Manual ISCO Signature
Manual ISCO Signature
Manual ISCO Signature
This instruction manual is designed to help you gain a thorough understanding of the operation of
the equipment. Teledyne Isco recommends that you read this manual completely before placing the
equipment in service.
Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a
malfunction. This manual may help in diagnosing and repairing the malfunction.
If a problem persists, call or e-mail Teledyne Isco technical support for assistance. Simple
difficulties can often be diagnosed over the phone. For faster service, please have your serial
number ready.
If it is necessary to return the equipment to the factory for service, please follow the shipping
instructions provided by technical support, including the use of the Return Material
Authorization (RMA) specified. Be sure to include a note describing the malfunction. This will
aid in the prompt repair and return of the equipment.
Teledyne Isco welcomes suggestions that would improve the information presented in this manual
or enhance the operation of the equipment itself.
Teledyne Isco is continually improving its products and reserves the right to change product
specifications, replacement parts, schematics, and instructions without notice.
Contact Information
Customer Service
Phone: (800) 228-4373 (USA, Canada, Mexico)
(402) 464-0231 (Outside North America)
Fax: (402) 465-3022
Email: [email protected]
Technical Support
Phone: Toll Free (800) 775-2965 (Syringe Pumps and Liquid Chromatography)
Email: [email protected]
Return equipment to: 4700 Superior Street, Lincoln, NE 68504-1398
Other Correspondence
Mail to: P.O. Box 82531, Lincoln, NE 68501-2531
Email: [email protected]
Signature® Flow Meter
Table of Contents
Section 1 Introduction
1.1 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2 Data Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.3 Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.4 Compatible Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.5 Identifying Signature Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.6 Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.6.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.6.2 Display and LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.3 Backlight and Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.4 Display Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.7 Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
5
Signature® Flow Meter
Table of Contents
6
Signature® Flow Meter
Table of Contents
7
Signature® Flow Meter
Table of Contents
8
Signature® Flow Meter
Table of Contents
List of Figures
1-1 Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 Multiple options can be used in any combination . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-3 Front and exterior component identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-4 Home screen and basic keypad functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1-5 Specification drawing: Signature Flow Meter with stand in side facing position 1-10
1-6 Specification drawing: Signature Flow Meter with stand in upward facing position
1-11
1-7 Specification drawing: Signature Bubbler Flow Meter, 1 of 2 . . . . . . . . . . . . . . . 1-12
1-8 Specification drawing: Signature Bubbler Flow Meter, 2 of 2 . . . . . . . . . . . . . . . 1-13
2-1 Flowlink Connect screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-2 Site Info tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-3 Devices tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-4 Program tab: Accessing the browser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-5 Home screen (normal operating mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-6 Character grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-7 Menu Tree: Top menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2-8 Menu Tree: Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2-9 Measurement parameters for each TIENet device . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-10 Modbus Input Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-11 Editing Modbus device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
2-12 Modbus Output Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2-13 Menu Tree: Configure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2-14 Menu Tree: Flow rate input setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2-15 Menu Tree: Volume Input Setup (total flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-16 Menu for Rainfall Input Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-17 Menu for Load Calculation Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2-18 Example of defining conditions & building equations. . . . . . . . . . . . . . . . . . . . . 2-26
2-19 Local alarm setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
2-20 SMS and Server alarm setup (modem required) . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2-21 Menu Tree: Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2-22 Menu Tree: USB Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2-23 Event Viewer in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35
2-24 Report file verifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
3-1 Open door and front panel to access interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2 Connector case, connectors, and fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 Connector Case cable entries for power and external devices . . . . . . . . . . . . . . . . 3-4
3-4 Strain relief 3/4 NPT Cord-grip fitting for TIENet devices . . . . . . . . . . . . . . . . . . 3-5
3-5 Cord-grip fittings installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-6 Diameter-seal plugs for unused ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-7 TIENet Device terminal strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3-8 Installing TIENet cable with a cord-grip fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-9 TIENet Device terminal connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-10 Attach wired terminal strip to connector case socket . . . . . . . . . . . . . . . . . . . . . 3-9
3-11 Insert the cable reference tubing into the case board reference port . . . . . . . . . 3-9
3-12 Position and secure the cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3-13 Location of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
9
Signature® Flow Meter
Table of Contents
10
Signature® Flow Meter
Table of Contents
List of Tables
1-1 Signature Flow Meter Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1-2 TIENet 330 Bubbler Module Technical Specifications . . . . . . . . . . . . . . . . . . . . . . 1-9
2-1 Modbus Setup Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
4-1 Battery Life Expectancy
(default measurement rate of 15 minutes) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-2 Battery Life Expectancy
( measurement rate of 5 minutes, all other parameters use default) . . . . . . . . . . 4-8
4-3 Default Method for Portable Signature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-4 Additional Average Current Draw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
5-1 Direction of arrows and mode types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
6-1 Recommended Maintenance (Accessible Locations) . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 Recommended Maintenance (Difficult-to-Access Locations) . . . . . . . . . . . . . . . . . 6-1
6-3 Troubleshooting: Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6-4 Troubleshooting: TIENet 300 Connector Case . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6-5 Troubleshooting: TIENet 301 pH/Temperature Device . . . . . . . . . . . . . . . . . . . . 6-11
6-6 Troubleshooting: TIENet 304 Contact Output Card . . . . . . . . . . . . . . . . . . . . . . 6-12
6-7 Troubleshooting: TIENet 307 Analog Input Card . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6-8 Troubleshooting: TIENet 306 Sampler Interface . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6-9 Troubleshooting: TIENet 308 4-20mA Analog Output . . . . . . . . . . . . . . . . . . . . . 6-14
6-10 Troubleshooting: TIENet 310 Ultrasonic Level Sensor . . . . . . . . . . . . . . . . . . . 6-15
6-11 Troubleshooting: TIENet 330 Bubbler Module . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
C-1 Modbus Output Registers for Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . C-4
11
Signature® Flow Meter
Table of Contents
12
Signature® Flow Meter
Safety
WARNING
Avoid hazardous practices! If you use this instrument in
any way not specified in this manual, the protection
provided by the instrument may be impaired.
AVERTISSEMENT
Éviter les usages périlleux! Si vous utilisez cet instrument
d’une manière autre que celles qui sont specifiées dans ce
manuel, la protection fournie de l’instrument peut être
affaiblie; cela augmentera votre risque de blessure.
Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts,
These three levels are described in the sample alerts below.
CAUTION
Cautions identify a potential hazard, which if not avoided, may
result in minor or moderate injury. This category can also warn
you of unsafe practices, or conditions that may cause property
damage.
WARNING
Warnings identify a potentially hazardous condition, which
if not avoided, could result in death or serious injury.
DANGER
DANGER – limited to the most extreme situations
to identify an imminent hazard, which if not
avoided, will result in death or serious injury.
v
Signature® Flow Meter
Safety
Hazard Symbols The equipment and this manual use symbols used to warn of
hazards. The symbols are explained below.
Hazard Symbols
Warnings and Cautions
The exclamation point within the triangle is a warning sign alerting you of
important instructions in the instrument’s technical reference manual.
The lightning flash and arrowhead within the triangle is a warning sign alert-
ing you of “dangerous voltage” inside the product.
Symboles de sécurité
Ce symbole signale l’existence d’instructions importantes relatives au
produit dans ce manuel.
Der gepfeilte Blitz im Dreieck ist ein Warnzeichen, das Sei vor “gefährlichen
Spannungen” im Inneren des Produkts warnt.
Advertencias y Precauciones
Esta señal le advierte sobre la importancia de las instrucciones del manual
que acompañan a este producto.
vi
Signature® Flow Meter
Section 1 Introduction
Note
Recent enhancements have been made to the Signature Flow
Meter. Many of these features are standard for all new Signa-
ture units which are identifiable by the serial label mounted on
the bottom of the unit (PN 60-4304-065).
1-1
Signature® Flow Meter
Section 1 Introduction
• Current displayed function •Clear last entry • Exit list • Close window
1-2
Signature® Flow Meter
Section 1 Introduction
1.2 Data Integrity What makes the Signature Flow Meter unique is its ability to
verify data integrity. This is accomplished by logging four special
event data types that cannot be altered, and are designed to alert
the user to any trends or anomalies, and to assess compliance.
This data can be downloaded from the flow meter and observed
in tabular or graphical format alongside regular site data, using
Flowlink software (see Section 2.10 Signature Data in Flowlink).
The data can also be downloaded onto a flash drive via the
micro-USB assembly on the front panel of the flow meter, then
imported into a spreadsheet or other viewing application (refer to
Section 2.9 USB Options for more information).
The four event data types are:
Program Report – Tracks changes to the Signature Flow Meter
configuration
Summary Report – Documents summaries of data measure-
ments (e.g., Min/Max/Avg)
Diagnostic Report – Tracks results from diagnostic tests
History Report – Tracks user and meter events (e.g., level
adjustments, data push, etc.)
1.3 Security The Signature Flow Meter can be secured shut with a padlock
with the hasp on the right side of the housing (refer to
Figure 1-3). The program settings and recorded data can be pro-
tected by a user-selected passcode (refer to Section 2.8.2 Set New
Passcode.
1.4 Compatible The Signature Flow Meter can interface with a variety of mea-
Equipment surement devices and other system components, depending on
site requirements.
Measurement devices for flow and water quality can be con-
nected to the same Signature Flow Meter and run simultane-
ously with TIENet® connectivity (up to nine TIENet devices).
The flow meter can also communicate with an optional Teledyne
Isco wastewater sampler and rain gauges.
For descriptions of interfacing and parameter sensing TIENet
devices, refer to Section 5 Equipment Options. Each external
TIENet device comes with its own user manual.
The Signature is capable of receiving data from devices using
Modbus ASCII or Modbus RTU protocol.
A variety of application-specific accessories are available from
Teledyne Isco. Refer to Appendix B Options and Accessories for a
complete list with ordering information.
A basic Signature system has one or more TIENet devices for
flow and/or parameter measurement connected to the Signature
Flow Meter (up to nine TIENet devices at once). Other configura-
tions may include an enclosure and additional internal or
external devices, including analog output cards, analog input
cards, contact output cards, a modem, up to two SDI-12 inputs,
and Modbus devices.
1-3
Signature® Flow Meter
Section 1 Introduction
1-4
Signature® Flow Meter
Section 1 Introduction
1.5 Identifying Signature The following are key components of the Signature Flow Meter:
Components
Injection molded UV protected
sturdy enclosure
Desiccant
Holder
(Used with TIENet 330
Bubbler, TIENet 350
AV Sensors, and 360
Laser Flow)
to bubbler air
supply
Padlock
Hasp
to reference
port
Signature Connection
ports for power,
TIENet devices,
antenna, bubble line
hose barb
1-5
Signature® Flow Meter
Section 1 Introduction
1.6 Controls and Before programming and operating the Signature Flow Meter,
Indicators become familiar with the keypad and standard screens.
1.6.1 Keypad Keys are identified in Figure 1-4; keypad functions are detailed
in Section 2.1.1.
Flow Meter
Total Flow
Status
line [ Status: 10/01/2011 15:20 ] Red/Green
Softkey SHORTCUTS MENU HELP
LED
functions indicator
navigation
Arrow keys (4) for screen
Micro-USB
indicator
Softkeys (3) Home Enter Delete
(Run screen) (Return key) (Undo)
1-6
Signature® Flow Meter
Section 1 Introduction
1.6.2 Display and LED The LED on the front panel is aligned next to the Status line on
the display screen.
• A green light indicates that new information is available
for viewing in the Status line.
• A flashing green light indicates low power mode when
the screen is off and data traffic on some channels.
• A red light indicates a warning or a user-defined alarm
condition, with further details viewable by pressing the
Alarm softkey (C).
Figure 1-4 shows the standard home screen, or run screen.
Displayed menus and programming steps are explained in
Section 2 Setup and Programming.
1.6.3 Backlight and Screen There is an option to turn off the backlight and screen.
1.6.4 Display Contrast The contrast of the LCD screen can be adjusted using the
keypad.
1.7 Technical Table 1-1 provides technical specifications for the Signature Flow
Specifications Meter. Table 1-2 provides technical specifications for the TIENet
330 bubbler module.
1-7
Signature® Flow Meter
Section 1 Introduction
1-8
Signature® Flow Meter
Section 1 Introduction
1-9
Signature® Flow Meter
Section 1 Introduction
1-10
Signature® Flow Meter
Section 1 Introduction
1-11
Signature® Flow Meter
Section 1 Introduction
1-12
Signature® Flow Meter
Section 1 Introduction
1-13
Signature® Flow Meter
Section 1 Introduction
1-14
Signature® Flow Meter
Section 2 Setup and Programming
2.1 User Interface The Signature Flow Meter can be set up, programmed, and inter-
rogated directly through the keypad and display screen, or
remotely using a computer equipped with Teledyne Isco’s
Flowlink® software, with either a USB cable or optional modem.
The Signature Flow Meter has its own browser, accessed via
Flowlink, that mirrors the physical keypad and display.
2.1.1 Keypad Functions The following briefly explains the basic function of each key.
Note
Enter is for selection only. The NEXT softkey is for selection
and advancement to a subsequent step.
2-1
Signature® Flow Meter
Section 2 Setup and Programming
The home key returns the flow meter to the home screen
from any other screen.
The delete key clears the last character entry, exits a pull
down list, or closes an open window.
2.1.2 Connecting to the With Flowlink software version 5.15.XX or later, you can set up,
Signature with program, and download data from the flow meter through its
Flowlink browser. Connection between the flow meter can be direct,
through the micro-USB assembly on the front panel, or remote,
with an internal cellular or ethernet modem.
USB Driver for Signature In order for your computer to connect to the Signature flow meter
through the micro-USB assembly, you must have the correct
driver installed. USB drivers for both 32-bit and 64-bit operating
systems are included with your flowlink program.
To install the driver:
After installing Flowlink, navigate to its program directory, and
then to the USB Driver folder, typically at C:\Program
Files\Flowlink 5.1\USB Driver.
Here you will find two drivers:
4300Driver_x64.msi for 64-bit operating systems and
4300Driver_x86.msi for 32-bit operating systems.
Without the Signature connected to your computer, begin
running the appropriate file for your operating system. When
prompted, connect the Signature to your computer’s micro-USB
assembly. You should see a message in the lower right corner
stating that new hardware has been found at the appropriate
com port number.
Ensure that the Signature flow meter is connected to the com-
puter before launching Flowlink.
Connect window You can connect with the Connect window, or if you have con-
nected with this site before, highlight the Signature site file in
the workspace (left column in Flowlink) and select
Item > Connect.
2-2
Signature® Flow Meter
Section 2 Setup and Programming
2-3
Signature® Flow Meter
Section 2 Setup and Programming
2-4
Signature® Flow Meter
Section 2 Setup and Programming
2-5
Signature® Flow Meter
Section 2 Setup and Programming
2-6
Signature® Flow Meter
Section 2 Setup and Programming
2.2 The Home Screen The home screen, or run screen, is displayed when the flow meter
is in normal operating mode. This screen shows the current
parameter readings and system status or alarm conditions.
