I2985 1.0en GEN2i User Manual
I2985 1.0en GEN2i User Manual
I2985 1.0en GEN2i User Manual
English
References made to the Perception software are for version 6.10 or higher
HBM GmbH
Im Tiefen See 45
64293 Darmstadt
Germany
Tel: +49 6151 80 30
Fax: +49 6151 8039100
Email: [email protected]
www.hbm.com/highspeed
Copyright © 2010
All rights reserved. No part of the contents of this book may be reproduced or
transmitted in any form or by any means without the written permission of the publisher.
2 I2985-1.0 en
GEN2i
I2985-1.0 en 3
4 I2985-1.0 en
GEN2i
I2985-1.0 en 5
GEN2i
6 I2985-1.0 en
GEN2i
I2985-1.0 en 7
GEN2i
8 I2985-1.0 en
GEN2i
1 Safety Messages
1.1 FCC and general
The first WARNING note below is required by the FCC and relates only to the
interference potential of this equipment. This message is a direct quotation.
WARNING
The equipment generates, uses, and can radiate radio frequency energy
and if not installed and used in accordance with the instructions manual,
may cause interference to radio communications. As temporarily
permitted by regulation, it has not been tested for compliance with the
limits for Class A computing devices pursuant to Subpart B or Part 15 of
FCC Rules, which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential area is
likely to cause interference, in which case the user at his own expense
will be required to take whatever measures may be required to correct the
interference.
The design of this instrument has been verified to EN 61010-1 for Class 1
(grounded use).
This manual contains information and warnings that must be observed to keep
the instrument in a safe condition. The instrument should not be switched on if
it is damaged and it should not be used under wet conditions.
For the correct and safe use of this instrument it is essential that both operating
and service personnel follow generally accepted safety procedures in addition
to the safety precautions specified in this manual.
Whenever it is likely that safety protection has been impaired, the instrument
must be made inoperative and secured against any unintended operation.
Qualified maintenance or repair personnel should be informed. Safety
protection is likely to be impaired if, for example, the instrument shows visible
damage or fails to operate normally.
I2985-1.0 en 9
GEN2i
1.2 Grounding
The instrument must be used with a protective ground connected via the
conductor of the supply cable. This is connected to the instrument before the
line and neutral connections when the supply connection is made. If the final
connection to the supply is made elsewhere, ensure that the ground connection
is made before line and neutral.
WARNING
For protection against electric shock, all external circuits or equipment shall
have a safe insulation. Therefore it is not permitted to connect peripheral
equipment to the system with a power supply without SELV (Separated Extra
Low Voltage) or Class II qualification.
WARNING
It is the responsibility of the user to ensure the safety of any accessories, such
as probes, used with the instrument.
CAUTION
Even low voltage inputs may contain high voltage fast transients (spikes),
which could damage the input. For this reason it is not safe, for instance,
to make direct connections to an AC line supply.
10 I2985-1.0 en
GEN2i
CAUTION
For input BNCs that are galvanically isolated from the chassis. Input
conductors including the BNC shell may carry hazardous voltages. Only
appropriate insulated BNC connectors should be used.
WARNING
All inputs are rated for IEC 61010 CAT I (Category 1) signals only. This
instrument should not be used to measure high-energy signals of
Categories II, III, and IV.
The covers protect the user from live parts and should only be removed by
suitably qualified personnel for maintenance and repair purposes.
I2985-1.0 en 11
GEN2i
CAUTION
Description of ESD
Static electricity is an electrical charge caused by the buildup of excess
electrons on the surface of a material. To most people, static electricity and
ESD are nothing more than annoyances. For example, after walking over a
carpet while scuffing your feet, building up electrons on your body, you may get
a shock - the discharge event - when you touch a metal doorknob. This little
shock discharges the built-up static electricity.
ESD-susceptible equipment
Even a small amount of ESD can harm circuitry, so when working with electronic
devices, take measures to help protect your electronic devices, including your
GEN2i Portable data recorder, from ESD harm. Although HBM has built
protections against ESD into its products, ESD unfortunately exists and, unless
neutralized, could build up to levels that could harm your equipment. Any
electronic device that contains an external entry point for plugging in anything
from cables to acquisition cards is susceptible to entry of ESD.
12 I2985-1.0 en
GEN2i
WARNING
The wrist strap is intended for static control only. It will not reduce or
increase your risk of receiving an electric shock from electrical
equipment. Follow the same precautions you would use without a wrist
strap.
I2985-1.0 en 13
GEN2i
1.4 Environment
The instrument should be operated in a clean, dry environment with an ambient
temperature of between 0 °C and +40 °C.
The instrument will operate with full, specified accuracy between +15 °C and
+35 °C.
Note Direct sunlight, radiators and other heat sources should be taken into account
when assessing the ambient temperature.
The instrument relies on forced air-cooling with fan and ventilation apertures.
Adequate ventilation can usually be achieved by leaving a 75 mm (3" gap)
around the instrument. Care should be taken to avoid restricting the airflow
around the fan holes at the side of the instrument.
To clean the instrument, disconnect all power sources and wipe the surfaces
lightly with a clean, soft cloth dampened with water.
14 I2985-1.0 en
GEN2i
The power connection of the GEN2i is via a standard IEC 320, EN 60320, C14
(male) appliance inlet.
To disconnect the instrument from the AC supply, unplug the IEC connector on
the top of the instrument. The instrument should be positioned to allow access
to the AC connector. The front power switch on the instrument is not a
disconnecting device. When the instrument is connected some power will be
consumed.
CAUTION
I2985-1.0 en 15
GEN2i
This instrument generates, accepts and can radiate radio frequency energy
and, if not installed and used in accordance with the operator manual, may
cause harmful interference to other equipment. However, there is no guarantee
that interference will not occur in a particular installation.
Immunity test: All immunity tests are done with the failure criterion being a
change of the instrument's control settings. Any of these tests may produce a
spurious trigger. Measurements are not valid during and immediately after the
immunity tests.
Supply cables, interface cables and probes should be kept as short as practical,
preferably a maximum of 1 m. Interface cables should be screened and
interface cables longer than 3 m are not acceptable in terms of interference port
immunity.
CAUTION
Electrical frequencies and equipment working in the range of 270 Khz can
interrupt the stability of the GEN2i Touch screen. Under these conditions,
it is possible that the touch sensitivity may become erratic or unusable.
If such interference occurs, please contact your local supplier for more
details.
16 I2985-1.0 en
GEN2i
For network interfacing the GEN2i supports Cat 5e cable. This cable is an
enhanced version of Cat 5 that adds specifications for far end crosstalk. The
tighter specifications associated with Cat 5e cable and connectors make it an
excellent choice for use with 1000BASE-T (1 Gigabit). Cat 5e cables are limited
to a maximum of 100 m (328 ft) in length (normal practice is to limit fixed
("horizontal") cables to 90 m to allow for up to 5 m of patch cable at each end).
I2985-1.0 en 17
GEN2i
WARNING
When the apparatus is connected to its supply, terminals may be live, and
the opening of covers for removal of parts is likely to expose live parts.
Whenever it is likely that the protection has been impaired, the apparatus shall
be made inoperative and be secured against any unintended operation.
The protection is likely to be impaired if, for example, the apparatus shows
visible damage or has been subjected to severe transport stresses.
WARNING
Proper use of this device depends on careful reading of all instructions and
labels.
WARNING
WARNING
This instrument and related accessories are not designed for biomedical
experimentation on humans and should not be directly connected to
human subjects or used for patient monitoring.
18 I2985-1.0 en
GEN2i
I2985-1.0 en 19
GEN2i
Where caution is required, this symbol refers to the User's Guide for
further information.
This symbol warns that high voltages are present close to this symbol.
20 I2985-1.0 en
GEN2i
CAUTION
The weight of the instrument may exceed 15 kg when fully loaded. Please
take appropriate actions before lifting the instrument.
Before lifting or carrying a heavy object, ask yourself the following questions:
I2985-1.0 en 21
GEN2i
1.11 Batteries
The GEN2i PC motherboard includes a removable, Lithium battery. Replace
only with same or equivalent type (CR2032).
