LCD20 Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 83

LCD20 User Manual

mantracourt.com

LCD20
Load Cell Amplifier In DIN Rail Format
Contents

LCD20 User Manual ......................................................................................................................................................3


Overview ...................................................................................................................................................................................... 3
How to Use This Manual ........................................................................................................................................................... 4
Using the Handheld Programmer/Display (LP2) ................................................................................................................... 5
Keypad and Display ............................................................................................................................................................................................................................. 5
Keys ..................................................................................................................................................................................................................................................... 6
Menus................................................................................................................................................................................................................................................. 7
Password Protection ...................................................................................................................................................................................................................... 7
Menu Tables ..................................................................................................................................................................................................................................... 7
Auto Calibration ..................................................................................................................................................................................................................................13
Procedure ........................................................................................................................................................................................................................................13
Notes on Auto calibration ...................................................................................................................................................................................................14
Using the LC Toolkit Software ................................................................................................................................................ 15
What Can It Do? ..................................................................................................................................................................................................................................15
Using the Software.............................................................................................................................................................................................................................15
Home Page ...........................................................................................................................................................................................................................................16
Information ...........................................................................................................................................................................................................................................17
Save & Restore ....................................................................................................................................................................................................................................19
Trend Chart ...........................................................................................................................................................................................................................................20
Logging..................................................................................................................................................................................................................................................22
Analogue Outputs ..............................................................................................................................................................................................................................25
Relays .....................................................................................................................................................................................................................................................28
Measurement .......................................................................................................................................................................................................................................29
Digital Inputs........................................................................................................................................................................................................................................31
Calibration.............................................................................................................................................................................................................................................32
Automatic Tab ...............................................................................................................................................................................................................................33
Table Tab .........................................................................................................................................................................................................................................34
Advanced Tab ................................................................................................................................................................................................................................35
Shunt Calibration Tab .................................................................................................................................................................................................................36
Cal Info Tab ....................................................................................................................................................................................................................................37
Filters ......................................................................................................................................................................................................................................................38
Installation................................................................................................................................................................................. 40
Mechanical............................................................................................................................................................................................................................................40
Front Panel......................................................................................................................................................................................................................................40
Connections .........................................................................................................................................................................................................................................41
Load cell Connections .................................................................................................................................................................................................................41
Connector Location ...............................................................................................................................................................................................................42
4 Wire .........................................................................................................................................................................................................................................42
6 Wire .........................................................................................................................................................................................................................................43
Barriers .......................................................................................................................................................................................................................................44
Digital Inputs..................................................................................................................................................................................................................................45
Connector Location ...............................................................................................................................................................................................................45
Connections .............................................................................................................................................................................................................................45
Analogue Output ..........................................................................................................................................................................................................................46
Connector Locations .............................................................................................................................................................................................................46
Voltage.......................................................................................................................................................................................................................................47
Current .......................................................................................................................................................................................................................................47
Relays ...............................................................................................................................................................................................................................................48
Contact Connections .............................................................................................................................................................................................................48
Programmer Module ...................................................................................................................................................................................................................49
LP2...............................................................................................................................................................................................................................................49
Programming Cable PGM1 .......................................................................................................................................................................................................50
Driver Settings .........................................................................................................................................................................................................................51
LED Indicator ........................................................................................................................................................................................................................................53
Detect Switch .......................................................................................................................................................................................................................................54
Appendix A – Calibration ........................................................................................................................................................ 55
Best practices .......................................................................................................................................................................................................................................55
Table Calibration Method ................................................................................................................................................................................................................55
Shunt Calibration ................................................................................................................................................................................................................................55
Using Shunt Calibration..............................................................................................................................................................................................................56
Appendix B – Measurement .................................................................................................................................................... 57
Overview................................................................................................................................................................................................................................................57
Zero Tracking .......................................................................................................................................................................................................................................60
System Zero..........................................................................................................................................................................................................................................60
Auto Tare...............................................................................................................................................................................................................................................60
Filters ......................................................................................................................................................................................................................................................61
Overview..........................................................................................................................................................................................................................................61
Simple Block Averaging..............................................................................................................................................................................................................61
Dynamic Digital Low Pass Filter ...............................................................................................................................................................................................61
Peak and Valley Values .....................................................................................................................................................................................................................64
Appendix C – Relays ................................................................................................................................................................. 65
Overview................................................................................................................................................................................................................................................65
Source ....................................................................................................................................................................................................................................................65
Setpoints ...............................................................................................................................................................................................................................................65
Inflights ..................................................................................................................................................................................................................................................66
Hysteresis ..............................................................................................................................................................................................................................................66
Inversion ................................................................................................................................................................................................................................................66
Latching .................................................................................................................................................................................................................................................66
Appendix D – Analogue Output ............................................................................................................................................. 67
Overview................................................................................................................................................................................................................................................67
Source ....................................................................................................................................................................................................................................................67
Scaling ....................................................................................................................................................................................................................................................67
User Adjustment .................................................................................................................................................................................................................................69
Appendix E – Digital Inputs .................................................................................................................................................... 70
Overview................................................................................................................................................................................................................................................70
Digital Input selection table ............................................................................................................................................................................................................70
Appendix F – Security Locking ............................................................................................................................................... 71
Menu Passwords .................................................................................................................................................................................................................................71
Handheld/Display Calibration Lock ..............................................................................................................................................................................................71
LC Toolkit Lock ....................................................................................................................................................................................................................................72
Specifications ............................................................................................................................................................................ 73
Electronics .............................................................................................................................................................................................................................................73
Loadcell Input ................................................................................................................................................................................................................................73
Analogue Output ..........................................................................................................................................................................................................................73
Power Supply .................................................................................................................................................................................................................................74
Relays ...............................................................................................................................................................................................................................................74
Environmental................................................................................................................................................................................................................................74
Regulatory ................................................................................................................................................................................. 75
CE Approvals ........................................................................................................................................................................................................................................75
European EMC Directive.............................................................................................................................................................................................................75
Low Voltage Directive .................................................................................................................................................................................................................75
Troubleshooting ....................................................................................................................................................................... 76
General connection and setup parameters. – No flashing status LED on power up. .............................................................................................76
Display shows (-1 or 1) continually, without a weight applied to the Strain Gauge...............................................................................................76
Display over ranges (-1 or 1) when, or before, the maximum required weight is applied to the Strain Gauge. ..........................................76
Display very noisy .........................................................................................................................................................................................................................76
Display operating in wrong direction ....................................................................................................................................................................................76
Unit will not auto calibrate ........................................................................................................................................................................................................76
Incorrect relay operation............................................................................................................................................................................................................77
Order Codes .............................................................................................................................................................................. 78
Accessories .....................................................................................................................................................................................................................................78
Configuration Record Sheet ................................................................................................................................................... 79
Warranty .................................................................................................................................................................................... 81

Page 2
Mantracourt Electronics Ltd LCD20 User Manual
LCD20 User Manual
Overview
The LCD20 In-Line Intelligent Strain Gauge Amplifier is a compact microprocessor-based unit
specifically designed to control and monitor weighing applications. Its flexibility of design allows for
the connection of most load cells and pressure or strain gauges over a wide range of sensitivities.
Housed in a DIN rail module.

The basic unit offers the following facilities:-


A simple auto calibration of the highest and lowest weights required, an easy auto tare setting and a
peak hold and valley facility. A menu password protects the configuration parameters.

DC analogue outputs of 4-20mA and 0-10V are standard with full scaling over any desired range and
the ability to invert these outputs if required.

Gain sensitivity is available in two ranges, 0.5 to 3.7mV/V and 3.7 to 7.8mV/V.

Offers two relay set points which can be inverted and/or latched. Hysteresis can be applied to both set
points together with individual In-Flight compensation. Both the relay and the Analogue Output have
a high level of electrical isolation.

Internal parameters can be set digitally in real engineering terms by means of an optional plug-in
programmer unit or LC Toolkit software.

The load cell input design has also taken advantage of new technology in the form of the delta-sigma
20 bit ADC. This device has a built-in low temperature drift, low noise amplification stage with
switchable gain. The appropriate range is selected from either the optional programmer keypad or via
the Toolkit software and PC communications cable removing the need to gain access to the
electronics.

The LCD20 is factory calibrated in mV/V terms to within 0.05% FS, this allows table entry of calibration
data straight from a load cell test certificate. An advantage of a factory calibration is that units can be
swapped in the field by copying across the calibration from one unit to another with calibration
accuracy maintained at 0.05%FS

Page 3
Mantracourt Electronics Ltd LCD20 User Manual
How to Use This Manual
There are two ways you may want to use the LCD20 and this manual has been arranged to best allow
the information required to be presented without having to read irrelevant information.

The two ways to interact with the LCD20 are as follows:


 Handheld programmer/display (LP2)
 LC Toolkit software (With PGM1 cable)

It may be that only one of the above methods is used or it may be possible that you will need to refer
to more than one section.

Page 4
Mantracourt Electronics Ltd LCD20 User Manual
Using the Handheld Programmer/Display (LP2)

Keypad and Display


The Programmer Unit - Is a small hand held unit together with a 2 metre connection lead known as
LP2 which plugs into an FFC68 socket accessed under the front panel door.

All user controls, displays and indicators are mounted on the front panel which comprises a 4.5 digit
LCD display and four flush mounted keys.

A flashing ‘’ symbol in the top left hand corner of the display indicates programming mode. This
symbol is also used to indicate there is a load cell integrity issue or that Shunt-Cal is switched on.

