l2.4mm <!l2.0mm <!l2.4mm
Inner diameter </l1.2mm </l0.9mm <!l1.2mm
R-404A, Charged q'ty 540g R-508S, Charged q'ty 370g
Refrigerant
n-Pentane (4wt%) 22.5g (36cc) n-Pentane (12.4%wt) 52.4g (83cc)
Dryer 3AXH-9, Char!]ed q'ty 18!] 4AXH-6, Char!]ed q'ty 58!]
Condensing fan Material: ASS, 4 blades, </l230mm -----
Condensing fan SV4-11AS5P
-----
motor runnin!] capacitor: 1.01JF
Thermostat, etc Thermistor, 502AT PT1000n
Heater ----- Capillary heater, 12W x 2P
Oil separator ----- Ze-NiUS L22A: 445cc (384.8g)
1111111111111 Refrigeration circuit 111111111111
-;p- -;p-
PRE-CONDENSER
<E-
Ii.I
FRAME PIPE
I PRE-CONDENSER
Ii.I YJ Ii. ~/\/\/\.r-----'
I
SUCTION CONDENSER
HEAT
«- EXCHANGER "*"" <E- I
"*"" YJ
~ '"w f CASCADE
0-
"''" CONDENSER
t --7>
~
w
d >:
=>
'-'
ffi
W :i
OIL ~ ~15~Wc~~2~sSOR
"'
'" (D SEPARATOR
~
C>
t ~ --7>
CAPILLARY
TUBE (L EX)
~ EXPANS ION TANK
--7>
CAPILLARY TUBEIL)
HEAT EXCHANGER
nnnPlntc::: on PCB
CN8
CNll #1-#2: Battery
#1-#3: Temp. sensor
CNl
CN7 #1-#3:Switching
#1-#2: Door switch
#5-#6: AT sensor
o power supply
CN9
#1-#2: H. Camp relay
CN6
Display PCB
CN4
#1-#2:Temp. control
relay
#3-#4: Heater relay
CN5 CN3
#1-#5: Switch PCB Remote alarm terminal
#l:COM.
~~~::2=rJ~~!::~::~o1-
~ 1F 0
#2: N.D.
#3:N.C.
o
Connections on PCB
CN1 #1, #3: Switching power supply To supply the power to PCB.
#1-#2: Battery(#1 :6V #2:Battery switch) To supply the power during power failure
CNS
#3-#4:Transformer
CN10 Unused
MDF-U76VA AC115V,60Hz
The following shows the temperature in thermal sensor (502AT-1) and its resistance value.
Temperature Resistance Temperature Resistance Temperature Resistance Temperature Resistance
ee) (kQ) ee) (kQ) ee) (kQ) ee) (kQ)
-50 154.5 -36 71.80 -22 35.65 0 13.29
-49 145.9 -35 68.15 -21 33.99 5 10.80
-48 137.8 -34 64.71 -20 32.43 10 8.84
-47 130.2 -33 61.48 -19 30.92 15 7.20
-46 123.1 -32 58.43 -18 29.50 20 6.01
-45 116.5 -31 55.55 -17 28.14 25 5.00
-44 110.2 -30 52.84 -16 26.87 30 4.17
-43 104.4 -29 50.23 -15 25.65 35 3.50
-42 98.87 -28 47.77 -14 24.51 40 2.96
-41 93.70 -27 45.45 -13 23.42 45 2.51
-40 88.85 -26 43.26 -12 22.39 50 2.13
-39 84.18 -25 41.19 -11 21.41 55 1.82
-38 79.80 -24 39.24 -10 20.48 60 1.56
-37 75.67 -23 37.39 -5 16.43 65 1.35
The following shows the temperature in thermal sensor (PT1 0000) and its resistance value.
