Manual Biometra Tadvanced PCR Systems Analytik Jena
Manual Biometra Tadvanced PCR Systems Analytik Jena
Manual Biometra Tadvanced PCR Systems Analytik Jena
Biometra TAdvanced
High-performance thermal cycler for
DNA amplification by PCR
Manufacturer Biometra GmbH (Analytik Jena GmbH)
Rudolf-Wissell-Str. 30
37079 Göttingen Germany
Phone + 49 551 50 68 6-0
Fax + 49 551 50 68 6-66
Email [email protected]
Copyrights and Microsoft and Windows are registered trademarks of Microsoft Corp.
Trademarks The identification with ® or TM is omitted in this manual.
Edition 01.21
Contents
1 Basic information ....................................................................................................... 9
1.1 User manual notes ............................................................................................................ 9
1.2 Intended use ................................................................................................................... 10
1.3 Warranty and liability ..................................................................................................... 10
6 Operation.................................................................................................................. 33
6.1 Power on self-test .......................................................................................................... 33
6.2 Login screen .................................................................................................................... 33
6.3 Login................................................................................................................................ 34
6.4 Home screen ................................................................................................................... 35
6.5 Block status ..................................................................................................................... 36
6.6 Quick start function ........................................................................................................ 37
6.7 Incubation mode ............................................................................................................. 40
Version 01.21 3
Contents Biometra TAdvanced Manual
11 Tools ......................................................................................................................... 85
11.1 Configuration .................................................................................................................. 85
11.1.1 Setting the date and time............................................................................................... 87
11.1.2 Automatic user logout .................................................................................................... 88
11.1.3 Configuring the beeper................................................................................................... 89
11.1.4 Network ........................................................................................................................... 89
11.1.5 Display brightness........................................................................................................... 91
11.1.6 Screen calibration ........................................................................................................... 92
11.1.7 Factory settings ............................................................................................................... 93
11.2 User management .......................................................................................................... 93
11.2.1 Creating a user ................................................................................................................ 95
11.2.2 Editing user settings ....................................................................................................... 96
11.2.3 Delete user ...................................................................................................................... 99
11.3 Documentation ............................................................................................................... 99
11.3.1 Run log file .................................................................................................................... 100
11.3.2 Power On Logfile .......................................................................................................... 102
11.3.3 Extended self-test log file ............................................................................................ 103
11.3.4 Error log file .................................................................................................................. 104
11.4 Extended self-test ......................................................................................................... 104
11.5 Service info file (SINF) .................................................................................................. 107
11.6 Backup ........................................................................................................................... 108
11.7 Cycler Information ........................................................................................................ 109
11.8 Contact........................................................................................................................... 110
12 Adaptation of programs ......................................................................................... 111
4 Version 01.21
Biometra TAdvanced Manual Contents
Figures
Fig. 1 The Biometra TAdvanced Thermal Cycler ........................................................ 15
Fig. 2 Voltage selector switch on the underside of the device .................................. 20
Fig. 3 Front view of the Biometra TAdvanced ............................................................ 22
Fig. 4 Rear view of the Biometra TAdvanced ............................................................. 23
Fig. 5 Quick block exchange ........................................................................................ 24
Fig. 6 Placement of sample tubes in different block systems ................................... 25
Fig. 7 Keyboard ............................................................................................................. 26
Fig. 8 Keypad ................................................................................................................ 27
Fig. 9 Start screen......................................................................................................... 33
Fig. 10 Login screen ................................................................................................... 34
Fig. 11 Select user screen ........................................................................................... 34
Fig. 12 Enter password screen ................................................................................... 35
Fig. 13 Home screen ................................................................................................... 35
Fig. 14 Block status/quick start button ..................................................................... 36
Fig. 15 Program quick start screen (Biometra TAdvanced Mono).......................... 38
Fig. 16 Program quick start screen (Biometra TAdvanced Twin) ........................... 39
Fig. 17 Multiblock selection query ............................................................................ 40
Fig. 18 Incubation mode screen (Biometra TAdvanced Mono) .............................. 41
Fig. 19 Incubation mode screen (Biometra TAdvanced Twin) ................................ 42
Fig. 20 Multiblock selection query (1) ...................................................................... 43
Fig. 21 Multiblock selection query (2) ...................................................................... 43
Fig. 22 Spreadsheet programming ............................................................................ 44
Fig. 23 Graphical programming................................................................................. 44
Fig. 24 Program overview screen .............................................................................. 46
Fig. 25 BLANK program template ............................................................................... 46
Fig. 26 Program template screen .............................................................................. 47
Fig. 27 Heated lid temperature setting screen ......................................................... 48
Version 01.21 5
Contents Biometra TAdvanced Manual
6 Version 01.21
Biometra TAdvanced Manual Contents
Version 01.21 7
Contents Biometra TAdvanced Manual
8 Version 01.21
Biometra TAdvanced Basic information
1 Basic information
The operating manual contains information about the design and operation of the
Biometra TAdvanced and provides personnel familiar with PCR technology with the
necessary know-how for the safe handling of the equipment and its components. The
operating manual further includes notes on the maintenance and service of the device
and potential causes and remedies of any faults.
These operating instructions apply to all Biometra TAdvanced as of software version
ME 1.70 - VRE 1.46 (see "Cycler Information" p.109)
Furthermore, sometimes the terms protocol and program are used synonymously,
these are one or more programmed temperature steps with different or the same
holding times and cycles.
Conventions Instructions for actions which occur in chronological order are numbered and
combined in action units.
Safety instructions are indicated by pictographs and signal words. The type and source
of the danger are stated together with notes on preventing the danger.
The elements of the control and analysis program are indicated as follows:
Program terms are identified with SMALL CAPS (e.g., Menu FILE).
Buttons are shown by square brackets (e.g., [OK] button)
Menu items are separated by arrows (e.g., FILE OPEN).
Symbols and signal words The operating manual uses the following symbols and signal words to indicate hazards
or instructions. The safety instructions are always placed before an action.
WARNING
Indicates a potentially hazardous situation which might cause fatal or very serious
injuries (deformities).
CAUTION
Indicates a potentially hazardous situation which might cause light or minor injuries.
NOTICE
Provides indications of potential material and environmental damage.
Version 01.21 9
Basic information Biometra TAdvanced
CE marking
The warranty duration and liability comply with the legal requirements and the
provisions in the general terms and conditions of Analytik Jena GmbH.
Deviations from the intended use described in this manual result in limitations of
warranty and liability in the event of damage.
Warranty and liability claims are excluded for personal injury and property damage
due to one or several of the following causes:
use of the Biometra TAdvanced other than intended
improper commissioning, operation and servicing of the device
modifications of the Biometra TAdvanced without prior consultation with Biometra
GmbH
unauthorized intervention in the Biometra TAdvanced
operation of the Biometra TAdvanced with faulty safety equipment
use of other than original spare parts, wearing parts or consumables
improper repairs
improper transport and storage of the Biometra TAdvanced
faults due to the non-observance of this manual
10 Version 01.21
Biometra TAdvanced Safety instructions
2 Safety instructions
For your own safety and to ensure error-free and safe operation of the Biometra
TAdvanced, please read this section carefully before commissioning.
Besides the safety instructions in this user manual and the local safety regulations that
apply to the operation of the device, the general applicable regulations regarding
accident prevention, occupational health and safety and environmental protection
have to be observed and complied with.
References to potential dangers do not replace the work protection regulations which
must be observed.
Damaged or missing safety symbols can cause incorrect actions leading to personal
injury or material damage. Do not remove safety symbols! Replace damaged safety
symbols immediately!
The following safety symbols are attached to the inside of the heated lid and the rear
of the device:
The Biometra TAdvanced has been built and certified according to safety standard
EN 61010-1. Do not modify the device in any way. Any modification of the device will
lead to a loss of warranty and the EN 61010-1 certificate and poses a potential risk.
Version 01.21 11
Safety instructions Biometra TAdvanced
2.6.1 General
The operator of the Biometra TAdvanced must make sure before each commissioning
that the condition of the device including the safety equipment is sound.
The following has to be observed:
Free access to the power switch on the back of the enclosure has to be ensured
during operation.
12 Version 01.21
Biometra TAdvanced Safety instructions
The ventilation fittings at the rear of the device must be unobstructed and
operational. Covered ventilation grilles or slits etc. may cause the device to break
down or may cause damage to it.
The use of oil between the samples and the sample block is not necessary to
achieve an improved heat exchange. However, if you still want to use oil, you
should use mineral oil. Do not use silicone oil.
Be careful to avoid crushing or pinching injuries when closing the device.
The Biometra TAdvanced may cause burns. The following has to be observed:
The thermal block, the samples and the heated lid reach high temperatures. There
is a risk of burns during contact.
Ensure that the lid is securely closed before starting the program! Do not touch the
heated lid!
The rapid heating of the thermal block can cause liquids to boil explosively. Always
wear safety goggles during operation!
Do not touch hot sample tubes or plates and do not open them or boiling liquid
may escape!
Only use plates and tubes that are suitable for high temperatures (up to 100°C), fit
well into the thermal block (no shaking) and whose lids seal tightly!
Work on the electrical components of the Biometra TAdvanced may only be performed
by a qualified electrical technician in accordance with the applicable electrical
engineering rules. Life-threatening electrical voltages may occur in the interior of the
Biometra TAdvanced!
The following has to be observed:
Any work on the interior of the device may only be carried out by the customer
service of Biometra GmbH and specially authorized technicians.
The electrical components must be checked regularly by a qualified electrician. Any
defects, such as loose connections, faulty or damaged cables, must be repaired
without delay.
Before opening the device it must be switched off at the power switch and the
mains plug must be disconnected from the mains outlet!
The Biometra TAdvanced must be switched off immediately using the power switch
(on the equipment back plate) and the mains plug disconnected from the power
supply if any faults occur in the electric components.
Version 01.21 13
Safety instructions Biometra TAdvanced
Do not operate the device near sources of strong electromagnetic radiation (for
example, an unshielded, deliberately operated high frequency source), as these
may affect proper operation of the device.
The operator is responsible for the selection of substances used in the process as well
as for their safe handling. This is particularly important for radioactive, pathogenic,
infectious, poisonous, corrosive or otherwise dangerous substances. For details contact
the safety officer responsible for your location. When handling dangerous substances
local safety codes and guidelines must be observed. The following general notes do
not replace the specific local regulations or the regulations in the EC safety data sheets
of the manufacturers for the auxiliary and operating materials.
The following has to be observed:
Protective goggles and rubber gloves have to be worn when handing reagents.
If only a few samples are treated, an (empty) tube of the same height must
additionally be placed at each corner position of the block. If the number of
samples in the block is too low, there is a risk of the tubes being damaged and
sample liquid leaking out.
