Dinamap 400 SM

Download as pdf or txt
Download as pdf or txt
You are on page 1of 96

DINAMAP ProCare Monitor Service Manual

This manual is for DINAMAP ProCare Monitors models 100, 200, 300, and 400, with or without printers. ProCare 100: BP, Pulse ProCare 200: BP, Pulse, and Temp ProCare 300: BP, Pulse, and SpO2 ProCare 400: BP, Pulse, Temp, and SpO2 The model of the Monitor determines which parameters are in your monitor. Please refer to applicable sections.

Reissues and Updates


Changes occurring between issues are addressed through Change Information Sheets, Addendums, and replacement pages. If a Change Information Sheet does not accompany this manual, it is correct as printed.

Errors and Omissions


If errors or omissions are found in this manual, please notify: GE Medical Systems Information Technologies Technical Publications 4502 Woodland Corporate Boulevard Tampa, FL 33614 1-877-274-8456 Part No. 2009381-001 A The content of this document including all figures and drawings is proprietary information of GE Medical Systems Information Technologies, provided solely for purposes of operation, maintenance or repair, and dissemination for other purposes or copying thereof is prohibited without prior written consent by GE Medical Systems Information Technologies, Tampa, Florida. Illustrations may show design models; production units may incorporate changes.

Hierarchy of Warnings and Cautions


A general warning is a statement that alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the misuse of the device. A warning relates to steps in a procedure. A general caution is a statement that alerts the user to the possibility of a problem with the device associated with its use or misuse. Such problems include device malfunction, device failure, damage to the device or damage to other property. A caution relates to steps in a procedure. GE Medical Systems Information Technologies 2002, TAMPA, FL 33614 Printed in the U.S.A. All rights reserved.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

United States GE Medical Systems Information Technologies 4502 Woodland Corporate Boulevard Tampa, FL 33614 Germany GE Medical Systems Information Technologies GmbH Munzinger Strasse 3, - 7911 Frieburg, Germany

ProCare Patient Monitor Service Manual 2009381-001

Revision A

DINAMAP ProCare Monitor


Service Manual

2009381-001

Revision A

NOTE: The information in this manual only applies to ProCare Monitor. It does not apply to earlier Monitors. Due to continuing product innovation, specifications in this manual are subject to change without notice.

GE Medical Systems Information Technologies, 2002. All rights reserved.

T-2

ProCare Monitor 2009381-001

Revision A

Table of Contents

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Scope of Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.2 Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3 Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.1 Service Contracts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.2 Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.3 Return to Factory Repair Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.4 Service Loaners . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.5 Repair Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.3.6 Replacement Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 5 5 5 6 6 7

1.4 Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.4.1 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2.2. Product Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 2.3. Controls, Indicators, and Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 2.3.1. ProCare Monitor Rear Panel Connections . . . . . . . . . . . . . . . . . . . . . . . 4 2.3.2. Front Panel Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 2.4. Host Communications Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.4.1. DB15 Connector Pin Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 2.5. Compatible Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 2.6. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 3.2 Overall Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 3.2.1 SpO2 (Model 300 and 400) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Revsion A

ProCare Patient Monitor Service Manual 2009381-001

Table of Contents 3.2.2 Cuff Blood Pressure (BP) and Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 3.2.3 Temperature (Model 200 and 400) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 3.2.4 Host Communication Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 3.3 Functional Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.1 Main Board PWA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.2 User Interface (UI) Board PWA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.3 SPO2 PWA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.4 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.5 Pneumatic Valve/Manifold (PVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3.6 Optical Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 6 6 7 7 7 7

Calibration & Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 1


4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 4.2 Configuring Your ProCare Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.1 Unpacking and Preparation for Installation . . . . . . . . . . . . . . . . . . . . . . . 4.2.2 Set the Date and the Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.3 Parameter Level Functional Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 4 4 5

4.3 Periodic Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 4.3.1 As Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 4.3.2 Annual Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 4.4 Care of the Storage Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 4.4.1 Battery Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 4.5 Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Notes on Electrical Safety Testing of the ProCare Monitor: . . . . . . . 9 4.5.1 Temp Circuit Leakage Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 4.5.2 SpO2 Circuit Leakage Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 4.6 ProCare Patient Monitor Parameter Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.1 SETUP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.2 Leakage Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.3 Pressure Transducer Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.4 Pressure Transducer Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.5 Overpressure Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.6 Button Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.7 LED Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.8 External DC Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.9 NIBP Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.10 NIBP Overpressure Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 11 12 12 13 14 14 15 15 15 16

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Table of Contents 4.6.11 Temperature (Perform if equipped with Temp module) . . . . . . . . . . . 4.6.12 SpO2 (Perform only if equipped with SpO2 module) . . . . . . . . . . . . . 4.6.13 Printer Output Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.6.14 Communication Port Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 17 18 18

4.7 Alarm Code Interpretation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 4.7.1 System Failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Appendix A - Test Results Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 Appendix B - Connectivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Appendix C - Display Cover: Removal, Installation . . . . . . . . . . . . . . . . . . . . . . . . . 25 Appendix D: Replacement Parts and Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Assembly Drawings & ProCare Schematics

Revsion A

ProCare Patient Monitor Service Manual 2009381-001

Table of Contents

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Introduction

Revision A

ProCare Patient Monitor Service Manual 2009381-001

1-1

1-2

ProCare Patient Monitor Service Manual 2009381-001

Revision A

1.1 Scope of Manual


This Service Manual provides service and parts repair information about the DINAMAP ProCare Monitor. This manual is intended for use by trained service technicians who are familiar with electromechanical devices and analog circuit techniques.

WARNING

To reduce the risk of electric shock, do not remove cover or back of any component. Refer servicing to qualified service personnel.

Only qualified technicians should perform repairs to this equipment. For information about operating the Monitor in a clinical environment, refer to the separate Operations Manual. This Service Manual consists of the following five sections: Section 1 describes this volume and tells you how to use it. Information is also provided about the physical and functional characteristics of the Monitor, and how to get assistance in the event the unit fails to function properly. Section 2 provides a general overview of the ProCare Monitor including the user controls, external connections and product/parameter specification. Section 3 presents principles of operation for the Monitor, including an overall system description and principles of operation at the component level. Section 4 provides information about periodic and corrective maintenance of the Monitor and part list, replacement part lists. Procedures include module performance testing, and calibration procedures. Information is also provided to facilitate isolating faults to the subassembly level. Section 5 provides component information about the Monitor, including assembly drawings and electrical schematics.

1-3

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Introduction: 1.2 Manual Changes

1.2 Manual Changes


If, in the normal course of using this manual, you notice errors, omissions, incorrect data, or if you can suggest comments that may help improve this manual, please send suggestion to: General Electric Medical Systems Information Technologies Technical Publications 4502 Woodland Corporate Boulevard Tampa, Florida, 33614 Changes to the Service Manual, either in response to user input or to reflect continuing product improvements, are accomplished through reissue. Changes occurring between reissues are addressed through Change Information Sheets and replacement pages. If a Change Information Sheet does not accompany your manual, the manual is correct as printed.

1-4

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Introduction: 1.3 Service Policy

1.3 Service Policy


The warranty for this product is two years Parts and Labor. In the event of a monitor malfunction the monitor must be returned to General Electric Medical Systems Information Technologies to be covered under warranty. All repairs on products under warranty must be performed or approved by Product Service personnel. Unauthorized repairs will void the warranty. Only qualified electronics service personnel should repair products not covered by warranty.

1.3.1 Service Contracts


Service contracts may be purchased on most products. Contact your Sales Representative for details and pricing.

1.3.2 Assistance
If the product fails to function properly, or if assistance, service or spare parts are required, contact Customer Support. Before contacting Customer Support, it is helpful to attempt to duplicate the problem and to check all accessories to ensure that they are not the cause of the problem. If you are unable to resolve the problem after checking these items, contact General Electric Medical Systems Information Technologies at 1-877-274-8456. Prior to calling, please be prepared to provide: product name and model number a complete description of the problem If repair parts or service are necessary, you will also be asked to provide: the product serial number the facility's complete name and address a purchase order number if the product is in need of non-warranty repair or to order spare parts the facility's General Electric Medical Systems Information Technologies account number the appropriate part number for spare or replacement parts

1.3.3 Return to Factory Repair Service


If your product requires warranty, extended warranty or non-warranty repair service, call Customer Support and a representative will assist you. Estimates for non-warranty repairs are provided pre-set flat rates to facilitate prompt service. In cases where the product has external chassis or case damage, please advise the Customer Support representative when you call. The Customer Support representative will record all necessary information and will provide you with a Return Merchandise Authorization Number (RMA). Prior to returning any product for repair, you must have a RMA number.

Revsion A

ProCare Patient Monitor Service Manual 2009381-001

1-5

Introduction: 1.3 Service Policy Contact General Electric Medical Systems Information Technologies at 1-877-274-8456 Monday through Friday, 8:00 a.m. to 6:00 p.m. EST, excluding holidays. Packing Instructions Follow these recommended packing instructions. Remove all hoses, cables, sensors, and power cords from the monitor before packing. Pack only the accessories you are requested to return; place them in a separate bag and insert the bag and the product inside the shipping carton. Use the original shipping carton and packing materials, if available. If the original shipping carton is not available: Place the product in a plastic bag and tie or tape the bag to prevent loose particles or materials from entering openings such as hose ports. Use a sturdy corrugated container to ship the product. Pack with 4 to 6 in. of padding on all sides of the product. Tape securely to seal the container for shipping. Insurance Insurance is at the customer's discretion. The shipper must initiate claims for damage to the product incurred during shipping. Shipping damage is not covered under warranty.

1.3.4 Service Loaners


Loaner units can be provided at no charge during the warranty period if your product needs to be returned for factory service. If requested, loaner units will be shipped within 48 hours to your facility. General Electric Medical Systems Information Technologies will pay outgoing charges for a loaner sent for product repairs under warranty. The customer is responsible for shipping charges to return the loaner unit to General Electric Medical Systems Information Technologies. All loaner units must be returned within 5 business days after receipt of your repaired unit. Loaner and/or Rental products are available to meet your needs in nonwarranty situations.

1.3.5 Repair Parts


Repair parts can be ordered from General Electric Medical Systems Information Technologies: Via phone: 1-877-274-8456, or Via FAX: 1-800-421-6841 Exchange replacement assemblies such as Circuit Board Assemblies are also available; ask your Customer Support representative for details.

1-6

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Introduction: 1.3 Service Policy Most orders ship on the same day if received before 3PM Eastern Time. Overnight freight options are available to meet your critical needs. All orders must include the following information: Facility's complete name, address, and phone number FAX number Your purchase order number Your General Electric Medical Systems Information Technologies account number (if available)

1.3.6 Replacement Accessories


Replacements such as hoses, sensors, etc. must be purchased from General Electric Medical Systems Information Technologies at 1-800-5585102. Please have the Product Code of the item you wish to order, your purchase order and account number available.

Revsion A

ProCare Patient Monitor Service Manual 2009381-001

1-7

Introduction: 1.4 Product Description

1.4 Product Description


The Monitor is described below. Refer to Table 1-1 for specifications.

1.4.1 General Description


The ProCare Monitor provides a small, portable, easy-to-use monitoring alternative for sub-acute hospital and non-hospital settings. The DCoperated Monitor offers noninvasive determination of systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate, oxygen saturation, and temperature. Monitors are available with or without integrated printers. ProCare Monitors are intended for use in various markets, from the physicians office to sub-acute triage and medical/surgical units. Indicators for external DC operation (from AC mains), battery operation, and battery charging are at the front of the unit. At the time of publication, the available functioning parameters included the following: NIBP Nellcor Pulse oximetry (SpO2) Masimo Pulse oximetry (SpO2) Alaris Oral and Rectal thermometry thermal recorder/printer Other DINAMAP ProCare features include: The ability to uses industry standard accessories Remote alarm capability Function keys for quick access to Alarm Silence, Monitor Menus, Print and NIBP Inflate/Stop The ProCare Monitor operates from either an external DC power supply, or from the internal lead-acid storage battery. When external DC power becomes available, the system rapidly switches from battery power to external power.

1-8

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Introduction: 1.4 Product Description

Revsion A

ProCare Patient Monitor Service Manual 2009381-001

1-9

Introduction: 1.4 Product Description

1-10

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Product Description

Revsion A

ProCare Partient Monitor Service Manual 2009381-001

2-1

For your notes

2-2

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.1 Introduction

2.1 Introduction
DINAMAP ProCare Monitors provide non-invasive determination of systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse rate, temperature, and oxygen saturation.

2.2. Product Configurations


Each ProCare Monitor is supplied with an accessory pack. The contents of the pack vary according to model. Unpack the items carefully, and check them against the checklists enclosed within the accessory boxes. If an accessory is missing or if an item is in a nonworking condition, contact General Electric Medical Systems Information Technologies Customer Service immediately. It is recommended that all the packaging be retained, in case the ProCare Monitor must be returned for service in the future.

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-3

Product Description: 2.3. Controls, Indicators, and Connectors

2.3. Controls, Indicators, and Connectors


Descriptions of the items shown are listed on the pages that follow. For symbol definitions, refer to page 2-8 of this section.

2.3.1. ProCare Monitor Rear Panel Connections

1 Speaker Grille. 2 Data interface connector: Host communications port (15 pin D-type RS232 serial port) for use only with equipment conforming to IEC 601-1, configured to comply with IEC 601-1. 3 Printer Door (optional.)

2-4

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.3. Controls, Indicators, and Connectors

2.3.2. Front Panel Controls and Indicators

Buttons 1 Silence button: Press to mute audible alarms. Any alarm active that is acknowledgeable is also removed whenever this key is pressed. When pressed after alarm sounds (silence active), the silence icon (bell) lights to indicate that audible alarms have been silenced for 2 minutes. 2 Alarms button: Press to view or adjust parameter alarm settings 3 +/- button (Plus/Minus): Press the + button to increment an adjustable setting and the - button to decrement an adjustable setting. This button is active only when a user-setting mode (limit or menu) is active. 4 Menu button: Press to access menu settings that can be adjusted while in clinical mode (i.e., ALARM VOLUME, PULSE VOLUME, INFLATE PRESSURE; refer to Operating Modes in this section for a description of clinical mode.) 5 SpO2 sensor connector: SpO2 sensor cable attaches here . 6 BP connector: BP cuff hose attaches here. 7 Inflate/Stop button: Press to start a manual BP determination or stop any BP determination. 8 Temperature probe holster: Temperature probe is stored here. 9 Cycle button: Press to start Auto Cycle or STAT mode. 10 Temperature probe cover storage: Box of probe covers is stored here 11 History button: Press to activate the history mode. When activated, it displays the most recent entries stored. Press and hold the button for 2 seconds to clear all entries stored. 12 Print button: Press to print currently displayed values or all stored entries when in history mode. 13 On/Off button: Controls on/off state of monitor; push for power on and push again for power off. 14 Temperature probe connector: Temperature probe cable attaches here.

