Librevna Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 53

LibreVNA User Manual

October 21, 2022

Contents

1 Overview 3
1.1 USB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2 External Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 RF ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.4 LEDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Reference Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.6 Reference Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 Getting started 4
2.1 GUI element types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Common GUI elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2.1 Reference Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.2 Device Log Dock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.3 File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.4 Device Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2.5 Window Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3 Vector Network Analyzer 7


3.1 Signal processing architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Sweep Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Acquisition Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4.1 Trace Data Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.4.2 Math operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5.2 Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.3 Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.4 Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5.5 Enabling a calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5.6 Saving/Loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5.7 Assigning a default calibration to a device . . . . . . . . . . . . . . . . . . . 24
3.5.8 Viewing error terms/calibration measurements . . . . . . . . . . . . . . . . 24
3.5.9 Electronic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6 De-embedding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.1 Port Extension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.6.2 Two Thru . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6.3 Matching Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.4 Impedance Renormalization . . . . . . . . . . . . . . . . . . . . . . . . . . 29

1
4 Signal Generator 30

5 Spectrum Analyzer 30
5.1 Acquisition Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2 Tracking Generator Toolbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

6 Graphs 31
6.1 General graph settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.2 Creating and positioning graphs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
6.3 Adding and removing traces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.4 Smithchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.5 XY-plot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
6.5.1 Limit lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6.6 Waterfall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.7 Polar Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

7 Markers 40
7.1 Creating and deleting markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.2 Marker Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.1 Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.2 Maximum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.3 Minimum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.4 Delta . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.2.5 Peak Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.2.6 Lowpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.2.7 Highpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.2.8 Bandpass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.2.9 TOI/IP3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
7.2.10 Phase noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.2.11 P1dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
7.3 Marker Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7.4 Linking markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.1 Creating a new linked group . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.4.2 Adding a marker to an existing linked group . . . . . . . . . . . . . . . . . 49
7.4.3 Removing a marker from a linked group . . . . . . . . . . . . . . . . . . . 49

8 Compound Device 49
8.1 Creating a compound device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.2 Connecting to a compound device . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.3 Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

9 Amplitude Calibration 51
9.1 Source Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
9.2 Receiver Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

10 Troubleshooting 53

2
1 Overview
Port 1
LEDs
Ready
Debug
Port1
Port2
Ext. Reference
Source unlocked
LO unlocked
FPGA unlocked
Booting
Power

External Power
(optional)

USB

Port 2 Reference
Output

Reference
Input

1.1 USB
The LibreVNA uses a USB-C connector as the power supply and for data transmission. The hardware
supports the USB power delivery standard1 , requires 5 V and draws up to 1.2 A of current.
The implemented USB device is limited to USB 2.0 Fullspeed and data transmission will work
with any USB 2.0 or 3.0 port (no USB-C required). However, the current consumption exceeds the
specifications for USB 2.0 and 3.0 ports and the LibreVNA may be unable to fully boot when no
external power is applied and the USB host limits the current.

1.2 External Power


Input jack for external DC power (5 V, 1.5 A, positive center pin). If a power source is connected,
no current is drawn from the USB port anymore.

1.3 RF ports
RF connector to connect the DUT. Both ports are identical and the device architecture support full
S-Parameter measurements (S11, S21, S12 and S22).
1
Work in progress, the device will not negotiate USB-PD yet

3
4
! The RF ports have no input protection, do not apply a signal with more than +10 dBm of power.

1.4 LEDs
Ten LEDs indicate the status:

Ready (green) Indicates fully booted LibreVNA, ready to take measurements

Debug (green) Not used at the moment

Port 1 (green) Active output signal on port 1

Port 2 (green) Active output signal on port 2

Ext. Reference (green) External reference in use

Source unlocked (red) The source PLL failed to lock (it may turn on briefly during a sweep, this
is not a problem)

LO unlocked (red) The 1.LO PLL failed to lock (it may turn on briefly during a sweep, this is not
a problem)

FPGA unlocked (red) The FPGA clock PLL failed to lock

Booting (red) Indicates boot process, also used for error codes

Power (green) USB power valid

1.5 Reference Output


Optional reference output port (user selectable 10 MHz or 100 MHz). The output power is approxi-
mately +3 dBm.

1.6 Reference Input


High impedance reference input port. When applying a 10 MHz input signal, it can be used to
overwrite the internal TCXO.

i Due to the high impedance input, the external input may detect a valid signal when picking up
external radiation. If this is a problem, either terminate the reference input connector or disable
automatic reference switching.

2 Getting started
1. Connect the LibreVNA to a computer using the USB port. After a couple of seconds, only the
"Power" and "Ready" LEDs should stay on. If the "Ready" LED does not turn on or the "Booting"
LED indicates an error code, there is a problem und the device will not function properly. See
section 10 for possible solutions.

4
2. Start the PC application to connect to the LibreVNA. The window should look similar to this:

2.1 GUI element types


The LibreVNA application is made up of six areas containing different types of control:
• The Trace Display Area shows the measured data in various diagrams.
• The Toolbars provide access to commonly used device settings. They can be rearranged or
disabled (see section 2.2.5).
• The Menu at the top left handles less common settings or more complex actions (often with
additional popup windows).
• The Docks mostly provide access to display related settings such as traces and markers. They can
be rearranged or disabled (see section 2.2.5).
• The Statusbar at the bottom show some basic informations about the connected device. If error
conditions (e.g. ADC overload) are present, they are indicated at the bottom right corner.
• The Modeswitch at the top right allows changing the LibreVNA operating mode. Although it
is primarily designed as a vector network analyzer, the hardware architecture allow some other
operating modes with certain limitations.
Depending on the current operating mode, the available toolbars, docks and menu entries change. It
is also possible to rearrange the docks and toolbars.

2.2 Common GUI elements


Certain control elements are always available, regardless of the selected mode. They control settings
of the LibreVNA that are not specific to any of the operating modes.

2.2.0.1 Entering values

Most entry fields support SI prefixes. To set a new value, type in the number followed by an SI
prefix (e.g. to set 3.2 GHz press 3 , . , 2 , + G ). The accepted prefixes vary depending on
the input (e.g. pico is a valid prefix for a capacitance while it will not be accepted for a frequency).

5
2.2.1 Reference Toolbar
The reference toolbar controls the external reference output as well as the input.

The reference can be set to "Int" (internal TCXO), "Ext" (external 10 MHz signal) or to automatic
mode. In automatic mode, the reference will switch to the external input when a signal is detected,
otherwise it will use the internal TCXO.
The external reference output may either be disabled or set to 10 MHz or 100 MHz.

2.2.2 Device Log Dock


This dock shows the debug output of the LibreVNA (the same messages as on the internal UART
port). It may help to identify problems, otherwise it is not required for operation.

