0200 2654
0200 2654
0200 2654
Content
Rosemount DIN-Style Temperature Sensor and Thermowells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2 Ordering Tables Rosemount DIN-Style Sensor and Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3 Series 65 and 185 Without Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3 Series 65 and 185 With Tubular Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6 Series 65 and 185 With Barstock Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 10 Rosemount DIN-Style Sensor and Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 14 Series 96 Barstock Thermowell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 14 Integral Mount Sensors and Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 18 Mounting Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 18 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 19 Hazardous Locations Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 20 Sensor-to-Transmitter Matching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 21 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 22 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 27 Thermowell Strength Calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 30
www.rosemount.com
Explore the benefits of a Complete Point Solution from Rosemount Temperature Measurement
An Assemble Sensor to Specific Transmitter option enables Emerson to provide a complete point temperature solution, delivering an installation-ready transmitter and sensor assembly Emerson has a complete portfolio of Single Point and High Density Temperature Measurement solutions, allowing you to effectively measure and control your processes with the reliability you trust from Rosemount products
Experience global consistency and local support from numerous worldwide Rosemount Temperature manufacturing sites
World-class manufacturing provides globally consistent product from every factory and the capacity to fulfill the needs of any project, large or small. Experienced Instrumentation Consultants help select the right product for any temperature application and advise on best installation practices. An extensive global network of Emerson service and support personnel can be on-site when and where they are needed.
Connection Head Standard C D 1 2 N Expanded G H J L Rosemount Stainless steel Rosemount Stainless steel GRA/BL (BUZ) Aluminum w/ Cable Gland TZA/BL (BUZH) Aluminum w/ Cable Gland
M20 x 1.5
1
68 68 65 65
M20 x 1.5
1/2-in. ANPT
M20 x 1.5 (with cable gland) M20 x 1.5 (with cable gland)
Sensor Lead Wire Termination Standard 0 2 3 Flying leads No springs on DIN plate Terminal block DIN 43762 Spring loaded adapter 1/2-in. ANPT Temperature Range Standard RTD, Single Element, 4-wire RTD, Dual Element, 3-wire RTD, Single Element, 4-wire RTD, Dual Element, 3-wire 50 to 450 C (58 to 842 F) 50 to 450 C (58 to 842 F) 196 to 600 C (321 to 1112 F) 196 to 600 C (321 to 1112 F) Standard
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
185 Only 03J1 03K1 05J1 05K1 Thermocouple, Type J, Single Element, Ungrounded Thermocouple, Type K, Single Element, Ungrounded Thermocouple, Type J, Dual Element, Isolated, Ungrounded Thermocouple, Type K, Dual Element, Isolated, Ungrounded 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F) 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F)
Expanded 03N1 05N1 Extension Standard D T F J N W L DIN Standard 12 x 1.5 DIN Standard 12 x 1.5 Nipple Union Nipple Nipple Union (M/F) No Extension Head Connection M24 x 1.5 No Extension Head Connection 1/2-in. ANPT M24 x 1.5 M24 x 1.5
1 1
Thermocouple, Type N, Single Element, Ungrounded Thermocouple, Type N, Dual Element, Isolated, Ungrounded Head Connection
40 to 1000 C (40 to 1832 F) 40 to 1000 C (40 to 1832 F) Instrument Connection Material Standard /2-in ANPT /2-in ANPT /2-in ANPT Stainless Steel Stainless Steel Stainless Steel Stainless Steel
M18 x 1.5
1 1
/2-in ANPT
no head
No Extension (use when ordering the sensor alone, only available with Extension Length (N) code 0000)
Extension Length (N) in Millimeters Standard 0000 0035 0080 0110 0135 0150 Expanded XXXX Non-standard extension length available from 35 to 500 mm No extension use with Extension Type code N 35 mm 80 mm standard for Extension Type code J 110 mm standard for Extension Type codes F and J 135 mm standard for DIN Extension used with Rosemount Connection Head Material codes C, D, G, H, 1, and 2 150 mm standard for DIN Extension used with Form B Connection Head Material codes J and L Standard
Sensor Length (L) in Millimeters Standard 0145 0205 0275 0315 0375 0405 0435 0555 Expanded XXXX Non-standard sensor length available from min. 100 mm, max. 9999 mm 145 mm 205 mm 275 mm 315 mm 375 mm 405 mm 435 mm 555 mm Standard
Product Certifications
Ground Screw
Extension Ring
Termination
Assemble To Option
VS System Calibration
To maintain IP 68 rating, use a suitable cable gland on the conduit connection thread. All threads must be sealed with a suitable sealing tape. For complete assemblies or as replacement sensor for N-series, component parts are not approved. If the transmitter is mounted in a connection head, the Sensor Lead Wire Termination code 0 (flying leads) is required. Not available with Connection Head Material Codes J and L. Not valid with E5, E7, ND, or E1 Approval. If ordering Assemble To Option XA with a transmitter, specify the same option on the transmitter model number.
