07.DC-3 Service Manual v1.0
07.DC-3 Service Manual v1.0
07.DC-3 Service Manual v1.0
Service Manual
© 2008 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All Rights Reserved.
Product Information
Product Name: Diagnostic Ultrasound System
Issued date of this manual: 2008-05
Version: 1.0
other countries. All other trademarks that appear in this manual are used only for editorial
purposes without the intention of improperly using them. They are the property of their
respective owners.
I-1
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
l All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
l The electrical installation of the relevant room complies with the applicable national and
local requirements;
l The product is used in accordance with the instructions for use.
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
l Any Mindray product which has been subjected to misuse, negligence or accident;
l Any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
l Any products of any other manufacturers.
I-2
Return Policy
Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
1. Return authorization: Contact the Customer Service Department and obtain a Customer
Service Authorization number. This number must appear on the outside of the shipping
container. Returned shipments will not be accepted if the number is not clearly visible.
Please provide the model number, serial number, and a brief description of the reason
for return.
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges).
3. Return address: Please send the part(s) or system to the address offered by the
Customer Service Department.
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, ShenZhen 518057, P. R. China
Tel: +86 755 26582888
I-3
Safety Precautions
CAUTION are signal words used to indicate safety and other important instructions. The
signal words and their meanings are defined as follows. Please understand their meanings
clearly before reading this manual.
S-1
3. Safety Precautions
Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the system.
WARNING: 1 Do connect the power plug of this system and power plugs of
the peripherals to wall receptacles that meet the ratings
indicated on the rating nameplate. Using an adapter or
multifunctional receptacle may affect the system’s grounding
performance, and cause the leakage current to exceed safety
requirements.
Besides, do connect the video printer to the auxiliary power
supply socket of this system, and use the printing cable
provided with this system to connect the printer. Otherwise, it
may cause electric shock.
2 Connect the grounding conductor only before turning ON the
system. Connect the grounding cable after turning OFF the
system. Otherwise, electric shock may result.
3 For the connection of power and grounding, follow the
appropriate procedures described in this operator’s manual.
Otherwise, there is risk of electric shock. Do not connect the
grounding cable to a gas pipe or water pipe; otherwise
improper grounding may result or a gas explosion may occur.
4 Before cleaning the system, disconnect the power cord from
the socket. System failure may result in electric shock.
5 This system is not water-proof. Do not use this system in any
place where water leakage may occur. If any water is sprayed
on or into the system, electric shock may result. If water is
accidentally sprayed on or into the system, please turn off
the system power immediately and contact Mindray
Customer Service Department or sales representative.
6 Store and use the transducers carefully. In case that a
scratched transducer surface is found, immediately stop
using the transducer and contact Mindray Customer Service
Department or sales representative. There is risk of electric
shock if using a damaged or scratched transducer.
7 Do not allow the patient to contact the live parts of the
ultrasound system or other devices, e.g. signal I /O ports. If
the system or other devices is defective, there is risk of
electric shock.
S-2
8 Do not use an aftermarket transducer other than those
specified by Mindray. The transducers may damage the
system, causing a profound failure, e.g. a fire in the worst
case.
9 Do not subject the transducers to knocks or drops. Use of a
defective transducer may cause an electric shock.
10 Do not open the covers and front panel of the system.
Otherwise, short circuit or electric shock may result.
11 Do not use this system simultaneously with equipments such
as an electrical knife, high-frequency therapy equipment, or a
defibrillator, etc., otherwise, electric shock may result.
12 Only use the ECG leads accompany with the ECG module;
otherwise electric shock may result.
13 Attentions in the transportation: when this system is moved,
please hold the handle. If other parts of the system are held, it
may cause damage due to the abnormal force. Do not push
the system from the left/right side, otherwise, it may be
toppled over.
14 Accessory equipments connected to the analog and digital
ports must comply with the relevant IEC standards (e.g., IEC
60950 information technology equipment safety standard and
IEC 60601-1 medical equipment standard). Furthermore, all
configurations must comply with the standard IEC60601-1-1.
It is the responsibility of the person, who connects additional
equipments to the signal input or output ports and configures
a medical system, to verify if the system complies with the
requirements of IEC60601-1-1. If you have any questions
regarding these requirements, consult your sales
representative.
15 Prolonged and repeated use of keyboards may result nerve
disorders in hand or arm for some individuals. Observe the
local safety or health regulations concerning the use of
keyboards.
S-3
2 Malfunctions due to radio wave:
(7) When move the system on the steps, please take care to
prevent the system from toppling.
S-4
6 Always keep the system dry. Avoid transporting this system
suddenly from a cold place to a warm place; otherwise
condensation or water droplets may form, which may result
short circuit.
S-5
14 Do not use the system to examine a fetus in the Doppler
mode for a long period of time.
4 Turn ON the system only after the power is turned OFF for more
than 20 seconds. If the system is turned ON immediately after
being turned OFF, the system may not reboot properly and may
lead to malfunction.
S-6
5 Do not disconnect a transducer that remains in a live imaging
state. This can damage the system and / or transducer. Press
[Freeze] or turn off the power of the system before connecting or
disconnecting a transducer.
6 Remove the ultrasound gel from the face of the transducer when
the examination is completed. Water in the gel may enter the
acoustic lens and adversely affect the performance and safety of
the transducer.
8 Do not apply external force to the control panel. The system may
be damaged.
13 Ensure that the current exam date and time are the same as the
system date and time.
4. Warning Labels
The warning labels are attached to this system in order to call your attention to potential
hazards.
The symbol on the warning labels indicates safety precautions. The warning labels use
S-7
the same signal words as those used in the operator’s manual.
Please refer to the operator’s manual for detailed information about the warning labels. Read
operator’s manual carefully before using the system.
The name, pattern and meaning of each warning label are described as follows:
S-8
Table of Contents
2.3.1 Turning...............................................................................................................2-7
C-1
3.3 Transmitting Board .....................................................................................................3-12
3.9 CW Board...................................................................................................................3-29
C-2
3.14 System Power Supply Distribution .............................................................................3-48
4.3.7 Replace the LCD Screen and LCD Control Board ..........................................4-15
C-3
4.3.19 Disassemble the Frontal Cover of the Main Unit.............................................4-29
4.3.35 Replace the Power Supply Main Board and Auxiliary Board ..........................4-52
C-4
5.2 Maintenance Personnel ...............................................................................................5-2
6 Checking 6-1
6.4.1 Checking the B/W Image Phantom Data and Image Recording.....................6-15
C-5
7.3.1 Self-diagnosis Interface...................................................................................7-14
8.1.2 Examine and Repair Flow When No Image Displayed in the Image Area........8-2
8.1.3 Examine and Repair Flow When Black Area Exists in the Image.....................8-3
8.1.6 Examine and Repair Flow When the System Can’t Start-up.............................8-4
C-6
1 System Introduction
1-1
System Structure
2 System Structure
<1>
<8>
<2>
<9>
<3>
<4> <10>
<5>
<6>
<7>
2-1
System Structure
<11>
<12>
<13>
<14>
2-2
System Structure
2-3
System Structure
Deflector rod
Shown as above, the deflector rod is located at the left lower corner at the back of the monitor.
When the deflector is at the rightmost, the monitor is at working status and the angle between
monitor and support arm can be changed at the range controlled by the deflector rod. The
monitor can be turned for 20° both forward and backward, shown as follows:
2-4
System Structure
In the process of transportation or moving the system, the deflector rod can be toggled left to
level the monitor, shown as follows:
2-5
System Structure
Put the support arm to horizontal position, the arm will be locked if the lock is at the locked
position; when the lock is located at the unlocked position, the arm can be adjusted from
upside to downside.
2-6
System Structure
Tuning screw
2.3.1 Turning
Hold and push the handle in the direction of front part of machine (located at the frontal part of
back keyboard ), turn the keyboard left and right within 45°, the keyboard will be locked at
where the handle be loosened.
2.3.2 Lifting
Handhold the handle of the keyboard with both hands, with the left hand hold the lifting handle,
press the lifting handle to a certain position, uplift or push down the keyboard with both hands,
then the keyboard can be elevated or lowered. The height can be adjusted is 150mm at most.
2-8
System Structure
<10> Separate video output port Used for composite video output, connect to VCR
output or video printer.
<11> Video input port Used for composite video input.
<12> Video output port Used for composite video signal output, connect
to VCR output or video printer.
2-9
System Structure
<1>
<2>
<5>
<3> <4>
2-10
System Structure
<1>
<2>
<3>
<4>
2-11
System Structure
2-12
System Structure
<11> Biopsy Biopsy Press to show or hide the biopsy guide line.
<12> Setup Preset Press to open/ close the preset dialog box.
<13> / Character key Same as the keys of PC.
Indicating light 1 The indicator light is on after the power is switched
<14> /
on; it is off after the system is turned on.
Indicating light 2 The indicator light flashes when the hard disk is
<15> /
reading or writing; and in other situations, it is off.
<16> F8 User-defined key The function is definable.
<17> F9 User-defined key The function is definable.
<18> F7 User-defined key The function is definable.
<19> TGC Slide bar To adjust the depth and gain.
<20> Volume Volume Rotate to increase or decrease spectrum volume.
To adjust the parameters, the functions are shown
Multi-function
<21> / by the sketch map displayed at the downside of the
knob 1
screen.
To adjust the parameters, the functions are shown by
Multi-function
<22> / the sketch map displayed at the downside of the
knob 2
screen.
To adjust the parameters, the functions are shown by
Multi-function
<23> / the sketch map displayed at the downside of the
knob 3
screen.
Four-direction The functions are shown by the sketch map displayed
<24> /
key at the downside of the screen.
To adjust the parameters, the functions are shown by
Multi-function
<25> / the sketch map displayed at the downside of the
knob 4
screen.
To adjust the parameters, the functions are shown by
Multi-function
<26> / the sketch map displayed at the downside of the
knob 5
screen.
<27> iTouch / Press to optimize the images.
Patient To enter into patient information input interface.
<28> Info
information
Transducer To switch the transducer and exam mode.
<29> Probe
switch
<30> Review Review To review the stored images.
<31> Report Diagnostic report Press to open/ close the report.
<32> End Exam End exam To end an exam.
<33> F10 Self-defining key The function is definable.
/ To switch the live image window in the multi-window
<34> Update
mode.
<35> CW / Reserved key (press to enter the CW mode).
<36> M / Press to enter the M mode.
<37> PW / Press to enter the PW mode.
To adjust the gain of the live image, the light is on;
<38> Gain Gain press to adjust the gain of the corresponding mode in
2D mode or 3D mode.
2-13
System Structure
2-14
System Principle
3 System Principle
IO Probe
connecting CPU connecting
board board board
IO port
board
Power 4D driver
board
Master board
USB TO IDE
DVD R/W
board
Ultrasound Transmitting
PC system
system board
Display and PCI
speaker
Power
AC/DC
supply
board
board
Power supply
AC input
connecting Breaker
connecter
board
Isolating
transformer
The system consists of main unit and power system, and the main unit consists of:
Master board: including ultrasound system, PC system, power supply management and
filtering system;
Ultrasound system: to carry out functions including echo receiving, zoom in, sampling, Beam
forming, signal processing, ultrasound scanning receiving control as well as data exchange
3-1
System Principle
between PCs;
PC system: post processing the echo data, communicate with users, to support many kinds of
peripherals;
Power management and filtering system: to ensure the normal working of power system,
support system shutdown, stand-by mode and dormancy mode;
Meanwhile, master board provides connecting sockets with each module;
Transmitting board: to realize delay focusing transmitting for N channels;
Control panel can communicate with the master board through USB;
CPU board;
Probe board module: including probe control board, probe board, probe connecting board, used
to connect the transducers, supports 4 transducer sockets, of which 3 sockets are live;
CW Board: processing CW signal and pencil transducer signal;
4D driver board (reserved);
The peripheral extension modules including:
I/O connecting board: connects the master board and IO front board, IO interface board,
provides power supply to the hard disk;
IO front board,USB HUB(one input, three outputs), provides external MIC port, USB port as well
as pencil probe port;
IO port board: USB HUB (one input, six outputs), video acquisition, and provides sockets to
some inner wires;
IO rear board: provides IO port for the rear part;
USB-to-IDE board: USB to IDE converting board, connects to DVD R/W;
ECG board;
Display module: including display and speaker;
The power system consists of (In fact, it also includes the power management system on the
master board and the filtering system):
Power supply board: provides power for the system;
Power supply connecting board: provides connection sockets between auxiliary output, breaker
and isolating transformer, to realize voltage input selection;
AC/DC board: to change the AC input into 12V DC.
