Luisa Manual Beatmung
Luisa Manual Beatmung
Luisa Manual Beatmung
LUISA
Ventilators
Contents
Contents
1 Introduction 3 5 Settings in the menu 15
1.1 Intended use .................................................... 3 5.1 Navigating in the menu ................................. 15
1.2 Description of function ..................................... 3 5.2 Menu structure for patient menu .................. 15
1.3 User qualification ............................................. 3
6 Hygiene treatment and servicing 17
1.4 Indications ........................................................ 3
6.1 Hygiene treatment ........................................ 17
1.5 Contraindications ............................................. 3
6.2 Function check .............................................. 19
1.6 Side effects ...................................................... 4
6.3 Checking alarms ............................................ 20
2 Safety 4 6.4 Servicing ....................................................... 20
2.1 Safety information ............................................ 4 6.5 Disposal ........................................................ 20
2.2 General information ......................................... 5
7 Alarms 21
2.3 Safety information in these instructions for
7.1 Sequence in which alarms are displayed ........ 21
use ...................................................................5
7.2 Muting alarms ............................................... 21
3 Product description 6 7.3 Configuring physiological alarms ................... 21
3.1 Overview .......................................................... 6
7.4 Technical alarms ............................................ 23
3.2 Control panel in display .................................... 7
7.5 Nurse call and remote alarm .......................... 27
3.3 Symbols in display ............................................ 8
3.4 Accessories (optional) ....................................... 8 8 Faults 27
3.5 Operating states ............................................... 9 9 Technical specifications 28
3.6 Batteries ........................................................... 9
10 Annex 33
3.7 Data management/compatibility ..................... 10 10.1 Pneumatic diagram ....................................... 33
4 Preparation and operation 11 10.2 System resistances ......................................... 34
4.1 Setting up and connecting device ................... 11 10.3 Emission of electromagnetic interference ....... 35
4.2 Connecting breathing tube ............................ 11 10.4 Electromagnetic interference immunity .......... 35
4.3 Before first use ............................................... 13 10.5 Markings and symbols ................................... 36
4.4 Switching on device ....................................... 13 10.6 Scope of delivery ........................................... 36
4.5 Starting therapy ............................................. 13 10.7 Accessories .................................................... 37
4.6 Ending therapy/switching off device ............... 13 10.8 Removable parts ............................................ 37
4.7 Performing tube test ...................................... 14 10.9 Warranty ...................................................... 38
4.8 Calibrating FiO2 cell ........................................ 14 10.10 Declaration of Conformity ........................... 38
|
2 EN
1 Introduction
1 Introduction Therapy data are stored in the device and can additionally be
loaded on a USB-C stick and analyzed by PC software.
EN |3
2 Safety
|
4 EN
2 Safety
Supplying oxygen without a special safety device can lead to • Connect only the licensed products and modules in
fire and injure people. accordance with these instructions for use. The products
Follow the instructions for use for the oxygen supply must meet the product standard applicable to them.
system. Non-medical equipment should be positioned out of the
Set up oxygen sources at a distance of over 1 m from patient's vicinity.
the device.
• Follow the section on hygiene treatment (see "6 Hygiene
The oxygen rate supplied in l/min must not exceed the
treatment and servicing", page 17) to avoid infection or
oxygen flow prescribed by the physician.
bacterial contamination.
The oxygen rate supplied in l/min may not exceed the
set HFT flow rate. • In the event of a power outage, all settings including
At the end of therapy, shut off the oxygen supply and alarm settings are retained.
allow the device to run on briefly to flush residual
oxygen out of the device. • In the EU: As a user and/or patient, you must report any
serious incident occurring in conjunction with the
product to the manufacturer and to the responsible
2.1.5 Transport authority.
Operating the device in any kind of carrying bag may impair
device performance and injure the patient. Water and dirt in
the device may damage the device. 2.3 Safety information in
Only operate the device in the associated LUISA mobility
bag. these instructions for use
Transport or store the device in the associated LUISA
protective bag. Indicates an unusually significant
hazardous situation. If you ignore this
2.1.6 Wireless module instruction, severe irreversible or fatal
injuries may result.
