EM00000364 Percussionaire Corporation - An IPV Operational & Service Manual Instructions For Use IPV
EM00000364 Percussionaire Corporation - An IPV Operational & Service Manual Instructions For Use IPV
EM00000364 Percussionaire Corporation - An IPV Operational & Service Manual Instructions For Use IPV
TABLE OF CONTENTS
PAGE
Indications for Use .......................................................................................................... 1
Contraindications for Use ............................................................................................... 1
Physiological Effects of IPV® ......................................................................................... 1
Clinical Limitations/Restrictions ……………………………………………………………. 1
Introduction ...................................................................................................................... 3
Key IPV® Flow Ventilators® ........................................................................................... 3
The IPV®-1C ..................................................................................................................... 3
The IPV®-2C ..................................................................................................................... 4
Clinician’s General Lung Recruitment Protocol ........................................................... 5
The Phasitron® ................................................................................................................ 7
Phasitron® Kits ………………………………………………………………………………… 8
The IPV® Cone Interface Adapter................................................................................... 9
Cleaning and Decontamination Procedures ................................................................. 10
Functional Evaluation ..................................................................................................... 11
Troubleshooting .............................................................................................................. 11
Appendices
Service and Repair ................................................................................................ 12
Storage .................................................................................................................. 13
Disposal of Equipment ......................................................................................... 13
Shipping Information ............................................................................................ 13
Glossary of Symbols ............................................................................................ 14
Glossary of Terms ................................................................................................ 15
Contact Information .............................................................................................. 13 & 20
3
INDICATIONS FOR USE
Use of Percussionaire® IPV® is indicated for the mobilization and raising of endobronchial secretions,
bronchodilation, reducing mucosal edema, and the resolution of diffuse patchy atelectasis in all patient
populations.
CONTRAINDICATIONS
The IPV® Flow Ventilators® have inherent risks similar to all positive pressure breathing devices. These
risks include but are not limited to under/over ventilation, under/over humidification, pneumothorax, and
pneumediastinum, PIE, pneumoperitoneum, and hemoptysis.
CLINICAL LIMITATIONS/RESTRICTIONS
Use of IPV Percussionators is limited to individuals who have received proper training in their use.
The use of standard therapeutic IPV® Percussionators will be restricted to a mouthpiece only. Under no
circumstances will a standard IPV® breathing head with a gated aerosol generator be used with other than
a mouthpiece (with lip seal only).
WARNING If an IPV® Flow Ventilator® is used on a patient with an indwelling airway (i.e.
endotracheal of tracheostomy tube), a clinician must be present so that a one to one
0 . relationship exists. These devices enhance secretion clearance. Patients must be assessed
pre and post treatment for a reduced vital capacity/FRC or the need for assistance in
clearing airway secretions. A patient cannot breathe through an obstructed airway! If an
artificial airway is in place, a special gated aerosol generator must be used to relieve any
transient/unwanted pressure rises.
WARNING - If the pulse frequency ratios are not properly calibrated, the clinical efficacy of the
device may be in jeopardy. As with all ventilators, calibration and other adjustments should
only be performed by individuals with appropriate factory training.
INTRODUCTION
Created by Dr. Forrest M. Bird in the mid 1980’s, Intrapulmonary Percussive Ventilation
(IPV®) was primarily conceived to recruit and therapeutically maintain the insidious loss
of bronchiolar blood supply in patients with progressive COPD.
MANUAL INSPIRATION-
Manual
Delivers a regulated source of Inspiration
Percussion
gas through the orifice of the
Phasitron Venturi. The longer
As with all IPV® Flow Ventilators®, the IPV®-1C is
the button is held depressed,
capable of percussively mobilizing bronchiolar
the greater the potential for
airways and their associated alveoli in all
tidal volume delivery.
cardiopulmonary patient populations with mixed
bronchiolar airway obstructions, from minimal toward
MASTER SWITCH- ON/OFF
total airway obstruction as well as associated diffuse
On position - patient receives
unobstructed PREFERENTIAL AIRWAYS without
percussive treatment.
barotraumatic hyperinflation of the alveoli served by
the unobstructed preferential bronchiolar airways.