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Model 0065 0185 Product Description Resistance Thermometer, Pt 100 Ohm RTD Standard with Tubular Thermowell Thermocouples, DIN EN 60584 (IEC 584) Class 1 with Tubular Thermowell IP Rating(1) Conduit / Cable Entry Standard Rosemount Aluminum Rosemount Aluminum Rosemount Aluminum with LCD Meter Cover Rosemount Aluminum with LCD Meter Cover No Connection Head 68 68 68 68 M20 x 1.5
1
Connection Head Standard C D 1 2 N Expanded G H J L Rosemount Stainless Steel Rosemount Stainless Steel GR A/BL (BUZ) Aluminum w/ Cable Gland TZA/BL (BUZH) Aluminum w/ Cable Gland
M20 x 1.5
1
68 68 65 65
M20 x 1.5
1/2-in. ANPT
M20 x 1.5 (with cable gland) M20 x 1.5 (with cable gland)
Sensor Lead Wire Termination Standard 0 2 3 Sensor Type Standard 1 65 Only 2 3 4 185 Only 03J1 03K1 05J1 05K1 RTD, Single Element, 4-wire RTD, Dual Element, 3-wire RTD, Single Element, 4-wire RTD, Dual Element, 3-wire Thermocouple, Type J, Single Element, Ungrounded Thermocouple, Type K, Single Element, Ungrounded Thermocouple, Type J, Dual Element, Isolated, Ungrounded Thermocouple, Type K, Dual Element, Isolated, Ungrounded 50 to 450 C (58 to 842 F) 50 to 450 C (58 to 842 F) 196 to 600 C (321 to 1112 F) 196 to 600 C (321 to 1112 F) 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F) 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F) Flying Leads No Springs on DIN plate Terminal block DIN 43762 Spring Loaded Adapter - 1/2-in. ANPT (available with 185 only) Temperature Range valid for tolerance Class B Pt 100 only Standard Standard
Expanded 03N1 05N1 Extension Standard Y Z Tubular, no Extension form GN Tubular, with Extension form GB, NAMUR, GC, FC Standard Thermocouple, Type N, Single Element, Ungrounded Thermocouple, Type N, Dual Element, Isolated, Ungrounded 40 to 1000 C (40 to 1832 F) 40 to 1000 C (40 to 1832 F)
Extension Length (N) in Millimeters Standard 0000 0050 0065 0105 0115 0130 0200 0250 Expanded XXXX Non-standard extension length available from 35 to 500 mm No extension use with Extension Type code Y 50 mm 65 mm 105 mm 115 mm 130 mm 200 mm 250 mm Standard
Table 2. Series 65 Platinum RTD and 185 Thermocouple With Tubular Thermowell
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Thermowell Material Standard D(2) Y 1.4404 (AISI 316L) 1.4571 (AISI 316Ti) Standard
Immersion Length (U) in Millimeters Standard 0050 0075 0100 0115 0130 0150 0160 0200 0220 0225 0250 0280 0300 0345 0400 Expanded XXXX Non-standard Immersion Length available from 50 to 2500 mm Process Connections Stem Style Standard Thread, Tapered Thread, Tapered Thread, Tapered Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Thread, Tapered Thread, Tapered Thread, Tapered Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Thread, Parallel Flange, RF Flange, RF Flange, RF Flange, RF Flange, RF Flange, RF Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 R 1/2-in. (1/2-in. BSPT) R 3/4-in. (3/4-in.BSPT) R 1-in. (1-in. BSPT) M27 x 2 G 1/2-in. (1/2-in. BSPF) G 3/4-in. (3/4-in.BSPF) G 1-in. (1-in. BSPF) M20 x 1.5 M33 x 2
1
Standard 50 mm 75 mm 100 mm 115 mm 130 mm 150 mm 160 mm 200 mm 220 mm 225 mm 250 mm 280 mm 300 mm 345 mm 400 mm
Mounting Style Standard G02 G04 G06 G13 G20 G22 G24 G91 G31 G38 G40 G42 G52 G92 G63 G94 G72 G95 L02 L08 L14 L20 L26 L32 H02 H08 H14
Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Straight, GN, D. 9 x 1 mm(4) Straight, GN, D. 9 x 1 mm(4) Straight, GN, D. 11 x 2 mm(4) Straight, GN, D. 11 x 2 mm(4) Straight, GB, D. 9 x 1 mm(4) Straight, GB, D. 9 x 1 mm(4) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3) Stepped, NAMUR(3)
/2-in. NPT
3/4-in. NPT
1-in. NPT G 1/2-in. (1/2-in. BSPF) M20 x 1.5 G 1/2-in. (1/2-in. BSPF) M20 x 1.5 G 1/2-in. (1/2-in. BSPF) M20 x 1.5 1-in. 150 lbs 1.5-in. 150 lbs 2-in. 150 lbs 1-in. 300 lbs 1.5-in. 300 lbs 2-in. 300 lbs DN 25 PN 16 DN 25 PN 25/40 DN 40 PN 16
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
H20 H26 Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 DN 40 PN 25/40 DN 50 PN 40 Stepped, NAMUR(3) Stepped, NAMUR(3)
Product Certifications
Ground Screw
Extension Ring
Material Certification
Special Cleaning
Assemble To Options
To maintain IP 68 rating, use a suitable cable gland on the conduit connection thread. All threads must be sealed with a suitable sealing tape. Process thread or process flange to be 316L material with a stem material of 316Ti. Not NAMUR compliant.