3-2
System Principle
System
Transmitting
board Master
board
FPGA1
Beamform
er Ultrasound
receiving
CW port 1
Signal port
Process 4D port Master board
power Power
port
Fan Tempe
port rature
sensor
Adapter
FPGA3 port
Video System
process monitor
Multi- DC3 Power
Extension function signal management
port port
COME INTERFACE
3-3
System Principle
Ultrasound receiving;
Signal process;
Power management;
System monitoring;
Video process.
And the master board provides connecting sockets for the following devices:
Probe board module;
Transmitting board;
CW board;
4D board (reserved);
CPU module;
IO connection board;
Power supply module;
Hard disk;
DC-3 power socket;
DC-3 signal socket;
Fan;
Speaker;
Network.
The layout of the above modules’ sockets on the master board is shown as Fig A-1.
3-4
System Principle
3-5
System Principle
Table 3-3 Definition of Signals between the Master Board and Probe Module Socket
Name Description
CW[9:0] CW echo signal
3-6
System Principle
PIN SIG
1 CLK_TMS
2 CLK_TDI
3 CLK_TDO
4 CLK_TCK
5 GND
6 VDD_CLK
PIN SIG
1 ROUTP
2 ROUTN
3 NC
4 LOUTP
5 LOUTN
3-7
System Principle
PIN SIG
1 TCK
2 TDO
3 TMS
4 TDI
5 3V3
6 GND
7 3V3
8 PLUGED
3-8
System Principle
3-9
System Principle
3-10
System Principle
3-11
System Principle
Sockets definition of transmitting board and master board is shown in Table A-1 and Table A-2.
The testing points related to the power supply of transmitting board is defined as follows:
3-12
System Principle
The N testing points are corresponding to the N transmitting channels. Of the 4 rows of testing
points, the upper two rows are even number points, while the lower two rows are odd number
points. The numbers are marked between the spaces, e.g. P65 refers to the 65th channel.
The tested wave of transmitting signal under B Mode is shown as follows:
Note: the amplitude, pulse width and pulse number vary with the transmitting parameters.
3-13
System Principle
Control signal
… ID signal of CPLD
Probe
probe
mode
A probe port switching
Control
signal
signal
Probe
……
Probe
Control
Probe power
signal
Probe port … power …
supply
control
…
Transmitting/ receiving
signal Signal switching
(Relay) Relay control
The probe connection board is applied to connect the master board and the probe board: the
probe control board, which is buckled with probe board, takes control on the switching relays on
the probe board, switches the working probes and A probe mode (probes including general
probes, 4D probe and TEE probe), and reads the probe ID and so on.
The above diagram shows the control circuit principle of a group of relays. Control signal
produced by CPLD turns into RELAY_EN1 after driven by 245, and the RELAY_EN1 signal will
then take control on the conduction and cut-off of N-MOSFET Q1. Q1 controls the contactor
position of relay K1 (whether toggle to normal open or to normal close). When the contactor is
normal close, signal KA2 and KA3 are of VCC level, both signals can act as the power supply of
the related relay, let the relay to be working status.
3-14
System Principle
Table 3-14 Sockets between Probe Connecting Board and Probe Board
Socket P1
No. Name No. Name No. Name No. Name No. Name
A1 Gnd B1 PE127 C1 Gnd D1 PE128 E1 Gnd
A2 PE124 B2 Gnd C2 PE125 D2 Gnd E2 PE126
A3 PE121 B3 Gnd C3 PE122 D3 Gnd E3 PE123
A4 Gnd B4 PE119 C4 Gnd D4 PE120 E4 Gnd
A5 PE116 B5 Gnd C5 PE117 D5 Gnd E5 PE118
A6 PE113 B6 Gnd C6 PE114 D6 Gnd E6 PE115
A7 Gnd B7 PE95 C7 Gnd D7 PE96 E7 Gnd
A8 PE92 B8 Gnd C8 PE93 D8 Gnd E8 PE94
A9 PE89 B9 Gnd C9 PE90 D9 Gnd E9 PE91
A10 Gnd B10 PE87 C10 Gnd D10 PE88 E10 Gnd
A11 PE84 B11 Gnd C11 PE85 D11 Gnd E11 PE86
A12 PE81 B12 Gnd C12 PE82 D12 Gnd E12 PE83
A13 Gnd B13 PE79 C13 Gnd D13 PE80 E13 Gnd
A14 PE76 B14 Gnd C14 PE77 D14 Gnd E14 PE78
A15 PE73 B15 Gnd C15 PE74 D15 Gnd E15 PE75
A16 Gnd B16 PE71 C16 Gnd D16 PE72 E16 Gnd
A17 PE68 B17 Gnd C17 PE69 D17 Gnd E17 PE70
A18 PE65 B18 Gnd C18 PE66 D18 Gnd E18 PE67
A19 Gnd B19 PE63 C19 Gnd D19 PE64 E19 Gnd
A20 PE60 B20 Gnd C20 PE61 D20 Gnd E20 PE62
A21 PE57 B21 Gnd C21 PE58 D21 Gnd E21 PE59
A22 Gnd B22 PE55 C22 Gnd D22 PE56 E22 Gnd
Socket P2
No. Name No. Name No. Name No. Name No. Name
A1 PE52 B1 Gnd C1 PE53 D1 Gnd E1 PE54
A2 PE49 B2 Gnd C2 PE50 D2 Gnd E2 PE51
A3 Gnd B3 PE47 C3 Gnd D3 PE48 E3 Gnd
A4 PE44 B4 Gnd C4 PE45 D4 Gnd E4 PE46
A5 PE41 B5 Gnd C5 PE42 D5 Gnd E5 PE43
A6 Gnd B6 PE111 C6 Gnd D6 PE112 E6 Gnd
A7 PE108 B7 Gnd C7 PE109 D7 Gnd E7 PE110
A8 PE105 B8 Gnd C8 PE106 D8 Gnd E8 PE107
A9 Gnd B9 PE103 C9 Gnd D9 PE104 E9 Gnd
A10 PE100 B10 Gnd C10 PE101 D10 Gnd E10 PE102
A11 PE97 B11 Gnd C11 PE98 D11 Gnd E11 PE99
A12 Gnd B12 PE39 C12 Gnd D12 PE40 E12 Gnd
3-15
System Principle
A13 PE36 B13 Gnd C13 PE37 D13 Gnd E13 PE38
A14 PE33 B14 Gnd C14 PE34 D14 Gnd E14 PE35
A15 Gnd B15 PE31 C15 Gnd D15 PE32 E15 Gnd
A16 PE28 B16 Gnd C16 PE29 D16 Gnd E16 PE30
A17 PE25 B17 Gnd C17 PE26 D17 Gnd E17 PE27
A18 Gnd B18 PE23 C18 Gnd D18 PE24 E18 Gnd
A19 PE20 B19 Gnd C19 PE21 D19 Gnd E19 PE22
A20 PE17 B20 Gnd C20 PE18 D20 Gnd E20 PE19
A21 Gnd B21 PE15 C21 Gnd D21 PE16 E21 Gnd
A22 PE12 B22 Gnd C22 PE13 D22 Gnd E28 PE14
Socket P3
No. Name No. Name No. Name No. Name No. Name
A1 PE9 B1 Gnd C1 PE10 D1 Gnd E1 PE11
A2 Gnd B2 PE7 C2 Gnd D2 PE8 E2 Gnd
A3 PE4 B3 Gnd C3 PE5 D3 Gnd E3 PE6
A4 PE1 B4 Gnd C4 PE2 D4 Gnd E4 PE3
A5 Gnd B5 Gnd C5 Gnd D5 Gnd E5 Gnd
A6 PROBEID B6 Gnd C6 RELAY_E D6 Gnd E6 EXP_PRE
_CS1 N1 SENT
A7 PROBEID B7 Gnd C7 RELAY_E D7 Gnd E7 PROBE_P
_CS0 N0 RESENT
A8 Gnd B8 Gnd C8 Gnd D8 Gnd E8 Gnd
A9 SPI_DIN/P B9 SPI_DOU C9 FLASH_P D9 FLASH_W E9 PROBE_I
ROBE_ID T/PROBE_ OWER/PR P/PROBE_ D7
1 ID2 OBE_ID4 ID5
A10 SPI_CLK/ B10 Gnd C10 SPI_CS/P D10 Gnd E10 PROBE_I
PROBE_I ROBE_ID D6
D0 3
A11 Gnd B11 Gnd C11 Gnd D11 Gnd E11 Gnd
A12 Gnd B12 Gnd C12 Gnd D12 Gnd E12 Gnd
A13 +5V B13 +5V C13 +5V D13 +5V E13 +5V
A14 +5V B14 +5V C14 +5V D14 +5V E14 +5V
A15 +5V B15 +5V C15 +5V D15 +5V E15 +5V
A16 Gnd B16 Gnd C16 Gnd D16 Gnd E16 Gnd
A17 Gnd B17 Gnd C17 Gnd D17 Gnd E17 Gnd
A18 4D-Gnd B18 4D-Gnd C18 4D-Gnd D18 4D-Gnd E18 4D-Gnd
A19 4D-RES1 B19 4D-RES2 C19 4D-RES3 D19 4D-RES4 E19 4D-RES5
Thermison Thermison TEE-Agnd Hall_signal 4D-ID
+ -
A20 4D-Gnd B20 4D-Gnd C20 4D-RES6 D20 4D-Gnd E20 4D-Gnd
TEE-Angle
A21 4D-MOTO B21 4D-MOTO C21 4D-Gnd D21 4D-MOTO E21 4D-MOTO
3-16
System Principle
3-17
System Principle
3-18
System Principle
Table 3-17 Signal Definition between I/O Connecting Board and CW Board Pencil Probe Port
Table 3-18 Socket Definition between the I/O Connecting Board and I/O Port Board
No. Name No. Name No. Name No. Name No. Name
A1 3V3 B1 3V3 C1 3V3 D1 3V3 E1 3V3
A2 GND B2 GND C2 GND D2 GND E2 GND
A3 5V B3 5V C3 5V D3 5V E3 5V
A4 5V B4 5V C4 5V D4 5V E4 5V
A5 GND B5 GND C5 GND D5 GND E5 GND
A6 12V B6 12V C6 12V D6 12V E6 12V
A7 GND B7 GND C7 GND D7 GND E7 GND
3-19
System Principle
No. Name
1 12V
2 GND
3 +5V
4 GND
5 +3.3V
6 GND
3.6.1 USB
The one-channel USB signal from the master board turns into 3 channels after processed by
HUB, of which, 1 channel is connected to ECG module and the other two channels are
connected to external standard USB ports. Meanwhile, the external USB port is turned into
VBUS that connects to U port by a 2042.
