1551A Ex/1552A Ex Intrinsically Safe "Stik" Thermometer: The New "Gold Standard" of Industrial Temperature Calibration
1551A Ex/1552A Ex Intrinsically Safe "Stik" Thermometer: The New "Gold Standard" of Industrial Temperature Calibration
1551A Ex/1552A Ex Intrinsically Safe "Stik" Thermometer: The New "Gold Standard" of Industrial Temperature Calibration
Features at a glance
Accuracy of 0.05 C ( 0.09 F) over full range Intrinsically safe (ATEX and IECEx compliant) Two models to choose from (-50 C to 160 C or -80 C to 300 C) User-configurable temperature trend/stability indicator
2
Display temperature in C or F Optional data logging to internal memory 300-hour battery life Percent battery-life and low-battery indicator NVLAP-accredited, NIST-traceable calibration included
1 Industrial reference thermometers available In order to minimize sensor on the market today, such as mercury-in-glass drift, the probe should always be protected thermometers (or ASTM thermometers) and from mechanical shock. portable electronic thermometers are useful, but both come with inherent problems. Although accurate and repeatable, mercury thermometers are fragile. The risk of a mercury spill poses potential hazards to the environment and to the health of personnel. Many of the United States and European Union countries have already banned their use in industrial applications. Some organizations have replaced their mercury thermometers with more durable portable electronic thermometers, only to discover that the RTD probes lack the repeatability and dependability required of a legitimate reference thermometer.
The digital display and probe are fixed together as a system. The hinge allows the display to rotate 90 to easily view the measurement from any angle. Easily power up with the press of a button and after a short selftest, start making measurements within seconds. The power key is slightly recessed inside the plastic molded case to protect against accidental power off. Max/Min/Trend/Up key allows you to see the maximum or minimum temperature since the most recent power up. The trend function tells you the positive or negative rate of temperature change in degrees per minute.
The Setup/Down key lets you step through submenus and adjust field values down. The Enter/Backlight key allows you to accept changes while in the setup menus or turn on the backlight. User-selectable resolution (0.1, 0.01, 0.001) Temperature displayed in C or F. The stability/trend indicator tells you if your temperature source is trending upward or downward (pictured trending up). A flat bar indicates that the temperature source and thermometer are stable within the user-selected stability window (0.01, 0.1, 1.0, 10.0).
6 7 8
5 5
1551A Ex -50 C to 160 C (-58 F to 320 F) C, F 100 thin-film RTD 10 mm 3 mm (0.1 in) 7 cm (2.8 in) SST
4.8 mm diameter probe (3/16 in): 14 seconds 6.35 mm diameter probe (1/4 in): 21 seconds 0.01 C Selectable: 0.1, 0.01, 0.001 (default 0.01) Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds) -10 C to 50 C (14 F to 122 F) -20 C to 60 C (-4 F to 140 F) 0 % to 95 % RH, non-condensing Up to 10,000 time-stamped readings stored to internal memory 2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds; 2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10) RS-232 stereo jack (access calibration parameters only) None Three AAA batteries, typical battery life of 300 hours without LCD backlight EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class B IP50 114 mm x 57 mm x 25 mm (4.5 in x 2.25 in x 1.0 in) 196 g (6.9 oz) NVLAP-accredited, NIST-traceable CVD ITS-90
Per ASTM E 644 See ordering information for optional data logging configurations
6 6
Ordering information
1551A Ex
1551A-9 Thermometer, Fixed RTD, -50 C to 160 C, 4.8 mm x 229 mm (3/16 in x 9 in) Thermometer, Fixed RTD, -50 C to 160 C, 6.35 mm x 305 mm (1/4 in x 12 in) Thermometer, Fixed RTD, -50 C to 160 C, 6.35 mm x 508 mm (1/4 in x 20 in)
1552A Ex
1552A-12 Thermometer, Fixed PRT, -80 C to 300 C, 6.35 mm x 305 mm (1/4 in x 12 in) Thermometer, Fixed PRT, -80 C to 300 C, 6.35 mm x 305 mm (1/4 in x 12 in) with Datalog
1551A-12
1552A-12-DL
1551A-20
Accessories
1551-CASE 1551-CBL Case, 1551-2, Carrying Cable, RS-232 155X
Standard accessories include: NVLAP-accredited report of calibration, Users Guide on CD-ROM, three AAA batteries Note: Model number appended with a -9, -12, -20 indicates probe sheath length in inches. All probe diameters are 6.35 mm (1/4 in) with the exception of the 1551A-9, which is 4.8 mm diameter (3/16 in).
1551A-12-DL
1551A-20-DL
7 7
Fluke Calibration.
Fluke Calibration PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
2010 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 11/2010 3947275A B-EN-N Pub-ID 11698-eng Modification of this document is not permitted without written permission from Fluke Corporation.