Manual Equipo Smartor
Manual Equipo Smartor
Manual Equipo Smartor
Contents
Preface… ............................................................................................................................ VII
Operation Cautions and Safety Guide ............................................................................. IX
Chapter 1 Basic Information ...................................................................................... 1
1.1 Appearance............................................................................................................ 1-1
1.2 Keys ....................................................................................................................... 1-3
1.3 Power Supply......................................................................................................... 1-4
1.3.2 Using the Battery ........................................................................................ 1-5
1.3.3 Charging the Battery................................................................................... 1-5
1.3.4 Battery Maintenance and Safety ................................................................ 1-6
1.3.5 Adapter for Power Supply ........................................................................... 1-7
1.3.6 Power Supply Mark .................................................................................... 1-8
1.4 Power On/ Off ........................................................................................................ 1-8
1.4.1 Power On.................................................................................................... 1-8
1.4.2 Power Off .................................................................................................... 1-8
1.5 Status Symbol ........................................................................................................ 1-9
1.6 Display Layout ....................................................................................................... 1-9
1.7 How to Assemble and Use System Straps .......................................................... 1-12
1.8 How to Use Back Strap........................................................................................ 1-16
Chapter 2 Conventional Ultrasound (UT)............................................................... 2-1
2.1 Menu Structure ...................................................................................................... 2-1
2.2 Menu Operation ..................................................................................................... 2-6
2.2.1 Main Menu Operation ................................................................................. 2-6
2.2.2 Submenu Operation ................................................................................... 2-6
2.2.3 Page Operation........................................................................................... 2-8
2.2.4 Menu adjustment ........................................................................................ 2-8
2.2.5 Custom function keys ................................................................................. 2-9
2.3 Functional Description ........................................................................................... 2-9
2.3.1 Detect ......................................................................................................... 2-9
2.3.2 Weld Wizard ............................................................................................. 2-17
2.3.3 Forging Wizard. ........................................................................................ 2-17
2.3.4 Plate Wizard ............................................................................................. 2-17
2.3.5 CrackMeas................................................................................................ 2-17
2.3.6 FFT ........................................................................................................... 2-17
2.3.7 B-Scan ...................................................................................................... 2-18
2.3.8 AWS .......................................................................................................... 2-20
2.3.9 CSC (Curved Surface Correction) ............................................................ 2-21
2.3.10 API 5UE .................................................................................................. 2-21
2.3.11 Manage ................................................................................................... 2-21
2.3.12 Setup ...................................................................................................... 2-23
2.4 Fast Adjustment ................................................................................................... 2-23
I
Contents
II
Contents
III
Contents
DCY2.781.Smartor/V1.0/F-E
IV
Preface
Preface
The Smartor digital ultrasonic flaw detector is designed for ultrasound testing of surface
and internal defects of metal and non-metallic materials in industrial areas, as well as and
b) Big Screen: 5.7" TFT high brightness LCD, with resolution 640×480 pixels, display
e) T/R front end - adjustable pulse width negative square emission, amplitude up to
f) UT functions: weld, plate and forging wizards; AGC (auto gain control), Peak memory,
DAC, AVG and video recording; Optional functions such as B-scan, TCG,
probe spectral analysis, CSC (curved surface correction), weld simulation, crack
compensation available.
VII
Preface
Tips:
a) The system screenshots in this manual are generated by effective software version,
b) Some of the features described in the manual are optional, and your system may not
have these features. If you have any questions, please contact the distributor.
c) To know system features and operation, as well as how to perform ultrasonic testing
with the system, please read this Operation Manual carefully before using the system.
Contact Information:
Website: http://www.siui.com
E-mail: [email protected]
VIII
Operation Cautions and Safety Guide
a) Please keep this Operation Manual and always operate the system by
b) Before operating the system, the operator shall have received formal training
on ultrasonic testing.
c) Do not try to dismantle and repair the system unless specifically described in
d) Do not place the system in the place of moisture or near radiator, strong
e) When the system is not in use for a long time, it should be powered up at
h) For cautions on using the battery, see 1.3.4 Battery Maintenance and Safety.
i) For cautions on using the adapter, see 1.3.5 Adapter Power Supply.
IX
Basic Information
1.1 Appearance
1-1
Basic Information
1-2
Basic Information
1 (R) Receive probe port (R) Ultrasound Receive probe port (R)
1.2 Keys
type.
1-3
Basic Information
functions.
The power of this system can be either supplied by the battery or the adapter.
When installing or removing the battery, loosen the screw for the battery compartment at
the system back, as shown in Fig. 1-4. Open the battery compartment, and then you may
1-4
Basic Information
If the battery is connected to the system, in power-on state, the upper left corner of the
【Note 1】: When battery power is very low, to avoid battery over-discharge, the
【Note 2】: When the battery is connected with the system, even if the system is
powered off, there is also a little current consumption. It is suggested to take out
the battery when the system is not in use for a long time.
Charging the battery should be in a dry and ventilated place under normal temperature.
Operation:
2) Connect the adapter to the power supply. Now the adapter indicator is green
1-5
Basic Information
3) Insert the adapter plug into the DC power socket on the system. The system
power indicator indicates the current power / charging status: Red for
charging (power-off state), Off for full charge (power-off state), or Green for
【Tip】: Battery recharge can be done either when the system is powered on or off.
If the system is powered on, the recharge time might be longer. Charging the
battery while operating the system for a long time may result in system temperature
a) The capacity and service life of the rechargeable battery depend on whether the
battery is used properly. Please recharge the battery with the configured adapter.
1) When charging the battery through the adapter, use stable and reliable mains
2) The system battery can only be recharged with the configured adapter and the
3) Please recharge the system battery in a cool and ventilated room, otherwise the
4) If the system is not in use for a long time, the battery should be discharged and
1-6
Basic Information
5) During storage or transportation, the battery should maintain about half of its
capacity.
6) When system battery comes to the end of its service life, in order to avoid
environmental pollution, do not discard, break or crush the battery at will. Please
standards.
toxic chemicals.
8) Do not charge the battery near fire source or in extremely hot environment.
10) Keep the battery from moisture and never throw it into water.
11) Do not store the battery at the temperature higher than 60℃ or lower than -20℃.
12) Do not use the battery if it has peculiar smelling, heat, distortion, changed color
The adapter may supply power for the system or recharge the battery. After connecting
the adapter to AC power supply, if the adapter indicator is green with no load (not
connected to the main unit for power supply or when recharged). Connect the adapter
output plug to the system adapter input socket for proper use. The adapter indicator turns
red if it is loaded.
【Note 1】: Do not use the adapter on a loosen-plug AC socket. Be noted that the
total load of adapter plugging into the socket shall not exceed the rated load.
【Note 2】:Do not try to open the adapter if the adapter makes abnormal sound, or
In power up state, the Power Supply mark is displayed at the upper left of the screen. For
indication)
1.4.1 Power On
Operation:
2) Long press Power key to boot up the system. The power/charging indicator
on the system panel is green, and the system enters work mode after
self-test.
Long press Power for about 1 second, and the system will enter the shutdown state and
turn off.
【Note】If any problem occurs during system power down, see Chapter 7.
1-8
Basic Information
In addition to the power supply symbol, the status identification information bar at the top
For the select buttons on the top-level desktop of UT (Ultrasonic Testing) and TG
1-9
Basic Information
Take the conventional ultrasound mode as an example, and the screen layouts of
The main menu screen display of conventional ultrasound (UT) is as shown in Fig. 1-6.
The menu screen display of Ultrasound Testing (UT) in the system is as shown in Fig.1-7.
3 System info: For displaying info such as battery, adapter status or system
5 Image area and scale: For displaying ultrasound image data and the
corresponding scales. (The x-axis is the A-scan sound path scale, and y-axis
1-11
Basic Information
1.7.1 How to assemble system straps (example for left- handed operation)
(1) Preparation: Place the straps on the left of the system, see Fig. 1-10. The targets
are two braided straps tied to the stainless-steel cylinders of the system corners.
(2) Reverse the two straps (relative position unchanged) 180 degrees, and keep the
1-12
Basic Information
(3) Pass the upper braided strap through the buckle, tighten it; do it in the same way
1-13
Basic Information
1.7.2 How to use system straps (example for left- handed operation)
(1) Hold the left of the system with your hand going through the straps. See Fig. 1-14
(2) Pass the belt through the buckle, see Fig. 1-15
1-14
Basic Information
(3) Tighten the belt and fasten it backward (Velcro), see Fig. 1-16
1-15
Basic Information
1-16
Conventional Ultrasound (UT)
【Note】: For the menus with the mark * in the menu structure, press Enter key
Main Detect
Menu
Submenu
MeasPnt -- EchoType -- --
Alarm -- -- -- --
2-1
Conventional Ultrasound (UT)
Main Detect
Menu
Magnify -- -- -- --
MeasLine -- -- -- --
Main Detect
Menu
-- -- -- Line1 --
-- -- -- Line2 --
-- -- -- Line3 --
-- -- -- Line4 --
-- -- -- Line5 --
-- -- -- Line6 --
Main Detect
Menu
1st
AboutAVG >> WeldSimu BEA -- --
Submenu
2-2
Conventional Ultrasound (UT)
Main Detect
Menu
Freq -- End -- --
Shape -- LineInd -- --
CrysLA -- -- -- --
2nd
CrysLB -- -- -- --
Submenu
BaseType -- -- -- --
BaseAper -- -- -- --
Atten. -- -- -- --
Mode -- -- -- --
NFL -- -- -- --
K1K2Corr -- -- -- --
Submenu
Delete -- -- Test --
Rename -- -- -- --
2-3
Conventional Ultrasound (UT)
Export -- -- -- --
StoreSet -- -- -- --
1st
System Reading Activate -- --
Submenu
-- Read1 Export -- --
-- Read2 Import -- --
2nd
-- Read3 -- -- --
Submenu
-- Read4 -- -- --
-- LargeR -- -- --
-- RefBW -- -- --
2-4
Conventional Ultrasound (UT)
-- -- -- Width* Rectify
-- -- -- Thresh Reject
-- -- -- MeasPnt EchoType
-- -- -- Alarm --
-- -- -- Magnify --
-- -- -- MeasLine --
Submenu
Frame Step
Gain Param>>
Range --
2nd DspDly --
Submenu AStart --
AWidth --
AThresh --
AMeasP. --
2-5
Conventional Ultrasound (UT)
BSta --
BWidth --
BThresh --
BMeasP. --
Submenu
Gate --
Logic --
2nd Start --
Submenu Width --
Thresh --
MeasLine --
The system main menus are displayed in desktop menus vertically at the center of the
screen.
