MenAMI Silica 03
MenAMI Silica 03
MenAMI Silica 03
Operator’s Manual
A-96.250.671 / 150517
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
E-mail: [email protected]
Document Status
ID: A-96.250.671
Revision Issue
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Restrictions for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.2 Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1 Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2 Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3 Install the Constant Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4 Connect Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.1 Sample Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.2 Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5 Install the 2nd Sample Stream Option . . . . . . . . . . . . . . . . . . . . . 21
3.6 Install the AMI Sample Sequencer . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6.1 Sample connection with 2nd sample stream. . . . . . . . . . . . . . . 25
3.7 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.7.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.7.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.8 Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.8.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.8.2 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.8.3 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.9 Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.9.1 Signal output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . 33
3.10 Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.10.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.10.2 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.10.3 HART Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.10.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.1 Activate the Peristaltic pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2 Prepare reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.3 Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.4 Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
A-96.250.671 / 150517 1
AMI Silica
5. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.1 Function of the Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.2 Measured Values and Symbols on the Display . . . . . . . . . . . . . . 40
5.3 Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.4 Changing Parameters and Values . . . . . . . . . . . . . . . . . . . . . . . . 43
5.5 Grab Sample Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.1 Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2 Stop of Operation for Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 45
6.3 Refill or replace Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.4 Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.5 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.6 Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.6.1 Disassemble the Constant Head . . . . . . . . . . . . . . . . . . . . . . . . 51
6.7 Cleaning the Photometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.8 Cleaning the solenoid valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.9 Tube Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.9.1 Replace the Pump Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.9.2 Tube Numbering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.10 Fill or Flush Reagent System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.11 Check Outputs and Relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.12 Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
6.13 Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
7. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.1 Slope Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
7.2. Grab Sample. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.3. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.1 Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
8.2 Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.3 Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.4 Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.5 Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
9. Program List and Explanations . . . . . . . . . . . . . . . . . . . . . . . . . 74
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
2 A-96.250.671 / 150517
AMI Silica
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
10. Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
10.1. Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
A-96.250.671 / 150517 3
AMI Silica
Safety Instructions
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important. Strictly
follow all safety instructions in this publication.
Target Operator: Qualified person who uses the equipment for its intended
audience purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location Keep the AMI Operator’s Manual in proximity of the instrument.
Qualification, To be qualified for instrument installation and operation, you must:
Training read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
4 A-96.250.671 / 150517
AMI Silica
Safety Instructions
DANGER
WARNING
CAUTION
Safety gloves
A-96.250.671 / 150517 5
AMI Silica
Safety Instructions
Corrosive
Harmful to health
Flammable
Warning general
Attention general
6 A-96.250.671 / 150517
AMI Silica
Safety Instructions
WARNING
WARNING
A-96.250.671 / 150517 7
AMI Silica
Safety Instructions
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual, as well as the
Material Safety Data Sheets (MSDS)
Reagent 1: Ammoniummolybdate consists of:
– Reagent 1a: Ammonium molydate Tetrahydrate
– Reagent 1b: Sodium hydroxide pellets
Reagent 2: Sulfuric acid 25%
Reagent 3: Oxalic acid
Reagent 4: Ammonium Ferrous Sulfate consists of:
– Reagent 4a: Sulfuric acid 25%
– Reagent 4b: Ammonium iron (II) sulfate hexahydrate
Silica Standard 100ppm 100ml
Download The current Material Safety Data Sheets (MSDS) for the above list-
MSDS ed Reagents are available for downloading at www.swan.ch.
8 A-96.250.671 / 150517
AMI Silica
Product Description
2. Product Description
Application The AMI Silica is a complete monitoring system for the automatic,
Range continuous measurement of the silica content in water of power
plants or demineralizer plants.
Silica The determination of silica is done by the photometric analysis of
Measurement molybdate blue at 810 nm.
Silica and ortho-phosphates react at low pH with ammonium mo-
lybdate to the yellow colored silico molybdic acid respectively phos-
pho molybdic acid. The phospho molybdic acid is destroyed with
oxalic acid before the silico molybdic acid is reduced with ammoni-
um iron(II) sulfate to the heteropoly-blue complex.
The required reagents are added in 3 steps to the sample in the pho-
tometer, where they allow a precise measurement of the silica con-
tent in the sample after the chemical reactions has been finished.
Grab Sample The measuring mode grab sample allows the measurement of
samples from remote locations.
Second If required the AMI Silica can be equipped with the optional second
Sample Stream sample stream module.
Sample If measurement of more than two sample streams is required, the
Sequencer AMI Silica can be connected to a Sample Sequencer which allows
to measure up to six sample streams.
Signal Two signal outputs programmable for measured values (freely scal-
Outputs able, linear or bilinear) or as continuous control output (control pa-
rameters programmable).
Current loop: 0/4–20 mA
Maximal burden: 510
Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable
via switch).
Relay Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A / 250 VAC
Operating The AMI Silica can be operated in the following 3 different modes:
Modes Internal
External
Fieldbus
Further information see Channel Selection, 5.1.6, p. 79.
