07.DC-3 Service Manual v1.0

Download as pdf or txt
Download as pdf or txt
You are on page 1of 229

DC-3 Series

Diagnostic Ultrasound System

Service Manual
© 2008 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All Rights Reserved.

Product Information
Product Name: Diagnostic Ultrasound System
Issued date of this manual: 2008-05
Version: 1.0

Intellectual Property Statement


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this manual. This
manual may refer to information protected by copyrights or patents and does not convey any
license under the patent rights of Mindray, nor the rights of others. Mindray does not assume
any liability arising out of any infringements of patents or other rights of third parties.
Mindray intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner without the written permission of
Mindray is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation or translation of this
manual, in any manner whatsoever without the written permission of Mindray, is strictly
forbidden.

, , , , BeneView, WATO, BeneHeart,

are the registered trademarks or trademarks owned by Mindray in China and

other countries. All other trademarks that appear in this manual are used only for editorial
purposes without the intention of improperly using them. They are the property of their
respective owners.

Responsibility on the Manufacturer Party


Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with
the furnishing, performance, or use of this manual.

I-1
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
l All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Mindray authorized personnel;
l The electrical installation of the relevant room complies with the applicable national and
local requirements;
l The product is used in accordance with the instructions for use.

It is important for the hospital or organization that employs


WARNING:
this system to carry out a reasonable service / maintenance
plan. Neglect of this may result in machine breakdown or
personal injury.

Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.

Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not
approved by Mindray or repairs by people other than Mindray authorized personnel.
This warranty shall not extend to:
l Any Mindray product which has been subjected to misuse, negligence or accident;
l Any Mindray product from which Mindray's original serial number tag or product
identification markings have been altered or removed;
l Any products of any other manufacturers.

I-2
Return Policy

Return Procedure
In the event that it becomes necessary to return this product or part of this product to
Mindray, the following procedure should be followed:
1. Return authorization: Contact the Customer Service Department and obtain a Customer
Service Authorization number. This number must appear on the outside of the shipping
container. Returned shipments will not be accepted if the number is not clearly visible.
Please provide the model number, serial number, and a brief description of the reason
for return.
2. Freight policy: The customer is responsible for freight charges when this product is
shipped to Mindray for service (this includes customs charges).
3. Return address: Please send the part(s) or system to the address offered by the
Customer Service Department.

Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building, Keji 12th Road South, Hi-tech Industrial Park,
Nanshan, ShenZhen 518057, P. R. China
Tel: +86 755 26582888

Fax: +86 755 26582680

I-3
Safety Precautions

1. Meaning of Signal Words

In this operator’s manual, DANGER, WARNING, CAUTION and

CAUTION are signal words used to indicate safety and other important instructions. The
signal words and their meanings are defined as follows. Please understand their meanings
clearly before reading this manual.

Signal Word Meaning

Indicates death or serious injury may occur imminently in this


DANGER hazardous situation if not avoided.

Indicates death or serious injury may occur potentially in this


WARNING hazardous situation if not avoided.

Indicates minor or moderate injury may occur potentially in


CAUTION this hazardous situation if not avoided.

Indicates property damage may occur potentially in this hazardous


CAUTION situation if not avoided.

2. Meaning of Safety Symbols


Symbol Description
Type-BF applied part.
The ultrasound transducers connected to this system are Type-BF applied
parts.
The ECG module connected to this system is also a Type-BF applied part.
“Attention” indicates the points requiring attention. Be sure to read the
operation manual concerning these points before using the equipment.

S-1
3. Safety Precautions

Please read the following precautions carefully to ensure the safety of the patient and the
operator when using the system.

Do not operate this system in an atmosphere containing flammable


DANGER:
or explosive gases such as anesthetic gases, oxygen, and
hydrogen because an explosion may occur.

WARNING: 1 Do connect the power plug of this system and power plugs of
the peripherals to wall receptacles that meet the ratings
indicated on the rating nameplate. Using an adapter or
multifunctional receptacle may affect the system’s grounding
performance, and cause the leakage current to exceed safety
requirements.
Besides, do connect the video printer to the auxiliary power
supply socket of this system, and use the printing cable
provided with this system to connect the printer. Otherwise, it
may cause electric shock.
2 Connect the grounding conductor only before turning ON the
system. Connect the grounding cable after turning OFF the
system. Otherwise, electric shock may result.
3 For the connection of power and grounding, follow the
appropriate procedures described in this operator’s manual.
Otherwise, there is risk of electric shock. Do not connect the
grounding cable to a gas pipe or water pipe; otherwise
improper grounding may result or a gas explosion may occur.
4 Before cleaning the system, disconnect the power cord from
the socket. System failure may result in electric shock.
5 This system is not water-proof. Do not use this system in any
place where water leakage may occur. If any water is sprayed
on or into the system, electric shock may result. If water is
accidentally sprayed on or into the system, please turn off
the system power immediately and contact Mindray
Customer Service Department or sales representative.
6 Store and use the transducers carefully. In case that a
scratched transducer surface is found, immediately stop
using the transducer and contact Mindray Customer Service
Department or sales representative. There is risk of electric
shock if using a damaged or scratched transducer.
7 Do not allow the patient to contact the live parts of the
ultrasound system or other devices, e.g. signal I /O ports. If
the system or other devices is defective, there is risk of
electric shock.

S-2
8 Do not use an aftermarket transducer other than those
specified by Mindray. The transducers may damage the
system, causing a profound failure, e.g. a fire in the worst
case.
9 Do not subject the transducers to knocks or drops. Use of a
defective transducer may cause an electric shock.
10 Do not open the covers and front panel of the system.
Otherwise, short circuit or electric shock may result.
11 Do not use this system simultaneously with equipments such
as an electrical knife, high-frequency therapy equipment, or a
defibrillator, etc., otherwise, electric shock may result.
12 Only use the ECG leads accompany with the ECG module;
otherwise electric shock may result.
13 Attentions in the transportation: when this system is moved,
please hold the handle. If other parts of the system are held, it
may cause damage due to the abnormal force. Do not push
the system from the left/right side, otherwise, it may be
toppled over.
14 Accessory equipments connected to the analog and digital
ports must comply with the relevant IEC standards (e.g., IEC
60950 information technology equipment safety standard and
IEC 60601-1 medical equipment standard). Furthermore, all
configurations must comply with the standard IEC60601-1-1.
It is the responsibility of the person, who connects additional
equipments to the signal input or output ports and configures
a medical system, to verify if the system complies with the
requirements of IEC60601-1-1. If you have any questions
regarding these requirements, consult your sales
representative.
15 Prolonged and repeated use of keyboards may result nerve
disorders in hand or arm for some individuals. Observe the
local safety or health regulations concerning the use of
keyboards.

CAUTION: 1 Precautions concerning clinical examination techniques:


(1) This system must be used only by qualified medical
professionals.

(2) This operator’s manual does not describe clinical


examination techniques. The clinician should select the
proper examination techniques based on specialized
training and clinical experience.

S-3
2 Malfunctions due to radio wave:

(1) Using devices transmitting RF signals in the vicinity of the


system may affect the system’s performance. Do not use
or take any devices transmitting RF signals (such as
cellular phones, transceivers and radio controlled
products) in the room placing the system.

(2) If a person brings a device that generates radio waves


near the system, ask him / her to immediately turn OFF the
device.

3 Precautions concerning moving the system:

(1) Make sure the system is installed on a plane with the


caster been locked, otherwise, movement of the system
may result in damage.

(2) Do not move the system laterally, which may result in


damage in case of toppling.

(3) Move the system slowly on the slope by two people,


otherwise, damage may result in case of unexpected
sliding.

(4) Do not sit on the system, which may result individual


falling in case of system moving.

(5) Object placed on the monitor may fall and injure an


individual.

(6) Fasten and fully secure any peripheral device before


moving the system. A loose peripheral device may fall
and injure an individual.

(7) When move the system on the steps, please take care to
prevent the system from toppling.

4 Do not expose the system to excessive vibration during the


transportation. Mechanical damage may result (such as
damage to the caster). If the system has to be moved on
uneven floor frequently, please contact Mindray Customer
Service Department or sales representative.

5 Do not connect this system to outlets with the same breakers


and fuses that control the current of devices such as
life-support systems. If malfunctions or over-current appear
on this system, or when there is an instantaneous current at
power ON, the breakers and fuses of the building’s supply
circuit may be tripped.

S-4
6 Always keep the system dry. Avoid transporting this system
suddenly from a cold place to a warm place; otherwise
condensation or water droplets may form, which may result
short circuit.

7 If the circuit protector is tripped on, it indicates that


malfunction appears on the system or a peripheral device.
You should not repair the system under this circumstance but
call the Mindray Customer Service Department or sales
representative for help.

8 There is no risk of high-temperature burns during normal


ultrasound examinations. It is possible for the surface
temperature of the transducer to exceed the body
temperature of a patient due to environmental temperature
and exam type combinations. If a patient complains of
abnormal heat from the transducer, immediately stop
scanning. To prevent patient burns, ensure there is no
surface damage to the transducer. Do not apply the
transducer to the same region on the patient for a long time.
Apply the transducer only for a period of time required for the
purpose of diagnosis.

9 The system and its accessories have not been disinfected or


sterilized prior to delivery. The operator is responsible for the
cleaning and disinfection of transducers and sterilization of
biopsy brackets according to the manuals before application.
All items must be thoroughly processed to completely
remove harmful residual chemicals, which will not only
harmful to the human body, but also damage the accessory.

10 It is necessary to press [End Exam] to end the current scan


and clear the current Patient Information field. Otherwise new
patient data may be mixed with the previous patient data.

11 Do not connect or disconnect the system’s power cord or its


accessories (e.g., a printer or a recorder) without turning OFF
the power first. This may damage the system and its
accessories or cause electric shock.

12 Do not turn OFF the power supply of the system during


printing, file storage or data invoking. An interrupted work
may not be completed afterwards, and the file may be lost or
corrupted.

13 If the system is powered off improperly during operation, it


may result in data damage to the system’s hard disk or
system failure.

S-5
14 Do not use the system to examine a fetus in the Doppler
mode for a long period of time.

15 Do not use a USB memory device (e.g., a USB flash drive, or


removable hard disk) which may include unsafe data.
Otherwise, system damage may result.

16 It is recommended to use the video devices specified in this


manual.

17 Read the Acoustic Output Principle in the operation manual


carefully before operate this system on clinical examination.

18 The auxiliary power output outlet in the system is used to


supply power for the recommended peripheral devices. Do
not connect other devices to the outlet, otherwise the rated
output power may be exceeded and failure may result.
Maximum output outlet for peripheral devices is 330V.

NOTE: 1 Do not use the system in the vicinity of strong electromagnetic


field (such as a transformer), which may affect the performance of
the system.

2 Do not use the system in the vicinity of high-frequency radiation


source, which may affect the performance of the system or even
lead to system failure.

3 To avoid damage to the system, do not use it in the following


environment conditions:

(1) Locations exposed to direct sunlight;

(2) Locations subject to sudden changes of environmental


temperature;

(3) Dusty locations;

(4) Locations subject to vibration;

(5) Locations near heat generators;

(6) Locations of high humidity.

4 Turn ON the system only after the power is turned OFF for more
than 20 seconds. If the system is turned ON immediately after
being turned OFF, the system may not reboot properly and may
lead to malfunction.

S-6
5 Do not disconnect a transducer that remains in a live imaging
state. This can damage the system and / or transducer. Press
[Freeze] or turn off the power of the system before connecting or
disconnecting a transducer.