2.2.3 File Menu


Changing the default setup of the application can require a lot of configuration, especially if some of
the more complicated math or de-embedding options are used. To simplify this process, setup-files
can be saved and opened, which perform these steps automatically. A setup file contains all settings
and configuration of these elements:
• Sweep and Acquisition settings (start/stop frequency, IF bandwidth, ...)
• The currently active mode (VNA, Signalgenerator or Spectrumanalyzer)
• Traces (Name, Color, Parameters, ...)
• Math operations applied to the traces
• Markers
• De-embedding options
• Graph setup and enabled traces

4
! If any part of the setup references another file (i.e. a trace imported from a touchstone file), only
the filename is saved. Opening a setup file after moving/deleting any of the referenced files will
result in missing data.

2.2.4 Device Menu


The device menu controls the connection state, allows low level hardware access and to perform
firmware updates:
• Update Device List: Scans the USB port for connected devices. Clicking this action simply
updates the list of devices, no connection is established.
• Connect to: Select which LibreVNA to connect to. Different devices are identified by their serial
number.
• Disconnect: Disconnect from the LibreVNA.

6
• Manual Control: Starts the manual control dialog in which all hardware settings are accessible
(mostly used for testing purposes):

• Firmware Update: Load a new microcontroller firmware and FPGA configuration into the
LibreVNA. Do not disconnect power while updating the firmware!

2.2.5 Window Menu


The window menu allows hiding not needed toolbars and docks. It also contains some application
preferences.

3 Vector Network Analyzer


In this operating mode, the LibreVNA takes S-parameter measurements. A source signal is generated
and alternately applied to the RF ports. The incoming signal at both RF ports is measured, resulting

7
in the four S-parameters S11 and S21 (when the source signal is routed to port 1) as well as S12 and
S22 (when it is routed to port 2).

3.1 Signal processing architecture


This section contains some details of the signal flow inside the LibreVNA-Application. Knowledge of
the signal flow is not required to operate the LibreVNA but may help to understand certain features
and limitations.

VNA Hardware

uncalibrated S parameter set (S11,S12,S21,S22) over USB

Calibration

calibrated S parameter set (S11,S12,S21,S22)

De-embedding

individual S parameters

Traces
imported files CSV/Touchstone export
Math

complex frequency/time domain data

Graphs

The central element of the signal flow is the concept of traces. A trace is created from a single S
parameter across frequency (either measured by the LibreVNA or imported from an external file).
Once the data is captured in a trace, optional math operations may be applied to it (e.g. performing a
TDR to transform the data into the time domain) and the result of these operations are passed on to
the graphs.
When capturing data from the LibreVNA, additional operations take place before the data is fed
to the traces. First, the LibreVNA measures the four S parameters between its two ports. The raw
measurements are transferred through USB. The first processing block is the calibration (see also
section 3.5), followed by optional de-embedding procedures. After the de-embedding, the four S
parameters are split into individual S parameters and the relevant traces are updated.
This concept has several consequences when using the LibreVNA, some of which are:
• You can always take calibration measurements. No matter how the de-embedding or math
operations in the traces are set up, they will never influence the calibration routine
• A valid calibration is required before using de-embedding options (otherwise, most de-embedding
options won’t be able to work properly)
• Applying a calibration or de-embedding to traces imported from files is not possible
• Once the data arrives in the traces, the S parameters are no longer associated (e.g. there is no
connection between S11 and S21 anymore. It is possible to "pause" the trace capturing S11 while
S21 continues to get updated)

8
3.2 Sweep Toolbar
This toolbar sets the swept frequency range.

The start/stop and center frequency as well as the span can be set directly.
Additionally, the sweep toolbar contains buttons for zooming in/out around the center frequency
and a preset to set the sweep to the full frequency range.

3.3 Acquisition Toolbar

• Level: The amount of power used for stimulus generation. The dynamic range decreases when
using smaller values. It is recommended to use the highest available settings when measuring
passive networks. When measuring active devices (e.g. amplifiers), decrease the stimulus power in
such a way that the input power into any port does not rise above −10 dBm to stay within the
linear range of the LibreVNA.
• Points: Amount of measurement points in one sweep. More points provide finer frequency
resolution but also increase sweep time.
• IF BW: Bandwidth of final IF measurement. Low bandwidths increase the sweep time but improve
the noisefloor. At higher frequencies (roughly above 3 GHz), the dynamic range is limited by
the isolation between the ports and decreasing the IF bandwidth does not improve the noisefloor
anymore.
• Averaging: Moving average across multiple sweeps.

3.4 Traces
Adding/configuring traces is handled by the Trace Dock (by default on the left side):

The list shows all available traces, by default the four S parameters S11, S12, S21 and S22. Different
icons in front of the trace names indicate the state of the trace. Clicking on one of the icons toggles
the state:
• / : Enable/disable global visibility: Shows/hides the trace an all graphs it is enabled on.
• / : Play/pause. Only available when the trace is fed from live VNA data, not when it was created
from a touchstone file. Pausing a trace freezes the current data, new measurements from the
LibreVNA will not update the trace.
• M/ M: Enable/disable math. Only available if at least one math operation is present.
Creating a new trace or double clicking on an existing trace opens the "edit trace" window:

9
On the left side, additional trace parameters can be changed, while the right side of the window
handles math operations that will be applied to the trace data. The available trace parameters are:
• Name
• Color
• Velocity factor (only relevant when transforming into the time domain)
• Data source (Live Capture/From File/From Math)

3.4.1 Trace Data Sources


Every trace can receive its data from one of three source:
• Live Capture: Sampled data from the LibreVNA are added to the trace in real time
• From File: Import static data from an external touchstone or CSV file
• From Math: Combine data from other traces with math operations to create a new trace

3.4.1.1 Live Capture

The trace is constantly updated by the data received from a connected and sweeping LibreVNA.
Available settings are the S parameter from which the data is updated and a simple Max Hold and
Min Hold function (based on signal magnitude).

10
3.4.1.2 From File

The trace data is extracted from a specified file. If the file contains data for multiple traces, the correct
parameter must be specified as well.

4
! When saving a setup with traces created from files, the actual trace data is not saved. Instead,
the path to the source file is added to the setup file. Opening the created setup file will result in
empty traces if the original source file is no longer present.

3.4.1.3 From Math

i This section is about creating a trace based on data from other traces and combining them with
math operations. If you are looking to change or transform the trace data itself without using
any data from other traces, see section 3.4.2.

11
The trace is created from other traces and user-defined math operations. To create a trace from math,
perform the following steps:
1. Select "From Math" as the trace source
2. Check all traces on which the trace should depend. Not all traces can be used, there are the
following limitations:
• All used traces must have the same output domain
• There can be no circular dependencies between "From Math" traces
3. (Optional) Change the variable name for the selected traces
4. Enter the formula using the chosen variables

3.4.2 Math operations

i This section is about transforming the already acquired trace data. If you are looking to create a
trace by combining data from other traces and math operations, see section 3.4.1.3.

