Manual ISCO Signature

Download as pdf or txt
Download as pdf or txt
You are on page 1of 230

Installation and Operation Guide

Manual #69-4333-004 of Assembly #60-4334-005


Copyright © 2012. All rights reserved, Teledyne Isco
Revision H, January 2016
Foreword

This instruction manual is designed to help you gain a thorough understanding of the operation of
the equipment. Teledyne Isco recommends that you read this manual completely before placing the
equipment in service.

Although Teledyne Isco designs reliability into all equipment, there is always the possibility of a
malfunction. This manual may help in diagnosing and repairing the malfunction.

If a problem persists, call or e-mail Teledyne Isco technical support for assistance. Simple
difficulties can often be diagnosed over the phone. For faster service, please have your serial
number ready.

If it is necessary to return the equipment to the factory for service, please follow the shipping
instructions provided by technical support, including the use of the Return Material
Authorization (RMA) specified. Be sure to include a note describing the malfunction. This will
aid in the prompt repair and return of the equipment.

Teledyne Isco welcomes suggestions that would improve the information presented in this manual
or enhance the operation of the equipment itself.

Teledyne Isco is continually improving its products and reserves the right to change product
specifications, replacement parts, schematics, and instructions without notice.

Contact Information
Customer Service
Phone: (800) 228-4373 (USA, Canada, Mexico)
(402) 464-0231 (Outside North America)
Fax: (402) 465-3022
Email: [email protected]
Technical Support
Phone: Toll Free (800) 775-2965 (Syringe Pumps and Liquid Chromatography)
Email: [email protected]
Return equipment to: 4700 Superior Street, Lincoln, NE 68504-1398
Other Correspondence
Mail to: P.O. Box 82531, Lincoln, NE 68501-2531
Email: [email protected]
Signature® Flow Meter
Table of Contents

Section 1 Introduction
1.1 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2 Data Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.3 Security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.4 Compatible Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1.5 Identifying Signature Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.6 Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.6.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.6.2 Display and LED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.3 Backlight and Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.6.4 Display Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.7 Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7

Section 2 Setup and Programming


2.1 User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1.1 Keypad Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1.2 Connecting to the Signature with Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2.2 The Home Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3 Shortcuts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3.1 Adjust Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3.2 Adjust Velocity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3.3 Purge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3.4 Histograph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2.3.5 Real-Time Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.3.6 Report View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.4 Programming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.4.1 Off-Screen Content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.4.2 Character Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
2.4.3 Pull Down Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2.5 Program Steps
(Menu Trees) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.6 Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.6.1 Smart Sensor Setup (TIENet) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.6.2 SDI-12 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
2.6.3 Modbus Input Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
2.6.4 Modbus Output Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.6.5 Modem Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2.7 Configure Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2.7.1 Site Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2.7.2 Measurement Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2.7.3 Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.7.4 Equation/Trigger Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.7.5 Data Storage/Push Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.7.6 Sampler Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.7.7 Inputs/Outputs/Alarms
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.7.8 Reset Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30

5
Signature® Flow Meter
Table of Contents

2.7.9 Reports/History Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30


2.8 Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2.8.1 Language Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.2 Set New Passcode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.3 Update Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.4 Sensor Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.5 Display Signature Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.6 Display License Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.8.7 Gather Fault Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
2.8.8 Restore to Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
2.8.9 Home . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
2.9 USB Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
2.9.1 Retrieve Text Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2.9.2 Retrieve Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2.9.3 Update Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2.9.4 Save Current Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.9.5 Load Existing Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.9.6 Save Signature Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.9.7 Gather Fault Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.9.8 None of the above . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.10 Signature Data in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.10.1 Event Viewer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34
2.11 Verifying Exported Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35

Section 3 Standard Installation


3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 Connecting External Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.3 Case Bottom Cable Entries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3.3.1 Cable Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.3.2 Connecting TIENet Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3.4 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.4.1 Hard Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.4.2 Line Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.4.3 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.5 Mounting the Signature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.6 Outdoor Recommendations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3.7 The Bubble Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7.1 Standard Bubble Lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7.2 Comparing Vinyl and PTFE Bubble Lines . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7.3 Bubble Line Length . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7.4 Attaching the PTFE Bubble Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.7.5 Attaching the Vinyl Bubble Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.7.6 Installing the Bubble Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
3.7.7 High-Velocity Flow Streams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.7.8 Stilling Wells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3.7.9 Flume Bubble Line Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3.7.10 Bubble Line Extensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
3.7.11 Open Channel Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.8 Setting the Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.9 TIENet Sensor Installation in a Hazardous Location. . . . . . . . . . . . . . . . . . . . . . 3-19

Section 4 Portable Installation


4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Portable Stand. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2.1 Adjust Signature to Upward Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4

6
Signature® Flow Meter
Table of Contents

4.3.1 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4


4.3.2 Outdoor Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.3.3 Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.4 Battery Life Expectancy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.4.1 Major Standard Equipment Settings Effecting Battery Life . . . . . . . . . . . 4-6
4.4.2 Major Optional Equipment Settings Effecting battery Life . . . . . . . . . . . . 4-7
4.4.3 Additional Considerations for Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.4.4 Calculating Average Current Draw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.4.5 Battery Life Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.4.6 Low Battery Cut Off and Battery Care . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.5 Connect to External Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
4.5.1 Connecting Devices to the TIENet Receptacle . . . . . . . . . . . . . . . . . . . . . . 4-12

Section 5 Equipment Options


5.1 AC Power Cord Kit /AC Wiring (Applies to Permanent Installation Only) . . . . . . 5-2
5.2 Battery Backup (Applies to Permanent Installation Only) . . . . . . . . . . . . . . . . . . 5-4
5.3 Power Loss Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5.4 Mechanical Totalizer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
5.5 External Desiccator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5.6 TIENet® Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5.6.1 Ultrasonic Level Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5.6.2 Bubbler Level Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5.6.3 Laser Doppler Velocity Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5.6.4 Continuous Wave Doppler Velocity Sensor . . . . . . . . . . . . . . . . . . . . . . . . 5-14
5.6.5 Sampler Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
5.6.6 pH and Temperature Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
5.6.7 TIENet Expansion Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
5.6.8 Contact Output Card
(TIENet 304) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5.6.9 Analog Input Card
(TIENet 307) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
5.6.10 Analog Output Card (TIENet 308) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.7 Reference Port Tubing Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5.8 Isco Flowlink Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
5.9 Ethernet Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
5.9.1 Ethernet Modem Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
5.9.2 Network Firewall Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-35
5.10 Cellular Modems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36
5.10.1 CDMA Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36
5.10.2 GSM Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37
5.10.3 Installing the Cellular Modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37
5.10.4 Cellular Modem Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40

Section 6 Maintenance and Servicing


6.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.3 Firmware Updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.4 Accessing the Interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6.5 Desiccant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.5.1 Internal Desiccator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.5.2 External Desiccator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.6 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.6.1 Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.6.2 TIENet 300 Connector Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.6.3 TIENet 301 pH/Temp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.6.4 TEINet 304 Contact Output Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12

7
Signature® Flow Meter
Table of Contents

6.6.5 TIENet 307 Analog Input Card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13


6.6.6 TIENet 306 Sampler Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.6.7 TIENet 308 Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.6.8 TIENet 310 USLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.6.9 TIENet 330 Bubbler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.7 330 Bubbler Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.7.1 Preparation: Non-Bubbler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6.7.2 Preparation: Existing Bubbler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6.7.3 Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6.8 Front Cover Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.9 System Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.10 Service and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25

Appendix A Replacement Parts


A.1 How to Order. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
A.2 Signature Flow Meter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
A.3 330 Bubbler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-14

Appendix B Options and Accessories


B.1 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.2 Signature Flow Meter Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B.3 Power Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
B.4 TIENet Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
B.4.1 Internal to Signature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
B.4.2 Sensor and Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
B.4.3 301 pH/Temperature Device with Signature Connection Ending in Untermi-
nated Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
B.4.4 301 pH/Temperature Device with Signature Connection Ending in TIENet
Plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
B.4.5 306 Sampler Interface Cable with Signature Connection Ending in Untermi-
nated Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
B.4.6 306 Sampler Interface Cable with Signature Connection Ending in TIENet
Plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
B.4.7 310 Ultrasonic Level Sensor with Signature Connection Ending in Untermi-
nated Leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
B.4.8 310 Ultrasonic Level Sensor with Signature Connection Ending in TIENet
Plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
B.5 330 Bubble Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.5.1 350 Area Velocity Sensor with Signature Connection Ending in Unterminated
Leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.5.2 350 Area Velocity Sensor with Signature Connection Ending in TIENet Plug.
B-5
B.5.3 360 LaserFlow Velocity Sensor with Signature connection ending in untermi-
nated leads. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B.5.4 360 LaserFlow Velocity Sensor with Signature Connection Ending in TIENet
Plug. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B.6 Modems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B.7 Manuals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B.8 Sensor Mounting Rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B.8.1 Spring Rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
B.8.2 Scissor Rings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7

Appendix C Modbus Output Protocol


C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
C.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1

8
Signature® Flow Meter
Table of Contents

C.3 Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2


C.4 Glossary of Terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
C.5 Signature ASCII or RTU Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
C.6 Register Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
C.6.1 Modbus Registers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4

Appendix D Material Safety Data Sheets

List of Figures
1-1 Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 Multiple options can be used in any combination . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-3 Front and exterior component identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-4 Home screen and basic keypad functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1-5 Specification drawing: Signature Flow Meter with stand in side facing position 1-10
1-6 Specification drawing: Signature Flow Meter with stand in upward facing position
1-11
1-7 Specification drawing: Signature Bubbler Flow Meter, 1 of 2 . . . . . . . . . . . . . . . 1-12
1-8 Specification drawing: Signature Bubbler Flow Meter, 2 of 2 . . . . . . . . . . . . . . . 1-13
2-1 Flowlink Connect screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-2 Site Info tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-3 Devices tab . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-4 Program tab: Accessing the browser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
2-5 Home screen (normal operating mode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-6 Character grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-7 Menu Tree: Top menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2-8 Menu Tree: Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2-9 Measurement parameters for each TIENet device . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-10 Modbus Input Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-11 Editing Modbus device parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
2-12 Modbus Output Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
2-13 Menu Tree: Configure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2-14 Menu Tree: Flow rate input setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2-15 Menu Tree: Volume Input Setup (total flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-16 Menu for Rainfall Input Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2-17 Menu for Load Calculation Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2-18 Example of defining conditions & building equations. . . . . . . . . . . . . . . . . . . . . 2-26
2-19 Local alarm setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
2-20 SMS and Server alarm setup (modem required) . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2-21 Menu Tree: Administration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2-22 Menu Tree: USB Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33
2-23 Event Viewer in Flowlink . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35
2-24 Report file verifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
3-1 Open door and front panel to access interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2 Connector case, connectors, and fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 Connector Case cable entries for power and external devices . . . . . . . . . . . . . . . . 3-4
3-4 Strain relief 3/4 NPT Cord-grip fitting for TIENet devices . . . . . . . . . . . . . . . . . . 3-5
3-5 Cord-grip fittings installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-6 Diameter-seal plugs for unused ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-7 TIENet Device terminal strips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
3-8 Installing TIENet cable with a cord-grip fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-9 TIENet Device terminal connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-10 Attach wired terminal strip to connector case socket . . . . . . . . . . . . . . . . . . . . . 3-9
3-11 Insert the cable reference tubing into the case board reference port . . . . . . . . . 3-9
3-12 Position and secure the cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3-13 Location of power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12

9
Signature® Flow Meter
Table of Contents

3-14 Signature Flow Meter mounted on wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12


3-15 Weather shield - recommended for outdoor installations
(Drawing & Photo courtesy of TRACOM, Inc.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
3-16 Positioning the Bubble Line in the Flow Stream . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3-17 Installing the Stainless Steel Bubble Line Extension . . . . . . . . . . . . . . . . . . . . 3-18
3-18 Level adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
3-19 Hazardous Location Installation Control Drawing-ATEX . . . . . . . . . . . . . . . . . 3-20
4-1 Signature on portable stand side facing (left) and upward facing (right) . . . . . . . 4-1
4-2 Pull out the handle latch pins on each side of the stand . . . . . . . . . . . . . . . . . . . . 4-2
4-3 One of the eight screws loosened. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4-4 Signature tilted up in preparation of removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-5 Signature in upward facing position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-6 Notch in stand for the power cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-7 946 lead-acid battery installed under the Signature Portable . . . . . . . . . . . . . . . 4-5
4-8 Battery hold down and thumb screw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-9 Example of defining conditions and building equations . . . . . . . . . . . . . . . . . . . . 4-6
4-10 Measuring flow meter current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4-11 Connector case, connectors, and fuses for Signature Portable Flow Meter . . . 4-12
4-12 How to connect a TIENet receptacle to the Signature Portable . . . . . . . . . . . . 4-13
4-13 Removing the plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
4-14 Connecting a rain gauge to the Signature receptacle (L) and then to the Signature
connector board (R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
5-1 AC Line cord with a cord-grip fitting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-2 Power supply mounting screw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-3 Power supply removal/replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5-4 Power supply terminal strip: AC Input and DC Output . . . . . . . . . . . . . . . . . . . . 5-3
5-5 Battery backup kit contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
5-6 Attach extension cable to the connector case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5-7 Installing the battery backup mounting plate . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5-8 Backup battery, installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
5-9 Disconnect and remove bubbler module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5-10 Remove totalizer window cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5-11 Optional non-resettable totalizer installation . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
5-12 External desiccator, installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
5-13 Remove red caps before installing external desiccant cartridge . . . . . . . . . . . . 5-13
5-14 Removing the external desiccant cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
5-15 Tubing configuration for optimal drying power (Exterior desiccator required) 5-15
5-16 TIENet Expansion Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
5-17 304 contact output card front and back view . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
5-18 View of two cards installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
5-19 304 contact output device configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20
5-20 304 contact output setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
5-21 307 analog input channel identification and terminal connections . . . . . . . . . 5-22
5-22 Direction of switch in active and passive modes . . . . . . . . . . . . . . . . . . . . . . . . 5-23
5-23 307 analog input device configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5-24 307 analog input setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
5-25 308 analog output channel identification and terminal connections . . . . . . . . 5-26
5-26 308 analog output device configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5-27 308 analog output setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
5-28 Reference port kit (full length not shown) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5-29 Reference port kit: Remove filter and barb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5-30 Reference port kit: Install filter and barb . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5-31 Reference port kit: Installed on flow meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5-32 Ethernet modem kit contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
5-33 Ethernet modem installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
5-34 Ethernet modem setup: Communication settings (default settings shown) . . . 5-34
5-35 Locating the NODE ID (MAC address) on the ethernet modem . . . . . . . . . . . . 5-35
5-36 Magnetic-mount cellular antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36
5-37 CDMA Cellular modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-36

10
Signature® Flow Meter
Table of Contents

5-38 GSM Cellular modem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37


5-39 FCC ID label on modem and location of label on the bottom of Signature case 5-37
5-40 Cellular modem kit contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-38
5-41 Cellular modem installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-39
5-42 Cellular modem setup: Communication settings . . . . . . . . . . . . . . . . . . . . . . . . 5-40
6-1 Locating firmware updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6-2 USB Micro adaptor cable (flash drive not included) . . . . . . . . . . . . . . . . . . . . . . . 6-3
6-3 Open door and front panel to access interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6-4 Removing the internal desiccant bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6-5 External desiccator, installed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6-6 Desiccant indicating saturation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6-7 Removing the external desiccant cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
6-8 Opening the desiccant cartridge chambers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
6-9 Plumbing diagram of bubbler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
6-10 Remove non-bubbler shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
6-11 Remove bubbler module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6-12 330 Bubbler assembly installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6-13 Remove extra tubing and fitting (if applicable) . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6-14 Routing and connections of 330 bubbler tubing . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6-15 Front cover (door) replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
C-1 Configuration example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2

List of Tables
1-1 Signature Flow Meter Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1-2 TIENet 330 Bubbler Module Technical Specifications . . . . . . . . . . . . . . . . . . . . . . 1-9
2-1 Modbus Setup Worksheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
4-1 Battery Life Expectancy
(default measurement rate of 15 minutes) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4-2 Battery Life Expectancy
( measurement rate of 5 minutes, all other parameters use default) . . . . . . . . . . 4-8
4-3 Default Method for Portable Signature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-4 Additional Average Current Draw . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
5-1 Direction of arrows and mode types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
6-1 Recommended Maintenance (Accessible Locations) . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 Recommended Maintenance (Difficult-to-Access Locations) . . . . . . . . . . . . . . . . . 6-1
6-3 Troubleshooting: Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6-4 Troubleshooting: TIENet 300 Connector Case . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6-5 Troubleshooting: TIENet 301 pH/Temperature Device . . . . . . . . . . . . . . . . . . . . 6-11
6-6 Troubleshooting: TIENet 304 Contact Output Card . . . . . . . . . . . . . . . . . . . . . . 6-12
6-7 Troubleshooting: TIENet 307 Analog Input Card . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6-8 Troubleshooting: TIENet 306 Sampler Interface . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6-9 Troubleshooting: TIENet 308 4-20mA Analog Output . . . . . . . . . . . . . . . . . . . . . 6-14
6-10 Troubleshooting: TIENet 310 Ultrasonic Level Sensor . . . . . . . . . . . . . . . . . . . 6-15
6-11 Troubleshooting: TIENet 330 Bubbler Module . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
C-1 Modbus Output Registers for Signature Flow Meter . . . . . . . . . . . . . . . . . . . . . . C-4

11
Signature® Flow Meter
Table of Contents

12
Signature® Flow Meter
Safety

Signature® Flow Meter


Safety

General Warnings Before installing, operating, or maintaining this equipment, it is


imperative that all hazards and preventive measures are fully
understood. While specific hazards may vary according to
location and application, take heed of the following general
warnings:

WARNING
Avoid hazardous practices! If you use this instrument in
any way not specified in this manual, the protection
provided by the instrument may be impaired.

AVERTISSEMENT
Éviter les usages périlleux! Si vous utilisez cet instrument
d’une manière autre que celles qui sont specifiées dans ce
manuel, la protection fournie de l’instrument peut être
affaiblie; cela augmentera votre risque de blessure.

Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts,
These three levels are described in the sample alerts below.

CAUTION
Cautions identify a potential hazard, which if not avoided, may
result in minor or moderate injury. This category can also warn
you of unsafe practices, or conditions that may cause property
damage.

WARNING
Warnings identify a potentially hazardous condition, which
if not avoided, could result in death or serious injury.

DANGER
DANGER – limited to the most extreme situations
to identify an imminent hazard, which if not
avoided, will result in death or serious injury.

v
Signature® Flow Meter
Safety

Hazard Symbols The equipment and this manual use symbols used to warn of
hazards. The symbols are explained below.

Hazard Symbols
Warnings and Cautions
The exclamation point within the triangle is a warning sign alerting you of
important instructions in the instrument’s technical reference manual.

The lightning flash and arrowhead within the triangle is a warning sign alert-
ing you of “dangerous voltage” inside the product.

Symboles de sécurité
Ce symbole signale l’existence d’instructions importantes relatives au
produit dans ce manuel.

Ce symbole signale la présence d’un danger d’électocution.

Warnungen und Vorsichtshinweise


Das Ausrufezeichen in Dreieck ist ein Warnzeichen, das Sie darauf
aufmerksam macht, daß wichtige Anleitungen zu diesem Handbuch
gehören.

Der gepfeilte Blitz im Dreieck ist ein Warnzeichen, das Sei vor “gefährlichen
Spannungen” im Inneren des Produkts warnt.

Advertencias y Precauciones
Esta señal le advierte sobre la importancia de las instrucciones del manual
que acompañan a este producto.

Esta señal alerta sobre la presencia de alto voltaje en el interior del


producto.

vi
Signature® Flow Meter
Section 1 Introduction

The Signature Flow Meter is designed for open channel flow


monitoring applications using any combination of flow and
parameter measurement technologies and sampling, depending
on what is required at the monitoring site.
The bubble line is anchored in the flow stream at the appropriate
measuring point in the weir, flume, or other open channel flow
situation. Air slowly bubbles out of the line into the flow stream.
The pressure in the bubble line is proportional to the liquid level
in the flow stream, and the flow meter measures this pressure,
sensing the liquid level.The Signature has built-in standard
level-to-flow conversions that cover the majority of open channel
flow measurement situations. Flow measurement is usually a
calculation based on a known relationship between liquid level
and flow rate. Additionally, the Signature can calculate flow
using standard open channel level-to-flow and area-velocity con-
versions, as well as equations, or data points, depending upon the
measurement device(s) attached to the meter and the program
specified by the user.

Note
Recent enhancements have been made to the Signature Flow
Meter. Many of these features are standard for all new Signa-
ture units which are identifiable by the serial label mounted on
the bottom of the unit (PN 60-4304-065).

Figure 1-1 Signature Flow Meter

1-1
Signature® Flow Meter
Section 1 Introduction

1.1 Quick Start

Numerical: , , ... Enter:

• Enter values • Type characters • Open list • Confirm selection • Activate


field
Arrows:
, ... • = Pull down list available
• Navigate up, down, left, right
• = Character grid available
Soft Keys:
Delete / Exit:

• Current displayed function •Clear last entry • Exit list • Close window

• Back = Return to last menu Decimal:

• Next = Confirm & advance to next • Navigate up one screen at a time


step
Plus-Minus:
Home:
• Navigate down one screen at a time
• Return to Home Screen from anywhere • Adjust screen contrast, when used with
the up or down arrow keys.

Hardware Setup Configure Options Administration


TIENet Devices Site: Clock, Name, Display, Units Language
SDI-12 Setup Measurement Setup: Set New Passcode
Level, Velocity, Flow, Volume
Modbus In Update Firmware
Modbus Out Adjust: Level, pH, Velocity Sensor Diagnostics
Equation / Trigger System Information
Modem
Data Storage Rate / Data Push License Information
Sampler Interface Setup Gather Fault Data
Outputs: Alarm, Analog Restore Defaults
Reset Totalizers
Reports / History

1-2
Signature® Flow Meter
Section 1 Introduction

1.2 Data Integrity What makes the Signature Flow Meter unique is its ability to
verify data integrity. This is accomplished by logging four special
event data types that cannot be altered, and are designed to alert
the user to any trends or anomalies, and to assess compliance.
This data can be downloaded from the flow meter and observed
in tabular or graphical format alongside regular site data, using
Flowlink software (see Section 2.10 Signature Data in Flowlink).
The data can also be downloaded onto a flash drive via the
micro-USB assembly on the front panel of the flow meter, then
imported into a spreadsheet or other viewing application (refer to
Section 2.9 USB Options for more information).
The four event data types are:
Program Report – Tracks changes to the Signature Flow Meter
configuration
Summary Report – Documents summaries of data measure-
ments (e.g., Min/Max/Avg)
Diagnostic Report – Tracks results from diagnostic tests
History Report – Tracks user and meter events (e.g., level
adjustments, data push, etc.)

1.3 Security The Signature Flow Meter can be secured shut with a padlock
with the hasp on the right side of the housing (refer to
Figure 1-3). The program settings and recorded data can be pro-
tected by a user-selected passcode (refer to Section 2.8.2 Set New
Passcode.

1.4 Compatible The Signature Flow Meter can interface with a variety of mea-
Equipment surement devices and other system components, depending on
site requirements.
Measurement devices for flow and water quality can be con-
nected to the same Signature Flow Meter and run simultane-
ously with TIENet® connectivity (up to nine TIENet devices).
The flow meter can also communicate with an optional Teledyne
Isco wastewater sampler and rain gauges.
For descriptions of interfacing and parameter sensing TIENet
devices, refer to Section 5 Equipment Options. Each external
TIENet device comes with its own user manual.
The Signature is capable of receiving data from devices using
Modbus ASCII or Modbus RTU protocol.
A variety of application-specific accessories are available from
Teledyne Isco. Refer to Appendix B Options and Accessories for a
complete list with ordering information.
A basic Signature system has one or more TIENet devices for
flow and/or parameter measurement connected to the Signature
Flow Meter (up to nine TIENet devices at once). Other configura-
tions may include an enclosure and additional internal or
external devices, including analog output cards, analog input
cards, contact output cards, a modem, up to two SDI-12 inputs,
and Modbus devices.

1-3
Signature® Flow Meter
Section 1 Introduction

When connected remotely via modem, the web interface of the


Signature Flow Meter provides remote control and data access.

Figure 1-2 Multiple options can be used in any combination

1-4
Signature® Flow Meter
Section 1 Introduction

1.5 Identifying Signature The following are key components of the Signature Flow Meter:
Components
Injection molded UV protected
sturdy enclosure

USB Cable USB


Connector

Desiccant
Holder
(Used with TIENet 330
Bubbler, TIENet 350
AV Sensors, and 360
Laser Flow)

to bubbler air
supply

Padlock
Hasp

to reference
port

Signature Connection
ports for power,
TIENet devices,
antenna, bubble line
hose barb

Signature Portable Connection


ports for power and TIENet devices

Figure 1-3 Front and exterior component identification

1-5
Signature® Flow Meter
Section 1 Introduction

1.6 Controls and Before programming and operating the Signature Flow Meter,
Indicators become familiar with the keypad and standard screens.

1.6.1 Keypad Keys are identified in Figure 1-4; keypad functions are detailed
in Section 2.1.1.

User-programmed name Non-resettable


of measurement site totalizer (optional) Numerical keys (12)
Main data type (Default =

Site Name Here


Total Flow)

Flow Meter

Total Flow

12345678 gal Scroll


Flow: 8.55 cfs bar
Additional
data types
Level B: 2.00 ft
pH: 7.00 pH

Status
line [ Status: 10/01/2011 15:20 ] Red/Green
Softkey SHORTCUTS MENU HELP
LED
functions indicator

navigation
Arrow keys (4) for screen
Micro-USB
indicator
Softkeys (3) Home Enter Delete
(Run screen) (Return key) (Undo)

Figure 1-4 Home screen and basic keypad functions

Below the display are three software configured function keys


(softkeys) that are used to make selections and navigate through
menus. Their specific functions are dependent upon what oper-
ation you are performing, and will appear in the display window.

1-6
Signature® Flow Meter
Section 1 Introduction

1.6.2 Display and LED The LED on the front panel is aligned next to the Status line on
the display screen.
• A green light indicates that new information is available
for viewing in the Status line.
• A flashing green light indicates low power mode when
the screen is off and data traffic on some channels.
• A red light indicates a warning or a user-defined alarm
condition, with further details viewable by pressing the
Alarm softkey (C).
Figure 1-4 shows the standard home screen, or run screen.
Displayed menus and programming steps are explained in
Section 2 Setup and Programming.

1.6.3 Backlight and Screen There is an option to turn off the backlight and screen.

1.6.4 Display Contrast The contrast of the LCD screen can be adjusted using the
keypad.

To adjust the contrast, hold down the ± key and repeatedly

press the down arrow (softer) or the up arrow


(sharper) to achieve the desired appearance.

1.7 Technical Table 1-1 provides technical specifications for the Signature Flow
Specifications Meter. Table 1-2 provides technical specifications for the TIENet
330 bubbler module.

Table 1-1 Signature Flow Meter Technical Specificationsa


Size (HxWxD) 22.6 x 31.0 x 20.9 cm (8.9 x 12.2 x 8.2 in)
with mounting bracket & external desiccator
Size with Portable Stand (HxWxD) 42.5 x 34.5 x 26.6 cm (16.74 x 13.58 x 10.48 in)
Weight
Flow Meter, No Bubbler Signature w/o options: 4.5 kg (10 lbs)
Signature w/ all interior options: 4.9 kg (10.7 lbs)
Signature w/ all int. options + mounted battery backup: 7.8 kg (17.3 lbs)

Bubbler Flow Meter Signature w/o options: 5.9 kg (13 lbs)


Signature w/ all interior options: 6.1 kg (13.5 lbs)
Signature w/ all int. options + mounted battery backup: 9.1 kg (20 lbs)
Weight (Portable Signature) Base Portable Signature w/ stand, TIENet Receptacle, DC Power Cable,
Desiccator: 8.0 Kg (17.5 Lbs.)

Optioned Portable Signature w/ Bubbler, stand, TIENet Receptacle, DC


Power Cable, Desiccator, CDMA Modem, 3 option card, rain gauge: 10.1
Kg (22.3 lbs.) Includes antenna.
Stand only: 3.9 Kg (8.7 lbs.)

1-7
Signature® Flow Meter
Section 1 Introduction

Table 1-1 Signature Flow Meter Technical Specificationsa (Continued)


Materials
Housing PPO Plastic (Noryl)
Window Polycarbonate
Hardware Stainless Steel
Enclosure NEMA4X/IP66
Power 100 to 240 VAC, 50/60Hz, 1.3A; Disconnect Device = Line Cord
12VDC (optional battery backup)b
12VDC 4.0A battery standalone power
Connections to Signature Flow Meter
External TIENet devices Bottom Cable entry, 1 to 4 position
(3/4" NPT user-supplied conduit or optional cord grips);
Pluggable screw terminals, 6-position;

Power supply Screw terminal, Wire 14-22 AWG


Parameter inputs Fixed terminals, 3-pin, Wire 14-30 AWG
Analog Input Pluggable screw terminal, 3-position, Wire 14-30 AWG
Analog Output Pluggable screw terminal, 3-position, Wire 14-30 AWG
Contact Output Pluggable screw terminal, 3-position, Wire 14-30 AWG
Cellular Modems Antenna Custom SMB connector
Ethernet RJ-45 connector
330 Bubbler module Internal, factory-installed
Flow Measurement Technologies Ultrasonic (TIENet 310)
Bubbler (TIENet 330)
Laser Doppler Velocity (TIENet 360 LaserFlow)
Continuous Wave Doppler Velocity (TIENet 350)
Flow Conversions Weir, Flume, British Flume, Metering Insert, Manning Formula, Equation,
LTF or LTA Data Points (up to 50 pairs), Area Velocity
Data Storage Non-volatile flash; retains stored data during program updates.
Interval: 15 or 30 seconds; 1,2, 5,15, or 30 minutes; or 1, 2, 4, 12, or 24 hrs
Capacity: 8M (180 days with 5 parameters logged at 1 minute intervals,
reports at 24-hour intervals)
Setup and Data Retrieval Serial connection to PC via USB; Cellular or Ethernet modem
Ambient Temperature Range
(Operation and Storage) -20 to 60 °C (-4 to 140 °F)c
NOTE — The operating ambient temperature range of the optional mechan-
ical totalizer (see Section 5.4) is -10 to 60 °C (14 to 140 °F).
Optional Teledyne Isco TIENet 306 device
Sampler Interfacing Output: Flow pacing, Enabling on trigger
Input: Event and bottle information
Optional 304 TIENet Contact Output:
Switching modes Normally open, Normally closed
Max Load 30 volts 1 amp
Isolation Galvanic Isolation
Outputs per card 2
Optional 307 TIENet Analog Input: Configurable either active (signature supplying loop power) or passive (rely-
ing on loop power, signature is not powering the loop).
Output voltage (in active mode) 17 VDC minimum
Range 4 to 20 mA
Isolation Galvanic Isolation
Maximum Load 400maximum (in passive mode at 20 mA)
Outputs per card 2

1-8
Signature® Flow Meter
Section 1 Introduction

Table 1-1 Signature Flow Meter Technical Specificationsa (Continued)


Optional 308 TIENet Analog Output:
Range 4 to 20 mA
Isolation Monolithic Isolation
Maximum Load 900
Outputs per card 2
Industry Standard Inputs Two SDI-12, RS485 Modbus ASCII & RTU, 4-20 mA Analog
Industry Standard Outputs 4-20 mA Analog, Modbus ASCII & RTU
Rain Gauge Connectiond Fixed terminals, 3-pin, Wire 14-30 AWG
Communication Options Direct USB Serial Connection,
CDMA (1XRTT), GSM (GPRS), and Ethernet

a. All specifications are subject to change without notice.


b. Power loss alarm options are available. Refer to 5.3 Power Loss Alarm in Section 5 Equipment Options.
c. Older model 69-4303-024 Connector Case circuit boards limit Ethernet ambient range to +40 °C (104 °F)
d. Optional industry standard rain gauge connector 60-4304-055.