For your own safety, please observe the potential infectious qualities of the
examined biological material.
In case of danger or accidents, immediately switch off the Biometra TAdvanced using
the main switch on the equipment back plate. Disconnect the mains plug from the
power supply!
14 Version 01.21
Biometra TAdvanced Function and setup
Model
Biometra TAdvanced 96 with aluminum block
Biometra TAdvanced 96 G with aluminum block with gradient function
Biometra TAdvanced 96 S with silver block
Biometra TAdvanced 96 SG with silver block with gradient function
Biometra TAdvanced 60 with aluminum block
Biometra TAdvanced 60 G with aluminum block with gradient function
Biometra TAdvanced 384 with aluminum block
Biometra TAdvanced 384 G with aluminum block with gradient function
Biometra TAdvanced Twin 48 with aluminum twinblock
Biometra TAdvanced Twin 48 G with aluminum twinblock with gradient function
Biometra TAdvanced Twin 30 with aluminum twinblock
Biometra TAdvanced Twin combi with aluminum twinblock
The device is controlled by a 7'' touchscreen display and an easy-to-use user interface.
PCR protocols and run log files can be saved to a connected USB stick. By using the
USB functions, PCR protocols can be exchanged easily between devices and run log
files can be stored for documentation reasons on a PC.
Version 01.21 15
Function and setup Biometra TAdvanced
The Biometra TAdvanced Thermal Cycler is available with different block formats and
with quick block exchange technology that allows easy exchange of block modules in a
few seconds. Most blocks are made of aluminum and offer high ramp rates for fast
protocol running times. Additionally, to achieve ultimate performance, a 96 well block
made of silver is available. Due to its excellent heat conductivity silver equilibrates
extremely quickly, thus providing maximum speed and temperature uniformity. To
protect the valuable silver blocks against corrosion, the block surfaces are gold plated.
A rubber seal on the heated lid forms an encapsulated space around the sample block
as soon as the lid is closed. The closed space serves to improve the temperature
uniformity of the sample block and avoids the formation of condensed water at the
final PCR cooling step. Furthermore all blocks are perfectly sealed to prevent
condensed water from penetrating into the Peltier elements underneath the sample
block and other parts of the electronics. The sealing protects the Peltier elements and
prolongs the lifetime of the device.
The Biometra TAdvanced heated lid has been optimized to heat twice as fast as with
older models and thus helps to shorten protocol run times. Moreover the heated lid
fulfills two other important functions: It prevents the formation of condensation at the
reaction tube portion located above the block surface level and it ensures reliable
contact between the reaction tubes and the thermoblock by applying constant
pressure. Thanks to the integrated clutch mechanism, the pressure applied to the
reaction tubes is always the same, regardless of the plasticware height. The
combination of the heated lid shape and the reliable contact pressure ensures even
temperature distribution between samples, thus significantly improving temperature
uniformity.
With one press on the front button, the heated lid gently swings open and locks in its
end position. This mechanism ensures that the lid cannot fall down and the heated lid
is out of the radius of action while the user inserts PCR tubes or plates. This prevents
any risk of injury from crushing or hot surfaces.
16 Version 01.21
Biometra TAdvanced Function and setup
3.3 Display
Many other thermal cyclers offer a so-called "List of latest programs used". In the list of
latest programs used a certain number of PCR programs is collected and offered for
quick start. The list is typically ordered only by date and the user has to search for the
desired program. In contrast the Biometra TAdvanced software creates a user-specific
list and only the latest programs started by the currently logged-in user are offered for
quick start. Even if a lab member does not use the device for a longer time, the
information on the latest programs started by this user will not get lost.
For maximum convenience the Biometra TAdvanced software additionally features a
program preview. Before the start of a program the protocol steps are summarized in a
clearly arranged table by the program preview tool. The preview therefore provides a
comprehensive overview of the protocol structure without the need to access
programming screens. The program preview is also available before the quick start of
programs.
Creating new PCR programs takes a lot of time if for each single step all parameters
have to be set manually. The Biometra TAdvanced offers two options for easy protocol
creation:
Pre-installed program templates
Protocol wizard
The Biometra TAdvanced software offers several pre-installed program templates for
different applications. The program templates provide a general protocol structure and
can be easily adapted for the current experiment.
The protocol wizard offers the comfort to create specific PCR programs by just a few
inputs. The protocol wizard is based on program templates for specific polymerases.
For up to eight different polymerases program templates can be saved in the Biometra
TAdvanced software and just the annealing temperature, cycle number and product
length has to be filled in to create a specific PCR program. Optionally two-step or
three-step PCR programs can be created by the software. Furthermore the protocol
wizard includes a primer annealing temperature calculator and the calculated
annealing temperature Ta value can be used for protocol creation.
Version 01.21 17
Function and setup Biometra TAdvanced
For optimizing new primer pairs in a single run the Biometra TAdvanced can optionally
be equipped with a gradient function. Finding the best primer annealing temperature
is crucial for the specificity and efficiency of PCR reactions. Often the optimization of
experiments is conducted to a limited extent only leading to unspecific byproducts or
reduced PCR sensitivity. By using the gradient function, new primer pairs with
unknown annealing temperatures can be tested quickly and optimized in a very short
time.
For maximum convenience the Biometra TAdvanced offers the Linear Gradient Tool.
For most other devices the temperature difference from row to row or from column to
column in a gradient step varies across the sample block. The Biometra TAdvanced
Linear Gradient Tool enables the programming of gradient steps with a defined
temperature increment between the columns or rows. The Linear Gradient Tool option
allows the comfortable creation of gradient steps and thus the best annealing
temperature can be easily transformed to a non-gradient protocol.
The software can manage up to 90 user accounts. With each system start the Biometra
TAdvanced performs an initial self-test and additionally the software offers an
extended self-test that can be initiated by the user. The results of the extended self-
test are summarized in a protocol and stored by the device. In addition to the self-test
functions, the device creates and stores log files for each single run. For long-term
storage, the log files and extended self-test protocols can be exported as a proprietary
file format.
18 Version 01.21
Biometra TAdvanced Function and setup
Block model Block type Gradient range Block exchange Order numbers
Block module 96 Aluminum – 846-070-231
Block module 96 G Aluminum 30 °C 846-070-221
Block module 96 S Silver – 846-070-271
Block module 96 SG Silver 40 °C 846-070-261
Block module 60 Aluminum – 846-070-230
Block module 60 G Aluminum 30 °C 846-070-220
Block module 384 Aluminum – 846-070-234
Block module 384 G Aluminum 24 °C 846-070-224
Twinblock module 48 Aluminum – 846-070-232
Twinblock module 48 G Aluminum 20 °C 846-070-222
Twinblock module 30 Aluminum – 846-070-233
Twinblock module combi Aluminum – 846-070-235
Base unit
TAdvanced base unit 846-x-070-280
Version 01.21 19
Installation and commissioning Biometra TAdvanced
Unpack and carefully check the device. Immediately report any damage to Biometra
GmbH. Do not attempt to operate the device if there is any visible damage.
Keep the original packaging for possible return. The Biometra TAdvanced must always
be transported in its original packaging. If the packaging is no longer available, please
contact Biometra GmbH.
WARNING
Danger of electric shock! Before commissioning the device, make sure that the voltage
selector setting on the underside of the device corresponds to the mains voltage in
your laboratory.
20 Version 01.21
Biometra TAdvanced Installation and commissioning
4.4 Commissioning
WARNING
Danger of electric shock! Prior to commissioning, please make sure that the operating
voltage set on the device is the same as the mains voltage (see section "Setting the
operating voltage" p. 20).
NOTICE
Do not cover the ventilation slits! Ensure that the ventilation slits are clear. Insufficient
ventilation may cause the device to overheat.
Make sure that there are no objects underneath the device (e.g., paper) that may
be sucked against the ventilation slot and thus interfere with ventilation.
Place the device at a minimum distance of 10 cm between the rear of the device
and the wall or other objects.
In case of several cyclers side by side running simultaneously we would recommend
to keep a distance of at least 10 cm between every thermal cycler.
Version 01.21 21
Control elements Biometra TAdvanced
5 Control elements
22 Version 01.21
Biometra TAdvanced Control elements
The Ethernet port, mains plug and mains switch are located at the rear of the
Biometra TAdvanced.
The option for quick block exchange is available for all model types.
To change the block:
1. Switch off the device.
2. Lift release lever on the backside (1 in Fig. 5) then pull block slowly to the back (2).
3. Insert the new block module carefully from the back in the base unit (2) until the
plugs at the front of the module are in contact with the socket inside the
instrument.
4. Then carefully push down the release lever (1).
5. Switch on instrument.
✓ During boot up the block module (block format and serial number) is
automatically recognized.
Version 01.21 23
Control elements Biometra TAdvanced
The Biometra TAdvanced has a height-adjustable lid for optimal contact pressure on
the samples.
1. Place the samples in the block, then close the lid until you feel the button in the
front click into position.
2. Turn the lid wheel clockwise until the clutch mechanism is activated (you will hear
a clicking noise).
NOTICE
The strength of the lid contact pressure has been designed for a fully populated block.
If only a few samples are to be used in the block, place an additional two (empty)
tubes of the same height in the four corner positions of the block (see Fig. 6).
Otherwise the sample tube may be damaged from excess contact pressure.
Never attempt to close the heated lid with force! This may damage the device.
24 Version 01.21
Biometra TAdvanced Control elements
NOTICE
Never open the lid under pressure!
Always follow the sequence below when opening the lid. If the lid is opened under
pressure, the locking mechanism may be damaged.
Reduce the contact pressure of the lid completely by turning the lid wheel
anticlockwise until there is no resistance.
Open the lid by pressing the button on the front of the lid.
✓ The lid opens and locks in a defined end position.
Version 01.21 25
Control elements Biometra TAdvanced
The Biometra TAdvanced thermal cycler features a 7" color touchscreen and a
convenient software for operation.
5.5.1 Keyboard
For some applications, names, addresses or passwords have to be defined. For this
purpose, the Biometra TAdvanced software offers a keyboard to enter characters,
digits and special characters.
Fig. 7 Keyboard
SPACE
TAB
SHIFT
BACK
FORWARD
Note: For some functions the maximum number of characters is limited. Usernames,
passwords and program names can have a maximum of 13 characters and user initials
a maximum of 3 characters.