This is accomplished in neonates through pediatrics to large adults WHILE MAINTAINING
A LUNG PROTECTIVE STRATEGY.
THE IPV®-2C IS FOR MORE FLEXIBLE LUNG CONDITIONING THERAPY
6
Programmable DEMAND
CONTINUOUS POSITIVE AIRWAY
D-CPAP Insp. Time PRESSURE (D-CPAP) – for reducing
the work of breathing.
Due to the therapeutic nature of these devices, they do not have alarms. Consequently,
the patient MUST be under continuous observation by a clinician.
Operating Temperature:
The IPV®-1C and IPV®-2C should not be used in temperatures lower than 10°C (50°F) and
higher than 40°C (104°F).
The IPV®-1C and IPV®-2C both run off any respiratory 50 - 80 psig (345 – 551 kPa) gas
supply.
1) IPV®-1C Connects to any respiratory 50 - 80 psig (345 – 551 kPa) gas power
source. IPV®-1C: 50 psi. (345 kPa) gas source. Master Switch - OFF.
A) Patient should be osculated for breath sounds, heart and respiratory rate or
follow institution guidelines.
4) Put prescribed medications into nebulizer and add diluent as directed by physician
to a total of 15ml. If no medications are prescribed, use normal saline or sterile
water as directed by physician. Assemble Phasitron® to Nebulizer.
6) With IPV-1C Gas Supply ON, Rotate OPERATIONAL PRESSURE control regulator
knob for an operating pressure of 30 to 35 psig (206 – 241 kPa).
7) If needed, allow patient to observe and feel pulses on their hand before either
connecting to the airway or breathing through the mouthpiece.
8
8) When using a mouthpiece, the patient should be instructed to inhale and exhale
through the pulses. Most patients will initially allow percussive bursts of air to leak
through their nose at the expense of an observable chest movement (wiggle). It
may be useful to ask if they have ever breathed through a snorkel, to teach the
patient how to breathe in and out of the mouthpiece.
Start to notice the chest movement (wiggle) as the patient exhales through the
mouthpiece. Advise the patient to relax taking normal (spontaneous) breaths
through the pulses whenever they desire. The objective is to perform a protracted
period of IPV breathing for a 15 to 20 minute treatment. Cheek puckering fatigue
may be an early consideration, however, this is soon eliminated.
9. When a patient has an artificial airway, the process is similar. The patient must
be observed carefully for signs of distress. While cheek fatigue will be less of a
consideration, pauses or breaks may still be necessary for the patient.
Note: The individual clinician teaching the patient how to use Therapeutic IPV should
have given themselves a treatment so that they are familiar with the sensations.
9) The patient should be instructed to keep lips and cheeks splinted to avoid nasal air
venting. As the patient learns to prevent air from leaking out of the lip seal around
the mouthpiece, the PERCUSSION control knob Arrow can be gradually rotated
(clockwise) toward the 12:00 index.
10) If the patient cannot understand instructions the therapist may have to use an
appropriate mechanical airway to administer therapy such as a mask.
11) After the ability to prevent the leaking of percussive air deliveries from the nose
and around the lips is learned, the entire Percussion frequency band should be
scanned by briefly rotating the PERCUSSION control knob Arrow from EASY to
9
HARD back and forth (a number of times) then back to the 12:00 Arrow position to
raise secretions from the Bronchial Airways, paying attention to patient comfort.
Tissue availability is advised.
As the learning period progresses the selected Source Pressure may be increased
for effective Endobronchial Percussion by assessing “chest percussion (shaking).
Nominal Operating Pressure is 40 psig (275 kPa).
13) When treatment is complete unit should be turned OFF. Phasitron® should be
removed, and then dismantled, washed and dried for next treatment on the SAME
patient. Percussionaire recommends cleaning per your institution’s approved
practice regarding respiratory care tubing sets and connections.
Notes:
A50095-D A50474-2D
1. Obtain specific IPV interfacing information for a specific “host ventilator” from
Percussionaire or appropriate authority.