Table 1-2 TIENet 330 Bubbler Module Technical Specificationsa


Operation and Storage Temperature -20 to 60 °C (-4 to 140 °F)
Level Measurement Range 0.003 to 3.05m (0.01 to 10 ft.)
Measurement Accuracy ±0.002m @ 22°C (±0.007 @ 72°F)
Temperature Compensation Range 0 to 60°C (32 to 140°F)
Temperature Coefficient ±0.0003 x Level (m) x Temperature deviation from 22 °C
(within compensated range) ±0.00017 x Level (ft) x Temperature deviation from 72 °F
1
Bubble Line Entry /8” Hose barb through bottom cable entry

a. All specifications are subject to change without notice.

1-9
Signature® Flow Meter
Section 1 Introduction

Figure 1-5 Specification drawing: Signature Flow Meter


with stand in side facing position

1-10
Signature® Flow Meter
Section 1 Introduction

Figure 1-6 Specification drawing: Signature Flow Meter


with stand in upward facing position

1-11
Signature® Flow Meter
Section 1 Introduction

Figure 1-7 Specification drawing: Signature Bubbler Flow Meter, 1 of 2

1-12
Signature® Flow Meter
Section 1 Introduction

Figure 1-8 Specification drawing: Signature Bubbler Flow Meter, 2 of 2

1-13
Signature® Flow Meter
Section 1 Introduction

1-14
Signature® Flow Meter
Section 2 Setup and Programming

The Signature Flow Meter is shipped from the factory with a


default program already configured. The Signature Portable is
shipped with different defaults than the Signature. Your par-
ticular installation will normally require different program set-
tings, specific to your monitoring site and application.
This section of the manual explains the Signature’s operating
modes, and provides instructions for site setup and pro-
gramming. Programming may be performed before transpor-
tation to the installation site.

2.1 User Interface The Signature Flow Meter can be set up, programmed, and inter-
rogated directly through the keypad and display screen, or
remotely using a computer equipped with Teledyne Isco’s
Flowlink® software, with either a USB cable or optional modem.
The Signature Flow Meter has its own browser, accessed via
Flowlink, that mirrors the physical keypad and display.

2.1.1 Keypad Functions The following briefly explains the basic function of each key.

The numerical keys are for entering values during


setup/programming.

The large softkeys (A, B, and C) perform whatever


function is currently displayed above them. Note that pressing
the BACK softkey (A) will discard any changes you have made
without saving.

The arrow keys are for navigating to different areas on-


and off-screen. The up/down arrows navigate a single line at a
time.

From the home screen, the Enter key is used to adjust or


configure the currently highlighted parameter.
From programming screens, the Enter key confirms selections
and entries you have made, opens the setup/programming screen
for a highlighted parameter, displays the character grid for
alphanumeric entry, displays a calendar for date selection, or dis-
plays the pull down menu for a highlighted field.

Note
Enter is for selection only. The NEXT softkey is for selection
and advancement to a subsequent step.

2-1
Signature® Flow Meter
Section 2 Setup and Programming

The home key returns the flow meter to the home screen
from any other screen.

The delete key clears the last character entry, exits a pull
down list, or closes an open window.

In addition to typing the decimal/period, this key can be


used to navigate up one screen at a time.

The ± key can be used to navigate down one screen at a


time, and also to adjust screen contrast, when used with the
up/down arrow keys.

2.1.2 Connecting to the With Flowlink software version 5.15.XX or later, you can set up,
Signature with program, and download data from the flow meter through its
Flowlink browser. Connection between the flow meter can be direct,
through the micro-USB assembly on the front panel, or remote,
with an internal cellular or ethernet modem.

USB Driver for Signature In order for your computer to connect to the Signature flow meter
through the micro-USB assembly, you must have the correct
driver installed. USB drivers for both 32-bit and 64-bit operating
systems are included with your flowlink program.
To install the driver:
After installing Flowlink, navigate to its program directory, and
then to the USB Driver folder, typically at C:\Program
Files\Flowlink 5.1\USB Driver.
Here you will find two drivers:
4300Driver_x64.msi for 64-bit operating systems and
4300Driver_x86.msi for 32-bit operating systems.
Without the Signature connected to your computer, begin
running the appropriate file for your operating system. When
prompted, connect the Signature to your computer’s micro-USB
assembly. You should see a message in the lower right corner
stating that new hardware has been found at the appropriate
com port number.
Ensure that the Signature flow meter is connected to the com-
puter before launching Flowlink.

Connect window You can connect with the Connect window, or if you have con-
nected with this site before, highlight the Signature site file in
the workspace (left column in Flowlink) and select
Item > Connect.

In the Connect window, select the Type of connection.

2-2
Signature® Flow Meter
Section 2 Setup and Programming

Figure 2-1 Flowlink Connect screen

Direct connection is made through the micro-USB assembly.


From the COM Port pull down list, select the port associated
with the Signature.
TCP connection is made from the computer to the Signature
flow meter’s optional CDMA, GSM, or Ethernet modem. Enter
the correct static IP and port number, separated by a colon, the
correct public domain address.
If you do not want the flow meter matched with an existing site
in the database, select the Create new site check box. (If this is
the first time the flow meter is connected to Flowlink, a new site
will be created automatically.)
Then click the large button under the heading Signature Series.

2-3
Signature® Flow Meter
Section 2 Setup and Programming

Site Screen The Signature site screen has three tabs:


Site Info contains information specific to this site. Enter all rel-
evant information into the Site Info fields, including the desired
Site Name, and save the information by clicking Apply.

Click Apply to save the


site information you
have entered.

Figure 2-2 Site Info tab

2-4
Signature® Flow Meter
Section 2 Setup and Programming

Devices lists the name, software version, and hardware version


of the flow meter for offline viewing of the site file.
This information, along with that of any connected TIENet
devices, can be viewed from the specific flow meter’s firmware
(refer to Sensor Diagnostics, on page 2-31).

Figure 2-3 Devices tab

2-5
Signature® Flow Meter
Section 2 Setup and Programming

Program is the portal through which you access the Signature


browser. The programming functions and displayed data in the
browser are functions of the flow meter firmware, not Flowlink.

The Browser Window


view (above) mirrors
t h e C o n t r o l Pa n e l
display (at left)

Figure 2-4 Program tab: Accessing the browser

2-6
Signature® Flow Meter
Section 2 Setup and Programming

2.2 The Home Screen The home screen, or run screen, is displayed when the flow meter
is in normal operating mode. This screen shows the current
parameter readings and system status or alarm conditions.
A scroll bar on the right of the screen indicates there are more
parameters off-screen that can be viewed by scrolling up or down.

Site Name Here


Total Flow

12345678 gal
Flow: 8.55 cfs
Level B: 2.00 ft
pH: 7.00 pH

[ Status: 02/06/2012 15:20 ]


SHORTCUTS MENU HELP
A B C
Figure 2-5 Home screen (normal operating mode)

2.3 Shortcuts The Shortcuts menu provides quicker access to most commonly
used commands, such as level adjustment or viewing data
recorded over a period of time. Not all menu items described in
this section will necessarily appear in your Shortcuts menu. The
selections available in the Shortcuts menu are determined by
what connected devices have been detected by the Signature flow
meter. To access your shortcuts, press SHORTCUTS ( ).

2.3.1 Adjust Level To set a new level, enter the value in the field next to Level, and
select Adjust. To update the current reading, select Update.

2.3.2 Adjust Velocity This selection will open the velocity grid with current readings
and laser controls for the TIENet LaserFlow velocity sensor. For
complete information about this device, refer to the TIENet 360
LaserFlow user manual.

2.3.3 Purge The Signature Bubbler flow meter allows you to manually purge
the bubble line if an obstruction is suspected.

2.3.4 Histograph The histograph displays the measurements taken of up to three


selected parameters in graphical format, beginning at your
selected date/time, and spanning one to 48 hours. Enter a value
in the Threshold field for a reference line. The measurements
available for graphing are determined by what measurements
are set up for data storage.

2-7
Signature® Flow Meter
Section 2 Setup and Programming

2.3.5 Real-Time The Real-Time Measurement will display the sensor/card options
Measurement that are available. Once one is selected, a table will be displayed
showing the different types of measurements the sensor/card is
collecting.

2.3.6 Report View Reporting is set up from the Configure Options menu. Summary
displays summaries of data measurements (i.e., Min/Max/Avg).
History tracks user and meter events. Program tracks changes
made to the flow meter’s program configuration.

2.4 Programming To access the setup/program menus, press MENU ( ).

When you press MENU, the four top menu options appear:
Hardware Setup detects all devices connected to the flow
meter, establishes proper communication with them, and
allows configuration of each device.
Configure Options sets up the measurement site and pro-
gram parameters.
Administration dictates operating preferences and perform
general housekeeping tasks.
Home returns to the home screen.
Additionally,
USB Options appears when a flash drive is connected to
the micro-USB assembly in the lower left corner of the con-
trol panel.
The program menus consist of steps and substeps. During pro-
gramming, available subordinate menu content and steps will be
determined by what you have previously entered, and what
optional equipment is connected to the Signature flow meter.

2.4.1 Off-Screen Content An arrow in the lower right corner of the flow meter’s screen (see
symbols at left) indicates that there is additional content on this
screen in the direction the arrow is pointing. Use the arrow keys
to access this content.

2.4.2 Character Grid A small grid icon in the lower right corner of the flow meter’s
screen (see symbol at left) indicates that the character grid is
available.
Whenever you need to enter characters, such as letters, numbers,
or punctuation, press Enter to display the character grid
(Figure 2-6).
Use the arrow keys to navigate to the desired character and
press Enter to select. When you are finished editing, select DONE
and press Enter.

2-8
Signature® Flow Meter
Section 2 Setup and Programming

I s c o Te s t S i t e
Done Cancel
A B C D E F G H I J K L M N
O P Q R S T U V W X Y Z a b
c d e f g h i j k l m n o p
q r s t u v w x y z / : !
@# $ % ^ & * ( ) - _ + = <
> ? , .
Figure 2-6 Character grid

2.4.3 Pull Down Menus Fields with a pull down arrow next to them (see example at left)
indicate a pull down list. Use the arrow keys to navigate between
Total Flow fields on the screen; when you highlight a pull down field, press
pull down Enter to display the items on the list. Then use the arrow keys
arrow and Enter to select from the list.

2-9
Signature® Flow Meter
Section 2 Setup and Programming

2.5 Program Steps The program steps in Figures 2-7 through 2-22 show the overall
(Menu Trees) menu sequencing.
An explanation of each program step, and information specific to
connected devices, is provided in Sections 2.6 Hardware Setup,
2.7 Configure Options, 2.8 Administration, and 2.9 USB
Options. This information can also be viewed by pressing the
Help softkey on your Signature flow meter or its browser screen.

Menu
1. Hardware Setup
2. Configure Options
3. Administration
4. Home

Hardware Setup Administration Options


1. Smart Sensor Setup (TIENet) 1. Language Options
2. SDI-12 Setup 2. Set New Passcode
3. MODBUS Input Setup 3. Update Firmware
4. MODBUS Output Setup 4. Sensor Diagnostics
5. Modem Setup 5. Display Signature Information
6. Display License Information
7. Gather Fault Data
8. Restore To Factory Defaults
Configure Options
1. Site Setup (Home)
2. Measurement Setup Returns to Home
3. Adjust screen
4. Equation/Trigger Setup
5. Data Storage/Push Setup
6. Sampler Setup
7. Inputs/Outputs/Alarms Setup
8. Reset Totalizers
9. Reports/History Setup

Figure 2-7 Menu Tree: Top menu

2-10
Signature® Flow Meter
Section 2 Setup and Programming

2.6 Hardware Setup From the Hardware Setup menus, the flow meter detects and
configures all connected devices.

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

Smart Sensor Setup (TIENet) MODBUS Output Setup


• View System Devices • Enter MODBUS address
• Configure Measurements from 2 to 247
• Perform Scan • Configure communication
protocol
All TIENet devices detected by the
most recent scan will be listed under For Modbus output register information,
S m a r t S e n s o r S e t u p, b y S e r i a l refer to:
Number and Device Type. Table C-1 Modbus Output Registers for
Signature Flow Meter, on page C-4.

SDI-12 Setup
• Scan for Connected Sondes
• Add/Remove
• Configure/
Manual Config
Modem Setup
Screen content determined by type
of modem installed.
MODBUS Input Setup Refer to Section 5.10.4 Cellular
1. MODBUS Input COM Settings Modem Configuration.
• Set up communication protocol
2. MODBUS Input Device Settings
• Configure or create device(s)

Figure 2-8 Menu Tree: Hardware Setup

2-11
Signature® Flow Meter
Section 2 Setup and Programming

2.6.1 Smart Sensor Setup This selection will display the most recently detected TIENet
(TIENet) devices connected to the Signature flow meter.
Perform Scan If TIENet devices have been added or removed from the system,
highlight Perform Scan and press Enter to detect the current
system configuration.
Sensor Differences If there are any differences in the device configuration since the
last scan, a list of Sensor Differences will appear.
Missing Sensors – The Missing Sensors list will indicate any
previously connected devices that are no longer detected. Select
Retain to keep the identification information for a previous
device; select Remove to delete it.
Replaced Sensors – The Replaced Sensors list displays any
newly added sensors that have replaced Missing Sensors that
have been Retained.
Additional Sensors – Displays any newly detected devices.
Following a scan, selecting NEXT from Sensor Differences will
navigate to Configure Measurements.
Configure Measurements Navigate to Configure Measurements to begin setting up mea-
surement parameters for each TIENet device detected.
To activate a measurement, highlight the radio button next to it
and press Enter. To change the name of the measurement, high-
light the name and press Enter.

Regardless of what you name them, the measurement


parameters for each device remain the same. For your ref-
erence, Figure 2-9 on the following page shows the position of
each measurement for each type of TIENet device.

2-12
Signature® Flow Meter
Section 2 Setup and Programming

300 Case Board:


Input Voltage Sense Voltage
Charge Voltage Charge Current
Case Humidity Reference Humidity
AC Power Rainfall
Case Temperature

301 pH / Temperature Device:


pH Temperature

304 Contact Output:


304 Digital Output 304 Digital Output-

306 Sampler Interface:


Sample Event / Bottle # Input Voltage

307 Analog Input Card(s):


Analog Input Analog Input-A

308 Analog Output Card(s):


Ch1 Current (Amps) Ch1 Percent of Scale (%)
Ch2 Current (Amps) Ch2 Percent of Scale (%)

310 Ultrasonic Level Sensor:


Distance Ultrasonic Level
Air Temperature Signal Strength

330 Bubbler Level Sensor:


Bubbler Level Bubble Air Humidity Case Temperature

Elapsed Time Elapsed Time-A

350 Area Velocity Sensor:


Level Sense Voltage Vel Spectrum Ratio
Velocity Temperature
Velocity Spectrum Velocity Signal

360 LaserFlow Sensor:

Distance Ultrasonic Signal Case Temperature Air Temperature


Velocity Sense Voltage X-Axis
Laser Temperature Doppler Power Laser Diode Current
Y-Axis Level Window Temperature
Figure 2-9 Measurement parameters for each TIENet device

2-13
Signature® Flow Meter
Section 2 Setup and Programming

2.6.2 SDI-12 Setup Sondes detected since the last scan are displayed, with the acti-
vated sondes in the top box. If SDI-12 devices have been added or
removed from the system, select Scan to detect the current
system configuration. Following the scan, add/remove sondes
from the Active list by selecting the sonde and clicking Add or
Remove. To begin using an Isco-Ready sonde with its configured
parameters, select Configure.

2.6.3 Modbus Input Setup Connect the external Modbus input device at one of the three
TIENet terminal connectors on the Signature case board as
described below:
Modbus In
D1 = Yellow (+)
D0 = Brown (–)
Gnd = Black

Table 2-1 on the following page provides an example of Modbus


settings for a connected DGH analog converter. The letters in the
left column correspond to the entry fields shown in Figures 2-10
and 2-11.

The multiplier and offset are used to scale the raw number
coming from the Modbus register(s) to represent the data in the
units of measure you specify, as expressed in the following
equation:
H in units of measure = (register value * J) + K.

In this example, the current input represents a flow rate where:


4mA = 0cfs, and 20mA = 10,000cfs
The D1252M documentation states that it produces a register
value of 0 at 0mA, and 65535 at 25mA. This means that at 4mA
the register will report 10485, and at 20mA it will report 52428.

The multiplier (J) is calculated as follows:


10,000/(52428 – 10485) = -2500

Before setting up the Modbus input function, it is recommended


that you print Table 2-1 and use the empty columns provided on
the right to fill in your own Modbus information.

2-14
Signature® Flow Meter
Section 2 Setup and Programming

Examples: Table 2-1 Modbus Setup Worksheet


Manufacturer DGH Manufacturer Manufacturer
Model D1252M Model Model
A Protocol (ASCII/RTU) ASCII Protocol (ASCII/RTU) Protocol (ASCII/RTU)
B Baud Rate 9600 Baud Rate Baud Rate
C Data Bits 8 Data Bits Data Bits
D Parity None Parity Parity
E Stop Bits 1 Stop Bits Stop Bits
F Device Name D1252M Device Name Device Name
G Address 11 Address Address
H Parameter Flow Rate X Parameter Parameter
I Addressa (Register) 30001 Address (Register) Address (Register)
J Multiplierb .238422 Multiplier Multiplier
K Offsetb -2500 Offset Offset
L Byte Orderc (Endian) Little Byte Order (Endian) Byte Order (Endian)
M Data Size (Format) Unsigned Word Data Size (Format) Data Size (Format)
N Data Type Flow Rate Data Type Data Type
O Units m3/s Units Units

a. For 2100 update interval in seconds must be written to register 26.


b. For assistance in calculating a multiplier and offset, contact Teledyne Isco.
c. Big Endian = Most significant register first; Little Endian = Least significant register first.

To begin configuring Modbus communication protocol and


devices, select MODBUS Input COM Settings and use the pull
down menus.
Add/Edit Device Select Modbus Input Device Settings. Enter the device name,
and the device address. For Request Timeout, enter a connection
retry interval in milliseconds, and the number of attempts before
a connection failure is determined.
Configure Modbus communication protocol using the pull down
menus. To add/edit parameter(s) for the device, select Edit
Parameters.

2-15
Signature® Flow Meter
Section 2 Setup and Programming

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

3.

MODBUS Input Setup


1. 1. MODBUS Input COM Settings 2.
2. MODBUS Input Device Settings

MODBUS Input Devices


MODBUS Input COM Settings
1. D1252M (device you have already created)
Protocol Type: ASCII (ASCII or RTU) A 2. Create New Device
Baud: 9600 (2400 bps to 115200) B
Bits: 8 (8, 7, or 6) C
Parity: NONE (none, even, or odd) D
Stop Bits: 1 (1 or 2) E

MODBUS Device - Create New Device [or select existing device]


Select NEXT to save device and
parameter changes.
Device Name D1252M (Rename or F
Enter New)
Address 15
G
Request Timeout 10 ms
Retries 3
Only appears
Continued in
Edit Parameters Remove for existing
Figure 2-11
devices

Figure 2-10 Modbus Input Setup

Note
External Modbus RTU devices cannot use addresses 1
through 10.

Add/Edit Parameters Select an existing parameter to edit, or select Add to add a new
parameter for the device. Enter a name (such as a data type), and
the register address. Use the pull down menus to select other
parameters.

2-16
Signature® Flow Meter
Section 2 Setup and Programming

Select Little if a multiple register parameter has the low-order


data in the first register; select Big if high-order. Select data size.
The available Units of Measure are determined by the data type
you select.
If necessary, enter a Multiplier and Offset so the register reports
a value in the units specified.

Edit Device Parameters - [parameter name]


Go BACK to the Modbus Device page in
order to save changes.
1. [Existing parameter]
2. Add

Edit Parameter
Name Flow Rate X H

Address 30001 I

Multiplier .238422 (to scale the register) J

(offset applied to K
Offset -2500 scaled register)
Byte Order LITTLE (little or big) L

Data Size Unsigned Word (bit, bytes, signed, M


unsigned, float)
Data Type Flow Rate N

Units m3/s (available UOM determined O


by data type selected)

Figure 2-11 Editing Modbus device parameters

2-17
Signature® Flow Meter
Section 2 Setup and Programming

2.6.4 Modbus Output Setup The Modbus RS-485 output function enables a SCADA system to
retrieve site data from the flow meter. Connection to the flow
meter is made via the RS-485 terminal on the Signature case
board (shown in Figure 2-12 below).
In the Device ID field, enter the Signature’s address (from 2 to
247) and configure the communication protocol.

CAUTION
Be careful not to assign the same address to more than one
flow meter.

For Modbus data register numbers and definitions, as well as a


general explanation of Modbus output protocol, refer to
Table C-1, in Appendix C Modbus Output Protocol.

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

4.

MODBUS Output Setup


Signature address
(Device ID from 2 to 247) Device ID:

Protocol Type: ASCII (ASCII or RTU)


Baud: 115200 (2400 bps to 115200)
Bits: 8 (8, 7, or 6)
Parity: NONE (none, even, or odd)
Stop Bits: 1 (1 or 2)

Figure 2-12 Modbus Output Setup

2.6.5 Modem Setup The menu choices displayed for modem setup depend on which
modem option is installed in the flow meter. For detailed infor-
mation about installation and operation of Ethernet, GSM, and
CDMA modems, refer to Sections 5.9 Ethernet Modem and 5.10
Cellular Modems.

2-18
Signature® Flow Meter
Section 2 Setup and Programming

2.7 Configure Options The Configure Options menu is used for setting up the mea-
surement site and setting the program parameters.

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Site Setup Options 1. Data Options


1. Set Clock 5. 1. Data Storage
2. Site Name 2. Data Push
3. Home Display
4. Default Units Sampler Setup
6.
5. Power Options 1. 306 <serial #> Sampler Interface

Measurement Setup Inputs/Output/Alarms Setup


2. 1. Alarm
1. Level Input Setup 7.
2. Analog Output
Bubbler - Set purge interval
Ultrasonic - Set blanking distance
2. Flow Rate Input Setup
Reset Totalizers
1. Total Flow
3. Rainfall Input Setup
2. Total Flow 2
4. Volume Input Setup
3. Total Flow 3
8.
4. Total Flow 4
Adjust Options 5. Rainfall
3.
1. Level
2. Velocity Report/History Setup
3. pH (301) 1. Reports 2. History
Report Setup 1. View History
Equations Setup 1. Report 1 2. Graphical
4. 9.
1. [Edit existing equation]
2. Report 2 Historical Display
2. Create New Equation

Figure 2-13 Menu Tree: Configure

2.7.1 Site Setup The Site Setup menu sets some basic operating characteristics
specific to the site.
Set Clock – Enter Year, Month, Day, Hour, Minute.

2-19
Signature® Flow Meter
Section 2 Setup and Programming

Site Name – Press Enter to display the character grid. Select


one character at a time to create the desired site name.
Home Display – The Home Display determines how current
measurement data is displayed on the Home screen.
From the Measurements Setup screen, select all measurement
parameters to be displayed. The parameters available in the pull
down menus will be determined by what devices are connected to
the Signature meter.
Default Units – To set units of measure for each parameter,
first select the parameter from the menu list.
The available units of measure that appear will be determined by
the parameter you have selected. Under Units, highlight the
units of measure and press NEXT. When finished, press NEXT
again to save and exit.
Power Options – There are two power options:
• Display: will allow you to shut off the backlight and the
screen.
• Low Battery Cutoff: will allow you to adjust the low
battery cutoff. Refer to Section 4-4 for more power infor-
mation.

2.7.2 Measurement Setup This menu is for setting up the level measurement (Level Input
Setup), flow conversion (Flow Input Setup), and flow volume
totalizer(s) (Volume Input Setup). Menu items that appear are
dependent on what equipment is connected to the Signature flow
meter.
Level Input Setup – Under Level Setup, select the level input.
Usually there will only be one listed, unless your system is using
more than one level measurement device.For the TIENet 310
ultrasonic sensor, the minimum blanking distance refers to the
maximum water level, and the maximum blanking distance
refers to zero water level in the channel. For detailed instructions
about 310 setup, refer to the TIENet 310 Installation and Oper-
ation Guide.
For the TIENet 330 bubbler, the purge function is a periodic
burst of air forced through the bubble line to keep it free of
debris. Select the purge interval between 15 minutes and 8 hours
from the pull down menu. –
Flow Rate Input Setup – Measurement settings and flow con-
version are programmed for the flow rate(s) from this menu
(refer to Figure 2-14 on the following page). If more than one flow
rate data set is being calculated, these settings are programmed
separately for each one.
1. Select the flow rate to configure or select "Add Flow Rate"
when two or more flow rates are present. Adding a flow
rate enables calculation of flow rates dependent on condi-
tion. The flow rate also allows for configuring standard
flow conversions from level inputs (e.g. level from Analog
Input).

2-20
Signature® Flow Meter
Section 2 Setup and Programming

2. For level-to-flow conversions, from Measurement Settings,


select the Level Input to be used in the flow calculation and
the Measurement Rate (interval). Enter the name for this
flow rate.
3. Select the flow conversion type to be used (Weir, Flume,
Metering Inserts, Manning Formula, Area Velocity, Equa-
tion, or Data Points); then set up the conversion.

Note
Additional information about flow conversions can be found in
the Isco Flow Measurement Handbook included with the Sig-
nature Flow Meter.

2-21
Signature® Flow Meter
Section 2 Setup and Programming

Configure Options

2. Measurement Setup
(example name)

Flow Rate Input Setup


1. Flow Rate-A
2. Flow Rate-B
3. Flow Rate-C
4. Add Flow Rate

Flow Rate-A
1. Measurement Settings
2. Flow Conversion

For level-to-flow conversions only.

Measurement Settings Flow Conversion


4. Manning Formula
(Select level input, Interval, Flow Rate-A
5. Area Velocity
& Flow measurement name.) 1. Weir
6. Flow Equation
2. Flume
7. Level to Flow Data Points
3. Flow Metering
8. Level to Area Data Points
Insert

1. 3. 4.
7, 8.

Weir Flow Conversion Metering Insert Manning Formula Edit Data Points
Flow Rate-A Flow Conversion Channel Flow Conversion • Level to Flow or Area
1. V-Notch Flow Rate-A 1. Round (Select units of measure,
2. Rectangular w/ End 1. V-Notch 2. U Shaped add/remove points.
Contractions 2. Round Orifice 3. Rectangular 3 to 50 pairs of points
3. Rectangular w/o End (enter size) 4. Trapezoidal may be used.)
Contractions
4. Cipolleti (enter dimensions)
5. Thel-Mar
5.
(enter size)
Area Velocity Channel Flow
2.
Conversion
Flume Flow Conversion
Flow Rate-A 4. Trapezoidal
Flow Rate-A 1. Round 5. Elliptical
1. Parshall 2. U Shaped (enter silt level
2. Palmer-Bowlus 3. Rectangular & dimensions)
3. Leopold-Lagco 6.
4. HS
5. H British Flume Flow Conversion Flow Equation
6. HL 1. Standard Flow Rate-A
7. Trapezoidal 2. Venturi (Select units of measure, enter
(enter size) 3. Khafagi one- or two-term polynomial
8. British Flume equation.)
(enter size)

Figure 2-14 Menu Tree: Flow rate input setup

Volume Input Setup – You can set up one to four Total Flow
measurements. Select the flow rate(s) used for total volume, the
totalizing method (Net, Positive, or Negative), and the interval at
which the total flow will be updated (between 30 seconds and 24
hours).

2-22
Signature® Flow Meter
Section 2 Setup and Programming

From the Resolution pull down menu, select the degree of reso-
lution required for your total flow (lower = fewer digits to right of
decimal; higher = more digits to right of decimal).

Configure Options NOTE:


Pres Next or Enter
to go to th next screen.

2. Measurement Setup

3. Volume Input Setup

Volume Setup
1. Total Flow
2. Total Flow 2
3. Total Flow 3
4. Total Flow 4

Total Flow
VOLUME
Input Flow Rate: Flow Rate-A

Totalizer: Net
Resolution: 9999999.99
NOTE:
Update Total Flow Every: Effects power
30 Seconds consumption
Figure 2-15 Menu Tree: Volume Input Setup (total flow)

Rainfall Input Setup – This option allows you to setup the per
tip options for the rain gauge. The Automatic roll-over can be set
to the time of day the rollover occurs. Once this number is
reached, the roll-over will go back to zero.

