USG-400
ISSUE 1
2011/11
USG-400
CONTENTS
Chapter 1 : Product Outline................................................................... 1-1
1 Outline ................................................................................................................................. 1-2
2 Features............................................................................................................................... 1-2
3 Limitations........................................................................................................................... 1-3
4 System chart ....................................................................................................................... 1-4
ISSUE1 1 Contents
USG-400
Contents 2 ISSUE1
USG-400
INTRODUCTION
Introduction
Repair and maintenance of this product requires highly specialized knowledge and techniques. We
recommend that you contact an Olympus service center in your area if a problem develops with the
product.
If repairs or modifications are made by personnel not authorized by Olympus, the warranty is void, and
Olympus shall not be liable for damage that occurs to or as a result of use of the modified product.
Applicable Unit
USG-400
Copyright
Trademarks
INTRODUCTION 1 ISSUE1
USG-400
1 Outline
(1) Purpose of Use
The THUNDERBEAT instruments are intended for use of coagulation/cutting of tissue and sealing/cutting of
vessels in combination with the ultrasonic generator, the electrosurgical generator and the THUNDERBEAT
transducer that are specified by Olympus in general surgery (including endscopic and microscopic surgery)
2 Features
(1) This product, which has two handpiece connectors, enables dissection and coaguration of body tissues by
plugging the handpieces int their corresponding handpiece connectors.
THUNDERBEAT Connector (Handpiece Connector 1) : SEAL & CUT output / SEAL output
SONICBEAT Connector (Handpiece Connector 2) : Ultrasonic output
(2) PLL resonance tracking control and constant current control enable stable ultrasonicvibration.
(3) Various setting are shown on the touch-panel LCD. They can be changed on the display.
(4) All the devices have their own hand switches that enable ON/OFF control of output.
(5) All the devices have their own foot switches that enable ON/OFF control of output.
(6) This product has the interface for output interlock and the output can be controlled.
Combination of ESG-400 and MAJ-1871 enable use of combine output.
Combination of UHI-2 or UHI-3 and MAJ-1871 or MAJ-1872 and MAJ-1873 enable use of automatic
fume control function.
Combination of UHI-4 and MAJ-1871 or MAJ-1872 enable use of automatic fume contril function.
(7) This product has the “setting selection” function that allows the user to arrange , and various settings
saved in advance can be called us easily.
(8) Its interface for PC communication enable use of the followingfunctions.(User unidentified)
Software Upgrading
Download and Upload of log data.
(9) The language shown on LCD can be selected from eight languages.
(10) As a result of introduction of the corporate product design, its desigh is consistent with the desighs of
products in hte same generation.
(11) This product can be connected to ENDOALPHA-3 system.
3 Limitations
(1) Use this product under the supervision of a doctor at a medical facility.
(2) Do not use this product in combination with the products other than those designated by Olympus.
(3) This product should be used, transported or stored in the following environment.
1)Usage environment
Ambient temperature 10 - 40°C
Relative humidity 30 - 85%
Atmospheric pressure 70 - 106kPa
Altitude 3000m or less
(Do not use this product in the combustible atmosphere)
2)Transportation/storage environment
Ambient temperature -25 - 70°C
Relative humidity 10 - 90%
Atmospheric pressure 70 - 106kPa
(4) Power source
1)Voltage, current
J type 100V, 3A
A type 120V, 3A
E type 220 - 240V, 1.5A
2)Frequency 50/60Hz
3)Voltage variation Within ±10%
4)Frequency variation Within ±1Hz
4 System chart
Chapter 2 : Specifications
1 Specifications ....................................................................................2-2
1 Specifications
Group 1,Class B
WEEE Directive
In accordance with European Directive 2002/96/EC on
Waste Electrical and Electronic Equipment, this symbol
indicates that the product must not be disposed of as
unsorted municipal waste, but should be collected
separately.
Prepare the ultrasonic generator and ancillary equipment for the intended purpose by referring to “System chart”
in the Product Outline. Install and connect the ultrasonic generator and ancillary equipment as described in the
following sections and also by referring to the instruction manuals for the ancillary equipment.
1 Installation workflow
The following is the equipment installation workflow. For detailson each step, read the corresponding description.