The device contains a wireless module. Operating the device Indicates a hazardous situation.
in the immediate vicinity of people and/or other antennas If you ignore this instruction, mild or
may injure people, damage the device or impair device moderate injuries may result.
performance. Indicates a harmful situation.
Set up the device at least 20 cm away from any people. If you ignore this instruction, material
Do not set up or operate the device with other damage may result.
antennas.
Indicates useful information within
procedures.
LMT 68651 11/2020
EN |5
3 Product description
3 Product description
3.1 Overview
1 2 3 4 5 6
23
22 12
21 7 8 9 10 11
13
18 17
14
20 16 15
19
1 External battery connection 13 Filter compartment with coarse dust filter and fine filter
2 Connection for monitor/prismaHUB 14 Compartment for internal battery
3 USB-C connection 15 Exhalation tube connection
4 Nurse call system connection 16 Device outlet port
5 Power supply indicator 17 Handle
6 Alarm acknowledgement key 18 On/off key
7 Pressure measuring tube connection 19 Power supply unit with power supply unit cable
8 Valve control tube connection 20 Power cable
9 SpO2 connection 21 O2 supply connection
10 CO2 connection (not in use) 22 Loudspeaker
11 Nebulizer connection (not in use) 23 Power supply unit connection
12 Breathing tube
LMT 68651 11/2020
|
6 EN
3 Product description
9 2
8 5
7 6
1
1 Status line - symbols indicate current device status (e.g. accessories connected, battery capacity).
2 Alarm acknowledgement key -
Press briefly: Acknowledges alarm. If the alarm persists, the alarm is muted for 120 seconds.
Press and hold: Mutes all acoustic alarms for 2 minutes.
Press briefly again: Suspends alarm muting.
3 Home key - switches the view back to the home screen.
4 Menu keys - provide access to the individual menus.
5 Display lock key - locks or unlocks the display, so that no settings can be changed as a result of incorrect contact.
6 Dimmer key - switches to night mode and the display goes dark. Touch the display to reactivate it.
Keep key depressed - opens the Display menu.
7 Program key - provides access to the ventilation programs.
8 Ventilation key - starts or stops ventilation.
9 Access key - locks or unlocks the expert menu.
LMT 68651 11/2020
EN |7
3 Product description
|
8 EN
3 Product description
EN |9
3 Product description
|
10 EN
4 Preparation and operation
4 Preparation and
operation
4.1 Setting up and connecting
device
EN | 11
4 Preparation and operation
1 2
1. Push the breathing tube onto the device outlet port. 1. Push the free end of the breathing tube 1 onto the
device outlet port.
2. Connect the invasive or non-invasive patient/ventilator
interface to the leakage circuit (see instructions for use 2. Push exhalation tube 3 onto the device inlet for exhaled
for the patient/ventilator interface). air.
3. Connect the pressure measuring tube 2 to
4.2.2 Connecting single circuit with connection P .
valve 4. Connect the patient/ventilator interface (e.g. mask) to
the Y-piece of the circuit.
1. Push the free end of the breathing tube onto the device
outlet port.
2. Connect mouthpiece to the tube (see instructions for use
1. Push the free end of the breathing tube onto the device for the patient/ventilator interface).
outlet port.
As an alternative to the leakage circuit, it is also
2. Connect the pressure measuring tube to possible to use a single circuit with valve or double
connection P . circuit for mouthpiece ventilation.
3. Connect the valve control tube to connection .
LMT 68651 11/2020
|
12 EN
4 Preparation and operation
EN | 13
4 Preparation and operation
|
14 EN
5 Settings in the menu
5.1 Navigating in the menu
ACTION FUNCTION ACTION FUNCTION
Function keys have a gray Opens range of values for setting
Press on value
background and the function is ventilation parameters
displayed on the key in text or as Move range of values
Decrease or increase value
a symbol (e.g. System, Start up or down
ventilation, or .
Press function key Confirm value
Symbols on a black background are
not function keys, but serve to Discard selection
provide information about device
status (see "3.3 Symbols in display", Switches the view back to the home
page 8). screen
Scroll in list Navigate up or down
Ventilation programs
Alarm list
Tube test
Event list
FiO2 cell
Alarm/event list Alarm volume
Export therapy data
Trends Display
p y
Flight mode Language
Parameter overview
Device settings Date and time
Device usage
Device status
5.2.1 Menu views in the patient menu In the On state: Remaining device life if
being supplied by battery
The Views menu shows 2 views. In the Standby state: Charging state of
the internal battery in percent assuming
a power supply
Parameters and set values for the
ventilation programs To switch to the next view in each case, tap the Views key
again. The horizontal lines on the Views key are the number
of available views.
