PACS Fundamentals
PACS Fundamentals
PACS Fundamentals
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Objectives
Define picture archiving and communication systems (PACS). Compare and contrast the various types of PACS display workstations. Differentiate between the different types of digital imaging workflow. Define system architecture and recognize the three major models. Summarize the common functions found on a PACS workstation. Describe the situations and users that might use advanced PACS workstation functions.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Key Terms
Archive Client/server-based system DICOM Display workstation Distributed system File room workstation Hanging protocol Navigation functions
PACS QC station Reading station Review station Soft copy System architecture Teleradiology Web-based system Workflow
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Introduction
PACS
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Fundamentals
Digital acquisition (Picture) Display workstations Storage devices (Archiving) Components are interconnected through an intricate network. (Communication)
PACS is the electronic version of the radiologists reading room and the file room.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 5
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Fundamentals
First PACS
Early 1980s Served one single modality Large research institutions Most developed by scientists in those institutions Vendors became more involved. Proprietary systems were developed.
Later
Standardization
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
DICOM
Digital imaging and communications in medicine. Universally accepted standard Laid the groundwork for the
future development of integrated PACSs. Now every modality and PACS communicates via DICOM. Each vendor and modality boasts DICOM compatibility.
Each DICOM statement must be read carefully to determine the extent of the compatibility.
http://medical.nema.org/
8
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
Components
Image Acquisition
Magnetic Resonance Imaging (MRI) As with CT as the numbers of images increased, reading
on a monitor became a good alternative.
Display Workstations
Most interactive part of a PACS. Used inside and outside of the radiology department.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 11
Display Workstations
Display station.
Receives images from archive or various radiology modalities Presents images to be viewed
Workstation has some sort of PACS application software. Some may have advanced software with additional image processing capabilities.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
12
Archive Servers
Database server or image manager Short-term and long-term storage Workflow manager
Central part of the PACS Houses all of the historic and current data May also serve as the centralized node that receives all images before interpretation
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 13
Workflow
Workflow
How a process is done step by step or how a task is completed How one completes an exam from order entry to transcribed report
Generic Workflow
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
15
System Architecture
American National Standards Institute and the Institute of Electrical and Electronics Engineers
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
16
System Architecture
In other words,
System architecture is the hardware and software infrastructure of the systems workflow. In PACS the system architecture normally consists of the following: Acquisition devices Storage and display workstations Image management system
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
17
PACS Architecture
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
18
Client/Server-Based Systems
Images are sent to archive server Display workstation functions as a client of the archive server
Accesses images based on a centralized worklist. Person at the client chooses a name from the list. Archive server sends the image data to the client. The image data is only on the client while viewing. Most systems allow basic image manipulation at the client. Changes are saved on the archive server.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
19
Client/Server-Based Systems
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
20
Client/Server-Based Systems
Advantages:
Any exam sent to the PACS is available anywhere without other interventions. Only one person can open the study with the intent to read it. Others that open the study receive a message that the study is open and being read. There is no need to pull or send historic images to a particular workstation because the old studies are available with the new on the archive.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
21
Client/Server-Based Systems
Disadvantages:
The archive server is seen as a single point of failure. If the archive goes down, the entire system is down and
no image movement can take place. Newly acquired images must remain at the modality until the archive is up and can receive the images.
System is network dependent. Images are flying back and forth between the archive and
the workstations. Network can become bogged down.
Archive server is handling many requests at once and can become bottlenecked.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 22
Distributed Systems
Distributed systems are also known as distributed or stand-alone models. Acquisition modalities send the images to a designated reading station and possibly to review stations.
In some systems the images are sent from the modality to the archive server, and the archive server distributes the images to the designated workstation.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 23
Distributed Systems
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
24
Distributed Systems
MRI may send to one station CT sends to another Or all cross-sectional neurologic images may be sent to one station, whereas all body imaging may be sent to another
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
25
Distributed Systems
Workstations can query and retrieve images from the archive. All images are locally stored. Images are then sent to the archive server once the images have been read. Images remain on local hard drive of workstation until they are deleted by user or system rules.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
26
Distributed Systems
Advantages:
If the archive server goes down, local reading at the workstations is not interrupted. After archive comes back up, the images that have been
changed and signed off are automatically forwarded to the archive to be saved.