A scroll bar on the right of the screen indicates there are more
parameters off-screen that can be viewed by scrolling up or down.
12345678 gal
Flow: 8.55 cfs
Level B: 2.00 ft
pH: 7.00 pH
2.3 Shortcuts The Shortcuts menu provides quicker access to most commonly
used commands, such as level adjustment or viewing data
recorded over a period of time. Not all menu items described in
this section will necessarily appear in your Shortcuts menu. The
selections available in the Shortcuts menu are determined by
what connected devices have been detected by the Signature flow
meter. To access your shortcuts, press SHORTCUTS ( ).
2.3.1 Adjust Level To set a new level, enter the value in the field next to Level, and
select Adjust. To update the current reading, select Update.
2.3.2 Adjust Velocity This selection will open the velocity grid with current readings
and laser controls for the TIENet LaserFlow velocity sensor. For
complete information about this device, refer to the TIENet 360
LaserFlow user manual.
2.3.3 Purge The Signature Bubbler flow meter allows you to manually purge
the bubble line if an obstruction is suspected.
2-7
Signature® Flow Meter
Section 2 Setup and Programming
2.3.5 Real-Time The Real-Time Measurement will display the sensor/card options
Measurement that are available. Once one is selected, a table will be displayed
showing the different types of measurements the sensor/card is
collecting.
2.3.6 Report View Reporting is set up from the Configure Options menu. Summary
displays summaries of data measurements (i.e., Min/Max/Avg).
History tracks user and meter events. Program tracks changes
made to the flow meter’s program configuration.
When you press MENU, the four top menu options appear:
Hardware Setup detects all devices connected to the flow
meter, establishes proper communication with them, and
allows configuration of each device.
Configure Options sets up the measurement site and pro-
gram parameters.
Administration dictates operating preferences and perform
general housekeeping tasks.
Home returns to the home screen.
Additionally,
USB Options appears when a flash drive is connected to
the micro-USB assembly in the lower left corner of the con-
trol panel.
The program menus consist of steps and substeps. During pro-
gramming, available subordinate menu content and steps will be
determined by what you have previously entered, and what
optional equipment is connected to the Signature flow meter.
2.4.1 Off-Screen Content An arrow in the lower right corner of the flow meter’s screen (see
symbols at left) indicates that there is additional content on this
screen in the direction the arrow is pointing. Use the arrow keys
to access this content.
2.4.2 Character Grid A small grid icon in the lower right corner of the flow meter’s
screen (see symbol at left) indicates that the character grid is
available.
Whenever you need to enter characters, such as letters, numbers,
or punctuation, press Enter to display the character grid
(Figure 2-6).
Use the arrow keys to navigate to the desired character and
press Enter to select. When you are finished editing, select DONE
and press Enter.
2-8
Signature® Flow Meter
Section 2 Setup and Programming
I s c o Te s t S i t e
Done Cancel
A B C D E F G H I J K L M N
O P Q R S T U V W X Y Z a b
c d e f g h i j k l m n o p
q r s t u v w x y z / : !
@# $ % ^ & * ( ) - _ + = <
> ? , .
Figure 2-6 Character grid
2.4.3 Pull Down Menus Fields with a pull down arrow next to them (see example at left)
indicate a pull down list. Use the arrow keys to navigate between
Total Flow fields on the screen; when you highlight a pull down field, press
pull down Enter to display the items on the list. Then use the arrow keys
arrow and Enter to select from the list.
2-9
Signature® Flow Meter
Section 2 Setup and Programming
2.5 Program Steps The program steps in Figures 2-7 through 2-22 show the overall
(Menu Trees) menu sequencing.
An explanation of each program step, and information specific to
connected devices, is provided in Sections 2.6 Hardware Setup,
2.7 Configure Options, 2.8 Administration, and 2.9 USB
Options. This information can also be viewed by pressing the
Help softkey on your Signature flow meter or its browser screen.
Menu
1. Hardware Setup
2. Configure Options
3. Administration
4. Home
2-10
Signature® Flow Meter
Section 2 Setup and Programming
2.6 Hardware Setup From the Hardware Setup menus, the flow meter detects and
configures all connected devices.
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
SDI-12 Setup
• Scan for Connected Sondes
• Add/Remove
• Configure/
Manual Config
Modem Setup
Screen content determined by type
of modem installed.
MODBUS Input Setup Refer to Section 5.10.4 Cellular
1. MODBUS Input COM Settings Modem Configuration.
• Set up communication protocol
2. MODBUS Input Device Settings
• Configure or create device(s)
2-11
Signature® Flow Meter
Section 2 Setup and Programming
2.6.1 Smart Sensor Setup This selection will display the most recently detected TIENet
(TIENet) devices connected to the Signature flow meter.
Perform Scan If TIENet devices have been added or removed from the system,
highlight Perform Scan and press Enter to detect the current
system configuration.
Sensor Differences If there are any differences in the device configuration since the
last scan, a list of Sensor Differences will appear.
Missing Sensors – The Missing Sensors list will indicate any
previously connected devices that are no longer detected. Select
Retain to keep the identification information for a previous
device; select Remove to delete it.
Replaced Sensors – The Replaced Sensors list displays any
newly added sensors that have replaced Missing Sensors that
have been Retained.
Additional Sensors – Displays any newly detected devices.
Following a scan, selecting NEXT from Sensor Differences will
navigate to Configure Measurements.
Configure Measurements Navigate to Configure Measurements to begin setting up mea-
surement parameters for each TIENet device detected.
To activate a measurement, highlight the radio button next to it
and press Enter. To change the name of the measurement, high-
light the name and press Enter.
2-12
Signature® Flow Meter
Section 2 Setup and Programming
2-13
Signature® Flow Meter
Section 2 Setup and Programming
2.6.2 SDI-12 Setup Sondes detected since the last scan are displayed, with the acti-
vated sondes in the top box. If SDI-12 devices have been added or
removed from the system, select Scan to detect the current
system configuration. Following the scan, add/remove sondes
from the Active list by selecting the sonde and clicking Add or
Remove. To begin using an Isco-Ready sonde with its configured
parameters, select Configure.
2.6.3 Modbus Input Setup Connect the external Modbus input device at one of the three
TIENet terminal connectors on the Signature case board as
described below:
Modbus In
D1 = Yellow (+)
D0 = Brown (–)
Gnd = Black
The multiplier and offset are used to scale the raw number
coming from the Modbus register(s) to represent the data in the
units of measure you specify, as expressed in the following
equation:
H in units of measure = (register value * J) + K.
2-14
Signature® Flow Meter
Section 2 Setup and Programming
2-15
Signature® Flow Meter
Section 2 Setup and Programming
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
3.
Note
External Modbus RTU devices cannot use addresses 1
through 10.
Add/Edit Parameters Select an existing parameter to edit, or select Add to add a new
parameter for the device. Enter a name (such as a data type), and
the register address. Use the pull down menus to select other
parameters.
2-16
Signature® Flow Meter
Section 2 Setup and Programming
Edit Parameter
Name Flow Rate X H
Address 30001 I
(offset applied to K
Offset -2500 scaled register)
Byte Order LITTLE (little or big) L
2-17
Signature® Flow Meter
Section 2 Setup and Programming
2.6.4 Modbus Output Setup The Modbus RS-485 output function enables a SCADA system to
retrieve site data from the flow meter. Connection to the flow
meter is made via the RS-485 terminal on the Signature case
board (shown in Figure 2-12 below).
In the Device ID field, enter the Signature’s address (from 2 to
247) and configure the communication protocol.
CAUTION
Be careful not to assign the same address to more than one
flow meter.
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
4.
2.6.5 Modem Setup The menu choices displayed for modem setup depend on which
modem option is installed in the flow meter. For detailed infor-
mation about installation and operation of Ethernet, GSM, and
CDMA modems, refer to Sections 5.9 Ethernet Modem and 5.10
Cellular Modems.
2-18
Signature® Flow Meter
Section 2 Setup and Programming
2.7 Configure Options The Configure Options menu is used for setting up the mea-
surement site and setting the program parameters.
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
2.7.1 Site Setup The Site Setup menu sets some basic operating characteristics
specific to the site.
Set Clock – Enter Year, Month, Day, Hour, Minute.
2-19
Signature® Flow Meter
Section 2 Setup and Programming
2.7.2 Measurement Setup This menu is for setting up the level measurement (Level Input
Setup), flow conversion (Flow Input Setup), and flow volume
totalizer(s) (Volume Input Setup). Menu items that appear are
dependent on what equipment is connected to the Signature flow
meter.
Level Input Setup – Under Level Setup, select the level input.
Usually there will only be one listed, unless your system is using
more than one level measurement device.For the TIENet 310
ultrasonic sensor, the minimum blanking distance refers to the
maximum water level, and the maximum blanking distance
refers to zero water level in the channel. For detailed instructions
about 310 setup, refer to the TIENet 310 Installation and Oper-
ation Guide.
For the TIENet 330 bubbler, the purge function is a periodic
burst of air forced through the bubble line to keep it free of
debris. Select the purge interval between 15 minutes and 8 hours
from the pull down menu. –
Flow Rate Input Setup – Measurement settings and flow con-
version are programmed for the flow rate(s) from this menu
(refer to Figure 2-14 on the following page). If more than one flow
rate data set is being calculated, these settings are programmed
separately for each one.
1. Select the flow rate to configure or select "Add Flow Rate"
when two or more flow rates are present. Adding a flow
rate enables calculation of flow rates dependent on condi-
tion. The flow rate also allows for configuring standard
flow conversions from level inputs (e.g. level from Analog
Input).
2-20
Signature® Flow Meter
Section 2 Setup and Programming
Note
Additional information about flow conversions can be found in
the Isco Flow Measurement Handbook included with the Sig-
nature Flow Meter.
2-21
Signature® Flow Meter
Section 2 Setup and Programming
Configure Options
2. Measurement Setup
(example name)
Flow Rate-A
1. Measurement Settings
2. Flow Conversion
1. 3. 4.
7, 8.
Weir Flow Conversion Metering Insert Manning Formula Edit Data Points
Flow Rate-A Flow Conversion Channel Flow Conversion • Level to Flow or Area
1. V-Notch Flow Rate-A 1. Round (Select units of measure,
2. Rectangular w/ End 1. V-Notch 2. U Shaped add/remove points.
Contractions 2. Round Orifice 3. Rectangular 3 to 50 pairs of points
3. Rectangular w/o End (enter size) 4. Trapezoidal may be used.)
Contractions
4. Cipolleti (enter dimensions)
5. Thel-Mar
5.
(enter size)
Area Velocity Channel Flow
2.
Conversion
Flume Flow Conversion
Flow Rate-A 4. Trapezoidal
Flow Rate-A 1. Round 5. Elliptical
1. Parshall 2. U Shaped (enter silt level
2. Palmer-Bowlus 3. Rectangular & dimensions)
3. Leopold-Lagco 6.
4. HS
5. H British Flume Flow Conversion Flow Equation
6. HL 1. Standard Flow Rate-A
7. Trapezoidal 2. Venturi (Select units of measure, enter
(enter size) 3. Khafagi one- or two-term polynomial
8. British Flume equation.)
(enter size)
Volume Input Setup – You can set up one to four Total Flow
measurements. Select the flow rate(s) used for total volume, the
totalizing method (Net, Positive, or Negative), and the interval at
which the total flow will be updated (between 30 seconds and 24
hours).
2-22
Signature® Flow Meter
Section 2 Setup and Programming
From the Resolution pull down menu, select the degree of reso-
lution required for your total flow (lower = fewer digits to right of
decimal; higher = more digits to right of decimal).
2. Measurement Setup
Volume Setup
1. Total Flow
2. Total Flow 2
3. Total Flow 3
4. Total Flow 4
Total Flow
VOLUME
Input Flow Rate: Flow Rate-A
Totalizer: Net
Resolution: 9999999.99
NOTE:
Update Total Flow Every: Effects power
30 Seconds consumption
Figure 2-15 Menu Tree: Volume Input Setup (total flow)
Rainfall Input Setup – This option allows you to setup the per
tip options for the rain gauge. The Automatic roll-over can be set
to the time of day the rollover occurs. Once this number is
reached, the roll-over will go back to zero.
300 Rainfall
SELECT RAINFALL PER TIP
0.01 inch
0.1 millimeter
User Entered 0.00 inches
Automatic roll-over: 00:00:00
2-23
Signature® Flow Meter
Section 2 Setup and Programming
2.7.3 Adjust Adjust levels and/or velocity measurement, and/or calibrate mea-
surement values for other connected TIENet devices. Level
adjustment instructions can be found in Section Setting the
Level, on page 3-18.
Note
For detailed instructions on calibration of a connected TIENet
301 pH device through this menu selection, refer to the 301
user manual.
Note
For detailed instructions on laser velocity measurement setup
through this menu selection, refer to the LaserFlow (360) user
manual.
2-24
Signature® Flow Meter
Section 2 Setup and Programming
2-25
Signature® Flow Meter
Section 2 Setup and Programming
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
Equations Setup
Equation that you have
1. Equation XYZ already created.
2. Create New Equation
1. 2.
A or B ( ) Bksp
AND & OR | NOT !
2.7.5 Data Storage/Push Data Storage – Set up data storage rates for a group of mea-
Setup surements, or separately for individual measurements. Scroll to
the bottom of the screen to set up the primary storage rate, and a
secondary one, if needed, with its trigger equation.
The display can be turned OFF and ON in the configure options,
site setup, and power options menu. Turning the display O FF
effects the way the data is recorded by the Signature meter.
2-26
Signature® Flow Meter
Section 2 Setup and Programming
2.7.6 Sampler Setup Program the flow meter to trigger and pace a sampler, and
receive sampling information from the sampler.
Note
For detailed instructions on configuring a connected TIENet
306 Sampler Interface through this menu selection, refer to the
306 user manual.
2.7.7 Inputs/Outputs/Alarms Alarm – Configure Local, SMS text, or Server alarms based on
Setup user-defined site conditions. Under Alarm, select an alarm from
the list or set up a new alarm.
Next, select an Alarm Trigger from the pull down list.
Note
The trigger(s) listed in the Alarm Trigger pull down list consist
of equations you have already created based on your defined
site conditions (refer to Section 2.7.4 Equation/Trigger Setup).
Alarms: Local Local alarms are viewed on the Signature Flow Meter itself.
When a programmed alarm condition becomes true, the LED on
the front panel glows red.
To view the alarm message, press the Alarm softkey ( ).
For local alarm setup, refer to Figure 2-19 on the following page.
Alarms: SMS / Server SMS and Server alarms require an optional internal modem. To
configure your modem for communication, refer to Section 5.9
Ethernet Modem or Section 5.10 Cellular Modems.
2-27
Signature® Flow Meter
Section 2 Setup and Programming
Note
Server alarms notify a specified list of contacts in the event
that a server running Flowlink Pro fails to receive pushed data
from a site within a specified duration.
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
7.