2. The Percussionaire Cone Interface Adapter replaces the “Y” in the CMV device
breathing circuit (heated wire or otherwise). The “Y” is removed from the ventilator
circuit and replaced with the Cone Interface Adapter. 15mm ET tube adapter is
attached to 15mm ID connection on cone.
3. Select and set up an appropriate IPV Flow Ventilator and Cone interface adapter
with the Phasitron inserted into the Cone Phasitron port.
c. Rotate the PERCUSSION control knob Arrow under the 12:00 index.
e. Wait for about fifteen minutes and evaluate saturation, PaO2 or PaCO2.
Appropriate pulse frequency selections can be evaluated over time.
All new Percussionaire® products are packaged clean. They should not be considered
sterile or decontaminated. Prior to use, and between uses on the same patient, it is
recommended that Phasitron components be disassembled then cleansed according to
local, site-specific, cleaning and infection control procedures.
The control unit may be cleaned with medical wipes approved by the facility for use on
ventilators. If using a sprayed agent, this should be sprayed onto a cloth or wipe, and not
directly onto the Flow Ventilator®.
Percussionaire® does not deliver sterile devices, which are appropriately labeled per
FDA.
WARNING
If the pulse frequency ratios are not properly calibrated, the clinical efficacy of the device
may be in jeopardy. As with all ventilators, calibration and other adjustments should only
be performed by individuals with appropriate factory training.
A mandated remanufacture (overhaul) (OH) is required every three (3) years after
the device is initiated into service or not later than four (4) years after first date of
purchase. A factory remanufacture consists of replacing all elastomeric seals,
sleeves, and diaphragms, with inspection of all components. The device is
factory calibrated and receives a functional evaluation, conformance certification,
and a one-year warranty on all parts installed during overhaul. If replacement
parts other than those specified for overhaul or preventive maintenance are
required for repair, the cost of the parts will be quoted to the customer in addition
to the cost of the Preventive Maintenance (PM) or Overhaul (OH). Cleaning time
allowed for OH or PM is fifteen (15) minutes, any extra cleaning time will be
charged at current hourly rate.
A device which has not received a mandated overhaul for a period of 10 years,
whether in use during that period or not, will be considered to be beyond
economic repair. If appropriate mandated preventive maintenance and overhauls
are conducted, a device may continue to be used. If, due to damage, lack of
mandated overhauls, voided warranty, or other misuse, a device is considered by
the Repair Department to be beyond economic repair, a letter will be sent
advising the owner of the device of the findings, and requesting disposition
instructions. Under no circumstances will a device considered by
Percussionaire® Corporation Repair Department to be beyond economic repair
be returned to active service.
Any device showing damage may be subject to additional charge if the repairs
require parts not normally replaced during remanufacturing.
STORAGE
The Percussionaire® units should be stored in a clean environment and covered when
not in use. Temperature should be maintained between 10°C to 25°C (50°F to 77°F).
Humidity range is 0 - 95% non-condensing.
DISPOSAL OF EQUIPMENT
At the end of useful life of a unit, disposal should be in accordance with local, state,
federal and international laws. The unit may also be packaged according to instructions
found within this manual and shipped to authorized maintenance centers below for
disposal.
SHIPPING INFORMATION
WEBSITE ADDRESS
www.percussionaire.com
16
GLOSSARY OF SYMBOLS
xxxx-xx-xx
17
GLOSSARY OF TERMS
CONVECTIVE TIDAL VOLUME DELIVERIES – The delivery into the pulmonary structure of
programmed volumes of a respiratory gas (measured in cubic centimeters) that exceed
the anatomical dead space, favoring the wash out of carbon dioxide.
PERCUSSION/BASELINE RATIO (I:E RATIO)- Expressing the ratio of the percussive sub
tidal (inspiratory) interval in relation to the time at baseline (expiratory) interval. A
method of describing the VDR® I/E ratio.
PHASING RATE- Describing the number of cyclic inspiratory/ expiratory intervals per
minute counted as returns to a programmed expiratory baseline.
PROXIMAL AIRWAY WAVE FORM ANALYSIS- Whereby proximal airway pressures are
directed against a transducer with sufficient capacities to relate the rapid (millisecond)
pressure changes associated with VDR®/IPV® scheduling. Therefore, a means for
presenting proximal airway pressure changes on a cathode ray tube (CRT) are enhanced.
