D5307013B Lifeline Connect and Connectplus Install and Prog Guide v9.7
D5307013B Lifeline Connect and Connectplus Install and Prog Guide v9.7
D5307013B Lifeline Connect and Connectplus Install and Prog Guide v9.7
Contents
Installation Guide................................................................................................3
Whats in the box ............................................................................................................ 3 Your home unit................................................................................................................ 4 Front view ..................................................................................................................... 4 Back view ..................................................................................................................... 4 End view....................................................................................................................... 4 For your safety - Installation advice............................................................................... 5 Quick start guide............................................................................................................. 6 Using the home unit........................................................................................................ 8 Making an alarm call ..................................................................................................... 8 Cancelling an alarm call ................................................................................................ 8 Answering calls remotely via the personal trigger .......................................................... 8 Status warnings .............................................................................................................. 9 Telephone line monitoring ............................................................................................. 9 Power failure monitoring................................................................................................ 9 The lights on the home unit indicate............................................................................ 10 The LED on the personal radio trigger indicates......................................................... 10 Programming a telecare sensor to the home unit ....................................................... 11 Range Test ................................................................................................................. 11
Technical Details...............................................................................................39
Environmental ............................................................................................................... 39 Standards ...................................................................................................................... 39 Declaration of Conformity............................................................................................. 39
Page 2 of 40
Installation Guide
Whats in the box
When you open the box for the first time, please ensure you have all of the following: Home unit Personal radio trigger Amie+ or Gem+
Amie+
Neckcord Neckcord
Belt clip Wrist strap Wrist strap Telephone lead (3 metre cable)
(not included)
Installation and User Guide If any of the above items are missing, please contact your supplier.
Page 3 of 40
Away Button (Yellow) Speaker Cancel Button (Green) Illuminated Alarm Button (Red)
Back view
Hardwired Input/Output
Microphone
Rubber Feet x 4
End view
LINE Telephone Lead Socket AUX TAPIT and Accessory Socket
- Telephone Adaptor
Page 4 of 40
All equipment requiring a link to the telephone line MUST be connected as follows:
Cordless phones
Ensure that the main base/charger which is registered to all other handsets in use, is connected directly to the home unit as above.
Safe Socket
Alternatively a Safe Socket (part number 36900/55) can be installed on all extensions used by other equipment, except the Lifeline, to ensure that alarm calls are raised even when the line is being used by another extension. Contact your supplier for more information.
Broadband
Please ensure a high quality ADSL filter is in use and the home unit is connected to the phone (analogue) socket on the filter. Please contact your supplier for further advice if necessary.
Dos
Keep the home unit connected to the mains power at all times. Connect the home unit to the first telephone point in the house with all other extensions wired into the unit to ensure proper operation even when another telephone is in use or off hook. Contact your supplier as soon as possible after the LED on the personal radio trigger indicates a low battery.
Donts
Expose the home unit to water or other liquids. Connect cables other than those supplied with the home unit. Place the home unit next to something that makes lots of noise, such as next to a television, radio or washing machine. Place the home unit close to a heat source e.g. cooker or large metal objects e.g. microwave.
Page 5 of 40
Step B Plug the telephone adaptor into the home unit socket labelled and then plug all required telephones / equipment into the telephone adaptor using a multi socket extension if required (not supplied). See page 5 for more details.
Step C Plug the mains adaptor into the home unit socket labelled AC and then connect to the mains power. Note ensure the mains power is switched on.
Step D Adjust speaker volume if required by depressing the volume control button on the underside of the home unit with a pen/pencil. After depressing for a few seconds, the home unit will emit a tone, release the pen/pencil when the tone reaches the required volume.
Page 6 of 40
Step 2 - Testing
Ensure the home unit is programmed to the correct telephone numbers (see page 15/16), then press the red alarm button on the home unit and ensure it raises a call through to the monitoring centre/personal recipient. Also remember to test the personal radio trigger by pressing its red button and ensuring a call is raised. The personal radio trigger test should be done at various points around the property to ensure the radio range provides sufficient coverage for the user to raise an alarm call using their personal radio trigger.
Wall mounting
Decide where you want to situate the home unit. Remember it should be within 2 metres of a mains power socket and the main telephone line socket. Then drill 2 holes 146mm apart, firmly attach screws (not supplied) leaving the screw heads protruding the surface and then locate the wall mounting points on the home unit with the screws. A full size template is available to download from the Tunstall website visit www.tunstall.co.uk/downloads, then search for Lifeline Connect/Connect+ wall mounting template.
146mm
NOTE: The diagram above is for illustrative purposes only and should not be used as a measuring tool i.e. it is not drawn to scale.
Page 7 of 40
Page 8 of 40
Status warnings
Telephone line monitoring
If the telephone line is faulty or becomes disconnected, the home unit will announce WARNING the telephone line is disconnected after 1 minute and the green LED flashes. This warning will be repeated every 30 seconds until the telephone line becomes available again.
To silence the warning, re-connect the telephone line. If the telephone line is connected and the warning continues, press the green cancel button. If the warning continues you should contact your telephone line supplier (e.g. BT) as the telephone line may be faulty.