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-5

Product Description: 2.3. Controls, Indicators, and Connectors

Front Panel 15 Silence icon: when Silence button is pressed after alarm sounds (silence active), silence icon (bell) lights to indicate that audible alarms have been silenced for 2 minutes. 16 Systolic window: 3-digit red LED indicates measured systolic Blood Pressure in mmHg. 17 Diastolic window: 3-digit red LED indicates measured diastolic Blood Pressure in mmHg. 18 Alarm volume indicator: lights to indicate you are making a change to the alarm volume. 19 Pulse volume: illuminates to indicate you are making a change to the pulse volume. 20 Inflate pressure: illuminates to indicate you are making a change to the inflation pressure. 21 Pulse Rate window: 3-digit yellow LED shows the pulse rate in beats per minute. 22 SpO2 pulse indicator: Red LED bar flashes to indicate that real-time pulse rate measurements are being derived from SpO2 signals. 23 SpO2 window: 3-digit red LED indicates oxygen saturation in %. 24 MAP/Cuff window: 3-digit red LED indicates measured MAP in mmHg and shows instantaneous cuff pressure during BP determination. 25 Min window: Displays the BP mode if manual or STAT is the cycle time when in Auto Cycle mode. 26 Battery power indicator: Green LED indicates the Monitor is operating on battery power. 27 Low battery power indicator: Yellow LED indicates LOW charge status of internal battery. 28 Charging indicator: Green LED indicates presence of external power source and battery charging. 29 Temperature window: 4-digit red LED indicates measured temperature.

2-6

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.3. Controls, Indicators, and Connectors

Right-Side Panel
30 External power socket: To be used with approved GE Medical Systems Information Technologies AC-DC power converter ONLY.

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-7

Product Description: 2.3. Controls, Indicators, and Connectors

Symbols
The following symbols are associated with the ProCare Monitor. Note: The model of the Monitor determines which symbols appear on it. Attention, consult accompanying documents Silence Alarms +/Menu Inflate/Stop Cycle History Print On/Off Battery Power External Communications Port Connector Charging Defibrillator-proof type BF equipment Class II equipment according to IEC 60536

Packaging label depicting the transportation and storage atmospheric pressure range of 500 to 1060 hPa.

2-8

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.4. Host Communications Connector

2.4. Host Communications Connector


All host port signals are NON-ISOLATED and should be connected to equipment conforming to IEC 601-1 ONLY. Where isolation of data communication is required, the isolated level converter should be used. If external alarm control is required, p/n 487208 (Isolated Remote Alarm Cable Assembly) should ALWAYS be used. Please refer to the Information Sheet included with the isolated remote alarm cable for operational details. Note: When using remote alarm, the ProCare Monitor should be considered the primary alarm source. The secondary alarm is used for secondary purposes only.

2.4.1. DB15 Connector Pin Assignments


Connection Details
Host Port Connector (rear panel) WARNING! Auxiliary equipment connected to the DINAMAP ProCare Monitor will result in the formation of an electromedical system and thus must comply with the requirements of EN 60601-1-1/ IEC 601-1. All host port signals are NON-ISOLATED and should be connected to equipment conforming to IEC-601-1, configured to comply with IEC 601-1-1 ONLY. Where isolation of data communication is required, GE Medical Systems Information Technologies part number ILC1926 should be used. If external alarm control is required, GE Medical Systems Information Technologies part number 487208 (Isolated Remote Alarm Cable Assembly) should ALWAYS be used. When a high-priority condition is displayed on the Monitor, the remote alarm signal becomes active within 0.5 seconds. The active state of the alarm signal is an open circuit. In the inactive state the alarm signal is connected to ground. Please refer to the Information Sheet included with the isolated remote alarm cable for operational details. Note: When using remote alarm, the ProCare Monitor should be considered the primary alarm source. The secondary alarm is used for secondary purposes only. Pin # 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Revision A

Function Common Inverted TTL Transmit Data Inverted TTL Receive Data +5 volts No connection No connection Common Remote Alarm No connection No connection RS232 Transmit Data (TxD) No connection RS232 Receive Data (RxD) No connection No connection
2-9

ProCare Partient Monitor Service Manual 2009381-001

Product Description: 2.5. Compatible Parts

2.5. Compatible Parts

Reorder Codes DINAMAP ProCare Monitor Operations Manual DINAMAP ProCare Monitor Service Manual Battery Power Cord Printer Paper (box of 10) DINAMAP Rolling Stand Pole Mounting Option Wall Mounting Option Hose Management Option Bed Rail Option Nurse Call NIBP OPTIONS: Airhose 12 ft Adult/Pediatric, Screw Connector Airhose 24 ft Adult/Pediatric, Screw Connector Airhose 12 ft Neonatal, Quick Disconnect Airhose 12 ft Adult/Pediatric, Quick Disconnect CUFF ASSORTMENT PACKS: CLASSIC-CUF Assortment pack Includes 1 each: Infant, Child, Small Adult, Adult, Large Adult, Thigh Cuff CLASSIC-CUF Assortment Pack, Neonate Includes 2 Neo #1, 3 Neo #2, 5 Neo #3, 5 Neo #4, 5 Neo #3 SOFT-CUF Assortment Pack Includes 1 each: Infant, Child, Small Adult, Adult, Large Adult, Thigh Cuff SOFT-CUF Assortment, Neonate Includes 2 Neo #1, 3 Neo #2, 5 neo #3, 5 Neo #4, 5 Neo #5 DURA-CUF Assortment Pack Includes 1 each: Infant, Child, Small Adult, Adult, Large Adult, Thigh Cuff DURA-CUF Assortment Pack, Adult Includes 1 each: Infant, CHild, Small Adult, Adult, Large Adult, Thigh Cuff

Product Code 2009360-001 2009381-001 633178CR 316579 089100 003215 2009762-001 2009763-001 2009764-001 2009765-001 487208CR

ECAT

E9050KP E9050JB

107365 107366 107368 88847

E9050LH E9050LJ E9050LK E9050KN

2692 2693 2695 2694 2699 2698

E2692J E2693J E2695J E2694J E2699J E2698J

2-10

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.5. Compatible Parts Reorder Codes DURA-CUF Assortment Pack, Child Includes 2 Infant, 3 Child, and 1 Small Adult Cuff Product Code 2697 ECAT E2697J

Additional Blood Pressure Cuffs are available through http://www.gemedicalsystems.com TEMPERATURE IVAC* Turbo Temp Oral Temperature Probe, Long Cord IVAC* Turbo Temp Rectal Temperature Probe, Long Cord IVAC* Temperature Probe Covers SpO2 Nellcor**: Pulse Oximeter Cable DOC-10 DuraSensor Adult Oxygen Sensor Masimo***: Adult Reusable Sensor, 1/Bx (NR125) Cable (PC08) * IVAC is a trademark of Alaris Medical Systems ** NELLCOR is a trademark of Mallinckrodt *** Masimo is a trademark of Masimo Corporation 2009745-001 2009743-001 2008773-001 DS100A 2008774-001 2008775-001 88015 E9050KK

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-11

Product Description: 2.6. Specifications

2.6. Specifications
0 0 8 6 IPX1 Power Requirements MAINS AC INPUT VOLTAGE Alternate Sources DC INPUT VOLTAGE Protection against electrical shock - Class II 120 VAC/60 Hz 24W Protection against electrical shock - Class II 24 VDC (nominal), at 1A The AC Mains adapter contains a nonresettable and nonreplaceable fuse. Battery Minimum Operation Time Fuses 6 volt, 3.3 amp-hours. Protected by internal auto-resetting fuse and thermal protection. 2 hrs (5 min cycle with adult cuff at 25 C, SpO2 active at 60 bpm, temp in monitor mode, printout of current values every 5 minutes) from full charge. The Monitor contains four fuses. The fuses are auto-resettable and mounted within the Monitor. The fuses protect the low voltage DC input, the battery, the remote alarm output and the +5V output on the host port connector. This product conforms with the essential requirements of the Medical Device Directive. Accessories without the CE Mark are not guaranteed to meet the Essential requirements of the Medical Device Directive. The ProCare Monitor is protected against vertically falling drops of water and conforms to the IEC-529 standard at level IPX1. Vertically falling drops shall have no harmful effects to the monitor.

Environmental Operating Temperature Storage Temperature Operating Atmospheric Pressure Range Humidity Range Radio Frequency +5 C to + 40 C (-41 F to +104 F) -20 C to +50 C (-4 F to +122 F) 700 to 1060 hecto Pascal 0% to 95% non-condensing Complies with IEC Publication 601-1-2 (April 1993) Medical Electrical Equipment, Electromagnetic Compatibility Requirements and Tests, and CISPR 11 (Group1, Class A) for radiated and conducted emissions.

Mechanical Dimensions Height: Width: Depth: Weight including battery Mountings Portability 7.8 lb (3.5 Kg) Self Supporting on rubber feet. Carried by built-in handle 9.8 in (25.0 cm) 9.8 in (24.8 cm) 6.9 in (17.5 cm)

2-12

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.6. Specifications Mechanical Classification Information Mode of Operation: Continuous degree of Protection against harmful ingress of water: Drip-Proof IPX1.

NIBP CUFF PRESSURE RANGE DEFAULT TARGET: CUFF INFLATION TARGET CUFF INFLATION ADJUSTMENT RANGE BLOOD PRESSURE DETERMINATION TIME PULSE RATE: RANGE BLOOD PRESSURE: MEASUREMENT RANGES Adult Neonate NIBP ACCURACY Adult Neonate Adult Neonate Adult Neonate Adult Neonate Adult Neonate Systolic mmHg 30 - 290 30 - 140 0 mmHg to 290 mmHg 0 mmHg to 145 mmHg 150 15 mmHg 110 15 mmHg 100 - 250 mmHg (5mmHg Steps) 100 - 140 mmHg (5 mmHg Steps) 120 seconds maximum 85 seconds maximum 30 - 200 BPM 3% 30 - 220 BPM 3% MAP mmHg 20 - 260 20 - 125 Diastolic mmHg 10 - 220 10 - 110

Meets AAMI/ANSI standard SP-10 AAMI/ANSI standard: 5 mmHg mean error Intra-arterial method: 8 mmHg standard dev.

Temperature - Alaris Thermometry SCALES RANGE Monitor Mode Accuracy Predictive Mode Accuracy Determination Time SpO2 - Nellcor Pulse Oximetry SpO2 Range and Accuracy Pulse Rate Range and Accuracy Saturation Pitch Indicator Sensor Connect/Disconnect from Patient adult/neonate: 70-100% 3.5 digits 30 - 250 BPM 3 BPM Pitch changes with saturation The Monitor detects the connection or disconnection or a sensor to the patient within 15 seconds Fahrenheit Max Min 0.1 C 0.6 C Less then 60 seconds Celsius 42.2 Celsius 108.0 Fahrenheit 31.6 Celsius 88.9 Fahrenheit 0.2 F (when tested in a calibrated liquid bath; meets ASTM E1112, Table 1 in range specified) 1.0 F

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-13

Product Description: 2.6. Specifications

SpO2 - Nellcor Pulse Oximetry Sensor Connect/Disconnect from Monitor Pulse Detection Loss of Pulse Nellcor Sensors - SpO2 Range 70% to 100% OxiMAX Sensor Models Single Patient Use MAX-A,* MAX-AL* MAX-N* MAX-P* MAX-I* MAX-R OxiCliq Sensor Models Single Patient Use OxiCliq A OxiCliq P OxiCliq N Reusable Sensor Models D-YS D-YS & D-YSE D-YS & DYSPD DS-100A OXI-A/N OXI-P/I 3 3.5 3.5 3 3 3 2.5 2.5 2.5 2 2 2 2 3.5 The monitor detects the attachment or disconnection of a sensor from the Monitor within 5 seconds. The Monitor detects a pulse of enters a no-signal state within 15 seconds of being attached to a patient. the monitor detects loss of pulse from patient and enters a no signal state within 10 seconds.

* The accuracy specification under motion conditions is 3. For a definition of motion contact GE Medical Services Technical Support. The MAX-N and the OxiCliq N were tested on patients > 40 kg. The accuracy specification has been determined between saturations of 80-100%

SpO2 - Masimo Oximetry Saturation Range Pulse Rate and Accuracy Perfusion Range 1% - 100% 25-240 BPM 3 digits 0.02 to 20%

2-14

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Product Description: 2.6. Specifications

SpO2 - Masimo Oximetry Accuracy and Motion Tolerance Without Motion - Adult/Ped* Without Motion - Neonate* With Motion - Adult/Ped/Neo** Low Perfusion Pulse Rate Without Motion With Motion 25 to 240 beats/min 3 digits normal physiologic range 25 to 240 beats/min 5 digits * The Masimo SET SpO2 parameter with LNOP-Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. **The Masimo SET SpO2 parameter with LNOP-Adt sensors has been validated for motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an amplitude of 1 to 2 cm and a nonrepetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. The Masimo SET SpO2 parameter with LNOP-Neo Pt sensors has been validated for neonatal motion accuracy in human blood studies on neonates while moving the neonates foot at 2 to 4 cm against a laboratory cooximeter and ECG monitor. This variation equals plus or minus, one standard deviation.Plus or minus one standard deviation encompasses 68% of the population. The Masimo SET SpO2 parameter with has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 stimulator and Masimos simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus, one standard deviation.Plus or minus one standard deviation encompasses 68% of the population. 70 to 100% 2 digits 70 to 100% 3 digits 70 to 100% 3 digits 70 to 100% 2 digits 0 to 69% unspecified

Revision A

ProCare Partient Monitor Service Manual 2009381-001

2-15

Product Description: 2.6. Specifications

2-16

ProCare Partient Monitor Service Manual 2009381-001

Revision A

Principles of Operation

Revision A

ProCare Patient Monitor Service Manual 2009381-001

3-1

For your notes

3-2

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Principles of Operation: 3.1 Introduction

3.1 Introduction
This section provides overall theory of operation and functional description of the ProCare Monitor. The ProCare Monitor comes in six different configurations:

ProCare 100 - Capable of monitoring Blood Pressure (BP) and Pulse ProCare 200 - Capable of monitoring Blood Pressure (BP), Pulse, and
Temperature

ProCare 300 Nellcor - Capable of monitoring Blood Pressure (BP), Pulse and SPO2 (Nellcor technology) ProCare 300 Masimo - Capable of monitoring Blood Pressure (BP), Pulse and SPO2 (Masimo technology) ProCare 400 Nellcor - Capable of monitoring Blood Pressure (BP), Pulse, SPO2 (Nellcor technology), and Temperature ProCare 400 Masimo - Capable of monitoring Blood Pressure (BP), Pulse,
SPO2 (Masimo technology), and Temperature The model of your monitor determines which parameters are in your monitor. Using the ProCare monitor, a clinician can view, print and recall data that is derived from each parameter. The Monitor is also capable of alerting the clinician to changes in the patients condition. All of the main operations of the ProCare Monitor are easy-to-use and only a button-touch away. Please review the factory default settings and, where applicable, enter settings appropriate for your use.