PACS data is less likely to be lost because of multiple copies in various locations. System is less dependent on network for speed. User can work on one exam while workstation is pulling
next exam to be read. Workstation can fetch historic images based on rules set up by the user. Fetching can be done while other exams are being read.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 27
Distributed Systems
Disadvantages:
Personnel rely heavily on system to perform image distribution correctly. If distribution is wrong, then prefetching of historic exams will
be wrong.
Each workstation has a different worklist; only one person at a time can work on that list. It can be inconvenient to read additional studies. Radiologist would have to move to another workstation to read
the images.
Users must depend on query and retrieve function when nonscheduled exams arrive at workstation. It is possible for two radiologists to be reading the same exam and not know it until they try to start dictation. Paper requisition becomes an important piece of information.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
28
Web-Based Systems
Web-based systems are similar to a client/server systems. Biggest differences are that images and application software are held centrally and loads to the client display.
Web-Based Systems
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
30
Web-Based Systems
Advantages:
Hardware at the client can be anything that will support an appropriate web browser.
This condition allows for greater flexibility with hardware. This can be a disadvantage because low-end PCs can be used and the image displays (monitors) may not be diagnostic quality.
The same application can be used on-site and at home in teleradiology situations.
Teleradiology is the reading of images from outside of the hospital walls.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
31
Web-Based Systems
Disadvantages:
System functionality may be limited because of software not being locally installed. Bandwidth of the network connection limits the amount of data
that can be transmitted. Some programs are large and cannot be transmitted via a network.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
32
Display Workstations
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
33
Display Workstations
Conventional film/screen radiography uses large multiviewer lightboxes. With early PACS, radiologists thought that they needed four to six monitors. Now, as radiologists have become more comfortable, the number of monitors has dropped to an average of two. Drop can be attributed to continued development of viewing software and better hardware.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 34
Display Workstations
Monitor
One of the most important elements Several types of monitors Cathode ray tube (CRT)
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
35
Display Workstations
Monitor, continued
CRT Heavy Puts off a lot of heat Very bright (good thing) Can view from most any angle
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
36
Display Workstations
Monitor, continued
LCD Dropped in price and has risen in quality. LCD will soon take over PACS display market because of its
size, resolution, and lack of heat production. LCD requires less maintenance. LCD gives more light. LCD can be used in areas with a high amount of ambient light.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
37
Display Workstations
Resolution and orientation of the monitor is a factor in determining which type of monitor is to be used.
Most cross-sectional imaging is read on a 1K square monitor. Most computed radiography (CR) and digital radiography (DR) images are read on at least a 2K portrait monitor.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
38
Display Workstations
More pixels: The higher the resolution, the more information that can be displayed. Resolution also is defined as the process or capability of distinguishing between individual parts of an image that are adjacent.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
39
Display Workstations
Pixels are arranged in a matrix. Common screen resolutions found on todays monitors are the following:
1280 1024 (1K) 1600 1200 (2K) 2048 1536 (3K) 2048 2560 (5K)
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
40
Display Workstations
Medical displays are generally higher quality than displays for other applications. Radiologists use highest-resolution monitors available for the modality that is being read:
Mammography requires a 5K or 5-megapixel resolution. Cross-sectional image only requires a 1K monitor. If referring physician is not the primary doctor reading the exams, a 1K monitor would be sufficient.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
41
Display Workstations
Primary reading stations for radiologists Review stations for referring physicians Technologist quality control (QC) station for technologist review of images Image management station for the file room personnel
Each has a specific, main purpose. Workstations are placed in strategic areas near the end-user of that particular need. Workstations may be made up of different hardware depending on demand and need of user and requirements of the software that will be used.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 42
Station is used by a radiologist to make a primary diagnosis. Station will have the highestquality hardware, including best monitor. Computer hardware used depends on the needs of the PACS vendor but usually is robust, with little downtime.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
43
Keyboard and mouse can be customized to needs of department. Many different styles of mice are available.