Operation:
(1) Up, Down, Left, Right for switching the cursor selected main menu
There are two levels of submenus: 1st submenu (horizontal), 2nd submenu (vertical).
When controlling a submenu, the cursor is on the selected 1st submenu, as shown in
2-6
Conventional Ultrasound (UT)
(1) Up, Down, Left, Right: turn pages of 1st submenu, and refresh the 2nd submenu
list.
(3) HOME: Exit the 1st submenu control and return to the desktop main menu
control.
When the 2nd submenu is controlled, the cursor is on the selected 2nd submenu, with the
corresponding 1st submenu marked □, as shown in Fig. 2-2. Now the keys are operated
as follows:
2-7
Conventional Ultrasound (UT)
b) Value type: to toggle the adjust state, such as "Range" when and only when
the menu value has a ^ or - flag (for fine /coarse adjust state).
c) Key type: Go to the next level menu list, for example, "Curve" -> "Adjust".
(4) HOME: Exit the 2nd submenu control and return to the 1st submenu control.
For the main menu that contains multiple pages of submenus, turning pages as follows.
Operation:
Select the "Page" submenu, press Left/ Right or Up/ Down key to turn pages.
Operation:
2-8
Conventional Ultrasound (UT)
2) Press Left/ Right to adjust the value of the 2nd submenu or the options for
【Note1】: When adjusting the value-type submenu, hold the adjust key to
【Note2】: For the submenus with coarse / fine adjustment, press Enter key
There are two custom function keys: F1 and F2, whose functions can be selected in the
The selectable functions for F1 and F2 keys in UT mode are: Freeze, PeakEnv, PeakEcho,
and AGC; the selectable functions in TG mode are: CalibZero, SingleSrch, CntinueSrch,
2.3.1 Detect
a) Basic
◆ FineGain: For fine adjustment of gain. The 4dB fine adjustment range has
40 steps. When the gain is fine tuned, the displayed gain value does not
change.
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
2-9
Conventional Ultrasound (UT)
◆ Gate: for setting the adjusted gate, i.e. GateA , GateB or GateC.
◆ Magnify: For magnifying echo in the gate to the full wave image range.
◆ MeasLine: For turning on / off the indicating line of gate measure result.
b) TR
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
c) WpPrb
2-10
Conventional Ultrasound (UT)
★Tip: "Angle" and "ValK" are associated; ValK =tan (probe angle).
d) WaveSet
◆ Coor: For setting coordinate type of horizontal axis in the image area (A
scan echo).
echo will serve as the background to compare with the dynamic wave.
the secondary wave area will be marked in the image area (A-scan echo).
2-11
Conventional Ultrasound (UT)
◆ FullScreen: enable/ disable the function to enlarge the image area to full
screen.
e) AutoC
◆ Calib: Select the calibration options, including Vel, PrbDly and Vel+PrbDly.
◆ Ref1: The reference position of the 1st record point when calibrating.
◆ Ref2: The reference position of the 2nd record point when calibrating, which
◆ Clear: To clear the calibration information, and restore to the default value
f) AngleC
◆ ReflDep: To set up the depth of the center point for the calibration reference
hole.
2-12
Conventional Ultrasound (UT)
◆ Calib: To calibrate the angle (ValK) based on the peak record point.
◆ Clear: To clear the calibration information and set up the default value.
g) Curve (DAC)
◆ Record: To record the peak record point as the curve record point.
◆ DelPnt: To enable the function of deleting the specified curve record point.
h) CurveCtrl (DAC)
◆ Comp: To adjust the compensation gain of the relative curve. The receiver
affect DAC / TCG curve is disabled, such as TR parameters, Vel and PrbDly.
◆ DACData: If enabled, the table displays the curve record point information.
◆ Crit.: To select the criterion to set up DAC dB value, or use a custom DAC.
◆ Grade: The menu is adjustable when a DAC criterion is selected. You may
◆ Line1: The menu is adjustable when a custom DAC curve is selected, and
◆ Line2: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥2, and you may set up the dB value of Line2.
◆ Line3: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥3, and you may set up the dB value of Line3.
◆ Line4: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥4, and you may set up the dB value of Line4.
◆ Line5: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥5, and you may set up the dB value of Line5.
◆ Line6: The menu is adjustable when a custom DAC curve is selected and
the curve No. ≥6, and you may set up the dB value of Line6.
47013 is selected as DAC criterion. You may select the proper block level for
as DAC criterion. You may select the level for testing the thickness of the
2-14
Conventional Ultrasound (UT)
flaw.
i) Curve (AVG)
◆ Record: To record the peak record point as the curve record point.
j) AboutAVG
◆ CrysLA: To set up the he probe crystal size A. It refers to the valid diameter
PrbShape is Round.
2-15
Conventional Ultrasound (UT)
◆ Mode: To calculate the AVG curve from the position of 3 or 0.1 times near
◆ NFL: To set up Near-Field Length (NFL), which is adjustable only for dual
probes.
k) AVG CurveCtrl
◆ Comp: To adjust the compensation gain of the relative curve. The receiver
echo on the V1 and V2 test blocks. Input the value of the probe AVG curve
directly.
l) WeldSimu
◆ DisE: To set up the distance from probe flank to the weld center.
m) BEA
2-16
Conventional Ultrasound (UT)
Enter the weld test wizard screen, and finish weld test settings step by step.
Enter the forging test wizard screen, and finish forging test settings step by step.
Enter the plate test wizard screen, and finish plate test settings step by step.
2.3.5 CrackMeas
◆ H1: The reflected echo height of the 1st endpoint recorded by the system.
◆ H2: The reflected echo height of the 2nd endpoint recorded by the system.
2.3.6 FFT
2-17
Conventional Ultrasound (UT)
◆ DspDly: For setting the ultrasonic pulse shift (display delay) on screen
display.
2.3.7 B-Scan
a) Ctrl
◆ Save: With the memory connected, to save scan files to the memory, with
◆ Curr: In TimeBase mode, " Curr " refers to the scan time that the probe is
coupled to the workpiece; In Encoder mode, " Curr " refers to the scan
◆ Loss: To calculate the total missed time or distance due to excessive scan
b) Setup
◆ Encoder: To select the encoder model when Encoder mode is selected and
◆ Step: To set up trigger time for scan when TimeBase is selected; To set up
◆ Dis: The menu is adjustable when Encoder is selected. It is for setting the
◆ StPos: The menu is adjustable when Encoder is selected. It is for setting the
c) Basic
d) TR
e) (Analyse) Setup
◆ Step: To set up the frame number for each move of scan line.
f) (Analyse) Meas
◆ Gate: for setting the adjusted gate, i.e. GateA , GateB or GateC.
2.3.8 AWS
value.
measure value.
measure value.
2-20
Conventional Ultrasound (UT)
◆ Start: To set up the start position of GateA, which can only be coarse
◆ Dr: For setting the reference indicator for calibration (the depth of a groove
◆ K: Correction factor.
2.3.11 Manage
a) Storage
2-21
Conventional Ultrasound (UT)
b) Wp
c) Probe
d) Encoder
2.3.12 Setup
a) System
To view and set up the system basic information, such as system S/N, time and
date.
b) Reading
To set up the display content of reading areas such as the Test screen.
c) Activate
2.4.1 Gain
Operation:
(1) Gain To enter the gain fast adjustment state, now the screen bottom displays
(2) Up/ Down To adjust the value of the selected item, where the gain highlighted
2.4.2 Gate
Operation:
2-23
Conventional Ultrasound (UT)
(1) Gate To enter the GAStart/Width fast adjustment state, now the screen bottom
2.5 Meas
2.5.1 Reading
The reading column below the information column shows 5 readings. See Fig. 2-4.
The first reading is for gain, in the format: the upper one is total gain (with the unit in the
In addition, the 4 readings can be selected by the user, where the last reading is a big
reading, which is highlighted in larger font size. See Table 2-2 for the options of reading.
Symbol Meaning
HxdB dB difference between echo measurement point amplitude and gate height
within gate x.
amplitude in the gate and curve point height of the same position on the
2-24
Conventional Ultrasound (UT)
Symbol Meaning
amplitude in the gate and curve point height of the same position on the
Rx Horizontal distance from the probe (wedge) flank to the echo measurement
A Ind (Auxiliary AWS computing) the gain value of the amplitude when recording
B Ref
Sab The sound path distance between measurement points in gate A and B
ERS The FBH equivalent of measuring point echo in gate A in AVG curve state.
2.5.2 MeasPnt
There are 3 methods for gate MeasPoint: Peak, Flank and JFlank, which are for sound
path and amplitude measurement for different modes. See Fig. 2-5.
2-25
Conventional Ultrasound (UT)
a) Peak: To calculate sound path (S3) and amplitude (H2) based on the echo peak
b) Flank: To calculate sound path (S1) based on the first intersection between the
echo and the gate, and to calculate amplitude (H2) based on the peak in the
gate range.
c) JFlank: To calculate sound path (S1) according to the first intersection between
the echo and the gate, and based on which calculate the amplitude (H1) by
searching for the echo peak where the intersection locates in the gate range.
2.5.3 Alarm
Alarm mode includes: off, Pos (the echo amplitude within the gate is higher than gate
thresh) and Neg (the echo amplitude within the gate is lower than gate thresh).
When giving alarm, the alarm symbol at the top of the screen is red and flashes, otherwise
When the system "Buzzer" is on, it supports sound alarm, and the Buzzer sounds when
giving alarm.
2.6 WaveSet
2.6.1 PeakEnv
If enabled, the echo amplitude peak values for all points in sound path direction are
recorded and displayed. When higher echo for one point appears, the recorded value for
Description: It is mainly used in the situations of angle probe testing positioning and
2-26
Conventional Ultrasound (UT)
2.6.2 PeakEcho
If enabled, the screen will record the peak echoes for all echoes (except for initial wave)
within the display range, and display these echoes in different colors. When higher echoes
appear, all echoes recorded will be refreshed, otherwise they remain unchanged.
2.7 B-Scan
B-Scan is to display A-scan echo in imaging mode, so as to achieve more intuitive test
2.7.1 Scan
Operation:
(1) Set up the system TR parameters according to the probe connected to the
system.
