LC-40D X3 CL Lc-40D XR CL: Instruction Manual
LC-40D X3 CL Lc-40D XR CL: Instruction Manual
LC-40D X3 CL Lc-40D XR CL: Instruction Manual
Sep. 2022
LC-40D X3 CL
LC-40D XR CL
Instruction Manual
Read this manual thoroughly before you use the product.
Keep this manual for future reference.
This page is intentionally left blank.
Introduction
Read this Instruction Manual thoroughly before using
the product.
Thank you for purchasing this product.
This instruction manual describes the basic operation, and accessories and options for this
product. Read this manual thoroughly before using the product and operate the product
in accordance with the instructions in this manual.
The following instruction manuals are included with the product in booklet form or in
PDF format. The PDF documentation is on the instruction manual DVD-ROM (Part No.
228-97193-41).
Important
• If the user or usage location changes, ensure that this Instruction Manual is always
kept together with the product.
• If this manual or a product warning label is lost or damaged, immediately contact
your Shimadzu representative to request a replacement.
• To ensure safe operation, read the accompanying booklet "Safety Guideline"
before using the product.
• To ensure safe operation, contact your Shimadzu representative if product
installation, adjustment, re-installation (after the product is moved), or repair is
required.
LC-40D X3 CL LC-40D XR CL i
Introduction
Notice
• Information in this manual is subject to change without notice and does not represent
a commitment on the part of the vendor.
• Any errors or omissions which may have occurred in this manual despite the utmost
care taken in its production will be corrected as soon as possible, although not
necessarily immediately after detection.
• All rights are reserved, including those to reproduce this manual or parts thereof in
any form without permission in writing from Shimadzu Corporation.
• Nexera and LabSolutions are trademarks of Shimadzu Corporation.
Windows is a registered trademark of Microsoft Corporation in the United States and/or
other countries.
Third-party trademarks and trade names may be used in this publication to refer to
either the entities or their products/services, whether or not they are used with
trademark symbol "TM" or "®".
Indication Meaning
Indication Meaning
ii LC-40D X3 CL LC-40D XR CL
Electromagnetic Compatibility
Electromagnetic Compatibility
Descriptions in this section apply only to the following models:
• 228-65076-55 LC-40D X3 CL
• 228-65000-55 LC-40D XR CL
This product complies with European standard EN61326, class B for electromagnetic
interference (Emissions) and industrial electromagnetic environment (Immunity).
n Calibration
n Quality Control
Implement quality control of the instrument by routinely measuring at least one each of
four types of quality-control samples (a sample with a normal concentration level, a
sample with a concentration level higher than normal as well as a sample lower than
normal, and a blank sample). Check the selectivity/specificity in the matrix used for the
actual sample, the accuracy, correctness, recovery rate, dynamic range, linearity, and lower
limit of quantitation.
Based on these assessment results, check that the results are within the permissible range.
If the assessment results are out of range, the measured data may be invalid, so do not
use analysis results obtained from the instrument in question until it can be confirmed
that it is functioning normally. When, for example, analyzing samples that contain
complex sample matrices such as serum, plasma or urine, it may be possible to obtain
stabilized data by using an appropriate pre-treatment or an internal standard.
Parts Identification 7
2 and Function
2.1 Front ..................................................................
iv LC-40D X3 CL LC-40D XR CL
Contents
LC-40D X3 CL LC-40D XR CL v
Contents
Technical 125
5 Information
5.1 Specifications ...................................................
vi LC-40D X3 CL LC-40D XR CL
Contents
1.1 Features
By reducing the discharge volume per plunger stroke to the microvolume level (10 µL),
and applying high-speed drive, the flow pulsation and pulsation period have been
reduced to levels significantly lower than those for other instruments.
LC-40D X3 CL/LC-40D XR CL can be used in both low- and high- pressure gradient modes.
The high-pressure gradient mode uses two pumps and is highly accurate with minimal
time lag. In low-pressure gradient mode, a single pump unit plus the gradient unit
provides a low cost option to combine up to four solvents.
There are two gradient modes available: fast LC mode or normal mode under control of
the system controller.
The fast LC mode for binary high-pressure gradient delivery supports ultra high speed
high separation analysis with minimized gradient control steps.
LC-40D X3 CL enables ultra high pressure solvent delivery up to 130 MPa supporting ultra
fast high separation analysis in combination with Shim-pack XR-ODS III or other ultra fast
high separation columns. For LC-40D XR CL, high-pressure solvent delivery up to 70 MPa
is possible so high speed high separation analysis is supported in combination with
Shim-pack XR-ODSⅡcolumns.
LC-40D X3 CL LC-40D XR CL 1
1 Overview
2 LC-40D X3 CL LC-40D XR CL
1.3 Optional Parts
(w/ mixer
Mixer installation kit Gradient mixer (40 µL internal capacity)
recognition
(MR40 LPGE, to be exclusively for low-pressure gradient analysis.
device)
built in) Fixed inside of the instrument.
228-65020-41
(w/o mixer
Mixer MR40 LPGE Gradient mixer (40 µL internal capacity)
recognition
(fixed inside column exclusively for low-pressure gradient analysis.
device)
oven) It can be installed inside of the CTO-40C CL.
228-45210-41
(w/ mixer
Gradient mixer (300 µL internal capacity)
recognition
exclusively for low-pressure gradient analysis.
device)
It can be installed inside of the CTO-40C CL.
228-72653-42
Mixer MR300 LPGE This option can be used for solvents that are
(w/o mixer
hard to mix, such as TFA (trifluoroacetic acid), or
recognition
for reducing the baseline fluctuation of the
device)
photo diode array detector.
228-45210-42
Gradient Mixer for
High-Pressure Gradient "Gradient Mixer for High-Pressure Gradient Analysis" P.5
Analysis
LC-40D X3 CL LC-40D XR CL 3
1 Overview
4 LC-40D X3 CL LC-40D XR CL
1.3 Optional Parts
The following gradient mixers for high-pressure gradient analysis are available for liquid
chromatographs.
1
Part Name Part No.
(w/ mixer recognition device) 228-72652-41
MIXER,MR 20 µL
(w/o mixer recognition device) 228-45209-41
(w/ mixer recognition device) 228-72652-42
MIXER,MR 40 µL
(w/o mixer recognition device) 228-45209-42
(w/ mixer recognition device) 228-72652-43
MIXER,MR 100 µL
(w/o mixer recognition device) 228-45209-43
(w/ mixer recognition device) 228-72652-44
MIXER,MR 180 µL Ⅱ
(w/o mixer recognition device) 228-45209-44
Recommended Mixers
The table below lists the mixers with the smallest volume among mixers that can be used
in individual combinations of detectors and mobile phases.
