TSHR 6000 User Manual V1.3
TSHR 6000 User Manual V1.3
TSHR 6000 User Manual V1.3
TSHR International
Strickledeweg 44
3044 EK Rotterdam
The Netherlands
www.tshrinternational.com
TSHR International | 6000 Series Systems - User Manual | v1.3
COPYRIGHT
TSHR International B.V. provides this document to its customers with a product purchase to use in the product operation.
This document is copyright protected and any reproduction of the whole or any part of this document is strictly
prohibited, except with the written authorization of TSHR International B.V.
The contents of this document are subject to change without notice. All technical information in this document is
for reference purposes only. System configurations and specifications in this document supersede all previous
information received by the purchaser.
TSHR International B.V. makes no representations that this document is complete, accurate or error-free and assumes no
responsibility and will not be liable for any errors, omissions, damage or loss that might result from any use of this
document, even if the information in the document is followed properly.
This document is not part of any sales contract between TSHR International B.V. and a purchaser. This document
shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of Sale shall
govern all conflicting information between the two documents.
Save to the extent permitted by law, the customer hereby undertakes not to alter or modify the whole or any part of
the software incorporated in the products in any way whatever nor permit the whole or any part of the software to be
combined with or become incorporated in any other programs nor decompile disassemble or reverse engineer the same
nor attempt to do any other such things.
In no event shall TSHR International B.V. have any obligation to repair or replace in accordance with any warranties
given with the product in whole or in part, as the result of:
• Causes external to the software such as, but not limited to, power failure or electrical power surges
• Use of the software in combination with equipment or software not supplied by TSHR International B.V. or with
which it is incompatible.
The warranty shall not apply to any software which the Customer alters or modifies.
TSHR International B.V. does not warrant that the software is error-free, will accomplish any particular result or is fit for
any particular purpose of or intended use by the customer and it is for the customer to satisfy themselves that the
software is so fit.
The photographs/illustrations used in this manual can differ from the actual module and set-up.
2
TSHR International | 6000 Series Systems - User Manual | v1.3
All photographs, drawings and diagrams are for illustrative purposes only. The
All efforts have been made to ensure the accuracy of the contents of this manual. However, should any errors be
detected then TSHR International B.V. would appreciate being informed.
The above notwithstanding TSHR International B.V. can assume no responsibility for any errors in this manual or
their consequences.
3
TSHR International | 6000 Series Systems - User Manual | v1.3
REGULATORY COMPLIANCE
TSHR International performs complete testing and evaluation of its products to ensure full compliance with applicable
domestic and international regulations. When the system is delivered to you, it meets all pertinent electromagnetic
compatibility (EMC) and safety standards as described below.
Changes that you make to your system may void compliance with one or more of these EMC and safety standards.
Changes to your system include replacing a part or adding components, options, or peripherals not specifically
authorized and qualified by TSHR International To ensure continued compliance with EMC and safety standards,
replacement parts and additional components, options, and peripherals must be ordered from TSHR International or
one of its authorized representatives.
CAUTION! Read and understand the various precautionary notes, signs, and symbols contained inside this
manual pertaining to the safe use and operation of this product before using the device.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment
(WEEE) Directive 2002/96/EC. It is marked with the following symbol:
TSHR International has contracted with one or more recycling or disposal companies in each European Union (EU)
Member State, and these companies should dispose of or recycle this product.
See www.tshrinternational.com for further information on TSHR International’s compliance with these Directives
and the recyclers in your country.
4
TSHR International | 6000 Series Systems - User Manual | v1.3
Inhoud
CHAPTER 1 GENERAL SAFETY INSTRUCTIONS ........................................................................................................................... 8
1.1 OPERATING MANUAL ............................................................................................................................................................... 9
1.2 INTRODUCTION TO THE 6000-SERIES ................................................................................................................................... 9
1.3 SERVICE AND MAINTENANCE ............................................................................................................................................. 10
1.4 SAFETY MARKS ON THE EQUIPMENT ................................................................................................................................ 11
1.5 SAFETY RULES FOR THE OPERATOR .................................................................................................................................. 12
1.6 OPERATIONAL COMPLIANCE RULES ................................................................................................................................. 13
1.7 WARNING SIGNS! ..................................................................................................................................................................... 14
Chapter 2 CONNECTIONS AND SPECIFICATIONS ......................................................................................................................... 15
2.1 FLOWPATH COULOMETRIC DETECTION ............................................................................................................................. 16
2.2 COULOMETRIC TITRATION DETECTORS............................................................................................................................ 17
2.3 PRINCIPLE OF HALIDE ANALYSIS ........................................................................................................................................ 17
2.4 PRINCIPLE OF SULFUR ANALYSIS ....................................................................................................................................... 20
Chapter 3 6000-SERIES START UP ..................................................................................................................................................... 24
3.1 INTRODUCTION TO THE 6000 SERIES SYSTEMS ............................................................................................................... 25
3.2 OVERVIEW TX6000 .................................................................................................................................................................. 26
TOP VIEW ..................................................................................................................................................................................... 27
BOTTOM VIEW.............................................................................................................................................................................. 1
3.3 TX6000 CONNECTIONS ............................................................................................................................................................. 2
3.4 GLASS PARTS USED WITH COULOMETRIC DETECTION (TX) ............................................................................................ 5
3.5 LIQUIDS SAMPLE INTRODUCTION MODULE ...................................................................................................................... 6
3.6 SOLIDS SAMPLE INTRODUCTION MODULE ........................................................................................................................ 7
3.7 FURNACE ..................................................................................................................................................................................... 8
3.8 TURBO COMBUSTION TUBE .................................................................................................................................................... 9
3.9 CONNECTION FURNACE SCRUBBER (CFS)............................................................................................................................ 9
3.10 OUTLET CONNECTION............................................................................................................................................................ 9
3.11 SULFURIC ACID SCRUBBER ................................................................................................................................................ 10
3.12 TITRATION CELLS ................................................................................................................................................................. 10
CHAPTER 4 TOTAL SULPHUR DETECTION (TS) .......................................................................................................................... 11
4.1 FLOWPATH TOTAL SULFUR DETECTION ............................................................................................................................ 37
4.2 TOTAL SULFUR DETECTION (TS) ........................................................................................................................................... 38
CHAPTER 5 CONNECTIONS AND SPECIFICATIONS TS6000 ...................................................................................................... 40
5.1 TS6000 CONNECTIONS ............................................................................................................................................................ 41
5.2 GLASS PARTS USED WITH UV-F DETECTION .................................................................................................................... 44
5.3 CLAMPS ...................................................................................................................................................................................... 44
5.4 THE PERMAPURE DRYER ....................................................................................................................................................... 45
5.5 THE GLASS FIBER FILTER ...................................................................................................................................................... 47
5.6 THE SULFUR DETECTOR ........................................................................................................................................................ 48
(TS-P-UV-F) ...................................................................................................................................................................................... 48
CHAPTER 6 TOTAL NITROGEN DETECTION (TN) ....................................................................................................................... 49
5
TSHR International | 6000 Series Systems - User Manual | v1.3
6.1 FLOWPATH TOTAL NITROGEN DETECTION ....................................................................................................................... 50
6.2 TOTAL NITROGEN DETECTION (TN) ..................................................................................................................................... 51
CHAPTER 7 CONNECTIONS AND SPECIFICATIONS TN6000 ..................................................................................................... 52
7.1 TN6000 CONNECTIONS ........................................................................................................................................................... 53
7.2 GLASS PARTS USED WITH CHEMILUMINESCENCE DETECTION .................................................................................. 55
7.3 CLAMPS ...................................................................................................................................................................................... 55
7.4 THE PERMAPURE DRYER ....................................................................................................................................................... 56
7.5 THE GLASS FIBER FILTER ...................................................................................................................................................... 58
7.6 THE MOLYBDENE CONVERTER ............................................................................................................................................ 59
7.7 THE NITROGEN DETECTOR (TN-CLD) .................................................................................................................................... 60
7.8 PHOTO MULTIPLIER TUBE (PMT) ......................................................................................................................................... 61
7.9 PRESSURE CONTROL VALVE ................................................................................................................................................ 61
7.10 THE OZONE KILLER .............................................................................................................................................................. 62
CHAPTER 8 HR7000 AUTOSAMPLER .............................................................................................................................................. 63
8.1 HR7000 AUTOSAMPLER .......................................................................................................................................................... 64
8.2 HR7000 CONNECTIONS ........................................................................................................................................................... 66
CHAPTER 9 GLASSWARE MAINTENANCE ................................................................................................................................... 67
9.1 VISIBLE INSPECTION OF GLASSWARE ............................................................................................................................... 68
CHAPTER 10 INSTALLATION GLASSWARE TX6000 ................................................................................................................... 70
10.1 INSTALLATION GLASSWARE TX6000 ............................................................................................................................... 71
10.2 GAS SCRUBBERS AND CONNECTIONS ............................................................................................................................. 72
10.3 TITRATION CELL DETECTOR .............................................................................................................................................. 74
CHAPTER 11 TX6000 START UP ....................................................................................................................................................... 76
11.1 TX6000 START UP.................................................................................................................................................................... 77
11.2 MATERIAL SAFETY DATA SHEETS (MSDS) ......................................................................................................................... 77
11.3 PREPARATION OF REAGENTS ............................................................................................................................................ 78
11.4 PREPARATION OF SAMPLES ............................................................................................................................................... 79
11.5 POWERING UP THE SYSTEM ............................................................................................................................................... 83
11.6 PREPARING THE TITRATION CELL (CHLORIDE) ................................................................................................................. 84
11.7 PREPARING THE TITRATION CELL (SULFUR) ..................................................................................................................... 86
11.8 FILLING THE SCRUBBER ...................................................................................................................................................... 88
11.9 GAS LEAKAGE TEST.............................................................................................................................................................. 90
11.10 TX6000 NORMAL OPERATION ........................................................................................................................................... 91
11.11 TX6000 STANDBY MODE ..................................................................................................................................................... 92
11.12 TX6000 SHUT DOWN ............................................................................................................................................................ 92
CHAPTER 12 TS / TN6000 start up ...................................................................................................................................................... 93
12.1 MATERIAL SAFETY DATA SHEETS (MSDS) ......................................................................................................................... 94
12.2 PREPARING STANDARDS ..................................................................................................................................................... 94
12.3 PERMAPURE CONNECTION ................................................................................................................................................. 95
12.4 MODULE CONNECTION ........................................................................................................................................................ 95
12.5 POWERING THE SYSTEM ..................................................................................................................................................... 96
12.6 STANDARD SOFTWARE SETTINGS ..................................................................................................................................... 96
CHAPTER 13 RUNNING AN ANALYSIS (general) ........................................................................................................................... 97
6
TSHR International | 6000 Series Systems - User Manual | v1.3
13.1 RUNNING AN ANALYSIS (GENERAL) ................................................................................................................................. 98
13.2 6000-SERIES STANDBY MODE ........................................................................................................................................... 103
13.3 6000-SERIES SHUT DOWN................................................................................................................................................... 103
13.4 BASIC SYSTEM SETTINGS .................................................................................................................................................. 104
CHAPTER 14 TROUBLESHOOTING TX6000 ................................................................................................................................. 105
TROUBLESHOOTING TX6000 ...................................................................................................................................................... 105
CHAPTER 15 TROUBLESHOOTING TS6000/TN6000 ................................................................................................................. 111
CHAPTER 16 TECHNICAL SPECIFICATIONS TS6000.................................................................... Error! Bookmark not defined.