Dictation system is near or is connected to PACS station. Many systems are integrating the RIS and dictation system within the PACS software. Grouping allows a more seamless workflow with little to no paper. Station streamlines the completion of the study.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
45
Station is a step-down model of the radiologists reading station. Station may have the same level of software but may reduce some advanced functions. One of the most important features is ability to view current and previous reports with images. Many vendors are integrating the RIS functions with PACS software.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
46
Most referring physicians want to read radiologists report along with viewing images. Many times, report is more important to them than the images. Software may be loaded on a stand-alone station that is dedicated to viewing images. Or, software may be delivered over a web browser on any personal computer within an office or floor.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
47
In high-volume areas such as the emergency room and intensive care unit, there are dedicated PACS workstations strictly for image viewing.
These may have the higher-end monitors, but many may have a lower-end monitor because of the costs.
One of the greatest advantages of a PACS is the ability to view the same set of images in multiple locations at one time.
Referring physician can pull up the patients images in the office and read the radiologists report and then call the radiologist on the phone and consult while both parties are viewing the same set of images. Continuity and speed of patient care has shown improvement with the use of PACS.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
48
Technologist QC Stations
Used to review images after acquisition but before sending them to the radiologist
May be used to improve or adjust image-quality characteristics May be used to verify patient demographic information Placed between the CR and DR acquisition modalities as a passthrough to ensure that the images have met the departmental quality standard
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
49
Technologist QC Station
Generally has a 1K monitor. Does not have the resolution capabilities of the radiologists reading station. Care required of technologist when manipulating images not to change the appearance too much from original acquired image. Technologist should consult frequently with the radiologist to ensure quality. Station can also be used to query and retrieve historic images to check previous pathologic conditions or body characteristics. Station can help with the selection of technical factors or procedural protocol. QC station can afford same benefit as pulling the film jacket.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
50
Before PACS, file room was a large open room with endless rows of shelves full of film jackets. Today, it may be as simple as several computers and a dry laser to make copies for outside needs. Workstation may be used to look up exams for a physician or to print copies of images for the patient to take to an outside physician.
Many hospitals are moving away from printing films because of the cost. Hospitals are moving toward burning compact disks (CDs) with the patients images. CDs can be made quickly and at a reduced cost compared with film.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 51
Common Functions
Navigation functions Image manipulation and enhancement functions Image management functions Advanced workstation functions
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
52
Navigation Functions
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
53
Navigation Functions
Use of grab bars on the right hand side of windows to scroll through a list Activation of the scroll wheel on the mouse to scroll through the list
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
54
Navigation Functions
Mouse is a useful navigation tool. Right mouse button offers many shortcut features in a menu of frequently used tasks and applications. Hanging protocols are available:
Each user has the ability to set up a custom hanging protocol. Protocol is defined as how a set of images will be displayed on the monitor.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
55
Navigation Functions
Hanging protocols
Can be viewed four images on each monitor. Can be viewed as one image on each monitor.
CR image is selected.
Protocol can also be specified to show the previous exam on one monitor and the current exam on the other. Once set, the most efficient study navigation is determined.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 56
Study Navigation
Study may consist of two or three single images such as the case with CR and DR. Study may contain several series of images such as the case with MRI. Images can be simply paged through with the scroll wheel or arrows on the keyboard.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
57
Study Navigation
Stack mode of scrolling through images made is called cine. Cine means to move through frame by frame of the series of
images.
Study may have an automatic setting that will run through the images at a preset pace. Cine function is used most often in cross-sectional imaging.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
58
Study Navigation
To move between a patients various studies To open the next unread patient file in the worklist To close patient or study icon Closes the active patient or study Pulls up the worklist or moves to the next unread patient
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
59
Tools
Window/level
By depressing and holding down the mouse button and moving the mouse up and down and left and right, the window and level can be adjusted.