NAMUR compliance only applicable with 316Ti material code Y. Minimum immersion length of 115 mm. For u < 115 mm, use straight thermowell, 8 mm Diameter. Not available with Thermowell Material code D. For complete assemblies or as replacement sensor for type N-series, component parts are not approved. If the transmitter is mounted in a connection head, the Sensor Lead Wire Termination code 0 (flying leads) is requested. Not available with Connection Head Material codes J and L. Not valid with E5, E7, ND, or E1 Approval. Not available with welded connection. If ordering Assemble To Option XA with a transmitter, specify the same option on the transmitter model number.
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Model 0065 0185 Product Description Resistance Thermometer, Pt 100 Ohm Standard with Barstock Thermowell Thermocouples, DIN EN 60584 (IEC 584) Class 1 with Barstock Thermowell IP Rating(1) Conduit / Cable Entry Standard Rosemount Aluminum Rosemount Aluminum Rosemount Aluminum with LCD Meter Cover Rosemount Aluminum with LCD Meter Cover No Connection head 68 68 68 68 M20 x 1.5
1
Connection Head Material Standard C D 1 2 N Expanded G H J L Rosemount Stainless Steel Rosemount Stainless steel GR A/BL (BUZ) Aluminum w/ Cable Gland TZA/BL (BUZH) Aluminum w/ Cable Gland
M20 x 1.5
1
68 68 65 65
M20 x 1.5 (with cable gland) M20 x 1.5 (with cable gland)
Sensor Lead Wire Termination Standard 0 2 3 Flying Leads No Springs on DIN plate Terminal Block DIN 43762 Spring Loaded Adapter 1/2-in. ANPT use with Extension Type codes J and N Temperature Range valid for tolerance Class B Pt 100 only Standard RTD, Single Element, 4-wire RTD, Dual Element, 3-wire RTD, Single Element, 4-wire RTD, Dual Element, 3-wire Thermocouple, Type J, Single Element, Ungrounded Thermocouple, Type K, Single Element, Ungrounded Thermocouple, Type J, Dual Element, Isolated, Ungrounded Thermocouple, Type K, Dual Element, Isolated, Ungrounded 50 to 450 C (58 to 842 F) 50 to 450 C (58 to 842 F) 196 to 600 C (321 to 1112 F) 196 to 600 C (321 to 1112 F) 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F) 40 to 750 C (40 to 1382 F) 40 to 1000 C (40 to 1832 F) Standard
Sensor Type Standard 1 65 Only 185 Only 2 3 4 03J1 03K1 05J1 05K1
Expanded 03N1 05N1 Extension Standard D T(2) F J N DIN Standard 12 x 1.5 DIN Standard 12 x 1.5 Nipple Union Nipple Nipple Union (MF) M24 x 1.5 M24 x 1.5
1/2-in ANPT 1
Thermocouple, Type N, Single Element, Ungrounded Thermocouple, Type N, Dual Element, Isolated, Ungrounded Head Connection Instrument Connection
40 to 1000 C (40 to 1832 F) 40 to 1000 C (40 to 1832 F) Materials Standard /2-in ANPT Stainless Steel Stainless Steel Stainless Steel Stainless Steel
M18 x 1.5
1/2-in ANPT 1 /2-in ANPT
(No Head)
Extension Length (N) in Millimeters Standard 0000 0035 0080 0110 0135 0150 Expanded XXXX Non-standard extension length available from 35 to 500 mm No extension use with Extension Type code N 35 mm 80 mm standard for Extension Type code J 110 mm standard for Extension Type codes F and J 135 mm standard for DIN Extension used with Rosemount Connection Head Material codes C, D, G, H, 1, and 2 150 mm standard for DIN Extension used with Form B Connection Head Material codes J and L Standard
10
Table 3. Series 65 Platinum RTD and 185 Thermocouple With Barstock Thermowell