If the power failure lasts for more than 1 hour, during the next hour the unit will automatically call the monitoring centre. A call will be raised every 4 hours to the monitoring centre until the power is restored. The battery provides 30 hours back-up.
Page 9 of 40
Lights Red alarm button on Red alarm button flashing (1 every 4 seconds) Red alarm button flashing (1 every second) Green LED flashing (2 every second) Green LED on Green LED flashing (in time with telephone ring) Yellow LED on Yellow LED off Yellow LED flashing (1 every 4 seconds) Yellow LED flashing (2 every second) No lights on
Home unit status Normal mode Normal mode running on battery (mains power off) Alarm mode Telephone line disconnected Telephone line in use Incoming call (telephone ringing) Function button in away mode Function button in home mode Low battery detected Intruder entry/exit time period Unit powered down (if power is on and connected then the unit may be faulty)
Page 10 of 40
Using the below quick code, the last assigned trigger can be given a location. 4zxx Set the last registered trigger for zone and location Must be done before exiting program mode where z = 0 for zone 1 and 1 for zone 2, xx = TT21 location code, see table 1
Range Test
The home unit has a range test feature that enables you to test the range of personal triggers without raising an alarm call. This is done by putting the home unit into programming mode (press and hold down the green cancel button until is bleeps). When in programming mode, press the required personal trigger if it is within range the home unit will bleep and announce the trigger or telecare sensor type.
Page 11 of 40
Programming guide
Programming of the home unit and its functions can be achieved using three different methods: PC Connect programming tool full programming can be achieved using a TAPIT programming tool linked to the home unit and a laptop running PC Connect software. Full help files are provided within the software. PNC software this method allows more in depth remote programming at the monitoring centre using custom designed screens within the PNC software (depending upon the software version) or via manual entry of parameters. Series telephone basic user programming can be achieved by using the keypad of a phone connected to the serial telephone socket of the home unit. This includes quick codes and manual entry of parameters. Instructions are included within this programming guide.
The following table provides an overview of which features can be configured using the above programming tools. For a full list of which features each Lifeline home unit can support, please see the Features at a glance section.
Icon Feature Telephone Numbers & IDs PC Connect Full PNC4 V2.5.1 Full Basic keypad/user option codes or Manual Entry Full Basic quick codes or Manual Entry Basic add/delete PNC5 Full Basic keypad/user option codes or Manual Entry Full Basic quick codes or Manual Entry Basic add/delete Series Telephone Basic keypad codes Basic keypad/user option codes
Inactivity Monitoring
Full
Call Sequences Intruder System and Away Options Radio Triggers Periodic Calls & AP (Auto Presence)* Hardwired Inputs Event Configuration
Full
Full
Full
Full
Full
Full
Full Non telecare sensor alarms/events only Yes Mains and telephone line failure only
Full Non telecare sensor alarms/events only Yes Mains and telephone line failure only
None
Full
Page 12 of 40
Reminders***
Manual Entry
Manual Entry
Full
Manual Entry
Manual Entry
None
Full
Manual Entry
Manual Entry
None
Time & Date Speech Configuration Output Configuration*** Remote Output Control
Full
Full
Full
Keypad code
Full
Manual Entry
Manual Entry
None
Full
Full
Full
None
None
None
Full
None
Full
Quick Codes
Full
None
None
None
* Auto Presence is not supported in the UK ** Programming home units using manual entry should only be done when advised by Tunstall. ***Lifeline Connect+ features only
Page 13 of 40
PNC4
In order to program the home unit from PNC4, version 2.5.1 is required, please contact Tunstalls customer satisfaction centre on 0844 415 2414 if your system requires updating. Older PNC4 software (version 2.5.0 and below) will recognise the home unit as a Lifeline 4000+ home unit however remote programming of the unit should not be done.
Step 2 Place the home unit in to programming mode by pressing and holding down the green cancel button until it bleeps (approx 5 seconds). The home unit announces Programming mode and the red alarm button flashes slowly.
Step 3 Lift the handset of the telephone handset and enter the quick codes listed on the following pages. Manual entry of parameters can also be completed via this method, however this method should only be used when advised by Tunstall.
NOTE: All programming not listed in this guide must be completed using the PC Connect programming tool or via the monitoring centre.
Page 14 of 40
This code means all functions are reset to default settings. The date and time remain unchanged. Resetting the home unit To reset all previous programmed information press:
Resetting erases all programmed telecare sensors and triggers and all functions are reset to default settings. The date and time remain unchanged. Setting the time and date There is a real time 24 hour clock in the home unit which automatically adjusts to BST. To set the clock press:
DDMMYY HH MM X Y
DD represents the day of the month (01-31) MM represents the months (01-12) YY represents the two digit year (00-99) HH represents hours 00-23; 24 cannot be programmed MM represents minutes 00-59; 60 cannot be programmed X represents the daylight saving time zone (0 = disabled, 1 = Europe, 2 = US) Y represents enable/disable auto CLI time update feature (0 = disable, 1 = enable) Telephone numbers The series telephone keypad supports the programming of 10 alarm numbers. By default, telephone numbers 1-4 are set to call control centres (CC) and telephone numbers 5-10 are set to a Personal Recipient (PR) destination. To change the destination from CC to PR or normal telephone (POTS) see the next section. Control centre numbers Control centre numbers are programmed by pressing: Tel. No. (max 20 digits)
Sets telephone number 1 To set telephone number 2 replace 00 with 01 To set telephone number 3 replace 00 with 07 To set telephone number 4 replace 00 with 08
Page 15 of 40
Changing telephone number destination to PR or POTS To program an existing telephone number to a PR or POTS destination, press:
XY
Where X represents the telephone number position (1-9 with 0 = 10) Where Y represents the destination type CC (0), PR (1) and POTS (2)
NOTE: It is important to set the correct destination type otherwise the recipient of the alarm call will not be able to deal with it correctly. A CC call expects a particular handshake from the control centre, a PR call requires a recipient with a touch tone telephone and a POTS call is a normal telephone call (i.e. fast dial button).