NOTE
Keep the instruction manual in an easily accessible place near
the ultrasonic generator.
CAUTION
Avoid applying excessive force on the communication cable.
Otherwise, wire disconnection or other failure may result.
The THUNDERBEAT performs cutting and coagulation using high frequency (RF
bipolar) and ultrasonic energy. When it is used, both the compatible
electrosurgical and ultrasonic generators are required.
To connect the two generators, align the docking connector at the bottom of the
compatible electrosurgical generator into the docking plug at the top of the
ultrasonic generator.
This section describes how to connect and disconnect the docking connector
and docking plug.
Figure 3.1
Connection
The docking fixture includes the tools for use in connection, including
2 fixture plates, 14 screws (12 pcs for the fixture plates and 2 pcs for the
docking plug cover) and a hex wrench.
Behind the docking plug cover is the docking plug for connection of the
compatible electrosurgical generator.
Figure 3.2
Figure 3.3
Figure 3.4
6. Connectiong
Align the rear panel of the compatible electrosurgical generator with the
fixture plates and then lower the front edge of the compatible
electrosurgicalgenerator slowly (see Figures 3.5 and 3.6).
Figure 3.5
Figure 3.6
Figure 3.7
Figure 3.8
Figure 3.9
Disconnection
Figure 3.10
Figure 3.11
CAUTION
Before connecting the power supply, make sure that it
conforms to the power specifications described on the rating
plate of the ultrasonic generator.
Figure 3.12
CAUTION
When using the automatic mist & smoke evacuation
system/function, also refer to the instruction manual for the
compatible high flow insufflation unit.
The 2-way cable for aeration and extension cable for aeration
are not allowed to use for the ultrasonic generator.
1. Preparation
Prepare the adapter for UHI-2/3 and the communication cable 10 m.
Figure 3.13
ISSUE1 3-13 Installation and Connection
USG-400
CAUTION
To prevent malfunction, connect either the compatible
electrosurgical or ultrasonic generator to the compatible high
flow insufflator. Do not connect both the compatible
electrosurgical and ultrasonic generators to the insufflator at
the same time.
When adding the compatible electrosurgical generator to the compatible high flow insufflation unit and the
ultrasonic generator that have been already connected, follow the below steps (see Figure 3.14).
Figure 3.14
WARNING
Do not connect any equipment other than the optional
footswitch to the footswitch socket. Otherwise, the system
cannot perform appropriately and injury and/or equipment
damage may also result.
CAUTION
Do not place a heavy object, such as a trolley (cart), on the
footswitch cable. Otherwise, cable disconnection may result.
Figure 3.15
Figure 3.16
Figure 3.17
Figure 3.18
Hold the section indicated by the arrow when unplugging the footswitch
plug. The plug cannot be unplugged by pulling the section close to the
footswitch cable.
Figure 3.19
Connect the transducer to the THUNDERBEAT or SONICBEAT instrument before connecting it to the
ultrasonic generator. For the connection of the THUNDERBEAT or SONICBEAT instrument to the transducer,
refer to the instruction manual for the THUNDERBEAT or SONICBEAT to be used.
WARNING
Do not connect an instrument other than the
THUNDERBEAT or SONICBEAT to the transducer socket.
Otherwise, the system will malfunction and injury and/or
equipment damage may also result.
CAUTION
Do not touch the transducer plug contacts. Static electricity
that was accumulated during autoclaving may cause an
electric shock.
Figure 3.20
Figure 3.21
1 Care
WARNING
After wiping the ultrasonic generator with liquid-moistened
gauze, allow it to dry completely before use. Otherwise, an
electric shock hazard may result.
CAUTION
Do not clean the THUNDERBEAT/SONICBEAT sockets,
other sockets and the AC power socket. Otherwise,
deformation or corrosion of the contacts may cause contact
failure as well as a malfunction of the ultrasonic generator
and accessories.
After using the ultrasonic generator, immediately clean it using the following steps. If cleaning is delayed, the
organic dirt may solidify and become difficult to remove. Clean the ultrasonic generator periodically.
4. Cleaning
Remove dirt, dust and stains from the ultrasonic generator and disinfect its
surface using gauze moistened with disinfecting alcohol.