Inputs/Outputs/Alarms Setup
1. Alarm
2. Analog
3. Input/Output
4. Contact Output
Alarm
1. Alarm 1 Select an alarm from the list or
2. Alarm 2 set up a new alarm.
3. Alarm 3
4. Alarm 4
ALARM 1
ALARM Select a trigger (equation) from
Alarm Trigger: 1 the list. You must have at least
Alarm Notification: one equation saved in order to
Local
trigger an alarm.
Delete Alarm
LOCAL
2-28
Signature® Flow Meter
Section 2 Setup and Programming
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
Inputs/Outputs/Alarms Setup
1. Alarms
2. Analog
3. Input/Output
4. Contact Output
Select a trigger
ALARM (equation) from the list.
ALARM You must have at least
Alarm Trigger: 1 one equation saved in
Alarm Notification:
SMS
order to trigger an
alarm.
SMS SERVER
Phone #5:
2-29
Signature® Flow Meter
Section 2 Setup and Programming
2.7.8 Reset Totalizers Select the flow volume totalizer(s) to be reset. Selection resets
the totalizer to zero.
2.7.9 Reports/History Setup Reports – Set up report interval and measurements to include
for one or two reports. To include all user and meter events in the
report, select the option Include history log.
History – Display user and meter events (i.e., log-ins, adjust-
ments, data push, etc.), and/or set up the graphical display for
selected measurements over a period of time.
Administration Options
1. Language Options 5. Display Signature Information
2. Set New Passcode 6. Display License Information
3. Update Firmware 7. Gather Fault Data / USB Options
4. Sensor Diagnostics 8. Restore To Factory Defaults
Device Information
5. Sensor Information
(Firmware & Serial #s)
Choose the language
for this device.
(Select from list.) Retrieve Diagnostics
(select TIENet device from list)
Set default units of measure
If no current
passcode: Please enter your new Review Diagnostics
passcode. (select TIENet device from list)
New Passcode: ____________
2.
(TIENet Device Name)
(select time from list)
To change Please enter your new
current passcode. License Information
passcode: 6.
Old Passcode: ____________ (Licensing & copyright details)
New Passcode: ___________
2-30
Signature® Flow Meter
Section 2 Setup and Programming
2.8.1 Language Options Select Device Language – From the list, select the default lan-
guage to be displayed by the Signature flow meter. Available lan-
guages include:
From this screen, you can also set general default Units Of
Measure to USA or International. You can still select units of
measure for individual measurements through Configure
Options > Site Setup.
2.8.2 Set New Passcode A numerical passcode can be set to protect access to pro-
g r a m m i n g a nd d a t a . B y d e f a u l t , t h e f l o w m e t e r i s n o t
passcode-protected. To change an existing passcode, enter the
current passcode, and then the new passcode. To remove the
passcode requirement, enter a new passcode of 0 (zero).
2.8.3 Update Firmware To use this function, first connect a flash drive loaded with the
correct firmware update file(s) to the micro-USB assembly. Refer
to Section 2.9 USB Options for complete instructions.
2.8.4 Sensor Diagnostics The Signature provides operating data for each connected
TIENet device upon request, for site evaluation or trouble-
shooting purposes.
To generate a new diagnostic report, select Retrieve Diag-
nostics. The available devices can then be selected from a list.
When the diagnostic data is available, press Enter to view it.
To view past diagnostic reports, select Review Diagnostics and
select from the list of devices. Each past diagnostic report will be
listed by date and time.
Diagnostic reports can also be downloaded to a connected USB
flash drive in the form of a text file. Refer to Retrieve Text
Reports, on page 2-33 for complete information.
Note
If you are running diagnostics on a 360 LaserFlow velocity
sensor, the Distance in the text report and graph are relative to
the face of the integral ultrasonic transducer, which is located
11.7" above the bottom of the sensor. Subtract 11.7" from the
Distance shown to obtain the actual value.
For more information about the LaserFlow sensor, refer to
Section 5.6.3 Laser Doppler Velocity Sensor and the Laser-
Flow user manual.
2.8.5 Display Signature Select this option to view the serial number, model number,
Information software revision, and hardware revision of the Signature flow
meter and any connected TIENet devices.
2.8.6 Display License This selection displays open-source licensing and copyright infor-
Information mation for the Signature firmware.
2-31
Signature® Flow Meter
Section 2 Setup and Programming
2.8.7 Gather Fault Data Fault data is a term describing the capture of any user and/or
flow meter activity prior to and during a system error or failure.
This data can be displayed to assist in troubleshooting. To
download the data to a flash drive, connect a flash drive to the
micro-USB assembly on the front panel of the Signature and
select Gather Fault Data from the USB Options menu that
appears (refer to Section 2.9 USB Options).
2.8.8 Restore to Factory This function returns the Signature Flow Meter to the program
Defaults that was installed the factory. This is an example program that
can be used for reference when designing your own program.
If the restore to portable defaults check box is selected, the power
savings options will be automatically turned on.
Be sure to record your own program settings and save all
data before restoring the factory defaults.
2.9 USB Options The USB Options menu will only appear when you connect a
flash drive to the micro-USB assembly on the front panel of the
Signature.
USB Options
Please choose the operation you want
to perform:
1. Retrieve Text Reports
2. Retrieve Data
3. Update Firmware
4. Save Current Program
5. Load Existing Program
6. Save Signature Information
7. Gather Fault Data
8. None of the above
2-32
Signature® Flow Meter
Section 2 Setup and Programming
2.9.1 Retrieve Text Reports Select this function to download the Signature data text reports
for sensor diagnostics and verification of data integrity.
Select “All reports,” or specify a start date or date range, and
press NEXT. The reports will be stored on the connected flash
drive in a folder called "ISCO."
This folder contains a sub-folder for each site, named with the
first eight characters of the site name. The site folder contains
one or more sub-folders, named by retrieval date (YYYYMMDD).
Each report file name has a prefix of one or two letters, followed
by four digits representing the time of day, e.g., PH0935.TXT is a
program report generated at 9:45 a.m.
Each report is contained in a text file, with a .TXT extension, or
in a Smart Sensor Diagnostic file, with a .SSD extension. Each
report file has a corresponding authentication file, with a .ath
extension, for verification purposes. For more information about
report verification, refer to Verifying Exported Reports, on page
2-35.
The four validation report types are:
Program Report (PH) – Tracks changes to the Signature
Meter’s configuration
Summary Report (R1 / R2) – Documents summaries of data
measurements (e.g. Min/Max/Avg)
Diagnostic Report – Tracks the occurrence of, and results
from, diagnostic tests
History Report (H) – Tracks user and meter events (e.g. level
adjustments, calibration, data push, etc.)
2.9.2 Retrieve Data The program settings and flow data can be downloaded onto your
flash drive in .ddp (data dump) format.
Select “All data,” or specify a start date or date range, and press
NEXT. The data will be stored on the connected flash drive in a
folder called “ISCO.”
This file can then be imported into Flowlink, where it can be
viewed in regular site file format, with the recorded data and
report/graphing capability.
There is a check box that, when checked, will export the .cvs file
onto a USB flash drive.
2.9.3 Update Firmware With a USB flash drive connected to the Signature’s front panel,
the Update Firmware option becomes active on both the Admin-
istration menu and the USP Options menu.
For step-by-step instructions for updating the firmware for either
the Signature flow meter or connected TIENet devices, refer to
Firmware Updates, on page 6-2.
2-33
Signature® Flow Meter
Section 2 Setup and Programming
2.9.4 Save Current Program Select this option to save a copy of the Signature’s current
program to your USB flash drive. You can also use this infor-
mation to program other meters with the same configuration.
2.9.5 Load Existing Program Select this option to load a saved program from your flash drive.
Note that selecting this option will cause the current program to
be overwritten with the one from the flash drive.
In order for the Signature to load the correct program, the name
of the site must match that of the site program that was saved.
2.9.6 Save Signature This option saves a snapshot of the firmware version(s) and
Information serial number(s) of the Signature and any connected TIENet
devices.
2.9.7 Gather Fault Data Fault data is a term describing the capture of any user and/or
flow meter activity prior to and during a system error or failure.
This data can be viewed and/or downloaded as a file to your flash
drive to assist in troubleshooting.
2.9.8 None of the above This item returns the screen to the top menu.
However, as long as the flash drive remains connected, the USB
Options menu will still be active, and can be reopened from the
Administration menu.
2.10 Signature Data in To download flow and event data from the Signature Flow Meter
Flowlink into the database with Flowlink, connect to the flow meter and
select Interrogate (F8), or Import ddp (data dump).
2.10.1 Event Viewer By default, Flowlink’s Event Viewer displays the four event data
types in tabular format, with a time stamp and short description
of the event for each entry (refer to Figure 2-23). Each event type
is represented by a graphical symbol, located in the first column:
Program
Diagnostic
2-34
Signature® Flow Meter
Section 2 Setup and Programming
Printing reports Select one or more rows in the viewer to be printed and then
select the Print button.
Exporting reports To save event data as a text report for future verification, high-
light the desired row(s) in the Event Viewer table and click
Export (or right-click and select Export).
The default destination is your My Documents folder; however,
you can change this to another preferred destination, including a
USB drive, if preferred. A message window will notify you when
the export is complete.
The files you exported are saved using the following hierarchy:
SITENAME \ MODULENAME \ DATE.
2.11 Verifying Exported Flowlink will only export already verified reports; they can also
Reports be verified after being retrieved from the Signature, either via
Flowlink export or USB flash drive download.
Note that this tool can also be used to verify data exported
directly to a USB drive using the USB Options menu (refer to
Section 2.9).
2-35
Signature® Flow Meter
Section 2 Setup and Programming
Data is
authentic.
2-36
Signature® Flow Meter
Section 3 Standard Installation
WARNING
The installation and use of this product may expose you to
hazardous working conditions that can cause serious or
fatal injury. Take all necessary precautions before entering
a worksite. Install and operate this product in accordance
with all applicable safety and health regulations and local
ordinances.
3.1 Introduction The Signature Flow Meter is used for permanent installations.
3.2 Connecting External External device cables and mains line cord are passed, usually
Devices via conduit or cord-grip fittings, through the port holes in the
bottom of the case and wired directly to the connector case. Cable
fittings are also available for Rain, Ethernet and option card cir-
cuits as a 3 hold cord grip.
Trimming unterminated wires prior to installation is recom-
mended.
Tools Required:
Small flat screwdriver (3.5mm)
#2 Phillips screwdriver
Channel locks
Soldering iron (for tinning wires)
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.
CAUTION
Before opening the case, disconnect the optional battery or
battery backup power, if used.
3-1
Signature® Flow Meter
Section 3 Standard Installation
Note
Before restoring mains power, ensure that the flow meter’s
USB connector does not have a cable attached.
Open the door to access the two large screws holding the front
panel on the connector case. Remove the two screws, then
reinsert them in the front panel and latch the lid so they will not
be misplaced.
3-2
Signature® Flow Meter
Section 3 Standard Installation
A External TIENet
Devices/
I
K A Modbus RS485 In
G D1 = Yellow (+)
D0 = Brown (–)
H Gnd = Black
B SDI-12 Input
I J
L Modbus
C RS485 Out
L D TTL Serial
J J Battery Backup/
A A E Battery for Portable
B D E F F 12.8 VDC
C
K K G Ethernet Modem
M
H Ethernet Port
B
Cellular Modem
I (power & serial)
4-20mA Input/Out-
put Card and
J Contact Output
Note: No power supply for Signature Portable Flow Meter
Card
K Fuse "T" 3.15A
M Rain Gauge
Note
The two TIENet connections satisfy most applications. If addi-
tional TIENet connections are needed, consider the two
TIENet Expansion Box versions for standard installations or
TIENet 'Y' cables for Portable Signature applications. See
Appendix B for ordering information.
3.3 Case Bottom Cable The connections made through the cable entries depend on the
Entries application, but their most common uses, in accordance with the
connector case layout, are depicted on the following page.
All optional cable entries must use appropriate ID conduit con-
nections or cord-grip fittings to retain the IP66 rating. If you are
using non-TIENet or non-Signature cables, you must supply the
appropriate ID conduit connections or cord-grip fittings.
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and
3-3
Signature® Flow Meter
Section 3 Standard Installation
TIENet
Cellular Device or Battery
Bubble Line Fitting Wired Backup or Main
Antenna or Power
Wired Options or TIENet Device Options Wired Options
Figure 3-3 Connector Case cable entries for power and external devices
3-4
Signature® Flow Meter
Section 3 Standard Installation
3.3.1 Cable Fittings Cord-grip fittings for TIENet devices, line cord, and battery
backup option are available from Teledyne Isco (see Appendix B
for ordering information). Cable fittings are also available for
rain, ethernet, and optional card circuits as a 3-hole cord grip.
The fitting for the line cord is a special strain-relief fitting, as
shown in Figure 3-5.
Figure 3-4 Strain relief 3/4 NPT Cord-grip fitting for TIENet devices
3-5
Signature® Flow Meter
Section 3 Standard Installation
Any unused cable entry holes should be sealed with plugs. Do not
overtighten the plugs. When a plug is flush against the outside of
the case and held in place by the metal nut inside, the hole is
sealed.
O-Ring
3-6
Signature® Flow Meter
Section 3 Standard Installation
3.3.2 Connecting TIENet The optional external TIENet devices compatible with the Sig-
Devices nature flow meter all connect in the same manner. Multiple
TIENet devices can be connected simultaneously to the same Sig-
nature Flow Meter.
Note
The steps that follow include instructions for installing cord-grip
fittings. Some applications will use user-supplied 3/4" ID con-
duit for cable routing.
3-7
Signature® Flow Meter
Section 3 Standard Installation
Lock Nut
(concave side Sealing Nut
facing wall)
Seal
Cable Nut (color may vary)
Note
The SHIELD wire is the bare drain emerging from the foil
s h i e l d a r o u n d t h e YE L L OW a n d B ROW N w i r e s. T h e
BRAID-DRAIN wire is the bare drain emerging from the sur-
rounding braided shield inside the cable jacket. It is not neces-
sary to prevent the two braids from coming into contact with
each other.
Shield D1 YELLOW
D0 BROWN
Shield SHIELD
Braid-Drain VP RED
Common BLACK
Chassis BRAID DRAIN
3-8
Signature® Flow Meter
Section 3 Standard Installation
Figure 3-11 Insert the cable reference tubing into the case board reference port
3-9
Signature® Flow Meter
Section 3 Standard Installation
6. Close the front panel and fasten it shut with the two Phil-
lips screws.
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit must be sealed to prevent harmful gases and moisture
from entering the Signature enclosure. Failure to seal conduit
could reduce equipment life.
3-10
Signature® Flow Meter
Section 3 Standard Installation
3.4 Power The Signature is in compliance with North American and Inter-
national safety standards while the input voltage remains within
100-240 VAC (50/60 Hz).
For external current protection, a 2 A slow-blow or time-lag fuse
between mains power and the Signature is recommended to
accommodate up to 40 A inrush current at power up in applica-
tions up to 230 VAC.