Unit ID numbers The home unit sends a unit ID number to the monitoring centre when an alarm is sent. The number identifies which home unit is sending the alarm. Unit ID number 1 must be programmed into the home unit (default 995) in order for an alarm to be sent. The unit ID number may be the same for all monitoring centres and personal recipients. If required the home unit can be configured to send a different unit ID to each telephone number it is programmed to call. Adding/Changing a unit ID Unit IDs can be programmed into the home unit by pressing: Unit ID (max 12 digits)
Sets Unit ID 1 To set Unit ID 2 replace 02 with 12 To set Unit ID 3 replace 02 with 13 To set Unit ID 4 replace 02 with 14 To set Unit ID 5 replace 02 with 15 To set Unit ID 6 replace 02 with 16 To set Unit ID 7 replace 02 with 17 To set Unit ID 8 replace 02 with 18 To set Unit ID 9 replace 02 with 19 To set Unit ID 10 replace 02 with 20
Page 16 of 40
Deletes Unit ID 1. Replace 02 with the numbers identified above to delete the appropriate Unit ID number.
NOTE: If no unit ID is linked to a telephone number, the first valid code will be used. The actual number of digits sent to the alarm receiver depends upon the type of monitoring centre being used. Please contact your monitoring centre for more information.
Prefix numbers A function can be enabled/disabled to ensure a prefix number is inserted before all dialled numbers from the home unit e.g. dialling 9 when using a PBX. This can be achieved by pressing: Prefix (max 8 digits) Suffix numbers To program a suffix please use PC Connect software. User Options Codes The following table provides a two digit code that enables you to set parameter 11 very simply. For example, to set a home unit to have; No inactivity monitoring, no line fail warnings (audio or visual) but with mains audio fail warning on, press;
Inactivity Period Line Disconnect Audio & Visual Warning? Disable Line Disconnect Audio Warning? Mains Fail Audio Warning? 65
12HRS
24HRS
64
61 60
25
24
21
20
75
74
71
70
35
34
31
30
45 44
41
40
05
04
01 00
Page 17 of 40
Quick Codes
Both the Lifeline Connect and Connect+ home units have a number of quick codes that can be entered into the series telephone when the home unit is in programming mode or remotely via PNC4 (v2.5.1) and PNC5.
Quick Code 2040 2050 2060 3000 4zxx Purpose Reset to default but retain radio triggers Reset to factory defaults Delete ALL radio triggers Delete the next radio trigger transmitted Set the last registered trigger for zone and location Remarks Time and date remain unchanged Time and date remain unchanged Restores to default Must activate trigger Must be done before exiting program mode where z = 0 for zone 1 and 1 for zone 2, xx = TT21 location code, see table 1 Where xx trigger type code, see table 2 x = 0 for disable, 1 = n/o, 2 = n/c y = 1 for zone 1 and 2 = zone 2 Where xx is TT21 location code, see table 1 Where x is telephone number 1 to 10 ( 0 = 10 ) y = 0 for CC, 1 for PR and 2 for POTS Where xx is 12 or 24 Default entry/exit time 30 secs Home unit must be in assign mode Home unit must be in assign mode Home unit must be in assign mode Home unit must be in assign mode Home unit must be in assign mode Home unit must be in assign mode Where xx is in seconds
Set hardwire input to trigger type number Set hardwired input sensor Set hardwired input location Set destination
51xx 6001 6002 6003 6004 6005 6006 6008 6009 61xx 6413 6403 6414 6404 6415 6405 9101 9108 9103
Enable inactivity monitoring for a period of 12 or 24 hrs Enable intruder Disable intruder Set last assigned trigger as a Zone 1 armer Set last assigned trigger to be a Bogus Caller Set last assigned trigger to be a Zone 1 and Zone 2 armer Set last assigned trigger to be Zone 1 and Zone 2 arm/disarmer Set last assigned trigger to start entry/exit tones on activation Set last assigned trigger to not start entry/exit tones on activation Enable intruder and set entry/exit tones Enable intruder disarm method of AWAY and personal trigger Disable intruder disarm method of AWAY and personal trigger Enable intruder disarm method by PIN Disable intruder disarm method by PIN Enable intruder disarm method by arm/disarm trigger Disable intruder disarm method by arm/disarm trigger Make all event calls silent and visual Make all event calls silent and non visual Restore all event calls to default states
Page 18 of 40
Installation and Programming Guide The following quick codes are only supported on the Lifeline Connect+
Quick Code 3011 3012 3013 3014 650x 6550 6551 7000 7010 7001 7002 7003 7004 7005 7006 7011 7012 7013 7014 7015 7016 Purpose Activate external relay for 2 seconds Activate external relay for all alarms when call selected and de-activate when calls cleared Activate external relay on radio smoke alarm and de-activate when cleared Disable external relay for all events Disable Virtual PES for x minutes Purge ADLife data Test call for ADLife Record PR message Delete PR message Record reminder message #1 Record reminder message #2 Record reminder message #3 Record reminder message #4 Record reminder message #5 Record reminder message #6 Delete reminder message #1 Delete reminder message #2 Delete reminder message #3 Delete reminder message #4 Delete reminder message #5 Delete reminder message #6 Remarks All alarms (Call Raised) Call Selected Call Cleared Smoke Alarm Call Cleared
Page 19 of 40
Features at a glance
The table below provides a full list of the features available in the Lifeline Connect and Lifeline Connect+.