LMT 68651 11/2020
EN | 15
5.2.2 Report menu in the patient menu 5.2.4 Device settings submenu
(usage data)
Information about the parameters in this menu can be found PARAMETER DESCRIPTION
in the table below. The patient can set the alarm level here.
1= very quiet, 2= quiet, 3= loud, 4=
PARAMETER DESCRIPTION Alarm volume
very loud
Lists the alarms which have occurred. You can test the alarms here.
The log is retained when the alarm You can set brightness, orientation,
system or the device is switched off. Display
and the screen background here.
The start and end of ventilation is
You can set the language and, if
recorded.
Language required, another language for the
The log is retained even if the device is
Alarm list alarm messages here.
disconnected from the power supply
and the batteries are removed. You can set the current date and time
Date and time
The log can store 1,000 alarms. Once here.
this capacity limit has been reached,
the oldest alarm is deleted and the 5.2.5 Ventilation menu
new alarm is saved.
Event list Lists the events that have occurred. The Ventilation menu shows the settings of the current
Lists the alarms and events which have ventilation parameters. The parameters which can be
Alarm/event list displayed and modified vary depending on the ventilation
occurred in chronological order.
mode set.
Displays daily summaries (median and
Trends In the Ventilation menu, you can
percentiles) for selected signals.
Lists all parameters and set values for • select a ventilation mode
Parameter overview the up to 4 ventilation programs
which can be configured. • set ventilation parameters
Obtain information here about the • set a circuit
patient’s therapy (duration, days used,
program proportions) and about • select the patient type: Child or adult
Device usage
device usage (operating time, internal
battery life remaining or charging 5.2.6 Alarms menu
state of internal battery in percent).
Obtain information here about the The Alarms menu shows the alarms and alarm limits set for
device (name, type, serial number of the selected ventilation mode. You can adjust the values.
Device status
device and components, firmware
version).
|
16 EN
6 Hygiene treatment and servicing
6.1 Hygiene treatment 6.1.2 Cleaning intervals
INTERVAL ACTION
Clean device (see "6.1.3 Subjecting device
Weekly
to a hygiene treatment", page 17).
Clean coarse dust filter (see " Cleaning
Risk of infection when the device is used again! coarse dust filter (gray filter)", page 18).
If the device is used by several patients, infections may be
Replace fine filter (see " Replacing fine filter
transmitted to the next patient and the device contaminated.
Monthly (white filter)", page 18).
Do not reuse disposables.
Clean filter for cooling air fan
Use the breathing system filter.
(see " Cleaning filter for cooling air fan",
page 19).
Every Replace coarse dust filter (see " Cleaning
6 months coarse dust filter (gray filter)", page 18).
Risk of injury due to contaminated or infected patient • Have the device subjected to a hygiene
circuit!
A contaminated or infected patient circuit may transmit treatment by the manufacturer or by an
contamination or infections to the next patient. authorized specialist dealer in line with
Do not reprocess disposable circuits. the service and repair instructions. The
Keredusy process can be used as an
6.1.1 General information alternative to manual disinfection.
On change
• Clean or replace the exhalation module.
of patient
• Wear appropriate safety gear (e.g. safety gloves) for the The black exhalation module (included
disinfecting process. in scope of delivery) is a disposable and
must be replaced. The black translucent
• Follow the instructions for use for the disinfectant used.
exhalation module (has to be ordered
Solutions containing alcohol (25 g ethanol (94 %-
separately) is suitable for autoclaving.
strength), 35 g propan-1-ol per 100 g) are suitable.
Set device to factory settings.
Recommended: Mikrozid AF liquid or perform advanced
Alcohol EP.
6.1.3 Subjecting device to a hygiene
• Following a hygiene treatment by the authorized
specialist dealer, the device is suitable for using again treatment
with other patients.
• The following gas route components may be
contaminated following use of the device:
Risk of injury from electric shock!