WARNING
22 I2985-1.0 en
GEN2i
SIKKERHEDSADVARSEL
Dette instrument skal anvendes med en beskyttelsesjordforbindelse via
netkablets jordledning til jordforbindelsen i instrumentets apparatkontakt eller -
hvis instrumentet er forsynet dermed - via sikkerhedsjordklemmen. Enhver
afbrydelse af sikkerhedsjordforbindelsen vil formentlig gøre instrumentet
berøringsfarligt. Bevidst afbrydelse er forbudt. Hvis et indgangssignal
overstiger 40 V spidsværdi, skal en ekstra signal jord forbindes.
Afbryd instrumentet fra lysnettet ved at fjerne IECstikket fra bagpanelet. Dette
instruments AC netafbryder er kun beregnet til funktionelle formål. Den er
hverken beregnet til eller egnet til afbrydelse af lysnettet.
Nederlands
VEILIGHEIDSWAARSCHUWING
Dit instrument mag uitsluitend worden gebruikt als een beschermende massa
(aarde) is aangesloten via de beschermende massageleider van de
voedingskabel, of - indien het instrument daarvan is voorzien - via de
veiligheids-massa-aansluiting. Als de beschermende massa, binnen of buiten
het instrument, wordt onderbroken, dan kan dat hierdoor uitermate gevaarlijk
worden. Het opzettelijk onderbreken van de massa, is verboden. Indien er een
signaal wordt aangeboden van meer dan 40 V (top-top) dan dient eveneens de
signaal aarde aangesloten te zijn.
Als de zekering doorbrandt als gevolg van een storing of een defect, dan is het
mogelijk dat de wisselstroom-voedingsschakelaar van het instrument
beschadigd is. Die schakelaar moet worden gecontroleerd door een deskundig
technicus.
I2985-1.0 en 23
GEN2i
Suomi
TURVAOHJEITA
Tätä laitetta käytettäessä sen tulee olla suojamaadoitettu joko verkkojohdon
suojajohtimen tai erillisen suojamaadoitusliitännän kautta, mikäli laitteeseen on
sellainen asennettu. Suojamaadoituksen katkaiseminen laitteen sisä-tai
ulkopuolelta tekevät siitä vaarallisen. Tahallinen katkaisu on kiellettyä. Lisäksi,
jos jokin tulosignaaleista ylittää 40 V peak, on signaalimaa kytkettävä.
Français
ATTENTION - DANGER!
Cet appareil doit impérativement être mis à la masse par le conducteur de terre
du câble d'alimentation ou, si l'instrument en comporte une, par la borne de
terre. Il peut être dangereux en cas de coupure du circuit de terre, que ce soit
à l'intérieur ou à l'extérieur de l'instrument. Il est formellement interdit de couper
intentionnellement le circuit de terre. De plus, une masse signal doit être
connectée si l'un quelconque des signaux d'entrée dépasse 40 V crête.
Le fait que le fusible d'alimentation saute par suite d'une anomalie risque de
détériorer l'alimentation secteur de l'instrument; dans ce cas, le faire contrôler
par un technicien qualifié.
24 I2985-1.0 en
GEN2i
Deutsch
WARNHINWEIS!
Dieses Gerät muß mit einer Schutzerde betrieben werden, die über den
Schutzleiter des Speisekabels oder über die Erdungsklemme des Gerätes (falls
vorhanden) anzuschließen ist. Bei einer Unterbrechung der Schutzerde
außerhalb oder innerhalb des Gerätes kann eine Gefahr am Gerät entstehen!
Eine beabsichtigte Unterbrechung ist nicht zulässig. Achtung! Bei
Signalspannungen über 40 V muß die Signalmasse angeschlossen sein.
Italiano
AVVISO DI SICUREZZA
Questo strumento deve esser utilizzato con un collegamento protettivo di
messa a terra tramite il filo di messa a terra del cavo di alimentazione o tramite
il terminale di messa a terra in sicurezza, nel caso in cui lo strumento ne sia
dotato. Qualsiasi interruzione della massa a terra protettiva, sia all'interno che
all'esterno dello strumento, lo renderà pericoloso. E'vietata qualsiasi
interruzione causata intenzionalmente. Inoltre, la connessione di terra deve
essere collegata se ad uno qualsiasi degli ingressi viene applicato un segnale
superiore a 40 V di picco.
I2985-1.0 en 25
GEN2i
Norsk
ADVARSEL!
Dette instrumentet må bare anvendes så lenge det er jordet via den
beskyttende jordlederen i strømkabelen, eller via jordingsklemmen, hvis
instrumentet har en. Eventuelle forstyrrelser i den beskyttende jordingen, inne
i eller utenfor instrumentet, vil sannsynligvis gjøre instrumentet farlig. Forsettlig
forstyrrelse er forbudt. I tillegg, signal jord må tilkobles dersom inngangs
signalet overstiger 40 V spissverdi.
Português
AVISO DE SEGURANÇA
Este aparelho deve ser operado com uma ligação terra ligado por um conductor
trifásico do cabo principal ou, se o instrumento já tiver um, via um terminal de
segurança. Qualquer interrupção do trifásico, dentro ou fora do aparelho, pode
tornar o aparelho perigoso. É proibida a interrupção intencional. Nota: O
terminal de terra deve ser ligado se o sinal de entrada a medir for superior a
40 V de pico.
26 I2985-1.0 en
GEN2i
Para desconectar este aparelho da fonte AC, retire o conector IEC do painel
trazeiro. Neste aparelho, o interruptor da fonte AC existe sómente por razões
funcionais. Não deve ser usado e nem é apropriado como dispositivo de
desconexão.
Español
I2985-1.0 en 27
GEN2i
Svenska
SÄKERHETSVARNING
Detta instrument måste drivas med en skyddande jordledning ansluten via den
skyddande jordledaren på matarkabeln eller, om instrumentet har sådan
monterad, via det jordade uttaget. Om jordanslutningen störs, inuti eller utanför
instrumentet, är det troligt att instrumentet kommer att utgöra en fara. Avsiktlig
störning är förbjuden. Dessutom måste en signaljord anslutas om någon av
ingångssignalerna överstiger 40 V topp.
English
SAFETY WARNING
This instrument must be operated with a protective ground (earth) connected
via the protective ground conductor of the supply cable or, if the instrument is
fitted with one, via the safety ground terminal. Any interruption of the protective
ground, inside or outside the instrument, is likely to make the instrument
dangerous. Intentional interruption is prohibited. In addition, a signal ground
must be connected if any input signal exceeds 40 V peak.
If the supply fuse blows as the result of a fault, it is possible that the instrument's
AC supply switch will be damaged and should be checked by a suitably qualified
engineer.
To disconnect this instrument from the AC supply, unplug the IEC connector
on the rear panel. The AC supply switch on this instrument is provided for
functional purposes only. It is not intended, or suitable, as a disconnecting
device.
28 I2985-1.0 en
GEN2i
日本語
安全上の警告
本機器の操作は、電源ケーブルの保護接地線で接地(アース)を施した上で
行ってください。また、安全接地用端子が存在する場合は、これを経由して
本機器を接地してください。機器の内部または外部にある保護接地線が遮断
されると、機器が危険な状態に陥る可能性があります。故意に保護接地線を
遮断することを禁止します。また、入力信号がピーク時に 40V を超える場合
は、信号接地線を接続してください。
カバーは取り外さないでください。
電源ヒューズが故障により飛んだ場合、機器の AC 電 源スイッチが損傷する
おそれがあるため、然るべき認 定を受けた適任者による点検を受けてくださ
い。
中文
安全警告
该仪器必须通过电源电缆的导线连接到保护接地(接地) ,如果该仪器已配备了
安全接地端子,则通过该端子接地。 断开仪器内外的任何保护接地可使仪器成
为危险设备。 严禁有意断开。 另外,如有任何输入信号超过 40 V 的峰值,还
必须连接信号接地。
不要取下保护盖。
如果电源保险丝因故障而熔断,则有可能损坏仪器的 交流电源开关并应由具备
资格的工程师检查。
I2985-1.0 en 29
GEN2i
30 I2985-1.0 en
GEN2i
WARNING
WARNING
CAUTION
CAUTION
HINT/TIP
The info icon indicates sections which give additional information about the
product. This information is not essential for correct operation of the instrument,
but provides knowledge to make better use of the instrument.
I2985-1.0 en 31
GEN2i
This manual only covers the basic Perception software and not the dedicated
Touch GUI interface referred to as the "Instrument Panel." This is described in
a separate manual: "The GEN2i Instrument Panel User Manual."