Unit Panel Layout

Page 5
Mantracourt Electronics Ltd LCD20 User Manual
Keys

The LCD20 has a multilevel menu system. The parameters available are dependent on the password
level that has been entered.

d Used to scroll through the menu items available at the current menu level.
When a level has been entered that holds parameters then this key scrolls
through the parameters available. At this point if the scroll key is not pressed
for 3 seconds then the current value of the parameter is displayed.

b Initially used to enter into the next menu level. When displaying a parameter
value then this key selects the display digit to be modified. The selected digit
will flash together with the program indicator.
When in programming mode it should be noted that the first digit in the
display may not be visible, but the program indicator ’’ will flash to indicate
that the instrument is in programming mode even though no digits can be
seen to be flashing.

c Used to move up a menu level. If already at the highest menu level PASS
then the menu system will be exited returning to displaying the input
variable.
When displaying a parameter value this key increments each selected display
digit 0-9. Pressing the c key under programming conditions will display the
leading digit as 1, -1, - or a blank display for zero.

Additionally, for items that can be selected from a list such as the digital
input function DIP1, then this key cycles through the available options. The
currently set value is the first shown data value.

a Resets the display to the input variable and enters new data in the LCD20
non-volatile memory. If, during the programming sequence no key is pressed
the display will revert to showing the input variable after 2 minutes and any
changes made will not be saved

If on pressing the d or a keys after a new value has been entered that is outside the allowed
range for that parameter then Err will be displayed until the d or a keys is pressed again.
Pressing the a key will exit the menu restoring the last good value to the parameter. Pressing
the d key will show the parameters mnemonic again followed by the last good value.

Page 6
Mantracourt Electronics Ltd LCD20 User Manual
Menus
The LCD20 menu system operates with multiple levels to allow easier navigation round the
parameters. Different levels and parameters being made available depending on the password level
entered. There are 3 password levels which can be considered as Operator, Engineer and
Configurator/Calibrator.

Password Protection
To enter the menu a 4 digit password number must be entered when 'PASS' is displayed. The factory
set number, 1111 allows access to the most common parameters.
Additional passwords are required to gain access to the remaining two levels.

Menu Tables
In the table below the hierarchal level of menu items is indicated by a prefix dot.

Item Pass Range Default Description


Level
PASS 1 1111 Security Password. Correct value required to proceed
further.
VALS 1 Values. View generated live values
GroS 1 - - Gross
nEt 1 - - Net
PEAK 1 - - Peak
VALY 1 - - Valley
SNAP 1 - - Snap (Held Value)
mVV 2 - - mV/V
CALV 2 - - Calibrated Value. Gross value before System zero, Zero-
Tracking
rLYS 1 Relay Menu
rLY1 1 Relay 1 Menu
SP1 1 ±19999 0000 Desired Set Point trip value for Relay 1
IF1 1 ±19999 0000 In-Flight compensation for SP1.
Relay 1 operates at = (SP1 - IF1)
HYS1 1 0-19999 0000 Hysteresis amount applied to SP1
InV1 1 No Invert SP1
Selection value Description
Yes Inverted SP1 operation
No Normal SP1 operation

Page 7
Mantracourt Electronics Ltd LCD20 User Manual
Item Pass Range Default Description
Level
Ltc1 1 No Latch SP1
Selection value Description
Yes Latching SP1 operation
No Normal SP1 operation

rLS1 2 nEt Relay 1 source taken from


Selection value Description
nEt Net Value
GroS Gross Value
PEAK Peak Value
VALY Valley Value
SNAP Snap (Held) value

rLY2 1 ±19999 0000 Relay 2 Menu


SP2 1 ±19999 0000 Desired Set Point trip value for Relay 2
IF2 1 ±19999 0000 In-Flight compensation for SP2.
Relay 2 operates at = (SP2 – IF2)
HYS2 1 0-19999 0000 Hysteresis amount applied to SP2
InV2 1 No Invert SP2
Selection value Description
Yes Inverted SP2 operation
No Normal SP2 operation

Ltc2 1 No Latch SP2


Selection value Description
Yes Latching SP2 operation
No Normal SP2 operation

rLS2 2 nEt Relay 2 source taken from


Selection value Description
nEt Net Value
GroS Gross Value
PEAK Peak Value
VALY Valley Value
SNAP Snap (Held) value

AnOP Analogue Output Menu


OPL 1 ±19999 0000 Output Low: display point for minimum Analogue Output
OPH 1 ±19999 0000 Output High: display point for maximum Analogue
Output

Page 8
Mantracourt Electronics Ltd LCD20 User Manual
Item Pass Range Default Description
Level
AnOP 2 nEt Sets what Value the Analogue Output operates from
Selection value Description
nEt Net Value
GroS Gross Value
PEAK Peak Value
VALY Valley Value
SNAP Snap (Held) value

AOPS 2 4-20 Selects Analogue Output, current or voltage.


Selection value Description
4-20 4-20mA current output
0-10 0-10V voltage output

InV 2 No Invert Analogue output. Note this can also be achieved by


using OPL greater than OPH.
Selection value Description
Yes Inverted analogue output
No Normal analogue output

Adj 3 Adjustment of Analogue output values Menu


4.20G 3 0.9 to 1.0 User adjustment for 4-20mA gain. Replace function of the
1.025 LCA15 gain potentiometer. Unit gain = 1.0
4.20o 3 -1000 to 0 User adjustment for 4-20mA offset. Replace function of
1000 the LCA15 offset potentiometer. Note the value is in PWM
counts not mA
0.10G 3 0.9-1.025 1.0 User adjustment for 0-10V gain. Replace function of the
gain LCA15 potentiometer. Unit gain = 1.0
0.10o 3 -1000 to 0 User adjustment for 0-10V offset. Replace function of the
1000 LCA15 offset potentiometer. Note the value is in PWM
counts not V

Page 9
Mantracourt Electronics Ltd LCD20 User Manual
diSP 1 Display parameters Menu
dA 1 0000 Display Averaging. Number of reading to block average.
Resultant samples per second dependant on rATE setting.
If rATE is set to 10 samples per second
Selection Description
Value
0 2.5 samples per second
1 1.25 samples per second
2 1.6 seconds per sample
3 3.2 seconds per sample
4 6.4 seconds per sample
5 12.8 seconds per sample
6 25.6 seconds per sample
7 10 samples per second

If rATE is set to 80 samples per second


Selection Description
Value
0 20 samples per second
1 10 samples per second
2 5 samples per second
3 2.5 samples per second
4 1.25 samples per second
5 1.6 seconds per sample
6 3.2 seconds per sample
7 80 samples per second

dP 1 0-5 0002 Decimal point position


rS 1 0-19999 0000 Display Resolution
ddiS 2 nEt Selects Defaults display source from
Selection value Description
nEt Net Value
GroS Gross Value
PEAK Peak Value
VALY Valley Value
SNAP Snap (Held) value

PEAk 2 nEt SelectsPeak value source


Selection value Description
nEt Net Value
GroS Gross Value

Page 10
Mantracourt Electronics Ltd LCD20 User Manual
VALY 2 nEt Selects Valley value source
Selection value Description
nEt Net Value
GroS Gross Value

SNAP 2 nEt Selects Snap value source.


Selection value Description
nEt Net Value
GroS Gross Value

OVrV 3 ±19999 19999 Over-range value. The Value at which the display will
indicate a Positive over-range condition and force the
relays and analogue output to behave accordingly.
(Display shows +1)
undV 3 ±19999 19999 Over-range value. The Value at which the display will
indicate a Negative under-range condition and force the
relays and analogue output to behave accordingly.
(Display shows -1)
LCiP 1 Load cell configuration menu
FSTP 1 0-255 0000 Low-pass Filter Steps
FLVL 1 0-19999 0000 Low-pass Filter Dynamic Level
SEnS 2 0.5-3 Load cell sensitivity
Selection value Description
0.5-3 0.5 to 3.7 mV/V range
3.7-7 3.7 to 7.8 mV/V range

rATE 2 10S Measurement rate 10SPS and 80SPS


Selection value Description
10S 10 samples per second
80S 80 samples per second

ACAL 2 Analogue Input Calibration Menu


AdCL 3 ±7.8 0000 mV/V Value captured for low calibration point CALL
AdCH 3 ±7.8 0000 mV/V Value captured for high calibration point CALH
CALL 2 ±19999 0000 Calibration Low: display value for low calibration point
CALH 2 ±19999 0000 Calibration High: display value for high calibration point
At 2 ±19999 0000 Auto Tare value. Value added to Gross to give Net
trbd 2 ±19999 0000 Zero Tracking Band
ZEro 2 ±19999 0000 System Zero
dIPS 2 Digital input menu

Page 11
Mantracourt Electronics Ltd LCD20 User Manual
dIP1 2 Prnt Digital Input 1 function
Selection Description
value
At Perform Tare
Lr Latch Relay Reset
Prnt Do Print
dEFt Display Default display set by “ddiS”
GroS Display Gross value
nEt Display Net value
PEAK Display Peak value
VALY Display Valley value
SNAP Display Snap (Held) value
PV-r Peak & Valley Reset
SCon Shunt-Cal On
SCoF Shunt-Cal Off
SCtG Shunt-Cal Toggle on/off
dSnP Do Snap (Hold current value)

dIP2 2 Lr Digital Input 2 function.


Selection Description
value
At Perform Tare
Lr Latch Relay Reset
Prnt Do Print
dEFt Display Default display set by “ddiS”
GroS Display Gross value
nEt Display Net value
PEAK Display Peak value
VALY Display Valley value
SNAP Display Snap (Held) value
PV-r Peak & Valley Reset
SCon Shunt-Cal On
SCoF Shunt-Cal Off
SCtG Shunt-Cal Toggle on/off
dSnP Do Snap (Hold current value)

Page 12
Mantracourt Electronics Ltd LCD20 User Manual
Auto Calibration
Using the Keypad; a simple two point calibration can be performed. Auto-calibration captures live
load cell input values in mV/V terms and relates them to required engineering unit’s values. The input
can be acquired by loading the load cells with known weights or by injecting known mV levels for the
required engineering values. These known mV/V levels may be taken from a load cell calibration sheet
or by measurement.
The two calibration points should be near the required lowest point (typically an unloaded structure
reading zero weight) and at least 75% of the highest point. This will help provide an accurate
extrapolation over the desired calibrated range.