The flow meter comes with the internal power supply wired to
the connector case, and held in place by a screw (see Figure 3-13
on the following page). Mains power is wired into the Signature’s
internal power supply, normally via a standard three-wire line
cord or hard-wiring through user-supplied conduit.
3.4.1 Hard Wiring Cable entries for hard wiring must use appropriate ID conduit
connections.
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and
moisture from entering the Signature enclosure. Failure to seal
conduit could reduce equipment life.
3.4.2 Line Cord If the instrument has been fitted with a line cord, ensure that its
installation is near a mains outlet for easy access to remove
power in the event of an emergency.
3.4.3 Power Connections Teledyne Isco offers an AC line cord kit that is installed at the
factory when ordered with the Signature. It is also sold sepa-
rately, and is easily installed by the user.
For installation instructions, refer to Section 5.1 AC Power Cord
Kit /AC Wiring (Applies to Permanent Installation Only). Note
that these instructions can also be used as guidance for
user-supplied line cord, hard wiring, or replacement
power supply.
3-11
Signature® Flow Meter
Section 3 Standard Installation
Power Supply
Mounting
Screw
Mains Power
Cord
Connector Case
3.5 Mounting the The Signature can lie flat on a horizontal surface, or be attached
Signature to a wall using the stainless steel bracket on the back of the case.
It can also be installed inside a console enclosure with other
system components. If a console enclosure is used, ensure that it
provided proper sealing to protect the flow meter and other
equipment from harsh environments and/or moisture.
Because it uses a bubble line, the Signature does not have to be
mounted directly above the primary device, or even particularly
close to the flow stream. You will need to mount the unit within
25 feet (7.6 m), or 50 feet (15.3 m) if you are using the 100 foot
bubble line. Distances greater than 100 feet are not recom-
mended.
The mounting location should allow for easy removal and rein-
stallation in the event that cleaning, testing, or replacement is
required. Refer to the dimensional drawings on the following
pages for physical installation specifications.
3-12
Signature® Flow Meter
Section 3 Standard Installation
3.6 Outdoor Where the Signature Flow Meter is mounted outdoors, a weather
Recommendations shield for protection from direct sunlight and rain is recom-
mended. The shield should accommodate the flow meter’s 18 inch
dimensions with the cover opened, as shown below.
FIBERGLASS PRODUCTS
3-13
Signature® Flow Meter
Section 3 Standard Installation
3.7 The Bubble Line Anchor the bubble line in the flow stream at the appropriate
measuring point in the weir, flume, or other open channel flow
situation. Air slowly bubbles out of the line into the flow stream.
The pressure in the bubble line is proportional to the liquid level
in the flow stream, and the flow meter measures this pressure,
sensing the liquid level.
3.7.1 Standard Bubble Lines Three different bubble lines are available for use with the Sig-
nature:
• (1/8" (0.32 cm) OD PTFE, 1/16" (0.17 cm) ID,
25 ft. (7.6 m) long)
• (1/4" (0.63cm) OD vinyl, 1/8" (0.32 cm) ID,
50 ft. (15.2 m) long)
• (1/4" (0.63 cm) OD vinyl, 1/8" (0.32 cm) ID,
100 ft. (30.5 m) long)
3.7.2 Comparing Vinyl and Wherever practical, Teledyne Isco recommends the vinyl line,
PTFE Bubble Lines which offers significant advantages over the PTFE line. The
vinyl line has a longer usable length than the PTFE line. This is
due to the small inside diameter of the PTFE tubing, which gen-
erates an undesirable friction head at lengths greater than 25
feet. Additionally, experience has shown that the larger ID vinyl
line is less likely to clog than the PTFE line when used in flow
streams with suspended solids. If the distance between the flow
meter and the measuring point exceeds 25 feet, you must use the
vinyl bubble line.
However, for certain installations, the smaller ID PTFE line also
has advantages. Due to the small inside diameter, the air volume
necessary (and, as a result, battery power) is minimized, a def-
inite advantage for battery-powered installations. Additionally,
almost no chemical can attack the PTFE line. Consequently, if
power consumption is critical, or there are known agents in the
flow stream that might attack the vinyl line, the PTFE line may
be more suitable.
3.7.3 Bubble Line Length The bubble line should be kept as short as possible. This will
minimize friction head effects in the line and will also minimize
the amount of line exposed to cuts, kinks, etc. Shorten the line as
necessary by cutting the tubing with a sharp knife.
Long Bubble Lines – For certain applications, you may need to
use a bubble line with an inside diameter larger than that of the
standard lines (for example, where the bubble line exceeds 50
feet, or where the flow stream is extremely dirty and the end of
the line might clog).
Consult the factory for specific recommendations regarding size
of line, special connectors required, etc. In no case should the
inside diameter of the bubble line exceed 1/4" (0.64 cm), and you
should recognize that a larger bubble line will result in increased
power consumption, a concern if you must power the flow meter
by battery.
3-14
Signature® Flow Meter
Section 3 Standard Installation
3.7.4 Attaching the PTFE The PTFE bubble line attaches to the flow meter with the bubble
Bubble Line line fitting on the bottom of the case and the silicone rubber
tubing connector. The tubing connector is a short length of 0.109
inch (0.20 cm) ID, 0.192 inch (0.49 cm) OD silicone tubing
attached to a barbed fitting on the case.
The 1/8" OD bubble line simply slips inside the silicone tubing,
forming a reliable union you can join and separate without tools.
Attachment To attach the PTFE bubble line, first slip the 2 inch (5.1 cm)
length of 1/4" (0.63 cm) ID vinyl tube included in the instrument
accessory package over the end of the bubble line. Grasp the sil-
icone tube to stiffen it and insert the end of the bubble line into
the silicone tube.
Slip the short length of vinyl tube over the union and force it over
the shoulder of the barbed fitting. The purpose of the short
length of vinyl tube is to support the union, preventing fatigue or
kinking of the silicone tube.
Removal To remove the bubble line from the tubing connector, first pull
the vinyl tube off of the barbed fitting and slip it down the bubble
line. Grasp the end of the silicone tube and then pull the bubble
line straight out of the silicone tube.
Several replacement lengths of the silicone tubing are included
in the instrument accessory package. Install them on the unit by
simply forcing the end of the tube over the barbed connector.
Note
Avoid placing tension on the silicone tube where it attaches to
the barbed fitting. Direct the PTFE tube straight away from the
fitting rather than at an angle. This reduces the likelihood of
tubing wear and leakage around the fitting.
3.7.5 Attaching the Vinyl The vinyl bubble line attaches directly to the barbed fitting.
Bubble Line Remove the silicone tube and slip the 1/8" (0.32 cm) tubing over
the fitting.
3.7.6 Installing the Bubble Install the bubble line at the recommended level measuring point
Line in the primary device or other open channel flow situation. If you
do not know where this is, consult the manufacturer of the
primary device.
If you are not using a fabricated device, consult the Isco Open
Channel Flow Measurement Handbook for suggestions. Many dif-
ferent devices are discussed there. Proper location of the bubble
line outlet is necessary for accurate measurement. Normally, the
bubble line is positioned in the flow stream with the end at a
right angle to the flow.
In many installations, it is not practical to locate the outlet of the
bubble line precisely at "zero" liquid level. Depending on the situ-
ation, the outlet end of the bubble line may be located anywhere
within ten feet (3 m) above or below the actual zero level of the
primary device. Then set the displayed level using Adjust
Options from the Configure menu. Refer to Section 3.8 Setting
the Level.
3-15
Signature® Flow Meter
Section 3 Standard Installation
Note
The Signature cannot accurately measure liquid levels that are
even with or below the bubble line outlet.
If you need to measure the liquid level down to the actual
“zero” level of the primary device, Teledyne Isco recommends
placing the bubble line outlet at least 1 to 2 inches (2.5 to 5.1
cm) below the primary device zero level to avoid measurement
failures when the liquid level is even with the outlet. Since the
flow meter can display negative measurements, you can com-
pensate when you set the level in the flow meter.
3.7.7 High-Velocity Flow The normal position of the bubble line in the flow stream is at a
Streams right angle to the flow. However, studies have indicated that at
relatively high flow stream velocities (greater than five feet per
second [1.5 meters per second]), this location may lead to a mea-
sured level lower than actual. This negative level offset is due to
localized areas of low pressure near the bubble line outlet; the
size of the offset depends upon the velocity of the flow stream and
the configuration of the flow stream channel. You can, of course,
adjust the level at the flow meter to compensate for the error.
However, this is only effective if the flow rate and velocity are
essentially constant. If the flow rate drops and the velocity
decreases, the level adjustment you made earlier will be incorrect
for the new, lower velocity and flow rate.
3-16
Signature® Flow Meter
Section 3 Standard Installation
3.7.8 Stilling Wells If the installation includes a stilling well in the primary mea-
suring device, installing the bubble line in the stilling well is rec-
ommended. Attach the line securely to the stilling well, using
stainless steel and/or plastic mounting hardware.
Not all stilling wells are suitable for bubble line installation. If
the well is subject to silting or buildup of foreign material, the
bubble line may have to be mounted in the flow stream proper.
3.7.9 Flume Bubble Line The large variety of primary measuring devices and installations
Fittings makes comprehensive bubble line installation instructions
impractical. However, valid general observations on bubble line
installation can be made. Flumes can be specified to include a
bubbler fitting. In new construction, this is highly recommended.
It may even be possible to modify an existing installation to
include a permanent bubbler fitting.
3.7.10 Bubble Line Teledyne Isco offers both stainless steel and copper bubble line
Extensions extensions. The metal extension may be easier to install in the
flow stream than the plastic bubble line because of its rigidity.
Two different extensions are available to match the two standard
bubble lines.
The extension for the PTFE bubble line includes a silicone rubber
tubing connector, and the tubing installation is as described
above. The vinyl bubble line attaches by simply slipping the vinyl
tube over the end of the extension.
3-17
Signature® Flow Meter
Section 3 Standard Installation
Vinyl Bubble
Metal Line Tubing
Extension
3.7.11 Open Channel If you do not use a stilling well, attach the bubble line to the side
Installation of the flow channel or flume. Make the attachment so it causes a
minimum amount of disturbance to the flow stream. If possible,
cut a groove in the side of the channel, place the bubble line in
the groove, and then grout over the groove.
Alternatively, you can attach the bubble line to the side of the
channel, and then grout over the line to form smooth, sloped
lead-in and lead-out surfaces. However, if neither of these
methods is practical, you may simply attach the bubble line to
the side of the channel or the upstream side of a weir using
stainless steel and/or plastic mounting hardware.
In any case, always install the bubble line so it causes as little
disturbance to the flow stream as possible. This usually means
an installation on or adjacent to a channel wall where there is a
condition of stagnant flow. For temporary survey applications,
you can attach the bubble line with waterproof tape or other tem-
porary means.
3.8 Setting the Level Although all other programming steps can be performed off-site,
the liquid level must be set at the measurement site following
installation.
Once the 330 bubble line is installed in the flow stream, or the
310 sensor is installed over the flow stream, measure the present
liquid level and enter this value for Level, under Configure >
Adjust Options. Highlight “Adjust” and press Enter to confirm.
From this screen, you can also update the display to show the
current level of the stream.
3-18
Signature® Flow Meter
Section 3 Standard Installation
Adjust Options
1. Level
Display depends on con-
nected TIENet devices
330 Level
LEVEL ADJUSTMENT
Display current
Last reading: X.XXX ft Update real-time level
Time of last
adjustment: MM/DD/YYYY TT:TT:TT
3.9 TIENet Sensor In applications where the TIENet sensor will operate within a
Installation in a hazardous area, the installation must be performed by trained
Hazardous Location and qualified personnel, according to the installation control
drawing provided in Figure 3-19, and in accordance with local
requirements.
Information about hazloc installation specific to the TIENet 310
Ultrasonic Level Sensor is provided in the sensor’s user manual.
3-19
Signature® Flow Meter
Section 3 Standard Installation
3-20
Signature® Flow Meter
Section 4 Portable Installation
4.1 Introduction The Signature Portable is equipped with one TIENet® receptacle.
In the event a connection must be made directly to the board
refer to the permanent Signature installation connection to
external devices in Section 3.2.
Where additional TIENet devices are deployed with a Signature
Portable, there is an optional TIENet 'Y' connection cable
available to expand the number of connections as needed. See
Appendix B for ordering information.
4.2 Portable Stand The portable stand allows the user to move the Signature to
various locations and without a permanent installation. This Sig-
nature has multiple power options and can be in either in an
upward facing or side facing position (Figure 4-1). The Signature
Portable is shipped from the factory in a side facing position.
Upward facing
Side facing
4-1
Signature® Flow Meter
Section 4 Portable Installation
4.2.1 Adjust Signature to To adjust the Signature to be in the upward facing position:
Upward Position 1. Move the handle into the down position by pulling out the
handle latch pins on each side of the stand (Figure 4-2)
Figure 4-2 Pull out the handle latch pins on each side of
the stand
4-2
Signature® Flow Meter
Section 4 Portable Installation
4-3
Signature® Flow Meter
Section 4 Portable Installation
4.3 Power The Signature Portable is designed to be used with 12 VDC lead
acid batteries.
4.3.1 Power Connections The standard battery connection on a Signature Portable accepts
Teledyne Isco model 948 or 946 lead acid batteries.
Adapting cables can be purchased to connect to a customer sup-
plied deep cycle marine battery.
WARNING
Do not set the portable stand directly on the terminals of a
battery. Serious injury or permanent damage may occur.
4.3.2 Outdoor The Signature Portable can be mounted to the portable stand in
Recommendations a side facing or upward facing position (Figure 4-1).
When the Signature Portable is used in an unprotected envi-
ronment it is recommended the screen is in the side facing
position so it does not collect water.
4.3.3 Battery Installation Common battery options for the Signature Portable:
• 946 lead-acid battery
• 948 lead-acid battery
• Lead-acid battery with solar panel
948 lead-acid battery When using the 948 lead-acid battery the portable stand will sit
directly on top of the battery case, where the battery is enclosed.
There is a notch in the back of the portable stand that will fit
over the power cable from the Signature to the battery. (Figure
4-6).
Handle
4-4
Signature® Flow Meter
Section 4 Portable Installation
When using the 946 lead-acid battery, this battery will fit
directly under the Signature inside of the portable stand (Figure
4-7).
Tab
4-5
Signature® Flow Meter
Section 4 Portable Installation
4.4 Battery Life This section will cover the average battery life as related to the
Expectancy most common applications. For more information about battery
life please consult the factory.
CAUTION
The following section on battery life is for information purposes
only and does not constitute a guarantee of service or war-
ranty of any kind.
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
Equations Setup
1. Equation XYZ
2. Create New Equation
1. 2.
A or B ( ) Bksp
AND & OR | NOT !
4-6
Signature® Flow Meter
Section 4 Portable Installation
4.4.3 Additional
Considerations for
Battery Life
Bubbler Tubing The longer the bubbler and sensor line the more current draw
the meter will have. In most cases the difference in current draw
will not be a factor, but exceptionally long lengths will be a factor.