When the wording “Click Start ...” is used, this refers to the Windows Start
button. In Windows Vista or Windows 7, the Start Menu has undergone some
significant changes. The taskbar icon is no longer labeled "Start" and is now
simply the pearl icon (of the window-frame in an orb).
Note If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
32 I2985-1.0 en
GEN2i
3 Introduction
3.1 Introducing the GEN2i
Welcome. You have made the right choice: your GEN2i Portable Data Recorder
is one of the most sophisticated and powerful systems in the marketplace and
demonstrates the quality HBM has to offer. The GEN2i is an all-in-one, field-
ready, feature-packed unit.
I2985-1.0 en 33
GEN2i
This instrument panel provides one-touch access to the most commonly used
features for everyday work.
Note Not all hardware features described in this manual are available through the
Instrument Panel.
For the sophisticated or more exotic features you can use the Perception
software that is accessible from the Instrument Panel via a single touch.
34 I2985-1.0 en
GEN2i
3.2 Hardware
The acquisition section of the GEN2i is based on the successful and proven
stand-alone GEN series Data Acquisition System.
I2985-1.0 en 35
GEN2i
3.3 Acquisition
The GEN2i is a multi-channel Portable data recorder. It provides real-time data
for waveform and meter displays. At a high streaming rate it allows unlimited
recording duration and file size. Statistics are performed in real-time. Its
extreme-performance signal conditioning includes both Bessel and Butterworth
anti-alias filters to provide excellent response.
3.3.1 StatStream®
Most PC-based DAQ systems can easily acquire megabytes of data. But even
the most powerful PC is poorly equipped to display and process files of
megabytes or gigabytes. In fact, most DAQ systems fail to display over 99% of
your live data! The exclusive StatStream® technology accelerates all aspects
of your measurement task with dedicated hardware and firmware.
36 I2985-1.0 en
GEN2i
When reviewing your stored files, the embedded StatStream® data enables an
accurate, detailed overview of any size file in seconds. Unlike competitive
systems, the GEN2i has no need to inspect gigabytes of information just to
display the last kilobyte. As you zoom in, more detail is displayed while always
maintaining the highest visible resolution.
I2985-1.0 en 37
GEN2i
38 I2985-1.0 en
GEN2i
Recorded files are standard Windows files with extension PNRF (Perception
Native Recording File).
I2985-1.0 en 39
GEN2i
3.6 PC Section
The GEN2i has a built-in industrial PC. This PC provides all the standard
features that you may expect at an industrial grade quality.
Features include:
l Low-Power AMD CPU
l Microsoft® Windows® 7 Operating System
l 4 GigaByte of RAM
l 256 GB Solid state drive
l On-board graphics engine with video out
l Full audio and internal speaker
l 8 USB 2.0 ports
l 1 Gigabit Ethernet
l Wireless LAN support at 54 Mbps
Refer to the specification section "General" on page 124 of this manual for full
details.
40 I2985-1.0 en
GEN2i
The GEN2i has 2 modes of operation. You can choose between the two modes
during operation.
Primary mode:
Secondary mode:
The Instrument Panel is the dedicated touch screen software interface which
has been specifically designed for the GEN2i as the normal mode of operation,
however you can switch to Standard Perception, for on-board post-processing
at any time.
I2985-1.0 en 41
GEN2i
l When in the Instrument panel, tap the Perception icon on the toolbar in the
lower left corner, to switch to Perception.
l When in Perception, Navigate to the menu item File and select Switch to
Instrument Panel and confirm your action to switch to Instrument Panel.
42 I2985-1.0 en
GEN2i
Note For information on the general usage of Perception you must use the separately
supplied Perception User Manual.
Note For information on the general usage of the Instrument Panel you must use the
separately supplied Instrument Panel Software For The GEN2i User Manual.
I2985-1.0 en 43
GEN2i
WARNING
The power inlet connects/disconnects the main power from the GEN2i system.
To disconnect the GEN2i completely from the AC supply, unplug the IEC
connector on the rear of the instrument. Plugging in the Unit will not switch on
the GEN2i instrument. Use the standby button on the front panel for this
purpose; see "Front panel control" on page 54.
44 I2985-1.0 en
GEN2i
4.2 PC connections
GEN2i has a PC motherboard inside. This PC motherboard has a lot of
connections that can be used to connect other devices to the GEN2i system.
I2985-1.0 en 45
GEN2i
46 I2985-1.0 en
GEN2i
CAUTION
CAUTION
1 Shut down the GEN2i system and remove the power input cable.
2 Loosen the small set screw on both ejectors on the module:
I2985-1.0 en 47
GEN2i
3 Press the inner grey button on each ejector to release the catch.
4 Press both ejectors outward to release the module. They act as levers to
gently pull the module from its backplane sockets
5 Slide the module out of the GEN2i unit.
For correct operation of the instrument, it is imperative that the board with
the lowest maximum sample rate is placed in the slot marked "A".
1 Shut down the GEN2i system and remove the power input cable.
2 Ensure the ejector levers are in the farthest outermost position, tilting away
from the module
3 Slide the module into its guide rails until the ejectors contact the perforated
metal strips at left and right.
4 Press both ejectors inward to seat the module. They act as levers to gently
pull the module into its backplane sockets. The gray button should snap to
its default position and lock the ejectors
48 I2985-1.0 en
GEN2i
4.4 Handle
The handle is used to carry the GEN2i system. Only carry the instrument when
the handle is in the upright position.
The handle can be turned to the top or rear side. When the handle is turned to
the rear side it can be used to lift the instrument front so the display angle is
towards the user. The angle between display and horizontal is about +40 °. For
other tilting methods see section "Feet" on page 53.
I2985-1.0 en 49
GEN2i
50 I2985-1.0 en
GEN2i
2a Rotate handle to the rear and release the handle buttons. The handle will
snap into one of its fixed positions (30° indexing).
I2985-1.0 en 51
GEN2i
52 I2985-1.0 en
GEN2i
4.5 Feet
GEN2i is standing in normal operation position on 4 feet. Two positioned at the
rear and two at the front of the instrument. The front feet can be used to lift the
instrument front so the display angle is towards the user. The angle between
display and vertical is about +8 ° in this position.
I2985-1.0 en 53
GEN2i
5 Getting Started
5.1 Front panel control
Standby
On the GEN2i the standby button is located on the front panel. When this button
is pressed, the instrument state toggles between operating and standby mode.
In standby mode some power will be consumed and the instrument is NOT
disconnected from the AC supply. To switch off the instrument completely from
the mains power disconnect the mains power cable from the instrument.
When you switch off the GEN2i using this standby button, the software will
display a confirmation dialog. Follow the on-screen instructions for a correct
shutdown of the system. When the system is not responsive you should press
the standby button for at least four seconds until the unit is forced to shut down.
USB Ports
Located on the bottom left are the two front mounted USB ports for easy access,
the remaining six are located on the back.
54 I2985-1.0 en
GEN2i
I2985-1.0 en 55
GEN2i
When you change nothing, this will be the default behavior each time you start
the system.
Each time you shut down, changes made to the preferences and setting will be
saved automatically. These setting are then loaded the next time you turn on
the GEN2i.
56 I2985-1.0 en
GEN2i
One of the wired network interfaces is used internally for the communication
between the PC and the acquisition system. The other network interface can
be used to connect the GEN2i integrated PC to a local network or the Internet.
1 Click Start
2 Select Control Panel
3 Open Network Connections
The local area connection unidentified network is the one used for internal
communication between the PC section and the GEN2i DAQ unit.
This network interface and the GEN2i acquisition system are organized to use
a fixed configuration. The IP configuration for the internal network interface is:
l IP Address 172.16.10.1
l Subnet mask: 255.255.0.0
I2985-1.0 en 57
GEN2i
WARNING
58 I2985-1.0 en
GEN2i
The adapter features full compliance with the IEEE 802.11 b/g wireless
standard offering the best compatibility and future-proof reliability. With data
rates up to 54 Mbps the adapter is also up to five times faster than previous
generation wireless connection standards.
Below is a quick start procedure to connect to your local network and a quick
start procedure to disconnect from a network.
For wireless network operation details refer to the Windows Help system. If you
are uncomfortable, or not yet familiar with these procedures, please contact
your IT department or a network knowledgeable person to assist you.
I2985-1.0 en 59
GEN2i
1 Click Start, click Control Panel, click Network and Internet, and click
Network and Sharing Center and select the task Manage Wireless
Networks.