CHAPTER 17 TECHNICAL SPECIFICATIONS TN6000 ................................................................... Error! Bookmark not defined.
CHAPTER 18 PRODUCT SPECIFICATIONS TS / TN / TX6000-SERIES ........................................ Error! Bookmark not defined.
CHAPTER 19 CONSUMABLES ........................................................................................................................................................ 123
7
TSHR International | 6000 Series Systems - User Manual | v1.3
Chapter 1
General safety instructions
8
TSHR International | 6000 Series Systems - User Manual | v1.3
Before using the 6000-series, carefully read through this operating manual. It is essential that any operator operating
the instrument or who come into direct contact with it must be fully aware of all its associated hazards. The operator
should closely follow the instructions contained in this operating manual in order to ensure the proper operation of the
instrument.
This manual makes use of warning signs in order to highlight remarks and instructions that require special attention. These
warning signs are listed below. The section “General Safety Instructions” details the warning signs appearing on the
outside and inside of the apparatus. All directions relating to these warning signs must be complied with at all times.
Should you encounter problems or have questions which are not discussed in this manual, please do not hesitate to
contact us.
TSHR International BV
Strickledeweg 44
3044 EK Rotterdam, The Netherlands
Phone: 0031 (0)10 2457701. Our telephone service is available from 09:00 to 17:00 (CET).
Email: [email protected]
ASTM D5453, ASTM D4629, ASTM D7183, ASTM D7184, ASTM D6667, ASTM D6069, EN ISO 20846, EN ISO
15486, UOP 936.
The 6000-series operates on the principle of oxidative combustion followed by pulsed UV fluorescence
(sulphur) and reduced pressure chemiluminescence (Nitrogen) detection.
Gases and LPG are introduced into the analyser using the GM 7000 gas/LPG sample introduction module. Liquids
are introduced into the analyser using the HR7000 auto sampler.
The analyser is unique in that it can directly inject liquids of both low and high boiling point samples into the furnace
without the requirement for boat introduction. In order to facilitate this introduction, the viscosity of the sample must
be reduced to a level whereby it can be easily injected via the auto sampler. For light and medium distillates such as
naphtha, formulated gasoline and diesel, no dilution is required and the sample can be directly injected. For heavy
hydrocarbons such as fuel oil (bunker oil), gas oil, crude oil and vacuum gas oil, the sample must be diluted. The
dilution factor depends on the sample viscosity and Nitrogen/sulphur concentration but typically a dilution of 50-
100x is recommended.
The ability to analyse heavy hydrocarbons without the use of boat introduction reduces the analysis time from
typically 10-12 minutes per sample repeat to 2-3 minutes per sample repeat. In addition, there is no requirement to
change the sample introduction module because of the sample type and light and heavy hydrocarbon samples can be
analysed in sequence with the benefit of full automation.
The 6000-series has several main components. These are:
Sample introduction
• Liquids Module (syringe injector) (light distillates, middle distillates)
9
TSHR International | 6000 Series Systems - User Manual | v1.3
• Solids Module (boat injection) (heavy distillates, solids like coal, catalyst)
• GM 7000 adaptor (gasses and LPGs)
• Horizontal furnace
Gas conditioning
• Membrane dryer
• Mass flow controllers
Detection
• Ozone generator for total Nitrogen detection
• Total Nitrogen detector (NO)
• Total Sulphur detector (SO2)
The 6000-series is controlled by an advanced software suite called Athena®. Athena is designed for use with Microsoft
Windows™ 7, 8, 8.1 and 10 (32 and 64 bit). Athena contains pre-loaded methods which ensure complete regulatory
method compliance. In addition to pre-loaded methods, users can create their own methods or edit existing methods
using intuitive method development tools.
Included with Athena is a module for evaluation of sample and calibration worklists.
10
TSHR International | 6000 Series Systems - User Manual | v1.3
Before using the 6000-series, carefully read through this operating manual. It is essential that any persons operating
the module or who come into direct contact with it must be fully aware of al its associated hazards. Always carry out
a LMRA (Last Minute Risk Analysis) before starting with the analyser.
We use the following safety colours and signs on the analytical instrument. These are based on the ISO 3864
international standards for safety colours and signs. These signs indicate those locations that require special attention
and/or constitute a danger to personnel or the instrument.
All personnel coming into either direct or indirect contact with the 6000-series should be fully acquainted with these
international safety-markings.
The signs and accompanying texts appearing on or inside the 6000-series are shown below.
IMPORTANT!
Any precautions and/or conditions imposed by national laws or company safety
regulations applicable to the space occupied by the 6000-series must be followed.
The management of your organization is responsible for communicating these
precautions and conditions and ensuring compliance. In the event of doubt, consult
your supervisor.
11
TSHR International | 6000 Series Systems - User Manual | v1.3
The apparatus must be installed by qualified technicians and factory trained Field Service Engineers in
accordance with the instructions provided by supplier. You should
WARNING! Before turning on the 6000-series for the first time, the voltage setting must be checked. The
small plate beneath the power lead should be positioned with the 230V indica- tion in the normal position
(not upside down) and is read from left to right when the operator is facing the back of the instrument. The
115 V indicators will be upside down. Should you need to switch from a 230 V to a 115 Volts supply, pull
the plate away from the instrument so that the prongs of the plug are fully withdrawn from the instrument
and re-insert the other way up.
ELECTRICAL HAZARD! Do not open the cabinet when the power cable is connected. Remove the
power cable from the wall socket and disconnect the cable from the instrument completely first. An
incorrectly grounded or ungrounded apparatus may cause electric shock.
SPECIAL ATTENTION! After supplying power to the instrument and starting the program, the furnaces
and heating elements will rise to the standard temperatures required by the con- nected module
(Initialization procedure). When no module is connected, the apparatus will go into standby mode i.e. no
furnace control.
12
TSHR International | 6000 Series Systems - User Manual | v1.3
• Refer to the manual in case of doubt about a procedure prior to starting up, running samples or shutting
down.
• Wear protective clothing, masks and spectacles in accordance with the laboratory requirements for
carrying out the measurements (see local PPE HSE regulations).
• Make sure that the apparatus is grounded as specified. Use a grounded power connector.
• The apparatus should be set up in such a way that it does not constitute a danger to the user during the
course of the work.
• Make sure that the doors of the analyser can be freely opened and closed without obstructions.
• Make regular checks on the connecting cables and gas piping for possible damage and leaks.
• The apparatus may only be used for the purpose for which it is intended and must be operated
by qualified personnel only.
• Never carry out any maintenance on the 6000-series which is not described in this manual.
• Never remove any covers, doors or panels which are fastened into position by means of screws.
• Keep all the analytical and other compartments provided with opening flaps shut during the
performance of analyses when this is required.