Safety features 869 MHz European Social Alarm frequency - compatible with Tunstalls full range of telecare sensors. Radio reliability - the EN300 220-2 (2007) Class 1 radio receiver ensures that signals from sensors are reliably received. Periodic calls - an automatic test call can be set up to ensure the unit is working properly. Backup battery monitoring - unit reports the status of its backup battery to the monitoring centre. Programming & Installation features 10 telephone numbers (Monitoring Centre, Personal Recipient, POTS) - allows a different number to be dialled depending on what generates the alarm by routing calls to the most appropriate recipient. Plug & Play registration - telecare sensors can be assigned quickly and easily, together with their location within a dwelling, reducing installation times. Local programming - basic parameters and quick codes can be carried out using a normal telephone handset. Range test and walk test features - allow the radio range of the pendant and telecare sensors to be easily tested. Advanced local programming - downloadable PC Connect software application for programming of home unit via a connected PC or laptop (requires TAPIT programming interface). Telecare sensor inputs - the number of sensors that can be linked to the home unit. TT92 protocol support provides the monitoring centre with additional telecare information such as the type of telecare sensor that generated the alarm call. STMF protocol an alternative protocol that is specifically designed for use on digital telephone networks including GSM networks. TT21 protocol support provides enhanced and more detailed information to the monitoring centre and offers flexibility for future use. Service Support features Intelligent Speech Switching - ensures the best quality of speech depending on how the alarm was generated (requires PNC4 or later monitoring centre). Automatic British Summer Time update - removes the need to manually adjust the clock on the unit. User features Telephone answering with personal trigger - users can answer incoming telephone calls hands-free by simply pressing their personal trigger. Configurable audible ringing - can be set up to get progressively louder the longer the call goes unanswered. Automatic audible warning alerts - the unit alerts the user to mains and telephone line failure/resumption with a visual or audible and visual signal. New ergonomic design - a contemporary design to fit into the modern home environment. Easy switching between hands-free and handset mode - hands-free calls can easily be made private by picking up the handset of an attached telephone. Optional local audible warnings - non-critical warnings such as mains failure can be turned off at night to avoid disturbing the user. Technical call queuing - if the telephone is in use, non-critical calls (e.g. low battery warning) will be queued until after the call has ended, thereby not interrupting your call. Ability to signal a beep - if the user is unable to speak in the event of an alarm, they can press their personal trigger during an alarm call to signal to the monitoring centre.
*Feature disabled by default
12
35
Page 20 of 40
Page 21 of 40
Features explained
Telephone numbers & IDs
Up to 10 telephone numbers can be entered in the boxes. The destination type has to be changed to the correct type for each telephone number. There are three different destination types: Control centre - this should be used for all telephone numbers used for control centre call handling Personal recipient - this should be used for sending an alarm call to a normal house phone or mobile phone POTS - this is used when setting up a fast dial button on the home unit e.g. the away button used a fast dial button
The home unit sends a unit ID number to the control centre when an alarm is sent. The number identifies which home unit is sending the alarm. The specific unit ID field enables you to enter a different unit ID for each telephone number
Call sequences
The call sequence consists of up to 10 telephone numbers that the home unit can be set to dial in any order with multiple attempts to each alarm number. The home unit will ring each number in the order set up via PC Connect or PNC. If the home unit reaches the end of a call sequence without the alarm being answered it will start again at the beginning of the sequence. There are a total of 10 call sequences.
Standard Home/Away - the Away button will suspend inactivity monitoring and arm the intruder alarm if it is enabled. Service Key - the Away button act as a fast dial button and call a designated telephone number when pressed. Check in/Out Button - Setting the Away button to a Check in/Check Out Button will raise a carer arrived event on the initial press and a carer departed event on the subsequent press.
Radio triggers
The Lifeline Connect supports up to 12 (Lifeline Connect+ = 35) telecare sensors/radio triggers. Using PC Connect the radio triggers can be set up with the correct trigger type, location code and the usage of the trigger e.g. whether it is used as part of a virtual sensor.