2 To change a setting right-click on the network's icon. On the context menu
that comes up:
l Select Remove Network to remove the network connection, or
l Select Properties. In the properties dialog that comes up make sure
that Connect automatically when this network is in range is
cleared.
When you have selected manual connection, you can still connect to the
network if you want to, but you'll need to do so manually. To do so right-click
the Windows Vista's networking icon in the task bar and select Connect to a
Network.
60 I2985-1.0 en
GEN2i
6 Input Modules
6.1 Available input modules
There are two acquisition slots available. This section provides more
information on the acquisition boards compatible with the GNE2i.
At printing time of this document the following input modules were available:
l The Basic Amplifier input module gives you 8 channels with single-ended
or isolated input on one board. With 200 kS/s or 1 MS/s digitizing rate at
16-bit resolution they are ideally suited for high definition transient
recording.
l The 200 kS/s and 1 MS/s Bridge Amplifier versions comprise everything
that you would expect from HBM. They offer isolated and differential inputs,
high gain, and extremely good SNR even at the maximum 400 kHz
bandwidth. Both versions come with on-board, software selectable shunt
calibration and bridge completion resistors, a bridge wizard for error-free
setup and fast auto balance.
l The unique high-end Universal Amplifier serves a variety of needs,
ranging from differential and/or isolated measurements to ICP-based
vibration or shunt-based current measurements. Apart from the differential
mode, the amplifier supports any type of 'constant current' vibration and
acceleration sensors. In “current mode” the built-in shunt can be used to
measure up to 1 ampère in a safe, isolated and fused manner, without the
need of external shunt resistors.
l As a special input board, the Binary Marker input enables up to 64 "digital"
channels to be recorded as well. This board can be used to record status
signals from the process or test like high/low, open/closed or left/right. In
addition 3 counter/timer channels are available.
l The Binary Marker HV board allows you to acquire 32 digital event signals
(markers) as well as 8 digital event signals that are optically isolated.
Although general purpose, this board is specifically suited for the medium/
high voltage market.
l For ultra fast signals, the HiSpeed 25 MS/s and 100 MS/s digitizers are
equipped with four channels sampling at incredible high speed. With
selectable anti-aliasing filtering and 14-bit (100 MS/s) or 15-bit resolution
(25 MS/s), these inputs turn the GEN2i into an extremely fast transient
recorder. Enhanced resolution mode increases input resolution for both
models to 16-bit. Inputs are single ended or differential. The 25 MS/s
digitizer replaces the 20 MS/s high speed digitizer that was produced before
2007.
I2985-1.0 en 61
GEN2i
l The Isolated Digitizer 6600 is used in the most demanding test lab
applications requiring high voltage isolation and safe and accurate
measurements in harsh electrical environments. It is available in four
different configurations. There are two mechanical variants, one for use in
medium and one for use in high voltage environments. Each model is
available with 25 MS/s or 100 MS/s digitizing speed per channel.
Please refer to the “Specifications” section on page 124 for more detailed
information on acquisition board.
62 I2985-1.0 en
GEN2i
The Isolated Digitizer 6600 comes in two models to best fit your requirements.
A high voltage / high power model that you can use in even the most demanding
applications and harsh environments, and a medium voltage / medium power
model. This last front-end can be connected directly to the mains. The built-in
power supply provides isolation up to 10 kV.
l Communication and data transfer between the system and the controlling
PC using the Ethernet interface;
l Provides various I/O connectors for acquisition support;
l Has space for options.
I2985-1.0 en 63
GEN2i
The 1 MS/s basic amplifier input module is also available in an isolated version
with plastic BNC connectors.
64 I2985-1.0 en
GEN2i
6.3.1 Specifications
Analog input sections
Acquisition
I2985-1.0 en 65
GEN2i
Acquisition
Anti-Alias filter 200 kS/s model: 7-pole Bessel 20 kHz, no wideband set-
ting
1 MS/s model: 7-pole Bessel 250 kHz
1 MS/s isolated model: 7-pole Bessel 220 kHz
Digital filter frequency-domain optimized, 12-pole Finite Impulse Re-
sponse (FIR) at sample rate divided by 4, 10, 20, 40
Triggering
Acquisition modes
66 I2985-1.0 en
GEN2i
Acquisition modes
Dual Combination of sweeps and continuous mode: recorder
type streaming to disk with simultaneously triggered
sweeps in RAM.
Note When mixing 200 kS/s and 1 MS/s input modules within a single acquisition
group in the Perception software, the maximum acquisition speed is limited to
200 kS/s.
I2985-1.0 en 67
GEN2i
68 I2985-1.0 en
GEN2i
IMPORTANT
Note When mixing 200 kS/s and 1 MS/s input modules within a single acquisition
group in the Perception software, the maximum acquisition speed is limited to
200 kS/s.
Input diagrams and typical connection diagrams for the bridge amplifiers are
shown on this and the following pages. For the maximum versatility, the
amplifiers allow a wide range of configurations. At minimum three wires are
necessary for a quarter- or half-bridge sensor and four wires for a full bridge.
Optional remote sensing of excitation voltage is supported for precision
transducer applications, which adds two wires. Remote shunt calibration is
possible with the addition of two or three more wires. Finally, both an isolated
common and a driven guard are provided for optional shielding.
I2985-1.0 en 69
GEN2i
70 I2985-1.0 en
GEN2i
Additional pins on the LEMO connector provide a precision 350 Ω resistor plus
an additional user-provided value for quarter-bridge completion. The user-
provided value is located on removable bridge completion card. The completion
sockets are designed for Vishay MicroMeasurements S-Type resistors but can
be used with other similar types. A diagram of the board layout on one of the
following page shows the location of each resistor.
100 KΩ 20 KΩ
BRIDGE 1000 Ω 350 Ω 120 Ω 1000 Ω 350 Ω 120 Ω
mV/V 2.4888 0.873 0.299 12.20 4.337 1.495
μstr full 1244 437 150 6098 2169 748
bridge
μstr 1/2 2488 873 300 12195 4337 1496
bridge
μstr 1/4 4975 1747 600 24390 8674 2991
bridge
I2985-1.0 en 71
GEN2i
72 I2985-1.0 en
GEN2i
A Strain gages
B Signal conductors
C Outer shield
Terminated near strain gages - signal source
D Inner shield
Driven guard at +Vcm
I2985-1.0 en 73
GEN2i
74 I2985-1.0 en
GEN2i
Figure 6.5: Reference card: LEMO FGG.2B.316 Connector, solder cup view
of male connector
I2985-1.0 en 75
GEN2i
Bridge completion
The Wheatstone bridge used in most strain gage measurement circuits usually
consists of (a) the gages for actively measuring the strains and (b) precision
resistors for completing the circuit. In the GEN2i bridge completion can be for
full, half and quarter bridge configurations. Completion resistors can be internal
(incorporated in the GEN2i) or external (if required).
76 I2985-1.0 en
GEN2i
To connect such a bridge you need a minimum of four wires. Refer to Figure
6.4 on page 74 for connection details. When using a full bridge you need to
inform the Perception software about this.
I2985-1.0 en 77
GEN2i
In this situation a sensor is used with two (out of four) resistors R4 and R3.
These resistors are placed between A-C and C-D.
You now have to provide the two additional resistors R1 and R2. To do this
there are two options:
In situation (1) you do not need to do anything from a hardware point of view.
In situation (2) you will need to add the two resistors to the plug-in module on
the locations marked HB+ and HB-. You will also need to remove the jumper
J1. Refer to for electrical/schematic details and to for mechanical/location
details.
To connect such a bridge you need a minimum of three wires. Refer to Figure
6.4 on page 74 for connection details. When using a half bridge you need to
inform the Perception software about this.
78 I2985-1.0 en
GEN2i
In this situation a sensor is used with only one resistor R4. This resistor is placed
between A and C.
I2985-1.0 en 79
GEN2i
You now have to provide the three additional resistors R1, R2, and R3 for
completion of the bridge. You do this by using the half-bridge completion as
described in the previous section and adding an additional resistor R3 between
C and D. To do this there are two options:
1 Use the standard 350 Ohm resistor inside the acquisition module.
2 Provide a resistor.
In situation (1) you do not need to do anything from a hardware point of view.