The sequence of calibration is not confined to capturing the Low calibration point CALL first. This is
useful if a vessel already has a known amount of product in and it is not practical to empty the vessel
before starting the calibration procedure.

For example: the calibration procedure can be initiated by capturing CALH with a full hopper if the
weight is known and CALL captured at a later date when the hopper is empty.

Please note that the LCD20 may not display the correct weight until the second data point is acquired.

The Sensitivity SEnS and Decimal Point position dP must be set before calibration takes place.

Procedure

Allow a warm up period of 20 minutes before carrying out the calibration procedure.

If It is intended that the calibration is permanent then please ensure the EEPROM has not been
disabled by the communications port. See Communications Protocols chapter.

The accuracy of the calibration depends upon the input being settled and stable.

This procedure is written assuming the low calibration point CALL is followed by the high calibration
point CALH but note that CALH could be done first.

1. With the load cells connected and the load cells settled at the low calibration point (this could
be the empty vessel) Press the d key until PASS appears.
2. Enter Level 2 password using b and c keys, then press d key.
3. Press the d key several times until ACAL appears. Press the b key to select and CALL
appears.

If CALL does not appear your LCD20 may be locked from calibration by your supplier!

4. Press the b key and check that the program symbol ‘’ flashes.

Page 13
Mantracourt Electronics Ltd LCD20 User Manual
Always ensure that the programmer indicator is flashing, even though the displayed value may
not need to change.

5. If the required value for CALL is already displayed then move onto to the next step else enter
the desired value that the input on the current load cells represents.
6. Ensure that the Strain Gauge has remained settled and is free from disturbance and press the
d key to capture and calibrate the CALL value.
7. If the CALH calibration point is to be done at a later date then the a can be pressed and the
values for CALL will be committed to memory. Else proceed to calibration high point CALH by
pressing the d.
8. Apply the input to the load cells that represents the calibration high point CALH and allow to
settle.
9. Press the b key and check that the program symbol ‘’ flashes.

Always ensure that the programmer indicator is flashing, even though the displayed value may
not need to change.

10. If the required value for CALH is already displayed then move onto to the next step else enter
the desired value that the input on the current load cells represents.
11. Ensure that the Strain Gauge has remained settled and is free from disturbance and press the
a key to capture and calibrate the CALH value and also commit these the values to memory.
12. On the display returning to indicating the load cell value and assuming the load has not
changed on the load cells the display will now be showing the required values.

Notes on Auto calibration


 If CALH is set to zero then the display value will be mV/V with decimal point position being
that set by dP.
 CALL can be greater than CALH. This will have the effect of for an increasing load cell input a
decreasing displayed value.
 The calibration can be manually adjusted by changing the mV/V values stored in AdCL and
AdCH. Care must be taken with changing these values and the resolution of entry is limited to
0.001mV/V.

Page 14
Mantracourt Electronics Ltd LCD20 User Manual
Using the LC Toolkit Software
The LC Toolkit software for Windows will allow a connection with the LCD20 via an optional
programming cable PGM1.

What Can It Do?


The toolkit allows all aspects of the LCD20 to be configured. Everything that can be achieved with the
plug-in programmer/display can be achieved with the toolkit plus a great deal more.
You can:
 Configure settings  Log data to a CSV file
 Save and restore settings to a file  View data on a chart
 Calibrate load cell  Use as a large display
 Calibrate analogue outputs

Using the Software

These items just display data and you cannot interact with them.

These items are editable in place. Click on them to be able to change the contents.

These items are dropdown lists. Click on them to drop down a selection of items that can be clicked to
select as shown below:

These are buttons. They will change colour dramatically when your mouse passes over them and look
like this:

Click to activate.

These items are tabs. The selected tab is not orange but you can click on the other tabs to select
them. Different tabs will show different control items when clicked.

Page 15
Mantracourt Electronics Ltd LCD20 User Manual
Home Page

This is the first page shown after the software launches.

Some of the pages shown here may not be available if your module has been locked by the
supplier.

The serial ports (COM ports) on the computer are detected and added to the drop down list on the
top left of the page. Clicking the drop down list will automatically re-detect available COM ports.
To detect the connected module:
 Connect the PGM1 cable to the LCD20. Plug the USB cable into the PC. (You may have to wait
for the drivers to install the first time this is plugged in).
 Power up the LCD20.
 Select the COM port that your PGM1 is allocated and click the Detect button.
 Press the Detect button on the LCD20 (Hold for 1 second). See Installation | Push Switch SW
for more information.)

If you cannot get physical access to the LCD20, to power cycle or to push the button, you can
connect manually by clicking the ‘Manual Connect’ text on this page. See Appendix G – Manual
Connect.

Page 16
Mantracourt Electronics Ltd LCD20 User Manual
Information
The Information page shows useful information about the module. The large display may prove useful
in simple monitoring projects.

Main Display – Displays the module value. If units have been specified at calibration then the units
will be shown.
Serial Number – Shows the serial number of the module.
Firmware Version – Shows the module firmware version.
Display Source – Indicates where the display (Parameter and LCD where fitted) gets its value.
Update Rate – Indicates the update rate of the module. This is the rate at which a new value is
generated.
Select Display Format – Drop down the list and select a display format for this software only. This
does not change any module settings.
More - Clicking this button displays more information as shown in the following screenshot.

Page 17
Mantracourt Electronics Ltd LCD20 User Manual
Relay 1 State – Indicates whether relay 1 is on or off.
Relay 2 State - Indicates whether relay 2 is on or off.
Digital Input 1 - Indicates whether digital input 1 is activated or not.
Digital Input 2 - Indicates whether digital input 2 is activated or not.
Shunt Calibration State – Allow the shunt calibration function to be activated.
Less – Clicking this button hides this extra information again.

See Also – Appendix A – Calibration | Shunt Calibration

Page 18
Mantracourt Electronics Ltd LCD20 User Manual
Save & Restore
This page allows the configuration of the module to be saved to a file for later restoration.

Save – Save the module configuration to a file on the computer.


Restore – Restore a previously saved configuration file to the connected module.
Firmware Update – If you have a firmware hex file you can click this to select the file and update
the firmware in the module.

Warning – A failed attempt at a firmware update will render the module inoperable until a
valid firmware update has been achieved.

Page 19
Mantracourt Electronics Ltd LCD20 User Manual
Trend Chart
This page allows you to view the module value on a scrolling chart. The chart will hold 10,000 points
of data and is updated at the update rate of the module (Displayed at the top right of the chart). This
is determined by the settings on the Filters page.
The chart Y axis is auto scaling but you can select the view into a stored history of 10,000 data points.
Just drag the handles of the range scrollbar under the chart to define the start and end point of the
view.
The date and time of the first and last point on the chart are displayed in the lower left and right area
under the chart. By clicking in the chart you can display a cursor and the date and time at the point of
the cursor, also the data point value will be displayed underneath the chart.

Pause Button – Pauses the chart. During a pause no more data will be added to the chart. You can
use the range scrollbar to change the view or click and drag on the chart to move the cursor. The
time and value of the data under the cursor will be displayed under the chart.
Range Scrollbar – By dragging the orange ‘thumb’ you can view different areas of the data history.
By dragging the left or right side of the ‘thumb’ you can increase or decrease the amount of data
shown thus zoom in and out.
Right Click Menu – Right clicking the chart will display a pop-up menu with three options:
Copy Chart Image – Copy a bitmap image of the chart to the clipboard.

Page 20
Mantracourt Electronics Ltd LCD20 User Manual
Copy Chart Data – Copy tab separated chart data to the clipboard in a suitable format for
pasting into MS Excel.
Clear Chart – Clear all data from the chart.

Page 21
Mantracourt Electronics Ltd LCD20 User Manual
Logging
The toolkit allows you to log data to a CSV (Comma Separated Value) file. This file can be opened for
analysis in various applications including MS Excel.
The logging session can be manually controlled or automatically stopped after a duration/number of
logs limit.

Interval

Log Interval - Choose a Log Interval in milliseconds between 12 and 32000. For example 20mS will
enable a log at 50Hz.
Duration

Select – Choose whether you want Manual control, so you can click the Start/Stop button
whenever you want, or choose Based on number of logs or Based on timed duration.
If you choose either of the last two options you can define the length of the log by entering either a
number of logs to record or the time to record in hours, minutes and seconds.
Number of Logs – if you choose Based on number of logs then you can enter the number of logs
here. The hours, minutes and seconds will update to show the actual duration based on the number
of logs and the log interval.
Hours:Minutes:Seconds - if you choose Based on timed duration then you can choose the hours,
minutes and seconds here. The number of logs will update based on the duration and the log
interval.

Page 22
Mantracourt Electronics Ltd LCD20 User Manual
Log Filename - Select a file to log to.

Each time the log starts it will erase any data already in the file.

Start - Clicking this button will start the log.


Stop - Clicking this button will stop the log.
View - When a log has just been stopped this button will launch the application associated with
CSV files.

The format of the log file is as follows:


DateTime, Elapsed, Value

Where:

DateTime is the date and time in long format


Elapsed is the time elapsed in milliseconds since the start of the log
Value is the numeric value logged

Example:

08/06/2009 15:41:08,0,-3.253001E-03
08/06/2009 15:41:08,10,-3.253001E-03
08/06/2009 15:41:08,43,-3.253001E-03
08/06/2009 15:41:08,60,-3.253001E-03

If the measurement rate of the module is less than the rate at which you have chosen to log you will
be informed with a message as shown below.
Either reduce the Log Interval or reduce the Measurement Rate.

Page 23
Mantracourt Electronics Ltd LCD20 User Manual
Once the logging starts the actual achieved log rate will be displayed at the bottom of the window.
This is useful for diagnostics such as when a heavily burdened PC is slowing the log rate.

Page 24
Mantracourt Electronics Ltd LCD20 User Manual
Analogue Outputs
This page allows the configuration and calibration of the analogue outputs.