Mobile Phase
Water (Including
Solvent Containing Solvent Containing Buffer Solution),
Trifluoroacetic Formic Acid, Organic Solvent,
Acid (TFA) Acetic Acid, etc. or Its Solvent
Mixture
Photo Diode Array
MR 180 µL MR 180 µL MR 180 µL
Detector
Detector
UV Detector MR 180 µL MR 100 µL MR 40 µL
LCMS, LCMSMS - MR 20 µL MR 20 µL
Using a mixer with a volume smaller than the recommendation shown above may cause
excessive fluctuation of the baseline.
When connecting multiple detectors, use a mixer with the largest volume among the
applicable mixers. For example, if using a Photo Diode Array and an LC/MS, the MR 180
µLⅡmixer is recommended.
A larger than recommended mixer can be used to improve mixing performance, but will
also increase the gradient delay volume.
LC-40D X3 CL LC-40D XR CL 5
1 Overview
The plunger can be rinsed with the automatic rinsing kit at the back part of the plunger
seal. LC-40D X3 CL/LC-40D XR CL has the automatic rinsing kit as standard. When using
buffer solution, be sure to attach the automatic rinsing kit.
See the table below for the necessity of the automatic rinsing kit and rinsing solution.
LC-40D X3 CL/LC-40D XR CL
Mobile Phase Recommended Rinsing
Necessity
Solution
Buffer Solution Necessary 10 % 2-propanol water
Normal-phase solvent such as hexane Not necessary -
Other Solutions Necessary 10 % 2-propanol water
• For the application system, refer to the instruction manual of the application
system.
• If the rinse solution is dirty, replace with new solution. If getting significantly
dirty, replace with new solution once a day. Even if it does not appear dirty,
replace with new solution once a week.
If using distilled water as rinsing solution, replace it once a day.
6 LC-40D X3 CL LC-40D XR CL
2 Parts Identification and Function
2.1 Front
LC-40D X3 CL LC-40D XR CL 7
2 Parts Identification and Function
9 Leak Sensor
8 LC-40D X3 CL LC-40D XR CL
2.3 Back
2.3 Back
*1 Normally use (power button) of the operation panel or system controller to turn ON/OFF the
power. If the system controller is connected and the main power switch is on, the user can use
(power button) of the system controller to turn ON/OFF the power from the front of the instrument.
If the instrument is not used for a long time, turn off the main power switch. Before turning off
the main power switch, be sure to turn OFF the power using the power button.
"2.5 Name and Functions of the Operation Panel" P.11
LC-40D X3 CL LC-40D XR CL 9
2 Parts Identification and Function
10 LC-40D X3 CL LC-40D XR CL
2.5 Name and Functions of the Operation Panel
When turning off the main power switch on the back of the instrument after 2
changing the parameters via panel operation, be sure to turn OFF the power
with the power button at the front of the instrument, and turn off the main
power switch. Otherwise, some of the changed parameters may return to their
original values.
LC-40D X3 CL LC-40D XR CL 11
2 Parts Identification and Function
*1 If CBM-40 CL is used, pressing the power button of the pump turns off the power of the entire
system.
Note, however, that the following will apply in the case of a system with multiple pumps.
- When using CBM-40 CL: The system power can be turned off using the power button of the
pump with the lowest optical link number.
12 LC-40D X3 CL LC-40D XR CL
2.5 Name and Functions of the Operation Panel
LC-40D X3 CL LC-40D XR CL 13
2 Parts Identification and Function
Use the keys on the front to operate or configure the instrument. Touch the key to make
the status LED brighter, accept input when you release the key.
14 LC-40D X3 CL LC-40D XR CL
2.5 Name and Functions of the Operation Panel
Regarding the acceptable part of the arrow keys on the touch panel.
When operating the arrow keys, it
responds by touching the part including
the upper and lower range of each key
indication on the operation panel.
No. Description 2
1 Acceptable part of [↑] key input.
2 Acceptable part of [→] key input.
3 Acceptable part of [↓] key input.
4 Acceptable part of [←] key input.
LC-40D X3 CL LC-40D XR CL 15
3 Operation
3.1 Settings for the Basic Operation
Before operating this instrument, it is necessary to set the flow rate and the pressure
limit to protect flow line parts such as columns.
See the product specifications before use.
"5.1.1 LC-40D X3 CL Specifications" P.125
"5.1.2 LC-40D XR CL Specifications" P.126
Touching the central part of the operation panel displays the operation keys that the
user can operate.
The following is the basic procedure for setting a flow rate [FLOW].
2 Press [Enter].
The cursor blinks at the input position indicating that the flow rate [FLOW] can be input.
3 Input the flow rate with the numeric keypad and press [Enter].
The flow rate [FLOW] is set and the display returns to the initial screen.
16 LC-40D X3 CL LC-40D XR CL
3.1 Settings for the Basic Operation
4 Input the compressibility [COMP] for the solvent to be used (0.45 for water) with 3
the numeric keypad.
"Setting the compressibility of solvent《COMP》" P.32
n Setting Procedure
The following is an example to set the maximum pressure [P.MAX] to 15.0 MPa.
2 Press [Enter].
The cursor blinks at the input position
indicating that [P.MAX] can be input.
LC-40D X3 CL LC-40D XR CL 17
3 Operation
The allowable maximum pressure is 130 MPa for LC-40D X3 and 70 MPa for
LC-40D XR CL. When connecting the instrument with other equipment, set the
maximum pressure to the lowest allowable maximum pressure of the equipment.
n Setting example
The following is an example to set the minimum pressure [P.MIN] to 2.0 MPa.
18 LC-40D X3 CL LC-40D XR CL
3.2 Operation
If [P.MIN] is set to "0", solvent delivery will not stop and an alarm will not
sound even if the pressure has dropped. The minimum pressure can be set to
130 MPa max. for LC-40D X3 CL and 60 MPa max. for LC-40D XR CL.
3.2 Operation 3
4 Put the end of the drain tubing into the waste container.
LC-40D X3 CL LC-40D XR CL 19
3 Operation
IF the drain valve knob is turned more than 180°, any mobile phase that drains
out may contain air bubbles. This is normal.
There are two basic operating modes; one is in constant flow delivery mode and the
other is in constant pressure delivery mode.
1 Turn the drain valve knob clockwise as far as possible to close the drain valve.
2 Press [CE].
The initial screen is displayed.
5 Check that the pump outlet pressure is increasing on the pressure indication
display.
20 LC-40D X3 CL LC-40D XR CL
3.2 Operation
Close the drain valve when not draining through the drain path.
If the drain valve is left opened, a pressure difference may cause the mobile
phase solvent to flow out of the drain outlet.
1 Turn the drain valve knob clockwise as far as possible to close the drain valve.
LC-40D X3 CL LC-40D XR CL 21
3 Operation
7 Press [Enter].
[press] becomes ready to accept input.