• LMT 31494 Device outlet port Ingress of liquids may lead to a short-circuit, injure the user
• LMT 31497 O2 measuring cell seal and damage the device.
Disconnect the device from the power supply before the
• LMT 31496 Flow sensor hygiene treatment.
• LMT 31505 Nonreturn valve, complete Do not immerse the device and components in liquids.
Do not pour liquids over the device and components.
• LMT 31530 Sound insulation case, pressure side
1. Wipe over the housing including the device outlet port,
• LMT 31490 Blower the power cable, and the display with a damp cloth.
Use water or mild detergent.
• LMT 31525 Sound insulation case, intake side
2. Replace the mask, breathing tube, coarse dust filter, fine
• LMT 31446 Central part of housing filter, filter for the cooling air fan, and the breathing
• WM 29389 Fine filter system filter.
3. Perform function check (see "6.2 Function check",
• LMT 31487 Coarse dust filter
page 19).
• LMT 31422 Filter holder
LMT 68651 11/2020
EN | 17
Cleaning coarse dust filter (gray filter) Cleaning exhalation module
|
18 EN
Cleaning filter for cooling air fan 10. Check the functionality of the batteries:
1. Open exhalation module compartment (see " Cleaning • Disconnect the device from the power supply.
exhalation module", page 18). The first external battery (if present) takes over energy
supply (watch what is shown in display).
• Disconnect the first external battery from the device.
The second external battery (if present) takes over
energy supply.
• Disconnect the second external battery from the
device.
The internal battery takes over energy supply.
11. Check the charging state of the batteries
(see "5.2.1 Menu views in the patient menu", page 15).
If the batteries are not charged, leave the device
connected to the power supply to charge batteries.
2. Remove filter for cooling air fan. 12. If a FiO2 measuring cell is in use: Perform FiO2 calibration
(see "5.2.3 Menu system in the patient menu",
3. Wash filter under running water.
page 16).
4. Allow filter to dry.
13. If one of the items is not OK or pressure deviates by
5. Replace filter in the holder. > 1 hPa: Do not use device and contact your specialist
6. Close exhalation module compartment. dealer.
14. If required: Check alarms (see "6.3 Checking alarms",
page 20).
6.2 Function check
Carry out a function check before using the device for the
first time, after every hygiene treatment and repair, but at
least every 6 months.
1. Check device for external damage.
2. Check connectors, cables, and breathing tube for
external damage.
3. Check accessories such as the breathing system filter,
external batteries, and SpO2 sensor for external damage.
Follow the associated instructions for use.
4. Check that components are connected to the device
correctly (see "4.2 Connecting breathing tube",
page 11).
5. Connect the device to the power supply
(see "4.1 Setting up and connecting device", page 11).
6. Switch on device (see "4.4 Switching on device",
page 13).
The device automatically performs a few function tests
on the sensor system. If the device is fully functional, the
home screen is displayed and the device switches to
standby.
7. Perform a tube test (see: System > Tube test menu).
If the tube test is failed, proceed according to the
troubleshooting table (see "8 Faults", page 27).
8. Seal the end of the tube and start ventilation. A brief
alarm signal must be audible on starting. The device
automatically performs a few function tests.
The alarm key lights up yellow and red.
LMT 68651 11/2020
EN | 19
6.3 Checking alarms
6.3.1 Non-specialist user
(patient or relatives)
ALARM ID NO. REQUIREMENT TEST
On a single circuit with valve:
Alarm limit is set to a value <150 l/m
With leakage ventilation: Leave breathing tube open at patient connection.
Leakage high
Alarm limit is set to a value <60 l/m Start ventilation.
(High level of 459
On a double circuit, 15 mm / 22 mm: Wait at least 30 seconds, more alarms may occur
leakage)
Alarm limit is set to a value <60 l/m during this period.
On a double circuit, 10 mm:
Alarm limit is set to a value < 35 l/min
Pressure low
(Low airway
Alarm limit is set to a Leave breathing tube open at patient connection.
pressure, low 457
value 6 hPa Start ventilation.
pressure on
inspiration)
Connect lung bag. Start ventilation.
Exhalation Single circuit with valve is connected. On a single circuit with valve: Seal exhalation opening
blocked 757 Alternatively of patient valve.
