User Manual EN

Download as pdf or txt
Download as pdf or txt
You are on page 1of 79

TECHNOS 5/3

Revision: E
Version: 26 October 2009
File: [201175-E-10-01.doc]
User’s Manual
WARNING: The information that is printed within this manual is vital for the correct use of the
equipment; please read it carefully before use.
User's Manual TECHNOS 5/3 Sias S.p.A.

This page is intentionally left blank

[File:201175-E-00-01.doc]
Sias S.p.A. TECHNOS 5/3 User's Manual

Table of contents

1. SAFETY AND COMPLIANCE ........................................................................................................... 2


1.1. Electrical safety ............................................................................................................................................. 3
1.2. Laser targeting devices safety ...................................................................................................................... 3
1.3. Mechanical safety ......................................................................................................................................... 3
1.4. Electromagnetic compatibility (EMC) ............................................................................................................ 4
1.5. Protection against ionizing radiation ............................................................................................................. 4
1.6. General disposal ........................................................................................................................................... 5
1.7. Interfaceability ............................................................................................................................................... 5
1.8. Copyright ....................................................................................................................................................... 5
1.9. Application and final destination.................................................................................................................... 6
1.10. Classification ................................................................................................................................................. 7
1.11. List of the Standards for the evaluation of the product compliance .............................................................. 7
1.12. Compliance ................................................................................................................................................... 7
2. COMPONENT IDENTIFICATION ...................................................................................................... 8
2.1. Overview ....................................................................................................................................................... 8
2.2. Mobile Stand ............................................................................................................................................... 10
2.3. Keyboard ..................................................................................................................................................... 12
2.4. Fluoroscopy control footswitch.................................................................................................................... 18
2.5. Handswitch for radiography / fluoroscopy control ....................................................................................... 18
2.6. Monitor Trolley ............................................................................................................................................ 19
2.7. Operative messages ................................................................................................................................... 20
2.8. Alarm messages ......................................................................................................................................... 21
2.9. Safety devices ............................................................................................................................................. 23
3. UNIT USE......................................................................................................................................... 24
3.1. Transport ..................................................................................................................................................... 24
3.2. Connection between monitor trolley – mobile stand ................................................................................... 25
3.3. Connection between Footswitch – Mobile Stand ........................................................................................ 26
3.4. Unit ON / OFF ............................................................................................................................................. 27
3.5. Positioning................................................................................................................................................... 29
3.6. Use mode .................................................................................................................................................... 30
3.7. Shutdown procedure ................................................................................................................................... 40
4. MAINTENANCE............................................................................................................................... 41
4.1. General warnings ........................................................................................................................................ 41
4.2. Checks and inspection by the user ............................................................................................................. 41
4.3. Cleaning ...................................................................................................................................................... 42
4.4. Disinfection.................................................................................................................................................. 42
5. TECHNICAL DATA.......................................................................................................................... 43
5.1. Labels and symbols .................................................................................................................................... 43
5.2. Environmental conditions ............................................................................................................................ 47
5.3. Electrical data.............................................................................................................................................. 48
5.4. Exposure mode: fluoroscopy....................................................................................................................... 58
5.5. Exposure mode: radiography ...................................................................................................................... 61
5.6. Image system .............................................................................................................................................. 62
5.7. Digital image processor............................................................................................................................... 65
5.8. Accessories................................................................................................................................................. 70
6. MECHANICAL DATA ...................................................................................................................... 73
6.1. C-arm unit dimensions with I.I. 9”............................................................................................................... 73
6.2. Axis and interventionist reference point ...................................................................................................... 73
6.3. “Base Trolley” monitor trolley dimensions ................................................................................................... 74
6.4. Dimensions of “High Configuration” monitor trolley..................................................................................... 75
6.5. Mechanical data, dimensions and weights.................................................................................................. 76
7. DOCUMENT STATUS ........................................................................................................................ I

[File:201175-E-00-01.doc] Rev. A - Pag. 1/76


User's Manual TECHNOS 5/3 Sias S.p.A.

1. SAFETY AND COMPLIANCE


The purpose of this user's manual is to provide a set of easy to use instructions for the
proper use of the system. All of the information contained herein is based on the current
version of the system. Sias S.p.A. reserves the right to improve and implement changes to
the information herein to reflect any changes necessitated by technological enhancements to
the system.

This x-ray unit must be used in strict compliance with the safety
instructions contained in this manual and must not be used for purposes
other those for which it was intended.

The x-ray unit may only be operated by skilled, properly trained personnel with the required
knowledge of x-ray safety practices and the proper use of x-ray equipment.
The operator is responsible for the use of the system in compliance with the applicable
standards concerning installation and use.

The system safety circuits and devices must not, for any reason, be moved, modified, or
omitted.
The monitor trolley key switch may only be activated by authorized personnel and only
during the use of the system. Once system use has been completed, the key must be
removed and stored in a safe place.

The unit must not be operated when electrical, mechanical, or radiological faults are present
or when any of the indicators or alarm devices are malfunctioning.

When used in conjunction with other apparatus, components, or modules, whose


compatibility is uncertain, it is necessary to ensure the absence of any danger to the patient
or operator. Consult Sias S.p.A. for information.

As with any technical apparatus, this x-ray unit must be used properly with periodic checks
and maintenance as specified in the chapter “Planned maintenance” of the Service Manual.

Sias S.p.A. is responsible for the safety of its products only when maintenance, repairs, or
modifications have been performed by its personnel or by personnel authorized by Sias
S.p.A. in writing.

Sias S.p.A. cannot be held liable for any malfunction, damage, or danger
resulting from improper use of the system or non-compliance with the rules
for proper maintenance.

Pag. 2/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

1.1. Electrical safety

Only trained service personnel authorized by Sias S.p.A. may remove the
unit covers and only in accordance with the instructions contained in the
Service Manual.

This x-ray unit may only be used in environments or medical rooms in compliance with the
applicable IEC standards.
The x-ray unit must not be used in areas where there exists a danger of explosion.
Cleaning and disinfecting agents, including those used on patients, may create an explosive,
gaseous mixture. Use only those products in compliance with the applicable rules.

1.2. Laser targeting devices safety

• Keep always a good lightening on the room.


• Never look through the output window of the laser targeting device.
• Never fix the reflections of the laser targeting devices.
• Before starting any examination, the patient must remove earrings, glasses, necklaces
and whatever could reflect the laser beam and be printed on the image.
• Don't clean the openings of the laser targeting devices with tools that could modify their
optics. Only the service personnel must perform possible cleaning actions.
• The min. distance between the laser source and the patient must not be lower than cm20.

The only purpose of the laser use is to reduce the patient dose to a
minimum, it doesn’t be considered as an absolute means of centering.
The activation of procedures other those listed above can cause the
emission of dangerous non-ionizing radiations.

1.3. Mechanical safety


The unit can be moved only by trained personnel by complying with the following safety
conditions:

• Move the unit by considering the conditions for the movement that are indicated in detail
in the paragraph 3.1.
• Don't move and use the unit on surfaces with inclination higher than 5°
• Only use the proper handles to move the unit.
• Avoid collision with obstacles.
• After positioning the unit, engage the parking brakes.

[File:201175-E-00-01.doc] Rev. A - Pag. 3/76


User's Manual TECHNOS 5/3 Sias S.p.A.

1.4. Electromagnetic compatibility (EMC)


This apparatus is in compliance with the standard IEC 60601-1-2 that defines the max.
allowed emission levels from electronic devices and the required immunity from interference
caused by externally generated electromagnetic fields.

It is not, however, possible to exclude radio signals coming from transmitters such as mobile
phones or similar mobile radio devices. These and other transmitting devices, including
those in compliance with the EMC standards, may influence the proper functioning of
medical apparatus when used in proximity and with a relatively high transmitting power.
Therefore, the use of radio equipment proximity to electronically controlled systems must be
avoided in order to eliminate any interference risk.

Any transmissions by mobile radio equipment must be avoided.


Mobile phones must be switched off in zones close to the unit.
These rules must be applied when the unit is switched on (that is to say
connected to the mains and ready for use).

1.5. Protection against ionizing radiation

Before any x-ray exposure, ensure that all the necessary protective
precautions have been taken.

During the use of x-rays, personnel present in the room must comply with the following rules
concerning protection against ionizing radiation:

When necessary, use protective shielding against radiation in addition to the shielding
already provided on the unit.
Use protective aprons containing a material equivalent to 0,35mm of lead. Material of this
nature reduces radiation at 50kV by 99,95% and at 100kV by 94,5%.
The best protection against radiation is distance. It is therefore recommended that you stay
as far as possible from the x-ray source and the exposure target. For this purpose, use all of
the cable length provided for the foot-switch.
Avoid walking or standing directly in the x-ray beam.
Always use the smallest possible field of exposure by closing properly the collimator
diaphragms. The scatter dose produced depends principally on the volume of the irradiated
object.

Never modify or disconnect the safety circuits or devices designed to


prevent accidental exposures.

Collimation system adjusted to a field of 30 x


30 cm2 at 1 m from the x-ray tube focus.
Exposure parameters: 100 kVp and 1 mA.
Values normalized to 1 min. of fluoroscopy
(µGy/min).
Equivalent water phantom 30x30x20cm.

The complete isodose measures are included


into the report 035-07 ISO KERMA MAPPING
(cod.200394) attached to the unit
documentation

Pag. 4/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

1.6. General disposal


Sias S.p.A. produces radiological systems that are advanced in terms of safety and
environmental protection. Assuming that the unit is properly used, there is no risk to people
or the environment.
In order to comply with applicable safety requirements, it is necessary to use materials that
may be harmful to the environment (for example: monobloc oil, protective lead, monitor
kinescope, boards and electronic components). Therefore, when necessary, proper disposal
methods, according to the regulations of the country where the unit is installed, should be
followed.

For this reason, the unit may not be disposed of along with industrial or
domestic waste and must be regarded as hazardous waste.

This symbol indicates that the waste of electrical and electronic equipment must
not be disposed as unsorted municipal waste and must be collected separately.
The proper differentiated collection for the following start of the unit disused to the
recycle, treatment and disposal, compatible with the environment, aid to prevent
possible negative effects on the environment and health and it favours the recycle
of materials that compose the unit.
The abusive disposal of the product from the user implies the application of
administrative sanctions according to the Standards in force of the unit installation
country.
For information concerning the dismantling modes of the units out of use, stick to
the local provisions or contact a representative authorized by the manufacturer.

For additional information, contact Sias S.p.A..

1.7. Interfaceability
The device does not forecast any interaction with devices for medication.
It is possible to interface the unit with certain devices such as DVD Recorder, thermal printer,
Network (DICOM System). Such devices must be in full compliance with the safety
requirements specified by 93/42/EEC Directive. The liability of the interface, if it has not been
evaluated and authorized by Sias S.p.A. in writing, is of the operator and/or of the person
who has performed this interface.

1.8. Copyright
The original release of this manual is in Italian language (file: 201175-X-00-00.doc). For
further information, please refer to the Italian version.
The software contained in the unit belongs to Sias S.p.A.. Upon receipt of the unit, the user
acquires the right to use the software. This right is neither exclusive nor transferable.
Written authorization to Technix S.p.A. is mandatory prior to any modifications for the unit
use with functions other than the ones foreseen.

[File:201175-E-00-01.doc] Rev. A - Pag. 5/76


User's Manual TECHNOS 5/3 Sias S.p.A.

1.9. Application and final destination


TECHNOS 5/3 is a "mobile x-ray unit with image intensifier system" and it has been
designed to be used for diagnosis. It must be operated exclusively by qualified, trained
personnel who have been informed of the risks linked to the use of ionizing radiation.

The system does not belong to the category of equipment designed for continuous
operation.

The system is not used in contact with the patient; however, accidental contact of
some unit parts with the patient and the operator is possible.

Contact with the patient is non-invasive.

Contact with the operator is strictly for reasons linked to the use of the equipment
(normal operation).

The unit is suitable to be used for x-ray examinations, and in particular for radioscopy,
radiography and diagnosis dedicated to:

• Traumatology
• Pediatrics
• Simple interventional radiology
• Pace Maker implantation
• Operating theater
• Intensive care
• Respiratory system
• Skeletal structure

This x-ray unit must not be used in areas where danger of explosion
exists.
For use in operating theater, it is necessary to use a sterile coverings to
protect the arm-monobloc-intensifier group from liquids seepage and a
footswitch type IP-X8 (like the standard one).

The patient support must not have an equivalent filtration higher than
2mmAl.
The patient must be placed as close as possible to the image intensifier.

The unit is available in two versions:


TECHNOS 5 with stationary anode monobloc, successively named
“Stationary anode version”,
TECHNOS 3 with rotating anode monobloc, successively named “Rotating
anode version”.
If not else specified, the technical characteristics are intended available for
both versions.