Revsion A

ProCare Partient Monitor Service Manual 2009381-001

3-3

Principles of Operation: 3.2 Overall Principles of Operation

3.2 Overall Principles of Operation


The following paragraphs provide a general system interface relationship. The general block diagram is located in Figure 3-1. The ProCare Monitor is a portable unit that receives input power from an internal rechargeable battery. When the ON/OFF button is pressed, the Main Board is brought out of a sleep mode and turns on the power regulators. The power regulators provide conditioned power from the Lead Acid Battery. The external DC source is used only to charge the Lead Acid Battery. Once the ProCare Monitor is energized, a self-test is performed. The self-test automatically tests the main functions of the ProCare Monitor. Failure of the self-test will set the ProCare Monitor into a fail-safe mode with an audio alarm. Under normal operating condition, the ProCare Monitor is ready to record the patient vital signs using three external attachments: the temperature probe, SPO2 sensor, and cuff. Interface with a central station or other device is accomplished through the host communication port on the back of the ProCare Monitor.

3.2.1 SpO2 (Model 300 and 400)


The SpO2 probe has a built-in sensor. When the SpO2 sensor is attached to the SpO2 connector and patient, the probe senses the heart rate and oxygen saturation. The analog signals are routed to the SpO2 PWA (Nellcor or Masimo). The analog signals are analyzed on the SpO2 PWA. The results are digitized and sent to the Main Board via opto couplers. The couplers provide patient isolation as well as serial data interface. The Main Board temporarily stores the data and routes it to the UI Board for display and/or printer. A reset signal to the SpO2 PWA is also provided so that the power up sequencing is corrected. If the SpO2 circuit quits communicating to the Main Board, the Main Board will attempt to reset the SpO2 PWA.

3.2.2 Cuff Blood Pressure (BP) and Pulse


When the cuff and hose are attached to the ProCare Monitor and a NonInvasive Blood Pressure (NIBP) determination is initiated, the pump inflates the cuff. Pressure transducers PT1 and PT2 monitor pressure information. The pneumatic manifold has one valve, which is used to deflate the cuff. Valve control is through the Main Board. Once determinations are made for the systolic BP and diastolic BP, the Main Board calculates the pulse rate/ Mean Arterial Pressure (MAP). The results are then displayed on the UI Board and sent to the printer (if specified). The Pneumatic Valve/Manifold (PVM) device is controlled by the secondary processor. The secondary processor monitors pressure information from PT2. If an over-inflation condition occurs, the OVERPRESSURE signal is routed to the PVM to release the air pressure. The Main Board also generates an alarm condition with the speaker sounding and error code message on the UI Board.
3-4 ProCare Patient Monitor Service Manual 2008381-001 Revision A

Principles of Operation: 3.3 Functional Description

3.2.3 Temperature (Model 200 and 400)


The ProCare Monitor uses Alaris Turbo Temp technology to measure patient temperature. The Turbo Temp probe contains a heating element that preheats the probe to reduce determination time. The heating function is controlled by the Main Board. The Turbo Temp probe also contains a thermistor that indicates the temperature. When the probe is attached to the temperature connector and patient, the signal generated by the thermistor is routed to the Main Board. The Main Board converts the thermistor signal along with status information (i.e ORAL or RECTAL probe indicators) to a DIGITAL signal. The Main Board then processes the DIGITAL signal and displays the patient temperature on the UI Board and printer in Celsius or Fahrenheit.

3.2.4 Host Communication Port


The Host Comm Port is used to interface the ProCare Monitor with other electronic devices (a central nurse's station or remote alarm device.) Signals can be sent to the ProCare Monitor to initiate blood pressure determinations and other functions. Patient data can also be retrieved through this port. For further information, reference the Dinamap ProCare Series Host Communication manual.

3.3 Functional Description


The following paragraphs provide the functional interface relationship. The ProCare Monitor contains a number of electrical & electro-mechanical assemblies. These assemblies are: * Main Board PWA * User Interface (UI) Board PWA * SPO2 PWA (optional) * Printer (optional) * Pneumatic Valve/Manifold (PVM) * Optical Switch (optional)

Revsion A

ProCare Partient Monitor Service Manual 2009381-001

3-5

Principles of Operation: 3.3 Functional Description

3.3.1 Main Board PWA


The ProCare Main Board is based on the Motorola MMC2107 integrated microprocessor. The microprocessor integrates Flash ROM, RAM, A/D converter with input multiplexor, SPI interface, and timers into one chip. This microprocessor is the primary processor for the ProCare Monitor. It services and controls the Patient Parameter Interface (PPI) devices, printer, UI Board, Real Time Clock, audio circuit, and host communication. The secondary processor controls the watchdog, pneumatic safety interlock, timing check, primary processor reset, and power supply control. The secondary processor is powered at all times. Independent software in the primary and secondary processor periodically communicate when the software systems are operating properly. When either system stops processing or detects an error, it stops communicating with the other. Either system, upon detecting a failure, can assert a safe state (herein called FAILSAFE) of the hardware. Upon entering a FAILSAFE condition, the Main Board will perform the following tasks: * Parameter monitoring disabled * Alarm tone sounding from speaker * Pneumatic FAILSAFE (deflate the cuff, pump off) * Normal communications interface disabled * Remote alarm control inactive * Hard keys except ON/OFF key inactive The ON/OFF key can reset the Monitor and end the FAILSAFE condition. The FAILSAFE condition will terminate automatically after 10 minutes to preserve battery power. All regulated DC power, isolated and non-isolated is generated on the Main Board from Battery supply. The external DC input is used to charge the battery via charging circuitry on the Main Board.

3.3.2 User Interface (UI) Board PWA


The UI Board is used as a message center. It displays patient vital signs, alarms status, monitor set-up, limit violation, BP cycle and the time the data was received. The primary processor on the Main Board controls the UI Board. When the primary processor reads the parameter signals, it decodes the signals and routes the display information to the UI Board. The UI assembly also provides hardkey switches for the ProCare Main Board. The primary processor asserts a HIGH on the 16 outputs of the 1-of16 decoder/demultiplexer one at a time and then reads at the signal on SW_MUX. A LOW on SW_MUX indicates that the switch is asserted.

3-6

ProCare Patient Monitor Service Manual 2008381-001

Revision A

Principles of Operation: 3.3 Functional Description

3.3.3 SPO2 PWA


The ProCare monitor can be configured for use with either a Nellcor or Masimo SPO2 PWA. The SPO2 PWA provides continuous readings of oxygen saturation and pulse rate. Additional circuitry on the Main Board provides power, data communications, and isolation between SPO2 PWA and primary processor. Patient data received from the finger sensor is filtered, amplified, and analyzed on the SPO2 PWA. The information is sent to the Main Board via the optically coupled electrically isolated serial connection. The primary processor receives the data and routes it to the UI board for display. The data is also sent to the printer if specified.

3.3.4 Printer
The printer receives power from the Main Board and communicates with the primary processor. Printer presence and print head temperature is indicated by PR_TH signal to the primary processor. When a print command is sent to the printer from primary processor, the following will occur: * PR_CLK signal - transfer the data into print head * PR_DI signal - serial dot to be printed * PR_LAT signal - latch the data stream into the head * PR_ST1-6 - cause the head to print various sections * PR_M1-4 signals - control power sequentially to the two stepper motor windings Together these signal (CONTROL DATA) cause the printer to print a graphic hardcopy of the patient vital sign values and trend data. It also causes the printer to print a hardcopy of error logging and service record data. The printer has a built-in sensor to monitor the printer paper presence. When the printer is out of paper, it sends a PAPER OUT signal to the primary processor.

3.3.5 Pneumatic Valve/Manifold (PVM)


The PVM assembly consists of a pump, a deflate valve, and a dump valve. The PVM inflates/deflates the cuff during BP determinations. During normal operation the PVM is controlled by the primary processor. If a failsafe mode or overpressure condition occurs, the secondary processor provides the appropriate control signals to insure a safe condition, where the cuff vents to ambient atmosphere pressure.

3.3.6 Optical Switch


The optical switch indicates whether the temperature probe is inserted in the probe holder or not. The Main Board powers the switch.

Revsion A

ProCare Partient Monitor Service Manual 2009381-001

3-7

Principles of Operation: 3.3 Functional Description

3-8

ProCare Patient Monitor Service Manual 2008381-001

Revision A

Optical Switch

Main Board
SpO2 Probe (Optional) SpO2 Circuit Isolated DC Power Supply

BP Cuff

tubing

Pump/Valve Manifold

Tubing Control

PT1/ PT2 NIBP Data

Pneumatic Control

Temp Probe (Optional)

Temp Control/ Data

UI Board
Temp Control Interface Switches

External DC 6V Battery Control / Data RS-232 Remote Alarm

Regulated DC Power Real Time Clock

Display LEDs

Host Comm Port (rear)

Audio Control

Printer Driver

Speaker

Printer

Control/ Data

ProCare Vital Signs Monitor System Block Diagram (1 of 2) page 3-9/10

Real Time Clock

Isolated Power Supply

External DC 6Volt Battery

Regulated DC Power Supply Printer Drivers

Speaker

Audio Control Pneumatic Valve Control PT1/PT2 NIBP Data SpO 2 Interface

Host Communications

Temp Probe (optional)

Temperature Control/Data

Temperature Control

ProCare Vital Signs Monitor Main Board Diagram (2 of 2) page 3-11/12

Calibration & Maintenance

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-1

For your notes

4-2

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.1 Introduction

4.1 Introduction
This section contains general Monitor service procedures, including alarm code interpretation, service mode operation, and periodic maintenance and battery care. Refer to Section 5 for disassembly and reassembly procedures and related component service information.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-3

Calibration & Maintenance: 4.2 Configuring Your ProCare Monitor

4.2 Configuring Your ProCare Monitor


4.2.1 Unpacking and Preparation for Installation
1.Unpack and identify the contents of all shipping materials. 2.Remove the ProCare monitor. 3.Unpack the external DC Power Input. 4.Insert the external DC Power Input connector into the DC input on the lower right side of the monitor. 5.Plug the DC Power Input transformer into a Hospital Grounded AC receptacle. The word CHARGING will illuminate green on the front of the monitor indicating that an external power source is available. Prior to usage, it is necessary to charge the monitor for 12 hours.

4.2.2 Set the Date and the Clock


Setting the Date and Time To set the date and time on the ProCare Monitor, you must first access the configuration mode. The following table illustrates the menus and the corresponding LCD display graphics.

Setting SpO2 Averaging SpO2 Temp Unit of Measure Year Month Day Hour Minute Procedures

Window Pulse Rate and SpO2

Text in Window

Temperature Systolic MAP/Cuff Diastolic min min

1. With the Monitor off, press and hold the Menu button at the same time as pressing the On/Off button for 3 seconds.

4-4

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.2 Configuring Your ProCare Monitor 2. The Monitor automatically switches on in configuration mode. 3. To set the date and time, press the Menu button to move from one setting to another. (You will need to press the Menu button several times until the year setting appears. Use the +/- buttons to increment or decrement the individual settings.) Once you are finished changing a setting, press the Menu button again to move to the next setting. Note: For the date and time to be saved, you must advance the menu through the min setting. 4. To exit config mode, press the On/Off button.

4.2.3 Parameter Level Functional Testing


After the initial configuration is complete, perform functional testing of each of the parameters, using the accessories supplied with the ProCare Monitor. Refer to the ProCare Operator's Manual for more detailed parameterspecific instructions. Perform a blood pressure by connecting the supplied hose and cuff together, then attaching to the front of the ProCare Monitor. Press the Inflate/Stop on the front to begin the NIBP cycle. Connect the supplied temperature probe to its corresponding connection. A predictive temperature will begin once the probe is removed from its holster. Replace the probe after completion of the Temp cycle. The SpO2 sensor is an assembly consisting of two parts: the DS-100A, and the extender cable DOC-10. Connect the cables prior to attaching to the monitor. An SpO2 reading will be displayed within moments of attaching the sensor to either a Nellcor simulator or to your finger.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-5

Calibration & Maintenance: 4.3 Periodic Maintenance

4.3 Periodic Maintenance


4.3.1 As Required
Perform the following maintenance procedures as required.

4.3.1.1 Integrity of Hoses and Cuffs


When the pneumatic integrity of any NIBP cuff and hose is in doubt, replace the cuff and hose, and discard the questionable accessories.

4.3.1.2 Cleaning of Monitor


CAUTION: Do not clean Monitor with isopropyl alcohol or other solvents. Wipe the exterior of the Monitor with a cloth slightly dampened with mild detergent or normal hospital bactericides. Use dishwashing detergents such as IVORY and JOY (registered trademarks of Procter & Gamble Corp.), or PALMOLIVE (registered trademark of Colgate-Palmolive Corp.) Do not immerse unit.