Periodic calls
The home unit allows a periodic call event to be generated either at a configurable period or at a fixed time. In the configurable period case, the period between events can range from seconds through to days. In the fixed time case, the period between events is a configurable number of days. When the unit is configured to generate periodic call events at a configurable period, an initial offset time can be specified which must elapse before the first periodic call event is generated. This feature allows a unit that is configured during the day to generate periodic call events at a more appropriate time i.e. during the night.
Page 22 of 40
Intruder monitoring
The home unit has the ability to provide a simple to use intruder alarm facility, which will alert the monitoring centre or personal recipient on detection of an intruder. When configured using the series telephone keypad, the intruder monitoring function is simplified and uses a number of default settings. These settings other than the entry/exit times period, can only be configured using the PC Connect programming tool or via the monitoring centre. By turning the function ON using the keypad, Intruder monitoring will use the following settings. Arm method press function button, unit announces Away and entry/exit tones will be heard for 30 seconds. Disarm method press away button followed by the personal radio trigger, the unit will announce Home and the entry/exit tones will stop.
If an intruder detection event is detected that is within the armed zone(s) and is from an entry/exit sensor then the entry period will commence and entry tones will sound. The user has until the entry period expires to disarm the intruder system otherwise an intruder alarm will be generated.
NOTE: The intruder function can be configured to meet the individual users need using either the PC Connect programming tool or via the monitoring centre. This enables more complex settings to be configured including: different arming methods, optional entry/exit tones, how the unit reacts to intruder detection events (event-based configuration), zoning etc.
Event configuration
This feature enables the home unit to react to each event in a different way and allows these events to be configured via PC Connect and PNC (non telecare sensor events only) based on whether they should; raise an alarm call, act as an intruder/inactivity system input, provide visual/audible reassurance, enable the microphone/speaker, operate the relay output plus much more. The events are split into the following categories: Buttons Virtual sensors (Lifeline Connect+ only) Faults Telephony Misc
Page 23 of 40
Hardwired input
The hardwired input is located on the underside of the unit with a green 2 wire sprung terminal block. Inputs can be normally open or normally closed volts free contacts.
To set an input you will need to configure the unit accordingly either with the serial telephone Quick Codes below, a PNC5 monitoring centre or via PC Connect.
45xx 46xy 47xx Set hardwire input to trigger type number Set hardwired input sensor Set hardwired input location Where xx trigger type code, see table 2 x = 0 for disable, 1 = n/o, 2 = n/c y = 1 for zone 1 and 2 = zone 2 Where xx is TT21 location code, see table 1
Fault monitoring
Fault monitoring enables the settings to be changed to ensure the home unit reacts in the required way when it senses a fault such as power, telephony and battery failures.
Speech configuration
This feature configures how the speech prompts programmed into the home unit are used during alarms, local warnings and programming. Please see the help files within PC Connect for more details. Lifeline Connect only supports basic reassurance speech configuration whereas Lifeline Connect+ offers complete speech configuration.
Page 24 of 40
Inactivity monitoring
The home unit can monitor movement around the home and send an alarm call to the monitoring centre if no movement is detected within a specific time period. Inactivity monitoring has three different modes, Lifeline Connect supports only simple inactivity monitoring (mode 1) whilst Lifeline Connect+ supports all three modes. Mode 1 - Simple generates an alarm if the user is inactive for a configurable 12 or 24 hour period (continuous period). (Lifeline Connect and Connect+) Mode 2 - Real Time generates an alarm if the user is inactive between a configurable start and end time (time window). Two time windows are supported e.g. 7am 10am and 4pm 7pm (Lifeline Connect+ only). Mode 3 - Elapsed generates an alarm if the user is inactive for a period of time within a time window or continuous period. Two monitoring windows are supported e.g. raise an alarm call if the user is inactive for any 1 hour period between 7am10am and any 40 minute period between 4pm 7pm (Lifeline Connect+ only). In all modes, before an inactivity alarm is raised an inactivity warning period will occur. This is fixed at 10 minutes for Mode 1 and is configurable between 0 and 9 minutes for Modes 2 and 3. This warning period is intended to inform the user that an inactivity alarm is about to be raised therefore giving them the opportunity to cancel the alarm. After an alarm has been raised, inactivity monitoring can either be suspended until further activity is detected (all Modes) or can optionally restart immediately (Modes 2 and 3 only). When configured using the quick code, inactivity monitoring is simplified using default settings. The following quick code can be used to enable simple (mode 1) inactivity monitoring.
51xx Enable inactivity monitoring for a period of 12 or 24 hrs Where xx is 12 or 24
Configuration of advanced inactivity monitoring must be done via the PC Connect programming tool or monitoring centre.
NOTE: To avoid false calls to the monitoring centre, inactivity monitoring should be de-activated when the user leaves their home. Activate (home mode) - press the yellow away button (unit announces Home and the yellow LED will turn off) De-activate (away mode) - press the yellow away button (unit announces Away and the yellow LED will turn on)
Page 25 of 40
Page 26 of 40
NO
NC COM
Its operation can be controlled via a series telephone using the quick codes below, or by setting the correct boxes in the Remote Output Control of PC Connect or remotely by a PNC5 monitoring centre.