5. Drying
After wiping with disinfecting alcohol, be sure to dry the ultrasonic generator
completely prior to use.
2 Storage
CAUTION
Do not store the ultrasonic generator in a place that is
exposed to direct sunlight, X-rays, radioactivity or strong
electromagnetic waves (e.g., the vicinity of a microwave
therapeutic device, short-wave therapeutic device, MRI,
wireless set, cellular phone, etc.). Also avoid storing in places
of high temperatures, high humidity and where
moisture/water is present. Otherwise, the instrument may be
damaged and/or could present an infection control risk.
1. Store location
Place the ultrasonic generator in a level posture on a stable surface. When it
is not placed on the energy cart.
2. Storge environment
Store the ultrasonic generator properly in a clean, dustless place in
compliance with the environmental conditions given in “Specifications” in the
Specifications.
3 Disposal
When disposing of the ultrasonic generator, be sure to observe your national and local laws and guidelines.
Chapter 5 : Inspection
1 Verification workflow ........................................................................5-2
2 Power verification .............................................................................5-3
3 Verification of communication cable connection between
compatible electrosurgical generator and ultrasonic generator
(when using the THUNDERBEAT) ....................................................5-6
4 Verification of touch-screen and push buttons ..............................5-7
5 Verification of the THUNDERBEAT and/ or SONICBEAT..............5-10
6 Verification of footswitch connection............................................5-13
7 Verification of the high-frequency(RF bipolar) output
(when using the THUNDERBEAT) ..................................................5-13
8 Verification of the alarm system ....................................................5-17
9 Procedure after verification ..........................................................5-19
10 Inspection check sheet...................................................................5-20
WARNING
Always inspect the ultrasonic generator as described in this
chapter before using. Also, inspect the ancillary equipment
used in combination with the ultrasonic generator according
to the instruction manuals. Should any irregularity be
observed, do not use the ultrasonic generator and take
proper measures by referring to Chapter 8
“Troubleshooting”. If the irregularity is still not resolved, do
not use the ultrasonic generator and contact Olympus. Using
irregular equipment may cause a malfunction and may also
result in an electric shock, burns and/or fire hazard.
Prepare the ultrasonic generator and ancillary equipment for the surgical procedure by referring to “System
chart” in Product Outline. Inspect the ultrasonic generator and ancillary equipment by referring to the respective
instruction manuals.
1 Verification workflow
The following is the equipment inspection workflow. For details on each step, read the corresponding
description.
2 Power verification
WARNING
Confirm that the power switch, touch-screen and push
buttons illuminate when the ultrasonic generator is turned
ON. If not, check that the power cord is connected properly. If
they still do not illuminate and turn the power switch to OFF,
unplug the power cord and contact Olympus. Continuing the
use of the ultrasonic generator in this condition may result in
a fire or electric shock hazard.
CAUTION
When using THUDNERBEAT, turn the compatible
electrosurgical generator ON before the ultrasonic generator.
Otherwise, an error tone will be generated.
Depress the power switch of the ultrasonic generator (see Figure 5.1).
Figure 5.1
Confirm that the start tone sounds at the same time that the “start screen” is displayed.
Figure 5.2
Figure 5.3
Figure 5.4
Figure 5.5
1. Confirm that the communication cable 0.25 m is connected properly to the generators.
2. Confirm that the power switches of the compatible electrosurgical and
ultrasonic generators are set to ON. If the power switches are not ON turn
the compatible electrosurgical generator ON first, then turn ON the ultrasonic generator.
CAUTION
To prevent malfunction, do not press more than one part on
the touch-screen simultaneously.
Dirt and dust between the touch-screen and front panel may
cause malfunction of the touch-screen. To prevent this, keep
the touch-screen clean.
Pressing and holding the “Plus or Minus button” varies the output level continuously.
2. Press the “Return button” on the bottom right of the screen (see Figure 5.6).
The screen changes from the “Set screen” to the “All screen”.
Figure 5.6
1. Press the “THUNDERBEAT or SONICBEAT button” on the screen (see Figure 5.7)
The screen changes from the “All screen” to the “Set screen”
Pressing and holding the “Plus or Minus button” varies the output level continuously.