The "math operations" section on the right contains the additional calculations that are performed
on the trace data before it is displayed. Initially, it contains only one line (representing the measured
data). New operations can be created using the buttons on the right. Each math operation adds a
new line to the list, consisting of three columns:
• Status: Visual indication whether the calculation succeeded. On warnings or errors, a mouse-over
text gives a hint about the problem.
• Description: Short description of the configured math operation. Double clicking will open an
(operation dependent) editing dialog providing more details.
• Output Domain: Domain (Time or Frequency) of the data after this math operation. Certain
math operations can only be applied to time domain data, others only to frequency domain data.
The output domain at the last line decides in which domain the trace is displayed on the graphs.
Populated with several math operations, the window could look like this:

In this example, time domain gating is performed:


• A median filter is applied to the data, smoothing the trace a little bit (not actually necessary for the
time domain gating)
• Next, a TDR transformation creates the time domain response of the measurement
• The impulse response is truncated at and above 1 m with a custom expression
• At last the time domain response is transformed back into the frequency domain with a DFT

12
3.4.2.1 Median Filter

The median filter filters values across frequency (or power or time, depending on acquisition settings
and previous math operations). For every point in a trace it takes the point and several surrounding
points, sorts them and uses the middle point of the sorted list as the filtered value. It is a great option
to get rid of fixed spikes (e.g. due to LO feedthrough or unwanted external signals). In comparison,
the averaging function from the toolbar will smooth captured traces over time but is not able to
remove static spikes.
Adjustable settings:
• Kernel size: Number of sample points that are used for the sorted list
• Sorting method: As most input values are complex, there are different options to sort the values:
– Absolute Value
– Phase
– Real part
– Imaginary part

3.4.2.2 TDR

The TDR takes frequency domain data and transforms it into time domain data. The number of
output samples as well as the time span is determined by the number of input samples and the selected

13
frequency span.
Adjustable settings:
• Mode: A TDR can be performed in two different modes:
– Lowpass: Input samples should start at near DC, i.e. the start frequency must be small compared
to the span. A step response can be calculated as the DC point can be extrapolated.
– Bandpass: Input samples may use any frequency and span. Only an impulse response can be
calculated as the DC point is unknown.
• Compute Step Response: Enable calculation of step response as well (only available in lowpass
mode)
• DC point: Chose between extrapolating the DC point from frequency data or specifying it
manually (only available in lowpass mode)
• Window: A window is applied before performing the transformation. Available windows are:
– Rectangular
– Gaussion
– Hann
– Hamming
– Blackman

3.4.2.3 DFT

The DFT is the inverse operation to the TDR and is used to transform time domain data back into
the frequency domain.
Adjustable settings:
• DC frequency: Depending on the original source data of the input time domain data, it may have
been calculated from a frequency range that does not start at DC. Transforming the time domain
data back into the frequency domain will result in frequency data at and around DC. This setting
allows to move the transformed frequency data to the correct frequency.

14
– Automatic: The DC frequency is determined automatically by looking at the last TDR operation
that resulted in the time domain data and extracting the correct frequency from that. Usually,
this results in the correct frequencies.
– Automatic: If for any reason a different frequency range must be used, the actual frequency of
the DC bin can be specified manually.
• Window: A window is applied before performing the transformation. Available windows are:
– Rectangular
– Gaussion
– Hann
– Hamming
– Blackman

3.4.2.4 Custom Expression

The custom expression allows user-specific calculations on the trace data. It can only use a single
data point as the input value. Access to other data points (e.g. at a lower/higher frequency or from
an earlier sweep) is not possible. Custom expressions can work with input data of any domain and
result in output data of the same domain. Depending on the input data domain, different variables
are available and shown in the dialog.
If the expression contains syntax errors, check the mouse-over text in the status column of the
trace edit dialog.

15
3.4.2.5 Time Gate

The time gate allows filtering data in the time domain. It can either be used as a bandpass (only
passing time domain data within a specified window) or as a bandstop (blocking time domain data
in a specified window). The displayed graph allows for easier adjustments. The green trace is the
time domain input data and is updated live. The red line is the currently selected gate filter.
The gate filter can be adjusted by manually specifying the start/stop or center/span times. As an
alternative, the filter edges can also be moved on the graph with the mouse. Adjustable settings:
• Window: The window function influences the filters pass- and stopband. The filter is constructed
by first calculating the ideal filter coefficients in the frequency domain. Afterward, the coefficients
are windowed to limit spectral leakage and finally transformed into the time domain (which is
displayed in the graph).
– Rectangular
– Gaussion
– Hann
– Hamming
– Blackman

3.5 Calibration
This section is about the VNA calibration which is used to remove the effect of connectors and cables
(as well as imperfections from the LibreVNA itself ). For the amplitude calibration see section 9.
Some general knowledge about the different calibration types is assumed, this manual only
explains how to use them with the LibreVNA.

3.5.1 Introduction
This section contains some implementation details of the calibration. It can be skipped but under-
standing them can be helpful when using the calibration.
Section 3.5 uses the following definitions:
• Calibration standard: A single calibration standard (e.g. an Open standard). Most standards can
either be defined by coefficients or a measurement file.
• Calibration kit: It contains the definition of the calibration standards. When the calibration error
terms are calculated, these definitions are taken into account.
• Measurement: Raw (uncorrected) measurement of a calibration standard. Depending on the type
of desired calibration, different measurements are required.
• Calibration type: The calibration type determines which errors are corrected by the calibration
and how the error terms are derived from the measurements.
The calibration type is followed by the ports for which the calibration corrects errors. For example,
when using a SOLT calibration on ports 1 and 2, the calibration type will be shown as "SOLT_12".

16
For the most accurate results, the calibration should be performed with the same settings (span,
number of points, stimulus level, ...) as the intended measurement. However, it is possible to change
these settings after the calibration. The calibration will be interpolated when a different span is used.

3.5.2 Types
Several different calibration types are available:

Type Applied corrections Required


measurements
Open and Short for each port
SOLT Full SOLT calibration, removes influences Load or SlidingLoad for each port
at all ports and also corrects transmission Through between all of ports
measurements Isolation(optional)
Through-
Normaliza- Only corrects for losses and phaseshift in Through between all of ports
tion through measurements. While this is not
as accurate as SOLT, it is useful e.g. when
measuring only the frequency response of
a filter because it requires far less calibra-
tion measurements.
Reflection for each port
TRL Full two-port calibration but with differ-
Through between all of ports
ent (less accurately defined) standards than
Line between all of ports
SOLT

3.5.3 Calibration Kit


The calibration kit contains the calibration standard definitions used in the calibration. It can be
edited through the menu: Calibration Edit Calibration Kit The same dialog can also be reached through
the calibration measurement dialog: Calibration Calibration Measurements Edit Calibration Kit
The calibration kit can consist of an arbitrary number of calibration standards.

17
The three text fields (Manufacturer, Serial number and Description) are meant for easy identifi-
cation of the correct calibration kit. None of the contents are used anywhere by the GUI and the
user is free to enter any text.
Calibration standards can be added, deleted and sorted with the buttons to the right of the
calibration standard list. Double-click a standard to edit its definition.