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Thermowell Material Standard D Y Code Standard 0065 0075 0115 0125 0150 0225 0300 0450 Expanded XXXX Code Standard T08 T10 T12 T26 T28 T30 T44 T46 T48 T93 T95 T98 F04 F10 F16 F22 F28 F34 F40 F46 F52 F58(3) F64(3) F70(3) F82 F88 D04 D10 D16 D22 D28 W10 W12 W14 W16 E01 Threaded Threaded Threaded Threaded Threaded Threaded Threaded Threaded Threaded Threaded Threaded Threaded Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flanged, RF Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 Flange, Form B1 according to EN 1092-1 Welded Welded Welded Welded D1 welded R 1/2-in. (1/2-in. BSPT) R 3/4-in. (3/4-in.BSPT) R 1-in. (1-in. BSPT) G 1/2-in. (1/2-in. BSPF) G 3/4-in. (3/4-in.BSPF) G 1-in. (1-in. BSPF)
1/2-in. ANPT 3
Standard 1.4404 (AISI 316L) 1.4571 (AISI 316Ti) Immersion Length (U) Standard 65 mm 75 mm 115 mm 125 mm 150 mm 225 mm 300 mm 450 mm
Non-standard immersion length available from 80 to 1000 mm in 5 mm increments Thermowell Mounting Style Process Connections Stem Style Standard Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered
/4-in. ANPT
1-in. ANPT M27 x 2 M33 x 2 M20 x 1.5 1-in. 150 lbs 1.5-in. 150 lbs 2-in. 150 lbs 1-in. 300 lbs 1.5-in. 300 lbs 2-in. 300 lbs 1-in. 600 lbs 1.5-in. 600 lbs 2-in. 600 lbs 1-in. 900/1500 lbs 1.5-in. 900/1500 lbs 2-in. 900/1500 lbs 1.5 in., 2500 lbs. 2 in. 2500 lbs. DN 25 PN 16 DN 25 PN 25/40 DN 40 PN 16 DN 40 PN 25/40 DN 50 PN 40
3
/4-in. pipe
11
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
E02 E04 E05 D2 welded D4 welded D5 welded 24h7 24h7 24h7 Tapered Tapered Tapered
Ground Screw
Extension Ring
Termination
Material Certification
Special Cleaning
NACE Approval
Weld Options
Wake Frequency
12
Table 3. Series 65 Platinum RTD and 185 Thermocouple With Barstock Thermowell
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
V Options (available with 65 only) Standard V10 V11 Standard XA(8) Assemble Sensor to Specific Temperature Transmitter (PTFE paste) Works Certificate Sensor Calibration from 50 to 450 C (58 to 842 F) with A, B, C, and Callendar-Van Dusen Constants Works Certificate Sensor Calibration from 0 to 100 C (32 to 212 F) with A, B, C, and Callendar-Van Dusen Constants Standard Standard
Assemble To Options
Temperature Range Option V Options (available with 65 only) Standard V10 V11 Standard MD1 MD2 MD3 LT MID Custody Transfer (-196 C to 0 C) MID Custody Transfer (-50 C to 100 C) MID Custody Transfer (50 C to 200 C) Special Material to meet extended Temperature range of -51 C 0065 G 2 2 D 0135 D 0225 F70 Q8 R01 R07 Works Certificate Sensor Calibration from 50 to 450 C (58 to 842 F) with A, B, C, and Callendar-Van Dusen Constants Works Certificate Sensor Calibration from 0 to 100 C (32 to 212 F) with A, B, C, and Callendar-Van Dusen Constants Standard Standard
VS System Calibration
To maintain IP 68 rating, use a suitable cable gland on the conduit connection thread. All threads must be sealed with a suitable sealing tape. Only available with Thermowell Mounting Style codes E01, E02, E04, and E05. Standard T-length is 80 mm, full penetration option R07 must be ordered. For complete assemblies or as replacement sensor for type N-series, component parts are not approved. If the transmitter is mounted in a connection head, the Sensor Lead Wire Termination code 0 (flying leads) is requested. Not available with Connection Head Material codes J and L. Not valid with E5, E7, ND, or E1 Approval Valid for thermowell material code D AISI 316L (1.4404) only. If ordering Assemble To Option XA with a transmitter, specify the same option on the transmitter model number.