3-20
System Principle
USB1 over-current
protection VBUS
USB2 over-current VBUS
protection 2042
NC
NC
USB1
USB2
Up -channel ECG_USB
USB GL850G NC
The chip is externally powered, the chip and peripheral circuits are powered by 3.3V, which is
adapted from 5V VCC by the adapter that built-in the chip; the drive level of HUB U port is
powered by on-board power supply. After powered on, the requested bus ID will be recorded in
the built-in memory by GL850G, so no external EEPROM or FLASH is required for memory. In
the PGANG position, interface to a pull-up resistor to enable the chip’s GANG mode, and we
reserve a 0ohm resistor position to make the chip work in independent mode. The
corresponding indicator light is used to indicate if the USB channel is OK. The relationship
between the indicator light and USB channel is as follows:
Item Function
D1 USB3 abnormal work indicator light
D2 USB3 normal work indicator light
D3 GANG mode indicator light
D4 USB1 normal work indicator light
D5 USB1 abnormal work indicator light
D6 USB2 normal work indicator light
D7 USB2 abnormal work indicator light
Table 3-21 Socket Definition between I/O Front Board and I/O Connecting Board
3-21
System Principle
Table 3-22 Socket Definition between I/O Front Board and Pencil Probe
No. Name
1 +5
2 USB-
3 USB+
4 GND
5 GND
6 GND
No. Name
1 +5V
2 +5V
3 USB-
4 USB+
5 Gnd
6 Gnd
² 1 MIC port (function reserved), defined as follows:
No. Name
1 GND
2 MIC_R
3 MIC_L
3-22
System Principle
into 2 channels S-VIDEO signal, 1 channel VGA signal into 2 channels, 1 channel USB signal
into 6 channels, and video/audio signal collection and so on.
Signals through I/O port board can be classified into two kinds: some signals are processed on
the board, including USB, VGA output, VIDEO input/output, S-VIDEO input, and left/right audio
input; other signals are not processed on the board, just transmitted, including parallel port,
serial port, left/right audio output, speaker output, HD of keyboard, BREAKER signal, soft switch
control signal and 2 channels remote signals.
3.7.1.1 USB
The one-channel USB signal from the master board is connected to the board through a 110PIN
socket, two one-to-four USB HUB CONTROLLER ICs are used to work in cascade mode to
realize USB 1-to-6, the diagram is shown as follows:
TP2042
Over-current
NC protection
NC
Over-current NC TP2042
NC NC
protection
Down-channel Down-channel
port 4 port 4 NC
3-23
System Principle
bandwidth, we assign the port which has more data onto the first level HUB. The corresponding
indicator lights are used to indicate if the USB channel is OK. The relationship between the
indicator lights and USB channels is as follows:
No. Name
D1 EM2860 working indicator light.
D3 HUB1working indicator light.
D4/D5 HUB1 1 channel (HUB2 up-channel) indicator light.
D6/D7 Working status indicator light of video/ audio collection module.
D8/D9 Working status indicator light of control panel U port.
D10/D11 Working status indicator light of DVD U port.
D12 Working status indicator light of HUB2.
D13/D14 Working status indicator light of standard USB port 2 on rear board.
D15/D16 Working status indicator light of printer U port.
D36/D37 Working status indicator light of standard USB port 2 on rear board.
Of which, the yellow light means abnormal work, and the green light means normal work.
Indicator light EM2860 can only be lightened when the collection software is running.
3-24
System Principle
Video
DV signal
COMP_In
Interface
Video Analog Y_In
Decoder switch
HUB C_In
Switch
USB Video I2C control Rear
Capture Device board
Line_In_R
AC97 Link Audio Codec Line_In_L
VGA RGB
ISL59830
VGA HS/VS
SVIDEO
ISL59830
Table 3-26 Socket between I/O Port Board and I/O Rear board
3-25
System Principle
Table 3-27 Socket between I/O Port Board and USB Port on DVD-RW Module
No. Name
1 +5V
2 +5V
3 DM6
4 DP6
5 GND
6 GND
² 1 power supply socket of DVD-RW module, which is defined as follows:
3-26
System Principle
Table 3-28 Socket between I/O Port Board and Power Supply of DVD-RW Module
No. Name
1 VGA_V
2 GND
3 VGA_H
4 GND
5 VGA_R
6 VGA_B
7 VGA_G
8 GND
9 GND
10 GND
² 1 socket of speaker output, which is defined as follows:
No. Name
1 LEFT+
2 LEFT-
3 RIGHT+
4 RIGHT-
² 1 socket of W/B video printer, which is defined as follows:
No. Name
1 VIDEO_OUT
2 VIDEO_OUT
3 GND
4 GND
5 RM_PRINT
6 RM_BUSY
² 1 socket between I/O port board and control panel, which is defined as follows:
Table 3-32 Socket between I/O Port Board and Control Panel
3-27
System Principle
Table 3-33 Socket between I/O Port Board and Power Soft Switch
No. Name
1 POWER_SSW
2 GND
² 1 socket of power supply auxiliary output control, which is defined as follows:
No. Name
1 +12V
2 +12V
3 GND
4 GND
² Socket connected to built-in USB port of digital W/B video printer, which is defined as
follows:
Table 3-35 Socket Connected the USB Port of Digital W/B Video Printer
No. Name
1 USB_DM1BUS
2 USB_DM1BUS
3 DM1
4 DP1
5 GND
6 GND
3.7.3.1 USB
The working status of USB channel can be seen from the indicator light on the IO port board. If
the light is still off after booting strap, it means no power on the board or the first USB received
no signal from the up-channel USB, or the chip didn’t be initialized. Check if there are clock
waves on C58, C59, if not, it means the chip or crystal is damaged.
3-28
System Principle
check the input frequency of crystal; If the input frequency of crystal is 14MHz, and the voltage
th rd
of 4 pin is 1.8V and the 3 pin is of 0V, then the chip should be replaced. Check if the signal
th
frequency of R28 is about 12.36MHz, if not, check if the frequency of 6 pin on EMP202 is
normal, if it is still abnormal, please replace the IC. If USB device can be recognized by PC,
then EM2860 is normal.
No. Name
1 SYS_RESET_FI
2 GND
² 1 reserved socket to the power supply of fan, which is defined as follows:
No. Name
1 +12V
2 GND
3 +12V
4 GND
3.9 CW Board
Principle of CW board is shown as follows. It mainly consists of a CW receiving channel and
pencil probe receiving & transmitting channel. It can be classified into several modules, which
include socket unit (power supply filtering and transforming), delay-line unit, pencil probe unit,
low pass filter unit, demodulation unit, wall filter wave unit, gain adjusting unit and A/D
conversion unit; power supply on the board can be classified into analog and digital one, the
former one includes A+5V, VCC(5V), A-5V, A+12V, A-12V, PHV, A+1v5, and the latter one
includes VDD(3.3V).
3-29
System Principle
+/-5V
3.3V
Single-end
CW to
0 differential
Connector 1
Connector 2
Single-end
Amplifier 2 to
I_V switching Amplifier Wall filter differential
Reference
LOCLK( level drive
differential)
Low noise amplifier 8MHZ
Pencil
probe
socket SW_CTRL
Passive Transm
Low- Drive
itting 2M square wave
pass Mos
drive
filter
Analog power: +5V, -5V, PHV,
+12V,
-12V (external power), +1.5V
digital power: +3.3V (external input)
+/-12V
PHV
3-30
System Principle
In 1.5V converting circuit, +5V is converted into 1.5V through voltage drop IC, and voltage
control is determined by the peripheral sampling resistors according to R1=R2*[(VOUT/0.9)-1].
3-31
System Principle
PHV
CW Pencil
CON2 MD1201 TC2320
probe
The function of receiving circuit is to amplify the signal received by the pencil probe, and then
send the signal to the front-end filter to share the after filtering demodulation and A/D sampling
circuit with the voltage wave output from CW, and before sharing, both signals will be switched
by analog switch.
3-32
System Principle
U1
In 0 3 VSS R6
1 1 50 4
2 2 45 5
VCC 99 10 C1
C3
VSS
C8 0
R8
R1=23.45k
C2=100nf
R2=31.24k
R3=25.0k
3-33
System Principle
Name Description
PIN Receive
POUT Transmit
PENPRESENT Presenting signal
3-34
System Principle
3-35
System Principle
3-36
System Principle
After powered by 3.3V and the reset signal output from R_Config Module became effectless, the
buckled board will read out data from SPI Flash and allocate FPGA, after the allocation is
completed, program Boot Loader saved in FPGA will copy the application program Boot Loader
to SDRAM, and the whole board starts to work.
Rotating
Trackball
encoder
Signal
Buzzer standardization LDO
circuit
3.3V
TGCadjus
Key LEDdrive
t(8 SPI ADI Board ID
matrix circuit
sections)
3.3V
3-37
System Principle
74LS0
7 L0 R
L1 R
F R
L2
P
G
A R0
R1 GND
R2
74HC14
3-38
System Principle
3-39
System Principle
The electrocardiosignal, after been amplified, filtered, and sampled, will be uploaded to the PC
for R-wave testing, the tested cardioelectrical trigger signal will be uploaded to PC through USB
port. Block diagram is shown as follows:
Right-
Electrode
leg
2
drive
Electrode
Difference High- Amplifica- Control signal
1 Input Pace DSC
Buffer amplifica- pass tion low- ECG serial port
Electrode protection rejection control
tion filter pass filter
2 Data serial system
port 2105
Testing signal 1 A
Lead off MCU
Buffer D
checking
Testing signal 2
PC
UART-to USB
2108
3-40
System Principle
3-41
System Principle
3 DP Differential signal
DP
4,5, GND Earth terminal of
6 main system
J5 Footswitch socket 1 ROW Common input
terminal of
footswitch
2 LINE1 Input of footswitch
cable 1
3 LINE2 Input of footswitch
cable 2
4 NC Not connected
J6 JTAG debugging 1 VCC 3.3V Applied in
socket 2 GND ECG module Debugging
floating ground
3 TRST JTAG — — TRST
signal
4 TCK JTAG — — TCK
signal
5 TDI JTAG——TDI signal
6 TDO JTAG — — TDO
signal
7 TMS JTAG — — TMS
singal
8 RTCK JTAG — — RTCK
signal
9 NRST JTAG — — NRST
signal
10 NC Not connected
J7 ARM BOOT 1 / To pin LPC213X Applied in
socket P0.14 repairing
2 GND ECG module
floating ground
J8 USB port 1 VBUS 5V VBUS signal
2 USB-
3 USB+ JTAG — — TRST
signal
4 TCK JTAG — — TCK
signal
3-42
System Principle
J8 USB port
(2109 exclusive)
J4 USB port
(2108 exclusive)
Footswitch socket
J5 (2105 exclusive)
3-43
System Principle
Crystal
Electrostatic
IDE USB
GL811S prevention,
socket port
filter
Power
5V supply VBUS
module
3-44
System Principle
The system master board and power supply master board are connected through a 64-pin
socket, signal is defined as Table A-8. In Table 3-20, the inputs of related signals on the system
master board and power supply to the power supply master board is finished through the 64-pin
socket. The power supply master board and auxiliary board output +12V, +5V, 5Vstb, 5Vstb-cpu,
3-45
System Principle
+3.3V, THV; the two boards are connected through 26-pin socket (signal definition is shown as
Table 3-44). The Vbus+, +5v, EDC_Power signals that input to the power supply master board
are transformed into -12V, -5V, +2.5V, and +1.2V outputs by the auxiliary board, the four outputs
are returned to the power supply master board through 26-pin socket, and then be output to
each electro-terminal on the system master board through 64-pin socket.