4.3.1.3 Cleaning of Accessories


Clean the adult cuffs supplied for use with the monitor by hand washing in warm, soapy water. However, take care to avoid entry of water into the cuff and hoses at any time. If water enters the cuff, dry the cuff by passing air through it. The neonatal cuffs are for single patient use - discard if they become soiled. Clean cuffs and hoses with a cloth slightly dampened with mild detergent. Do not immerse hoses. Do not immerse cuffs without prior application of cuff hose caps. Clean SpO2 sensor surface before and after each patient use. Clean SpO2 sensor with a cloth slightly dampened with a mild detergent. Wipe SpO2 sensor to ensure all detergent residue has been removed. Compatible cleaning and disinfecting solutions are: Chlorine bleach disinfectant, 5.25%, 0.75 cup per gallon of water. CAUTION: Do not apply isopropyl alcohol to the Monitor - some parts can become marred and cracked.

4-6

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.3 Periodic Maintenance Cidex Formula 7 (registered trademark of Johnson & Johnson Medical Products, Inc.) or pHisoHex (registered trademark of Winthrop-Breon Laboratories.) Quaternary-based germicidal detergents like VESTAL INSURANCE (registered trademark of the Vestal Corp.), HI-TOR PLUS (registered trademark of the Huntington Corp.), or VIREX (registered trademark of S.C. Johnson & Son Corp.) For the above, follow manufacturers' recommendations for dilution rate and use. These recommendations are not an endorsement of the manufacturers or of the effectiveness of these materials for cleaning or disinfecting.

4.3.1.4 Long-Term Storage


If it becomes necessary to store the Monitor for an extended period of time, remove all attached accessories. Attach the original packing inserts, and place the monitor into the original shipping container. Long-term storage of a discharged lead-acid battery can permanently degrade its storage capacity. Therefore, the battery should be fully charged before storage. Batteries in storage should be charged every six months to maintain their capacity. Elevated temperatures will shorten battery life and can lead to permanent battery damage.

4.3.2 Annual Procedures


Perform the test procedures described in section 4.6 every twelve months, or whenever the accuracy of of the monitor is in doubt.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-7

Calibration & Maintenance: 4.4 Care of the Storage Battery

4.4 Care of the Storage Battery


The Monitor uses one Lead-Acid storage battery. The battery can be charged at any time without reducing the charging capacity.

4.4.1 Battery Charging


The Monitor charges the Lead-Acid battery whenever the AC power supply is in use. The Monitor automatically senses if the battery needs recharging. Battery charging will continue as long as the Monitor is connected to the AC power supply, even when the Monitor is turned off. Batteries should be charged before first use or after long periods of storage. The battery should be charged for 8 hours before use, as a charged battery loses some charge when left in storage. It is possible to use the Monitor while the Monitor is charging. The battery should be charged at room temperature (59 F - 86 F; 16 C 30 C). It is normal for the battery to become warm during charging or after use. Batteries can be charged or topped-off at any time. It is not necessary to wait until they are fully discharged. If the monitor is idle for extended periods, it should be fully charged once a month to ensure optimum performance.

4-8

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.5 Safety Testing

4.5 Safety Testing


To adequately test the safety and integrity of the ProCare Monitor, the following test equipment is recommended:

12VDC Power Supply DMM (Fluke 8842 or equivalent) NIBP Analyzer (DNI Nevada "Cufflink" or Equivalent) Adult BP Cuff, Neonate BP Cuff, hose, inflation bulb, and mandrel Inflation bulb and associated tubing Calibration Kit, p/n 320246 available through GE Medical Systems SPO2 Simulator (for appropriate SPO2 type if SPO2 is installed) SPO2 Cable (for appropriate SPO2 type, if SPO2 is installed) TE 1811 Temperature Probe Simulator (if TEMP is installed.) The
Temperature Simulator for the Alaris System is available from Alaris Medical Systems, Inc. (619) 458-7000

Printer Paper (if PRINTER is installed)


CAUTION! Calibration equipment should always be kept dry and free of particulate matter. Moisture or foreign substances introduced to the pneumatic system will likely cause damage to the monitor and/or the accessories. Complete the Test Record (Appendix A) as tests are performed. Note: This test is written so that a knowledgeable technician who is familiar with the ProCare monitor and the test equipment and will be able to follow the test procedure. Note: To enter service mode press and hold the CYCLE buttons while pressing the ON/OFF button for 3 seconds.

Notes on Electrical Safety Testing of the ProCare Monitor:


The DINAMAP ProCare Monitor is designed and tested to meet electrical safety standard IEC 601-1. Requirements in this standard parallel requirements in NFPA99 relating to electrical safety. This product meets the requirements of NFPA99 section 7-5.1.2.2. (grounding of appliances) through the use of double insulation in the external power supply module (Power Brick.) The absence of a ground connection on the power brick obviates the need for any ground resistance test, such as that described in NFPA99 section 75.1.3.2. (Some non-USA Power bricks are equipped with a non-conductive

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-9

Calibration & Maintenance: 4.5 Safety Testing pin at the ground pin location to aid mechanical retention of the Power Brick into the receptacle.) Dielectric strength (Hi-Pot) testing of the double (reinforced) mains insulation (4000 V RMS, 60 Hz) may be done by disconnecting the Power Brick from the Monitor and applying the Hi-Pot tester as follows: Connect one lead to the two mains power prongs. Connect the other lead to the conductive sleeve on the DC output connector at the end of the cord. Hi-Pot testing of the patient connections (1500 VRMS, 60 Hz) may be done by connecting one lead of the Hi-Pot tester to the outer shell of a spare DC connector plug inserted into the DC input connector on the Monitor. The other lead of the tester should be connected to the patient connection being tested. WARNING: Dont electrocute yourself or anyone else when performing HiPot testing. NOTE: Hi-Pot testing is done on every unit at the factory and should not be repeated unnecessarily nor performed more often than required.

4.5.1 Temp Circuit Leakage Test


1. Setup an IEC 601-1 approved leakage tester to apply 240 VAC to an isolated circuit. 2. Plug temp probe Hi Pot adapter into the temp jack. 3. Record and verify the temp circuit leakage current.

4.5.2 SpO2 Circuit Leakage Test


1. Setup an IEC 601-1 approved leakage tester to apply 240 VAC to an isolated circuit. 2. Plug an SpO2 cable into the SpO2 connector on the front of the unit. 3. Plug SpO2 probe Hi Pot adapter in to the DB9 jack at the end of the SpO2 cable. 4. Record and verify the SpO2 circuit leakage current.

4-10

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests

4.6 ProCare Patient Monitor Parameter Tests


Complete the Test Record (Appendix A) as tests are performed. Note: This test is written so that a technician who is familiar with the ProCare monitor and the test equipment will be able to successfully complete the test procedures. To enter service mode press and hold the CYCLE button while pressing the ON/OFF button for 3 seconds. Saving Changes into Flash Memory: Modifications made to the Monitors settings/configuration must be saved prior to placing the Monitor into service. The Monitor must be in service mode for any changes to be saved. Events that require you to manually save to the Monitors memory include calibration attempts and enabling/disabling the Monitors parameters (refer to section 4.6.4.) It is not necessary to manually save following a pressure transducer calibration verification or simple calibration check. The number of times the flash memory can be saved is limited. The fatal alarm error 975 may be issued after this process is performed more than 90 times. Following issuance of this fatal alarm, the monitor must be returned to the GE Medical Systems Information Technologies Service Center to have the flash memory reset so that future saves can be performed.

4.6.1 SETUP
1. Connect your NIBP Analyzer to the ProCare Monitor. 2. T an inflation bulb into the pneumatic setup 3. Consult the following diagram for pneumatic setup guidelines.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-11

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests

4.6.2 Leakage Testing


Note: To enter service mode press and hold the CYCLE button while pressing the ON/OFF button 1. Turn the Monitor ON and enter Service Mode. 2. Press CYCLE button and the min display should change to a 1 3. Close the valve on the inflation bulb 4. Set the NIBP Analyzer to "Leak test" function 5. Press the "Pump" button on the NIBP Analyzer 6. After the system inflates to 200mmHg, wait for the "Wait" on the NIBP analyzer message to be cleared 7. Press the "Start" button on the NIBP Analyzer to begin the leakage test 8. After 60 seconds the NIBP Analyzer will display the leakage rate 9. Record and verify the leakage rate. 10. Turn the ProCare Monitor off

4.6.3 Pressure Transducer Verification


Note: To enter service mode press and hold the CYCLE button while pressing the ON/OFF button 1. Turn the Monitor ON and enter Service Mode. 2. The HISTORY LED should display 0. 3. Set NIBP Analyzer to "Manometer" function and press "Zero Pressure" 4. Press CYCLE button, HISTORY display should change to a 1. 5. Use the inflation bulb to inflate the cuff, hose and pressure indicator setup to 200mmHg 6. Record and verify the pressure reading on the top LED display (SYSTOLIC) 7. Record and verify the pressure reading on the bottom LED display (DIASTOLIC) 8. Use the valve on the bulb to reduce pressure to 150mmHg 9. Record and verify the pressure reading on the top LED display (SYSTOLIC) 10. Record and verify the pressure reading on the bottom LED display (DIASTOLIC) 11. Use the valve on the bulb to reduce pressure to 100mmHg
4-12 ProCare Patient Monitor Service Manual 2009381-001 Revision A

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests 12. Record and verify the pressure reading on the top LED display (SYSTOLIC) 13. Record and verify the pressure reading on the bottom LED display (DIASTOLIC) 14. Use the valve on the bulb to reduce pressure to 50mmHg 15. Record and verify the pressure reading on the top LED display (SYSTOLIC) 16. Record and verify the pressure reading on the bottom LED display (DIASTOLIC) If any of the tests fail, continue to section 4.6.4. Otherwise, continue to section 4.6.5

4.6.4 Pressure Transducer Calibration


Perform only if Pressure Transducer Verification is out of tolerance as specified in Appendix A. Note: To enter service mode press and hold the CYCLE button while pressing the ON/OFF button. 1. Turn the ProCare Monitor ON and enter Service Mode. 2. The min window should display 0. 3. Open valve on bulb to open pressure system to atmosphere 4. Set NIBP Analyzer to "Manometer" function and press "Zero Pressure" 5. Press CYCLE button until the min window shows 1. 6. Close valve on bulb and manually inflate pressure to 200 mmHg (using the CuffLink manometer as reference). 7. Press MENU button when pressure reads 200 mmHg to save calibration setting. 8. Two steps are necessary for burning changes into flash: a. Press the CYCLE button repeatedly until the number 6 is displayed in the min window. Note: the number of remaining changes to the flash memory available are displayed in the MAP/Cuff window. Following each true calibration, the Monitor decrements by 1 the number of remaining calibrations available. If the MAP/Cuff window is displaying a figure of less than 10 and the accuracy of the Monitor is in doubt, contact General Electric Medical Systems Information Technologies Technical Support at 1-877-274-8456. b. Press and hold the MENU button until two beeps are heard (one when

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-13

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests the button is pressed and another after the button has been depressed long enough for the process to complete.) 9. Turn the ProCare Monitor off

4.6.5 Overpressure Verification


Note: To enter service mode press and hold the CYCLE button while pressing the ON/OFF button. 1. Remain in service mode. 2. Use the inflation bulb to inflate pressure until valve opens. 3. Record and verify pressure at which valve opens. 4. Press CYCLE button so that the min window changes to a 2. 5. Use the inflation bulb to inflate pressure until valve opens. 6. Record and verify pressure at which valve opens. 7. Turn unit off.

4.6.6 Button Testing


1. Turn the Monitor on. 2. Record software revision as shown in the systolic and diastolic display. 3. Press START/STOP button. 4. Verify an NIBP determination has been initiated. 5. Block pump port and verify "E80" alarm. 6. Press SILENCE button, verify alarm has been silenced. 7. Verify flashing red indicator of Silence. 8. Press the SILENCE button, verify alarm condition is removed 9. Press ALARM button several times, verify unit cycles through all alarm settings (i.e SYS, DIA, SPO2). 10. Turn the Monitor off

4-14

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests

4.6.7 LED Tests


1. Power on the ProCare Monitor. 2. During the power-up self-test verify all 7 segment LED display segments and all discrete LEDs (except CHARGING LEDs) illuminate and are the correct color. LED COLOR MATRIX RED Systolic Diastolic F C HIGH (all) LOW (all) Alarm Bell Icon SpO2 Strength Meter 3. Repeat power up cycle until all LEDs are checked. YELLOW Pulse Rate LOW GREEN AUTO CYCLE HISTORY BATTERY BATTERY ICON ALARM VOLUME PULSE VOLUME INFLATE PRESSURE NOT LIT CHARGING

4.6.8 External DC Verification


1. Plug the DC power cable into the monitor. 2. Verify that the CHARGING indicator is illuminated.

4.6.9 NIBP Determination


1. Set NIBP Analyzer to Adult Mode: SYS/DIA = 120/80 MAP = 90 BPM = 80 2. Press START/STOP button on the Monitor to begin determination 3. Record and verify systolic, diastolic, map and heart rate from the monitor display 4. Press CYCLE button to initiate a determination in "Auto BP" mode.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-15

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests 5. Record and verify systolic, diastolic, map and heart rate from the monitor display 6. Press CYCLE button until is displayed on the HISTORY LED to initiate a determination in STAT mode. 7. Record and verify systolic, diastolic, map and heart rate from the monitor display 8. Press Start/Stop button, end STAT mode.

4.6.10 NIBP Overpressure Verification


1. Restrict airflow through cuff hose port. 2. Press "Inflate/Stop" to begin NIBP determination 3. Verify that alarm sounds is displayed on the SYSTOLIC display and an audible

4. Remove the air restriction 5. Press "Inflate/Stop" and verify that the pump does not start 6. Press the "Silence" button 7. Press the "Silence" button again 8. Verify the alarm condition is cleared from the SYSTOLIC display

4.6.11 Temperature (Perform if equipped with Temp module)


The Temperature Simulator for the Alaris System is available from Alaris Medical Systems, Inc. (619) 458-7000. 1. Disconnect the temp probe from the Monitor. 2. Connect the probe simulator to the Monitor. 4. Set the probe simulator to 80.2 F. 5. Record and verify the reading in the TEMP display is 80.2 F 0.2 F. 6. Set the probe simulator to 102 F. 7. Record and verify the reading in the TEMP display is 102 F 0.2 F. 8. Set the probe simulator to B.P. and verify reading is 106.0 F 0.2 F. Press broken probe button down and verify the Monitor displays .