(2) Adjust the scan mode and parameters in Setup according to the scan.
(4) During the scanning process, the "Curr" menu is updated to the time /position of
the scan.
(5) When the scan is finished, enable "Save" function (which is disabled before
(6) Press Right or Enter to go to “Analyse”, and to browse the scan file just finished.
【Note 】
(1) Due to the limited storage space, B-Scan supports only 1000 frames of data,
(2) The system will automatically compress/ stretch the scan image to full
2-27
Conventional Ultrasound (UT)
2.7.2 Analyse
Entering Analyse mode and the scan file will be played automatically. Now:
(1) The scan lines (marked in red on the B-Scan image) scroll to show the frame
position.
(2) The waveform image area shows the waveform data information for the frames.
(3) The 2nd menu area displays the information such as "Gain", "Range", "Delay"
In the 1st menu area, there are button menus of "Play / Pause", "Stop" "Forward",
"Backward" and "Finish". Select one and press Enter to perform the operation.
AutoC is for calibrating material velocity and probe delay (probe zero, the acoustic time
from the probe crystal to the incident point) for the system, so as to eliminate the impact of
Vel and PrbDly on the measurement values of sound path, horizontal distance and vertical
distance.
The user may select either material velocity calibration or probe zero calibration, or both
a) VelC: Calculate material velocity by using echo propagation time difference of the
b) DelayC: Calculate probe delay by using the difference between the echo
propagation time of one reflector with known sound path, and the echo time
Example condition:
2-28
Conventional Ultrasound (UT)
(1) The calibration block is as shown in Fig. 2-6, the sound path of the first reflector
Operation:
2-29
Conventional Ultrasound (UT)
2) In “WpPrb”, set up probe “Type” as Angle, “Angle “ is 60.0, now the system will
3) In “AutoC”, set up “Calib” to Vel+PrbDly, and set up Ref1 and Ref2 sound path of
the calibration block according to test block parameters setup (set to 50mm and
100mm here).
4) Place the probe on the calibration block, as shown in Fig. 2-6, and get the echo
When the "Ref1" value is adjusted, the gate center position is automatically adjusted to its
set value. If the parameter setup and probe placement are correct, the gate may record
the desired echo. If the gate position is different from the echo position, adjust "Start"
menu, or press Gate key to quickly adjust (see 2.3.2) and enable recording the echo.
Move the probe to get peak position of the reference point, press Right or Enter in
2-30
Conventional Ultrasound (UT)
【Note】The red circle on the echo indicates the echo peak position, which can be
5) Move the probe to get Ref2 echo, as shown in Fig. 2-10. Refer to step 4) for the
operation, and the system gets Ref2 peak position, press Right or Enter in
2-31
Conventional Ultrasound (UT)
6) The system updates the values of "Vel" and "PrbDly", and displays them in the
upper right corner of the waveform area. √ Indicates calibration for the item is
finished.
7) At the screen bottom, it prompts to input the actual probe flank value according to
2-32
Conventional Ultrasound (UT)
AngleC is for measuring the angle of an angle probe with a reflector (e.g. side drilled hole)
【Note】
(2) When performing angle measurement, the sound path between the probe
and the reflection hole shall be greater than 2 times of NFL, so as to avoid
Example condition:
(1) The calibration block is as shown in Fig. 2-12, ReflDep 30mm, ReflApe 50mm.
2-33
Conventional Ultrasound (UT)
Operation:
1) Finish Vel+PrbDly calibration. See 2.7, or set up the parameters according to the
3) Place the probe on the test block, as shown in Fig. 2-12. The system gets the
When adjusting "ReflDep" value, the gate center position is automatically adjusted to its
set value. If the parameter setting and the probe are placed properly, the gate may record
the desired echo. If the gate position is different from the echo position, adjust "Start"
menu, or use Gate key to quickly adjust (see 2.3.2), enabling the gate to record the echo.
Move the probe to get the peak of the reference point, press Right or Enter in “Calib”
2-34
Conventional Ultrasound (UT)
Note: The red circle on the echo indicates the echo peak position, which can be
4) The system updates the values of "Angle" and "ValK ", and displays them in the
upper right corner of the waveform area as shown in Fig. 2-14 Finish angle
2-35
Conventional Ultrasound (UT)
If you are not satisfied with the calibration result, turn on "Clear" and return to step 1) to
calibrate again.
The Distance Amplitude Correction (DAC) curve is used to indicate the amplitude variation
of the reflector signal with the same size but different distance from the probe. The
purpose of the DAC curve is to compensate material attenuation, near field effect, beam
DAC Curve is for guiding the user to make DAC curves based on the detection
requirement.
Operation:
2-36
Conventional Ultrasound (UT)
2) Adjust “Start” menu, or press Gate key to adjust quickly (see 2.3.2), enabling the
3) Adjust “Gain” menu, or press Gain key to adjust quickly (see 2.3.1), to make the
4) Press Right or Enter key in “Record” menu (or press Enter key in “Start” menu
to record quickly), to finish the setup of one DAC curve record point, “Record”
5) Repeat steps 2)~4) to continuously increase DAC curve record points till the
In this process, if any wrong record point is selected, enable “DelPnt” to delete the
specified record point. See the operation steps of DelPnt below. You can also enable
2-37
Conventional Ultrasound (UT)
If the record point position is correct, but the amplitude is deviated, you can enable
"AdjPnt" function to adjust the recorded record amplitude. See the AdjPnt operation
below.
DelPnt Operation:
(2) The operating tips are shown at the screen. See Fig. 2-16.
【Note】The green circle on the curve (with the numbers above to indicate the
(3) If “DelPnt” is enabled, press Left/ Right key to select the point to be deleted.
Press Enter key to remove the selected point from the DAC curve.
(4) After the deletion is finished, press HOME key to disable "DelPnt" function.
2-38
Conventional Ultrasound (UT)
AdjPnt Operation:
(2) The operating tips are shown at the screen. See Fig. 2-17.
【Note】The green circle on the curve (with the numbers above to indicate the
(3) If ”AdjPnt” is enabled, press Left/ Right key to select the point to be adjusted.
Press Up/ Down key to adjust the amplitude of the selected point.
(4) After the adjustment is finished, press HOME key to disable "AdjPnt" function.
2.11 CurveCtrl_DAC
After making the baseline of DAC curves, you may adjust the curves according to the
standard.
2-39
Conventional Ultrasound (UT)
2.11.1 TCG
After making the curves, the default is DAC. You may adjust "Kind" menu and switch to
TCG.
Time-Changed Gain (TCG) is to compensate the same factor as DAC, but with different
representation. DAC is drawn with reference to the reflector peak position when beams
attenuate, and TCG is to set up a function based on time (distance) amplified gain,
enabling the reference reflector peak reaches the same screen height (the amplitude of
the first DAC point), where the curve is displayed as a straight line through the screen, as
2.11.2 ParamLock
When the DAC Curve is finished, "ParamLock" function is turned on by default. In this
case, the adjustment of the parameter menu (see note (1)) which may cause the
2-40
Conventional Ultrasound (UT)
However, taking into account of probe wear or other situations, and the inspectors may
need to fine tune the restricted parameters, you can turn off the "ParamLock" function and
【Note】
2) To turn off the "ParamLock" function, the inspector should understand the
2.11.3 DACData
When the "DACData" function is turned on, the gain value and the attenuation value of the
recorded points in the DAC curves (BaseL, Line1, Line2 and Line3) at 80% amplitude are
2-41
Conventional Ultrasound (UT)
【Note】
(2) When the DAC criterion is applied, the table will display the selected
criterion and the reflection hole size of the standard test block for the DAC
2.11.4 Crit.
(1) Custom: The built-in criterion is not used. It is up to the user to adjust the number
(2) GB/T 11345: In the standard, the DAC curve requirements the same as GB/T
29712.
Grade2
Thickness T
8≤T<15 15≤T<100
(mm)
Flaw Length
L≤T L>T L≤0.5T 0.5T<L≤T L>T
L
Grade3
Workpiece
(mm)
2-42
Conventional Ultrasound (UT)
Flaw Length
L≤T L>T L≤0.5T 0.5T<L≤T L>T
L
Line1 (dB) 0 -6 4 -2 -6
When the "FlawBlock" menu value is off, this time, the most stringent curve criterion for
the grade and ThkBlock is created (ie, the curve when L> T). When the FlawBlock" menu
value is not off, the curve is created according to the “Grade”, "ThkBlock" and “FlawBlock”.
Line1 (dB) -4 2 6
Line1 (dB) 2 5 10
Line2 (dB) -6 -3 0
Line1 (dB) 3 5 --
Line2 (dB) -7 -5 --
2-43
Conventional Ultrasound (UT)
Line1 (dB) -4 3 6
Select the "ThkBlock" menu value according to the tested thickness. The curve is created
2.12 Curve_AVG
In the actual ultrasonic testing, due to different shapes and nature of defects, it is difficult
to determine the real size of the defect. Therefore using "equivalent method" may quantify
the defects.
The equivalent method is a method of determining the size of the defect, as compared
with the artificial defect of regular shapes, that is, when the perceived defect wave height
is equal to the artificial defect wave height under the same testing condition, the size of the
The artificial defect usually used to represent the equivalent size is FBH (flat bottom hole).
For the regular reflector (including artificial defects), under certain assumptions, the
relationship between the echo sound path, the amplitude (dB) and the defect equivalent
can be expressed in a regular way that is called the AVG (DGS) curve, which is typically
used for the area equivalent calculation of the defect for forging.
Where A represents the distance in the unit of NFL; V represents the amplitude in the unit
2-44
Conventional Ultrasound (UT)
of dB, with the initial wave height 0 dB, and G is the defect equivalent, in the unit of probe
size.
AVG Curve is to guide the user to make AVG curves according to the test requirements.
Example condition:
(1) The test block is a 100-Φ2 forging block, i.e. Thickness 100mm, FBH Φ2.
(2) The probe is a normal (straight-beam) probe with frequency 2.5MHz, diameter
20.0mm.