Temperature Resistance Temperature Resistance Temperature Resistance
(Oe) (kQ) ee) (kQ) ee) (kQ)
-140 450.83 -70 729.99 0 1000.0
-130 491.47 -60 769.02 10 1038.0
-120 531.83 -50 807.87 20 1076.0
-110 571.92 -40 846.58 30 1113.8
-100 611.76 -30 885.13 40 1151.4
-90 651.38 -20 923.55 50 1189.0
-80 690.78 -10 961.84 60 1226.4
'-J
..,., --- BATTERY
rltl
~
ttl [OPTION
BK
3:(J\
I MTR-480
MTR-L03 rtl.
l4J BATTERY
3P
TERMINAL
t:J I
..,.,
SWITCH
~ ~
L. ::::l.
jfi'!J
IT
'"0
IN. ::J
[
-I>- (0
C -I>- c:::J~ R BL BK
--,J C-
(J\ >-'
I eN2 eNG eNs al
<
J>
-I>- (0
"'I
>-'
I \Jl al
-0 I
J> :1 l.b.!!L. 1 13
0
0
[
0
~
I---l
10
I---l
Z
Q BL
-;:Jj
o
I---l
BL I -..:;: I
DOOR SWITCH
L.BL
J>
Q L.BL
CAPITUBE HEATER
10 L,BL
J> STARTING, CAP.L' STARTING CAP.L
:s:
/ il:'
r
;,
AT
SENSOR Gl
(';;" y
'II I~
~ FILTER BK
, ~lJ ~
CONDENSING HARNESS PN r
R F'ANMOTOR CAPITUBE , V VI SENSOR S BK ~
S FM HEATER ~r;!
(,1:;::
~
-
~ (t)
BK
'~;:g
!il" Pl
~
-<
~
PK CAPITUBE ~
Cl r
tD
BR
HEATER
''if ~~
n'"
TERMINAL
~ ~
r :t:!;l :J ~
bf=~ ~
@
@ ©-=-1"'1 ~ g
"oioiH-,,""I'L-- BK ","0 GNIY
r C'l '::t'l
r x~
00
~~ ~ <:::
(11""1
" "
READ WIRE ASSY rAN DTDR 1"1~
>n
~3: Z -< ~E
~ ~
("1 Pl
COMPRESSOR H ,..,
5k;
'" <
5t
~Cl
:r"
"
:to
"
"
:t: ~~RNESS
'"
@~
L,"O
...........
BL FUSE
HOLDER BK c~
R
YL4P V
~
S (VI)
OPTION
REMOTE
ALARM,
YL3
('yI L ~8
... D
TERMINAL BATTERY ~
GN/Y
EARTH
~;:gr-
.,r '"
r
IlK IlK ....
~~
~
PO'w'ER
MTR-480
(- HARNESS EC
HARNESS CH
HARNESS P
PK I I TT I
rJI eN
Y R BL
eN
r-1.r.1
I
III
[
I~I~I~~
PK BK BK BK LBL L.BL LBL L,BL
1 2 3 -4 5 6 7 8 9 10 11 ,12
III I~
~
I~TH T]VVV[Vl
PK BR BK BK BR BK W LBL LBL L.BL LBL
GNlYi
I 1III I
III •
'PK
,.
I IVVVVI
I :0
CN12
I:[~
FR'tNE
EARTH LBL
NESS H
L.BLI vi BK ~
1L
eN'
I L- eN· o
~ ,!8 .........
~-~
.... STARTING' BK BK
m
1:1
'-
BK STARTING
Cltdb:lt:d RELAY L RELAY H
:;u;:orr
~~fl~
:t:
ili.cb.
,,""" I. 2D- ...
~
10,,0...(
"0 '"
~ ~ ~
FANMOTrlR
CAP,
2
1
01
3
;Ilj-
<;;
<
0
:0
CN13
~
::O"l]CC'I
1:C~
fTl .... ::O;:o
t::Jz·;r:. ;r,.
7\~-<
Iv N
'"
Y~
~ ~Ni
e1 k IY
3[[: I:NI
r
;,
I....BL Ii" CNS S:W'ITCHING EARTH
r«-< 1 POWER
................. SUPPLY
r<:(..-(
"...