4-16

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests 9. Calibration verification is complete. Disconnect the probe simulator and install the temperature probe. If the Monitor does not pass the calibration verification, then the Monitor needs repair.

4.6.12 SpO2 (Perform only if equipped with SpO2 module)


Note for Monitors equipped with Nellcor SpO2: Nellcor is aware that various Nellcor oximetry platforms have some interesting behaviors when used with various pulse simulators. On occasion when testing the integrity of the Nellcor oximetry system, abnormal results may occur when introducing large changes in the pulse rate and/or pulse amplitude. Extreme changes in rate sent to the Nellcor sensor by the SpO2 simulator may cause the SpO2 algorithm to completely miss in finding the pulse rate. This is an expected result. To workaround this, incrementally step up or down the settings on your SpO2 simulator and allow for the Monitor to detect and display the new pulse rate or saturation. Nellcor recommends use of the SRCmax Portable Tester for use with ProCare Monitors equipped with the Nellcor SpO2 system. Masimo recommends BIO-TEK SpO2 simulators. 1. Connect the appropriate SpO2 simulator and cable to the SpO2 connector. 2. Verify the unit displays a: Pulse value Saturation value Signal Strength Bar Graph 3. Disconnect the SpO2 cable 4. Verify the unit generates a 5. Press the SILENCE button. 6. Verify the sound has stopped but the error display remains 7. Re-connect the SpO2 sensor 8. Verify the unit displays a: Pulse Value Saturation value Signal Strength bar Graph alarm and speaker is sounding

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-17

Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests

4.6.13 Printer Output Test


1. Load Thermal Paper into the print mechanism 2. Press PRINT button 3. Verify the printer outputs a record and print quality is good

4.6.14 Communication Port Test


1. Connect unit to a PC terminal emulator. 2. Turn unit on. 3. Type " NC0!E", press "Enter" 4. Verify response from unit " NC+!@" 5. Verify that within 40 seconds the pump starts. 6. Type " ND!5", press "Enter" 7. Verify response from unit " ND+!A" 8. Verify the pump stops.

4-18

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: 4.7 Alarm Code Interpretation

4.7 Alarm Code Interpretation


If any other alarms appear that are not listed in the paragraphs that follow, record the error message and report the failure to Customer Support. Refer to the Operation Manual for information about patient alarms and general procedural alarms.

4.7.1 System Failures


When a system failure is encountered, the error code is displayed on the screen for five seconds and the system enters failsafe mode. The error code is logged in the history log. General system error codes are listed below.

Alarm Conditions and Error Codes


When responding to a Monitor alarm, always CHECK THE PATIENT FIRST and then check the Monitor, cuff, hose and sensors. Press SILENCE to reset patient alarm conditions.

NIBP
Alarm E89 E84 E85 E83 E82 E80 Definition NIBP No Determination Timeout Timeout Timeout: Inflation Excess Air in Cuff Overpressure Possible Cause Unable to make an NIBP determination due to insufficient signal. Determination time > 2 minutes. Motion Artifact. One cuff pressure at > 1 minute. Motion artifact. Inflation time is > 40 seconds Possible air leak is being detected. Residual air in cuff above threshold for successful auto-zero. Overpressure condition detected

SpO2
Alarm E20 Definition SpO2 No Sensor Possible Cause SpO2 sensor not connected. No sensor code detected. Sensor failure SpO2 sensor or cable possibly defective. Cable not connected properly.

E21

SpO2 Replace Sensor

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-19

Calibration & Maintenance: 4.7 Alarm Code Interpretation E23 E25 SpO2 Sensor off finger SpO2 No Signal No SpO2 signal, check or reposition the sensor No or very low SpO2 signal, reposition sensor

Temperature / Printer / Miscellaneous


Alarm E63 E61 E66 E10 Definition Temp Disconnected or Wrong Probe Type Temp Probe Broken Temp probe Too Hot Printer No Paper Possible Cause Incorrect type of temperature probe: use TurboTemp-type temperature probe Bad temperature probe Temp probe not properly connected Bad temperature probe Verify monitor with known good accessories Printer is out of paper Printer door open Printer may be defective

E11 E12 E00 900999

Printer Too Hot Recorder cannot print Memory Lost Internal Memory Errors Main battery voltage is too low to operate Refer Monitor to Customer Service Refer Monitor to Customer Service

4-20

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix A - Test Results Form

Appendix A - Test Results Form


Step 4.6.2 4.6.3 Leakage Leakage Result (mmHg) Pressure Transducer Verification Pressure reading at 200mmHg, top display - Systolic Pressure reading at 200mmHg, bottom display - Diastolic Pressure reading at 150mmHg, top display - Systolic Pressure reading at 150mmHg, bottom display - Diastolic Pressure reading at 100mmHg, top display - Systolic Pressure reading at 100mmHg, bottom display - Diastolic Pressure reading at 50mmHg, top display - Systolic Pressure reading at 50mmHg, bottom display - Diastolic 4.6.5 Overpressure Verification Overpressure threshold, Adult (mmHg) Overpressure threshold, Neonate (mmHg) 4.6.6 Buttons Software revision in systolic display (top) Software revision in systolic display (bottom) NIBP alarm initiated "E80" displayed on SYSTOLIC display Audible alarm can be silenced "Silenced" LEDs flash Overpressure alarm can be cleared Alarm button is functioning 4.6.7 Display All 7-Segment LEDs Light Correct Color All Discrete LEDs Light, Correct Color 4.6.8 4.6.9 4.6.8 External DC Detection Charging indicator LED illuminated NIBP Determination Systolic reading (mmHg) 107 133 on off 305 155 325 159 197 197 147 147 97 97 47 47 203 203 153 153 103 103 53 53 0 6 Description Min Max Actual Pass-Fail-N/A

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-21

Calibration & Maintenance: Appendix A - Test Results Form

Step 4.6.9

Description NIBP Determination (continued) Diastolic reading (mmHg) MAP reading (mmHg) Heart rate reading (bpm) Systolic reading (mmHg) Diastolic reading (mmHg) MAP reading (mmHg) Heart rate reading (bpm) Systolic reading (mmHg) Diastolic reading (mmHg) MAP reading (mmHg) Heart rate reading (bpm)

Min 67 85 76 107 67 85 76 107 67 85 76

Max 93 95 84 133 93 95 84 133 93 95 84

Actual

Pass-Fail-N/A

4.6.10

NIBP Overpressure "E80" displayed on SYSTOLIC display Pump will not start Overpressure alarm can be cleared

4.6.11

Temperature Test Temperature reading at 80.2 F Temperature reading at 102.0 F Temperature reading at BP verify reads E61

80.0F 79.9 F 101.8 F N/A

80.4F 80.5 F 102.2 F N/A

4.6.12

SpO2 Pulse Value Displayed Saturation Value Displayed Signal Strength Bar Graph Displayed "E23" displayed on SpO2 display Alarm is silenced, error display remains Pulse Value Saturation Value Displayed Signal Strength Bar Graph Displayed

4.6.13 4.6.14

Printer Test Printout is generated cleanly Communication Port Test "_NC+!@" is displayed on terminal Pump stops

4-22

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix B - Connectivity

Appendix B - Connectivity
CHANT

Provides HL7 output for electronic Patient Medical Records Uses ILCs (Isolated Level Convertors) where required along with the
CHANT software to communicate with the HIS System Compatible Monitors: DINAMAP XL, Compact, MPS (Select and Portable), PRO 100-400 series, ProCare Series, and PRO 1000. Nurse Call System

When the DINAMAP monitor alarms, the Nurse Call System is triggered. Uses a Nurse Call Cable to attach to the System. P/N 487208CR.
Compatible Monitors: Compact, PRO 100-400 Series, ProCare Series, and PRO 1000. Alarm View

Wireless transmitter attached to the Monitor. Sends alarms and vital signs results to a pager
Compatible Monitors: Compact, MPS (Select and Portable) PRO 100-400 series, ProCare Series, and PRO 1000. CIC

Provides Central Alarm Notification & Data Management. Compatible with StatView for paging.
Compatible Monitors: PRO 1000 Version 2 and Dash 2000 (both hardwired) ApexPro

Provides Central Alarm Notification and data management. Transmits data wirelessly. Pro 100-400 Series and ProCare (requires P/N IPC-1931)
PatientNet

Provides central alarm notification and data management. Lethal arrhythmia at the Central Station. WMTS frequency hopping spread spectrum.
Note: Dash 2000 is incompatible with PatientNet. ILC-1926

Interfaces the PRO 100-400 series or ProCare monitor to ANY hardwired


connection.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-23

Calibration & Maintenance: Appendix B - Connectivity Requires Cables: -PRO 100-400, ProCare Series, Compact, Plus: cable p/n 683235 -XL monitor series: p/n 682234 ILC-1927

Interfaces the PRO 1000 monitor to any hardwire connection. Installs into the Communications bay at the rear of the monitor.
IPC1928

Interfaces the PRO 1000 Monitor to the Unity network, hardwire. Installs into the Communications bay at the rear of the monitor.
IPC1931 Connects DINAMAP Monitors to the CIC system. Transfers vital signs results from the Monitor to the CIC Requirements for use: - Plug DINAMAP into ApexPro using the DINALINK cable.

4-24

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix B - Connectivity - Apex version 2 - CIC version 3

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-25

Calibration & Maintenance: Appendix C - Display Cover: Removal, Installation

Appendix C - Display Cover: Removal, Installation


To remove the plastic display cover: 1. Place the Monitor on a stable surface. 2. Insert a flat-head screwdriver between the top of the display cover and the body of the monitor.

3. Twist the screwdriver so that the display cover is slightly separated from the body of the monitor. 4. Using two hands, grasp the top-middle edge of the display cover and pull away and down.

4-26

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix C - Display Cover: Removal, Installation 5. The fascia and displays are now visible.

Installing the Display Cover: 1. Place the Monitor on a flat, stable surface. 2. Insert the tabs at the base of the display cover into the slots underneath the bottom row of LEDs. 3. Using two hands, apply firm pressure to the outside edges, while flexing the upper middle-edge of the display cover away from the monitor. It is imperative that even pressure is applied as the plastic retention tabs can break free from the display cover.

4. Apply steady, even pressure to the opposite sides of the display cover until an audible click is heard.

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-27

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

Appendix D: Replacement Parts and Assemblies


ID numbers reference the explode drawings in Section 5.

ID
1 2 3 4 5 6

FRU Number
2009910-001 2009911-001 2009912-001 2009913-001 2009914-001 2009915-001

Description
Main PWA Main PWA- Ausc UI PWA Nellcor Module Masimo Module Plastics kit

Content/comments
Supplied with Arterial ref product s/w Supplied with Auscultatory version of Monitor

Supplied with mounting spacers. Supplied with mounting spacers. 1. 2. 3. 4. 5. 6. 7. 8. 9. Case, front & rear Handle front & rear with h/w Bulkheads 4 types & NIBP insert. Nellcor & Masimo inside labels Std. labels, DC input, No temp side label. Battery door & foam. Help card guide and rubber feet. Printer door (and blanking plate when available). Cable tie hardware

2009916-001 7 2009918-001

Screw kit Temperature

All internal hardware, screws, washers etc 1. Housing front & rear. 2. Mounting bracket and plastic guide. 3. Sensor and cable. 4. Mounting hardware 1. Printer 2. Printer chassis and screw. 3. Printer door with roller and label. Door with label and roller Pump assembly Dump valve and adhesive tape to apply. Set, left and right keypads Include mounting bracket With mounting h/w UK plug 240V input Euro plug 230V input US plug 120V input 1. keypad labels set (LH & RH) 2. Fascia 3. Display cover For language variants reference following pages All 3 guides and for each language

2009917-001

Printer

9 10 11 12 13 14

2009919-001 2009920-001 2009921-001 2011645-001 2009922-001 2009923-001 2008538-001 2008539-001 2009460-001

Printer door Pneumatics Dump valve Kit, Keypads Speaker Host comms cable Power brick UK Power brick EUR Power brick US English 400 lbl kit English 300 lbl kit English 200 lbl kit English 100 lbl kit Help card guides

15

2010429-001 2010478-001 2010479-001 2010480-001

16

4-28

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

2010429-001

Label Kit

English 400 printer

P/N (-001) 2010429-001 2010478-001 2010479-001 2010480-001 2010481-001 2010482-001 2010483-001 2010484-001 2010485-001 2010486-001 2010487-001 2010488-001 2010489-001 2010490-001 2010491-001 2010492-001 2010493-001 2010494-001 2010495-001 2010496-001 2010497-001 2010498-001 2010499-001 2010500-001 2010501-001 2010502-001

Language English English English English English English English English English English English English English English English English French French French French French French French French French French

Model 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300

Is Printer Installed? Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed

Method of NIBP Determination (Auscultatory is labeled on left keypad) Traditional Traditional Traditional Traditional Auscultatory Auscultatory Auscultatory Auscultatory Traditional Traditional Traditional Traditional Auscultatory Auscultatory Auscultatory Auscultatory

Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed

Traditional Traditional Traditional Traditional Auscultatory Auscultatory Auscultatory Auscultatory Traditional Traditional

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-29

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

2010503-001 2010504-001 P/N (-001) 2010505-001 2010506-001 2010507-001 2010508-001 2010509-001 2010510-001 2010511-001 2010512-001 2010513-001 2010514-001 2010515-001 2010516-001 2010517-001 2010518-001 2010519-001 2010520-001 2010521-001 2010522-001 2010523-001 2010524-001 2010525-001 2010526-001 2010527-001 2010528-001 2010529-001 2010530-001 2010531-001

French French Language French French French French German German German German German German German German German German German German German German German German Spanish Spanish Spanish Spanish Spanish Spanish Spanish

200 100 Model 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Is Printer Installed?