10 Observe that the pump outlet pressure rises and pressure display stabilizes at
about 2.0 MPa.
While monitoring the pressure, increase the flow rate until the pressure reaches the set
value. If increasing the flow rate does not increase the pressure, stop increasing the flow
rate at 1 mL/min.
22 LC-40D X3 CL LC-40D XR CL
3.3 Types of Screens
Pressing [→] on the initial screen displays the screen to select the auxiliary functions
screen [FUNCTION] or VP functions screen [VP]. Move the cursor to the desired option
with [↓] or [↑] and press [→] to display the screen. Press [←] to return to the previous
screen. And Press [CE] to return to the initial screen.
LC-40D X3 CL LC-40D XR CL 23
3 Operation
DEGAS PRESSURE To show the vacuum pressure of the degassing unit. P.26
To show the status of the connection between the
PC/WS*1 P.27
instrument and the PC/WS.
n Displaying the pump ID and the port of the flow path switching valve connected to
the instrument during remote control《MONITOR ID/SV》
24 LC-40D X3 CL LC-40D XR CL
3.4 Monitoring Screens
On the right side of the screen, the open ports of the flow path switching valve
connected to the instrument are displayed. In the previous figure, the connection address
of the system controller is set to A. When the low-pressure gradient valve is used, the
set ports open only during delivery. While delivery is stopped, the screen displays
"OFF".
n Displays of Condition《CONDITION》
Auto-Diagnostics issues a
warning if a pulse exceeds a
specific threshold.
Note that very slight pressure
variation may not be detected.
The table shows a guideline of the conditions under which the Auto-Diagnostics function
can correctly detect a delivery failure.
Item Specification
Constant flow pumping
Pumping Methods "Switching the delivery mode《MODE CHANGE》(Only for
LC-40D XR CL)" P.38
0.1 to 2 mL/min
Flow Rate * Pumps should satisfy the above flow rate range in the high
-pressure gradient mode.
LC-40D X3 CL LC-40D XR CL 25
3 Operation
Item Specification
LC-40D X3 CL: 5 to 130 MPa
Delivery Pressure
LC-40D XR CL: 5 to 70 MPa
Isocratic, two-solvent high-pressure gradient, two-solvent
Pumping Modes
low-pressure gradient
[Press Increase] and [Press Decrease] are also displayed during normal operation,
such as when the pumps start and stop.
If [Press Dec Right] is displayed, it may indicate one of the following:
• Air remains inside the right pump head.
• There is a leak in the right pump head's seal.
• The right pump head's inlet check valve or the left pump head's outlet check
valve are not operating properly.
If [Press Dec Left] is displayed, it may indicate that conditions described above
have occurred for the left pump head.
26 LC-40D X3 CL LC-40D XR CL
3.4 Monitoring Screens
LC-40D X3 CL LC-40D XR CL 27
3 Operation
Setting Group
Description
Name
PARAMETER The group allows the user to set analysis parameters.
CONTROL The group enables specific operation.
The group allows the user to set parameters related to the system such
SYSTEM
as parameters of connection with external equipment.
The group allows the user to set parameters not related to the analysis
UTILITY
result such as display and buzzer parameters.
EDIT Time programs can be edited.
FILE Not used in CL model.
28 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
LC-40D X3 CL LC-40D XR CL 29
3 Operation
n [EDIT] group
Use this group to create a time program. For details, see "3.7 Creating Time Program"
P.69.
This function is not used because events are not used in the CL model.
30 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
About files:
The instrument can store up to 10 files of flow rate and other parameter values, and
time programs in memory. File No. 0 to 9 are assigned to each file. The auxiliary
functions except [INT.SV PORT] and [EXT.SV PORT] are common functions.
Use [FILE NUMBER] of the file operation group to switch the files.
Setting Range
Setting Step
LC-40D X3 CL LC-40D XR CL
1.0 to 130.0 MPa 1.0 to 70.0 MPa 0.1 MPa
10 to 1326 kgf/cm 10 to 714 kgf/cm 1 kgf/cm
10 to 1300 bar 10 to 700 bar 1 bar
142 to 19000 psi 142 to 10157 psi 1 psi
Setting Range
Setting Step
LC-40D X3 CL LC-40D XR CL
0 to 130.0 MPa 0 to 60.0 MPa 0.1 MPa
0 to 1326 kgf/cm2 0 to 612 kgf/cm2 1 kgf/cm2
0 to 1300 bar 0 to 600 bar 1 bar
0 to 19000 psi 0 to 8706 psi 1 psi
LC-40D X3 CL LC-40D XR CL 31
3 Operation
Single solvent
Set the compressibility according to the table.
1 Connect a resistance tube (or column) that can create pressure similar to the
desired delivery pressure at 1.0 mL/min or less and wait until delivery becomes
stable.
32 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
3
n Setting the port of the internal solenoid valve《INT.SV PORT》
The value cannot be input if the instrument has no internal solenoid valve.
LC-40D X3 CL LC-40D XR CL 33
3 Operation
n Setting the flow path of the mobile phase switching valve connected to the back of
the instrument《EXT.SV PORT CH1 / CH2 / CH3》
34 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
The value can be set in the range from 0.1 mL/min to 10.0 mL/min in 0.1 mL/min
increments.
The flow rate set here is approximate for replacing the mobile phase in the
pump.
LC-40D X3 CL LC-40D XR CL 35
3 Operation
• The approximate delivery amount necessary for replacing the rinse solution in the rinsing
flow path is approximately 10 mL.
• To replace the rinse solution in the rinsing flow path, select "4" and operate the pump
approximately 3 minutes.
36 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
n Setting the time to reach the set flow rate after the start of delivery《FLOW SLOPE》
Set the reference value for the zero point of the pressure sensor.
2 Press [Enter].
LC-40D X3 CL LC-40D XR CL 37
3 Operation
Stop the pump before switching the delivery mode. Switching the delivery mode
while the pump is operating is not possible.
This group is relevant to the system configuration such as connection with external
equipment.
Address Function
1 to 12 A channel number for connecting with the system controller
38 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
LC-40D X3 CL LC-40D XR CL 39
3 Operation
The system functions normally only with the above values. Values other than the
above cannot be set.
To clear a [P.MAX] error, press [CE]. The alarm stops and the flow rate returns to that
before the error.
40 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
3
Set Value Function
0 Do not use FCV.
1 Not used in CL model.
2 Not used in CL model.
n Inputting the serial number of the external flow path switching valve《EXT.SV SERIAL》
LC-40D X3 CL LC-40D XR CL 41
3 Operation
Set
Unit of Displayed Value
Value
0 MPa
1 kgf/cm2
2 bar
3 psi
42 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
Tube for buffer solution SUS Tubing (ID 0.3 x 600) 355 µL
(OUT tube, 400 mm)
3
SUS Tubing (ID 0.3 x 300) 335 µL
Use the Mixer installation kit
SUS Tubing (ID 0.3 x 185) 0 µL
(228-65020-41)
• The values above are intended when using the mixer recognition device. When the mixer
recognition device is not used, add the capacity depending on the mixer.