A-96.250.671 / 150517 9
AMI Silica
Product Description
10 A-96.250.671 / 150517
AMI Silica
Product Description
A
Level
S B
Q C
P
O
D
N 6
5 E
1 7
4 F
2 G
3
H
I
M
L
K
J
A-96.250.671 / 150517 11
AMI Silica
Product Description
12 A-96.250.671 / 150517
AMI Silica
Product Description
1
2 150 sec
Y 3 90 sec
4 90 sec
5
6 X
1 2 3 4 5 6 7 8 9 10 11 12 13 14
A-96.250.671 / 150517 13
AMI Silica
Product Description
14 A-96.250.671 / 150517
AMI Silica
Product Description
400 mm / 15¾”
13 mm / ½”
374 mm / 14¾”
13 mm / ½”
AMI Silica
824 mm / 32 7/16”
850 mm / 33½”
4 x dia. 10 mm / 3/8”
A-96.250.671 / 150517 15
AMI Silica
Product Description
A
J
K
L
B M
C
N
D
O
E
F
G
H P
I
16 A-96.250.671 / 150517
AMI Silica
Installation
3. Installation
3.1 Installation Check List
Electrical Wir- NOTICE: Do not switch on the Instrument until all electrical
ing connections are made.
Connect all external devices like limit switches, current loops
and pumps.
Install the 2nd Sample Stream Option, p. 21 (if available)
Install the AMI Sample Sequencer, p. 24 (if available)
Connect the power cord, see Power Supply, p. 29
A-96.250.671 / 150517 17
AMI Silica
Installation
18 A-96.250.671 / 150517
AMI Silica
Installation
E
F
1 Push the overflow tube [C] through the flow cell block [E] as far
as it ends the waste funnel.
2 Fit the constant head tube [B] into the flow cell block [E].
3 Put the constant head cover [A] onto the constant head tube.
4 Adjust the overflow tube [C] so that it is at the lower level mark.
A-96.250.671 / 150517 19
AMI Silica
Installation
D
E
Connect the 1/2” tube [E] to the hose nozzle [D] and place it into a
pressure free drain with sufficient capacity.
20 A-96.250.671 / 150517
AMI Silica
Installation
E
F
X Level
G
N
Exit Enter
M
AMI SILICA
SiO2 Process Analyzer A
D
L
K
J
I
A-96.250.671 / 150517 21
AMI Silica
Installation
3 Unscrew and remove the flow regulating valve [E] from the flow
cell block [F].
4 Screw the blind plug [D] into the flow cell block.
5 Remove the sample inlet tube [N] from the sample inlet (elbow
union) [G].
6 Remove the elbow union from the flow cell block.
7 Put one of the O-rings [C] into the bore of the block housing [A]
and the other one into the bore of the flow cell block [F].
8 Insert the connecting piece [B] into the bore of the flow cell
block.
9 Slide the block housing [A] over the connecting piece and press
it against the flow cell block [F] while screwing the block hous-
ing with the 2 fixing screws [L] onto the panel.
10 Push a 15 x 20 mm tube [K] (not included in the installation set)
over the ½” hose nozzle of the block housing and put the end of
the tube into a pressure free drain.
11 Connect the sample stream 1 [I] and the sample stream 2 [J] to
the corresponding sample inlets of the block housing according
to section Sample Inlet, p. 20.
Connect the
solenoid valve
WARNING
Electrical shock hazard!
Before opening the AMI Transmitter switch power off.
Use one of the PG7 cable glands to feed the cable of the solenoid
valve into the AMI transmitter housing.
22 A-96.250.671 / 150517
AMI Silica
Installation
A B
Firmware After the 2nd sample stream option is installed and connected, set
settings 2nd the firmware according to your requirements.
sample stream 1 Navigate to menu <Installation>/
Sensors 5.1.5 <Sensors>.
Ref.Verification 0.235
Standard 100 ppb 2 Navigate to <Channels> with
Blank 0.0 ppb the[ ] or[ ] key.
Meas. interval 10 Min 3 Press [Enter].
Channels 1
4 Set <Channels> to 2 with the
Sensors 5.1.5 [ ] key.
Ref.Verifikation
Save?0.235
Standard Yes 100 ppb 5 Confirm with [Enter].
Blank No 0.0 ppb 6 Press [Exit], choose <Save> “Yes”
Meas. interval 10 Min Confirm with [Enter].
Channels 2
A-96.250.671 / 150517 23
AMI Silica
Installation
With 2nd sample stream option the AMI Silica can be operated in
the following 3 different modes.
internal
fieldbus
external
1 Navigate to menu <Installation>/
Sensors 5.1.6
Channel Selection <Sensors>.
Standard 100 ppb
Blank internal 0.0 ppb 2 Navigate to <Channel Selection>
fieldbus
Meas. interval
external 10 Min
with the [ ] or[ ] key.
Channels 2
3 Press [Enter].
Channel Selection Intern
Internal The switch over between the two sample streams is controlled by the
AMI Transmitter according to the programmed measuring interval.
Fieldbus The switch over between the two sample streams is controlled by
Profibus or Modbus.
External The switch over between the two sample streams is controlled via
an external device connected to the input.
