Multi-Slice

CT Scanner System
Brand
Type
Application
Origin

Neusoft
Neuviz 64i
Applicable for imaging diagnostic incl. Cardiac Analysis
CHINA

Standard Configuration
Components
Gantry

Qty.
1

Low voltage Slip-ring rotating gantry


50 KW High Voltage Generator
Dual focus oil-cooled Tube
High sensitivity GOS Solid-state Detector
Patient Table
Operating Console
19" LCD Monitor
Neuviz 64 software
Cardiac Scan System (software)
ClearView Iterative Recon
Workstation (Hardware)
Workstation (Software)
2D/3D Viewer
MPR Viewer
CPR Viewer
MIP/MinIP
Volume Rendering
Vessel Analysis
Calcium Scoring
Cardiac Function Analysis
Coronary Analysis
Tumor Evaluation
Report
3D Software Package
Accessories
Operation Manual
Main AC Cable
Computer Table
64 Slice Cradle Cushion
NMS Head Holder Assembly
Head Holder Cushion
Mindray PM-9000 Express
Connection Cable
Tool Box
UPS (console)

1
1

1
1

1
1

Technical Specification
Item

Specification

I Component Performance
1 Gantry
Gantry aperture
Scan Field
Gantry tilt
Rotation Time (360rotation)
Partial Scan Time (240)
Focus-to-isocenter Distance
Focus-to-detector Distance

70 cm
50 cm
30
0.5s, 0.6s, 0.8s, 1s, 1.5s, 2s
0.32s, 0.39s, 0.52s, 0.65s, 0.97s, 1.3s
570 mm
1040 mm

Multi-Slice
CT Scanner System
Brand
Type
Application
Origin
2 Data Acquisition System
Max. number of slices/rotation
Number of Detector Rows
Number of detector electronic channels (DAS)
utilized for 64 slices/rotation acquisition
Number of Detector Elements
Total Channels per Slice
Number of Projections
Sequence Acquisition Modes
Spiral Acquisition Modes

Neusoft
Neuviz 64i
Applicable for imaging diagnostic incl. Cardiac Analysis
CHINA

64
32
64
672 x 32
1344
4640
64x0.625, 32x0.625, 16x0.625, 8x0.625, 4x0.625, 2x0.625
64x0.625, 32x0.625, 16x0.625

3 X-ray Tube Assembly


Tube
Tube Current Range
Tube Voltage
Tube Anode Heat Storage Capacity
Cooling Rate
Focal Spot Size

CTR2250
30mA ~ 420mA
80 kV, 100 kV, 120 kV, 140 kV
5.0 MHU
815 KHU/min
0.6 x 1.2 (small) / 1.1 x 1.2 (large)

4 Filter System
Equivalent
Beam Limiting Device

Al Equivalent Tube: 1.5mm Al


Equivalent to 6.68mm Al

5 X-Ray Generator
Max. Power

50 kW

6 Patient table
Max. table load
Table feed speed
Vertical table / Travel range
Vertical travel speed
Scannable range

205 kg/ 452 lbs


1 mm/s - 160 mm/s
430 mm - 970 mm
9 mm/s - 15 mm/s
1750 mm

7 Host computer system


The host computer workplace provides an intelligent and reliable workflow for data acquisition, image reconstruction, and
routine post processing at the CT scanner.
High-performance Computer
Host : 1 x Quad Core Intel (R) Xeon (R)
Recon : 1 x Quad Core Intel Xeon 2.40 GHz processor
Standard Monitor
Flat Screen Monitor 19" (48 cm)
1,280 x 1,024 Resolution
1,024 x 1,024 Image Display Matrix
0.29 mm Pixel Size
RAM Storage
36 GB
Image Storage
250 + 500 GB; 960,000 Uncompressed Images
Additional Storage
CD-R 700 MB
1,100 Images
DVD DICOM Drive 4.7 GB DVD Media
8.400 Images
Write-RW/ +RW/ -DL/ Read
DICOM Viewer
Included on each CD;
Automatically started on the viewer's PC

Multi-Slice
CT Scanner System
Brand
Type
Application
Origin

Neusoft
Neuviz 64i
Applicable for imaging diagnostic incl. Cardiac Analysis
CHINA

8 AVW Workplace system


AVW workplace provides the unique advantage of an efficient multi-modality diagnostic workflow at a single workplace.
It manages the clinical diagnostic workflow anywhere within the clinical environment.
High-performance computer
Dell Precision
Standard monitor
Flat Screen Monitor 19"
RAM storage
> 16 GB
Image storage
> 700 GB; 1,400,000 Uncompressed images
Additional storage
CD-R 700 MB
1,100 Images
DVD DICOM Drive 4.7 GB DVD Media
8,4000 Images
Write-RW/ +RW/ - DL/ Read
DICOM Viewer
Included on each CD;
Automatically started on the viewer's PC

II System Performance
1 System Software
Protocols
Topogram
Length
Scan times
Views
Real-time topogram
Sequence Acquisition
Reconstructed slice widths
Dynamic Multi-Scan:

Multi-slice Spiral Acquisition


Reconstructed Slice Widths
Slice Increment
Spiral Scan Time
Scan Length
Pitch Factor
Automatic clustering of scans.