(Obstruction) Double circuit is connected. On a double circuit: Take the exhalation tube off the
device and seal the connection on the tube.
Start ventilation.
Tidal volume low
Disconnect breathing tube at the exhalation
(Low volume 450 Double circuit: Alarm limit is set.
connection of the device.
exhaled)
Wait 3 breaths.
FiO2 low O2 cell is fitted and activated.
Start ventilation.
(Oxygen 494 Alarm limit is set.
concentration) No external oxygen supply available.
Start ventilation until the internal battery has
Battery capacity Device is not connected to the power
551 15 minutes’ life remaining before it discharges
low supply.
completely.
Start ventilation until the internal battery has
Battery capacity Device is not connected to the power
550 5 minutes’ life remaining before it discharges
critical supply.
completely.
Energy supply via Disconnect power cable from device. Disconnect the
584 None
internal battery cable for the external batteries from the device.
|
20 EN
7 Alarms
Pressure high
456 Maximum pressure exceeded. Check therapy and alarm settings.
Frequency low
452 Minimum respiratory frequency undershot. Check therapy and alarm settings.
LMT 68651 11/2020
EN | 21
7 Alarms
settings.
Oxygen sensor incorrectly calibrated. Calibrate oxygen sensor.
|
22 EN
7 Alarms
Computer module
temperature high Ambient temperature too high. Operate device at an ambient temperature
264
of 5 °C to 40 °C.
Disconnection, device
outlet port Circuit is not connected to the device
460 Check circuit and tube connections.
correctly or is not connected at all.
Disconnection, airway
Pressure measuring tube is not
pressure
461 connected to the device correctly or is Check pressure measuring tube.
not connected at all.
Disconnection,
Circuit and/or exhalation module is not
exhalation module Check circuit, tube connections, and exhalation
463 connected to the device correctly or is
module.
not connected at all.
Temperature of battery
Battery will switch off due to temperature.
E2 critically high
548 External battery 2 too warm. Operate device at an ambient temperature
of 5 °C to 40 °C.
No internal battery
Contact your specialist dealer. Have internal
553 No internal battery.
LMT 68651 11/2020
battery inserted.
EN | 23
7 Alarms
Internal battery
Battery has switched off due to temperature.
overheated
556 Internal battery overheated. Operate device at an ambient temperature
of 5 °C to 40 °C.
Unable to charge
internal battery Contact your specialist dealer.
558 Internal battery defective.
Have battery replaced.
Temperature of internal
battery high Operate device at an ambient temperature
559 Internal battery too warm.
of 5 °C to 40 °C.
Temperature of internal
battery low Operate device at an ambient temperature
560 Internal battery too cold.
of 5 °C to 40 °C.
Temperature of battery
E1 high Operate device at an ambient temperature
568 External battery 1 too warm.
of 5 °C to 40 °C.
Temperature of battery
E2 high Operate device at an ambient temperature
569 External battery 2 too warm.
of 5 °C to 40 °C.
Temperature of battery
E1 low Operate device at an ambient temperature
570 External battery 1 too cold.
of 5 °C to 40 °C.
LMT 68651 11/2020
|
24 EN
7 Alarms
Internal battery
communication fault Internal battery defective.
572 Contact your specialist dealer.
Device defective.
Battery E1
communication fault External battery 1 defective.
573 Contact your specialist dealer.
Device defective.
Battery E2
communication fault External battery 2 defective.
574 Contact your specialist dealer.
Device defective.
Battery E1 fault
External battery 1 defective.
575 Contact your specialist dealer.
Battery E2 fault
External battery 2 defective.
576 Contact your specialist dealer.
Internal battery
temperature fault Operate device at an ambient temperature
577 Ambient temperature too high.
of 5 °C to 40 °C.
Battery E1 temperature
fault Operate device at an ambient temperature
578 Ambient temperature too high.
of 5 °C to 40 °C.
Battery E2 temperature
fault Operate device at an ambient temperature
579 Ambient temperature too high.
of 5 °C to 40 °C.