Isola-
Primary winding
ting tap
Mains input
transfor Power
mer
Secondary
winding tap
supply
connec-
ting
board
Peripherals output
AC output
(3 channels)
Power
supply
adapter
Edc_power Edc_power Edc_power
Power_on Power_on
CW_mode
5Vstb_cpu_en
THV_range
TEMP_D+
TEMP_D- start
+12V VBUS+
+5V +5V
+5Vstb
Power Power
System
supply supply
master +5Vstb_cpu
master auxiliary
board
+3.3V board board
THV
+2.5V +2.5V
+1.5V +1.5V
-5V -5V
-12V -12V
3-46
System Principle
Table 3-44 Definition of Signals between the Power Supply Master Board and Auxiliary Board
3-47
System Principle
3-48
System Principle
3.15 LCD
3.15.1 Inverter
Voltage
feedback
Voltage-up
Switch power PWM control MOSFET
transfer
Switch-
AC ing
power
input
Output to the
light
On/Off
Luminance
control Current
feedback
3-49
System Principle
3-50
System Principle
CN4
CN5 CN1
CN100
AC power input
CN2
CN3
Lights drive output sockets CN2, CN3, CN4, and CN5 are defined as follows:
AC power input socket CN100 (socket model: JST, B03P-VL600V20A) are defined as follows:
3-51
System Principle
Pin Symbol
1 L
2 N
3 GND
LVDS
Panel
Power
Module&
Inverter
Cry
Scaler EEPROM
stal
MCU
Cry IIC
stal
After VGA signal is input to RTD2620, MCU will carry out functions including signal recognition,
zoom, color process and transform signal into LVDS signal and then display it to LCD. U6
(MTV412) is MCU, functions of MCU include: controlling on RTD2620 and external related
operation, such as operation on the keys and IIC. U1, U3, U4 carry out DC-DC voltage
converting, thus providing normal voltages to each circuit.
3-52
System Principle
U5 24C16 EEPROM
U6 MTV412 MCU
U8 K4D263238I GDDR
3-53
System Principle
J2 Pin definition
Pin1 L-SDBX+
Pin2 L-SDBX-
Pin3 R-SDBX+
Pin4 R-SDBX-
J3 Pin definition
Pin1 VSYNC Pin2 GND
Pin3 HSYNC Pin4 GND
Pin5 RED Pin6 BLUE
Pin7 GREEN Pin8 GND
Pin9 GND Pin10 GND
J4 Pin definition
Pin1 L-SDBX+
Pin2 L-SDBX-
Pin3 R-SDBX+
Pin4 R-SDBX-
J5 Pin definition
Pin1 R_LED
Pin2 G_LED
Pin3 AD0
Pin4 AD1
Pin5 GND
Pin6 GND
Pin7 12V
Pin8 5V
3-54
System Principle
J6 Pin definition
Pin1 ADJ(Backlight Adjust)
Pin2 EN(Backlight ON/OFF)
Pin3 GND
Pin4 GND
Pin5 12V
Pin6 12V
Pin7 PS(Power Saving)
3-55
System Structure and Assembly / Disassembly
4-1
System Structure and Assembly / Disassembly
No. Material
Name No. Material number Name
number
2109-30-76182 Main unit left 15 Monitor module
1 2109-30-76336
cover
2109-30-76194 DVD-R/W 16 2109-30-76189 Up/down air spring
2
module module
2109-20-76214 Isolating 17 2109-20-76078 Rotating axis barrel
3
transformer
2109-30-76181 Main unit rear 18 2109-30-76178 Front cover of main
4
cover unit module
2109-30-76199 Main unit box 19 2109-30-76183 Main unit right
5
module cover
6 2109-30-76188 IO module 20 2109-30-76193 ECG module
2109-30-76187 Power input 21 2109-30-76258 Probe board
7
module
2109-21-76337 Power adapter 22 2109-30-76179 Main unit frontal
8 and connecting cover
cable
2109-30-76177 Main unit top 23 2109-20-76091 Cover of printer
9
cover compartment
2109-20-76059 Cast aluminium 24 2109-30-76180 Footswitch
10
-51 keyboard base
2109-30-76169 Support arm 25 2109-30-76185 Main unit housing
11
base module
2109-20-76071 Cover of cast 26 2109-20-76145 Dustproof net
12
aluminium base bracket
2109-30-76261 Control panel 27 2109-30-76184 System soft switch
13
module
14 2109-30-76176 Support arm
4-2
System Structure and Assembly / Disassembly
8 11
10 12
13
6
9
5
4
3
2
14
1
No. Material
Name No. Material number Name
number
Cover of lower Cover of upper
1 2109-20-76057 8 2109-20-76058-51
support arm support arm
Upper arm
2 2111-20-73313 turning auxiliary 9 2111-20-73285-51 Decorative cover
pressing block
Dish-shaped
rd
3 M6T-020002--- spring, the 3 10 2111-20-73325 Damp axis
series D=50
Upper arm
4 2111-20-73296 turning auxiliary 11 2111-20-73288 Monitor rear cover
washer
Monitor
Lower support
5 2109-30-76170 12 2109-20-76334 components-install
arm
board module
Upper support
6 2109-30-76174 13 2111-30-73334 Frontal cover
arm
Bottom
Monitor cable lights(included in
7 2111-20-73289 14 2111-30-73335
cover the frontal cover
module)
4-3
System Structure and Assembly / Disassembly
4-4
System Structure and Assembly / Disassembly
4-5
System Structure and Assembly / Disassembly
Material
No. Name No. Material number Name
number
Power and
Power supply main
1 2109-20-76326 signal cable of 16 2109-30-76310
board
SATA
HDD bracket Power supply
2 2109-20-76305 17 2108-20-65745
separating board
Hard disk 40G
Power supply
3 0000-10-11267 5400rpm, 2.5 18 2109-30-76312
auxiliary board
inch SATA port
Bottom cover of
4 2109-20-76150 Main box 19 2108-20-65744
power supply box
IO connecting Shielding cover of
5 2109-30-76201 20 2109-20-76158
board port board
Probe
6 2109-30-76211 connecting 21 2109-30-76224 IO port board
board
Board guide-pin Transmitting board
7 2109-20-76155 22 2109-20-76149
socket shielding cover
Shielding frame
of probe
8 2109-20-76154 23 2109-30-76456 Transmitting board
connecting
board
Shielding cover CW board shielding
9 2109-20-76472 24 2108-20-65750
of CPU board cover
10 0000-10-11188 CPU board 25 2108-30-65899 CW Doppler board
CPU heat
11 2109-20-76329 26 2109-30-76294 Master board
conducting mat
Lower shielding
12 2109-20-76473 CPU radiator 27 2109-20-76061 cover of master
board
CPU fan and
13 2109-20-76238 connecting the 28 2109-20-76153 Air inlet fan frame
cable
Cover of main Air inlet fan and the
14 2109-20-76151 29 2109-20-76231
unit box connecting cable
Top cover of
15 2108-20-65743 power supply
box
4-6
System Structure and Assembly / Disassembly
3) Remove the two M4X12 screws which are used to secure the monitor cable, and then
remove the monitor cable cover aslant.
4-7
System Structure and Assembly / Disassembly
4) Remove the 1 M4X12 screw securing the ground wire holder, take out the cable plug
and remove the 6 M4X12 screws that are used to secure the monitor.
4-8
System Structure and Assembly / Disassembly
5)Hold the lateral sides of the monitor, raise up the monitor to separate it with the support
arm hooker when the monitor is 20° from the vertical direction, and then remove the monitor.
4-9
System Structure and Assembly / Disassembly
1.Frontal Cover of the Monitor 2. Rear Cover of the Monitor 3.M4X12 Screw
4-10
System Structure and Assembly / Disassembly
3) Pull out the speaker connecting cable plug and cut off the tie which is used to secure the
cable, remove the self-driving screws which are used to secure the speaker (4 screws at
each side), and then remove the speaker.
1. Connecting Cable Plug 2. OSD Board 3. PT3X8 Screw 4.Connnecting Cable Plug 5. Monitor
4-11
System Structure and Assembly / Disassembly
1. Bottom Light Module 2. Connecting Cable Plug 3. Screw PT3X8 4. Monitor Front Cover
4) Remove the 2 M3X8 sunk screws securing the transparent cover, and remove the
cover.
5) Take off the two light tubes.
6) Remove the 3 self-driving screws PT3X8 which are used to secure the monitor light
board, and then remove the board.
4-12
System Structure and Assembly / Disassembly
4-13
System Structure and Assembly / Disassembly
1. M3X8 Screw 2. PCB Shielding Cover 3. Connecting Cable Plug of OSD Board
4. Connecting Cable Plug of Speaker Output
Fig 4-15 Replace the Power Supply Inverter Board of the Monitor (1)
4)Pull out the 4 LCD connecting cable plugs at the left side of LCD power supply converter,
and pull out the connecting cable plug at the right side between the LCD power supply board
and the driving board.
5)Remove the 5 M3X8 screws which are used to secure the LCD power supply inverter, and
then remove the power supply inverter.
1. LCD Power Supply Cable Plug 2. M3X8 Screw 3. LCD Power Supply inverter Board
4. Cable connecting LCD Power Supply Board with Drive Board 5. LCD Control Board
Fig 4-16 Replace the Power Supply Inverter Board of LCD (2)
4-14
System Structure and Assembly / Disassembly
Fig 4-17 Replace the LCD Screen and LCD Control Board (1)
3)Remove the 10 M3X8 screws which are used to secure the PCB shielding cover, and then
remove the cover.
Fig 4-18 Replace the LCD Screen and LCD Control Board (2)
4-15
System Structure and Assembly / Disassembly
4)Pull out the 4 LCD connecting cable plugs at the left side of LCD power supply inverter,
and pull out the connecting cable plug on the LCD control board between the LCD power
supply board and drive board, pull out the LCD LVDS connecting cable plug.
5)Cut off the 2 ties which are used to secure the connecting cable, remove the 5 M3X8
screws which are used to secure LCD power supply inverter and the 5 M3X8 screws which
are used to secure LCD control board, and then take away the inverter board and LCD
control board.
1. Power Supply Converter Board 2. LCD Connecting Cable Board 3. M3X8 Screw
4. Connecting Cable between LCD Power Supply Board and Drive Board
5. M3X8 Screw 6.LCD LVDS Connecting Cable
Fig 4-19 Replace the LCD Screen and LCD Control Board (3)
6)Remove the 4 M3X8 screws which are used to secure LCD screen, take away the LCD
screen, and then finish the replacement of LCD screen and LCD control board.
4-16
System Structure and Assembly / Disassembly
Fig 4-20 Replace the LCD Screen and LCD Control Board (4)
4-17
System Structure and Assembly / Disassembly
1. Upper Support Arm Cover 2. Upper Support Arm Module 3. Caster Brake
4) Lock the upper support arm by the up/down spanner of upper support arm. (Refer to
2.2.3 ).
5) Push down the upper support arm by left hand, screw out the securing pins and remove
them. And then leave the arm carefully.
4-18
System Structure and Assembly / Disassembly
NOTE: Please drop some fixing glue onto the tail of the pin when install the pin.
6) Remove the 3 M4X8 screws which are used to secure the air spring, screw out the air
spring and connection base at the position of lifting ear of the air spring (use the clamp
to remove the cementing agent at the lifting ear).