6 Remove the ultrasound gel from the face of the transducer when
the examination is completed. Water in the gel may enter the
acoustic lens and adversely affect the performance and safety of
the transducer.

7 To ensure safety of the data, please back up all the data to a


secure external storage media, as data stored to the system’s
internal memory may be lost due to system failure, or improper
operation.

8 Do not apply external force to the control panel. The system may
be damaged.

9 If the system is used in a small room, the room temperature may


rise. Please ensure good ventilation.

10 To dispose the system or any part, please contact Mindray


Customer Service Department or sales representative. Mindray
would bear no responsibility for the damages resulting from
disposal of this system without consulting Mindray.

11 Electrical and mechanical performance may be degraded due to


long usage (such as current leakage or distortion and abrasion).
To ensure optimal system operations, it is recommended that you
maintain the system under a Mindray service agreement. Various
aspects of system performance and operation can be maintained
under the professional supervision of a service representative.

12 The iScape feature constructs a single extended image from a


series of continuous individual image frames scanned on the
interesting field. The quality of the final image is user-dependent
and thus requires skills to efficiently apply the feature and
technique. Exercise caution when measurements are performed
under an iScape mode.

13 Ensure that the current exam date and time are the same as the
system date and time.

4. Warning Labels

The warning labels are attached to this system in order to call your attention to potential
hazards.

The symbol on the warning labels indicates safety precautions. The warning labels use

S-7
the same signal words as those used in the operator’s manual.
Please refer to the operator’s manual for detailed information about the warning labels. Read
operator’s manual carefully before using the system.
The name, pattern and meaning of each warning label are described as follows:

No. Label Meaning

<1> (a) Do not sit on the system.

(b) There is explosion risk if the system is


applied around flammable gas.

(c) Before using the system, be sure to


carefully read the relevant contents of this
operator’s manual.

<2> (a) Place the system on a plane.


(b) (b) Do not open the cover, because the
high voltage inside may result in electric
shock.

S-8
Table of Contents

1 System Introduction............................................................................. 1-1

1.1 Intended Use ................................................................................................................1-1

2 System Structure.................................................................................. 2-1

2.1 Appearance of the System...........................................................................................2-1

2.2 LCD Monitor .................................................................................................................2-3

2.2.1 Contrast& Brightness Control Key of the Monitor .............................................2-4

2.2.2 Up/down Deflector Rod of Monitor ....................................................................2-4

2.2.3 Up/down Lock of Upper Support Arm................................................................2-6

2.2.4 Tuning Screw of Air Spring on Support Arm ......................................................2-6

2.3 Turning and Lifting the Keyboard .................................................................................2-7

2.3.1 Turning...............................................................................................................2-7

2.3.2 Lifting .................................................................................................................2-8

2.4 I/O Panel ......................................................................................................................2-9

2.5 Power Supply Panel ...................................................................................................2-10

2.6 ECG Panel ................................................................................................................. 2-11

2.7 Control Panel..............................................................................................................2-12

3 System Principle .................................................................................. 3-1

3.1 System Principle...........................................................................................................3-1

3.2 Working Principles of Each Module in the Hardware System......................................3-3

3.2.1 Master Board .....................................................................................................3-3

3.2.2 Ultrasound Receiving ........................................................................................3-4

3.2.3 Signal Processing..............................................................................................3-4

3.2.4 Power Management ..........................................................................................3-5

3.2.5 System Monitoring.............................................................................................3-5

3.2.6 Video Processing...............................................................................................3-6

3.2.7 Sockets on Master Board ..................................................................................3-6

3.2.8 Testing Points on the Master Board ..................................................................3-9

C-1
3.3 Transmitting Board .....................................................................................................3-12

3.4 Probe Board Module ..................................................................................................3-13

3.4.1 Principle of Probe Board Module ....................................................................3-13

3.4.2 Definition of Sockets........................................................................................3-15

3.5 I/O Connecting Board.................................................................................................3-19

3.5.1 Definition of Sockets on I/O Connection Board...............................................3-19

3.6 IO Front Board............................................................................................................3-20

3.6.1 USB .................................................................................................................3-20

3.6.2 Sockets Definition of IO Front Board...............................................................3-21

3.7 IO Port Board .............................................................................................................3-22

3.7.1 Functional Modules .........................................................................................3-23

3.7.2 Sockets Definition of I/O Port Board................................................................3-25

3.7.3 Introduction of Testing Points on IO Port Board ..............................................3-28

3.8 IO Rear Panel.............................................................................................................3-29

3.8.1 Sockets Definition of IO Rear Panel................................................................3-29

3.9 CW Board...................................................................................................................3-29

3.9.1 Functional Modules Introduction .....................................................................3-30

3.9.2 Sockets Definition............................................................................................3-34

3.9.3 Layout of CW Board ........................................................................................3-35

3.10 Control Panel..............................................................................................................3-36

3.10.1 The Buckled Board on the Control Panel........................................................3-36

3.10.2 Control Panel...................................................................................................3-37

3.11 ECG Module (optional)...............................................................................................3-39

3.11.1 Principle and Functions ...................................................................................3-39

3.11.2 Outlet Definition ...............................................................................................3-40

3.11.3 Plug-in Diagram and Components ..................................................................3-42

3.12 USB-to-IDE Port Board ..............................................................................................3-43

3.12.1 Socket Definition .............................................................................................3-44

3.13 Power Supply System ................................................................................................3-45

3.13.1 Basic Functions of Power Supply System.......................................................3-45

3.13.2 Basic Principle of Power Supply System ........................................................3-45

C-2
3.14 System Power Supply Distribution .............................................................................3-48

3.15 LCD ............................................................................................................................3-49

3.15.1 Inverter ............................................................................................................3-49

3.15.2 AD Controller Board ........................................................................................3-52

4 System Structure and Assembly / Disassembly................................ 4-1

4.1 Overall Exploded View .................................................................................................4-1

4.2 Exploded View of Each Module ...................................................................................4-3

4.2.1 Exploded View of LCD Display and Support Arm .............................................4-3

4.2.2 Exploded View of Monitor Components-install Board .......................................4-4

4.2.3 Exploded View of Main Unit Box Module ..........................................................4-5

4.2.4 Exploded View of Probe Board Module ............................................................4-7

4.3 Overall Disassemble and Parts Replace......................................................................4-7

4.3.1 Disassemble the Monitor Module ......................................................................4-7

4.3.2 Disassemble the Bottom Lights of the Monitor..................................................4-9

4.3.3 Replace the Speaker on the Monitor...............................................................4-10

4.3.4 Replace the OSD Board of Monitor................................................................. 4-11

4.3.5 Replace the Bottom Light Board of the Monitor ..............................................4-12

4.3.6 Replace the Power Supply Inverter Board of the LCD....................................4-13

4.3.7 Replace the LCD Screen and LCD Control Board ..........................................4-15

4.3.8 Replace the Air Spring of Support Arm............................................................4-17

4.3.9 Disassemble the Support Arm.........................................................................4-20

4.3.10 Disassemble the Control Panel .......................................................................4-21

4.3.11 Replace the Encoder Connecting Board.........................................................4-23

4.3.12 Replace the Trackball......................................................................................4-24

4.3.13 Disassemble the Control Panel Buckled Board ..............................................4-25

4.3.14 Replace the Control Panel ..............................................................................4-25

4.3.15 Disassemble the Casting Aluminium Base Cover ...........................................4-27

4.3.16 Disassemble the Up/ Down Handle ................................................................4-28

4.3.17 Disassemble the Turning Handle Module .......................................................4-28

4.3.18 Disassemble the Turning Block Cover ............................................................4-29

C-3
4.3.19 Disassemble the Frontal Cover of the Main Unit.............................................4-29

4.3.20 Disassemble the Rear Cover of the Main Unit ................................................4-31

4.3.21 Disassemble the IO Module ............................................................................4-31

4.3.22 Disassemble the Power Supply Input Module.................................................4-32

4.3.23 Disassemble the DVD-R/W Module ................................................................4-33

4.3.24 Disassemble the DVD and USB-to-IDE Port Board ........................................4-34

4.3.25 Disassemble the Power Supply Adapter .........................................................4-35

4.3.26 Disassemble the IO Frontal Board and ECG Board........................................4-36

4.3.27 Disassemble the Probe Board.........................................................................4-37

4.3.28 Replace the Air Spring.....................................................................................4-40

4.3.29 Disassemble the Isolating Transformer...........................................................4-43

4.3.30 Replace the HDD ............................................................................................4-44

4.3.31 Disassemble the Main Unit Box Module .........................................................4-45

4.3.32 Replace the IO Port Board ..............................................................................4-47

4.3.33 Replace the IO Connecting Board ..................................................................4-49

4.3.34 Disassemble the Transducer Connecting Board.............................................4-51

4.3.35 Replace the Power Supply Main Board and Auxiliary Board ..........................4-52

4.3.36 Replace the Transmitting Board ......................................................................4-54

4.3.37 Replace the CW Board....................................................................................4-55

4.3.38 Replace the CPU Board ..................................................................................4-56

4.3.39 Replace the Master Board...............................................................................4-58

4.3.40 Replace the Air Inlet Fan .................................................................................4-59

4.3.41 Disassemble the Dust Net...............................................................................4-60

4.3.42 Replace the Caster..........................................................................................4-60

4.4 Installation of Peripherals ...........................................................................................4-61

4.4.1 Installation of W/B Video Printer......................................................................4-61

5 Maintenance Requirements................................................................. 5-1

5.1 Tools Used in Maintenance ..........................................................................................5-1

5.1.1 Tools, Measurement Devices, Consumables ....................................................5-1

5.1.2 Standard Configuration of Tools and Consumables ..........................................5-2

C-4
5.2 Maintenance Personnel ...............................................................................................5-2

6 Checking 6-1

6.1 System Status Checking ..............................................................................................6-1

6.1.1 Checking Plan ...................................................................................................6-1

6.1.2 Confirmation before Checking...........................................................................6-1

6.2 Checking Functionality .................................................................................................6-2

6.2.1 Flow of Functionality Checking..........................................................................6-2

6.2.2 Details of Functionality Check ...........................................................................6-2

6.3 Safety Checking ...........................................................................................................6-4

6.3.1 Electrical Safety Checking.................................................................................6-4

6.3.2 Mechanical Safety Checking ...........................................................................6-12

6.4 Image Checking .........................................................................................................6-15

6.4.1 Checking the B/W Image Phantom Data and Image Recording.....................6-15

6.4.2 Checking Phantom Data .................................................................................6-15

6.4.3 Final Operation Checking and Image Archiving ..............................................6-16

6.4.4 Color flow image check ...................................................................................6-16

7 System Maintenance............................................................................ 7-1

7.1 System Cleaning ..........................................................................................................7-1

7.1.1 Cleaning Flow....................................................................................................7-1

7.1.2 Details for Cleaning ...........................................................................................7-2

7.2 Software Maintenance..................................................................................................7-4

7.2.1 Confirming the System Starting.........................................................................7-4

7.2.2 Preset ................................................................................................................7-5

7.2.3 Viewing System Information..............................................................................7-5

7.2.4 Preset Data Management .................................................................................7-6

7.2.5 Maintenance Status ...........................................................................................7-7

7.2.6 Software Upgrading...........................................................................................7-8

7.2.7 Model Configuration ........................................................................................ 7-11

7.2.8 Install the Software of Optional Devices..........................................................7-12

7.3 System Self-diagnosis................................................................................................7-14