• Always use authorized TSHR materials and accessories. Unapproved materials and accessories may
cause the malfunction of the instrument or give rise to hazardous situations. Never use materials for a
purpose other than that for which they were originally intended.
• The use of parts, which are not manufactured/approved by us, could result into exclusion of warranty
arrangements.
• In case of a malfunction that could endanger safety, turn off the instrument immediately.
• Never sprinkle the housing with water or other liquids. To clean, wipe with a moist cloth.
• Never close off the ventilation openings and check them regularly to ensure they are free from blockages
such as tissues.
• Never place combustible materials on top of the analyser e.g. tissues, solvents, etc.
13
TSHR International | 6000 Series Systems - User Manual | v1.3
SYMBOL CAUTION
ELECTRIC SHOCK!
This instrument uses high voltages that can cause personal injury. Before
servicing, shut down the instrument and disconnect the instrument from line
power. Keep the top cover on while operating the instrument. Do not
remove protective covers from PCBs.
HEAT!
Before servicing the instrument, allow any heated
components to cool.
FIRE!
Use care when operating the system in the presence of
flammable gases.
GENERAL HAZARD!
A hazard is present that is not included in the above categories. Also,
this symbol appears on the instrument to refer the user to instructions in
this manual.
CRUSH HAZARD!
Keep hands clear during operation!
14
TSHR International | 6000 Series Systems - User Manual | v1.3
Chapter 2
Connections and specifications
15
TSHR International | 6000 Series Systems - User Manual | v1.3
2.1 FLOWPATH COULOMETRIC DETECTION
1. A sample is introduced (Liquids or Solids) into the is removed. From there it goes to the sulfuric acid
furnace a boat driver or injection syringe. During scrubber.
the transfer into the furnace the sample is carried 4. The gas flows into the titration cell where the
with an Argon flow. The volatile components are reaction takes place. Silver-ions (Ag+) react with
transported directly into the furnace. halide-ions (X- ).
2. In the furnace the sample is oxidized with a 5. A signal appears on the PC screen and the area
flow of O2. generated by this signal is integrated.
The software (Athena) calculates the actual
3. The gas flow from the furnace is carried
concentration and stores the result.
through a heated connection where the water
1. LIQUIDS
Titration curve
PC
Module
O2 connector
COULOMETER
5 Ar
FURNACE
ZONE II ZONE I
LIQUIDS MODULE
O2 Ar (backside)
CFS
ANALYTE
2
4 3 1
Sample Liquids
LIQUIDS MODULE
SCRUBBER
TITRATION CELL
Titration curve
PC
2. SOLIDS
COULOMETER
5
FURNACE
ZONE II ZONE I
SOLIDS MODULE
CFS ANALYTE
O2
Ar
2
4 3 1
Sample Solids
SCRUBBER Boat driver
TITRATION CELL
16
TSHR International | 6000 Series Systems - User Manual | v1.3
Ei-r is governed by the concentration of silver ions in solution as stated by Nernst's Law. A change in
the concentration of silver ions will therefore give rise to a change in Ei-r. The concentration of silver ions changes when
halide ions enter the solution. The halide ions tend to remove silver ions from the solution by the formation of silver
halides, which are notoriously insoluble. This can for example occur by the reaction:
When Ei-r changes, the coulometer passes current (1) through the generator circuit, thus forming silver ions at the anode.
Hydrogen is then formed at the cathode. The silver ions react with halide ions until these have disappeared or, in other
words, until the concentration of silver ions has returned to the same level as the initial concentration of roughly 10-7 M.
The solubility product indicates the relationship between the concentrations of silver ions and halide ions in a saturated
solution. For silver chloride this is:
As long as Ei-r continues to change current will pass through the generator circuit and silver ions will be generated. As E i-r
approaches its initial potential the current will fall off. Once Ei-r has reached its initial value again, i.e. when all the halide
ions have been converted, the generator curve (1) will again be zero.
There is a difference in potential between anode and cathode of:
E i-r + Ebias
Ebias is a fixed potential of -315 mV that serves, among other things to compensate the electrical resistance
of the cell. The total of Ei-r + Ebias is equal to about 0.5 V.
Coulometry is a form of titration. The substance to be titrated is converted by means of an electrical charge. It should
therefore be remembered that it is not the substance itself, which is being titrated but an electrical charge.
17
TSHR International | 6000 Series Systems - User Manual | v1.3
The figure below shows a TX titration curve. Initially the current increases.
The speed with which this takes place depends on the speed with which the hydrogen halides enter the titration cell,
or in other words the way in which the analysis sample is introduced. After reaching a peak, the current strength
returns to zero (speed of titration depends on gain and cell condition,
see figure below).
Example
titration
curve
The quantity of Ag+ ions which are generated by the flow of the current through the cell and
consequently the quantity of deposited halide ions, behave in accordance with Faraday’s Law:
t
1 Q
m= ∫ I . dt n. f
n. f 0
where:
Integrating the current over time provides the quantity of charge used for titrating the halide.
The quantity of charge (area counts) which has been titrated is shown in ىC in the queue manager.
The quantity of halide present in the cell will then be automatically calculated. The
calculations are carried out automatically by the Athena software.
18
TSHR International | 6000 Series Systems - User Manual | v1.3
AI G
R
C
where:
O = Outlet
G = Gas inlet tube
ACID! The titration cell contains a stabilized electrolyte solution of 75% acetic acid in water. The
concentration of silver ions in solution is maintained at a constant level of about 10 -7 M.
Where:
COULOMETER
V = Voltmeter
Amp = Currentmeter/Ampmeter
RΩ = Variable resistance
R = Reference electrode
I = Indicator electrode
A = Anode Amp
C = Cathode RΩ
R I A C
19
TSHR International | 6000 Series Systems - User Manual | v1.3
Measuring pair:
• Anode A (silver).
• Cathode C (platinum).
• The measuring circuit - Here the difference in potential, Er-i, between the indicator and reference electrodes
is measured.
• The generator circuit - When there is a change in Er-i, the coulometer passes a current through the
anode/cathode circuit. Ag ions are generated at the anode as a result.
The concentration of I- changes when sulphur dioxide dissolves in the electrolyte. Sulphur
dioxide reacts with I3- as follows:
SO + I -+ 6 H 0 SO 2- + 3 I- + 4 H O+
2 3 2 4 3
When Ei-r changes, the coulometer passes current (1) through the generator circuit. This results in the
conversion of I- into I2 at the platinum anode, followed by the formation of I3- as indicated below:
2 I- I2 + 2e
I 2 + I- I3-
This process will continue until Ei-r no longer changes, i.e. until the concentration of I 3 has returned- to the value it
had before the addition of SO2. The generator current will then be equal to zero and all the SO2 will have
been converted into sulphate.
20
TSHR International | 6000 Series Systems - User Manual | v1.3
Here too the difference in potential between anode and cathode is:
The quantity of I- converted can be calculated from the charge needed for the C
Where:
R C A/I G
ACID!
The titration cell contains a solution of 0.1% KI, 0.1% NaN3 and 0.5 % acetic acid in de-ionized
water. The NaN3 (sodium azide) is added to prevent possible disturbances to the measurement
caused by nitrous oxides forming in the cell.
TOXIC!
Sodium Azide is extremely toxic. Read the instructions of the chemical safety sheet
carefully and handle accordingly.
21
TSHR International | 6000 Series Systems - User Manual | v1.3
Where:
COULOMETER
V = Voltmeter
Amp = Ampmeter
RΩ = Variable resistance Amp
R = Reference electrode RΩ
I = Indicator electrode
V
A = Anode
C = Cathode
Measuring pair:
Generation pair:
• Anode A (platinum)
• Cathode C (platinum)
The measuring circuit - Here the difference in potential E, between the indicator and reference electrodes
is measured.
The generator circuit - When there is a change in El-R the coulometer passes a current through the
anode/cathode circuit. At the anode, I- ions are converted into I2 at the anode.
23
TSHR International | 6000 Series Systems - User Manual | v1.3
Chapter 3
6000-series start up
24
TSHR International | 6000 Series Systems - User Manual | v1.3
The 6000-series is designed for fast and efficient analysis of total (organic) halogens (X), Sulphur (S) and/or
Nitrogen (N). It supports the use of a number of different modules, which are used for the introduction of
various sample matrices into the furnace.
25
TSHR International | 6000 Series Systems - User Manual | v1.3
FRONT VIEW
FURNACE
26
TSHR International | 6000 Series Systems - User Manual | v1.3
TOP VIEW
MODULE CONNECTOR
27
TSHR International | 6000 Series Systems - User Manual | v1.3
POWER SWITCH
BOTTOM VIEW
CABLE GROMMETS
COOLING FANS
VIBRATION DAMPERS
1
TSHR International | 6000 Series Systems - User Manual | v1.3
Power Connection
A The power cord has to be connected to the 6000-series.
Communication Connection
B The USB cable has to be connected to the 6000-series on the USB-connection. The other
side of this cable has to be connected on the USB-port of the computer. The TX600 has
two Ethernet Ports.