300 Rainfall
SELECT RAINFALL PER TIP
0.01 inch
0.1 millimeter
User Entered 0.00 inches
Automatic roll-over: 00:00:00

Figure 2-16 Menu for Rainfall Input Setup

2-23
Signature® Flow Meter
Section 2 Setup and Programming

Load Calculation Setup – This option allows you to setup the


.....

Figure 2-17 Menu for Load Calculation Setup

2.7.3 Adjust Adjust levels and/or velocity measurement, and/or calibrate mea-
surement values for other connected TIENet devices. Level
adjustment instructions can be found in Section Setting the
Level, on page 3-18.

Note
For detailed instructions on calibration of a connected TIENet
301 pH device through this menu selection, refer to the 301
user manual.

Note
For detailed instructions on laser velocity measurement setup
through this menu selection, refer to the LaserFlow (360) user
manual.

2.7.4 Equation/Trigger Conditions are sets of site-specific, user-defined parameters.


Setup Refer to Figure 2-18 on the following page.
Equations are created from various site conditions that can be
used to generate alarms, log or push data at secondary rates,
trigger a connected sampler, or conserve power by turning on
equipment only when needed.
Types of conditions There are five types of conditions provided:

2-24
Signature® Flow Meter
Section 2 Setup and Programming

Range – TRUE when a measured parameter value is inside or


outside specified upper and lower limits.
Rate of Change – TRUE when a measured parameter changes
by a specified amount over a specified time duration.
Sensor Error – TRUE when a sensor error is present for a spec-
ified time duration.
Threshold – TRUE when a measured parameter reaches or
exceeds a user-defined setpoint.
Time Table – TRUE when the flow meter’s internal clock is
within a defined time duration. This may be a weekly, daily, or
specific one-time stop/start time.
Rain Event – TRUE when threshold is met over a period of
time. Meeting threshold during the dry period will reset the Dry
Period timer.
Defining conditions To define a condition:
1. Highlight the desired condition in the lower left corner of
the screen (Conditions A-F).
2. Highlight Edit Condition and press Enter.
3. Scroll down and press Enter to select the type of condition
(listed above). Press NEXT to go to the configuration
screen for that condition type. Press NEXT when complete.
The condition in the left-hand corner of the screen will now show
the condition type.
Building equations Build or modify the equation by navigating to the desired condi-
tions and operators. Highlight Select Condition and press Enter
again to add it to the equation. Press Enter to add a highlighted
operator.
Press NEXT when complete.
Measurement Interval When the screen is off and measurement interval is less than the
data storage rate, the readings will be averaged and the battery
life will slightly decrease.

2-25
Signature® Flow Meter
Section 2 Setup and Programming

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Equations Setup
Equation that you have
1. Equation XYZ already created.
2. Create New Equation

1. 2.

Modify Equation XYZ Modify the equation.


Edit
Name: XYZ
Delete

A or B ( ) Bksp
AND & OR | NOT !

Condition A (Sensor Error) Measurement Interval:


Condition B (Threshold) 5 minutes
Condition C
Select Condition
NOTE: Condition D
Power consumption is effect by Edit Condition
measurement intervals and alarm Condition E
triggering with the screen is off. Condition F

Figure 2-18 Example of defining conditions & building


equations.

2.7.5 Data Storage/Push Data Storage – Set up data storage rates for a group of mea-
Setup surements, or separately for individual measurements. Scroll to
the bottom of the screen to set up the primary storage rate, and a
secondary one, if needed, with its trigger equation.
The display can be turned OFF and ON in the configure options,
site setup, and power options menu. Turning the display O FF
effects the way the data is recorded by the Signature meter.

2-26
Signature® Flow Meter
Section 2 Setup and Programming

• When the display is ON the Signature takes an average


of the measurements over the data storage interval and
records it.
• When the display is OFF the Signature takes a reading
at the reading interval set in the data storage screen (no
average is taken) unless alarms, triggers, or secondary
measurements are active.
• When the display is OFF and if alarms, triggers, or
secondary measurements are set up and the intervals
are shorter than the data storage rate, the readings
recorded by the meter will be an average of readings
taken when the alarms, triggers, or secondary measure-
ments were taken.
Data Push – Set up the flow meter to push data to a server
running Isco Flowlink Professional software (internal modem
required).

2.7.6 Sampler Setup Program the flow meter to trigger and pace a sampler, and
receive sampling information from the sampler.

Note
For detailed instructions on configuring a connected TIENet
306 Sampler Interface through this menu selection, refer to the
306 user manual.

2.7.7 Inputs/Outputs/Alarms Alarm – Configure Local, SMS text, or Server alarms based on
Setup user-defined site conditions. Under Alarm, select an alarm from
the list or set up a new alarm.
Next, select an Alarm Trigger from the pull down list.

Note
The trigger(s) listed in the Alarm Trigger pull down list consist
of equations you have already created based on your defined
site conditions (refer to Section 2.7.4 Equation/Trigger Setup).

Alarms: Local Local alarms are viewed on the Signature Flow Meter itself.
When a programmed alarm condition becomes true, the LED on
the front panel glows red.
To view the alarm message, press the Alarm softkey ( ).
For local alarm setup, refer to Figure 2-19 on the following page.

Alarms: SMS / Server SMS and Server alarms require an optional internal modem. To
configure your modem for communication, refer to Section 5.9
Ethernet Modem or Section 5.10 Cellular Modems.

2-27
Signature® Flow Meter
Section 2 Setup and Programming

Note
Server alarms notify a specified list of contacts in the event
that a server running Flowlink Pro fails to receive pushed data
from a site within a specified duration.

For SMS and Server alarm setup, refer to Figure 2-20.

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup
7.

Inputs/Outputs/Alarms Setup
1. Alarm
2. Analog
3. Input/Output
4. Contact Output

Alarm
1. Alarm 1 Select an alarm from the list or
2. Alarm 2 set up a new alarm.
3. Alarm 3
4. Alarm 4

ALARM 1
ALARM Select a trigger (equation) from
Alarm Trigger: 1 the list. You must have at least
Alarm Notification: one equation saved in order to
Local
trigger an alarm.
Delete Alarm

LOCAL

Configure ALARM 1 Your message will appear on the screen


LCL if the alarm is triggered and you press
Message: message here the Alarm softkey (C).

Figure 2-19 Local alarm setup

2-28
Signature® Flow Meter
Section 2 Setup and Programming

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Inputs/Outputs/Alarms Setup
1. Alarms
2. Analog
3. Input/Output
4. Contact Output

Select an alarm from the list


Alarm or set up a new alarm.
1. Alarm

Select a trigger
ALARM (equation) from the list.
ALARM You must have at least
Alarm Trigger: 1 one equation saved in
Alarm Notification:
SMS
order to trigger an
alarm.

SMS SERVER

Configure ALARM Configure ALARM


SMS SERVER
Message: message here Message: message here
Retry Interval: 0 Minutes Retry Interval: 0
Retries: 0 Retries: 0
Server Group: 1
Phone #1: Phone #2:

Phone #3: Phone #4:

Phone #5:

Figure 2-20 SMS and Server alarm setup (modem required)

Analog Output – Select the output to configure, then select and


configure the measurement the output will represent. TIENet
308 option card required (see Section 5.6.10).
Analog Input – Select the input to configure, then select and
configure the measurement the input will represent. TIENet 307
option card required (see Section 5.6.9).
Contact Output – Select the output to configure, then select
and configure the measurement the output will represent.
TIENet 304 option card required (see Section 5.6.8).

2-29
Signature® Flow Meter
Section 2 Setup and Programming

2.7.8 Reset Totalizers Select the flow volume totalizer(s) to be reset. Selection resets
the totalizer to zero.

2.7.9 Reports/History Setup Reports – Set up report interval and measurements to include
for one or two reports. To include all user and meter events in the
report, select the option Include history log.
History – Display user and meter events (i.e., log-ins, adjust-
ments, data push, etc.), and/or set up the graphical display for
selected measurements over a period of time.

2.8 Administration Administrative settings (see following page) dictate operating


preferences and perform general housekeeping tasks.

Administration Options
1. Language Options 5. Display Signature Information
2. Set New Passcode 6. Display License Information
3. Update Firmware 7. Gather Fault Data / USB Options
4. Sensor Diagnostics 8. Restore To Factory Defaults

1. Sensor Diagnostic Options


Language Options 4.
1. Retrieve Diagnostics
1. Select Device Language
2. Review Diagnostics

Device Information
5. Sensor Information
(Firmware & Serial #s)
Choose the language
for this device.
(Select from list.) Retrieve Diagnostics
(select TIENet device from list)
Set default units of measure

x - 3xx Diagnostics Information


Diagnostic Text
(diagnostic results shown)

If no current
passcode: Please enter your new Review Diagnostics
passcode. (select TIENet device from list)
New Passcode: ____________
2.
(TIENet Device Name)
(select time from list)
To change Please enter your new
current passcode. License Information
passcode: 6.
Old Passcode: ____________ (Licensing & copyright details)
New Passcode: ___________

Fault Data Options


Firmware Update Options 7.
1. Display Fault Data
3.
Connect flash drive to front panel
to use this function.
About to restore factory
8.
defaults
(Erases all data; press Back to cancel)

Figure 2-21 Menu Tree: Administration

2-30
Signature® Flow Meter
Section 2 Setup and Programming

2.8.1 Language Options Select Device Language – From the list, select the default lan-
guage to be displayed by the Signature flow meter. Available lan-
guages include:

English (USA) Dansk Português (Brasil) Deutsch


English (International) Nederlands Español (México)
Français Türk Svenskt

From this screen, you can also set general default Units Of
Measure to USA or International. You can still select units of
measure for individual measurements through Configure
Options > Site Setup.

2.8.2 Set New Passcode A numerical passcode can be set to protect access to pro-
g r a m m i n g a nd d a t a . B y d e f a u l t , t h e f l o w m e t e r i s n o t
passcode-protected. To change an existing passcode, enter the
current passcode, and then the new passcode. To remove the
passcode requirement, enter a new passcode of 0 (zero).

2.8.3 Update Firmware To use this function, first connect a flash drive loaded with the
correct firmware update file(s) to the micro-USB assembly. Refer
to Section 2.9 USB Options for complete instructions.

2.8.4 Sensor Diagnostics The Signature provides operating data for each connected
TIENet device upon request, for site evaluation or trouble-
shooting purposes.
To generate a new diagnostic report, select Retrieve Diag-
nostics. The available devices can then be selected from a list.
When the diagnostic data is available, press Enter to view it.
To view past diagnostic reports, select Review Diagnostics and
select from the list of devices. Each past diagnostic report will be
listed by date and time.
Diagnostic reports can also be downloaded to a connected USB
flash drive in the form of a text file. Refer to Retrieve Text
Reports, on page 2-33 for complete information.

Note
If you are running diagnostics on a 360 LaserFlow velocity
sensor, the Distance in the text report and graph are relative to
the face of the integral ultrasonic transducer, which is located
11.7" above the bottom of the sensor. Subtract 11.7" from the
Distance shown to obtain the actual value.
For more information about the LaserFlow sensor, refer to
Section 5.6.3 Laser Doppler Velocity Sensor and the Laser-
Flow user manual.

2.8.5 Display Signature Select this option to view the serial number, model number,
Information software revision, and hardware revision of the Signature flow
meter and any connected TIENet devices.

2.8.6 Display License This selection displays open-source licensing and copyright infor-
Information mation for the Signature firmware.

2-31
Signature® Flow Meter
Section 2 Setup and Programming

2.8.7 Gather Fault Data Fault data is a term describing the capture of any user and/or
flow meter activity prior to and during a system error or failure.
This data can be displayed to assist in troubleshooting. To
download the data to a flash drive, connect a flash drive to the
micro-USB assembly on the front panel of the Signature and
select Gather Fault Data from the USB Options menu that
appears (refer to Section 2.9 USB Options).

2.8.8 Restore to Factory This function returns the Signature Flow Meter to the program
Defaults that was installed the factory. This is an example program that
can be used for reference when designing your own program.
If the restore to portable defaults check box is selected, the power
savings options will be automatically turned on.
Be sure to record your own program settings and save all
data before restoring the factory defaults.

2.8.9 Home HOME returns to the home screen.

2.9 USB Options The USB Options menu will only appear when you connect a
flash drive to the micro-USB assembly on the front panel of the
Signature.

USB Options
Please choose the operation you want
to perform:
1. Retrieve Text Reports
2. Retrieve Data
3. Update Firmware
4. Save Current Program
5. Load Existing Program
6. Save Signature Information
7. Gather Fault Data
8. None of the above

1. Select the reports you 4. Save Current Program


Saves the Signature’s current program settings
want to retrieve
to your flash drive. Press NEXT when prompted.
• All reports
• All reports since:
5. Load Existing Program
[Enter date]
• All reports in range: Select program to load and press Enter.
[Enter dates] Overwrites current Signature program.

6. The information was stored as


2. Select the data you 1:/device.htm
want to retrieve Saves device firmware version(s)
and serial number(s).
• All data
• All data since:
[Enter date] 7. Fault Data Options
• All data in range: 1. Display Fault Data
[Enter dates] 2. Retrieve Fault Data (saves fault
• Additionally export as .csv data to flash drive)

3. Firmware Update Options 8. None of the Above


1. Update Signature Firmware Returns to top menu. Note:
2. Update Smart Sensor (TIENet) Firmware USB Options menu still accessible To view all Administration
through Administration menu. menu selections, refer to
Firmware Update File igure 2-19.
1. (select firmware file from pull down
list and press NEXT)

Smart Sensor (TIENet) Firmware


2. Update
(select all TIENet devices to update
and press NEXT)

2-32
Signature® Flow Meter
Section 2 Setup and Programming

Figure 2-22 Menu Tree: USB Options

2.9.1 Retrieve Text Reports Select this function to download the Signature data text reports
for sensor diagnostics and verification of data integrity.
Select “All reports,” or specify a start date or date range, and
press NEXT. The reports will be stored on the connected flash
drive in a folder called "ISCO."
This folder contains a sub-folder for each site, named with the
first eight characters of the site name. The site folder contains
one or more sub-folders, named by retrieval date (YYYYMMDD).
Each report file name has a prefix of one or two letters, followed
by four digits representing the time of day, e.g., PH0935.TXT is a
program report generated at 9:45 a.m.
Each report is contained in a text file, with a .TXT extension, or
in a Smart Sensor Diagnostic file, with a .SSD extension. Each
report file has a corresponding authentication file, with a .ath
extension, for verification purposes. For more information about
report verification, refer to Verifying Exported Reports, on page
2-35.
The four validation report types are:
Program Report (PH) – Tracks changes to the Signature
Meter’s configuration
Summary Report (R1 / R2) – Documents summaries of data
measurements (e.g. Min/Max/Avg)
Diagnostic Report – Tracks the occurrence of, and results
from, diagnostic tests
History Report (H) – Tracks user and meter events (e.g. level
adjustments, calibration, data push, etc.)

2.9.2 Retrieve Data The program settings and flow data can be downloaded onto your
flash drive in .ddp (data dump) format.
Select “All data,” or specify a start date or date range, and press
NEXT. The data will be stored on the connected flash drive in a
folder called “ISCO.”
This file can then be imported into Flowlink, where it can be
viewed in regular site file format, with the recorded data and
report/graphing capability.
There is a check box that, when checked, will export the .cvs file
onto a USB flash drive.

2.9.3 Update Firmware With a USB flash drive connected to the Signature’s front panel,
the Update Firmware option becomes active on both the Admin-
istration menu and the USP Options menu.
For step-by-step instructions for updating the firmware for either
the Signature flow meter or connected TIENet devices, refer to
Firmware Updates, on page 6-2.

2-33
Signature® Flow Meter
Section 2 Setup and Programming

2.9.4 Save Current Program Select this option to save a copy of the Signature’s current
program to your USB flash drive. You can also use this infor-
mation to program other meters with the same configuration.

2.9.5 Load Existing Program Select this option to load a saved program from your flash drive.
Note that selecting this option will cause the current program to
be overwritten with the one from the flash drive.
In order for the Signature to load the correct program, the name
of the site must match that of the site program that was saved.

2.9.6 Save Signature This option saves a snapshot of the firmware version(s) and
Information serial number(s) of the Signature and any connected TIENet
devices.

2.9.7 Gather Fault Data Fault data is a term describing the capture of any user and/or
flow meter activity prior to and during a system error or failure.
This data can be viewed and/or downloaded as a file to your flash
drive to assist in troubleshooting.

2.9.8 None of the above This item returns the screen to the top menu.
However, as long as the flash drive remains connected, the USB
Options menu will still be active, and can be reopened from the
Administration menu.

2.10 Signature Data in To download flow and event data from the Signature Flow Meter
Flowlink into the database with Flowlink, connect to the flow meter and
select Interrogate (F8), or Import ddp (data dump).

2.10.1 Event Viewer By default, Flowlink’s Event Viewer displays the four event data
types in tabular format, with a time stamp and short description
of the event for each entry (refer to Figure 2-23). Each event type
is represented by a graphical symbol, located in the first column:

Program

Summary Reports 1 & 2

Diagnostic

History (User / Meter Actions) Meter Actions include Data

Push - start/fail/complete, and Power Up/Down; all other events


are user events (i.e., calibration, changes to the program,
totalizer reset, etc.).

2-34
Signature® Flow Meter
Section 2 Setup and Programming

If the data for that event is unaltered, a green check mark

appears next to it. If the data cannot be verified as authentic, a


red slash appears next to it.

Figure 2-23 Event Viewer in Flowlink

Printing reports Select one or more rows in the viewer to be printed and then
select the Print button.
Exporting reports To save event data as a text report for future verification, high-
light the desired row(s) in the Event Viewer table and click
Export (or right-click and select Export).
The default destination is your My Documents folder; however,
you can change this to another preferred destination, including a
USB drive, if preferred. A message window will notify you when
the export is complete.
The files you exported are saved using the following hierarchy:
SITENAME \ MODULENAME \ DATE.

2.11 Verifying Exported Flowlink will only export already verified reports; they can also
Reports be verified after being retrieved from the Signature, either via
Flowlink export or USB flash drive download.

Verification of exported data reports is done using the Report


Verification tool, a small application installed separately when
Flowlink was installed. This tool is located in the Flowlink
program folder, normally at C:\Program Files\Flowlink 5.1, and

is identified with a traffic light icon .

Note that this tool can also be used to verify data exported
directly to a USB drive using the USB Options menu (refer to
Section 2.9).

2-35
Signature® Flow Meter
Section 2 Setup and Programming

Use the top Browse... button to navigate to the desired report


(*.txt or *.SSD) file. Use the bottom Browse... button to navigate
to its corresponding authentication (*.ath) file. Click Verify. The
application will quickly return a message showing the verifi-
cation result.

Data is
authentic.

Data has been


altered.

Figure 2-24 Report file verifier

2-36
Signature® Flow Meter
Section 3 Standard Installation

This section contains physical preparation procedures and per-


manent mounting methods for the Signature Flow Meter and
associated Teledyne Isco equipment. Section 4 will contain the
installation information for the Signature Portable Flow Meter.

WARNING
The installation and use of this product may expose you to
hazardous working conditions that can cause serious or
fatal injury. Take all necessary precautions before entering
a worksite. Install and operate this product in accordance
with all applicable safety and health regulations and local
ordinances.

3.1 Introduction The Signature Flow Meter is used for permanent installations.

3.2 Connecting External External device cables and mains line cord are passed, usually
Devices via conduit or cord-grip fittings, through the port holes in the
bottom of the case and wired directly to the connector case. Cable
fittings are also available for Rain, Ethernet and option card cir-
cuits as a 3 hold cord grip.
Trimming unterminated wires prior to installation is recom-
mended.
Tools Required:
Small flat screwdriver (3.5mm)
#2 Phillips screwdriver
Channel locks
Soldering iron (for tinning wires)

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.

CAUTION
Before opening the case, disconnect the optional battery or
battery backup power, if used.

3-1
Signature® Flow Meter
Section 3 Standard Installation

Note
Before restoring mains power, ensure that the flow meter’s
USB connector does not have a cable attached.

Open the door to access the two large screws holding the front
panel on the connector case. Remove the two screws, then
reinsert them in the front panel and latch the lid so they will not
be misplaced.

Figure 3-1 Open door and front panel to access interior

Open the front panel to access the connector case. Connectors on


the board are identified in Figure 3-2.

3-2
Signature® Flow Meter
Section 3 Standard Installation

A External TIENet
Devices/
I
K A Modbus RS485 In
G D1 = Yellow (+)
D0 = Brown (–)
H Gnd = Black
B SDI-12 Input
I J
L Modbus
C RS485 Out

L D TTL Serial
J J Battery Backup/
A A E Battery for Portable

B D E F F 12.8 VDC
C
K K G Ethernet Modem
M
H Ethernet Port
B
Cellular Modem
I (power & serial)
4-20mA Input/Out-
put Card and
J Contact Output
Note: No power supply for Signature Portable Flow Meter
Card
K Fuse "T" 3.15A

L Fuse "T" 4.00A

M Rain Gauge

Figure 3-2 Connector case, connectors, and fuses

Note
The two TIENet connections satisfy most applications. If addi-
tional TIENet connections are needed, consider the two
TIENet Expansion Box versions for standard installations or
TIENet 'Y' cables for Portable Signature applications. See
Appendix B for ordering information.

3.3 Case Bottom Cable The connections made through the cable entries depend on the
Entries application, but their most common uses, in accordance with the
connector case layout, are depicted on the following page.
All optional cable entries must use appropriate ID conduit con-
nections or cord-grip fittings to retain the IP66 rating. If you are
using non-TIENet or non-Signature cables, you must supply the
appropriate ID conduit connections or cord-grip fittings.

CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and

3-3
Signature® Flow Meter
Section 3 Standard Installation

moisture from entering the Signature enclosure. Failure to seal


conduit could reduce equipment life.

TIENet
Cellular Device or Battery
Bubble Line Fitting Wired Backup or Main
Antenna or Power
Wired Options or TIENet Device Options Wired Options

See Section 5.10.3 for


cellular modem instal- See Section 5.2 for battery
lation instructions. backup installation
instructions.

See Section 6.7 for bubbler


installation/replacement
instructions.

Figure 3-3 Connector Case cable entries for power and external devices

3-4
Signature® Flow Meter
Section 3 Standard Installation

3.3.1 Cable Fittings Cord-grip fittings for TIENet devices, line cord, and battery
backup option are available from Teledyne Isco (see Appendix B
for ordering information). Cable fittings are also available for
rain, ethernet, and optional card circuits as a 3-hole cord grip.
The fitting for the line cord is a special strain-relief fitting, as
shown in Figure 3-5.

Figure 3-4 Strain relief 3/4 NPT Cord-grip fitting for TIENet devices

Figure 3-5 Cord-grip fittings installed

3-5
Signature® Flow Meter
Section 3 Standard Installation

Any unused cable entry holes should be sealed with plugs. Do not
overtighten the plugs. When a plug is flush against the outside of
the case and held in place by the metal nut inside, the hole is
sealed.

O-Ring

Plug is flush against


outer wall.

Figure 3-6 Diameter-seal plugs for unused ports

3-6
Signature® Flow Meter
Section 3 Standard Installation

3.3.2 Connecting TIENet The optional external TIENet devices compatible with the Sig-
Devices nature flow meter all connect in the same manner. Multiple
TIENet devices can be connected simultaneously to the same Sig-
nature Flow Meter.

Note
The steps that follow include instructions for installing cord-grip
fittings. Some applications will use user-supplied 3/4" ID con-
duit for cable routing.

1. Remove one of the 6-position plug-in terminal strip connec-


tors from the connector case.

Figure 3-7 TIENet Device terminal strips

2. If using a cord-grip fitting, install the cable nut in the


appropriate opening on the bottom of the Signature enclo-
sure, securing it to the wall with the lock nut (concave side
facing wall).
3. Feed the TIENet device cable end through the sealing nut
and seal, and through the cable nut. Lightly tighten the
sealing nut, just enough to hold the cable in place while
installing the connector.

3-7
Signature® Flow Meter
Section 3 Standard Installation

Lock Nut
(concave side Sealing Nut
facing wall)
Seal
Cable Nut (color may vary)

Figure 3-8 Installing TIENet cable with a cord-grip fitting

4. Attach the wire ends to the terminal strip as shown in


Figure 3-9, then press the terminal strip back down into its
socket on the case board, as shown in Figure 3-10, taking
care not to strain any wire connections. Gently tug each
wire when finished, to verify secure connection to the
screw terminals.

Note
The SHIELD wire is the bare drain emerging from the foil
s h i e l d a r o u n d t h e YE L L OW a n d B ROW N w i r e s. T h e
BRAID-DRAIN wire is the bare drain emerging from the sur-
rounding braided shield inside the cable jacket. It is not neces-
sary to prevent the two braids from coming into contact with
each other.

Shield D1 YELLOW
D0 BROWN
Shield SHIELD
Braid-Drain VP RED
Common BLACK
Chassis BRAID DRAIN

All optional cable entries must use


appropriate ID conduit connections
or cord-grip fittings to retain the
I P 6 6 r a t i n g. I f y o u a r e u s i n g
non-TIENet or non-Signature
cables, you must supply the appro-
priate ID conduit connections or
cord-grip fittings.

Figure 3-9 TIENet Device terminal connections

3-8
Signature® Flow Meter
Section 3 Standard Installation

Figure 3-10 Attach wired terminal strip to connector case socket

a. Systems using the LaserFlow or Area Velocity


350 Sensor:
Insert the reference tubing into the REF AIR port on
the case board, pushing it down inside the silicon tub-
ing. Be careful not to kink the reference tubing.

Figure 3-11 Insert the cable reference tubing into the case board reference port

3-9
Signature® Flow Meter
Section 3 Standard Installation

5. Tighten the cord grip sealing nut.

Figure 3-12 Position and secure the cable

6. Close the front panel and fasten it shut with the two Phil-
lips screws.

CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit must be sealed to prevent harmful gases and moisture
from entering the Signature enclosure. Failure to seal conduit
could reduce equipment life.

3-10
Signature® Flow Meter
Section 3 Standard Installation

3.4 Power The Signature is in compliance with North American and Inter-
national safety standards while the input voltage remains within
100-240 VAC (50/60 Hz).
For external current protection, a 2 A slow-blow or time-lag fuse
between mains power and the Signature is recommended to
accommodate up to 40 A inrush current at power up in applica-
tions up to 230 VAC.
The flow meter comes with the internal power supply wired to
the connector case, and held in place by a screw (see Figure 3-13
on the following page). Mains power is wired into the Signature’s
internal power supply, normally via a standard three-wire line
cord or hard-wiring through user-supplied conduit.

3.4.1 Hard Wiring Cable entries for hard wiring must use appropriate ID conduit
connections.

CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and
moisture from entering the Signature enclosure. Failure to seal
conduit could reduce equipment life.

If the instrument has been hard-wired for power, ensure that a


switch or mains circuit breaker is installed near the instrument
for easy access to remove power in the event of an emergency.

3.4.2 Line Cord If the instrument has been fitted with a line cord, ensure that its
installation is near a mains outlet for easy access to remove
power in the event of an emergency.

3.4.3 Power Connections Teledyne Isco offers an AC line cord kit that is installed at the
factory when ordered with the Signature. It is also sold sepa-
rately, and is easily installed by the user.
For installation instructions, refer to Section 5.1 AC Power Cord
Kit /AC Wiring (Applies to Permanent Installation Only). Note
that these instructions can also be used as guidance for
user-supplied line cord, hard wiring, or replacement
power supply.

3-11
Signature® Flow Meter
Section 3 Standard Installation

Power Supply

Mounting
Screw

Mains Power
Cord

Connector Case

Figure 3-13 Location of power supply

3.5 Mounting the The Signature can lie flat on a horizontal surface, or be attached
Signature to a wall using the stainless steel bracket on the back of the case.
It can also be installed inside a console enclosure with other
system components. If a console enclosure is used, ensure that it
provided proper sealing to protect the flow meter and other
equipment from harsh environments and/or moisture.
Because it uses a bubble line, the Signature does not have to be
mounted directly above the primary device, or even particularly
close to the flow stream. You will need to mount the unit within
25 feet (7.6 m), or 50 feet (15.3 m) if you are using the 100 foot
bubble line. Distances greater than 100 feet are not recom-
mended.
The mounting location should allow for easy removal and rein-
stallation in the event that cleaning, testing, or replacement is
required. Refer to the dimensional drawings on the following
pages for physical installation specifications.

Figure 3-14 Signature Flow Meter mounted on wall

3-12
Signature® Flow Meter
Section 3 Standard Installation

3.6 Outdoor Where the Signature Flow Meter is mounted outdoors, a weather
Recommendations shield for protection from direct sunlight and rain is recom-
mended. The shield should accommodate the flow meter’s 18 inch
dimensions with the cover opened, as shown below.

FIBERGLASS PRODUCTS

Figure 3-15 Weather shield - recommended for outdoor installations


(Drawing & Photo courtesy of TRACOM, Inc.)

3-13
Signature® Flow Meter
Section 3 Standard Installation

3.7 The Bubble Line Anchor the bubble line in the flow stream at the appropriate
measuring point in the weir, flume, or other open channel flow
situation. Air slowly bubbles out of the line into the flow stream.
The pressure in the bubble line is proportional to the liquid level
in the flow stream, and the flow meter measures this pressure,
sensing the liquid level.

3.7.1 Standard Bubble Lines Three different bubble lines are available for use with the Sig-
nature:
• (1/8" (0.32 cm) OD PTFE, 1/16" (0.17 cm) ID,
25 ft. (7.6 m) long)
• (1/4" (0.63cm) OD vinyl, 1/8" (0.32 cm) ID,
50 ft. (15.2 m) long)
• (1/4" (0.63 cm) OD vinyl, 1/8" (0.32 cm) ID,
100 ft. (30.5 m) long)

3.7.2 Comparing Vinyl and Wherever practical, Teledyne Isco recommends the vinyl line,
PTFE Bubble Lines which offers significant advantages over the PTFE line. The
vinyl line has a longer usable length than the PTFE line. This is
due to the small inside diameter of the PTFE tubing, which gen-
erates an undesirable friction head at lengths greater than 25
feet. Additionally, experience has shown that the larger ID vinyl
line is less likely to clog than the PTFE line when used in flow
streams with suspended solids. If the distance between the flow
meter and the measuring point exceeds 25 feet, you must use the
vinyl bubble line.
However, for certain installations, the smaller ID PTFE line also
has advantages. Due to the small inside diameter, the air volume
necessary (and, as a result, battery power) is minimized, a def-
inite advantage for battery-powered installations. Additionally,
almost no chemical can attack the PTFE line. Consequently, if
power consumption is critical, or there are known agents in the
flow stream that might attack the vinyl line, the PTFE line may
be more suitable.