Operation:
(1) To ensure the accuracy of curve making, auto calibration is required before
(3) Press Right key to enter the “AboutAVG” menu list, and set AboutAVG
parameters according to the probe and test block. In this example, "Type" is
2-45
Conventional Ultrasound (UT)
(4) Place the probe on the calibration block and get the FBH echo. Adjust “Start”
menu, or Gate key to adjust quickly (see 2.5.2), enabling the gate to record the
echo. Move the probe to get the peak of FBH, then press Right or Enter key in
2-46
Conventional Ultrasound (UT)
2.13 CurveCtrl_AVG
2.13.1 Comp
If the echo amplitude of the test block in use or the workpiece does not comply with the
theoretical value, Adjust “Comp” menu to compensate the difference, which is commonly
Since BW echo amplitude usually cannot reach the theoretical difference with the FBH
echo amplitude of the same distance, taking the BW amplitude as the base will result in
incorrect FBH equivalent calculated in practical testing. Therefore, in the situation that BW
echo is used as the base, to make the calculation more accurate, it is suggested to use
the error between the actual difference [of base BW echo amplitude and the echo
amplitude of FBH test block of the same distance (different distance theory)] and the
2-47
Conventional Ultrasound (UT)
2.13.2 Unit dB
When the "Unit dB" function is turned off, the AVG curve position is set according to the
equivalent value (mm/ inch); when the "Unit dB" function is turned on, the base line is 0dB.
The personal setup wizard will help users configure the professional operation process,
enabling more professional testing. Each wizard includes 2 modes: Basic and
Profession. Basic mode is for primary users, with experience parameter presets based
on field testing; Profession mode is for users of certain test experience, with rich
2.15 AWS
Welding grade is calculated as per AWS (American Welding Society) steel structural
welding code (D1.1) and bridge welding code (D1.5). Before operation, select the
appropriate probe and standard test block according to AWS D1.1 / D1.5.
Operation:
(1) Select "AWS" from the desktop and press Enter to enter the function.
(2) Place the probe on the test point of the standard test block, and ensure good
coupling.
Move the probe back and forth to get the highest echo to make the echo amplitude to
about 80%. Adjust “Start” menu, or Gate key to adjust quickly (see 2.5.2), enabling the
(3) Press Right or Enter key in “Record” menu, and “B_Ref” menu (B
2-48
Conventional Ultrasound (UT)
(4) Place the probe on the tested work piece and ensure good coupling. The
same as step (2), obtain the highest echo of the flow in the workpiece.
The system will automatically refresh “AInd.”, “CAtt.” and “DRating” measure values.
“AInd.”, A Indication Level, which indicates the gain value of the amplitude when the
【Note】
(1) All calculations are based on AWS D1.1-2006 and AWS D1.5-2002.
(2) When recording B Reference Level in step 3), the values of A, C, D are reset
to 0.
2.16 CSC
CSC (Curved Surface Correction) is for depth and horizontal distance correction when
Operation:
(1) To ensure the accuracy of the curve, Vel and auto calibration are required
(2) Select CSC from the desktop and enter the function.
(3) According to the tested workpiece, input "O.D." and "Thickness" values.
(4) Depending on the probe in use, set up the probe parameter menu values.
2-49
Conventional Ultrasound (UT)
(5) If the inner surface of the workpiece is tested, set “Correct “ to “I.D.”; If the
outer surface of the workpiece is tested, set “Correct “ to “O.D.” Now Depth
and Hor measure values will be corrected according to the surface settings
before display.
A quantitative method for flaw depth according to API (American Petroleum Institute).
Operation:
(2) “Dr” menu is set up as a reference indicator for calibration, i.e. the depth of
(3) Place the probe on the standard test block and ensure good coupling.
API 5UE.
(5) Enter “K” menu, and the system will calculate and refresh “K” value.
(6) Place the probe on the tested work piece and ensure good coupling. Move
the probe to bring on screen a dynamic echo envelope required by API 5UE.
(8) Enter “Di” menu, and the system will calculate and refresh “Di” value.
2.18 WeldSimu
WeldSimu is to simulate the weld shape and use the dynamic beam tracking function to
assist the user to quickly determine the location of flaws in the weld.
2-50
Conventional Ultrasound (UT)
Operation:
(1) Set up "WpPrb" parameter based on the tested workpiece and the
connected probe.
(3) Turn on "WeldSimu" function, and the waveform image area is displayed
(4) Place the probe on the workpiece and move the probe back and forth to find
the waveform with high amplitude. Now do not move the probe, and use a
ruler to measure the distance between the probe flank and the weld center,
2-51
Conventional Ultrasound (UT)
(5) Finish settings for "WeldSimu", and enter "Basic" for testing.
Adjust “Gate”to record the echo. Now the end of the beams indicates the flaw in the
workpiece.
Adjust "Gate", trap echo, then the end of the line that defects in the workpiece position.
【Note】The color of beams is the same as the waveform, where the section with
2.19 CrackMeas
To measure crack height with endpoint peak echo method of angle probe.
When incident waves enter the crack tip, some of them are reflected along their original
path, which is called endpoint reflected echoes. The endpoint reflected echo method is to
calculate crack height by measuring the sound path of endpoint peak echo produced
Operation:
8) Now the system will automatically calculate the recorded depth difference
between “H1” and “H2” items, with the result displayed in “Result” menu.
2-52
Conventional Ultrasound (UT)
FFT is to obtain the spectrum based on received signals, and further acquire the measure
Operation:
(3) Now "Range" and "Delay" are automatically magnified according to the gate,
(5) Turn on “MeasFrq” function, and you may select “-6dB” “-12dB” or “-20dB”.
Now the measure values of “Peak (Freq)”, “C.Freq (Freq)”, “L.Edge (Freq)”,
“H.Edge (Freq)” and “RefBW” are refreshed real time. See the figure.
2-53
Thickness Gauge (TG)
Note: The Thickness Gauge (TG) feature is optional. Please contact the supplier for
purchase if necessary.
【Note】: For the menus with the mark * in the menu structure, press Enter key
Main StdMode
Menu
Submenu
-- -- -- ShowWave Vel*
3-1
Thickness Gauge (TG)
Main StdMode
Menu
Submenu
AvgNum -- -- --
MeasFreq -- -- --
Freeze -- -- --
Main CoatMode
Menu
Submenu
Submenu
3-2
Thickness Gauge (TG)
Status
-- LowLimit* -- ShowWave --
-- UpLimit* -- -- --
-- Buzzer -- -- --
Submenu
AvgNum -- -- -- --
MeasFreq -- -- -- --
Freeze -- -- -- --
3-3
Thickness Gauge (TG)
Main ECHOMode
Menu
Submenu
-- LowAlarm -- Freeze --
-- UpAlarm -- ShowWave --
-- LowLimit -- -- --
-- UpLimit -- -- --
-- Buzzer -- -- --
Main ECHOMode
Menu
Submenu
3-4
Thickness Gauge (TG)
AvgNum -- -- --
MeasFreq -- -- --
Freeze -- -- --
Main VelMeas
Menu
Submenu
-- -- -- ApplyParam --
-- -- -- Freeze --
3-5
Thickness Gauge (TG)
-- -- -- ShowWave --
Main VelMeas
Menu
Submenu
DspDly* AvgNum -- -- --
PrbDly* MeasFrq -- -- --
Vel* Freeze -- -- --
Main MultiMode
Menu
Submenu
-- -- Rectify Save --
3-6
Thickness Gauge (TG)
Main MultiMode
Menu
Submenu
-- -- AutoSrch ApplyParam --
-- -- MaxSGain* Freeze --
-- -- AGC ShowWave --
-- -- Gain -- --
-- -- Step -- --
-- -- MeasPnt -- --
Main MultiMode
Menu
Submenu
DspDly* MeasFrq -- -- --
PrbDly* Freeze -- -- --
Vel* -- -- -- --
3-7
Thickness Gauge (TG)
Total -- -- Gate --
2nd -- -- -- Logic --
Submenu -- -- -- Start* --
-- -- -- Width* --
-- -- -- Thresh --
-- -- -- MeasPnt --
-- -- -- Alarm --
-- -- -- Magnify --
-- -- -- MeasLine --
Submenu
Frame Step
Gain
Range* --
Delay* --
2nd
GAStart* --
Submenu
GAWidth* --
GAThresh --
GBStart* --
GBWidth* --
3-8
Thickness Gauge (TG)
GBThresh --
Submenu
2nd MeasLine --
Submenu
3.2.1 StdMode
a) Probe
◆ VpathSel: Select Dual, and then select a V-PATH curve from the curve
the default curve suffix is "* _echo". For Single probe, the VpathSel is fixed
to "None”.
◆ CalibZero: A hot key button. “Ref1” and “Vel” in calibration function are used
for probe delay calibration. That is to perform fast calibration of probe delay
according to the valid echo signals in GateA based on the known material
PrbDly is refreshed.
3-9
Thickness Gauge (TG)
b) Gate
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
【Note 3】If the Alarm function is not turned on or does not meet the alarm
c) Meas
3-10
Thickness Gauge (TG)
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC is
◆ AGC: to enable/ disable AGC function. This function can not be enabled with
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ MeasPnt: To set up the measure point value position of the GateA, which
supports two modes: JFlank (the first intersection between the echo and the
gate in gate range) and Peak (the intersection between the peak wave
d) Data
(press Right to enter the detailed attribute setup screen, press Left to return
increment/decrement.
◆ CurRow: To set up the row location of data Record, Delete, DataRcl for the
◆ CurCol: To set up the column location of data Record, Delete, DataRcl for
◆ CurPnt: To set up the point location of data Record, Delete, DataRcl for the
3-11
Thickness Gauge (TG)
◆ To save the current MeasPnt to Data CurRow, CurCol, CurPnt location. The
◆ Delete: Delete the MeasPnt data at the CurRow, CurCol, CurPnt position.
◆ Save: In Storage menu of Manage, select the file type as "Data", and you
may set up the "Storage mode". If it is "Auto", click Save. If the file matches
the save condition, it will be named and saved to the SD card based on the
automatically; If it is "Manual", click Save, and it will pop up the name edit
◆ DataRcl: To recall the saved MeasPnt data from the current data (specific
location by CurRow/ CurCol/ CurPnt), and display the saved measure value
apply the saved MeasPnt parameter to the curren use after unfrozen.
Off: Apply the saved point parameter state to the current use after unfrozen.
On: After unfrozen, return to the parameter state before recall operation (that
【Note】 The maximum points (all rows, columns and points of each Data) that can
e) Calib
◆ Calib: Select the items for calibration, which includes Vel, PrbDly and
3-12
Thickness Gauge (TG)
Vel+PrbDly.
◆ Ref1: The reference position of the 1st record point when calibrating.