The column oven automatically recognise the capacity of the mixer when using the mixer
recognition device.
• The values above are intended for the recommended tubes. When using a tube not listed
above, increase or decrease the capacity depending on the tube.
Example
1) Use the standard tube + SUS Tubing (ID 0.3 x 600) + Mixer MR40 LPGE (without the
mixer recognition device)
→Set value: 240 (200 µL (from the table above) + 40 µL (mixer capacity))
2) Using the tube for buffer solution + SUS Tubing (ID 0.3 x 600) + Mixer MR40 LPGE
(with the mixer recognition device)
→Set value: 355(from the table above) (The capacity of the mixer will be
automatically recognised by the column oven)
3) Use the mixer instllation kit (without the mixer recognition device)
→Set value: 40 (0 µL (from the table above) + 40 µL (mixer capacity))
LC-40D X3 CL LC-40D XR CL 43
3 Operation
n Setting the time to turn off the operation panel automatically《DISP OFF TIME》
The setting range is from 0 to 10 (min), in which the step can be set by minute.
44 LC-40D X3 CL LC-40D XR CL
3.5 Auxiliary Functions Screen (FUNCTION)
Setting "0" minutes causes the display screen to be always on. If the display
screen is always on, some dots frequently lighting will deteriorate in brightness
in a shorter time.
When the optional low-pressure gradient kit or reservoir switching valve is installed inside,
pressing [PURGE] when starting purging does not display the confirmation screen. The
confirmation screen, however, is displayed when purging is stopped.
LC-40D X3 CL LC-40D XR CL 45
3 Operation
This function is not used because events are not used in the CL model.
Input the file number with the numeric keypad and press [Enter].
46 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
3
3.6.1 List of VP Functions
*1 Normally one solvent can be set for the mobile phase level monitoring function. Only in the
low-pressure gradient mode, four solvents (A to D) can be set.
LC-40D X3 CL LC-40D XR CL 47
3 Operation
* 1 In the calibration support group, an invalid password prevents the user from going further than
[OPERATION MODE] even if the user presses [Enter].
* 2 Do not change the factory default value.
48 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
n Displaying the delivery amount of the left plunger seal《L SEAL USED》
If the parts management tool (PARTS MGMT TOOL) is not used, press [del] after replacing
the seal. The total delivery amount is reset to "0" and the reset date is recorded in the
maintenance log.
• The plunger seal on the left when viewed from the front is shown as "L".
• When PARTS MGMT TOOL is active, the delivery amount cannot be reset on the operation
panel. Reset it from PARTS MGMT TOOL. For the operation procedure, refer to the PARTS
MGMT TOOL instruction manual.
n Displaying the delivery amount of the right plunger seal《R SEAL USED》
LC-40D X3 CL LC-40D XR CL 49
3 Operation
If the parts management tool (PARTS MGMT TOOL) is not used, press [del] after replacing
the seal. The total delivery amount is reset to "0" and the reset date is recorded in the
maintenance log.
• The plunger seal on the right when viewed from the front is shown as "R".
• When PARTS MGMT TOOL is active, the delivery amount cannot be reset on the operation
panel. Reset it from PARTS MGMT TOOL. For the operation procedure, refer to the PARTS
MGMT TOOL instruction manual.
50 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
LC-40D X3 CL LC-40D XR CL 51
3 Operation
The table shows the messages displayed in the second row of a maintenance log.
• To activate this function, supply power from the instrument to the degassing
unit and make sure that the pressure signal cable is connected correctly. If
more than one pump is used, be sure to connect the power cable and pressure
signal cable to the same pump. For how to connect, refer to the System Guide.
• The shutdown function of CBM-40 CL or LC-Workstation can stop the degassing
unit. For details, refer to the instruction manual of the system controller or
LC-Workstation.
• Approximate time to replace the vacuum pump of the degassing unit is 3 years
or 8000 hours of operation time. Contact your Shimadzu representative if you
need to replace the vacuum pump.
52 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
The group allows the user to check whether the equipment is operating correctly.
n Entering date《DATE》
n Entering time《TIME》
LC-40D X3 CL LC-40D XR CL 53
3 Operation
3 Press [CE].
The result display will be cleared.
If the result is [No Good], adjust the detection level with [LEAK THOLD] function in the
Calibration Support Group.
"Setting the operation level of leak sensor《LEAK THOLD》" P.66
After wetting and testing the leak sensor, wipe away the water on the tray
completely. When wiping away the water, do not apply pressure to the leak
sensor, the sensitivity decreases if the sensor unit touches the wall.
54 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
When checking operation, connect a resistance tube to the pump outlet to create 5 MPa
to 10 MPa pressure at 1 mL/min during delivery.
1 Press [Enter].
• When setting 0.20 MPa as the acceptance criterion, input a value as shown in the figure.
Standard of the acceptance criterion (MPa) : 0.20 MPa or less (When connecting the
resistance tube 4 m x 0.1 mm I.D and delivery of water at 1 mL/min.)
LC-40D X3 CL LC-40D XR CL 55
3 Operation
When checking operation, connect a resistance tube to the pump outlet to create 5 MPa
to 10 MPa pressure during delivery.
According to your measurement method, measure the flow rate by following the
instructions below.
2 After confirming the delivery stabilizes, measure the time required for the
solvent to accumulate to 5 mL in a measuring flask.
56 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
LC-40D X3 CL LC-40D XR CL 57
3 Operation
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
58 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
For example, if the set flow rate is 1 mL/min and the criterion is ±2 %, input 0.02 mL.
3
7 Input the flow rate measured with a
flow meter.
LC-40D X3 CL LC-40D XR CL 59
3 Operation
n Checking operation of the maximum and minimum pressure functions 《PRESS LIMIT
CHECK》
When checking operation, connect a resistance tube to the pump outlet to create 5 MPa
to 10 MPa pressure at 1 mL/min during delivery.
1 Press [Enter].
60 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
3
Be sure to input the five numbers. The default password is [00000].
When the password matches, the [OPERATION MODE] screen below appears on the screen.
If the password does not match, the user cannot go further.
LC-40D X3 CL LC-40D XR CL 61
3 Operation
62 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
When the setting is changed, the change is recorded in the operation log.
"Displaying the operation log《OPERATION LOG》" P.51
3
n Changing the password《PASSWORD SETTING》
1 Press [Enter].
The input screen is displayed.
LC-40D X3 CL LC-40D XR CL 63
3 Operation
If CBM-40 CL is not connected (it is not set the link address), pressing [→] does not bring
the user to the CBM parameter setting screen.