3.7.3 Bubble Line Length The bubble line should be kept as short as possible. This will
minimize friction head effects in the line and will also minimize
the amount of line exposed to cuts, kinks, etc. Shorten the line as
necessary by cutting the tubing with a sharp knife.
Long Bubble Lines – For certain applications, you may need to
use a bubble line with an inside diameter larger than that of the
standard lines (for example, where the bubble line exceeds 50
feet, or where the flow stream is extremely dirty and the end of
the line might clog).
Consult the factory for specific recommendations regarding size
of line, special connectors required, etc. In no case should the
inside diameter of the bubble line exceed 1/4" (0.64 cm), and you
should recognize that a larger bubble line will result in increased
power consumption, a concern if you must power the flow meter
by battery.

3-14
Signature® Flow Meter
Section 3 Standard Installation

3.7.4 Attaching the PTFE The PTFE bubble line attaches to the flow meter with the bubble
Bubble Line line fitting on the bottom of the case and the silicone rubber
tubing connector. The tubing connector is a short length of 0.109
inch (0.20 cm) ID, 0.192 inch (0.49 cm) OD silicone tubing
attached to a barbed fitting on the case.
The 1/8" OD bubble line simply slips inside the silicone tubing,
forming a reliable union you can join and separate without tools.
Attachment To attach the PTFE bubble line, first slip the 2 inch (5.1 cm)
length of 1/4" (0.63 cm) ID vinyl tube included in the instrument
accessory package over the end of the bubble line. Grasp the sil-
icone tube to stiffen it and insert the end of the bubble line into
the silicone tube.
Slip the short length of vinyl tube over the union and force it over
the shoulder of the barbed fitting. The purpose of the short
length of vinyl tube is to support the union, preventing fatigue or
kinking of the silicone tube.
Removal To remove the bubble line from the tubing connector, first pull
the vinyl tube off of the barbed fitting and slip it down the bubble
line. Grasp the end of the silicone tube and then pull the bubble
line straight out of the silicone tube.
Several replacement lengths of the silicone tubing are included
in the instrument accessory package. Install them on the unit by
simply forcing the end of the tube over the barbed connector.

Note
Avoid placing tension on the silicone tube where it attaches to
the barbed fitting. Direct the PTFE tube straight away from the
fitting rather than at an angle. This reduces the likelihood of
tubing wear and leakage around the fitting.

3.7.5 Attaching the Vinyl The vinyl bubble line attaches directly to the barbed fitting.
Bubble Line Remove the silicone tube and slip the 1/8" (0.32 cm) tubing over
the fitting.

3.7.6 Installing the Bubble Install the bubble line at the recommended level measuring point
Line in the primary device or other open channel flow situation. If you
do not know where this is, consult the manufacturer of the
primary device.
If you are not using a fabricated device, consult the Isco Open
Channel Flow Measurement Handbook for suggestions. Many dif-
ferent devices are discussed there. Proper location of the bubble
line outlet is necessary for accurate measurement. Normally, the
bubble line is positioned in the flow stream with the end at a
right angle to the flow.
In many installations, it is not practical to locate the outlet of the
bubble line precisely at "zero" liquid level. Depending on the situ-
ation, the outlet end of the bubble line may be located anywhere
within ten feet (3 m) above or below the actual zero level of the
primary device. Then set the displayed level using Adjust
Options from the Configure menu. Refer to Section 3.8 Setting
the Level.

3-15
Signature® Flow Meter
Section 3 Standard Installation

Note
The Signature cannot accurately measure liquid levels that are
even with or below the bubble line outlet.
If you need to measure the liquid level down to the actual
“zero” level of the primary device, Teledyne Isco recommends
placing the bubble line outlet at least 1 to 2 inches (2.5 to 5.1
cm) below the primary device zero level to avoid measurement
failures when the liquid level is even with the outlet. Since the
flow meter can display negative measurements, you can com-
pensate when you set the level in the flow meter.

Vertical location of outlet end of bubble line ranges


If the outlet is above the zero from ½" (1.3 cm) above. To 6" below the zero level.
level, the flow meter cannot (The preferred location is at least 1" to 2" [2.5 to 5 cm]
measure this head range. below the zero level.)

Attach bubble line to the side of


Bubble the flow stream. Bubble line Zero Liquid Level
Line of Weir or Flume

Flow Distance as Prescribed for


Primary Device

End View Side View

Figure 3-16 Positioning the Bubble Line in the Flow Stream

3.7.7 High-Velocity Flow The normal position of the bubble line in the flow stream is at a
Streams right angle to the flow. However, studies have indicated that at
relatively high flow stream velocities (greater than five feet per
second [1.5 meters per second]), this location may lead to a mea-
sured level lower than actual. This negative level offset is due to
localized areas of low pressure near the bubble line outlet; the
size of the offset depends upon the velocity of the flow stream and
the configuration of the flow stream channel. You can, of course,
adjust the level at the flow meter to compensate for the error.
However, this is only effective if the flow rate and velocity are
essentially constant. If the flow rate drops and the velocity
decreases, the level adjustment you made earlier will be incorrect
for the new, lower velocity and flow rate.

3-16
Signature® Flow Meter
Section 3 Standard Installation

The best way to overcome this problem is to completely isolate


the bubble line from the flow stream velocity by placing it in a
stilling well, as described below. If this is not possible, you can
perhaps create a cavity in the bottom of the channel, and locate
the outlet of the bubble line in the depression, again isolating it
from the flow stream velocity.
In flow streams carrying large amounts of solids, however, this
may cause problems because of silt collecting in the depression
and restricting the bubble line. A third alternative is to put a 90
degree bend in the end of the bubble line, forming a horizontal
leg approximately two inches (5 cm) long, and orienting this hori-
zontal leg downstream, parallel with the flow. Tests have shown
that this orientation of the bubble line minimizes the effect of
flow stream velocity.

3.7.8 Stilling Wells If the installation includes a stilling well in the primary mea-
suring device, installing the bubble line in the stilling well is rec-
ommended. Attach the line securely to the stilling well, using
stainless steel and/or plastic mounting hardware.
Not all stilling wells are suitable for bubble line installation. If
the well is subject to silting or buildup of foreign material, the
bubble line may have to be mounted in the flow stream proper.

3.7.9 Flume Bubble Line The large variety of primary measuring devices and installations
Fittings makes comprehensive bubble line installation instructions
impractical. However, valid general observations on bubble line
installation can be made. Flumes can be specified to include a
bubbler fitting. In new construction, this is highly recommended.
It may even be possible to modify an existing installation to
include a permanent bubbler fitting.

3.7.10 Bubble Line Teledyne Isco offers both stainless steel and copper bubble line
Extensions extensions. The metal extension may be easier to install in the
flow stream than the plastic bubble line because of its rigidity.
Two different extensions are available to match the two standard
bubble lines.
The extension for the PTFE bubble line includes a silicone rubber
tubing connector, and the tubing installation is as described
above. The vinyl bubble line attaches by simply slipping the vinyl
tube over the end of the extension.

3-17
Signature® Flow Meter
Section 3 Standard Installation

Metal Silicone PTFE Bubble


Extension Tube Line Tubing

Vinyl Bubble
Metal Line Tubing
Extension

Figure 3-17 Installing the Stainless Steel Bubble Line Extension

3.7.11 Open Channel If you do not use a stilling well, attach the bubble line to the side
Installation of the flow channel or flume. Make the attachment so it causes a
minimum amount of disturbance to the flow stream. If possible,
cut a groove in the side of the channel, place the bubble line in
the groove, and then grout over the groove.
Alternatively, you can attach the bubble line to the side of the
channel, and then grout over the line to form smooth, sloped
lead-in and lead-out surfaces. However, if neither of these
methods is practical, you may simply attach the bubble line to
the side of the channel or the upstream side of a weir using
stainless steel and/or plastic mounting hardware.
In any case, always install the bubble line so it causes as little
disturbance to the flow stream as possible. This usually means
an installation on or adjacent to a channel wall where there is a
condition of stagnant flow. For temporary survey applications,
you can attach the bubble line with waterproof tape or other tem-
porary means.

3.8 Setting the Level Although all other programming steps can be performed off-site,
the liquid level must be set at the measurement site following
installation.
Once the 330 bubble line is installed in the flow stream, or the
310 sensor is installed over the flow stream, measure the present
liquid level and enter this value for Level, under Configure >
Adjust Options. Highlight “Adjust” and press Enter to confirm.
From this screen, you can also update the display to show the
current level of the stream.

3-18
Signature® Flow Meter
Section 3 Standard Installation

Adjust Options
1. Level
Display depends on con-
nected TIENet devices

Adjust Level Setup


1. 330 Level
2. 310 Level

330 Level
LEVEL ADJUSTMENT

Level: ft Adjust Reset level

Display current
Last reading: X.XXX ft Update real-time level

Time of last
adjustment: MM/DD/YYYY TT:TT:TT

Figure 3-18 Level adjustment

3.9 TIENet Sensor In applications where the TIENet sensor will operate within a
Installation in a hazardous area, the installation must be performed by trained
Hazardous Location and qualified personnel, according to the installation control
drawing provided in Figure 3-19, and in accordance with local
requirements.
Information about hazloc installation specific to the TIENet 310
Ultrasonic Level Sensor is provided in the sensor’s user manual.

3-19
Signature® Flow Meter
Section 3 Standard Installation

Figure 3-19 Hazardous Location Installation Control Drawing-ATEX

3-20
Signature® Flow Meter
Section 4 Portable Installation

This section contains physical preparation procedures and por-


table mounting methods for the Signature Flow Meter and asso-
ciated Teledyne Isco equipment. Section 3 will contain the
permanent installation information for the Signature Flow
Meter.

4.1 Introduction The Signature Portable is equipped with one TIENet® receptacle.
In the event a connection must be made directly to the board
refer to the permanent Signature installation connection to
external devices in Section 3.2.
Where additional TIENet devices are deployed with a Signature
Portable, there is an optional TIENet 'Y' connection cable
available to expand the number of connections as needed. See
Appendix B for ordering information.

4.2 Portable Stand The portable stand allows the user to move the Signature to
various locations and without a permanent installation. This Sig-
nature has multiple power options and can be in either in an
upward facing or side facing position (Figure 4-1). The Signature
Portable is shipped from the factory in a side facing position.

Upward facing
Side facing

Figure 4-1 Signature on portable stand side facing (left)


and upward facing (right)

4-1
Signature® Flow Meter
Section 4 Portable Installation

Tools #2 Phillips screw driver

4.2.1 Adjust Signature to To adjust the Signature to be in the upward facing position:
Upward Position 1. Move the handle into the down position by pulling out the
handle latch pins on each side of the stand (Figure 4-2)

Figure 4-2 Pull out the handle latch pins on each side of
the stand

2. Loosen all of the screws in the top of the stand (2 panhead


and 2 flathead per side) with a #2 Phillips screw driver.
(Figure 4-3). These screws are captivated, so loosen until
they stop turing.

Figure 4-3 One of the eight screws loosened.

4-2
Signature® Flow Meter
Section 4 Portable Installation

3. Once the four flathead screws are loosened, rotate the


meter toward the back and lift straight up to remove it
from the stand (Figure 4-4).

Figure 4-4 Signature tilted up in preparation of removal

4. Align the notches in the Signature mount with the pan-


head screws on the stand.
5. Engage the slots and push the meter toward the back until
it stops.
6. Tighten all 8 screws (4 on each side) (Figure 4-5).

Figure 4-5 Signature in upward facing position

4-3
Signature® Flow Meter
Section 4 Portable Installation

4.3 Power The Signature Portable is designed to be used with 12 VDC lead
acid batteries.

4.3.1 Power Connections The standard battery connection on a Signature Portable accepts
Teledyne Isco model 948 or 946 lead acid batteries.
Adapting cables can be purchased to connect to a customer sup-
plied deep cycle marine battery.

WARNING
Do not set the portable stand directly on the terminals of a
battery. Serious injury or permanent damage may occur.

4.3.2 Outdoor The Signature Portable can be mounted to the portable stand in
Recommendations a side facing or upward facing position (Figure 4-1).
When the Signature Portable is used in an unprotected envi-
ronment it is recommended the screen is in the side facing
position so it does not collect water.

4.3.3 Battery Installation Common battery options for the Signature Portable:
• 946 lead-acid battery
• 948 lead-acid battery
• Lead-acid battery with solar panel
948 lead-acid battery When using the 948 lead-acid battery the portable stand will sit
directly on top of the battery case, where the battery is enclosed.
There is a notch in the back of the portable stand that will fit
over the power cable from the Signature to the battery. (Figure
4-6).

Handle

Notch for power cable Battery Case


Figure 4-6 Notch in stand for the power cable

4-4
Signature® Flow Meter
Section 4 Portable Installation

When using the 946 lead-acid battery, this battery will fit
directly under the Signature inside of the portable stand (Figure
4-7).

Tab

Battery Battery hold down Tab Stand

Figure 4-7 946 lead-acid battery installed under the


Signature Portable

946 lead acid battery To install the 946 lead-acid battery:


1. Remove the battery hold down of the portable stand by
loosening the thumb screw.
2. Thread the power cable through to the left side of the por-
table stand and slide the battery in, between the metal
tabs, after it. The battery will fit snugly.
3. Replace the battery hold down and tighten the thumb
screw.
4. Plug the cable into the Signatures power connector.

Tabs the battry


fits between

Battery hold down Thumb screw

Figure 4-8 Battery hold down and thumb screw

4-5
Signature® Flow Meter
Section 4 Portable Installation

4.4 Battery Life This section will cover the average battery life as related to the
Expectancy most common applications. For more information about battery
life please consult the factory.

CAUTION
The following section on battery life is for information purposes
only and does not constitute a guarantee of service or war-
ranty of any kind.

4.4.1 Major Standard


Equipment Settings
Effecting Battery Life
Display backlight and When the front screen is powered off the Signature is no longer
power taking near continuous readings. The reading interval will be
based on the data storage rate and any actively used equations.
Data storage rate Many options effect the storage rate as related to battery life,
see. See 2.7.5 Data storage for a complete explanation.
Equations Equations used by alarms, triggers, or secondary measurements
can effect the battery life of the equipment. The minimum mea-
surement setup in the equation should be the data storage
interval used for battery calculations.
For example, the data storage rate is set to 15 minutes, an
equation is set with a measurement interval of 5 minutes. For
battery life calculations it would be more accurate to use the
battery life expectancy of 5 minutes.

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Equations Setup
1. Equation XYZ
2. Create New Equation

1. 2.

Modify Equation XYZ Modify the equation.


Edit
Name: XYZ
Delete

A or B ( ) Bksp
AND & OR | NOT !

Condition A (Sensor Error) Measurement Interval:


Condition B (Threshold) 5 minutes
Condition C
Select Condition
Condition D
Condition E Edit Condition
Condition F

Figure 4-9 Example of defining conditions and building


equations

4-6
Signature® Flow Meter
Section 4 Portable Installation

4.4.2 Major Optional


Equipment Settings
Effecting battery Life
Cell Modem The longer the call window is open the more it will drain battery.
LaserFlow The LaserFlow has additional battery life considerations. Please
see the Laser 360 manual for more information
Analog Output Card The analog output card is not a typical component on a battery
operated device. The current draw over time is great and hard to
predict.
Ethernet Card With a very high average current, the ethernet card is not recom-
mended for most applications using batteries.
Bubbler Bubbler purges draw a large amount of current. Lengthen the
Purge Interval for improved battery life.

4.4.3 Additional
Considerations for
Battery Life
Bubbler Tubing The longer the bubbler and sensor line the more current draw
the meter will have. In most cases the difference in current draw
will not be a factor, but exceptionally long lengths will be a factor.
Reading Retries If the sensor is in a location where reading errors occur, the unit
will attempt to retry until a good reading is taken.
Battery Capacity If the battery is damaged (decreased battery capacity) the Sig-
nature will not run as long as the charts and calculations in this
section suggest. Batteries loose capacity:
• as they age
• as temperatures decreases
• when stored
• discharged below default or recommended levels

Table 4-1 Battery Life Expectancy


(default measurement rate of 15 minutes)
Ultra Sonic LaserFlow A/V Bubbler
Average current draw in ma 65 68 68 67
Model 946 lead-acid battery, 12 vCD, 6.5 100 hours 96 hours 96 hours 97 hours
amp-hours
Model 948 lead-acid battery, rechargeable, 692 hours 662 hours 662 hours 672 hours
12 vCD, 45 amp-hours

4-7
Signature® Flow Meter
Section 4 Portable Installation

Table 4-2 Battery Life Expectancy


( measurement rate of 5 minutes, all other parameters use default)
Ultra Sonic LaserFlow A/V Bubbler
Average current draw in ma 65 70 69 67
Model 946 lead-acid battery, 12 vCD, 100 hours 93 hours 94 hours 97 hours
6.5 amp-hours
Model 948 lead-acid battery, rechargeable, 692 hours 643 hours 652 hours 672 hours
12 vCD, 45 amp-hours

Table 4-3 Default Method for Portable Signature


Setting Duration/Setting
Laser (if present) Single point measurement
Optical clarity (if laser used) enabled (but not active)
Bubbler purge (if bubbler is present) 4 hours
Measurement Rate 15 minutes (for all flow technologies)
Backlight Key press time-out (30 second timeout enabled)
Display shutoff Key press time-out (5 minute timeout enabled)
Cellular Modem call Window (if modem is present) Open daily for 1hr
Equations used by alarms, triggers. None set
Secondary Data storage rate
If present during a TIENet scan the Signature will configure flow technologies, modem and laser as shown in the
chart above. If none are present the options will not be configured and will not affect power settings.

Table 4-4 Additional Average Current Draw


Purge interval 15min (if bubbler is present) 7
Backlight always on, display always on 48
301 PH sensor (15min data storage interval) 27
306 Sampler interface assuming No significant increase
307 Analog input card (active) 45
307 Analog input card (passive) No significant increase
304 Contact Output 10
Totalizer 1000 count per hour No significant increase
Ethernet communication module 88

4-8
Signature® Flow Meter
Section 4 Portable Installation

4.4.4 Calculating Average If the conditions listed in the average battery life tables and
Current Draw default method do not match your application, you can use the
following procedure to calculate the current draw.

Case Fluke® 87
(or other current-
averaging meter)

FLUKE 87 TRUE RMS MULTIMETER

Board
00I5
V- V+

mA
A

Black wire
Door White wire
A mA A Com V

+ lead

Battery, 12 Volt

+ clip
+
A good quality, adjustable, regulated DC power supply
can be substituted for the 12-volt battery. The power
supply should have at least 3 Amperes output, prefer-
ably more, and capable of overcurrent surges.

Figure 4-10 Measuring flow meter current

Note
Do not attempt this procedure unless you have the proper
equipment available and know how to make electrical mea-
surements.

Many of the power using functions in the Signature vary over


time. To measure current for a varying load requires a
more-sophisticated type of multimeter, one that is capable of
averaging high and low readings over a period of time. The
Fluke® 87 Multimeter is one example of this type of meter.
You should set the meter on MIN/MAX and let it run with your
program for several hours or more. Other manufacturers’ meters
are also acceptable, but only if they are capable of averaging
current draw. The current test should be run long enough to
capture all periodic usage. The longer you run the test, the more
accurate the average will be.

4-9
Signature® Flow Meter
Section 4 Portable Installation

4.4.5 Battery Life To calculate battery life expectancy for an installation, you must
Calculations know two things:
• The capacity of the battery you are using
• The average current draw of the flow meter or (other
device) powered
Battery capacity is expressed in ampere-hours. The battery man-
ufacturer provides this information for each battery. This value is
the product of a load current times an arbitrary time period,
twenty hours for lead-acid types. Isco 946 batteries lead acid bat-
teries are rated for 6.5 ampere-hours.
To determine battery life for a Signature running on a 946 lead
acid battery, convert the battery capacity into milliam-
peres/hours and then divide the ma/hrs by the avg current draw.
This will give you a number in hours. Divide that figure by 24,
and you will have the number of days.

Note
The published ampere-hour figures do not mean that you can
expect to draw 6.5 amperes from the lead-acid battery for one
hour.

To convert ampere-hours to milliamperes, multiply by 1,000.


Examples 6.5 ampere-hours × 1,000 = 6,500 mAh
If you divide this figure by the average current of the flow meter,
say 65 mA, you will have:
6,500 ÷ 65 = 100 hours

Divide this number by twenty-four to get days:


100 hours ÷ 24 = 4.1 days

For considerations of safety, we suggest you subtract 10% from


this number (100% – 5% for 95% capacity and 5% for a reserve at
the end of discharge).
4.1 – .4 = 3.7 days

The 3.7 days is the battery expectancy for a lead-acid battery


with a 65 mA continuous average drain, with a 10% derating
factor. Remember if the battery fails there will be a period of time
during which no measurements will be taken and no data stored
(if you are also using Flowlink® software).

Note
Always operate these batteries with a reserve factor.

4-10
Signature® Flow Meter
Section 4 Portable Installation

4.4.6 Low Battery Cut Off Batteries are considered fully discharged well before the ter-
and Battery Care minal voltage drops to zero volts. Operating lead acid batteries
below the fully discharged point decreases their capacity and
damages the battery. Lead acid batteries, under a constant rate
of discharge, are considered fully discharged at 11.5 volts,
however; the Signature does not discharge the batteries at a con-
stant rate.
Time durations will allow you to get the maximum amount of life
and readings out of the battery without damaging it. It can also
prevent a short term power dip from causing a premature
shutdown. By default, the Signature will turn itself off when the
battery voltage is at 11.5v for 2 min. To prevent the unit from
cycling OFF and ON the unit will not recover (turn back ON) until
the battery voltage is 13.0 for 2 minutes.

WARNING
Adjusting the low battery cutoff improperly will damage the
batteries. Teledyne Isco assumes no liability for damage
done to Teledyne Isco or third party vendors batteries.
Adjust the low battery cut off at your own risk. User
assumes all liability for damage done to batteries by
altering the low battery cut off.

Using disposable batteries may allow for greater flexibility and


possible usage. The low battery cutoff point can be adjusted by
the user. The duration the battery must remain at this voltage
for shutdown to occur is fixed at 5 min. If the low battery cut off
is altered it must remain at the adjusted level for 5 min before
the unit will shut down. Altering the low battery cutoff does not
alter the recovery settings. No matter what the low battery cutoff
is set to the unit will not recover or turn back on until the battery
voltage is at 13.0 for 2min.

4-11
Signature® Flow Meter
Section 4 Portable Installation

4.5 Connect to External Connection to TIENet socket.


Devices

A External TIENet
Devices/
H
K F A Modbus RS485 In
D1 = Yellow (+)
D0 = Brown (–)
G Gnd = Black
B SDI-12 Input
H I
K Modbus
C RS485 Out
K D TTL Serial
I I Battery Backup/
A A E Battery for Portable

B D E F Ethernet Modem
C
G Ethernet Port
L J J
Cellular Modem
H (power & serial)
B
4-20mA
Input/Output Card
I and
Contact Output
Note: No 12.8VDC power supply for Signature Portable Card
J Fuse "T" 3.15A

K Fuse "T" 4.00A

L Rain Gauge

Figure 4-11 Connector case, connectors, and fuses for


Signature Portable Flow Meter

4.5.1 Connecting Devices to The optional external TIENet devices compatible with the Sig-
the TIENet Receptacle nature Portable all scan in the hardware in the same manner.
Multiple TIENet devices can be connected simultaneously to the
same Signature Flow Meter. The following TIENet smart
sensors/cables will attach to the TIENet receptacle:
• Ultrasonic Level Sensor
• Area Velocity Sensor
• Bubbler
• pH/Temperature
• LaserFlow
• Sampler

4-12
Signature® Flow Meter
Section 4 Portable Installation

Connecting a TIENet To connect the TIENet plug from the sensor to the TIENet Recep-
receptacle to the Signature tacle:
Portable 1. Align the connectors and push together (Figure 4-12).
2. After the physical connection is made, a scan must be per-
formed (see section 2.6.1) for the device to be recognized.
For additional TIENet connections, use the TIENet Y-cable or
alternately an expansion box.
O-Ring and Lubrication 1. Coat the O-ring’s sealing surface with a silicone lubricant.
for the TIENet receptacle
CAUTION
Do not use petroleum-based lubricants. Petroleum-based
lubricants will cause the O-ring to swell and eventually deterio-
rate. Aerosol silicone lubricant sprays often use petro-
leum-based propellents. If you are using an aerosol spray,
allow a few minutes for the propellent to evaporate before pro-
ceeding.

2. Align and insert the connector. The sensor release will


“click” when the sensor connector is fully seated.
3. Connect the two caps together.

Figure 4-12 How to connect a TIENet receptacle to the


Signature Portable

Three-Hole Cord Grip The following device wires/cables are normally connected thru
the 3-hole cord grip:
• Rain Gauge
• SDI-12
• Modbus output
The wire/cable diameter can be .200 in to .230 in.
Installing a Cable Through 1. Remove plug to open hole for an additional cable.
a Three-Hole Cord Grip
Note
Gland nut must be loose to remove this plug.

4-13
Signature® Flow Meter
Section 4 Portable Installation

Figure 4-13 Removing the plug

Note
When replacing the plug, the rounded side needs to be pushed
into the hole, with the flange facing out (see figure below).

2. Thread the cable through the open hole and various con-
nectors.

3. Connect the three-hole cord grip to the Signature.

4. Attach the three-hole cord grip to the Signature port and


attach the wires to the appropriate screw terminals.

4-14
Signature® Flow Meter
Section 4 Portable Installation

CAUTION
Drilling holes in the Signature cases is not recommended and
may result in water damage. The material of the Signature
case is constructed of Noryl and does not lend itself to post
molding alterations.

Connecting the Rain The following shows how to connect the rain gauge to the Sig-
Gauge nature Portable and how the cables should look when properly
connected (Figure 4-14).
1. Remove the rain gauge cap and connect it to the TIENet
receptacle.

Figure 4-14 Connecting a rain gauge to the Signature


receptacle (L) and then to the Signature
connector board (R)

4-15
Signature® Flow Meter
Section 4 Portable Installation

4-16
Signature® Flow Meter
Section 5 Equipment Options

Optional equipment is designed to be user-installable. Internal


options, when ordered at time of purchase, are installed in the
Signature meter at the factory. This section describes each option
and provides instructions for its installation and operation.
All optional cable entries must use appropriate ID conduit con-
nections or cord-grip fittings to retain the IP66 rating. If you are
using non-TIENet or non-Signature cables, you must supply the
appropriate ID conduit connections or cord-grip fittings.

Note
Installation and operation of exterior TIENet devices is covered
in detail in the user manual for that technology.

Many options require interior access for installation. For con-


nector case interior access and TIENet device wiring instruc-
tions, refer to Connecting External Devices, on page 3-1 and
Connecting TIENet Devices, on page 3-7.

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit and any
battery power is disconnected.

Pa rt numbers for ordering accessories ar e provided in


Appendix B Options and Accessories. Optional equipment from
Teledyne Isco includes:

AC Power Cord Kit /AC Wiring (Applies to Permanent Installation Only), on page 5-2
Battery Backup (Applies to Permanent Installation Only), on page 5-4
Mechanical Totalizer, on page 5-8 Sampler Interface, on page 5-15
External Desiccator, on page 5-12 TIENet Expansion Box, on page 5-16
Ultrasonic Level Sensor, on page 5-13 Reference Port Tubing Kit, on page 5-29
Bubbler Level Sensor, on page 5-14 and 6-19 Ethernet Modem, on page 5-31
Contact Output Card (TIENet 304), on page 5-17 Cellular Modems, on page 5-36
Analog Input Card (TIENet 307), on page 5-22 Laser Doppler Velocity Sensor, on page 5-14
Analog Output Card (TIENet 308), on page 5-26 Continuous Wave Doppler Velocity Sensor, on page 5-14
pH and Temperature Device, on page 5-16
Sampler Interface, on page 5-15
pH and Temperature Device, on page 5-16

5-1
Signature® Flow Meter
Section 5 Equipment Options

5.1 AC Power Cord Kit /AC The AC power cord kit includes a line cord with a strain relief
Wiring (Applies to cord-grip fitting. If ordered with the Signature Flow Meter, it will
Permanent be shipped from the factory already installed.
Installation Only) Instructions for user installation are provided in this section.
Note that these instructions can also be used as guidance
for user-supplied or replacement line cord, hard wiring,
or replacement power supply.
Open the flow meter housing, following all warnings and instruc-
tions provided in Connecting External Devices, on page 3-1.
1. If a cord-grip fitting is already installed, loosen the sealing
nut on the AC line cord. If conduit is installed, remove the
sealing material around the AC line cord where it enters
the housing. This is to free the cabling for movement or
removal.

Lock Nut
Sealing Nut
Cable Nut
& O-Ring

Figure 5-1 AC Line cord with a cord-grip fitting

2. Remove the mounting screw and lift the power supply out
of its molded niche, taking care not to strain the wires
going to the board.

Power Supply

Mounting
Screw

AC Line Cord

Connector Case

Figure 5-2 Power supply mounting screw

5-2
Signature® Flow Meter
Section 5 Equipment Options

Figure 5-3 Power supply removal/replacement

3. Remove the clear plastic shield protecting the power sup-


ply terminals (Figure 5-4). Note that the Signature ground
wire ends in a ring terminal so the line cord ground wire
can easily be connected to the same terminal.
4. If replacing the power supply, label all wires according to
their terminal connections (i.e., L, N/L, Gnd, -V, +V), and
disconnect them.