Gas Connections
C Gases must be provided by the customer and need to be
PREFERRED PRESSURE
SETTINGS.
connected to the 6000-series. The gas specifications For Argon/Helium
for Argon/Helium and Oxygen are: advised input pressure is 1-3 bar.
For Oxygen advised input
• Argon/Helium (Ar) (He) 99.998% pressure is 1-3 bar.
• Oxygen (02) 99.6%
• Maximum pressure 3 bar, minimum 1 bar
• Gas connection 1/8” Swagelok
BACK VIEW
230V
US
B
A B INPUT
P-max 3 bar
C ARGON
OXYGE
N2
2
TSHR International | 6000 Series Systems - User Manual | v1.3
MODULE
CONNECTOR
Liquids Module connection
O2
Ar
3
TSHR International | 6000 Series Systems - User Manual | v1.3
Connector
Sulfuric Acid Scrubber
connection
4
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Syringe
2. Sample introduction module (liquids in
combination with the HR7000 Autosampler)
3. Sample introduction module (liquids)
4. Sample introduction module (solids) 1
5. Furnace tube
6. Connection Furnace Scrubber
2
(CFS)
7. Sulfuric Acid Scrubber
8. Outlet Connection
9. Outlet Connection (ball-joint)
10. Halogen Titration Cell
11. Sulfur Titration Cell
5 4
8
7
9 6
11 10
5
TSHR International | 6000 Series Systems - User Manual | v1.3
analysed.
The two liquids modules used are:
The liquid module has two connections in which the
carrier gases are injected.
Argon (red) and Oxygen (blue). 1. Liquids module in combination with the HR7000
Autosampler. (On top the sample is directly injected
in the liquids module by the HR7000 autosampler
which holds the syringe).
1 Sample
injected with
syringe
Sample
injected with
syringe
Ar entrance
O2 entrance
Ar entrance
O2 entrance
6
TSHR International | 6000 Series Systems - User Manual | v1.3
FURNACE
ZONE II ZONE I
SOLIDS MODULE
ANALYTE
O2
Ar
2
1
Sample Solids
Boat driver
Solids module
The solids module is used for the analysis of Total
halogen (TX) or Total sulphur (TS) content in solid
samples. The module is meant for the analysis of halogen
and sulphur compounds in solid samples and liquids
containing large molecular compounds with boiling
temperatures above 450°C. These are carried into the
furnace by a flow of Argon and burned in Oxygen.
1 2
Ar entrance
O2 entrance
7
TSHR International | 6000 Series Systems - User Manual | v1.3
3.7 FURNACE
The TX6000 uses a furnace to combust the samples. According to ASTM requirements, the furnace is divided into two
temperature zones. The temperature in each zone can be adjusted individually, up to 1250°C, however operating
temperature is between 800°C/1000°C. The maximum temperature
difference between the zones is approximately 250°C.
TEMPERATURES MODULES
WARNING! The combustion tube, out- let, • Sample introduction module 500 °C
modules, joints, etc. become very hot during • Furnace / Furnace tube 1000 °C
the operation of the 6000-series.
• Connection furnace scrubber 325 °C
Carelessness may result in burns. Avoid contact with bare
hands. Use special protective gloves and/or tools for • Outlet connections 320 °C
“hot tube removal”. The safest way to remove the
combustion tube is to first allow it to cool down to room
temperature.
8
TSHR International | 6000 Series Systems - User Manual | v1.3
Turbo O2
9
TSHR International | 6000 Series Systems - User Manual | v1.3
The sulfuric acid scrubber is designed with a scrubber buffer which can be
filled with 97-98% sulfuric acid. The sulfuric acid dries the gasses and
removes the possible soot particles.
It has two stopcocks which can be removed. To shut the stopcock of the
scrubber, the red handles must be in horizontal position.
The upper glass stopper prevents the sulfuric acid from splashing.
Stopcock
The titration cells are designed for long term automatic analyses. The signals from the titration cell go directly to the IOC-
board in the computer, where the data is collected and processed.
The Halogen
The Sulphur Titration Cell
Titration Cell
10
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 4
Total Sulphur detection (TS)
TSHR International | 6000 Series Systems - User Manual | v1.3
1. The sample is introduced into the furnace by 6. The fluorescent light is detected by the Photo
the HR7000 autosampler. Multiplier Tube during the relaxation of the
2. The sample is combusted and oxidized to form
decaying SO2* back to SO2. The PMT signal
SO2.
measured is proportional to the concentration of
3. H2O formed by combustion and present in
sulphur in the gas.
the combustion gases is removed by the
Permapure dryer. 7. The PC processes the PMT signal and
generates a graphical display.
4. Soot and undesired particles
are trapped in the filter assembly.
5. In the reaction chamber from the pulsed Xenon
source, UV light is used to excite the SO2 molecules
present in the gas flow.
ZONE II ZONE I 2
37
TSHR International | 6000 Series Systems - User Manual | v1.3
The sulphur detector (TS-P-UV-F) detects the presence of SO2 formed during combustion of the sample in the furnace.
During the combustion process, the following reaction occurs:
R-S + O SO + CO + H O
2 2 2 2
Detection is based on the principle that SO2 molecules absorb ultraviolet (UV) light and become excited, then decay to
a lower energy state emitting UV light at specific wavelengths which are separated in wave- length from the incident
wavelength. The process of absorption and fluorescence of an SO 2 molecule is depicted in the Figure below.
Intersystem
crossing
Internal &
external
conversion
The pulsed UV light is generated by a xenon discharge lamp operating at 10 Hz. The light from the Xe flash- lamp is
filtered and shaped such that only light which is sensitive to SO 2 molecules is incident on the sample gas cloud.
After excitation, the excited SO2 molecules decay to a lower energy state and UV light is emitted whose intensity is
proportional to the SO2 concentration. This light is detected by a photomultiplier tube (PMT) oriented perpendicular
to the axis of propagation of the xenon flash lamp light.
38
TSHR International | 6000 Series Systems - User Manual | v1.3
A bandpass filter, placed between the sample gas cloud and the photomultiplier tube, ensures that only the fluorescent
radiation is passed to the photomultiplier tube for detection. The emission intensity of the xenon flash lamp is monitored
by a UV photodiode. Through signal processing electronics, fluctuations in the emission intensity of the flash lamp are
compensated for and they do not affect the stability of the instrument. The amount of light detected by the PMT is
proportional to the concentration of sulphur contained in the combusted sample. The exact concentration contained
within a given sample can be determined by calibrating the instrument using samples of known concentration.
The layout of the total sulphur detector is presented in the Figure below.
Mirror Assembly
Condensing Lens
Electronics
PMT
Photodetector
Reaction
Chamber
PMT
Excited state
SO2*
SO2 SO2+hv.
UV (8.78Hz)
Ground state Lower energy state
WARNING! UV light from the Xenon flash lamp is harmful for your eyes.
During operation, do not remove the socket and do not look directly into the source.
39
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 5
Connections and specifications
TS6000
TSHR International | 6000 Series Systems - User Manual | v1.3
Power Connection
A The powercord has to be connected to the
Sulfur Module Compartment.
B Connection
230V
C Quick lock 1/4" Loop in upper
B C connector only
D
D Quick lock 1/4"Gas inlet UV-F
detector in lower position.
A
41
TSHR International | 6000 Series Systems - User Manual | v1.3
Power Connection
E The powercord has to be connected to the 6000-series.
Communication Connection
F The USB cable has to be connected to the 6000-series on the USB-connection. The other
side of this cable has to be connected on the USB-port of the computer. The TX600 has
two Ethernet Ports.
Gas Connections
G Gases must be provided by the customer and need to be PREFERRED PRESSURE
connected to the 6000-series. The gas specifications SETTINGS.
for Argon/Helium and Oxygen are: For Argon/Helium
advised input pressure is 1-3 bar.
• Argon/Helium (Ar) (He) 99.998% For Oxygen advised input
• Oxygen (02) 99.6% pressure is 1-3 bar.
• Maximum pressure 3 bar, minimum 1 bar
• Gas connection 1/8" Swagelok
230V
US
B
E F INPUT
P-max 3 bar
G ARGON
OXYGE
N2
42
TSHR International | 6000 Series Systems - User Manual | v1.3
MODULE
CONNECTOR
Liquids Module connection
O2
Ar
43
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Syringe
2. Universal Sample introduction module (liquids)
3. Sample introduction module (liquids in 1
combination with the HR7000 Autosampler)
4. Furnace tube
5. Outlet Connection
4 3
2
5.3 CLAMPS
1. Clamp fork NS 29.2 (Secures the Stopper with the Solids Module).
2. Clamp No.35 (Secures the Liquids or Solids Module with the 2
Combustion Tube).
3. Clamp No.12 (Secures Ball joints of the furnace tube with the CFS. The
Sulfuric Acid Scrubber, Outlet Connections and Titration Cells).
44
TSHR International | 6000 Series Systems - User Manual | v1.3
Conditioning
The gas flow coming from the furnace travels through a Permapure membrane dryer. The principle of the Nafion
Permapure membrane dryer is diffusion of water from the gas stream via co-polymers with sulfuric acid ligands.