In situation (2) you will need to add the resistor on the plug-in module on the
location marked QB. Refer to for electrical/schematic details and to for
mechanical/location details.
Additional wiring: you need to add wiring for the quarter bridge completion
resistor. Depending on the selection you made, connect either pin 10 (350
Ohm) or pin 9 (user) to the bridge connection marked C in the diagram, or
directly to pin 6 (-amp in) of the connector. Refer to Figure 6.4 on page 74 for
connection details.
When using a quarter bridge you need to inform the Perception software about
this.
80 I2985-1.0 en
GEN2i
5 When you select quarter bridge completion, switch S3 in Figure 6.1 will be
closed.
Excitation
The following options are provided for bridge excitation:
You make all these selections in the Perception application. However, when
using sense lines you will need to do additional wiring:
I2985-1.0 en 81
GEN2i
Excitation on/off: Provision for separately switching off the bridge voltage
while the remainder of the measuring circuit remains operational is an important
and useful feature, particularly when measuring dynamic strains. Any output
observed when the bridge voltage is switched off must be due to electrical
noise, as the output cannot possibly be the result of resistance changes in the
measuring circuit when a bridge voltage is not present. The ability to turn off
the bridge power is therefore a useful diagnostic tool for establishing whether
electrical noise is a problem.
Voltage and current excitation: For the balanced bridge it doesn’t matter if
the power supply is of the constant-voltage or constant-current variety. In both
cases the output will be zero for the resistively balanced state. However,
resistive balance circuits may be used with constantcurrent excitation to obtain
an initial zero balance of the instrument output when the bridge itself is
unbalanced.
Sense lines: Remote sense or, more correctly, remote sensing of excitation
voltage, is commonly recommended for use with precision, commercial
transducers to prevent leadwire resistance changes (due to changes in either
temperature or length) from affecting transducer span, or sensitivity.
In Perception
To make the required settings in the Perception software do the following:
82 I2985-1.0 en
GEN2i
4 To switch the excitation on or off do one of the following (this will open/close
the switch marked S2a and S2b in the following (this will open/close the
switch marked S2a and S2b Figure 6.1):
l In the spreadsheet style matrix double-click in the correct row(s) on the
Excitation column.
You can select between one of the following excitation type options:
I2985-1.0 en 83
GEN2i
l Voltage Sense: Voltage excitation with sense. When you select this option
the sense lines are used: this will put the switch marked S1a and S1b in
Figure 6.1 into the sense position. Use the Voltage procedure to set the
required voltage. You can also use the sense check boxes (marked D in
Figure 6.6 on page 76) to toggle the sense lines.
l Current: Current excitation. Now a constant current is fed into the bridge.
To set the current level do one of the following:
l In the spreadsheet style matrix in the Excitation current column enter
the required current.
84 I2985-1.0 en
GEN2i
Additional wiring
When using remote calibration / shunt verification you will need to add the
following wiring:
For an example refer to Figure 6.4 "Full, half and quarter bridge configurations"
on page 74.
In Perception
To make the required settings in the Perception software do the following:
I2985-1.0 en 85
GEN2i
or type the value of the CAL resistor. This selection operates switch S8
in Figure 6.1.
l When external is chosen type the correct value of the external resistor
in the Shunt value column.
5 Select the bridge arm to operate switch S5 in Figure 6.1:
l In the sheet use the Active gage column to select between Positive
(A-C) or Negative (C-D).
86 I2985-1.0 en
GEN2i
The actual shunt verification is done using the shunt verification dialog.
I2985-1.0 en 87
GEN2i
H Channel select
I Shunt active gage select (S5 in Figure 6.1)
Shunt verification
To do the actual shunt verification:
Bridge balance
The bridge circuit is only in balance (has no output when the bridge voltage is
applied) provided that R1 / R2 = R4 / R3. Taking into account the various
resistance tolerances on the strain gage(s), resistors and leadwires, an initial
unbalance is invariably present. Adjustment of initial balance so that at zero
strain there is zero output is achieved by bridge balancing.
HINT/TIP
When doing a bridge balance, the GEN2i acquisition card measures the input
value at the connector of the acquisition card. This means it cannot “see” if a
bridge is actually connected or not. When no voltage is present this can be since
the bridge is balanced or that no bridge is connected.
88 I2985-1.0 en
GEN2i
To balance a bridge
To balance a bridge in Perception do the following:
I2985-1.0 en 89
GEN2i
90 I2985-1.0 en
GEN2i
Note When mixing 200 kS/s and 1 MS/s input modules within a single acquisition
group in the Perception software, the maximum acquisition speed is limited to
200 kS/s.
For the purpose of this document we only describe attenuating probes within
two categories: 1X Probes and 10X Probes.
6.5.2 1X Probes
1X probes, also known as 1:1 (one-to-one) probes, simply connect the input of
the instrument to the circuit being measured. They are designed for minimum
loss and easy connection, but otherwise they are equivalent to using a cable to
connect the instrument. Figure 6-12 shows the circuit diagram for an instrument
input connected to a circuit under test. The circuit under test is modelled as a
voltage source with a series resistor. The 1X probe (or cable) will introduce a
significant amount of capacitance that appears in parallel with the input of the
instrument. A 1X probe may have around 40 to 60 pF of capacitance.
I2985-1.0 en 91
GEN2i
The impedance of the circuit and the input impedance of the instrument together
produce a low-pass filter. For very low frequencies, the capacitor acts as an
open circuit and has little or no effect on the measurement. For high
frequencies, the capacitor's impedance becomes significant and loads down
the voltage seen by the instrument. Figure 6-13 shows this effect in the
frequency domain. If the input is a sine wave, the amplitude tends to decrease
with increasing frequency and the phase is shifted.
Frequency
92 I2985-1.0 en
GEN2i
(EQ1)
The loading due to the input impedance of the instrument and the probe
capacitance is twofold: resistive loading and capacitive loading.
The resistive loading actually reduces the voltage delivered to the instrument:
(EQ2)
(EQ3)
I2985-1.0 en 93
GEN2i
If R1C1 = R2C2, then this circuit has the result that the effect of both capacitors
exactly cancel. The capacitor is usually made adjustable and can be tweaked
for a near perfect match. Under these conditions, the relationship of VS to VIN
is:
(EQ 4)
R2 is the input resistance of the instrument's high input impedance (1 MΩ) and
R1 = 9R2. From the previous equation, this results in:
(EQ 5)
So the final result is a probe / instrument input combination that has a much
wider bandwidth than the 1X probe, due to the effective cancellation of the two
capacitors. However, the instrument now sees only one-tenth of the original
voltage (hence the name 10X probe). Also notice that the circuit being
measured sees a load impedance of R1 + R2 = 10 MΩ, which is much higher
than with the 1X probe.
94 I2985-1.0 en
GEN2i
If R1C1 = R2C2, then this circuit has the result that the effect of both capacitors
exactly cancel. The capacitor is usually made adjustable and can be tweaked
for a near perfect match. Under these conditions, the relationship of VS to VIN
is:
(EQ 4)
R2 is the input resistance of the instrument's high input impedance (1 MΩ) and
R1 = 9R2. From the previous equation, this results in:
(EQ 5)
So the final result is a probe / instrument input combination that has a much
wider bandwidth than the 1X probe, due to the effective cancellation of the two
capacitors. However, the instrument now sees only one-tenth of the original
voltage (hence the name 10X probe). Also notice that the circuit being
measured sees a load impedance of R1 + R2 = 10 MΩ, which is much higher
than with the 1X probe.
IMPORTANT
I2985-1.0 en 95
GEN2i
96 I2985-1.0 en
GEN2i
6.6.1 Specifications
General
I2985-1.0 en 97
GEN2i
General
Counter / timer 3 channels, providing:
l Up / down counter
l Frequency / RPM measurement
l Quadrature (position) measurement
Memory 512 MegaByte on board (4096 Megabit)
all binary channels (64): 64 MSamples per channel
plus 1 counter channel: 32 MSamples per channel
plus 2 counter channels: 20 MSamples per channel
plus 3 counter channels: 16 MSamples per channel
Sample rate 1 MS/s
Power output 5V @ 0.5 A typical @ 20 °C (ambient PCB)
5V @ 0.35 A typical @ 60 °C (ambient PCB)
Inputs
Input type TTL active low, pull-up resistor to enable activation by re-
lais or shortcircuit to ground
Pull-up 25.5 kΩ @ 5 Volt
Voltage range TTL compatible, 30 V maximum
Voltage levels logic treshold voltage levels:
–30 V to +0.7 V = logical “0”
+2 V to +30 V = logical “1”
Protection ± 30 V continuous
Connectors four 26-pin SubD type connectors with 16 events per con-
nector
98 I2985-1.0 en
GEN2i
Inputs
Type KF66-A26P-N
Counter
Frequency
Quadrature
The reset functionality of the counter/timer is under software control and can
be set for each channel separately.