26 Version 01.21
Biometra TAdvanced Control elements
5.5.2 Keypad
In some screens of the Biometra TAdvanced software digits have to be entered. For
this purpose, the software features a keypad integrated in some screens:
Fig. 8 Keypad
The keypad can be used to enter digits, dots, minus signs and infinite. In addition the
following special keys are available:
TAB
INFINITE
CONFIRM
CANCEL
Version 01.21 27
Control elements Biometra TAdvanced
Command Icon
HOME
NEXT
BACK
RIGHT
LEFT
UP
DOWN
CONFIRM
CANCEL
LOGIN
USER MANAGEMENT
USER MANAGEMENT
ON/OFF
NEW USER
DELETE USER
EDIT USER
PROTOCOL WIZARD
NEW POLYMERASE
EDIT POLYMERASE
DELETE POLYMERASE
28 Version 01.21
Biometra TAdvanced Control elements
Command Icon
ANNEALING
TEMPERATURE
CALCULATOR
EDIT PRIMER
CONCENTRATION
EDIT PRIMER
CALCULATE PRIMER
SAVE ANNEALING
TEMPERATURE TA
INCUBATION
START INCUBATION
PROGRAMS
NEW PROGRAM/NEW
FROM TEMPLATE
OPEN TEMPLATE
COPY PROGRAM/COPY
ALL PROGRAMS
DELETE PROGRAM/
DELETE ALL PROGRAMS
EDIT PROGRAM
START PROGRAM
QUICK SELECTION
(PROGRAM OR USER)
STOP PROGRAM
DONE
Version 01.21 29
Control elements Biometra TAdvanced
Command Icon
SAVE PROGRAM
SAVE AS
PAUSE PROGRAM
START PROGRAM/
CONTINUE PROGRAM
SKIP STEP
BLOCK STOP
GRAPHICAL VIEW
SPREADSHEET VIEW
GRADIENT DISPLAY
EDIT STEP
INSERT STEP
DELETE STEP
TEMPERATURE
INCREMENT
TIME INCREMENT
HEATING RATE
GRADIENT
STANDARD GRADIENT
PROGRAMMING
30 Version 01.21
Biometra TAdvanced Control elements
Command Icon
PREHEATING OFF
PREHEATING ON
TOOLS
SETTINGS
AUTOMATIC USER
LOGOUT
BEEPER
NETWORK
DISPLAY BRIGHTNESS
DISPLAY DARKER
DISPLAY BRIGHTER
TOUCHSCREEN
CALIBRATION
FACTORY SETTINGS
DOCUMENTATION
Version 01.21 31
Control elements Biometra TAdvanced
Command Icon
VIEW PROGRAM
EXTENDED SELF-TEST
BACKUP
CYCLER INFORMATION
CONTACT
32 Version 01.21
Biometra TAdvanced Operation
6 Operation
After the Biometra TAdvanced is switched on, the device performs an automatic self-
test (power on self-test). During the test the device shows the following screen:
For the power on self-test a log file is created and stored in the thermal cycler memory
(see section "Power On Logfile" p. 102).
After the Biometra TAdvanced has performed the power on self-test, the login screen
is displayed. The login screen offers the following functions:
Version 01.21 33
Operation Biometra TAdvanced
6.3 Login
After starting the Biometra TAdvanced, the login screen is displayed. To log in as an
existing user, press
LOGIN
Note: The position of the buttons can change. The button for the most recently logged
in user is always shown in the first position at the top left of the screen. All the other
buttons become arranged from top left to bottom right according to the date and time
of the latest user login. If more than 6 user accounts are available, you might need to
use the LEFT or RIGHT cursor keys to turn the page (see section "Cursor keys" p. 27).
Touch the PASSWORD input field and enter the password using the Biometra TAdvanced
keyboard (see section "Keyboard" p. 26). The default password for the ADMIN
administrator is "Admin". Please note that the password query is case sensitive.
34 Version 01.21
Biometra TAdvanced Operation
After user login the Biometra TAdvanced software shows the home screen. The home
screen offers the following functions:
Version 01.21 35
Operation Biometra TAdvanced
In the login screen (see section "Login screen" p. 33) and in the home screen (see
section "Home screen" p. 35) a special button with information on the current block
status is shown. The sample block status can be FREE, RUNNING or PAUSE and is
indicated by the color of the button and the textbox in the middle of the button (2 in
Fig. 14). In the line above the textbox the current block temperature is shown (1)- If
the temperature is higher than 70°C, the characters are shown in red and the warning
Hot plus the corresponding warning symbol are displayed. In the line below the
textbox the remaining runtime is shown (3). The sample block number is shown above
the button.
1 Block temperature
2 Block status
3 Remaining run time / pause time
The following table summarizes the different button colors and sample block statuses.
When the button is pressed, it depends on the sample block status which function is
activated in the Biometra TAdvanced software. If the status is RUNNING or PAUSE, a
touch of the button will lead to the program view (see section "Display during a run"
p. 69). If the status is FREE, the quick start function is activated (see section "Quick start
function" p. 37).
36 Version 01.21
Biometra TAdvanced Operation
Biometra TAdvanced Note: Programs can only be started by users with the right START/STOP PROGRAMS
Mono activated (see section "Editing user settings" p. 96).
The Biometra TAdvanced software offers a quick start function for previously used
programs. To use the quick start function, press the block status display button in the
login screen (see section "Login screen" p. 33) or in the home screen (see section
"Home screen" p. 35).
Note: The quick start function can only be activated if the current block status is FREE
(see section "Quick start function" p. 37). If the status is RUNNING or PAUSE, a touch of
the button will lead to the program view screen instead.
If the block status button is used in the login screen before a user is logged in, a list of
the programs most recently used on the thermal cycler are offered for quick start. If, on
the other hand, the button is used in the home screen after user login, the software
will show a user-specific list of the latest programs used for quick start.
Version 01.21 37
Operation Biometra TAdvanced
Select the program for quick start in the table and press the [START] button to start the
selected program. If necessary, the program can be edited before the start or a new
program can be created based on a template (see section "Programming modes"
p. 44).
START
Note: Programs offered for quick start can only be edited or deleted by users with the
right WRITE/DELETE OWN PROGRAMS activated (see section "Editing user settings" p. 96).
Press the corresponding button to activate the programming mode or select a program
from the list of templates:
EDIT
Biometra TAdvanced Note: Programs can only be started by users with the right START/STOP PROGRAMS
Twin activated (see section "Editing user settings" p. 96).
The Biometra TAdvanced software offers a quick start function for previously used
programs. To use the quick start function, press the block status display button in the
38 Version 01.21
Biometra TAdvanced Operation
login screen (see section "Login screen" p. 33) or in the home screen (see section
"Home screen" p. 35).
Note: The quick start function can only be activated if the current block status is FREE
(see section "Quick start function" p. 37). If the status is RUNNING or PAUSE, a touch of
the button will lead to the program view screen instead.
If the block status button is used in the login screen before a user is logged in, a list of
the programs most recently used on the thermal cycler are offered for quick start. If, on
the other hand, the button is used in the home screen after user login, the software
will show a user-specific list of the latest programs used for quick start.
Select the program for quick start in the table and press the [START] button to start the
selected program. If necessary, the program can be edited before the start or a new
program can be created based on a template (see section "Programming modes"
p. 44).
START ON BLOCK …
To switch between the sample blocks press the button with the corresponding number.
SAMPLE BLOCK 1
SAMPLE BLOCK 2
Version 01.21 39
Operation Biometra TAdvanced
By activating the checkboxes select the sample blocks for program start.
Note: Programs offered for quick start can only be edited or deleted by users with the
right WRITE/DELETE OWN PROGRAMS activated (see section "Editing user settings" p. 96).
Press the corresponding button to activate the programming mode or select a program
from the list of templates:
EDIT
Biometra TAdvanced Note: Programs can only be started by users with the right START/STOP PROGRAMS
Mono activated (see section "Editing user settings" p. 96).
INCUBATE
40 Version 01.21
Biometra TAdvanced Operation
ENDLESS TIME
START
STOP
PAUSE
CONTINUE
SKIP
Note: To stop the incubation, use the corresponding command. Do not switch off the
device while the incubation is running. The device will react like after a power failure
and will restart the incubation step when it is switched on next time.
Biometra TAdvanced Note: Programs can only be started by users with the right START/STOP PROGRAMS
Twin activated (see section "Editing user settings" p. 96).
Version 01.21 41
Operation Biometra TAdvanced
INCUBATE
ENDLESS TIME
START ON BLOCK …
To switch between the sample blocks press the button with the corresponding number.
SAMPLE BLOCK 1
SAMPLE BLOCK 2
42 Version 01.21
Biometra TAdvanced Operation
By activating the checkboxes select the sample blocks for program start.
STOP
BLOCK STOP
PAUSE
CONTINUE
SKIP
Note: To stop the incubation, use the corresponding command. Do not switch off the
device while the incubation is running. The device will react like after a power failure
and will restart the incubation step when it is switched on next time.
After you have pressed the [BLOCK STOP] button, choose the block to stop incubation in
the following query.
Version 01.21 43
Creating, editing and saving programs Biometra TAdvanced
GRAPHICAL VIEW
SPREADSHEET VIEW
44 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
Both programming modes show the same information but in different ways. In the
spreadsheet mode all parameters are summarized in a table. In the graphical mode
symbols are displayed for some parameters (see the table below). Toggle to the
spreadsheet mode to view the settings for these parameters.
If a program has more than 6 steps, use the cursor keys to navigate in the spreadsheet
or the graphical display.
Only when right WRITE/DELETE ALL PROGRAMS is activated, users can edit programs
created by other users.
1. To create a new program or open a program template, first log in as an existing
user.
Version 01.21 45
Creating, editing and saving programs Biometra TAdvanced
PROGRAM
3. In the program overview screen, press the [NEW FROM TEMPLATE] button:
4. The program template screen opens (see Fig. 26). Search for a program template
using the cursor keys and press the corresponding button to select it.
UP
DOWN
Note: The Biometra TAdvanced offers several program templates for different purpo-
ses. You can either use a template and modify it as desired or select the option BLANK.
BLANK is a program with no predefined steps (see Fig. 25).
46 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
Select BLANK whenever a new program should be created and select another template
from the list whenever an existing program should be edited.
The program structure can be controlled in the program preview (see Fig. 26). If a
template is selected, the program steps and some additional information are displayed
in spreadsheet form. By using the program preview function, templates can be checked
before they are opened.
5. To open a program template, press the [OPEN TEMPLATE] button.
OPEN TEMPLATE
Each template has a specific name which is superimposed in the programming screen.
To enter a name, touch the NAME input field and enter a new name using the Biometra
TAdvanced keyboard (see section "Keyboard" p. 26).
Note: If a character is pressed, the pre-displayed name will be deleted and a new
program name can be set. If the program name should be modified, please first press
another button like delete or space.
To set the heated lid temperature, touch the LID input field. A small window opens in
the middle of the screen with a keyboard (see Fig. 27). Enter a value between 30 and
110 using the Biometra TAdvanced keyboard.
Note: The default setting for the heated lid temperature is 99 °C. 99 °C is sufficient for
typical PCR programs.
Use the corresponding buttons to switch the heated lid on or off.