3011 3012 3013 3014 Activate external relay for 2 seconds Activate external relay for all alarms when call selected and de-activate when calls cleared Activate external relay on radio smoke alarm and de-activate when cleared Disable external relay for all events All alarms (Call Raised) Call Selected Call Cleared Smoke Alarm Call Cleared
Page 27 of 40
Installation and Programming Guide IVR reminder menu Step 1 Use a normal telephone (or mobile phone) to call the home unit. Step 2 Answer the call using the personal trigger or cancel key. If the call is answered by the user on their normal telephone, you must ask them replace the handset and answer the next call using their personal trigger or cancel key. Then call the home unit again. Step 3 - When answered correctly, press on the telephone keypad Step 4 - You will then be prompted to key in the PIN (default 1234) Step 5 - The time currently held on the home units internal clock will then be confirmed. Step 6 - You will then be given the below menu options. Firstly alter the time* if incorrect (menu option 3) and then follow the menu to configure and record each message.
MENU To add a reminder, press 1 INSTRUCTIONS Please type in the hour and then press *. Please type in the minute and then press * To repeat this reminder once only, press 1, to repeat this reminder daily press 2. Please record the reminder message now. Then return to main menu. Each reminder will be replayed followed by: To save this reminder, press 1. To remove this reminder, press 2. To set the time, press 3 Then return to main menu. Please type in the hour and then press *. Please type in the minute and then press *. The time will then be confirmed.
Recording reminder messages via a series telephone keypad To record a reminder message, press: Then record the message. Note: (1-6) represents the message slot number. If a message is already recorded under the number entered, this key sequence will replay the message. To re-record a message, the existing number must be deleted first. To delete a reminder message, press:
Note: (1-6) represents the message slot number. Configuring reminder messages PC Connect software is required to program reminder messages such as the time of the reminder message, the duration of the reminder bleep, whether the unit should announce Reminder or bleep and the regularity of the reminder (e.g. one off or every day). Page 28 of 40
Page 29 of 40
Page 30 of 40
Page 31 of 40
Installation and Programming Guide Configuring ADLife PC Connect is used to configure ADLife. The following steps are required: 1. Setting the PNC5 number - If the Lifeline Connect+ is required to send ADLife data to a different PNC5 (than normal alarm calls) then it is necessary to set up the alarm numbers and call sequence for ADLife calls. This is done in the standard way using PC Connect: a. Click on the Telephone Numbers and ID icon (on the main PC Connect screen) and enter the number(s) of the PNC5 that will receive ADLife calls in this case numbers 3 and 4. Click OK to exit the screen.
PNC alarm number 1 PNC alarm number 2 PNC ADLife number 1 PNC ADLife number 2
b. Click on the Call Sequences icon (on the main PC Connect screen) and configure Call Sequence 10 to an appropriate call sequence for the PNC5 numbers required to receive ADLife data calls. Click OK to exit the screen.
PNC alarm number 1 PNC alarm number 2 PNC ADLife number 1 PNC ADLife number 2
PNC ADLife PNC ADLife PNC ADLife PNC ADLife PNC ADLife PNC ADLife
number 1 (1) number 2 (1) number 1 (1) number 2 (1) number 1 (1) number 2 (1)
Page 32 of 40
Installation and Programming Guide 2. Configuring the ADLife Data Event - The next stage is to configure the Lifeline Connect+ home units response to the ADLife Data event. Click on the Alarms and Events icon (on the main PC Connect screen), the Alarms Event screen will open. Click the Misc tab and select the ADLife - Data option and then click configure. This will open the ADLife Data event configuration window. Ensure the Alarm Behaviour box is configured as shown and that the correct Call Sequence is selected. Click OK to exit the screen.
Page 33 of 40
Installation and Programming Guide 3. Configuring the ADLife Parameter Parameter 94 enables/disables ADLife, specifies the ADLife I.D. and the time to send ADLife data. An ADLife icon is available on the main PC Connect screen (V1.17 or later) to access this parameter. a. Click on the ADLife icon (on the main PC Connect screen) which will open the ADLife Configuration screen.
b. Tick the Enable ADLife box c. Enter the ADLife ID (supplied by Tunstall) d. Enter the time at which the ADLife data will be sent to the PNC5 monitoring centre. This is recommended to be between 03:00 and 04:00 (24 hour format) e. Enter the randomise send time. This programs the unit to send its data +/- 0, 1 or 2 hours from the actual send time previously set. f. Click OK to confirm the changes g. Set the correct time/date on the unit using the Time and Date icon (on the main PC Connect screen) h. Click Write, then either Write only modified parameters or Write all parameters except date and time. 4. Completing the ADLife installation a. Use the quick code (6550) to clean out any ADLife data from the unit. This will remove any data collected as a result of testing sensors etc. The unit will send ADLife data (within the time window specified by Parameter 94) and this can be viewed the next day on the ADLife Server. b. Test the setup a quick code (6551) is also available to force the unit to make an ADLife data call immediately, this data will then be presented on the ADLife website within 5 minutes. If the data doesnt appear check that the installation and configuration process has been followed correctly.