C-5
7.3.1 Self-diagnosis Interface...................................................................................7-14

7.3.2 Description of Testing Items ............................................................................7-16

7.3.3 Details of the Testing Items .............................................................................7-17

8 Troubleshooting ................................................................................... 8-1

8.1 Malfunction ...................................................................................................................8-1

8.1.1 Examine and Repair Flow of Blank Screen.......................................................8-1

8.1.2 Examine and Repair Flow When No Image Displayed in the Image Area........8-2

8.1.3 Examine and Repair Flow When Black Area Exists in the Image.....................8-3

8.1.4 Abnormal Image in the Image Area...................................................................8-3

8.1.5 CW Image Abnormity ........................................................................................8-4

8.1.6 Examine and Repair Flow When the System Can’t Start-up.............................8-4

8.2 Modules and Boards Malfunction.................................................................................8-5

8.2.1 LCD Display.......................................................................................................8-5

8.2.2 General Malfunctions of the Control Panel Unit ................................................8-7

8.2.3 General Malfunctions of the Probe Board .........................................................8-8

8.2.4 General Malfunctions of CW Doppler Board .....................................................8-9

8.2.5 General Malfunctions of USB-to-IDE Port Board ............................................ 8-11

8.2.6 PC System Related Malfunctions....................................................................8-12

8.2.7 Power Supply Malfunction...............................................................................8-13

8.3 Software Troubleshooting ..........................................................................................8-18

8.3.1 Troubleshooting in the Case of HDD Damaged ..............................................8-18

8.3.2 XP Operation System Troubleshooting ...........................................................8-19

8.3.3 Ultrasound Software System Troubleshooting ................................................8-20

8.3.4 Patient Databank Troubleshooting ..................................................................8-22

C-6
1 System Introduction

1.1 Intended Use


DC-3/DC-3T diagnostic ultrasound system is applicable for clinical ultrasound diagnostic
examination.
There is one difference between DC-3 and DC-3T, that is, DC-3 has an orthopedics
measurement package, but DC-3T does not.
DC-3Vet is intended for diagnosis of animals.

1-1
System Structure

2 System Structure

2.1 Appearance of the System

<1>

<8>
<2>

<9>

<3>

<4> <10>

<5>

<6>

<7>

2-1
System Structure

<11>

<12>

<13>

<14>

No. Part Function

<1> Monitor Displays the images and parameters during scanning.


<2> Control panel Operator-system interface, operation control.
<3> DVD-RW DVD-RW drive.
<4> Power switch Used for turning on/ off the power supply.
Compartment for placing Used for placing B/W video printer.
<5>
video printer
Sockets connecting transducers and the main unit. There
<6> Transducer socket
are 3 active sockets and 1 parking socket.
<7> Caster Used for fixing or moving the system.
<8> Transducer& gel holder Used for placing transducers and gel temporarily.

2-2
System Structure

<9> Handle Used for pushing and moving the system.


Used for connecting the ECG cable, USB devices and
<10> ECG panel
footswitch.
Used for supporting and adjusting the height and position of
<11> Monitor support arm
monitor.
<12> Table for placing objects Used for placing articles and instruments and so on.
<13> I/O panel Port panel for input and output signals.
<14> Power panel Electrical port panel.

2.2 LCD Monitor

2-3
System Structure

2.2.1 Contrast& Brightness Control Key of the Monitor

Shown as figure at the right side:


<1>, <2> refer to the brightness control keys with
a sun marked at the top side; key <1>, which
marked a “-” on the top side, can be used to
decrease the brightness; while key<2>, which
marked a “+” on the top side, can be used to
increase the brightness
<3>, <4> refer to the contrast control keys with a
lune marked at the top side; key <3>, which
marked a “-” on the top side, can be used to
decrease the contrast, while key <4>, which
marked a “+” on the top side, can be used to
increase the contrast.

2.2.2 Up/down Deflector Rod of Monitor

Deflector rod

Shown as above, the deflector rod is located at the left lower corner at the back of the monitor.
When the deflector is at the rightmost, the monitor is at working status and the angle between
monitor and support arm can be changed at the range controlled by the deflector rod. The
monitor can be turned for 20° both forward and backward, shown as follows:

2-4
System Structure

In the process of transportation or moving the system, the deflector rod can be toggled left to
level the monitor, shown as follows:

2-5
System Structure

2.2.3 Up/down Lock of Upper Support Arm

Put the support arm to horizontal position, the arm will be locked if the lock is at the locked
position; when the lock is located at the unlocked position, the arm can be adjusted from
upside to downside.

2.2.4 Tuning Screw of Air Spring on Support Arm


Turn the monitor and the support arm to the same position, adjust the tuning screw by
screwdriver.
Try to adjust the screw when the monitor rises and falls automatically, and the air spring
needs to be replaced if it can not meet the requirement even when the screw is at the tightest
or loosest status. (For the replacement of the air spring, please refer to 4.3.8 ).

2-6
System Structure

Tuning screw

2.3 Turning and Lifting the Keyboard

2.3.1 Turning
Hold and push the handle in the direction of front part of machine (located at the frontal part of
back keyboard ), turn the keyboard left and right within 45°, the keyboard will be locked at
where the handle be loosened.

Hold and push the handle in the direction of Handle


front part of machine to turn the keyboard.
2-7
System Structure

2.3.2 Lifting
Handhold the handle of the keyboard with both hands, with the left hand hold the lifting handle,
press the lifting handle to a certain position, uplift or push down the keyboard with both hands,
then the keyboard can be elevated or lowered. The height can be adjusted is 150mm at most.

Push in this way Lifting handle

2-8
System Structure

2.4 I/O Panel

<1> <5> <9> <13>

<2> <6> <10> <14>

<3> <7> <11> <15>

<8> <12> <16>


<4>

No. Port Function


<1>, <2> USB port Used to connect USB devices.
<3> Reset System reset.
<4> Ethernet port Used to connect to the network.

<5> Connects to the video printer control port.


Remote control port
<6> Connects to the serial port device.
Serial port
<7> Connects to the external display devices such as
VGA output port
monitor and projector.
<8> Connects the parallel port device.
Parallel port
<9> Separate video input port Separates the input video signal.

<10> Separate video output port Used for composite video output, connect to VCR
output or video printer.
<11> Video input port Used for composite video input.
<12> Video output port Used for composite video signal output, connect
to VCR output or video printer.

2-9
System Structure

<13> Audio input port Used for audio signal input.


<14> Audio input port Used for audio signal input.

<15> Audio output port Used for audio signal output.


<16> Audio output port Used for audio signal output.

2.5 Power Supply Panel

<1>

<2>

<5>

<3> <4>

No. Name Function


<1> Equipotential terminal Connects the equipotential points

<2> Power inlet AC power inlet


<3>, <4> Power output Supply power for optional peripheral devices (e.g. VCR)
<5> Circuit breaker Switches off the power supply

2-10
System Structure

2.6 ECG Panel

<1>

<2>

<3>

<4>

No. Name Function


<1> USB ports (2) Connects the USB devices.
<2> Reserved for future use.
Mic In port
(used for connecting microphone to record vocal
comments).
<3> ECG lead signal input port Connects to ECG leads, to directly obtain the
electrocardiosignal of the patient.
Reserved for future use.
<4> Pencil probe port
(used for connecting a pencil probe)

2-11
System Structure

2.7 Control Panel

No. Symbol Name Function


<1> Esc Escape Press to exit the current status to the previous status.
Help Press to open or close the accompanying help
<2> Help
documents.
/ Press to enter or exit patient information management
<3> iStation
system.
<4> F1 User-defined key The function is definable.
<5> F2 User-defined key The function is definable.
<6> F3 User-defined key The function is definable.
<7> F4 User-defined key The function is definable.
<8> F5 User-defined key The function is definable.
<9> F6 User-defined key The function is definable.
Quad-split Press to enter/ exit quad-split screen mode.
<10> Quad
screen

2-12
System Structure

<11> Biopsy Biopsy Press to show or hide the biopsy guide line.
<12> Setup Preset Press to open/ close the preset dialog box.
<13> / Character key Same as the keys of PC.
Indicating light 1 The indicator light is on after the power is switched
<14> /
on; it is off after the system is turned on.
Indicating light 2 The indicator light flashes when the hard disk is
<15> /
reading or writing; and in other situations, it is off.
<16> F8 User-defined key The function is definable.
<17> F9 User-defined key The function is definable.
<18> F7 User-defined key The function is definable.
<19> TGC Slide bar To adjust the depth and gain.
<20> Volume Volume Rotate to increase or decrease spectrum volume.
To adjust the parameters, the functions are shown
Multi-function
<21> / by the sketch map displayed at the downside of the
knob 1
screen.
To adjust the parameters, the functions are shown by
Multi-function
<22> / the sketch map displayed at the downside of the
knob 2
screen.
To adjust the parameters, the functions are shown by
Multi-function
<23> / the sketch map displayed at the downside of the
knob 3
screen.
Four-direction The functions are shown by the sketch map displayed
<24> /
key at the downside of the screen.
To adjust the parameters, the functions are shown by
Multi-function
<25> / the sketch map displayed at the downside of the
knob 4
screen.
To adjust the parameters, the functions are shown by
Multi-function
<26> / the sketch map displayed at the downside of the
knob 5
screen.
<27> iTouch / Press to optimize the images.
Patient To enter into patient information input interface.
<28> Info
information
Transducer To switch the transducer and exam mode.
<29> Probe
switch
<30> Review Review To review the stored images.
<31> Report Diagnostic report Press to open/ close the report.
<32> End Exam End exam To end an exam.
<33> F10 Self-defining key The function is definable.
/ To switch the live image window in the multi-window
<34> Update
mode.
<35> CW / Reserved key (press to enter the CW mode).
<36> M / Press to enter the M mode.
<37> PW / Press to enter the PW mode.
To adjust the gain of the live image, the light is on;
<38> Gain Gain press to adjust the gain of the corresponding mode in
2D mode or 3D mode.

2-13
System Structure

<39> Power / Press to enter the Power mode.


/ To enter Dual mode in Non-Dual mode;
<40> Dual
or to switch the live window in Dual mode.
<41> Color / Press to enter Color mode.
<42> B / Press to enter B mode.
<43> Body Mark Body mark To enter/ exit Body mark mode.
<44> Comment Comment To enter/ exit the character comment mode.
Clear To clear the comments and measurement caliper and
<45> Clear
so on.
<46> Arrow Arrow To enter/ exit arrow comment.
<47> Cine Cine review To enter/ exit Cine review.
Press to switch to the depth/zoom function, the
<48> Depth/Zoom Depth/ Zoom corresponding light will be on; turn the knob to adjust
the depth or zoom factor.
<49> Print Print Press to print the contents which have been set.
<50> Save Save Save the single frame image.
<51> Exit Exit To exit the current status.
<52> Measure Measure To enter/ exit the general measure mode.
To change the active point of the caliper during
<53> Change Change
measurement.
<54> Caliper Caliper To enter/ exit the general measure mode.
Multi-function Status dependent.
<55> /
knob
Main menu To invoke or close the menu corresponding to the
<56> Menu
current status.
<57> Freeze Freeze Press to freeze or unfreeze an image.
Press to return to the previous operation or to delete
<58> Back Back
the last project.
Trackball Rotate to change the position of the cursor on the
<59> /
screen.
To confirm an operation, equivalent to the left-button
<60> Set Set
of the mouse.