Version 01.21 47
Creating, editing and saving programs Biometra TAdvanced
NOTICE
The heated lid automatically switches off at a temperature difference of >75 °C
between the block and the heated lid. At these low block temperatures, sample
condensation on the tube lid is no longer to be expected. In addition, it supports a
longer service life of the Peltier elements in the block, since no unnecessary cooling
against the heating cover is required.
The heated lid can be set to preheating ON or OFF using the corresponding buttons:
If the preheating mode is activated, the heated lid is first heated up to the
programmed temperature while the sample block is held constant at 25°C. After the
heated lid reaches the target temperature, a 40 second equilibration phase follows.
Then the sample block starts to heat up from 25°C to the first programmed target
temperature.
Note: For most applications it is recommendable to set the preheating mode to "ON".
48 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
When editing program steps, the following parameters can be entered in the Biometra
TAdvanced software:
Time increment (e.g., for long range PCR) by which the time is increased with each cycle. The
time increment can be between 0s and 240s.
Average heating and cooling rate between steps. The heating and cooling rate can be set
from 0.1°C/s to max. in tenths of a degree. (The maximum ramp rate is sample bock
specific). The ramp rate specifies the speed at which the selected step is reached with.
It is possible that in some models this display may show a slightly higher maximum ramping
rate than is achieved according to the technical specifications of the device. This is technically
justified and achieves optimum control of the maximum value. In fact, the maximum heating
and cooling rate is achieved as specified in the technical specifications.
Establishes a temperature gradient across the sample block. The maximum gradient
temperature range and the temperature range that can be used for the gradient will depend
on the installed sample block.
The Biometra TAdvanced offers two options for editing a step in a PCR program:
Change all parameters for a step on a screen
Change individual parameters directly in the spreadsheet or graphical display
EDIT
In the graphical view you can also access the screen for editing a program step by
touching the currently activated step again.
Version 01.21 49
Creating, editing and saving programs Biometra TAdvanced
Note: To program a step at least the temperature and the hold time must be defined.
3. To set the temperature, touch the corresponding input field and enter a value
between 3.0°C and 99.0°C. Use the [TAB] button or activate the next input field and
enter a hold time in H:MM:SS.
The Biometra TAdvanced software allows several program steps to be programmed
subsequently (Multi Step Programming).
4. Use the cursor keys to navigate between program steps.
NEXT STEP
PREVIOUS STEP
The currently activated step is highlighted in blue in the table above the input fields
(see Fig. 28 on page 48).
Note: By navigating between program steps it is not necessary to confirm the settings
for each single parameter or step and to leave the screen for editing program steps.
Instead, you can fill in the parameters for all steps subsequently and confirm your
entries after all steps are programmed.
Programming loops A typical PCR program consists of repetitive steps for denaturation, annealing and
extension. For the repetition of steps the Biometra TAdvanced software allows the
programming of loops. For a loop two parameters have to be set:
The step number to which the program should return (GO TO)
The number of repetitions (CYCLES)
1. To program a loop, select the last step of the loop and enter the step number to
which the program should return in the GO TO input field (see Fig. 28).
2. To define the number of repetitions, enter a number in the CYCLES field
(see Fig. 28).
For example, to program the loop for the following protocol, select step 4 and enter
the value 2 in the GO TO input field and the value 35 in the CYCLES input field.
50 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
In this example, the device first executes steps 1 to 4, then repeats step 2 to step 4 for
34 times, performs the final extension at 72°C and holds the temperature indefinitely
at 16°C in the last step.
Programming a For some applications, for example touch down PCR, temperature increments are
temperature increment used. The target temperature is increased or decreased in each cycle by the set value.
To program a temperature increment, touch the [T] button (see Fig. 28) and
enter a value between +20°C and -20°C in the corresponding input field.
Note: Programmed temperature increments only become active if they are in loops. If
a temperature increment has been programmed, the small symbol will be shown
at the corresponding step in the graphical programming mode.
Programming a time For some applications, for example long range PCR, time increments are used. The
increment time is increased in each cycle by the set value.
To program a time increment, touch the t button (see Fig. 28) and enter a value
between 0 s and 240 s in the corresponding input field.
Note: Time increments only become active if they are in loops. If a time increment has
been programmed, the small symbol will be shown at the corresponding step in
the graphical programming mode.
Version 01.21 51
Creating, editing and saving programs Biometra TAdvanced
Adjusting the ramp rate The average ramp rate can be adjusted for each step. This adjustment can be useful if
PCR programs are transferred from slower devices to the Biometra TAdvanced or for
special PCR applications, for example, telomerase PCR. The average ramp rate
specifies the speed at which the selected step is reached with.
To adjust the ramp, press the following button (see Fig. 28):
Then enter a value between 0.1°C/s and max. in tenths of degree Celsius.
It is also possible that the entered ramp rate is applied to all steps in the PCR
program. For this the following checkbox to the right of the input field for the ramp
rate must be activated.
Note: The maximum average ramp rate depends on the sample block installed in the
Biometra TAdvanced. If the ramp rate has been adjusted, the small symbol will be
shown at the corresponding step in the graphical programming mode.
In a specific column all wells have the same temperature but the temperature differs
from column to column. The temperature gradient can be applied in both directions;
the highest temperature can be either in column 1 or column 12.
Sample replicates are distributed across the columns of the sample block and tested for
the temperature giving the optimal result. To program a gradient step press the
following key (see Fig. 28)
52 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
GRADIENT
There are two options to program a gradient. To toggle between the two options,
press the corresponding button on the Biometra TAdvanced gradient programming
screen (see Fig. 31):
STANDARD GRADIENT
PROGRAMMING
To program a gradient using the Linear Gradient Tool, enter the desired annealing
temperature (ANNEALING TEMP.) and define a temperature increment (GRAD INCR.)
between the individual columns of the sample block. The effective temperatures for all
the other columns are automatically calculated and displayed as a bar chart (see Fig.
31).
Note: The annealing temperature is set for a specific column in the sample block. For
example, the annealing temperature in a 96 well sample block corresponds to the
temperature in column 6. The column number is indicated in parenthesis next to the
corresponding input field (see Fig. 31).
When setting a positive increment, the temperature in column 1 is the lowest and the
temperature in column 12 is the highest. Use the minus button to enter a negative
increment. When a negative value is used, the temperature in column 1 is the highest
and the temperature in column 12 is the lowest.
Version 01.21 53
Creating, editing and saving programs Biometra TAdvanced
In a specific column all wells have the same temperature but the temperature differs
from column to column. The temperature gradient can be applied in both directions;
the highest temperature can be either in column 1 or column 8.
Sample replicates are distributed across the columns of the sample block and tested for
the temperature giving the optimal result. To program a gradient step press the
following key (see Fig. 28)
GRADIENT
54 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
There are two options to program a gradient. To toggle between the two options,
press the corresponding button on the Biometra TAdvanced gradient programming
screen (see Fig. 31):
STANDARD GRADIENT
PROGRAMMING
To program a gradient using the Linear Gradient Tool, enter the desired annealing
temperature (ANNEALING TEMP.) and define a temperature increment (GRAD INCR.)
between the individual columns of the sample block. The effective temperatures for all
the other columns are automatically calculated and displayed as a bar chart (see Fig.
34).
Note: The annealing temperature is set for a specific column in the sample block. For
example, the annealing temperature in a 48 well Twinblock corresponds to the
temperature in column 4. The column number is indicated in parenthesis next to the
corresponding input field (see Fig. 34). When setting a positive increment, the
temperature in column 1 is the lowest and the temperature in column 8 is the highest.
Use the minus button to enter a negative increment. When a negative value is used,
the temperature in column 1 is the highest and the temperature in column 8 is the
lowest.
Version 01.21 55
Creating, editing and saving programs Biometra TAdvanced
Note: Programs with gradient steps can only be started on blocks with gradient
function. Please note that for twinblock models only the left block (block 1) features
the gradient function. If you attempt to start a gradient program on a block without
gradient function, the following message appears:
56 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
To edit program parameters in the graphical view, press the corresponding input field
for the parameter that you want to edit for the step (see Fig. 38).
Program options (loops, temperature and time increment, as well as the ramp rate)
must be activated using a button to the left of the graphical view (see Fig. 38). The
loop setting is predefined.
The activated program option below the graphical view can then be edited by pressing
the input field.
Version 01.21 57
Creating, editing and saving programs Biometra TAdvanced
When you press a cell in the spreadsheet view or an input field in the graphical view, a
small window opens in the middle of the screen with a keyboard (see Fig. 39).
This window allows you to enter the desired value for the parameter using the
keyboard.
58 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
If the sample block is gradient-enabled, you also have the option of programming a
gradient step.
To do this, when editing the temperature, press the GRAD button.
GRADIENT
Version 01.21 59
Creating, editing and saving programs Biometra TAdvanced
INSERT
When you press the [INSERT] button, the screen for editing program steps opens (see
section "Editing steps" p. 49). Enter the desired parameters for the new step.
Note: The new step will be inserted at the selected program step. If, for example, step
3 is selected, the new step will be inserted at position 3. The existing step 3 is not
overwritten; instead step 3 and all the following steps are moved one position higher.
This means the existing step 3 becomes step 4, step 4 becomes step 5 and so on.
Note: If a new step is inserted in an existing loop, the number of steps within the loop
increases, whereas the number of cycles remains the same.
DELETE
Note: If a step is deleted, all the following steps will be moved one position lower. If,
for example, step 3 is deleted, step 4 will become step 3, step 5 will become step 4 and
so on.
Note: If a step is deleted in an existing loop, the number of steps within the loop
decreases, whereas the number of cycles remains the same.
DONE
60 Version 01.21
Biometra TAdvanced Creating, editing and saving programs
Note: If a new program is to be saved, the first free storage location in the directory of
the currently logged in user is automatically selected. However, the storage location
can be changed using the cursor keys above and below the lists. If the selected storage
location is already in use, the stored program will not be overwritten; instead this
program and all other following programs will be moved one position higher. The
storage location number for these programs will increase by +1 and the new program
is saved at the selected location.
There is also the option of selecting a user directory or a program location using the
quick select function.
Touch the icon above the user or program view (see Fig. 41).
RIGHT
LEFT
To select a user or program, press the corresponding button with the user or program
name. The user or the program will be selected and the program overview screen (see
Fig. 41) appears again.
Version 01.21 61
Creating, editing and saving programs Biometra TAdvanced
Before saving the program there is still the possibility of changing the program name.
To do this, press the [SAVE AS] button and enter the new program name using the
Biometra TAdvanced keyboard (see section "Assigning program names" p. 47).
SAVE AS
After selecting the user directory and the storage location and possibly changing the
program name, press the [SAVE] button on the Biometra TAdvanced program overview
screen (see Fig. 41) to save the program.
