BioChemistry 1021 Lab - Protein Determination
BioChemistry 1021 Lab - Protein Determination
BioChemistry 1021 Lab - Protein Determination
Methodology: The experiment was executed as outlined in the BioC 1021 lab Manual.
Questions:
1. From the information provided, calculate the protein concentration for each sample.
(20 marks)
3. Using your graph and the measured OD at 595 nm, determine the concentration of your
According to the graph, the concentration of the unknown protein after extrapolation is
4. Think about whether or not it should start at the origin and give your reasons. (5 marks)
The best fit curve should start at the origin (0,0) because the blank sample does not have
any of the protein samples included in it. Hence, when placed in the spectrophotometer, an
absorbance of 0.0 OD at 595 nm with a protein concentration of 0.0mg/mL, which would have
been the first recorded and plotted point of the graph. Beer-Lambert’s law is represented by a
accordance with Beer-Lambert's Law, as it is evident that the absorbance and the concentration
Accuracy can be defined as the degree to which the result of a measurement conforms to
the correct value or a standard and it essentially relates to how close a measurement is to its
results of high degrees of accuracy, precautions need to be adhered to and sources of errors need
to be little to none. In this experiment to determine the concentration of protein in the different
samples, the apparatus used such as the spectrophotometer and micropipette were ensured to be
properly used in order to avoid errors and get accurate results. The disposable pipette tips were
changed when necessary (for example, when touched by surfaces such as other solutions, the
hand, the lab table among others) to avoid contamination of other samples, it was ensured that
light passed through the sample as much as possible by wiping off any smudges that might have
been on the transparent sides of the cuvette and avoiding scratches that would have caused any
deflection. Also, it can be concluded that the results obtained some degree of accuracy as it
reflects the principle of Beer-Lambert’s Law which is defined as the absorbance being directly
proportional to the concentration (c) of the solution and the light path length (l) (Clarke, 2019).
This equation can be represented by a linear line or curve which illustrates direct proportionality
and proves the results gathered after plotting the graph. Finally, the spectrophotometer was
blanked by using a solution that contained water and Bradford reagent and no protein (BSA), so
that when the absorbance of the sample was measured, it is known that the light passing through
is deflected only by the protein and that the absorbance readings only account for the BSA
present. Accuracy also reflects how close a value is to a standard, which is an accepted value.
extent to which a tool or equipment can yield the same result when used repeatedly by different
individuals, using the same input variables and methodological steps at various locations under
the same analysis conditions. Reproducibility is pertinent to this experiment because a duplicate
of each protein determination sample was made. This was to see if the absorbance reading of 1B
would be in the same range of values as 1A, and such was observed for most of the samples
because the same conditions and procedures were applied, for example, the incubation time and
the volumes of reagents used were identical and the cleaning of the cuvettes and their appropriate
Thence, it can be concluded that the results from this experiment were accurate and
reproducible.
6. Name TWO (2) alternate methods that could be used to determine protein concentration
in a sample. In your answer, briefly describe the principle of each method. (10 marks)
In addition to the Bradford Assay, the Ultraviolet Absorption Method and the Lowry
The aromatic amino acid residues (tyrosine and tryptophan) in a protein exhibit an
absorption maximum at a wavelength of 280 nm. Since the proportions of these aromatic amino
acids in proteins vary, the extinction coefficients for individual proteins also vary. However, for
the majority of proteins the extinction coefficient lies in the range 0.4–1.5; so for a complex
mixture of proteins it is a fair approximation to say that a solution with an absorbance at 280 nm
and, unlike colorimetric methods, this technique is non-destructive, as having made the
measurement, the sample in the cuvette can be recovered and further used.
This is useful especially when working with small amounts of protein that cannot be
allowed to go to waste. Even so, the method is subject to interference by the presence of other
compounds that absorb at 280 nm, such as nucleic acids, as they fall into this category having an
absorbance as much as 10 times that of protein at this wavelength. Hence the presence of only a
small percentage of nucleic acid can greatly influence the absorbance at this wave-length.