2-14
System Principle

3 System Principle

3.1 System Principle

The block diagram of DC-3 system is shown as follows:

IO rear IO front ECG Probe


Control panel Probe
board board board control
board
board

IO Probe
connecting CPU connecting
board board board
IO port
board
Power 4D driver
board

Master board
USB TO IDE
DVD R/W
board

Ultrasound Transmitting
PC system
system board
Display and PCI
speaker

Black and white CW board


video printer

Power
AC/DC
supply
board
board

Power supply
AC input
connecting Breaker
connecter
board

Isolating
transformer

Fig 3-1 Block Diagram of DC-3

The system consists of main unit and power system, and the main unit consists of:
Master board: including ultrasound system, PC system, power supply management and
filtering system;
Ultrasound system: to carry out functions including echo receiving, zoom in, sampling, Beam
forming, signal processing, ultrasound scanning receiving control as well as data exchange

3-1
System Principle

between PCs;
PC system: post processing the echo data, communicate with users, to support many kinds of
peripherals;
Power management and filtering system: to ensure the normal working of power system,
support system shutdown, stand-by mode and dormancy mode;
Meanwhile, master board provides connecting sockets with each module;
Transmitting board: to realize delay focusing transmitting for N channels;
Control panel can communicate with the master board through USB;
CPU board;
Probe board module: including probe control board, probe board, probe connecting board, used
to connect the transducers, supports 4 transducer sockets, of which 3 sockets are live;
CW Board: processing CW signal and pencil transducer signal;
4D driver board (reserved);
The peripheral extension modules including:
I/O connecting board: connects the master board and IO front board, IO interface board,
provides power supply to the hard disk;
IO front board,USB HUB(one input, three outputs), provides external MIC port, USB port as well
as pencil probe port;
IO port board: USB HUB (one input, six outputs), video acquisition, and provides sockets to
some inner wires;
IO rear board: provides IO port for the rear part;
USB-to-IDE board: USB to IDE converting board, connects to DVD R/W;
ECG board;
Display module: including display and speaker;
The power system consists of (In fact, it also includes the power management system on the
master board and the filtering system):
Power supply board: provides power for the system;
Power supply connecting board: provides connection sockets between auxiliary output, breaker
and isolating transformer, to realize voltage input selection;
AC/DC board: to change the AC input into 12V DC.

3-2
System Principle

3.2 Working Principles of Each Module in the Hardware

System

3.2.1 Master Board


The block diagram of the master board is shown as follows:

4D port 1 Probe module

Transmitting
board Master
board
FPGA1

Beamform
er Ultrasound
receiving
CW port 1

CW port 2 DC3 power


FPGA2

Signal port
Process 4D port Master board
power Power
port
Fan Tempe
port rature
sensor
Adapter
FPGA3 port
Video System
process monitor
Multi- DC3 Power
Extension function signal management
port port

COME INTERFACE

USB HDD Audio Network

Fig 3-2 Block Diagram of the Master Board

Shown as above, the function modules of the master board are:

3-3
System Principle

Ultrasound receiving;
Signal process;
Power management;
System monitoring;
Video process.
And the master board provides connecting sockets for the following devices:
Probe board module;
Transmitting board;
CW board;
4D board (reserved);
CPU module;
IO connection board;
Power supply module;
Hard disk;
DC-3 power socket;
DC-3 signal socket;
Fan;
Speaker;
Network.
The layout of the above modules’ sockets on the master board is shown as Fig A-1.

3.2.2 Ultrasound Receiving


Ultrasound receiving consists of high voltage isolating & receiving channel selection,
voltage-control gain amplifier and ADC. The receiving channel selection is controlled by two
CPLDs, the analog control voltage of the amplifier comes from a DAC, and SPI signals of DAC,
the amplifier as well as the ADC all come from beamformer FPGA1.

3.2.3 Signal Processing


Signal processing is completed by two FPGAs, they are FPGA1 and FPGA2.
FPGA1 mainly controls the front-end chip (channel selection CPLD, VGA gain control DAC,
ADC and the transducers), produces transmitting sequence pulse, carries out wave combination,
quadrature demodulation and so on.
FPGA2 is used to process signal, upload data, call real-time parameter scanning and so on,
data is uploaded through PCI socket, and the USB port is a reserved port.

3-4
System Principle

3.2.4 Power Management


Power management is basically completed by CPLD.
According to the power on sequence, the system power can be classified into: STANDBY power
supply, CPU STANDBY power supply and normal working power supply.
STANDBY power supply is the power supply under POWERDOWN status. It includes +5VSTB,
+3V3STB and +1.8VSTB. They power the comparator that generated POWER OK, crystal of
6M, power supply management CPLD, power supply drive and 244 status drive. Just to connect
to AC power and switch on the breaker.
CPU STANDBY is the power supply when PC is in standby mode, it stops working when the
system is turned off, and it works first when the system begins to work.
The normal working power supply refers to the power required during normal work of the system.
Except STANDBY and CPU STANDBY mode, others all belong to normal working power supply,
of which the indicator lights are shown as follows:

Table 3-1 Indicator Lights of the Master Board Power Supply

No. LED number Meaning


1 D7 12V
2 D25 3.3V indicator light
3 D26 5V indicator light
One indicator light on the DC-3 control panel indicates the working status of the breaker. The
related lights descriptions are described as Table 3-2:

Table 3-2 Description of Power Supply Indicator Lights

Single color(green), indicates the working status of the


breaker after mains power connected; the light is off
Indicator light of breaker status
when the breaker is switched off;
1 (refer to the 14th item marked
The light is green when the breaker is switched on, it
on the control panel figure)
flashes during start-up process, and it is off when the
back light of the keyboard is on.

3.2.5 System Monitoring


System monitoring including monitor on the voltage, temperature, fan malfunction detection and
control on the fan speed;
Monitor on the voltage: direct measurement, including measurement of 12V, VCC(+5V),
VDD(+3.3V), 2.5V, 1.5V, PHV, A+3, -5V, -12V;
Monitor on the power supply module and temperature of ultrasound front-end;
Monitor and control on fans and ventilation unit.

3-5
System Principle

3.2.6 Video Processing


Video signal that outputs from the PC module is converted into standard S_Video and video
signal post processed by FPGA3 and video code chip.

3.2.7 Sockets on Master Board


The standard sockets, which have no definition in the master board sockets, are required to be
designed by the related standards, the sockets include: dual-USB port, SATA socket, Ethernet
socket, SATA hard disk socket of notebook, COM EXPRESS socket, button-shaped battery
socket and S-Video socket.

3.2.7.1 Socket between Master Board and Probe Module


Signals between the probe board module and master board socket are defined as Table A-1.
Descriptions of signals are described as follows:

Table 3-3 Definition of Signals between the Master Board and Probe Module Socket

Signal name Description


SPI_CLK
SPI_DIN
Commands and data port to the probe module.
SPI_DOUT
SPI_CS
FLASH_POWER Reserved
FLASH_WP Reserved
PROBE_ID[7:6]
PROBEID_CS[1:0] Reserved signal
RELAY_EN[1:0]
PROBE_PRESENT Presents probe signal
EXP_PRESENT Connected to earth (in the probe unit)
PE[1:N] N channel transmitting and receiving signal

3.2.7.2 Socket between Master Board and Transmitting Board


Signals between the master board and transmitting board are defined in Table A-1 and Table
A-2.

3.2.7.3 Socket between CW Board and Master Board


Signals between the CW board and master board sockets are defined as Table A-3 and Table
A-4. Of which, the signals of CW CON1 are defined as follows:

Table 3-4 Signals Definition at CW CON1 Socket

Name Description
CW[9:0] CW echo signal

3-6
System Principle

12V +12 power supply


-12V -12V power supply
5V +5V power supply
-5V -5V power supply

3.2.7.4 JTAG Socket Definition of Clock Chip


The JTAG socket is defined as follows:

Table 3-5 Definition of JTAG Socket

PIN SIG
1 CLK_TMS
2 CLK_TDI
3 CLK_TDO
4 CLK_TCK
5 GND
6 VDD_CLK

3.2.7.5 Socket between 4D Board and Master Board


Socket design between 4D board and master board is reserved currently, and the signals are
defined as Table A-5.

3.2.7.6 Socket between Speaker and Master Board


Socket between speaker and master board is defined as follows:

Table 3-6 Definition of Sockets between Speaker and master Board

PIN SIG
1 ROUTP
2 ROUTN
3 NC
4 LOUTP
5 LOUTN

3.2.7.7 Socket between Fan and Master Board


Socket between fan and master board is defined as follows;

Table 3-7 Definition of Socket between Fan and Master Board

PIN SIG PIN SIG


1 FANSPEED4 2 12V
3 PWM45 4 12V
5 FANSPEED5 6 CPU_FAN_PN
7 FANSPEED2 8 CPU_FAN_SPEED
9 PWM23 10 12V
11 FANSPEED3 12 12V

3-7
System Principle

3.2.7.8 Socket of Master Board CPLD JTAG Socket


Master board CPLD JTAG socket is defined as follows:

Table 3-8 Definition of Master Board CPLD JTAG Socket

PIN SIG
1 TCK
2 TDO
3 TMS
4 TDI
5 3V3
6 GND
7 3V3
8 PLUGED

3.2.7.9 Signal Socket


Signal socket is defined as follows:

Table 3-9 Definition of Signal Socket

No. Name No. Name


1 I^C_DA 2 GND
3 I^C_CK 4 GND
5 GND 6 Gnd
7 Power_ON 8 HDD_Status
9 SYS_RESET 10 Gnd
11 C_Print 12 Breaker_Status
13 C_Busy 14 Gnd
15 Gnd 16 Y_Svideo
17 Power_SSW 18 C_Svideo
19 Gnd 20 Gnd

3.2.7.10 Power Supply Socket


Power supply socket is defined as follows:

Table 3-10 Definition of Power Supply Socket

No. Name No. Name


1 +12V 2 +12V
3 +12V 4 +12V
5 Gnd 6 Gnd
7 Gnd 8 Gnd
9 Gnd 10 Gnd
11 +5V 12 +5V
13 +5V 14 +5V
15 +5V 16 +5V
17 +5V 18 +5V

3-8
System Principle

No. Name No. Name


19 Gnd 20 Gnd
21 Gnd 22 Gnd
23 +3.3V 24 +3.3V

3.2.7.11 Socket between Adapter and Master Board


Socket between adapter and master board is defined as follows:

Table 3-11 Definition of Socket between Adapter and Master Board

No. Name Direction


1 Gnd \
6 Gnd \
2 Gnd \
4 Gnd \
3 +12V In
7 +12V In
5 +12V In
8 +12V In

3.2.7.12 Socket between Power Supply Module and Master Board


Definition of socket between power supply module and master board is shown as Table A-7, and
the signals description is shown in Table A-8. The signal direction is defined relative to the
power supply module.

3.2.7.13 Signal Definition between Master Board and Extended Socket


Signal definition between master board and extended socket is shown as Table A-9.