3. Press the “Return button” on the bottom right of the screen (see Figure 5.7).
The screen changes from the “Set screen” to the “All screen”.
Figure 5.7
Figure 5.8
2. Press the “Cancel button” on the bottom left of the screen (see Figure 5.9).
The confirmation window (probe check) disappears and the “All screen”
is displayed again.
Figure 5.9
WARNING
After the transducer plug is connected to the ultrasonic
generator and the ultrasonic generator is turned ON, the
generator touch-screen will show the same symbol that is
indicated on the transducer plug. If these symbols are not
identical, the ultrasonic generator or transducer may
malfunction. If this occurs, immediately stop using the system
and contact Olympus.
Figure 5.10
Figure 5.11
The displayed “Set screen” changes to the “All screen” at the moment
the other THUNDERBEAT or SONICBEAT is connected (see Figure
5.12).
Figure 5.12
Figure 5.13
WARNING
The output should always be inspected outside the body
cavity. Otherwise, tissues may be burnt.
CAUTION
Do not touch the probe tip during output. Otherwise, the
high-frequency (RF bipolar) current may cause burns.
If the output tone does not sound or the output screen is not
displayed even when the handswitch or footswitch pedal is
pressed, immediately stop using the system and turn the
system OFF. The ultrasonic generator, the THUNDERBEAT
or footswitch may be defective. Take proper measures as
instructed in Chapter 8, “Troubleshooting”. Continuing the
use of the ultrasonic generator in this condition may result in
burns of the surgeon, surgical staff and/or patient.
If the output tone does not cease to sound even when the
handswitch or footswitch pedal is pressed, immediately stop
using the system and turn the system OFF. The ultrasonic
generator, the THUNDERBEAT or footswitch may be
defective. Take proper measures as instructed in Chapter 8,
“Troubleshooting”. Continuing the use of the ultrasonic
generator in this condition may result in burns of the surgeon,
surgical staff and/or patient.
If you retry the verification and the window and tone persist,
stop using the system and take remedial measures by
referring to Chapter 8, “Troubleshooting” in the instruction
manual for the ultrasonic generator.
If the output level is erroneous, press the “Plus or Minus button” to set
the correct output level.
1. Preparation
Prepare a sterile container (like a bowl) with dimensions similar to those
shown in Figure 5.14. Fill it with saline.
Figure 5.14
Figure 5.15
Figure 5.16
3. Activation
Press and hold the SEAL button (blue) of the handswitch or the SEAL pedal
(right pedal: blue) of the footswitch for the high-frequency (RF bipolar)
output.
Figure 5.17
5. Activation stop
After the confirmation, release the SEAL button of the handswitch or the SEAL pedal of the
footswitch.
6. Cleaning
Wipe any remaining saline on the probe tip with dry, sterile gauze.
WARNING
Before generating an alarm, be sure to confirm that the
transducer plug is unplugged from the transducer socket of
the ultrasonic generator. If the handswitch is depressed while
the transducer plug is connected, an unintended output may
cause burns..
Figure 5.18
Figure 5.19
The error is cleared and the error window disappears from the touch-screen.
Figure 5.20
Day 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
USG-400
Inspection
Power verification
Inspector
5-20
Day 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Power verification
Inspector
ISSUE1
USG-400
1 Check procedures
1-1 Grounding Resistance confirmation
○ Criteria
The resistance between an accessible metal part of the USG-400 and the grounding
terminal of the meet must be 0.1Ω or less.
○ Inspection procedures
(1) Measure the resistance value between the inlet grounding terminal the exterior chassis
screw farthest from the inlet to apply 25A (non-loaded condition : 6A or less) for five to ten
seconds.
Chassis screw
USG-400
Chassis screw
○ Inspection procedures
(1) Make measurements at the grounding terminal of the inlet.
○ Inspection procedures
(1) Make measurement at the exterior chassis screw farthest from the inlet.
Chassis screw
○ Inspection procedures
(1) Turn off the power switch of USG-400.
(2) Connect the Leakage current measuring jig (SS0499) to USG-400, and then connect the
SONICBEAT Transducer with the SONICBEAT instrument (SB-0545PC or SB-0545IC)
attached to it to the opposite side.