Pag. 6/76 - Rev. B [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

1.10. Classification
Protection against electrical hazards ................ Class I
Protection against direct and indirect contact ... Unit Type B with Type B applied part
Protection against water penetration................. Common protection (IPXO)
Fluoroscopy footswitch protected against the
submersion effects (IPX8)
Use condition protection.................................... Continuous working with temporary load

1.11. List of the Standards for the evaluation of the product compliance

Reference Description

MDD 93/42/EEC class IIB according to Medical Devices Directive (EC mark)
Annex IX rule 10.
IEC 60601-1 1st edition Medical devices safety
IEC 60601-1-2 1st edition Electromagnetic compatibility
IEC 60601-1-3 1st edition Protection against ionizing radiation
IEC 60601-2-7 2nd edition HV generators
IEC 60601-2-28 1st edition Tube-housing groups
IEC 60601-2-32 1st edition Mechanical safety aspects
ISO 14971:2000 Risk analysis
CEI EN 60825-1 2nd edition Laser equipment safety

TECHNOS 5/3 with radio-protection according to the Standard CEI EN 60601-1-3 (1995)

Inverter-monobloc group
(IN-9040-5 HF + I-40S 3,5 RF, IN-9040-5 HF + I-40R 5 RF) EN60601-2-7:1998

X-ray group for diagnostics TECHNOS 5/3 IEC 601-2-28 (1993)

Complementary unit TECHNOS 5/3 IEC 601-2-32

1.12. Compliance

This x-ray unit is in compliance with the Electromedical Devices Directive


93/42 EEC class IIb and with the Annex IX rule 10.

For any further information concerning the The manufacturer (according to the European
compliance please contact: Directive 93/42/EEC) of the unit TECHNOS 5/3
is:
SIAS S.p.A. Sias S.p.A.
Via MInghetti, 9/11 Via E. Fermi, 45
40057 Cadriano di Granarolo dell’Emilia, BO 24050 Grassobbio, BG (ITALY)
(ITALY) Tel: +39 (0)35-3846611
tel. +39 051 60 20 211 Fax: +39 (0)35-335675
fax. +39 051 60 20 218
Web: http:www.sias-spa.it
E-mail: [email protected]

[File:201175-E-00-01.doc] Rev. E - Pag. 7/76


User's Manual TECHNOS 5/3 Sias S.p.A.

2. COMPONENT IDENTIFICATION
2.1. Overview
The unit TECHNOS 5/3 is made up of two different parts: Mobile Stand and Display Station.
Mobile stand consisting of:
“C” Arm, Monobloc, Image
Intensifier, Control Panel (Fig. 1).

Fig. 1

Pag. 8/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Display Station consisting of:


• nr.1 orientable LCD monitor
17”, directly on mobile stand for
digital systems with CCD 0,5K x
0,5K (Fig. 2). *
• “Low Profile Base Trolley” with
two LCD monitors 19” for video
systems 0,5K2 (Fig. 3).
• “High Profile Base Trolley” with
two monitors 18” for video
systems 1K2 (Fig. 4).
• “High Configuration” trolley with
two monitors 19”/18” for video
systems 0,5K2 / 1k2 (Fig. 5).

Fig. 2

Fig. 3 Fig. 4 Fig. 5


* Note: the number of the monitors and the
monitor trolley change according to the
image system choice.

[File:201175-E-00-01.doc] Rev. A - Pag. 9/76


User's Manual TECHNOS 5/3 Sias S.p.A.

2.2. Mobile Stand

Image intensifier

Emergency push-button for


the UP/DOWN movement

Handles for the


Handle for the horizontal
arm sliding
arm group sliding

Control panel

Handles for the movement


of the mobile image
Collimator intensifier.

Handswitch for radiography /


fluoroscopy control

Connector for the connecting


cable between mobile image
intensifier and monitor trolley

Monobloc

Focus position

Fig. 6

Cable reel

Printer for dosimeter


(optional)

ON key

Outlet for the


fluoroscopy footswitch
connector

Outlet for External


Interblocks (optional)

Mains cable

Fig. 7

Pag. 10/76 - Rev. B [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

2.2.1. Handles and brakes position for manual movements

The brake handle is take-up type. In order to avoid 1. Brake for the arm
collision dangers with the unit or the user, always put the rotation around the
handle again in horizontal position. horizontal axis.
Movement:±270°

2. Brake for horizontal


sliding of the arm group.
Movement: 220mm

1
3. Brake for the orbital arm
3 rotation.
4 Movement:123°
2 (+90°÷-33°)

4. Brake for the overview


movement of the arm
group.
Movement: ±12,5°
Fig. 8

1. Driving handle and rear wheels


braking

Fig. 9

Pos.1: brake ON

Pos.2: oblique movement

Pos.3: free movement

Pos. 4: oblique movement


1 5
Pos.5: right-left movement

2 4

[File:201175-E-00-01.doc] Rev. B - Pag. 11/76


User's Manual TECHNOS 5/3 Sias S.p.A.

2.3. Keyboard

2.3.1. Unit
All keys are membrane type. The keyboard group can rotate ±45° in respect of the central position
for its simpler use.

Fig. 10

Area 1
Movimentazione verticale Spegnimento / Accensione
dell’arco centratore laser

Spegnimento Spia segnalazione


dell'apparecchiatura emissione raggi X

Accensione
Segnalazione allarmi
dell'apparecchiatura

Area 2
Rotazione digitale Rotazione diaframma a
1
dell’immagine lamelle

Chiusura / Apertura
Ingrandimento immagine
diaframma a lamelle

Inversione orizzontale Chiusura / Apertura


immagine diaframma a iride

Rafforzamento dei bordi di


un’immagine
ENTER + F1: selezione
modalità “Utility Mode”
Motion Detector
(funzione non abilitata)

Commutazione tra
Scorrimento immagini
immagine LIVE e immagine
memorizzate
MEM su singolo monitor.
Trasferimento immagine da
monitor lavoro a monitor
memoria

Decremento / Incremento
Decremento / Incremento kV
mA/mAs

1
Per memorie serie SBFM78, DIP e HRC: premendo contemporaneamente i due pulsanti di rotazione per
quattro secondi, si ha l’azzeramento della rotazione digitale dell’immagine
Pag. 12/76 - Rev. A [File:201175-E-00-01.doc]
Sias S.p.A. TECHNOS 5/3 User's Manual

2.3.2. “Base Monitor Trolley”

Memories series SBFM /HRC

The memory series SBFM76 has not the keyboard.

The memories series SBFM78 have only the alphanumeric keyboard for the patient data input. The
keyboard can be placed under the unit keyboard (unit with monitor aboard) or on the highest shelf
of the “Base Trolley” in “Low Profile” configuration.

The memories series HRC have the alphanumeric keyboard and mouse for the patient data input.
Keyboard and mouse are placed on the highest shelf of the “Base Trolley” in “High Profile”
configuration.

2.3.3. High Configuration trolley

On the High Configuration trolleys, there is always a small remote keyboard that duplicates some
controls present on the unit keyboard.
All keys are membrane type.

F11 F12 LINE

Fig. 11

1^ line 2^ line

Image enlargement Digital image rotation

Image edges enhancement Stored images scrolling

Motion Detector Image transfer from the live


(unavailable function) monitor to the memory monitor

Image reversal on vertical


n.a.
axis

Shutters diaphragm
Shutters diaphragm rotation
closing/opening

Iris diaphragm
n.a.
closing/opening

[File:201175-E-00-01.doc] Rev. A - Pag. 13/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Memories series DIP

The memories series DIP have a keyboard dedicated to the post processing.
All keys are membrane type.

KEYBOARD ACQUISITION

Esc F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 Del


DOSE

~ ! @ # $ % ^ & * ( ) _ + M
` 1 2 3 4 5 6 7 8 9 0 = = L

Q W E R T Y U I O P { } ANGIOGRAPHY DICOM
[ ]

A S D F G H J K L : "
; '
Enter ON
Z X C V B N M < > ? ¦
, . I.R.
VIEW OVERVIEW

Shift Ctrl Space Alt Shift

Fig. 12
For an easier understanding they have been divided into 7 different groups: KEYBOARD,
ACQUISITION, PROCESSING, ANGIOGRAPHY, DICOM, VIEW, OVERVIEW

KEYBOARD

New exam Patients list

Dose information VCR prearrangement


DOSE

Search start backwards Search start onwards

Rotation + Rotation -

ACQUISITION

SNAPSHOT control prearrangement Image reversal on vertical axis

PULSED FLUOROSCOPY control


Image rotation resetting
prearrangement

MEM image and FLUOROSCOPY


Recursive filter selection
L alternation

SMART filter activation / deactivation Image storage key


L

Pag. 14/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

PROCESSING
It allows the images visualization in SHARP / SMOOTH / OFF edges
negative processing

Key to increase the brightness Key to increase the contrast

Key to decrease the brightness Key to decrease the contrast

ANGIOGRAPHY
Angiographic acquisition rate selection Key for the mode activation of the
1 image per second subtraction between images

Angiographic acquisition rate selection


Mask acquisition key
3 images per second

Angiographic acquisition rate selection


SHIFTING PIXEL function activation
6 images per second

MAX OP control activation / deactivation LAND MARK control activation

DICOM

Single image sending to DTU Whole exam sending to DTU

Current RUN sending to DTU DTU monitor commutation

VIEW

Memory number decrease Memory number increase

ALTERNATE MEMORY procedure Key to delete the images stored in the


control buffer (active only in DISKLESS mode)

OVERVIEW

Stored image zoom Stored images overview

[File:201175-E-00-01.doc] Rev. A - Pag. 15/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Memories series HRP

The memories series HRP have the alphanumeric keyboard and the mouse for the patient data
input.
All keys are membrane type.

Esc F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 F11 F12

! @ # $ % &
* ( )

<
+ _
Num Scroll Print

` 1 2 3 4 5 6 7 8 9 0 =

{
-

} |
Lock

7
Lock

8
Pause

9
Screen

Q W E R T Y U I O P [ ] \ Home Pag *
: "
4 5 6
Caps-lock A S D F G H J K L ; '
Ent er -

> ? 1 2 3
Z X C V B N M , . / End Pag +
0 .
Ctrl Alt Alt Gr Ctrl Ins Del Ent er

Pag. 16/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

2.3.4. Display

Alphanumeric touch-screen display 5” for x-ray parameters and warning/error messages.

The keys, so-called at “retention”, are displayed:

in positive on white background for the non-active function

in negative on black background for the active function

Note: the box with the dose indication is displayed only if the dosimeter is installed and in working
condition.

2.3.5. Audible signals

1 BEEP 2 Sound signal when any key is pressed.


2 BEEPS Storage OK.
A LONG BEEP Alarm, malfunction.
(about 1sec)

2
It is possible to deactivate the audible signals or modify the volume (§3.6.9 Utility Mode).

[File:201175-E-00-01.doc] Rev. A - Pag. 17/76


User's Manual TECHNOS 5/3 Sias S.p.A.

2.4. Fluoroscopy control footswitch

The footswitch is enabled only in fluoroscopy mode.

The fluoroscopy control footswitch consists of a double-step pedal and two single-step ones.
The functions in the different modes are:

1. Left pedal (two steps):

Fluoroscopy Mode:
1° step: fluoroscopy control.
2° step: image storage; in continuous 3
fluoroscopy the current image is stored. 2

2. Right pedal (a step):

Pulsed fluoroscopy mode.

3. Third pedal (a step):

Fluoroscopy mode: 1
"snapshot" control (high-contrast
exposure with reduced background noise).
Fig. 13

2.5. Handswitch for radiography / fluoroscopy control


The control handswitch is made of a two-
steps switch.

Radiography mode:
1° step: preparation control.
2° step: emission control.
In radiography mode it is possible to press
immediately the exposure control (2° step),
with an emission delay due to the anode
starting phase (only for rotating anode version).

It is possible to control Fluoroscopy and


Storage even with the radiography
handswitch.
Fluoroscopy mode:
1° step: fluoroscopy control.
2° step: image storage.

Normally the function is enabled. In order


to disable it, it is necessary Service Fig. 14
intervention.

Pag. 18/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

2.6. Monitor Trolley

Unit “TECHNOS 5/3” with monitor aboard Unit “TECHNOS 5/3” + “Low Profile Base Trolley”

Unit “TECHNOS 5/3” + “High Profile Base Trolley” Unit “TECHNOS 5/3” + “High Configuration” trolley

[File:201175-E-00-01.doc] Rev. A - Pag. 19/76


User's Manual TECHNOS 5/3 Sias S.p.A.