Plastic shield

Figure 5-4 Power supply terminal strip: AC Input and DC Output

5-3
Signature® Flow Meter
Section 5 Equipment Options

5. If installing a cord-grip fitting (refer to Figure 5-1):


a. Remove the lock nut from the cable nut.
b. Install the cable nut through the line cord cable entry
(closest to power supply location) in the bottom of the
connector case and secure it to the Signature case wall
with the lock nut.
c. Feed the line cord end through the sealing nut and then
through the cable nut, into the case.
d. Lightly tighten the sealing nut, just enough to hold the
line cord in place while connecting it to the power sup-
ply.
6. Connect the line cord wires to the power supply, as shown
in Figure 5-4, and then reinstall the plastic protective
shield.

Note
Double-check to ensure that the terminals labeled Gnd, -V,
and +V on the Signature circuit board are wired to their corre-
sponding terminals on the power supply.

7. Reinstall the plastic shield over the wiring connections.


8. When seating the power supply into its niche, guide the
attached wires around in front of the mounting standoff
and through the molded slot, so they are not strained or
damaged (see Figure 5-3).
9. If using a line cord, gently tug the line cord to remove any
slack within the enclosure, taking care not to stress the
connection, and tighten the cord grip sealing nut.

CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires in the conduit must be sealed to prevent harm-
ful gases and moisture from entering the Signature enclosure.
Failure to seal conduit could reduce equipment life.

10. Close the front panel and fasten it shut with the two Phil-
lips screws.

5.2 Battery Backup The battery backup option consists of a Teledyne Isco Model 946
(Applies to Permanent lead-acid battery pack and extension cable, with special
Installation Only) hardware to mount it on the top of the Signature Flow Meter, or
on a wall. The unterminated power cable normally enters the
connector case through the second port from the right.

Note
An optional external power loss alarm is available. See Section
5.3 Power Loss Alarm.

5-4
Signature® Flow Meter
Section 5 Equipment Options

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.

CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.

CAUTION
Do not substitute another battery type for this option. Use only
the Model 946 Lead-Acid battery.

T15 Torx
screws (3)

Figure 5-5 Battery backup kit contents

CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit and wires must be sealed to prevent harmful gases and
moisture from entering the Signature enclosure. Failure to seal
conduit could reduce equipment life.

Installation 1. Remove line power from the Signature Flow Meter and
open the case as previously described in Section 3.2.

CAUTION
Do not connect the extension cable to the battery cable until all
other steps are completed.

5-5
Signature® Flow Meter
Section 5 Equipment Options

2. At the LEAD-ACID BATTERY terminal strip, connect the


extension cable’s black wire to the +12 terminal, and the
white wire to the ground terminal.

Figure 5-6 Attach extension cable to the connector case

3. Install the mounting plate, either on top of the flow meter


case using the Torx screws provided, or on the wall nearby.

Mounting screw
holes

Figure 5-7 Installing the battery backup mounting plate

4. Place the 946 battery on the mounting plate and secure it


in place using the two black rubber draw catches.
5. Connect the battery cable to the extension cable.

5-6
Signature® Flow Meter
Section 5 Equipment Options

Figure 5-8 Backup battery, installed

Note
Be sure to unplug the battery when intentionally disconnecting
from AC power.

5.3 Power Loss Alarm The Signature Flow Meter offers different options to notify you of
line power loss.
Contact Output Card The Contact Output Card is an optional card that closes or opens
(TIENet 304) relays at specific thresholds set within the Signature Flow Meter
programming. For power loss indication, it is recommended to
configure the Signature to activate the contact output to CLOSE
when the 300 Sense Voltage drops below 12 VDC. The output
relay will deactivate when the 300 Sense Voltage drops below 12
VDC or if the power drops all together.
Analog Output Card The Analog Output Card provides an analog output signal pro-
(TIENet 308) portional to a configured measurement input. For power loss
indication, it is recommended to configure the Signature so the
analog output is proportional to the 300 Sense Voltage. The
external device monitoring the analog output should alarm when
the analog signal drops below the equivalent of 12 VDC reported
by 300 Sense Voltage.
12 Volt DC Output The Signature system operates on 12 VDC. A voltage level of 12
VDC at the POWER 12.8V terminal (Figure 5-6) will indicate
adequate power supplied to the Signature meter, whereas a
voltage level of less than 12 VDC will indicate a power loss.
Signature Power Loss This device energizes a relay based on the Signature operating
Alarm Box (60-5324-024) voltage of 12 VDC. When the 12 VDC supply drops below 9 VDC
the relay becomes de-energized and switches the relay contact
indicating a power loss. It has NORMALLY OPEN and
NORMALLY CLOSED outputs.

5-7
Signature® Flow Meter
Section 5 Equipment Options

5.4 Mechanical Totalizer The mechanical totalizer is a seven-digit, non-resettable


mechanical counter installed in the front panel. It increments
according to programmed totalizer resolution and units of
measure. The totalizer can be viewed once the metal shield or
bubbler module is removed.
The volume represented by the mechanical totalizer is always
the primary Total Flow programmed in Measurement Setup >
Volume Input Setup (refer to Configure Options, on page 2-19).
The mechanical totalizer increments with the third significant
digit of the selected resolution (see Figure 2-15 Menu Tree:
Volume Input Setup (total flow)), e.g.:

Resolution 999999999 = Increment every 100 units;


Resolution 99999999.9 = Increment every 10 units;
Resolution 9999999.99 = Increment every 1 unit, etc.

The Signature permits a maximum 300 counts per minute; if


totalized flow exceeds this rate, remaining volume will be
buffered until it can be counted, although buffering over
extended time periods is not recommended.
Installation 1. Remove line power from the Signature Flow Meter and
open the case as previously described in Section 3.2.
2. Remove the four mounting screws holding the metal shield
in place and set it aside.
3. To ensure that the bubble line tubing is reconnected cor-
rectly, label the tubing ends as shown below in Figure 5-9,
then remove the four mounting screws holding the 330
bubbler in place and set it aside.

5-8
Signature® Flow Meter
Section 5 Equipment Options

Intake

Connecto
Air ce

Humidity
ren
fe
Re

r
Outlet

330 Mounting
Screws (4)

Figure 5-9 Disconnect and remove bubbler module

4. Using a razor blade or utility knife, carefully cut the six


tabs in the control panel label to detach the totalizer win-
dow cover.

Note
Be sure to cut all six tabs. Pulling on the cover with some of the
tabs still attached will damage the control panel label.

5-9
Signature® Flow Meter
Section 5 Equipment Options

Figure 5-10 Remove totalizer window cover

Referring to Figure 5-11:


5. Remove the two screws above the totalizer cutout provided
in the Main CBA. These screws will be used for mounting
the totalizer.
6. Install the totalizer in the cutout, attaching the two
mounting tabs with the two screws, so that the numbers
appear in the window.
View the totalizer through the window and adjust the posi-
tion, if necessary, before tightening the screws.
7. Plug the totalizer connector into P10 on the Main CBA.
This two pin connector is keyed to prevent incorrect attach-
ment.

5-10
Signature® Flow Meter
Section 5 Equipment Options

Before installation

After installation

Figure 5-11 Optional non-resettable totalizer installation

5-11
Signature® Flow Meter
Section 5 Equipment Options

5.5 External Desiccator For Signature systems using the 330 or 360 Bubbler Module
and/or the TIENet 350 Area Velocity Sensor, the desiccator vents
the reference port for a pressure transducer, and the air intake
port for the bubbler system air pump, keeping the interior of the
flow meter case dry, as well as the sensor reference line.

Note
The desiccant is standard on the portable version of the 330,
350, and 360 Bubbler module. It is not standard on the
non-portable 310 Bubbler module.

To air intake

To reference port

Figure 5-12 External desiccator, installed

Remove the two red protective end caps from the ports before
installing a new cartridge.

5-12
Signature® Flow Meter
Section 5 Equipment Options

Figure 5-13 Remove red caps before installing external


desiccant cartridge

The desiccant cartridge is held in place by a spring tab on the


side of the flow meter. Press against the front of the cartridge to
disengage it from the unit.

Figure 5-14 Removing the external desiccant cartridge

The desiccant cartridge requires periodic maintenance. Refer to


Section 6.5.2 External Desiccator for instructions.

5.6 TIENet® Devices Teledyne Isco’s proprietary TIENet connectivity allows for the
combination of multiple flow measurement technologies and
other devices with the Signature flow meter.

5.6.1 Ultrasonic Level The TIENet 310 Ultrasonic Level Sensor mounts directly over
Sensor the flow stream. The sensor measures level by transmitting an
ultrasonic pulse toward the liquid surface and then measuring
the time it takes for the echo to return. The 310 is normally used
with some type of primary device (typically a weir or flume) to
measure flow in an open channel.
The 310 Sensor is ATEX-approved for use in potentially
explosive atmospheres when specific conditions are met. Refer to
the 310 user manual, as well as the control installation drawing
in Section 3.9.
For complete installation and operation procedures, refer to the
TIENet 310 sensor’s user manual.

5-13
Signature® Flow Meter
Section 5 Equipment Options

5.6.2 Bubbler Level Sensor The factory-installed TIENet 330 Bubbler is normally used with
some type of primary device (typically a weir or flume) to
measure flow in an open channel.
The amount of pressure required to force bubbles from the end of
a submerged bubble line is directly dependent on the hydrostatic
pressure of the flow stream over the end of the bubble line. A
pressure transducer inside the module senses this pressure and
converts it into a level signal that the flow meter uses to cal-
culate flow rate and total flow.
In order to operate with the 330 Bubbler, the Signature must
have an external desiccator installed. For installation of the
external desiccator, refer to Section 5.5.
Because the 330 Bubbler is a standard component in bubbler Sig-
nature meters, installation instructions are located in Section 6
Maintenance and Servicing, under 330 Bubbler Installation.

5.6.3 Laser Doppler Velocity The TIENet 360 LaserFlow™ velocity sensor remotely measures
Sensor flow in open channels with non-contact Laser Doppler Velocity
technology and non-contact Ultrasonic Level technology. The
sensor uses advanced technology to measure velocity with a laser
beam at single or multiple points below the surface of the waste-
water stream.
To operate with the LaserFlow, the Signature requires firmware
version 1.18 or later. Firmware update instructions are provided
in Section 6.3 Firmware Updates.
For complete installation and operation procedures, refer to the
LaserFlow sensor’s user manual.

5.6.4 Continuous Wave The TIENet 350 Area Velocity Sensor measures flow stream
Doppler Velocity average area velocity and liquid level. The Signature uses this
Sensor information to calculate the flow rate and total flow of the
stream. To operate with the 350 sensor, the Signature requires
firmware version 1.18 or later. Firmware update instructions are
provided in Section 6.3 Firmware Updates.
The sensor is mounted in the flow stream, normally at the
bottom of the channel. It measures average velocity using con-
tinuous ultrasonic sound waves to produce a Doppler effect. The
sensor measures liquid level using an internal differential
pressure transducer.
In order to operate with the 350 AV sensor, the Signature must
have an external desiccator installed (see Section 5.5 External
Desiccator). Signature bubbler systems will already have a desic-
cator installed. If you are adding a 350 AV sensor to a
non-bubbler system, you will also need to add an external desic-
cator.
For complete sensor installation and operation procedures, refer
to the TIENet 350 sensor’s user manual.

5-14
Signature® Flow Meter
Section 5 Equipment Options

Optimal drying power Some Signature flow meters have a single piece of tubing
installed between the reference port and the humidity connector,
and a cap plug on the intake port. While a Signature with this
tubing configuration will operate satisfactorily with the 350 AV
sensor in most situations, you can configure the tubing to utilize
both chambers of the external desiccator to increase drying
power and extend the desiccant service interval.
Items required:
• Plastic ‘Y’ Fitting (Part #209-0167-49)
• 0.25 x 0.125 silicone tubing (2 pcs, Part #029-1353-02)

Procedure:
1. Remove the cap plug.
2. Disconnect the tubing from the humidity connector and
reroute it behind the ribbon cable.
3. Connect both the reference port tubing and the intake port
tubing to the ‘Y’ connector.
4. Connect the ‘Y’ connector to the humidity connector
(Figure 5-15).

Intake port
tubing
Reference port
tubing

Humidity connector
tubing

Figure 5-15 Tubing configuration for optimal drying power


(Exterior desiccator required)

5.6.5 Sampler Interface The TIENet 306 Sampler Interface connects the Signature Flow
Meter to a Teledyne Isco wastewater sampler. Through this con-
nection, the Signature can enable the sampler based on
user-specified conditions, pace the sampling routine based on
flow volume, and receive sample and bottle information from the
sampler.
For complete installation and operation procedures, refer to the
TIENet 306 device’s user manual.

5-15
Signature® Flow Meter
Section 5 Equipment Options

5.6.6 pH and Temperature The TIENet 301 pH sensor measures the acidity or alkalinity of
Device an aqueous solution by determining the relative quantity of dis-
sociated hydrogen ions in the solution. The normal scale for pH
runs from 0 to 14, with 0 being most acidic and 14 being the most
alkaline.

For complete installation and operation procedures, refer to the


TIENet 301 device’s user manual.

5.6.7 TIENet Expansion Box The water-tight expansion box connects to a TIENet terminal
strip like other TIENet devices, and contains three additional
strips inside, for connecting more devices. The expansion box can
be daisy-chained with each box providing two additional cord
grip ports. Additionally, the expansion box contains a TIENet
connection for an option card (such as the 308 Analog Output
option, Section 5.6.9). If an option card is used in an expansion
box there is room for just one TIENet cable cord grip exiting the
box.

Note
For applications requiring reference air moisture protection,
please contact the factory.

Note
The expansion box is not rated for use in hazardous locations.

5-16
Signature® Flow Meter
Section 5 Equipment Options

Enclosure Rating:
IP67 (NEMA4X, 6)

All optional cable entries must use


appropriate ID conduit connections or
cord-grip fittings to retain the IP67 rating.
If you are using non-TIENet or non-Signature cables
you must supply the appropriate ID conduit connections
or cod-grip fittings.

Desiccant bag
TIENet device
terminals

(Cord-grip fittings are ordered separately)

Mounting plate Option card


location

Figure 5-16 TIENet Expansion Box

5.6.8 Contact Output Card The contact output card provides a contact closure that can be
(TIENet 304) enabled based off of conditions set in the equation builder.
The Signature accepts up to three internal, user-installed
TIENet option cards. The 304 contact output card provides two
contact outputs per card for connection between the Signature
meter and non-Isco process control equipment that requires a
contact output.

WARNING
The 304 contact output card is not recommended for
voltages above 60vDC or 48vAC.

5-17
Signature® Flow Meter
Section 5 Equipment Options

CAUTION
Use proper static dissipation when handling circuit boards.

Programming menus and data display distinguish each output


by serial number and channel number.
Tools required T-15 Torx driver
To install a card:
1. Remove power from the Signature flow meter and open the
case, as previously described in Connecting External
Devices, on page 3-1.
2. The option card includes a mounting screw. Remove the
tubing retainer from the screw.
3. Remove the 3-pin header clip from its socket on the board.
4. Connect the receiving wires to the terminals according to
their labeling (positive and ground).
Note that Contact 1 and Contact 2 are identified on the
back of the board.

Contact 1
Contact 2

CH2
Back of card with Normally Open (NO),
Common (C), and Normally Closed (NC)

Figure 5-17 304 contact output card front and back view

Note
• Normally Open (NO): When the condition is met, the circuit
will be closed.
• Common (C): There is always be a connection to the com-
mon.
• Normally Closed (NC): When the condition is met, the circuit
will be closed.

5. Gently press the card down so that the 4-pin connector P4


plugs into one of the three analog output jacks on the board
(Item ‘J’ in Figure 3-2).

5-18
Signature® Flow Meter
Section 5 Equipment Options

6. Secure the card in place by tightening its mounting screw


with the T-15 Torx driver. Do not overtighten.

Two cards are shown here.


The Signature accepts up
to three cards at once, for a
possible six simultaneous
input/output channels.

Figure 5-18 View of two cards installed

7. Feed the clip with receiving wires through the appropriate


port on the bottom of the meter, and press the clip down
into its socket on top of the card.

Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.

Configuration The 304 measurement configuration screen includes the 304


digital output option.
Home Display

5-19
Signature® Flow Meter
Section 5 Equipment Options

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup
With initial installation, begin by per-
forming a hardware scan to add the 304.

Smart Sensor Setup (TIENet)


• Configure Measurements
• Perform Scan

Configure Measurements
1 - XXX Press Enter for a list of sensors.
Scroll to the 304 and press
XXX Parameter XXX Parameter Enter to select.

XXX Parameter

Configure Measurements Scroll with arrow keys to high-


<serial #> Relay Output light / select / deselect any dis-
played parameter or edit its
304 Digital Output 304 Digital Output- name.

Smart Sensor Configuration Press NEXT to confirm


Smart Sensor Setup (TIENet) configuration. There may be a
The sensors are being
The sensors have been configured. slight delay.
configured...

Figure 5-19 304 contact output device configuration

5-20
Signature® Flow Meter
Section 5 Equipment Options

Configure Options
1. Site Setup 5. Data Storage/Push Setup
2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Inputs/Outputs/Alarms Setup
1. Alarms
2. Digital Outputs

Digital Output Setup


1. 304 Digital Output
2. 304 Digital Output -A

304 Digital Output


DIGITAL OUTPUT
Measurement Name: 304 Digital Output
Equation: --none--

Figure 5-20 304 contact output setup

5-21
Signature® Flow Meter
Section 5 Equipment Options

5.6.9 Analog Input Card The 307 analog input card allows the Signature to record an
(TIENet 307) analog signal as a number of data types from several different
units.
The Signature accepts up to three internal, user-installed
TIENet 307 option cards. The 307 analog input card provides two
4-20mA inputs a piece for connection between the Signature
meter and non-Isco process control equipment or other
equipment that outputs a 4-20mA current signal.

CAUTION
Use proper static dissipation when handling circuit boards.

Programming menus and data display distinguish each output


by serial number and channel number.
Tools required T-15 Torx driver
To install a card:
1. Remove power from the Signature flow meter and open the
case, as previously described in Connecting External
Devices, on page 3-1. If the LED is green, a current is flow-
ing.
2. The option card includes a mounting screw. Remove the
tubing retainer from the screw.
3. Remove the 3-pin header clip from its socket on the board.
4. Connect the receiving wires to the terminals according to
their labeling (positive and ground).
Note that Channel 1 and Channel 2 are identified on the
back of the board.

Channel 1 Channel 2

Back of board
CH2

Figure 5-21 307 analog input channel identification and


terminal connections

5-22
Signature® Flow Meter
Section 5 Equipment Options

5. Gently press the card down so that the 4-pin connector P4


plugs into one of the three analog output jacks on the board
(Item ‘J’ in Figure 3-2).
6. Secure the card in place by tightening its mounting screw
with the T-15 Torx driver. Do not overtighten.
7. Feed the clip with receiving wires through the appropriate
port on the bottom of the meter, and press the clip down
into its socket on top of the card.

Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.

Channels The orientation of the wires is dependent on the setting of the


channel. If the channel is set to active, the positive wire is con-
nected to the out arrow. The connection will be reversed for the
passive mode.

Table 5-1 Direction of arrows and mode types


Out
In
(arrow pointing
Switch (arrow is pointing
away from the
toward the board)
board)
Switch is in the Positive Negative
active mode (A)
Switch is in the Negative Positive
passive mode (P)

NOTE: Active mode supplies


power to the loop. Passive
relies on another device
powering the loop.

Active Mode Passive Mode

Figure 5-22 Direction of switch in active and passive modes

5-23
Signature® Flow Meter
Section 5 Equipment Options

Configuration The 307 measurement configuration screen includes a reading at


0% (minimum or 4 ma) and a reading at 100% (maximum or
20ma) as well as a scaling option.

Hardware Setup
1. Smart Sensor Setup (TIENet) With initial installation, begin by per-
2. SDI-12 Setup forming a hardware scan to add the 307.
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

Smart Sensor Setup (TIENet)


• Configure Measurements
• Perform Scan

Configure Measurements Press Enter for a list of sensors.


1 - XXX Scroll to the 307 and press
Enter to select.
XXX Parameter XXX Parameter

XXX Parameter

Configure Measurements Scroll with arrow keys to high-


<serial #> Analog Inout light / select / deselect any
displayed parameter or edit its
CH1 Current - A CH1 Percent - A name.
CH2 Current - A CH2 Percent - A

Smart Sensor Configuration


Smart Sensor Setup (TIENet)
The sensors are being
The sensors have been configured.
configured...

Figure 5-23 307 analog input device configuration

5-24
Signature® Flow Meter
Section 5 Equipment Options

1. Site Setup 5. Data Storage/Push Setup


2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Inputs/Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Input Options
1. Alarms
2. Analog

Analog Input Setup


1. 307 Analog Input
2. 307 Analog Input-A
Press Enter for the list of
parameters.
307 Analog Input-A
ANALOG INPUT
Measurement Name: 307 Analog Input-A To edit units of measure,
Data Type: --none-- highlight and press Enter.
Percent and humidity UOM
Units: --none-- are % only.
4mA:
20mA:

Figure 5-24 307 analog input setup

5-25
Signature® Flow Meter
Section 5 Equipment Options

5.6.10 Analog Output Card The 307 analog output cards convert digital information from the
(TIENet 308) flowmeter to a variable analog output current ranging from 4 to 20
milliamperes. When a parameter measured by the flow meter is
converted into an analog output, 4 mA becomes the 0%, or
baseline, for the parameter, while 20 mA becomes the 100%, or
full-scale, of the parameter. For basic programming steps, refer to
Figures 5-18 and 5-19.
The Signature accepts up to three internal, user-installed
TIENet option cards. The 308 analog output card provides two
scalable 4-20mA outputs per card for connection between the Si
nature meter and non-Isco process control equipment or other
equipment that accepts a 4-20mA current signal.

CAUTION
Use proper static dissipation when handling circuit boards.

Programming menus and data display distinguish each output


by serial number and channel number.
Tools required T-15 Torx driver
To install a card:
1. Remove power from the Signature flow meter and open the
case, as previously described in Connecting External
Devices, on page 3-1.
2. The option card includes a mounting screw. Remove the
tubing retainer from the screw.
3. Remove the 3-pin header clip from its socket on the board.
4. Connect the receiving wires to the terminals according to
their labeling (positive and ground).
Note that Channel 1 and Channel 2 are identified on the
back of the board.

Channel 1
Channel 2

Back of board
CH2

Figure 5-25 308 analog output channel identification and


terminal connections

5-26
Signature® Flow Meter
Section 5 Equipment Options

5. Gently press the card down so that the 4-pin connector P4


plugs into one of the three analog output jacks on the board
(Item ‘J’ in Figure 3-2).
6. Secure the card in place by tightening its mounting screw
with the T-15 Torx driver. Do not overtighten.
7. Feed the clip with receiving wires through the appropriate
port on the bottom of the meter, and press the clip down
into its socket on top of the card.

Note
Cabling is user-supplied. Shielded cable is recommended. For
cord-grip fitting options, refer to Appendix B, Section B.3. If
conduit is used, the conduit and wiring must be sealed to pre-
vent entry of harmful gases and/or moisture.

Configuration The 308 measurement configuration screen includes a minimum


and maximum reading as well as a scaling option.

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

Smart Sensor Setup (TIENet) With initial installation, begin by performing a


• Configure Measurements
hardware scan to add the 308.
• Perform Scan

Configure Measurements
1 - XXX Press Enter for a list of sensors. Scroll
to the 308 and press Enter to select.
XXX Parameter XXX Parameter

XXX Parameter

Configure Measurements
<serial #> Analog Inout Scroll with arrow keys to highlight /
select / deselect any displayed
CH1 Current - A CH1 Percent - A parameter or edit its name.
CH2 Current - A CH2 Percent - A

Smart Sensor Configuration


Smart Sensor Setup (TIENet)
The sensors are being
The sensors have been configured.
configured...

Figure 5-26 308 analog output device configuration

5-27
Signature® Flow Meter
Section 5 Equipment Options

1. Site Setup 5. Data Storage/Push Setup


2. Measurement Setup 6. Sampler Setup
3. Adjust 7. Outputs/Alarms Setup
4. Equation/Trigger Setup 8. Reset Totalizers
9. Reports/History Setup

Output Options
1. Alarms
2. Analog

Analog Ouput Setup


1. CH1 Percent
2. CH2 Percent
3. CH1 Percent
4. CH2 Percent

CH1 Percent Press Enter for the list of


parameters.
ANALOG OUTPUT
Measurement Input: 1 - XXX

To edit units of
CH1 Percent measure, high-
ANALOG OUTPUT light and press
Enter. Percent
Measurement Input: 330 Level (example)
and humidity
Low Limit: 0 feet (ft)
High Limit: 0
Channel Name: <serial # of 308 card> Channel #

Figure 5-27 308 analog output setup

5-28
Signature® Flow Meter
Section 5 Equipment Options

5.7 Reference Port Tubing For Signature Bubbler monitoring sites where the flow meter
Kit mounting location does not provide adequate reference to atmo-
spheric pressure at the measuring point, the reference port
tubing kit includes tubing and adaptors to relocate the reference
port.
An extra kit can also be used for the bubbler intake, if the air at
the flow meter mounting location is excessively humid.
The 25-foot, vinyl 1/4 " ID, 3/8 " OD tubing should be cut to the
shortest length practical for your installation.

Figure 5-28 Reference port kit (full length not shown)

Installation The tubing attaches to the reference port of the external desic-
cator. If using two kits, the second tube attaches in the same
manner to the intake port of the external desiccator.
To install:
1. Unscrew the hydrophobic filter cap from the reference
chamber (smaller chamber) of the external desiccator.

Figure 5-29 Reference port kit: Remove filter and barb

5-29
Signature® Flow Meter
Section 5 Equipment Options

2. Screw the hydrophobic filter into the end of the reference


port tubing connector.
3. Screw the hose barb fitting with o-ring into the reference
port of the desiccator.

Figure 5-30 Reference port kit: Install filter and barb

4. Push the open end of the tubing down over the hose barb
fitting.

Figure 5-31 Reference port kit: Installed on flow meter

5-30
Signature® Flow Meter
Section 5 Equipment Options

5.8 Isco Flowlink Software Flowlink® is Teledyne Isco’s proprietary software system for
data acquisition, storage, retrieval, and analysis. Using the
interface of Microsoft Windows, Flowlink can be used to remotely
program the Signature Flow Meter, retrieve data from the flow
monitoring system, present site data graphically, and generate
statistical information from the site data.
Flowlink helps ensure data integrity by displaying the Sig-
nature’s tracked configuration changes, data measurement sum-
maries, diagnostic test results, and user events in the program.
With these tools, Flowlink provides assurance that the data has
not been altered.
USB drivers for computer direct connection to the Signature
Flow Meter are included on the Flowlink CD, and must be loaded
prior to direct connection between the computer and the Sig-
nature.
See Section Connecting to the Signature with Flowlink, on page
2-2 for instructions on how to connect to the Signature meter
with Flowlink software.
From Flowlink, the event data can be exported and saved in the
form of text reports on your computer, searchable by site name,
m o d u l e, a n d d a t e. Fo r c o m p l e t e i n f o r m a t i o n , r e f e r t o
Section Signature Data in Flowlink, on page 2-34.

5.9 Ethernet Modem Setup and data retrieval, as well as alarm output configuration,
can be accomplished remotely via TCP/IP communication pro-
tocol with a static address, using Flowlink software and the eth-
ernet modem to access the Signature’s web browser. The ethernet
modem is factory-installed on the connector case.
The ethernet modem can also be installed by the user. Remove
line and/or optional battery power from the Signature Flow
Meter and open the case as previously described in Section 3.2.

CAUTION
Always use proper static dissipation methods when handling
circuit boards.

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.

CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.

5-31
Signature® Flow Meter
Section 5 Equipment Options

Ethernet
modem

Mounting
hardware

Figure 5-32 Ethernet modem kit contents

5-32
Signature® Flow Meter
Section 5 Equipment Options

1. Press the modem assembly down


into its socket on the connector case
(item G in Figure 3-2 Connector
case, connectors, and fuses), with
the row of dots along the bottom left
and right edges aligned with the
row of circles on the board to ensure
proper orientation.

2. Place the plastic spacer over the


screw hole by the bottom right cor-
ner of the modem.
3. Place the lock washer on top of the
spacer.

4. Place the rectangular retainer over


the lockwasher, with the counter-
sink facing up, and attach with the
screw.

Figure 5-33 Ethernet modem installation

In order to communicate with the Signature Flow Meter using


the ethernet modem, your network must have TCP/IP services
installed. A static IP address must be reserved for the Signature,
and client network computers must be allowed to access the
static IP address.

5-33
Signature® Flow Meter
Section 5 Equipment Options

5.9.1 Ethernet Modem When installation is complete and power restored, wait one
Configuration minute for the Signature to recognize the modem before pro-
gramming.

Note
The Signature does not support Dynamic Host Configuration
Protocol (DHCP). The network communication information (IP,
gateway, and subnet mask) must come from your network
administrator and be entered manually into the flow meter.

When you select Modem Setup from the Hardware Setup menu,
the type of modem installed determines what screen is displayed.

To configure the Signature for ethernet


communication, you must have the Hardware Setup
following information on hand prior to 1. Smart Sensor Setup (TIENet)
Hardware Setup:
2. SDI-12 Setup
IP Address – An Internet Protocol (IP)
address is the unique numerical label 3. MODBUS Input Setup
assigned to each device (e.g., computer, 4. MODBUS Output Setup
printer, flow meter, etc.) on a computer 5. Modem Setup
network for interface identification and
location addressing. The Signature’s Eth-
ernet modem requires a Static IP address
for remote communication.
Modem Setup
TCP Port – The default port setting is
1700. This is the communication port
Modem Model: MULTITECH MT 100 SEM
associated with the static IP address on
your network. Static IP Address: 000.000.000.00
Gateway Address – The gateway is the TCP Port: 1700
point of communication that joins two
different networks with different base IPs. Gateway Address: 000.000.000.000
Subnet Mask – T h i s i s t h e u m b r e l l a Subnet Mask: 255.255.255.0
location that allows multiple nodes to
communicate within the network.
It designates a subnetwork within the
larger network. Traffic between
s u b ne t w o r ks i s e x ch a n g e d o r r o u t e d
through the Gateway.
Special network access may be required to
configure these settings. For further
assistance, contact your network
administrator.