◆ Ref2: The reference position of the 2nd record point when calibrating, which
situation.
◆ Clear: To clear the calibration information, and restore to the default value
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
f) Display
◆ Mode: For setting the display layout and combination on the screen, with
A/BVa, A/Ba/ BVa, Ba/ BVa as options. "A" refers to the A-scan display;
“BVa” refers to the reading display screen at the central area; “Ba” refers to
the data.
3-13
Thickness Gauge (TG)
5) The system parameters such as Range, Delay and Gate can not be
adjusted.
operators):
2) Couple to the system test block (4mm, Vel 5920m/s), and perform
CalibZero;
◆ MainVal: To set up the display content for measuring MainVal, with the
options available: Std, MinVal, MaxVal, Avg, Diff (See 3.12 for the
◆ LstRead: To set up whether to keep the last reading or display blank when
If in “Hold” mode, the reading is the measure value for the last valid coupling
control of the latest reading cache display on the right of MainVal, which is
calculation restarted if the probe lost coupling time (LostCoupT) reaches the
setup value. For example, when the measurement mode is "MinVal” and the
last valid measure value is 50mm, if the measured value is 100mm this time,
the displayed measure value should still be 50mm. But if the time interval
between the two measurements exceeds the time of "LostCoupT", then the
◆ DiffRef: For Diff measurement, the thickness value is the measured value
◆ AvgNum: For Avg measurement, the thickness value is the average value of
g) TDG
◆ TDG: For turning on/off TDG function. TDG-Time Distance Gain curve. Gain
When TDG is turned on, the screen displays a TDG curve, which
axis direction; while in the amplitude direction, the start gain and the target
gain are compressed in the screen for display, which does not represent the
StGain+CmpGain ≤ 110dB.
◆ StTime: The valid start time of TDG curve (the sound time counting from
zero)
3-15
Thickness Gauge (TG)
h) TempComp
calibration test block is different from that of the test part, you may turn on
-1.0~1.0.
3.2.2 CoatMode
a) Probe
b) Gate
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
【Note 3】If the Alarm function is not turned on or does not meet the alarm
c) Meas
◆ Rectify: To set up rectification mode of receiving echo. There are four Rectify
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC is
3-17
Thickness Gauge (TG)
◆ AGC: to enable/ disable AGC function. This function cannot be enabled with
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ CoatThk: For turning on/off coat thickness measurement. If turned on, when
known for measuring the thickness of the coat itself. The measure value is
the distance between the first wave and the peak wave in the gate A range.
Peak (the highest wave intersection). This adjustment of this item may be
used to set up the gate value mode of GateA and GateB at the same time.
◆ Coat Status: Thin / Thick. Based on “Coat Status” setting during AutoSrch,
GateB accurate positioning. Thin is for the coat smaller than or equal to
d) Alarm
e) Data Save
f) Calib
3-18
Thickness Gauge (TG)
g) Display
h) TDG
i) TempComp
3.2.3 ECHOMode
a) Probe
b) Gate
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
the measure value of GateA. The target is the positioning of the next echo
after GateA.
3-19
Thickness Gauge (TG)
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
L.Edge, The alarm symbol on the right side of system information column
【Note 3】If the Alarm function is not turned on or does not meet the alarm
c) Meas
d) Alarm
e) Data Save
f) Calib
g) Display
3-20
Thickness Gauge (TG)
h) TDG
i) TempComp
3.2.4 VelMeas
a) Probe
◆ WorkMode: For selecting the work mode. Three modes are available:
◆ PrbSel: For selecting the TG probe connected to the system, Std and
Custom are optional. The TG probe model series manufactured by SIUI are
◆ VpathSel: Select DualProbe, and then select a V-PATH curve from the curve
object.
◆ CalibZero: A hot key button. “Ref1” and “Vel” in custom calibration are used
for probe delay calibration. That is to perform fast calibration of probe delay
according to the valid echo signals in GateA based on the known material
PrbDly is refreshed.
b) Gate
c) Meas
d) Alarm
e) Data Save
f) Calib
g) Display
h) TDG
i) TempComp
3.2.5 MultiMode
a) Probe
b) MeasParam
◆ MeasObj: For setting the object of MultiMode, i.e. for selecting multi-layer
◆ LVel1: For setting MtVel of the 1st layer. When MultiMode object is Vel, this
◆ LVel2: For setting MtVel of the 2nd layer. When MultiMode object is Vel, this
◆ LVel3: For setting MtVel of the 3rd layer. When MultiMode object is Vel, this
◆ LVel4: For setting MtVel of the 4th layer. When MultiMode object is Vel, this
c) LMeas
◆ AreaSw: For turning on/off CurArea and the measurement gate for the area.
3-23
Thickness Gauge (TG)
◆ Delay: For setting the ultrasonic pulse shift (display delay) on screen display.
rectification mode of receiving echo. There are four Rectify modes for
◆ MaxSGain: For setting the max search gain value when AutoSrch or AGC is
◆ AGC: to enable/ disable AGC function. This function cannot be enabled with
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
Peak (the highest wave intersection). This adjustment of this item may be
used to set up the gate value mode for each layer at the same time.
d) Data
e) Alarm
【Note 1】When UpAlarm is on, and the measurement value exceeds the
H.Edge, The alarm symbol on the right side of system information column
【Note 2】 When LowAlarm is on, and the measurement value exceeds the
3-24
Thickness Gauge (TG)
L.Edge, The alarm symbol on the right side of system information column
【Note 3】If the Alarm function is not turned on or does not meet the alarm
f) Data Save
g) Calib
h) Display
i) TDG
j) TempComp
3.2.6 B-Scan
a) Ctrl
◆ Save: With the memory connected, to save scan files to the memory, with
3-25
Thickness Gauge (TG)
◆ Curr: In TimeBase mode, " Curr " refers to the scan time that the probe is
coupled to the workpiece; In Encoder mode, " Curr " refers to the scan
◆ Loss: To calculate the total missed time or distance due to excessive scan
b) Setup
◆ Encoder: To select the encoder model when Encoder mode is selected and
the menu is adjustable.
◆ Step: To set up trigger time for scan when TimeBase is selected; To set up
trigger distance precision when Encoder is selected.
◆ Dis: The menu is adjustable when Encoder is selected. It is for setting the
max. length of scan, which is mutual restricted by “Step”.
◆ StPos: The menu is adjustable when Encoder is selected. It is for setting the
start position of scan.
c) Basic
◆ LosMode: To set up the scan stops or continues if the probe loses coupling,
3-26
Thickness Gauge (TG)
◆ FrzMode: For setting the position of cursor line stays in the scanned image
when pressing the Freeze button. CurVal/ MinVal/ MaxVal for option.
◆ Gain: For setting receiving amplifier gain. The higher the gain, the more
◆ DspDly: For setting the ultrasonic pulse shift (display delay) on screen
display.
In the reading column there is a coupling symbol with an inverted arrow. In StdMode, a
solid coupling symbol appears when gate A and B are coupled at the same time, or a
hollow coupling symbol appears when one of gate A/B is coupled; In MultiMode, a solid
coupling symbol appears when any gate is coupled. There is no such symbol in other
situation.
The image information column of ultrasound images is as shown in Table 3-2, which is for
Symbol Meaning
3-27
Thickness Gauge (TG)
Symbol Meaning
The gate is in valid coupling with the echo. (One gate in good
coupling in MultiMode)
CoatMode/ ECHOMode.
3.3.1 Reading
The reading column below the information column shows 5 readings. See Fig. 3-1.
The first reading is for gain, in the format: the upper one is total gain (with the unit in the
In addition, 3 readings can be selected by the user, where the last reading is a big value
(BVa), which is highlighted in larger font size. See Table 2-2 for the options of reading.
Symbol Meaning
HxdB dB difference between echo measurement point amplitude and gate height
within gate x.
MaxVal The Max thickness/ VelMeas in the valid measure time in non B-Scan;
3-28
Thickness Gauge (TG)
Symbol Meaning
MinVal The minimum thickness/ VelMeas in the valid measure time in non B-Scan;
B-Scan;
3.4.1 StdMode
If WorkMode is StdMode, the thickness reading is the measure value between the
calibration zero and the GateA meas point (JFlank: the zero-crossing point of the first peak
exceeding gate thresh. Peak: the zero-crossing point of the peak exceeding gate thresh).
The small inverted triangle identifier automatically tracks gate measurement position.
3-29
Thickness Gauge (TG)
Operation:
(1) In Probe menu, set up PrbSel, ProbeDly and MtVel. If required, perform
(2) In Gate menu, adjust the gate position, set up measurement parameters and turn
on the over-limit alarm function.
3.4.2 CoatMode
If WorkMode is CoatMode, the thickness reading is the measure value between the gate
A MeasPnt and gate B MeasPnt. (MeasPnt can be selected. JFlank: the zero-crossing
point of the first peak exceeding gate thresh. Peak: the zero-crossing point of the peak
exceeding gate thresh). The small inverted triangle identifier on the gate automatically
3-30
Thickness Gauge (TG)
If CoatThk is enabled, the coating thickness is measured when the base below the
of the first peak exceeding aThresh is measured by the software, and the coating
Operation:
(1) In Probe menu, set up PrbSel, PrbDly and MtVel. Generally, perform CalibZero
for the known material velocity in Probe menu of StdMode. If required, perform
CalibZero in this menu again, and then enter the measurement state.
(2) In Gate menu, adjust the gate position, set up measurement parameters and turn
on the over-limit alarm function.
(3) In Meas menu, set up Rectify and AutoSrch. If CoatThk measurement is enabled,
3-31
Thickness Gauge (TG)
set up the known CoatVel, and the coating thickness can be measured at the
same time.
3.4.3 ECHOMode
If WorkMode is ECHOMode, the thickness reading is the measure value between gate A
MeasPnt and gate B MeasPnt. (MeasPnt can be selected. JFlank: the zero-crossing point
of the first peak exceeding gate thresh. Peak: the zero-crossing point of the peak
In ECHOMode, Gate A is the trigger gate. Adjust Gate A to record the first bottom wave
(by ensuring the measurement position in the gate is the peak position of the correct first
bottom wave, which can be determined by adjusting aStart and aWidth, and the position of
the inverted triangle in the gate line), According to the peak position in gate A, bStart is
shifted automatically in proportion to record the second bottom wave. The difference of the
peak in gate B and the peak in gate A is the thickness value of Echo to Echo (also by
ensuring the measurement position in gate B is the peak position of the correct second
bottom wave, which can be determined by adjusting bStart and bWidth, and the position of
ECHOMode is suitable for fast measurement of B1-B2 with different thickness. When
adjusting the position 2 gates for manual measurement is required, use CoatMode. By
adjusting the Start and Width of the two gates, the Bm-Bn value (such as B2-B3, or B3-B4)
is measured.