2.7. Operative messages

Display in fluoroscopy mode

1 8

2 7

5 6

1 Indication of the dose value (only if the dosimeter is present and in working condition /
MESSAGES and/or ERRORS Area)
2 Indication of the FLUOROSCOPY / RADIOGRAPHY phase.
3 Indication of the available thermal units (HUT), expressed in percentage from 0% (indicator
completely white and spent HUT) to 100% (indicator completely black and available HUT).
When the residual HUT value allows a use of the unit below five minutes, under the
indicator will appear also the residual time indication (t = …min), valued in function of the
employed power. Simultaneously it’s issued an acoustic alarm not modifiable.
When the residual HUT reach 0%, the x-rays emission block intervenes.
It’s possible to exceed this limit and proceed with the x-rays emission (only in fluoroscopy
mode) till intervention of monobloc thermal security.
This choice must be performed by Service personnel and on express request of the final
user.
As the monobloc temperature decreases, the HUT will be regenerated. Residual time
indication will disappear, when thermal units allow a use exceeding five minutes.
4 Selection of the Anatomic Curve in Fluoroscopy (Standard / Paediatric / Pelvis-Head /
Standard for sturdy patient (only for rotating anode with camera 1K2)).
5 Indication of the set kV value. It changes in Automatic mode or in Manual mode by using
the kV+ and kV- keys of the control panel.
6 Indication of mA value concerning kV and the selected curve. During the exposure it
indicates the measured mA value.
7 Indication of the AUTOMATIC / MANUAL mode.
8 Dose printing key. It appears if the dosimeter is present and in working condition and if the
printing mode in Utility Mode is selected (§3.6.9).

Pag. 20/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

2.8. Alarm messages


In case of alarm condition, the alarm messages are displayed in an appropriate box with
audio-visible signals.
The key for the dose visualization resetting changes in ALARM RESET.
When an error/warning appears, press the key “ALARM RESET” to reset the audible signal,
delete the message and go on to work.

ALARM RESET

The table indicates how to proceed in case of faults and anomalies.

Text Meaning Intervention

The dosimeter does not reply. Press the “ALARM RESET” key in
order to go on to work.
Check that it is not selected when
DOSIMETER NOT OK
it is not present on the unit.
Call Service to perform the
necessary checks.
At least eleven months are passed Press the “ALARM RESET” key in
from the last performed planned order to go on to work.
MAINTENANCE REQUIRED
maintenance. Call Service to perform the
planned maintenance.
Press the “ALARM RESET” key in
order to go on to work.
Warning! This function results
particularly important for the
REAL TIME CLOCK ERROR The system clock does not work.
calculation of the thermal units
recovery.
Call Service to perform the
necessary checks.
The thermal capacity of the Wait for the thermal units
EXHAUSTED THERMAL UNIT
monobloc is exhausted. recovery.
Continuous fluoroscopy time Release the footswitch and go on
EXHAUSTED XR TIME
exhausted (max 10min) with the operation.
Switch the unit OFF, wait for some
The generator does not seconds and switch ON again.
GENERATOR OFFLINE
communicate with the central unit. In case the error persists, call
Service.
Switch the unit OFF, wait for some
seconds and switch ON again.
CAN-BUS ERROR Error in the field bus
In case the error persists, call
Service.
Switch the unit OFF, wait for some
MEMORY CONTROLLER The memory does not seconds and switch ON again.
OFFLINE communicate with the central unit. In case the error persists, call
Service.

[File:201175-E-00-01.doc] Rev. B - Pag. 21/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Press the “ALARM RESET” key in


MOTOR DRIVER
order to go on to work (it is possible to
COMMUNICATION * The collimator motor does not
go on with XR, if the I.I. field results to
communicate with the central unit.
be free).
* Iris – Rotation – Shutters
In case the error persists, call Service.
Press the “ALARM RESET” key in
Problems found in the filament order to go on to work.
FILAMENT ERROR
management. If the error repeats and persists also at
the next start-up, call Service.
Press the “ALARM RESET” key in
order to go on to work.
KV ERROR Error in the kV generation
If the error repeats and persists also at
the next start-up, call Service.
Press the “ALARM RESET” key in
Error in the mA reading (over the order to go on to work.
mA OVERLOAD
max. allowed limit) If the error repeats and persists also at
the next start-up, call Service.
The temperature of the monobloc has Wait for the tube cooling
THERMIC ALARM reached the max. allowed value.

Pressed UP/DOWN emergency push- Reset the correct working of the


EMERGENCY UP/DOWN
button. emergency push-button by turning it
PUSHED
clockwise.
Switch the unit OFF, wait for some
POWER SUPPLY ERROR Incorrect secondary power supply seconds and switch ON again.
In case the error persists, call Service.
Faulty or damaged footswitch or Disconnect and reconnect the
FOOTSWITCH OR
handswitch for the x-ray control footswitch.
HANDSWITCH ERROR
In case the error persists, call Service.
STARTER FAULT Error in the circuit of the rotating Switch the unit OFF, wait for some
anode. seconds and switch ON again.
Only rotating anode In case the error persists, call Service.
For the units provided with the door Check that the door is closed
OPEN DOOR WARNING control, the door to enter the room is correctly.
open. In case the error persists, call Service.
Switch the unit OFF, wait for some
seconds and switch ON again.
The video signal is not present or it is Check that the trolley is connected to
VIDEO SIGNAL NOT OK
not bright enough. the unit and that the BNC cables are
properly connected to the monitors.
In case the error persists, call Service.
Switch the unit OFF, wait for some
UNAVAILABLE INVERTER
Unavailable inverter power supply seconds and switch ON again.
POWER SUPPLY
In case the error persists, call Service.
The max. exposure time has been Press “ALARM RESET” key to go on,
MAX. X-RAY TIME reached. repeat x-rays.
In case the error persists, call Service.
The preparation handswitch is held Release the preparation handswitch
TIMEOUT down for more than 15 seconds and repeat the operation.
without performing x-rays.
The x-ray hand switch has been Press “ALARM RESET” key to go on
MANUAL X-RAY STOP released before the end of the and repeat the exposure.
exposure.
Press the “ALARM RESET” key in
MISSING PULSES Control pulses lack from the memory order to go on to work.
In case the error persists, call Service.
Press the “ALARM RESET” key in
CCA BUSY Busy central unit control order to go on to work.
In case the error persists, call Service.

Pag. 22/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

2.9. Safety devices


TECHNOS 5/3 is protected against short circuits through a magneto-thermic switch placed
on the unit (pos. 1).
In case the magneto-thermic switch intervenes, in order to reactivate the working of the unit,
place the magneto-thermic switch control in “ON” condition.
Furthermore, the unit has got a mushroom emergency push-button placed on the console
(pos. 2). It allows to stop the UP-DOWN arm movement in case of danger.
In order to restore the UP-DOWN movement, rotate clockwise the push-button body.

A removable operating key (pos. 3), that is necessary for the unit setup, prevents the unit
use from non-authorized personnel.

In case of “High Configuration” trolley, the magneto-thermic switch (pos. 1) and the safety
key (pos. 3) are not on the unit but on the trolley. Furthermore, on the “High Configuration”
trolley there is a mushroom-head emergency push-button that stops the power supply to the
whole unit (pos. 4).

3
2
4

Fig. 15

[File:201175-E-00-01.doc] Rev. A - Pag. 23/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3. UNIT USE
3.1. Transport
Don’t move the unit on planes with inclination higher than 5°.
The unit must be moved only in the condition defined “of transport” (Fig. 16)
and with all the arm movement brakes on.
The uncontrolled arm movement could cause damages to the operator, patient
and personnel that are near the unit.

Before moving the unit, ensure that it is in safety position as shown in Fig. 16:
The sequence described below shows how to put the unit in safety position:

1. Overview position at 0°

2. arm rotation at 0°

3. Horizontal position of the arm group. It is all back.

4. Orbital position of the arm at 0°.

5. Vertical position of the arm group. It is all down.

6. All the safety systems and the brakes on. (cfr.§2.2.1).

Fig. 16

The unit must be OFF, the mains plug removed from the outlet and the cable winded-up on the
cable reel.
The monitor trolley must be disconnected from the Mobile Stand and the cables winded around the
reel.

• check that the arm movement brakes are all ON (locked) (see Fig. 8).
• don't move the unit on planes with inclination higher than 5°.
• move the unit only after unlocking the parking brake and use only the proper
handles (see Fig. 9 pos.3).
• in order to make the movement easier, use the rotation of the rear wheels of the mobile
image intensifier.

Pag. 24/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.2. Connection between monitor trolley – mobile stand


Only for version with monitor trolley

The connection between these parts of TECHNOS 5/3 is performed through a multipolar
connector with bayonet connection.
On “Base Trolleys”, the cable is fixed permanently from the monitor trolley side, whilst it is
sectionable from the unit side.
On “High Configuration” trolley, the cable is sectionable both from the unit side and the
trolley side.

Handle the connectors with care.


Don't unplug by pulling the cable but grip the plug body.

FOR A CORRECT CONNECTOR INSERTION:


• Hold THE SOCKET CONNECTOR IN ORDER TO MAKE LINED UP THE YELLOW ARROWS ON THE
PARTS TO BE JOINED (END TO END).
• TAKE CARE OF INSERTING THE SOCKET CONNECTOR IN ORDER TO BE THE MOST POSSIBLE
PARALLEL TO THE FIXED CONNECTOR. AVOID OBLIQUE INSERTIONS.
• TURN CLOCKWISE THE RUBBER RING NUT OG THE SOCKET CONNECTOR TILL HEARING THE
CLICK THAT ENSURES THE SEAL.

Fig. 17

[File:201175-E-00-01.doc] Rev. B - Pag. 25/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.3. Connection between Footswitch – Mobile Stand


The connection between these parts of TECHNOS 5/3 is performed through a connector
with press connection.

Handle the connectors with care.


Don't unplug by pulling the cable but grip the plug body.

• insert the connector of the footswitch in the appropriate outlet present on the mobile
image intensifier (Fig. 18 pos. A);

Fig. 18

3.3.1. Connection for External Interlocks (optional)

It is possible to use the unit as fixed position so that some warning signals are repeated outside the
operative room.
In detail the repeated warning signals are for:
• a lamp that indicates the unit ON
• a lamp that indicates the x-ray emission
• a door opening contact that stops the x-ray emission.
The connection is performed through a multipolar connector with locking on ring nut (Fig. 18 pos.
B). The prearrangement of the external signals wiring must be performed by the technical
personnel.
For a correct connection it is necessary to:
• ensure that the unit is OFF;
• remove the terminal board closing cap;
• insert the connector of “external interlocks” in the proper socket present on the mobile
image intensifier;
• ensure that the connector is screwed tight.
In order to remove the connection it is necessary:
• unscrew and remove the cable connector
• apply again the closing cap in the terminal board.

Pag. 26/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.4. Unit ON / OFF


Unit with monitor aboard or with “BASE monitor TROLLEY”

The unit TECHNOS 5/3 has got a switching-ON key with three car-type positions.

Key extraction position.


0 position OFF
Unit OFF
Stable position.
I position WAIT
Unit supplied by the mains
Instable position.
II position START
Unit ON

Connect the mobile image intensifier to the monitor trolley


.
(only for the version with monitor trolley)
Connect the unit to the mains and put the magneto-thermic switch in “I”
position.
Insert the ON key and rotate it clockwise in “WAIT” position
Rotate clockwise the ON key in “START” position and release it (push-button
led for the mobile image intensifier console switching-ON lit).
Press the unit ON key.
After the test phase of the internal circuits and the software, the display
shows the initial screen:

Activate the intended exposure mode according to what described in §3.6


“Use mode”.
In order to turn the unit OFF, press OFF key, rotate the key in “OFF” position,
put the magneto-thermic switch in “0” position.

[File:201175-E-00-01.doc] Rev. A - Pag. 27/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Unit with “HIGH CONFIGURATION monitor TROLLEY”

The unit TECHNOS 5/3 has got a switching-ON key with three car-type positions mounted
on the monitor trolley.

Key extraction position.


Position 0 OFF
Unit OFF
Stable position.
Position I WAIT
Unit supplied by the mains
Instable position.
Position II START
Unit ON

Connect the trolley to the mains and put the magneto-thermic switch in “I” position.
Insert the switching-ON key and turn it clockwise in “WAIT” position.
Turn the switching-ON key clockwise in “START” position and release it (the
switching-ON led will light up on the trolley control panel). The trolley is powered
and it can work even without the unit connected.
Connect the mobile image intensifier to the monitor trolley. The connecting cable
can be connected/disconnected even with the powered trolley.
Press the unit switching-ON push-button.
After the test phase of the internal circuits and the software, the display shows the
initial screen:

Activate the intended exposure mode according to what described in §3.6 “Use
mode”.
In order to turn the unit OFF, press OFF key, rotate the key in “OFF” position (on
the trolley), put the magneto-thermic switch in “0” position (on the trolley).

Pag. 28/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.5. Positioning

Don't move the mobile image intensifier or the C-arm with brakes ON.

For movements use the proper handles (see §2.2.1).

Turn the unit ON (ref. §3.4)

Adjust the arm height (motorized movement).

Use the control keys to change the position inside the travel.
Use the Down key with the “Down Enable” key to perform the extra-travel
downwards.