Reset modes
Conditional functionality
I2985-1.0 en 99
GEN2i
Conditional functionality
Trigger modes: off, rising edge active, falling edge active;
combination: each event trigger is OR-ed with all other trig-
ger sources
Qualifier modes: off, active high, active low
combination: each event qualifier is AND-ed with all other
qualifier sources
Alarm modes: off, active low, active high
100 I2985-1.0 en
GEN2i
I2985-1.0 en 101
GEN2i
In the Perception software the event bits are combined within one channel and
labelled as CH1_1 through CH1_64. The counter/timer channels are referred
to as CH2 through CH4.
Counter input The counter input is the actual signal input. The counter value
will be modified on each rising edge of this signal. The maximum input rate is
10 Mhz.
102 I2985-1.0 en
GEN2i
Reset The reset signal will reset the counter to zero. The reset enabling as well
as the active level is determined under software control.
I2985-1.0 en 103
GEN2i
In the Perception software the event bits are combined within one channel and
labelled as CH1_1 through CH1_64. The counter/timer channels are referred
to as CH2 through CH4.
Counter input The counter input is the actual signal input. The counter will be
incremented on each rising edge of this signal. The maximum input frequency
is 10 MHz.
104 I2985-1.0 en
GEN2i
In the Perception software the event bits are combined within one channel and
labelled as CH1_1 through CH1_64. The counter/timer channels are referred
to as CH2 through CH4.
Clock input (A) The clock input is the actual signal input. The counter will be
incremented on each rising edge of this signal if the direction input is low (“0”).
The counter will be decremented on each rising edge of this signal if the
direction input is high (“1”).
I2985-1.0 en 105
GEN2i
Direction input (B) The direction signal determines if the counter will be
incremented (direction = “0”), or decremented (direction = “1”) on each rising
edge of the counter input.
Reset The reset signal will reset the counter to zero. The reset enabling as well
as the active level is determined under software control.
The most common type of incremental encoder uses two output channels (A
and B) to sense position. Using two code tracks with sectors positioned 90
degrees out of phase, the two output channels of the quadrature encoder
indicate both position and direction of rotation. If A leads B, for example, the
disk is rotating in a clockwise direction. If B leads A, then the disk is rotating in
a counter-clockwise direction.
By monitoring both the number of pulses and the relative phase of signals A
and B, you can track both the position and direction of rotation.
Some quadrature encoders also include a third output channel, called a zero
or index or reference signal, which supplies a single pulse per revolution. This
single pulse is used for precise determination of a reference position.
106 I2985-1.0 en
GEN2i
The counter/timer functionality uses up to 3 event bits per channel. These event
bits also keep their original functionality. You can, for example, use a
quadrature encoder and at the same time look at the quadrature signals
separately.
6.7.1 Specifications
General
I2985-1.0 en 107
GEN2i
General
Counter / timer 3 channels, providing:
l Up / down counter
l Frequency / RPM measurement
l Quadrature (position) measurement
Memory 512 MegaByte on board (4096 Megabit)
all binary channels (64): 64 MSamples per channel
plus 1 counter channel: 32 MSamples per channel
plus 2 counter channels: 20 MSamples per channel
plus 3 counter channels: 16 MSamples per channel
Sample rate 1 MS/s
Power output 5V @ 0.5 A typical @ 20 °C (ambient PCB)
5V @ 0.35 A typical @ 60 °C (ambient PCB)
Non-isolated Inputs
Input type TTL active low, pull-up resistor to enable activation by re-
lais or shortcircuit to ground
Pull-up 25.5 kΩ @ 5 Volt
Voltage range TTL compatible, 30 V maximum
Voltage levels logic treshold voltage levels:
–30 V to +0.7 V = logical “0”
+2 V to +30 V = logical “1”
Protection ± 30 V continuous
Connectors four 26-pin SubD type connectors with 16 events per con-
nector
108 I2985-1.0 en
GEN2i
Non-isolated Inputs
Type KF66-A26P-N
Fiber-optic I/O
Counter
Frequency
Quadrature
I2985-1.0 en 109
GEN2i
Quadrature
Frequency 10 MHz maximum
The reset functionality of the counter/timer is under software control and can
be set for each channel separately.
Reset modes
Conditional functionality
110 I2985-1.0 en
GEN2i
I2985-1.0 en 111
GEN2i
6.9.1 Specifications
Analog input sections
112 I2985-1.0 en
GEN2i
Acquisition
I2985-1.0 en 113
GEN2i
Triggering
Miscellaneous
Acquisition modes
114 I2985-1.0 en
GEN2i
Note When mixing 100 MS/s and 25 MS/s input modules within a single acquisition
group in the Perception software, the maximum acquisition speed is limited to
25 MS/s.
I2985-1.0 en 115
GEN2i
116 I2985-1.0 en
GEN2i
6.10.1 Installation
One board is required in the master mainframe and one board is required per
slave mainframe.
I2985-1.0 en 117
GEN2i
118 I2985-1.0 en
GEN2i
The CPU Reset Switch can be used to reset the controller/interface in the rare
event of a system malfunction. To reset the unit carefully press the recessed
switch with a small screwdriver or equivalent.
l External Timebase In This input can be used to provide a timebase for the
ADC other than the internal one. Typically used in combination with rotating
machinery where the ADC clock is synchronized with the revolutions. In the
Perception software the selection between external and internal timebase
is made in the Mainframe section of the Settings. (under development)
l External Event Out This output is software selectable between Alarm
Out and Recording Active Out. When alarm is selected, the output is
driven by channel alarm detectors. For details see “Channel alarm” . When
recording active is selected, the output is “high” when a recording is in
progress.
l External Trigger In / Out This input and output are related to the recorder
trigger logic. For details see “Recorder and system trigger” .
Note To make sure that all potential triggers are seen at the same time without
dependency from board or sample rate there exists a set constant delay of 514
μs from the original system trigger on this output.
6.11.2 Options
The GEN2i Portable data recorder can be equipped with a variety of options.
These options are factory-installed, i.e. you must choose an option at ordering
time or return the instrument to a qualified service point for upgrade.
The controller/interface board has one expansion slot that can be used for one
of the following options:
I2985-1.0 en 119
GEN2i
The following option does not use the expansion slot and can therefore be
combined with one of the above options:
120 I2985-1.0 en
GEN2i
7 GEN2i Options
7.1 Introduction
Your GEN2i Portable data recorder can be equipped with a variety of options.
These options are factory-installed, i.e. you must choose an option at ordering
time or return the instrument to a qualified service point for upgrade.
The system controller board of the GEN2i mainframe has one expansion slot
that can be used for one of the following options:
The following option does not use the expansion slot and can therefore be
combined with one of the above options:
I2985-1.0 en 121
GEN2i
7.2 Options
At printing time of this manual the following options were available.
122 I2985-1.0 en
GEN2i
I2985-1.0 en 123
GEN2i
A Specifications
A.1 General
This section describes the specifications pertinent to the GEN2i mainframe,
including the PC section. For detailed specifications of the various data
acquisition input modules refer to the section "Available input modules" on
page 61. For detailed information on the Perception software refer to the
separately supplied User Manual for the Perception software.