3.5.3.1 Open Standard

The Open standard can either be defined by its coefficients or by a touchstone measurement file. If
defined by a touchstone file, the file limits the span of the calibration for which the standard can be
used.
Coefficients:
• Z0: Impedance of the transmission line from the reference plane to the open termination
• Offset delay: One-way length of the transmission line
• Offset loss: Loss/attenuation of the transmission line
• C0–C3: Parasitic capacitance of the open termination

18
3.5.3.2 Short Standard

The Short standard can either be defined by its coefficients or by a touchstone measurement file. If
defined by a touchstone file, the file limits the span of the calibration for which the standard can be
used.
Coefficients:
• Z0: Impedance of the transmission line from the reference plane to the short termination
• Offset delay: One-way length of the transmission line
• Offset loss: Loss/attenuation of the transmission line
• L0–L3: Parasitic inductance of the short termination

19
3.5.3.3 Load Standard

The Load standard can either be defined by its coefficients or by a touchstone measurement file. If
defined by a touchstone file, the file limits the span of the calibration for which the standard can be
used.
Coefficients:
• Resistance: Resistance of the termination resistor
• Z0: Impedance of the transmission line from the reference plane to the termination resistor
• Offset delay: One-way length of the transmission line
• Offset loss: Loss/attenuation of the transmission line
• Parallel C: Parasitic capacitance in parallel to the termination resistor
• Series L: Parasitic inductance in series to the termination resistor
• Load Parameter Model: Two different Load models are implemented, with the order of the
parasitic elements swapped:
– Series L first:
Z0 Series L

P arallel C Resistance

– Shunt C first:

20
Z0 Series L

P arallel C Resistance

3.5.3.4 Reflect Standard

The Reflect standard doesn’t have any coefficients as it is an unknown standard. The only available
choice is whether it is assumed to be a short or an open.

3.5.3.5 Through Standard

The Through standard can either be defined by its coefficients or by a touchstone measurement file.
If defined by a touchstone file, the file limits the span of the calibration for which the standard can be
used.
Coefficients:
• Z0: Characterisitc impedance of the Through standard (fixed to 50 Ω for now)
• Delay: One-way length of the transmission line
• Loss: Loss/attenuation of the transmission line

21
3.5.3.6 Line Standard

The Line standard is an unknown standard and is not accurately defined by coefficients. The only
parameter is its delay to determine the usable span when using this standard in a TLR calibration.
The delay mustn’t be specified with high accuracy either.

3.5.4 Measurements

The calibration measurements can be found in Calibration Calibration Measurements . Initially, the list
is empty. Measurements can be added and removed with the buttons on the right of the list. For
convenience, the required measurements for common calibrations can also be created automatically
(select the desired calibration type from the drop-down list on the top).

Each measurement has a few settings, depending on the measurement type:

3.5.4.1 Open Measurement

• Select the calibration standard to be used with this measurement.


• Select the port

22
3.5.4.2 Short Measurement

• Select the calibration standard to be used with this measurement.


• Select the port

3.5.4.3 Load Measurement

• Select the calibration standard to be used with this measurement.


• Select the port

3.5.4.4 Sliding Load Measurement

• No calibration standard is required


• Select the port
• At least three Sliding Load measurements must be used per port, each with a different setting on
the sliding load. More measurements may result in more accurate results and there is no upper
limit on measurements

3.5.4.5 Reflect Measurement

• Any open, short or reflect standard can be selected


• Select the port

3.5.4.6 Through Measurement

• Select the calibration standard


• Select the ports
• When the Through standard is definded by a touchstone file and the port order in that file does
not match the port order in the measurement, check the "Reversed" checkbox. Through standards
defined by coefficients are always symmetrical and the checkbox has no effect

3.5.4.7 Line Measurement

• Select the calibration standard


• Select the ports
• As Line standards are always symmetrical, the checkbox has no effect

3.5.4.8 Isolation Measurement

• No settings available

23
3.5.4.9 Taking measurements

To take a measurement, select the corresponding measurement and press Measure . Multiple measure-
ments may be taken at the same time if they use different ports.

i All measurements required by a calibration type should have the same start and stop frequencies
as well as number of points. Do not change the span between measurements.

3.5.5 Enabling a calibration


After all required measurements have been taken, the calibration can be enabled. There are two
different options available for this:
• Select the calibration type in the calibration toolbar and enable the checkbox. If all required mea-
surements have already been taken, the calibration is activated immediately. If some measurements
are missing or have been taken with different span settings, applying the calibration is not possible.
In that case, the calibration measurement dialog (see section 3.5.4) opens instead.
• Select the calibration type in the calibration measurement dialog and click "Activate". Only
calibration types for which all required measurements are available can be activated.

3.5.6 Saving/Loading

Once a calibration is active, it can be saved by selecting Calibration Save . All raw measurements are
saved together with the calibration type and the calibration kit.
Similarly, a saved calibration can be opened by selecting Calibration Load . It is applied immediately
after opening.

3.5.7 Assigning a default calibration to a device


Once a calibration has been saved, it can be selected as the default calibration for a specific device.
To do so, first connect to the device and then select Device Default Calibration Assign... . Once a default
calibration has been assigned, it will automatically be opened and applied every time the application
connects to that specific device.

3.5.8 Viewing error terms/calibration measurements


The error terms calculated from the calibration measurements as well as the raw measurements
used to derive these error terms can be imported by selecting Calibration Import error terms as traces or
Calibration Import measurements as traces . This feature is mostly intended to debug calibration problems
and is not normally required.

3.5.9 Electronic Calibration


Performing a calibration, especially when multiple ports are used, can require a lot of manual
steps. The LibreVNA supports an automatic electronic calibration, which takes all the required
measurements for a SOLT calibration on its own. To use the electronic calibration, the LibreCAL2
device is required.
To start the electronic calibration, select Calibration Electronic Calibration or Calibration Calibration Measurements
Electronic Calibration . There is no need to create a calibration kit first, as the electronic calibration will
use the calibration coefficients stored in the LibreCAL.

2
https://github.com/jankae/LibreCAL

24
4
! The electronic calibration will delete and replace any currently active calibration and also the
calibration kit standards. Please store any unsaved calibration data before starting the electronic
calibration.

Steps required to perform the electronic calibration:


1. Connect the LibreCAL via USB to the same machine on which the LibreVNA-GUI is running
2. Connect every port of the LibreVNA that should be calibrated to one of the ports on the LibreCAL
3. Start the electronic calibration dialog
4. Select the correct LibreCAL device and the coefficients:
• Device: Serialnumber of the LibreCAL. Usually, only one should be connected and the
serialnumber will be automatically selected when the dialog is first opened
• Coefficients: Select the coefficient set to be used. The LibreCAL can store multiple sets of
coefficients. Please the manual3 of the LibreCAL for detailed information on coefficient sets.
5. Check and adjust the port assignments. Ports between the LibreVNA and LibreCAL can be
connected in any order. Match the ports in the dialog to your physical setup.
6. Press the "Start" button. This will perform the following steps:
• Fill the calibration kit based on the coefficients extracted from the LibreCAL
• Take an open, short and load measurement for every port of the LibreVNA
• Take a through measurement for any combination of ports
• Apply the SOLT calibration. For the FACTORY coefficient set, the reference plane will be at
the ports of the LibreCAL´
3
https://github.com/jankae/LibreCAL/blob/main/Documentation/manual.pdf

25
3.6 De-embedding
The de-embedding options are available under Tools De-embedding . The GUI works similar to the
math operations for traces (see section 3.4.2) but the de-embedding is performed before the data
reaches the traces (compare with section 3.1).