3.2.8 Testing Points on the Master Board


The testing points on the master board are shown as follows:

3-9
System Principle

Fig 1 Testing Points

3-10
System Principle

Fig 2 Testing Points

3-11
System Principle

Table 3-12 Testing Points on the Master Board

No. Symbol Location


1 1V2_BF FPGA1 kernel voltage (1.2V)
2 1V2_DSP FPGA2, FPGA3 kernel voltage (1.2V)
3 2V5_BF FPGA1 IO voltage (2.5V)
4 3V3_BF FPGA1 IO voltage (3.3V)
5 12V 12V voltage
6 A+3 Changeable gain amplifier voltage (3V)
7 A+5V_REV Voltage of receiving module (4.65V)
8 A-5V_REV Voltage of receiving module (-4.65V)
9 A+3V3_REV Voltage of receiving module (3.3V)
10 A1V8 Voltage of receiving module (1.8V)
11 A3V3_SW Voltage of receiving module (3.3V)
12 MMF_VREF FPGA3 DDR voltage (1.25V)
13 MMF_VTT FPGA3 DDR voltage (1.25V)
14 PHV Remote control high voltage 5-10V(CW), 20-140V(B\C\D)
15 TP3 STB voltage (3.3V)
16 TP6 Working voltage (5V)
17 TP17 Left audio input
18 TP18 Right audio input
19 TP20 Power supply management CPLD clock (6M)
20 TP30 Power for CPLD power supply management (1.8V)
21 VDR0 FPGA2 DDR0 voltage (1.25V)
22 VTT0 FPGA2 DDR0 voltage (1.25V)
23 VDR1 FPGA2 DDR1 voltage (1.25V)
24 VTT1 FPGA2 DDR1 voltage (1.25V)
25 A_N3V3_REV Voltage of receiving module (-3.3V)

3.3 Transmitting Board


The function of transmitting board is to convert the low voltage pulse from the master board into
high voltage pulse that required by transmitting. The block diagram is shown as follows:

Fig 3-3 Ultrasound Transmitting Diagram

Sockets definition of transmitting board and master board is shown in Table A-1 and Table A-2.
The testing points related to the power supply of transmitting board is defined as follows:

3-12
System Principle

Table 3-13.Testing Points Related to Power Supply of Transmitting Board

No. Symbol Location


1 PHV Remote control high voltage 5-10V(CW),20-140V(B\C\D)
2 A+12 11.4V voltage
3 VOE 3.3V voltage

The N testing points are corresponding to the N transmitting channels. Of the 4 rows of testing
points, the upper two rows are even number points, while the lower two rows are odd number
points. The numbers are marked between the spaces, e.g. P65 refers to the 65th channel.
The tested wave of transmitting signal under B Mode is shown as follows:

Fig 3-4 Tested Wave of Transmitting Signal in B Mode

Note: the amplitude, pulse width and pulse number vary with the transmitting parameters.

3.4 Probe Board Module

3.4.1 Principle of Probe Board Module


Functions of the probe are carried out by the probe connecting board, probe control board as
well as probe board. Probe control board are buckled with probe board, probe board and master
board are combined by probe connection board. Signals between the three panels and the
control structure are shown as follows:

3-13
System Principle

Probe connecting board Probe board Probe control board


Power
supply 5V 3.3V
LDO

Control signal

… ID signal of CPLD
Probe
probe
mode
A probe port switching

Control
signal
signal
Probe
……

Probe

Control
Probe power

signal
Probe port … power …
supply
control


Transmitting/ receiving
signal Signal switching
(Relay) Relay control

Fig 3-5 Electrical Block Diagram of Probe Board Module

The probe connection board is applied to connect the master board and the probe board: the
probe control board, which is buckled with probe board, takes control on the switching relays on
the probe board, switches the working probes and A probe mode (probes including general
probes, 4D probe and TEE probe), and reads the probe ID and so on.

Fig 3-6 Principle Diagram of Relay Control Circuit

The above diagram shows the control circuit principle of a group of relays. Control signal
produced by CPLD turns into RELAY_EN1 after driven by 245, and the RELAY_EN1 signal will
then take control on the conduction and cut-off of N-MOSFET Q1. Q1 controls the contactor
position of relay K1 (whether toggle to normal open or to normal close). When the contactor is
normal close, signal KA2 and KA3 are of VCC level, both signals can act as the power supply of
the related relay, let the relay to be working status.

3-14
System Principle

3.4.2 Definition of Sockets

Table 3-14 Sockets between Probe Connecting Board and Probe Board

Socket P1
No. Name No. Name No. Name No. Name No. Name
A1 Gnd B1 PE127 C1 Gnd D1 PE128 E1 Gnd
A2 PE124 B2 Gnd C2 PE125 D2 Gnd E2 PE126
A3 PE121 B3 Gnd C3 PE122 D3 Gnd E3 PE123
A4 Gnd B4 PE119 C4 Gnd D4 PE120 E4 Gnd
A5 PE116 B5 Gnd C5 PE117 D5 Gnd E5 PE118
A6 PE113 B6 Gnd C6 PE114 D6 Gnd E6 PE115
A7 Gnd B7 PE95 C7 Gnd D7 PE96 E7 Gnd
A8 PE92 B8 Gnd C8 PE93 D8 Gnd E8 PE94
A9 PE89 B9 Gnd C9 PE90 D9 Gnd E9 PE91
A10 Gnd B10 PE87 C10 Gnd D10 PE88 E10 Gnd
A11 PE84 B11 Gnd C11 PE85 D11 Gnd E11 PE86
A12 PE81 B12 Gnd C12 PE82 D12 Gnd E12 PE83
A13 Gnd B13 PE79 C13 Gnd D13 PE80 E13 Gnd
A14 PE76 B14 Gnd C14 PE77 D14 Gnd E14 PE78
A15 PE73 B15 Gnd C15 PE74 D15 Gnd E15 PE75
A16 Gnd B16 PE71 C16 Gnd D16 PE72 E16 Gnd
A17 PE68 B17 Gnd C17 PE69 D17 Gnd E17 PE70
A18 PE65 B18 Gnd C18 PE66 D18 Gnd E18 PE67
A19 Gnd B19 PE63 C19 Gnd D19 PE64 E19 Gnd
A20 PE60 B20 Gnd C20 PE61 D20 Gnd E20 PE62
A21 PE57 B21 Gnd C21 PE58 D21 Gnd E21 PE59
A22 Gnd B22 PE55 C22 Gnd D22 PE56 E22 Gnd
Socket P2
No. Name No. Name No. Name No. Name No. Name
A1 PE52 B1 Gnd C1 PE53 D1 Gnd E1 PE54
A2 PE49 B2 Gnd C2 PE50 D2 Gnd E2 PE51
A3 Gnd B3 PE47 C3 Gnd D3 PE48 E3 Gnd
A4 PE44 B4 Gnd C4 PE45 D4 Gnd E4 PE46
A5 PE41 B5 Gnd C5 PE42 D5 Gnd E5 PE43
A6 Gnd B6 PE111 C6 Gnd D6 PE112 E6 Gnd
A7 PE108 B7 Gnd C7 PE109 D7 Gnd E7 PE110
A8 PE105 B8 Gnd C8 PE106 D8 Gnd E8 PE107
A9 Gnd B9 PE103 C9 Gnd D9 PE104 E9 Gnd
A10 PE100 B10 Gnd C10 PE101 D10 Gnd E10 PE102
A11 PE97 B11 Gnd C11 PE98 D11 Gnd E11 PE99
A12 Gnd B12 PE39 C12 Gnd D12 PE40 E12 Gnd

3-15
System Principle

A13 PE36 B13 Gnd C13 PE37 D13 Gnd E13 PE38
A14 PE33 B14 Gnd C14 PE34 D14 Gnd E14 PE35
A15 Gnd B15 PE31 C15 Gnd D15 PE32 E15 Gnd
A16 PE28 B16 Gnd C16 PE29 D16 Gnd E16 PE30
A17 PE25 B17 Gnd C17 PE26 D17 Gnd E17 PE27
A18 Gnd B18 PE23 C18 Gnd D18 PE24 E18 Gnd
A19 PE20 B19 Gnd C19 PE21 D19 Gnd E19 PE22
A20 PE17 B20 Gnd C20 PE18 D20 Gnd E20 PE19
A21 Gnd B21 PE15 C21 Gnd D21 PE16 E21 Gnd
A22 PE12 B22 Gnd C22 PE13 D22 Gnd E28 PE14
Socket P3
No. Name No. Name No. Name No. Name No. Name
A1 PE9 B1 Gnd C1 PE10 D1 Gnd E1 PE11
A2 Gnd B2 PE7 C2 Gnd D2 PE8 E2 Gnd
A3 PE4 B3 Gnd C3 PE5 D3 Gnd E3 PE6
A4 PE1 B4 Gnd C4 PE2 D4 Gnd E4 PE3
A5 Gnd B5 Gnd C5 Gnd D5 Gnd E5 Gnd
A6 PROBEID B6 Gnd C6 RELAY_E D6 Gnd E6 EXP_PRE
_CS1 N1 SENT
A7 PROBEID B7 Gnd C7 RELAY_E D7 Gnd E7 PROBE_P
_CS0 N0 RESENT
A8 Gnd B8 Gnd C8 Gnd D8 Gnd E8 Gnd
A9 SPI_DIN/P B9 SPI_DOU C9 FLASH_P D9 FLASH_W E9 PROBE_I
ROBE_ID T/PROBE_ OWER/PR P/PROBE_ D7
1 ID2 OBE_ID4 ID5
A10 SPI_CLK/ B10 Gnd C10 SPI_CS/P D10 Gnd E10 PROBE_I
PROBE_I ROBE_ID D6
D0 3
A11 Gnd B11 Gnd C11 Gnd D11 Gnd E11 Gnd
A12 Gnd B12 Gnd C12 Gnd D12 Gnd E12 Gnd
A13 +5V B13 +5V C13 +5V D13 +5V E13 +5V
A14 +5V B14 +5V C14 +5V D14 +5V E14 +5V
A15 +5V B15 +5V C15 +5V D15 +5V E15 +5V
A16 Gnd B16 Gnd C16 Gnd D16 Gnd E16 Gnd
A17 Gnd B17 Gnd C17 Gnd D17 Gnd E17 Gnd
A18 4D-Gnd B18 4D-Gnd C18 4D-Gnd D18 4D-Gnd E18 4D-Gnd
A19 4D-RES1 B19 4D-RES2 C19 4D-RES3 D19 4D-RES4 E19 4D-RES5
Thermison Thermison TEE-Agnd Hall_signal 4D-ID
+ -
A20 4D-Gnd B20 4D-Gnd C20 4D-RES6 D20 4D-Gnd E20 4D-Gnd
TEE-Angle
A21 4D-MOTO B21 4D-MOTO C21 4D-Gnd D21 4D-MOTO E21 4D-MOTO

3-16
System Principle

ROUT1A ROUT1B ROUT2B ROUT2A


Sin+ Sin- Cos- Cos+
A22 4D-MOTO B22 4D-MOTO C22 4D-Gnd D22 4D-MOTO E22 4D-MOTO
ROUT1A ROUT1B ROUT2B ROUT3A
Sin+ Sin- Cos- Cos+

Table 3-15 Definition of 4D Reserved Socket on Probe Connecting Board

PIN No. Name PIN No. Name


1 4D_MOTOROUT1A 2 4D_MOTOROUT1A
3 4D_MOTOROUT1B 4 4D_MOTOROUT1B
5 4D_MOTOROUT2A 6 4D_MOTOROUT2A
7 4D_MOTOROUT2B 8 4D_MOTOROUT2B
9 4D_GND 10 4D_GND
11 4D_GN D 12 4D_GND
13 4D_RES1/Thermison+ 14 4D_RES2/Thermison-
15 4D_RES3/TEE_AGND 16 4D_RES4/4D_Hall_angle
17 4D_RES5/4D_ID 18 4D_RES6/TEE_angle_out
19 4D_GND 20 4D_GND