20
......... :::J:,..,
G'lo ... ' "
n" "
nc
'" " lii=l' 1t
..J:I>
:cr-lr"
-If'TI"lC
",'" <:
"'"
"'"
"O~ ~~ ".z
"OZ
n"
""~ .,," ""z" ~:;! nC 1
rz
~
r~
f-* '"
:0"
:t: z
'"
:-o-a:
!J:~ :t::z BATTERY : HS
IBK IR
'"
. 111111111111111 Circuit diagram 111111111111111
~
...
-
,
~
;~ ~ .
i ui ~ ~
l 0~ !~ ~
"
~
< ~
0...
...
",'-'1.-.1
..::.:, ~,
O.Ol"~ L_1'I1.;-z--] !~~~
0g *E *0: 0:~
~
,:::::~,ya
0,01.1" ~
t;: I'JU ."'~
.m.,.L
SET
[)(>
..'-'" t- ;
: h
Ir
""...Lo--
~-u ~
:l'ff'1 DC +12V
~
. ~. I~ ""~ BUZZER
,---
.~. ,-1.0-
,--1 0- ALAR\,{ TEST
:
I ~~ ~ ,I :
•• m•• """'"
~. I
1'3.8 :, ,,}- ,. .. -- -- -- - --- _______ ,
!S28£~~~~-s : .
P3.5
,
1'3 ••
I::: 1m
,
~i f\ i
0
" . : o
£8Z0!!UC-,V:
L---SOZZe-fn·wif----J
0
.?
~ tl'IO~2UHl •.. ~11
"o,-.!.+.r--h-1r-rrH-+1'-H:'i , 1'2.0
~wP1
1 H 1
~ ~;¥~;~;;~;t If: ~
1111111
1'2.1
P2.2
'"'' ""
'" '"''
"'"
"';;::'
Pl!.7
,
,
1 • JJIJ
to It lli'-~ lli~ I- I" Ii-
0 0
~
2 ~ ~
2 0 W W
~
u
0
oc
<
J
~.
c;
~
~
~
~
0
0
0
..,
~
~
~
J < z
0
u
DISPLA.Y PWB
I I '"
,I
!ila~5
"'0-"'0-
~
0
0
e~
0
0 g
~ ~
IIIIIII Control specifications IIIIIII
1. Key and Switch
BUZZER In alarm condition, audible alarm will silence when this key is pressed, but
remote alarm will be still active and message will not be eliminated.
When a power failure is occurred (during battery back-up), press this key to call
a chamber temperature display for 5 seconds.
ALARM TEST When this key is pressed, unit will step to Alarm Test mode and ALARM lamp
will blink, audible alarm will sound intermittently, digital LED display will go off
and remote alarm will be activate.
A chamber temperature display will appear after approx. 90seconds elapse.
(Auto return)
If Alarm Test is performed when a battery switch is in off position, "E09" will
flash.
SET Press this key to step to setting mode and the 2"' digit in LED display will blink.
Press this key again to memorize a value. Also used to decide and memorize
value in each function and setting mode.
STATUS Press this key during STATUS lamp is on to display occurring status code
("--1", "--2", "--3"). Unused when STATUS lamp is off.
~ In setting mode, press this key to move curser among 1" digit - 3'" digit.
(Digit shift key) If this key is pressed for 5 seconds when a chamber temperature is displayed,
"L_O" will be displayed and go into key lock setting mode.
It] In setting mode, press this key to add numerical value in a displayed digit.
(Numerical value If this key is pressed for 5 seconds when a chamber temperature is displayed,
shift key) "FOO" will be displayed and go into function mode. (use this key and digit shift
key to input function code and press SET key to go into each function mode)
2. Temperature control
Setting range -50°C--90°C
Display range -180-50
How to set Press SET key to step to setting mode and change a value by using It] key
temperature and ~key.
Press SET key to store the value in non-volatile memory.
Unacceptable setting If a value which deviates from setting range is input and SET key is pressed,
value error sound come out and keep setting mode.
4. Function mode
Setting range 00-50
Display range 00-59
00, 16 and 33-43, 44-49, 51-59 are unused.
How to set Function In chamber temperature display, press It] key for 5 seconds to step to
mode: function mode and "FOO" is displayed.
Change a value by using It] key and ~ key.