7)Dispatch the shield ring by the straight screwdriver, take out the air spring pin from the
4-19
System Structure and Assembly / Disassembly
NOTE: After the support arm (including the monitor) is installed, please check if
the monitor can stay at any height, if can’t, then: a. If it keeps rising, first
adjust the screw anti-clockwise to be still by the straight screwdriver, and
then check again, if that situation continuous, replace the air spring; b. If
it keeps lowering down, first adjust the screw clockwise to be still by the
straight screwdriver, and then check again, if that situation continuous,
replace the air spring.
1. Adjusting Screw
4-20
System Structure and Assembly / Disassembly
3) Cut off the tie on the monitor signal cable, and pull out the cable from the joint of the
support arm.
4) Take off the plastic cover from the support arm using the straight screwdriver, and then
take off the 3 internal hexagonal screws M5X16 as well as the washer, at last separate
the support arm module from the base module.
4-21
System Structure and Assembly / Disassembly
Fig 4-28 Disassemble the Control Panel Module (1) (Bottom View)
2) Slightly pull out the control panel module forward in the frontal handle direction, lift the
frontal part of the panel and draw out the external cable plug connected to the control
panel module.
3) Remove the 1 M3X8 screw that jointed with the base module, and separate the control
panel module from the base.
4-22
System Structure and Assembly / Disassembly
3) Pull out the connecting cable plug on the control panel, remove the 6 M3X8 screws
which are used to secure the encoder 6-to-1 connecting board, then replace the board.
4-23
System Structure and Assembly / Disassembly
3) Pull out the connecting cable plug of the single encoder connecting board from the
control panel, remove the 2 M3X8 screws and replace the connecting board.
4-24
System Structure and Assembly / Disassembly
2) Pull out the connecting cable plug connected to the trackball, remove the 2 M3X8
screws which are used to secure the trackball, and take away the trackball.
4-25
System Structure and Assembly / Disassembly
2) Pull out the 8 TGC knobs and 9 encoder knobs upward from the control panel. (Refer to
4.3.8 )
3) Take off the encoder 6-to-1 connecting board, each single encoder connecting board
and the control panel buckled board.(Refer to 4.3.8 and 4.3.10 )
4) Pull out the connecting cable plug of the buzzer from the control panel.
5) Remove the 26 PT3X10 screws on the control panel and the 1 M3X8 screw which is
used to secure the grounding wire.
6) Take out the control panel and the keys from the keyboard cover, and then separate
them to finish the replacement.
4-26
System Structure and Assembly / Disassembly
1. M3X8 Screw 2. Encoder 6-to-1 Connecting Board 3. Single Encoder Connecting Board
4. M3X8 Screw 5. Control Panel Buckled Board 6. PT3X10 Self-driving Screw
7. Control Panel 8. Silica Gel Key 9. Keyboard Cover
3) Remove the 15 M4X12 screws which are used to secure the casting alumilinium cover
at the back of the cover, and then remove the cover.
4-27
System Structure and Assembly / Disassembly
1. M4X12 Screw 2. Large Flat Washer 3. Up/Down Handle 4. Turning Axis Socket
4-28
System Structure and Assembly / Disassembly
1. M4X8 Screw 2. Turning Block Wearing Piece 3. Turning Block 4. Spring 5. Dragline
2) Pull out the footswitch board module horizontally from the machine.
4-29
System Structure and Assembly / Disassembly
Fig 4-43 Disassemble the Frontal Cover of the Main Unit (1)
3) Pull out the plugs (three at each side) that are filled in the left and right lateral sides,
and then remove the screws (three at each side) which are installed on the two sides of
the rack.
4) Remove the 4 M4X8 screws which are used to secure the rack below the main unit
cover, remove the cover downward slightly and take off the frontal cover when the
cover is separated from the slot at the top side.
Fig 4-44 Disassemble the Frontal Cover of the Main Unit (2)
4-30
System Structure and Assembly / Disassembly
4-31
System Structure and Assembly / Disassembly
3) Turn the module 90°to insert its two lugs into the rectangular hole at the lower part of
the rack.
4) If the module has to be disassembled, first remove the 5 M4X8 screws securing the
rack base, and then pull out the cables.
4-32
System Structure and Assembly / Disassembly
4) When assemble the module, put the edge fold of the module totally on the elastic sheet
of main unit, and close the bottom guarding edge to the main unit rack, meanwhile, try to
press down the DVD button and make some changes if necessary.
4-33
System Structure and Assembly / Disassembly
3) Remove the 2 M3X8 screws that connecting the USB-to-IDE port board with the DVD
rack, take off the port board and DVD crooked-angle part.
4-34
System Structure and Assembly / Disassembly
4) Remove the 2 M3X8 screws which are used to secure the DVD patching board
crooked-angle part, disassemble and replace the USB-to-IDE port board.
4-35
System Structure and Assembly / Disassembly
3) Remove the 4 M3 X8 screws on the enclosure of ECG module, take off the internal
cables and disassemble the ECG base module and ECG installation cover module.
Fig 4-56 Disassemble the IO Frontal Board and ECG Board (1)
4) Pull out the internal cables, remove the 4 M3X8 screws securing the IO frontal board,
and then take off the I/O frontal Board.
4-36
System Structure and Assembly / Disassembly
Fig 4-57 Disassemble the I/O Frontal Board and ECG Board (2)
5) Remove the 4 M3X8 screws securing the ECG board, and then take off the ECG. (When
configured with ECG module.)
4-37
System Structure and Assembly / Disassembly
3) Remove the 10 M4X8 screws which are used to secure the probe board shielding cover
and then take off the cover.
1. Probe Board Shielding Cover 2. Shielding Plate of Probe Board 3. M4X8 Screw
4) Remove the 6 M3X8 screws which are used to secure the probe control board, vertically
pull out the board, and then slightly separate the 3 connectors from the board.
5) Remove the 19 M3X8 screws which are used to secure the shielding cover of the probe
board and aluminium shielding frame, then remove the 7 M3X8 screws and 2 M3X7
bolts securing the probe board and the probe board shielding cover, and at last take
away the 4 gaskets, then the probe board shielding cover can be taken off.
4-38
System Structure and Assembly / Disassembly
6) Remove the M3X8 sunk screws on the aluminium shielding frames and shielding plates
of the transducer socket (6 for each), then remove the 3 shielding frames of the
transducer. (Do not move the pseudo transducer at the leftmost).
7) Remove the M3×8 sunk screws securing the probe installing bracket and the aluminium
probe shielding cover (there are 2 sunk screws on each part), then remove the 2
M2.5X8 panhead screws securing the probe port and probe installing bracket, and then
4-39
System Structure and Assembly / Disassembly
8) Remove the 12 M3X8 screws securing the transducer board and aluminium shielding
plate, then remove the 4 M3X7+8-6 stud screws and 8 gaskets, and then separate the
shielding plate from the transducer board.
4-40
System Structure and Assembly / Disassembly
3) Pull out the lower cross-pin of the air spring from the guide holder.
4) Draw the spanner of the retainer cross-pin outwardly, turn it 90°clockwise, and the
spanner will be locked after releasing. (There are lock and unlock symbols marked on
the retainer).
(Note: Remove the ECG module first if it is an obstacle for the operation of the spanner).
5) Hold the keyboard handle by both hands, press down the up/down handle on the
keyboard handle and raise the keyboard. When you hearing “Cluck”, it means the
cross-pin is inserted into the hole of the back-up block, the keyboard is vertically locked
with the main unit.
6) Remove the 6 M5X16 inner hexagon screws, the flat washer, and elastic washer by the
inner hexagon spanner, and then take out the air spring module from the machine.
4-41
System Structure and Assembly / Disassembly
1. M5X16 Inner Hexagon Screw 2. Elastic Washer 3. Flat Washer 4. Air Spring Module
7) Disassemble the retaining ring from the upper cross-pin of air spring by the cross
screwdriver, and then take off the rotary gear.
8) Screw out the upper connecting link and lower connecting link of the air spring, and then
finish the disassembly of the air spring.
4-42
System Structure and Assembly / Disassembly
4-43
System Structure and Assembly / Disassembly
1. M5X10 Inner Hexagon Screw 2. Elastic Washer 3. Flat Washer 4. Isolating Transformer
4-44
System Structure and Assembly / Disassembly
3) Take out the HDD bracket and pull out the HDD plug.
4) Take off the HDD and its bracket, remove the 4 M3X4 panhead screws securing the
HDD, and then finish the replacement.
4-45
System Structure and Assembly / Disassembly
4) Pull out the cable plugs on the IO port board, take off the IO module, and remove the 4
M4X8 screws securing the main unit box module from the back part of the machine.
1. Main Unit Box Module 2. M4X8 Screw (used to fix the cable)
3. M4X8 Screw (To fix the main unit box)
5) Hold the handle of the main unit box, slowly pull out the main unit box module, pull out
the plugs of network cable and power supply adapter (when the plugs can be seen on
4-46
System Structure and Assembly / Disassembly
the top part of the main unit box), and then finish the disassembly.
6) When assemble the main unit box module, make sure to align the guide-pin socket with
the guide-pin, and align the plug of probe connecting board with the socket on the probe
board.
1. Transducer Board Module 2. Guide Pin 3. Guide-pin Socket 4. Main Unit Box module
4-47
System Structure and Assembly / Disassembly
2) Pull out all the connecting cables on the IO port board, and remove the 4 M3X8 screws
securing the IO port board.
If it is not convenient to remove the screws above, you can first remove the 6 M4X8 screws
securing the power supply input module, turn the module 90°, insert the two supporting
lugs into the rectangular hole. (Please refer to procedure 1, 2, 3 as described in section
4.3.19 ).
3) Pull the two spanners to separate the socket of IO port board from the socket of IO
connecting board, and then take out the IO port board upward.
4) Remove the 4 M3X8 screws securing the IO port board shielding cover and then finish
the replacement.
4-48
System Structure and Assembly / Disassembly
4-49
System Structure and Assembly / Disassembly
4) Remove the 6 M3X8 screws securing the IO connecting board, pull out the IO
connecting board outward, and draw out the cable plugs on the IO connecting board,
then finish the replacement.
4-50
System Structure and Assembly / Disassembly
3) Remove the 3 M3X8 screws, pull the 2 spanners to separate the board from the sockets
and then pull out the probe connecting board.
4) Remove the 4 M3X8 screws securing the guide-pin socket, and then replace the
connecting board.
4-51
System Structure and Assembly / Disassembly
Board
1) Remove the main unit box cover. (Please refer to the procedure 1, 2, 3 as described in
section 4.3.30 ).
2) Remove the 4 M3X6 panhead screws securing the power supply module, draw out the
cables on the power supply module and take out the module.
Fig 4-87 Replace the Main Board and Auxiliary Board of Power Supply (1)
3) Remove the 2 M3X6 panhead screws securing the power supply box, take off the cover;
remove the 4 M3X6 screws securing the power supply main board; take out the main
board, the power supply box clapboard and power supply auxiliary board assembly.
4-52
System Structure and Assembly / Disassembly
Fig 4-88 Replace the Power Supply of Main Board and Auxiliary Board (2)
4) Raise the power supply main board to separate it from the connecting socket of the
auxiliary board, and then take off the power supply main board.
Fig 4-89 Replace the Main Board and Auxiliary Board of Power Supply (3)
5) Remove the 4 M3X6 panhead screws securing the power supply auxiliary board, and
then take off the auxiliary board.