Table 3-16 Definition of Sockets on Probe Board Module

Probe control board socket J1


PIN No. Name PIN No. Name
1 KA3 2 KA2
3 KA7 4 KA1
5 SPI4/FLASH_POWER 6 GND
7 PROBEID_CS1 8 SPI5/FLASH_WP
9 RELAY_EN0 10 SPI7/PROBE_ID7
11 GND 12 PROBE_PRESENT
13 PROBEID_CS0 14 RELAY_EN1
15 SPI2/SPI_DOUT 16 EXP_PRESENT
17 SPI1/SPI_DIN 18 GND
19 SPI6/PROBE_ID6 20 A+5V
21 SPI3/SPI_CS 22 A+5V
23 SPI0/SPI_CLK 24 A+5V
25 GND 26 A+5V
27 4D_GND 28 GND
29 4D_RES4/4D_Hall_angle 30 4D_RES5/4D_ID
31 4D_RES1/Thermison+ 32 4D_RES3/TEE_AGND
33 4D_RES2/Thermison- 34 4D_RES6/TEE_angle_out
35 4D_GND 36 4D_GND
37 4D_MOTOROUT1A 38 4D_MOTOROUT2B
39 4D_MOTOROUT1B 40 4D_MOTOROUT2A

3-17
System Principle

Probe control board socket J2


PIN No. Name PIN No. Name
1 KD2 2 DID4
3 KD3 4 DID7
5 KD7 6 KD4
7 KD1 8 KD8
9 GND 10 GND
11 DID1 12 KD6
13 DID2 14 KD5
15 D_ON 16 GND
17 DID8 18 KA8
19 DID6 20 KA6
21 DID3 22 KA5
23 DID5 24 KA4
25 GND 26 GND
27 4D_RES3/TEE_AGND 28 PA_6S
29 4D_RES6/TEE_angle_out 30 PA_5S
31 A_ON 32 4D_GND
33 4D_ON 34 PA_6P
35 4D_GND 36 PA6R
37 PA_5M 38 PA5R
39 PA_5N 40 PA5P
Probe control board socket J3
PIN No. Name PIN No. Name
1 C_ON 2 CID3
3 CID1 4 CID5
5 CID2 6 CID4
7 CID7 8 CID6
9 GND 10 GND
11 CID8 12 KC4
13 KC3 14 KB4
15 KC1 16 KC5
17 KB7 18 KC6
19 GND 20 GND
21 KC7 22 KC8
23 KC2 24 KB5
25 KB2 26 KB6
27 KB3 28 KB8
29 GND 30 GND
31 DID7 32 BID8
33 KB1 34 BID6
35 B_ON 36 BID5
37 BID1 38 BID4

3-18
System Principle

PIN No. Name PIN No. Name


39 BID2 40 BID3

3.5 I/O Connecting Board


Function of I/O connection board includes sending signals from the master board to port board
and front board, providing power supply to SATA hard disk and so on.

3.5.1 Definition of Sockets on I/O Connection Board


² 1 connecting socket to the master board 50Pin I/O extended ports, signals definition is
shown as Table A-10.
² 1 connecting socket to the master board 20Pin reserved socket, signals definition is shown
as Table 3-9.
² 1 connecting socket to the master board 24Pin power reserved socket, signals definition is
shown as Table 3-10.
² 1 connecting socket to the CW board pencil probe port, the signals definition is shown as
follows:

Table 3-17 Signal Definition between I/O Connecting Board and CW Board Pencil Probe Port

PIN NUM SIGNAL


1 PENCIL_R
2 AGND
3 AGND
4 PENCIL_T
² 1 connecting socket to the DC-3 master board speaker socket, socket definition is shown
as Table 3-6.
² 1 connecting socket to the DC-3 I/O front board socket, the socket definition is shown as
Table 3-21.
² 1 connecting socket to the DC-3 I/O port board socket, the socket definition is shown as
follows:

Table 3-18 Socket Definition between the I/O Connecting Board and I/O Port Board

No. Name No. Name No. Name No. Name No. Name
A1 3V3 B1 3V3 C1 3V3 D1 3V3 E1 3V3
A2 GND B2 GND C2 GND D2 GND E2 GND
A3 5V B3 5V C3 5V D3 5V E3 5V
A4 5V B4 5V C4 5V D4 5V E4 5V
A5 GND B5 GND C5 GND D5 GND E5 GND
A6 12V B6 12V C6 12V D6 12V E6 12V
A7 GND B7 GND C7 GND D7 GND E7 GND

3-19
System Principle

A8 C_SVIDEO B8 Y_SVIDEO C8 VIDEO D8 VGA_GREEN E8 GND


A9 GND B9 GND C9 GND D9 GND E9 VGA_RED
A10 RIGHT- B10 LEFT- C10 AUDIO_L D10 GND E10 GND
A11 RIGHT+ B11 LEFT+ C11 AUDIO_R D11 GND E11 VGA_BLUE
A12 GND B12 GND C12 GND D12 GND E12 GND
A13 USB-5 B13 UART_RX0 C13 POWER_SSW D13 VGA_VS E13 VGA_HS
A14 USB+5 B14 UART_TX0 C14 BREAKER_STATUS D14 C_BUSY E14 C_PRINT
A15 GND B15 GND C15 GND D15 GND E15 GND
A16 D2 B16 HD_STATUS C16 SYS_RESET D16 POWER_ON E16 D0
A17 RM_PRINT B17 GND C17 I^2C_CK D17 I^2C_DA E17 D1
A18 GND B18 GND C18 GND D18 GND E18 GND
A19 ACK_N B19 BUSY C19 RM_BUSY D19 D3 E19 D4
A20 D5 B20 D6 C20 D7 D20 PE E20 AUTOFD_N
A21 SEL B21 ERROR_N C21 INIT_N D21 STROBE_N E21 SELIN_N
A22 GND B22 GND C22 GND D22 GND E22 GND

Power socket definition of SATA hard disk is shown as follows:

Table 3-19 Power Supply Socket Definition of SATA Hard Disk

No. Name
1 12V
2 GND
3 +5V
4 GND
5 +3.3V
6 GND

3.6 IO Front Board


The main function of I/O front board is to extend 1 channel USB signal into 3 channels, and
transmit some signals from the master board at the same time.

3.6.1 USB
The one-channel USB signal from the master board turns into 3 channels after processed by
HUB, of which, 1 channel is connected to ECG module and the other two channels are
connected to external standard USB ports. Meanwhile, the external USB port is turned into
VBUS that connects to U port by a 2042.

3-20
System Principle

USB1 over-current
protection VBUS
USB2 over-current VBUS
protection 2042

NC
NC
USB1

USB2
Up -channel ECG_USB
USB GL850G NC

The chip is externally powered, the chip and peripheral circuits are powered by 3.3V, which is
adapted from 5V VCC by the adapter that built-in the chip; the drive level of HUB U port is
powered by on-board power supply. After powered on, the requested bus ID will be recorded in
the built-in memory by GL850G, so no external EEPROM or FLASH is required for memory. In
the PGANG position, interface to a pull-up resistor to enable the chip’s GANG mode, and we
reserve a 0ohm resistor position to make the chip work in independent mode. The
corresponding indicator light is used to indicate if the USB channel is OK. The relationship
between the indicator light and USB channel is as follows:

Table 3-20 Function of LED Indicator Lights on I/O Front Board

Item Function
D1 USB3 abnormal work indicator light
D2 USB3 normal work indicator light
D3 GANG mode indicator light
D4 USB1 normal work indicator light
D5 USB1 abnormal work indicator light
D6 USB2 normal work indicator light
D7 USB2 abnormal work indicator light

3.6.2 Sockets Definition of IO Front Board


² 1 socket connected to I/O connecting board, defined as Table 3-21:

Table 3-21 Socket Definition between I/O Front Board and I/O Connecting Board

No. Name No. Name


1 GND 2 AGND
3 MIC_L 4 PENCIL_T
5 MIC_R 6 AGND
7 GND 8 PENCIL_R
9 USB-4 10 GND
11 USB+4 12 3.3V
13 GND 14 GND

3-21
System Principle

No. Name No. Name


15 +5V 16 +5V
17 +5V 18 +5V
19 GND 20 GND
² 1 pencil probe port (reserved), defined as follows:

Table 3-22 Socket Definition between I/O Front Board and Pencil Probe

PIN NUM SIGNAL


1 PENCIL_R
2 AGND
3 AGND
4 PENCIL_T
² 2 USB ports that provide external output, defined as follows:

Table 3-23 Standard USB Ports for External Output

No. Name
1 +5
2 USB-
3 USB+
4 GND
5 GND
6 GND

Table 3-24 USB Port for ECG Module

No. Name
1 +5V
2 +5V
3 USB-
4 USB+
5 Gnd
6 Gnd
² 1 MIC port (function reserved), defined as follows:

Table 3-25 MIC Port Definition

No. Name
1 GND
2 MIC_R
3 MIC_L

3.7 IO Port Board


I/O port board mainly provides connections for master board, I/O rear board, the console, the
printer, LCD, DVD-RW and so on; other functions including extending 1 channel VIDEO signal

3-22
System Principle

into 2 channels S-VIDEO signal, 1 channel VGA signal into 2 channels, 1 channel USB signal
into 6 channels, and video/audio signal collection and so on.
Signals through I/O port board can be classified into two kinds: some signals are processed on
the board, including USB, VGA output, VIDEO input/output, S-VIDEO input, and left/right audio
input; other signals are not processed on the board, just transmitted, including parallel port,
serial port, left/right audio output, speaker output, HD of keyboard, BREAKER signal, soft switch
control signal and 2 channels remote signals.

3.7.1 Functional Modules

3.7.1.1 USB
The one-channel USB signal from the master board is connected to the board through a 110PIN
socket, two one-to-four USB HUB CONTROLLER ICs are used to work in cascade mode to
realize USB 1-to-6, the diagram is shown as follows:

TP2042
Over-current
NC protection
NC
Over-current NC TP2042
NC NC
protection

Up-channel Down-channel Down-channel


USB port 1 port 1 Printer
Down-channel Rear board standard U
Down-channel
port 2 port 2 port 1
GL850G GL850G
Down-channel Down-channel Rear board standard U
port 3 port 3 port 2

Down-channel Down-channel
port 4 port 4 NC

Video/ audio collection


Control panel
DVD(usb2ide)
The chip is externally powered, the chip and peripheral circuits are powered by 3.3V level,
which is adapted from 5V VCC by the adapter that built-in the chip; the drive level of HUB U port
is powered by on-board power supply. After powered on, the requested bus ID will be recorded
in the built-in memory by GL850G, so no external EEPROM or FLASH is required for memory.
In the PGANG position, interface to a pull-up resistor to enable the chip’s GANG mode, and we
reserve a 0Ohm resistor position to make the chip work in independent mode. There is no great
difference between the two modes. In the assignment of pins of the HUB, considering the

3-23
System Principle

bandwidth, we assign the port which has more data onto the first level HUB. The corresponding
indicator lights are used to indicate if the USB channel is OK. The relationship between the
indicator lights and USB channels is as follows:
No. Name
D1 EM2860 working indicator light.
D3 HUB1working indicator light.
D4/D5 HUB1 1 channel (HUB2 up-channel) indicator light.
D6/D7 Working status indicator light of video/ audio collection module.
D8/D9 Working status indicator light of control panel U port.
D10/D11 Working status indicator light of DVD U port.
D12 Working status indicator light of HUB2.
D13/D14 Working status indicator light of standard USB port 2 on rear board.
D15/D16 Working status indicator light of printer U port.
D36/D37 Working status indicator light of standard USB port 2 on rear board.
Of which, the yellow light means abnormal work, and the green light means normal work.
Indicator light EM2860 can only be lightened when the collection software is running.