Display Description
SERIAL NUMBER Displays the serial No. of CBM.
S/W ID Displays the program version No. of CBM.
INTERFACE Sets the transmitting protocol to data processing unit.
ETHERNET SPEED Sets the transmitting speed of Ethernet.*1
USE GATEWAY Sets usage of default gateway or DHCP server.*1
IP ADDRESS Sets IP address of CBM.*1
SUBNET MASK Sets subnet mask.*1
DEFAULT GATEWAY Sets default gateway.*1, *2
64 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
LC-40D X3 CL LC-40D XR CL 65
3 Operation
Do not set values other than the set value above. Otherwise the instrument will
not work properly.
1 Dip the sensor unit at the bottom of the leak sensor into water.
Upon detection of water, the leak sensor value (Actual Level) in the screen increases.
2 Read the maximum value of the leak sensor value (Actual Level).
Example) The maximum value here is "160".
66 LC-40D X3 CL LC-40D XR CL
3.6 VP Functions Screen
When wiping away the water, do not apply pressure to the sensor unit of the
leak sensor. The sensitivity decreases if the sensor unit touches the instrument
wall surface.
4 Calculate the threshold of the leak sensor based on the value read in step 2.
Formula to calculate the leak sensor threshold:
Threshold = (Max. value-100)×0.7+100
Example) The value obtained here is "142".
3
5 Press [Enter], input the value obtained
in step 4, and press [Enter].
The new "threshold" is stored.
LC-40D X3 CL LC-40D XR CL 67
3 Operation
"Displaying the delivery amount of the left plunger seal《L SEAL USED》" P.49
"Displaying the delivery amount of the right plunger seal《R SEAL USED》" P.49
68 LC-40D X3 CL LC-40D XR CL
3.7 Creating Time Program
• The function of the time program is not available when controlled by the
system controller. Configure from the workstation. See " system guide 2.2
Editing Methods"
• Refer to the following procedure when using this function in local mode.
The instrument can run a time program with a flow rate and other parameters. A created 3
time program is saved in a file. Check for file numbers in use before creating a time
program.
See "Specifying the file number《FILE NUMBER》" P.46 of the auxiliary function to view
or specify file numbers.
Com-
Description Setting Range Remark Ref.
mand
Sets flow rate (Available
0 to Minimum unit:
FLOW only in constant flow P.20
10.0000 mL/min 0.0001 mL/min
solvent delivery mode)
Sets delivery pressure.
Minimum units:
(Available only in 10 to 60.0 MPa
0.1 MPa
constant pressure solvent 10 to 612 kgf/cm2
PRESS 1 kgf/cm2 P.21
delivery mode. This can 10 to 600 bar
1 bar
be set only for 142 to 8706 psi
1 psi
LC-40D XR CL)
Specifies concentration of
mobile phase B Minimum unit: *2
BCONC 0 to 100 %
Only active when [PUMP 0.1 %
CONFIG(SYS)] = 4*1
Specifies concentration of
mobile phase C Minimum unit:
CCONC 0 to100 %*2
Only active when [PUMP 0.1 %
CONFIG(SYS)] = 4*1
Specifies concentration of
mobile phase D Minimum unit:
DCONC 0 to100 %*2
Only active when [PUMP 0.1 %
CONFIG(SYS)] = 4*1
LC-40D X3 CL LC-40D XR CL 69
3 Operation
Com-
Description Setting Range Remark Ref.
mand
Opens or closes the
solenoid valve of the
INT.SV low-pressure gradient unit 1, 2, 3, 4 - P.32
or internal reservoir
switching valve (optional)
Opens or closes the
0, 1, 2, 3, 12, 13,
EXT.SV solenoid valve of Not used in CL model. -
23, 123
FCV-11A/FCV-11ALS
Sets ON/OFF of the event
output (relay contact) on
EVENT 0, 1, 2, 12 Not used in CL model. -
the back of the
instrument
0 to 255
LOOP Repeats program * 0 means 256 - P.75
times.
1 Press [CE].
The initial screen is displayed.
70 LC-40D X3 CL LC-40D XR CL
3.7 Creating Time Program
3 Press [↓] several times until [EDIT] is displayed in the screen, move the cursor
to [EDIT], and press [→].
The time program editing screen is
displayed. The figure shows an example of
a time program that has ten steps already
created and 310 steps left available.
Display Description
USED The number of created steps in the selected file
LEFT The number of steps left available
Displays Description
TIME Time (minute) that elapsed from the start of the step
FUNC Command name
VALUE Set value
LC-40D X3 CL LC-40D XR CL 71
3 Operation
2 Press [Enter].
The time (minute) setting screen is
displayed.
To set step 1
4 Press [Enter].
The [FLOW] command is selected, which
enables the user to input the flow rate.
To set step 2
7 Press [Enter].
The [FLOW] command is selected, which enables the user to input the flow rate.
To set step 3
72 LC-40D X3 CL LC-40D XR CL
3.7 Creating Time Program
11 Press [Enter].
When inputting multiple steps, the user does not need to input the data in order of time
because they are automatically sorted.
2 Press [Enter] and input the time with the numeric keypad.
LC-40D X3 CL LC-40D XR CL 73
3 Operation
Press [↓] several times in the same manner to display [CCONC] and [DCONC].
The [BCONC] command is selected, which enables the user to set the solvent B
concentration.
5 Input the concentration with the numeric keypad and press [Enter].
2 Press [del].
The first step of the program is deleted
and the second step is displayed.
74 LC-40D X3 CL LC-40D XR CL
3.7 Creating Time Program
This section explains about the commands that can be set for time programs only.
3
n Setting a loop count for a time program《LOOP》
Settings shown in the table repeat steps 1 and 2 three times at intervals of 30 minutes.
TIME VALUE
Step FUNC
(Minute) (mL/min)
1 15.00 FLOW 2.0000
2 20.00 FLOW 5.0000
3 30.00 LOOP 3
When the time program is executed continuously, do not set a [STOP] command
at the end of the program.
LC-40D X3 CL LC-40D XR CL 75
3 Operation
1 Press [CONC].
The concentration setting screen
is displayed.
76 LC-40D X3 CL LC-40D XR CL
3.8 Delivery in the Low-Pressure Gradient Mode
2 Press [→].
The compressibility setting screen
is displayed.
Press [↓] several times to display
the [COMP B], [COMP C], and
[COMP D] setting screens.
2 Press [PURGE].
The flow path selection screen is
displayed.
3 Select the flow path to purge with the numeric keypad and press [Enter].
[PUMP] on the display is highlighted and purging starts.