Energy outage
580 Power supply outage. Use alternative ventilation option.
Pressure permanently
low
755 Mask leakage too high. Check and correct position of mask.
Tidal volume
permanently low
756 Settings implausible. Check therapy and alarm settings.
Exhalation blocked
757 Exhaled air outlet is blocked. Check exhalation valve and exhalation module.
LMT 68651 11/2020
EN | 25
7 Alarms
Pressure measuring tube Valve control tube and pressure Check circuit and tube connections are
and valve control tube measuring tube switched. connected correctly (see 4.2.3, p. 12).
switched 760 Check valve control tube for blockages and
Valve control tube kinked. damage.
If necessary: Replace circuit.
FiO2 sensor fault
Contact your specialist dealer.
770 FiO2 sensor defective.
Replace FiO2 sensor.
No FiO2 cell
Contact your specialist dealer.
771 No FiO2 cell fitted.
Have FiO2 cell fitted.
Blower temperature
Cool device immediately or therapy will end.
high Blower temperature too high. Cooling
789 Check cooling air filter. If necessary: Have
air filter blocked.
cooling air filter replaced by specialist dealer.
SpO2 signal weak 792 SpO2 sensor not connected to the Check connection with the finger. If alarm
finger correctly. persists: Contact your specialist dealer.
Therapy ended
794 Device is switched off. Switch device back on.
|
26 EN
7 Alarms
8 Faults
FAULT CAUSE ACTION
No running noise, no No power supply Check power cable is securely connected.
display on screen. present. Check function of socket.
Clean coarse dust filter. If necessary:
Coarse dust filter soiled.
Replace filter (see 6, p. 17).
Adjust headband so that the mask does not leak
Device does not reach (see instructions for use for the mask).
Breathing mask leaking.
set If necessary: Replace defective mask or patient/ventilator
target pressure. interface.
Check circuit and eliminate leaks.
Circuit leaking.
If necessary: Replace circuit.
Device defective. Contact your specialist dealer.
Dark display does not react
to display being touched. Device is switched off. Switch on device (see 4.4, p. 13).
Display remains dark.
LMT 68651 11/2020
EN | 27
9 Technical specifications
SPECIFICATION DEVICE
Product class to 93/42/EEC IIb
Dimensions W x H x D in cm 30 x 13 x 21
Weight 3.8 kg
Temperature range
- Operation +5 °C to +40 °C
- Transport and storage -25 °C to +70 °C
- Transport and storage at +70 °C Allow to cool to room temperature for 4 hours before
starting up.
- Transport and storage at -25 °C Allow to heat to room temperature for 4 hours before
starting up.
Relative humidity 10 % to 90 %, no condensation
Permitted humidity for operation, transport and storage
> 35 °C to 70 °C at a water vapor pressure up to 50 hPa
700 hPa to 1100 hPa,
Air pressure range corresponds to an altitude of 3,000 m above mean sea level
Connection diameter for breathing tube Standard 22 mm tapered connector to ISO 5356-1
Maximum air flow at 20 hPa > 220 l/min
3 VDC/0.2 A
System interface When connecting the prismaHub device: 24 VDC/0.2 A
USB-C interface
Maximum power output 5 V/1.1 A
No power input
Power consumption on standby without battery charging 230 VAC/0.07 A
Screen brightness 90 % 48 VDC/0.3 A
24 VDC/0.61 A
12 VDC/1.21 A
Nurse call Maximum 60 V/1 A
Power consumption during ventilation without battery
charging
Screen brightness 90 %
Ventilation settings:
Mode: T
Configuration: Adult
Breathing tube with leakage circuit, 15 mm
Additional accessories: Breathing system filter, WilaSilent
exhalation system
IPAP: 40 230 VAC/0.18 A
EPAP: 4 48 VDC/0.81 A
24 VDC/1.61 A
F: 26.5 12 VDC/2.86 A
Ti: 1.1
Pressure increase: 1
Pressure reduction: 1
Test lung bag
Electrical connection, device maximum
48 VDC/2.7 A
24 VDC/5.4 A
12 VDC/7.0 A
Power supply unit 100 – 240 VAC/2.1 A
Input voltage/maximum current 50 – 60 Hz
Input frequency 48 VDC/2.7 A
Output voltage/maximum current Tolerance -20 % + 10 %
LMT 68651 11/2020
|
28 EN
SPECIFICATION DEVICE
Internal/external battery
Type Li-ion
Nominal capacity 3200 mAh
29.3 V
Nominal voltage 93.7 Wh
Energy 500 charging cycles
Typical discharge cycles Battery capacity is reduced when the device is operated at low
temperatures.