PC
Component Unit Description Unit Value
Memory DDR2 RAM 4 GB
PC Board AMD based Low Power AMD M690T
Processor AMD Turiontm 64 Dual 2.1 GHz
core TL-62
Ethernet Copper RJ45 Ethernet 1 Gbit/s
Connection
Wireless LAN Embedded 801.11b/g 54 Mbit/s
USB Connectors USB 2.0 6 on back + 2 on front
Internal Storage MLC Solid State Drive 256 GB*
Data streaming to drive ≈ 20 MB/s direct
16 Channels x 500 kS/s
@ 16 bits
Display TFT SXGA Touch Screen 17” / 1280x1024 resolu-
tion
Video connection DVI and VGA 1x DVI, 1x VGA
Multiple Monitors Support Clone Mode and Exten-
ded Mode
Speaker Internal speaker 1
Line Out Jack plug 3.5 mm 1
Microphone Jack plug 3.5 mm 1
Accessories Keyboard / Mouse 1/1
Software
Component Unit Description Unit Value
Operating system Microsoft® Windowstm 7
124 I2985-1.0 en
GEN2i
Software
Component Unit Description Unit Value
GEN series firmware Includes High Speed Up to 20 MB/s (aggregate
Streaming option, data transfer rate)
enabling direct-to-disk
acquisitions
DAQ software Perception standard Refer to Perception
package, 6.10 or higher specification sheet for
details
Instrument panel Setup of instrument
Touch Interface Acquisition control
(Fully touch-optimized) Display Data: Live / Review
Basic Measurements
Export + Archiving
Basic Reporting
Power
Component Unit Description Unit Value
Power Inlet Primary Power Inlet 47-63 Hz 100-240 Vac
Power Total Power of unit 250 VA
(maximum)
Hardware
Component Unit Description Unit Value
Backplane Acquisition slots – See also 2
Acquisition boards
Optional Interfaces IRIG 1x Optional interface only
IRIG/GPS timing
SCSI (Perception only storage)
I/O Alarm out TTL, Level active
Trigger in TTL, 50 ns resolution,
minimum pulse width
500 ns (Max non
destructive voltage: 30 V)
Trigger out/Recording out TTL, edge active, pulse
width 10 μs/recording
duration (software
selectable)
Timebase in TTL
Timebase Accuracy < 30 ppm for internal
sources
Base Binary, decimal
I2985-1.0 en 125
GEN2i
Hardware
Component Unit Description Unit Value
External Up to 500 kS/s or 1MS/s
depending on input
module (5 MS/s as
special); programmable
divider per module
Physical/Environmental
Component Unit Description Unit Value
Weight GEN2i Mainframe (exc. 9.5 kg (add ≈ 1 kg per acq
acquisition boards) board)
(20.9 lbs (add ≈ 2.2 lbs
per acq board))
Dimensions Height/Height with 34.2 cm/39.2 cm
handle (13.5"/15.4")
Width 43.6 cm (18.2")
Depth 18.6 cm (7.3")
Physical/Environmental
Component Unit Description Unit Value
Shipping Carton Weight GEN2i + shipping 18.2 kg (40.1 lbs)
(Gross) container
Shipping Carton Height 59.0 cm (23.2")
dimensions Width 49.7 cm (19.6")
126 I2985-1.0 en
GEN2i
Physical/Environmental
Component Unit Description Unit Value
Depth 55.6 cm (21.9")
Shipping Case With wheels and a 1x Optional
handle, tested according
to ASTM D4169-04 Level
1 (drop), and ASTM
D4728 E (Vibration and
shock)
Shock Operational 10 g @ 11 ms half sine
According to IEC-68-2-27 Non-Operational 25 g @ 6 ms half sine
Vibration Operational 1 g rms
According to Non-Operational 2 g rms
IEC-60068-2-34
Temperature range Operating 0 °C to +40 °C (+32 °F to
+ 104 °F)
Storage -20 °C to +60 °C (-4 °F to
+ 140 °F)
Humidity Relative humidity (non- 0 % to 80 %
condensing)
Altitude Operational altitude 2000 m (6100 ft)
(maximum)
Acoustic Noise The total A-weighted SPL 60 dBA @ 0.6 m
of the GEN2i Maximum
Temperature Sensors Temperature for Multiple
monitoring and Air Flow
Control
Cooling Fans Provides Air Flow 2
Handle Carrying handle, also 1 adjustable
used for higher tilt angles
Tilting Feet Retractable feet for small 2
tilt angles
Grounding Banana plug 1
Casing Aluminum/plastic cover
Accessories Soft carry case with strap 1
for transportation
included with hardened
front and back for
protection, and storage
pouches for mouse and
keyboard
Rack Mount Kit 19 inch rack mount kit 1 optional
I2985-1.0 en 127
GEN2i
Acquisition Boards
Model Input Isolation Sample Resolu- Memory No. of
type rate tion Chan-
(Max) nels
Basic200 Single no 200 kS/s 16 bit 64 MS 8
Ended
Basic200 Unbal- yes 200 kS/s 16 bit 64 MS 8
XT iso anced Diff
Basic1M Single no 1 MS/s 16 bit 128 MS 8
Ended
Basic1M Unbal- yes 1 MS/s 16 bit 256 MS 8
iso anced Diff
Basic1M Unbal- yes 1 MS/s 16 bit 256 MS 8
XT iso anced Diff
Bridge200 Bridge/ yes 200 kS/s 16 bit 64 MS 4
iso Diff
Bridge1M Bridge/ yes 1 MS/s 16 bit 256 MS 4
iso Diff
Uni200 iso Diff/ICP/ yes 200 kS/s 16 bit 64 MS 4
Shunt
Uni1M iso Diff/ICP/ yes 1 MS/s 16 bit 256 MS 4
Shunt
HiSpeed SE/Diff no 25 MS/s 15 bit 64 MS 4
25M
HiSpeed SE/Diff no 100 MS/s 14 bit 900 MS 4
100M
IsoDig MV Unbal- yes 25/100 14/15 bit 100-900 1t / 4r*
anced Diff MS/s MS
IsoDig HV Unbal- yes 25/100 14/15 bit 100-900 1t / 4r*
anced Diff MS/s MS
Marker1M Binary no 1 MS/s 1 bit 512 MB 64
Marker1M Binary yes/no 1 MS/s 1 bit 512 MB 8 / 32
HV
128 I2985-1.0 en
GEN2i
The GEN2i by itself cannot directly be mounted into a 19-inch rack. For this you
need to use the optional 19-inch rack mount kit.
I2985-1.0 en 129
GEN2i
There are only 2 screws which are fastened to the upper portion of
both sides of the brackets as seen in Figure B.2.
130 I2985-1.0 en
GEN2i
C System Restore
C.1 Restoring the GEN2i
Your GEN2i comes with an internally based System Image Restore. This USB
contains a complete software image of the system as delivered.
You might need to restore the original image because Windows won’t boot, or
because your system drive died, or for other possible reasons.
1 Insert the GEN2i System Image Restore USB into a spare USB port.
2 Turn on the system and press the F11 key on the keyboard until the Boot
Device selection window is shown.
3 From the list of devices select the USB Unknown device and press the
Enter key to continue.
4 When the USB is booted the following dialog is shown:
Read the text carefully and select the “I understand the consequences”
check box to enable the Yes button. Click Yes to restore the image. To
cancel the process click No. Wait until the image is restored.
WARNING
Do not turn off or reboot the system while restoring the image.
I2985-1.0 en 131
GEN2i
5 After the image has been restored, remove the USB from the port and
press any key to reboot the system.
After the system has rebooted, the system will automatically continue to finalize
the restore process. The following is the list of steps to complete.
1 Language From the dialog, select the appropriate user interface language
and click Next.
132 I2985-1.0 en
GEN2i
2 Regional and keyboard The regional and keyboard settings must be se-
lected.
Select the correct settings from each of the drop-down combo boxes and
click Next to proceed.
I2985-1.0 en 133
GEN2i
3 Activation In the next step, the Windows 7 product key must be entered.
The product key sticker is attached to the back of the system.
When you re-image the GEN2i, the system hard disk is erased. Therefore
you need to activate the current copy of your Microsoft Vista again. You
can activate either online or by phone. To activate online you need an
Internet connection. Activating by phone requires interacting with an au-
tomated phone system. When you have an Internet connection, activation
will be done automatically after the re-image process.
134 I2985-1.0 en
GEN2i
4 License In the next step, the license terms are shown. You must accept
these before the system can continue.
I2985-1.0 en 135
GEN2i
The original system name is the serial number of the system which starts
with IEX and can be found on the back of the system.
136 I2985-1.0 en
GEN2i
6 Review The last configuration step is used to review and configure the
time and date settings.
Verify the date and the time and select the correct time zone from the Time
zone box. Click Next to continue.
WARNING
Using the incorrect time zone results in incorrect date and time informa-
tion in the recordings.
I2985-1.0 en 137
GEN2i
7 Ready The system is now ready to be used. Click Start to start using the
system.