LC-40D X3 CL LC-40D XR CL 77
3 Operation
Setting
Set Value Remark
Parameter
"Selecting the local/remote mode《LOCAL MODE》"
LOCAL MODE 0: Remote
P.38
LINK ADDRESS Link address "Setting the link address《LINK ADDRESS》" P.38
(LC-40D X3 CL)
A max. of 4 pump units can be connected to CBM-40 CL. Workstation can set flow rate
(pressure), maximum pressure limit, minimum pressure limit, solenoid valves, pumping
ON/OFF, time program, etc., and can control binary and ternary high-pressure gradient
analysis depending on the system configuration. Also, by combining the SIL-40 CL series
autosampler, the auto purging function is available. For details, refer to the system guide.
In the CL model, do not connect another external device with the black-white
event cable.
Wiring is done by service personnel during installation.
78 LC-40D X3 CL LC-40D XR CL
4 Maintenance
4.1 Periodic Inspection and Maintenance
It is necessary to perform periodic inspections of this instrument to ensure its safe use.
It is possible to have these periodic inspections performed by Shimadzu service personnel
on a contractual basis.
For information regarding the maintenance inspection contract, contact your Shimadzu
representative.
Turn off the main power switch on the back of the instrument and unplug
the power cable prior to maintenance when instructed.
Fire, electric shock or malfunction may occur.
4
For parts replacement, use parts and tools listed in "1.2 Component Parts" P.2
and "5.2 Maintenance Parts" P.130.
If any other parts or tools are used, part damage, injury, and malfunction may
occur.
LC-40D X3 CL LC-40D XR CL 79
4 Maintenance
the tubes bound at the front right from the tube holder to prevent them from
interfering with the work.
"4.2 Removing the Right Panel Cover and the Filter Holder" P.84
Note that in addition to the following tools and parts are also required. The tools and
parts required for the maintenance of each part are listed at the beginning of each
section.
80 LC-40D X3 CL LC-40D XR CL
4.1 Periodic Inspection and Maintenance
LC-40D X3 CL LC-40D XR CL 81
4 Maintenance
The maintenance and replacement periods listed in this table are presented only
as guidelines. These are not guarantee periods.
These will vary depending on usage conditions.
Inspection/Maintenance Replacement
Remark Ref.
Item Periods
• Sealing efficiency decreases when seals
are worn.
• As a guideline, seals should be
replaced after delivery of the
respective volumes listed below. (VP
fuction, [L(R)SEAL USED], shows the
total delivered volume.)
(LC-40D X3 CL)
82 LC-40D X3 CL LC-40D XR CL
4.1 Periodic Inspection and Maintenance
Inspection/Maintenance Replacement
Remark Ref.
Item Periods
Replacement of Air Filter
1 year - P.120
(for Right Panel)
Replacement of Air Filter
1 year Contact your Shimadzu representative. -
(for FAN)
Pump ASSY lubrication 3 years Contact your Shimadzu representative. -
Replacement of Fuse 3 years Contact your Shimadzu representative. -
Cleaning of the
1 year - P.123
Automatic Rinsing Kit
Pumping mobile phase mixtures
containing acids such as trifluoroacetic
acid or formic acid can cause leakage
due to internal damage in the pump
Replacement of Pump head. In such cases, replace the pump
(2 years) -
Head head every two years, as a general
guideline.
(Pump heads do not need to be
replaced if only mobile phases without
acid are pumped.)
Inspection of Leak Sensor 1 year - P.54
4
n Optional Parts
Replacement
Inspection/Maintenance Item Remark Ref.
Periods
Replacing the Air Filter of Contact your Shimadzu
1 year -
Low Pressure Gradient Kit representative.
When the parts management tool (PARTS MGMT TOOL) is not used, input the
part number when replacing a part not registered in the PARTS MGMT TOOL.
LC-40D X3 CL LC-40D XR CL 83
4 Maintenance
4.2 Removing the Right Panel Cover and the Filter Holder
84 LC-40D X3 CL LC-40D XR CL
4.3 Removing/Installing the Pump Heads
Necessary tools
LC-40D X3 CL LC-40D XR CL 85
4 Maintenance
5 Press [Enter].
The cursor blinks at the input position.
7 Follow the instructions below to remove the tube (SUS pipe R) connecting the
check valve OUT and the pressure sensor.
(LC-40D X3 CL)
86 LC-40D X3 CL LC-40D XR CL
4.3 Removing/Installing the Pump Heads
(LC-40D XR CL)
LC-40D X3 CL LC-40D XR CL 87
4 Maintenance
11 Place the removed cap (B) and rinse solution bottle on the reservoir tray on the
top of the instrument.
88 LC-40D X3 CL LC-40D XR CL
4.3 Removing/Installing the Pump Heads
1 Set the plunger position by following steps 2 to 6 in "Removing the pump head"
P.85.
4
Tighten the left and right hexagonal socket screws 90 degrees at a time, and
then firmly tighten using the long side of the Allen wrench.
LC-40D X3 CL LC-40D XR CL 89
4 Maintenance
4 Follow the instructions below to attach the tube (SUS pipe R) connecting the
check valve OUT and the pressure sensor.
90 LC-40D X3 CL LC-40D XR CL
4.3 Removing/Installing the Pump Heads
LC-40D X3 CL LC-40D XR CL 91
4 Maintenance
Necessary tools
Necessary parts
1 Remove the pump head, SUS pipe, and inlet tube by following the instructions
in "Removing the pump head" P.85.
92 LC-40D X3 CL LC-40D XR CL
4.4 Replacing the Plunger Seal
2 Perform the following steps to replace the plunger seal and backup ring.
The plunger seal and the backup ring are installed inside the pump head. Use the
provided seal installer/remover to remove them.
LC-40D X3 CL LC-40D XR CL 93
4 Maintenance
94 LC-40D X3 CL LC-40D XR CL
4.4 Replacing the Plunger Seal
10 Place the pump head into a clean beaker containing 2-propanol, and clean it for
10 minutes using an ultrasonic bath.
LC-40D X3 CL LC-40D XR CL 95
4 Maintenance
17 Soak a new plunger seal and a new backup ring in 2-propanol in a clean beaker.
Be sure that the plunger seal is moistened with 2-propanol when putting it
on the seal installer/remover.
96 LC-40D X3 CL LC-40D XR CL
4.4 Replacing the Plunger Seal
Be sure that the backup ring is moistened with 2-propanol when putting it
on the seal installer/remover.
LC-40D X3 CL LC-40D XR CL 97
4 Maintenance
3 Attach the pump head, SUS pipe, and inlet tube by following the instructions in
"Installing the pump head" P.89.
98 LC-40D X3 CL LC-40D XR CL
4.4 Replacing the Plunger Seal
When PARTS MGMT TOOL is active, the delivery amount cannot be reset on the
operation panel. Reset it from PARTS MGMT TOOL.
3 Press [del].
The delivery amount of the plunger seal
before replacement is reset and the reset
date is recorded in the maintenance log.