24 A-96.250.671 / 150517
AMI Silica
Installation
B
C
D
2 1
E
F
A-96.250.671 / 150517 25
AMI Silica
Installation
WARNING
A B C
26 A-96.250.671 / 150517
AMI Silica
Installation
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
WARNING
WARNING
A-96.250.671 / 150517 27
AMI Silica
Installation
CAUTION
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.
28 A-96.250.671 / 150517
AMI Silica
Installation
WARNING
A
B
C
A-96.250.671 / 150517 29
AMI Silica
Installation
3.8.1 Input
NOTICE: Use only potential-free (dry) contacts.
The total resistance (sum of cable resistance and resistance of
the relay contact) must be less than 50 Ω.
Terminals 16/42
If signal output is set to hold, measurement is interrupted if input is
active.
For programming see menu 5.3.4, p. 91
1) usual use
30 A-96.250.671 / 150517
AMI Silica
Installation
Relay Jumper
config. Terminals pos. Description Relay configuration
Normally 6/7: Relay 1 Inactive (opened) during
Open 8/9: Relay 2 normal operation and 6
loss of power.
Active (closed) when a 0V
programmed function is 7
executed.
Normally 6/7: Relay 1 Inactive (closed) during
Closed 8/9: Relay 2 normal operation and 6
loss of power.
Active (opened) when a 0V
programmed function is 7
executed.
A
B
A-96.250.671 / 150517 31
AMI Silica
Installation
CAUTION
Inductive load Small inductive loads (max 0.1A) as for example the coil of a power
relay can be switched directly. To avoid noise voltage in the
AMI Transmitter it is mandatory to connect a snubber circuit in par-
allel to the load.
A snubber circuit is not necessary if an AMI relaybox is used.
A A AC or DC power supply
B C B AMI Transmitter
C External power relay
D E D Snubber
E Power relay coil
Resistive load Resistive loads (max. 1A) and control signals for PLC, impulse
pumps and so on can be connected without further measures
A B A AMI Transmitter
B PLC or controlled pulse pump
C C Logic
Actuators Actuators, like motor valves, are using both relays: One relay con-
tact is used for opening, the other for closing the valve, i.e. with the
2 relay contacts available, only one motor valve can be controlled.
Motors with loads bigger than 0.1A must be controlled via external
power relays or an AMI relay box.
A A AC or DC power supply
B C B AMI Transmitter
C Actuator
32 A-96.250.671 / 150517
AMI Silica
Installation
A-96.250.671 / 150517 33
AMI Silica
Installation
A
Third signal output 0/4–20 mA PCB
A Operating mode selector switch
ON
34 A-96.250.671 / 150517
AMI Silica
Installation
USB Interface
A USB Interface PCB
B Third signal output 0/4 - 20 mA PCB
A-96.250.671 / 150517 35
AMI Silica
Instrument Setup
4. Instrument Setup
36 A-96.250.671 / 150517
AMI Silica
Instrument Setup
CAUTION
A Overflow tube
B Flow regulating valve
Level
A-96.250.671 / 150517 37
AMI Silica
Instrument Setup
4.4 Programming
Programming Program all parameters for external devices (interface, recorders,
etc.) Program all parameters for instrument operation (limits,
alarms). See Program Overview, p. 69 and for explanations, see
Program List and Explanations, p. 74
Configuring Input:
38 A-96.250.671 / 150517
AMI Silica
Operation
5. Operation
A B C D
Exit Enter
Program
RUN 14:10:45 Main Menu 1
Access, Exit Enter
R1 Messages
Diagnostics
R2 0.42 ppb Maintenance
Exit Operation
25.4°C 32 B/s Installation
A-96.250.671 / 150517 39
AMI Silica
Operation
R1 15:10:52
F
R2 0.0 ppb
I 24.2°C 31 B/s
G
H
40 A-96.250.671 / 150517
AMI Silica
Operation
Display when
A B C D E
Operating with
two Sample
Streams
RUN 15:20:18
R1 15:10:52
Si1 0.0 ppb
F
15:20:52
R2
Si2 0.0 ppb ~
I 24.2°C 31 B/s
G
H
A-96.250.671 / 150517 41
AMI Silica
Operation
Main Menu 1
Messages
Diagnostics
Maintenance
Operation
Installation
Menu Messages 1
Messages 1.1
Reveals pending errors as well as an event history
Pending Errors
(time and state of events that have occurred at an
Maintenance List
Message List earlier point of time).
It contains user relevant data.
Menu Diagnostics 2
Diagnostics 2.1
Identification Provides user relevant instrument and sample data.
Sensors
Sample
I/O State
Interface
Menu Maintenance 3
Maintenance 3.1
For instrument calibration, relay and signal output
Calibration
Service
simulation, and to set the instrument time.
Simulation It is used by the service personnel.
Set Time 01.01.05 16:30:00
Menu Operation 4
Operation 4.1
User relevant parameters that might need to be
Grab Sample
Sensors modified during daily routine. Normally password
Relay Contacts protected and used by the process-operator.
Logger Subset of menu 5 - Installation, but process-related.