Figure 5-34 Ethernet modem setup: Communication settings (default settings shown)

5-34
Signature® Flow Meter
Section 5 Equipment Options

5.9.2 Network Firewall In order for your network administrator to identify the Signature
Settings in the network firewall setup, it must have a node ID (also known
as the MAC address). This is the NODE ID printed on the eth-
ernet modem’s serial tag (refer to Figure 5-35).

Figure 5-35 Locating the NODE ID (MAC address) on the ethernet modem

5-35
Signature® Flow Meter
Section 5 Equipment Options

5.10 Cellular Modems Setup and data retrieval through the Signature’s web browser, as
well as alarm outputs, can be accomplished remotely with one of
the available cellular modems. The whip-style antenna has a
magnetic mounting base.

Figure 5-36 Magnetic-mount cellular antenna

5.10.1 CDMA Modem The Code Division Multiple Access (CDMA) modem can automat-
ically push data to a secure server running Isco Flowlink Pro
software, with 1xRTT data transmission.

Figure 5-37 CDMA Cellular modem

5-36
Signature® Flow Meter
Section 5 Equipment Options

5.10.2 GSM Modem The Global System Mobile (GSM) modem can automatically push
data to a secure server running Isco Flowlink Pro software, with
GPRS data transmission.
Your service parameters, or provider, can be changed by
replacing the removable Subscriber Information Module (SIM)
card in your modem.

Note
The modems shown do not necessary represent the modems
installed.

SIM card slot

Figure 5-38 GSM Cellular modem

5.10.3 Installing the Cellular The modem kit includes the modem, power cable, DB9 serial
Modem cable, and coaxial antenna plug cable.

Note
Before installing the modem, remove the top label (with the
FCC ID on it) taped to the modem and adhere it to the outside
of the Signature case on the bottom of the unit, on the left side,
in the largest of the three sections (Figure 5-39). This is
required by the FCC.

FCC ID Label

Label on
bottom of
Sigature case

Figure 5-39 FCC ID label on modem and location of label


on the bottom of Signature case

5-37
Signature® Flow Meter
Section 5 Equipment Options

1. Remove line power from the Signature Flow Meter and


open the case as previously described in Section 3.2.

Antenna Plug
Cable
Serial Cable

Power Cable

Modem

Figure 5-40 Cellular modem kit contents

2. Install the plug in the preferred port (far left most com-
monly used). Route the antenna plug cable under any other
cabling, and install the plug in any open port.
3. Connect the three cables to the modem.
4. Remove the screw retainers and fasten the modem’s
mounting bracket against the connector case, as shown
below, using the two mounting screws.
5. Plug the serial and power cables into their respective con-
nectors on the board.

5-38
Signature® Flow Meter
Section 5 Equipment Options

Figure 5-41 Cellular modem installation

5-39
Signature® Flow Meter
Section 5 Equipment Options

5.10.4 Cellular Modem When installation is complete and power restored, wait one
Configuration minute for the Signature to recognize the modem before pro-
gramming.
When you select Modem Setup from the Hardware Setup menu,
the type of modem installed determines what screen is displayed.
To set up the Signature Flow Meter for cellular communication,
select either internet (IP) or CSD for the connection.
For the GSM modem, if an Access Point Name (APN) is used,
select SINGLE for the operating mode and enter the name.
For portable applications the modem can be set up with the call
in the time window. To conserve power, the time window does not
effect the push/outgoing connections. The time window restricts
when the meter can be acceded via a modem connection.

Hardware Setup
1. Smart Sensor Setup (TIENet)
2. SDI-12 Setup
3. MODBUS Input Setup
4. MODBUS Output Setup
5. Modem Setup

GSM CDMA

Modem Setup Modem Setup


Modem Model: WAVECOM FXT002 Modem Model: WAVECOM Q26 ELITE
Connection Mode: IP Connection Mode: IP
APN Operation Mode: NONE Incoming Port: 1700
Incoming Port: 1700 Modem Model: --none--
Access Point Name: Wireless Signal: 0%
User Name:
Password:

Figure 5-42 Cellular modem setup: Communication


settings

5-40
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.1 Maintenance The following tables are recommended maintenance checks to


ensure proper operation. As site conditions may vary, increase
the frequency of inspections as needed.

Table 6-1 Recommended Maintenance (Accessible Locations)


Recommended
Action Location
Frequency
Check desiccant for appropriate colora Monthly or following On-site
humidity error
Check bubble line for plugs or obstructions Monthly On-site
Check bubble line for kinks Monthly On-site
Check pump run time during manual purgeb Monthly On-site
Check for level measurement errors Weekly Via Flowlink application

a. When dry, the desiccant appears orange in color.


b. If the run time is >15 seconds, replace the intake gore filter.

Table 6-2 Recommended Maintenance (Difficult-to-Access Locations)


Recommended
Action Location
Frequency
Check desiccant for appropriate colora Every 6 months or On-site
following humidity
error
Check bubble line for plugs or obstructions Every 6 months On-site
Check bubble line for kinks Every 6 months On-site
Check pump run time during manual purgeb Every 6 months On-site
Check for level measurement errors Weekly Via Flowlink application

a. When dry, the desiccant appears orange in color.l


b. If the run time is >15 seconds, replace the intake gore filter.

6-1
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.2 Cleaning The Signature flow meter may be cleaned with water and a mild
detergent. For hard to remove stains, isopropyl alcohol may be
used. If the instrument is in an isolated area and the case is
sealed closed, it may be cleaned using a water hose.

6.3 Firmware Updates Signature and TIENet device firmware updates are provided in
the form of .bin files, which will be available for download from
the Teledyne Isco website. Note that firmware updates do not
remove any program settings or delete data.
To install an update:
1. Create a folder in the top directory of a flash drive, and
name it BINFILE.
2. Download the .bin file to be installed. To find your correct
.bin file, go to www.isco.com and click on Software/Firm-
ware Updates in the lower left corner. Select Open Channel
Flow Measurement.

Figure 6-1 Locating firmware updates

3. Save the new .bin file(s) to the BINFILE folder you created
on the flash drive.

6-2
Signature® Flow Meter
Section 6 Maintenance and Servicing

4. Using the flash drive adaptor cable provided with the Sig-
nature, connect the flash drive to the micro-USB assembly
on the flow meter’s front panel.

Figure 6-2 USB Micro adaptor cable (flash drive not included)

5. The USB Options menu appears on the display. Select


option #3, Update Firmware.
6. You will be prompted to select either Signature or TIENet
firmware. Select the appropriate .bin file from the pull
down menu and press NEXT.
a. Update Signature Firmware
The update will load for approximately three minutes.
During this time, do not make any changes to the Sig-
nature. When the firmware load is complete, the Signa-
ture will prompt you to remove the USB drive. An
automatic reboot then occurs over a period of approxi-
mately five minutes, during which the green LED sig-
nals that an internal operation is in progress. Do not
unplug the flow meter or press any keys until the
Home screen appears. In the event that the upload
fails, contact Teledyne Isco.
b. Update TIENet (Smart Sensor) Firmware
Select the radio button next to each device to be
updated and press NEXT. The progress of the sensor
firmware update(s) will be displayed. A confirmation
screen will appear when the update is complete.
c. Bootcode
This option is only used when the update(s) failed and
the sensor is no longer responding. Select the “Boot-
code” option, then select the appropriate file from the
drop-down menu to be updated. All other updates are
inactive when this option is selected.

6-3
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.4 Accessing the Interior Some maintenance or servicing tasks require opening the Sig-
nature housing to access the interior. Always refer to this section
prior to doing so.

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.

CAUTION
Before opening the case, disconnect the optional battery
backup power, if used (refer to Figure 5-8 Backup battery,
installed).

Note
Before restoring mains power, ensure that the flow meter’s
USB connector does not have a cable attached.

Open the door to access the two large screws holding the front
panel on the connector case. Remove the two screws, then
reinsert them in the front panel and latch the lid so they will not
be misplaced.

Figure 6-3 Open door and front panel to access interior

6-4
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.5 Desiccant The inside of the flow meter housing must be kept dry at all
times to prevent moisture damage to the internal components.
All Signature flow meters have an internal desiccant bag to
absorb moisture. Signature flow meters using a 330 bubbler also
require an external desiccator.
If increased humidity is indicated by either the humidity reading
of the flow meter or the color of the external desiccant, the des-
iccant must be renewed or replaced before damage occurs.
If this occurs more frequently than expected, inspect the seals of
cord-grip fittings and conduit, if used.

Humidity alarm The humidity of the case interior, reference (ambient) air, and
bubble intake air (if a 330 bubbler is installed) are all param-
eters that can be selected as conditions to trigger an alarm, noti-
fying you when it is time to renew or replace your desiccant.
The suggested alarm setting is a threshold condition of 40%. For
detailed instructions about setting up conditions and alarms,
refer to Sections Equation/Trigger Setup, on page 2-24, and
Inputs/Outputs/Alarms Setup, on page 2-27.

6.5.1 Internal Desiccator Saturated internal desiccant bags must be replaced; unlike the
external desiccant, they are not renewable.
The desiccant bag is held in place by a metal bracket. Remove
the two screws holding the bracket.

Figure 6-4 Removing the internal desiccant bag

6-5
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.5.2 External Desiccator The desiccator vents the reference port for a pressure transducer,
and the air intake port for the bubbler system air pump, keeping
the interior of the flow meter case dry.

To air intake

To reference port

Figure 6-5 External desiccator, installed

When dry, the loose silica gel desiccant inside the chambers is
orange or yellow. When the desiccant becomes saturated with
moisture, it turns green or blue, indicating that the intake air
and reference line are no longer protected from humidity.

Note
Teledyne Isco recommends checking the desiccant at least every
6 months, and changing/renewing the desiccant before the entire
compartment has changed color.
Dry desiccant
(good)

Saturated
desiccant (bad)

Figure 6-6 Desiccant indicating saturation

The desiccant cartridge is held in place by a spring tab on the


side of the flow meter. Press against the front of the cartridge to
disengage it from the unit.

6-6
Signature® Flow Meter
Section 6 Maintenance and Servicing

Figure 6-7 Removing the external desiccant cartridge

Unscrew the two black caps and carefully pour the desiccant out.
If removal is difficult, screw the caps back in and unscrew again.
Gently knock the caps and the cartridge against a hard surface to
free any small particles in the threads, as these can hinder
proper sealing and cause wear.
Using a funnel, fill both chambers with dry desiccant, replace the
caps, ensuring that they are fully engaged. Press the cartridge
back into place on the side of the flow meter.

Use a funnel Note


to refill the
If this is a new desiccant cartridge, remove the two red protec-
desiccator.
tive end caps from the ports before installing a new cartridge.

Remove red caps


before installing cartridge.

Figure 6-8 Opening the desiccant cartridge chambers

Renewing loose desiccant To renew the desiccant, spread it in a single layer on a flat metal
tray. Place in a vented, circulating forced air, conventional oven
in a well ventilated room, and heat at 100 - 175°C (212 - 350°F)
for about three hours, or until the color has returned to orange or
yellow.
MSDS (Material Safety Data Sheets) for silica gel chemicals are
provided in Appendix C.

6-7
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.6 Troubleshooting The tables in the following section provide troubleshooting infor-
mation to help in determining the causes of problems that may
occur with the Signature flow meter or TIENet devices.
The troubleshooting tables cover the flow meter and each TIENet
device separately. Note that the 300 TIENet device (Table 6-4) is
the internal connector case.

Note
Any time a circuit board is replaced or a sensor disconnected,
you MUST perform a hardware scan and SDI-12 scan (if con-
nected) before resuming operation.
6.6.1 Signature Flow Meter
Table 6-3 Troubleshooting: Signature Flow Meter
Symptom Cause Action Parts
Display repeatedly goes Power setting is set to If key is not pressed within 5 minutes, press
blank turn display off any key to turn the display on.
Contrast is out of Adjust the display contrast by repeatedly pressing the up or down
Blank Display but audi- adjustment arrow while holding down the +/- key.
ble beep when a key is
pressed Faulty Display Replace with known good display. Display
130-0602-06

6-8
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-3 Troubleshooting: Signature Flow Meter (Continued)


Symptom Cause Action Parts

Replace 4A/250V/5X20mm Slo Blo fuse


(Figure 3-2 Connector case, connectors, and
fuses, Item L). If the fuse opens again, check 4A Fuse
Open Fuse F3
for devices that may be shorting the supply, 411-9901-84
such as an external connection, modem or
option card.

60-3004-106
Dead Lead-Acid Replace or recharge the main battery con-
Model 946
Battery nected to the internal “Lead-Acid” terminals.
Lead-Acid Battery

Blank display and no


Check for proper AC voltage. If proper AC DC Power Supply
beep when a key is
voltage is present, replace DC power supply. 60-4304-037
pressed

DC power supply not


supplying 12.8 VDC
output. Service check: Disconnect the internal power
supply wires (Red +/pos, Black –/neg) from
the power terminals (Figure 3-2 Connector
case, connectors, and fuses, Item F). Con-
nect an Isco adaptor cable to the power termi- Isco Adaptor Cable
nals (Black +/pos, White –/neg). Then 69-4304-034
connect an Isco power supply (Model 913,
914, 923, or 924) to the adaptor cable.
If the Signature then functions properly,
replace the internal power supply.

Broken or loose wire


from power supply
Repair connections (Red +/pos, Black –/neg).
module to the connec-
tor case.

6-9
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-3 Troubleshooting: Signature Flow Meter (Continued)


Symptom Cause Action Parts

Ribbon cable loose or Check or cycle connections, then replace with Ribbon Cable
damaged known good cable. 69-4304-032

Blank display and no


beep when a key is
pressed (Con’t.)
Keypad
Defective keypad Substitute a known working keypad.
69-4303-009

Faulty or missing SD SD Card


Reinstall or replace SD card on Main CBA.
card 250-3000-66
Main CBA
Main CBA faulty Substitute with known good Main CBA.
60-4304-042

Programming error -
Check measurement configuration of level, flow rate, and volume
Zero flow rate or aster-
input for the Total Flow parameter.
isk (*)

Nonresettable totalizer
Broken wire connection Check wire connections for the totalizer on the Main CBA.
does not advance

Mechanical
Defective totalizer Replace totalizer Totalizer
60-4304-015

Flash drive encrypted


Try a different USB Flash drive
or defective
USB Adaptor
USB device not recog-
Adaptor cable defective Replace cable Cable
nized - No USB Options
480-2946-02
screen
Micro-USB
Micro-USB Assembly
Replace port Assembly
damaged
60-4304-053
The necessary files are Load the firmware from our website onto the
www.isco.com
not on the flash drive. flash drive, into a folder named BINFILE.
Cannot update soft- SD Card not functional Verify that the SD card includes a BINFILE Micro-USB Card
ware / Read flash drive or missing files. folder with the appropriate bin file(s) inside. 250-300-066
Main CBA
Faulty Main CBA Replace the Main CBA.
60-4304-042

6-10
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.6.2 TIENet 300 Connector


Case

Table 6-4 Troubleshooting: TIENet 300 Connector Case


Symptom Cause Action Part

Device not configured for display Add the parameters to the Home Display. Refer to
on the Home Display. Section 2.7.1 Site Setup.

Perform a hardware scan from TIENet Setup or SDI-12


Device has not been scanned. Setup. Refer to Sections 2.6.1 Smart Sensor Setup
(TIENet) and 2.6.2 SDI-12 Setup.

Device is not wired correctly. Rewire connector following label on the case circuit board.

TIENet or SDI12 Check fuse FU-T 3.15A (F1, F4, F5).


devices not appear- Open Fuse Replace if open. Refer to Figure 3-2 3.15A Fuse
ing on display for Connector case, connectors, and fuses, 411-0212-70
configuration Item K.
Refer to Section
2.6.1& 2.6.2. Defective TIENet or SDI12 Substitute a known working device and rescan. If it now
device. works, replace the faulty device.

Case circuit board faulty. 300 Connector


Case CBA
Substitute with known working board. 60-4304-041

Main CBA
Main CBA faulty.
60-4304-042

Check or cycle connections, then substi- Ribbon Cable


Ribbon cable damaged or loose.
tute with known working cable. 69-4304-032

6.6.3 TIENet 301 pH/Temp

Table 6-5 Troubleshooting: TIENet 301 pH/Temperature Device


Symptom Cause Action Part
No sensor connected to the 301 Connect pH probe

Rescan device in Hardware Setup


301 module not recognized
pH Will not calibrate Check TIENet wire connections. Follow wiring code
silk-screened on circuit board.
3.15A Fuse
TIENet connection fuse open Replace if open
411-0212-70
pH Probe
Probe defective Replace probe
60-9004-126

6-11
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-5 Troubleshooting: TIENet 301 pH/Temperature Device (Continued)


Symptom Cause Action Part

Buffers contaminated or wrong


Use new/correct pH buffer solution.
buffer used.

Temperature is not being read. Replace probe


Incorrect pH read- pH Probe
ings / slow response Probe bulb is contaminated Clean probe and recalibrate. If readings 60-9004-126
are still incorrect, replace probe.

Calibrated before reading stabi- Recalibrate and allow the readings to stabilize before con-
lized. tinuing with calibration.

6.6.4 TEINet 304 Contact


Output Card

Table 6-6 Troubleshooting: TIENet 304 Contact Output Card


Symptom Cause Action

A TIENET scan must be pre-


formed after the card is physi- Preform a TIENet scan.
The 304 card is
cally installed.
not a selectable
option in the
software after
the card is
installed. In the HARDWARE SETUP | SMART
SENSOR SETUP | CONFIGURE In the HARDWARE SETUP | SMART SENSOR SETUP | CONFIGURE
MEASUREMENTS, no options are MEASUREMENTS, select one or more of the options.
selected.

Contact closure
is not being The Contact Output card is not
made (verified configured correctly in the soft- See section 2.7.7 to configure the 304 card properly.
with an ohm ware.
meter)

6-12
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.6.5 TIENet 307 Analog


Input Card

Table 6-7 Troubleshooting: TIENet 307 Analog Input Card


Symptom Cause Action

A TIENet scan must be pre-


formed after the card is physi-Preform a TIENet scan.
The 307 card is
cally installed.
not a selectable
option in the
software after
the card is
installed. In the HARDWARE SETUP | SMART
SENSOR SETUP | CONFIGURE In the HARDWARE SETUP | SMART SENSOR SETUP | CONFIGURE
MEASUREMENTS, no options are MEASUREMENTS, select one or more of the options.
selected.

See section 2.7.7 to configure the 307 card properly. Use the
The analog input card is not green and yellow LED lamps to verify the hardware using the
configured correctly in the soft- set-up menus.
ware.

Verify the setting your device requires and change the active or
The passive/active setting is
passive settings on the 307 card. Verify the setting from the
incorrect or indeterminate.
set-up menus.

Verify the wires are attached to the correct channels and in the
The 307 card is correct polarity. Use the labels and LED lamps to verity the hard-
Improper connection.
not reading 4-20 ware using the setup menus. The green LED lamp will light when
proper current is flowing in the analog circuit.

External device is faulty or incor-


To verify, check the output with an ampmeter.
rectly configured.

Over-current protection device Disconnect power from the 307 card, the Signature, and the
in the 307 has been tripped. external analog loop. Wait 30 seconds.

6-13
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.6.6 TIENet 306 Sampler


Interface

Table 6-8 Troubleshooting: TIENet 306 Sampler Interface


Symptom Cause Action

Assign the correct sensor to the correct flow rate to the correct
Incorrect pacing Incorrect flow total selected for total flow.
interval pacing Example: Needed to pace from the 330 bubbler, but programmed
to pace from the 310 USLS.

Connect a different sampler, or test the existing sampler by short-


No sampler pac- Sampler’s flow pulse input not
ing pins A and C on the sampler’s Flow Meter port, while the pro-
ing working
gram is running. The displayed pulse count should count down.

6.6.7 TIENet 308 Analog


Output

Table 6-9 Troubleshooting: TIENet 308 4-20mA Analog Output


Symptom Cause Action
Incorrect wiring Rewire per connector diagram

Disconnect external equipment and test the output with


Excessive load VOM. If OK then reduce load resistance (maximum
900 ) or add isolated power to the current loop.
4-20 output is missing, or
zero current output

Use the other output channel on the 308 circuit board. If


current is still 0 mA, replace circuit board. If the VOM
Analog circuit board failure
reads 4mA or greater, reprogram to use that output or
replace the circuit board. Part #60-4304-006

Wires on incorrect output


Move connector to proper output and verify using the
(wired to output 2 instead of out-
yellow LED lamp indications from the setup menu.
put 1)
Analog percent is not selected in In the HARDWARE SETUP | SMART SENSOR SETUP | CON-
4-20 only reads 4mA the TIENet HARDWARE SETUP | FIGURE MEASUREMENTS, select one or more of the
SMART SENSOR SETUP | CONFIG- options. See sections 2.6.1 Smart Sensor Setup
URE MEASUREMENTS (TIENet) and 2.7.7 Inputs/Outputs/Alarms Setup.

Improper parameter set for the Verify/change the settings/range to the proper parame-
output. ter.

6-14
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-9 Troubleshooting: TIENet 308 4-20mA Analog Output (Continued)


Symptom Cause Action

Disconnect external equipment and test the output with


Excessive load VOM. If OK then reduce load resistance (maximum
900 ) or add isolated power to the current loop.

4-20 reading incorrectly Improper module/parameter set Verify/change the settings/range to the correct mod-
for the output ule/parameter.

Connected to incorrect output;


Move connector to proper output and verify using the
e.g., wired to output 2 instead of
yellow LED lamp indications from the setup menu.
output 1

The 308 option card is not


a selectable option in the TIENet 308 is not properly con- Verify the TIENet configuration contains analog percent
software after the card is figured readings
installed.
The output must have a load resistance (maximum
Measurement error for No load applied to the output cir- 900 ). For verification, this can be accomplished by
analog current cuit, or open circuit wiring. connecting the current meter leads to the terminals of
the 308 card.

External device not Over-current protection device in Disconnect power from the 307 card, the Signature, and
reading 4-20 the 307 has been tripped. the external analog loop. Wait 30 seconds.

6.6.8 TIENet 310 USLS

Table 6-10 Troubleshooting: TIENet 310 Ultrasonic Level Sensor


Symptom Cause Action

Not scanned Perform a smart sensor scan

Not able to achieve signal lock (misalignment,


Adjust mounting or place over a solid
loose mounting, turbulence, foam, or debris in
surface.
the water)

Invalid level, display has


asterisk (*) by level reading Level outside of the Blanking distances Adjust min/max blanking distances

Not wired correctly Check/repair wiring


Replace fuse FU-T 3.15A and rescan.
Open fuse Part #411-0212-70. Refer to Figure 3-2
Item K.
Failed sensor Replace with known good sensor

No level reading on the dis- Parameter not selected to be displayed on Add the parameter to the Home Dis-
play Home Display play. Refer to Section 2.7.1 Site Setup.

6-15
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-10 Troubleshooting: TIENet 310 Ultrasonic Level Sensor (Continued)


Symptom Cause Action
Level not adjusted properly Readjust level
Sensor misaligned Realign sensor

Incorrect level reading Adjust min/max blanking distances


Objects in the path of the signal
and/or reposition sensor.
Install sunshade. Refer to Appendix B
Sensor exposed to direct sunlight
Options and Accessories.

6.6.9 TIENet 330 Bubbler


Air Supply of Bubbler The bubbler’s air supply is brought into the air pump thru a des-
iccator, filter, humidity sensor, and into the intake of the air
pump.
Output of Air Pump Past the output of the air pump is a check valve that allows pres-
surized air from the air pump to enter the air tank. The air pump
is programmed to maintain a pressure in the air tank measured
by a pressure transducer (P4 referenced thru P3 vented to ref-
erence air).
Normal Operation During normal operation, the air purge solenoid is shut as air
passes thru the filter, bubble orifice, and to the bubble out. Air
passes thru the autozero valve to the pressure transducer P1(ref-
erenced to P2 which is vented to reference air). The pressure
measurement taken at P1 is used to calculate the level reading.
Purge Cycle During the purge cycle, the purge valve is opened allowing all the
pressure in the tank to vent though the bubble out.
Autozero The Signature automatically performs an autozero. During an
autozero, the autozero valve is activated tying P1 and P2
together so an autozero of the pressure transducer can be per-
formed.

6-16
Signature® Flow Meter
Section 6 Maintenance and Servicing

Figure 6-9 Plumbing diagram of bubbler

• P1 is the pressure measurement for the bubble line.


• P2 is the atmopheric reference for the bubble line.
• P4 is the pressure measurement for tank pressure.
• P4 is the atmopheric reference for the bubble line.
• Autozero Valve normally has the bottom port going to P2
open and top port going to the 4-way manifold (closed)
during the autozero cycle.

6-17
Signature® Flow Meter
Section 6 Maintenance and Servicing

Table 6-11 Troubleshooting: TIENet 330 Bubbler Module


Error Type Additional
Symptom Cause Action
(How to Clear) Information
Measurement error Bubble Pressure Flag error, no actions
(Level) <0.4PSI or >12 PSI prevented
Measurement error Temp/Humidity Flag error, no actions
(Temp/Humid.) reading failed prevented
Measurement out of Measured Level > Flag error, no actions
range High 3.2m prevented
Measurement out of Measured Level < Flag error, no actions
range Low 0.002m prevented
Plugged Bubble Line Tank Pressure Clears when Tank
>12PSI after purge Pressure <6.5PSI
Pump/Pressure Error Recharge time > 4s Continue level
readings if bubbling
Pump Motor Error Motor current sensed Flag error, no actions
too high or low prevented
Level not reading Blocked inlet/outlet Bubbler Basic check See section XXXX
correctly line not connected procedure
No Bubble out Blocked inlet/outlet Bubbler Basic check See section XXXX
line not connected procedure

6-18
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.7 330 Bubbler The TIENet 330 bubbler device is factory-installed for Signature
Installation bubbler flow meters. It can also be installed by the user to
convert a Signature flow meter into a bubbler, or to replace an
old 330 device.
Preparation steps are provided in Section 6.7.1 (converting a
non-bubbler unit) and Section 6.7.2 (replacing an existing
bubbler device). Instructions for installing the new bubbler are in
Section 6.7.3.

DANGER
Before opening the case, first ensure that mains
power is disconnected from the unit.

CAUTION
Before opening the case, disconnect the optional battery
backup power, if used.

Note
In order to work with the 330 Bubbler module, the Signature
must have an external desiccator installed. Refer to External
Desiccator, on page 5-12.

6.7.1 Preparation: Open the case, as described in Section 3.2.


Non-Bubbler If no bubbler was previously installed, remove the cover shield
over the main CBA. This will not be used again, since the 330
bubbler assembly includes its own cover shield.

Shield Mounting
Screws (4)

Figure 6-10 Remove non-bubbler shield

6-19
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.7.2 Preparation: Existing Open the case, as described in Section 3.2.


Bubbler When replacing an existing 330 bubbler, to ensure that the
bubble line tubing is reconnected correctly, label the tubing
ends, then remove the four mounting screws holding the 330
bubbler module in place (see Figure 6-11).

330 Mounting
Screws (4)

Figure 6-11 Remove bubbler module

6-20
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.7.3 Installation Procedure Referring to Figures 6-12, and 6-14, perform the following steps.
1. Place the bubbler assembly on top of the main CBA, ensur-
ing that the four screw holes line up in the case, and the
10-pin connector engages correctly in its socket.
2. Attach the bubbler board to the control panel using the
four self-tapping screws previously (part #231-6149-07).
Do not overtighten.

10-pin
connector

Figure 6-12 330 Bubbler assembly installation

Note
There are three pieces of tubing on the bubbler that must be
correctly connected for operation.
If any tubing is damaged, please purchase 1 meter of part
#029-1353-02 and cut to the length needed for replacement.
For additional information about tubing connections, refer to
Section A.3 330 Bubbler in Appendix A Replacement Parts.

6-21
Signature® Flow Meter
Section 6 Maintenance and Servicing

3. Replace the port plug in the bottom of the case (usually


second from the left) with the bubble line fitting.
4. Route the reference line tubing (with fitting on the end)
through the bushing and press the fitting into the refer-
ence connector on the connector case.
5. Route the intake tubing (the shorter of the two open-ended
tubes) through the bushing and behind the ribbon cable,
and connect it to the intake port in the case wall.
6. Route the other end of the short tubing through the bush-
ing and connect it to the humidity connector on the board.
7. Connect the bubble tubing (the longer of the two
open-ended tubes) to the bubble line port in the bottom of
the case.

Some Signature flow meters have a ‘Y’ fitting connecting both the
reference and intake ports to the humidity connector.
When installing a 330 bubbler in a unit with this tubing configu-
ration, remove the ‘Y’ fitting and two shorter pieces of tubing.
Bring the reference port tubing out from behind the ribbon and
reroute it directly to the humidity connector.

Intake port
tubing
(remove)
‘Y’ Fitting
(remove)

Humidity
connector tubing
(remove)

Reference port Before


tubing
After

Figure 6-13 Remove extra tubing and fitting (if applicable)

6-22
Signature® Flow Meter
Section 6 Maintenance and Servicing

Reference line

Intake
Port

Humidity
Connector

Reference
Line

Reference
Port

Bubble
Tubing

Figure 6-14 Routing and connections of 330 bubbler tubing

6-23
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.8 Front Cover A replacement front cover (door) comes with latches attached,
Replacement and two new hinge pins.
Align the hinges of the front cover front panel. Press the pins into
the hinge barrels, with each flange facing inward (refer to
Figure 6-15), until it is flush against the hinge surface.
Using a vice grip or other tool, spread and flatten the outward
facing ends of the pins so they cannot be removed from the
hinges.

Spread ends of pins

Pin flanges
flush.

Figure 6-15 Front cover (door) replacement

6.9 System Reset In the event that the Signature Flow Meter becomes unre-
sponsive, operation may be restored by removing and then
restoring line power.
If the problem persists, operation may be restored by performing
a hard reset.

CAUTION
A hard reset erases site data and restores the program to fac-
tory default settings.

To perform a hard reset, first remove line power and any external
battery power and wait 30 seconds. Then, while holding down
both the Home key and the Delete key, restore line power.