In order to avoid collisions between the monobloc (0,5K x 0,5K version) or


I.I. tube (1K x 1K version) with the front unit leg, the vertical travel is limited
downwards. It is possible to exceed this limit by pressing at the same time
the down key placed on the control panel and the “Enable Down” push-
button placed near the red emergency key. During this operation be careful
to the monobloc or I.I. position in order to avoid collisions and damages to
the unit.

Adjust the horizontal position (manual movement).

Control the possible overview rotation of the C-arm (manual movement).

Adjust the C-arm rotation around the horizontal axis, the goniometric scale
will indicate the rotation angle (manual movement).

Adjust the orbital position of the C-arm (manual movement).

When the positioning has been completed, check that all the braking systems
are activated (see Fig. 8, Fig. 9).

[File:201175-E-00-01.doc] Rev. A - Pag. 29/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Before performing an exposure, ensure that all the necessary precautions


against radiation have been activated.
Before performing an examination, ensure that the residual exposure time
indication is enough to the complete realization of diagnostic investigation.

If the display shows the “TUBE SEASONING” message, it means that the
unit is idle for three months or more and it is necessary to proceed with the
tube seasoning, in order to avoid high voltage discharges that could be
harmful for the x-ray tube.
This operation requires the intervention of a skilled technician.

3.6. Use mode


The foreseen use modes are:

• Continuous fluoroscopy (automatic and manual)


• Pulsed fluoroscopy (automatic and manual)
• Snap Shot
• Radiography
• Utility Mode

Pag. 30/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.6.1. AUTOMATIC CONTINUOUS FLUOROSCOPY

Turn the unit ON according to what described in §3.4 “Unit ON / OFF”. After the
test phase of the internal circuits and the software, the unit is prearranged for the
working in automatic fluoroscopy. The display appears as showed below:

Selection of the RADIOGRAPHY mode (§3.6.5)

Selection of the PULSED FLUOROSCOPY frequency. By pressing the key, the


selection and the indication aside change (according to the installed memory).

FILTER selection. By pressing the key, the selection and the indication aside
change (according to the installed memory).

STANDARD ANATOMICS curve selection. When the function is selected, aside the key,
the keys concerning the anatomic curves that can be selected appear.
ANATOMY curve for Fine Anatomic
Parts or paediatrics. When the key is selected, it appears
in “negative”. By selecting one of
ANATOMY curve for lungs. these curves, the unit proposes
proper kV-mA couples. It is not
ANATOMY curve for head, column, possible to modify the parameters.
pelvis for sturdy patient.
(1/2) mA curve selection. When the key is selected, the STANDARD curve with the
halved mA values is used.

Selection of the MANUAL mode (§3.6.2).

Indication of the residual Fluoroscopy time and the relative resetting key.
The time indicates the effective x-ray passage and it is updated every second.
When 5 minutes are expired, the key is crossed and the audible indicator intervenes.
At any minute, it is possible to press the reset key by taking the time back to the initial 5
minutes and by stopping the audible signal, if it is in working condition.
After 10 minutes of continuous fluoroscopy, the x-ray emission is interrupted. This
supplementary time is resetting automatically at every release of the fluoroscopy
footswitch.
Key for the resetting of the dose value indicated in the central box. The key is active only
if the dosimeter is installed and in working condition.
Hold down for at least two seconds to reset the displayed value.
Press the left pedal of the x-ray control footswitch.
After activating the x-ray control, the kV/mA values can be adapted automatically to the
patient under examination in order to get the best possible image.
By pressing completely the pedal (second step), the image is stored.

X-ray emission LEDs on the keyboard and on the monitor trolley lit.

[File:201175-E-00-01.doc] Rev. A - Pag. 31/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.6.2. MANUAL CONTINUOUS FLUOROSCOPY

By entering MANUAL FLUOROSCOPY mode, the display appears in the following


way:

Selection of the RADIOGRAPHY mode (§3.6.5)

Selection of the PULSED FLUOROSCOPY frequency. By pressing the key, the


selection and the indication aside change (according to the installed memory).

FILTER selection. By pressing the key, the selection and the indication aside
change (according to the installed memory).

Selection of the AUTOMATIC mode (§3.6.1).

Indication of the residual Fluoroscopy time and the relative resetting key.
The time indicates the effective x-ray passage and it is updated every second.
When 5 minutes are expired, the key is crossed and the audible indicator
intervenes.
At any minute, it is possible to press the reset key by taking the time back to the
initial 5 minutes and by stopping the audible signal, if it is in working condition.
After 10 minutes of continuous fluoroscopy, the x-ray emission is interrupted. This
supplementary time is resetting automatically at every release of the fluoroscopy
footswitch.
Key for the resetting of the dose value indicated in the central box. The key is
active only if the dosimeter is installed and in working condition.
Hold down for at least two seconds to reset the displayed value.

Press the left pedal to control the exposure.


After activating the x-ray control, the kV/mA values can be set by the operator. The
exposure parameters variation allows to modify the image quality at the discretion
of the operator. The automatic image control system is disabled.
By pressing completely the pedal (second step), the image is stored.
X-ray emission LEDs on the keyboard and on the monitor trolley lit.

The same functions of Fluoroscopy, both Automatic and Manual, and Storage can
be performed also with the x-ray handswitch (if it is prearranged):
1° step: fluoroscopy control.
2° step: image storage.

Pag. 32/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.6.3. AUTOMATIC AND MANUAL PULSED FLUOROSCOPY

Selection of the PULSED FLUOROSCOPY frequency. By pressing the key, the


selection and the indication aside change (according to the installed memory).

Press the right pedal to control the exposure.


After the kV balance time, the exposure with the frequency set on the display is
performed (the frequency changes according to the features of the installed
memory).

In MANUAL mode, the kV and mAs value must be set by the operator. The
exposure parameters variation allows to modify the image quality at the discretion
of the operator.
X-ray emission LEDs on the keyboard and on the monitor trolley lit.

3.6.4. “SNAP SHOT”

No selection from the keyboard must be performed.

The “Snapshot” can be performed in Automatic and Manual Fluoroscopy mode


without performing any additional choice on the keyboard.

Press the pedal of the X-ray control footswitch.


After the kV balance time, an exposure of about 1 sec controlled directly from the
unit is performed. This exposure allows to get an image that is at high contrast
and without background noise.

X-ray emission LEDs on the keyboard and on the monitor trolley lit.

[File:201175-E-00-01.doc] Rev. A - Pag. 33/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.6.5. RADIOGRAPHY

Select Radiography. The display appears as in the figure below:

Key for the FLUOROSCOPY mode selection (§3.6.1)

Key for the APR 1 mode


selection. (Head) The chosen APR program proposes the default
exposure parameters (kV, mAs, focus). The
Key for the APR 2 mode proposed data can be modified according to the
selection. (Spinal column) requirements.
Key for the APR 3 mode
selection. (Pelvis)
HiPower Radiography selection It appears only if the set mAs are lower than
key 4mAs.

Put the cassette holder at the input of the I.I. tube by coupling the two fixed locks
and by locking the knob, insert the loaded x-ray cassette.
Set the kV and mAs parameters by operating on the relative increase and decrease
keys placed under the indication of the value selected on the display.
Use the whole length of the extensible cable to move away as much as
possible from the irradiated area.
Press the x-ray handswitch and hold it down till the end of the exposure.

X-ray emission LEDs on the keyboard and on the monitor trolley lit.

At the end of the exposure the performed exposure time is displayed.

Pag. 34/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

APR radiography

By selecting the wished APR program, the relative key appears in negative and, on
the right side of the screen, the keys concerning the selection of “slim patient” and
“sturdy patient” appear.

Indication of the selected APR program.

Key to select the APR values for slim patient (visible only if an APR program is
selected).

Key to select the APR values for sturdy patient (visible only if an APR program is
selected).
When the prearrangement for slim or sturdy patient is selected, the icon of the key
in “negative” appears.
The proposed data can be modified at pleasure
by the user. If the storage function is enabled in
the USER SETUP menu (§3.6.9), the modified
Data storage key values can be saved in memory through the
relative key. In case the modified data are not
stored, at next return in Radiography mode, the
default data will be re-proposed.

HiPower radiography

By selecting HiPower radiography, the relative key appears in negative, the keys
concerning the APR programs disappear and the mAs are limited to a value lower
than 4mAs. The kV can be set without limitations.

In this mode it is possible to perform a radiography at 5kW with rotating anode and
at 3,5kW with fixed anode.

[File:201175-E-00-01.doc] Rev. A - Pag. 35/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.6.6. DURING THE EXAM

During every fluoroscopy exam, it is possible:

Select a different image enlargement.

Rotate the image position.

Overturn horizontally the image.

Enlarge or reduce in size the irradiated area through the iris diaphragm
or the parallel shutters diaphragm.

Open completely the iris diaphragm automatically.


+

Open completely the shutter diaphragm automatically.


+

Rotate the position of the parallel shutters diaphragm

Alternate the various fluoroscopy modes simply by passing from a pedal


to the other one of the fluoroscopy footswitch.
Store the current image in continuous fluoroscopy, by pressing the
second pedal of the fluoroscopy footswitch.

Store the current image in pulsed fluoroscopy, by pressing the storage


key.

3.6.7. AFTER THE EXAM

After every exam of fluoroscopy it is possible:

Review the stored images.

Rotate the image position.

Overturn horizontally the image.

For additional images processing, refer to the user’s manual of the memory installed on the unit.

Pag. 36/76 - Rev. B [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

3.6.8. IMAGES TRANSFER TO EXTERNAL NETWORK

In case of unit provided with DTU (Dicom Transmission Unit), it is possible to transfer the images to
a network printer or to a central file for the possible filing or processing. In this case it is necessary
to connect the unit to the network through a connecting cable (it is not supplied with the unit). The
DTU, RJ-45 plug for connector, is present only on the trolleys showed in the following figure.

For the operation of transfer or printing, refer to the DTU User’s Manual.

[File:201175-E-00-01.doc] Rev. A - Pag. 37/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.6.9. Utility Mode


Utility Mode is a particular use mode to set some unit parameters. It is accessible to the final
user at any minute from the Automatic Continuous Fluoroscopy mode (§3.6.1). You enter
Utility Mode by pressing on the keyboard the F1 key and, by holding it down, press the
ENTER key.
The display appears as below:

In this mode it is possible to set the following parameters:

System date (yyyy/mm/dd format).


DATE
It can be modified with the keys INC-DEC aside the icon.
System time (hh:mm:ss format).
TIME
It can be modified with the keys INC-DEC aside the icon.
Display brightness adjustment (from 0 to 100%).
BRIGHTNESS
It’s modifiable by the INC-DEC keys, beside the icon.
Display contrast adjustment (from 0 to 100%).
CONTRAST
It’s modifiable by the INC-DEC keys, beside the icon.
Intensity of the buzzer volume (from 0 to 100%).
VOLUME It can be modified with the keys INC-DEC aside the icon.
It is not possible to modify the volume of the x-ray passage signals.
Storage of the values associated to the APR exams.
APR
The activated function is indicated by the key in negative.

Enabling of the sound matched to the press of every key.


CLICK
The activated function is indicated by the key in negative.

Enabling of the dose printing.


PRINT
The activated function is indicated by the key in negative.
Selection of the commutating time between LIVE image and MEM
FLIP image on single monitor (only for SBFM). It’s modifiable from 1s to
5s, beyond infinite time (“…s”) for applications with DVD recorder.

SERVICE Area reserved to service personnel.

Page with the firmware parameters display, serial number and data
REPORT
for service (see next page).

OK Key of modifications confirmation and exit from USER SETUP.

ANNULLA Key to exit from UTILITY MODE without saving the modifications.

INC-DEC Keys to increase and decrease the value displayed on their left.

Pag. 38/76 - Rev. B [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Utility Mode - Report


The report screen is a particular screen accessible to the user in only reading. It does not
include the data necessary to the normal unit use.

Firmware Version Indication of the firmware versions loaded on the unit.


Serial Number Indication of the monobloc and x-ray tube serial number.
Counter of the total time (expressed in minutes) of the exposures in
Tot. Time Xray – min.
fluoroscopy mode with relative resetting key.
Counter of the total exposures number in Radiography mode with
Tot. Rad Esposure
relative resetting key.
Unit installation data with relative resetting key. Datum inserted at the
Install date
moment of the unit installation.
Limit date foreseen for the unit maintenance with relative resetting key.
Maintenance Date Datum automatically calculated by the unit according to the installation
date or the date of the last maintenance (foreseen yearly).
Key of resetting/activation. The activation/resetting keys are not
accessible to the final user. For the possible data resetting it is
necessary the intervention of Service.
Key to exit from the display.

[File:201175-E-00-01.doc] Rev. A - Pag. 39/76


User's Manual TECHNOS 5/3 Sias S.p.A.

3.7. Shutdown procedure


This section describes the system shutdown procedure in order to ensure operators and
patients' safety and a long unit life.