Operating hours of internal battery assuming following
settings: Double circuit, PCV mode, f=20/min, Ti=1 s,
PEEP=off, Vt=800 ml 6 hours
Passive lung: Resistance R = 5 hPa/(l/s);
Compliance C = 50 ml/hPa
Duration of complete battery charge < 6 hours
Duration of 80 % battery charge < 5 hours
Classification to IEC 60601-1-11:
Class of protection against electric shock Protection class II
Degree of protection against electric shock Type BF
Protection against harmful ingress of solids and water IP22
Classification to IEC 60601-1:
Continuous duty
Operating mode
Patient/ventilator interface (e.g. mask, endotracheal tube,
Applied part tracheal cannula), breathing tube, breathing system filter,
SpO2 sensor
Medical electrical devices may only be installed and
Electromagnetic compatibility (EMC) to EN 60601-1-2
commissioned in a defined electromagnetic environment with
regard to emission and radio interference immunity.
More information, including test parameters and limit values,
can be obtained from the manufacturer if required.
EN 55011 B
Radio interference immunity
IEC 61000-4 Parts 2 to 6, Part 11, Part 8
IEC 61000-3 Parts 2 and 3
Heating of respiratory air Maximum +3 °C
Mean sound pressure level/operation to ISO 80601-2-72
at 500 ml 38.5 dB(A) ±3 dB(A), sound power level 46.5 dB(A) ±3 dB(A)
at 150 ml 37 dB(A) ±3 dB(A), sound power level 45 dB(A) ±3 dB(A)
at 30 ml 41 dB(A) ±3 dB(A), sound power level 49 dB(A) ±3 dB(A)
Level 1
Low priority: 59 dB(A)
Medium priority: 63 dB(A)
High priority: 68 dB(A)
Sound pressure level of acoustic alarm to IEC 60601-1-8 ±3 dB(A)
for all alarm conditions (high, medium, low priority)
Tolerance Level 4
Low priority: 78 dB(A)
Medium priority: 83 dB(A)
High priority: 87 dB(A)
±3 dB(A)
4 hPa – 50 hPa
IPAP pressure range Most disadvantageous circuit for leakage ventilation:
Breathing tube WM 29988, bacteria filter WM 27591
4 hPa – 60 hPa
Most disadvantageous circuit for valve ventilation:
Breathing tube LMT31383, bacteria filter WM27591
EN | 29
SPECIFICATION DEVICE
EPAP pressure range 4 hPa - 25 hPa
Most disadvantageous circuit for leakage ventilation:
Breathing tube WM 29988, bacteria filter WM 27591
0 hPa - 25 hPa
PEEP pressure range Most disadvantageous circuit for valve ventilation:
Breathing tube LMT31383, bacteria filter WM 27591
Accuracy of airway pressure ±(2 hPa + 4 % of the set value)
±(2 cmH2O + 4 % of the set value)
CPAP operating pressure range 4 hPa to 20 hPa
Most disadvantageous circuit for leakage ventilation:
Breathing tube WM 29988, bacteria filter WM 27591
±(2 hPa + 4 % of the set value)
Tolerance ±(2 cmH2O + 4 % of the set value)
Pressure increment 0.2 hPa
Maximum pressure in the event of a fault < 90 hPa
Respiratory frequency, adult 2 – 60 bpm
Respiratory frequency, child 5 – 80 bpm
Accuracy ± 0.5 bpm
Increment 0.5 bpm
Ti min, Ti max, Ti timed 0.2 s (child)
0.5 s (adult)
auto (Ti timed only)
0.05 s
Accuracy 0.05 s from 0.2 s to 0.8 s
Increment 0.1 s from 0.8 s to 4 s
Target volume/tidal volume/minute volume
(averaged over previous 5 breaths)
30 ml to 400 ml (child)
100 ml to 3,000 ml (adult)
Accuracy
Most disadvantageous circuit < 50 ml: < 50 ml: ±(4 ml + 20 % of current value)
Breathing tube LMT 31383
Most disadvantageous circuit 50 ml: 50 ml: ±(4 ml +15 % of current value)
Breathing tube LMT 31382
5 ml from 30 ml to 100 ml
Increment 10 ml from 100 ml to 3,000 ml
Trigger level
Inspiration 1 (high sensitivity) to 10 (low sensitivity) (step 1)
Exhalation 95 % to 5 % of maximum flow in 5 % steps
The trigger on inspiration is triggered when patient flow
exceeds the trigger threshold.