Window represents the range of gray values. Level represents the center value of the range. A change in the window and level appears to change the
brightness and contrast of the image.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 60
Tools, continued
Annotations
Tools, continued
Tools, continued
Pan, zoom, and magnify Tools are used primarily by the radiologist
to increase the size of an area on the image. Magnify usually magnifies a square area of the image.
Square can be moved around the image to quickly see various areas magnified.
Image is first zoomed up to the desired magnification level. Pan icon is activated. Zoomed image can be moved around to view the different areas of the image.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
63
Tools, continued
Tools, continued
Measurement tool determines the pixel intensity of a certain area. Each type of tissue or fluid has a different intensity of reading. Radiologist can make a determination whether something is solid or fluid.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
65
Patient demographics
Patient demographics must be correct. If demographics are not correct at the archive level, the images
could be lost.
Changes should only be made when the information is absolutely known to be wrong. Many hospitals allow only certain persons the access to change demographics just to keep the errors to a minimum.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 66
Query/retrieve icon
Used to retrieve on demand any studies from the archive Allows user to query a study on multiple fields Patients name or identification Date of service Modality Diagnosis code or comment field
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
67
CD burning option
Feature may only be available in the file room to control the CDs that are sent out. HIPAA (Health Insurance Portability and Accountability Act) compliance must also be maintained.
Function is used with the web-based systems when creating presentations for conferences. Patient information must be removed from the image before it is placed into a presentation.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
68
Print films
Printing is usually only done in the file room so that control can be maintained over the printed films for HIPAA purposes and cost reasons. Workstations may be connected to paper printers for quick consults and for medical records.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
69
Advanced functions are usually placed on specialty workstations for the radiologist, and some are found on the technologist QC station to further enhance the images. The following is a bulleted list of some of the most common advanced functions:
Multiplanar reconstruction (MPR) Maximum intensity projection and minimum intensity projection (MIP and MinIP) Volume rendering technique Shaded surface display Stitching
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
70
Maximum intensity projection and minimum intensity projection (MIP and MinIP) Function is used to
visualize vessels (MIP) and air-filled structures (MinIP). Function is commonly performed after injection of contrast medium on CT and MRI studies. Contrast medium shows areas of strictures and blockages within the vessels.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
72
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
73
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
74
CR manufacturers have developed a 3-foot CR cassette that contains multiple image plates (IPs).
Each of the IPs is scanned through the reader, and the individual images are sent to the QC workstation. Software interpolates images and connects them using known markers from the IPs. Technologist has the ability to adjust the connection of the images.
Long leg images are used for leg-length discrepancy. If the special 3-foot cassettes are not available, a radiopaque ruler can be used to ensure that the images are stitched at the right area.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
75
Image Postprocessing
Many other advanced workstation functions are available to be added to the PACS workstation. This is a growing field with advancements coming each year. Specific information about how to perform these procedures can be found in the vendors user manual.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc.
76
Summary
A PACS consists of digital acquisition, display workstations, and storage devices interconnected through an intricate network. Digital imaging and communications in medicine (DICOM) is a universally accepted standard for exchanging medical images between the modality, viewing stations, and the archive. A display workstation is any computer that a health care worker uses to view a digital image, and it is the most interactive part of a PACS. The archive is the central part of the PACS and houses all of the historic data along with the current data being generated. Workflow is how a process is done step by step or how a task is completed.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 77
Summary
System architecture is the basic organization of a system, come to life in its components, their associations to each other and the environment, and the principles leading its design and development, or in other words, system architecture can be defined as the hardware and software infrastructure of the systems workflow. Common system architectures found with a PACS are client/server-based systems, distributed or stand-alone systems, and webbased systems.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 78
Summary
Display stations can be categorized by their means of use, such as primary reading stations for radiologists, review stations for referring physicians, technologist QC stations for technologist review of images, and image management stations for the file room personnel. Many functions are available on a PACS workstation, and each set of functions can be broken down into four categories: navigation functions, image manipulation and enhancement functions, image management functions, and advanced workstation functions.
Elsevier items and derived items 2008 by Mosby, Inc., an affiliate of Elsevier Inc. 79