Up to 10,000 protocols can be edited, modified, and stored

50 - 1700 mm
1.5 - 18 s
A.P., Lateral, Dual
Yes

0.625, 1.25, 2.5, 5, 10 mm


Multiple (continuous) sequence scanning without table movement for
fast dynamic contrast studies with maximum slice thickness of 20 mm

0.625, 0.8, 1, 1.25, 1.5, 2, 2.5, 3, 4, 5, 6, 7, 8, 9, 10 mm


0.1 20 mm
Max. 100 s
Max. 1700 mm
0.13 - 2.0

2 Image reconstruction
Real-Time Display
Scan Field
Recon Field
Recon Time
Recon Matrix
HU Scale

Real-time image display during spiral acquisition.


50 cm
5 50 cm
Up to 20 images/s with full cone beam reconstruction
512x512, 768x768, 1024x1024
3,2768 to + 3,2767

3 CINE Display
Max. Image Rate

30 frames/s

4 Image transfer/Networking
Interface for transfer of medical images and information using the DICOM standard.
5 Raw Data
Capacity

800 GB

Multi-Slice
CT Scanner System
Brand
Type
Application
Origin

Neusoft
Neuviz 64i
Applicable for imaging diagnostic incl. Cardiac Analysis
CHINA

6 Application
SSD (Surface Shaded Display)
Volume Rendering (VR)
Maximum Intensity Projection (MIP)
Minimum Intensity Projection (MinIP)
Average Intensity Projection (AIP)
Multi-Planar Reconstruction (MPR)
Curve Multi-Planar Reconstruction (CPR)
Auto Voice
Auto Film
CT Angiography (CTA),
Virtual Endoscopy (VE),
Pediatric Protocols
Networking
Bolus Tracking
CD Viewer
ClearView
Remote Service System
Volume Calculation
Vessel Measurable Analysis
Bone Remove
7 Heart Scan System
Heart-scan CT with ECG-synchronized true isotropic volume acquisition using prospective ECG triggered or retrospective
ECG-gating mode.
The ECG signal used for gating the CT images is acquired by an ECG device. The ECG signal is displayed on the ECG device
and the scan interface. Temporal resolution can reach down to 65 High/ 83 Low ms.
8 Installation
Outline Dimension & Weight
Gantry Dimensions
Gantry Weight
Gantry Package
Dimensions
Couch Dimensions
Couch Weight
Couch Package Dimensions
Console Table Dimensions
Power Supply Requirements
Power Capacity
Input Voltage

Voltage Variation
3-phase Unbalance
Frequency
Grounding Resistance
Min. Area of Scanning room
Min. Area of Operating Room
Operating Room
Recommended Room Size
Min. Height of Ceiling
Temperature of Scanning Room
Humidity of Scanning room
Atmospheric Pressure
Temperature of Transportation and Storage
Humidity of Transportation and Storage
Running Noise

2244mm (L) x 890mm (W) x 1920mm (H)


1800 Kg
2370 mm (L) x 1030 mm (W) x 2250 mm (H)

2400mm (L) x 610mm (W) x 1055mm (H)


350 Kg
2570mm (L) x 970mm (W) x 1230mm (H)
1400mm (L) x 800mm (W) x 760mm (H)

80 KVA (Low configuration)


380 / 400 VAC
3-phase 4-line
Power supply from below op ons :
190/ 200/ 208/ 220/ 230/ 240/ 380/ 400/ 415/ 440/ 460/ 480 VAC
10%
5%
50/60Hz 1Hz
4 (independent grounding system); 1 (complex grounding system)
5550 mm 3650 mm
1700 mm 3650 mm

Opera ng Room : 3000 mm 4600 mm


Scanning Room : 6000 mm 4600 mm
2010 mm
Scan room 18 ~ 24 ;
Control room 18 28
Scan room 30 % ~ 60 %;
Control room 20 % ~ 80 %
70 kPa ~ 106 kPa
-20 ~ +55
10 % ~ 90 %, no-condensing