Traditional Traditional Method of NIBP Determination (Auscultatory is labeled on left keypad) Auscultatory Auscultatory Auscultatory Auscultatory Traditional Traditional Traditional Traditional Auscultatory Auscultatory Auscultatory Auscultatory Traditional Traditional Traditional Traditional Auscultatory Auscultatory Auscultatory Auscultatory

4-30

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

2010532-001 P/N (-001) 2010534-001 2010535-001 2010536-001 2010537-001 2010538-001 2010539-001 2010540-001 2010541-001 2010542-001 2010543-001 2010544-001 2010545-001 2010546-001 2010547-001 2010552-001 2010553-001 2010554-001 2010555-001 2010556-001 2010557-001 2010588-001 2010559-001 2010560-001 2010561-001 2010578-001 2010589-001 2010590-001 2010592-001 2010593-001

Spanish Language Italian Italian Italian Italian Italian Italian Italian Italian Swedish Swedish Swedish Swedish Swedish Swedish Swedish Swedish Dutch Dutch Dutch Dutch Dutch Dutch Dutch Dutch Chinese Chinese Chinese Chinese Chinese

100 Model 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Printer Installed Is Printer Installed? Printer Installed Printer Installed Printer Installed Printer Installed Method of NIBP Determination (Auscultatory is labeled on left keypad)

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-31

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

2010594-001 P/N (-001) 2010595-001 2010596-001 2010597-001 2010598-001 2010599-001 2010600-001 2010601-001 2010602-001 2010603-001 2010604-001 2010605-001 2010606-001 2010607-001 2010608-001 2010609-001 2010610-001 2010611-001 2010612-001 2010613-001 2010614-001 2010615-001 2010616-001 2010617-001 2010618-001 2010619-001 2010620-001 2010621-001 2010622-001 2010623-001
4-32

Chinese Language Chinese Chinese Japanese Japanese Japanese Japanese Japanese Japanese Japanese Japanese Danish Danish Danish Danish Danish Danish Danish Danish Norwegian Norwegian Norwegian Norwegian Norwegian Norwegian Norwegian Norwegian Korean Korean Korean

300 Model 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 Printer Installed Printer Installed Printer Installed
Revision A

Is Printer Installed?

Method of NIBP Determination (Auscultatory is labeled on left keypad)

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

ProCare Patient Monitor Service Manual 2009381-001

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

P/N (-001) 2010624-001 2010625-001 2010626-001 2010627-001 2010628-001 2010629-001 2010630-001 2010631-001 2010632-001 2010633-001 2010634-001 2010635-001 2010636-001 2010638-001 2010639-001 2010640-001 2010641-001 2010642-001 2010643-001 2010644-001 2010645-001 2010646-001 2010647-001 2010648-001 2010649-001 2010650-001 2010651-001 2010652-001 2010653-001

Language Korean Korean Korean Korean Korean Finnish Finnish Finnish Finnish Finnish Finnish Finnish Finnish Portuguese Portuguese Portuguese Portuguese Portuguese Portuguese Portuguese Portuguese Russian Russian Russian Russian Russian Russian Russian Russian

Model 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100

Is Printer Installed? Printer Installed

Method of NIBP Determination (Auscultatory is labeled on left keypad)

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-33

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

P/N (-001) 2010656-001 2010657-001 2010658-001 2010659-001 2010660-001 2010661-001 2010662-001 2010663-001 2010664-001 2010665-001 2010666-001 2010667-001 2010668-001 2010669-001 2010670-001 2010671-001 2010672-001 2010673-001 2010674-001 2010675-001 2010676-001 2010677-001 2010678-001 2010679-001 2010680-001 2010681-001 2010682-001 2010683-001 2010684-001

Language Hungarian Hungarian Hungarian Hungarian Hungarian Hungarian Hungarian Hungarian Czech Czech Czech Czech Czech Czech Czech Czech Slovak Slovak Slovak Slovak Slovak Slovak Slovak Slovak Polish Polish Polish Polish Polish

Model 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400 300 200 100 400

Is Printer Installed? Printer Installed Printer Installed Printer Installed Printer Installed

Method of NIBP Determination (Auscultatory is labeled on left keypad)

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

Printer Installed Printer Installed Printer Installed Printer Installed

4-34

ProCare Patient Monitor Service Manual 2009381-001

Revision A

Calibration & Maintenance: Appendix D: Replacement Parts and Assemblies

P/N (-001) 2010685-001 2010686-001 2010687-001 2010688-001 2010689-001 2010690-001 2010691-001 2010692-001 2010693-001 2010694-001 2010695-001

Language Polish Polish Polish Greek Greek Greek Greek Greek Greek Greek Greek

Model 300 200 100 400 300 200 100 400 300 200 100

Is Printer Installed?

Method of NIBP Determination (Auscultatory is labeled on left keypad)

Printer Installed Printer Installed Printer Installed Printer Installed

Revision A

ProCare Patient Monitor Service Manual 2009381-001

4-35

Assembly Drawings & ProCare Schematics: Appendix D: Replacement Parts and Assemblies

Assembly Drawings & ProCare Schematics

4-36

ProCare Patient Monitor Service Manual 2009381-001

Revision A

or

see Appendix C in Section 4 for Display Cover removal instructions.

or

Exploded Assembly Drawing ProCare Monitor (1 of 2) page 5-1/2

or

PNEUMATIC TUBING ROUTING

Exploded Assembly Drawing ProCare Monitor (front case) page 5-3/4

MAIN PROCESSOR
P5A C47 C50
25V 100nF

P3V3 C45 C48


100nF 25V

100nF 25V

C44 C46
100nF 25V

100nF 25V

P3V3 P3V3 R1 R117 100K 100K


R104

100nF 25V

13 VDD1

22 VDD2

55 VDD3

91 VDD4

97 VDD5

PR_TH I_CHG V_DC_MON PUMP_I PT1 PT2 FPT V_BAT_MON_PRI

R84 10K0

78 77 76 75 74 73 72 71

1K00 D_FROM_COMM D_FROM_SPO2 D_TO_COMM D_TO_SPO2 SPI_MISO SPI_MOSI SPI_CLK T_DE T_CLK T_TDI T_TDO T_TMS T_TRST RESET_TEST-0 TP39 TP40 TP41
R119

P3V3 1K00
R108

1K00
R106

80 VRH 79 VRL 48 RXD1 46 RXD2 47 TXD1 45 TXD2

PQA0 PQA1 PQA3 PQA4 PQB0 PQB1 PQB2 PQB3

U9 MMC2107

INT0 INT1 INT2 INT3 INT4 INT5 INT6 INT7

49 50 53 56 57 58 60 61

1K00

TP34 TP33

R116 100K

PUMP_ON DUMP_ON LATCHED_OVP DUMP2_ON ADULT-0 PNEU_RESET

R206 75R0 TP26

R201 75R0

63 MISO 62 MOSI 65 SCK 66 SS 99 92 93 94 95 98 DE TCLK TDI TDO TMS TRST

7 PA0 6 PA1 5 PA2 4 PA3 3 PA4 2 PA5 1 PA6 100 PA7 PB0 PB1 PB2 PB3 PB4 PB5 PB6 PB7 PC0 PC1 PC2 PC3 PC4 PC5 PC6 PC7 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 17 16 15 14 11 10 9 8

PR_DI PR_CLK PR_ST1 PR_ST2 PR_ST3 PR_ST4 PR_ST5 PR_ST6 PR_VHON PR_M4 PR_M2 PR_M3 PR_M1 PR_EOP_EN PR_LAT PR_EOP 1K00
R113

Place near U9 R103 0R0 R90 R120 R115100K 100K 1K00

C32 10P C33 10P XL2 R62 10M0

8.0MHZ

R63 1K00

RESET_PRI-0 P3V3

ADC_PWM CS_DISPLAY PWM_VOL PWM_PITCH P5D

90 CLKOUT 87 EXTAL 86 XTAL 83 RESET 84 RSTOUT 43 ICOC10 42 ICOC11 41 ICOC12 40 ICOC13 39 ICOC20 38 ICOC21 37 ICOC22 36 ICOC23 59 70 82 81 52 51 85 88 64 44 VPP VDDH VDDA VSSA VDDF VSSF VDDSYN VSSSYN VSTBY TEST

1K00
27 26 25 24 23 20 19 18 35 34 33 32 31 30 29 28 R114

SHDN_TO_PRI SW_MUX RTC_DATA RTC_CLK SPO2_RST-0 IPS_ON CS_SEC-0 FSP_FROM_PRI FAST_OVP-0 DUMP_ON_MON D_FROM_TEMP D_TO_TEMP REM_ALARM-0 TEMP_PROBE_IN DUMP2_ON_MON FSP_TO_PRI R190 10K0

TP42

R112

1K00

C12 10N0

100NF 25V

C49

P5D

P3V3 C53

P3V3 C3 10U0 C54 P3V3SEC

69 PE5 68 PE6 67 PE7 VSS1 VSS2 VSS3 VSS4 VSS5

R107 TP30 R98 P3V3 10K0 PE7 needs to be high at start-up R118 100K 1K00

C1 10U0

C14 10N0

100NF 25V

C52

P5A C2 10U0 C5

C15 10N0

100NF 25V

C16 10N0

100NF 25V

R95 1K00

12

21

54

89

96

100NF 25V

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (1 of 10) page 5-5/6

+3.3V SECONDARY POWER


PBAT

P3V3SEC C141
100nF 25V

RESET
16 VCC

P3V3SEC
1 VIN U22 3 ON/OFFL LP2985 4 BYPASS GND 2 VOUT 5

U27
Q14 Q13 Q12 Q10 Q9 Q8 Q7 Q6 Q5 Q4

C115 R137 10K0 R141 49K9


100nF 25V

9 OSCOUT2 10 OSCOUT1 11 OSCIN 12 RESET 8 GND

3 2 1 15 13 14 6 4 5 7

DTC143ZKA
Logic

Q14

C106 2U2 C13 10N0 16V

C105 10U0

REAL TIME CLOCK


P3V3SEC C56
25V 100nF

P3V3

74LV4060

SECONDARY PROCESSOR

P3V3SEC R37 1K00 C103 12P0 XL3 32.768KHZ


1 OSCI 2 OSCO 3 INT VSS 4 8 VDD

U21 PCF8563T
CLKOUT 7 SCL 6 SDA 5

R2 4K99

R3 4K99 RTC_CLK RTC_DATA

PT2_SEC R174 10K2 R176 20K0 P3V3SEC P3V3SEC R43 1K00


1 MCLR 20 VDD RB7 RB6 RB5 RB4 RB3 RB2 RB1 RB0/INT0 28 27 26 25 24 23 22 21

C55
100nF 25V

V_BAT_MON_SEC I_CHG P5A2 R175 10K2 R177 20K0

TP1

2 3 4 5 6

RA0/AN0 RA1/AN1 RA2/AN2 RA3/AN3/VREF RA4

FAST_CHG LATCHED_OVP FAST_OVP-0 PNEU_OVERRIDE-0 RESET_PRI-0 SW_PWR-0

C35 22P0 C34 22P0


4.0MHZ

9 OSC1

XL1

U3 PIC16LC73B
RC0 11

R35 0R0

10

OSC2 RC7/RX RC6/TX RC5/SDO RC4/SDI RC3/SCK RA5/AN4/SS VSS 19 18 17 16 15 14 7

FSP_TO_PRI SEC_ALARM

12 RC1/CCP2 13 RC2/CCP1

AGND 8

FSP_FROM_PRI PNEU_RESET ADULT-0 SPI_MOSI SPI_CLK CS_SEC-0

SHDN_TO_PRI PS_ON SPI_MISO

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (2 of 10) page 5-7/8

P6VMIN C66
100nF 25V

P5A
1 VIN

LP2980-5V U7
GND 2 VOUT 5

PNEUMATICS
C37 2U2
16V

Pump Connector
J7
1 2 3 4 5 6

P5D P5D

PBAT C39 220U0

PBAT
3 1

3 ON/OFF

PBAT PUMP P5D DUMP P5D DUMP2

BAS16

D9

C65 P5D
100nF 25V
XFPM-050

TP37
PBACK_IN

TP38
PBACK_OUT

C140 33U0
25V

C28 33U0
25V

1 3 3 1

5V1 D1 D10
3 1

PT1

VCC VOUT GND CAP

3 2 1 6

R36 R180 C20 0R0

R51 2K0 C67


100NF 25V

PT1

BAS16

BAS16

D27

1K00 680P

Spare Parts
DUMP_ON_MON R202 100K
SI2306DS 3

U14
HCT04

P5A R9 10K0
3
TLC2272

P5A C60
100nF 25V

Q7

5
3

U14
HCT04

DUMP_ON

U14
HCT04

8 4

R10 10K0

FPT

DUMP2_ON_MON R203 100K DUMP2_ON R204 100K R205 100K R173 100K R121 100K
SI2306DS
3 1 1

SI2306DS
3

Q24

11

U14
HCT04

U13 C17 10N0 R6 10K0

10

13
2

U14
HCT04

R83 649K C64 10N0 P5A C61


100nF 25V

12

Q17

PNEU_OVERRIDE-0

P6VMIN
5 6 7 8 4 2 3

R29 4K99 C58


100nF 25V

R28 C43 470N 4K99

VCC 14 2 HCT04 GND 7

Q9 PUMP_ON

U14

ADC_PWM

SI9410DY

P5A2 LP2980-5V U6
GND 2 VOUT 5

1 VIN

Q8 FAST_OVP-0
4

5 6 7 8

3 ON/OFF

C36 2U2
16V

SI9410DY

R44
2 3

1K00 R75 0R10

C63 C59
100nF 25V
XFPM-050

R5 10K0

R99 10K0

R100 10K0

100NF 25V

PT2

VCC VOUT GND CAP

3 2 1 6

R50 R181 1K00 C21 680P 2K0 C57


100NF 25V

PT2

PUMP_I

PT2_SEC

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (3 of 10) page 5-9/10

BATTERY CHARGER
DC Input Power
J3
1 2 3

V_IN FS1
100nF

PBAT
2 VIN

L4 900NH_0A7

D5

MIC4681-ADJ U17
GND 6 7

C104

2A5

SW 3 FB 4 8

BYG22

68U0 L1

D3
BYG22

R96 301K V_DC_MON R87 100K C90


100nF

D8
SM6T39CA

C29 33U0
25V

C69
100nF

1 SHDN 5

D4
BYS10

C40 220U0
25V

C70

100NF 25V

R25 49K9

BC_5V
8

MC78L05
IN GND 2 3 6 7

PBAT VR1 50K R150 0R10

R149 0R10

U2

OUT

R39 301K

C68

100NF 25V

CW

VF = 6.9 V VC = 7.2 V

R42 10K0

Battery Connector
J2
1 2

BC_5V PBAT 4A0 FS4 R13 10K0 C71


100nF 25V

MBR2030CTL

D14

C98 33U0
25V

3 2 U15

8 OPA2340 1 4

R72 4K99

D12

BAS16

R11 R138 10K0 R92 100K R147 100K R76 0R10 R182 0R10 R61 C91 C118 R64 4K99 V_BAT_MON_SEC V_BAT_MON_PRI I_CHG FAST_CHG 249K C117
100nF 25V
5 U15 6 OPA2340 7