USG-400
(3) Connect the Leakage current measuring jig (SS0499) to the Leakage current tester.
USG-400
(5) Turn on the power switch of USG-400, and then measure the leak current when the
SONICBEAT is activating at level 3( either handswitch can be used)
2 Check Card
Date of check
Product SN
Inspected by Checked by
Measurement
Inspection item Criteria result
Equipment No.
Grounding Resistance The resistance must be 0.1Ω Ω
Ground leakage Normal Condition 500 μA μA
current Single Fault 1000 μA μA
Normal Condition 100 μA μA
Enclosure leakage
current 500 μA
Single Fault μA
300 μA (UL Deviation)
AC 10 μA μA
Normal Condition
Patient leakage DC 10 μA μA
current AC 50 μA μA
Single Fault
DC 50 μA μA
1 Block Diagram
LINK-IN LINK-OUT
connector connector
DC/AC Control
Power inlet converter Touch-screen
circuit
circuit
Oscillation
control circuit
THUNDERBEAT
footswitch socket
Ultrasonic
Transducer socket
Amplifier output (THUNDERBEAT)
SONICBEAT
footswitch socket circuit
Transducer socket
(SONICBEAT)
Chapter 8 : Troubleshooting
1 Troubleshooting ................................................................................8-2
2 Error screen.......................................................................................8-3
3 Troubleshooting guide......................................................................8-3
4 Returning the ultrasonic generator for repair...............................8-22
1 Troubleshooting
WARNING
If the ultrasonicIf the ultrasonic generator does not work after the
touch-screen displays irregularity, turn the ultrasonic
generator OFF and then turn it ON again. Confirm that the
output level setting returns to the normal state.
If any irregularity is observed with the equipment, follow the below steps.
1. Immediately stop the procedure and remove any equipment from inside the
body cavity.
2. Immediately stop the procedure and remove any equipment from inside the
body cavity.
2 Error screen
The error window is configured as shown below in Figure 8.1
Figure 8.1
3 Troubleshooting guide
Types of error codes
The error windows are displayed in one of the following six patterns.
Pattern 5 Target error code: U509, U510, U511, U512, U514, U601
Turn the ultrasonic generator OFF and ON again. If the error window is still
displayed, disconnect the power supply and contact Olympus.
Turn the ultrasonic generator OFF and ON again. If the error window is
still displayed, disconnect the power supply and contact Olympus.
If the error code is still displayed, disconnect the power supply and
contact Olympus (see Figure 8.2).
Figure 8.2
When the “OK button” is pressed, the error window disappears and the
display returns to the screen displayed before the error window was
generated (see Figure 8.3).
Figure 8.3
Energy output is not available when the error window (U504: Probe damage
error) is displayed. In this case, perform the probe check. The output will become
available when the confirmation window (Activation OK) is displayed.
Remedial actions:
Figure 8.4
Figure 8.5
To cancel the probe check before the execution, press the “Cancel button” on the
bottom left of the confirmation window (Probe Check).
The display returns from the confirmation window (Probe Check) to the
“All screen” or “Set screen”.The confirmation window (Probe Check) is displayed.
Figure 8.6
The confirmation window (Activation OK) disappears and the “All screen” or
“Set screen” is displayed (see Figure 8.7).
Figure 8.7
Figure 8.8
Do not press the “OK button”, but proceed to the remedial actions on the
error window or of “What to do when an error code is displayed”.
If the “OK button” is pressed without taking the remedial actions, the
error window disappears even when the error is not cleared.
When the error window is not displayed after inspection, the error has
been cleared.
If the error window is still displayed, the error is not cleared. Take the
next remedial action.
Figure 8.9
Do not press the “OK button”, but pressing the “Next button”, perform to
the remedial actions on the error window.
If the “OK button” is pressed without taking the remedial actions, the
error window disappears even if the error is not cleared.
When the error window is not displayed after inspection, the error has
been cleared.
If the error window is still displayed, the error is not cleared. Take the
next remedial action.
Figure 8.10
If the error window is still displayed, take the next remedial action.
The error U505 has two pages for the remedial actions.