SAVE
62 Version 01.21
Biometra TAdvanced Starting, coping and deleting programs
Note: Programs can only be started by users with the right START/STOP PROGRAMS
activated (see section "Editing user settings" p. 96). Additionally the right READ ALL
PROGRAMS defines, if a user is allowed to read all programs or if the access is limited to
own programs. A user who is not allowed to read all programs cannot start programs
of others even if the right START/STOP PROGRAMS is activated.
The Biometra TAdvanced software offers a quick start function for previously started
programs (see section "Quick start function" p. 37).
This section describes the process of selecting and starting stored programs from a
user account. To select a program to be started, press the [PROGRAMS] button on the
Biometra TAdvanced home screen (see section "Home screen" p. 35).
PROGRAMS
The program overview screen opens with the directory of the logged in user
preselected (see Fig. 43).
Version 01.21 63
Starting, coping and deleting programs Biometra TAdvanced
Touch the corresponding buttons to select a user directory and a stored program. If the
number of user directories or programs is higher than the maximum number that can
be displayed in the list, the buttons above and below the lists can be used to scroll up
and down.
There is also the option of selecting a user directory or a program location using the
quick select function.
For the selected program a preview opens with general information on the program
and the steps listed in a table (see Fig. 44). The program preview can be used to check
the program before it is started. To start the selected program, press the [START]
button.
START
For twinblock models select the sample block to start and confirm your selection. The
program will be started on the selected sample blocks.
64 Version 01.21
Biometra TAdvanced Starting, coping and deleting programs
Note: Users can only copy and save their own programs when the right WRITE/DELETE
OWN PROGRAMS is activated. Programs of others can be copied and saved with the right
WRITE/DELETE ALL PROGRAMS activated (see section "Editing user settings" p. 96).
To copy a program, touch the corresponding buttons to select a user directory and a
program on the Biometra TAdvanced program overview screen (see Fig. 43 p. 63). If
the number of user directories or programs is higher than the maximum number that
can be displayed in the list, the buttons above and below the lists can be used to scroll
up and down. To copy the selected program, press the [COPY] button.
COPY
Touch the corresponding buttons to select a user directory and a storage location on
the Biometra TAdvanced program overview screen (see Fig. 46). The program to be
copied is highlighted in green. If the number of user directories or programs is higher
than the maximum number that can be displayed in the list, the buttons above and
below the lists can be used to scroll up and down. To save the selected program, press
the [SAVE] button.
SAVE
Note: If the selected storage location is already in use, the stored program will not be
overwritten; instead this program and all other following programs will be moved one
position higher. The storage location number for these programs will increase by +1
and the new program is saved at the selected location.
Version 01.21 65
Starting, coping and deleting programs Biometra TAdvanced
Note: Users can only copy and save their own programs when the right WRITE/DELETE
OWN PROGRAMS is activated. Programs of others can be copied and saved with the right
WRITE/DELETE ALL PROGRAMS activated (see section "Editing user settings" p. 96).
To copy all programs, touch the corresponding button to select a user directory on the
Biometra TAdvanced program overview screen (see Fig. 43 p. 63). If the number of
user directories or programs is higher than the maximum number that can be
displayed in the list, the buttons above and below the lists can be used to scroll up and
down. To copy all the programs, press the [COPY ALL] button.
COPY ALL
Touch the corresponding button to select a user directory on the Biometra TAdvanced
program overview screen (see Fig. 46 p. 66). If the number of user directories or
programs is higher than the maximum number that can be displayed in the list, the
buttons above and below the lists can be used to scroll up and down. To save the
programs, press the [SAVE] button.
SAVE
66 Version 01.21
Biometra TAdvanced Starting, coping and deleting programs
Note: The programs will be copied into the selected user directory beginning at the
first free storage location.
Note: Users can only delete their own programs when the right WRITE/DELETE OWN
PROGRAMS is activated. Programs of others can be deleted with the right WRITE/DELETE
ALL PROGRAMS activated (see section "Editing user settings" p. 96).
To delete a program, touch the corresponding buttons to select a user directory and a
program on the Biometra TAdvanced program overview screen (see Fig. 43 p. 63). If
the number of user directories or programs is higher than the maximum number that
can be displayed in the list, the buttons above and below the lists can be used to scroll
up and down. To delete the selected program, press the DELETE button.
DELETE
Version 01.21 67
Starting, coping and deleting programs Biometra TAdvanced
Note: If a program is deleted, all the following programs will be moved one storage
location lower. The storage location number for these programs will decrease by -1.
Note: Users can only delete their own programs when the right WRITE/DELETE OWN
PROGRAMS is activated. Programs of others can be deleted with the right WRITE/DELETE
ALL PROGRAMS activated (see section "Editing user settings" p. 96).
To delete all programs, touch the corresponding buttons to select a user directory on
the Biometra TAdvanced program overview screen (see Fig. 43 p. 63). If the number
of user directories is higher than the maximum number that can be displayed in the
list, the buttons above and below the lists can be used to scroll up and down. To delete
the programs, press [DELETE ALL].
DELETE ALL
68 Version 01.21
Biometra TAdvanced Running, pausing, continuing and stopping programs
SAMPLE BLOCK 1
SAMPLE BLOCK 2
After a program has been started (see section "Starting a program" p. 63) the Biometra
TAdvanced software can display the currently running program in a spreadsheet or
graphical view. In addition, the curve of the gradient can be displayed in gradient
steps. Use the following buttons to toggle between the different views.
GRAPHICAL VIEW
SPREADSHEET VIEW
GRADIENT VIEW
Biometra TAdvanced
Mono
Version 01.21 69
Running, pausing, continuing and stopping programs Biometra TAdvanced
Fig. 52 Display during the run in a graphical, spreadsheet and gradient view
70 Version 01.21
Biometra TAdvanced Running, pausing, continuing and stopping programs
Biometra TAdvanced
Twin
Version 01.21 71
Running, pausing, continuing and stopping programs Biometra TAdvanced
Fig. 55 Display during the run in a graphical, spreadsheet and gradient view
72 Version 01.21
Biometra TAdvanced Running, pausing, continuing and stopping programs
To pause a program, press the [PAUSE] button on the Biometra TAdvanced run display
screen (see Fig. 52 p. 44).
PAUSE
The word "Pause" is shown and the pause time is counted in the REMAIN. TIME field.
Biometra TAdvanced
Mono
Version 01.21 73
Running, pausing, continuing and stopping programs Biometra TAdvanced
Biometra TAdvanced
Twin
Note: Pressing the [PAUSE] button requires manual interaction. The Biometra
TAdvanced software also allows pauses to be programmed (see section "Editing steps"
p. 49), then the pause step is automatically started by the device.
Note: If the device is switched off during a pause step, this will be recognized as a
power failure (a corresponding error message is written to the run log file). If, for
example, the last step in a program is a pause, the program has to be stopped by
pressing the [STOP] button (see section "Stopping the program" p. 75), before
switching off the Biometra TAdvanced.
74 Version 01.21
Biometra TAdvanced Running, pausing, continuing and stopping programs
To continue a paused program, press the CONTINUE button on the Biometra TAdvanced
run display screen (see Fig. 57 p. 73).
CONTINUE
The Biometra TAdvanced will continue the program from exactly the point at which it
was paused.
To skip a program step, press the [SKIP] button on the Biometra TAdvanced run display
screen (see Fig. 52 p. 70).
SKIP
The Biometra TAdvanced will continue with the next program step and skip the
current step.
Note: Programs can only be stopped by users with the right START/STOP PROGRAMS
activated (see section "Editing user settings" p. 96).
To stop a program, press the [STOP PROGRAM] button on the Biometra TAdvanced run
display screen (see Fig. 52).
STOP PROGRAM
Version 01.21 75
Running, pausing, continuing and stopping programs Biometra TAdvanced
After confirmation of the confirmation prompt the Biometra TAdvanced stops the
current program.
Biometra TAdvanced If two sample blocks are running the Biometra TAdvanced offers the option to stop
Twin only one block or to stop both blocks simultaneously. Therefore, press the following
button:
BLOCK STOP
After confirmation of the confirmation prompt the Biometra TAdvanced stops the
program on the selected blocks.
76 Version 01.21
Biometra TAdvanced Protocol Wizard
10 Protocol Wizard
The protocol wizard is a helpful tool for automatic creation of PCR programs and
primer annealing temperature calculation. To start the protocol wizard press the
following button in the Biometra TAdvanced home screen (see section "Home screen"
p. 35):
PROTOCOL WIZARD
Pre-installed in the software is the program template for Analytik Jena Innupure
polymerases. To create a PCR program select a polymerase from the list and press the
[NEXT] button.
NEXT
In the next screen (see section "Enter a new polymerase or edit a polymerase" p. 78)
enter values for the requested parameters to continue with the automatic program
creation.
Version 01.21 77
Protocol Wizard Biometra TAdvanced
To enter program templates for new polymerases or to edit preset program templates
for polymerases press the corresponding button:
NEW POLYMERASE
EDIT POLYMERASE
In the next screen (see Fig. 64) touch the corresponding input boxes to enter a name
for the polymerase and set parameters for all requested program steps. With two
exceptions for most program steps the temperature and the hold time must be
defined:
The temperature for the annealing step depends on the calculated primer melting
temperature (Tm value) and has to be set individually for each primer pair (see
section "Primer Annealing Temperature Calculator" p. 82).
The hold time for the elongation step becomes calculated by the expected product
length.
The corresponding input boxes for the annealing temperature and elongation time are
shown in grey and cannot be activated.
After creating the program template for a new polymerase or editing the program
template of a polymerase press the [DONE] button.
DONE
In the next screen (see Fig. 65) touch the corresponding button to select a storage
place for the program template (polymerase). If the number of polymerases exceeds
the number of polymerases than can be displayed in the list, the cursor key function
above and below the polymerase list can be used to scroll up and down.
78 Version 01.21
Biometra TAdvanced Protocol Wizard
Note: If a new or edited polymerase is to be saved, the first free storage place in the
polymerase list is pre-selected. However, the storage place can be modified by using
the cursor key function above and below the list.
To save the new or edited polymerase press the [SAVE] button.
SAVE
Note: If the selected storage place is already in use, the stored polymerase will be
overwritten. Confirm the security query to overwrite the existing polymerase:
Version 01.21 79
Protocol Wizard Biometra TAdvanced
DELETE POLYMERASE
For automatic calculation of PCR programs in the protocol wizard parameter screen
(see Fig. 67) some parameters have to be set by the user:
1. Touch the corresponding checkbox to select between two step and three step
programs.