Just like with the math operations, a list shows the de-embedding options. Once a new option
is added (or is double clicked) a window opens which allows to configure the option. Depending
on the de-embedding option, there might be some user adjustable parameters (e.g. the length and
attenuation of a port extension) or the option needs a measurement from the LibreVNA to calculate
the de-embedding parameters (e.g. take a measurement to automatically measure the port extension
length). If a measurement is required at any point, a new window opens with two choices to provide
the measurement data:

The first possibility is to click the measure button (with the LibreVNA connected) which starts a new
measurement. The LibreVNA connections must have been setup properly before (e.g. in the case of
the port extension: extension cable connected and terminated either into an open or short. The exact

26
setup depends on the de-embedding option). The other possibilty is to provide the measurement by
using traces. This is useful if the same measurement has already been taken but the physical setup has
changed since then. Select the correct traces for all required S parameters and click the "OK" button.

3.6.1 Port Extension

The port extension allows moving the reference plane relative to the plane used during the calibration.
The port extension is defined by two characteristics:
• Distance/Time: The one-way distance or time the reference plane will be moved. Setting either
parameter will automatically calculate the other by using the velocity factor.
• Loss: The additional loss introduced by the physical component that caused the reference plane to
move. It is further split into two parameters:
– at DC: One-way loss at 0 Hz
– at a specific frequency: One-way loss at a chosen frequency, typically higher than the DC loss
All these parameters can also be calculated automatically by measuring the port extension with
either an open or short connected instead of the DUT.
If a port extension is required at multiple ports, add this option once for each port.

27
3.6.2 Two Thru

The two thru option is an advanced de-embedding option for when there are additional fixtures
between the reference plane and the DUT that can’t be easily calibrated out by either a SOLT
calibration or a port extension. The are a few prerequisites for using this option:
• The LibreVNA must be calibrated at a reference place which does not include the fixtures
• The fixtures (everything between the reference plane and the DUT) must be identical for both
ports
• It must be possible to connect the fixtures directly to each other as well as directly to the DUT

To create a two thru de-embedding, perform the following steps:


1. Select the two ports for which the de-embedding will be applied in the configuration section of
the dialog
2. Connect both fixtures directly to each other (without the DUT)
3. Take the mandatory 2xThru measurement
4. If the fixtures have a different characteristic impedance than 50 Ω:
• Insert the DUT between the fixtures
• Take the optional Fixture-DUT-Fixture measurement
5. Compensate the effect of the fixtures by clicking the "Calculate" button in the dialog

28
3.6.3 Matching Network

The matching network allows the user to (de-)embed lumped components and arbitrary transmission
lines between the DUT and a measurement port.
Available components are:
• Series R/L/C: Lumped component in series between the port and the DUT
• Parallel R/L/C: Lumped component from the signal path to GND
• Defined Through: Any two-port network as defined by a touchstone file
Any combination of these components can be dragged into the signal path between the port of
theLibreVNA and the DUT.
Further settings:
• Operation: Chose between embedding and de-embedding
• Port: Port of the LibreVNA
If a matching network is required at multiple ports, add this option once for each port.

3.6.4 Impedance Renormalization

The LibreVNA is a 50 Ω system. S parameter measurements are reported with a 50 Ω reference


impedance. With the impedance renormalization, the acquired data can be transformed into what
it would look like if the system would have a different reference impedance (e.g. 75 Ω). This
transformation influences all traces captured by the LibreVNA.

29
4 Signal Generator
In the signal generator mode, measurements are stopped and the LibreVNA only outputs a CW
signal.

The hardware is not optimized for signal generation which means that the output level is not very
accurate, especially at higher frequencies (also see section 9). Also, the signal will contain higher
levels of harmonics than one would normally expect from a signal generator.

5 Spectrum Analyzer
Although the LibreVNA hardware is not designed to be used as a spectrum analyzer, the general
hardware architecture of a spectrum analyzer is similar enough to that of a VNA to implement basic
two-channel spectrum measurements. This is in no way on the same performance level as a dedicated
spectrum analyzer but for simple measurements it might suffice if no other equipment is available.
The main differences to a real spectrum analyzer are:
• No input attenuator or pre-amplifier: This means that the measurement range is essentially
fixed to approximately −110 dBm to −10 dBm.
• No amplitude calibration: The displayed signal level is not very accurate, especially at higher
frequencies (also see section 9).
• No image rejection filters: This is probably the most severe limitation, because it means that for
every real signal several other signals will show up in the spectrum that are not actually present at
the input. Most of the images can be filtered out by signal identification but some will remain (at
lower amplitude) in the reported data.
• Highest resolution bandwidth is quite low: The sweep speed is too slow to cover the complete
frequency range of 1 MHz to 6 GHz in an acceptable time.
Most control elements (Frequencies, Span, Traces, Marker, ...) are indentical to the vector
network analyzer mode.

5.1 Acquisition Toolbar

• RBW: Resolution bandwidth. Lower values allow differentiating between signals at closer fre-
quencies. Lower values also result in a reduced noisefloor but significantly increase sweep time.
For low RBWs, a faster DFT acquisition method is available. Instead of tuning the LOs to look at
each frequency point by itself, several points can be measured at once by performing a DFT of the
final IF. Due to limitations in the hardware and the available processing power, the DFT is limited
to 96 points and can only cover up to 30 kHz at a time. Whether the DFT is used (and at which
RBWs) can be selected in Window Preferences
• Window: Window type used in the DFT of the final IF. If in doubt, use Flat Top.
• Detector: For every displayed point, several measurements are taken. The detector type determines
which one of these measurement will be displayed:

30
Detector Displayed data
+Peak Highest amplitude
-Peak Lowest amplitude
Sample Only uses the middle measurement
Normal Highest amplitude on odd points, lowest amplitude on even points
Average Average all measurements for each point

• Signal ID: Signal identification. This can help to determine whether a displayed signal is actually
present or the result from internal imaging. When enabled, the LibreVNA changes the LO
frequencies for every measurement point and observes how the final IF signal is affected by that.
This removes almost all of the mirror images but at the cost of increased sweep time.
The following example shows the effect of signal ID. For both measurements the only signal at the
input was a 1 GHz tone with a level of −10 dBm. On the left, signal ID is turned off, resulting in a
lot of extra tones. On the right, signal ID has removed most of these tones:

5.2 Tracking Generator Toolbar

A tracking generator at either port is available in the spectum analyzer. If the tracking generator
is active at a port, measurements from that port are still available but meaningless. The tracking
generator also includes an optional frequency offset. To get accurate results, a Source/Receiver
Calibration (section 9) should have been performed when using the tracking generator.
DFT acquisition is not available when the tracking generator is active. Also, due to hardware
limitations, the tracking generator is unable to reach every frequency exactly. For narrow spans
this could result in "drops" in the spectrum where the signal of the tracking generator is outside of
the RBW filters passband. The frequency resolution is frequency dependent. A warning message
appears if this could be a problem with the selected span and stop frequency.