Reading Retries If the sensor is in a location where reading errors occur, the unit
will attempt to retry until a good reading is taken.
Battery Capacity If the battery is damaged (decreased battery capacity) the Sig-
nature will not run as long as the charts and calculations in this
section suggest. Batteries loose capacity:
• as they age
• as temperatures decreases
• when stored
• discharged below default or recommended levels
4-7
Signature® Flow Meter
Section 4 Portable Installation
4-8
Signature® Flow Meter
Section 4 Portable Installation
4.4.4 Calculating Average If the conditions listed in the average battery life tables and
Current Draw default method do not match your application, you can use the
following procedure to calculate the current draw.
Case Fluke® 87
(or other current-
averaging meter)
Board
00I5
V- V+
mA
A
Black wire
Door White wire
A mA A Com V
+ lead
–
Battery, 12 Volt
+ clip
+
A good quality, adjustable, regulated DC power supply
can be substituted for the 12-volt battery. The power
supply should have at least 3 Amperes output, prefer-
ably more, and capable of overcurrent surges.
Note
Do not attempt this procedure unless you have the proper
equipment available and know how to make electrical mea-
surements.
4-9
Signature® Flow Meter
Section 4 Portable Installation
4.4.5 Battery Life To calculate battery life expectancy for an installation, you must
Calculations know two things:
• The capacity of the battery you are using
• The average current draw of the flow meter or (other
device) powered
Battery capacity is expressed in ampere-hours. The battery man-
ufacturer provides this information for each battery. This value is
the product of a load current times an arbitrary time period,
twenty hours for lead-acid types. Isco 946 batteries lead acid bat-
teries are rated for 6.5 ampere-hours.
To determine battery life for a Signature running on a 946 lead
acid battery, convert the battery capacity into milliam-
peres/hours and then divide the ma/hrs by the avg current draw.
This will give you a number in hours. Divide that figure by 24,
and you will have the number of days.
Note
The published ampere-hour figures do not mean that you can
expect to draw 6.5 amperes from the lead-acid battery for one
hour.
Note
Always operate these batteries with a reserve factor.
4-10
Signature® Flow Meter
Section 4 Portable Installation
4.4.6 Low Battery Cut Off Batteries are considered fully discharged well before the ter-
and Battery Care minal voltage drops to zero volts. Operating lead acid batteries
below the fully discharged point decreases their capacity and
damages the battery. Lead acid batteries, under a constant rate
of discharge, are considered fully discharged at 11.5 volts,
however; the Signature does not discharge the batteries at a con-
stant rate.
Time durations will allow you to get the maximum amount of life
and readings out of the battery without damaging it. It can also
prevent a short term power dip from causing a premature
shutdown. By default, the Signature will turn itself off when the
battery voltage is at 11.5v for 2 min. To prevent the unit from
cycling OFF and ON the unit will not recover (turn back ON) until
the battery voltage is 13.0 for 2 minutes.
WARNING
Adjusting the low battery cutoff improperly will damage the
batteries. Teledyne Isco assumes no liability for damage
done to Teledyne Isco or third party vendors batteries.
Adjust the low battery cut off at your own risk. User
assumes all liability for damage done to batteries by
altering the low battery cut off.
4-11
Signature® Flow Meter
Section 4 Portable Installation
A External TIENet
Devices/
H
K F A Modbus RS485 In
D1 = Yellow (+)
D0 = Brown (–)
G Gnd = Black
B SDI-12 Input
H I
K Modbus
C RS485 Out
K D TTL Serial
I I Battery Backup/
A A E Battery for Portable
B D E F Ethernet Modem
C
G Ethernet Port
L J J
Cellular Modem
H (power & serial)
B
4-20mA
Input/Output Card
I and
Contact Output
Note: No 12.8VDC power supply for Signature Portable Card
J Fuse "T" 3.15A
L Rain Gauge
4.5.1 Connecting Devices to The optional external TIENet devices compatible with the Sig-
the TIENet Receptacle nature Portable all scan in the hardware in the same manner.
Multiple TIENet devices can be connected simultaneously to the
same Signature Flow Meter. The following TIENet smart
sensors/cables will attach to the TIENet receptacle:
• Ultrasonic Level Sensor
• Area Velocity Sensor
• Bubbler
• pH/Temperature
• LaserFlow
• Sampler
4-12
Signature® Flow Meter
Section 4 Portable Installation
Connecting a TIENet To connect the TIENet plug from the sensor to the TIENet Recep-
receptacle to the Signature tacle:
Portable 1. Align the connectors and push together (Figure 4-12).
2. After the physical connection is made, a scan must be per-
formed (see section 2.6.1) for the device to be recognized.
For additional TIENet connections, use the TIENet Y-cable or
alternately an expansion box.
O-Ring and Lubrication 1. Coat the O-ring’s sealing surface with a silicone lubricant.
for the TIENet receptacle
CAUTION
Do not use petroleum-based lubricants. Petroleum-based
lubricants will cause the O-ring to swell and eventually deterio-
rate. Aerosol silicone lubricant sprays often use petro-
leum-based propellents. If you are using an aerosol spray,
allow a few minutes for the propellent to evaporate before pro-
ceeding.
Three-Hole Cord Grip The following device wires/cables are normally connected thru
the 3-hole cord grip:
• Rain Gauge
• SDI-12
• Modbus output
The wire/cable diameter can be .200 in to .230 in.
Installing a Cable Through 1. Remove plug to open hole for an additional cable.
a Three-Hole Cord Grip
Note
Gland nut must be loose to remove this plug.
4-13
Signature® Flow Meter
Section 4 Portable Installation
Note
When replacing the plug, the rounded side needs to be pushed
into the hole, with the flange facing out (see figure below).
2. Thread the cable through the open hole and various con-
nectors.
4-14
Signature® Flow Meter
Section 4 Portable Installation
CAUTION
Drilling holes in the Signature cases is not recommended and
may result in water damage. The material of the Signature
case is constructed of Noryl and does not lend itself to post
molding alterations.
Connecting the Rain The following shows how to connect the rain gauge to the Sig-
Gauge nature Portable and how the cables should look when properly
connected (Figure 4-14).
1. Remove the rain gauge cap and connect it to the TIENet
receptacle.
4-15
Signature® Flow Meter
Section 4 Portable Installation
4-16
Signature® Flow Meter
Section 5 Equipment Options
Note
Installation and operation of exterior TIENet devices is covered
in detail in the user manual for that technology.
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit and any
battery power is disconnected.
AC Power Cord Kit /AC Wiring (Applies to Permanent Installation Only), on page 5-2
Battery Backup (Applies to Permanent Installation Only), on page 5-4
Mechanical Totalizer, on page 5-8 Sampler Interface, on page 5-15
External Desiccator, on page 5-12 TIENet Expansion Box, on page 5-16
Ultrasonic Level Sensor, on page 5-13 Reference Port Tubing Kit, on page 5-29
Bubbler Level Sensor, on page 5-14 and 6-19 Ethernet Modem, on page 5-31
Contact Output Card (TIENet 304), on page 5-17 Cellular Modems, on page 5-36
Analog Input Card (TIENet 307), on page 5-22 Laser Doppler Velocity Sensor, on page 5-14
Analog Output Card (TIENet 308), on page 5-26 Continuous Wave Doppler Velocity Sensor, on page 5-14
pH and Temperature Device, on page 5-16
Sampler Interface, on page 5-15
pH and Temperature Device, on page 5-16
5-1
Signature® Flow Meter
Section 5 Equipment Options
5.1 AC Power Cord Kit /AC The AC power cord kit includes a line cord with a strain relief
Wiring (Applies to cord-grip fitting. If ordered with the Signature Flow Meter, it will
Permanent be shipped from the factory already installed.
Installation Only) Instructions for user installation are provided in this section.
Note that these instructions can also be used as guidance
for user-supplied or replacement line cord, hard wiring,
or replacement power supply.
Open the flow meter housing, following all warnings and instruc-
tions provided in Connecting External Devices, on page 3-1.
1. If a cord-grip fitting is already installed, loosen the sealing
nut on the AC line cord. If conduit is installed, remove the
sealing material around the AC line cord where it enters
the housing. This is to free the cabling for movement or
removal.
Lock Nut
Sealing Nut
Cable Nut
& O-Ring
2. Remove the mounting screw and lift the power supply out
of its molded niche, taking care not to strain the wires
going to the board.
Power Supply
Mounting
Screw
AC Line Cord
Connector Case
5-2
Signature® Flow Meter
Section 5 Equipment Options
Plastic shield
5-3
Signature® Flow Meter
Section 5 Equipment Options
Note
Double-check to ensure that the terminals labeled Gnd, -V,
and +V on the Signature circuit board are wired to their corre-
sponding terminals on the power supply.
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires in the conduit must be sealed to prevent harm-
ful gases and moisture from entering the Signature enclosure.
Failure to seal conduit could reduce equipment life.
10. Close the front panel and fasten it shut with the two Phil-
lips screws.
5.2 Battery Backup The battery backup option consists of a Teledyne Isco Model 946
(Applies to Permanent lead-acid battery pack and extension cable, with special
Installation Only) hardware to mount it on the top of the Signature Flow Meter, or
on a wall. The unterminated power cable normally enters the
connector case through the second port from the right.
Note
An optional external power loss alarm is available. See Section
5.3 Power Loss Alarm.
5-4
Signature® Flow Meter
Section 5 Equipment Options
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.
CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.
CAUTION
Do not substitute another battery type for this option. Use only
the Model 946 Lead-Acid battery.
T15 Torx
screws (3)
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and
moisture from entering the Signature enclosure. Failure to seal
conduit could reduce equipment life.
Installation 1. Remove line power from the Signature Flow Meter and
open the case as previously described in Section 3.2.
CAUTION
Do not connect the extension cable to the battery cable until all
other steps are completed.
5-5
Signature® Flow Meter
Section 5 Equipment Options
Mounting screw
holes
5-6
Signature® Flow Meter
Section 5 Equipment Options
Note
Be sure to unplug the battery when intentionally disconnecting
from AC power.
5.3 Power Loss Alarm The Signature Flow Meter offers different options to notify you of
line power loss.
Contact Output Card The Contact Output Card is an optional card that closes or opens
(TIENet 304) relays at specific thresholds set within the Signature Flow Meter
programming. For power loss indication, it is recommended to
configure the Signature to activate the contact output to CLOSE
when the 300 Sense Voltage drops below 12 VDC. The output
relay will deactivate when the 300 Sense Voltage drops below 12
VDC or if the power drops all together.
Analog Output Card The Analog Output Card provides an analog output signal pro-
(TIENet 308) portional to a configured measurement input. For power loss
indication, it is recommended to configure the Signature so the
analog output is proportional to the 300 Sense Voltage. The
external device monitoring the analog output should alarm when
the analog signal drops below the equivalent of 12 VDC reported
by 300 Sense Voltage.
12 Volt DC Output The Signature system operates on 12 VDC. A voltage level of 12
VDC at the POWER 12.8V terminal (Figure 5-6) will indicate
adequate power supplied to the Signature meter, whereas a
voltage level of less than 12 VDC will indicate a power loss.
Signature Power Loss This device energizes a relay based on the Signature operating
Alarm Box (60-5324-024) voltage of 12 VDC. When the 12 VDC supply drops below 9 VDC
the relay becomes de-energized and switches the relay contact
indicating a power loss. It has NORMALLY OPEN and
NORMALLY CLOSED outputs.
5-7
Signature® Flow Meter
Section 5 Equipment Options
5-8
Signature® Flow Meter
Section 5 Equipment Options
Intake
Connecto
Air ce
Humidity
ren
fe
Re
r
Outlet
330 Mounting
Screws (4)
Note
Be sure to cut all six tabs. Pulling on the cover with some of the
tabs still attached will damage the control panel label.
5-9
Signature® Flow Meter
Section 5 Equipment Options
5-10
Signature® Flow Meter
Section 5 Equipment Options
Before installation
After installation
5-11
Signature® Flow Meter
Section 5 Equipment Options
5.5 External Desiccator For Signature systems using the 330 or 360 Bubbler Module
and/or the TIENet 350 Area Velocity Sensor, the desiccator vents
the reference port for a pressure transducer, and the air intake
port for the bubbler system air pump, keeping the interior of the
flow meter case dry, as well as the sensor reference line.
Note
The desiccant is standard on the portable version of the 330,
350, and 360 Bubbler module. It is not standard on the
non-portable 310 Bubbler module.
To air intake
To reference port
Remove the two red protective end caps from the ports before
installing a new cartridge.
5-12
Signature® Flow Meter
Section 5 Equipment Options
5.6 TIENet® Devices Teledyne Isco’s proprietary TIENet connectivity allows for the
combination of multiple flow measurement technologies and
other devices with the Signature flow meter.
5.6.1 Ultrasonic Level The TIENet 310 Ultrasonic Level Sensor mounts directly over
Sensor the flow stream. The sensor measures level by transmitting an
ultrasonic pulse toward the liquid surface and then measuring
the time it takes for the echo to return. The 310 is normally used
with some type of primary device (typically a weir or flume) to
measure flow in an open channel.
The 310 Sensor is ATEX-approved for use in potentially
explosive atmospheres when specific conditions are met. Refer to
the 310 user manual, as well as the control installation drawing
in Section 3.9.
For complete installation and operation procedures, refer to the
TIENet 310 sensor’s user manual.
5-13
Signature® Flow Meter
Section 5 Equipment Options
5.6.2 Bubbler Level Sensor The factory-installed TIENet 330 Bubbler is normally used with
some type of primary device (typically a weir or flume) to
measure flow in an open channel.
The amount of pressure required to force bubbles from the end of
a submerged bubble line is directly dependent on the hydrostatic
pressure of the flow stream over the end of the bubble line. A
pressure transducer inside the module senses this pressure and
converts it into a level signal that the flow meter uses to cal-
culate flow rate and total flow.
In order to operate with the 330 Bubbler, the Signature must
have an external desiccator installed. For installation of the
external desiccator, refer to Section 5.5.
Because the 330 Bubbler is a standard component in bubbler Sig-
nature meters, installation instructions are located in Section 6
Maintenance and Servicing, under 330 Bubbler Installation.
5.6.3 Laser Doppler Velocity The TIENet 360 LaserFlow™ velocity sensor remotely measures
Sensor flow in open channels with non-contact Laser Doppler Velocity
technology and non-contact Ultrasonic Level technology. The
sensor uses advanced technology to measure velocity with a laser
beam at single or multiple points below the surface of the waste-
water stream.
To operate with the LaserFlow, the Signature requires firmware
version 1.18 or later. Firmware update instructions are provided
in Section 6.3 Firmware Updates.
For complete installation and operation procedures, refer to the
LaserFlow sensor’s user manual.
5.6.4 Continuous Wave The TIENet 350 Area Velocity Sensor measures flow stream
Doppler Velocity average area velocity and liquid level. The Signature uses this
Sensor information to calculate the flow rate and total flow of the
stream. To operate with the 350 sensor, the Signature requires
firmware version 1.18 or later. Firmware update instructions are
provided in Section 6.3 Firmware Updates.