Figure 8.11
The ultrasonic generator The (push) buttons are not Release the footswitch or handswitch to
does not react when a available during activation. stop the activation.
(push) button on the front
panel is pressed.
No sound is audible The volume is set to an inaudible Increase the volume either on the
during activation. level (e.g. due to high touch-screen within the “Menu screen” or
environmental noise). use the volume control on the rear panel of
the ultrasonic generator.
Ultrasonic generator malfunction Stop using the ultrasonic generator and
press the power switch to turn OFF the
ultrasonic generator. Contact Olympus.
The error volume can not The error volume is fixed. No action required.
be adjusted.
Automatic mist & smoke The settings are erroneous. Correct the settings of the compatible high
Evacuation system/function flow insufflation unit.
does not work.
The communication cable is not Connect the communication
connected. cable. ”Automatic mist & smoke
evacuationsystem/function (when using
the compatible high flow insufflation unit)”
The connection of the Reconnect the communication cable.
communication cable is “Automatic mist & smoke evacuation
erroneous. system/function (when using the
compatible high flow insufflation unit)”.
Compatible high flow insufflation Contact Olympus.
unit malfunction.
The transducer is heated. Transducer temperature is too Stop output until the temperature cools down.
high due to continuous output.
Difficult to grasp tissue. Operation is not appropriate. Operate the instrument properly referring to
Chapter 4, “Operation” in the instruction
manual for THUNDERBEAT or SONICBEAT.
The treated site is not appropriate. Treat the adequate site referring to Chapter
4, “Operation” in the instruction manual for
THUNDERBEAT or SONICBEAT.
The grasping surface (white Teflon If the grasping surface (white Teflon surface)
surface) has been worn out. has been worn out, replace THUNDERBEAT
or SONICBEAT with a new one.
Carbonized tissue is present on the Remove the carbonized tissue from the
grasping section or the probe tip. grasping section or the probe tip with sterile
gauze or soft brush.
THUNDERBEAT or SONICBEAT The THUNDERBEAT or SONICBEAT
instrument malfunction. instrument is defective. Remove the
THUNDERBEAT and/or SONICBEAT from
the body cavity and press the power switch
to OFF. Replace the THUNDERBEAT or
SONICBEAT instrument.
Opening and closing ofthe Carbonized tissue is present onthe Remove the carbonized tissue from the
grasping section is grasping section or the probe tip. grasping section or the probe tip with sterile
stiff/can't open. gauze or soft brush.
THUNDERBEAT or SONICBEAT The THUNDERBEAT or SONICBEAT
instrument malfunction. instrument is defective. Remove the
THUNDERBEAT and/or SONICBEAT from
the body cavity and press the power switch
to OFF. Replace the THUNDERBEAT or
SONICBEAT instrument.
Irregularity
Error Code Possible cause Remedial Actions
Description
U001 USG-400 Ultrasonic generator malfunction Turn the ultrasonic generator OFF
generator error due to the fail of the internal and ON again. If the error window is
(Configuration circuitry of the ultrasonic still displayed, contact Olympus.
malfunction) generator.
U002 USG-400 Same as the above. Same as the above.
generator error
(FPGA (software)
malfunction)
U007 USG-400 Ultrasonic generator malfunction Turn the ultrasonic generator OFF
generator error due to the fail of the internal and ON again. If the error window is
(Memory damage) circuitry of the ultrasonic still displayed, contact Olympus.
generator.
Irregularity
Error Code Possible cause Remedial Actions
Description
U009 USG-400 Ultrasonic generator malfunction Turn the ultrasonic generator OFF
generator error due to the fail of the internal and ON again. If the error window is
(Calibration circuitry of the ultrasonic still displayed, contact Olympus.
malfunction) generator.
U010 Overcurrent error Ultrasonic generator malfunction. 1.Turn the ultrasonic generator OFF
then ON again. Confirm instrument
activation.
Transducer malfunction.
2. If no output, replace transducer.
repeat step 1.
3. If the error window is still displayed,
contact Olympus.
U011 Cooling fan error Foreign object is caught in rear 1. Confirm a foreign substance object
cooling fan. is not caught in rear vent holes,
then turn the ultrasonic generator
OFF and ON again.