SCHEMATIC EXPLANATION
OFWATER REMOVAL
OutletPermapure
Vented gasses
Inlet Permapure
Outlet Connection
45
TSHR International | 6000 Series Systems - User Manual | v1.3
To replace and connect the Permapure dryer, the following steps are necessary.
1. Connect the inlet of the Permapure to the outlet glass connector.
2. Connect the tube from gas outlet coming from the vacuum pump to the outside inlet (for drying).
3. Connect the tube from outlet of the Permapure to the 1/4" bulkhead connection going to the blue
glass fibre filter which is located under the Furnace compartment .
1 2 3
Tube to vacuum pump
Glass fibre filter
Inlet Permapure
Outlet Permapure
SPECIAL ATTTENTION!
When installing a new PermaPure dryer, care must be taken not to rotate the end pieces of the PermaPure
about the axis of the 1/4" tubing by more than 10°. Rotation of the end pieces will cause the headers of the
inner Nafion tubing to rotate resulting in permanent damage.
46
TSHR International | 6000 Series Systems - User Manual | v1.3
White housing
Blue housing
47
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Chamber inlet
2. PMT
3. Flash lamp TS
4. Coupling
5. Lens Mount Assembly
Gas connections with Sulphur detector
48
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 6
Total Nitrogen Detection (TN)
TSHR International | 6000 Series Systems - User Manual | v1.3
50
TSHR International | 6000 Series Systems - User Manual | v1.3
R-O + O2 NO + CO 2 + H 2O
The water produced in this reaction is removed by the PermaPure dryer and the NO and CO2 formed pass
through a pressure control valve on their way to the NO engine where the detection takes place. The following
sections describe the function of the detection components.
NO + O NO * + O
3 2 2
2
NO2* NO + hν. NO
O3
Ozone input
The reaction of ozone with NO forms NO2 in
an excited state which then decays by
emitting a chemiluminescent photon. The
number of chemiluminescent photons Reaction PMT
detected is proportional to the concentration Chamber
of nitrogen contained within the combusted
sample.
The exact concentration of nitrogen
contained within a given sample can again
be determined by calibrating the
instrument with samples of known
concentration.
NO NO2+hv.
PMT
Ground state Lower energystate
51
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 7
Connections and
specifications TN6000
TSHR International | 6000 Series Systems - User Manual | v1.3
Power Connection
A The powercord has to be connected to the 6000-series.
Communication Connection
B The USB cable has to be connected to the 6000-series on the USB-connection. The other
side of this cable has to be connected on the USB-port of the computer. The TX600 has
two Ethernet Ports.
Gas Connections
C Gases must be provided by the customer and need to be
PREFERRED PRESSURE
SETTINGS.
connected to the 6000-series. The gas specifications
For Argon/Helium
for Argon/Helium and Oxygen are:
advised input pressure is 1-3 bar.
For Oxygen advised input
• Argon/Helium (Ar) (He) 99.998%
• Oxygen (02) 99.6% pressure is 1-3 bar.
• Maximum pressure 3 bar, minimum 1 bar
• Gas connection 1/8" Swagelok
230V
US
B
A B INPUT
P-max 3 bar
C ARGON
OXYGE
N2
53
TSHR International | 6000 Series Systems - User Manual | v1.3
MODULE
CONNECTOR
Liquids Module connection
O2
Ar
54
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Syringe
2. Universal Sample introduction module (liquids) 1
3. Sample introduction module (liquids in
combination with the HR7000 Autosampler)
4. Furnace tube
5. Outlet Connection
4 3
2
7.3 CLAMPS
1. Clamp fork NS 29.2 (Secures the Stopper with the Solids Module).
2. Clamp No.35 (Secures the Liquids or Solids Module with the
Combustion Tube).
2
3. Clamp No.12 (Secures Ball joints of the furnace tube with the CFS. The
Sulfuric Acid Scrubber, Outlet Connections and Titration Cells).
55
TSHR International | 6000 Series Systems - User Manual | v1.3
Conditioning
The gas flow coming from the furnace travels through a PermaPure membrane dryer. The principle of the Nafion
PermaPure membrane dryer is diffusion of water from the gas stream via co-polymers with sulfuric acid ligands.
SCHEMATIC EXPLANATION
OFWATER REMOVAL
OutletPermapure
Vented gasses
Inlet PermaPure
Outlet Connection
56
TSHR International | 6000 Series Systems - User Manual | v1.3
To replace and connect the PermaPure dryer, the following steps are necessary.
1. Connect the inlet of the PermaPure to the outlet glass connector.
2. Connect the tube from gas outlet coming from the vacuum pump to the outside inlet (for drying).
3. Connect the tube from outlet of the PermaPure to the 1/4" bulkhead connection going to the blue
glass fibre filter which is located under the Furnace compartment .
1 2 3
Tube to vacuum pump
Glass fiber filter
Inlet Permapure
Outlet Permapure
SPECIAL ATTTENTION!
When installing a new PermaPure dryer, care must be taken not to rotate the end pieces of the PermaPure
about the axis of the 1/4" tubing by more than 10°. Rotation of the end pieces will cause the headers of the
inner Nafion tubing to rotate resulting in permanent damage.
57
TSHR International | 6000 Series Systems - User Manual | v1.3
White housing
Blue housing
58
TSHR International | 6000 Series Systems - User Manual | v1.3
59
TSHR International | 6000 Series Systems - User Manual | v1.3
➈
➀ ➆
➅
➄ ➂
➃
60
TSHR International | 6000 Series Systems - User Manual | v1.3
Electrical
Incoming Photon connectors
Focusing
electrode Multiplied Electrons
61
TSHR International | 6000 Series Systems - User Manual | v1.3
62
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 8
HR7000 Autosampler
TSHR International | 6000 Series Systems - User Manual | v1.3
2
3 5
3
TOP VIEW
4
3
1
3
Closed
Open 3
64
TSHR International | 6000 Series Systems - User Manual | v1.3
65
TSHR International | 6000 Series Systems - User Manual | v1.3
E 5 GC connector:
for other devices
(GC, analyzer....)
66
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 9
Glassware maintenance
TSHR International | 6000 Series Systems - User Manual | v1.3
9.1 VISIBLE INSPECTION OF GLASSWARE
All glassware should be checked at least once a month for scratches, cracks and blockages. After
inspection, a gas leak test should be performed. The frequency of inspection depends strongly on the
sample matrix. Samples containing high concentrations of salts or metals reduce the service life of quartz
and the number of inspections should be increased in this case.
Glassware maintenance
Furnace tube
Soot in the furnace tube is a cause of memory effects. Mineralization can also cause memory effects as the tube
becomes slightly porous. Soot deposited at the end of the furnace tube is relatively easy to remove.
1. Remove the clamp between the furnace tube outlet and the PermaPure dryer.
2. Remove the clamp for the “turbo Oxygen” at the furnace tube outlet.
3. Remove the clamp which connects the sample introduction module to the inlet of the furnace tube and remove
the sample introduction module.
4. Remove the stainless steel fork between the furnace tube and the sample inlet module.
5. Pull back the furnace tube halfway so that outlet tubes are inside the furnace, use the clamp nr. 35 to hold the
combustion tube in position. Set the furnace temperature to 1000°C,
it should be clean in 10 minutes.
The furnace tube can become opaque on the inside and sometimes on the outside. This is
recognized as mineralization of the furnace tube.
WARNING!
PULL BACK THE Removing the soot in the
TUBE HALFWAY combustion tube should
only be carried out by
trained Field Service
Engineers.
SET FURNACE (1000 °C)
Soot
67
TSHR International | 6000 Series Systems - User Manual | v1.3
ACID! Take the necessary PPE precautions to avoid contact with sulfuric acid.
Wear protective gloves, goggles and acid proof clothing, according safety procedures on the
laboratory. Failure to do so may result in physical injury.
1 7 8
20
50
min 0 90°C
68
TSHR International | 6000 Series Systems - User Manual | v1.3
Liquids module
During the use of the “liquids module” analysis, the inner part and the injection port can become polluted.
Cleaning procedure:
1. Replace the septum when necessary but at least once every week.
2. Reversing the connected 02 and Ar/He gas flows will burn off this pollution.
3. This can be monitored on the display (base line going up > pollution, baseline coming down again
is clean). A temporary increase of the “liquids module” temperature (500 > 600ºC) accelerates this process.
1
2
REPLACE
SEPTUM!
REPLACE
SEPTUM!
Reverse the 02
(blue tube) and
Ar/He (red tube)
clean
NOTE! Before commencing analysis, make sure the gasses have been reconnected in the correct
order!
69
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 10
Installation glassware TX6000
TSHR International | 6000 Series Systems - User Manual | v1.3
10.1 INSTALLATION GLASSWARE TX6000
This chapter will guide you through the correct installation procedure for the TX7000. Always check if
the surface and table, used for the installation, are strong enough to support the weight of the system.
Keep in mind that chemicals might be spilled, affecting the nature and structure of the support. Acids
like “Sulfuric acid” are substances used in the system.
Check for chemical resistance of the surface before final installation and if necessary take measures.
WARNING!