Note: The protocol wizard offers to create two step or three step PCR programs. A
three step program consists of separated steps for denaturation, annealing and
extension. In a two-step program the annealing and extension steps are combined
to one single step. In consequence for a three step protocol different temperatures
for the annealing and extension steps are used and for a two-step program the
primer annealing temperature Ta (see below) is used for the annealing and
extension step.
2. Enter the PCR product length.
Note: The product length is used by the software to calculate the elongation time.
The software uses the following equation:
Elongation time [s] = product length [bases]/1000 [bases]*60 s
For very short products the minimum elongation time is set to 6 s.
3. Enter the primer annealing temperature
Note: In general the annealing temperature (Ta) should be set approximately 5 °C
below the calculated mean primer pair melting temperature (T m). The theoretical
Ta value can also be calculated by the Biometra TAdvanced protocol wizard ("Primer
Annealing Temperature Calculator" p. 82).
4. Enter the number of cycles.
For gradient enabled 5. Optional: To define a temperature gradient for the annealing step press the [GRAD]
TAdvanced models button.
DELETE POLYMERASE
An additional input box GRAD. INCR. opens. Set the value for the temperature
gradient increment. For more information on the gradient temperature increment
see section "Editing steps" p. 49.
80 Version 01.21
Biometra TAdvanced Protocol Wizard
6. To set the parameters touch the corresponding field in the protocol wizard
parameter screen and enter the desired values.
NEXT
The next screen lists the all steps of the program that will be created by the
software:
Press the [CREATE PROGRAM] button to enter the TAdvanced programming screen (see
section "Creating, editing and saving programs" p. 44). In the programming screen the
program is displayed in graphical or spreadsheet mode and can be edited. To save the
Version 01.21 81
Protocol Wizard Biometra TAdvanced
program created by the protocol wizard follow the instructions in section "Saving the
program" p. 60.
CREATE PROGRAM
Note: Programs have to be saved before they can be started. If the program has been
created but should not be saved confirm the following security query to discard the
program:
Based on the primer sequences the Biometra TAdvanced protocol wizard can calculate
the theoretical primer annealing temperature (Ta). Press the [CALC TA] button in the
TAdvanced protocol wizard parameter screen (see Fig. 67 p. 81) to access the primer
annealing temperature calculator screen (see Fig. 70).
CALC TA
In the annealing temperature calculator screen enter the sequences for the forward
and reverse primer using the button below the input fields. The sequence length is
counted by the software and displayed in brackets left to the input fields.
82 Version 01.21
Biometra TAdvanced Protocol Wizard
To adjust the primer or salt concentration press the [EDIT CONCENTRATION] button.
EDIT CONCENTRATION
In the next screen the primer concentration, salt concentration and magnesium
concentration can be adjusted. Touch the corresponding input field and enter values
using the keypad. The concentrations have the following default values and ranges:
Version 01.21 83
Protocol Wizard Biometra TAdvanced
Press the [EDIT PRIMER] button to toggle to the primer annealing temperature
calculator screen (see Fig. 70) or press the [CALC. PRIMER] button to calculate the
theoretical primer annealing temperature (Ta).
EDIT PRIMER
CALC. PRIMER
The calculated values for the forward primer melting temperature (T m), reverse primer
melting temperature (Tm), mean primer melting temperature (Tm) and calculated
annealing temperature (Ta) are displayed.
Press the [SAVE TEMP.] button to transfer the calculated primer annealing temperature
to the protocol wizard (see Fig. 67 p. 81). The calculated Ta value will be displayed in
the input box for the primer annealing temperature.
SAVE TEMP.
Press the [INFO] button to get information on the algorithms used for annealing
temperature calculations.
INFO
For short sequences of 14 nucleotides or less the software calculates the primer
annealing temperature (Ta) based on the primer Tm values as follows:
1. For each primer the melting temperature is calculated using the following equation
according to Wallace et al. (1):
Tm = ((w*A + x*T)*2) + ((y*G + z*C)*4)
Where w, x, y and z are the number of the bases A, T, G and C in the primer
sequence, respectively.
2. The average Tm value is calculated from the Tm values for both primer sequences.
3. The annealing temperature Ta is calculated based on the average Tm value by
applying the following rules:
If the difference between the primer Tm values is 4 °C or less, the Ta is
calculated as the average Tm value minus 5 °C.
If the difference between the primer Tm values is higher than 4 °C, the Ta is
calculated as the lower Tm value plus 2 °C.
For long primer sequences of 15 nucleotides or more the software calculates the
primer annealing temperature (Ta) based on the nearest neighbor method according
to SantaLucia (2) and von Ahsen et al. (3).
1 Wallace et al., Nucleic Acids Res. 6, 3543, 1979
2 SantaLucia, Proc Natl Acad Sci U S A. 1998 Feb 17;95(4):1460-5
3 von Ahsen et al., Clin Chem. 1999 Dec;45(12):2094-101
84 Version 01.21
Biometra TAdvanced Tools
11 Tools
The Biometra TAdvanced software provides the following tools:
11.1 Configuration
Note: System settings can only be modified by users with the right SYSTEM
CONFIGURATION activated (see section "Editing user settings" p. 96). The system
configuration affects all functions listed below expect of the user management.
Version 01.21 85
Tools Biometra TAdvanced
DATE AND TIME Sets the date and time See section "Setting the date and time"
p. 87
AUTOMATIC USER LOGOUT Switches automatic user logout on or off and See section "Automatic user logout"
time settings p. 88
BEEPER Switches the beeper on or off See section "Configuring the beeper"
p. 89
NETWORK Defines network settings See section "Network" p. 89
DISPLAY BRIGHTNESS Adjusts the display brightness See section "Display brightness" p. 91
TOUCH SCREEN CALIBRATION Calibrates the touchscreen See section "Screen calibration" p. 92
FACTORY SETTINGS Resets all the settings See section "Factory settings" p. 93
86 Version 01.21
Biometra TAdvanced Tools
The Biometra TAdvanced software allows the user to set the date and time. The date
has the format DAY:MONTH:YEAR and the time the format HOURS:MINUTES:SECONDS.
Activate the corresponding input field and use the keypad on the right side of the date
and time entry screen to enter digits. The following buttons may also be helpful when
completing the fields:
Note: The following buttons are inactive on the date and time entry screen and cannot
be used:
Version 01.21 87
Tools Biometra TAdvanced
Note: The automatic user logout can only be configured by users with the right SYSTEM
CONFIGURATION activated (see section "Editing user settings" p. 96).
The Biometra TAdvanced features an automatic user logout function. The automatic
user logout can be switched ON or OFF by using the corresponding button:
AUTOMATIC LOGOUT ON
If automatic user logout is switched on, the time after which the user is logged out
from the system can be set. You can enter values between 30 and 60 minutes.
88 Version 01.21
Biometra TAdvanced Tools
Note: The beeper can only be configured by users with the right SYSTEM CONFIGURATION
activated (see section "Editing user settings" p. 96).
The Biometra TAdvanced has a beeper which, if switched on, will issue an acoustic
signal when a PCR program is finished. The beeper can be switched ON or OFF by
activating the corresponding checkbox.
11.1.4 Network
The NETWORK menu item can be used to configure basic settings or show a list of
network users. Press the corresponding button
NETWORK SETTINGS
NETWORK USER
Network settings Note: The network settings can only be configured by users with the right SYSTEM
CONFIGURATION activated (see section "Editing user settings" p. 96).
The Biometra TAdvanced can handle static and dynamic IP addresses. Activate the
corresponding checkbox to choose between dynamic (DHCP) and static IP address
management.
Version 01.21 89
Tools Biometra TAdvanced
If DHCP (Fig. 78) is activated, the device automatically receives the network
configuration settings. If the static checkbox is activated, you can enter the IP address
and the subnet mask using the keypad on the right side of the screen.
The port number can entered manually in both setting types (static and DHCP).
The following buttons may be helpful when completing the fields:
Note: The network cable must be at least performance class Cat 5e and the cable
configuration has to be STP.
90 Version 01.21
Biometra TAdvanced Tools
Network users The software stores the users who access the Biometra TAdvanced via the network
and lists them in chronological order in a table. The function makes it possible to check
if only authorized persons have network access to the device.
PCR control App The app is available for iOS (Apple App Store) and Android (Google Play Store) driven
systems.
The PCR Control App offers the following functions:
Live monitoring of PCR runs on network-integrated thermal cyclers- Starting and
stopping PCR protocols are stored on the thermal cycler
Notification system for monitoring status changes on selected blocks of a thermal
cycler
Store PCR protocols on the terminal. This can be stored on a PC using a USB stick
Copy PCR protocols from the terminal to a selected thermal cycler
Read device-specific parameters
Read out thermal cycler error messages
Note: The display settings can only be edited by users with the right SYSTEM
CONFIGURATION activated (see section "Editing user settings" p. 96).
The display brightness can be adjusted to the local light conditions. To adjust the
display brightness, use the corresponding buttons or the slider:
DARKER
BRIGHTER
Version 01.21 91
Tools Biometra TAdvanced
Note: The touch screen can only be calibrated by users with the right SYSTEM
CONFIGURATION activated (see section "Editing user settings" p. 96).
To calibrate the screen, follow the instructions and touch the screen at the center of
the displayed circle. This procedure is repeated three times at different positions on
the display. If, at the end of the procedure, a small cross is shown within the large
circle, the calibration was successful.
92 Version 01.21
Biometra TAdvanced Tools
Note: Factory settings can only be reset by users with the right SYSTEM CONFIGURATION
activated (see section "Editing user settings" p. 96).
Press the [FACTORY SETTINGS] button and confirm the following confirmation prompt:
Using the factory settings tool will delete all users and programs! Use a backup file to
restore the system (see section "Backup" p. 108).
Note: The [USER MANAGEMENT] button is only available for the user group
ADMINISTRATOR. Only an administrator can switch on or off the Biometra TAdvanced
Version 01.21 93
Tools Biometra TAdvanced
user management. Also with the user management switched off the administrator
ADMIN has to log in.
In the next screen activate the corresponding checkbox to configure the Biometra
TAdvanced user management:
If the user management is switched off, all new users have all rights and can access all
software functions.
With the user management switched on, the Biometra TAdvanced Thermal Cycler
supports 3 different user groups with pre-defined rights:
94 Version 01.21
Biometra TAdvanced Tools
WRITE/DELETE OWN x x
PROGRAMS
* An administrator can create users of all groups and modify their rights settings. A
user with generic rights is only allowed to create users with generic or limited rights
but no administrator. In addition, the user with generic rights cannot modify the rights
settings of other users.
The settings for each user can be modified by the administrator and will be stored by
the Biometra TAdvanced software.
Note: If the user management is switched off, user accounts and programs are not
protected and can be modified or deleted by all users.