R81
200K

10K0
V = 0.191 V @ 75mA

C116
100nF 25V

R148 150K

R93 49K9

100NF 25V

R146 100K

Q15
Logic

100NF 25V

DTC143ZKA

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (4 of 10) page 5-11/12

+3.3 V SUPPLY
PBAT MIC4680-3V3 U12
GND 6 7

P3V3
1

2 VIN

SW 3 FB 4 8
1

68U0 L2

R101 10K0

C30 33U0
25V

100NF 25V

C74

1 SHDN 5

D6
BYG22

C41 220U0
25V

C75

100NF 25V

PS_ON

DTC143ZKA
Logic 1

Q12

SI2307DS
2

Q18

ANALOG VOLTAGE SUPPLY (+6.5 V)


1

68U0 L3

P6VMIN

5 VIN

MIC3172 U16

VSW 7

D7

C31 33U0
25V

C93

4 EN SGND 1

100NF 25V

FB 3 GND COMP 8 6 2

BYG22

R27 12K1

C96 220U0
25V

C97 220U0
25V

C76

100NF 25V

R65 1K00 C25 1U0

R33 2K87

PROGRAM CONNECTOR
J1
1 2 3 4 5

16V

PGM_PS_ON-0

+5.0 V SUPPLY
2 VIN

P5D

MIC39101-5V U5
GND

VOUT 3 FLG 4

C38 2U2
16V

C72

100NF 25V

1 EN 5 6

C73

7 8

100NF 25V

C4 10U0

RESET_TEST-0

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (5 of 10) page 5-13/14

P3V3 C79 R123 R59 100K 100K PR_DI PR_ST6 PR_ST5 PR_ST4 PR_ST3 PR_ST2 PR_ST1 PR_CLK R58 100K R57 100K R56 10K0 R55 100K R54 100K R122 100K
100nF 25V

P5D

PRINTER INTERFACE
16 1Y1 14 1Y2 12 1Y3 3 2Y0 5 2Y1 7 2Y2 9 2Y3

20 19 VCC 2OE 2 1A0 4 1A1 6 1A2 8 1A3 17 2A0 15 2A1 13 2A2 11 2A3 1OE 1

U11 1Y0 18

HCT240
GND 10

P5A P5A PR_EOP P3V3 PR_EOP_EN Q11


Logic

P5A R60 30K1

P5D

Printer Connector
J9
1 2 3 4 5 6 7 8 9 10 11 12 13 14

R97 1M00
7

8 4

TLC2272

R91 20K0

5 6

U13

R89 30K1

DTA143XKA

R73 PBAT R14 10K0 100R

SI2307DS

Q5

15 16
2

R15 10K0

17 18

PBAT_PR
3

19 20 21 22

PR_VHON

Q3

DTC143ZKA
Logic

C6 33U0
25V

P5A R26 15K0

23 24 25 26 27

PR_TH P5D

R53 2K0

PR_LAT

Q4

DTC143ZKA
Logic

C77
25V 100nF

P3V3

C78
100nF 25V

PR_M1 PR_M3 PR_M2 PR_M4

2 13 5 VCC VS2 VS1 3 IN1 OUT1 4

7 IN2 15 IN3

LB1831M U4

OUT2 6 OUT3 14

11 IN4 OUT4 12 17 19 FG7 FG5 18 20 FG6 FG8 FG1 FG2 FG3 FG4 1 8 9 16

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (6 of 10) page 5-15/16

AUDIO ALARMS
P5D P5D
3 C 1 B

R49 1R00 C7 10U0


16V

PWM_VOL

R82 10K0 C26 1U0


16V

Q27 MMBT3904LT1
E 2

R24 100K R111 10K0 C11 10N0

C10 10N0

R41 4K99 R71 15K0

U10 LM386
3 2

1 4 7

C42
5

J4
1 2

C87

100NF 25V

C8 10U0
16V

R74 2R70 C86


100NF 25V

220U0 C88
100NF 25V

Speaker Connector

PWM_PITCH

D18

BAS16

R94 49K9 C89 10N0 P5D P5D R178 1K00 R179 2K0

P5D R124 10K0

SEC_ALARM

R46 1K00

D15 5V1

Q26 D_TO_COMM Q25


Logic Logic

DTC143ZKA

C85

P5D

D16 5V1

DTC143ZKA
Logic

Q28

P5D

DTC143ZKA

HOST COMM

100nF 25V

Comms Connector
GND TTL_TX
1 2

FS2

0A3

100NF 25V

C83

2 C1+ 4 C15 C2+ 6 C211 T1IN 9

15 VCC

12 16 FON FOFF

V+ 3

C84
100nF 25V

J6

U8 ICL3221
V- 7

TTL_RX 3 5V_FUSED 4 GND 5 REMOTE_ALARM-0 6 RS-232_TX 7 RS-232_RX 8

C81
100nF 25V

100NF 25V

C82

T1OUT 13 R1IN 8 1 INVALID GND 14 0

R1OUT

External DB15 Connector Mapping J6


1 2 3 4 5 6 7 8
1

1 EN

NAME GND TTL_TX TTL_RX 5V_FUSED GND REMOTE_ALARM-0 RS-232_TX RS-232_RX

DB15 1 2 3 4 7 8 11 13

D_FROM_COMM Q13
Logic

P5D

FS3

0A3

DTA143XKA

R110 10K0

NDT3055L

Q6

REM_ALARM-0 R102 10K0

D2
SM6T15CA

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (7 of 10) page 5-17/18

P3V3

Test Connector
R19 10K0 R18 10K0 R17 10K0
100NF 25V

R22 10K0

C92

T_TDI T_TDO T_CLK RESET_TEST-0 T_TMS T_DE T_TRST R21 1K00

1 2 3 4 5 6 7 8 9 10 11 12 13 14

J10

P3V3 R125 10K0

P5D

PBAT

Display Connector

1 2

J5

SW_MUX CS_DISPLAY SW_PWR-0 SPI_MOSI SPI_CLK V_IN R127 100K R126 100K

3 4 5 6 7 8 9

R133 499

R134 499

R135 499

CHARGING_LED 10

ISOLATED POWER SUPPLY


L12 C147 330P R200 150 MCB1608S102E D19
2 1

2 VIN

MIC39101-5V U29
GND

P5TEMP

VOUT 3 FLG 4

BYG22

C122 10U0

C148
100NF 25V

1 EN 5 6

C121 10U0

7 8

ISOLATION BARRIER

PBAT
1

SP

D21
2

R185 4K99

C124 10U0

C149
100NF 25V

C126 10U0
2 GND 5 -VIN

MIC5270-5.0 U28-VOUT
4

SP

N5VSP

BYG22

DO NOT POPULATE

100NF 25V

C151

C129 33U0
25V

C128 680P R153 75R0

T1 6395-TO12
7 2 2 8

P5VSP D20
1

2 VIN

MIC39101-5V U30
GND

VOUT 3 FLG 4

12

IPS_ON C127 1N0


PKG_TYPE=0603

15 SD-0 6 SYNC 4 VC 5 RREF

VIN

SW 11 FB 3 ROC 14 RCC 13

L13

11 5 12

L11 MCB1608S102E

BYG22

C123 10U0

C150
100NF 25V

1 EN 5 6

C125 10U0

R151 13K7 0.1%

7 8

MMZ2012R102A

SP

R152 3K01

LT1425CS U31

9 G3 16 G4 7 SG

G1 1 G2 8 PG 10
C101 1N0
3K15V

C102 1N0
3K15V

C145 1N0
3K15V

C146 1N0
3K15V

SP

ISOLATION BARRIER
SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (8 of 10) page 5-19/20

SpO2 Module Connector 10 pin


J11
1 2 3 4 5 6 7 8 9 10 11 12 13 14

Allow 4mm clearance between these signal and all other board signals
1 2 3 4

SpO2 Patient Connector


J8
NELLCOR
LED + RCAL - RETURN

FL3

8 7 6 5

MASIMO
IR LED CATHODE

1 2 3 4 5 6 7 8 9 3 4 10 2 1 11 12 13 14 15 2 1 16

DET + A GND DIGICAL DET - C

Z>200/100MHZ

INNER SHIELD DET ANODE LED RCAL DIGICAL

INNER SHIELD DET CATHODE RED LED CATHODE

TO J1 ON NELLCOR MP506

GND N/C LED N/C LED + GND

FL1

NC
3 4

Z>200/100MHZ FL2

ANALOG GND DET CATHODE

OUTER SHIELD DET ANODE

Z>200/100MHZ

J16
DET + A DET - C INNER SHIELD INNER SHIELD

1 2 3 4 5 6 7 8 9 10
3

P5VSP
NC
2

TO J1 ON MASIMO MS-5

N/C OUTER SHIELD IR LED CATH OUTER SHIELD RED LED CATH N/C

D35
BAV199

NC
1

P3V3 Q19
Logic

P5VSP R183 10K0


C B CTH E 5 6 4

SP

P3V3

DTA143XKA

R144 249

U23 4N35
1 2 AN

SpO2 Module Connector 14 pin


N5VSP P5VSP J12 P5VSP
1 2 3 4

R129 10K0

SP

NELLCOR
C-LOCK GND GND RESET GND +12V (N/A) ANALOG OUT (N/A) TRANSMIT (MP506 OUT) RECEIVE (MP506 IN) +5 ANALOG CTS (MP506 IN) GND +5V DIGITAL GND

C119 SPO2_RST-0
100nF 25V

P5D
U25 HCPL-2601
8 7 6 5 VCC VE VO GND AN CTH 2 3

C 3

Q16 MMBT3904LT1
2 E

1 B

R142 3K01

5 6 7 8 9 10 11 12 13 14

TO J2 ON NELLCOR MP506

D_FROM_SPO2

R136 499 P3V3 P5VSP


SP

R184 1K00
SP SP

D_TO_SPO2

DTA143XKA
Logic

Q20

U26 HCPL-2601

C120
8 7 6 5

R145 249

100nF 25V
SP

R139 1K00

J17
1 2 3 4 5 6 7 8 9 10

MASIMO
+5V GND GND GND RESET RECEIVE (MS-5 IN) TRANSMIT (MS-5 OUT)

2 3

AN CTH

VCC VE VO GND

TO J3 ON MASIMO MS-5

SP

ISOLATION BARRIER
SCHEMATIC

SP

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (9 of 10) page 5-21/22

P5TEMP P5TEMP R166 1K00 U36


TL431 1 8 2 3 6 7

C137 R186 10K0 C130 10U0


16V
17 AVDD 11 12 13 14 AD0 AD1 AD2 AD3 DVDD 1 DIN 20 DOUT 21 SCLK 22
100nF 25V
T

P5TEMP

D_TO_T_ADC D_FROM_T_ADC SCLK_TO_T_ADC DRDY_FROM_T_ADC P5TEMP R156 10K0 R155 10K0


T

6 5 4 3 2 1

Temp PGM Connector

P5TEMP P5TEMP R193 10K0 L14


1 2 3 4 5 6 3

R172 13K7 0.1% CLK_TADC R167 100R

DRDY-0 23 15 AINCOM U34 CS-0 19 9 REF+ 10 REF3 4 5 X_IN X_OUT RESET-0

DO NOT POPULATE

ADS1240
DSYNC-0 POL BUFEN PDWN-0 6 18 24 7

J15

Temperature Probe Connector


J14

BAS16

D24

16 AGND

DGND1 2 DGND2 8

R157
3

V_THERM
T T

ISOLATION BARRIER

BLM11B221SB L9 BLM11B221SB

BAS16
T T

D23

100NF 25V

C152

1K00

C136
100NF 25V

P5TEMP C138 P5TEMP C133


100nF 25V 100nF 25V
T

P3V3

U33 4N35
1 AN CTH C B E 5 6 4

R154 10K0 D_FROM_TEMP

P5TEMP P5TEMP
3

R168
T

L7 BLM11B221SB

BAS16

D28

R194 R195 R196 30R 30R 30R

R197 R164 30R 499K

SI2307DS
2 1

Q21

R192 10K0

R171 1K00

15 VDD1 16 VDD2 4 MCLR-0 19 20 1 2 3 18 17

BAS16

D29

R160 10K0
100NF 25V

MON_HTR C134

RB7 RB6 RB5 RB4/PGM RA0/AN0 RB3 RB2/TX RA1/AN1 8 RA2/AN2/VREF RB1/RX 7 RA3/AN3/CMP1 RB0/INT RA4/T0CKI/CMP2 OSC1

14 13 12 11 10 9

R191 2K21

2K0

DTA143XKA
Logic

Q23

D_TO_TEMP R169 1K00

U32 4N35
5 6 4 C B E CTH AN 1 2

R198 10K0
T

R188 100K
T

U35 PIC16F628
VSS1 5 VSS2 6

P5TEMP
T

OSC2/OUT

R187 10K0
T

C132 22P0 C131 22P0 XL4


4.0MHZ

L10 BLM11B221SB

BAS16

D22

R162 0R0

BAS16
3

D25

SI2306DS
1

Q22

TEMP_PROBE_IN
T

P3V3 EN_HTR V_JUMPER C142 2U2


16V
T

R158 10K0

R189 499K

R132 10K0

Temp Probe Switch Connector


P5D
1 2 3

J13

100NF 25V

C135

R165 499K
T

SCHEMATIC

PWA, MAIN BOARD, PROCARE

Schematic - Main Board P/N 2008855 Rev B (10 of 10) page 5-23/24

P5D
C3

PBAT

P5D
C2 100NF 20 VCC R127 100R 11 12 8 2 DS0 CLK Q0 OE1 OE2 S0 S1 CLR
9

100NF 24 VCC 7 13 6 14 5 15 4 16 17 18 R60 10K0 1% 1 23 EL E HEF4514 U2 Y0 Y1 A0 A1 A2 A3 Y2 Y3 Y4 Y5 Y6 Y7 Y8 Y9 Y10 Y11 Y12 Y13 Y14 Y15 11 9 10 8 7 6 5 4 18 17 20 19 14 13 16 15 PDG_L0 PDG_L1 PDG_L2 PDG_L3 PDG_L4 PDG_L5 PDG_L6 PDG_L7 DG11 DG10 DG0 DG1 DG2 DG3 DG4 DG5 DG6 DG7 DG8 PDG_L9 PDG_L10 PDG_L11 PDG_L12 DG9 PDG_L8 PDG_L11 R1 10K0 1% R13 10K0 1% R4 10K0 1% R3 10K0 1% R5 10K0 1% R6 10K0 1% R7 10K0 1% R8 10K0 1% R9 10K0 1% R10 10K0 1% R11 10K0 1% R12 10K0 1%