3-32
Thickness Gauge (TG)
Operation:
(1) In Probe menu, set up PrbSel, PrbDly and MtVel. Generally, perform CalibZero
for the known material velocity in Probe menu of StdMode. If required, perform
CalibZero in this menu again, and then enter the measurement state.
(2) In Gate menu, adjust the gate position, set up measurement parameters and turn
(1) The measured waveform amplitude or gate should be adjusted according to the
measurement requirements for each mode, enabling the signal at the measured
AutoSrch or AGC cannot be achieved, please cancel the automatic function and
3-33
Thickness Gauge (TG)
and stable measurement reading. For the general wide-pulse dual probe,
ensure the cursor position in the 2 gates corresponds to signal in the 2 gates
and better measurement accuracy. You may manually adjust the gate start,
width and thresh of the 2 gates to properly locate the measurement position.
(3) If Rectify is RF, and GateThresh is default to negative, it measures the negative
to the first peak of negative signal in the gate exceeding gate thresh. The
first peak of positive signal in the gate exceeding gate thresh. Due to the weak
peak signal, the zero-crossing after the peak is not stable, which is not
RF (GateThresh is set Neg) or Neg for Rectify, so as to get the most stable
measurements displayed.
3.5 VelMeas
Subject to the selected WorkMode, the material velocity is calculated and displayed based
Before exiting VelMeas, choose whether to apply the velocity measurement to the current
material.
3-34
Thickness Gauge (TG)
Operation:
(1) In Probe menu, set up WorkMode, PrbSel, PrbDly and the known thickness. If
(2) In Gate menu, adjust the gate position and set up the measurement parameters.
3.6 MultiMode
If the workpiece has multilayer materials which are linked closely, you can use MultMode
to measure, provided that the material velocity for each layer is known. The actual
measurement effect is dependent on the multi-layer material. If the thickness for each
layer is known, the velocity for each layer can also be measured. For MultMode
For MultMode measurement, up to 4 layers can be measured at the same time. There are
4 measurement gates L1, L2, L3, L4 in total. The measurement for the 1 st layer is the
sound path difference between the first wave peak in gate L1 to zero; The measurement
for the 2nd layer is the sound path difference between the first wave peak in gate L2 and L1;
The measurement for the 3rd layer is the sound path difference between the first wave
peak in gate L3 and L2; and the measurement for the 4th layer is the sound path difference
(The description example above takes Peak as "MeasPnt". In the practice, you may also
set it as JFlank, the measure value at the intersection of the gate and the first wave. )
The main reading display for MultMode is the sum of layer measurements, with the
3-35
Thickness Gauge (TG)
Operation:
(3) In LMeas, adjust the gate position for the measurement area, and set up the
3.7 B-Scan
B-Scan function is to perform thickness measurement based on the set time interval or the
3-36
Thickness Gauge (TG)
encoder, and display the measurement reading of scan process in B-mode images, which
is for thickness measurement image analysis and recording for the scanned workpiece.
Operation:
regular parameters and calibrate zero. Return to the desktop and enters B-Scan
screen. B-Scan uses the current work mode for measurement. It does not
(2) Set up ScanMode, Step, TimeL in Setup menu. Select the encoder, scan
(3) In “Basic”, set up Basic parameters, thickness display range, the mode for lost
3-37
Thickness Gauge (TG)
(4) In “Ctrl”, you can start or finish B-Scan. After scanning, save the record or
Calibration wizard is for MtVel and PrbDly calibration. The user may choose to calibrate
PrbDly (1-point calibration), or PrbDly and MtVel (2-point calibration) at the same time.
3.8 Calibration
1- point Calibration is to calibrate ProbeDly only with known velocity. Before calibration,
Operation:
(1) Select the probe and the curve to enter the calibration menu (the "Calib" 1 st
up Basic parameters. Select “PrbDly” as the calibration mode, and set up the test
block thickness for Ref1. Adjust to the appropriate Range and DspDly.
(2) Couple the probe on the test block 1 to be calibrated. Adjust “GAStart”,
“GAWidth” and “GAThresh” to bring Gate A to record the first wave. Set up the
(3) Tap “Record” to finish calibration. For 1-point Calibration, only 1 point is recorded
for PrbDly calibration. If not satisfied with the result, use “Clear” to restore to the
【Note】 After the 1-point Calibration, the ProbeDly value is refreshed by the
calibrated value.
3-38
Thickness Gauge (TG)
2- point Calibration is to calibrate PrbDly and Vel at the same time. Before calibration,
Operation:
(1) Select the probe and the curve to enter the calibration menu (the "Calib" 1 st
Basic parameters. Select “PrbDly” as the calibration mode, and set up the the test
block thickness for Ref1 and Ref2. Adjust to the appropriate Range and DspDly.
"Ref1" and "Ref2" test blocks at the same time for PrbDly and Vel calibration.
(2) Couple the probe on the test block 1 to be calibrated. Adjust “GAStart”,
“GAWidth” and “GAThresh” to bring Gate A to record the first wave. Set up the
(3) Tap “Record”. If the coupling is good, it prompts “Please record echo for Ref2”.
(4) Couple the probe on the test block 2 to be calibrated. Tap “Record”, and it
prompts “Calibration finished”. If not satisfied with the result, use “Clear” to
【Note】 After the 1-point and 2-point Calibration, the PrbDly and Vel values are
"VelMeas" and "MultiMode" corresponding to the 2nd menu hot key button “CalibZero”, and
after setting the hotkeys F1and F2 to "CalibZero", you may perform 1-point calibration to
3-39
Thickness Gauge (TG)
3.9 TG Probe
TG probes include standard /custom. The standard TG probes are made by SIUI, with the
TG probe series as options. The NotStd TG probes are non-standard model or user
defined probe types. The probe specification series include: DualDelay, SingleDly, and
Operation:
The probe list is on the left, and the probe attributes on the right. Select one and press
3.10 V-PATH
All the original dual probes have a set of default V-PATH calibration curves. Users can
make a set of UserVpath curves for a specific probe. Generally you need to identify
whether it is StdMode, CoatModeor or ECHOMode, and then make the UserVpath for that
3-40
Thickness Gauge (TG)
WorkMode. CoatMode and ECHOModee may share the same V-PATH curve.
The VPATH curve for the current WorkMode of the probe in use is displayed on the left,
and the Basic information such as Type; PntNum; point attributes of the selected curve is
AutoSrch function may search and adjust waves quickly based on the received wave
signals. According to the wave signals of the tested thickness, AutoSrch quickly adjusts
the system automatically, including the adjustment of DspDly, Range, Gain and the
material, thickness and surface condition. If AutoSrch cannot capture and measure
properly, please turn off AutoSrch and use manual adjustment for measurement.
If the probe type is Dual, in StdMode, CoatMode or WorkMode, AutoSrch shall include
auto adjustment of searching waves, range, gain and gates. If WorkMode is StdMode, the
software search will bring gate A to record the first wave and adjust AGC of the first peak
to 80%. If the workmode is CoatMode or ECHOMode, the software search will bring gate
A to record the first wave and adjust AGC of the first peak to 80%, and gate B
automatically records the second wave. AutoSrch in MultiMode and B-Scan is equivalent
If the probe type is Single, AutoSrch in all work modes is equivalent to auto gain A only.
AutoSrch can be set Single auto search, i.e. to auto search once if enabled once, or
The maximum gain auto adjustment of AutoSrch is bound by the set value of “MaxSGain“.
If “MaxSGain“ is set too high, it may result in searching abnormal noise signals; If
“MaxSGain“ is set too low, it may result in the desired signals not measured. The specific
“Coat Status” parameter in “Meas” menu. “Coat Status” includes “Thin” and “Thick”. When
the system is in calculating GateB position, it will refer to this parameter to adjust the
position.
【Note】" Coat Status " is suggested to set as below: set to “Thin” if the coat
Operation:
Select “Continue” or “Single” or “Off” in "AutoSrch" of Meas menu, or press the hot
key F1or F2 (when the hot key is defined as AutoSrch), to enable or disable this
function.
If AGC is enabled, select to perform auto gain control for gate A or gate B (for CoatMode
and ECHOMode only). For StdMode, the first peak amplitude in gate A is automatically
adjusted to 80% height; For CoatMode and ECHOMode, the maximum peak amplitude in
AGC and AutoSrch functions cannot be enabled at the same time. The maximum gain
Operation:
Select “GateA”, “GateB” or “Off” in “AGC “of Meas menu to enable or disable this
function.
3-42
Thickness Gauge (TG)
3.12 TDG
TDG- Time Distance Gain curve. Gain compensation based on time and distance, which
may compensate the loss of echo amplitude due to propagation of sound path.
When TDG function is turned on, the screen displays a TDG curve, which corresponds to
the actual signal position of gain compensation in the time axis direction; while in the
amplitude direction, the start gain and the target gain are compressed in the screen for
StGain+CmpGain ≤ 110dB.
◆ StTime: The valid start time of TDG curve (the acoustic time counting from
PrbDly)
Take the 2D data (4 rows, 5 columns) as an example for the operation. The figure when
The upper part of the center is the Ba of the data, and the lower part of the center is the
A-scan figure. In this mode, the small data column of the upper part displays an enlarged
reading, i.e. the current measure value. "Std" means the reading is set to Standard (which
3.13.1 Save
Operation:
(1) Set up DNum, RowNum, ColNum and PntNum in "Format" (total point
number exceeding 1000.) If the storage file has been saved with data, you
cannot modify the content in "Format”, but only set up again after "Clear"
operation.
(2) Exit the "Format" set up screen and return to the "Data" operation screen.
Select the desired "CurRow", "CurCol" and "CurPnt." Couple the probe
3-44
Thickness Gauge (TG)
(3) Press Record key to save, or set up the Store key in “System” as the hotkey
(4) After saving properly, the measure value will be displayed on the
3.13.2 DataRcl
Operation:
(1) Adjust the "CurRow / CurCol / CurPnt" to the position to recall the data, and
run "DataRcl".