4-53
System Structure and Assembly / Disassembly
Fig 4-90 Replace the Main Board and Auxiliary Board of Power Supply (4)
4-54
System Structure and Assembly / Disassembly
4-55
System Structure and Assembly / Disassembly
4-56
System Structure and Assembly / Disassembly
1. M3X35 Screw 2. Fan and Connecting Cable 3. M3X8 Screw 4. CPU Module
3) Remove the 2 M3X8 screws and 2 M2.5X16 screws securing the CPU radiator and
shielding cover, take off the cover; remove the 2 M2.5X8 screws securing the CPU on
the radiator, and then replace the CPU board. (Heat outlet washer has to be replaced.)
Pay attention to the direction of the fan during the assembly, and take care of the memory
bank on the board when replace the CPU.
4-57
System Structure and Assembly / Disassembly
1. Memory Bank
4-58
System Structure and Assembly / Disassembly
4-59
System Structure and Assembly / Disassembly
3) Cut off the tie securing the magnetic ring, remove the 8 screws securing the fan covers,
take off the fan after removing the cover.
1. Screw 2. Fan Cover 3. Fan Bracket 4. Magnetic Ring 5. Fan and the Cables
3) Place a piece of wood at the middle of front/ rear bottom. (150 mm ´ 150 mm ´ 220 mm,
W´L´H).
4-60
System Structure and Assembly / Disassembly
4) Remove the 4 M8X16 screws securing the caster by the inner hexagon spanner, take
out the flat washers, and then disassemble the caster for replacement.
4-61
System Structure and Assembly / Disassembly
2) Place the printer into the compartment and grab it with the frontal cover of the main unit.
(To avoid slippage)
4-62
Maintenance Requirements
5 Maintenance Requirements
Consumables
Consumable Qty. Description
Aluminum foil 1
Physiological saline (0.85 - 0.95%) To fill the whole container.
Glass cleaner 1 To clean the display
Neutral cleanser
Soft cloth
5-1
Maintenance Requirements
Consumables
5-2
Checking
6 Checking
Check times
Model 6 months after 12 months after Remarks
installation installation
DC-3 series - 1
If the insulation feature of the system degrades as the service time increases or due to the
system malfunctions, the deflection range of measurement result may increase as the
humidity increases.
6-1
Checking
½
¯
Check initial operations
½
¯
Check general operations
½
¯
Check keys on the control panel
½
¯
Check the trackball
½
¯
Check the transducers
½
¯
Check the fans
6-2
Checking
l Check all keys on the control panel from left to right and from up to down to see if
they can work properly.
l If a video printer is connected to the system, check if the [Print] key can work
normally.
l Press the [Freeze] key to enter the image freeze status. Press the [Measure] key
to enter the measurement status.
l Measure the vertical distance and horizontal distance to see if the trackball can
work normally.
l Check if the transducers can be pulled out properly. Check if the transducers can
work normally after been connected to the main unit.
l Check if the system fan (fan in the main unit) can work normally after the main
unit is powered on.
l Check if the fans make any abnormal noise when they are working.
6-3
Checking
· Flows of checking
· Performing steps:
6-4
Checking
2) To measure earth leakage current: Press the shortcut key 4 on the instrument. In each
status specified in the Table I below, press the soft key 4 to select AP-Earth and No
AP-Earth to measure the earth leakage current.
4) To measure patient leakage current: Press the shortcut key 6 on the instrument.
In each status specified in the Table II above, press the soft key 4 to select
6-5
Checking
NOTE: 1 Be careful when perform the test. Voltage on the applied part
should be 10% higher than the mains voltage.
2 There must be physiological saline between the transducer and the
aluminum foil.
Table III Testing Status of Patient Leakage Current (110% mains voltage on the applied
part)
6) To measure patient auxiliary current: Press the shortcut key 8 on the instrument.
In each status specified in the Table II above, press the soft key 4 to select
RA-ALL (to measure the patient auxiliary current).
6-6
Checking
See Fig A.
6-7
Checking
(1) Pack the transducer with metal foil and then AC less than or
put the transducer into the container filled equal to 100 uA;
with physiological saline. Connect the DC less than or
aluminium foil to the RA input terminal (B2) equal to 10 uA.
of the instrument with the cable provided by
the instrument, and also plug the 3-way Single-failure
power plug of DUT into the auxiliary power status:
Patient socket (B1) of the instrument.
leakage AC less than or
current (2) Connect ECG electrode to (V1/V2/V3) (B3, equal to 500 uA;
B4, B5) input terminals of the instrument by DC less than or
the input clip. equal to 50 uA.
6-8
Checking
6-9
Checking
6-10
Checking
6-11
Checking
· Checking flow
½
¯
Check the connection of the caster
½
¯
Check the handle
½
¯
Check the rotating part and strutting piece of
keyboard
½
¯
Check the structure of up/ down system
½
¯
Check the monitor and structure of support arm
½
¯
Check the status of the peripheral instruments
½
¯
Check the cable connection
½
¯
Check the transducer appearance
½
¯
Check other mechanical structures
If the check result is “fail”, the system is in abnormal status. Stop the system and adopt proper
6-12
Checking
measures.
6-13
Checking
6-14
Checking
Recording
2) After obtaining an optimal image, freeze the image and record it.
Condition: system preset parameters.
6.4.2.2 Penetration
1) Put some gel on the phantom and then scan the phantom with a transducer.
2) Adjust the gain to make the soft tissue spot displayed at the deepest position.
3) Measure the depths of noise and of the soft tissue boundary, and record the images for
the measurement.
6-15
Checking
½
¯
Image operations checking
½
¯
Image archiving
3) Image archiving
Print images obtained through the steps above and archive them.
6-16
System Maintenance
7 System Maintenance
½
¯
Disconnect the accessories
½
¯
Disassemble the enclosure of main unit
½
¯
Disassemble the box of main unit
½
¯
Clean the interior of main unit and the fans
½
¯
Clean the covers and control panel
½
¯
Clean the screen
½
¯
Clean the peripherals and accessories
½
¯
Clean the cable and the power cord
7-1
System Maintenance
Disassemble the enclosure of main unit, and the relative shielding covers to check the
PCBA and power supply part. If the PCBA have no abnormality, do not disassemble
them.
Put the mouth of the vacuum dust-collector into the plastic bag and turn on the
dust-collector to collect the dust.
Open a small hole on the top of the plastic bag to insert the mouth of the air blower. Then
turn on the blower to remove the dust.
Use neutral cleanser to remove the dust on the enclosure of main unit and on the control
panel. If it is difficult to clean the control panel, disassemble the keys first and then use
neutral cleanser to clean it.
Make sure the display is not skew and the fixing mechanism is secured.
4) Transducers
Remove the dust on the transducer. Check if it has any abnormality such as crack. Then
connect it to the main unit. Use a soft brush to clean the transducer holder.
5) Video printer
Clean the thermal head according to the user manual of the printer.
6) Cable / cord
Use neutral cleanser to clean the cable and power cord.
7-2
System Maintenance
Rotary ball
b)Cleaning
Clean the two long shafts, the bearing and the rotary ball with soft dry cloth or paper.
Long shaft
c)Installing
Put the rotary ball back in the trackball and then align the clamping ring click with the
top cover notch. Press the bulges on the ring with both hands and turn the ring about
45°counterclockwise until the ring clicks. As the bulges are flush with the top cover,
the ring is secured.
7-3
System Maintenance
Top cover
notch
NOTE: Do not turn off the power supply in the process of upgrading,
otherwise, it will lead to upgrading failure and can’t be restarted.
7-4
System Maintenance
7.2.2 Preset
Press [Setup] key, pops up the preset menu.
7-5
System Maintenance
NOTE Be sure to confirm the system information before and after the software
maintenance.
7-6
System Maintenance
A B
D
C D E
Mark Function
A To export the selected presets, and save them in different files.
B To import the preset parameters according to the selection, or to
restore it to the original factory default.
C To export all the preset parameters, and save them in a file.
D To import all the preset parameters at one time.
E To restore all preset parameters to the default value.
7-7
System Maintenance
4. Input the correct password, select “OK”, and press [Set] to enter to “Maintenance” menu.
7-8
System Maintenance
7.2.6.1 Upgrading
This upgrading is done through an upgrading package, the content to be upgraded is the
content changed in the last version prior to the upgrading package.
1. Enter into maintenance status (Please refer to 7.2.5 );
2. Select “Upgrade”, press [Set] key;
3. Select the correct file in the Load File dialog box, and click “OK”, the system will start the
upgrading, and the progress will be displayed on the bottom of the screen.
4. After the upgrading is complete, the successful prompt appears. Restart the system
according to prompts.
7-9
System Maintenance
NOTE: Select the proper upgrading program (package) whether or not ECG
module is configured.
3. Select the item to be upgraded, and press [Set], select the right file in the Load File dialog
box, and click “OK” then the system begins to upgrade the file and the progress will be
displayed at the bottom of the screen.
4. After the upgrading is completed, the system will give out related information and prompt
the user to restart the system.
7-10
System Maintenance
NOTE: Please shutdown the system and turn off the power supply before
restarting the system after completing all the upgrading.
5. Move the cursor to the “Load Config Script”, press Set, and the following dialogue box pops
up.
7-11
System Maintenance
6. Select correct files, and click “OK”. After the operation is complete, the successful prompt
appears. Restart the system according to prompts.
7-12
System Maintenance
4. In the optional list, select the desire software function, and click “Install”;
5. Select the corresponding file with suffix “key” in the dialogue box, and then click “OK”.
The files and the devices are of one-to-one correspondence, install the software.
6. After successful installation, return to the system preset interface, the corresponding
optional modules should be shown as installed. Click “OK” to close the system preset dialog
box.
7-13
System Maintenance
Item
Select
Area
Diagram
Area
Status
7-14
System Maintenance
7-15
System Maintenance
DC-3
Testing Item
Manufacturer Maintainece
Register Read MainBoard ID Read
DSP Logic Version Read
MultiFunction Logic Version Read
Interconnect Test The Connection of BF and DSP(System Control
Bus)
The Connection of BF and DSP(RF Data Port)
The Connection of CPU and DSP(PCI)
The Connection of CPU and MultiFunction (PCI)
Memory Test DataBuffer SSRAM0 Test
DataBuffer SSRAM1 Test
Frame Correlation SSRAM Test
RealTime Data Upload DDR1 Test
Scan Parameter Setting DDR Test
MuliFunction DDR Test
Power Circuit Test Battery A Test
Battery B Test
Scan Mode and PHV Control Test
System Monitor Test Voltage Monitor Test
FrontEnd Circuit Test ADC Digital Port SelfTest
CW Circuit Port Test
Adjust Module Connunication Test
Prober Port Test
Prober Expension Module Port Test
KeyBoard Communication Test
Port Test
Serial Port Test
Video Test
S-Video Test
Audio Test
ECG Module Port Test
Note: : means this testing item exists in the system self-diagnosis software of this version.
: means this testing item doesn’t exist in the system self-diagnosis software of this
version.
7-16
System Maintenance
7-17
System Maintenance
7-18
System Maintenance
7-19
System Maintenance
The circuits concerned including the connecting circuit between Beamformer (U302) and DA
chip (U297), U279, connecting circuit between DA output of U297 to the power supply module,
as well as PHV power circuit.
NOTE: It may take about 3 minutes to finish this test, please wait.
7-20
System Maintenance
Testing function:
It is AD controller board communication test.
Details description:
It is to test if SM bus between PC module and AD controller board is normal, and if AD
controller board is normal.
3)Probe Port Test
Testing function:
To test the ID communication channel between the probe board and the probe is normally
working.
Details description:
Plug the probe 7L4A into the A socket of the probe board before testing the DC-3 series
instruments.
7-21
System Maintenance
Click [Start] to start the testing, each control on the simulating keyboard is corresponding to
the one control on the actual keyboard.