Available users are listed on the Biometra TAdvanced user management screen (see
Fig. 82) sorted by the date and time of the last login and the username, the initials
and the group assignment are shown. Depending on the settings a user can be
grouped as:
Administrator User with limited rights
User with generic rights User defined
If more than 8 users have been created, you can use the cursor keys to navigate
through the table with the list of users (see Fig. 82).
UP
DOWN
Note: New users can only be created by users with the right CREATE USER activated.
An administrator can create users of all groups and modify their rights settings. A user
with generic rights is only allowed to create users with generic or limited rights but no
administrator. In addition, the user with generic rights cannot modify the rights
settings of other users.
Version 01.21 95
Tools Biometra TAdvanced
NEW USER
In the next screen enter the using the Biometra TAdvanced keyboard (see section
"Keyboard" p. 26)
USERNAME (up to 13 characters or numbers)
INITIALS (2 to 3 characters)
PASSWORD (up to 13 characters, numbers or special characters)
REPEAT PASSWORD
Set the language by activating the corresponding checkbox.
Note: User rights can only be edited by users with the right EDIT USER activated.
To edit the user settings, select a user on the Biometra TAdvanced user management
screen (see Fig. 82 p. 94) and press the [EDIT USER ACCOUNT] button.
96 Version 01.21
Biometra TAdvanced Tools
On the next screen the user data and the language settings can be changed.
Note: For the administrator ADMIN the user name and the initials ADM cannot be
modified. It is recommended to change the password for the administrator ADMIN after
the first system start and to create a backup of the system. Folders, programs, users
and rights settings can be restored from a backup file.
To modify the user rights press the [USER PROPERTIES] button.
USER PROPERTIES
Version 01.21 97
Tools Biometra TAdvanced
Right Function
SYSTEM CONFIGURATION Provides access to several system settings like date and time,
screen saver, beeper, network settings, display brightness,
touch screen calibration and factory settings
EDIT USER Allows to modify the user name, user initials, password,
language setting and user rights
CREATE USER Allows to create new user accounts. A user with generic rights
can only create generic or limited users and is not allowed to
modify the rights settings. An administrator is allowed to
create all user types and to modify the rights settings for each
user.
DELETE USER Allows to delete existing user accounts
WRITE/DELETE PROGRAMS OF Allows to edit/copy and delete all programs of all users
OTHER USERS
READ ALL PROGRAMS OF OTHER If inactive the access is limited to own programs. Programs of
USERS others cannot be viewed (if a program cannot be viewed it
also cannot be edited or started by the user).
START/STOP PROGRAMS Allows to start or stop programs
WRITE/DELETE OWN PROGRAMS Allows to edit/copy and delete own programs but not those of
other users.
SHOW YOUR OWN PROGRAMS TO Allows users to hide their own programs from other users.
OTHER USERS
Note: The settings for the administrator ADMIN cannot be changed. The administrator
always has all rights since at least one user needs to have all rights.
Press the [USER DATA] button to go back to the user data screen (see Fig. 85 p. 97).
The settings for each user will be stored by the Biometra TAdvanced software and are
shown in the user management screen.
98 Version 01.21
Biometra TAdvanced Tools
Note: User accounts can only be deleted by users with the right DELETE USER activated.
To delete a user, select the user on the user management screen (see section "User
management" p. 93) and press the [DELETE USER ACCOUNT] button.
Note: The administrator ADMIN cannot be deleted because at least one user must be
the administrator of the Biometra TAdvanced.
Confirm the confirmation prompt to delete the user:
After confirmation the user and all of his programs will be deleted.
11.3 Documentation
RUN LOG FILE Documentation of PCR runs See section "Run log file" p. 100
POWER ON LOG FILE Result of the latest initial self-test See section "Power On Logfile" p. 102
SELF TEST LOG FILE Result of the extended self-test See section "Extended self-test log file
p." 103
ERROR LOG FILE List of all the error messages detected by the device See section "Error log file" p. 104
Version 01.21 99
Tools Biometra TAdvanced
The Biometra TAdvanced lists up to 24 run log files for the programs that were last
started.
The run log files are sorted by date and time. The run log file for the program that was
last started is shown at the first position in the table. The initials of the user who
started the program, the number of the sample block being used, the program name,
the directory and the program number are also shown.
If more than 6 run log files have been created, you can use the following cursor keys to
navigate through the table with the run log files (see Fig. 89).
UP
DOWN
To view the collected data, select a run log file by activating the corresponding row in
the table and press the [VIEW RUN LOG FILE] button.
VIEW PROGRAM
The Biometra TAdvanced offers the option to show the program in graphical from or
as a spreadsheet (see section "Programming modes" p. 44). To toggle between the
graphical and spreadsheet view, use the buttons:
GRAPHICAL VIEW
SPREADSHEET VIEW
Use the buttons to toggle between the overview screen, the run log file screen and the
view program screen.
SHOW PROGRAM
To export all the saved run log files on USB, press the [SAVE ALL RUN LOG FILES ON USB]
button. To export a specific run log file, select a log file from the list on the run log file
overview screen (see Fig. 89 p. 100) and press the [SAVE RUN LOG FILE ON USB] button.
With each system start, the Biometra TAdvanced performs an initial self-test and
stores the results in a power on logfile.
In the power on logfile screen the following information is listed:
Power-on by date and time
Last power-on by date and time
During the extended self-test the Biometra TAdvanced checks several functions and
components like cooler, thermal tracking, heating and cooling rate, refrigeration,
gradient, heated lid and regulation (see section "Extended self-test" p. 104). Press on
the corresponding button to see the results for a specific test summarized in
spreadsheet form.
The test result can be PASSED or FAILED and is listed in the last row of the table for each
test.
The Biometra TAdvanced records all errors by date and time. The latest error is shown
at the first position in the error table. The error code and the message are also listed.
NOTICE
The Extended Self-Test must be performed under the following conditions to achieve
reproducible and meaningful results:
Ambient temperature in the range 20 °C to 25 °C
Mains voltage depending on the voltage type of the device: 100 V/115 V/230 V,
±10 % each
The device must be cooled to the ambient temperature of 20 °C to 25 °C after the
last application.
During the extended self-test the Biometra TAdvanced checks several functions and
components:
Test Function
Cooler Incubates the sample block to 4°C and checks if the
temperature in the block is reached and can be held for a
longer time.
Thermal tracking Checks the synchronism of the control circuits and if they work
together in a coordinated way.
Heating and cooling rate Checks if the specified average heating and cooling rate is
reached.
Refrigeration Checks if the heat sink and the fans are working properly
together.
Test Function
Gradient1 Checks if the sample block reaches the set gradient
temperatures.
Heated lid Tests if the heated lid reaches the set temperature and can
hold it for a longer time.
Regulation Tests if the sample block is controlled properly.
1
for gradient-capable models only
96 well sample block (Biometra TAdvanced Mono) 48 well sample block (Bioemtra TAdvanced Twin)
Fig. 95 Sample block loaded with plasticware for the extended self-test
When the plasticware has been properly inserted and the heated lid is closed, the
extended self-test can be started. To start the extended self-test press:
The software provides a progress overview during the extended self-test. When a test
has finished, the result PASSED or FAILED is displayed. For the currently running test, the
status RUNNING is shown and all other tests for which no result is shown are not
finished yet.
Note: The extended self-test will take approx. 30 minutes. The device is blocked while
the test is running and no other program can be started. To stop the extended self-
test, press the [CANCEL SELF TEST] button.
If the extended self-test is canceled, either no log file or an incomplete log file will be
created.
The results of the extended self-test are summarized in a log file, stored by the device
and can be viewed by the user (see section "Extended self-test log file" p. 103).
The service info file is a helpful tool for remote fault diagnosis by the Biometra service
department.
To create a service info file, connect a USB stick to the Biometra TAdvanced and press
the [INFO FILE TO USB] button.
The service info file can be sent by e-mail to the Biometra service department. For
contact details, refer to the inside front cover.
11.6 Backup
The backup function is used to synchronize the memory contents between Biometra
TAdvanced thermal cyclers. With the backup all folders, programs, users and user
settings can be stored to USB and transferred to another Biometra thermal cycler of
the same type.
Connect a USB stick to the device and press the [SAVE BACKUP FILE] button to create a
backup file.
To load a backup file, select it in the list and press the [LOAD BACKUP FILE] button.
Backup files that are no longer used can be deleted using the functions:
The Cycler Info screen provides general information about the device like:
Cycler type Revision number power board
Company Software version power board
Serial number cycler Software version power board logic
Block type Revision number processor board
Block serial number Software version processor board
Software version Revision number IO
Protocol version Software version IO
The information is summarized in two screens. Press the [SYSTEM/CYCLER INFO] button
to toggle between the two screens.
System/Cycler Info
11.8 Contact
12 Adaptation of programs
If programs are exchanged between different thermal cycler models or the sample
block is exchanged, programs might need to be adapted to the installed sample block
type prior to the start or during editing. The following table provides an overview of
possible program adaptations:
The maximum gradient A program for the 96 well silver block The temperature range of the gradient will be
range is exceeded. with a programmed gradient range of reduced based on the average temperature of the
>30°C is to be started on a 96 well maximum permitted range. If, for example, the
aluminum block and exceeds the average temperature of the programmed gradient
block's maximum permitted gradient is 60°C and the gradient range is ± 20°C, the
range. gradient will be reduced to 60°C ± 15°C.
A program with a gradient The program with the gradient step is The gradient will be deleted and the average
step is transferred to a started on a sample block without temperature or annealing temperature will be used
non-gradient-enabled gradient function. for this step. If the average temperature or
device. annealing temperature of the programmed
gradient is, for example, 60°C, this value will be
used for the step.
The gradient is outside the A program for the 96 well silver block The lower temperature is increased to the
permitted limit. with a gradient temperature below minimum permitted value. If, for example, a
20°C is to be started on a 96 well gradient is programmed from 5°C to 25°C, the
aluminum block and falls below the gradient is changed to 20°C to 25°C.
minimum permitted temperature.
A program with a A TOS program is transferred from The average value of the temperatures
temperature optimization the Biometra TRIO to another cycler programmed in the TOS step for the left and right
step (TOS) is transferred to type. sample block is used.
a device that does not
offer this function.
The heating and cooling A program for a 96 well silver block is The heating and cooling rate will be adjusted to the
rate is exceeded. to be started on a 96 aluminum block. maximum possible value for the installed sample
block type. If the programmed heating and cooling
rate is, for example, 8°C/s, it will be reduced for the
aluminum block to 6°C/s.
The heating and cooling A program for a 96 well aluminum The heating and cooling rate will be increased to
rate is not reached. block is to be started on a 96 silver the maximum possible value for the installed
block. sample block type. If the programmed heating and
cooling rate is, for example, 6°C/s, it will be
increased for the silver block to 8°C/s.
User-defined heating and The heating and cooling rate is lower The user-defined value is maintained and not
cooling rate. than the maximum value for a adjusted.
program step.