Windows will start to check the computer’s performance. This will take ap-
proximately 2-3 minutes.
138 I2985-1.0 en
GEN2i
If servicing is ever needed on your GEN2i, contact the factory with the model
and serial numbers, a description of the problem, and your contact information.
You will be issued a Return Material Authorization (RMA) number. Attach this
number to the unit and/or the accompanying paperwork.
During the warranty period, the customer pays for shipping to HBM. HBM will
pay for the return of the equipment in the same fashion as it was received.
Outside the warranty period, a quote will be given. A purchase order must be
received before work can be performed.
I2985-1.0 en 139
GEN2i
HBM recommends that the GEN2i system should be tested and if necessary,
calibrated, at one year intervals or after any major event that may effect
calibration. When in doubt consult your local supplier.
140 I2985-1.0 en
GEN2i
D.3 Cleaning
To clean the instrument, disconnect all power sources and wipe the surfaces
lightly with a clean, soft cloth dampened with water.
The GEN2i does not require additional routine cleaning. If the cooling inlets on
the side and rear of the instrument become clogged with dust, use a small brush
and/or vacuum cleaner to remove the dust.
I2985-1.0 en 141
GEN2i
Index
A Connectors
Binary marker HV module ............................. 110
Acquisition ............................................................ 36 Binary marker module ........... 100, 101, 102, 105
Alarm Bridge .............................................................. 71
Output .................................................... 118, 119 Reference card (bridge) .................................. 75
Controller ................................................ 35, 63, 118
B Counter ......................................................... 97, 107
CPU reset ........................................................... 119
Backplane ............................................................. 35 Current input, see Universal amplifier .................. 90
Balancing, bridge .................................................. 88
Basic amplifier ................................................ 61, 64
D
Batteries ............................................................... 22
Binary marker HV input ................................ 61, 107 Data storage ......................................................... 39
Binary marker input ........................................ 61, 97 Declaration of conformity ...................................... 30
Board calibration .................................................. 47 Differential
Bridge amplifer ~ input ............................................................. 90
Configuration ................................................... 73 ~ measurements ............................................. 96
Bridge amplifier .............................................. 61, 68 Differential high speed digitizer .......................... 112
Balancing ......................................................... 88 Driven guard ......................................................... 72
Completion ...................................................... 71
Completion/Full ............................................... 76
Completion/Half ............................................... 78 E
Completion/Jumper ......................................... 71
EC Declaration of conformity ................................ 30
Completion/Quarter ......................................... 79
EMC ...................................................................... 16
Configuration ................................................... 69
Ethernet ...................................................... 121, 123
Connector reference card ................................ 75
Event out ............................................................ 119
Connectors ...................................................... 71
Events, see Binary marker input .......................... 97
Excitation ......................................................... 81
Excitation, bridge .................................................. 81
Sense ........................................................ 81, 82
Expansion, see Options ..................................... 121
Shunt calibration ........................................ 71, 85
External alarm .................................................... 119
External recording active .................................... 119
C External timebase ....................................... 118, 119
External trigger ................................................... 118
Calibration (shunt) .......................................... 71, 85
Calibration (system) ............................................. 47
Cautions ............................................................... 31 F
CE ........................................................................ 30
FCC ........................................................................ 9
Cleaning ............................................................. 141
Feet ...................................................................... 53
Completion (bridge) .............................................. 71
Front panel control
Jumper ............................................................ 71
USB Ports ........................................................ 54
Conditioning, see Signal conditioning .................. 38
Fuse ..................................................................... 18
Configuring and using the bridge amplifier ........... 75
Connecting power ................................................ 44
142 I2985-1.0 en
GEN2i
G StatStream ...................................................... 36
IRIG and IRIG/GPS .................................... 121, 122
Gate-time ............................................................ 104 Isolated digitizers .................................................. 63
Guard, see Driven guard ...................................... 72
J
H
Jumper (bridge completion) .................................. 71
Handle .................................................................. 49
Hardware .............................................................. 35
High speed digitizer .............................. 61, 111, 112 L
High speed Ethernet ................................... 121, 123
LICENSE and WARRANTY ................................... 3
High voltage modules ........................................... 63
Linearity verification .............................................. 71
I
M
ICP, see Universal amplifier ................................. 91
Master/Slave .......................................... 35, 63, 116
Input modules ....................................................... 35
Master/Slave Operation ...................................... 117
Basic amplifier ........................................... 61, 64
Motherboard connections
Binary marker .................................................. 61
Audio ............................................................... 45
Binary marker HV .......................................... 107
Ethernet ........................................................... 46
Binary marker HV/Connectors ....................... 110
Serial Port ........................................................ 46
Binary marker/Connectors ............................. 100
USB ................................................................. 45
Bridge amplfier/Configuration .................... 69, 73
Mount GEN2i in a 19-inch rack .......................... 129
Bridge amplfier/Connector reference card ...... 75
19 inch brackets installation .......................... 129
Bridge amplifer ................................................ 61
Bridge amplifier ............................................... 68
Bridge amplifier/Completion ............................ 71 N
Bridge amplifier/Connectors ............................ 71
Bridge amplifier/Shunt calibration .................... 71 Network interfacing ............................................... 57
High speed digitizers ....................................... 61
High speed digitizers (differential) ................. 112 O
High speed digitizers (single ended) ............. 111
High voltage modules ...................................... 63 Options
Universal amplifier ..................................... 61, 90 High speed Ethernet .............................. 121, 123
Interface module ........................................... 35, 118 IRIG and IRIG/GPS ......................... 35, 121, 122
Introducing the GEN2i SCSI ................................................ 35, 121, 123
Software .......................................................... 33 Overvoltage .......................................................... 19
Transient recorder ........................................... 33
Introduction
Acquisition ....................................................... 36 P
Backplane ........................................................ 35
PC connections .................................................... 45
Controller ......................................................... 35
Digital Display connector ................................. 45
Data storage .................................................... 39
PC Section ........................................................... 40
Hardware ......................................................... 35
Perception Software ............................................. 41
Input modules .................................................. 35
Position measurement ........................................ 104
Interface module .............................................. 35
Power and frequency ........................................... 15
Master/slave .................................................... 35
Preventive maintenance ..................................... 140
Signal conditioning .......................................... 38
I2985-1.0 en 143
GEN2i
Probes .................................................................. 96 U
Protective cover
Feet/Turning feet in ......................................... 53 Unbalanced differential ......................................... 67
Feet/Turning feet out ....................................... 53 Universal amplifier .......................................... 61, 90
Handle/Turning handle .................................... 50
V
Q
Verification (shunt) ............................................... 84
Quadrature ......................................................... 104 Verification of linearity .......................................... 71
R W
Safety
Current ............................................................ 19
EMC ................................................................ 16
FCC and general ............................................... 9
Fuse ................................................................ 18
International warnings ..................................... 23
Overvaltage ..................................................... 19
Power and frequency ...................................... 15
SCSI ........................................................... 121, 123
Sense, bridge ................................................. 81, 82
Service and maintenance
General .......................................................... 139
Shield ................................................................... 72
Shunt calibration ................................................... 71
Signal conditioning ............................................... 38
Single ended high speed digitizer ...................... 111
StatStream ........................................................... 36
Storage, see Data storage ................................... 39
System calibration ................................................ 47
Timebase
External ................................................. 118, 119
Timer ............................................................ 97, 107
144 I2985-1.0 en
Head Office
HBM
Im Tiefen See 45
64293 Darmstadt
Germany
Tel: +49 6151 8030
Email: [email protected]
France
HBM France SAS
46 rue du Champoreux
BP76
91542 Mennecy Cedex
Tél:+33 (0)1 69 90 63 70
Fax: +33 (0) 1 69 90 63 80
Email: [email protected]
Germany
HBM Sales Office
Carl-Zeiss-Ring 11-13
85737 Ismaning
Tel: +49 89 92 33 33 0
Email: [email protected]
UK
HBM United Kingdom
1 Churchill Court, 58 Station Road
North Harrow, Middlesex, HA2 7SA
Tel: +44 (0) 208 515 6100
Email: [email protected]
USA
HBM, Inc.
19 Bartlett Street
Marlborough, MA 01752, USA
Tel : +1 (800) 578-4260
Email: [email protected]
PR China
HBM Sales Office
Room 2912, Jing Guang Centre
Beijing, China 100020
Tel: +86 10 6597 4006
Email: [email protected]