6 Graphs
In VNA and spectrum analyzer mode, the trace data in displayed on graphs. Several different graph
types are available and most can be customized as well.

6.1 General graph settings


Common settings for the behavior and appearance of all graphs can be found in Window Preferences
Graphs :

31
Most of the settings are related to graph colors and sizes, allowing adjustments to personal preferences
and different monitor resolutions.
Each graph type also has some specific settings. They can be reached by right-clicking on an
existing graph and selecting Setup... or Axis Setup... .

6.2 Creating and positioning graphs


Graphs use an underlying tile system. Each tile can either:
• be empty
• contain one graph
• be split horizontally into two new empty tiles
• be split vertically into two new empty tiles
For split tiles, the position of the split can be adjusted manually by using the cursor.
To create new graphs, an empty tile must be available. If all tiles are showing graphs (the default
when starting the GUI), an empty one must be created first. This can be done by right-clicking a
graph and either selecting close to remove it from its tile or selecting Add tile... and chosing in which
direction the new tile should be added.
From the empty tile, all graph types are available:

32
i For a quick enlarged view of a graph, double-click it. The graph will take up the whole tile
area. Another double-click returns the graph to its original size.

6.3 Adding and removing traces


When a new graph is created, it is empty by default and traces must be added to show up. There are
two ways to add traces:
• Drag-and-drop traces from the trace list onto the graph
• Right-click a graph and select the trace to add. Traces can also be removed be deselecting them in
the context menu

6.4 Smithchart

The smithchart is the default graph for reflection measurements (S11, S22,...). Adding through
measurements (S12, S21,...) is not possible.
There are a few smithchart specific settings in the setup dialog:

33
The left side of the dialog contains settings on how data is displayed:
• Display mode (Frequency) There are three options on how much of a trace is visible in the
frequency range:
– Limit to current span Shows all traces only in the frequency range of the currently selected
span, even when the trace contains data outside of the span
– Show complete traces Always shows the complete trace, regardless of the selected span
– Manual frequency range Allows selection of an abitrary frequency range to display
• Display mode (Impedance) There are two options on how much of a trace impedance is visible:
– Limit to visible are Clips the trace data at the edge of the smithchart circle
– Show complete traces Always shows the complete trace, even when the trace data falls outside
of the smithchart circle
• Zoom Allows to change the center position and zoom factor of the smithchart. Zooming is also
possible by using the mousewheel on the graph
The right side of the dialog allows the user to add lines of constant values to the smithchart for a
quick visual comparisation of trace data to these values. Four types of lines are available:
• Constant VSWR
• Constant resistance
• Constant reactance
• Constant Q
For each line, the color and constant value must be manually selected.

34
6.5 XY-plot

The XY-plot is the most versatile graph. It can display every trace type if configured correctly. The
XY-plot consists of up to three axes:
• X-axis Always present
• Primary Y-axis Optional, on the left side of the graph
• Secondary Y-axis Optional, on the right side of the graph
Different types can be used for the X-axis and the selected type must match the output domain of all
displayed traces. Available types are:

X-axis type Supported trace domains Description


Frequency Frequency S-parameter data plotted across frequency
S-parameter data plotted across stimulus
Power Power
power
S-parameter data plotted across time at a
Time (Zerospan) Time (Zerospan)
fixed frequency
Time-domain data (transformed
Time Time
S-parameter data)
Time-domain data (transformed
Distance Time S-parameter data), converted to distances
according to the traces velocity factor

Different types are also available for the Y-axes. The Y-axis type determines how the trace data
is plotted on the graph. The underlying trace data is not affected by the Y-axis type, the graph itself
performs the required calculations to display the correct data. Available types are:

35
Y-axis type Description
Disabled No Y-axis displayed
Magnitude Magnitude in dB (or dBm in spectrum analyzer mode)
Magnitude (dBuV) Magnitude in dBuV
Magnitude (linear) Magnitude of the complex S-parameter
Phase Phase in degrees, confined to ±180°
Unwrapped phase Unwrapped phase in degrees
VSWR Standing wave ratio
Real Real part of the complex S-parameter
Imaginary Imaginary part of the complex S-paramter
Series Resistance Real part of the impedance from a reflection measurement
Reactance Imaginary part of the impedance from a reflection measurement
Capacitance Extracted capacitance from a reflection measurement
Inductance Extracted inductance from a reflection measurement
Quality Facotr Quality Factor of the impedance from a reflection measurement
Group Delay Group Delay of a transmission measurement
Impulse Response (real) Real part of the impulse response from a TDR
Impulse Response (magnitude) Magnitude of the impulse response from a TDR
Step Response Step response from a TDR
Impedance Impedance as calculated from the step response from a TDR

Not every Y-axis type can be combined with every X-axis type. The possible combinations are:
In VNA mode:

X-axis Y-axis
Magnitude
Magnitude (linear)
Phase
Unwrapped Phase
VSWR
Frequency Real
Power Imaginary
Time (Zerospan) Series Resistance
Reactance
Capacitance
Inductance
Quality Factor
Group Delay
Magnitude
Time Impulse Response (real)
Distance Impulse Response (Magnitude)
Step Response
Impedance

In spectrum analyzer mode:

X-axis Y-axis
Magnitude
Frequency
Magnitude (dBuV)

The axes types can be selected in the setup dialog for the XY-plot:

36
For every axis, the displayed range can either be determined automatically or manually. Additionally,
the X-axis can be displayed either in linear or logarithmic mode. Logarithmic mode for the Y-axes
is not supported yet.

6.5.1 Limit lines


The XY-plot also supports custom limit lines:

Limit lines can be added and removed in the bottom part of the Axis Setup Dialog. Each line is a list
of manually defined points, a selectable color and can be configured either as a high limit, low limit
or just as a constant line without any limit checking:

37
6.6 Waterfall

The waterfall graph can display the amplitude trend of a single trace over multiple sweeps. Its specific
settings are mostly a subset of the XY-plot, the Y-axis configuration is almost identical.

Other settings include:

38
• Direction: Determines whether the waterfall graph is filled from the top or bottom
• Pixels per line: Provides a simple vertical zoom feature
• Mode: Determines whether sweep data that doesn’t fit onto the graph anymore is kept or deleted
(a resize of the graph would reveal kept data)
• Number of sweeps: The maximum number of sweeps kept if mode is set to "Keep offscreen data"
• Alignment: The waterfall can be aligned to a XY-plot of equal size and peaks in the trace data of
both plots will line up.