Output Select – Choose between 4-20mA current output or 0-10V voltage output.
Source – Select the source value that the analogue output is based on.
Inverted – Select this to invert the output. The original LCA15 did not allow Output Low to be
higher than Output High so this option had the same effect as entering a higher value for Output
Low than for Output High. This new module will allow you to do this in both modes so this option is
for backwards compatibility.
Manual Tab – This tab allows you to manually specify the values that relate to the minimum and
maximum analogue output levels.
Output Low – Enter the value (of the source selected above) at which the analogue output will
be at the minimum level (4mA or 0V).
Output High – Enter the value (of the source selected above) at which the analogue output
will be at the maximum level (20mA or 10V).
Value – This shows the current selected source value.
Calculated Analogue Output – Shows the calculated analogue output for the current input
value. Note that this calculated value may be beyond what the output is capable of providing.

Page 25
Mantracourt Electronics Ltd LCD20 User Manual
Automatic Tab – If you do not know what input value relate exactly to the minimum and maximum
analogue output levels but you do know the input values for other output values use this tab.

Output Low – Shows the new Output Low value after you click the Calculate button.
Output High – Shows the new Output High value after you click the Calculate button.
At Input Value – Enter an input value that you know the required analogue output value for.
There are one of these for the low level and one for the high level.
Required Output Value Is - Enter the required analogue output value for the input value you
have specified. There are one of these for the low level and one for the high level.
Calculate – Click this button to calculate new Output Low and Output High values based on
the information you have entered.

Page 26
Mantracourt Electronics Ltd LCD20 User Manual
Advanced Trimming Tab – Sometimes when the analogue outputs are correct the device you are
feeding the analogue output into does not (because of inaccuracies in that device) read the correct
value. The gain and offset virtual knobs allow you to tweak the gain and offset of the analogue
output to achieve fine control rather than perform measurements and recalculate Output Low and
Output High Values. Also tweaking the gain and offset is the only way to push the analogue output
past the min and max values (0V, 4mA and 10V, 20mA).

Gain – Alter the analogue output gain by clicking and dragging the pointer on the virtual
knob.
Offset – Alter the analogue output offset by clicking and dragging the pointer on the virtual
knob.
Default – Clicking this button will return the gain and offset to default values.

The gain and offset are interactive. You may need to alternate between the adjustments to
achieve the desired effect.

Page 27
Mantracourt Electronics Ltd LCD20 User Manual
Relays
This page allows you to configure the relays.

Source – Select the source of the value that the relays will act upon.
Setpoint – The target value at which the relay operation will function. The actual point at which
contacts may changeover may be affected by Inflight and hysteresis value.
Inflight – Compensates for amount of delivered product still falling even after valves are closed.
Hysteresis – This is used to stop relay chatter and affects the actual point that relay contact
changeover occurs.
Inverted – Whether the operation of the relay is inverted.
Latched – Once contacts have changed this option can be used to latch the relays. A serial
command or digital input is then required to reset the relay contacts. Relays only latch in
deactivated state.
Reset – Click this button to reset any latched relays.

See Also – Appendix C - Relays for more information.

Page 28
Mantracourt Electronics Ltd LCD20 User Manual
Measurement
This page shows how the core measurement values relate to each other.

Calibrated Value – Shows the calibrated value. Calibration turns the input mV/V value into this
value.
System Zero – Enter a value here that is added to the Calibrated Value to produce the Gross
value. System Zero is used to remove permanent values from an installed system such as the weight
of a vessel or weighing platform to give a zero Gross value.
Zero Tracking – Zero tracking is applied after the system zero and compensates for small changes
in the input to allow for a zero Gross value. See Appendix B – Zero Tracking for more information.
Gross – The Gross value of the system.
Net – The Net value of the system is generated from the Gross value plus the Tare value.
Tare – The Tare value is added to the Gross value to generate the Net value.
Peak – Indicates the highest value measured since power up or last Peak/Valley Reset.
Peak Source – Select which source the peak value is based on.
Valley – Indicates the lowest value measured since power up or last Peak/Valley Reset.
Valley Source – Select which source the valley value is based on.
Reset Button – Reset the peak and valley values to the current source value.

Page 29
Mantracourt Electronics Ltd LCD20 User Manual
Snap – Shows the last snap value recorded.
Snap Source – Select which source the snap value is based on.
Snap Button – Perform a snap recording now. (This control is included for demonstrating the
function. Snap would normally be triggered by a digital input to ‘Hold’ a value.
Display – Shows the main value that is indicated on the optional display. Note that the actual value
shown on the LCD may differ from this floating point value due to the Resolution parameter
setting.
Display Source – Select the source for the display value.
LCD – Mimics the actual display that would be shown on the optional LCD display.
Resolution – This parameter value sets the block size that the optional LCD display will increment
and decrement by. This setting does not affect anything other than the LCD display.

Page 30
Mantracourt Electronics Ltd LCD20 User Manual
Digital Inputs
Allows the function of digital inputs to be set and indicates the current state of the inputs.

For each of the three digital inputs the following settings are available:

Action – Select the action to execute when the digital input is activated.
State – Indicates the current state of the digital input.

See Also – Appendix E - Digital Inputs for more information.

Page 31
Mantracourt Electronics Ltd LCD20 User Manual
Calibration
The calibration page offers three ways to calibrate the module and also allows configuration of shunt
calibration.

The Input Sensitivity and Decimal Point parameters should be decided on before calibration
takes place. Changing either after calibration may void that calibration.

Input Sensitivity – Choose the appropriate range for the load cell sensitivity that is connected.
Decimal Point – This setting determines the maximum range of the engineering units for the
module. This must be selected to allow for the maximum value likely to be encountered.
Underrange Value – Enter the engineering unit value of the Gross value below which the
underrange flags will be set. The output value will change to -20001 with the decimal point as
defined above. The optional LCD display will show -1 with no other digits shown.
Overrange Value – Enter the engineering unit value of the Gross value above which the overrange
flags will be set. The output value will change to +20001 with the decimal point as defined above.
The optional LCD display will show +1 with no other digits shown.
mV/V Input – The mV/V value measured at the input.
Calibrated Value – The mV/V value after having the calibration applied.
Units – Optional 6 characters of units that will be displayed on the main Information page.

Page 32
Mantracourt Electronics Ltd LCD20 User Manual
The following sections will deal with each of the tabs available on the calibration page.

Automatic Tab
Select this tab to calibrate the module by applying known loads.

Each time you enter this page it will hold the last data entered or will be blank ready to accept
values. i.e. The page is used to generate the calibration data seen in the Advanced Tab but does
not reflect any data held in the module.

Measurement Points – Enter between 2 and 10 measurement points. These are the points at which
weights will be applied. This process will linearise as it calibrates.
The following pair of controls will be available for the number of measurement points selected.

Applied Weight – Enter the known weight applied at this measurement point. The weights must be
applied in ascending order.
Acquire Button – Once the input has been given enough time to settle click this button to acquire
the point and start calculating calibration values. You can then redo the calibration at this point
again or move on to the next calibration point.
Start Again – Click this button to restart the calibration process.

Page 33
Mantracourt Electronics Ltd LCD20 User Manual
Table Tab
Select this tab if the mV/V input values are known for 2 or more known weights. This can be used to
calibrate based on load cell manufacturer certificates and there is no need to apply actual weights.

Each time you enter this page it will hold the last data entered or will be blank ready to accept
values. i.e. The page is used to generate the calibration data seen in the Advanced Tab but does
not reflect any data held in the module.

Calibration Points – Enter the number of known mV/V and engineering unit value pairs.

The following pair of controls are repeated for each calibration point.

mV/V – Starting from the lowest mV/V value enter the known value for this point.
Engineering Units – Enter the known engineering unit weight for the stated mV/V value.
Calibrate – Click the Calibrate button to calculate the calibration settings based on the information
entered.

Page 34
Mantracourt Electronics Ltd LCD20 User Manual
Advanced Tab
This gives access to the internal gain and offset values. This is considered advanced and care should
be taken in this tab.

The data shown on this page is the actual calibration data held in the module. Use with caution!

Calibration Points – Shows the number of calibration points in the calibration.


The following triplets of controls are repeated for each calibration point.
mV/V – Between this mV/V value and the next the following gain and offset will be used to convert
the value to the Calibrated Value.
Gain – The gain (multiplication factor) applied to the mV/V value.
Offset – The offset added to the mV/V * Gain value to give the Calibrated Value.
Defaults – Click this button to revert all calibration data to factory defaults. The Calibrated Value will
then be mV/V.

Page 35
Mantracourt Electronics Ltd LCD20 User Manual
Shunt Calibration Tab
This tab allows the calibrating of the shunt calibration function. After a load cell has been connected
to the module and all wiring and connectors have been finalised the Shunt Calibration can be
calibrated so that the Shunt Calibration output always indicates 100.00 if nothing has changed on the
load cell or wiring. This value will indicate 100.0 regardless of the weight on the load cell at the time
the shunt calibration is activated. This allows very quick checking of calibration and hardware integrity.

Calibrate – Click this button to calibrate the shunt to the current load cell attached.
Shunt Calibration State – Allows the control of switching on or off the shunt calibration.
Shunt Calibration Value – This is the special shunt calibration value that should read 100.0 when
shunt calibration is activated and the integrity of the calibration and hardware is correct.

Page 36
Mantracourt Electronics Ltd LCD20 User Manual
Cal Info Tab
This tab shows the last date the Toolkit was used to perform calibration and shows the Calibration
Counter parameter that the module increments whenever a calibration related parameter is written to.

Last Calibrated Date – Indicates the last date that calibration was performed by the Toolkit.
Calibration Counter – Shows the internal calibration counter.

Page 37
Mantracourt Electronics Ltd LCD20 User Manual
Filters
This page allows the selection of filter settings that are applied to generate the Gross value.