The sensor is mounted in the flow stream, normally at the
bottom of the channel. It measures average velocity using con-
tinuous ultrasonic sound waves to produce a Doppler effect. The
sensor measures liquid level using an internal differential
pressure transducer.
In order to operate with the 350 AV sensor, the Signature must
have an external desiccator installed (see Section 5.5 External
Desiccator). Signature bubbler systems will already have a desic-
cator installed. If you are adding a 350 AV sensor to a
non-bubbler system, you will also need to add an external desic-
cator.
For complete sensor installation and operation procedures, refer
to the TIENet 350 sensor’s user manual.
5-14
Signature® Flow Meter
Section 5 Equipment Options
Optimal drying power Some Signature flow meters have a single piece of tubing
installed between the reference port and the humidity connector,
and a cap plug on the intake port. While a Signature with this
tubing configuration will operate satisfactorily with the 350 AV
sensor in most situations, you can configure the tubing to utilize
both chambers of the external desiccator to increase drying
power and extend the desiccant service interval.
Items required:
• Plastic ‘Y’ Fitting (Part #209-0167-49)
• 0.25 x 0.125 silicone tubing (2 pcs, Part #029-1353-02)
Procedure:
1. Remove the cap plug.
2. Disconnect the tubing from the humidity connector and
reroute it behind the ribbon cable.
3. Connect both the reference port tubing and the intake port
tubing to the ‘Y’ connector.
4. Connect the ‘Y’ connector to the humidity connector
(Figure 5-15).
Intake port
tubing
Reference port
tubing
Humidity connector
tubing
5.6.5 Sampler Interface The TIENet 306 Sampler Interface connects the Signature Flow
Meter to a Teledyne Isco wastewater sampler. Through this con-
nection, the Signature can enable the sampler based on
user-specified conditions, pace the sampling routine based on
flow volume, and receive sample and bottle information from the
sampler.
For complete installation and operation procedures, refer to the
TIENet 306 device’s user manual.
5-15
Signature® Flow Meter
Section 5 Equipment Options
5.6.6 pH and Temperature The TIENet 301 pH sensor measures the acidity or alkalinity of
Device an aqueous solution by determining the relative quantity of dis-
sociated hydrogen ions in the solution. The normal scale for pH
runs from 0 to 14, with 0 being most acidic and 14 being the most
alkaline.
5.6.7 TIENet Expansion Box The water-tight expansion box connects to a TIENet terminal
strip like other TIENet devices, and contains three additional
strips inside, for connecting more devices. The expansion box can
be daisy-chained with each box providing two additional cord
grip ports. Additionally, the expansion box contains a TIENet
connection for an option card (such as the 308 Analog Output
option, Section 5.6.9). If an option card is used in an expansion
box there is room for just one TIENet cable cord grip exiting the
box.
Note
For applications requiring reference air moisture protection,
please contact the factory.
Note
The expansion box is not rated for use in hazardous locations.
5-16
Signature® Flow Meter
Section 5 Equipment Options
Enclosure Rating:
IP67 (NEMA4X, 6)
Desiccant bag
TIENet device
terminals
5.6.8 Contact Output Card The contact output card provides a contact closure that can be
(TIENet 304) enabled based off of conditions set in the equation builder.
The Signature accepts up to three internal, user-installed
TIENet option cards. The 304 contact output card provides two
contact outputs per card for connection between the Signature
meter and non-Isco process control equipment that requires a
contact output.
WARNING
The 304 contact output card is not recommended for
voltages above 60vDC or 48vAC.
5-17
Signature® Flow Meter
Section 5 Equipment Options
CAUTION
Use proper static dissipation when handling circuit boards.
Contact 1
Contact 2
CH2
Back of card with Normally Open (NO),
Common (C), and Normally Closed (NC)
Figure 5-17 304 contact output card front and back view
Note
• Normally Open (NO): When the condition is met, the circuit
will be closed.
• Common (C): There is always be a connection to the com-
mon.
• Normally Closed (NC): When the condition is met, the circuit
will be closed.
5-18
Signature® Flow Meter
Section 5 Equipment Options
Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.
5-19
Signature® Flow Meter
Section 5 Equipment Options
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
With initial installation, begin by per-
forming a hardware scan to add the 304.
Configure Measurements
1 - XXX Press Enter for a list of sensors.
Scroll to the 304 and press
XXX Parameter XXX Parameter Enter to select.
XXX Parameter
5-20
Signature® Flow Meter
Section 5 Equipment Options
Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
Inputs/Outputs/Alarms Setup
1. Alarms
2. Digital Outputs
5-21
Signature® Flow Meter
Section 5 Equipment Options
5.6.9 Analog Input Card The 307 analog input card allows the Signature to record an
(TIENet 307) analog signal as a number of data types from several different
units.
The Signature accepts up to three internal, user-installed
TIENet 307 option cards. The 307 analog input card provides two
4-20mA inputs a piece for connection between the Signature
meter and non-Isco process control equipment or other
equipment that outputs a 4-20mA current signal.
CAUTION
Use proper static dissipation when handling circuit boards.
Channel 1 Channel 2
Back of board
CH2
5-22
Signature® Flow Meter
Section 5 Equipment Options
Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.
5-23
Signature® Flow Meter
Section 5 Equipment Options
Hardware Setup
1. Smart Sensor Setup (TIENet) With initial installation, begin by per-
2. SDI-12 Setup forming a hardware scan to add the 307.
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
XXX Parameter
5-24
Signature® Flow Meter
Section 5 Equipment Options
Input Options
1. Alarms
2. Analog
5-25
Signature® Flow Meter
Section 5 Equipment Options
5.6.10 Analog Output Card The 307 analog output cards convert digital information from the
(TIENet 308) flowmeter to a variable analog output current ranging from 4 to 20
milliamperes. When a parameter measured by the flow meter is
converted into an analog output, 4 mA becomes the 0%, or
baseline, for the parameter, while 20 mA becomes the 100%, or
full-scale, of the parameter. For basic programming steps, refer to
Figures 5-18 and 5-19.
The Signature accepts up to three internal, user-installed
TIENet option cards. The 308 analog output card provides two
scalable 4-20mA outputs per card for connection between the Si
nature meter and non-Isco process control equipment or other
equipment that accepts a 4-20mA current signal.
CAUTION
Use proper static dissipation when handling circuit boards.
Channel 1
Channel 2
Back of board
CH2
5-26
Signature® Flow Meter
Section 5 Equipment Options
Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
Configure Measurements
1 - XXX Press Enter for a list of sensors. Scroll
to the 308 and press Enter to select.
XXX Parameter XXX Parameter
XXX Parameter
Configure Measurements
<serial #> Analog Inout Scroll with arrow keys to highlight /
select / deselect any displayed
CH1 Current - A CH1 Percent - A parameter or edit its name.
CH2 Current - A CH2 Percent - A
5-27
Signature® Flow Meter
Section 5 Equipment Options
Output Options
1. Alarms
2. Analog
To edit units of
CH1 Percent measure, high-
ANALOG OUTPUT light and press
Enter. Percent
Measurement Input: 330 Level (example)
and humidity
Low Limit: 0 feet (ft)
High Limit: 0
Channel Name: <serial # of 308 card> Channel #
5-28
Signature® Flow Meter
Section 5 Equipment Options
5.7 Reference Port Tubing For Signature Bubbler monitoring sites where the flow meter
Kit mounting location does not provide adequate reference to atmo-
spheric pressure at the measuring point, the reference port
tubing kit includes tubing and adaptors to relocate the reference
port.
An extra kit can also be used for the bubbler intake, if the air at
the flow meter mounting location is excessively humid.
The 25-foot, vinyl 1/4 " ID, 3/8 " OD tubing should be cut to the
shortest length practical for your installation.
Installation The tubing attaches to the reference port of the external desic-
cator. If using two kits, the second tube attaches in the same
manner to the intake port of the external desiccator.
To install:
1. Unscrew the hydrophobic filter cap from the reference
chamber (smaller chamber) of the external desiccator.
5-29
Signature® Flow Meter
Section 5 Equipment Options
4. Push the open end of the tubing down over the hose barb
fitting.
5-30
Signature® Flow Meter
Section 5 Equipment Options
5.8 Isco Flowlink Software Flowlink® is Teledyne Isco’s proprietary software system for
data acquisition, storage, retrieval, and analysis. Using the
interface of Microsoft Windows, Flowlink can be used to remotely
program the Signature Flow Meter, retrieve data from the flow
monitoring system, present site data graphically, and generate
statistical information from the site data.
Flowlink helps ensure data integrity by displaying the Sig-
nature’s tracked configuration changes, data measurement sum-
maries, diagnostic test results, and user events in the program.
With these tools, Flowlink provides assurance that the data has
not been altered.
USB drivers for computer direct connection to the Signature
Flow Meter are included on the Flowlink CD, and must be loaded
prior to direct connection between the computer and the Sig-
nature.
See Section Connecting to the Signature with Flowlink, on page
2-2 for instructions on how to connect to the Signature meter
with Flowlink software.
From Flowlink, the event data can be exported and saved in the
form of text reports on your computer, searchable by site name,
m o d u l e, a n d d a t e. Fo r c o m p l e t e i n f o r m a t i o n , r e f e r t o
Section Signature Data in Flowlink, on page 2-34.
5.9 Ethernet Modem Setup and data retrieval, as well as alarm output configuration,
can be accomplished remotely via TCP/IP communication pro-
tocol with a static address, using Flowlink software and the eth-
ernet modem to access the Signature’s web browser. The ethernet
modem is factory-installed on the connector case.
The ethernet modem can also be installed by the user. Remove
line and/or optional battery power from the Signature Flow
Meter and open the case as previously described in Section 3.2.
CAUTION
Always use proper static dissipation methods when handling
circuit boards.
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.
CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.
5-31
Signature® Flow Meter
Section 5 Equipment Options
Ethernet
modem
Mounting
hardware
5-32
Signature® Flow Meter
Section 5 Equipment Options
5-33
Signature® Flow Meter
Section 5 Equipment Options
5.9.1 Ethernet Modem When installation is complete and power restored, wait one
Configuration minute for the Signature to recognize the modem before pro-
gramming.
Note
The Signature does not support Dynamic Host Configuration
Protocol (DHCP). The network communication information (IP,
gateway, and subnet mask) must come from your network
administrator and be entered manually into the flow meter.
When you select Modem Setup from the Hardware Setup menu,
the type of modem installed determines what screen is displayed.
Figure 5-34 Ethernet modem setup: Communication settings (default settings shown)
5-34
Signature® Flow Meter
Section 5 Equipment Options
5.9.2 Network Firewall In order for your network administrator to identify the Signature
Settings in the network firewall setup, it must have a node ID (also known
as the MAC address). This is the NODE ID printed on the eth-
ernet modem’s serial tag (refer to Figure 5-35).
Figure 5-35 Locating the NODE ID (MAC address) on the ethernet modem
5-35
Signature® Flow Meter
Section 5 Equipment Options
5.10 Cellular Modems Setup and data retrieval through the Signature’s web browser, as
well as alarm outputs, can be accomplished remotely with one of
the available cellular modems. The whip-style antenna has a
magnetic mounting base.
5.10.1 CDMA Modem The Code Division Multiple Access (CDMA) modem can automat-
ically push data to a secure server running Isco Flowlink Pro
software, with 1xRTT data transmission.
5-36
Signature® Flow Meter
Section 5 Equipment Options
5.10.2 GSM Modem The Global System Mobile (GSM) modem can automatically push
data to a secure server running Isco Flowlink Pro software, with
GPRS data transmission.
Your service parameters, or provider, can be changed by
replacing the removable Subscriber Information Module (SIM)
card in your modem.
Note
The modems shown do not necessary represent the modems
installed.
5.10.3 Installing the Cellular The modem kit includes the modem, power cable, DB9 serial
Modem cable, and coaxial antenna plug cable.
Note
Before installing the modem, remove the top label (with the
FCC ID on it) taped to the modem and adhere it to the outside
of the Signature case on the bottom of the unit, on the left side,
in the largest of the three sections (Figure 5-39). This is
required by the FCC.
FCC ID Label
Label on
bottom of
Sigature case
5-37
Signature® Flow Meter
Section 5 Equipment Options
Antenna Plug
Cable
Serial Cable
Power Cable
Modem
2. Install the plug in the preferred port (far left most com-
monly used). Route the antenna plug cable under any other
cabling, and install the plug in any open port.
3. Connect the three cables to the modem.
4. Remove the screw retainers and fasten the modem’s
mounting bracket against the connector case, as shown
below, using the two mounting screws.
5. Plug the serial and power cables into their respective con-
nectors on the board.
5-38
Signature® Flow Meter
Section 5 Equipment Options
5-39
Signature® Flow Meter
Section 5 Equipment Options
5.10.4 Cellular Modem When installation is complete and power restored, wait one
Configuration minute for the Signature to recognize the modem before pro-
gramming.
When you select Modem Setup from the Hardware Setup menu,
the type of modem installed determines what screen is displayed.
To set up the Signature Flow Meter for cellular communication,
select either internet (IP) or CSD for the connection.
For the GSM modem, if an Access Point Name (APN) is used,
select SINGLE for the operating mode and enter the name.
For portable applications the modem can be set up with the call
in the time window. To conserve power, the time window does not
effect the push/outgoing connections. The time window restricts
when the meter can be acceded via a modem connection.
Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
GSM CDMA
5-40
Signature® Flow Meter
Section 6 Maintenance and Servicing
6-1
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.2 Cleaning The Signature flow meter may be cleaned with water and a mild
detergent. For hard to remove stains, isopropyl alcohol may be
used. If the instrument is in an isolated area and the case is
sealed closed, it may be cleaned using a water hose.
6.3 Firmware Updates Signature and TIENet device firmware updates are provided in
the form of .bin files, which will be available for download from
the Teledyne Isco website. Note that firmware updates do not
remove any program settings or delete data.
To install an update:
1. Create a folder in the top directory of a flash drive, and
name it BINFILE.
2. Download the .bin file to be installed. To find your correct
.bin file, go to www.isco.com and click on Software/Firm-
ware Updates in the lower left corner. Select Open Channel
Flow Measurement.
3. Save the new .bin file(s) to the BINFILE folder you created
on the flash drive.
6-2
Signature® Flow Meter
Section 6 Maintenance and Servicing
4. Using the flash drive adaptor cable provided with the Sig-
nature, connect the flash drive to the micro-USB assembly
on the flow meter’s front panel.
Figure 6-2 USB Micro adaptor cable (flash drive not included)
6-3
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.4 Accessing the Interior Some maintenance or servicing tasks require opening the Sig-
nature housing to access the interior. Always refer to this section
prior to doing so.
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.
CAUTION
Before opening the case, disconnect the optional battery
backup power, if used (refer to Figure 5-8 Backup battery,
installed).
Note
Before restoring mains power, ensure that the flow meter’s
USB connector does not have a cable attached.