PDG_L[0:12]

IO0 IO1 IO2 IO3 IO4 IO5 IO6 IO7 Q7 DS7

DIG_MUX0 DIG_MUX1 DIG_MUX2 DIG_MUX3


P5D

2 3 21 22

P5D
19

3 1 19

HCT299 U5

Q15 2SB1188 R41 180R R53 180R

Q16 2SB1188 R52 180R


PDG_L0

Q17 2SB1188 R43 180R


PDG_L1

Q18 2SB1188 R45 180R


PDG_L2

Q19 2SB1188 R44 180R


PDG_L3

Q20 2SB1188 R49 180R


PDG_L4

Q21 2SB1188 R48 180R


PDG_L5

Q22 2SB1188 R47 180R


PDG_L6

Q23 2SB1188 R46 180R


PDG_L7

Q24 2SB1188 R51 180R


PDG_L8

Q25 2SB1188 R50 180R


PDG_L9

Q26 2SB1188

P5D

GND 10

Q1 DTC143ZKA

PDG_L10

Q3 DTC143ZKA

Q4 DTC143ZKA

Q5 DTC143ZKA

Q6 DTC143ZKA

Q7 DTC143ZKA

Q8 DTC143ZKA

Q9 DTC143ZKA

Q10 DTC143ZKA

Q11 DTC143ZKA

Q12 DTC143ZKA

Q14 DTC143ZKA

D1

P5D

HCT132

P5D

R61 10K0 1%

GND 12

DG[0:12]
R2 10K0 1%

U1

GND

BAT54C

R39 200R
C1 R40 470K 100NF

Q2 2SB1188 R42 180R


DG12 SA0 SA1 SA2
L

LED13 DMR20A-1 RED


2 15 13 11 9 5 SA4 7 3 14 10 4 12 17 DG0 A B C D1 D2 E1 E2 F G DP COM COM COM e d
DP

SYSTOLIC
SA0 2 15 13 11 9 5 SA4 SA5 SA6 7 3 14 10 4 12 17 DG1 SA1

MAP
LED10 DMR20A-1 RED
SA0 a f g e d
DP

PULSE
LED28 LTC4624Y
S0 R111 R105 200 R106 200 R107 200 200 R110 200 R109 200 R108 200 12 11 3 4 2 15 a
L3

LED16 DMR20A-1 RED


2 15 13 11 9 5 SA4 SA5 SA6 7 3 14 10 4 12 17 DG2 SA1 A B C D1 D2 E1 E2 F G DP COM COM COM e d
DP

PDG_L12

A B C D1 D2 E1 E2 F G DP COM COM COM

LED29 LTC2621JD
S8 R104 R98 620 R99 620 R100 620 620 R103 620 R102 620 R101 620 15 12 6 1 4 16 S9 S10 S11 S12 S13 S14 A/L1 B/L2 C/L3 D E f g c d
DP

RED
a b g e L2 d
DP

YELLOW
a b f g e
L2

a f g c b

a f g c b

50V

SA2 SA3

Q13 DTC143ZKA

SA3

SA2 SA3

a b f

S1 b S2 S3 S4 c S5 S6

A/L1 B/L2 C/L3 D E f g c d


DP

a b
L3

L1

f g

L1

f g

CONN_S_2X5_HDRF
V_IN 10 PBAT 9 SPI_CLK 8 SPI_MOSI 7 SW_PWR 6 CS_DISPLAY 5 P5D 4 SW_MUX 3 GND 2

PBAT

P5D
C9 100NF HCT132 R55

HCT132

P5D

SA5 SA6

P5D

U1

GND

HCT132P5D S0 U1 GND S1 S2 S3 S4 S5 S6 R23 330 R25 330 R27 330 R14 330 R24 330 R26 330 R28 330 SA0 SA1 SA2 SA3 SA4 SA5 SA6 SA0 SA1 SA2 C4 100NF SA4 SA5 IO0 IO1 IO2 IO3 IO4 IO5 IO6 IO7 Q7 DS7 7 13 6 14 5 15 4 16 17 18 S8 S9 S10 S11 C5 100NF 20 S12 S13 S14 S15 R29 330 R30 330 R31 330 R32 330 R36 330 R35 330 R34 330 R33 330 DG7 DG8 DG9 DG10 14 16 13 3 5 11 15 SA3

F e 7 G 3 DP DG0 2 DIG1 DG1 5 DIG2 DG2 138 DIG3 L1,L2,L3

e d

DP

F e 8 G 6 DP DG6 1 DG1 DG7 5 DG2 DG8 7 DIG3 14 L1,L2,L3

c d
DP

e d

DP

R131 P5D 100R

100R U1

GND

SA[0:6]

J1

LED17 DMR20A-1 RED


2 15 13 11 9 5 7 3 14 10 4 12 17 DG3 A B C D1 D2 E1 E2 F G DP COM COM COM DG4 e d
DP

DIASTOLIC
SA0 2 15 13 11 9 5 SA4 SA5 SA6 7 3 14 10 4 12 17 SA1 A B C D1 D2 E1 E2 F G DP COM COM COM e d f g a

LED18 DMR20A-1 RED


SA0 SA1 b SA2 SA3 2 15 13 11 9 c 5 SA4 SA5 SA6 7 3 14 10 4 12 17 DG5

LED19 DMR20A-1 RED


A B C D1 D2 E1 E2 F G DP COM COM COM e d
DP

SPO2
LED27 LTC4624JD
S0 R97 R37 330 R57 330 R58 330 330 R96 330 R95 330 R94 330 DG9 DG10 DG11
1 5 7 14

LED9 1 2 3
b g c e d
DP DP

GURX10D

RED

20 19 18 17 16 15 14 13 12 11
200 R119 R120 200 S15 S14 S13 S12 S11 S15 S[0:15]

RED
a b f g e
L2

12 11 3 4 2 15 8 6

P5D
needs pull up on main board

a f g c b

a f g c b

SA2 SA3

S1 S2 S3 S4 S5 S6

A/L1 B/L2 C/L3 D E F G DP DG1 DG2 DIG3 L1,L2,L3 DG11 e d


DP

a f g c

a b
L3

L1

4 5 6 7 8 9 10
DG6

R56 100R

20

P5D
11 12 8 2 3
19

VCC DS0 CLK Q0 OE1 OE2 S0 S1 CLR


9

PS0 PS1 PS2 PS3 PS4 PS5 PS6 PS7 LED30 LTC4627JD
A B C D E F G e d
DP

SA6

DP

R121 200

R122 200

R123 200

R124 200

HCT299 U4

19

TEMP
RED
a b f g b f g c d
DP

AUTO/TIME
LED11 LTC2623JD
S8 S9 R112 620 R113 620 R114 620 a b f g c d
DP

P5D

GND 10

RED
a b f g b
L1

R118 620 R117 620 R116 620 R115 620

13 15 2 6 5 16

A/L1 B/L2 C/L3 D E f g c d


DP

a f g c e

a b

a f g c e
L2

S10 S11 S12 S13 c S14 DG3 DG4 DG5 DG6 DG11

L3

a f g b

P5D

e d

P5D

VCC 11 12 8 DS0 CLK Q0

IO0 IO1 IO2 IO3 IO4 IO5 IO6 IO7 Q7 DS7

7 13 6 14 5 15 4 16 17 18

PS8 PS9 PS10 PS11 PS12 PS13 PS14 PS15

HCT299 2 OE1 U3 3 OE2 1 S0 19 19 S1 CLR


9

7 DP 1 DIG1 2 DIG2 6 DIG3 8 DIG4

DP

F e 7 G 3 DP 1 DIG1 14 DIG2 11 DIG3 8 DIG4 12 L1,L2,L3

e d

c d

e d

DP

DP

DP

P5D

GND 10

PS[0:15]

DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor DTD123EK is a digital transistor with a 2K2 base resistor and a 2K2 base pull-down resistor
S10 S9 S11 S12 S13 S1 S2 S3 S4 S5 S6 S7 S0 S8

S[0:15]
S14 S15

Q27 DTD123EK

Q28 DTD123EK

Q29 DTD123EK

Q30 DTD123EK

Q31 DTD123EK

Q32 DTD123EK

Q33 DTD123EK

Q34 DTD123EK

Q35 DTD123EK

Q36 DTD123EK

Q37 DTD123EK

SCHEMATIC
L

Q38 DTD123EK

Q39 DTD123EK

Q40 DTD123EK

Q41 DTD123EK

Q42 DTD123EK

PROCARE USER INTERFACE BOARD

Schematic User Interface Board P/N 2008741 Rev B (1 of 2) page 5-25/26

PS10

PS11

PS12

PS13

PS14

PS15

PS0

PS1

PS2

PS3

PS4

PS5

PS6

PS7

PS8

PS9

PCB BOARD SWITCH LOCATIONS


SILENCE
PBAT P5D

2 1 7 6 4

5 3 0 8 9

INFLATE CYCLE HISTORY PRINT ON/OFF


V_IN DG3
LED37 APTD3216QYC GREEN CHARGING LED21 APTD3216QYC GREEN R126 10K0 1%

ALARM
C7
16V
47U0

C12
100NF

C11
100NF

C10
100NF

C8
16V
47U0

C6
16V
47U0

50V

50V

50V

+ MENU

DG[0:12]

PDG_L[0:11]

DG4

R135 10K0 1% LED56 APTD3216SYCK YELLOW BELL LED60 APTD3216SYCK YELLOW LED54 AP3216SURC RED

DG5
R136 10K0 1%

SW_PWR

SW9
A B

ON/OFF
CHARGING

LED66 APTD3216QYC GREEN

BATT LOW LEFT LED62 AP3216SURC RED LED61 AP3216SURC RED

BATT LOW RIGHT LED69 APTD3216SYCK YELLOW

BATTERY RIGHT LED55 AP3216SURC RED

BATTERY MIDDLE LED68 APTD3216QYC GREEN

2 1

LED15 APTD3216QYC GREEN

BATTERY LEFT

LED1 APTD3216QYC GREEN

degF
LED65 APTD3216QYC GREEN

degC

LED57 APTD3216QYC GREEN

LED5 APTD3216QYC GREEN

SW_MUX

R20 38R3

R18 38R3

R133 38R3

R75 22R

R77 22R

R134 38R3

R16 22R

R74 22R

Q43 DTD123EK

D2

SW0
A

S7
B

S15

S7

S7

S15

S15

S15

S7

HISTORY

PDG_L0

BAT54C

SW1
A B

ALARM

PDG_L1

DG12

SW2 D6
A B

SILENCE

PDG_L2
minus

LED67 AP3216SURC RED R125 10K0 1%

PULSE LOW

LED53 AP3216SURC RED

PULSE HIGH

LED40 AP3216SURC RED

SpO2 LOW

LED52 AP3216SURC RED

SpO2 HIGH

LED50 AP3216SURC RED

DIA LOW

LED49 AP3216SURC RED

DIA HIGH

LED46 AP3216SURC RED

SYS LOW

LED44 AP3216SURC RED SYS HIGH LED45 AP3216SURC RED

LED42 AP3216SURC RED

BAT54C
A

SW3
B

PDG_L3

CYCLE SW4
A B

R64 38R3

LED38 AP3216SURC RED

LED39 AP3216SURC RED

LED41 AP3216SURC RED

LED51 AP3216SURC RED

LED48 AP3216SURC RED

LED47 AP3216SURC RED

LED43 AP3216SURC RED

R79 22R
PDG_L4

R65 22R

R63 22R

R38 22R

R21 22R

R19 22R

R17 22R

R15 22R

D5

MENU

S8

S7

S6

S5

S4

S3

S2

S1

S0

BAT54C

SW5
A B

INFLATE

PDG_L5

DG0

SW6 D4
A B

DG1 DG2 PDG_L6 DG12

MINUS

BAT54C
A

SW7
B

PDG_L7

AL VOL RIGHT

LED3 APTD3216QYC GREEN

AL VOL MIDDLE

LED2 APTD3216QYC GREEN AL VOL LEFT

LED58 APTD3216QYC GREEN

PUL VOL LEFT

LED64 APTD3216QYC GREEN PUL VOL MIDDLE

LED7 APTD3216QYC GREEN

PLUS SW8
A B LED8 APTD3216QYC GREEN LED6 APTD3216QYC GREEN LED59 APTD3216QYC GREEN LED63 APTD3216QYC GREEN LED20 APTD3216QYC GREEN

PUL VOL RIGHT

LED22 APTD3216QYC GREEN HISTORY LED4 APTD3216QYC GREEN

LED35 APTD3216QYC GREEN AUTO RIGHT LED36 APTD3216QYC GREEN LED70 APTD3216QYC GREEN

LED34 APTD3216QYC GREEN AUTO LEFT LED14 APTD3216QYC GREEN

LED32 APTD3216QYC GREEN INF PRES RIGHT LED33 APTD3216QYC GREEN

LED26 APTD3216QYC GREEN INF PRES MIDDLE LED31 APTD3216QYC GREEN

2 1

LED25 APTD3216QYC GREEN INF PRES LEFT LED12 APTD3216QYC GREEN

LED24 APTD3216QYC GREEN

LED23 APTD3216QYC GREEN

D3

PRINT

PDG_L8

R70 22R BAT54C

R72 22R

R22 22R

R62 22R

R68 22R

R80 22R

R90 22R

R137 38R3

R89 22R

R87 22R

R85 22R

R84 22R

R82 22R

S7

S15

S15

S7

S7

S15

S15

S14

S14

S13

S12

S11

S10

PROCARE USER INTERFACE BOARD

Schematic User Interface Board P/N 2008741 Rev B (2 of 2) page 5-27/28

2009829-001 A

You might also like