Rate – Here the basic selection of A/D measurement rate can be selected. The slower rate will have
an improved noise free resolution.
Filtering – This section allows the choice between simple block averaging or digital low pass
filtering. Choosing block averaging will slow down the generation of new values which will be
noticeable on the optional display or via the communications bus.
The digital low pass filter will result in new values being generated at the A/D measurement rate.
Block Averaging – Select the required rate at which new readings are generated. This is in the
form of samples per second or seconds per sample.
Filter Level – If the difference between the newest unfiltered value and the previous filtered
value is greater than this filter level value then the filter is bypassed and the gross will jump
straight to the unfiltered value. If the difference is less than the filter level value then the filter
will be operational.
Filter Steps – Set the fractional part of the difference between the newest and previous values
that is added to the previous value to generate the next gross. i.e. if filter steps is 10 then only
a 10th of the change is applied to the gross value.
Update Rate – This indicates the samples per second or rate at which new values are generated.

Page 38
Mantracourt Electronics Ltd LCD20 User Manual
3db Down Point – When using the digital low pass filter there is an attenuation of higher
frequency components on the input signal. This indicates the frequency at which the output value
would be half of what it would be at a fixed level.

See Also – Appendix B – Measurement | Filters for more information.

Page 39
Mantracourt Electronics Ltd LCD20 User Manual
Installation

Mechanical
The LCD20 is designed to connect to a ‘Top Hat’ DIN rail.

Front Panel
Access to various connectors and buttons is gained by opening the front panel door. This is hinged at
the opposite side to the LEDs.

Page 40
Mantracourt Electronics Ltd LCD20 User Manual
Connections

Load cell Connections


The LCD20 offers a direct connection to a range of low level (foil) strain gauge sensors.

Gain sensitivity is optimised for two ranges, 0.5 to 3.7mV/V and 3.7 to 7.8mV/V selected by the
Sensitivity parameter in the menu.

A stable 5VDC excitation supply at 150mA is provided which is sufficient to power up to four 350 ohm
load cells connected in parallel.

Ratiometric strain gauge measurements are taken which cancel out errors caused by changes or drift
in Excitation voltage. The LCD20 facilitates 6 wire measurement which allows for voltage drops caused
by long cable lengths between the LCD20 and the load cells, where barriers are used or if the load
cells are powered from another source. Where possible, use 6-wire connections to the load cells.

Page 41
Mantracourt Electronics Ltd LCD20 User Manual
For load cell connections we recommend the use of 3 pairs of twin twisted pair cable such as Alpha
Wire 5123C SL005 available from Farnell 130-2249.

The pairs should be arranged as follows:


Excitation pins 1 & 6
Sense pins 2 & 5
Signal pins 3 & 4

Terminate all screens at ‘SCR’. The screens should not be connected at the transducer end of
the cables. However, if the transducer has an EMC gland then the screen should be connected
there and not at ‘SCR’ on the LCD20.

Connector Location

4 Wire
If the load cell is supplied with a 4 wire cable and long cable lengths are required it is recommended
that any extension cabling required is 6 wire to compensate for voltage drops along the cable.
Once a 4-wire system is calibrated any changes to the cable length will require that the system is re-
calibrated.

Please refer to load cell manufacturer’s instructions.

Page 42
Mantracourt Electronics Ltd LCD20 User Manual
When 4-wire connectivity is used it is important that the respective Excitation and Sense
connections are linked together.

6 Wire
Where load cells are provided with 6 wires then it is recommended to continue the 6 wires
connectivity throughout the system.

Page 43
Mantracourt Electronics Ltd LCD20 User Manual
Barriers
The LCD20 can be used with isolating barriers. This allows the load cells to be situated in the
hazardous area whilst the LCB is in the safe area.
6 wire connectivity must be used.

The example below uses Pepperl + Fuchs isolated barriers.

KFD2-VR2-Ex1.50M is used to isolate the Output signal from the load cell.
KFD2-VR-Ex1.18 is used to isolate the Sense connections.
KFD0-SD2-EX1.1065 used to provide excitation for up to 4 350 ohm load cells.

Because the barriers are fully isolated, the REF connection MUST be made to ensure that the
LCD20 input stage is biased correctly.

If galvanic isolation is NOT required, passive non-isolated barriers such as the MTL77xx series
can be used instead. In this case the REF connection should NOT be made.

Page 44
Mantracourt Electronics Ltd LCD20 User Manual
Digital Inputs
The digital inputs are for switch contacts or volt free transducers such as relays. Another common type
that is suitable for the LCD20 digital inputs is a NPN open-collector transistor.

The digital input internal circuitry has 1k ohm pull-up with a 300ms de-bounce filter.
The common of the digital input marked ‘0V’ has a 47 ohm series resistor.

The Digital Inputs are not isolated from the load cell input

Do not apply voltages to these digital inputs

Connector Location

Connections

Page 45
Mantracourt Electronics Ltd LCD20 User Manual
Analogue Output
There are two analogue output available on the LCD20, a 0-10V voltage and a 4-20mA current. They
are both driven from a single Pulse Width Modulated (PWM) source meaning they track each other
and use the same scaling factors.

The analogue outputs are isolated from the load cells, digital inputs and power source but are
common to each other.

We recommend using a cable with a screened twisted pair such as ALPHA WIRE - 5121C available
from Farnell stock code 130-2298.

The screen should be connected at either the LCD20 or the remote device but not both thus avoiding
ground loops.

Connector Locations

Page 46
Mantracourt Electronics Ltd LCD20 User Manual
Voltage
For voltage output the cable lengths should be kept to less than 20m. Voltage outputs are more
susceptible to electrical interference so where long cable lengths are required the current output
should be used.
If the screen is to be connected at the LCD20 then it should connect to the terminal marked SCR.

Minimum load impedance for the voltage output is 5k ohms

Current
The current is suitable for driving long cable lengths and in areas where higher electrical interference
is likely.
If the screen is to be connected at the LCD20 then it should connect to the terminal marked SCR.

The maximum load impedance for the current output is 1k ohm

Page 47
Mantracourt Electronics Ltd LCD20 User Manual
Relays
The Relays provide for 3 A switching at 240 VAC or 3 A switching at 30 Vdc. Suitable cable for the
installation should be selected.

Care should be taken as any connections already made by the installer may be at mains
potential.

Contact Connections

The LEDs marked SP1 and SP2 will illuminate when the relay is energised.

Page 48
Mantracourt Electronics Ltd LCD20 User Manual
Programmer Module

LP2
The handheld programmer comes with 2m of cable. Power should be switched off before the
connector is inserted or removed.

Page 49
Mantracourt Electronics Ltd LCD20 User Manual
Programming Cable PGM1
For communicating with the module using the LC Toolkit for configuration or firmware updates using
the PGM1 programming cable.
The programming cable plugs into a 6 way row of pins above the LEDs on the LCD20 module and the
other end into the PC USB socket.

Please ensure correct orientation as marked by colour coding above.

This will appear as a virtual RS232 port (COM port) on the PC.

The PGM1 cable is available from Mantracourt or can be purchased from Farnell or RS. Please see the
following information:

Manufacturer: FTDI Chip. http:/ftdichip.com Part number TTL-232R-5V


Supplier: RS 687-7770
Driver: http://www.ftdichip.com/Drivers/VCP.htm (Select the Virtual COM Port Drivers)

A driver may need to be installed for the PGM1 cable to work correctly. This will create a virtual
RS232 serial port (COM Port) which your own software or the LC Toolkit will connect to.

Page 50
Mantracourt Electronics Ltd LCD20 User Manual
Driver Settings
If you are using the communications to read or log data at high speed (This includes viewing the
Trend Chart using the LCD20 Toolkit) you may find the desired speed is unachievable even with high
baud rates. This could be due to the default settings of the Latency Timer. Open the Control Panel
and launch the Device Manager.
Once Device Manager is open you need to locate the virtual COM port assigned to this interface. This
will be listed under the Ports (COM & LPT) section of the hardware tree. You may need to unplug
and re-attach the interface cable from the computer to identify which entry to use.

Double click the item to open the Serial Port Properties dialog window.

Click the Advanced button to display the Advanced Settings

Page 51
Mantracourt Electronics Ltd LCD20 User Manual
You need to ensure that the Latency Timer setting is set to 1.

Page 52
Mantracourt Electronics Ltd LCD20 User Manual
LED Indicator
Status LED
The red LED is used to indicate a number of functions.

 Under normal healthy operating conditions the red LED should be OFF flashing ON every
second.
 If the LED is ON Flashing OFF every second then this indicates that an error condition exists
and the reasons for this are as follows:
o There is a Load cell integrity problem
o The Shunt-calibration has been activated.
o An Over-range or under-range condition exists.
o The unit has had a ‘watch-dog’ reset. Cleared only by power cycling.
 The LED is permanently ON indicates the unit is either attempting to connect with the toolkit
or already in communications with the Toolkit.
 Led Flashing approximately 10 times a second indicates the unit is in boot-loader mode either
waiting to be re-flashed or because an incorrect or no Flash file exists in the processor.

See Also – Using The LC Toolkit Software | Save & Restore for more information
regarding updating the firmware.

Page 53
Mantracourt Electronics Ltd LCD20 User Manual
Detect Switch
This switch is used to connect to the LC Toolkit without the need to power cycle. See Using The LC
Toolkit Software section for operation. The LC Toolkit software searches for modules connected to the
selected serial port. Push the switch for 1 second after selecting ‘Detect’ in the LC Toolkit.

The button needs to be held in for at least 1 second.

Page 54
Mantracourt Electronics Ltd LCD20 User Manual
Appendix A – Calibration

Best practices
 Before calibrating the LCD20 it is advised that the device be powered on for 20 minutes to
allow the electronics and load cells to stabilise.
 Set the Sensitivity and Decimal Point Parameters before calibrating
 For multi-point linearization / calibration we recommend the use of the toolkit as this
simplifies the process.
 The best calibration is achieved by using the complete system using actual known weights.
 When using the multi-point linearization load the complete system in increments that allow a
graph of non-linearity to be plotted. From this graph the best point can be determined for the
calibration points.
 When multiple load cells are connected together in parallel ensure that they have equal mV/V
outputs. Alternatively use an Active Junction module such as a JBA.