4
! Both plots are still completely independent of each other. For the alignment to work, they
must have the same width and one must be placed exactly below the other.

6.7 Polar Chart

The Polar Chart looks similar to the smithchart but doesn’t perform the transformation from S-
paramter to impedance. Furthermore, through measurements can be displayed as well. The available
settings are identical to the smithchart but the Polar Chart does not support adding custom constant
lines:

39
7 Markers
Markers provide an easy read-out of trace data at specific points. Each marker is assigned to one
trace and will show up on any graph that show the trace at the marker position.
Some general marker settings are available in the preferences:

• Show data on graphs: Marker data (X-coordinate, trace value) will be shown by default to the
right of any graph the marker is visible on
• Show data in all available formats: All available formats for the marker data will be shown by
default
• Positioning: When moving markers by hand, the can either snap to the individual trace points
or be interpolated along the plotted trace
• Sort order on graphs: Defines the order of marker data to the right of a graph
• Symbol style: Various symbol styles for drawing markers on the graphs are available

40
Example of a marker with all data formats shown:

The marker dock provides a quick overview of all markers:

• Marker #: The marker number as shown on the graphs


• / : Enable/disable global visibility: Shows/hides the marker an all graphs
• : Indicates linked markers. Markers with the same number are linked and all of them move when
one changes its position
• Trace: The trace name the markers is assigned to
• Type: Various marker types are available, see section 7.2
• Settings: Depends on the marker type, allows configuration of the marker
• Restrict: Forces the marker to a specific frequency range (or power/time range if the trace uses
another domain)
• Data: The trace data at the marker position. Can be displayed in different formats
Various other settings are also available in the context menu. The context menu can be reached
by right-clicking the marker, either in the marker dock or on the marker symbol in any graph.

7.1 Creating and deleting markers


There are three ways to create a marker:
• Right-click a trace on a graph and select "Add marker here"
• Use the "Add marker" button in the marker dock
• Use the "Add markers to all traces" button in the marker dock to create a marker for each trace.
The created markers are linked by default (see section 7.4)
Markers can be deleted by:
• Selecting "delete" in the context menu

41
• Selecting the marker in the marker dock and either pressing DEL or using the "Delete marker"
button

7.2 Marker Types


Every marker is of a specific type. The type determines how the marker position is calculated and
influences the available marker data formats. Some marker types also include helper markers for
showing additional trace points. The marker type can be changed in the marker dock or in the
context menu.

7.2.1 Manual

This is the default marker type. It can be positioned at any position by the user and will never move
on its own.

7.2.2 Maximum

The marker snaps to the maximum amplitude of the assigned trace (within the limits of the "restrict"
column).

7.2.3 Minimum

The marker snaps to the minimum amplitude of the assigned trace (within the limits of the "restrict"
column).

7.2.4 Delta

The delta marker can be positioned by the user. It will show the trace data relative to its assigned
reference marker. The reference marker is selected by its marker number in the "type" column.

42
7.2.5 Peak Table

The peak table lists all peaks in the signal above a specified signal level. It will create a helper marker
for each peak. The peak threshold level can be adjusted in the "settings" column.

7.2.6 Lowpass

43
The lowpass marker determines the cutoff frequency of a lowpass filter. At first the filter attenuation is
calculated by finding the maximum signal amplitude. Afterward, the cutoff frequency is determined
by finding the frequency at which the signal level has dropped by a specified amount. The threshold
for this signal level drop can be adjusted in the "settings" column.
This marker type is only available for through measurements.

7.2.7 Highpass

The highpass marker determines the cutoff frequency of a highpass filter. At first the filter attenuation
is calculated by finding the maximum signal amplitude. Afterward, the cutoff frequency is determined
by finding the frequency at which the signal level has dropped by a specified amount. The threshold
for this signal level drop can be adjusted in the "settings" column.
This marker type is only available for through measurements.

7.2.8 Bandpass

44
The bandpass marker determines the center frequency and bandwidth of a highpass filter. At first
the filter attenuation is calculated by finding the maximum signal amplitude. Afterward, the cutoff
frequency in each direction is determined by finding the frequency at which the signal level has
dropped by a specified amount. The threshold for this signal level drop can be adjusted in the
"settings" column. Finally, the center frequency is set to the middle of the higher and lower cutoff
frequency.
This marker type is only available for through measurements.

7.2.9 TOI/IP3

This marker type calculated the third-order intercept point. It is only available in spectrum analyzer
mode. Initially a peak search is executed to find the two highest peaks. The frequencies of the
intermodulation products are calculated from these peak frequencies and helper markers placed on
each of these frequencies. Finally, the third-order intercept point as well as the tone and distortion
signal amplitudes are calculated.

45
7.2.10 Phase noise

This marker type calculates the phase noise of a signal. Is is only available in spectrum analyzer mode.
Initially a peak search is executed to determine the signal frequency. Afterwards, a helper marker
is placed at a specified offset. The offset can be configured in the "settings" column. By using the
signal amplitudes from the peak and the offset marker, the phase noise is calculated.

7.2.11 P1dB

This marker type is only available for through measurements on power sweeps. It calculates the 1dB
compression point of amplifiers.

46
7.3 Marker Data
The trace data at the marker position can be displayed in the marker dock and on the graphs in
various formats. The available formats depend on the marker type as well as the domain of the trace
data. Only one of the available formats can be displayed in the marker dock at a time. On graphs,
any amount of formats can be displayed at once. The shown formats can be selected in the context
menu of the marker.

47
Trace domain Marker type Available data formats
dB
Manual
Time Real/Imaginary
Delta
Impedance (if step response available)
Manual dBm
Delta dBuV
Time (zero span)
Maximum Noise
in SA mode
Minimum
Peak Table
dB
dB + angle
Manual Real/Imaginary
Delta Impedance

Time (zero span)

Maximum VSWR



in VNA mode


Minimum Series Resistance Only for reflection

Peak Table Capacitance 


 measurements
Inductance




Quality factor

Manual dBm
Delta dBuV
Maximum Noise
Minimum
Frequency Peak Table
in SA mode Phase Noise
Phase Noise
dB
third-order intercept
TOI/IP3 Average tone level
Average modulation product level
dB
dB + angle
Manual Real/Imaginary
Delta Impedance


Maximum VSWR





Minimum Series Resistance Only for reflection

Frequency
Peak Table Capacitance  measurements
in VNA mode

Inductance




Quality factor

Center and bandwidth


Bandpass
Insertion loss
Lowpass Cutoff frequency
Highpass Insertion loss
1dB compression point (only for P1dB type)
dB
dB + angle
Manual
Real/Imaginary
Delta
Impedance

Power Maximum

VSWR



Minimum


Series Resistance Only for reflection

P1dB
Capacitance 
 measurements
Inductance




Quality factor

48
7.4 Linking markers
Normally, markers can be moved individually and are only connected to one trace. This creates a
problem when reading out trace data from multiple traces at the same position. If the position is
changed, all markers need to be moved manually to the new position. Linked markers provide a
solution to this problem: All markers within a linked group always use the same position. If one
marker is moved, the others move as well. The linked group of each marker is indicated in the
marker dock (empty if the marker does not belong to any linked group.
Only markers that are movable can be added to linked groups. If a marker type performs
automatic positioning of the marker (e.g. the maximum marker type), it can not be added.