13
Thermowell Material(1) Standard D Y 1.4404 (AISI 316L) 1.4571 (AISI 316Ti) Standard
Immersion Length (L) in Millimeters Standard 0065 0075 0115 0125 0150 0225 0300 0450 Expanded XXXX Non-standard immersion length Process Connections Stem Style Standard Thread Thread Thread Thread Thread Thread Thread Thread Thread Thread Thread Thread Flange, RF Flange, RF Flange, RF R 1/2-in. (1/2-in. BSPT) R 3/4-in. (3/4-in.BSPT) R 1-in. (1-in. BSPT) G 1/2-in. (1/2-in. BSPF) G 3/4-in. (3/4-in.BSPF) G 1-in. (1-in. BSPF)
1/2-in. ANPT 3
Standard 65 mm standard length for weld-in thermowells, E01 and E04 75 mm 115 mm 125 mm standard length for weld-in thermowells, E02 and E05 150 mm 225 mm 300 mm 450 mm
Thermowell Mounting Style Standard T08 T10 T12 T26 T28 T30 T44 T46 T48 T93 T95 T98 F04 F10 F16
Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered
/4-in. ANPT
1-in. ANPT M27 x 2 M33 x 2 M20 x 1.5 1-in. 150 lbs 1.5-in. 150 lbs 2-in. 150 lbs
14
Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered Tapered
/4-in. pipe
1-in. pipe 11/4-in. pipe 11/2-in. pipe 24h7 24h7 24h7 24h7
Lagging Length Standard T040 T060 T075 T080 T135 Expanded TXXX Non Standard Lagging Length 40 mm valid for Mounting style codes T26, T28, T30, T93, T95, and T98 60 mm 75 mm valid for weld-in thermowells codes E01 and E02 80 mm valid for flanged thermowells codes F58, F64, F70 135 mm valid for weld-in thermowells codes E04 and E05 Standard
Standard
M18 x 1.5 valid for weld-in thermowells codes E01, E02, E04, and E05
15
The Standard offering represents the most common options. The starred options () should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Special Cleaning Standard R04 Standard R05
(3)
Standard Thermowell Special Cleaning Standard Thermowell NACE Approval Standard Stainless Steel Plug and Chain Standard Full Penetration Weld - for flanged thermowells only Standard Ring Joint Flange Face Standard Wake Frequency Thermowell Strength Calculation Standard Assemble Sensor to Specific Temperature Transmitter (PTFE paste) Standard Special Material to meet extended Temperature range of -51 C 0096 D 0300 F04 T060 D Q8 R01
NACE Approval
Plug/Chain Standard R06 Standard R07 Flange Type Standard R16 Standard R21 Standard XA
(4)
Weld Options
Wake Frequency
Assemble To Options
Additional materials are available upon request. Standard T-length is 80 mm, full penetration option R07 must be ordered. Not available with Thermowell Material code Y. If ordering Assemble To Option XA with a transmitter, specify the same option on the transmitter model number.
16
Overview
Rosemount integral mount temperature sensors, accessory hardware, and assemblies constitute a complete line of industrial temperature-sensing instruments. A variety of RTD and thermocouple sensors are available alone, or as complete assemblies including connection heads, thermowells, and extension fittings. Emerson Process Management (Emerson) offers complete temperature measurement assemblies including Rosemount Smart and Programmable Temperature Transmitters. Please ask your Emerson representative for details. Series 65 Platinum RTD Temperature Sensors are highly linear and have a stable resistance versus temperature relationship. These sensors are used primarily in industrial environments where high accuracy, durability, and long-term stability are required. Series 65 sensors are designed to meet the most critical parameters of international standards: DIN EN 60751 incorporating Amendments 1 and 2, DIN 43760, and BS 1904.(1) This standardization provides sensor interchangeability without the need for transmitter circuitry adjustment. Enhanced performance and optimal temperature measurement accuracy is available for Series 65 sensors coupled with a range of smart temperature transmitters through calibration schedules and Callendar van Dusen constants. Series 185 Thermocouple Temperature Sensors conform to IEC 584 and are available in types J, K and N. Series 185 sensors are available single ungrounded, or dual ungrounded, isolated. All sensors are available in a variety of lengths(2) and ranges with flying lead, terminal block, or 1/2-inch ANPT spring-loaded adapter lead wire terminations. In addition to complete assemblies, Emerson offers a selection of separate accessory hardware including connection heads and thermowells.
81 5 C
30 20 10 0
Pro ces sT Pr emp oc es erat sT ure emper ature 250 C P rocess Temperatur e
C
75
100
125
150
175 200
225
Example The rated ambient temperature specification for the transmitter is 85 C. If the maximum ambient temperature is 40 C and the temperature to be measured is 540 C, the maximum allowable housing temperature rise is the rated temperature specification limit minus the existing ambient temperature (85 40), or 45 C. As shown in Figure 1, an N dimension of 90 mm will result in a housing temperature rise of 22 C. An N dimension of 100 mm would therefore be the minimum recommended length, and would provide a safety factor of about 25 C. A longer N dimension, such as 150 mm, would be desirable in order to reduce errors caused by transmitter temperature effect, although in that case the transmitter may require extra support.