6-24
Signature® Flow Meter
Section 6 Maintenance and Servicing

6.10 Service and Repair Service tasks described in this manual may be performed on site
by properly trained personnel. Other service and repairs must be
performed at the factory. If your Teledyne Isco equipment
requires repair, contact Teledyne Isco technical support.
Teledyne Isco
Technical Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA

Phone:866 298-6174
402 464-0231
FAX:402 465-3085

E-mail:
[email protected]
Speaking with a Teledyne Isco Technical Service representative
can often resolve the problem without the need to return the
item. If the issue cannot be resolved by phone or email, you will
receive a Return Authorization Number (RAN) and information
on returning the equipment to the factory.

6-25
Signature® Flow Meter
Section 6 Maintenance and Servicing

6-26
Signature® Flow Meter
Appendix A Replacement Parts

Replacement parts are called out in illustrations in this section.


Reference the call-outs in the accompanying tables to determine
the part number for the item.

A.1 How to Order Replacement parts can be purchased by contacting Teledyne


Isco’s Customer Service Department.
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA

Phone: 800 228-4373


402 464-0231
FAX: 402 465-3022

E-mail:[email protected]

A-1
Signature® Flow Meter

A.2 Signature Flow Meter

A-2
Signature® Flow Meter

A-3
Signature® Flow Meter

A-4
Signature® Flow Meter

A-5
Signature® Flow Meter

A-6
Signature® Flow Meter

A-7
Signature® Flow Meter

A-8
Signature® Flow Meter

A-9
Signature® Flow Meter

A-10
Signature® Flow Meter

A-11
Signature® Flow Meter

A-12
Signature® Flow Meter

A-13
Signature® Flow Meter

A.3 330 Bubbler

A-14
Signature® Flow Meter

A-15
Signature® Flow Meter

A-16
Signature® Flow Meter

A-17
Signature® Flow Meter

A-18
Signature® Flow Meter
Appendix B Options and Accessories

B.1 Ordering Information Options and accessories can be purchased by contacting Teledyne
Isco’s Customer Service Department.
Teledyne Isco
Customer Service Dept.
P.O. Box 82531
Lincoln, NE 68501 USA

Phone: 800 228-4373


402 464-0231
FAX: 402 465-3022

E-mail: [email protected]

Note
For options and accessories exclusive to the external TIENet
devices, refer to the appropriate TIENet user manuals (found
at www.isco.com and listed in Section B.7 Manuals).

B.2 Signature Flow Meter


Accessories
Cord grip fitting for TIENet cable ........................................................................................... 209-0073-12
3/4" NPT .375/.437" OD

Cord grip fitting for Battery backup cable .............................................................................. 209-0073-11


3
/4" NPT .250/.375" OD
Cord grip fitting with flexible strain protection ..................................................................... 60-4307-022
Cord grip fitting for rain gauge and option card circuits, 3 hole ........................................... 60-4304-080
Exterior desiccator - Required for use with 330 and 350 TIENet devices ............................ 60-4354-019
Silica gel desiccant, 1.5-lb container ....................................................................................... 099-0011-03
Model 674 rain gauge connect cable for Signature................................................................. 60-4304-055
TIENet connection cable for Signature ................................................................................... 60-4304-056
TIENet connection cable for Signature, cut to length ............................................................ 60-4304-068
TIENet ‘Y’ connection cable ..................................................................................................... 60-4304-066
TIENet Header 6 screw clamp plug ........................................................................................ 69-4303-102
Analog Output 3 screw clamp header clip .............................................................................. 60-4304-085
Analog Input 3 screw clamp header clip ................................................................................. 60-4304-083
Contact Output 3 screw clamp header clip ............................................................................. 60-4304-084
Power Supply Assembly with wiring harness......................................................................... 60-4304-037
TIENet Expansion Box w/ 10 ft cable ..................................................................................... 60-4307-023
TIENet Expansion Box w/ 10 ft cable and reference air support60-4357-018Bulk TIENet Cable, cut to
Length ....................................................................................................................................... 60-4304-050
USB Flash drive, 2GB.............................................................................................................. 250-3000-42

B-1
Signature® Flow Meter

Adaptor cable, USB Micro to USB-A....................................................................................... 480-2946-02


USB connect cable, Signature to PC ....................................................................................... 480-2946-01
USB Flash drive, 2GB .............................................................................................................. 250-3000-42
Option wire 3Conductor 22 GA (use with 604304080) ........................................................... 480-1446-01
ProHanger SST Suspension bracket for 18 - 24in. manhole shaft ........................................ 69-2003-599
Spreader bar for suspension of sensor or flow meter in manhole shaft ................................ 60-3004-110
Suspension harness................................................................................................................. 60-1704-016
7-Digit, non-resettable mechanical totalize ........................................................................... 60-4304-015
Portable Stand Assembly ........................................................................................................ 60-4304-079
Cord Grip 3-Hole for option wiring......................................................................................... 60-4304-080
Rain Gauge 0.1in ...................................................................................................................... 60-3284-001
Rain Gauge 0.1 mm .................................................................................................................. 60-3284-006
Model 674 rain gauge connect cable for Signature................................................................. 60-4304-055
Reference port tubing kit ......................................................................................................... 60-4307-017
Dri-Can Desiccant ................................................................................................................... 099-0012-00
8oz. Desiccant bag .................................................................................................................... 099-0002-00
Gore Filter ................................................................................................................................ 209-0093-93
External Desiccator Assembly ................................................................................................. 60-4304-013

B.3 Power Accessories


Battery Backup: 946 Lead-acid battery pack, connect cable, and
battery mounting hardware..................................................................................................... 60-4307-015
DC Power Cable with cord grip ............................................................................................... 60-4304-086
117V Power cord kit ................................................................................................................. 60-4304-044
Includes strain relief cord-grip fitting
117V Power cord ....................................................................................................................... 60-1683-216
240V Power cord kit ................................................................................................................. 60-4304-045
Includes strain relief cord-grip fitting
240V Power cord....................................................................................................................... 48-0124-101
946 Lead Acid Battery ............................................................................................................. 60-3004-106
948 Lead Acid Battery 45A-H ................................................................................................ 68-3000-948
948 Batttery Charger 12V 6A.................................................................................................. 34-1011-812
913 Power Pack 117V............................................................................................................... 60-1684-088
914 Power Pack w/ Battery Backup 117 V ............................................................................. 68-3004-130
923 Power Pack 240V............................................................................................................... 68-3004-190
924 Power Pack w/ Battery Backup 240 V ............................................................................. 68-3004-160
Cord grip w/protection ............................................................................................................. 60-4307-022
965 Battery Charger 5 station 117V ...................................................................................... 68-3000-965
966 Battery Charger 5 station 117V ...................................................................................... 68-3000-966
963 Battery Charger 1 station 117/240V ............................................................................... 68-3004-198

B-2
Signature® Flow Meter

Signature Power Loss Alarm Box............................................................................................ 60-5324-024


Contact the Teledyne Isco Special Product Applications
department for ordering information
240V Power cord ....................................................................................................................... 48-0124-101

B.4 TIENet Accessories

B.4.1 Internal to Signature


Analog Output Option Card..................................................................................................... 60-4304-006
Analog Input Option Card ...................................................................................................... 60-4304-054
Contact Output Card ............................................................................................................... 60-4304-069

B.4.2 Sensor and Cabling


TIENet connection cable for Signature .................................................................................. 60-4304-056
TIENet connection cable for Signature, cut to length ............................................................ 60-4304-068
CA Assembly TIENet Y w/ connector...................................................................................... 60-4304-066
TIENet Expansion Box (includes 10 ft TIENet cable and 2 cord grips)................................ 60-4307-023
TIENet Expansion Box with reference line support ............................................................. 60-4357-018
Cord grip fitting, 3/4" NPT, for TIENet cable ........................................................................ 209-0073-12
Bulk TIENet Cable, cut-to-length, order by the foot ............................................................. 60-4304-050

B.4.3 301 pH/Temperature For use with Signature 6 position plug-in (green) terminal
Device with Signature strip.
Connection Ending in
Includes cord grip, combination pH probe with built-in exposed
Unterminated Leads
temperature probe and 25 ft. probe
cable, and one package of each buffer and rinse solution for probe
calibration.

10 m cable* .............................................................................................................................. 60-4307-018


23 m cable* .............................................................................................................................. 60-4307-019
Cut-to-length cable* ................................................................................................................ 60-4307-025
*Cable lengths from Signature to TIENet 301 device

B.4.4 301 pH/Temperature For use with portable Signature TIENet receptacle.
Device with Signature Includes combination pH probe with built-in exposed temperature
Connection Ending in probe and 25 ft. probe cable and one package of each buffer and
TIENet Plug
rinse solution for probe calibration
10 m cable* .............................................................................................................................. 60-4307-070
23 m cable* .............................................................................................................................. 60-4307-071
Cut-to-length cable* ................................................................................................................ 60-4307-072
*Cable lengths from Signature to TIENet 301 device

B-3
Signature® Flow Meter

B.4.5 306 Sampler Interface For use with Signature 6 position plug-in (green) terminal
Cable with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads below).
306 Sampler Interface w/ 10 m (32.8 ft) cable ........................................................................ 60-4304-007
306 Sampler Interface w/ 23 m (75 ft) cable ........................................................................... 60-4304-008
306 Sampler Interface w/ CTL cable ....................................................................................... 60-4304-088
Assembly Model 306 5m (5800/4700 OLNY) ......................................................................... 60-5804-178
Assembly Model 306 10m (5800/4700 OLNY) ........................................................................ 60-5804-179
Assembly Model 306 23m (5800/4700 OLNY) ....................................................................... 60-5804-180

B.4.6 306 Sampler Interface For use with portable Signature TIENet receptacle.
Cable with Signature
Connection Ending in
TIENet Plug.

306 Sampler Interface w/ connector and 10m (32.8 ft) cable ................................................. 60-4304-076
306 Sampler Interface w/ connector and 23 m (75 ft) cable ................................................... 60-4304-077
306 Sampler Interface w/ connector and CTL cable............................................................... 60-4304-078
4700/5800 Sampler Interface cable ........................................................................................ 60-5314-697

B.4.7 310 Ultrasonic Level For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads.
below).
310 Ultrasonic sensor w/ 10m cable ....................................................................................... 60-4314-005
310 Ultrasonic sensor w/ 23m cable ....................................................................................... 60-4314-006
310 Ultrasonic sensor Cut-to-length* .................................................................................... 60-4314-014
. *Cable lengths can go up to 150 m with an expansion box.

B.4.8 310 Ultrasonic Level For use with portable Signature TIENet receptacle.
Sensor with Signature Includes cord grip and sensor with cable. (See cable lengths
Connection Ending in below).
TIENet Plug.
310 Ultrasonic sensor w/ connector and 10m cable ............................................................... 60-4314-009
310 Ultrasonic sensor w/ connector and 23m cable ............................................................... 60-4314-011
310 Ultrasonic sensor w/ connector Cut-to-length* ............................................................... 60-4314-013
*Cable lengths can go up to 150 m with an expansion box.

B-4
Signature® Flow Meter

B.5 330 Bubble Options


and Accessories
330 Bubbler internal sensor ................................................................................................... 60-4334-003
Bubble line, PTFE, 1/16" x 25ft ............................................................................................. 60-1873-051
Bubble line, vinyl, 1/8" x 50ft ................................................................................................. 60-1873-044
Bubble line, vinyl, 1/8" x 100ft................................................................................................ 60-1700-003
SST Bubble tube, 4ft long - for PTFE, 1/16" Line.................................................................. 60-1704-018
SST Bubble tube, 4ft long - for PTFE, 1/8" Line.................................................................... 60-1873-043
Bubble line carrier - attach to Isco Mounting Ring ............................................................... 60-3204-007

B.5.1 350 Area Velocity For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
Connection Ending in Includes cord grip and sensor with cable. (See cable lengths
Unterminated Leads.
below).
350 AV Sensor w/ 10m Cable .................................................................................................. 60-4354-013
350 AV Sensor w/ 23m Cable .................................................................................................. 60-4354-014
350 Ultrasonic sensor Cut-to-length ...................................................................................... 60-4354-026
350 Ultrasonic sensor Cut-to-length ...................................................................................... 60-4354-026

B.5.2 350 Area Velocity For use with portable Signature TIENet receptacle.
Sensor with Signature
Connection Ending in
TIENet Plug.
350 AV Sensor w/ connector and 2ft connector..................................................................... 60-4354-017
350 AV Sensor w/ connector and 10m cable........................................................................... 60-4354-023
350 AV Sensor w/ connector and 23m cable........................................................................... 60-4354-024
350 AV Sensor w/ connector Cut-to-length ............................................................................ 60-4354-025

B.5.3 360 LaserFlow Velocity For use with Signature 6 position plug-in (green) terminal
Sensor with Signature strip.
connection ending in Includes cord grip and sensor with cable. (See cable lengths
unterminated leads.
below).
360 Laserflow sensor w/ 10m cable ........................................................................................ 60-4364-056
360 Laserflow sensor w/ 23m cable ........................................................................................ 60-4364-057
360 Laserflow sensor Cut-to-length*...................................................................................... 60-4364-076
*Cable lengths can go up to 150 m
360 Laserflow sensor w/ 10m cable + USNC ......................................................................... 60-4364-062
360 Laserflow sensor w/ 23m cable + USNC ......................................................................... 60-4364-063
360 Laserflow sensor Cut-to-length* + USNC....................................................................... 60-4364-078
*Cable lengths can go up to 150 m

B-5
Signature® Flow Meter

B.5.4 360 LaserFlow Velocity For use with portable Signature TIENet receptacle.
Sensor with Signature
Connection Ending in
TIENet Plug.
360 Laserflow sensor w/ connector and 10m cable ................................................................ 60-4364-058
360 Laserflow sensor w/ connector and 23m cable ................................................................ 60-4364-060
360 Laserflow sensor w/ connector Cut-to-length*................................................................ 60-4364-077
*Cable lengths can go up to 150 m

360 Laserflow sensor w/ connector and 10m cable +USNC ................................................. 60-4364-064
360 Laserflow sensor w/ connector and 23m cable +USNC .................................................. 60-4364-066
360 Laserflow sensor w/ connector Cut-to-length* +USNC ................................................. 60-4364-079
*Cable lengths can go up to 150 m

B.6 Modems
CDMA Digital cellular modem
(Cellular service not included.) ................................................................................................ 60-4307-074
Magnetic mount antenna for CDMA ....................................................................................... 60-2004-550
GSM Digital cellular modem (Outside North America)
(Cellular service not included.) ................................................................................................ 60-4307-090
GSM Digital cellular modem (North America)
(Cellular service not included.) ................................................................................................ 60-4307-075
Magnetic mount antenna for GSM.......................................................................................... 60-2004-551
Ethernet modem ....................................................................................................................... 60-4307-016

B.7 Manuals
Signature Flow Meter .............................................................................................................. 69-4303-070
Signature Bubbler Flow Meter ................................................................................................ 69-4333-004
TIENet 301 pH/Temperature Device....................................................................................... 69-4303-071
TIENet 306 Sampler Interface ................................................................................................ 69-4303-072
TIENet 310 Ultrasonic Sensor................................................................................................. 69-4313-010
TIENet 350 Area Velocity Sensor ............................................................................................ 69-4353-024
TIENet 360 LaserFlow Sensor ................................................................................................ 69-4363-043
Mounting Rings ........................................................................................................................ 60-3203-061
Flow Metering Insert ............................................................................................................... 60-3234-064
Flowlink 5.1 Software .............................................................................................................. 69-2543-213
Isco Open Channel Flow Measurement Handbook, 6th Ed. .................................................. 60-3003-041

B.8 Sensor Mounting


Rings

B.8.1 Spring Rings


Probe Mounting Ring for 6" pipe ............................................................................................. 68-3200-007
Probe Mounting Ring for 8" pipe ............................................................................................. 68-3200-008
Probe Mounting Ring for 10" pipe ........................................................................................... 68-3200-009
Probe Mounting Ring for 12" pipe ........................................................................................... 68-3200-010
Probe Mounting Ring for 15" pipe ........................................................................................... 68-3200-011

B-6
Signature® Flow Meter

B.8.2 Scissor Rings


Base Section (with tabs for mounting up to five probes) ........................................................ 60-3004-169
Scissors Assembly..................................................................................................................... 60-3004-170
Extension 1 (9.0") ..................................................................................................................... 60-3004-172
Extension 2 (21.5") ................................................................................................................... 60-3004-173
Extension 3 (31.5") ................................................................................................................... 60-3004-174
Extension 4 (41.5") ................................................................................................................... 60-3004-175
16-24" Pipe ID........................................................................................................................... 68-3000-042
26-38" Pipe ID .......... ………………………………………………………………………………... 68-3000-043
38-44" Pipe ID ..........………………………………………………………………………………… 68-3000-044
44-48" Pipe ID ..........………………………………………………………………………………… 68-3000-045
60" Pipe ID ........... …………………………………………………………………………………… 68-3000-046
72" Pipe ID ........... …………………………………………………………………………………… 68-3000-047
16-60" Pipe ID .......... ………………………………………………………………………………... 68-3000-048
Base Section (with plastic ties and manual). …………………………………………………… 60-3004-169

Note
Scissor Mounting Ring Assemblies will require a base and
scissors section for all sizes. Each scissors ring includes a
base section, scissors mechanism, extensions, plastic ties,
and a manual). Sizes from 16" to 80" will also require two or
more extension sections.

B-7
Signature® Flow Meter

B-8
Signature® Flow Meter
Appendix C Modbus Output Protocol

Sections C.1 through C.3 give an overview of the basic capabil-


ities and operation of Modbus output protocol as it applies to the
Isco Signature Flow Meter.
For a Glossary of Terms and Common Acronyms, see Section C.4.

C.1 Introduction Modbus is a simple command/response mechanism to read from


and write to specific memory locations called registers. A register
is a holding place for a piece of digital information within the
equipment. The Signature uses Modbus ASCII and Modbus RTU
protocols, providing a standard protocol for real-time data
retrieval. The data can be sent to a central computer for display,
data collection, or process control. Modbus cannot be used to
retrieve historical data from the Signature’s memory.
This section describes the overall capabilities and operation of
Modbus.

C.2 Operation There are many standard, third party Modbus drivers and OPC
servers that may be used to link a remote Modbus device, such as
the Signature Flow Meter, to SCADA or process control software,
such as Wonderware®1 or Intellution®2. The OPC server commu-
nicates with the flow meter and accesses registers. The definition
of what information is contained and where (the register number,
or address) is set by Teledyne Isco.
The Signature register addresses, and what parameters are held
where, are available in Table C-1 in Section C.6.1.
By accessing these registers you can obtain the current value of
whatever parameter you desire. The reading(s) can then be dis-
played or stored wherever you designate as a destination; for
example, a process control computer.
Not all registers are limited to read-only data storage. You can
also use two of the registers for control purposes. For example,
writing a “1” value to register 25 (“TakeReadingFlag” register),
tells the Signature to update its readings.

1. Wonderware® is a registered trademark of Wonderware Software


Development Corporation.
2. Intellution® is a registered trademark of Intellution, Inc.

C-1
Signature® Flow Meter

C.3 Configurations A variety of configurations can be made with Modbus, either


through direct connection or through a modem.
In the example shown in Figure C-1, you are direct-connecting a
server PC to two individual Signature sites through Modbus,
using the COM ports on the OPC Server, which are directly con-
nected to the remote sites.
Connection to the flow meter is made via the RS-485 terminal on
the Signature case board (refer to Figure 3-2 Connector case,
connectors, and fuses).

Signature

OPC Process
Server Control

Signature

Figure C-1 Configuration example

The operation sequence for the example above can be summa-


rized in the following steps:
1. Signatures take readings from sensors.
2. Signatures store readings (level, velocity, flow rate, etc.) in
their specified registers.
3. The user requests data through Process Control.
4. Process Control asks the OPC server to gather informa-
tion.
5. OPC connects to the specified Signature flow meter
through the cable (direct connection), retrieves site data
and populates the OPC server's holding index.
6. Process Control takes the data from the OPC server's hold-
ing index and gives the data to the user.
Note that Process Control can be either manual or automated in
this example, and that the OPC server and Process Control may
be located physically on the same computer.

C-2
Signature® Flow Meter

C.4 Glossary of Terms Address – An address is a digital location specified for a device
(such as the Signature Flow Meter).
ASCII – Short for American Standard Code for Information
Interchange, ASCII is a code that represents English characters
with numbers. Most computers represent text with ASCII code,
making it possible for one computer or device to share data with
another.
DCS – Distributed Control Systems
Dedicated Line – A telecommunications path reserved for com-
munication between two specified points and not shared among
multiple points.
Modbus Protocol – Modbus Protocol is a messaging structure
used to establish master-slave/client server communications
between intelligent devices. Modbus is a simple
command/response mechanism to read from and write to reg-
isters.
MTU – Master Terminal Unit
OPC – OPC (OLE for Process Control) means open connectivity
via open (free for use) standards. It is a series of software stan-
dards specifications that fill a need in automation (like printer
drivers did for Windows), acting as a translator for data trans-
mission and process control.
The specification defines a standard set of objects, interfaces, and
methods for use in process control and manufacturing auto-
mation applications to facilitate interoperability. There are hun-
dreds of OPC Data Access servers and clients.
PLC – Programmable Logic Controller
Registers – A register is a location in memory for a specific data
type. The definition of what data is contained and where (the reg-
istry number, or address) is set by the equipment manufacturer
(in this case Teledyne Isco).
RTU – Short for Remote Terminal Unit (or Remote Telemetry
Unit), RTU is a code that represents data using a compact binary
format.
SCADA – SCADA (Supervisory Control And Data Acquisition)
is a computer system for gathering and analyzing real-time data.
SCADA systems are used to monitor and control plant operation,
or equipment in industries.
The SCADA system transfers the information (for example,
where a leak has occurred in a pipeline), back to a central site,
alerting the home station of the leak, performing necessary
analysis and control (such as determining if the leak is critical),
and displaying the information in a logical and organized
manner.
SCADA systems can be relatively simple, such as one that mon-
itors the environmental conditions of a small office building, or
very complex, such as a system that monitors all the activity in a
nuclear power plant or a municipal water system.
TCP/IP – Transmission Control Protocol/Internet Protocol

C-3
Signature® Flow Meter

C.5 Signature ASCII or The Signature’s address (Device ID) is user-programmable


RTU Address between 2 and 247.

CAUTION
Be careful not to assign the same address to more than one
flow meter.

C.6 Register Definitions The register definitions for the Signature flow meter are pro-
vided in the following table.
Where no other Unit Of Measure exists for a parameter, percent
(%) can be used in most situations.

C.6.1 Modbus Registers

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40025 TakeReadingFlag Word R/W Set to 1 to update readings, 2 for automatic
update
40026 UpdateInterval Word Seconds R/W The reading update interval in seconds
40027 Activeflags Word R The bit fields to indicate which sensors are
active
40040 Level Float R Level
Meters
40041
40042 Levelstatus Word R The level status - non-zero is an error
40043 - 48 Leveltime Word R The last level reading time,
sec-min-hour-day-month-year
40055 Level1 Float R Level 1
Meters
40056
40057 Level1status Word R The level 1 status - non-zero is an error
40058 - 63 Level1time Word R The last level 1 reading time,
sec-min-hour-day-month-year
40070 Level2 Float R Level 2
Meters
40071
40072 Level2status Word R The level 2 status - non-zero is an error
40073 - 78 Level2time Word R The last level 2 reading time,
sec-min-hour-day-month-year
40085 Level3 Float R Level 3
Meters
40086
40087 Level3status Word R The level 3 status - non-zero is an error
40088 - 93 Level3time Word R The last level 3 reading time,
sec-min-hour-day-month-year
40100 Velocity Float R Velocity
Meters/Sec
40101

C-4
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40102 Velocitystatus Word R 0 = Good, 1 = Failed to Measure, 4 = below min
Depth
40103 - 08 Velocitytime Word R The last velocity reading time,
sec-min-hour-day-month-year
40115 Velocity1 Float R Velocity 1
Meters/Sec
40116
40117 Velocity1status Word R 0 = Good, 1 = Failed to Measure, 4 = below min
Depth
40118 - 23 Velocity1time Word R The last velocity 1 reading time,
sec-min-hour-day-month-year
40130 Velocity2 Float R Velocity 2
Meters/Sec
40131
40132 Velocity2status Word R 0 = Good, 1 = Failed to Measure, 4 = below min
Depth
40133 - 38 Velocity2time Word R The last velocity 2 reading time,
sec-min-hour-day-month-year
40145 Velocity3 Float R Velocity 3
Meters/Sec
40146
40147 Velocity3status Word R 0 = Good, 1 = Failed to Measure, 4 = below min
Depth
40148 - 53 Velocity3time Word R The last velocity 3 reading time,
sec-min-hour-day-month-year
40160 Flowrate Float Cubic R Flow rate
40161 Meters/Sec
40162 Flowratestatus Word R The flow rate status - non-zero is an error
40163 - 68 Flowratetime Word R The last flow rate reading time,
sec-min-hour-day-month-year
40175 Flowrate1 Float Cubic R Flow rate 1
40176 Meters/Sec
40177 Flowrate1status Word R The flow rate 1 status - non-zero is an error
40178 - 83 Flowrate1time Word R The last flow rate1 reading time,
sec-min-hour-day-month-year
40190 Flowrate2 Float Cubic R Flow rate 2
40191 Meters/Sec
40192 Flowrate2status Word R The flow rate 2 status - non-zero is an error
40193 - 98 Flowrate2time Word R The last flow rate 2 reading time,
sec-min-hour-day-month-year
40205 Flowrate3 Float Cubic R Flow rate 3
40206 Meters/Sec
40207 Flowrate3status Word R The flow rate 3 status - non-zero is an error
40208 - 13 Flowrate3time Word R The last flow rate 3 reading time,
sec-min-hour-day-month-year

C-5
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40220 Temperature Float Degrees R Temperature
40221 Celsius
40222 Temperaturestatus Word R The temperature status - non-zero is an error
40223 - 28 Temperaturetime Word R The last temperature reading time,
sec-min-hour-day-month-year
40235 Temperature1 Float Degrees R Temperature 1
40236 Celsius
40237 Temperature1status Word R The temperature 1 status - non-zero is an error
40238 - 43 Temperature1time Word R The last temperature 1 reading time,
sec-min-hour-day-month-year
40250 Temperature2 Float Degrees R Temperature 2
40251 Celsius
40252 Temperature2status Word R The temperature 2 status - non-zero is an error
40253 - 58 Temperature2time Word R The last temperature reading time,
sec-min-hour-day-month-year
40265 Temperature3 Float Degrees R Temperature 3
40266 Celsius
40267 Temperature3status Word R The temperature 3 status - non-zero is an error
40268 - 73 Temperature3time Word R The last temperature 3 reading time,
sec-min-hour-day-month-year
40280 Volume Float Cubic R Volume
40281 Meters
40282 Volumestatus Word R The volume status - non-zero is an error
40283 - 88 Volumetime Word R The last volume reading time,
sec-min-hour-day-month-year
40295 Volume1 Float Cubic R Volume 1
40296 Meters
40297 Volume1status Word R The volume 1 status - non-zero is an error
40298 - 303 Volume1time Word R The last volume 1 reading time,
sec-min-hour-day-month-year
40310 Volume2 Float Cubic R Volume 2
40311 Meters
40312 Volume2status Word R The volume 2 status - non-zero is an error
40313 - 18 Volume2time Word R The last volume 2 reading time,
sec-min-hour-day-month-year
40325 Volume3 Float Cubic R Volume 3
40326 Meters
40327 Volume3status Word R The volume 3 status - non-zero is an error
40328 - 33 Volume3time Word R The last volume 3 reading time,
sec-min-hour-day-month-year
40340 Voltage Float R Voltage
Volts
40341

C-6
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40342 Voltagestatus Word R The voltage status - non-zero is an error
40343 - 48 Voltagetime Word R The last voltage reading time,
sec-min-hour-day-month-year
40355 Voltage1 Float R Voltage 1
Volts
40356
40357 Voltage1status Word R The voltage 1 status - non-zero is an error
40358 - 63 Voltage1time Word R The last voltage 1 reading time,
sec-min-hour-day-month-year
40370 Voltage2 Float R Voltage 2
Volts
40371
40372 Voltage2status Word R The voltage 2 status - non-zero is an error
40373 - 78 Voltage2time Word R The last voltage 2 reading time,
sec-min-hour-day-month-year
40385 Voltage3 Float R Voltage 3
Volts
40386
40387 Voltage3status Word R The voltage 3 status - non-zero is an error
40388 - 93 Voltage3time Word R The last voltage 3 reading time,
sec-min-hour-day-month-year
40400 Analog/% Float 4-20mA/ R Analog output or percentage
40401 0-100%
40402 Analog/status Word R The Analog output or percentage status -
non-zero is an error
40403 - 08 Analog/time Word R The last analog or percentage reading time,
sec-min-hour-day-month-year
40415 Analog/1 Float 4-20mA/ R Analog 1 output or percentage
40416 0-100%
40417 Analog/1status Word R The Analog 1 output or percentage status -
non-zero is an error
40418 - 23 Analog/1time Word R The last analog 1 or percentage reading time,
sec-min-hour-day-month-year
40430 Analog/2 Float 4-20mA/ R Analog 2 output or percentage
40431 0-100%
40432 Analog/2status Word R The Analog 2 output or percentage status -
non-zero is an error
40433 - 38 Analog/2time Word R The last analog 2 or percentage reading time,
sec-min-hour-day-month-year
40445 Analog/3 Float 4-20mA/ R Analog 3 output or percentage
40446 0-100%
40447 Analog/3status Word R The Analog 3 output or percentage status -
non-zero is an error
40448 - 53 Analog/3time Word R The last analog 3 or percentage reading time,
sec-min-hour-day-month-year