7.4.1 Creating a new linked group


Select multiple markers in the marker dock, right-click and select "link selected".

7.4.2 Adding a marker to an existing linked group


Open the markers context menu, chose "Add to linked group" and select the linked group the marker
should be added to.

7.4.3 Removing a marker from a linked group


• Open the markers context menu, chose "Remove from linked group"
• Select multiple linked markers in the marker dock, right-click and select "Break links" to remove
all selected markers from their linked groups.

8 Compound Device
The LibreVNA supports combining multiple hardware units to a "Compound Device". When used,
the configured physical LibreVNAs are combined into a virtual device with more ports. When
connected to such a device, additional measurement paramaters are available (e.g. measuring S23 in
VNA mode or port 3 in spectrumanalyzer mode).

4
! When connected to a compound device and in VNA mode, there is no phase information for
through measurements between different physical devices. This is a hardware limitation and
the phase is set to zero in the software. Through measurements within a physical device and all
reflection measurements retain their phase information.

8.1 Creating a compound device


Compound devices must be configured in the preferences: Window Preferences Compound Devices

49
Create and remove compound devices with the buttons on the right. Edit an existing compound
device by double-clicking it:

Required steps when creating a compound device:


1. Assign a name to the new compound device
2. Select the synchronization method between devices. At the moment, only USB synchronization
is supported but future hardware versions might support faster synchronization via dedicated
trigger ports
3. Drag-and-drop a LibreVNA symbol into the configuration area for every physical device in the
compound device
• At least two physical devices must be used

50
• At most four physical devices can be combined with a maximum of eight virtual ports
4. Assign the serialnumbers for each physical device. Serialnumbers for all currently connected
devices are available as suggestions but it is also possible to enter a different serialnumber manually
5. Assign the virtual ports of the compound device:
• Each port number must appear only once
• Port numbers must start with port 1
• Port numbers must be consecutive, e.g. port 1,2 and 4 is not allowed because port 3 is missing
• Physical ports may be left unused

8.2 Connecting to a compound device


Configured compound devices appear in the device list with their name when all required physical
devices are connected. Once connected, new measurements or ports are available depending on the
number of configured ports in the compound device.

i Through measurements between physical devices depend on precise matching of the stimulus
frequency. Depending on the accuracy of the internal frequency source and the selected IF
bandwidth, the stimulus signal from the generating device may fall outside of the IF bandwidth
of the receiving device, resulting in a reported lower amplitude than actually present. Either
align both internal oscillators precisely with the frequency calibration or (recommended) switch
to using the external reference input when using a compound device.

8.3 Limitations
Certain features are not available when connected to a compound device. To use them, disconnect
from the compound device and connect directly to the individual physical devices:
• Firmware update
• Manual control
• Source calibration
• Receiver calibration
• Frequency calibration

9 Amplitude Calibration
This section is about calibrating the source output level and spectrum analyzer level. It does not affect
VNA measurements at all. For the VNA calibration, see section 3.5.
Both the internal signal source and the receiver paths used in the generator or spectrum analyzer
mode have no means of ensuring accurate amplitudes (as imperfections are removed by the VNA
calibration, the hardware is not optimized for absolute amplitude accuracy). In order to improve
the accuracy of both the generator and the spectrum analyzer (especially with enabled tracking
generator), frequency dependent amplitude calibrations are available. They are permanently stored
in the device and are only required to be performed once. The Source Calibration corrects the
ouput of the (tracking) generator, while the Receiver Calibration corrects the measurement of the
spectrum analyzer.
For both calibrations, other test equipment is required, ideally covering the whole range of 1 MHz
to 6 GHz (a calibration of only the partial frequeny range is also possible but signals/measurements
outside the calibrated range will be inaccurate). The Soure Calibration requires a way to measure
the output of the LibreVNA (e.g. a spectrum analyzer or power meter), the Receiver Calibration
requires a signal source (e.g. a signal generator). Once either calibration has been completed, the
LibreVNA can also infer the other one with reasonable accuracy.

51
9.1 Source Calibration
Select Device Source Calibration . A window showing the current source calibration opens:

These are the necessary steps:


1. Set up the calibration frequencies: Add or remove points according to your needs. Ideally,
they should be evenly spaced across the range to calibrate, maybe with a slight emphasis on higher
frequencies (signal levels generally vary more at higher frequencies). Additionally, it is advisable
to set one point at 25 MHz and one just below (at 25 MHz a band is switched, resulting in an
amplitude discontinuity). Up to 64 points can be used.
2. Select the calibration mode: Either calibrate both ports separately (best accuracy but more
work) or assume that they are identical and use the measurement at one port for both.
3. Perform the calibration: For each frequency and port, select the corresponding cell in the table.
The LibreVNA will generate a signal at the selected frequency and port with (what it thinks are)
−20 dBm. Use external equipment to determine the actual output level and enter the measured
value into the selected cell.
4. Save the calibration: Once all measurements have been taken, the "Save to Device" button
becomes available. There is no way to save intermediate results, you have to calibrate all frequencies
you have chosen in step 1.
If the Receiver Calibration is already done, it is also possible to infer the Source Calibration from
that. Press Automatic measurement and connect both ports of the LibreVNA directly to each other.
This connection should be as ideal as possible (short, low loss cable) as every imperfection directly
translates into calibration errors. During the automatic calibration, the LibreVNA uses the already
established frequency points from the Receiver Calibration and measures the output amplitude by

52
itself to determine the coefficients of the Source Calibration.

9.2 Receiver Calibration


Select Device Receiver Calibration . The Receiver Calibration works almost identically to the Source
Calibration. The only difference is in step 3:
3. Perform the calibration: For each frequency and port, select the corresponding cell in the table
and apply a signal with known level at the selected frequency and port. The signal level should
be in the range of −40 dBm to −10 dBm. Higher signal levels might already compress in the
receiver while lower levels unnecessarily increase calibration errors due to noise. The LibreVNA
tunes its receiver to the selected frequency and measures the amplitude. Apply the signal for at
least 2 s and enter its actual amplitude into the selected cell while the signal is still active.

10 Troubleshooting
If there was an error in the boot process, the "Booting" LED will blink an error code repeatedly:

Error
Meaning Possible solution
Code
1 Failed to detect FLASH memory Hardware failure, check soldering
2 Invalid FLASH content Perform firmware update
3 Failed to configure FPGA Check FPGA bitstream generator settings
Verify that FPGA bitstream and uC
4 Hardware initialization failed
firmware are created from the same release

53

You might also like