Press down the key on the actual keyboard, the corresponding key on the simulating
keyboard will flash and change into green, flashing means the key is being tested, while
green means the key past the test.
Scroll the trackball, the cursor on the interface will be shifted accordingly.
Encoder rotating function test: on the simulating keyboard represents the encoder on
the actual keyboard, when the rotating function of the encoder is being tested, the ball
corresponding to the encoder will flash and change into green. Flashing means the encoder is
being tested, while green means the encoder past the tested. Meanwhile, with the rotating of
the encoder, the green ball around will rotate in the same direction with the encoder.
Encoder key test: press down the key of the encoder, the ball corresponding to the encoder
will flash and change into green. Flashing means the encoder being tested, while green
means the encoder past the test. Meanwhile, the central key of will change with the
TGC test: the 8 TGCs on the keyboard are corresponding to controls on the
simulating keyboard respectively. Slide the TGC on the keyboard, the TGC on the simulating
7-22
System Maintenance
If all parts of keyboard are in normal conditions, press down [KeyBoard is OK and Quit] button,
otherwise, press down [KeyBoard is Bad and Quit] button.
5)Serial Port Test
Testing function:
Serial port test, the main function is to test if the serial ports are normally working.
Details description:
Connect a tester with the serial port before the testing.
6)Audio Test
Testing function:
The main function is to test if the two buzzers and the related circuits are normally working.
7-23
System Maintenance
Details description:
During the test, the left side of machine will play out “Ding~~”, and pops up the following
dialog box to ask if you can hear “Ding~~” playing out of the left side machine. If you didn’t
hear it, please click “Retry” to play it again, and click “No, I Can’t”, if you still can’t hear it after
trying many times. And if you heard it, please click “Yes, I Can”.
After that, the program will control the right side of the machine to play out “Ding~~”, and pop
up the same dialog box, please do the same operation as described above.
7)ECG Module Port Test
Test function:
The main function is to test if the ECG module is normally working.
Details description:
After the ECG module is connected to the machine, it will communicate the system
self-checking module during the self-checking process, and the ECG module can be judged if
it is normally working according to the feedback information.
NOTE: If no ECG module is configured, this item is a Fail item during the
self-checking.
7-24
Troubleshooting
8 Troubleshooting
8.1 Malfunction
Blank Y
Blank screen?
screen
Replace master
N Can the machine be Replace CPU
board
started?
Machine Y
can’t Y
be started
Replace IO
Any display? N
port board
Replace monitor
Y Blank screen?
module
Over
8-1
Troubleshooting
Over No image
Y
Y
Noise-free
Remove all probes
Image?
Follow the
flows to
check blank
screen
Fig 8-2 Examine and Repair Flow When No Image Displayed in the Image Area
8-2
Troubleshooting
the Image
Black area
Over in the
image
Y
Y
Y N
Black area
Replace probe
exists with all N Black area?
control board
probes
Replace Y
Y
transmitting board
Replace probe
board
Over
Fig 8-3 Examine and Repair Flow When Black Area Exists in the Image
8-3
Troubleshooting
Start-up
Machine
can’t be
started
Any output in
N Replace adapter
adapter?
Y N
Replace IO
Can't be started? Y Over
port board
N
Over N Can’t be started? Y Replace CPU
Replace master
Y Can’t be started?
board
Fig 8-4 Examine and Repair Flow When the System Can’t Start-up
8-4
Troubleshooting
8.2.1.1 No Displaying in the LCD and the Indicator Lights are OFF
Check if CN1 is
Y
properly
connected
Replace the
N
power supply
Y
board
Reconnect the
cable
Note: LCD control board should match with the screen, and them should be replaced together if required.
8-5
Troubleshooting
8-6
Troubleshooting
8-7
Troubleshooting
8-8
Troubleshooting
8-9
Troubleshooting
normal, then test the output of U12, U30, U31, if error still exists,
and U14 to check if the output signal replace U13; if U30, U31 and
is normal and if there are distortion U13 are normal but error
in the signal. exists in output signals of
U12, U14, first check if the
peripheral resistors and
frequency
power supply of U12, U14
error.
are OK, if they are normal,
then replace U12, U14.
The pre-demodulation low-pass filter Check if there is damage to
is damaged. the filtering peripheral units,
and replace the damaged
units.
2.5V voltage input/output error in U2, Replace U6, U2 or U4.
and U4.
U1,U5 is damaged, which can be Replace U1 or U5.
concluded from the same input but
Amplitude
different output.
inconsistency
3 of the CW U13 is damaged, which can be Replace U13.
output signal I, concluded from the output
Q. inconsistency.
There are amplifier or peripheral Test and compare the testing
devices be damaged in the points, and then replace the
amplification stage or filtering unit of damaged units.
I and Q.
U13 is damaged. Replace U13
The pre-demodulation low-pass filter Check if there is damage to
is damaged. the filtering peripheral units,
and replace the damage
units.
High harmonic
Output signal saturation distortion in Check if any peripheral unit
wave appears
4 the amplification stage of I, Q. in the amplification stage is
in the CW
damaged, if any, please
output signal.
replace it.
I, Q wall filter output distortion. Check if there is damage to
the filtering peripheral units,
and replace the damage
units.
5 No CW output Electronic switch is damaged. Replace U11.
signal The low-pass filter is damaged. Check if any amplifier in the
amplification stage is
damaged, if any, replace it.
Demodulation IC U13 is damaged. Replace U13.
Amplifier damaged in the Replace the damaged units.
amplification stage and wall filter of I,
8-10
Troubleshooting
Q.
Damage to ADC IC U6 or signal end Replace U6, U2 or U4.
to difference circuit.
To judge the USB communication malfunction, you can test if L1 is working normally, if not
(the conducting resistor is less than 1Ohm), it means malfunction exists in L1. And if L1 is
normal, please replace IP4220CZ6 to judge the malfunction. At last, test should be done to
U3 and the peripheral circuit if the former method can’t determine the malfunction.
8-11
Troubleshooting
8-12
Troubleshooting
normally. is damaged.
testing of Multi-function FPGA is normal, but If the testing of voltage,
system voltage, ADT7462 can’t work normally. temperature and fan rotating
temperature, as speed is abnormal while the
well as fan multi-function FPGA is
rotating speed. normal, then the malfunction
must exist in ADT7462.
Multi-function FPGA can’t work Check if multi-function FPGA
normally. is damaged.
Malfunction in Multi-function FPGA is normal, but If Video/SVideo output is
7 Video & the video converting IC 25874 abnormal while multi-function
S-Video output can’t work normally. FPGA is normal, then the
malfunction must exist in
25874.
Check if the bus current to the If the USB port can’t be used
USB port can USB is more than 1A. because of surge, just to turn
8
not be applied. off the system and then
restart it.
8-13
Troubleshooting
Periphe
rals
90~140Vac Primary windings are
Main
unit
connected in parallel, which
increased the load capability
Periphe
capability
Main
rals
unit
Fig 8-7 Connections of the Transformer Winding
When the malfunction exists in the adapter of the power supply system or the DC power, the
malfunction can be determined according to the flows in Fig 8-8.
8-14
Troubleshooting
M a k e s u re th e b o a rd s
a re p ro p e rly c o n n e c te d
C o n n e c t A C a d a p te r a n d
AC pow er
M a lfu n c tio n in T H V , o r
N o rm a l s h o rt -c irc u it / o v e r-c u rre n t A b n o rm a l If T H V is n o rm a l?
in th e lo a d
If e d c _ p o w e r o n
M a lfu n c tio n in a d a p te r o r
p o w e r s u p p ly N o rm a l
s h o rt c irc u it o n p o w e r A b n o rm a l
m a s te r b o a rd is
s u p p ly m a s te r b o a rd
n o rm a l?
R e p a ir + 2 .5 V a n d 1 .5 V
If + 2 .5 V a n d 1 .5 V
u n its o n th e p o w e r A b n o rm a l
a re n o rm a l?
N o rm a l s u p p ly a u x ilia ry b o a rd
R e p a ir p o w e r s u p p ly N o rm a l
If 5 V s tb is
m a s te r b o a rd U 9 re la te d A b n o rm a l
n o rm a l?
u n it
R e p a ir -5 . 3 V a n d -1 2 V
If -5 .3 V a n d
u n its o n th e p o w e r A b n o rm a l
N o rm a l -1 2 V a re n o rm a l?
s u p p ly a u x ilia ry b o a rd
W hen N o rm a l
R e p a ir p o w e r s u p p ly
5 V s tb _ c p u _ e n is
m a s te r b o a rd 5 v s tb _ c p u A b n o rm a l
0 , if 5 V s tb _ c p u is
re la te d u n it
n o rm a l?
O ver
N o rm a l
R e p a ir th e p o w e r s u p p ly W h en P ow er_ o n
m a s te r b o a rd + 1 2 V A b n o rm a l is 0 , if + 1 2 V is
re la te d u n it n o rm a l?
N o rm a l
R e p a ir th e p o w e r s u p p ly
m a s te r b o a rd + 5 V A b n o rm a l If + 5 V is n o rm a l ?
re la te d u n it
N o rm a l
R e p a ir th e p o w e r s u p p ly
If + 3 .3 V is
m a s te r b o a rd + 3 .3 V A b n o rm a l
n o rm a l?
re la te d u n it
N o rm a l
R e p a ir th e p o w e r s u p p ly
If o u tp u t o f U 1 (-
m a s te r b o a rd U 1 re la te d A b n o rm a l
1 2 V ) is n o rm a l?
u n it
8-15
Troubleshooting
8-16
Troubleshooting
8-17
Troubleshooting
with VBUS.
Test the voltage of PIN5 (the reference voltage pin), which should be 1.25V.
Test the voltage of PIN8 (over-voltage/ over-current protection control pin), which should be
higher than 0.5V.
Test PIN14 (drive pin), which should output drive pulse.
8-18
Troubleshooting
Changes--Save CMOS To Flash; at last, click【Exit Saving Changes】 on【Exit】 page to exit.
All the data on the hard disk will be deleted after formatting!
WARNING
Note: To make a new hard disk, you have to re-install the XPE system (see XP Operation
System Troubleshooting), and run the user service ultrasound software restore program (see
Ultrasound Software System Troubleshooting); and then configure the machine according to
the model, at last, configure the optional functions according to the function CD-ROM owned
by the user. (see 7.2.7 and 7.2.8 ).
8-19
Troubleshooting
4. Change the BIOS setting after the hard disk is repaired. Enter into BIOS, cancel the
function to boot from USB-CDROM by setting the 【Boot Order】 on 【Boot】 page, (select the
USB CDROM by up arrow or down arrow, press the ‘Shift” key while pressing, at each
pressing of “-”, the USB CDROM will move down a step, we should put the item below the
HDD), set 【CMOS Restore Condition】 as Always on 【Exit】 page, then click Save
Changes--Save CMOS To Flash; at last, click【Exit Saving Changes】 on【Exit】 page to exit.
8-20
Troubleshooting
restore software.
1)The ultrasound system software must be setup in XP interface, if the XPE system is setup
just for a short time, the system can be restored by the ultrasound system restore software; if
the machine is still running under the ultrasound system, some simple operations are required
before restoring the system by the ultrasound system restore software.
1.1) To restore the ultrasound system software in the case that the XPE system is setup just
for a short time: put the prepared ultrasound system restore software CD-ROM to the
CD-ROM drive, the system software setup will be automatically started. When the interface
prompts “System need reboot!Please input Enter key to quit the Program!”, press [Enter] on
the control panel and the system will return to the Windows desk. Take out the CD-ROM and
press the system soft switch to turn off the main machine and then switch off the breaker.
After the system is booted, it will enter into the ultrasound system.