Don't disconnect the cable between the mobile image intensifier and the
monitor trolley if the mobile image intensifier is not off.
Don't remove the connector from the mains outlet if the unit is not off.

When finished the examination, do the following:

If the unit is provided with DTU (Dicom Transmission Unit), in order to


avoid the malfunctions due to the sudden lack of mains supply, it is
necessary to perform first the correct shutdown of the DTU system.
Only after the DTU shutdown, proceed with the unit switching-OFF.

1. Position the mobile image intensifier in parking position (down, with the mechanical
brakes ON).
2. Turn the mobile image intensifier OFF by operating on the "unit off" key of the control
panel. Rotate the key in OFF position, remove and store it in a safe place.
3. Coil the footswitch cable and put it in the proper place.
4. Disconnect the mains cable and coil it again on the proper supports of the unit.
5. Disconnect the cable of the mobile image intensifier/monitor trolley and coil it on the
proper supports of the monitor trolley
(only for version with monitor trolley)
6. Park the monitor trolley by engaging the wheels brakes
(only for version with monitor trolley)

Pag. 40/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

4. MAINTENANCE
4.1. General warnings
As with any technical device, this system requires:

o proper use;
o regular checks by the user;
o regular maintenance by the authorized personnel

By taking these precautions, the working and the operative reliability of the unit are ensured.
The user of a x-ray unit is obligated to adopt such precautions in compliance with the accident-
prevention standards, the standards on the medical products and other regulations.

As users of x-ray units it is necessary to take these precautions in


compliance with the prevention standards formulated by the laws
concerning the medical equipment.

The unit needs regular checks and maintenances. The purpose of the following warnings is
to keep a good operating and safety level.
The unit includes mechanical parts that are subjected to wear in function of the use. After a
long period of use, it is possible that the safety of the system may decrease due to the parts
wear.
Regular checks and maintenance are necessary to protect the patient and the operator from
damage due to the mechanical parts breakage.
The correct adjustment of the electro-mechanical and electronic modules is essential, as this
has a direct influence on the unit operation, the image quality, the electrical safety and the
exposure level of radiation to which the medical - nursering personnel is subjected.
The maintenance plan includes checks and prevention measures to be done by expressly
authorized personnel and at the unit owner's charge.

In the replacement of any parts that can affect the units safe operation, use
only original spare parts.

4.2. Checks and inspection by the user


The user has to check the possible presence of defects on the x-ray unit (see the following table).
In case some operative defects or other deviations from the standard working occur, the user has
to switch the unit OFF and inform the Service. The working of the x-ray unit can restart only when
it has been repaired. The use of defected components can cause a risks increase for the safety or
a high exposure to radiations.

If a faulty or malfunctioning unit is used, risks to the operators and


patients can increase.

[File:201175-E-00-01.doc] Rev. A - Pag. 41/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Summary of the periodical checks


Frequency Reference Method
Daily Stability tests
Inspection
Daily Faulty leds, damaged components, plates and warning signals
Inspection
Weekly All cables and terminals (damages/breakages).
Oil drippings and unusual noises in high voltage generator.

Inspection
Half-yearly Operation of the arm movement limit switch.
Accessories for the centering of the x-ray group and the image acquisition
unit (marks, lock devices, contacts).

Yearly Contact the technical after-sale service to perform the constancy and Inspection
reproducibility tests, as indicated by IEC 61223-1 and IEC 61223-2-11
standards, as well as the other operating tests of the unit, as instructed in the
planned maintenance plan.

4.3. Cleaning
Please take the following information into consideration before choosing a detergent:
• To clean plastic surfaces, simply use water and soap. If other detergents are used (e.g. with a
high alcoholic content), the material will tend to break or opacify.
• Never use corrosive substances, abrasive solvents or detergents.

For the cleaning, please take the following actions:


• Unplug the unit from the mains before cleaning it.
• Ensure that no liquid seeps into the unit, so as to avoid short-circuiting or corroding
the electrical and electromechanical parts.

• To clean enameled and aluminium surfaces, simply rub them with a wet cloth and a delicate
detergent, after that rub them with a dry wool cloth.
• To clean chromium-plated surfaces, only rub them using dry wool clothes.

Do not use direct water jets or detergent for cleaning. The unit protection level
for liquids is IPXO.

4.4. Disinfection
The used disinfection method must be in compliance with the standards and the directives in force
concerning the disinfection and protection against explosions.

Never use corrosive substances, abrasive solvents or detergents.

In cases where there is a danger that disinfection products may form


inflammable or explosive gaseous mixtures, always ensure that such gases
have dispersed before re-using the equipment.

Before disinfecting the unit, unplug it from the mains.


Disinfect the unit parts, including accessories and connecting cables, by using only a wet cloth
with a disinfecting substance.
Do not use disinfectant spray; it might penetrate the system.
To disinfect with a disinfectant spray the room where the unit is installed, turn it OFF, let cool it and
cover it carefully with plastic sheets. Once the disinfectant gases have dispersed, remove the
sheets and proceed with the disinfection with a cloth.

Pag. 42/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5. TECHNICAL DATA
5.1. Labels and symbols

5.1.1. Unit labels


Pos. 1: S/N Label Pos. 2: WEEE Label

Fixed Anode Version All versions

Rotating Anode Version

2
1

Fig. 19

[File:201175-E-00-01.doc] Rev. E - Pag. 43/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.1.2. Monitor trolley labels

Pos.1: S/N label (present only on versions Pos. 2: WEEE label


with monitor trolley)

2
2 1
1

Fig. 20
Fig. 21

5.1.3. Monobloc labels


Pos. 1: S/N label
Fixed Anode Version

Rotating Anode Version

Fig. 22

Pag. 44/76 - Rev. E [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.1.4. IMAGE INTENSIFIER LABEL


Applied outside the image intensifier

I.I. Tube type A1234BCD * s/n 12345


Power supply type AB12345 ** s/n 123456

Codes present on the label in relation to the installed I.I. tube type:

Brand THALES TOSHIBA


TH 9428 HP2
* I.I. Tube type 9”/6”/4” E5764SDS-T1
H542 VR13
** Power supply type TH 7195 UP 33350

5.1.5. ANTISCATTERING GRID LABEL

X-RAY GRID

Focus: 80 cm
Ratio 8:1 36 L/cm

[File:201175-E-00-01.doc] Rev. A - Pag. 45/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.1.6. UNIT SYMBOLS

ON key label. Emergency stop

Label of “Enable Down”


push-button, enabled
arm extra-travel
downwards

MAIN label

Equipotential node

Brakes lock/release
WHEEE label
label.

Monitor connection
External connections (only version with monitor
aboard)

Hands squashing
Footswitch
hazard

Fluoroscopy and Clutch of the trolley -


radiography control unit connection
(yellow inside)

C-arm rotation degrees


Dose meter printer connection. indication
(black inside)

Label concerning the


warning for the
handle horizontal
positioning.

Pag. 46/76 - Rev. B [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Additional DAP meter filtration label (only in case there is


the dose meter accessory)

Warning about laser radiation hazard (only in case there is


the laser accessory).

5.1.7. MONITOR SYMBOLS

Image contrast

Image brightness

on/off push

5.1.8. FLUOROSCOPY FOOTSWITCH SYMBOLS

Fluoroscopy control +
Image storage

Pulsed fluoroscopy control

SNAPSHOT Control

5.2. Environmental conditions

Environmental Factor In normal use Warehouse and transport

Temperature from 10°C to 40°C from -25°C to 70°C


from 10% to 90% non-
Relative Humidity from 30% to 75% non-condensing
condensing
Pressure from 700hPa to 1060hPa from 500hPa to 1060hPa

[File:201175-E-00-01.doc] Rev.E - Pag. 47/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.3. Electrical data

Description Fixed Anode Version Rotating Anode Version


Voltage 230 Vac ±10% standard monophase
105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request
Frequency 50 Hz standard - 60 Hz on request
Absorbed current in fluoroscopy 4,5A @ 230Vac 5A @ 230Vac
7,5A @ 115Vac 10A @ 115Vac
Absorbed current in radiography 20A @ 230Vac 20A @ 230Vac
28A @ 115Vac 30A @ 115Vac
Absorbed current in stand-by 1,0A @ 230Vac 1,0A @ 230Vac
1,7A @ 115Vac 2,0A @ 115Vac
Line compensation Automatic
Line resistance <0,4Ω @230V
<0,2Ω @115V
Standard mains plug 16A @230Vac
Classification EN60601-1 par.5:
Protection against electrical hazards Class I
Protection against direct and indirect contacts equipment with applied part Type B
Protection against water seepage common protection (IPXO)
Use conditions Continuous operating with intermittent load
The unit is not suitable to be used where there exist mixtures inflammable with air or nitrous oxide

5.3.1. RADIOLOGICAL DATA

Description Fixed Anode Version Rotating Anode Version


Generator, power in DC current 2,5kW @100kV
(IEC 60601-2-7) (100kV, 25mA, 100ms)
Generator, max power in DC current 3,5kW (100kV, 35mA, 91ms) 5kW (100kV, 50mA, 64ms)
mAs < 4 @230Vac; mAs < 4 @230Vac;
2,5kW (100kV, 25mA, 128ms) 3,5kW (100kV, 35mA, 91ms)
mAs < 4 @115Vac) mAs < 4 @115Vac
Max high voltage (fluoroscopy and
110kVp 120kVp
radiography)
Inverter frequency 40 kHz
Max. current in continuous fluoroscopy 8,0mA (STANDARD curve)
Max. current in “SNAPSHOT” fluoroscopy 12mA (0,5K camera)
10mA
30mA (1K camera)
Max. current in radiography 25mA @230Vac
18mA @115Vac
Max. mAs in radiography 125mAs @230Vac
90mAs @115Vac

Description Fixed anode version Rotating anode


version
Max. fluoroscopy time at 75W (75kV-1mA) continuous continuous
Max. fluoroscopy time at 280W (80kV-3,5A) 44 min 54 min
Max. fluoroscopy time at 400W (100kV-4mA) 29 min 38 min
Max. fluoroscopy time at 525W (70kV-7,5mA) 24 min 31 min
Max. fluoroscopy time at 550W (110kV-5mA) 21 min ---
Max. fluoroscopy time at 600W (120kV-5mA) --- 28 min
Max. fluoroscopy time H.U Safety after 21 min H.U Safety after 28 min
of fluoroscopy @110kV, of fluoroscopy @120kV,
5mA (550W). 5mA (600W).
Cooling at 30° 300 min 420 min

Pag. 48/76 - Rev. D [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Fluoroscopy time for fixed anode version

The values indicated in the graph refer to the trends calculated from the software safety procedures that
are preventive concerning the bimetallic contact inside the monobloc that indicates the achievement of
57°C (134,60°F).

Recovery temperature and percentage

[File:201175-E-00-01.doc] Rev. A - Pag. 49/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Fluoroscopy time for rotating anode version

The values indicated in the graph refer to the trends calculated from the software safety procedures that
are preventive concerning the bimetallic contact inside the monobloc that indicates the achievement of
57°C (134,60°F).

Recovery temperature and percentage

Pag. 50/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.3.2. COLLIMATOR

Description Data

Model With iris:


mod.R605/027A/DASM
With iris and parallel shutters:
Mod.R605/027/DASM
Fields dimension (DF 100cm- 39”):
round field delimitation 5±23cm
elliptical field delimitation 0±23cm
Leakage radiation <40mR/h
(EN60601.1.3 par.29.204.3)
Inherent filtration (EN60601.1.3) 0mm
Classification EN60601-1 par.5:
Protection against electrical hazards Class I
Protection against direct and indirect equipment with applied part Type B
contacts
Protection against water seepage common protection (IPXO)

[File:201175-E-00-01.doc] Rev. A - Pag. 51/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.3.3. X-RAY GROUP


X-ray housing data

Description Fixed Anode Version Rotating Anode Version


Monobloc model I-40S 3,5 RF I-40R 5 RF
Max. power (100kV – 35mA) (IEC 601-1) 3,5 kW
Max. tube voltage 110KV 120kV
Ripple at the max. power <2%
kV Rise time at max. power <1ms
Mechanical housing features
Half-value layer @75kV 2,0mmAl
Min. inherent filtration @75kV 1,4mmAl
Weight 15,5 kg (34,17 lb) 19,0 kg (41,89 lb)
Thermal housing features
Thermal capacity 500kJ (670kHU) 600kJ (800 kHU)
Thermal safety 57°C ±3°C (134,60 °F ± 37,40 °F)
Compensation lung 410cm3 (25.015 cubic inch)
Continuous thermal dissipation 75W, 102 HU/sec, 6120 HU/min
Max. housing temperature 60°C (140°F)
X-ray tube filament power supply
Max. current (rms) 500mA 500mA
Leakage radiation (CEI EN 60601-1-3)
<100mR/h a 100cm (39,37 in)

In case it is not used for more than three months, proceed to the tube reset in the following way:

Working parameters Exposure time


To be repeated
kV mA/mAs time ON time OFF
Fluoros 70 3 5' 5' 5 times
Increase of 5kV till
Radiog 70 12,5mAs 0,5 sec 30 sec
kVmax.