Table Calibration Method


If the load cell manufacturer’s calibration certificate is available then it is possible to calculate gains
and offsets that can be entered into the linearization table parameters. A simple way of achieving this
is provided by the LC Toolkit.
The LC Toolkit’s Calibration Advanced Tab is where you can view the actual linearization table
parameters from either the Auto-Cal or Table calibration methods.

Shunt Calibration
The LCD20 is fitted with a 100k ohm resistor which, when commanded from a digital input or over the
communications port, is placed across the Load Cell Negative Input and the Excitation Negative. This
has the effect of changing the mV/V by a fixed amount in the positive direction.

The change in mV/V can be calculated as follows

ΔmV/V = 1000Rg / (4Rs + 2Rg)

Where Rg is the Load Cell impedance


And Rs is the Shunt Calibration Resistor impedance.

The Shunt Calibration is for indication and integrity purposes and should not be used to
perform calibrations from.

Page 55
Mantracourt Electronics Ltd LCD20 User Manual
Using Shunt Calibration
The Shunt Calibration facility allows a known value to be displayed thus allowing the integrity of the
LCD20, load cell wiring and load cell to be checked. The load cell input should be constant before and
during this Integrity check.
The Shunt Calibration can be switched on, off or toggled by a digital input. It can also be commanded
on and off from the communications port. When Shunt Calibration is switched on the Status
parameter will flag this and also flag that a Load Cell Integrity condition exists. The LCD display will
indicate this by the PGM Arrow segment being on.

When the Shunt Calibration is commanded to switch on; but before the Shunt Calibration resistor is
placed across the load cell; the current mV/V value is recorded in a Pre Shunt Calibration Value
parameter. This is used when calculating the Shunt Calibration Scaling factor parameter see following.

 
 
 Required known value 
Shunt Calibration Scaling Parameter    * 10
Post Shunt Calibration mV/V -
 
 Pre Shunt Calibration Value parameter 
 

Note: When Shunt Calibration is switched on the LCD20 values such as Gross, Net, Peak, Valley ETC
are also affected. Consequently the Analogue out and relays are also affected.

Page 56
Mantracourt Electronics Ltd LCD20 User Manual
Appendix B – Measurement

Overview

Page 57
Mantracourt Electronics Ltd LCD20 User Manual
1. The measurement process starts by converting the load cell analogue values to digital values.
The Analogue to Digital Converter (ADC) has a bipolar input allowing for load cells to be used
in both tension and compression.
The ADC measures the 5V excitation voltage using the sense inputs and through an amplifier
the output of the load cells from these two measurements the mV/V output from the load cells
is calculated.
This amplifier has two gain settings allowing for two ranges, ±3.7mV/V and ±7.8mV/V. These
gains setting are selected by the Sensitivity parameter.
The ADC can also take measurements at two different rates, 10 samples per second (sps) and

Page 58
Mantracourt Electronics Ltd LCD20 User Manual
80sps, selected by the Rate parameter. The output from the ADC is a signed 20bit result in
ADC counts terms.

2. When a new ADC result is available; which is determined by the Rate parameter; the process is
checked to see if block averaging is required. This is determined by the value set in the Display
Averaging parameter and the digital low pass filter parameter Filter Steps. The block averaging
is bypassed if Display Averaging parameter is set to 7 or the Filter Steps parameters is non
zero. Block averaging simply adds a number of results together then divide the result by the
number of results added (known as the mean). For Display Averaging values of 0-6 the number
of result to average is calculated as….

2 to the power of (Display Average + 2)

Once a new block average value is ready the Filtered ADC Counts value is updated and the
process continues.

3. The Filtered ADC Counts Value is then converted using the factory calibration to a mV/V value
to within 0.05% accuracy.

4. This mV/V value is then tested to see if in fact the input has been over or under ranged and if
so an internal input over or under range flag is set that is checked later in the process (see 8).

5. The current mV/V value is then converted into the user desired engineering units using the 9
point linearization programme. This programme searches the mV/V levels for each point to
find what Gain and Offset values to apply. The number of points to search through is set by
the Number of Calibration Points parameter. If the current mV/V is above or below the last or
first mV/V table level then the gain and offset values to be used are extrapolated from the last
or first two points.
When the 2 point Auto-Calibration routine is used the same principle applies. The auto-cal
calculates the values for the first two points in the table using the mV/V values captured for
Calibration Low and Calibration High parameters.
The output from this process is termed the Calibrated Value.

6. The System Zero parameter is then added to the Calibrated value to give an intermediate
internal value known as Pre-Gross Value.

7. If block average has not been selected as determined in (2) above and the Filter Steps
parameter is non-zero then the Low Pass Filter algorithm is implemented. This is covered in
Appendix B – Measurement | Filters. The Pre-Gross value is updated with the result of the Low
Pass Filter.

Page 59
Mantracourt Electronics Ltd LCD20 User Manual
8. The Pre Gross Value is checked against the Over and Under Range parameters and if exceeded
or if the internal input over or under range flag is set then the Status and Flag Parameter is
updated accordingly. The Flag parameter being a non-volatile latched flag requiring user
intervention using the communications port to be reset.
The Gross and Net values are set to over or under range values also. This ensures that the
display, relays, analogue output and communication ports all detect and recognise this error
condition.

9. Pre Gross Value is now adjusted with the Zero Tracking value. Zero Tracking is enabled when
the Zero Tracking Band parameter is non-zero.

10. Net is calculated by adding the Auto-Tare parameter to the Gross.

11. The Peak value is checked against the source selected by the Peak Selection parameter which
could select either Gross or Net values.

12. The Valley value is checked against the source selected by the Valley Selection parameter
which could select either Gross or Net values.

Zero Tracking
Zero tracking is applied to the Gross value.
A Tracking Band parameter sets a band about the Pre Gross absolute zero value. If a Tracking Band
parameter has a value of 10.0 then the band is calculated as between + 10.0 and -10.0. If the Pre
Gross value is in between these two calculated values for more than 1 second then the value of Pre
Gross is stored as the new zero tracking value.
This Zero Tracking value is then subtracted from the Pre Gross value to give a Gross value of zero.
Zero Tracking can be disabled by setting the Tracking Band Parameter to zero. The Zero tracking value
is volatile therefore reset to zero on power up.

System Zero
System Zero is used to remove permanent values from an installed system such as the weight of a
vessel or weighing platform to give a zero Gross value.
The System Zero parameter is then added to the Calibrated value to give an intermediate internal
value known as Pre-Gross Value.

Auto Tare
The Auto Tare facility enables the generation of the Net value. The Net Value is calculated as the
Gross value added to the value set in the Auto Tare parameter. An update of the Auto Tare value can
be commanded from a digital input or from the communications port. Additionally an Auto Tare value

Page 60
Mantracourt Electronics Ltd LCD20 User Manual
can be entered into this parameter from the keypad or again using the communication port. The Auto
Tare parameter is non-volatile and is therefore saved through power fail.

Filters
Overview
The least significant digits of the display or values over the communications port may change
excessively in certain applications due to vibration or ‘noise’ on the input. Filtering can be applied to
smooth the display by two methods.
The simple block average filter uses averaging and therefore generates new values at the
Measurement Rate which is slower than the A/D Measurement Rate.
The dynamic digital low pass filter uses software to generate new values at the A/D Measurement
Rate.

Simple Block Averaging


Block averaging delays the generation of a new output value until a number of new values generated
at the A/D Measurement Rate have been averaged. This provides a very steady display or output but
at the expense of a slow update rate. The range of choices of the final update rate is dependent on
the selected A/D Measurement Rate.

A/D Measurement Rate is set to 10 samples A/D Measurement Rate is set to 80 samples
per second per second
10 samples per second 80 samples per second
2.5 samples per second 20 samples per second
1.25 samples per second 10 samples per second
1.6 seconds per sample 5 samples per second
3.2 seconds per sample 2.5 samples per second
6.4 seconds per sample 1.25 samples per second
12.8 seconds per sample 1.6 seconds per sample
25.6 seconds per sample 3.2 seconds per sample

Dynamic Digital Low Pass Filter


The dynamic filter is a digital low pass filter implemented in software which behaves like an electronic
‘RC’ circuit. It has two user settings, a Filter Level set in the calibrated engineering units and the
maximum number of steps, Filter Steps 1 to 255. To bypass the dynamic filter set Filter Steps to 0.

Instead of outputting every new value, a fraction of the difference between the new input value and
the current filtered value is added to the current filtered value to produce the filtering action.

If this difference is less than the value set in Filter Level then the fractional amount added each time
is decremented until it reaches the minimum level set by Filter Steps i.e. Filter Steps is the limit of
the divisor.

Page 61
Mantracourt Electronics Ltd LCD20 User Manual
Example: if Filter Steps = 4, initially the whole difference between the new value and the current
filtered value is added to the current filtered value. The next value displayed is the result of adding
one half of the difference. The following update adds one third while subsequent updates continue to
add one quarter of the difference, the limit of the divisor having been reached
i.e. 1/ Filter Steps = 1/4.
If Filter Steps were increased to 10, the process would loop for ten times then continue to add just
10% of the difference resulting in a longer settling time but deeper filtering effect.

If a rapid or step change in the input causes the difference between the new input value and the
current filtered value to exceed the amount set in Filter Level, the filter acquires the new value
immediately by resetting the divisor to 1. The filtering process repeats as above on subsequent
updates by adding one half of the difference, one third, one quarter etc. until 1/ Filter Steps is
reached. This allows the Filter to respond rapidly to fast moving input signals but still retain its
smoothing characteristic as the input settles.