(1) (2)
100 at 0 C, = 0.00385 x C/ Sensors over two meters long will be supplied coiled unless otherwise requested.
17
Ambient temperature range addresses the connection head only, and requires suitable cable glands and field wiring provisions to meet the temperature requirements below -40 C.
Figure 2. Series 65 RTD Lead Wire Configuration
Series 65 RTD Flying Leads and Spring-Loaded Adapter-Termination Codes 0 or 3 only Single Element Dual Element
White White
MOUNTING CONFIGURATIONS
Series 65 Platinum RTDs and Series 185 Thermocouples You may order the Series 65 RTDs and the Series 185 Thermocouples with flying leads, a terminal block, or a 1/2-inch ANPT spring-loaded adapter. Ordered with flying leads, the sensors are designed to be used with a head-mount temperature transmitter attached directly to the sensor. The flying lead configuration allows the removal of the sensor and transmitter as one assembly. The BUZH connection head allows terminal block style sensors and transmitters to be mounted together. The transmitters in these assemblies will be mounted in the cover of the BUZH connection head. The sensors with a 1/2-inch ANPT spring-loaded adapter are used with directly mounted 3144P field-mount temperature transmitters or through the use of Rosemount connection heads. This assembly requires a terminal block to be mounted inside the head. Hazardous area approvals are available with all three types of sensors, but they are dependent on the configuration of the entire temperature measurement assembly (see Hazardous Locations Certification on page 20). Temperature Considerations Ambient temperature limits for the connection head are -40 C to +85 C. The LT Option may be extended down to a range of -51 C to +85 C.
Red Red
Series 65 RTD Terminal Block Termination Code 2 Single Element Dual Element
Green
4 3 1
1 3 4
3 2
Red Red
2
1 White
1 3 4 6
3(+) 1 (-)
1 3
3(+) 1(-)
1 6 3 4
6(-) 4(+)
18
Temperature Range
50 to 450 C or 196 to 600 C depending on type
Self Heating
0.15 K/mW when measured per method defined in DIN EN 60751:1996
Sheath Material
Rosemount thermocouples are made of a mineral insulated cable design with a variety of sheath materials available to suit both the temperature and the environment. For temperature up to 800 C in air, AISI 321 is standard. For temperatures from 800 to 1100 C in air, Inconel 600 is standard. For temperatures above 1100 C, precious metal or ceramic protective sheaths are available upon request. For strongly oxidising or reducing atmospheres, please consult your local Emerson representative.
Immersion Error
60 mm minimum usable depth of immersion when tested according to IEC 751
Insulation Resistance
1,000 M minimum insulation resistance when measured at 500 Vdc and at room temperature
Sheath Material
316 SST / 321 SST with mineral-insulated cable construction
Lead Wires
Thermocouple, internal 18 SWG (16 AWG) solid wire (max), 19 SWG (18 AWG) solid wire (min.). External extension leads, Type J and K 0.8 mm minimum stranded wire, PTFE insulated. Color coded per IEC 584. See Figure 3 for wire configuration.
Lead Wire
PTFE insulated, silver-coated copper wire. See Figure 3 for wire configuration
Identification Data
The model and serial numbers are marked on each sensor.
Identification Data
The model and serial numbers are marked on each sensor
Insulation Resistance
1 000 M minimum insulation resistance when measured at 500 Vdc and at room temperature.
Type J K N
Alloys (wire colour) Fe (+ black), CuNi ( white) NiCr (+ green), NiAl ( white) NiCrSi (+ rose), NiSi (- white)
Temp. Range (C) 40 to 375, 375 to 750 40 to 375, 375 to 1000 40 to 375, 375 to 1000
Tolerance Class 1 1 1
19
ATEX/CENELEC Type n Approval ATEX Marking II 3 G Certification Number BAS00ATEX3145 [Ex nL II T5 (Tamb = 40 to 70 C)] The ATEX/CENELEC EExn Approval allows equipment, which under normal conditions does not emit sparks or produce hot surfaces, to be installed in Zone 2 areas. The Ex n integrity is achieved by the design and construction that maintains a minimum of IP 54 protection. Individual items are not approved. The Rosemount type n approval applies to complete assemblies only. This approval applies to every combination of thermowell, connection head, extension, and sensor detailed in this Product Data Sheet, except the 1/2-inch ANPT spring-loaded style. In addition, certain thermowells, not designed by Rosemount Inc., are acceptable for the EEx n assembly provided they conform exactly to Rosemount specifications. ATEX/IBExU Intrinsic Safety Approval ATEX Marking II 2 G Ex ia IIC T6 (Tamb = 51 to 60 C) The Intrinsically Safety Approval is valid for Series 65 RTD Sensors and Series 185 thermocouples. These certified sensors can only be applied in Zone 1. The marking of intrinsically safe circuits is affected by color codes or print, Ex i. A mounted-on connection head is provided with a ground screw for earth connection and a blue painted cable gland. FM Explosion-Proof Explosion-Proof for Class I, Division 1, Groups B, C, D. Dust-Ignition Proof for Class I, III, Division 1, Groups E, F, G. Ambient temperature Limits: -40 to 245 C When installed per Rosemount Drawing 000068-0013 NEMA Enclosure Type 4X. IECEx Flameproof Approval Ex d IIC T6 (Tamb = 40 to 65 C) The Rosemount Series 65 RTD and Series 185 Thermocouple Temperature Sensors are approved for direct mount to the Rosemount 248H, 644, and 3144P Smart Temperature Transmitters. To ensure approval compliance, specify the E7 option of both the sensor and the transmitter at the time of ordering.