Page 34 of 40
Text Unit First resident personal Second resident personal Third resident personal Unspecified location Unspecified location Bogus Caller CO Detector activation CO Detector auto low bat Intruder tamper Arm/Disarm auto low bat Flood Detector 2 Flood Detector 3 Flood Detector 4 Flood Detector 5 Bed/chair not in by Bed/chair not up by Bed/chair absent Bed/chair other Bed/chair auto low bat Bedroom not specified Master bedroom Second bedroom Other bedroom Other bedroom ROM 1 event 1 ROM 1 event 2 ROM 1 event 3 ROM 1 event 4 ROM 1 auto low bat Bathroom WC not specified Main bathroom Second bathroom Downstairs WC Outside WC ROM 2 event 1 ROM 2 event 2 Electrical Usage auto low bat ROM 2 event 4 ROM 2 auto low bat Kitchen not spec Main kitchen area Second kitchen area Other kitchen area Other kitchen area ROM 3 event 1 ROM 3 event 2 ROM 3 event 3 ROM 3 event 4 ROM 3 auto low bat Living room area not spec
Text Living room Dining room Study Second living room ROM 4 event 1 ROM 4 event 2 Door usage auto low bat ROM 4 event 4 ROM 4 auto low bat Hall/stairs not spec Hall Landing Stairs Bath high level Med reminder no ack Auto presence failed Incontinence event Incontinence auto low bat Bath high temp Bath low temp Garage 1 Garage 2 Front garden Back garden Epilepsy Epilepsy spare Epilepsy auto low bat Carer arrived Carer not arrived Bath auto low bat Dose missed Med dispenser fault Med dispenser auto low bat CO end of life CO fault Temp rise Low temp Temp sensor fault High temp Temp sensor auto low bat Fall detector auto low bat Fall detector button press Fall detector fall Flood detector 1 Gas detector activated Gas detector auto low bat Door left open Property exit Property exit auto low bat
9 3 8 6 8 6 or 6 or 6 or 6 or 6 6 6 6 8
6 6 8 6 8
8 8 8 8
8 8 6 8 6 6
6 6 6 6 8
6 6 8 6 8
6 6 8 8 6 8 6 6 8 6 6 3 6 8 3 8 8 2 or 5 2 or 5 6 or 8 3 8 9 9 8
6 6 6 6 8
Page 35 of 40
Location
Blank Resident 1 Resident 2 Resident 3 Unspecified Unspecified Unspecified Unspecified Unspecified Unit Kettle Television Stove Microwave Toaster Vacuum Appliance 1 Appliance 2 Appliance 3 Appliance 4 Bedroom Master bedroom Second bedroom Third bedroom Fourth bedroom Other bedroom Other bedroom Other bedroom Other bedroom Other bedroom Bathroom /WC Main bathroom Second bathroom Downstairs WC Outside toilet En- suite Shower Other bathroom Other bathroom Other bathroom Kitchen Main Kitchen Second kitchen Other kitchen Other kitchen Other kitchen Other kitchen Other kitchen Other kitchen Other kitchen
TT21 Code
50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99
Location
Living area Living room Dining room Study Living room 2 Living area Living area Living area Living area Living area Hall/stairs Hall Landing Stairs Hall/stairs Hall/stairs Hall/stairs Hall/stairs Hall/stairs Hall/stairs Garden/garage Garage 1 Garage 2 Front garden Back garden Shed Garden/garage Garden/garage Garden/garage Garden/garage Front door Back door Fridge door Medicine cabinet Wardrobe door Food cupboard Other door 1 Other door 2 Other door 3 Other door 4 Unspecified Unspecified Unspecified Unspecified Unspecified Unspecified Unspecified Unspecified Unspecified Unspecified
Type Code
01 02 03 04 05 06 07 08 09 10 11 12 13 14
Trigger
Carer Trigger ROM #1 ROM #2 ROM #3 ROM #4 Nat Gas Property Exit Arm/Disarm Bogus Caller Electrical Usage Medication Dispenser Bath Sensor Epilepsy Zoning Trigger
Type Code
15 16 17 18 19 20 21 22 23 24 26 27 28 29
Page 36 of 40
Event
4 Digit Code
Event
Call Code
CR Cr D6 D7 CS Cs CT Ct D8 D9 CU Cu CV Cv E0 E1 CW Cw CX Cx CY Cy E2 E3 CZ Cz DA Da DB Db DC Dc E4 E5 DD Dd DE De DF Df DG Dg E6 E7 DH Dh Cn CO Co D4 CN D5 Di DJ Dj E8 E9 DK Dk DL Dl DM Dm F0
4 Digit Code