Press SET key to go into each function mode. If "00" or "16" is input and
press SET key, return to chamber temperature display.
Unacceptable setting f a value which deviates from setting range is input and SET key is pressed,
value: error sound come out and keep setting mode.
5. Error codes
E01: Temp. sensor is open circuited
E02: Temp. sensor is short circuited
E03: Cascade sensor is open circuited
E04: Cascade sensor is short circuited
E05: Filter sensor is open circuited
E06: Filter sensor is short circuited
EO?: AT sensor is open circuited
E08: AT sensor is short circuited
E09: Battery switch is in off position or battery is unconnected
E10: Compressor temperature is abnormal
(4) AT sensor
Open circuit (EO?): If an AT sensor temperature is lower than -60°C, EO? and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Regardless of ambient temperature, operate warm up starting of L side
compressor when AT sensor error is happened.
Press BUZZER key to silence audible alarm.
Short circuit(E08): If an AT sensor temperature is higher than 60°C, E08 and chamber
temperature will be displayed alternately, audible alarm will sound
intermittently and remote alarm contact will be active.
Regardless of ambient temperature, operate warm up starting of L side
compressor when AT sensor error is happened.
Press BUZZER key to silence audible alarm.
(5) Battery SW is off If ALARM TEST key is pressed when battery switch is off position or battery
position or battery is unconnected, E09 will be displayed.
is unconnected(E09):
(6) Compressor If a filter sensor temperature is higher than 54°C, it regards as fan motor
abnormal failure or abnormal compressor temperature.
temperature (E1 0): E10 and chamber temperature will be displayed alternately and high stage
side compressor will be forced to turn off.
Press BUZZER key to silence audible alarm.
If a value which is obtained from filter sensor temperature subtracts an
ambient temperature is equal or lower than 10°C, compressor will turn on.
6. Warning function
High temperature If a chamber temperature is equal or higher than set temperature + high
alarm: temp. alarm set temperature +1°C, ALARM lamp and LED display will blink,
audible alarm will sound intermittently after 10 minutes of delay, and remote
alarm will be active.
If a chamber temperature is equal or lower than set temperature, ALARM
lamp will be off, LED display will blink, audible alarm will silence and remote
alarm will turn off.
Press BUZZER key is to silence audible alarm, but remote alarm will not be
inactive.
Low temp. alarm If a chamber temperature is equal or lower than set temperature - low temp.
alarm set temperature _1°C, ALARM lamp and LED display will blink, audible
alarm will sound intermittently after 10 minutes delay, and remote alarm will
be active.
If a chamber temperature is equal or higher than set temperature, ALARM
lamp will be off, LED display will blink, audible alarm will silence and remote
alarm will turn off.
Press BUZZER key to silence audible alarm, but remote alarm will not be
inactive.
Door alarm If an outer door is left open, DOOR lamp (DP54: red) will turn on. Audible
alarm will sound after 1-15 minutes (default: 2 minutes) of delay. Audible
alarm and remote alarm are not synchronized. Audible alarm will keep
silence by pressing BUZZER key (no ring back function).
Power failure alarm If power supply switch is off or power interrupt during battery switch is in ON
position, it regards as power failure condition.
If a power failure condition keeps for 3 seconds, ALARM lamp will blink,
audible alarm will sound intermittently and remote alarm will be active.
If a power returns within 3 seconds, a unit will start operation at default
settings and remote alarm will be off.
Press BUZZER key to silence audible alarm, but remote alarm will not be
inactive. Remote alarm will keep active until chamber temperature is
stabilized after the power interruption.
Press BUZZER key during power failure condition to display chamber
temperature for about 5 seconds.
7. Status function
When the unit become following 3 conditions, STATUS lamp (DP52: orange) turns on. Press STATUS
key during this condition to display each status code.
Status 1: When AT sensor temperature is lower than O'C or higher than 35'C, display "--1 ".
Status 2: Decrease power supply voltage (between TP7 and TP3: lower than CD2.01V), display
"__2",
Status 3: Running rate is over 95%, display "--3".
Display method
When more than one status is occurred at same time, press STATUS key to display status code in
following order.