Menu Installation 5
Installation 5.1
Sensors For initial instrument set up by SWAN authorized
Signal Outputs person, to set all instrument parameters. Can be
Relay Contacts protected by means of password.
Miscellaneous
Interface
42 A-96.250.671 / 150517
AMI Silica
Operation
Yes is highlighted.
Logger 4.1.3
Log interval
Save ?
6 Press [Enter] to save the new pa-
Clear logger no rameter.
Yes
No The system reboots, the new
parameter is set.
A-96.250.671 / 150517 43
AMI Silica
Operation
<Enter> to continue
4 Wait until the measuring has been
Grab Sample 4.1.5 finished.
Wait till measuring is
completed
Progress
44 A-96.250.671 / 150517
AMI Silica
Maintenance
6. Maintenance
WARNING
A-96.250.671 / 150517 45
AMI Silica
Maintenance
WARNING
Health hazard
For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS).
Only persons trained in handling dangerous chemicals are al-
lowed to prepare the reagents.
Canister setup A Suction lance without
A
level detector
(containers 1–3)
B B Suction lance with
level detector
(container 4)
C Level detector
D 2 L mark
E Container 1:
C Ammonium
molybdate
F Container 2:
D 2L 2L 2L 2L Sulfuric Acid
E G Container 3:
Oxalic Acid
OXYCON ON-LINE SILCA
AMI SILICA REAGENT 1
OXYCON ON-LINE SILCA
AMI SILICA REAGENT 2
OXYCON ON-LINE SILCA
AMI SILICA REAGENT 3
OXYCON ON-LINE SILCA
AMI SILICA REAGENT 4 H Container 4:
F Ammonium Ferrous
G Sulfate
H I Holder
I
46 A-96.250.671 / 150517
AMI Silica
Maintenance
Reagent The 2 liter reagent canister will last for 1 month of operation with
consumption default measurement interval of 10 minutes. The provided reagent
set for 3 canister fillings therefore lasts for 3 months of operation.
A-96.250.671 / 150517 47
AMI Silica
Maintenance
WARNING
All canisters
Always replace the reagent filters (included with each reagent set)
when preparing new reagents.
Insert suction lances into the containers. Make sure that the num-
bers on the suction lances correspond to the numbers on the con-
tainers.
48 A-96.250.671 / 150517
AMI Silica
Maintenance
6.4 Verification
The “Verification kit for AMI Photometer” is available as
an accessory. An optical window with a precisely deter-
mined absorbance value is placed into the light beam
of the photometer. The actual measured absorbance
will be compared to the reference value labeled on
each kit.
Set reference Prior to performing the verification the reference value, e.g. 0.235,
value: needs to be set in menu <Installation \ Sensors \ Ref. Verification>
5.1.2*) Verification reference value.
Verification Basically follow the dialog in menu <Maintenance> \ <Service> \
procedure: <Verification> 3.2.1*)
1 Stop sample flow by closing the flow regulating valve. Wait for
next prompt: Constant head will be drained and an automatic
zero will be defined.
2 Open cuvette of the photometer and insert the verification filter.
[Enter] to continue.
3 Adjust for minimal absorbance (see AMI Display).
4 Press [Enter] to save the verification measurement. The verifi-
cation is successful if the difference is within the limits. <Enter>
to continue.
5 Remove filter, close cuvette and open regulating valve. <Enter>
to finish and [Exit] to the main display.
Verification Can be reviewed in menu Verification History
history: <Diagnostics \ Sensors \ FOME Sensor \ Ver. History> 2.2.1.5*
A-96.250.671 / 150517 49
AMI Silica
Maintenance
6.5 Calibration
Select Menu 3.1 (Maintenance/Calibration) and follow the instruc-
tions on the display.
Relay status during calibration:
Signal outputs are on hold
All limits are switched off
1 Close the flow regulating valve.
Calibration 3.1.1
- Close sample tap 2 Prepare 1 l of Standard.
<Enter> to continue
4 Wait until the calibration has been
Calibration 3.1.1 finished.
Wait till calibration is
completed
Progress
Calibration 3.1.1
Calibration completed
Wait till flow cell is empty
<Enter> to save
50 A-96.250.671 / 150517
AMI Silica
Maintenance
CAUTION
A-96.250.671 / 150517 51
AMI Silica
Maintenance
E
F
52 A-96.250.671 / 150517
AMI Silica
Maintenance
A-96.250.671 / 150517 53
AMI Silica
Maintenance
54 A-96.250.671 / 150517
AMI Silica
Maintenance
A-96.250.671 / 150517 55
AMI Silica
Maintenance
CAUTION
Pollution of reagents possible.
If the occlusion frames are opened during operation, already
mixed reagents will flow back into the reagent canisters and pol-
lute the reagents.
Never open the occlusion frames if the instrument is in opera-
tion.
Proceed according to Stop of Operation for Maintenance, p.
45 before opening the occlusion frames.