Ultrasonic generator malfunction. 2. If the error window is still displayed,
contact Olympus.
U012 Overheated circuit Fan entrance/exit is blocked. 1. Confirm vent holes are not covered
error by an obstacle, and then turn the
ultrasonic generator OFF and ON
again.
Ultrasonic generator malfunction. 2. If the error window is still displayed,
contact Olympus.
U013 Touch-screen error Touch-screen was pressed when 1. Turn the ultrasonic generator OFF
the ultrasonic generator was and ON again without touching the
turned ON. touch-screen.
U014 Push button error A push button was pressed when 1. Turn the ultrasonic generator OFF
the ultrasonic generator was and ON again without touching the
turned ON. push button.
U501 Handswitch error Handswitch was pressed when 1. Turn the ultrasonic generator OFF
the ultrasonic generator was and ON again without touching the
turned ON. handswitch.
Irregularity
Error Code Possible cause Remedial Actions
Description
U502 Footswitch error Footswitch pedal was pressed 1. Turn the ultrasonic generator OFF
when the ultrasonic generator is and ON again without pressing the
turned ON. footswitch pedal.
U503 System connection A unit with overlapped ID is 1. Confirm the system does not have
error connected. Example: When two multiple same generators.
compatible electrosurgical
generators are connected.
U507 Unsupported ESG The software version of the Contact Olympus (Connected ESG
error connected compatible not supported by USG-400).
electrosurgical generator is not
supported.
Irregularity
Error Code Possible cause Remedial Actions
Description
U508 SEAL & CUT short Output is generated while a 1. Confirm no metal object is grasped
circuit error metallic object is grasped. between jaw and probe, then
confirm activation.
U509 Ultrasonic level error Hard tissue (e.g., bone) is 1. Do not grasp the hard tissue.
caught.
Irregularity
Error Code Possible cause Remedial Actions
Description
U510 Ultrasonic Hard tissue (e.g., bone) is 1. Do not grasp the hard tissue.
frequency error caught.
(which is detected
during surgery) Excessive load is applied to the 2. Reduce force applied to distal end
distal end. of the probe.
Connection of the transducer 3. Check if the THUNDERBEAT or
with THUNDERBEAT or SONICBEAT instrument and
SONICBEAT instrument is loose. transducer are connected firmly. If
not, connect them firmly.
Irregularity
Error Code Possible cause Remedial Actions
Description
U513 Time out error (SEAL Output time period is longer than The maximum output time period has
& CUT, MAX, VAR) allowable by the system been expired. Re-grasp the tissue and
(maximum output time: 60 s). confirm activation.
U514 Time out error The specified output time period Remove tissue or fluids from jaw and
(SEAL incomplete) has expired (maximum output probe. Re-grasp tissue and confirm
time: 15 s). activation.
U601 Probe damage Cracked probe. 1. Confirm that the probe tip is not
error (which is broken. Press the PROBE CHECK
detected through push button, open the grasping
PROBE CHECK) section and confirm the normal
operation. If the probe is broken,
replace the THUNDERBEAT or
SONICBEAT instrument.
Excessive load is applied to the 2. Reduce force applied to distal end
distal end. of the probe.
Connection of the transducer 3. Check if the THUNDERBEAT or
with THUNDERBEAT or SONICBEAT instrument and
SONICBEAT instrument is loose. transducer are connected firmly. If
not, connect them firmly.
Transducer plug is not 4. Confirm that the transducer plug is
connected o the connected to the THUNDERBEAT
THUNDERBEAT or ONICBEAT or SONICBEAT socket on the front
socket properly. panel of the ultrasonic generator. If
not, connect it properly.
Transducer temperature is too 5. Allow the transducer temperature
high due to use immediately after cool prior to use.
autoclaving.
Transducer temperature is too 6. Stop output until the temperature
high due to continuous output. cools down.
Transducer malfunction. 7. Replace the transducer.
When returning the ultrasonic generator for repair, contact Olympus. With the
ultrasonic generator, include a description of the malfunction or damage and the
name and telephone number of the individual at your location who is most
familiar with the problem. Include a repair purchase order.
Engineering 2
Service Engineering Department
Service Division