" ___ " ---+ "--1" ~ "--2" ---+ "--3" -+ back to "--1"
If there is no status notice, previous status code is displayed (interval is about 3 seconds)
" ___ " ---+ "--1" --+ "--2" --+ "--2" --+ back to "--1"
(when status 1 and 2 are occurred, status 2 is displaying for about 6 seconds)
Return to chamber temperature display after about 90 seconds or all status is disappear.
8. Running rate
Running rate = (ON time I (ON time + OFF time)) x 100%
Measurement of running rate:
It regards as 'cycle start' when a compressor turns on after it turned off once
chamber temperature was lower than set temperature.
Running rate should be measured on and after 2 hours then.
ON time (Min.) = Time until P3.1 in IC5 first reaches from LOW to HIGH
OFF time (Min.) = Time until P3.1 in IC5 reaches from HIGH to LOW
Settemp.
Compressor OFF (1st) 8 hours stabilized Diagnosis is done with every 8 hours
2 hours
Reach to set temp. (Cycle run) until chamber temp is higher than
Set temp + 4.0
Calculation for diagnosed running rate
=(-(Set temp.)XO.9)+((ATXO.9-4.5°C))-((Set temp.+85°C) 110))
9. Other functions
Auto Return: If there is no key operations for 90 seconds in setting mode, Key lock mode
and Function mode, return to chamber temperature display.
Note) Auto return does not work in FOg and F10.
Ring Back: In alarming condition, audible alarm will sound again if someone presses
BUZZER key to silence audible alarm. This is to prevent that buzzer sound is
stopped by third person and the operator cannot notice buzzer.
Ring Back time can be set in F25.
Forcible ON/OFF F18: If you input 'ODD', compressor will turn off and capillary heater will be
operation in capillary forcibly turned on.
heater: If you input 'O~~' during capillary heater turns on, capillary heater will be
forcibly turned off.
If you input '001', capillary heater will be forcibly turned off.
Compressor will be turned off every 18hours, but capillary heater will
be kept turned off.
10, Function mode
FOO Unused: return to chamber temperature display
F01 Setting of high temperature alarm set temperature
F02 Setting of low temperature alarm set temperature
F03 Display of battery accumulation time
F04 Setting of door alarm delay time
F05 Setting of compressor delay time
F06 Setting of service code (Code: 384), Reset of accumulation time
F07 • Temperature sensor zero calibration
F08 • Cascade sensor zero calibration
F09 • Compressor continuous running mode ... Factory use
F10 • Program running mode ... Factory use
F11 • PCB test mode ... Factory use
F12 • Display of temp. sensor temperature
F13 • Display of cascade sensor temperature
F14 • Display of filter sensor temperature
F15 • Display of AT sensor temperature
F16 Unused: return to chamber temperature display
F17 • Setting/Display of model code (Initialization of non-volatile ROM and memory)
F18 • ON/OFF control of capillary heater
F19 • Setting of capillary heater ON time
F20 • Setting of diagnosed value of overload running rate
F21 Setting of communication 10
F22 Setting of communication mode
F23 • Display of power supply voltage
F24 Linkage between remote alarm and buzzer
F25 Setting of Ring Back time
F26 • Display of actual operation rate
F27 • Display of diagnosed value for overload running rate
F28 • Display of delay time for measuring running rate (2 hrs timer)
F29 • Display of delay time for measuring running rate (8 hrs timer)
F30 • Display of ROM version
F31 • Setting of filter alarm operation ON/OFF
F32 Display of condensing fan motor accumUlation time
F33-F43 Unused
F44 • Adjustment of display of power supply voltage
F45-F49 Unused
F50 Setting of alarm delay time
F51-F59 Unused
• Input service code '384' in F06 prior to use function codes which are marked with "
To cancel service code, input "000" in F06 or turn the power off.
Setting In chamber temperature display, press III key for 5 seconds to display "FOO".
procedure: Input Function code by pressing III key and ~ key.
Press SET key to go into each function mode.
FOO: <Purpose> No function to be set in order for user who enter by mistake to return.