56 A-96.250.671 / 150517
AMI Silica
Maintenance
A-96.250.671 / 150517 57
AMI Silica
Maintenance
Level
6
5 5
1 7
6 7
4
2
3
1 2 3 4
J K L M
Tube
No. from to
1 Reagent container J 6-way valve port 1
2 Reagent container K 6-way valve port 2
3 Reagent container L 6-way valve port 3
4 Reagent container M 6-way valve port 4
5 6-way valve port 5 Photometer
6 6-way valve port 7 Peristaltic pump input
7 Peristaltic pump output Photometer reagent input
58 A-96.250.671 / 150517
AMI Silica
Maintenance
<Enter> to stop
Progress
Done
Press 4 x [Exit]
A-96.250.671 / 150517 59
AMI Silica
Maintenance
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
– relay 1
– relay 2
– alarm relay
When a fuse has blown, find out the cause and fix it before
replacing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
A B C D E F G
60 A-96.250.671 / 150517
AMI Silica
Maintenance
A-96.250.671 / 150517 61
AMI Silica
Troubleshooting
7. Troubleshooting
This chapter provides some hints to make troubleshooting easier.
For any detailed information how to handle or clean parts please
see Maintenance, p. 45. For any detailed information how to pro-
gram the instrument please see Program List and Explanations, p.
74.
62 A-96.250.671 / 150517
AMI Silica
Troubleshooting
Check pump A low calibration factor may be caused by leaking tube connec-
delivery tions. The delivery volume of the peristaltic pump is approximately
volume 4.2 g/min.
B
6
5 5
1 7
6 7
4 C
2
3
1 2 3 4
D
2.1 g
E
J K L M
A Flow regulating valve D Beaker
B Peristaltic pump E Balance
C Tube fitting
1 Close the flow regulating valve [A] to stop the sample flow.
2 Put a beaker [D] onto a balance [E] and set the balance to zero.
3 Unscrew and remove the tube fitting [C] from the photometer to
disconnect tube No. 7.
The sample contained in the photometer will flow out.
4 Put the tube into the beaker.
5 Navigate to menu <Maintenance>/<Service>/<Manual Filling>
and press [Enter].
6 Select <Position> and press [Enter].
7 Set the 6-way valve to position 1 with the [ ] or [ ] key.
The positions 1 to 4 of the 6-way valve are assigned to the 4
canisters.
A-96.250.671 / 150517 63
AMI Silica
Troubleshooting
8 After the 6-way valve position has been set, select <Pump> and
press [Enter].
9 Set the pump to <on> for a half minute.
The reagent is pumped into the beaker.
10 Read the weight on the display of the balance.
The weight must be within 2.0 to 2.3 g.
11 Repeat step 6 to 10 and set the 6-way valve to position 2, 3,
and 4 to check the tubes of the canisters 2, 3 and 4.
64 A-96.250.671 / 150517
AMI Silica
Troubleshooting
Main Menu 1
Reagent level low
Messages Indicates the remaining reagent in per-
Diagnostics cent
Maintenance Navigate to menu Messages.
Operation Press [ENTER].
Installation
A-96.250.671 / 150517 65
AMI Silica
Troubleshooting
66 A-96.250.671 / 150517
AMI Silica
Troubleshooting
A-96.250.671 / 150517 67
AMI Silica
Troubleshooting
68 A-96.250.671 / 150517
AMI Silica
Program Overview
8. Program Overview
For explanations about each parameter of the menus see Program
List and Explanations, p. 74.
A-96.250.671 / 150517 69
AMI Silica
Program Overview
70 A-96.250.671 / 150517
AMI Silica
Program Overview
A-96.250.671 / 150517 71
AMI Silica
Program Overview
72 A-96.250.671 / 150517
AMI Silica
Program Overview
A-96.250.671 / 150517 73
AMI Silica
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active,
acknowledged). If an active error is acknowledged, the alarm relay
is active again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5 Demands necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1 Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation: Designation of the instrument.
Version: Firmware of instrument (e.g. V6.20-08/16)
2.1.3 Peripherals:
2.1.3.1 PeriClip: Firmware of peristaltic pump (e.g. 1.06)
RoValve: Firmware of rotary valve (6-way valve) (e.g. 1.60)
2.1.4 Factory Test: Test date of the Instrument and Motherboard
2.1.5 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 FOME Sensor:
Current value: Shows the actual photometer signal in ppb.
Raw value: Shows the actual photometer signal in V.
Absorbance: Process value, depends on sample.
2.2.1.4 Cal. History: Review diagnostic values of the last calibrations.
Factor (slope).
Max. 64 data records are memorized. One calibration step
corresponds to one data record:
Slope photometer: 0.5–2.0
74 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
A-96.250.671 / 150517 75
AMI Silica
Program List and Explanations
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
3 Maintenance
3.1 Calibration
3.1.5 Performs a calibration using the standard solution. Follow the
instruction on the screen. See Calibration, p. 50
3.2 Service
3.2.1 Verification: Performs a verification using the verification kit.
Follow the instruction on the screen. See Verification, p. 49.
3.2.2 Fill System: Activates the reagent pump and fills all tubes from the
container to the 6-way valve.
3.2.3 Manual Filling: In the menu <Manual filling> the position of the
6-way valve can be set manually and the peristaltic pump can be
switched on and off.