When a step change occurs which does not exceed Filter Level, the new filtered value is calculated as
follows:
 InputValue  Current Filter OutputValue  
New Filter OutputValue  Current Filter OutputValue   
 Filter Steps 

Note: To retain the filtering effect but defeat the step change response set Filter Steps appropriately
and Filter Level to 0.

The time taken to reach 63% of a step change input (which is less than Filter Level) depends on the
rate at which values are passed to the dynamic filter, set in A/D Measurement Rate (10 or 80 samples
per second (SPS)), multiplied by Filter Steps.

The table below gives an indication of the response to a step input which is less than Filter Level.

% of Final Value Time To settle


63% Measurement Period * Filter Steps
99% Measurement Period * Filter Steps * 5
99.9% Measurement Period * Filter Steps * 7

For example, if A/D Measurement Rate is set to 80 samples per second (12.5ms period) and Filter
Steps is set to 30 then the time taken to reach a percentage of the step change value is as follows:

% Of Final Value Time To settle


63% 0.0125 x 30 = 0.375 seconds
99% 0.0125 x 30 x 5 = 1.875 seconds
99.9% 0.0125 x 30 x 7 = 2.625 seconds

Frequency Response:

Page 62
Mantracourt Electronics Ltd LCD20 User Manual
For a sinusoidal input the -3dB point, i.e. the frequency where the output is 0.71 of the input is given
by:
  SampleRate( SPS)  1 
Frequency      
  Filter Steps  6.3 

To calculate the number of steps for a -3dB frequency point use:

 SampleRate( SPS) 
Filter Steps   
 Frequency  6.3  

Page 63
Mantracourt Electronics Ltd LCD20 User Manual
Peak and Valley Values
The LCD20 incorporates a Peak/Valley function which continuously monitors the Gross or Net value,
selected by the Peak and Valley Selection Parameter, and stores the minimum (valley) and maximum
(peak) values. These values are volatile and reset on power up.

Peak and Valley values can be used as the default display value, analogue output and relay source
values. That is to say, the display can be set to show Peak value, Relay 1 source value could be Peak,
Relay 2 could be Valley and the analogue output source value also set for Peak.
Peak and Valley values can also be selected to view by using the digital inputs. For example Digital
input 1 could be used to display Gross value, Digital input 2 to display Peak value.
Peak and Valley values can be reset from a digital input or from the LC Toolkit.

Page 64
Mantracourt Electronics Ltd LCD20 User Manual
Appendix C – Relays

Overview
The relays provide output control signals which can be used for switching functions such as ON/OFF
control and alarm indications. The relays are activated by the values programmed for the Set Points.

Relay Functions
The LCD20 can be programmed so that the relay output module reacts to all or any of the following
functions:
 Source for relay such as Gross, Net, Peak, Valley and Snap
 Setpoints
 Inflight compensation
 Hysteresis
 Relay inversion
 Relay latching

The two relays provide Single Pole Change-Over (SPCO) contacts and are rated at 240 VAC at 3 A or
30 Vdc at 3 A.

Source
For LCD20 mode the source value for each relay can be selected from the following values; Net, Gross,
Peak, Valley and Snap.

Setpoints
Setpoints are used to produce relay signals at any required value so that the operation of the
monitored process can be maintained to pre-set levels. Any excursion beyond the Setpoint(s) will
normally activate the relay or relays to provide an alarm or initiate control as required.

Two relays can be programmed to suit different applications. The actions of either or both relays can
be inverted if required.
For normal operation the relay is energised until the input reaches the Setpoint level. In this
condition, when the input value is less than the Setpoint, the setpoint LED is on and the output relay
is energised producing a closed circuit on a normally open contact. When the Setpoint value is
reached, the setpoint LED turns off and the relay is de-energised producing an open circuit on a
normally open contact.
For an Inverted operation the reverse operations apply.
Normal and inverted action is determined by the direction of the input value as it changes.

A High-High operation allows for a rising input value to operate on two relays to define an acceptable
quantity, weight or band of operation.

Page 65
Mantracourt Electronics Ltd LCD20 User Manual
A Low-Low operation operates on a falling value.

A High-Low operation will operate on a rising or falling value, setting a 'band' by one relay operating
normally and the other being inverted.

Inflights
The setting of an Inflight value causes the relays to automatically adjust to control the flow of the
material being weighed.
For example, if a relay is used to control a flow, a certain amount of product will be 'Inflight' between
the supply point and receiving point causing a positive error when the required weight is reached. The
Inflight compensation value is adjusted by the user to reduce the Setpoint value to prematurely stop
the flow, allowing the Inflight amount to make up the required total set by the associated Setpoint.

Hysteresis
When Hysteresis is applied to relays with normal output action, the input is allowed to rise to the
Setpoint value and the relay is then turned off. The relay is held off until the input value has dropped
to the Setpoint minus the Hysteresis value.
For inverted action the input drops to the Setpoint and the relay goes off and comes on again when
the input rises to the Setpoint plus the Hysteresis value.

For LCA15 mode Hysteresis will be applied to both Setpoint values. It is effective for both normal
and Inverted action.

Inversion
A normal, non-inverted relay is defined as being in the energised state when the source value is below
the resultant setpoint.

An inverted relay is defined as being in the energised state when the source value is above the
resultant setpoint.

Note the SP1 and SP2 LED’s visible on the front panel denote whether a relay is energised or de-
energised. The LED is ON for an energised relay.

Latching
The latching facility allows the relay module output to be held until reset externally via the digital
input or via the communications port. Latching is applied to the ‘off’ status of the relay.
Latched relays are reset by operating a closing contact connected between 'LR' terminal and '0' volts,
on the field terminals (this assumes that Digital Input 2 is set to its default function ‘LR’

See Also – Appendix E - Digital Inputs for more information.

Page 66
Mantracourt Electronics Ltd LCD20 User Manual
Appendix D – Analogue Output

Overview
Two analogue outputs are available offering a DC current range (4 to 20mA) and a DC voltage range
(0 to 10V).
They are fully scalable, optically isolated and both generated from a single selected variable (see
source below)
Both outputs are pre-calibrated to an accuracy of 0.15% FR and have a resolution of approximately 13
bits or 1 part in 8000.

WARNING:
Maximum load on voltage modules is 2mA
Maximum drive voltage available for the current range is 15V

Source
The value the selected analogue output, 4-20mA or 0-10V, calculates its output from can be sourced
from Net, Gross, Peak, Valley or Snap values.

Scaling
Output scaling factors are set by the user and determine the display range over which the analogue
module operates.
Output Low - This sets the displayed value at the analogue outputs minimum i.e. 4mA or 0V.
Output High - This sets the displayed value at analogue outputs maximum output i.e. 20mA or 10V

If the display is outside the range defined by Output Low and Output High scale factors the
analogue output will remain constant at its minimum or maximum output value.

So for example, if the 4-20mA output is selected and Output Low scale factor is set to 00.00 whilst
the Output High scale factor is set to 100.00 then if the source value is at 00.00 or less the 4-20mA
output will be at 4mA level and consequently if the source value is 100.00 or greater the 4-20mA
output would be at 20mA.

Inversion of the analogue output can be set by the Output Action parameter or transposing Output
Low and Output High scale factors.

Calculating the Scale Factors


Example: Assume a 4-20mA output module is required to provide an output of 4mA for 1000Kg and
20mA for 6500Kg.
Set Output Low to 1000 and Output High to 6500

Page 67
Mantracourt Electronics Ltd LCD20 User Manual
It will be necessary to determine Output Low and Output High scale factors by graphical or
mathematical means if the known source values do not coincide with the minimum and/or maximum
analogue output.

Method of Calculating Output Low & Output High from known output values

 Display Span * ( Low Output  Min Output) 


Output Low  Low Display   
 ( High Output  Low Output) 

 Display Span * ( Max Output  High Output) 


Output High  High Display   

 ( High Output  Low Output) 
Where:
Low Output Required analogue output at low display value
High Output Required analogue output at high display value
Min Output Lowest value of analogue output (4mA or 0V)
Max Output Highest value of analogue output (20mA or 10V)
Display Span Required High display value minus required Low display value.

Example
To scale the output from 6mA at a display value of 400 to 18mA at a display value of 1100:

Display Span  1100  400   700

 700 * ( 6  4 ) 
Output Low  400   
 (18  6 ) 
1400 
Output Low  400     ( 400  116 .66 )
 12 
Output Low  283 .33

Page 68
Mantracourt Electronics Ltd LCD20 User Manual
 700 * ( 20  18 ) 
OutputHigh  1100   
 (18  6 ) 
1400 
OutputHigh  1100     (1100  116 .66 )
 12 
OutputHigh  1216 .66

Note 1: The output can be inverted by setting Output Low more positive than Output High.
Note 2: If Output Low or Output High is greater than ± 19999 then divide both Output Low and
Output High by 10, this will give less resolution.

User Adjustment
The LCD20 has no Potentiometers for trimming the 4-20mA and 0-10V outputs as did its predecessor.
Instead the output can be ‘trimmed’ digitally using parameters that can adjust both the 4-20mA and
the 0-10V outputs for gain and offset. This is useful for when the input of the device that the LCD20 is
connected to is not accurately calibrated itself or, in the case of the voltage output, to compensate for
small voltage drops in the cabling.

As an example, setting the 4-20mA and 0-10V Gain parameters to 0.5 re-scales the analogue outputs
to 4-12mA and 0-5V respectively.

An offset can be introduced by entering a value for 4-20mA and 0-10V Offset parameters with one
digit nominally representing 2.03uA and 1.254mV respectively.

The factory calibration can be re-instated by resetting the offsets to 0 and the gains to 1.