Trigger device The trigger on exhalation is triggered when patient flow on
inspiration drops to the percentage value of maximum patient
flow on inspiration.
Speed of pressure increase adult Level 1: 100 hPa/s
Level 2: 80 hPa/s
Level 3: 50 hPa/s
Level 4: 20 hPa/s
Speed of pressure increase child Level 1: 135 hPa/s
Level 2: 100 hPa/s
Level 3: 80 hPa/s
Level 4: 50 hPa/s
Speed of pressure increase MPV mode Level 1: 60 hPa/s
Level 2: 45 hPa/s
Level 3: 30 hPa/s
Level 4: 15 hPa/s
LMT 68651 11/2020
|
30 EN
SPECIFICATION DEVICE
Speed of pressure reduction (in leakage ventilation only)
Level 1: -100 hPa/s
Adult Level 2: -80 hPa/s
Level 3: -50 hPa/s
Level 4: -20 hPa/s
Dead space: 26 ml
Breathing system filter Flow resistance: 2.0 cm H2O
at 60 l/min
EN | 31
The software of this device contains code which is subject to
the GPL. You can obtain the source code and the GPL on
request.
|
32 EN
10 Annex
10 Annex
10.1 Pneumatic diagram
10.1.1 Leakage ventilation
Leakage Circuit
O2 pathway
O2 supply
Patient Interface
inspiration
Pressure sensor
for patient
pressure
exhalation
Spontaneous
ambient air breathing & Optional Optional: Optional:
Air filter Blower Flow sensor non-return Patient circuit Leakage system
O2- sensor bacteria filter humidifier
valve unit
leakage / exhalation
Pressure sensor
for patient
pressure
Pressure sensor
for monitoring
outlet
cooling air
ambient air
Patient Interface
inspiration
Pressure sensor
valve control Pressure sensor
for patient
Control valves pressure
for exhalation
system exhalation
Spontaneous
ambient air breathing & Optional Optional: Optional: Exhalation
Air filter Blower Flow sensor non-return Patient circuit
O2- sensor bacteria filter humidifier system
valve unit
ambient air
Pressure sensor
for patient
pressure
Pressure sensor
for monitoring
outlet
cooling air
LMT 68651 11/2020
EN | 33
10 Annex
Exhalation
Flow sensor Patient circuit Patient Interface
system
exhalation
inspiration
Pressure sensor
valve control Pressure sensor
for patient
Control valves pressure
for exhalation
system
Spontaneous
ambient air breathing & Optional Optional: Optional:
Air filter Blower Flow sensor Patient circuit
non-return O2- sensor bacteria filter humidifier
valve unit
Pressure sensor
for patient
pressure
Pressure sensor
for monitoring
outlet
cooling air
|
34 EN
10 Annex
EN | 35
10 Annex
Output
Fragile. Do not throw or drop
|
36 EN
10 Annex
EN | 37
10 Annex
10.9 Warranty
Löwenstein Medical Technology gives the purchaser of
a new original product and of a spare part fitted by
Löwenstein Medical Technology a limited manufacturer
warranty in accordance with the warranty conditions
applicable to the product in question and in accordance with
the warranty periods from date of purchase listed below.
The warranty conditions are available on the manufacturer’s
website. We will also send you the warranty conditions on
request.
In the event of a claim under warranty, contact your
specialist dealer.
PRODUCT
WARRANTY
PERIODS
Devices including accessories
2 years
(Exception: Masks)
Masks including accessories, batteries
(unless quoted differently in the
6 months
technical documentation), sensors,
circuits
Disposable products None
|
38 EN
LMT 68651 11/2020 EN
Löwenstein Medical
Technology GmbH + Co. KG
Kronsaalsweg 40
22525 Hamburg, Germany
T: +49 40 54702-0
F: +49 40 54702-461
www.loewensteinmedical.de