I1
E7 Special conditions for safe use (X) For information on the dimensions of the flameproof joints the manufacturer shall be contacted. ND ATEX / CENELEC Dust Ignition-Proof Certification Number: KEMA99ATEX8715 ATEX Marking: II 1 D 1180 T95 C (-40 C Tamb 85 C) IP66 The ATEX/CENELEC Dust Ignition-Proof approval is dependent on the Rosemount Integral Sensor Connection Head assembled with a Rosemount RTD or thermocouple temperature sensor (see Figure 4). The captive flame arrestor insert must be fully engaged into the connection head for compliance with this approval. Special conditions for safe use (X) For information on the dimensions of the flameproof joints the manufacturer shall be contacted.
20
SENSOR-TO-TRANSMITTER MATCHING
Significant measurement accuracy improvements can be attained using a temperature sensor that is matched to a temperature transmitter. This process involves identifying the relationship between resistance and temperature for a specific RTD sensor. This relationship, approximated by the Callendar-van Dusen equation, is described as:
Rt = Ro + Ro[t (0.01t 1)(0.01t) (0.01t 1)(0.01t)3],
where:
Rt = Resistance (ohms) at Temperature t (C) Ro = Sensor-Specific Constant (Resistance at t = 0 C) =Sensor-Specific Constant =Sensor-Specific Constant =Sensor-Specific Constant (0 at t > 0 C)
The exact values for the Callendar-van Dusen constants (Ro, , , are specific to each RTD sensor and are established by testing each individual sensor at various temperatures. Series 65 RTD sensors can be ordered with the Calibration Option codes V10 or V11, where the values of all four sensor-specific constants are supplied with each sensor. To utilize the unique, built-in sensor-matching capability of the Rosemount 644, and 3144P transmitters, the Callendar-van Dusen constants can be programmed into the transmitter at the factory or in the field using a Field Communicator. The transmitter uses the Callendar-van Dusen constants to generate a sensor curve that describes the relationship between resistance and temperature for this particular sensor and transmitter assembly. By using the sensors actual resistance-vs.temperature curve, there is a 3- or 4-fold improvement in temperature measurement accuracy for the total system.
As in the Callendar-van Dusen method, Ro, A, B, C are specific to each RTD and are established by testing each sensor at various temperatures. The actual values for A, B, and C differ in magnitude from the Callendar-van Dusen constants (Ro, , , ), while Ro is the same in both equations. Either methodology yields the same result in any sensor-to-transmitter matching scenario, since one equation is a simple mathematical interpretation of the other.
Table 6. Series 65 Interchangeability
Standard Series 65 IEC-751 Class B 0.80 C (1.44 F) 0.30 C (0.54 F) 0.80 C (1.44 F) 1.80 C (3.24 F) 2.30 C (4.14 F) Series 65 with IEC-751 Class A Option 0.35 C (0.63 F) 0.15 C (0.27 F) 0.35 C (0.63 F) 0.75 C (1.35 F) 0.95 C (1.71 F) Temperature -100 C (-148 F) 0 C (32 F) 100 C (212 F) 300 C (572 F) 400 C (752 F) Temperature -100 C (-148 F) 0 C (32 F) 100 C (212 F) 300 C (572 F) 400 C (752 F)
21
Transmitter: Rosemount 3144 (has built-in sensor matching capabilities), span of 0 to 200 C, accuracy = 0.1 C) Sensor: Series 65 RTD Callendar van Dusen Option: V10 Process Temperature: 150 C
System Uncertainty Comparison at 150 C:
Standard 65 Sensor
1.05 C
Rosemount 3144P: Standard Series 65 RTD: Total System(1): 65 Sensor with V10 Option Rosemount 3144P: Calibrated Sensor Series 65 RTD: Total System(1):
Standard
(1) Calculated using RSS statistical method:
System accuracy = =
CALIBRATION
Sensor calibration may be required for input to quality systems, or for control system enhancement. More frequently, it is used to improve the overall temperature measurement performance by matching the sensor to a temperature transmitter. Sensor matching is available for RTD sensors used with Rosemount Smart transmitters where the inherent stability and repeatability of the RTD technology is well established.