(2) If data is correct, the stored measure value is displayed in the reading
column. The A-scan figure is restored to the case where the data is saved,
the recall operation (ie, the DataRcl operation is for preview only).
3.13.3 Other
Save: after running the save function, the data is automatically named in the named
3-45
Thickness Gauge (TG)
If the workmode is StdMode, see Fig. 3- for the flow chart of general thickness gauge
application.
3-46
Thickness Gauge (TG)
For most materials, such as carbon steel, the general flow chart above is practical and
effective. On how to use the system, especially on calibration and auto search, see 3.11.
When testing a thickness close to the critical measurement range, sometimes AutoSrch
may need manual adjustment and judgment of the operator to ensure the measurement
accuracy. For coarse grain, high-attenuation material (such as ductile iron), special
measure. Now it is very important that the operator be acquainted with certain ultrasound
3-47
Management Operation
This chapter describes the operation and technique for storage, work piece, probe,
The screen layout of storage management is shown in Fig. 4-1. For functional description,
1 Functional menu
4-1
Management Operation
The Smartor has the ability to store thickness values measured during inspection or
thickness measurement tasks. The user has the option to create a type of file depending on
their needs (No. of rows and columns needed, type of component, etc.), Smartor has 8
types of file format, these files are structured according to the number of readings that the
user is going to obtain, and a file name defined by the user is assigned.
Before taking the step to create a file, it is important to take into account the characteristics
When you are going to create a data file in the DR menu, the following 8 options will be
available:
1.- Linear
2.- Sequential
7.- Boiler
In the file list, it displays the selected file type only. Select "Type" menu, press Enter to
toggle between different options, including "Image", "Param", "Record" and "Data" (for TG
only).
◆ Param file: For saving system setup parameters. The formats include: UT
includes: UT CineRec, "*.urec" for B-scan file, and "*.tgrec" for TG B-Scan file.
4-2
Management Operation
◆ Data file: For TG mode ohly, which is for saving TG measure date, in the format
“*.tgdat”.
4.1.2 Save
In the storage management mode, only "Save" of "Param" file type is valid (for the storage
of other file types, use the save operation if F1 or F2 is set for related function).
Operation:
(3) Save the current ParamFile under the specified path (see 4.1.7.2).
4.1.3 Open
Method 1:
(1) Select "Open" menu. Press Right or Enter to enter the file list control;
(2) Press Up or Down to select the desired file to open, and press Enter to open it.
4-3
Management Operation
Note: If the type is Image, the file cannot be opened with this method.
Method 2:
(1) Select "Type" menu. Press Enter to select the type. When the selection is
(2) Select a file and press Enter, Follow the screen prompt to select the responsive
◆ Param: Apply the saved parameter information to the system, and exit the
4.1.4 Preview
Turn on "Preview" and enter the preview mode. You may view part of the information for
4.1.5 Delete
Turn on “Delete”, and you may delete one, multiple or all files.
Operation:
(1) Select "Delete" menu. Press Right or Enter to enter the file list control.
(2) Select one or all files, press Enter, and follow the screen prompt to delete the
selected file(s).
Note:
4.1.6 Rename
Operation:
(1) Select "Rename" menu. Press Right or Enter to enter the file list control.
(2) Select one file, press Enter, and you may rename the file in the pop-up file
naming window.
4.1.7 StoreSet
When saving a file (not including image type), it requires to input the file name, Now the
system will provide a default file name to improve the user's operating efficiency. There
are 2 types of file name and naming method: Default and Incr.
YYMMDD_hhmmss.
◆ Incr: To set up the format, you may specify the incremental start bit, and the
increase.
4.1.7.2 StorePath
To set up the path for file storage. The user may choose "Default" or "Custom" for
StorePath.
a) If "StorePath" is set default, the system automatically creates a default store path
4-5
Management Operation
b) If "StorePath" is set to "Custom", the user may customize the file store path. How
to “Custom” StorePath:
(1) Set "StorePath" menu to "Custom" and select the StorePath information below
the menu. Press Right or Enter to enter the file list control.
(2) Press Up, Down or Enter to switch the directory, and set up StorePath by
4.1.7.3 StoreName
To set up whether to bring up the file name input screen when saving the file.
a) If "StoreName set "Off", the file will be saved directly according to the naming rule
for the set stored file and StorePath when saving a file.
b) If "StoreName" is set "On", the user can modify the file name in the pop-up file
Note: If the storage type is "Image", no file naming window is popped up when
storing.
4.1.7.4 NewFLR
Operation:
(1) Select "NewFLR" menu, press Right or Enter to to enter file list control.
(2) Press Up or Down to switch the current directory to the specified one.
(3) Press Enter and finish NewFLR operation by following the on-screen prompt.
4-6
Management Operation
Encoder management is for naming, calibrating and deleting the crawler encoders
connected with the system. The screen interface is shown in Fig. 4-2. See Table 4-2 for
functional description.
1 Submenu
2 Encoder in use
4-7
Management Operation
4.2.1 Edit
4.2.2 New
Operation:
(2) Adjust “Precisn” menu and input the known precision of the encoder.
a) Adjust "CalibD." menu to set up the moving distance for encoder calibration.
b) Enable "Calib". At the bottom of the screen it shows the operation tip.
c) Place the encoder behind the start position and press Enter.
d) Move the encoder to the finish position, i.e. the actual calibrated distance,
then press Enter. If the encoder moves forward during this process, the
"Move" menu will refresh the current moving distance in real time (based on
the vale calculated from encoder precision), otherwise, the prompt is invalid.
f) After moving the encoder from the finish position backward to the start
position, press Enter. In this process, the “CurrMov” will refresh its moving
distance (the value calculated from the new encoder precision) in real time.
the "CurrMov” menu display should≈ the input moving distance in "CalibD."
menu), if it is normal, follow the tip and press Enter, and the calibrated new
4-8
Management Operation
encoder precision will be set to the current setting, otherwise press HOME to
cancel.
(4) After finishing the step, press Right or Enter, input the encoder name and save it
4.2.3 Delete
This function enables the user to delete the created encoder. After the "Delete" function is
enabled, operate by following the operation tip at the bottom of the screen.
4.2.4 Rename
This function enables the user to rename the created encoder. After the "Rename"
function is enabled, operate by following the operation tip at the bottom of the screen.
4.2.5 Test
Operation:
(3) Press Enter to start the test. Now the "move" menu is cleared at the same time.
(4) While the user moves the encoder, observe if the "move" menu value complies
(5) When the moving is finished, return to step (2) to continue testing, or press
4-9
Management Operation
The VPATH management interface is as shown in Fig. 4-3. The user may create a new
The middle list shows all the VPATH curves (including the built-in and user defined curves)
default curve suffix is “* _sys"; in CoatMode and ECHOMode, the default curve
suffix is "* _echo". For SingleProbe, the Vpath selection is fixed at "None".
4-10
Management Operation
4.3.1 Edit
Operation:
(2) Press Up/ Down to select the curve to be edited, then press Enter.
(3) Adjust "CurPnt" menu to specify the point on the curve to be modified. Then
"BlkThk" and "MeasRes" are updated to the stored value of the CurPnt.
(4) To adjust the CurPnt data, adjust "BlkThk" and "MeasRes" menus.
the "BlkThk" values between two points before and after the current points. Such
(5) To insert a point, you can turn on "InsertUp" or "InsertDn" function. One point is
automatically created before or after the current point. "BlkThk" and "MeasRes"
(8) If you are not satisfied with most of the point data in the current curve, turn on
(9) Turn on "Finish" to confirm whether to save the modified curve information.
4.3.2 New
Operation:
4-11
Management Operation
(2) Press Up / Down key to select a curve as a reference for the new curve, that is to
create a new curve exactly the same as the specified curve for modification, and
(4) Turn on "Finish" to enter the input state. Input the curve name and confirm
4.3.3 Delete
See 4.2.3.
4.3.4 Rename
See 4.2.4.
【Note】 The default V-PATH curve is for browsing the record points only, while
edit and management functions can be performed on the custom curve. The
V-PATH curve has up to 32 measurement location points, and the selection of the
32 points shall focus on an actual measure range that requires smaller measure
error, such as the range lower than 10mm (steel longitudinal wave), or 20-40mm
(steel longitudinal wave) range, but not a random measure point between the
minimum to the maximum range (generally the default V-PATH is fine). For high
UserVpath curves.
The Wp management screen is as shown in Fig. 4-4. The user may create a new work
4-12
Management Operation
The user-defined workpiece can be called from "Wp", and the information such as
thickness and velocity of the workpiece after being called up will be applied to the current
parameter settings.
The probe management interface is as shown in Fig. 4-5. The user may create a new
The user-defined probe can be called from “Probe”, and the information such as probe
angle (ValK) and PrbDly after being called up will be applied to the current parameter
settings.
4-13
Management Operation
4-14
System Setup
In "Setup" menu, you can view and set up the basic information of the system, set up the
display content of Reading column in the test interface and view the activated optional
functions.
5.1 System
Operation:
The system setup consists of "System" and "UT/ TG", wherein "System" is for setting
system parameters, shared by UT and TG modules and consistent; UT and TG are valid
For the operation, you may refer to the prompts at the bottom of the screen.
Enter key is for quickly toggling the option menu value; Gate and Gain keys are for
5-1
System Setup
settings, while some parameters that affect the operation (e.g. language,
◆ Update: Connect the memory card containing the update package to the
automatically enter the update process, and shut down when finished.
Operation:
(2) The display after pressing Enter key is subject to the adjust object. 17-01-01
◆ Time: Adjust system hours/ minutes. Refer to "Date" for the operation.
◆ DispDate: Turn on/off the date displayed at the upper right of the screen.
◆ Unit: Set up the units used in the system, including "mm" and "inch".
【Note】 Lower screen brightness can extend the life of the display, and
the ambient brightness is high, the higher screen brightness may improve
◆ Outdoor: When the user operates the system outdoors, turn on "Outdoor" to
help the user to view the system menu interface and operate more clearly.
◆ VGA: Set up VGA video output. When the user connects VGA video out port
display screen to the external display. Now the "VGA" symbol will be
◆ Buzzer: used with "Alarm" function. If the alarm requirements are met and
the function is enabled, the buzzer in the system will give sound alarm.
◆ L/R Op: As the system function keys are all on one side, to facilitate the use
of either left or right hand easily, turning on and off this function to rotate the
screen display 180 °, and all the display will be rotated 180 °.