WARNING
Before initial operation check the correct operating voltage is set on the voltage
selection switch on the underside of the device. Do not open the housing of the device
unless you are authorized to do so. Danger of electric shock!
CAUTION
The sample block and the heated lid can reach high temperatures during operation.
Both the sample block and the heated lid can burn you. The samples are heated up
quickly and may come to an explosive boil. Wear goggles when handling hot samples.
Make sure that the lid is closed securely before starting the program!
NOTICE
The use of oil between samples and sample block is not necessary to achieve an
improved heat exchange. If you wish to use oil, never use silicone oil. Use mineral oil.
Ensure that the ventilation slits are clear. Insufficient ventilation may cause the device
to overheat. Tip: Clean the air inlets on the underside of the device at regular intervals
to ensure optimum airflow.
Important: Release the lid wheel before opening the lid! Adjust the lid wheel for
every individual run!
Login screen for Biometra TAdvanced Mono for Biometra TAdvanced Twin
Press [LOGIN] to log in as an existing user.
Press [QUICK START] for the non-user-specific quick start of programs.
Press [EN/DE/CN] to change the language setting of the login screen.
Home screen After login the Biometra TAdvanced home screen opens.
Home screen for Biometra TAdvanced Mono for Biometra TAdvanced Twin
Editing, saving and Press [PROGRAMS] on the home screen to open the program overview:
copying programs
To edit a template, press [NEW FROM TEMPLATE], select a template from the list and
press [OPEN TEMPLATE].
To edit a stored program, select a user directory and a program and press [EDIT].
Alternatively, a user or a program can be selected via the quick selection.
The Biometra TAdvanced programming screen opens. Use the following buttons to
toggle between the different programming modes:
Use the cursor keys above the keyboard to switch between program steps. The
number of the currently activated step is displayed on a blue background in the
table at the top of the screen.
Touch the TEMPERATURE input field and enter a temperature [°C].
Touch the HOLD TIME input field and enter a time. To program a pause step press .
To program a loop, touch the GO TO input field and enter the step number to which
the program should return. Then touch the CYCLES input field and enter how often
the loop should be repeated.
To reduce or increase the annealing temperature from step to step within a loop,
press ∆T and set a negative or positive temperature increment.
To reduce or increase the hold time from step to step within a loop, press ∆T and
set a time increment.
To adjust the heating rate, press ∆R and enter a value between 0.1 °C and max. You
can accept the heating rate for all steps in the program by activating the [APPLY
HEATING/COOLING RATE TO ALL STEPS] checkbox.
For gradient-capable To program a gradient, press GRAD. On the next screen you can choose between
devices only standard [STD.] and linear gradient programming [LIN.]. For standard programming
enter a temperature for both the first and last row in the sample block. For linear
programming enter the primer annealing temperature and set a temperature
increment. The gradient curve is displayed. Press ✓ to confirm your entries.
14 Fault removal
If the administrator has forgotten his password, use the FACTORY SETTINGS function (see
section "Factory settings" p. 93) to reset the system and download a backup file for
recovery (see section "Backup" p. 108).
The Biometra TAdvanced is equipped with strong fans that remove the heat from the
heat sink. The air supply to these fans is located on the underside of the device. Make
sure that the inlet is not clogged by dust or other material (e.g., a piece of paper under
the device may be sucked into the fan and restrict its functionality). Dirt should be
frequently removed from the air supply of the fan with a vacuum cleaner or a brush.
The Biometra TAdvanced features an auto restart function. If a power failure occurs
during a run, the device will continue the run at the same point when the power
returns. In case of long-term power failure (longer than 30 minutes) the device keeps
the sample block at 4°C (freeze step) and the user can decide whether the device
should continue the run or whether the samples should be discarded.
Note: After an auto restart the user has to log in again. In addition information can be
displayed for a limited time.
High voltage fluctuations may lead to a restart of the Biometra TAdvanced. The device
will then react similar to a power failure.
The following messages are displayed (see section "Auto restart" p. 116):
A power failure occurred during the run.
Time and step at which the program was restarted.
To avoid high voltage fluctuations, do not connect the Biometra to a socket shared by
other, very strong power consumers (such as refrigerators or centrifuges).
Note: After an auto restart the user has to log in again. In addition information can be
displayed for a limited time.
Since the Biometra TAdvanced thermal cycler is a very fast device, it may be necessary
to reduce the heating and cooling ramp when running protocols from other (slower)
thermal cyclers (see section "Adjusting the ramp rate" p. 52). Alternatively, the hold
times may be extended.
When the heated lid is in the highest or lowest position, it may happen that the lid
wheel is uncoupled. In this situation the clutch mechanism is active in both directions
(there is a clicking noise in both directions).
To unlock the lid wheel, press down the metal pin on the top of the lid wheel with a
pen. This overrides the clutch mechanism. Take care not to apply excessive pressure.
Lid blocked in the top position:
Important: When the clutch mechanism is active (= optimal contact pressure reached)
do not, under any circumstances, press the metal pin again to increase the contact
pressure further. This will damage the samples and the device.
For example:
- Using 96 °C then please set 96 °C lid temperature, independent from sample volume.
- Using 94 °C at air humidity > 70 %, please set 95 °C for sample volumes < 10 µl resp.
set 94 °C for sample volumes > 10 µl.
Even with these reduced lid temperatures, condensation will not be a problem. It is
recommended to reduce the lid temperature according to the reaction volume.
Otherwise sample volumes or amplification yields could be reduced.
WARNING
Risk of electric shock! Prior to commencing any maintenance or cleaning work, switch
off the device and pull out the mains plug. After cleaning, wait until the Biometra
TAdvanced is completely dry before recommissioning it.
WARNING
If you are handling infectious and pathogenic materials observe the relevant safety
regulations. Decontaminate the device in accordance with section "Disinfecting the
device" p. 120.
The professional decontamination of radioactive contaminations depends on the type
of substance used. Please contact your radiation protection officer on this matter and
observe the relevant safety regulations!
NOTICE
The use of alcohol, organic solvents or abrasives can damage the paintwork. When
cleaning, please also pay attention to the notes in section "Safety instructions –
maintenance and repair" p. 14.
Only wipe the housing of the Biometra TAdvanced with a soft clean cloth which
may be slightly wetted with a commercial neutral detergent, if necessary.
Clean the ventilation slits on the underside and at the rear of the device with a
vacuum cleaner.
WARNING
Biological hazard! Clean the Biometra TAdvanced with particular care after analyzing
potentially infectious material. Wear suitable protective equipment, e.g., protective
gloves.
NOTICE
The only suitable cleaning method for the housing is wipe disinfection.
When using spray disinfectants there is a risk that the liquid enters the sensitive
electronic system through the ventilation slits. If the disinfectant has a spray nozzle,
apply disinfectant to a suitable cloth before using it on the device.
Please contact the Biometra customer service department or your local dealer for
instructions on how to upgrade the Biometra TAdvanced firmware.
16 Transport
If you have any problems with the device, please contact the customer service
department or your local Biometra dealer. For the address of the Biometra customer
service department refer to the inside front cover of this operating manual.
Please observe the return information (see section "Device return information" p. 121)
if you wish to return a device to Biometra.
WARNING
Risk of damage to health due to improper decontamination! Perform a professional
decontamination before returning the device to Analytik Jena AG.
NOTICE
Biometra GmbH must refuse acceptance of contaminated devices. The sender may be
liable for any damage caused by inadequate decontamination of the device.
Biometra uses an extra designed packaging system with inserts of foamed material.
NOTICE
The thermal cycler is only protected from transport damage if the original packaging is
used and the packing instructions are followed. Biometra is not liable for transport
damage due to incorrect packaging.
17 Disposal
The user of the Biometra TAdvanced must dispose of any waste materials that occur
during measurements (sample materials) in accordance with statutory and local
regulations.
At the end of its service life, the Biometra TAdvanced and all its electronic components
must be disposed of as electronic waste in accordance with valid regulations.
18 Specifications
Software User-specific quick start option for the five most recent programs;
program preview before start; option for toggling between spreadsheet
and graphical programming mode; Linear Gradient Tool1, service info
file (SINF) generation; expanded self-test; adjustable ramp rates;
Ethernet-based PC control; comprehensive user administration tool;
Protocol Wizard
Program memory Total capacity of 350 programs à 6 steps in up to 90 user directories
Language English, German, Chinese
Display 7" color touchscreen
Automatic restart after power failure Yes
High Performance Smart Lid (HPSL) technology Yes
Power consumption 850 watt
Operating voltage 100, 115, 230 Volt, 50-60 Hz
Noise emissions Very low, max. 45 dB
Interfaces2 USB A, Ethernet
Environmental conditions 15 °C to 35 °C, 70 % air humidity, max. 2,000 m above sea level
Dimensions (WxDxH) 277 mm x 457 mm x 264 mm
Dimensions (WxDxH) with open lid 277 mm x 457 mm x 414 mm
Weight 15 kg
2
The network cable must be at least performance class Cat 5e and the cable configuration has to be STP
19 Declaration of conformity
20 Index
Auto restart 116 Heated lid temperature 47
Backup 108 Loop 50
Block module exchange 23 Name 47
Block status 36 New 45
Care 119 Pause 73
Cleaning the housing 119 Preheating 48
Configuration 85 Ramp rate 52
Declaration of conformity 131 Save 60
Decontamination 120 Spreadsheet programming 44
Disinfection instructions 120 Start 63
Display Stop 75
Run 69 Temperature increment 51
Disposal 123 Template 45
Documentation 99 Time increment 51
Error log file 104 Program preview 64
Extended self-test 103 Protocol Wizard 77
Power on logfile 102 Quick reference guide 112
Run log file 100 Quick start function 37
Emergency 14 Safety instructions 11
Firmware update 120 Settings
Heated lid 24, 117 Automatic user logout 88
Home screen 35 Beeper 89
Icons 28 Date and time 87
Incubation 40, 42 Display brightness 91
Liability 10 Factory settings 93
Login 34 Network 89
Login screen 33 Screen calibration 92
Maintenance 119 User management 93
Manual Step
Symbols 9 Skip 75
Operating voltage 20 Technical specifications
Packaging 122 Data for all models 130
Password Model specific data 124
Administrator 34 Tools 85
Forgotten 116 Contact 110
User 96 Cycler Info 109
Personnel 12 Extended self-test 104
Power on self-test 33 Service info file 107
Program Touch screen
Continue 75 Calibration 92
Copy 65 User management 93
Copy all 66 Create a user 96
Cycles 50 Delete user 99
Delete 67 Edit user settings 96
Delete all 68 Switch on or off 93
Editing steps 49 User properties 96
Gradient 52 User rights 94
Graphical programming 44 Warranty 10