3.7.1.2 Video/ Audio Collection


Solution of EMPIA is applied for video/ audio collection.
For audio collection, one EMP202, which change the audio data into serial data (the data will be
sent to EM2860 through SDI signal and then to the master board by USB bus via the
connecting board), is involved for signal encoding and decoding.
For video data collection, TVP5150AM1 of TI is applied for video decoding. The S-VIDEO or
VIDEO signal, which is sampled by TVP5150AM1 after being sent to the port board via the rear
board, will be transferred to 8 bits digital signal and be output in parallel to EM2860, and then to
be sent to the master board through USB bus. An analog switch is required in the process of
S-VIDEO and VIDEO signal to be input into the port board, this is because TVP5150AM1 can
encode only one signal at a time. ADG774 is applied here, and one channel is selected for
collection. The analog switch is controlled under GPCL signal of TVP5150AM1.
After powered on, the control signal will be sent to TVP5150AM1 by USB port from EM2860 via
I2C bus. Then BIT5 and BIT6 of Miscellaneous Control Register, (When BIT5 is high, enable
GPCL signal output, the value of GPCL is reflected on BIT6), as well as BIT0, BIT1 of Input
Source Selection Register will be set. (When BIT0 is high, AIPIA/AIPIB signal receives SVIDEO
signal, and when BIT0 is low, COMPOSITE VIDEO signal will be received; BIT1 controls the
input of AIPIA/AIPIB, when it is low, input AIPIA, when it is high, input AIPIB). At the same time,
a high level pass through a PMOS to control the reset time of EM2860 costs less time than
EMP202.
The signals from EMP202 and TVP5150AM1 will be gathered by EM2860, and then be sent to
the master board by USB bus, at the same time, it is a way for the master board to control video

3-24
System Principle

collection via USB bus.


Block diagram of video collection is shown as follows:

Video
DV signal
COMP_In
Interface
Video Analog Y_In
Decoder switch
HUB C_In
Switch
USB Video I2C control Rear
Capture Device board
Line_In_R
AC97 Link Audio Codec Line_In_L

3.7.1.3 Signal Extension


Signal extension is driven by ISL59830, which will extend one signal into two signals, the block
diagram is shown as follows:

VGA RGB
ISL59830
VGA HS/VS

SVIDEO
ISL59830

3.7.2 Sockets Definition of I/O Port Board


² 1 socket connected to I/O connecting board, which is defined as Table 3-18.
² 1 socket connected to I/O rear board, which is defined as follows:

Table 3-26 Socket between I/O Port Board and I/O Rear board

No. Name No. Name


A1 GND B1 GND
A2 D0 B2 D1
A3 GND B3 D2
A4 D3 B4 D4
A5 D5 B5 D6
A6 GND B6 D7
A7 AUTOFD_N B7 PE
A8 BUSY B8 ACK_N

3-25
System Principle

No. Name No. Name


A9 GND B9 SELIN_N
A10 STROBE_N B10 INIT_N
A11 SEL B11 ERROR_N
A12 GND B12 GND
A13 UART_RX0 B13 UART_TX0
A14 USB_1- B14 USB_1+
A15 USB_1BUS B15 USB_1BUS
A16 GND B16 GND
A17 USB_2- B17 USB_2+
A18 USB_2BUS B18 USB_2BUS
A19 GND B19 GND
A20 C_PRINT B20 C_BUSY
A21 SYS_RESET B21 GND
A22 VGA_HS B22 VGA_VS
A23 GND B23 GND
A24 Y_SVIDEO_IN B24 C_SVIDEO_IN
A25 GND B25 VIDEO_IN
A26 GND B26 GND
A27 VGA_RED2_FI B27 GND
A28 VGA_GREEN2_FI B28 GND
A29 VGA_BLUE2_FI B29 GND
A30 GND B30 GND
A31 AUDIO_L_IN B31 AUDIO_R_IN
A32 GND B32 GND
A33 Y_SVIDEO B33 C_SVIDEO
A34 GND B34 VIDEO_OUT2_FI
A35 GND B35 GND
A36 AUDIO_R B36 AUDIO_L
73 GND 74 GND
75 GND 76 GND
77 +12V 78 +12V
79 +12V 80 +12V
² 1 socket connected to the USB port on DVD-RW module, which is defined as follows:

Table 3-27 Socket between I/O Port Board and USB Port on DVD-RW Module

No. Name
1 +5V
2 +5V
3 DM6
4 DP6
5 GND
6 GND
² 1 power supply socket of DVD-RW module, which is defined as follows:

3-26
System Principle

Table 3-28 Socket between I/O Port Board and Power Supply of DVD-RW Module

No. Name No. Name


1 +5V 2 GND
3 +12V 4 GND

² 1 socket connected to VGA port of LCD, which is defined as follows:

Table 3-29 Socket Connected to VGA Port of LCD

No. Name
1 VGA_V
2 GND
3 VGA_H
4 GND
5 VGA_R
6 VGA_B
7 VGA_G
8 GND
9 GND
10 GND
² 1 socket of speaker output, which is defined as follows:

Table 3-30 Speaker Output Port

No. Name
1 LEFT+
2 LEFT-
3 RIGHT+
4 RIGHT-
² 1 socket of W/B video printer, which is defined as follows:

Table 3-31 Socket of W/B Video Printer Output

No. Name
1 VIDEO_OUT
2 VIDEO_OUT
3 GND
4 GND
5 RM_PRINT
6 RM_BUSY
² 1 socket between I/O port board and control panel, which is defined as follows:

Table 3-32 Socket between I/O Port Board and Control Panel

No. Name No. Name


1 DM5 2 HD_STATUS

3-27
System Principle

No. Name No. Name


3 DP5 4 BREAKER_STATUS
5 GND 6 GND
7 +12V 8 +12V
9 GND 10 GND
11 +5V 12 +5V
² 1 socket between I/O port board and power supply soft switch, which is defined as follows:

Table 3-33 Socket between I/O Port Board and Power Soft Switch

No. Name
1 POWER_SSW
2 GND
² 1 socket of power supply auxiliary output control, which is defined as follows:

Table 3-34 Socket of Power Supply Auxiliary Output Control

No. Name
1 +12V
2 +12V
3 GND
4 GND
² Socket connected to built-in USB port of digital W/B video printer, which is defined as
follows:

Table 3-35 Socket Connected the USB Port of Digital W/B Video Printer

No. Name
1 USB_DM1BUS
2 USB_DM1BUS
3 DM1
4 DP1
5 GND
6 GND

3.7.3 Introduction of Testing Points on IO Port Board

3.7.3.1 USB
The working status of USB channel can be seen from the indicator light on the IO port board. If
the light is still off after booting strap, it means no power on the board or the first USB received
no signal from the up-channel USB, or the chip didn’t be initialized. Check if there are clock
waves on C58, C59, if not, it means the chip or crystal is damaged.

3.7.3.2 Video Collection


th
The video collection part tests the 9 pin of TVP5150 to check if it is working at 27MHz, if not,

3-28
System Principle

check the input frequency of crystal; If the input frequency of crystal is 14MHz, and the voltage
th rd
of 4 pin is 1.8V and the 3 pin is of 0V, then the chip should be replaced. Check if the signal
th
frequency of R28 is about 12.36MHz, if not, check if the frequency of 6 pin on EMP202 is
normal, if it is still abnormal, please replace the IC. If USB device can be recognized by PC,
then EM2860 is normal.

3.8 IO Rear Panel


IO rear panel is installed on the back of the machine, it provides all kinds of ports to the external
devices (most of them are standard ports), the main signals are from IO port board as well as
network ports of the master board.

3.8.1 Sockets Definition of IO Rear Panel


The definition is only concerned on the un-standard sockets:
² 1 socket connected to I/O port board, which is defined as Table 3-26.
² 1 socket connected to RESET button, which is defined as follows:

Table 3-36 Definition of Socket Connected to RESET Button

No. Name
1 SYS_RESET_FI
2 GND
² 1 reserved socket to the power supply of fan, which is defined as follows:

Table 3-37 Socket Definition of Fan Power Supply

No. Name
1 +12V
2 GND
3 +12V

4 GND

3.9 CW Board
Principle of CW board is shown as follows. It mainly consists of a CW receiving channel and
pencil probe receiving & transmitting channel. It can be classified into several modules, which
include socket unit (power supply filtering and transforming), delay-line unit, pencil probe unit,
low pass filter unit, demodulation unit, wall filter wave unit, gain adjusting unit and A/D
conversion unit; power supply on the board can be classified into analog and digital one, the
former one includes A+5V, VCC(5V), A-5V, A+12V, A-12V, PHV, A+1v5, and the latter one
includes VDD(3.3V).

3-29
System Principle

+/-5V

3.3V

Single-end
CW to
0 differential
Connector 1

Analog I_V switching Amplifier Wall filter Amplifier 2


delay- Reference
line Electro Quadrature level drive
nic demudula- ADC
I-V switching
switch tor
Low-pass
CW filter
9

Connector 2
Single-end
Amplifier 2 to
I_V switching Amplifier Wall filter differential
Reference
LOCLK( level drive
differential)
Low noise amplifier 8MHZ
Pencil
probe
socket SW_CTRL
Passive Transm
Low- Drive
itting 2M square wave
pass Mos
drive
filter
Analog power: +5V, -5V, PHV,
+12V,
-12V (external power), +1.5V
digital power: +3.3V (external input)

+/-12V

PHV

Fig 3-7 Principle Diagram of CW Panel

3.9.1 Functional Modules Introduction

3.9.1.1 Interface Circuit


The interface circuit module consists of analog signal socket (connect 1), digital signal socket
(connect 2), power supply filtering circuit and 1.5V transforming circuit, for the detailed definition
of pins, please refer to definition of analog and digital signal sockets.
Filtering module is used to filter the power supply that provided by the main control board, and
there are indicator lights for +5V and 3.3V power supply, the return circuit of indicator lights is
connected in series with current-limiting resistance of 510 Ohm, the diagram is shown as follows:

3-30
System Principle

Table 3-8 Diagram of Power Indicator Light

In 1.5V converting circuit, +5V is converted into 1.5V through voltage drop IC, and voltage
control is determined by the peripheral sampling resistors according to R1=R2*[(VOUT/0.9)-1].

3.9.1.2 Delay-line Circuit


The main function of delay-line module is to combine CW signal into a current signal after
processed by the delay-line. But in order to adapt to the later quadrature demodulation, filtering,
amplification and A/D conversion, the current signal should to be transformed into voltage signal.
So the delay-line unit consists of two parts: the one is delay-line circuit and the other is I-V
conversion circuit.

3.9.1.3 Pencil Probe Unit


Pencil probe unit consists of two parts: transmitting circuit and receiving circuit. Of which, the
transmitting circuit is shown as follows:

3-31
System Principle

PHV

CW Pencil
CON2 MD1201 TC2320
probe

Fig 3-9 Transmitting and Receiving Circuits of Pencil Probe

The function of receiving circuit is to amplify the signal received by the pencil probe, and then
send the signal to the front-end filter to share the after filtering demodulation and A/D sampling
circuit with the voltage wave output from CW, and before sharing, both signals will be switched
by analog switch.