3.2.3.1 Position: Set the position of the 6-way valve.
Pos. 1 Reagent 1
Pos. 2 Reagent 2
Pos. 3 Reagent 3
Pos. 4 Reagent 4
Pos. 5 Loop
Pos. 6 –
76 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
3.3 Simulation
Select the alarm relay, relay 1 and 2, signal output 1 and 2, input,
magnetic valve 1 and 2 or stirrer with the up/down keys. Then press
<Enter> to change the setting or value. After confirming the setting
with the Enter key, the value is simulated by the relay/signal output.
Alarm Relay: Active or inactive
Relay 1 and 2: Active or inactive
Input: Open or closed
Signal Output 1 and 2: Actual current in mA
Signal Output 3: Actual current in mA (if option is installed)
Magnetic valve 1 Active or inactive
Magnetic valve 2 Active or inactive
Stirrer Active or inactive
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated
values will be reset.
3.4 Set Time
Adjust date and time.
A-96.250.671 / 150517 77
AMI Silica
Program List and Explanations
4 Operation
4.1 Grab Sample
4.1.5 Starts a grab sample measurement. Follow the instruction on the
screen. See Grab Sample Measurement, p. 44
4.2 Sensors
4.2.1 Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
4.2.2 Hold after Cal: Delay permitting the instrument to stabilize again
after calibration. During calibration, verification or grab sample plus
hold-time, the signal outputs are frozen (held on last valid value),
alarm values, limits are not active.
Range: 0–6‘000 Sec
78 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1 Ref. Verification: Set absorbance value of verification kit according
to label.
Range: 0.200–0.600
5.1.2 Standard: Set the value in ppb of the standard you use for
calibration.
Range: 50 ppb–5 ppm
5.1.3 Blank: If the reagents contain a known amount of Silica, “Blank” can
be used to shift the zero point. The entered Blank value then is
subtracted from the measuring value.
Range: 0.0 ppb–10.0 ppb
5.1.4 Meas. Interval: The measuring interval can be set to 10, 15, 20 or
30 minutes.
5.1.5 Channels: Choose between 1 channel or 2 channel measurement.
If a Sample Sequencer is connected, up to 6 channels can be
chosen.
If 2 channels are selected, the transmitter switches automatically
between the channels after each measurement.
5.1.6 Channel Selection: The following 3 operating modes can be set:
Internal
Fieldbus
External
Mode Internal In the Mode Internal, the AMI Silica works as a master.
2nd Sample Stream Option
The AMI Silica switches automatically between channel 1 and 2.
Sample Sequencer
The AMI Silica sequentially measures each single sample stream of
the Sample Sequencer.
Via an external PLC it can be defined which sample streams should
not be measured. In the example below, only the samples streams
2, 4 and 6 are measured, whereas the samples streams 1, 3 and 5
are switched off. Sample streams which are switched off are
marked with an “x” behind the measuring value on the AMI Silica
display.
A-96.250.671 / 150517 79
AMI Silica
Program List and Explanations
CH1 9
1 mA sink, 5 V open CH2 10
6 Inputs from PLC
CH3 11
Processor
CH4 12
CH5 13
CH6 14
GND 15
SEQUENCER
Mode Fieldbus The AMI Silica is controlled via fieldbus.
Mode External In the mode External, the AMI Silica works as a slave.
2nd Sample Stream Option
The 2nd Sample Stream Option is switched between sample
stream 1 and 2 via input, see Input 5.3.4, p. 91.
Sample Sequencer
The AMI Silica is controlled by the Sample Sequencer. The Sample
Sequencer in turn is controlled via an external PLC. Each sample
stream to be measured has to be activated by closing the respective
contact.
Example:
If sample stream 1 of the Sample Sequencer is active, the AMI Silica
measures the sample stream 1 until the Sample Sequencer changes
to the next programmed channel. In the example below, the sample
stream 3 (CH3) highlighted in green will be measured as soon as the
AMI analyzer has finished the previous measurement.The current
measurement is completed before the channel is switched.
CH1 9
CH2 10
1 mA sink, 5 V open
6 Inputs from PLC
CH3 11
Processor
CH4 12
CH5 13
CH6 14
GND 15
SEQUENCER
80 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
20
(0 - 20 [mA])
(4 - 20 [mA])
B A
10 12
0/4 X
0.0 0.1 0.2 0.3 0.4 0.5
A-96.250.671 / 150517 81
AMI Silica
Program List and Explanations
[mA]
20
(0 - 20 [mA])
(4 - 20 [mA])
10 12
0/4 X
2 4 6
1 10 100 1’000 10’000
0 1 2 3 4
82 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
As control Signal outputs can be used for driving control units. We distinguish
output different kinds of controls:
P-controller: The controller action is proportional to the devia-
tion from the setpoint. The controller is characterized by the
P-Band. In the steady-state, the setpoint will never be
reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset
time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change
of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
Ziegler-Nichols method for the optimization of a PID controller:
Parameters: Setpoint, P-Band, Reset time, Derivative time
Y
B A
X
a
The point of intersection of the tangent with the respective axis will
result in the parameters a and L.