Before turning the TX7000 on for the first time check the voltage. The selected
voltage should match the “standard” voltage supplied to the laboratory. Unplug the
powercord from the TX7000! E.G: to switch from a 230 V supply. The readable
voltage is the selected voltage! In case of doubt, please contact your local
representative.
ELECTRICAL HAZARD!
Do not open the cabinet when the power cable is connected. Remove the power cable
from the wall socket and disconnect the cable from the instrument completely first.
WARNING!
Wrong connections could lead to malfunction of the instrument or computer.
SPECIAL ATTENTION!
After connecting to the electricity supply the furnaces and heating elements will rise to the
safe stand-by temperatures required by the connected module.
71
TSHR International | 6000 Series Systems - User Manual | v1.3
This chapter describes the installation of the parts used to transport the gas from furnace tube to
detection cell.
1. Open the the outlet section (blue door) and the scrubber part (white door).
3. Connect the tube to the shorter outlet of the furnace tube. (Use clamp no. 12)
1 2 3
Connect the
4 flowmeter
here! 5 6
25 mL/min
POWER GAS
72
TSHR International | 6000 Series Systems - User Manual | v1.3
1 2 3 CLAMP NO. 12
4 5
73
TSHR International | 6000 Series Systems - User Manual | v1.3
SPECIAL ATTENTION! Before the titration cell is positioned, remove all spilled acetic acid from the
compartment. Check if the drip tray is placed in the detector compartment and the stirrer motor
positioned.
(Red) Anode
C (Black) Cathode Gas inlet
Gas outlet/
injection
1 2 3
5 I
74
TSHR International | 6000 Series Systems - User Manual | v1.3
1 2 3
4 6 7
75
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 11
TX6000 start up
TSHR International | 6000 Series Systems - User Manual | v1.3
Next chapter describes the steps how to start up the TX6000 and preparation of the glassware before running an
analysis. The following paragraphs should be read carefully, otherwise the measurement could be worthless.
Please note that the software manual must be read before switching on the TX6000.
NOTE!
Read the Athena software manual before operating the TX7000.
Before operating the TX6000 the operator has to be familiar with the format and understand the contents
of the MSDS.
A Material Safety Data Sheet (MSDS) gives health and safety information on chemical products. All
operators who will be using the material must be aware of its hazards, proper work procedures, and personal
protective equipment requirements.
77
TSHR International | 6000 Series Systems - User Manual | v1.3
All necessary solutions needed for a good analysis are described below. For some
specific applications refer to the application notes.
78
TSHR International | 6000 Series Systems - User Manual | v1.3
AOX
Preparation of organic halogen sample adsorbed onto activated carbon is essential for analyses performed with
the TX6000. The sample can be prepared either by the batch or the column method. The following diagram
illustrates how the samples should be prepared, prior to analysis.
79
TSHR International | 6000 Series Systems - User Manual | v1.3
AOX SAMPLING
Sampling
Storage at 4°C.
DOC analysis
Choose method
pH 2-3
pH 2-3
AOX analysis
Washing by 25 ml water
AOX analysis
80
TSHR International | 6000 Series Systems - User Manual | v1.3
Sampling
50 g sample
100ml extract in a
separating funnel
Add 200mlwater
to 100ml extract
EOX analysis
81
TSHR International | 6000 Series Systems - User Manual | v1.3
Store at 4 degrees.C
pH =2
Water layer pH = 9
Separation of organic phase Extract with 100ml petroleum ether
Separation of organic phase
Collect extracts
Concentrate to 5 ml by evaporation
Concentrate to 1ml
EOX analysis
EOX analysis
82
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Switch on the computer. Make sure the USB cable is connected with the TX6000.
2. Switch on the TX6000.
3. Check if the introduction module is properly connected and make sure the TX6000 and the
introduction module is connected to the gas cylinders (O2 and Ar).
1 2 3 O
2
A
r
230V
4. Start the TX analyser software by clicking on the software icon and wait 10 seconds.
5. Click system-status.
6. Start-up settings must be:
83
TSHR International | 6000 Series Systems - User Manual | v1.3
While the system is heating up you can continue with the next steps.
1. Clean the cell thoroughly with demineralized water and/or a tissue paper.
2. Rinse the cell with a small quantity of ammonia in order to remove any excess water and Silver ions.
(Never use organic solvents to clean the cell.)
3. Fill the cell with the correct electrolyte solution up to the level mark
ACID! The titration cell contains a stabilized electrolyte solution of 75% acetic acid in water. The
concentration of silver ions in solution is maintained at a constant level of about 10 -7 M.
4. Fill the reference electrode with the electrolyte solution by using a bulb with the fingers to produce a vacuum
for aspiration. Release once the bulb is completely deflated and the reference electrode is filled.
5. Add the reference electrode.
6. Place the teflon stirrer in the titration cell.
7. Place the Electrode container and add the three electrodes and Gas inlet.
8. The inner part of the Cathode must be filled with the electrolyte solution. If the Cathode is clogged with Silver
ions, place it in a cup filled with ammonia to remove the Silver ions deposit.
9. Place the cell on the stirrer motor (Connect the 5-pole DIN-connector first.)
10. Connect the leads to the corresponding electrodes.
11. Go to Athena software and Set stirrer: 45. See if he stirrer is moving correctly.
12. Set gain: 10. Chloride ions are titrated with silver ions generated at the silver anode.
A peak in the signal can be observed, after automatic calibration the signal will decline to 0mQ.
1 2 3
75%
ACET
IC
ACID
84
TSHR International | 6000 Series Systems - User Manual | v1.3
4 5 6
7 8 9
CATHODE
FILLED WITH
ELECTROLYTE
SOLUTION
ELECTRODE
CONTAINER
10 C A 11
µA 12
0,4
0,35
0,3
0,25
0,2
0,1
0,05
0
85
TSHR International | 6000 Series Systems - User Manual | v1.3
1. Rinse the cell with a small quantity of ammonia in order to remove any excess water.
(Never use organic solvents to clean the cell)
2. Clean the cell thoroughly with demineralized water and/or a tissue paper.
3. Fill the cell with the correct electrolyte solution up to the level mark
4. Fill the reference electrode with the electrolyte solution by using a bulb with the fingers to produce a vacuum
for aspiration. Release once the bulb is completely deflated and the reference electrode is filled.
ACETIC ACID! The titration cell contains a diluted sulfur electrolyte solution.
The concentration of silver ions in solution is maintained at a constant level of about 10 -7 M.
1 2 3
SULFUR
ELECTROL
YTE
SOLUTIO
N
1x rinse 3x rinse
LEVEL MARK
86
TSHR International | 6000 Series Systems - User Manual | v1.3
4 5 6
7 8 9
CATHODE
FILLED WITH
ELECTROLYTE
SOLUTION
10 R I
A
11 C
R
I A
12 µA
0,4
NOTE! I If the signal goes
directly negative, disconnect
13
0,35 the Cathode!
0,3
0,25
0,2
0,1
0,05
0
87
TSHR International | 6000 Series Systems - User Manual | v1.3
ACID! Take the necessary PPE precautions to avoid contact with sulfuric acid.
Wear protective gloves, goggles and acid proof clothing, according safety procedures
on the laboratory. Failure to do so may result in physical injury.
SPECIAL ATTENTION! Before the scrubber is positioned, remove all spilled acetic acid from the
compartment. Check if the drip tray is placed in the detector compartment.
10. Check that the lower stopcock of the scrubber is shut, before filling the absorber of the scrubber to
the red line by opening the upper stopcock (red handle in vertical position).
11. Close upper stopcock (red handle in horizontal position).
Replace upper glass stopper and the connection from the scrubber to the coulometric cell.
12. Now you can place and connect the tubing Scrubber / Titration Cell!
(see page 73 Tubing Scrubber / Titration Cell)
88
TSHR International | 6000 Series Systems - User Manual | v1.3
1 2 3a
HORIZONTAL=CLOSED
3b 3c 4
5 6 7
95%
SULFU
3/ RIC
ACID
4
8 9 3 bar(g)
10
O2 Ar
VERTICAL=OPEN
11 12 25ml
HORIZONTAL=CLOSED
HORIZONTAL=CLOSED
89
TSHR International | 6000 Series Systems - User Manual | v1.3
For reliable and accurate analysis the TX6000 needs to be free of leakage. Therefore a
flow test has to be performed. You can find below the flow-settings:
Measure
CAUTION!
The combustion tube, outlet, modules, joints, etc. may become very hot during the operation
of the TX7000. Carelessness may result in burns.
Procedure
1. If there are no bubbles visible, a gas leakage has to be performed. Adjust flows to the “normal” levels.
(setting according current method) in the software.
2. Disconnect the gas stream between furnace output and the “connection furnace scrubber”.
3. Manually obstruct the gas stream on the outlet (using a septum) and observe the flow meters.
4. Removing the septum from the output should clearly release pressure from the tube. This indicates
that no gas-flow leakage is present in the connectors, Module, Furnace-tube and other joints. The leakage
has to be further down the flow path. (leakage could be found in the septum of the module, any gas
connector, glass joints, loose clamps, mineralized furnace tube, etc.)