Desirable pressure scales and sweep speeds can be selected, allowing the clinician to
program and interpret proximal airway pressure potentials as they may affect
physiological parameters. Additionally, proximal airway pressure tracings can be
documented on strip chart recorders.
PULSE FREQUENCY- Describing the number of pulsatile sub tidal volume deliveries per
minute.
VDR “I/E” RATIO- Describing the ratio between the length of time (in seconds) that sub
tidal volumes are intrapulmonarily delivered (oscillatory inspiratory interval) to the length
of time a scheduled interruption at baseline (expiratory interval) is scheduled. Oscillatory
Inspiratory interval/expiratory interval.
LIMITED WARRANTY
Model No:________________________________________
Serial No:___________________________________________
The products of Percussionaire® Corporation are warranted, for a period of one year from date of purchase, to be free from
defects in materials and workmanship and to meet the published specifications. THE FOREGOING IS IN LIEU OF ANY OTHER
WARRANTY, EXPRESS OR IMPLIED, INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF MERCHANTABILITY,
except as to title, and can be amended only in writing by a duly authorized representative of Percussionaire® Corporation. The
liability under this warranty is limited solely to replacing, repairing or issuing credit, at the discretion of Percussionaire®
Corporation, for the parts that become defective or fail to meet published specifications during the warranty period; provided
that, Percussionaire® Corporation will not be liable under this warranty unless (I) Percussionaire® Corporation is promptly notified
in writing by Buyer upon discovery of defects or failure to meet specifications; (II) the defective unit or part is returned to
Percussionaire® Corporation, transportation charges prepaid by Buyer; (III) the defective unit or part is received by
Percussionaire® Corporation for adjustment no later than four weeks following the last day of the warranty period; and (IV)
Percussionaire® Corporation's examination of such unit or part shall disclose, to its satisfaction, that such defects or failures
have not been caused by misuse, neglect, improper installation, unauthorized repair or alteration or accident. Any authorization of
Percussionaire® Corporation for repair or alteration by the Buyer must be in writing to prevent voiding warranty. In no event shall
Percussionaire® Corporation be liable to Buyer for loss of profits, loss of use, consequential damages or damages of any kind
based upon a claim for breach of warranty, other than the purchase price of any defective product covered hereunder.
Percussionaire® Corporation warranties as hereinabove set forth shall not be enlarged, diminished or affected by, and no
obligation or liability shall arise or grow out of the rendering of technical advice or service by Percussionaire® Corporation or its
agents in connection with Buyer's order or the products furnished hereunder.
PLEASE NOTE: In the event operation of a Percussionaire® product is in any way adversely affected by the use of
components other than those designed, manufactured, or approved by Percussionaire® Corporation, Percussionaire®
Corporation shall not be liable under this warranty with respect to such product.
Percussionaire® Corporation recommends an annual preventive maintenance (PM) for each device. A mandated
remanufacture/overhaul (OH) is required every three (3) years after the device is initiated into service or not later than four (4)
years after the first date of purchase.
24
Percussionaire® Corporation
130 McGhee Rd., Ste. 109
Sandpoint, ID 83864 USA
MDSS GmbH C
Schiffgraben 41 0123
Telephone: 208.263.2549 30175 Hannover
Fax: 208.263.0577 Germany
Percussionaire® Corp. 2009
www.percussionaire.com
US and Foreign Patents Issued and Pending
Document ID: P123-1A REV M
ADVISORY NOTICE #2021-09
Mandatory Overhaul Discontinuation
Device: IPV®-1C, IPV®-2C
June 2, 2021
Dear Customer,
This Advisory Notice is informing you that the three-year mandatory overhaul
requirement for the IPV®-1C and IPV®-2C has been discontinued.
NOTE:
• The IPV®-1C and IPV®-2C can be serviced as needed.
• Pre-use check is recommended before each use.
• Preventive maintenance is still recommended annually and now can be
performed by any biomedical technician.
Best
Regards,
Percussionaire® Corporation