Open the door to access the two large screws holding the front
panel on the connector case. Remove the two screws, then
reinsert them in the front panel and latch the lid so they will not
be misplaced.
6-4
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.5 Desiccant The inside of the flow meter housing must be kept dry at all
times to prevent moisture damage to the internal components.
All Signature flow meters have an internal desiccant bag to
absorb moisture. Signature flow meters using a 330 bubbler also
require an external desiccator.
If increased humidity is indicated by either the humidity reading
of the flow meter or the color of the external desiccant, the des-
iccant must be renewed or replaced before damage occurs.
If this occurs more frequently than expected, inspect the seals of
cord-grip fittings and conduit, if used.
Humidity alarm The humidity of the case interior, reference (ambient) air, and
bubble intake air (if a 330 bubbler is installed) are all param-
eters that can be selected as conditions to trigger an alarm, noti-
fying you when it is time to renew or replace your desiccant.
The suggested alarm setting is a threshold condition of 40%. For
detailed instructions about setting up conditions and alarms,
refer to Sections Equation/Trigger Setup, on page 2-24, and
Inputs/Outputs/Alarms Setup, on page 2-27.
6.5.1 Internal Desiccator Saturated internal desiccant bags must be replaced; unlike the
external desiccant, they are not renewable.
The desiccant bag is held in place by a metal bracket. Remove
the two screws holding the bracket.
6-5
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.5.2 External Desiccator The desiccator vents the reference port for a pressure transducer,
and the air intake port for the bubbler system air pump, keeping
the interior of the flow meter case dry.
To air intake
To reference port
When dry, the loose silica gel desiccant inside the chambers is
orange or yellow. When the desiccant becomes saturated with
moisture, it turns green or blue, indicating that the intake air
and reference line are no longer protected from humidity.
Note
Teledyne Isco recommends checking the desiccant at least every
6 months, and changing/renewing the desiccant before the entire
compartment has changed color.
Dry desiccant
(good)
Saturated
desiccant (bad)
6-6
Signature® Flow Meter
Section 6 Maintenance and Servicing
Unscrew the two black caps and carefully pour the desiccant out.
If removal is difficult, screw the caps back in and unscrew again.
Gently knock the caps and the cartridge against a hard surface to
free any small particles in the threads, as these can hinder
proper sealing and cause wear.
Using a funnel, fill both chambers with dry desiccant, replace the
caps, ensuring that they are fully engaged. Press the cartridge
back into place on the side of the flow meter.
Renewing loose desiccant To renew the desiccant, spread it in a single layer on a flat metal
tray. Place in a vented, circulating forced air, conventional oven
in a well ventilated room, and heat at 100 - 175°C (212 - 350°F)
for about three hours, or until the color has returned to orange or
yellow.
MSDS (Material Safety Data Sheets) for silica gel chemicals are
provided in Appendix C.
6-7
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.6 Troubleshooting The tables in the following section provide troubleshooting infor-
mation to help in determining the causes of problems that may
occur with the Signature flow meter or TIENet devices.
The troubleshooting tables cover the flow meter and each TIENet
device separately. Note that the 300 TIENet device (Table 6-4) is
the internal connector case.
Note
Any time a circuit board is replaced or a sensor disconnected,
you MUST perform a hardware scan and SDI-12 scan (if con-
nected) before resuming operation.
6.6.1 Signature Flow Meter
Table 6-3 Troubleshooting: Signature Flow Meter
Symptom Cause Action Parts
Display repeatedly goes Power setting is set to If key is not pressed within 5 minutes, press
blank turn display off any key to turn the display on.
Contrast is out of Adjust the display contrast by repeatedly pressing the up or down
Blank Display but audi- adjustment arrow while holding down the +/- key.
ble beep when a key is
pressed Faulty Display Replace with known good display. Display
130-0602-06
6-8
Signature® Flow Meter
Section 6 Maintenance and Servicing
60-3004-106
Dead Lead-Acid Replace or recharge the main battery con-
Model 946
Battery nected to the internal “Lead-Acid” terminals.
Lead-Acid Battery
6-9
Signature® Flow Meter
Section 6 Maintenance and Servicing
Ribbon cable loose or Check or cycle connections, then replace with Ribbon Cable
damaged known good cable. 69-4304-032
Programming error -
Check measurement configuration of level, flow rate, and volume
Zero flow rate or aster-
input for the Total Flow parameter.
isk (*)
Nonresettable totalizer
Broken wire connection Check wire connections for the totalizer on the Main CBA.
does not advance
Mechanical
Defective totalizer Replace totalizer Totalizer
60-4304-015
6-10
Signature® Flow Meter
Section 6 Maintenance and Servicing
Device not configured for display Add the parameters to the Home Display. Refer to
on the Home Display. Section 2.7.1 Site Setup.
Device is not wired correctly. Rewire connector following label on the case circuit board.
Main CBA
Main CBA faulty.
60-4304-042
6-11
Signature® Flow Meter
Section 6 Maintenance and Servicing
Calibrated before reading stabi- Recalibrate and allow the readings to stabilize before con-
lized. tinuing with calibration.
Contact closure
is not being The Contact Output card is not
made (verified configured correctly in the soft- See section 2.7.7 to configure the 304 card properly.
with an ohm ware.
meter)
6-12
Signature® Flow Meter
Section 6 Maintenance and Servicing
See section 2.7.7 to configure the 307 card properly. Use the
The analog input card is not green and yellow LED lamps to verify the hardware using the
configured correctly in the soft- set-up menus.
ware.
Verify the setting your device requires and change the active or
The passive/active setting is
passive settings on the 307 card. Verify the setting from the
incorrect or indeterminate.
set-up menus.
Verify the wires are attached to the correct channels and in the
The 307 card is correct polarity. Use the labels and LED lamps to verity the hard-
Improper connection.
not reading 4-20 ware using the setup menus. The green LED lamp will light when
proper current is flowing in the analog circuit.
Over-current protection device Disconnect power from the 307 card, the Signature, and the
in the 307 has been tripped. external analog loop. Wait 30 seconds.
6-13
Signature® Flow Meter
Section 6 Maintenance and Servicing
Assign the correct sensor to the correct flow rate to the correct
Incorrect pacing Incorrect flow total selected for total flow.
interval pacing Example: Needed to pace from the 330 bubbler, but programmed
to pace from the 310 USLS.
Improper parameter set for the Verify/change the settings/range to the proper parame-
output. ter.
6-14
Signature® Flow Meter
Section 6 Maintenance and Servicing
4-20 reading incorrectly Improper module/parameter set Verify/change the settings/range to the correct mod-
for the output ule/parameter.
External device not Over-current protection device in Disconnect power from the 307 card, the Signature, and
reading 4-20 the 307 has been tripped. the external analog loop. Wait 30 seconds.
No level reading on the dis- Parameter not selected to be displayed on Add the parameter to the Home Dis-
play Home Display play. Refer to Section 2.7.1 Site Setup.
6-15
Signature® Flow Meter
Section 6 Maintenance and Servicing
6-16
Signature® Flow Meter
Section 6 Maintenance and Servicing
6-17
Signature® Flow Meter
Section 6 Maintenance and Servicing
6-18
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.7 330 Bubbler The TIENet 330 bubbler device is factory-installed for Signature
Installation bubbler flow meters. It can also be installed by the user to
convert a Signature flow meter into a bubbler, or to replace an
old 330 device.
Preparation steps are provided in Section 6.7.1 (converting a
non-bubbler unit) and Section 6.7.2 (replacing an existing
bubbler device). Instructions for installing the new bubbler are in
Section 6.7.3.
DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.
CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.
Note
In order to work with the 330 Bubbler module, the Signature
must have an external desiccator installed. Refer to External
Desiccator, on page 5-12.
Shield Mounting
Screws (4)
6-19
Signature® Flow Meter
Section 6 Maintenance and Servicing
330 Mounting
Screws (4)
6-20
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.7.3 Installation Procedure Referring to Figures 6-12, and 6-14, perform the following steps.
1. Place the bubbler assembly on top of the main CBA, ensur-
ing that the four screw holes line up in the case, and the
10-pin connector engages correctly in its socket.
2. Attach the bubbler board to the control panel using the
four self-tapping screws previously (part #231-6149-07).
Do not overtighten.
10-pin
connector
Note
There are three pieces of tubing on the bubbler that must be
correctly connected for operation.
If any tubing is damaged, please purchase 1 meter of part
#029-1353-02 and cut to the length needed for replacement.
For additional information about tubing connections, refer to
Section A.3 330 Bubbler in Appendix A Replacement Parts.
6-21
Signature® Flow Meter
Section 6 Maintenance and Servicing
Some Signature flow meters have a ‘Y’ fitting connecting both the
reference and intake ports to the humidity connector.
When installing a 330 bubbler in a unit with this tubing configu-
ration, remove the ‘Y’ fitting and two shorter pieces of tubing.
Bring the reference port tubing out from behind the ribbon and
reroute it directly to the humidity connector.
Intake port
tubing
(remove)
‘Y’ Fitting
(remove)
Humidity
connector tubing
(remove)
6-22
Signature® Flow Meter
Section 6 Maintenance and Servicing
Reference line
Intake
Port
Humidity
Connector
Reference
Line
Reference
Port
Bubble
Tubing
6-23
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.8 Front Cover A replacement front cover (door) comes with latches attached,
Replacement and two new hinge pins.
Align the hinges of the front cover front panel. Press the pins into
the hinge barrels, with each flange facing inward (refer to
Figure 6-15), until it is flush against the hinge surface.
Using a vice grip or other tool, spread and flatten the outward
facing ends of the pins so they cannot be removed from the
hinges.
Pin flanges
flush.
6.9 System Reset In the event that the Signature Flow Meter becomes unre-
sponsive, operation may be restored by removing and then
restoring line power.
If the problem persists, operation may be restored by performing
a hard reset.
CAUTION
A hard reset erases site data and restores the program to fac-
tory default settings.
To perform a hard reset, first remove line power and any external
battery power and wait 30 seconds. Then, while holding down
both the Home key and the Delete key, restore line power.
6-24
Signature® Flow Meter
Section 6 Maintenance and Servicing
6.10 Service and Repair Service tasks described in this manual may be performed on site
by properly trained personnel. Other service and repairs must be
performed at the factory. If your Teledyne Isco equipment
requires repair, contact Teledyne Isco technical support.
Teledyne Isco
Technical Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
Phone:866 298-6174
402 464-0231
FAX:402 465-3085
E-mail:
[email protected]
Speaking with a Teledyne Isco Technical Service representative
can often resolve the problem without the need to return the
item. If the issue cannot be resolved by phone or email, you will
receive a Return Authorization Number (RAN) and information
on returning the equipment to the factory.
6-25
Signature® Flow Meter
Section 6 Maintenance and Servicing
6-26
Signature® Flow Meter
Appendix A Replacement Parts
E-mail:[email protected]
A-1
Signature® Flow Meter
A-2
Signature® Flow Meter
A-3
Signature® Flow Meter
A-4
Signature® Flow Meter
A-5
Signature® Flow Meter
A-6
Signature® Flow Meter
A-7
Signature® Flow Meter
A-8
Signature® Flow Meter
A-9
Signature® Flow Meter
A-10
Signature® Flow Meter
A-11
Signature® Flow Meter
A-12
Signature® Flow Meter
A-13
Signature® Flow Meter
A-14
Signature® Flow Meter
A-15
Signature® Flow Meter
A-16
Signature® Flow Meter
A-17
Signature® Flow Meter
A-18
Signature® Flow Meter
Appendix B Options and Accessories
B.1 Ordering Information Options and accessories can be purchased by contacting Teledyne
Isco’s Customer Service Department.
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA
E-mail: [email protected]
Note
For options and accessories exclusive to the external TIENet
devices, refer to the appropriate TIENet user manuals (found
at www.isco.com and listed in Section B.7 Manuals).
B-1
Signature® Flow Meter
B-2
Signature® Flow Meter
B.4.3 301 pH/Temperature For use with Signature 6 position plug-in (green) terminal
Device with Signature strip.
Connection Ending in
Includes cord grip, combination pH probe with built-in exposed
Unterminated Leads
temperature probe and 25 ft. probe
cable, and one package of each buffer and rinse solution for probe
calibration.
B.4.4 301 pH/Temperature For use with portable Signature TIENet receptacle.
Device with Signature Includes combination pH probe with built-in exposed temperature
Connection Ending in probe and 25 ft. probe cable and one package of each buffer and
TIENet Plug
rinse solution for probe calibration
10 m cable* .............................................................................................................................. 60-4307-070
23 m cable* .............................................................................................................................. 60-4307-071
Cut-to-length cable* ................................................................................................................ 60-4307-072
*Cable lengths from Signature to TIENet 301 device
B-3
Signature® Flow Meter
B.4.5 306 Sampler Interface For use with Signature 6 position plug-in (green) terminal
Cable with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads below).
306 Sampler Interface w/ 10 m (32.8 ft) cable ........................................................................ 60-4304-007
306 Sampler Interface w/ 23 m (75 ft) cable ........................................................................... 60-4304-008
306 Sampler Interface w/ CTL cable ....................................................................................... 60-4304-088
Assembly Model 306 5m (5800/4700 OLNY) ......................................................................... 60-5804-178
Assembly Model 306 10m (5800/4700 OLNY) ........................................................................ 60-5804-179
Assembly Model 306 23m (5800/4700 OLNY) ....................................................................... 60-5804-180
B.4.6 306 Sampler Interface For use with portable Signature TIENet receptacle.
Cable with Signature
Connection Ending in
TIENet Plug.
306 Sampler Interface w/ connector and 10m (32.8 ft) cable ................................................. 60-4304-076
306 Sampler Interface w/ connector and 23 m (75 ft) cable ................................................... 60-4304-077
306 Sampler Interface w/ connector and CTL cable............................................................... 60-4304-078
4700/5800 Sampler Interface cable ........................................................................................ 60-5314-697
B.4.7 310 Ultrasonic Level For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads.
below).
310 Ultrasonic sensor w/ 10m cable ....................................................................................... 60-4314-005
310 Ultrasonic sensor w/ 23m cable ....................................................................................... 60-4314-006
310 Ultrasonic sensor Cut-to-length* .................................................................................... 60-4314-014
. *Cable lengths can go up to 150 m with an expansion box.
B.4.8 310 Ultrasonic Level For use with portable Signature TIENet receptacle.
Sensor with Signature Includes cord grip and sensor with cable. (See cable lengths
Connection Ending in below).
TIENet Plug.
310 Ultrasonic sensor w/ connector and 10m cable ............................................................... 60-4314-009
310 Ultrasonic sensor w/ connector and 23m cable ............................................................... 60-4314-011
310 Ultrasonic sensor w/ connector Cut-to-length* ............................................................... 60-4314-013
*Cable lengths can go up to 150 m with an expansion box.
B-4
Signature® Flow Meter
B.5.1 350 Area Velocity For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads.
below).