◆ LogExport: Enable this function, and the key operation logs during the use of
【Note】 The logs are saved in the SMARTORLOG folder of the root
◆ StorePos: Select to save all data and screenshots inside/outside the system.
5-3
System Setup
◆ AGC value (UT only): Set up the target amplitude of auto gain.
◆ Storage key (TG only): For selecting the storage method, including QuickSv
➢ QuickSv: Quickly save the point date of dataset. Now the valid coupling
"current row / column / point", and the current line / column / point is
➢ Choose: For the user to select the storage type, including: PrintScreen,
ParamFile, DataRecord.
◆ Precisn (TG only): set up the display precision of the measurement data.
5.2 Reading
Enter the "Reading" function, and you can select the display object of the reading column.
5.3 Activate
Operation:
(1) Connect the memory to the system till the system prompts the detection of
the memory.
5-4
System Setup
(3) Enable "Export" function, and save the file to the memory.
【Note】 The file storage path defaults to the directory with the SN as the
directory.
(5) Copy the license file obtained from the vendor into the memory.
(6) Connect the memory to the system till the system prompts the detection of
the memory.
(8) The screen displays the list identified as a directory or ini file in the memory.
Through the need and prompt to switch the directory, find the target license
(9) If the function is activated successfully, the software function status, the
5-5
Operator Training and System Maintenance
To ensure a safe and proper operation, the operator shall receive appropriate training on
ultrasound testing, especially training on phased-array ultrasound testing, and read this
Operation Manual before operating this system. The lack of relevant knowledge on
ultrasound testing may lead to unpredictable wrong testing results. Therefore, to acquire
information on ultrasonic testing training and qualification, please contact relevant NDT
Please use the factory configured accessories and options, such as probe cable and
communication cable. Incompatible cables may result in system internal circuit failure or
6.2.2 Cleaning
DO NOT drop any liquid into the system. The system enclosure, glass panel and other
parts may be cleaned with a piece of cloth soaked with a little water or neutral household
cleanser.
【Note】: DO NOT use any solvent to clean the system. The plastic parts may be
6.3 Service
The system adopts the latest technologies and is made of high quality components. The
process inspection, the in-process testing and the quality management system are all
In case any failure is found in the system, please shut off the system and take out the
re-calibrate the system. The contact information is listed in the Preface of the Operation
Manual. The warranty does not cover any unauthorized repair or disassembly.
Under the following situations, the safe operation of the system is not ensured:
b) The system is stored in adverse environment for a long time (eg. abnormal
and humidity-proof.
6-2
Common Failures and Troubleshooting
suggested to power the system with the adapter or use the battery with ≥30% power
1) Copy the software update package to the root directory of the SD card.
port.
3) Turn on the system. In the desktop button screen, enter the "Setup" menu
screen, select System" 1st menu, press Enter in the "Update" menu, and
a) If the power supply is from the battery, press Power and the system has no
response:
2) Remove the battery from the system, and wait for about 10 seconds before
installing the battery again in the system. Try to turn on the system.
3) Remove the battery from the system, and try to use the adapter for power
supply.
7-1
Common Failures and Troubleshooting
b) If the power supply is from the adapter, press Power and the system has no
response:
1) Check if the system is installed with battery. If yes, remove the battery, and
use the adapter for power supply separately. Try to turn on the system.
4) Disconnect the adapter from the main unit. Wait for about 10 seconds before
5) If the battery is not used on the system, try to use the system with a full-charge
c) The system can be turned on, but cannot enter the application properly.
1) Turn off the system, and disconnect all the peripherals connected to the
system, including USB device, crawler and network cable. Try to turn on the
system.
2) If step 1) does not work, try to update the system with the update package.
d) If the failure still exists after trying the solution above, please contact the
distributor or SIUI.
If normal shutdown cannot power off the system, long press Power for about 10 seconds
to force the system off, or unplug the adapter and/ or remove the battery to cut off power
supply. Restart the system to see whether it is in normal operation. If not, try to contact the
7-2
Common Failures and Troubleshooting
【Note】For updating software, it would be better to restart the system with the
7.5 No Echo
a) Check if the S/DPrb setting is correct, whether it is set Dual, but with a single
b) Check if the probe cable is in bad contact. Try to replace another probe cable.
c) Check if the setup of Gain, Range, DspDly, PrbDly and Reject are correct.
d) Set the system to Default, use a new probe cable to connect a good probe and
scan on a test block. If echoes appear, the failure is due to improper setting.
e) If the failure still exists after trying the solutions above, please contact the
distributor or SIUI.
a) Check if AVG curves are made. DAC curves cannot be created if AVG curves are
made.
c) Check if the echo amplitude recorded by the gate is not within the range of 20%
~100%.
7-3
Common Failures and Troubleshooting
a) Check if DAC curves are made. AVG curves cannot be created if DAC curves are
made.
c) Check if the echo amplitude recorded by the gate is not within the range of 20%
~100%.
c) Check whether the gate records the echo to be tested or there are clutters with
higher echo amplitude. By measuring PeakEcho within gate, only when the
echo being recorded by the gate properly, can accurate calculation result comes
out.
refers to the horizontal distance between the incident point of probe and the flaw,
otherwise, it refers to the horizontal distance between the probe flank and the
flaw.
e) Check whether the gate records the echo to be tested or there are clutters with
7-4
Common Failures and Troubleshooting
higher echo amplitude. By measuring PeakEcho within gate, only when the
echo being hitched by the gate properly, can accurate calculation result comes
out.
between the flaw (beam reflection point) and the probe, rather than between the
round-hole center and the probe. Otherwise, the bigger the round-hole diameter
a) Check if the adapter AC power input plug is connected properly or comes loose.
b) Check if the adapter indicates normal operation. The adapter indicator is green
If the failure still exists after trying the solution above, please contact the distributor or
SIUI.
7-5
System Technical Specifications
Display Screen 5.7''TFT high brightness LCD screen, resolution 640 × 480
Operation
Keys
Method
Unit mm/inch
Model DC-A010
Capacity 5.0Ah
Battery
Voltage 7.4V
Quantity 1pc
Operation
≥8 hours
Time
Model CD-92A
Power 36W
A-1
System Technical Specifications
Network Port ——
Encoder
1 pc (4-core)
Connector
TG calibration
1 pc
block
IP Rating ≥IP66
A-2
System Technical Specifications
Max. supporting
2 pcs
elements
Number of sampling
1024 points, 16bit/point
points
11 levels:
5MHz/10MHz/13MHz/15MHz/20MHz
A-3
System Technical Specifications
and B
Measurement
Cursor: 2 cross cursors, for measuring horizontal and vertical
optional B-Scan)
Number of gates: 2
Curve Function curve, compliant with NB/T47013, GB/T 11345 and GB/T29712.
A-4
System Technical Specifications
Detection sensitivity
≥65dB (P2.5-20 probe)
surplus
StdMode (R-B1, measurement from transmit pulse to the 1st bottom wave);
by adopting zero-crossing measurement
Work Mode TG ECHO mode: ECHO to ECHO TG mode (B1 - B2, which is measured
by automatically tracking the 2nd bottom wave according to the 1st bottom
wave); CoatThk measurement can also be achieved.
Indication
0.001/0.01/0.1 mm (0.0001/0.001/0.01 inch)
resolution
A-5
System Technical Specifications
0.80~9.99mm ± 0.05mm
Equipped with TGM5-10L
Indication
10.00~99.99mm ± (1‰H + 0.04)mm probe; H is the measured
accuracy error
thickness value
100.0~400.0mm ± 3‰H mm
Pipe wall
Equipped with TGM5-10L probe, it can measure the wall thickness of steel
thickness
pipe with diameter ≥Φ20mm and wall thickness ≥2.0mm.
measurement
Range 5~1000mm
Pulse Type Negative square, negative spike (auto-match according to probe type)
PRF 200Hz
Operating
0.5~20MHz (auto match to the probe type)
Frequency
0~110dB
A-6
System Technical Specifications
off/on; If this function is activated, the display range, gain and gate position
AutoSrch can be automatically adjusted properly according to the measured
waveform signals, so as to improve measurement efficiency.
Measurement
Std, MinVal, MaxVal, Avg, Diff
reading mode
Display
A/BVa, A/Ba/SVa, Ba/BVa
Combination
Refresh rate of
4/8/16/32Hz
measurements
Data post Parameters, data sets, printscreen files can be played back and analyzed
processing with SuporUp software, and a report can be generated.
A-7
Compatible TG Probe Specifications
Measurement
Range
(Having
Freque Application Element Contact measurement
Probe Type Remark
ncy Type Diameter Diameter on 45# steel in
Normal mode
without special
instruction)
Dual-element delay type
TGM5-10L 5MHz Dual-element 10mm 11.8mm Standard: With probe
delay type, 0.8~200mm cable
Thru-coating Thru-coating
or (Bm-Bn):
EchoToEcho 3~100mm
Mode
TG5-10L 5MHz Dual-element 10mm 11.8mm 0.8~400mm With probe
delay type, cable
standard,
commonly
use
TG7.5-6L 7.5MHz Dual-element 6mm 9mm 0.8~100mm With probe
delay type, cable
standard, thin
plate
TG2-12L 2MHz Dual-element 12mm 16mm Steel: 3~ With probe
delay type, 600mm cable
high Nodular cast
penetration
Iron:
3~40mm
TG5-10HL 5MHz Dual-element 10mm 12mm 3~300mm With probe
delay type cable
High
temperature:
200°C
TG5-6L 5MHz Dual-element 6mm 10mm 0.8~200mm With probe
delay type, cable
standard
B-1
Compatible TG Probe Specifications
B-2
Compatible TG Probe Specifications
【Note 1】The table above applies to the thickness measurement (TG) module.
【Note 2】When using a dual probe for measurement in CoatMode or ECHO mode, it
is recommended to use a TGM5-10L wideband narrow-pulse probe for high
accuracy and stable measurement readings.
【Note 3】If a low frequency (less than 2.5MHz) or high temperature probe is
adopted, it is recommended to select precision as 0.1mm or 0.01 in.
【Note 4】For measuring coarse or high attenuation materials such as cast iron, it is
recommended to select precision as 0.1mm or 0.01 in.
B-3
Compatible TG Probe Specifications
【Note 5】Using other probes (including the user-defined probes) except the
thickness probe supported by SIUI, the measurement range and precision cannot
be predictable and guaranteed.
B-4