Personal Trigger activation Personal Trigger activation (LB) Personal Trigger AP Personal Trigger AP (LB) Personal Trigger ALB Personal Trigger AP Fail TES High Temp TES High Temp (LB) TES Fault TES Fault (LB) TES Low Temp TES Low Temp (LB) TES Temp Rise TES Temp Rise (LB) TES AP TES AP (LB) TES ALB TES AP Fail Flood Detector activation Flood Detector activation (LB) Flood Detector AP Flood Detector AP (LB) Flood Detector ALB Flood Detector AP Fail CO Detector activation CO Detector activation (LB) CO Detector EOL CO Detector EOL (LB) CO Detector fault CO Detector fault (LB) CO Detector AP CO Detector AP (LB) CO Detector ALB CO Detector AP Fail Smoke Detector activation Smoke Detector activation (LB) Smoke Detector AP Smoke Detector AP (LB) Smoke Detector ALB Smoke Detector AP Fail Door Usage opening Door Usage opening (LB) Door Usage closing Door Usage closing (LB) Door Usage AP Door Usage AP (LB) Pressure Mat activation Pressure Mat activation (LB) Pressure Mat AP Pressure Mat ALB Pressure Mat AP Fail Pullcord activation Pullcord activation (LB) Pullcord AP Pullcord AP (LB) Pullcord activation Pullcord activation (LB) Enuresis Sensor activation Enuresis Sensor activation (LB) Enuresis Sensor ALB Bed/Chair Not in Bed/Chair Not in (LB) Door Usage ALB Door Usage AP Fail
4143 4163 4144 4164 4145 4165 4146 4166 4147 4167 4132 4133
PES AP PES AP (LB) PES ALB PES AP Fail Arm/Disarm activation Arm/Disarm activation (LB) Arm/Disarm AP Arm/Disarm AP (LB) Arm/Disarm ALB Arm/Disarm AP Fail Bogus Caller activation Bogus Caller activation (LB) Bogus Caller AP Bogus Caller AP (LB) Bogus Caller ALB Bogus Caller AP Fail Electrical Usage Sensor ON Electrical Usage Sensor ON (LB) Electrical Usage Sensor OFF Electrical Usage Sensor OFF (LB) Electrical Usage Sensor AP Electrical Usage Sensor AP (LB) Electrical Usage Sensor ALB Electrical Usage Sensor AP Fail Medication Dispenser Dose Missed Medication Dispenser Dose Missed (LB) Medication Dispenser Device Fault Medication Dispenser Device Fault (LB) Medication Dispenser Dose Taken Medication Dispenser Dose Taken (LB) Medication Dispenser AP Medication Dispenser AP (LB) Medication Dispenser ALB Medication Dispenser AP Fail Bath Sensor High Level Bath Sensor High Level (LB) Bath Sensor High Temp Bath Sensor High Temp (LB) Bath Sensor Low Temp Bath Sensor Low Temp (LB) Bath Sensor AP Bath Sensor AP (LB) Bath Sensor ALB Bath Sensor AP Fail Epilepsy Sensor activation Epilepsy Sensor activation (LB) Natural Gas Detector activation (LB) Natural Gas Detector AP Natural Gas Detector AP (LB) Natural Gas Detector ALB Natural Gas Detector activation Natural Gas Detector AP Fail Epilepsy Sensor other (LB) Epilepsy Sensor AP Epilepsy Sensor AP (LB) Epilepsy Sensor ALB Epilepsy Sensor AP Fail Zoning Trigger Zone 1 Zoning Trigger Zone 1 (LB) Zoning Trigger Zone 2 Zoning Trigger Zone 2 (LB) Zoning Trigger AP Zoning Trigger AP (LB) Zoning Trigger ALB
Page 37 of 40
4253 4273 4254 4274 4255 4275 4256 4276 4257 4277
Page 38 of 40
Technical Details
Weight: Dimensions: Mains power: Stand-by battery: Back-up time: Radio frequency: REN: External connections: 650g 195 x 215 x 36mm (WxLxD) 230v ac 13A electrical socket 1200mAhr capacity (continually internally charged) 30 hours of stand-by operation with one 30 minute alarm call (minimum expected at date of purchase and when fully charged) 869.2125MHz, compliant with the European Social Alarm frequency band 1 3m telephone line cord with type BS6312 plug Plug top transformer with 3m cable
Environmental
Temperature: Humidity: Operating temperature (to perform to full specification) = 0 to 45 storage = -10 to 50 C C, C C Operating relative humidity (non condensing to perform to full specification) = 0 to 80%, storage relative humidity (non condensing) = 0 to 93%
Standards
EMC: EN55022: 1998, EN55024:1998, EN50130-4: 1995, ETSI EN 301 489-1: 2000 Safety: EN 60950:2000 Radio: ETSI EN 300 220-2 (2007) Class 1 CE: Compliant Social alarm: EN50134-1:2002, EN50134-2:1999 (trigger device), EN50134-3:2001 Design, Manufacture, Installation and Service: ISO9001:2000
Declaration of Conformity
We, Tunstall declare that this social alarm equipment is in compliance with the essential requirements and other relevant provisions of the R&TTE Directive 1999/5/EC.
Page 39 of 40
www.tunstallhealth.com A member of the Tunstall Healthcare Group Our policy of continual development means that product specification and appearance may change without notice. Tunstall does not accept any responsibility for any errors or omissions contained in this document. 2008 Tunstall Group Ltd. TUNSTALL, LIFELINE and AMIE are registered trademarks Tunstall Group Limited, Whitley Lodge, Whitley Bridge, Yorkshire DN14 0HR Tel: 01977 661234 Fax: 01977 662450 Email: [email protected]