If during the procedure some working irregularities or anomalies are found, it is necessary to stop it
for at least half an hour and start it again from the beginning.

Pag. 52/76 - Rev. E [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Monobloc RX I-40S 3,5 RF

dimensions

[File:201175-E-00-01.doc] Rev. D - Pag. 53/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Data of x-ray fixed anode insert

X-RAY INSERT MODEL CEI OX/110-5 PHU 0,5/1,5


Max. peak voltage 110kVP
Max. filament current 4,5A
Nominal foci dimension:
Small focus 0,5mm
Large focus 1,5mm
Nominal anodic power:
Small focus 800W
Large focus 4000W
Anode material Tungsten
Anode inclination angle 12°
Thermal anode capacity 40kJ (54kHU)
Max. continuous thermal dissipation 400W (536 HU/sec)
Min. inherent filtration 0,5mm Al
Tube material Glass

Mechanical dimensions Thermal anode feature

Load curves ( - 3~)

Pag. 54/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Monobloc I-40R 5 RF

dimensions

[File:201175-E-00-01.doc] Rev. E - Pag. 55/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Data of the x-ray rotating anode insert

X-RAY INSERT MODEL IAE X20P 0,3-0,6


Max. peak voltage 130kVP
Max. filament current 5,4A
Nominal foci dimension:
Small focus 0,3mm
Large focus 0,6mm
Nominal anodic power:
Small focus 5kW
Large focus 17kW
Anode material Rhenium / Tungsten / Molybdenum
Anodic diameter 64mm
Anode inclination angle 10°
Thermal anode capacity 150kJ (200kHU)
Max. continuous thermal dissipation 300W
Min. inherent filtration 0,7mm Al
Tube material glass

Anode heating and cooling curves


Mechanical dimensions

0,3mm focus load curves 0,6mm focus load curves

Pag. 56/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.3.4. TOTAL UNIT HALF-VALUE LAYER

Rotating Anode
Description Fixed Anode Version
Version
Half-value layer @75kV 2,8mmAl 2,8 mmAl
Half-value layer @110kV - -
Half-value layer @120kV - -

5.3.5. TOTAL UNIT FILTRATION

Description without DAPMeter with DAPMeter


Monobloc 1,4 mmAl 1,4 mmAl
Collimator 0 mmAl 0 mmAl
Permanent additional filter 1 mmAl 1 mmAl
Monobloc cover 0,1 mmAl 0,1 mmAl
Permanent additional DAPMeter filtration --- 0,3 mmAl
Total unit filtration @75kV 2,5 mmAl 2,8 mmAl

Required min. filtration


≥ 2,5 mmAl eq.
According to the IEC 60601-1-3 par. 29.201.5

[File:201175-E-00-01.doc] Rev. C - Pag. 57/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.4. Exposure mode: fluoroscopy

5.4.1. CONTINUOUS FLUOROSCOPY

Description Fixed Anode Version Rotating Anode Version


Small focus (IEC 336) 0,5mm 0,3mm
kV variation range 40÷110kV 40÷120kV
mA variation range 0,5÷8mA 0,5÷8mA
kV-mA relationship Standard: Standard:
40kV / 0,5mA 40kV / 0,5mA
80kV / 7,6mA 80kV / 7,6mA
110kV / 8mA 120kV / 8mA
Last Image Hold
Safety timer Audible alarm after 5’
X-ray stop after 10’ of uninterrupted fluoroscopy (IEC
60601-2-7 §29.1.104).

5.4.2. PULSED FLUOROSCOPY

Camera 0,5K2
Description Fixed Anode Version Rotating Anode Version

Small focus (IEC 336) 0,5mm 0,3mm


Exposure frequency (selectable from console) 2imm/sec, 1imm/sec, 1imm/2sec, 1imm/3sec,
1imm/5sec.
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

Camera 1K2
Description Fixed Anode Version Rotating Anode Version

Small focus (IEC 336) 0,5mm 0,3mm


Exposure frequency (selectable from console) 2/sec, 1/sec, 1/2sec
Xr flash time Min. time for the best image quality
Other features as per Continuous Fluoroscopy

5.4.3. SNAPSHOT FLUOROSCOPY

0,5K2 Camera
Description Fixed Anode Version Rotating Anode Version

mA variation range 1÷10mA 1÷12mA


Xr flash time < 1sec
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy

1K2 Camera
Description Fixed Anode Version Rotating Anode Version

mA variation range 1÷10mA 2,5÷30mA


Xr flash time < 1sec
Acquisition obtained from the integration of 16 continuative readings.
Automatic storage in non-volatile RAM memory according to the memory.
Other features as per Continuous Fluoroscopy

Pag. 58/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.4.4. (1/2) mA FLUOROSCOPY


Description Fixed Anode Version Rotating Anode Version

mA variation range 0,25÷4mA 0,25÷4mA


kV-mA relationship 40kV / 0,25mA 40kV / 0,25mA
80kV / 3,8mA 80kV / 3,8mA
110kV / 4mA 120kV / 4mA
Other features as per Continuous Fluoroscopy

5.4.5. APR FLUOROSCOPY


Camera 0,5K2

Description Fixed Anode Version Rotating Anode Version

APR 1 (Fine anatomic parts or for 40kV / 0,7mA 40kV / 0,7mA


paediatric use) 80kV / 6,4mA 80kV / 6,4mA
110kV / 6,6mA 120kV / 6,6mA
APR 2 (lungs) 40kV / 0,5mA 40kV / 0,5mA
80kV / 7mA 80kV / 7mA
110kV / 6mA 120kV / 5,5mA
The curves cannot be modified by the operator.

Camera 1K2

Description Fixed Anode Version Rotating Anode Version

APR 1 (Fine anatomic parts or for 40kV / 0,7mA 40kV / 0,7mA


paediatric use) 80kV / 6,4mA 80kV / 6,4mA
110kV / 6,6mA 120kV / 6,6mA
APR 2 (lungs) 40kV / 0,5mA 40kV / 0,5mA
80kV / 7mA 80kV / 7mA
110kV / 6mA 120kV / 5,5mA
APR 3 (Head, spinal column, pelvis) --- 40kV / 1mA
70kV / 15mA
120kV / 9mA
The curves cannot be modified by the operator.

[File:201175-E-00-01.doc] Rev. A - Pag. 59/76


User's Manual TECHNOS 5/3 Sias S.p.A.

kV-mA relationship of camera 0,5K2

kV mA (Fixed Anode Version)

Standard mA(1/2) Snapshot

40 0.50 0.25 0.70 0.50 1.00


50 2.50 1.25 2.40 1.00 4.00
60 5.00 2.50 4.20 3.00 7.00
70 7.50 3.75 5.60 5.00 10.0
80 7.60 3.80 6.40 7.00 10.0
90 7.70 3.85 6.40 6.50 9.00
100 7.80 3.90 6.40 6.25 8.00
110 8.00 4.00 6.60 6.00 8.00
kV mA (Rotating Anode Version)

Standard mA(1/2) Snapshot

40 0.50 0.25 0.70 0.50 1.00


50 2.50 1.25 2.40 1.00 4.00
60 5.00 2.50 4.20 3.00 7.00
70 7.50 3.75 5.60 5.00 10.0
80 7.60 3.80 6.40 7.00 12.0
90 7.70 3.85 6.40 6.50 12.0
100 7.80 3.90 6.40 6.25 11.0
110 8.00 4.00 6.60 6.00 10.0
120 8.00 4.00 6.60 5.50 9.00

kV-mA relationship of camera 1K2

kV mA (Fixed Anode Version)

Standard mA(1/2) Snapshot

40 0.50 0.25 0.70 0.50 1.00


50 2.50 1.25 2.40 1.00 4.00
60 5.00 2.50 4.20 3.00 7.00
70 7.50 3.75 5.60 5.00 10.0
80 7.60 3.80 6.40 7.00 10.0
90 7.70 3.85 6.40 6.50 9.00
100 7.80 3.90 6.40 6.25 8.00
110 8.00 4.00 6.60 6.00 8.00
kV mA (Rotating Anode Version)

Standard mA(1/2) Snapshot

40 0.50 0.25 0.70 0.50 1.00 2.50


50 2.50 1.25 2.40 1.00 4.00 7.50
60 5.00 2.50 4.20 3.00 10.0 15.0
70 7.50 3.75 5.60 5.00 15.0 25.0
80 7.60 3.80 6.40 7.00 13.0 30.0
90 7.70 3.85 6.40 6.50 12.0 25.0
100 7.80 3.90 6.40 6.25 11.0 20.0
110 7.90 3.95 6.60 6.00 10.0 15.0
120 8.00 4.00 6.60 5.50 9.00 10.0

Pag. 60/76 - Rev. C [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.5. Exposure mode: radiography


Two points technique, kV and mAs selection

Description Fixed Anode Version Rotating Anode Version

Large Focus (IEC 336) 1,5mm 0,6mm


kVp range 40÷110kV 40÷120kV
mA range @ 230Vac @ 230Vac
25 mA fixed from 40kV to 100kV; 25 mA fixed from 40kV to 100kV;
22 mA at 110kV 20 mA at 120kV
@ 115Vac @ 115Vac
18 mA fixed from 40kV to 100kV; 18 mA fixed from 40kV to 100kV;
16 mA at 110kV 15 mA at 120kV
mAs range @ 230Vac @ 230Vac
1 – 125 mAs from 40 to 100kV 1 – 125 mAs from 40 to 100kV
1 – 100 mAS from 101 to 110kV 1 – 100 mAS from 101 to 120kV
in 42 steps, curve R’20 in 42 steps, curve R’20
@ 115Vac @ 115Vac
1 – 90 mAs from 40 to 100kV 1 – 90 mAs from 40 to 100kV
1 – 80 mAS from 101 to 110kV 1 – 71 mAs from 101 to 120kV
HiRad mA range HiRad @ 230Vac @ 230Vac
(mAs < 3,6) 35 mA fixed from 40kV to 100kV; 50 mA fixed from 40kV to 100kV;
31 mA at 110kV 30 mA at 120kV
@ 115Vac @ 115Vac
25 mA fixed from 40kV to 100kV; 35 mA fixed from 40kV to 100kV;
22 mA at 110kV 20 mA at 120kV
HiRad Exposure times range @ 230Vac @ 230Vac
28÷90 msec 20÷64 msec
@ 115Vac @ 115Vac
40÷128 msec 28÷90 msec
Exposure times range 0,04÷5 sec
Duty cycle Calculated according to the anode dissipation

5.5.1. APR RADIOGRAPHY

Description Norm

APR 1 (Head) 77kV 56mAs 74kV 45mAs 80kV 71mAs

APR 2 (Lungs) 110kV 11mAs 107kV 9mAs 110kV14mAs

APR 3 (Pelvis) 85kV 22mAs 82kV 28mAs 88kV 18mAs

The curves can be modified by the operator.