"Displaying the delivery amount of the right plunger seal《R SEAL USED》" P.49 4
To replace the left plunger seal, display [L SEAL USED] and press [del].
If the above occurs even after replacing the plunger seal, it is possible that there
are scratches on the surface of the plunger. If there are scratches on the plunger,
replace it with a new plunger.
LC-40D X3 CL LC-40D XR CL 99
4 Maintenance
Necessary tools
Necessary parts
n Prior to the inspection and the maintenance Perform the preparatory work.
4
3 Gradually loosen the 2 hexagonal
socket screws (B) of the head holder
(A) using an Allen wrench (M4) by
alternating between the left and right
screw.
When replacing the plunger, be sure to replace the plunger seal and the
diaphragms.
1 Perform the following steps to clean or replace the plunger and diaphragm.
2 Check that there are no visible scratches on the plunger (A) or the diaphragm (B).
• If scratches are found through visual inspection, replace with a new plunger.
Proceed to the next procedure 3.
• If no scratches are found through visual inspection, proceed to step 2.
Confirm the
direction of the
diaphragm when
attaching.
2 In the [PLUNGER SET] screen of the control setting group, press [2] and [Enter].
(To replace the left plunger, press [1] and [Enter].)
[PUMP] on the display is highlighted and after a while stops being highlighted. The right
plunger is now at the backmost position.
"Setting the plunger stop position《PLUNGER SET》" P.37
4
3 Attach the head holder to the pump
body.
4 Attach the 2 hexagonal socket screws to the head holder and tighten them with
an Allen wrench.
Tighten the right and left hexagon socket screws alternately in 90° increments.
Finally hold the longer part of the hex wrench and tighten them securely.
5 Attach rinse tubes to the top and bottom of the head holder.
8 In the [PLUNGER SET] screen, press [2] and [Enter]. (To attach the left plunger
seal, backup ring and pump head, press [1] and [Enter].)
[PUMP] on the display is highlighted and after a while stops being highlighted. The right
plunger is now at the backmost position.
"Setting the plunger stop position《PLUNGER SET》" P.37
When replacing the plunger, be sure to replace the plunger seal and the
diaphragms."4.4 Replacing the Plunger Seal" P.92.
9 Attach the pump head, SUS pipe, and inlet tube and set the seal rinsing kit back
in place by following the instructions in "Removing the pump head" P.85.
Necessary tools
Necessary parts
1 After using buffer solution as the mobile phase, replace the solvent in the
reservoir bottle with water.
4
After using buffer solution as the mobile phase, do NOT replace with an
organic solvent such as 2-propanol without first replacing with water.
When using a buffer solution in the reservoir bottle, replace with distilled
water or purified water, and after pumping at least 20 mL into the instrument,
replace with new 2-propanol. If the buffer solution is directly replaced with
2-propanol, crystals will be precipitated, which can cause damage to the
plunger seal or cause problems with the check valve.
System Guide "Precautions When Replacing the Mobile Phase"
8 Press [PUMP].
[PUMP] on the display is highlighted and delivery of 2-propanol starts at 2 mL/min.
Continue delivery at least 1 hour.
12 Replace the solvent in reservoir bottle with the original mobile phase.
System Guide "Precautions When Replacing the Mobile Phase"
Necessary tools
1 Remove the SUS pipe and inlet tube by following the instructions in "Removing
the pump head" P.85.
3 Immerse the inlet check valve (A) and outlet check valve (B) into 2-propanol, and
clean for 5 minutes using an ultrasonic bath.
5 Attach the SUS pipe and inlet tube by following the instructions in "Installing
the pump head" P.89.
Necessary tools
Necessary parts
2 Remove the check valve and install a new one by following the instructions in
Step 2 of "4.7 Performing an Ultrasonic Cleaning of the Check Valve" P.110.
3 Attach the SUS pipe and inlet tube by following the instructions in "Installing
the pump head" P.89.
7 Press [PUMP] and confirm the solvent continues to flow out from the drain valve.
[PUMP] on the display is highlighted and the solvent flows out of the drain tube.
There is no problem with the valve if the solvent continues to flow out.
Necessary tools
2 Pour water into a reservoir bottle and set the flow rate to 1 mL/min.
When a mobile phase additive such as trifluoroacetic acid is used, a ghost peak
may be generated from the line filter. In this case, perform the following
procedure to rinse the line filter.
• Deliver a 17% phosphoric acid solution at a flow rate of 1 mL/min for about
30 minutes. (Pumping pressure: 1 MPa to 5 MPa)
For example, attach a stainless tube with ID0.1 x 2000mm (Part No.:
228-57908-41).
• Deliver purified water at a flow rate of 10 mL/min for about 30 minutes.
(Pumping pressure: 1 MPa to 5 MPa)
For example, attach a stainless tube with ID0.3 x 2000mm*1.
*1: Stainless tubing ID0.3 x 1000mm (Part No.: 228-53184-96) 2pc and Coupling
(Part No.: 228-16004-13) 1pc can be substituted.
Necessary tools
Necessary parts
Necessary Parts
Use the SUCTION FILTER with tube and fittings for pump inlet (Part No. 228-45708-91 ) or
SUCTION FILTER with tube and fittings for the low pressure gradient kit or the degassing
unit (Part No. 228-61905-49 ) as necessary.
After using the buffer solution as the mobile phase, do NOT replace with an
organic solvent such as 2-propanol without first replacing with water.
When using a buffer solution, replace it with distilled water or purified water,
and after pumping at least 20 mL into the instrument, replace it with new
2-propanol. If the buffer solution is directly replaced with 2-propanol, crystals
will be precipitated, which can cause damage to the plunger seal or cause
problems with the check valve.
System Guide "Precautions When Replacing the Mobile Phase"
4 Replace the solvent in reservoir bottle with water, and place the suction tubing
back into the bottle.
If Ghost peaking occurs, rinse the suction filter. Refer to System Guide "5.3.2
Symptoms Presumably Due to an Error, Based on a Chromatogram".
Necessary tools
Necessary parts
1 Rotate the drain valve ASSY counter-clockwise at least 3 complete rotations and
pull it off straight.
Necessary parts
If a leak occurred, wipe off the leak in the leak tray completely as below.
Leak of mobile phase solution used as buffer solution may dry and crystallize,
clogging the leak tray. If the leak tray is contaminated by such crystallized buffer
solution, wipe off the leak tray with paper wiper soaked in water in the same
manner as below.
1 Wipe off leak around the leak sensor or on the leak tray completely with wiper
paper.
If the exterior of the instrument is dirty, clean it with a dry soft cloth or tissue paper.
If it is very dirty, clean it as below.
1 Clean with a cloth soaked in diluted neutral detergent and wrung tightly.
2 Wipe off detergent with a cloth soaked in water and wrung tightly and then
wipe off water with a dry cloth.