However, if the absorbances (A) at 280 and 260 nm wavelengths are measured, it is possible to
continuously, for example in chromatographic column effluents. Even greater sensitivity can be
obtained by measuring the absorbance of ultraviolet light by peptide bonds. The peptide bond
absorbs strongly in the far ultraviolet, with a maximum at about 190 nm. However, because of
the difficulties caused by the absorption by oxygen and the low output of conventional spectro-
photometers at this wavelength, measurements are usually made at 205 or 210 nm. Most proteins
have an extinction coefficient for a 1 mg cm3 solution of about 30 at 205 nm and about 20 at 210
nm. Clearly therefore measuring at these wavelengths is 20 to 30 times more sensitive than
measuring at 280 nm, and protein concentration can be measured to less than 1 µg cm-3.
buffers and other buffer components commonly used in protein studies also absorb strongly at
protein concentration determination. The Lowry method is reasonably sensitive, detecting down
to 10 µg cm-3 of protein, and the sensitivity is moderately constant from one protein to another.
When the Folin reagent (a mixture of sodium tungstate, molybdate and phosphate), together with
a copper sulphate solution, is mixed with a protein solution, a blue-purple colour is produced
which can be quantified by its absorbance at 660 nm. The Folin reagent is reduced as the cuprous
ions are oxidized, creating Molybdenum Blue, a blue complex that absorbs light at 750 nm.
Using the Lowry method, this complex is what is used to calculate the protein concentration. The
Lowry Assay Method works on the following principle: a darker blue solution forms when the
absorbance at 750 nm increases along with the intensity of the solution. The solution’s increasing
strength is a sign that the amount of Molybdenum Blue is rising. More Molybdenum Blue in a
solution indicates more reduced Folin reagent and hence more cuprous ions in the solution. It is
deduced that cuprous ions only result when the cuprous ions react with the peptide bonds.
Therefore, the rise in cuprous ions implies that more cupric ions are being reduced as a result of
their interactions with more peptides. The more peptide linkages present, the greater the
concentration of protein.
Like the majority of colorimetric assays, care must be taken that other compounds that
interfere with the assay are not present. For this method, this includes Tris, zwitterionic buffers
such as Pipes and Hepes, and EDTA. The method is essentially based on both the Biuret
reaction, where the peptide bonds of proteins react with Cu22+ under alkaline conditions
producing Cu+, which reacts with the Folin reagent, and the Folin–Ciocalteau reaction, which
by the copper-catalysed oxidation of aromatic amino acids. The resultant strong blue colour is
therefore partly dependent on the tyrosine and tryptophan content of the protein sample. Though
the Lowry Assay Method is more sensitive than the Biuret Assay Method, it takes much longer
and is more susceptible to the effects of chemicals. Additionally, the Lowry Assay has other
disadvantages such as its pH-which is in the range of 10.0 to 10.5- as it exclusively operates
References
https://egyankosh.ac.in/bitstream/123456789/71248/1/Experiment-2.pdf
Aliouche, Hidaya (2022, July 15). What is Reproducibility? News Medical Life Sciences.
Reproducibility.aspx
Dahal, Prashant (2023, April 16), Biuret Test for Protein- Principle, Procedure, Results, Uses
Department of Basic Medical Sciences, Mona Campus, U.W.I. BIOC 1021 Lab Manual
Practical Biochemistry
Helmenstine, Anne Marie, Ph.D. (2020, August 25), Accuracy Definition in Science.
accuracy-in-science-604356
Kielkopf, Clara L. (2020 Apr 1), Methods for Measuring the Concentrations of Proteins from
PubMed
https://pubmed.ncbi.nlm.nih.gov/32238598/#:~:text=Determining%20the%20concentration
%20of%20protein,Lowry%2C%20or%20BCA%20assays)
Mackenzie, R.J. (2019, March 25), Repeatability v.s. Reproducibility. Informatics from
https://www.technologynetworks.com/informatics/articles/repeatability-vs-
reproducibility317157
Wilson, Keith and Walker, John (2010), Protein Determination, Principles and Techniques of
Biochemistry and Molecular Biology, Seventh edition
Retrieved on September 17, 2023.