<Operation> Press SET key in "FOO" display to return to chamber temperature
display.
F05: <Purpose> Setting of compressor delay time when a power is supplied (a power
returns from power failure).
<Operation> Input F05 and press SET key to display '003' (Factory default).
Setting range is '003'-'015'. (Unit: Minute)
Change a value by pressing ~ key and ~ key.
Press SET key to store the vaJue and to return to chamber temperature
display.
<Cancellation>
Input F06 again and press SET key to display '384'.
Change a value to '000' by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.
A display will be also changed to '000' if you turn a power off and on,
but it will not be stored in non-volatile memory.
F07: <Purpose> To match a temperature in temp. sensor with 1/2H air temperature
<Operation> Input service code in F06 prior to use this mode,
Input F07 and press SET key to display '00.0' (Factory default).
Setting range is '-4.9'-'04.9'.
Change a value by pressing III key and ~ key.
Press SET key to store the value and to return to chamber temperature
display.
F08: <Purpose> To calibrate a temperature in cascade sensor
<Operation> Input service code in F06 prior to use this mode,
Input F08 and press SET key to display '00,0' (Factory default),
Setting range is '-9,9'-'09,9',
Change a value by pressing III key and ~ key,
Press SET key to store the value and to return to chamber temperature
display,
Note) Setting value will not be changed by operating control panel when
control mode is set in 'Remote'.
F44: <Purpose> Adjustment for difference between display and actual power source
<Operation> Input F44 and press SET key to display '000' (Factory default).
Setting range is '000'-'003'.
Press SET key to return to chamber temperature display.
Note) Off period counting operates when the unit power supply is ON or battery switch is ON and
micro computer is operating during power failure. Therefore, counting does not work when the unit
power supply is OFF or power supply returns before power failure alarm is on after micro computer
reset by power failure.
10sec
=;tIC=>
II more than
II 9sec
1min
morethan
3min II
:H ~
I I
: : :<===0>;1 :
ON
Power supply
OFF
H compo operation
ON
OP6: turn on
OFF
L compo operation
ON
DP2
I I OFF
6. L compo
normal operation
1. Apply power supply 2. H compo ON
4. LCOJ.t 6. H compo ON
3. H compo OFF
ON I
5. L compo OFF
4. L compo ON 4. Lcomp. ON
1 min after H compo
is ON and cascade
1 min after H compo 5. L compo OFF sensor temp. is
is ON and cascade less than _34 DC.
sensor temp. is less 5. L compo OFF
than _34 DC. or
Note) Compressor L does not turn on until 1 minute after compressor H turned on, if a temperature
in cascade sensor is lower than -34'C. Compressor L does not also turn on for 1 minute after
compressor H turn on even if temperature in cascade sensor is lower than -34'C.
Set temp.
Set temp. -OAoC
Settemp. _1.4°C
------------1'~ . . . . .:'f-r_
Cascade Temp.
=-34°C
ON
OFF
ON
OFF
Compressor H protection:
Compressor H turns off to prevent it from being warmed up that is caused by fan motor locked.
CN3
1 - 2 (N.O.) 1 - 3 (N.C.)
Normal Open Close
In alarm Close Open
<Control PCB>
DP1: Orange lamp
Turn on: no alarm
Turn off: High/low temp. alarm (15min. delay), sensor error,
power failure
DP2: Green lamp
Turn on: L side compressor is ON
Turn off: L side compressor is OFF
DP3: Red lamp
Turn on: Capillary heater is ON
Turn off: Capillary heater is OFF
DP4: Yellow lamp (Unused)
DP6: Green lamp
Turn on: H side compressor is ON
Turn off: H side compressor is OFF
Examples of display:
r.n,ntrnl panel
Step-up transformer
/
..J~ Handle
Door latch H/L side compressor
Expansion
AT sensor
Filter
Battery switch
Power s,,"tCrL_ _
Running capatiror H
Starting capaCitor H
'Following data are the reference only, so they do not assure product's performance.
20 ~
\
~
P
~ 0 1\
...