C-7
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40460 Analog/4 Float 4-20mA/ R Analog 4 output or percentage
40461 0-100%
40462 Analog/4status Word R The Analog 4 output or percentage status -
non-zero is an error
40463 - 68 Analog/4time Word R The last analog 4 or percentage reading time,
sec-min-hour-day-month-year
40475 Analog/5 Float 4-20mA/ R Analog 5 output or percentage
40476 0-100%
40477 Analog/5status Word R The Analog 5 output or percentage status -
non-zero is an error
40478 - 83 Analog/5time Word R The last analog 5 or percentage reading time,
sec-min-hour-day-month-year
40490 Analog/6 Float 4-20mA/ R Analog 6 output or percentage
40491 0-100%
40492 Analog/6status Word R The Analog 6 output or percentage status -
non-zero is an error
40493 - 98 Analog/6time Word R The last analog 6 or percentage reading time,
sec-min-hour-day-month-year
40505 Analog/7 Float 4-20mA/ R Analog 7 output or percentage
40506 0-100%
40507 Analog/7status Word R The Analog 7 output or percentage status -
non-zero is an error
40508 - 13 Analog/7time Word R The last analog 7 or percentage reading time,
sec-min-hour-day-month-year
40520 Analog/8 Float 4-20mA/ R Analog 8 output or percentage
40521 0-100%
40522 Analog/8status Word R The Analog 8 output or percentage status -
non-zero is an error
40523 - 28 Analog/8time Word R The last analog 8 or percentage reading time,
sec-min-hour-day-month-year
40535 Analog/9 Float 4-20mA/ R Analog 9 output or percentage
40536 0-100%
40537 Analog/9status Word R The Analog 9 output or percentage status -
non-zero is an error
40538 - 43 Analog/9time Word R The last analog 9 or percentage reading time,
sec-min-hour-day-month-year
40550 Analog/10 Float 4-20mA/ R Analog 10 output or percentage
40551 0-100%
40552 Analog/10status Word R The Analog 10 output or percentage status -
non-zero is an error
40553 - 58 Analog/10time Word R The last analog 10 or percentage reading time,
sec-min-hour-day-month-year
40565 Analog/11 Float 4-20mA/ R Analog 11 output or percentage
40566 0-100%

C-8
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40567 Analog/11status Word R The Analog 11 output or percentage status -
non-zero is an error
40568 - 73 Analog/11time Word R The last analog 11 or percentage reading time,
sec-min-hour-day-month-year
40580 Analog/12 Float 4-20mA/ R Analog 12 output or percentage
40581 0-100%
40582 Analog/12status Word R The Analog 12 output or percentage status -
non-zero is an error
40583 - 88 Analog/12time Word R The last analog 12 or percentage reading time,
sec-min-hour-day-month-year
40595 Analog/13 Float 4-20mA/ R Analog 13 output or percentage
40596 0-100%
40597 Analog/13status Word R The Analog 13 output or percentage status -
non-zero is an error
40598 - 603 Analog/13time Word R The last analog 13 or percentage reading time,
sec-min-hour-day-month-year
40610 Analog/14 Float 4-20mA/ R Analog 14 output or percentage
40611 0-100%
40612 Analog/14status Word R The Analog 14 output or percentage status -
non-zero is an error
40613 - 18 Analog/14time Word R The last analog 14 or percentage reading time,
sec-min-hour-day-month-year
40625 Analog/15 Float 4-20mA/ R Analog 14 output or percentage
40626 0-100%
40627 Analog/15status Word R The Analog 15 output or percentage status -
non-zero is an error
40628 - 33 Analog/15time Word R The last analog 15 or percentage reading time,
sec-min-hour-day-month-year
40700 Fluoresence Float % R
40701
40702 Fluoresencestatus Word R
40703 Fluoresencetime Word R
40715 Fluoresence1 Float % R
40716
40717 Fluoresence1status Word R
40718 - 23 Fluoresence1time Word R
40730 - 31 Fluoresence2 Float % R
40732 Fluoresence2status Word R
40733 - 38 Fluoresence2time Word R
40745 - 46 Fluoresence3 Float % R
40747 Fluoresence3status Word R

C-9
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40748 - 53 Fluoresence3time Word R
40760 Battery Float Volts R
40761
40762 Batterystatus Word R
40763 - 68 Batterytime Word R
40775 Battery1 Float Volts R
40776
40777 Battery1status Word R
40778 - 83 Battery1time Word R
40790 Battery2 Float Volts R
40791
40792 Battery2status Word R
40793 - 98 Battery2time Word R
40805 Battery3 Float Volts R
40806
40807 Battery3status Word R
40808 - 13 Battery3time Word R
40820 Dissolved Gas Float mmHg R
40821
40822 Dissolved Gasstatus Word R
40823 - 28 Dissolved Gastime Word R
40835 Dissolved Gas1 Float mmHg R
40836
40837 Dissolved Gas1status Word R
40838 - 43 Dissolved Gas1time Word R
40850 Dissolved Gas2 Float mmHg R
40851
40852 Dissolved Gas2status Word R
40853 - 58 Dissolved Gas2time Word R
40865 Dissolved Gas3 Float mmHg R
40866
40867 Dissolved Gas3status Word R
40868 - 73 Dissolved Gas3time Word R
40940 Photosyn Rad Float umol s1 m2 R
40941
40942 Photosyn Radstatus Word R
40943 - 48 Photosyn Radtime Word R
40955 Photosyn Rad1 Float umol s1 m2 R
40956

C-10
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
40957 Photosyn Rad1status Word R
40958 - 63 Photosyn Rad1time Word R
40970 Photosyn Rad2 Float umol s1 m2 R
40971
40972 Photosyn Rad2status Word R
40973 - 78 Photosyn Rad2time Word R
40985 Photosyn Rad3 Float umol s1 m2 R
40986
40987 Photosyn Rad3status Word R
40988 - 93 Photosyn Rad3time Word R
41000 Transmissivity Float % R
41001
41002 Transmissivitystatus Word R
41003 - 08 Transmissivitytime Word R
41015 Transmissivity1 Float % R
41016
41017 Transmissivity1status Word R
41018 - 23 Transmissivity1time Word R
41030 Transmissivity2 Float % R
41031
41032 Transmissivity2status Word R
41033 - 38 Transmissivity2time Word R
41045 Transmissivity3 Float % R
41046
41047 Transmissivity3status Word R
41048 - 53 Transmissivity3time Word R
41060 Conductivity Float uS/cm R
41061
41062 Conductivitystatus Word R
41063 - 68 Conductivitytime Word R
41075 Conductivity1 Float uS/cm R
41076
41077 Conductivity1status Word R
41078 - 83 Conductivity1time Word R
41090 Conductivity2 Float uS/cm R
41091
41092 Conductivity2status Word R
41093 - 98 Conductivity2time Word R

C-11
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41105 Conductivity3 Float uS/cm R
41106
41107 Conductivity3status Word R
41108 - 13 Conductivity3time Word R
41120 Specific Conductance Float uS/cm R
41121
41122 Specific Conductan- Word R
cestatus
41123 - 28 Specific Conduc- Word R
tancetime
41135 Specific Float uS/cm R
41136 Conductance1
41137 Specific Word R
Conductance1status
41138 - 43 Specific Word R
Conductance1time
41150 Specific Float uS/cm R
41151 Conductance2
41152 Specific Word R
Conductance2status
41153 - 58 Specific Word R
Conductance2time
41165 Specific Float uS/cm R
41166 Conductance3
41167 Specific Word R
Conductance3status
41168 - 73 Specific Word R
Conductance3time
41180 Dissolved Solid Float mg/l R
41181
41182 Dissolved Solidstatus Word R
41183 - 88 Dissolved Solidtime Word R
41195 Dissolved Solid1 Float mg/l R
41196
41197 Dissolved Word R
Solid1status
41198 - 203 Dissolved Solid1time Word R
41210 Dissolved Solid2 Float mg/l R
41211
41212 Dissolved Word R
Solid2status
41213 - 18 Dissolved Solid2time Word R

C-12
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41225 Dissolved Solid3 Float mg/l R
41226
41227 Dissolved Word R
Solid3status
41228 - 33 Dissolved Solid3time Word R
41240 Salinity Float mg/l R
41241
41242 Salinitystatus Word R
41243 - 48 Salinitytime Word R
41255 Salinity1 Float mg/l R
41256
41257 Salinity1status Word R
41258 - 63 Salinity1time Word R
41270 Salinity2 Float mg/l R
41271
41272 Salinity2status Word R
41273 - 78 Salinity2time Word R
41285 Salinity3 Float mg/l R
41286
41287 Salinity3status Word R
41288 - 93 Salinity3time Word R
41300 Dissolved Oxygen Float mg/l R
41301
41302 Dissolved Oxygensta- Word R
tus
41303 - 08 Dissolved Oxygen- Word R
time
41315 Dissolved Oxygen1 Float mg/l R
41316
41317 Dissolved Word R
Oxygen1status
41318 - 23 Dissolved Word R
Oxygen1time
41330 Dissolved Oxygen2 Float mg/l R
401331
41332 Dissolved Word R
Oxygen2status
41333 - 38 Dissolved Word R
Oxygen2time
41345 Dissolved Oxygen3 Float mg/l R
41346

C-13
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41347 Dissolved Word R
Oxygen3status
41348 - 53 Dissolved Word R
Oxygen3time
41360 pH Float pH R
41361
41362 pHstatus Word R
41363 - 68 pHtime Word R
41375 pH1 Float pH R
41376
41377 pH1status Word R
41378 - 83 pH1time Word R
41390 pH2 Float pH R
41391
41392 pH2status Word R
41393 - 98 pH2time Word R
41405 pH3 Float pH R
41406
41407 pH3status Word R
41408 - 13 pH3time Word R
41420 ORP Float Volts R
41421
41422 ORPstatus Word R
41423 - 28 ORPtime Word R
41435 ORP1 Float Volts R
41436
41437 ORP1status Word R
41438 - 43 ORP1time Word R
41450 ORP2 Float Volts R
41451
41452 ORP2status Word R
41453 - 58 ORP2time Word R
41465 ORP3 Float Volts R
41466
41467 ORP3status Word R
41468 - 73 ORP3time Word R
41480 NH4 Nitrogen Float mg/l R
41481
41482 NH4 Nitrogenstatus Word R

C-14
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41483 - 88 NH4 Nitrogentime Word R
41495 NH4 Nitrogen1 Float mg/l R
41496
41497 NH4 Nitrogen1status Word R
41498 - 503 NH4 Nitrogen1time Word R
41510 NH4 Nitrogen2 Float mg/l R
41511
41512 NH4 Nitrogen2status Word R
41513 - 18 NH4 Nitrogen2time Word R
41525 NH4 Nitrogen3 Float mg/l R
41526
41527 NH4 Nitrogen3status Word R
41528 - 33 NH4 Nitrogen3time Word R
41540 NO3 Nitrogen Float mg/l R
41541
41542 NO3 Nitrogenstatus Word R
41543 - 48 NO3 Nitrogentime Word R
41555 NO3 Nitrogen1 Float mg/l R
41556
41557 NO3 Nitrogen1status Word R
41558 - 63 NO3 Nitrogen1time Word R
41570 NO3 Nitrogen2 Float mg/l R
41571
41572 NO3 Nitrogen2status Word R
41573 - 78 NO3 Nitrogen2time Word R
41585 NO3 Nitrogen3 Float mg/l R
41586
41587 NO3 Nitrogen3status Word R
41588 - 93 NO3 Nitrogen3time Word R
41600 Turbidity Float NTU R
41601
41602 Turbiditystatus Word R
41603 Turbiditytime Word R
41608
41615 Turbidity1 Float NTU R
41616
41617 Turbidity1status Word R
41618 - 23 Turbidity1time Word R

C-15
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41630 Turbidity2 Float NTU R
41631
41632 Turbidity2status Word R
41633 - 38 Turbidity2time Word R
41645 Turbidity3 Float NTU R
41646
41647 Turbidity3status Word R
41648 Turbidity3time Word R
41653
41660 Chloride Float mg/l R
41661
41662 Chloridestatus Word R
41663 - 68 Chloridetime Word R
41675 Chloride1 Float mg/l R
41676
41677 Chloride1status Word R
41678 - 83 Chloride1time Word R
41690 Chloride2 Float mg/l R
41691
41692 Chloride2status Word R
41693 - 98 Chloride2time Word R
41705 Chloride3 Float mg/l R
41706
41707 Chloride3status Word R
41708 - 13 Chloride3time Word R
41720 Resistivity Float Ohm-cm R
41721
41722 Resistivitystatus Word R
41723 Resistivitytime Word R
41728
41735 Resistivity1 Float Ohm-cm R
41736
41737 Resistivity1status Word R
41738 - 43 Resistivity1time Word R
41750 Resistivity2 Float Ohm-cm R
41751
41752 Resistivity2status Word R
41753 - 58 Resistivity2time Word R
41765 Resistivity3 Float Ohm-cm R
41766

C-16
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41767 Resistivity3status Word R
41768 - 73 Resistivity3time Word R
41780 Pressure Float mmHg R
41781
41782 Pressurestatus Word R
41783 - 88 Pressuretime Word R
41795 Pressure1 Float mmHg R
41796
41797 Pressure1status Word R
41798 - 803 Pressure1time Word R
41810 Pressure2 Float mmHg R
41811
41812 Pressure2status Word R
41813 - 18 Pressure2time Word R
41825 Pressure3 Float mmHg R
41826
41827 Pressure3status Word R
41828 - 33 Pressure3time Word R
41840 Reserved Float R
41841
41842 Reservedstatus Word R
41843 - 48 Reservedtime Word R
41855 Generic Float R
41856
41857 Genericstatus Word R
401858 - 63 Generictime Word R
401870 Generic1 Float R
401871
41872 Generic1status Word R
41873 - 78 Generic1time Word R
41885 Generic2 Float R
41886
41887 Generic2status Word R
41888 - 93 Generic2time Word R
41900 Generic3 Float R
41901
41902 Generic3status Word R
41903 - 08 Generic3time Word R

C-17
Signature® Flow Meter

Table C-1 Modbus Output Registers for Signature Flow Meter


Register Data Units of Read/
Name Description
Number Type Measure Write
41915 Generic4 Float R
41916
41917 Generic4status Word R
41918 - 23 Generic4time Word R
41930 Generic5 Float R
41931
41932 Generic5status Word R
41933 - 38 Generic5time Word R
41945 Generic6 Float R
41946
41947 Generic6status Word R
41948 - 53 Generic6time Word R
41960 Generic7 Float R
41961
41962 Generic7status Word R
41963 - 68 Generic7time Word R
41975 Generic8 Float R
41976
41977 Generic8status Word R
41978 - 83 Generic8time Word R
41990 Wireless Power Float R
41991
41992 Wireless Powerstatus Word R
41993 - 98 Wireless Powertime Word R

C-18
Signature® Flow Meter
Appendix D Material Safety Data Sheets

This appendix provides Material Safety Data Sheets for the des-
iccant used by the Signature Flow Meter.
Teledyne Isco cannot guarantee the accuracy of the data. Specific
questions regarding the use and handling of the products should
be directed to the manufacturer listed on the MSDS.
SORB-IT® is a registered trademark of N. T. Gates Company.

D-1
Signature® Flow Meter

D-2
Signature® Flow Meter

D-3
Signature® Flow Meter

D-4
Signature® Flow Meter

D-5
Signature® Flow Meter

D-6
Signature® Flow Meter

D-7
Signature® Flow Meter

D-8
Signature® Flow Meter

D-9
Signature® Flow Meter

D-10
Signature® Flow Meter
Index

Numerics GSM, 5-37


301 pH/temperature device, 5-16 installation, 5-37
306 sampler interface, 5-15 setup, 5-40
310 ultrasonic sensor, 5-13 character grid, 2-8
330 bubbler module, 5-14 clock, 2-19
installation, 6-19 compliance, 2-35
350 area velocity sensor, 5-14 components, 1-5
360 laser velocity sensor, 5-14 conditions, 2-24
configure measurements, 2-12
configure options, 2-19
A adjust, 2-24
AC wiring, 5-2 data storage/push, 2-26
accessories, B-1 equation/trigger setup, 2-24
adjust display, 1-7 flow settings, 2-22
adjust levels, 2-24, 3-18 measurement setup, 2-20
administration, 2-30 outputs, 2-27
display license information, 2-31 reports/history, 2-30
display signature information, 2-31 reset totalizers, 2-30
language options, 2-31 sampler, 2-27
restore to factory defaults, 2-32 site, 2-19
sensor diagnostics, 2-31 connecting TIENet devices, 3-7
set new passcode, 2-31 connecting via Flowlink, 2-2
update firmware, 2-31 connecting via USB, 2-2
alarm connections, 3-1, 3-3
local, 2-27 power, 3-11
SMS & server, 2-27 contact Teledyne Isco, 6-25
alarms, 2-27 control panel
alphanumerics, 2-8 display, 1-7
analog output, 2-27 keypad, 1-7, 2-1
programming, 5-19, 5-24, 5-27 LED indicator, 1-7
area velocity sensor, 5-14 cord-grip, 3-5

B D
battery backup, 5-4 data
browser window, 2-6 compliance, 1-3, 2-35
Bubble Line download, 2-34
Extensions, 3-17 dump, 2-33
Flume Fittings, 3-17 events, 1-3
High-Velocity Flow Streams, 3-16 export, 2-35
Open Channel Installation, 3-18 integrity, 1-3, 2-33, 2-34, 2-35
Stilling Wells, 3-17 push, 2-26
bubbler, 5-14, 6-19 storage, 2-26
default settings, 2-32
C default units, 2-19
case board, 3-1 description, 1-1
layout, 3-3 desiccant, 6-5
CDMA, 5-36 external, 5-12, 6-6
cellular modems, 5-36 internal, 6-5
CDMA, 5-36 device info, 2-5

Index-1
Signature® Flow Meter
Index

diagnostics, 2-31 introduction, 1-1


display, 1-7
display contrast, 1-7 K
display equipment info, 2-31 keypad, 1-6, 1-7, 2-1
display license info, 2-31
drop down menus, 2-9
dump data, 2-33 L
language, 2-31
laser velocity sensor, 5-14
E level adjustment, 2-7, 3-18
equations, 2-24
equipment options, 4-1, 5-1
ethernet modem, 5-31 M
configuration, 5-34 maintenance, 6-1
MAC address, 5-35 bubbler replacement, 6-19
event viewer, 2-34 desiccant, 6-5
export data, 2-35 firmware updates, 6-2
events, 1-3 Material Safety Data Sheets, D-1
expansion box, 5-16 measurement setup
flow input, 2-20
level input, 2-20
F volume input, 2-20
factory defaults, 2-32 measurements, 2-12
firmware, 2-31, 2-33 mechanical totalizer, 5-8
firmware updates, 6-2 menus, 2-9, 2-10
flash drive, 2-32, 6-2 Modbus
flow conversions, 2-22 configurations, C-2
flow measurement, 2-22 glossary, C-3
Flowlink, 2-34, 5-31 Modbus input setup, 2-14
browser, 2-6 Modbus output setup, 2-18
connecting, 2-2 modem
devices, 2-5 cellular, 5-36
event viewer, 2-34 ethernet, 5-31
program window, 2-6 modem setup, 2-18
site info, 2-4 mounting options, 3-12
MSDS, D-1
G
GSM, 5-37 O
off-screen content, 2-8
H optional equipment, 1-3, 4-1, 5-1, B-1
hardware scan, 2-12 AC power cord, 5-2
hardware setup, 2-11 battery backup, 5-4
Modbus input, 2-14 cellular modems, 5-36
Modbus output, 2-18 ethernet modem, 5-31
modem, 2-18 expansion box, 5-16
SDI-12, 2-14 external desiccator, 5-12
smart sensor, 2-12 Flowlink, 5-31
histograph, 2-7 mechanical totalizer, 5-8
history, 2-30 reference port tubing kit, 5-29
home display, 2-19 TIENet devices, 5-13
home screen, 2-7 order parts, A-1, B-1
outputs
I alarm, 2-27
Installation analog, 2-27
Bubble Line, 3-15
installation, 3-1, 4-1 P
mounting, 3-12 passcode, 2-31
outdoor, 3-13 pH/temperature device, 5-16
interior, 3-1 port holes, 3-1, 3-3

Index-2
Signature® Flow Meter
Index

fittings, 3-5 smart sensor setup, 2-12


plugs, 3-6 software, 2-2, 5-31
power, 3-11 specifications, 1-7
battery backup, 5-4 system reset, 6-24
power supply replacement, 5-2
program window, 2-6 T
programming, 2-8 Technical Service, 6-25
administration, 2-30 technical specifications, 1-7
alphanumerics, 2-8 TIENet devices, 5-13
analog output, 5-19, 5-24, 5-27 301 pH/temperature device, 5-16
cellular modem, 5-40 306 sampler interface, 2-27, 5-15
character grid, 2-8 310 ultrasonic level sensor, 5-13
configure, 2-19 330 bubbler module, 5-14, 6-19
general rules, 2-8 350 area velocity sensor, 5-14
hardware setup, 2-11 360 laser velocity sensor, 5-14
menus, 2-10 connecting, 3-7
navigation, 2-8 expansion box, 5-16
pull down menus, 2-9 totalizer reset, 2-30
scrolling, 2-8 totalizer, mechanical, 5-8
top menu, 2-10 triggers, 2-24
USB options, 2-32 typing characters, 2-8
pull down menus, 2-9
pushed data, 2-26
U
ultrasonic level sensor, 5-13
Q units, 2-19, 2-31
quick start, 1-2 update firmware, 2-31, 2-33
USB drivers, 2-2
R USB options, 2-32
reference port tubing kit, 5-29 dump data, 2-33
repairs, 6-25 none of the above, 2-34
replacement parts, A-1 retrieve text reports, 2-33
report view, 2-8 update firmware, 2-33
reports, 2-30, 2-33 user interface, 2-1
export, 2-35 browser, 2-2, 2-6
verification, 2-35 Flowlink, 2-2
reset, 6-24 keypad, 2-1
reset passcode, 2-31
V
S verification, 2-33, 2-35
safety, 1-v verification tool, 2-35
sampler interface, 2-27, 5-15
scan, 2-12
scrolling, 2-8
SDI-12, 2-14
security, 1-3
select operation, 2-10
sensor diagnostics, 2-31
sensor differences, 2-12
servicing, 6-1, 6-25
setup and programming, 2-1
shield, 3-13
shortcuts, 2-7
histograph, 2-7
level adjustment, 2-7
report view, 2-8
site information, 2-4
site name, 2-19
site setup, 2-19

Index-3
Signature® Flow Meter
Index

Index-4
Compliance Statements

ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞
Name and amount of Hazardous Substances or Elements in the product
㦘㹡㦘⹂䓸德㒥⏒侯
捷ↅ⚜䱿 Hazardous Substances or Elements
Component Name 杔 㻭 柘 ⏼ↆ杻 ⮩䅃勣啾 ⮩䅃ℛ勣啾
(Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE)
兎恾㨎
X O O O O O
Circuit Boards
㣍䯉
X O O O O O
Display
㘴兎
O O O O O X
Wiring
␔捷䟄冕
O O O O O X
Internal Cables
䦃㿐䟄㧉
X O O O O X
DC Motor
㘴⯃
O O X O O O
Connectors
䟄㻯
X X X O O O
e

Battery
bl
Ta

䟄䭐梏
X O O O O X
at

Solenoid valve
M
az
H

ℶ❐₼㦘㹡㦘⹂䓸德㒥⏒侯䤓⚜䱿♙⚺摞᧶Name and amount of Hazardous Substances or Elements in


the product
O: 嫷䯉年㦘㹡㦘⹂䓸德⦷年捷ↅ㓏㦘⧖德㧟㠨₼䤓⚺摞⧖⦷ST/ 㪖⑕屓⸩䤓棟摞尐㻑ⅴₚᇭ
O: Represent the concentration of the hazardous substance in this component’s any homogeneous pieces is
lower than the ST/ standard limitation.
X᧶嫷䯉年㦘㹡㦘⹂䓸德咂⺠⦷年捷ↅ䤓㩟₏⧖德㧟㠨₼䤓⚺摞怔⒉ST/ 㪖⑕屓⸩䤓棟摞尐㻑ᇭ
(←₩♾⦷㷳⮓᧨㫈㗽⸭棔㍔⑄⺈ₙ嫷₼㓢“X” 䤓㔏㦾☮⥯扪嫛扪₏㷴広㢝ᇭ)
X: Represent the concentration of the hazardous substance in this component’s at least one homogeneous
piece is higher than the ST/ standard limitation.
(Manufacturer may give technical reasons to the “X”marks)
䘾≬∎䞷㦮䟀兞洛䫽⸩ᇭ
The Environmentally Friendly Use Period (EFUP) was determined through experience.
䞮ℶ㡴㦮嬺冥䪐⦷侊⒦⚆䪐₼ᇭⓜₘ⇜㟿ⷦ⃉䞮ℶ㄃(207 ⅲ嫷 2007 ㄃) ᇭ椞⚝䤓₏₹ⷦ㹜ⅲ嫷㦗↌᧶
A ⃉₏㦗᧨B ⃉ℛ㦗᧨䷘䷘ᇭ
The date of Manufacture is in code within the serial number. The first three numbers are the year of
manufacture (207 is year 2007) followed by a letter for the month. "A" is January, "B" is February and so on.

Hazmat Table Signature Meter, Sensors and Accessories 60-4302-091 Rev.A


Warranty

DECLARATION OF CONFORMITY
Application of Council Directive: 2004/108/EC -The EMC Directive
2012/19/EC– The WEEE Directive
2006/95/EC – The Low Voltage Directive
Manufacturer's Name: Teledyne Isco
Manufacturer's Address: 4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Phone: +1 (402) 464-0231
FAX: +1 (402) 465-3799
Equipment Type/Environment: Laboratory Equipment for Light Industrial/Commercial Environments
Trade Name/Model No: Signature Flow Meter (AC or DC power) with 350, 360, 301 sensors, 306 Sampler
Interface, RS485 and Ethernet Modem options
Year of Issue: 2014
Standards to which Conformity is Declared: EN 61326:2006 & IEC 61326-1:2012 EMC Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use
EN 61010-1 2nd edition Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN60529 Special Protection offered by the Signature’s Enclosure: IP-66
Standard Description Severity Applied Performance Criteria
EN61000-4-2:2008 Electrostatic Discharge (AC or DC Level 2 - 4kV contact discharge A
power) Level 3 - 8kV air discharge
EN61000-4-3:2006 Radiated RF Immunity (AC or DC 10 V/m:80MHz-1GHz, 3 V/m:1.4GHz-2GHz,
/A1:2007 /A2:2010 power) 1 V/m: 2GHz-2.7GHz, 80%, AM, 1KHz rate,1 A
sec dwell
EN61000-4-4:2004 Electrical Fast Transient (EFT) on ± 2 kV on AC lines, ±1 kV on (AC & DC) Signal A
/A1:2010 Mains and I/O Cables
EN61000-4-5:2005 Surge on AC Lines ±1 kV L-L /±2 kV L-PE A

EN61000-4-6:2008 Conducted RF Immunity on Mains and 3 Vrms on AC Mains & (AC & DC) Signal A
I/O Cables
EN61000-4-11:2004 Voltage Dips 0.5,1,25 Cycle/0,0,70% A

CISPR11/ RF Emissions Radiated, below 1GHz Group 1, Class A Industrial, Scientific, and PASS
EN 55011:2009 and Conducted, AC Mains Medical Equipment
/A1:2010
EN61000-3-2:2006 AC Harmonics, Flicker PASS
/A1:2009 /A2:2009
IEC 61000-3-2:2014
EN61000-3-3:2008

We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of August 18. 2014.

USA Representative

_______________________________
Vikas V. Padhye, Ph. D.
Executive Vice President of Sales and Marketing and
Product Line General Management
4700 Superior Street
Lincoln, Nebraska 68504
Phone: 402-464-0231
Fax: 402-464-0318

60-4302-021
Rev. C
Teledyne Isco One Year Limited Factory Service Warranty*
This warranty exclusively covers Teledyne Isco The warranty for any instrument is the one in effect on date
instruments, providing a one-year limited warranty of shipment. The warranty period begins on the shipping
covering parts and labor. date, unless Teledyne Isco agrees in writing to a different
Any instrument that fails during the warranty period due to date.
faulty parts or workmanship will be repaired at the factory Excluded from this warranty are normal wear; expendable
at no charge to the customer. Teledyne Isco’s exclusive items such as charts, ribbon, lamps, tubing, and glassware;
liability is limited to repair or replacement of defective fittings and wetted parts of valves; and damage due to
instruments. Teledyne Isco is not liable for consequential corrosion, misuse, accident, or lack of proper maintenance.
damages. This warranty does not cover products not sold under the
Teledyne Isco will pay surface transportation charges both Teledyne Isco trademark or for which any other warranty is
ways within the 48 contiguous United States if the specifically stated.
instrument proves to be defective within 30 days of No item may be returned for warranty service without a
shipment. Throughout the remainder of the warranty period, return authorization number issued by Teledyne Isco.
the customer will pay to return the instrument to Teledyne This warranty is expressly in lieu of all other warranties
Isco, and Teledyne Isco will pay surface transportation to and obligations and Teledyne Isco specifically disclaims
return the repaired instrument to the customer. Teledyne any warranty of merchantability or fitness for a
Isco will not pay air freight or customer’s packing and particular purpose.
crating charges. This warranty does not cover loss, damage, The warrantor is Teledyne Isco, 4700 Superior, Lincoln, NE
or defects resulting from transportation between the 68504, U.S.A.
customer’s facility and the repair facility.

* This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer.
Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their
Teledyne Isco dealer for warranty service.

Before returning any instrument for repair, please call, fax, or e-mail the Teledyne Isco Service
Department for instructions. Many problems can often be diagnosed and corrected over the
phone, or by e-mail, without returning the instrument to the factory.
Instruments needing factory repair should be packed carefully, and shipped to the attention of
the service department. Small, non-fragile items can be sent by insured parcel post. PLEASE
BE SURE TO ENCLOSE A NOTE EXPLAINING THE PROBLEM.

Shipping Address: Teledyne Isco - Attention Repair Service


4700 Superior Street
Lincoln, NE 68504 USA
Mailing Address: Teledyne Isco
PO Box 82531
Lincoln, NE 68501 USA
Phone: Repair service: (800) 775-2965 (lab instruments)
(866) 298-6174 (samplers & flow meters)
Sales & General Information: (800) 228-4373 (USA & Canada)
Fax: (402) 465-3001
Email: [email protected]

February 28, 2012 P/N 60-1002-040 Rev G