Consult the manual of the control unit for connecting and
programming details. Choose control upwards or downwards.
A-96.250.671 / 150517 83
AMI Silica
Program List and Explanations
Control upwards/downwards
Setpoint: User-defined process value (Measured value or flow)
P-Band: Range below (upwards control) or above (downwards
control) the set-point, within the dosing intensity is reduced from
100% to 0% to reach the set-point without overshooting.
5.2.1.43 Control Parameters: Temperature
5.2.1.43.10 Setpoint: -30 °C to + 120 °C
5.2.1.43.20 P-Band: 0 °C to + 100 °C
5.2.1.43 Control Parameters: Sample Flow
5.2.1.43.11 Setpoint: 0 –600 B/s
5.2.1.43.21 P-Band: 0 –200 B/s
5.2.1.43 Control Parameters: Si 1
5.2.1.43.12 Setpoint: 0 ppb –5 ppm
5.2.1.43.22 P-Band: 0 ppb –5 ppm
5.2.1.43.3 Reset time: The reset time is the time till the step response of a
single I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9’000 sec
5.2.1.43.4 Derivative time: The derivative time is the time till the ramp
response of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9’000 sec
5.2.1.43.5 Control timeout: If a controller action (dosing intensity) is constantly
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min
5.2.4 Signal Sequencer: Only visible if a Sample Sequencer is
connected.
Assign process value, the current loop range and a function to the
signal output.
5.2.4.1 Parameter: Only the parameter “Si Sequencer” is available.
5.2.4.2 Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
84 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
5.2.4.3 Scaling: Define the scaling of the signal output used to transmit a
process value. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 81
5.3 Relay Contacts
5.3.1 Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact is inactive at:
Power loss
Detection of system faults like defective sensors or electronic
parts
High case temperature
Lack of reagents
Process values out of programmed ranges.
Program alarm levels, hysteresis values and delay times for the
following parameters:
Alarm Si 1
Sample Temp.
Sample Flow
Case Temp. high
Case Temp. low
5.3.1.1 Alarm Si 1
5.3.1.1.1 Alarm High: If the measured value rises above the alarm high
value, the alarm relay is activated and E001 (E003), is displayed in
the message list.
Range: 0.00 ppb –5.00 ppm
5.3.1.1.25 Alarm Low: If the measured value falls below the alarm low value,
the alarm relay is activated and E002 (E004) is displayed in the
message list.
Range: 0.00 ppb –5.00 ppm
5.3.1.1.35 Hysteresis: Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value
fluctuates around the alarm value.
Range. 0.00 ppb –5.00 ppm
5.3.1.1.45 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
A-96.250.671 / 150517 85
AMI Silica
Program List and Explanations
86 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
5.3.2 and 5.3.3 Relay 1 and 2: The contacts can be set as normally open or
normally closed with a jumper. See Relay 1 and 2, p. 31. The
function of relay contacts 1 or 2 are defined by the user.
NOTICE: The navigation in the menu <Relay 1> and <Relay 2>
is equal. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
- End of Batch (relay 2 only)
- Channel Selection (relay 2 only)
2 Then enter the necessary data depending on the selected func-
tion.
5.3.2.1 Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20 Parameter: select a process value
5.3.2.300 Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter Range
Temperature -30 °C to + 120 °C
Sample Flow 0 –600 B/s
Si 1 0 ppb –5 ppm
5.3.2.401 Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter Range
Temperature 0 °C –100 °C
Sample Flow 0 –200 B/s
Si 1 0 ppb –5 ppm
5.3.3.50 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec
A-96.250.671 / 150517 87
AMI Silica
Program List and Explanations
88 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
A-96.250.671 / 150517 89
AMI Silica
Program List and Explanations
5.3.2.24 daily
The relay contact can be activated daily, at any time of a day.
5.3.2.341 Start time: to set the start time proceed as follows:
1 Press [Enter], to set the hours.
2 Set the hour with the [ ] or [ ] keys.
3 Press [Enter], to set the minutes.
4 Set the minutes with the [ ] or [ ] keys.
5 Press [Enter], to set the seconds.
6 Set the seconds with the [ ] or [ ] keys.
Range: 00:00:00–23:59:59
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/Control: see Interval
5.3.2.24 weekly
The relay contact can be activated at one or several days, of a
week. The daily starting time is valid for all days.
5.3.2.342 Calendar:
5.3.2.342.1 Start time: The programmed start time is valid for each of the
programmed days. To set the start time see 5.3.2.341, p. 90.
Range: 00:00:00–23:59:59
5.3.2.342.2 Monday: Possible settings, on or off
to
5.3.2.342.8 Sunday: Possible settings, on or off
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/Control: see Interval
5.3.2.1 Function = Fieldbus:
The relay will be switched via the Profibus input. No further
parameters are needed.
90 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
5.3.4 Input: The functions of the relays and signal outputs can be
defined depending on the position of the input contact, i.e. no
function, closed or open.
If <Channel Selection> in Menu < Installation>/<Sensors> is set to
“external”, the Input is set to “Active = no” and can be used to
switch the 2nd Sample Stream Option via an external device.