...."::l
...'" -20
\
"E
Co
"
f- -40 1\
-60 ~
~ I'--.
-80
-100
o 2 3! 4 5 6 7 8 9 10 11 12 13 14 15
Time [h]
-90 /
o 1 2 3 4 5 6 7 8 9 10 11 12
Time [h]
MDF-U76VA H/L EVA Temperature AT30°C, 115V60Hz
40
-1/2H Centre
20
f\\ - H EVA OUT
r-
;' '---\ L EVA IN
\
f? -20 I
'"
~
Q)
\
'-
::l
"til'- -40
It \ \.
I~ \
Q)
c. :
E
Q) -60
~
f-
-80
:
-- --_. '"- . . ",
--- -. ~
--- . -' ..
i~ ---x
-~"-
---"-'-' .. -'",~'-
-'--'., .,.,.. -- - ---.-.-
-100
o 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Time [h]
MDF-U76VA Pressure AT35°C, 115V60Hz
2.8
-HPd
2.4 r---~----~--__-+____~____~____ +-____r-~-HPs
······LPd
-LPs
2.0
~
II)
-g 1.6
~
a.'"
61.2 I---+--+---I---+--===i=-- .... _.... ..... ...- .----1
e
'~" 0.8
II)
a.'-
0.4
0.0
o 2 4 6 8 10 12 14 16 18
Time [h]
18 1.8
16 1.6
14 1.4
12 ~ ~
-...... 1.2
'-....-... ~
1.0 ...,
0.8 ..5'c."
-A
6 W 0.6
4 0.4
2 0.2
o 0.0
o 2 4 6 8 10 12 14 16 18
Time [h]
111111111 Temperature uniformity . 17points measuring 1111111111
Back
/'
Front
./
./ i
/~~op ofi~~.=~i~~..
~ ... (
.. ", @
I
1
f--.-c:7..."-....-....-...+i..-.. - - - - -...-....-...-....-..,.-.:1"<: . ..1 70mmH ru"-p-p-e-r-a-re-a-m-ea-s-u-r:-in-g-p-o.,-in-ts------,
(j) ®
(j)-@: 70mm(H) from the top of interior
@ I each
27mm(O) and 42.6mm(W) from
A-------++--®-9--±-,.i7.:: :::·27mmO
~ CD @
• I [BJ Middle area measuring points
fi&?l.~'~®~!$iM~id?ld~le?ls?iih~e~l1iilf~®!fi!.r4'..::::::.\ Om H ®--®: 10mm(H), 27mm(O), 42.6mm(W)
from each corner on the middle shelf
Mullion
@: 10mm(H) from the middle shelf
® ..::::::: 1o~mH
Lower shelf
Bottom area measuring points
..•......•..•.................... ..!,.......•••••....:". @
."."=-----------+-+--,7. "--=--.;----'- ®-@:70mm(H)fromthebottomofinterior
/// ® • @ / 27mm(O) and 42.6mm(W) from
./ each
....................•...••.••......................................•: ..... .
corner on the middle shelf
@ Bottom of interior ® 70mmH
- -
@ 65mm(H) from the bottom of interior
Left Right
MDF-U76VA Internal Temperature Uniformity (Reference Data)
<Conditions>
Ambient temperature: 23/30°C
Source: 115V 60Hz
Load: Unloaded
<Distribution data>
Temperat ure 0 fth e eye I --80°C • al'r t emperat ure
e 'In e ach area (SV-
AT 23 C AT 30 C
Maximum Minimum I M~~([~ of Differential Maximum Minimum
Middle of
Differential
~- .I!!
::J Center
Left front
77.5
-78.4
-81.0
-80.0
-79.3
-79.2
+1.8
+0.8
-77.4
-78.6
80.9
-80.2
-79.2
-79.4
+1.8
+0.8
-\}2~ . Left back -80.0 -82.5 -81.3 +1.3 -80.2 -82.6 -81.4 +1.2
~ "
'5
Rightfront -78.8 -80.5 -79.7 +0.9 -79.0 -80.7 -79.9 +0.9
~(j)-
®
..