5. Install the splash bulb again including the tubing and repeat the “obstruct” procedure.
6. The final check will be the number 3 step from procedure 1.
7. Following these steps should clearly lead to the “problem” and solution in the flow path.
1 2
NO BUBBLES
90
TSHR International | 6000 Series Systems - User Manual | v1.3
3 4
USE A
SEPTUM!
500 mL/min
0 mL/min
POWER GAS
POWER GAS
NOTE! Don't forget to (re)place the septums in the introduction modules. This is
often the cause of gas leakage!
Standard setpoints
The software controls most of the setpoints. These parameters are related to a method. These
methods describe all parameters necessary for running some applications.
1. EOX / EOS Extractable organic halogen or organic Sulphur analysis method for liquid applications
2. AOX / TOX Absorbable or Total organic halogen analysis method for liquid applications
3. POX Purgeable organic halogen analysis method for liquid applications
4. TX / TS-liquids Total halogen or sulphur analysis method for liquid applications
5. TX / TS-gas Total halogen or sulphur analysis method for gas applications
6. TX / TS solids Total halogen or sulphur analysis method for solid applications
Other methods are custom made by TSHR International or created by customers themselves. The
software manual describes how these methods can be created/changed.
An example of setpoints is given in the basic system settings on page 67.
91
TSHR International | 6000 Series Systems - User Manual | v1.3
Furnace 2 700 °C 2
Inlet 50 °C
Outlet 25 °C VERTICAL=OPEN
1. Close the header gas supply or close the flow meters by turning the black knob clockwise.
2. Shut down the software. Switch off the TX6000 (do not disconnect the power plug in order to keep
cooling fans running).
92
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 12
TS / TN6000 start up
TSHR International | 6000 Series Systems - User Manual | v1.3
Before operating the 6000-series the operator has to be familiar with the format and understand the
contents of the MSDS.
A Material Safety Data Sheet (MSDS) gives health and safety information on chemical products.
All operators who will be using the material must be aware of its hazards, proper work procedures, and personal
protective equipment requirements.
94
TSHR International | 6000 Series Systems - User Manual | v1.3
Sulphur Standards
SPECIAL ATTENTION! Advised thiophene in xylene and to order certified standards from Ultra
Scientific. www.ultra-sci.com
A 1000 mg/l mixed TNTS standard is prepared by adding dibenzothiophene (1.437 g) and carbazole (2.986
g) to a 250 ml graduated flask. Pure ACS grade xylene is added to the flask to dissolve the solid. The con- tents of the
flask are swirled to dissolve any remaining solids before being made up to the calibration line with xylene. A stopper is
applied to the flask which is then thoroughly shaken. A range of standards can then be generated from the 1000 mg/l
standard using a 5 ml, 10 ml and 25 ml pipette together with 100 ml volumetric flask. It is important to thoroughly mix,
by shaking each of the mixtures before diluting further.
250 mg/l 25 ml of 1000 mg/l solution made up to 100 ml with pure xylene 100
mg/l 10 ml of 1000 mg/l solution made up to 100 ml with pure xylene 50
mg/l 5 ml of 1000 mg/l solution made up to 100 ml with pure xylene 25
mg/l 10 ml of 250 mg/l solution made up to 100 ml with pure xylene 10
mg/l 10 ml of 100 mg/l solution made up to 100 ml with pure xylene 5 mg/l
10 ml of 50 mg/l solution made up to 100 ml with pure xylene
2.5 mg/l 10 ml of 25 mg/l solution made up to 100 ml with pure xylene 1
mg/l 1 0 ml of 10 mg/l solution made up to 100 ml with pure xylene 0 mg/l
Pure xylene
Nitrogen Standards
95
TSHR International | 6000 Series Systems - User Manual | v1.3
NOTE! The detectors need 30 minutes to reach operational temperatures and optimal
performance conditions.
The UV detector produces a ticking sound after switching on; this is due to the UV flasher lamp and is normal. The TN
detector reaction chamber heats up, PMT cools down as soon as the unit is switched on and produces no sound.
3. Check in the “System status screen” that communication is established between the PC and the 6000-series. If
there is no communication, check that the correct “com port” has been selected and connected. Furnace
temperatures and system parameters can be adjusted and monitored.
The software controls the operation of the 6000-series. Methods are described in the Athena software manual. During
normal operation, regular maintenance procedures need to be carried out and these include:
96
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 13
Running an analysis
(general)
Running an analysis general
TSHR International | 6000 Series Systems - User Manual | v1.3
Starting Athena®
Athena is already installed on the computer when supplied with the system and a USB is also included in the shipment.
If no computer was included, Athena can be installed on the computer connected to
the analyser by inserting the Thumbdrive into the USB port. Installation will begin automatically. Please
refer to the installation manual for more details.
98
TSHR International | 6000 Series Systems - User Manual | v1.3
NOTE! If the number of samples entered in the ‘Number of samples’ box is greater than 1
then ‘Begin sample position’ box denotes start position for the first sample in the queue.
Dilution: Use this box to enter the dilution factor if the sample has been diluted prior to
analysis.
NOTE! By double clicking on this box, a dilution calculation wizard is activated. Please
refer to the software manual for more details.
Density: To compensate for density when the density is not 1. E.g. if the sample is dissolved
in toluene then the density for this solution will be 0.85.
This value is used for calculating the analyzed concentration.
Method: Depending on the configuration of the system, one or more methods
are available. Select the correct method from a pull down menu.
NOTE! A method is a sequence of action steps necessary for performing an analysis. Refer to
the software manual for more details.
99
TSHR International | 6000 Series Systems - User Manual | v1.3
NOTE! This will only work when an autosampler is present. Also see Sample start position.
Analysis per sample: In this box, the number of times this sample needs to be analyzed is
entered.
By clicking the Add button again, the next sample can be added. Repeat this
action until all desired samples have been added to the queue. When finished, click
the analyse button. A similar window will appear:
100
TSHR International | 6000 Series Systems - User Manual | v1.3
If in the ‘calibration line’ box, the option new is selected, a new calibration set will be created. The
following box will be displayed:
A new name, the calibration order, signal and through zero must be entered. For the coulometer
(TX and TS), the order chosen must be ‘Absolute’. In most other cases, a first order (linear fit) is applied.
If the system has more than one detector, the desired signal has to be selected. ‘Through
zero’ is used to force the calibration line through the origin.
An analysis is started in the same way as a sample by clicking the ‘analyse’ button. The
calibration standards are measured and a calibration curve produced.
101
TSHR International | 6000 Series Systems - User Manual | v1.3
Before shipment to its final destination, the system undergoes a factory calibration. This instrument specific information
can be found in the Test Report which is sent together with the analyser. The report shows test results and PMT specific
data. The data shows the generated area for a specific standard. This value may be used as an “absolute reference”
for system performance.
By injecting standards at an injection speed of 1μl/s and using the universal “liquids module”,
a comparable result should be obtained at the customer site.
NOTE! The results in terms of area counts strongly depends on the used standards, gas quality
and local contamination levels.
Minimum specifications for a calibration line in the range of 0-50 mg/L, when
If the values above are obtained, the system is functioning correctly and calibration and analysis may be carried out.
Areas lower than ±10% of the values above indicate system problems. In this case, a gas test and overall system
check is required.
102
TSHR International | 6000 Series Systems - User Manual | v1.3
WARNING! Whenever the vacuum pump is shut down, shut off the Oxygen supply to the
ozonator. This will prevent the production of ozone, which could affect the performance of
several devices in the 6000-series.
Furnace 1 700 °C
Furnace 2 700 °C
Ozonator off
103
TSHR International | 6000 Series Systems - User Manual | v1.3
METHOD FURN 1 FURN INLET OUTLET ARGON/HEL OXYGEN OXYGEN EXT. OXYGEN
NORM/S 2 NORM/S NORM/ TURBO
TD NORM/S TD STD
SORT CELCIUS CELCIUS CELCIUS CELCIUS ML/MIN ML/MIN ML/MIN ML/MIN
TD
EOX/CI 850/700 850/700 550/100 300/100 40/35 50 50 20
Tips
1. Take the cell out of the system when you don’t use it for more than a week, remove
the two silver electrodes and leave the rest in the cell.
2. Every day you have to refresh the sulphuric acid in the NOx-scrubber and the scrubber when you get a negative
baseline or whether is an increase in the volume.
3. Unused reference electrodes can be hold in good condition when you store them in demi water.
104
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 14
Troubleshooting tx6000
Troubleshooting TX6000
TSHR International | 6000 Series Systems - User Manual | v1.3
No cell current. Gain control on zero. Set the gain control at 10.
Cell current deviates greatly Electrical connection between cell Fill the titration cell with
from zero or fluctuates. and microcoulometer is broken. electrolyte solution and check
whether electrolyte solution is
present in the side arm of the
titration cell.
Cell current is large and Cl- is leaking from the reference Change the reference electrode.
positive. electrode into the cell. Replace the reference electrode.
Cell current is large and The system is severely Change the electrolyte solution.
negative. contaminated with silver ions.
Cell current is unstable. Irregular stirring, stirrer element out Check the stirrer element and
of phase; the cell is not firmly stirrer motor.
located on the stirrer motor.