[File:201175-E-00-01.doc] Rev. C - Pag. 61/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.6. Image system

5.6.1. IMAGE INTENSIFIER (1)

Description Triple 9/6/4”

Brand Thales
Model TH 9428 HP2
H542 VR13
Fields Number 3
Nominal Input Diameter 230 mm
Output Image Diameter 20 mm
Output Window Thickness 3,6 mm
Useful Input Field Size 215/160/120 mm
Typical Resolution (Central) 48/56/64 lp/cm
Conversion Factor (Cd/m-2/mR/s-1) 240/120/60
Contrast Ratio 23:1/25:1/30:1
Integral Image Distortion 4% / 2% / 1%
Differential Distortion at 90% radius 15% / 6% / 3%
DQE at 59.5 kV 65%
"All metal" Technology Yes
Input Screen "Hi-Res." Yes
MTF at 10 Lp/cm 60/65/70%
MTF at 20 Lp/cm 25/30/40%
Low frequency drop LDF 7/6/5%
Fixed Antiscattering Grid Ratio 8:1, 36 shutters/cm,
focus 80 cm, material
aluminum

(1) According to IEC standard IEC from 1262-1 to 1262-6

Pag. 62/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.6.2. TV CHAIN AND MONITOR


TV chain with CCD 0,5K x 0,5K

Description Data
Camera
Camera technology CCD at low persistence of ½” (470.000pixels)
Video standard CCIR 625/50Hz interlaced with matrix 752x 582 pixels
Aspect ratio 4:3 interlaced scanning
Band width 20 MHz ± 3dB
Signal-noise ratio 65 dB
Resolution 20 lines-pairs (on 6” image intensifier)
Gamma correction 0,4 o 1
Automatic video level Yes
compensation
Dynamic contrast Shading Yes
Video output A/D converter 10 bit
Power supply 24Vdc ±20% 20W
Control Unit Dimensions 226x120x46 mm, 0,550Kg
Camera Head Dimensions Ø85x87mm, standard
Camera Head Weight 0.460Kg, standard

TV chain with CCD 1K x 1K

Description Data

Camera
Camera technology CCD interlain progressive scanning
Active pixels 1024x1024
Acquisition Matrix 1024x1024 pixel, 10bit, rate 25fps
Resolution contrast 1024 grey levels
Sensitivity 0,2 LUX (PB20 light)
Noise signal ratio 60db
Weight (board + optics) 1400gr
Camera head power supply 24Vdc 200mA
Receiver power supply 12Vdc 400ma

[File:201175-E-00-01.doc] Rev. A - Pag. 63/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Monitors

Monitor 17” 18” 19” B/W 19”Color


Standard for unit with Standard for camera Standard per camera Standard for camera
2
0,5K2, opzional for 2
monitor aboard 1K , 0,5K , optional for
optional for camera camera 1K2. camera 1K2.
0,5K2
Model color base LCD 17” monochromatic LCD monochromatic LCD color base LCD 19”
TFT 18.1” TFT 19” TFT B/N TFT
Display angle 180° 170° 178° 178°
Contrast Ratio 1000:1 400:1 typical 700:1 800:1
Resolution 1280 x 1024 1280 x 1024 1280 x 1024 1280 x 1024
Pixel dimensions 0,264 x 0,264 mm 0,280 x 0,280mm 0,294 x 0,294 mm 0,294 x 0,294 mm
Grays scale 256 x 3 =768 levels 256 x 3 = 768 levels 256 x 3 = 768 levels 256 x 3 = 768 levels /
16,7 Mil colors
2
Brightness Max. 350cd/m 750 cd/m2 typical 800 cd/m2 typical 250 cd/m2 typical
luminance
Aspect Ratio 4:3 4:3 4:3 4:3
Response time 25ms 40ms 18ms 18ms
Video Standard CCIR 625/50Hz 625 lines / 100Hz 4:3 VGA / DVI VGA / DVI
EIA 525/60 interlaced
1049/60 1249 lines / 50Hz 4:3
625/100Hz interlaced
1023/60 1249 lines/60Hz 1:1
525/120 interlaced
Connectors Nr.2 Connectors Connector BNC D-Sub 15p. HD D-Sub 15p. HD;
BNC 75ohm 75ohm DVI-D DVI-D;
VGA connector BNC 75ohm per BNC 75ohm per
S-Video connector video compositi (B/N) video compositi.
RCA connector
Brightness / Dedicate knobs OSD menu Control push-buttons Control push-buttons
Contrast
Power supply 115-230Vac, 60/50Hz 100-240Vac, 60/50Hz 100-240Vac, 60/50Hz 100-240Vac, 60/50Hz
Absorption 55W 60W 60W 42W
Weight 4kg (8,82 lb) 9,2kg (20,28 lb) 9,8kg (21,61 lb) 9,8kg (21,61 lb)
Mounting interface VESA 100x100 VESA 100x100 VESA 100x100 VESA 100x100

Pag. 64/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

5.7. Digital image processor


TV chain and Monitor with CCD 0,5K x 0,5 K

Features SBFM 76 SBFM 78/330 SBFM 78/2700

Model SBFM76 SBFM 78/330 SBFM 78/2700


Images number L.I.H. L.I.H. +330 L.I.H. + 2700
Only Ram image (non volatile images) (non volatile images)
Image format 768x576x12 bits 50Hz 768x576x12 bit 50Hz 768x576x12 bit 50Hz
256 gray level 256 grey levels 256 grey levels

A/D Converter 8 bits 10 bit 10 bit


D/A Converter 8 bits 8 bit 8 bit
Sampling frequency 15 MHZ 15 MHZ 15 MHZ
Monitors number 1 2 2
Recursive Filter, noises YES YES YES
reduction, OFF,2,4,8,16 0,2,4,8,16 0,2,4,8,16 0,2,4,8,16
Digital rotation in real time YES YES YES
Vertical image Inversion YES, in combination YES, in combination YES, in combination
with the rotation with the rotation with the rotation
Horizontal image inversion YES YES YES
Grey scale Inversion NO YES YES
Frame rate fps acquisition Single store image Single store image Single store image
Image edge (EDGE) YES YES YES
Patient data editing NO YES YES
Video input Standard CCIR 1 Vpp Composite Video Composite Video
composite video signal signal signal
75 Ohm termination standard CCIR 1Vpp standard CCIR 1Vpp
end 75 Ohm end 75 Ohm
Video output BNC, Standard CCIR 1 BNC, Composite BNC, Composite
Vpp composite video Video signal Video signal
signal standard CCIR 1Vpp standard CCIR 1Vpp
75 Ohm termination end 75 Ohm end 75 Ohm
Power supply 20V to 36V dc 10Vdc ÷ 36Vdc 10Vdc ÷ 36Vdc
Dimensions BOX (26x10,2x6,5) BOX (26x10,2x6,5) BOX (26x10,2x6,5)
Weight .... ...

[File:201175-E-00-01.doc] Rev. A - Pag. 65/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Features DIP3000 A

Model DIP 3000 A


Number of images L.I.H.+192 image buffer ram
Number of images in Hard Disk About 40.000
Video signal CCIR625/50, EIA 525/60, 1Vpp
75 Ohm
Image format working memory 576x576x12bit
Image format 576x576x8bit
A/D converter 10 bit
D/A converter 8 bit
Sampling rate 15 MHZ

Number of monitors 2
Flicker free YES
Smart Filter, motion detector YES
algorithms
Virtual shutter YES
Frame acquisition rate YES, 1,3,6 fps
Recursive Filter, YES 2,4,8,16
Noise reduction, OFF,2,4,8,16
Left and Right image inversion YES
Real time Digital rotation without X- YES
Ray on memory
Cine loop review YES
Programmable frame rate acquisition Yes, 1, 3, 6 fps
Zoom YES, by 2
Overview YES, 4/16 images
Contrast enhancement YES
Edge enhancement YES
Smooth, Normal, Sharp in real time
Digital adjustment for YES
contrast/brightness
Positive/negative image display YES
Text editing YES
Overlay note YES

Directory patients memory YES


Max Opac. YES
Road-mapping, capability for catheter YES
Placement
Real time Subtraction (DSA) YES
Shifting Pixel YES
Land Marking YES

Connection with Dicom YES, optional


Infrared Remote control Optional

Pag. 66/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

TV chain and Monitor with CCD 1K2

Features HRC 1000 type

Images acquirement 1024 x 1024 x 16 bit


Images storage 1024 x 1024 x 10 bit up to 256 frames (option).
D/A converter 8 bits
Video input Digital LVDS type 10 bits
Video output 2x 1249/1049 lines 50/60Hz interlaced
1x 625/525 lines 50/60Hz interlaced
Processing in real time
Images number L.I.H Last Image Hold + 256 images, output for 2
monitors
Images acquirement rates Single image storage up to 256 images in ram or
50.000 images on hard disk (dicom).
Gamma correction Yes, digital
Recursive filter / smart Yes
Digital image rotation LIH Yes
Edge enh 3x3 Yes
Dynamic range 1:2 1:3 1:4 1:5
Post Processing
Grey scale inversion Yes
Brightness and contrast Yes
Overview, Windowing Yes 4-9-16 images, 1+5 or 1+7 images
H/V inversion Yes
Electronic zoom Yes
Step rotation 90° Yes
Continuous fluoroscopy Yes
LIH Yes
Pulsed fluoroscopy Yes
Snapshot Yes
LIH Autostore Yes
LIH marker Yes
Image numeration Yes
Overview 4/9/16 images Yes
Patients file Yes
Images subtraction Yes
Electronic diaphragms Yes
Measures
Distances calculation Yes
Angles Yes
Dicom Options
Dicom VERIFY (SCU/SCP)
Dicom STORAGE (SCU)
Dicom WORK LIST (SCU)
Dicom PRINT (SCU)
Dicom CDR/DVD
Dicom QUERY/RETRIEVE (SCU)

[File:201175-E-00-01.doc] Rev. E - Pag. 67/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Features HRP 2000


Acquirement
Images acquirement From digital CCD 1024x1024 up to 12 bits
Pulsed at high dose up to 12 images/second
1024x1024x12bit
Fluoroscopy 25 images/second 1024x1024x8bit
Fluoroscopy with recursive filter and movement
holding
Linear integration up to 16 images in pulsed
acquirement
Images storage 1024 x 1024 x 12 bit
Images storage on H.D SCSI type 36 Gbyte,
36.000 images with 1kx1k 8 bit
18.000 images with 1kx1k 12bit
144.000 images with 512 x512x8 bit
Video input Digital, LVDS type 10 bits
Video output D/A converter 10 bit, VGA 1280x1024 60Hz,
interlaced 325 lines 50Hz only for printer.
Electronic rotation at 1° steps Yes
Horizontal and vertical inversion Yes
Grey scale inversion Yes
Brightness and contrast Yes
Max. opacity fluoroscopy acquirement Yes
Subtraction in real time with manual/automatic mask Yes
Programmed acquirement sequences Yes, 3, 6, 12, 25 fps

Post Processing
Image horizontal and vertical inversion Yes
Brightness and contrast Yes
Grey scale inversion Yes
Edges enh Yes
Cineloop Yes
Electronic collimator (rectangular and circular) Yes
Shifting pixels Yes
Electronic zoom factor from 1,2 to 3 Yes
Electronic lens factor from 1,2 to 3 Yes
Images sum for noise reduction Yes
Images composition for research of the max. contrast Yes
liquid concentration
Land marking Yes
Overview 4-9-16 images Yes
Image subtraction Yes

Measures
Distances calculation Yes
Angles Yes
Stenosis Yes
Text overwriting Yes
Dicom options
Dicom VERIFY (SCU/SCP)
Dicom STORAGE (SCU)
Dicom WORK LIST (SCU)
Dicom PRINT (SCU)
Dicom CDR/DVD
Dicom QUERY/RETRIEVE (SCU)
Dicom MPPS (CPU)
Dicom STORAGE COMMITMENT (SCU)

Pag. 68/76 - Rev. E [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

DICOM2, DICOM3 Digital Imaging and Communications in Medicine (Option)

Description Data
Network connection features: ESIDIC2, ESIDIC3
Speed Ethernet autosense 10/100 M bit/s
Protocol TCP/IP
IP address Personalizable in 4 byte
IP mask Personalizable in 4 byte
Dicom connection features, Functions performed:
Dicom VERIFY (scu/scp) Supported
Dicom STORAGE (SCU) Supported
Dicom WORKLIST (SCU) Supported
Dicom PRINT (SCU) Supported
Dicom CDR/DVD (media Interchange) Supported
Dicom MPPS (SCU) Not available
Dicom STORAGE COMMITMENT (SCU) Not available
Dicom QUERY/RETRIEVE (SCU) Not available for ESIDIC 2.
Available for ESIDIC 3.
Dicom port address Settable
Video Output on live LCD monitor High resolution 1249/50 Hz, 1049 60 Hz
Video output VGA, 1280/1024 60 Hz
Hard disk 80 GB or more

DICOM Terminology

Terminology Description
Dicom VERIFY (SCU/SCP) It allows the connection check, in both ways, with
Dicom units present on the network.
Dicom STORAGE (SCU) It allows to send the images to a Dicom server for
filing.
Dicom WORKLIST (SCU) It queries and receives from a Dicom server the
patients list to be examined on the acquirement
system.
Dicom PRINT (SCU) It sends to Dicom printer the images to be printed in
panoramic way and through film composer.
Dicom CDR/DVD (media Interchange) It burn on Cd or DVD the patients images with the
possibility to add a display program.
Dicom MPPS-Modality Performed Procedure Step It informs the server that the examination is in
(SCU) operation and then that it has been completed by
sending the reference indication to the images and, if
available, the total acquirement dose.
Dicom STORAGE COMMITMENT (SCU) It asks and wait for the confirmation to the storage
server that the images sent are saved in a safe way
by allowing the cancellation of the acquirement unit
from the file.
Dicom QUERY/RETRIEVE (SCU) It queries and receives from a server the images of a
patient data for consultation.

SCU (SERVICE CLASS UNIT), unit that asks a Dicom service to a unit SCP (SERVICE CLASS PROVIDER)
that is able to supply such service.

[File:201175-E-00-01.doc] Rev. E - Pag. 69/76


User's Manual TECHNOS 5/3 Sias S.p.A.