Q)
~
Left front
Left back
-68.2
-69.0
-72.6
-74.4
cvcle
-70.4
-71.7
+2.2
+2.7
-67.7
-68.6
-72.3
-74.4
c cle
-70.0
-71.5
+2.3
+2.9
~ Right front -68.2 -72.8 -70.5 +2.3 -67.9 -72.2 -70.1 +2.2
~
@- ::J""-
"- Right back -68.5 -74.1 -71.3 +2.8 -68.2 73.2 70.7 +2.5
@
JID: I!! . Center
Left front
-68.5
-67.5
-73.5
-70.0
-71.0
-68.8
+2.5
+1.3
-68.2
-67.4
-73.2
-70.0
-70.7
-68.7
+2.5
+1.3
I~ .. Left back -70.1 -73.8 -72.0 +1.9 -70.0 -73.8 -71.9 +1.9
~~ '5" Right front -67.7 -70.1 -68.9 +1.2 -67.7 70.2 69.0 +1.3
Right back -70.0 -73.9 -72.0 +2.0 -70.0 . -73.9 -72.0 +2.0
I®
@) ":!! Center -69.0 -71.4 -70.2 +1.2 -69.1 -71.4 -70.3 +1.2
CJ@- .. I!!
E
0
Left front
Left back
Right front
-64.4
-65.7
-64.2
-66.5
-68.2
-66.2
65.5
-67.0
-65.2
+1.1
+1.3
+1.0
64.3
-65.7
-64.2
66.5
68.2
-66.3
-65.4
-67.0
-65.3
+1.1
+1.3
+1.1
(It
-~~ ==0 Right back -64.3 -67.7 -66.0 +1.7 -64.2 -67.7 -66.0 +1.8
@ III Center -64.9 -67.1 -66.0 +1.1 -65.0 -67.2 -66.1 +1.1
(3-76
L
ru
5-78
~
l'~-80
E
~-82
-;;;
E-84
ru
~
c
---86
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]
0-76
L
ru
5-78
§
~
ru
~-80
"E
~-82
-;;;
E-84
ru
~
c
-- -86
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]
§
~
~-68
~
~-70
ru
?
~-72
~
c "'(//
a;-74
~
c
---76
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]
§
~
"E-68
~
~-70
~-72
ru
~
~
c
10-74
~
c
---76
0 10 20 30 40 50 60 70 80 90 100 110 120
Time scale[min]
Sample load test
<Condition>
500ml water x 240bottles (Total:120L)
Measuring points:a, b, C, d, H EVA OUT, L EVA OUT
Ambient temperature: 30°C
~ ::-....
20
~ ,,~
o
~ l'-
IILoad bl
\
-...... ~oad~1
ILEVA OUTI I'--- -"'\ ~ ILoad1'-
J3'
~-20 \
dl
e \ .'!\
~ -40
Q) HH EVAOU1,
"-
i~,·
"-
,,'" -
~~
0. -~-
E "'-...,
Q) ILoad al
I-
-60
.~
-80
-100
o 6 12 18 24 30 36 42 48 54 60 66 72 78 84
Time [h]
V
~
~-50
E
"
f-- -60
.-
-70 / / /
-80 '.
V
,----
j '- ---- --. _. ,-"-, . . l . -- .-"~-
V
• _•• -"# • -
-90
o 10 20 30 40 50 60 70 80
Time [min]
I /
-20
j J
E-30
~
\
V ---- V V
r
-- J
.i3 -40
~
~-50
E
"
f-- -60
/
/ ---- /
/'
--- 1/
./ f-1
-70
-80
-90
o 10 20 30 40 50 60 70 80 90
Time [h]
SV -80°C Pull-up data
40
20
rH EVAOUTI
o v .. ,"
---I---
~
-_.--'-'
e
k k~ ~
j....--- ~
I.---- rLoad ILoad dl
~ -40 :/'
Q)
.' -' ..,..~
/
~~
c. 'iP
E IrLoad ,;j /
~
-60
... / / V
~
t:/'
-80 -.~
-100
o 6 12 18 24 30 36 42 48 54 60 66 72 78 84
Time [h]