The silver surface of the indicator Allow the electrode to stabilize for
electrode has not yet stabilized a period of time or replace with a
e.g. after polishing or renewal. stabilized electrode.
106
TSHR International | 6000 Series Systems - User Manual | v1.3
Cell current is unstable. Changing exposure to light of the Make sure that the cell cabinet
cell. is properly shut.
No response to Chloride Incorrect bias potential, making Check the potential difference
dosage. the system insensitive to chloride between the indicator
and silver ions. electrode pair. The bias setting
should be –315 mV.
Titration time is too long. The cell is severely After removing the electrodes,
Contaminated with AgCI. clean the cell with concentrate
ammonia, large amounts of
demineralised water and
electrolyte solution.
The composition of the electrolyte is Replace the solution with new solution
incorrect. with the correct solution.
107
TSHR International | 6000 Series Systems - User Manual | v1.3
Titration time is too slow. Large volume of titration solution. Increase the gain or drain off
some titration solution.
Titration time is too fast. Gain level is set too high. Reduce the gain.
High baseline due to high current Sulfuric acid solution is severely Purge the sulfuric acid solution
Cell current is unstable contaminated. for 15 minutes or replace it with
clean solution.
Long titration time. Gain is set too low. Increase gain level.
Gas flows too low. Check the flow setting and the
apparatus for blockages in the
scrubber or titration cell inlet.
Check the apparatus for leaks
(including the septum of the EOX
and POX modules).
Recalibrate the MFC's.
108
TSHR International | 6000 Series Systems - User Manual | v1.3
Recovery too low. Furnace temperature is incorrect. Check the oven temperature.
Recovery is too high. High blank values for the solvent Check the solvent, if necessary
used for the Liquid analysis. clean by distillation.
Irregular or repeating spikes. Air bubbles in the injection Clean the syringe and check if
syringe (EOX, EOS). there still is enough wash solution
available.
Condense in the splash bulb Dry the splash bulb and check
(HCl absorption). the O2 and Ar/He flows.
109
TSHR International | 6000 Series Systems - User Manual | v1.3
Irregular or repeating spikes. A defect in the scrubber heating or Increase the temperature to
low temperature setting, (HCI normal levels.
absorption) due to condenses in the
heater section. Contact the TSHR
Technical Support.
In correct baseline. If the baseline does not behave as Contact the TSHR
outlined above, this may be Technical Support
caused by a number of errors.
Generally too high baseline. Cell problems. Check whether the baseline
drops to an acceptable level
when the cell is disconnected
from the scrubber.
110
TSHR International | 6000 Series Systems - User Manual | v1.3
CHAPTER 15
Troubleshooting
TS6000/TN6000
TSHR International | 6000 Series Systems - User Manual | v1.3
The TS6000 and the TN6000 has been designed to achieve a high level of reliability. In case of
Recovery is too low or too high. Permapure has melted Replace the Permapure Clean
Soot in the Permapure. it or replace it with a new
Permapure.
Recovery is too low. Furnace temperature is incorrect Check the furnace temperature.
Recovery is too high. High blank values for the Check the solvent and if necessary
solvent used for the liquid clean by distillation. Check the gas
analysis. quality. Make sure that it's according
to the specifications.
112
TSHR International | 6000 Series Systems - User Manual | v1.3
Incorrect baseline. If the baseline does not behave as Contact our Service Department.
outlined above, this may be
caused by a number of factors.
Generally too high baseline. Gas problems Check gas supply and the sample
introduction section.
When possible, connect the gas
bottles separately, so it is possible to
circumvent the gas supply network.
No ticking sound in the Xenon UV lamp is at it's end of Replace the Xenon UV lamp.
TS-detector. lifetime.
113
TSHR International | 6000 Series Systems - User Manual | v1.3
Furnace Temperature Sensor 2 x Ni-Cr/Ni Computer Intel Core i3 / AMD Phenom or better
Furnace configuration Dual temperature controlled Software Athena
Furnace Temperature 1250 °C Max Regulatory compliance ASTM, IP, UOP, ISO
FACILITY REQUIRMENTS
Voltage 115/230 V, 50/60 Hz Gasses O2(99,6%) medical grade 2.6
Power 1200W Ar or He (99,998%) technical grade 4.8
Gas connector 1/8” swagelok Gas pressure 2 – 3 Bar (30-45 psi)
Ambient temperature 5 – 35 °C (41 – 95 °F)
1128 mm 600 mm
372 mm
372 mm
117
TSHR International | 6000 Series Systems - User Manual | v1.3
17. TECHNICAL SPECIFICATIONS TN6000
Parameter 6000 series Detection Principle Chemiluminescence (extra
Dimensions 939 (W) x 372 (H) x 600 (D) mm options: Coulometrics and
37 (W) x 15 (H) x 24 (D) in Pulsed UV-Fluorescence)
Furnace Voltage 2 x 24 V, 50/60 Hz PC operating system Windows 7 or higher
Furnace Power 2 x 300 W Computer Intel Core i3 / AMD Phenom or
Furnace Temperature Sensor 2 x Ni-Cr/Ni better
Furnace configuration Dual temperature controlled Software Athena
Furnace Temperature 1250 °C Max Regulatory compliance ASTM, IP, UOP, ISO
FACILITY REQUIRMENTS
Voltage 115/230 V, 50/60 Hz Gasses O2(99,6%) medical grade 2.6
Power 1200W Ar or He (99,998%) technial grade 4.8
Gas connector 1/8” swagelok Gas pressure 2 – 3 Bar (30-45 psi)
Ambient temperature 5 – 35 °C (41 – 95 °F)
939 mm 600
372 mm
372 mm
119
TSHR International | 6000 Series Systems - User Manual | v1.3
Instrument
Properties 6000-series TS6000 TN6000 TX6000
Dimensions (W x D x H) 601 x 580 x 1096 mm 601 x 580 x 1096 mm 601 x 580 x 1096 mm 601 x 580 x 1096 mm
Weight 55 kg 50 kg 50 kg 45 kg
Septum Type 12 mm HT 12 mm HT 12 mm HT 12 mm HT
Operating Humidity 5% @ 15°C - 60% @ 35°C 5% @ 15°C - 60% @ 35°C 5% @ 15°C - 60% @ 35°C 5% @ 15°C - 60% @ 35°C
Sample Injection Speed 0,8 μl/s 0,8 μl/s 0,8 μl/s 0,8 μl/s
Working Range 30 ppb - 10,000 ppm 30 ppb - 10,000 ppm 30 ppb - 10,000 ppm 100 ppb - 10,000 ppm
Method Compliance ASTM D4629, ASTM ASTM D7183, ASTM ASTM D4629, ASTM ASTM D4929, ASTM
D6069, ASTM D7184, D5453, ASTM D6667, D6069, ASTM D7184, D5808, ASTM D5194,
ASTM D7183, ASTM ISO 9001 UOP 936, ISO 9001 ASTM D3120, ISO
D5453, ASTM D6667, 9001
UOP 936, ISO 9001
Voltage 230 / 115 VAC 230 / 115 VAC 230 / 115 VAC 230 / 115 VAC
121
TSHR International | 6000 Series Systems - User Manual | v1.3
Minimum Oxygen Grade 2.6 (99.6 %) 2.6 (99.6 %) 2.6 (99.6 %) 2.6 (99.6 %)
Oxygen Consumption 500 - 550 ml/min 400 - 450 ml/min 500 - 550 ml/min 400 - 450 ml/min
Oxygen Consumption 500 - 550 ml/min 400 - 450 ml/min 500 - 550 ml/min 400 - 450 ml/min
Argon Consumption 0 - 150 ml/min 0 - 150 ml/min 0 - 150 ml/min 0 - 150 ml/min
Helium Consumption 150 ml/min 150 ml/min 150 ml/min 150 ml/min
HR7000
6000-series TS6000 TN6000 TX6000
Autosampler
Dimension (W x H x D) 280 x 570 x 320 mm 280 x 570 x 320 mm 280 x 570 x 320 mm 280 x 570 x 320 mm
Weight 8 kg 8 kg 8 kg 8 kg
Vial Size 2 ml 2 ml 2 ml 2 ml
122
TSHR International | 6000 Series Systems - User Manual | v1.3
Chapter 19
Consumables
TSHR International | 6000 Series Systems - User Manual | v1.3
PGENE0002
1 Syringe 100μL 71mm PS5 GA22, EURO2 1
PGENE0004
2 Needles 71mm PS5 GA22, 3 pieces 1
PANAL0025
3 Septum diameter 12mm, 10 pieces 1
742312010071
4 Particle filters (PTFE) 10pcs. 1
(TN/TS)
PMECH0060
5 Union ¼”x ¼” OD Nylon 1
PMECH0061
6 Connector nylon elbow ¼”x ¼” 1
742312010061
7 Perma pure dryer 1
1. 2. 3. 4.
5. 6. 7.
TSHR International | 6000 Series Systems - User Manual | v1.3
TECHNICAL SUPPORT
Phone 0031 (0)10 2457701
E-mail [email protected]
TSHR International
Strickledeweg 44
3044 EK
Rotterdam The
Netherlands
www.tshrinternational.com