OMMXTS-W00064 - rev 4
AMOT does not accept warranty and liability claims neither upon this publication
nor in case of improper treatment of the described products.
The document may contain technical inaccuracies and typographical errors. The
content will be revised on a regular basis. These changes will be implemented in
later versions. The described products can be improved and changed at any time
without prior notice.
© Copyright
All rights reserved.
USA
FCC notice:
This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential instal-
lation. This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the installation manual, may
cause harmful interference to radio communications. However, there is no gua-
rantee that interference will not occur in a particular installation. If this equip-
ment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the following measures:
Caution:
Any changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate this device.
CANADIAN
ICES-003 notification:
OMMXTS-W00064 - rev 4 2
TABLE OF CONTENTS
HARDWARE
1. GENERAL 4
1.1 Symbol assertion 4
1.2 Safety instructions 4
1.3 Environmental information 4
2. PRODUCT DESCRIPTION 5
3. INSTALLATION 5
3.1 Installation of the housing 5
3.2 Installation of the probe 5
3.2.1 General Safety instructions for installation 5
3.2.2 Installation of the probe directly in the process 6
3.2.3 Installation of the probe by utilizing the ball valve set 6
4. ELECTRICAL CONNECTIONS 7
4.1 Connection diagram 7
5. MAINTENANCE 8
5.1 Sensor cleaning 8
5.2 Self-diagnostics and error messages 8
7. TECHNICAL DATA 9
OMMXTS-W00064 - rev 4 3
1. GENERAL
The manual is a part of the scope of supply and serves to ensure proper handling and optimum
functioning of the instrument. For this reason, the manual must be read before start-up.
This manual is for all personnel who require knowledge concerning transport, setup, operati-
on, maintenance and repair.
The manual must not be used for the purpose of competition without written consent and
must also not be forwarded to third parties. Copies for personal use are permitted.
All information, technical data and illustrations contained in these instructions are based on
information available at the time of publication.
Safety instructions for use of the alarm module with voltages >50V
• To insulate the optional alarm module from the low-voltage side of the transmitter, the
partition provided for this purpose must be fitted in the lower section.
• During operation of the instrument the modular housing must be completely closed.
• The protection class of an opened housing corresponds to IP00 and direct contact with
components carrying dangerous voltages is therefore possible. In general, work on live
components should be avoided and when absolutely necessary, should be performed
by qualified personnel only.
Safety instructions for use of the integrated power supply (option V01)
• During operation of the instrument the modular housing must be completely closed.
• The protection class of an opened housing corresponds to IP00. In general, work on
live components should be avoided and when absolutely necessary, should be
performed by qualified personnel only.
The equipment is developed with due consideration to all resultant environmental issues.
When you dispose of the equipment you should avoid environmental pollution. For disposal of
the transmitter the individual components must be sorted with care. The housing consists of
recyclable polycarbonate or metal (aluminium, Al Si 9 Cu 3). The electronics must be collec-
ted as electronic scrap and disposed of according to the regulations in force.
OMMXTS-W00064 - rev 4 4
2. PRODUCT DESCRIPTION
Transmitter Series S4600-WIO-001 is specially designed for the measurement of water con-
tent in oil. The measured value is sent to the SPU over an RS485 link.
The modular housing enables an user-friendly operation and a quick replacement of the sen-
sor unit for service purposes. The construction of the transmitter makes field and local loop
calibration an easy task.
3. INSTALLATION
The necessary dimensions for the mounting holes can be found in the drawings below.
metal housing:
)
(6”
150
(2.2”)
(3.5”)
57
90
fig 1
3”)
(5.
135
Select a location with environmental conditions that permit an optimal measurement of the
process. The measuring medium (e.g.: oil) shall be as clean as possible i.e. without conta-
mination.
fig 2
OMMXTS-W00064 - rev 4 5
3.2.2 Installation of the probe directly in the process
For direct probe installation, a stop valve should be provided on both sides of the probe
insert. This allows the sensor probe to be removed for maintenance and calibration wit-
hout any problems.
If the sensor probe is installed in a pressure chamber, make sure that the pressure in
the chamber and the ambient pressure are in equilibrium before you remove the probe.
The ball valve set allows for the removal of the probe without interrupting the process.
Install the balve valve perpendicular to the direction of flow. The pressure of the process
must be less than 10 bar (145psi). The two metal sealing rings (see fig 4) should be
replaced every time prior to re-installing the probe.
• 4th step:
To ensure a secure installation of the probe, the lock nut
inside diameter
must be tightened to a defined torque of 30 Nm. If no ≥ 13 mm (0.55”)
torque-spanner is available tighten the lock nut by hand
as far as possible. Continue to turn with an open-ended
spanner ~50°.
fig 4
OMMXTS-W00064 - rev 4 6
A too low torque results in a smaller tension force (fixing force) on the clamping sleeve. There is the risk
of injury due to sudden expulsion of the sensing probe. A too high torque can lead to permanent deforma-
tion of the clamping sleeve and the sensing probe. This can make the removal and re-installation difficult
or impossible.
• 1st step:
Firmly hold sensing probe. (Attention: do not bend con-
nection cable)
clamping
• 2nd step: sleeve
Slowly loosen the lock nut with a spanner (spanner
width 24) until the expulsion force acts on the probe. lock nut
sealing
In the installed state, never completely remove the element 2
lock nut, only unscrew it as much as necessary!
sealing
adapter
• 3rd step: element 1
body
After the sensor head has been pulled out of the pro-
cess up to the stop, close the ball valve.
• 4th step:
Probe can now be completely removed.
4. ELECTRICAL CONNECTIONS
4.1 Connection diagram
supply
fig 6
OMMXTS-W00064 - rev 4 7
5. MAINTENANCE
Cleaning of the measuring head is recommended before emerging in other oil and before a
calibration.
• Red LED
constantly lit ⇒ Humidity sensor element damaged
flashing ⇒ Humidity sensor element accruing moisture (condensation!)
• Transmitter failure
no supply voltage
⇒ Check wiring and supply voltage
⇒ only green LED is illuminated continuously ⇒ Electronics defect
⇒ contact AMOT
OMMXTS-W00064 - rev 4 8
7. TECHNICAL DATA
Measuring values:
Water activity:
Temperature dependence
of electronics: typ. ± 0.0001 [1/°C] (typ. ± 5.6 * 10-5 [1/°F])
Temperature dependence
of sensing probe: typ. ± (0.00002 + 0.0002 x aw) x ∆T [°C] ∆T = T - 20°C
Temperature:
Accuracy:
∆°C
°C
Temperature dependence
of electronics: typ. ± 0.005°C/°C
Adjustable measurement
range:
General:
Supply voltage: 8 ..35V DC
Pressure range sensing probe: 0.01 ...10 bar 0.15 ... 145 psi
Sensor fluid connection: 1/2" NPT male standard (ISO 228-1/G 3/4" adaptor also provided)
OMMXTS-W00064 - rev 4 9
Europe, Middle East and Africa Americas
AMOT Russia
#34 Shabolovka Street
Building 2
Moscow 115419
Russia
Tel +7 495 617 12 93
Fax +7 495 913 97 65
Email [email protected]
AMOT Shanghai
Rm 308, Building No A8
Jiahua Business Center