Ordering Information
Use the formats presented below to order a calibrated Series 65 RTD. If you fail to specify all of the necessary calibration-related information when you place your order, Emerson will contact you for the information and your order may be delayed slightly.
(1) 22
Calibration Options
The X8 option calibrates the sensor to a customer-specific temperature range. The Callendar van Dusen, and A, B, and C-constants are supplied with a works certificate. Option X8: Sensor Calibrated to a Customer-Specified Temperature Range (see Temperature Range) When you order an RTD with the X8 option, the temperature range the sensor needs to be calibrated must be specified. Take note of the sensor temperature limits as shown below.
Typical Model Number Model 0065 Connection Head C Lead Wire Termination 2 Sensor Type 1 Extension Type D Extension Length 0135 Thermowell Material D Immersion Length 0225 Mounting Style T12 Additional Options X8
23
644
248
40 mm
25 mm
SENSOR WITH FLYING LEADS, TERMINAL BLOCK, OR SPRING-LOADED ADAPTER 2 x 1/2-in. ANPT
L 11 mm L
16 mm N
EXTENSIONS
N N 11 mm
Terminal Block
Flying Leads
Terminal Block
Flying Leads
Adapter
41
41
41
33
41
25
10
112 L
Series 65 Single Element 65 Dual Element 185 Single Element 185 Dual Element
Number of Leads 4 6 2 4
Lead Wire Length (Flying Leads) Element 1 100 100 100 100 Element 2 200 200
Lead Wire Length (Spring Loaded) Element 1 150 150 150 150 Element 2 200 200
24
40 mm
25 mm
N N
U U U
For straight threading, N dimension references bottom of hex. For tapered threading, N dimension references thread engagement point (bottom of thread)
NAMUR
NAMUR
GB
GN
NAMUR
25
IP 68 OR IP 65 CONNECTION HEADS
3144
40 mm
25 mm
SENSOR WITH FLYING LEADS, TERMINAL BLOCK, OR SPRING-LOADED ADAPTER 2 x 1 /2-in. ANPT
11 mm L L
16 mm
11 mm
STAND-ALONE EXTENSIONS
11 mm
11 mm
40 mm
60 mm
60 mm
60 mm
N dimension measures from thread engagement point. This dimension is 80 mm for 1500# and 2500# flanges.
26
Process Connection
1
/2-inch ANPT /2-inch ANPT /2-inch ANPT /2-inch ANPT /2-inch ANPT
M24 x 1.5
1
M24 x 1.5
1
M24 x 1.5
1
M24 x 1.5
1
M24 x 1.5
1/2-inch ANPT
Option Codes 1, 2
104
78
LCD Display
78
100
GR-A/BL (BUZ)
Cable Entry
110
Cable Entry
84
27
T = Lagging Length
Dimensions in millimeters
Flanged Barstock Thermowell Tapered
Flange Size
D 19 26.5 26.5
d 12.5 18 18
T 60 60 80
7.0
1-in. 150 1500 lbs, DN 25 11/2 to 2-in. 150 600 lbs, DN40 - 50 1.5 to 2-in. 900/1500
5.0
Note: Flanged thermowells generally conform to the specifications of ASME B 16.5 (ANSI) and DIN EN 1092-1.
5.0 7.0 D D1 d
Parallel Thread Size
1/2-in. BSPF (G1/2);
D 17 19 26.5 19
D1 26 32 39 29
M20 x 1.5
3/4-in. BSPF (G3/4)
40
U
Threaded Barstock Thermowell Tapered Thread
60
U
Tapered Thread Size D d 12.5 12.5 18
17 19 26.5
1-in. ANPT
7.0 5.0
28
Dimensions in millimeters
Weld-in Barstock Thermowell (codes W10, W12, W14, W16)
60
Socket Size
/4-in.
5.0 S D
7.0
D1 140 65 75
D2 200 125 75
D4 200 65 135
29
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms_of_sale 2011 Rosemount, Inc. Emerson Process Management Rosemount Inc. 8200 Market Boulevard Chanhassen, MN 55317 USA T (U.S.) 1-800-999-9307 T (International) (952) 906-8888 F (952) 906-8889 www.rosemount.com Emerson Process Management Heath Place Bognor Regis West Sussex PO22 9SH England Tel 44 (1243) 863 121 Fax 44 (1243) 867 554 Emerson Process Management Asia Pacific Private Limited 1 Pandan Crescent Singapore 128461 T (65) 6777 8211 F (65) 6777 0947 [email protected]