Page 69
Mantracourt Electronics Ltd LCD20 User Manual
Appendix E – Digital Inputs

Overview
The LCD20 has 2 digital inputs. These inputs are volt free which require either a contact closure or
open collector type output which connects the particular input to the terminal marked ‘0V’. The input
has an internal 1000 Ohm resistor pulling the input to +5Vdc. The input requires to be closed for a
minimum of 300mS due to the de-bounce circuitry. Note that these inputs are not isolated from the
load cell input.

WARNING - Do not apply voltages to the digital inputs.

The two digital inputs can be individually configured to provide the following functionality.

Digital Input selection table

Description Notes
Perform Tare Perform a tare action.
Latch Relay Reset Reset any latched relays.
Display Default display Switch the display to the original display source value.
Display Gross value Switch the display to Gross value.
Display Net value Switch the display to Net value.
Display Peak value Switch the display to Peak value.
Display Valley value Switch the display to Valley value.
Display Snap value Switch the display to Snap value.
Peak & Valley Reset Reset the peak and valley capture values.
Shunt-Cal On Turn Shunt calibration on. The Shunt Calibration Value available
over the communications bus and optional display will show 100 if
integrity of load cell and wiring is OK and the Shunt Calibration
value has been calibrated. See Using the LCA Toolkit Software |
Calibration | Shunt Calibration section of the manual for more
information.
Shunt-Cal Off Turn off shunt calibration.
Shunt-Cal Toggle on/off Toggle between shunt calibration on and off.
Do Snap Store the current value of the snap source to the snap value.

Page 70
Mantracourt Electronics Ltd LCD20 User Manual
Appendix F – Security Locking
By default the LCD20 arrives from the factory unlocked in terms of requiring changes to hardware to
gain access to certain functionality.

Menu Passwords
Access to menu sections using the wired handheld programmer/display is controlled by passwords.
The user is asked for a password and depending on the password entered gains access to the
appropriate section and all sections beneath.
These passwords are fixed as follows:

Level 1 Password 1111 Operator


Level 2 Password 2111 Engineer
Level 3 Password 3111 Advanced Calibration /
Configure

Handheld/Display Calibration Lock


There is also a hardware link on the main board that if removed will hide the CALL, CALH, ADCL, ADCH
and SENS parameters from the handheld menu and thus prevent calibration.

LK1 Jumper Link On Calibration Visible


LK1 Jumper Link Off Calibration Hidden

Page 71
Mantracourt Electronics Ltd LCD20 User Manual
LC Toolkit Lock
LK2 on the main board controls access to the Lock page in the Toolkit. When the jumper link is fitted
all pages are visible including this Lock Page.

Certain pages and functionality can be hidden or disabled using this page. Once the LK2 jumper link is
removed and the module is connected to the LCA Toolkit this page and all other selected pages and
functions will be hidden.

LK2 Jumper Link On All Pages Visible


LK2 Jumper Link Off Lock Page & Selected Pages Hidden

Page 72
Mantracourt Electronics Ltd LCD20 User Manual
Specifications

Electronics
Loadcell Input
Physical Parameter Minimum Typical Maximum Units
Strain Gauge Excitation System 6 Wire
Strain Gauge Excitation Voltage 4.75 5.0 5.25 V
Strain Gauge Drive Capability 85 ohms
Strain Gauge Sensitivity Range 1 ±3.7 mV/V
Strain Gauge Sensitivity Range 2 ±7.8 mV/V
Offset Stability @ 2.5mv/V FS ±2 ppm/°C
Gain Stability ±4 ppm/°C
Offset Stability With Time ±3 ppm FR
st
Gain Stability With Time 1 Year ±6 ppm FR
Integral Non Linearity Before ±10 ppm
Linearisation
Internal Resolution 1:1000,000/20 Resolution/Bits
10hz Noise Free Resolution 1:180,000/17.5 Resolution/Bits
80hz Noise Free Resolution 1:130,000/17 Resolution/Bits

Analogue Output
Physical Parameter Minimum Typical Maximum Units
Current Output 4 20 mA
Current Output Drive Capability 1 K Ohm
Voltage Output 0 10 V
Voltage Output Drive Capability 5 K Ohm
Phase Delay – Sine Wave 55 ms
3db Down Point 8.5 Hz
Phase Delay – Step Change 26 ms
Rise/Fall Time (10% To 90% FR) 40 ms
Rise/Fall Time (1% To 99% FR) 85 ms
Zero Stability 0.005 0.01 %FS/°C
Gain Stability 0.005 0.01 %FS/°C
Resolution 1:8000/13 Resolution/Bits
Isolation ±130V RMS or DC to any other port

Above specification is based on a Measurement Rate of 80 samples per second and no averaging or
filtering.

Page 73
Mantracourt Electronics Ltd LCD20 User Manual
Power Supply
Physical Parameter Minimum Typical Maximum Units
Supply Voltage (DC) Isolated 9 32 V DC
Power 3 W

Relays
Physical Parameter Minimum Typical Maximum Units
Contact Type 2 X SPNO
Contact Rating 240V AC @ 3A or 30 Vdc @ 3 A

Environmental
Physical Parameter Minimum Typical Maximum Units
Operating temperature range -20 +50 °C
Storage temperature -20 +70 °C
Humidity 0 95 %RH
Protection IP20

Page 74
Mantracourt Electronics Ltd LCD20 User Manual
Regulatory

CE Approvals
European EMC Directive
2014/30/EU
BS EN 61326-1:2013
BS EN 61326-2-3:2013

Low Voltage Directive


2006/95/EC
BS EN 61010-1:2001 Rated for Basic Insulation, Normal Condition, Pollution Degree 2, Permanently Connected,
Insulation Category lll

Page 75
Mantracourt Electronics Ltd LCD20 User Manual
Troubleshooting
This chapter is designed to assist in the identification of problems relating to the installation and
setting up of the LCD20.

General connection and setup parameters. – No flashing status LED on power up.
 Check supply is present at the LCD20 terminals.
 Remove all other connections to the LCD20 except the power
 If supply is correct contact distributor.

Display shows (-1 or 1) continually, without a weight applied to the Strain Gauge.
 Check input connections to the LCD20 from the Strain Gauge.
 If connecting a 4 wire device ensure terminals 1&2 and 5&6 are linked.
 Check Strain Gauge output between input terminals 3&4 of the LCD20.
 Check that the CALH weight is applied and is not the same or lower than CALL.

Display over ranges (-1 or 1) when, or before, the maximum required weight is applied to the Strain
Gauge.
 Check that the Sensitivity parameter is set to the appropriate range in the menu.

Display very noisy


 If using a 4 wire device ensure terminals 1&2 & 5&6 are linked.
 Check output voltage of Strain Gauge.

Display operating in wrong direction


 Check connections to input terminals 3&4 are correct way round.
 Check the type of Strain Gauge - compression or tension.

Unit will not auto calibrate


 Check that CALH is not zero and its weight is greater than CALL.
 Check that input is not over-ranged on CALH weight.

Page 76
Mantracourt Electronics Ltd LCD20 User Manual
Incorrect relay operation
 Check setpoint, in flight and hysteresis values are correct.
 Check latching and inversion settings are correct.
 Check connections to output terminals.

Page 77
Mantracourt Electronics Ltd LCD20 User Manual
Order Codes

Description Code
Load Cell Amplifier In DIN Rail Format LCD20

Accessories
Description Code
Programming Cable (For Toolkit Software) PGM1
Handheld Programmer LP2

Page 78
Mantracourt Electronics Ltd LCD20 User Manual
Configuration Record Sheet

Product
Product Code
Serial No.
Tag No.
Date
Location
Measurement type, range & engineering units
Calibration Date
Calibration Interval

Parameter Value
PASS
tYPE
VALS
GroS
nEt
PEAK
VALY
SNAP
mVV
CALV
rLYS
rLY1
SP1
IF1
HYS1
InV1
Ltc1
rLS1
rLY2
SP2
IF2
HYS2
InV2
Ltc2
rLS2

Page 79
Mantracourt Electronics Ltd LCD20 User Manual
Parameter Value
AnOP
OPL
OPH
AnOP
AoPS
InV
Adj
4 20G
4 20o
0 10G
0 10o
diSP
dA
dP
rS
ddiS
PEAk
VALY
SNAP
OVrV
UndV
LCiP
FSTP
FLVL
SEnS
rATE
ACAL
AdCL
AdCH
CALL
CALH
AT
trbd
ZEro
diPS
dIP1
dIP2

Page 80
Mantracourt Electronics Ltd LCD20 User Manual
Warranty
All LC products from Mantracourt Electronics Ltd., ('Mantracourt') are warranted against defective
material and workmanship for a period of three (3) years from the date of dispatch.
If the 'Mantracourt' product you purchase appears to have a defect in material or workmanship or fails
during normal use within the period, please contact your Distributor, who will assist you in resolving
the problem. If it is necessary to return the product to 'Mantracourt' please include a note stating
name, company, address, phone number and a detailed description of the problem. Also, please
indicate if it is a warranty repair.
The sender is responsible for shipping charges, freight insurance and proper packaging to prevent
breakage in transit.
'Mantracourt' warranty does not apply to defects resulting from action of the buyer such as
mishandling, improper interfacing, operation outside of design limits, improper repair or unauthorised
modification.
No other warranties are expressed or implied. 'Mantracourt' specifically disclaims any implied
warranties of merchantability or fitness for a specific purpose. The remedies outlined above are the
buyer’s only remedies. 'Mantracourt' will not be liable for direct, indirect, special, incidental or
consequential damages whether based on the contract, tort or other legal theory.

Any corrective maintenance required after the warranty period should be performed by 'Mantracourt'
approved personnel only.

Page 81
Mantracourt Electronics Ltd LCD20 User Manual
Document Title: LCD20 User Manual
Applies To: LCD20
Part Number: 517-939
Issue Number: 1.0b
Dated: 9th August 2016

In the interests of continued product development, Mantracourt Electronics


Limited reserves the right to alter product specifications without prior notice.

www.mantracourt.com

Page 82
Mantracourt Electronics Ltd LCD20 User Manual