Do NOT leave any part wet nor use thinner for cleaning.
Doing so may cause rust or discoloration.
The gas sensor may defect volatile gas generated from organic solvent.
2 Clean with a cloth soaked in Aqueous ethanol solution (80%) or aqueous sodium
hypochlorite solution (0.05%) and wrung tightly.
Necessary parts
If the 10% 2-propanol water in the rinse solution bottle is dirty, replace with
new solution. If getting significantly dirty, replace with new solution once a day.
Even if it does not appear dirty, replace with new solution once a week.
1 Remove the rinse cap of the automatic rinsing kit from the rinse solution bottle.
2 Pour distilled water into the rinse solution botte and clean the inside of the rinse
solution bottle using a brush, etc.
3 Pour distilled water into the rinse solution bottle and rinse the inside of the
rinse solution bottle for about 5 minutes using an ultrasonic bath.
4 Pour new 10% 2-propanol water into the rinse solution bottle, and return the
cap and rinse solution bottle back to their original positions.
Item Specification
Micro-volume double plunger pump
Pump Type
(approx. 10 µL/stroke)
Constant flow pumping and constant pressure
Pumping Methods
pumping
Allowable Maximum Pressure 130 MPa
Flow Rate Setting
0.0001 to 10.0000 mL/min
Range
130 MPa
Maximum Delivery (0.0001 to 3.0000 mL/min)
Pressure 80 MPa (5.0000 mL/min)
Constant Flow 22 MPa (10.0000 mL/min)
Pumping ±1 %
(1 mL/min when water is pumped at 80 MPa,
Flow Rate Accuracy
with water and room temperature constant
between 20 to 30 °C)
*1 When water is pumped at 10 to 40 MPa, with water and room temperature constant between 20
to 30 °C.
*2 When water is pumped at 40 to 60 MPa, with water and room temperature constant between 20
to 30 °C.
Item Specification
Micro-volume double plunger pump
Pump Type
(approx. 10 µL/stroke)
Constant flow pumping and constant pressure
Pumping Methods
pumping
Allowable Maximum Pressure 70 MPa
Flow Rate Setting
0.0001 to 10.0000 mL/min
Range
70 MPa
(0.0001 to 3.0000 mL/min)
Maximum Delivery 44 MPa
Pressure (3.0001 to 5.0000 mL/min)
Constant Flow 22MPa
Pumping (5.0001 to 10.0000 mL/min)
Larger value of either ±1 % or ±2 µL/min
(0.01 mL/min to 3 mL/min)
Flow Rate Accuracy
Larger value of either ±2 % or ±2 µL/min
(0.01 mL/min to 3 mL/min)
No more than the large value calculated with
Flow Rate Precision
either 0.06 % RSD or 0.02 min SD
Pressure Setting 1.0 to 60 MPa
Constant Pressure Range (in steps of 0.1 MPa)
Pumping
Pressure Accuracy Larger value of either ±10 % or 1.5 MPa
Pressure Display Accuracy Larger value of either ±2 % or ±1.0 MPa
Suction Filter 10 µm
Line Filter 5 µm
Commands for flow rate, pressure, [EVENT],
Time Program [LOOP] (for program repetition), 320 steps (total
of 10 program files)
Specification
Item
LC-40D X3 CL LC-40D XR CL
Number of Solvents Mixed 2 or 3
Step, linear and exponential functions (Only when
Gradient Profile
controlled by LC workstation) possible at multiple levels
Maximum Program Steps 400 steps (total of 20 program files)
Program Duration 0.01 to 9999.90 minutes (in steps of 0.01 minutes)
Range of Set Concentrations 0 % to 100 % (in steps of 0.1 %)
±0.5 %
±0.5 %
Concentration Accuracy*1 (0.05 to 3 mL/min,
(1 mL/min, 80 MPa)
1 to 40 MPa)
Flow Rate Possible 0.0001 mL/min to 10 mL/min
Item Specification
Number of Solvents Mixed Max. 4
Gradient Profile
Step, linear and exponential functions (Only when controlled by 5
LC workstation) possible at multiple levels
Maximum Program Steps 400 steps (total of 20 program files)
Program Duration 0.01 to 9999.90 minutes (in steps of 0.01 minutes)
Range of Set
0 % to 100 % (in steps of 0.1 %)
Concentrations
±0.5 %(1 mL/min, 10 MPa)
Concentration Accuracy
(for binary gradient with water/caffeine solution)
Flow Rate Possible 0.0001 mL/min to 10 mL/min
Item Specification
SUS316L, PEEK, ruby, sapphire, HastelloyC,
Liquid-Contacting Part Materials
polyethylene
Operating Temperature Range 4 °C to 35 °C
Humidity Range 20 % to 85 %
Available pH Range 1 to 14
W260 mm ´ H140 mm ´ D500 mm (Excluding
Dimensions
protruding parts)
Mass 10 kg
Power Supply Voltage 100 V AC to 240 V AC
(indicated on the * Mains supply voltage fluctuations are not to
instrument) exceed 10 % of the nominal supply voltage.
Power Consumption 150 VA
Item Specification
Flow Line Switching
Flow Line Switching with the five-way solenoid valve
Method
Number of Flow Lines 1 (with four built-in two way solenoid valve)
Solenoid Valve Operating
0 kPa to 50 kPa
Pressure
Materials for Parts
PTFE, PEEK, fluororubber
Exposed to Liquid
Operating Temperature 4 °C to 35 °C
C P
Calibration Support Information Group [PARAMETER] Setting Group 31
(CALIBRATION) 61 password 63
CBM parameter 64 Periodic Inspection 79
Check Valve 107, 110, 112 Power Button 11
Component Parts 2 Product Information Group (PRODUCT INFO) 49
Compressibility 16, 32 Purge 14
Constant Flow Delivery Mode 20
Constant Pressure Delivery Mode 21 R
Consumable Parts 131 Replacing the Plunger Seal 92
[CONTROL] Setting Group 37
S
D Shipping Screws 10
Display Area 13 Specifications 125
Drain Valve 119 Status LED 11
Suction Filter 117
F System Controller 78
[FILE] Operation Group 46 [SYSTEM] Setting Group 38
Filter Holder 84
T
G Time Program 69
Gradient Mixer for High-Pressure Gradient Analysis tube volume 43
5
U
H [UTILITY] Setting Group 44
High-Pressure Gradient System 127
V
I Validation Support Information Group
Initializing 62 (VALIDATION) 53
VP Functions Screen 47
L
LC-40D XR CL Specifications 126
Leak Sensor 8
Line Filter 114, 115
Liquid-Contacting Part Materials 128
Low-Pressure Gradient 76
Low-Pressure Gradient System 127
M
Main Power Switch 9
Maintenance Information Group (MAINTENANCE)
49
Maintenance Kit 133