1.2)To restore the ultrasound system software in the case that the machine is still running
under the ultrasound system:
Press “Ctrl + Shift + =” on the control panel, and enter the correct password in the popped out
dialogue box, open the online debugger, and enter “shellapp off” in the debugging column,
press [Enter], turn off the machine and then reboot it. After the machine is booted, it will
directly enter into the Windows desk; delete the M5, and PatientBak file folders in C disk, and
the PADIENTDATA_2108 folder in D disk; put the prepared ultrasound system restore
CD-ROM into the CD-ROM drive, the setup will be automatically started, when it prompts
“System need reboot!Please input Enter key to quit the Program!”, press [Enter] and the
system will return to Windows desk. Take out the CD-ROM and press the system soft switch
to turn off the main machine and then switch off the breaker. After the system is started, it will
enter into the ultrasound system.;
2)If the ultrasound system is restored, the machine has to be configured, also, the optional
functions should be configured according to the function CD-ROM owned by the user. (see
7.2.7 and 7.2.8 )
After the ultrasound software system can be applied on the
WARNING:
machine, the original ultrasound software, the related
information (including the user preset files and backup
information in patient data bank ), as well as the patient
information in the D disk will be deleted when restore the
system of the machine, so remember to backup all the
information before deletion.
8-21
Troubleshooting
8-22
Definition of Commonly Used Sockets and Functions in
System Maintenance
Figure A-1 Layout of Each Module and Socket on the Master Board
Ports of probe module
CW board
Receiving unit
T ra n sm ittin g
b o a rd
CPU module
Dual-USB
SVIDEO
Network
Adapter
port
port
port
1
Definition of Commonly Used Sockets and Functions in System Maintenance
Table 8-1-1 Sockets between the Master Board and Probe Board
CON1 CON2
PIN PIN PIN
PI NAM PI NAM PI NAM PI PIN PI PIN PI PIN
N E N E N E N NAME N NAME N NAME
1 GND 2 PE1 3 GND 1 PE69 2 PE70 3 PE71
4 PE2 5 GND 6 PE3 4 PE72 5 PE73 6 PE74
7 GND 8 PE4 9 PE5 7 PE75 8 GND 9 PE76
10 PE6 11 PE7 12 GND 10 GND 11 PE77 12 PE78
13 PE8 14 GND 15 PE9 13 PE79 14 PE80 15 GND
PE1 PE1
16 GND 17 0 18 1 16 PE81 17 GND 18 PE82
PE1 PE1
19 2 20 3 21 GND 19 GND 20 PE83 21 PE84
PE1 PE1
22 4 23 GND 24 5 22 PE85 23 PE86 24 GND
PE1 PE1
25 GND 26 6 27 7 25 PE87 26 GND 27 PE88
PE1 PE1
28 8 29 9 30 GND 28 GND 29 PE89 30 PE90
PE2 PE2
31 0 32 GND 33 1 31 PE91 32 PE92 33 GND
PE2 PE2
34 GND 35 2 36 3 34 PE93 35 GND 36 PE94
PE2 PE2
37 4 38 5 39 GND 37 GND 38 PE95 39 PE96
PE2 PE2
40 6 41 GND 42 7 40 PE97 41 PE98 42 GND
PE2 PE2
43 GND 44 8 45 9 43 PE99 44 GND 45 PE100
PE3 PE3 PE10
46 0 47 1 48 GND 46 PE101 47 2 48 PE103
PE3 PE3 PE10
49 2 50 GND 51 3 49 PE104 50 5 51 GND
PE3 PE3
52 GND 53 4 54 5 52 PE106 53 GND 54 PE107
PE3 PE3 PE10
55 6 56 7 57 GND 55 GND 56 8 57 PE109
PE3 PE3 PE11
58 8 59 GND 60 9 58 PE110 59 1 60 PE112
PE4 PE4
61 GND 62 0 63 1 61 PE113 62 GND 63 PE114
PE4 PE4 PE11
64 2 65 3 66 GND 64 PE115 65 6 66 PE117
PE4 PE4 PE11
67 4 68 GND 69 5 67 PE118 68 9 69 GND
2
Definition of Commonly Used Sockets and Functions in System Maintenance
PE4 PE4
70 GND 71 6 72 7 70 PE120 71 GND 72 PE121
PE4 PE4 PE12
73 8 74 9 75 GND 73 GND 74 2 75 PE123
PE5 PE5 PE12
76 0 77 GND 78 1 76 PE124 77 5 78 PE126
PE5 PE5
79 GND 80 2 81 3 79 PE127 80 GND 81 PE128
PE5 PE5 SPI_C
82 4 83 5 84 GND 82 SPI_CLK 83 S 84 SPI_DIN
PE5 PE5 SPI_DOU FLAS
85 6 86 GND 87 7 85 T 86 H_WP 87 GND
PE5 PE5 FLASH_P PROBE_
88 GND 89 8 90 9 88 OWER 89 GND 90 ID6
PE6 PE6 PROB PROBEI
91 0 92 1 93 GND 91 GND 92 E_ID7 93 D_CS0
RELA
PE6 PE6 PROBEID Y_EN RELAY_
94 2 95 GND 96 3 94 _CS1 95 0 96 EN1
PROBE_
PE6 PE6 EXP_PRE PRESEN
97 GND 98 4 99 5 97 SENT 98 GND 99 T
10 PE6 10 PE6 10 10 10 10
0 6 1 7 2 GND 0 GND 1 GND 2 GND
10 10 10 PE6 10 10 10
3 GND 4 GND 5 8 3 VCC 4 VCC 5 VCC
CON1
3
Definition of Commonly Used Sockets and Functions in System Maintenance
CON2
4
Definition of Commonly Used Sockets and Functions in System Maintenance
CON1
5
Definition of Commonly Used Sockets and Functions in System Maintenance
CON2
6
Definition of Commonly Used Sockets and Functions in System Maintenance
SIGNA SIGNA
PIN L PIN L
1 GND 2 GND
3 CW0 4 CW1
5 GND 6 GND
7 CW2 8 CW3
9 GND 10 GND
11 CW4 12 CW5
13 GND 14 GND
15 CW6 16 CW7
17 GND 18 GND
19 CW8 20 CW9
21 GND 22 GND
23 GND 24 GND
25 12V 26 12V
27 GND 28 GND
29 -12V 30 -12V
31 GND 32 GND
7
Definition of Commonly Used Sockets and Functions in System Maintenance
33 5V 34 5V
35 GND 36 GND
37 -5V 38 -5V
39 GND 40 GND
PIN
NUM Signal
name Note
1 AD_NRST AD reset signal( Alreadyposited to the right position, can be reserved)
2 RESERVED Reserved, not be used temporarily
3 RESERVED Reserved, not be used temporarily
4 RESERVED Reserved, not be used temporarily
5 GND
6 GND
7 LOCLK_P Quadrature demodulation local oscillation input (If it is difference, it is positive end)
8 RESERVED Reserved, not be used temporarily
9 LOCLK_N Quadrature demodulation local oscillation input (If it is difference, it is negative end)
10 RESERVED Reserved, not be used temporarily
11 GND
12 GND
13 RESERVED Reserved, not be used temporarily
14 RESERVED Reserved, not be used temporarily
15 RESERVED Reserved, not be used temporarily
16 AD_HPFD AD internal high-pass filter enable control
(Already posited to normal open, can be reserved)
17 GND
18 GND
19 AD audio serial port left/ right output clock
AD_LRCK (Audio serial port left/ right (or word) clock)
20 AD audio serial port output data
AD_DATA (Audio serial port left and right channel PCM data)
21 AD audio serial port output bit clock
AD_BCK (Audio serial port bit clock)
22 SW_CTRL Analog switch channel selection control signal
23 GND
24 GND
25 TPU Input of pencil probe drive signal
26 RESERVED Reserved, not be used temporarily
27 GND
28 GND
29 RESERVED Reserved, not be used temporarily
30 RESERVED Reserved, not be used temporarily
8
Definition of Commonly Used Sockets and Functions in System Maintenance
Table A-5 Signal Definition between the Master Board and 4D Board
Table A-6 Pins Definition between the Master Board and AD Control Board
9
Definition of Commonly Used Sockets and Functions in System Maintenance
9 12V 10 12V
11 GND 12 GND
13 GND 14 GND
15 GND 16 GND
17 GND 18 GND
19 5V 20 5V
21 5V 22 5V
23 3.3V 24 3.3V
25 3.3V 26 3.3V
27 3.3V 28 3.3V
29 3.3V 30 3.3V
31 GND 32 GND
33 GND 34 VGA_SDA
35 GND 36 GND
37 GND 38 VGA_SCL
39 GND 40 GND
41 GND 42 VGA_VSYNC
43 GND 44 GND
45 GND 46 VGA_HSYNC
47 GND 48 GND
49 GND 50 VGA_BLUE
51 GND 52 GND
53 GND 54 VGA_GREEN
55 GND 56 GND
57 GND 58 VGA_RED
59 GND 60 GND
Table A-7 Pins Definition between the Master Board and Power Supply Module
10
Definition of Commonly Used Sockets and Functions in System Maintenance
25 Temperature_D+ 26 CW_Mode
27 Temperature_D- 28 +5VStb_CPU_En
29 Gnd 30 Gnd
31 +5VStb 32 +5VStb_CPU
33 +12V 34 +12V
35 Gnd 36 Gnd
37 GND 38 GND
39 -12V 40 -5V
41 Gnd 42 Gnd
43 +5V 44 +5V
45 Gnd 46 Gnd
47 +2.5V 48 +2.5V
49 Gnd 50 Gnd
51 Gnd 52 NC
53 +1.5V 54 +1.5V
55 Gnd 56 Gnd
57 3.3V 58 3.3V
59 Gnd 60 Gnd
61 Gnd 62 Gnd
63 THV 64 THV_Range
Table A-8 Signal Definition between the Master Board and Power Supply Module
11
Definition of Commonly Used Sockets and Functions in System Maintenance
Table A-9 Signal Definition between Mater Board and Extended Socket
PIN Signal Name PIN Signal Name PIN Signal Name PIN Signal Name
A1 USB+4 B1 USB-4 C1 USB+5 D1 USB-5
A2 VBUS_SYS45 B2 GND C2 VBUS_SYS45 D2 GND
A3 GND B3 VGA_HS C3 VGA_RED D3 VGA_GREEN
A4 UART_RX0 B4 VGA_VS C4 GND D4 GND
A5 UART_TX0 B5 GND C5 VGA_BLUE D5 D0
A6 GND B6 AUDIO_L C6 GND D6 D1
A7 RM_PRINT B7 AUDIO_R C7 D2 D7 GND
A8 RM_BUSY B8 GND C8 D3 D8 AUTOFD_N
A9 GND B9 D7 C9 D4 D9 BUSY
A10 MIC_L B10 PE C10 GND D10 GND
A11 MIC_R B11 STROBE_N C11 D5 D11 ACK_N
A12 VIDEO B12 INIT_N C12 D6 D12 SELIN_N
\ \ B13 SEL \ \ D13 ERROR_N
Table A-10 Signal Definition between Master Board and Keyboard Board
12
Definition of Commonly Used Sockets and Functions in System Maintenance
11 +5V 12 +5V
13 Gnd 14 Gnd
15 Gnd 16 Gnd
17 Gnd 18 Gnd
19 Work_Status_O 20 Work_Status_G
21 Battery_Status_O 22 Battery_Status_G
23 Gnd 24 EDC_Status_G
25 USB_DN 26 LCD_SW
27 USB_DP 28 Power_BTN
29 Gnd 30 Gnd
13