350 AV Sensor w/ 10m Cable .................................................................................................. 60-4354-013
350 AV Sensor w/ 23m Cable .................................................................................................. 60-4354-014
350 Ultrasonic sensor Cut-to-length ...................................................................................... 60-4354-026
350 Ultrasonic sensor Cut-to-length ...................................................................................... 60-4354-026
B.5.2 350 Area Velocity For use with portable Signature TIENet receptacle.
Sensor with Signature
Connection Ending in
TIENet Plug.
350 AV Sensor w/ connector and 2ft connector..................................................................... 60-4354-017
350 AV Sensor w/ connector and 10m cable........................................................................... 60-4354-023
350 AV Sensor w/ connector and 23m cable........................................................................... 60-4354-024
350 AV Sensor w/ connector Cut-to-length ............................................................................ 60-4354-025
B.5.3 360 LaserFlow Velocity For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
connection ending in Includes cord grip and sensor with cable. (See cable lengths
unterminated leads.
below).
360 Laserflow sensor w/ 10m cable ........................................................................................ 60-4364-056
360 Laserflow sensor w/ 23m cable ........................................................................................ 60-4364-057
360 Laserflow sensor Cut-to-length*...................................................................................... 60-4364-076
*Cable lengths can go up to 150 m
360 Laserflow sensor w/ 10m cable + USNC ......................................................................... 60-4364-062
360 Laserflow sensor w/ 23m cable + USNC ......................................................................... 60-4364-063
360 Laserflow sensor Cut-to-length* + USNC....................................................................... 60-4364-078
*Cable lengths can go up to 150 m
B-5
Signature® Flow Meter
B.5.4 360 LaserFlow Velocity For use with portable Signature TIENet receptacle.
Sensor with Signature
Connection Ending in
TIENet Plug.
360 Laserflow sensor w/ connector and 10m cable ................................................................ 60-4364-058
360 Laserflow sensor w/ connector and 23m cable ................................................................ 60-4364-060
360 Laserflow sensor w/ connector Cut-to-length*................................................................ 60-4364-077
*Cable lengths can go up to 150 m
360 Laserflow sensor w/ connector and 10m cable +USNC ................................................. 60-4364-064
360 Laserflow sensor w/ connector and 23m cable +USNC .................................................. 60-4364-066
360 Laserflow sensor w/ connector Cut-to-length* +USNC ................................................. 60-4364-079
*Cable lengths can go up to 150 m
B.6 Modems
CDMA Digital cellular modem
(Cellular service not included.) ................................................................................................ 60-4307-074
Magnetic mount antenna for CDMA ....................................................................................... 60-2004-550
GSM Digital cellular modem (Outside North America)
(Cellular service not included.) ................................................................................................ 60-4307-090
GSM Digital cellular modem (North America)
(Cellular service not included.) ................................................................................................ 60-4307-075
Magnetic mount antenna for GSM.......................................................................................... 60-2004-551
Ethernet modem ....................................................................................................................... 60-4307-016
B.7 Manuals
Signature Flow Meter .............................................................................................................. 69-4303-070
Signature Bubbler Flow Meter ................................................................................................ 69-4333-004
TIENet 301 pH/Temperature Device....................................................................................... 69-4303-071
TIENet 306 Sampler Interface ................................................................................................ 69-4303-072
TIENet 310 Ultrasonic Sensor................................................................................................. 69-4313-010
TIENet 350 Area Velocity Sensor ............................................................................................ 69-4353-024
TIENet 360 LaserFlow Sensor ................................................................................................ 69-4363-043
Mounting Rings ........................................................................................................................ 60-3203-061
Flow Metering Insert ............................................................................................................... 60-3234-064
Flowlink 5.1 Software .............................................................................................................. 69-2543-213
Isco Open Channel Flow Measurement Handbook, 6th Ed. .................................................. 60-3003-041
B-6
Signature® Flow Meter
Note
Scissor Mounting Ring Assemblies will require a base and
scissors section for all sizes. Each scissors ring includes a
base section, scissors mechanism, extensions, plastic ties,
and a manual). Sizes from 16" to 80" will also require two or
more extension sections.
B-7
Signature® Flow Meter
B-8
Signature® Flow Meter
Appendix C Modbus Output Protocol
C.2 Operation There are many standard, third party Modbus drivers and OPC
servers that may be used to link a remote Modbus device, such as
the Signature Flow Meter, to SCADA or process control software,
such as Wonderware®1 or Intellution®2. The OPC server commu-
nicates with the flow meter and accesses registers. The definition
of what information is contained and where (the register number,
or address) is set by Teledyne Isco.
The Signature register addresses, and what parameters are held
where, are available in Table C-1 in Section C.6.1.
By accessing these registers you can obtain the current value of
whatever parameter you desire. The reading(s) can then be dis-
played or stored wherever you designate as a destination; for
example, a process control computer.
Not all registers are limited to read-only data storage. You can
also use two of the registers for control purposes. For example,
writing a “1” value to register 25 (“TakeReadingFlag” register),
tells the Signature to update its readings.
C-1
Signature® Flow Meter
Signature
OPC Process
Server Control
Signature
C-2
Signature® Flow Meter
C.4 Glossary of Terms Address – An address is a digital location specified for a device
(such as the Signature Flow Meter).
ASCII – Short for American Standard Code for Information
Interchange, ASCII is a code that represents English characters
with numbers. Most computers represent text with ASCII code,
making it possible for one computer or device to share data with
another.
DCS – Distributed Control Systems
Dedicated Line – A telecommunications path reserved for com-
munication between two specified points and not shared among
multiple points.
Modbus Protocol – Modbus Protocol is a messaging structure
used to establish master-slave/client server communications
between intelligent devices. Modbus is a simple
command/response mechanism to read from and write to reg-
isters.
MTU – Master Terminal Unit
OPC – OPC (OLE for Process Control) means open connectivity
via open (free for use) standards. It is a series of software stan-
dards specifications that fill a need in automation (like printer
drivers did for Windows), acting as a translator for data trans-
mission and process control.
The specification defines a standard set of objects, interfaces, and
methods for use in process control and manufacturing auto-
mation applications to facilitate interoperability. There are hun-
dreds of OPC Data Access servers and clients.
PLC – Programmable Logic Controller
Registers – A register is a location in memory for a specific data
type. The definition of what data is contained and where (the reg-
istry number, or address) is set by the equipment manufacturer
(in this case Teledyne Isco).
RTU – Short for Remote Terminal Unit (or Remote Telemetry
Unit), RTU is a code that represents data using a compact binary
format.
SCADA – SCADA (Supervisory Control And Data Acquisition)
is a computer system for gathering and analyzing real-time data.
SCADA systems are used to monitor and control plant operation,
or equipment in industries.
The SCADA system transfers the information (for example,
where a leak has occurred in a pipeline), back to a central site,
alerting the home station of the leak, performing necessary
analysis and control (such as determining if the leak is critical),
and displaying the information in a logical and organized
manner.
SCADA systems can be relatively simple, such as one that mon-
itors the environmental conditions of a small office building, or
very complex, such as a system that monitors all the activity in a
nuclear power plant or a municipal water system.
TCP/IP – Transmission Control Protocol/Internet Protocol
C-3
Signature® Flow Meter
CAUTION
Be careful not to assign the same address to more than one
flow meter.
C.6 Register Definitions The register definitions for the Signature flow meter are pro-
vided in the following table.
Where no other Unit Of Measure exists for a parameter, percent
(%) can be used in most situations.
C-4
Signature® Flow Meter
C-5
Signature® Flow Meter
C-6
Signature® Flow Meter
C-7
Signature® Flow Meter
C-8
Signature® Flow Meter
C-9
Signature® Flow Meter
C-10
Signature® Flow Meter
C-11
Signature® Flow Meter
C-12
Signature® Flow Meter
C-13
Signature® Flow Meter
C-14
Signature® Flow Meter
C-15
Signature® Flow Meter
C-16
Signature® Flow Meter
C-17
Signature® Flow Meter
C-18
Signature® Flow Meter
Appendix D Material Safety Data Sheets
This appendix provides Material Safety Data Sheets for the des-
iccant used by the Signature Flow Meter.
Teledyne Isco cannot guarantee the accuracy of the data. Specific
questions regarding the use and handling of the products should
be directed to the manufacturer listed on the MSDS.
SORB-IT® is a registered trademark of N. T. Gates Company.
D-1
Signature® Flow Meter
D-2
Signature® Flow Meter
D-3
Signature® Flow Meter
D-4
Signature® Flow Meter
D-5
Signature® Flow Meter
D-6
Signature® Flow Meter
D-7
Signature® Flow Meter
D-8
Signature® Flow Meter
D-9
Signature® Flow Meter
D-10
Signature® Flow Meter
Index
B D
battery backup, 5-4 data
browser window, 2-6 compliance, 1-3, 2-35
Bubble Line download, 2-34
Extensions, 3-17 dump, 2-33
Flume Fittings, 3-17 events, 1-3
High-Velocity Flow Streams, 3-16 export, 2-35
Open Channel Installation, 3-18 integrity, 1-3, 2-33, 2-34, 2-35
Stilling Wells, 3-17 push, 2-26
bubbler, 5-14, 6-19 storage, 2-26
default settings, 2-32
C default units, 2-19
case board, 3-1 description, 1-1
layout, 3-3 desiccant, 6-5
CDMA, 5-36 external, 5-12, 6-6
cellular modems, 5-36 internal, 6-5
CDMA, 5-36 device info, 2-5
Index-1
Signature® Flow Meter
Index
Index-2
Signature® Flow Meter
Index
Index-3
Signature® Flow Meter
Index
Index-4
Compliance Statements
ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞
Name and amount of Hazardous Substances or Elements in the product
㦘㹡㦘⹂䓸德㒥⏒侯
捷ↅ⚜䱿 Hazardous Substances or Elements
Component Name 杔 㻭 柘 ⏼ↆ杻 ⮩䅃勣啾 ⮩䅃ℛ勣啾
(Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE)
兎恾㨎
X O O O O O
Circuit Boards
㣍䯉
X O O O O O
Display
㘴兎
O O O O O X
Wiring
␔捷䟄冕
O O O O O X
Internal Cables
䦃㿐䟄㧉
X O O O O X
DC Motor
㘴⯃
O O X O O O
Connectors
䟄㻯
X X X O O O
e
Battery
bl
Ta
䟄䭐梏
X O O O O X
at
Solenoid valve
M
az
H
DECLARATION OF CONFORMITY
Application of Council Directive: 2004/108/EC -The EMC Directive
2012/19/EC– The WEEE Directive
2006/95/EC – The Low Voltage Directive
Manufacturer's Name: Teledyne Isco
Manufacturer's Address: 4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Phone: +1 (402) 464-0231
FAX: +1 (402) 465-3799
Equipment Type/Environment: Laboratory Equipment for Light Industrial/Commercial Environments
Trade Name/Model No: Signature Flow Meter (AC or DC power) with 350, 360, 301 sensors, 306 Sampler
Interface, RS485 and Ethernet Modem options
Year of Issue: 2014
Standards to which Conformity is Declared: EN 61326:2006 & IEC 61326-1:2012 EMC Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use
EN 61010-1 2nd edition Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN60529 Special Protection offered by the Signature’s Enclosure: IP-66
Standard Description Severity Applied Performance Criteria
EN61000-4-2:2008 Electrostatic Discharge (AC or DC Level 2 - 4kV contact discharge A
power) Level 3 - 8kV air discharge
EN61000-4-3:2006 Radiated RF Immunity (AC or DC 10 V/m:80MHz-1GHz, 3 V/m:1.4GHz-2GHz,
/A1:2007 /A2:2010 power) 1 V/m: 2GHz-2.7GHz, 80%, AM, 1KHz rate,1 A
sec dwell
EN61000-4-4:2004 Electrical Fast Transient (EFT) on ± 2 kV on AC lines, ±1 kV on (AC & DC) Signal A
/A1:2010 Mains and I/O Cables
EN61000-4-5:2005 Surge on AC Lines ±1 kV L-L /±2 kV L-PE A
EN61000-4-6:2008 Conducted RF Immunity on Mains and 3 Vrms on AC Mains & (AC & DC) Signal A
I/O Cables
EN61000-4-11:2004 Voltage Dips 0.5,1,25 Cycle/0,0,70% A
CISPR11/ RF Emissions Radiated, below 1GHz Group 1, Class A Industrial, Scientific, and PASS
EN 55011:2009 and Conducted, AC Mains Medical Equipment
/A1:2010
EN61000-3-2:2006 AC Harmonics, Flicker PASS
/A1:2009 /A2:2009
IEC 61000-3-2:2014
EN61000-3-3:2008
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of August 18. 2014.
USA Representative
_______________________________
Vikas V. Padhye, Ph. D.
Executive Vice President of Sales and Marketing and
Product Line General Management
4700 Superior Street
Lincoln, Nebraska 68504
Phone: 402-464-0231
Fax: 402-464-0318
60-4302-021
Rev. C
Teledyne Isco One Year Limited Factory Service Warranty*
This warranty exclusively covers Teledyne Isco The warranty for any instrument is the one in effect on date
instruments, providing a one-year limited warranty of shipment. The warranty period begins on the shipping
covering parts and labor. date, unless Teledyne Isco agrees in writing to a different
Any instrument that fails during the warranty period due to date.
faulty parts or workmanship will be repaired at the factory Excluded from this warranty are normal wear; expendable
at no charge to the customer. Teledyne Isco’s exclusive items such as charts, ribbon, lamps, tubing, and glassware;
liability is limited to repair or replacement of defective fittings and wetted parts of valves; and damage due to
instruments. Teledyne Isco is not liable for consequential corrosion, misuse, accident, or lack of proper maintenance.
damages. This warranty does not cover products not sold under the
Teledyne Isco will pay surface transportation charges both Teledyne Isco trademark or for which any other warranty is
ways within the 48 contiguous United States if the specifically stated.
instrument proves to be defective within 30 days of No item may be returned for warranty service without a
shipment. Throughout the remainder of the warranty period, return authorization number issued by Teledyne Isco.
the customer will pay to return the instrument to Teledyne This warranty is expressly in lieu of all other warranties
Isco, and Teledyne Isco will pay surface transportation to and obligations and Teledyne Isco specifically disclaims
return the repaired instrument to the customer. Teledyne any warranty of merchantability or fitness for a
Isco will not pay air freight or customer’s packing and particular purpose.
crating charges. This warranty does not cover loss, damage, The warrantor is Teledyne Isco, 4700 Superior, Lincoln, NE
or defects resulting from transportation between the 68504, U.S.A.
customer’s facility and the repair facility.
* This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer.
Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their
Teledyne Isco dealer for warranty service.
Before returning any instrument for repair, please call, fax, or e-mail the Teledyne Isco Service
Department for instructions. Many problems can often be diagnosed and corrected over the
phone, or by e-mail, without returning the instrument to the factory.
Instruments needing factory repair should be packed carefully, and shipped to the attention of
the service department. Small, non-fragile items can be sent by insured parcel post. PLEASE
BE SURE TO ENCLOSE A NOTE EXPLAINING THE PROBLEM.