5.3.4.1 Active: Define when the input should be active:
The measurement is interrupted during the time the input is active.
No: Input is never active.
When closed: Input is active if the input relay is closed
When open: Input is active if the input relay is open
5.3.4.2 Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Cont.: Signal outputs continue to issue the measured
value.
Hold: Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off: Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
5.3.4.3 Output/Control: (relay or signal output):
Cont.: Controller continues normally.
Hold: Controller continues on the last valid value.
Off: Controller is switched off.
A-96.250.671 / 150517 91
AMI Silica
Program List and Explanations
5.3.4.4 Fault:
No: No message is issued in pending error list and the
alarm relay does not close when input is active.
Message E024 is stored in the message
list.
Yes: Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.
5.3.4.5 Delay: Time which the instrument waits, after the input is
deactivated, before returning to normal operation.
Range: 0–6‘000 Sec
92 A-96.250.671 / 150517
AMI Silica
Program List and Explanations
5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Language
German
English
French
Spanish
5.4.2 Set defaults: Reset the instrument to factory default values in three
different ways:
Set defaults
no
Calibration
In parts
Completely
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All
other values are set back to default values.
Completely: Sets back all values including communication
parameters.
5.4.3 Load Firmware: Firmware updates should be done by instructed
service personnel only.
Load Firmware
no
yes
5.4.4 Password: Select a password different from 0000 to prevent unau-
thorized access to the following menus:
5.4.4.1 Messages
5.4.4.2 Maintenance
5.4.4.3 Operation
5.4.4.4 Installation.
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representative.
5.4.5 Sample ID: Identify the process value with any meaning full text,
such as KKS number.
5.4.6 Line Break Detection: If activated, error message E028 is shown in
case of line break on signal outputs 1 and 2.
A-96.250.671 / 150517 93
AMI Silica
Program List and Explanations
5.5 Interface
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
94 A-96.250.671 / 150517
AMI Silica
Material Safety Data sheets
10.1. Reagents
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 1a for AMI Silica
Ammonium heptamolybdate tetrahydrate
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 1b for AMI Silica
Sodium hydroxide pellets
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 2 for AMI Silica
Sulfuric acid 25 %
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 3 for AMI Silica
Oxalic acid dihydrate
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 4a for AMI Silica
Sulfuric acid 25 %
Catalogue No.: Part of Article No. A - 85.420. 560
Product name: OXYCON ON-LINE SILICA
Reagent 4b for AMI Silica
Ammonium iron(II) sulfate hexahydrate
Catalogue No.: A85.142.400
Product name: Silica Standard 100ppm 100ml
Download The current Material Safety Data Sheets (MSDS) for the above list-
MSDS ed Reagents are available for downloading at www.swan.ch.
A-96.250.671 / 150517 95
AMI Silica
Default Values
96 A-96.250.671 / 150517
AMI Silica
Default Values
A-96.250.671 / 150517 97
AMI Silica
Index
12. Index
A Interface
Alarm Relay . . . . . . . 10, 30 HART . . . . . . . . ... 35
Ammonium Ferrous Sulfate Interface Modbus . . . . ... 34
Reagent 4 . . . . . . . . 48 Interface Profibus DP . ... 34
Ammoniummolybdate Interface RS485 . . . . 34–35
Reagent 1 . . . . . . . . 48 Interface USB . . . . . . . . . 35
Application Range . . . . . . . 9 Interrupt measurement . . . 38
C L
Cable thicknesses ...... 26 Log Interval .......... 78
Calendar . . . . . . ...... 90 Logger . . . .......... 78
Changing values . ...... 43
Constant Head . . ...... 51 M
Manual Filling . . . . . . ... 76
D Measurement cycle . . ... 12
Default Values ........ 96 Measurement interval . ... 47
Messumformer . . . . . ... 16
E Mode External . . . . . ... 80
Electrical wiring . ....... 17 Mode Fieldbus . . . . . ... 80
Error List . . . . . ....... 62 Mode Internal . . . . . . ... 79
Mounting requirements ... 18
F
Fill System . . . . . . .... 76 O
Filter Time Constant . .... 78 On-site requirements . 15, 17
Fluidics . . . . . . . . .... 10 Oxalic Acid
Reagent 3 . . . . . 46, 48
G
Grab Sample ......... 44 P
Power Supply . . . . . . 14, 29
Power-up . . . . . . . . . . . 17
H Program Access . . . . . . . 39
Hold after Cal . ........ 78 Programming . . . . . . . . . 38
I R
I/O State . . . . . . . . . . . . 76 Reagent 1
Input . . . . . . . . . 10, 30, 91 Ammoniummolybdate . 48
Instrument Setup . . . . . . . 17
98 A-96.250.671 / 150517
AMI Silica
Index
S W
Safety Features ....... 10 Wire . ............. 26
Sample Flow
Establish. . ....... 37
A-96.250.671 / 150517 99
AMI Silica
Notes
13. Notes
SWAN
is represented worldwide by subsidiary companies
and distributors.
SWAN Products
Potable Water
Cooling Water
Made in Switzerland