5.8. Accessories

Description Standard Optional

X-ray handswitch with extensible cable •


Fluoroscopy control by triple footswitch •
Cassette-holder for radiography on cassette 24x30cm •
Cassette-holder for radiography on cassette 18x24cm •
Cassette-holder for radiography on cassette 10x12inch •
Sterile drapes set •
Thermal dry film printer Sony UP 970 / UP 990 or equivalent •
Medical Image Capture Device MediCap USB200 •
USB Data: 512 MB, 350 Kb per image, about 1450 image •
USB Data: 2 GB, 350 Kb per image, about 5700 image •
DVD Recorder Sony DVO 1000 MD. Automatic start- stop with DIP3000 •
DVD Recorder for Fluoro acquisition, Only manual Star Stop •
Dicom Utility ESIDIC2 (only for DIP3000 memory) •
Dicom Utility ESIDIC3 (only for HRC memory) •
Laser Targeting device (only on I.I. tube) •
Dose meter with ionization chamber (DAP) Diamentor CM-T •
Printer for dose meter S Sprint s (on unit) •
Collimator with iris and parallel shutter •
Equipotential node cable lg.300 •
Extractable alphanumeric keyboard (only for SBFM78) •
External interlocks kit •
Movement handles on “C” arm •
Movement handle on I.I. tube •

Laser targeting device, for dose reduction (only on I.I. tube)

Description Data
Wave length 635nm
Optical power of the laser diode < 4 mW
Optical power of the collimated beam < 3,5 mW
Divergence 6,67 mRad
Optical power on the operation plane < 3 mW
Safety class 1A

Dose meter with ionizing chamber PTW, DIAMENTOR CM-T ( (DAPMeter)

Description Data
Measurement amplifier
Measurement category Dose area product
2
Measuring unit μGy m
2
Digital resolution 0,01 μGy m
Measurement category
Measurement category Dose area product rate
2
Measuring unit μGy m /s
2
Digital resolution 0,01 μGy m /s
Nominal useful range
Dose rate range 0,10 to 400 mGy/sec
Max field dimension φ 6 cm
Min field dimension φ 0,6 cm
2
Dose area product range rate 0,01 to 4500 μGy m /s
Tube voltage 40 to 120 kV
Chamber: Type TA34037
Dimension of chamber field φ 7,2 cm
Max energy dependence ± 8% according to IEC60580, Table 6
Hardening equivalent value < 0,3 mmAl / 70 kV according to IEC60522/1999

Pag. 70/76 - Rev. E [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

Thermal Dose Meter Printer Custom S print s type

Printing method Thermal line printing


Resolution 203 dpi
Printing speed >50mm/sec (it depends on the printing typology and the environment
temperature)
Paper width (mm 58mm
Roll dimensions (mm) 57.5 ±1
Print area 48mm
Interface RS-232
Power Supply 18±24VDC / 0,6A
Power consumption (print) 925mA
Operating Temperature 0±50°C, -20±70°C, without paper roll
Storage
Humidity 10±85%, there must be no condensation
Operating Storage 10±90%, there must be no condensation, without paper roll
Dimensions (WxDxH) 146 x 88 x 65mm
Weight 370gr ( without paper roll )
Safety EN60950+A1+A2+A3+A4

Thermal printer Sony UP970CE or UP990CE

Description Data
Print method Thermal printer with multi image picture ( UP 960 only 2 images )
Sublimation print, not available
Thermal head 1280 elements
Grey levels 256 levels
Sheet format Amplitude from 210 mm
Print dimensions UP990CE Standard mode CCIR 187 x 138 mm ; EIA 187 x 140 mm
Side mode CCIR 186 x 249 mm ; EIA 188 x 249 mm
Print dimensions UP970CE Standard mode CCIR 190 x 142 mm ; EIA 190 x 144 mm
Side mode CCIR 181 x 243 mm ; EIA 184 x 243 mm
Printer memory UP960CE 2048 x 1024 x 8 bit; UP980CE 2048 x 2048 x 8 bit:
Printer time Standard CCIR about 12 sec.; Standard EIA about 10 sec.
Inputs to the printer CCIR or EIA videocomposed BNC 1.0 Vpp, 75 ohm high imped
Output from printer CCIR or EIA videocomposed BNC 1.0 Vpp, 75 ohm high imped
Power supply 120 ÷ 240V 50 / 60 Hz
Size 316L x 132H x 305D mm
Weight 8 Kg
Paper Thermal paper, dry blue film only for UP 980
Accessories Remote control

[File:201175-E-00-01.doc] Rev. A - Pag. 71/76


User's Manual TECHNOS 5/3 Sias S.p.A.

Sony DVD Recorder DVO 1000MD

Description Data
General data:
Recording System DVD recording, NTSC/PAL, Switchable
Recording Format Video: MPEG-2 compression
Audio: Dolby Digital Format
Recordable media DVD+RW ( 2,4x and 4x speed )
Recording time HQ mode: 60 minutes
SP mode 120 minutes
LP mode 180 minutes
Medical Standards EN60601-1, EN60601-1-2, UL60601-1,CSAC22.2 N° 60601-1
Power Requirements From 100 to 240 Vac 50/60 Hz
Power Consumption 35 W
Operating temperature +5° to +40° C
Storage temperature -20° to +60° C
Operating Humidity From 20 to 80 %
Size 212 x 128,5 x 382 cm
Mass 6 Kg
Input / Output
Analog Composite Input BNCx2, with loop-through, unbalanced 1.0 Vpp 75 Ohm
S-Video Input 4 pin DIN x2, with loop- through
Y: 1Vpp, 75 Ohm unbalanced
C: 0,286 Vpp ( NTSC)/ 0,3 Vpp (PAL), 75 Ohm, unbalanced
Analog Composite Output BNC x1, 1±0,2 Vpp, 75 Ohm, unbalanced
S Video Output 4 pin DIN x1
Y: 1Vpp, 75 Ohm unbalanced
C: 0,286 Vpp ( NTSC)/ 0,3 Vpp (PAL), 75 Ohm, unbalanced
Analog Audio Input RCA Pin x2 /L/R), 2 Vrms /full bit), input impedance 47 kOhm
Analog Audio Output RCA Pin x2 /L/R), 2 Vrms /full bit), load impedance 47 kOhm
Monitor Audio Output Monitor RCA Pin x1 /L/R), 2 Vrms /full bit), load impedance 47 kOhm
ILINK (DV IN) ILINK 6-pin x1 IEEE1394
Remote Control RS-232C x1 D-sub 9-pin
USB 2.2 x1 (Full Speed )
Remote 2 x1, stereo mini jack for connection with optional FS-20 foot
switch

Medical Image Capture Device MediCap USB200

Description Data
Model MediCap USB200. Digital medical Image capture device.
Saves digital images and video sources to a USB flash drive or USB
hard drive.
Equipped with USB Data: 2GB, 350 Kb per image, about 5700 image
Case Rugged metal case with high-impact plastic faceplate
Buttons Sealed membrane, fluid resistant
Video PAL / NTSC (switch selectable)
S-video or composite BNC connections for both input and output
Foot Switch Hands-free capture. Standard 3.5 mini jack
Mechanical Size: 240 x 200 x 63mm / 9.5 x 9.7 x 2.5” Weight: 1.7 Kg / 3lb
Environmental Storage: -40° to +85° C - Operating: -20°n to +40° C
Power 100/240Vac 50-60Hz 20W 3-prong AC jack
Media USB flash drives (MediCapture brand recommended) or external USB
hard drive
Image Formats DICOM, JPEG, TIFF, PNG
Video Formats MPEG2 PS, DVD-quality. Quality Levels:
Low (DVD-): 352 x 240 pixels, 1.5 Mbps Variable Bit Rate (VBR)
Normal (DVD): NTSC: 720 x 480 pixels, 3.5 Mbps VBR
PAL: 720 x 576 pixels, 3.5 Mbps VBR
High (DVD+): NTSC: 720 x 480 pixels, 8 Mbps VBR
PAL: 720 x 576 pixels, 8 Mbps VBR
Audio format MPEG1 L2, 2 channels, 16 bits/channel, 48 kHz sampling 256 kbps
bitrate
Image Size 1024 x 768 pixels, 800 x 600 pixels, 640 x 480 pixels
Certifications IEC, UL, CSA, FCC

Pag. 72/76 - Rev. D [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

6. MECHANICAL DATA
6.1. C-arm unit dimensions with I.I. 9”

6.2. Axis and interventionist reference point

300 [11,81

Interventionist
reference point

Reference axis

[File:201175-E-00-01.doc] Rev. A - Pag. 73/76


User's Manual TECHNOS 5/3 Sias S.p.A.

6.3. “Base Trolley” monitor trolley dimensions

max.
admissible
weights on the
single shelves:
2 kg.

10 kg.

In the version
“Low Profile
Base Trolley” for
systems 0,5K2,
the two lower
shelves are not
available.

Pag. 74/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

6.4. Dimensions of “High Configuration” monitor trolley

[File:201175-E-00-01.doc] Rev. A - Pag. 75/76


User's Manual TECHNOS 5/3 Sias S.p.A.

6.5. Mechanical data, dimensions and weights

“Base Monitor Trolley” “High Configuration”


Description C-arm unit
Monitor Trolley
Weight 340 kg (unit with fixed “Low Profile” “High Profile” 150Kg
749,57 lb anode) 67kg 97 kg 330,69 lb
360kg (unit with 147,71 lb 213,85 lb
793,66 lb rotating anode)
(1) Width 859 mm (33,82 in) 869 mm 869 mm 872 mm
34,21 in 34,21 in 34,33 in
(2) Depth in movement 2000 mm (78,74 in) 855 mm 855 mm 653 mm
position 33,66 in 33,66 in 25,71 in
(3) Height in movement 1800 mm (70,87 in) 1606 mm 1606 mm 1603 mm
position (camera 1k) 63,23 in 63,23 in 63,11 in
1780 mm (70,08 in)
(camera 0,5k)

(A) Vertical motion Total travel 500 mm (19,69 in) , motorized in 60 sec

(B) Horizontal motion Manual, 210 mm (8,27 in)

(C) C-arm rotation around the


Manual, ±270°
horizontal axis

(D) Orbital rotation Manual, 123° (+90° ÷ - 33°)

(E) C-arm group overview Manual, ±12,5°

(F) Useful space 770 mm (30,31 in)

(G) C-arm depth 690 mm (27,17 in)

971 mm (38,23 in) Rotating Anode


(H) S.I.D.
988 mm (38,90 in) Fixed Anode
200 mm (7,87 in) Rotating Anode
(I) Focus-Skin Distance
218 mm (8,58 in) Fixed Anode
152 mm (5,98 in) Rotating Anode
(L) Minimum distance from the floor
172 mm (6,77 in) Fixed Anode
357 mm (14,06 in) Rotating Anode
(M) Distance Floor - Focus
339 mm (13,35 in) Fixed Anode
(N) Distance Floor - Skin 557 mm (21,93 in)
(O) Min. distance between I.I. center – floor in
954 mm (37,56 in)
oblique projection
C-arm unit:
Rear driving wheels with manual control by the operator,
Unit movement front castor. Manual parking brake.
Monitor trolley:
Front and rear castors. Brake on the front wheels
Rear 125 x 40 mm (4,92 x 1,57 in)
Unit
Front 80 x 30 mm (3,15 x 1,18 in)
Wheels Diameter 80 x 35 mm (nr.5)
Base Trolley
(3,15 x 1,38 in)
Trolley
125 x 30 mm (nr.4)
High Configuration Trolley
(4,92 x 1,18 in)
Protection against the cables squashing Core hitch on all the C-arm unit wheels.

The measures are referred to the figure at page 73 (§6.1 -C-arm unit dimensions with I.I. 9”)

Pag. 76/76 - Rev. A [File:201175-E-00-01.doc]


Sias S.p.A. TECHNOS 5/3 User's Manual

7. DOCUMENT STATUS
Rev. Date Page/s Modification description
- 30/10/2007 - Document approval
A 28/04/08 All General document revision.
6, Note addition on the unit versions.
10, 11, Note addition about the use of the overview movement handle.
21 “Error of the field bus” alarm addition.
B 28/11/08 25, Note addition about the unit-trolley connection cable insertion.
36, Added automatic opening of the collimator diaphragms.
38, Modified “User Setup” page.
46, 47 Added warning label.
Unification of the SNAPSHOT function wordings, added data of
14,18, 20, HiRad exposure time range, modification to the small focus from 0,6
C 06/04/09 44,47,48,54,5 to 0,5 on the OX/100-5 insert concerning the fixed anode monobloc
7,58,60,61 and various corrections in the document. Updated description at
point 3 of the “Operative Messages”
7, 44, 48,
Introduced the new model of fixed anode monobloc and inverter.
D 10/09/09 52, 53.
Added DICOM function, new Medicap USB200 model.
68, 69, 70, 72
Introduced the new model of rotating anode monobloc and inverter.
7, 43, 44, 47,
Upgrading of manufacturer address and relative S/N labels
E 26/10/09 52, 54, 55, 61,
Upgrading of data table HRP HRC DICOM and laser targeting device
67, 68, 69, 70
data.
F

[File: 201175-E-10-01.doc] Pag. I/I

You might also like