3.9.1.4 Low-pass Filter


Two functions are involved in this unit, the one is to make channel selection for CW voltage
wave and signals received by the pencil probe; the other is to filter the signal input from analog
switch. In application, channel selection is controlled by signals from the main control board, S1,
S2 are connected to CW voltage signal and the signal received by the pencil probe respectively.
Controlling signal SW_CTRL is produced by FPGA, at the same time, a pull-down resistance is
connected to the pin of analog switch to ensure the default input channel is CW input. For the
filter, two active filtering units are involved to make a four–scale low-pass filter. The low-pass
filter is applied to filter out the high frequency signals (above 2.5M). The principle diagram is
shown as follows:

Fig 3-10 Principle Diagram of Analog Switch

The block diagram of low-pass filter unit is shown as follows:

3-32
System Principle

U1
In 0 3 VSS R6
1 1 50 4
2 2 45 5
VCC 99 10 C1
C3
VSS
C8 0

R8

Fig 3-11 Principle Diagram of Low-pass Filter

3.9.1.5 Demodulation Circuit


Quadrature demodulation unit is to convert the analog signal sent from the front filter into I/Q
baseband signal by quadrature demodulating, the I/Q baseband signal is to be applied by the
sampling unit.

3.9.1.6 Wall Filter Wave Circuit


The I and Q voltage signals have to be processed by low and high pass filter after quadrature
demodulation. The main object of high-pass filter in CW Doppler system receiving circuit is to
eliminate DC level caused by interference and the low frequency signal resulted from slow
movement tissue, then dynamic range of ADC can be fully used. The structure is shown as
follows. Low-pass filter of CW receiving circuit acts to filter the wide-range high frequency
harmonic wave caused by frequency mixing and to filter the wide-band noise.
C1=100nf

R1=23.45k

C2=100nf

R2=31.24k

R3=25.0k

Fig 3-12 Structure of High-pass Filter (Two-scale Bessel)

3.9.1.7 Gain Adjustment Circuit


This unit can be divided into two parts: before the high-pass filter and after the low-pass filter
respectively. The current design is to amplify 13dB before the filter, the gain is controlled by a
feedback resistance that the gain varies with the resistance. The position of first-order gain
circuit is reserved for adjustment of circuit gain mode according to the system requirements in
the final period. The current way is to short-circuit 0 Ohm resistance. The purpose to amplify the
gain is to adjust I, and Q signal to the full range of A/D converter.

3-33
System Principle

3.9.1.8 A/D Converting Circuit


The I and Q signals, after the gain is adjusted by AD8671, will finally be sampled by A/D
converter of high resolution. In ADC, the analog input is differential monopole signal, while
signal output at stage of amplification is single-end signal. So, a single-end to differential circuit
is required, and a DC biasing voltage of 2.5V is input into ADC. So a drive circuit of reference
level is added to this unit.

3.9.2 Sockets Definition


Sockets between the board and system can be divided into three parts: analog signal socket,
digital signal socket and pencil probe port.

3.9.2.1 Analog Signal Socket


Signals at analog signal socket J1 include the input of CW current signal and power input of
analog circuit. The definition is shown in Table A-3. For the definition of signal, please refer to
Table 3-4.

3.9.2.2 Digital Signal Socket


Signals at digital signal socket J2 include all the digital signals between each board and the
main control board, as well as power input digital signal, for the definition, please refer to Table
A-4.

3.9.2.3 Pencil Probe Port


The pencil probe port J3 includes the receiving and transmitting sockets of the probe.

Table 3-38 Pencil Probe Port

PIN NUM SIGNAL


1 PIN
2 PENPRESENT
3 AGND
4 POUT

Table 3-39 Definition of Pencil Probe Port

Name Description
PIN Receive
POUT Transmit
PENPRESENT Presenting signal

3-34
System Principle

3.9.3 Layout of CW Board

Fig 3-13 Components Layout of CW Front Board

3-35
System Principle

Fig 3-14 Components Layout of the Rear Control Board

3.10 Control Panel

3.10.1 The Buckled Board on the Control Panel


This buckled board is the control center of the whole control panel, it consists of FPGA, SDRAM,
SPI Flash, reset circuit and USB control chip; FPGA consists of soft kernel NiosII and other
necessary logic circuits, the block diagram of the buckled board is shown as follows:

3-36
System Principle

Fig 3-15 Block Diagram of the Buckled Board

After powered by 3.3V and the reset signal output from R_Config Module became effectless, the
buckled board will read out data from SPI Flash and allocate FPGA, after the allocation is
completed, program Boot Loader saved in FPGA will copy the application program Boot Loader
to SDRAM, and the whole board starts to work.

3.10.2 Control Panel


The peripheral components of control panel include encoders, keys, trackball, LED, buzzer,
TGCs and so on. The control panel is powered by 5V and 12V external power supply. The
encoders and trackball are 5V powered, LEDs and drive circuit are 12V powered, and other
circuits are 3.3V powered.
The block diagram of control panel is shown as follows:
5V

Rotating
Trackball
encoder

Signal
Buzzer standardization LDO
circuit
3.3V

Control panel buckled board socket JI

TGCadjus
Key LEDdrive
t(8 SPI ADI Board ID
matrix circuit
sections)

Ref V LEDscanning 12V


(2.5V) matrix

3.3V

Fig 3-16 Block Diagram of the Control Panel

3-37
System Principle

3.10.2.1 LED Drive Circuit


The LED matrix consists of 4 arrays, the signal in row and signal in array are controlled
differently. The principle is show as Table 3-16.
Signals in array are connected to the LED anode, driven by EL7212, when the control signal is
of high level (input pin of EL7212), LED anode is connected to 5V power, LED is array enable.
Signals in row are connected to the LED cathode, when 74LS07 outputs low level under the
control of FPGA, LED is row enable.
When both the row and array control signals of a LED are enable, this LED will be lightened.

3.10.2.2 Key Scanning Circuit


The key matrix consists of 11 rows and 12 arrays, the signal in row and signal in array are
controlled differently. Testing signal will be output from FPGA row by row, when press a certain
key, the signal in row will corresponding to a signal in array. The location of the pressed key can
be known from the number of row and array.

74LS0
7 L0 R

L1 R
F R
L2
P
G
A R0
R1 GND
R2

74HC14

Fig 3-17 Principle Diagram of Key Scanning

3.10.2.3 Socket Definition

Table 3-40 Socket Definition of the Control Panel

Socket Name Pin definition Remarks


number
on PCB No. Pin name Specification

J2, J6, Single-encoder 1 GND Signal grounding


J9 socket 2 EDA Encoder signal A
3 KLx Key scanning signal
in row
4 LRy Key scanning signal
in array
5 VCC 5V power supply

3-38
System Principle

6 EDB Encoder signal B


1 VCC 5V power supply
2 GND System grounding
Power supply 3 GND System grounding
J3
socket 4 GND System grounding
5 VPP 12V power supply
6 VPP 12V power supply
1 DP USB-DP signal
2 DN USB-DN signal
Power supply 3 GND System grounding
J7
socket 4 HD_Status Signal of HDD status
indicator light
5 Break_Status Break
1 ED1A Encoder 1 signal A
2 ED1B Encoder 1 signal B
3 ED2A Encoder 2 signal A
4 ED2B Encoder 2 signal B
5 ED3A Encoder 3 signal A
6 ED3B Encoder 3 signal B
7 ED4A Encoder 4 signal A
8 ED4B Encoder 4 signal B
9 ED5A Encoder 5 signal A
10 ED5B Encoder 5 signal B
11 ED6A Encoder 6 signal A
12 ED6B Encoder 6 signal B
Multi-encoder
J5 13 KR9 Key scanning signal
socket
in array 9
14 KL8 Key scanning signal
in row 8
15 KR10 Key scanning signal
in array 10
16 KL9 Key scanning signal
in row 9
17 KR11 Key scanning signal
in array 11
18 NC Not connected
19 VCC 5V power supply
20 GND System grounding

3.11 ECG Module (optional)

3.11.1 Principle and Functions


The function of the ECG board contains: ECG signal detection; ECG waveform display;
ultrasound image reference signal; real-time 2-D image and color flow image synchronous
signal.

3-39
System Principle

The electrocardiosignal, after been amplified, filtered, and sampled, will be uploaded to the PC
for R-wave testing, the tested cardioelectrical trigger signal will be uploaded to PC through USB
port. Block diagram is shown as follows:

Right-
Electrode
leg
2
drive

Electrode
Difference High- Amplifica- Control signal
1 Input Pace DSC
Buffer amplifica- pass tion low- ECG serial port
Electrode protection rejection control
tion filter pass filter
2 Data serial system
port 2105

Testing signal 1 A
Lead off MCU
Buffer D
checking
Testing signal 2
PC
UART-to USB
2108

Power Footswitch DC Bias


DC_IN
supply port IN adjust

3.11.2 Outlet Definition

Table 3-41 Socket Definition of ECG Module

Socket Name Pin definition Remarks


number
on No. Pin name Specification
PCB
J1 DC_IN socket 1 FGND ECG module
floating ground
2 FGND ECG module
floating ground
3 DC_IN DC IN input
4 DC_IN DC IN input
5 FGND ECG module
floating ground
6 FGND ECG module
floating ground
J2 Lead cable socket 1 Lead_F_IN Lead cable, Green
2 Lead_R_IN Lead cable Red
3 Lead_N_DR Lead cable, Black
(Fixed to the right
leg)
4 NC Not connected

3-40
System Principle

5 C_SHIELD Cable shielding level


drive
6 NC Not connected
J3 Main system 1 GND Earth terminal of
socket main system
2 +12V Power supply of the
main system
3 GND Earth terminal of
main system
4 +12V Power supply of the
main system
5 ECG_CTRLT Serial control
transmitting
6 +12V Power supply of the
main system
7 ECG_CTRLR Serial control
receiving
8 GND Earth terminal of
main system
9 GND Earth terminal of
main system
10 GND Earth terminal of
main system
11 ECG_DATAT Serial data
transmitting
12 GND Earth terminal of
main system
13 ECG_DATAR Serial data receiving
14 GND Earth terminal of
main system
15 GND Earth terminal of
main system
16 GND Earth terminal of
main system
17 ROW Common input
terminal of
footswitch
18 GND Earth terminal of
main system
19 LINE1 Input of switch cable
1
20 LINE2 Input of switch cable
2
J4 USB port 1 VBUS 5V input
2 DM Differential signal
DM

3-41
System Principle

3 DP Differential signal
DP
4,5, GND Earth terminal of
6 main system
J5 Footswitch socket 1 ROW Common input
terminal of
footswitch
2 LINE1 Input of footswitch
cable 1
3 LINE2 Input of footswitch
cable 2
4 NC Not connected
J6 JTAG debugging 1 VCC 3.3V Applied in
socket 2 GND ECG module Debugging
floating ground
3 TRST JTAG — — TRST
signal
4 TCK JTAG — — TCK
signal
5 TDI JTAG——TDI signal
6 TDO JTAG — — TDO
signal
7 TMS JTAG — — TMS
singal
8 RTCK JTAG — — RTCK
signal
9 NRST JTAG — — NRST
signal
10 NC Not connected
J7 ARM BOOT 1 / To pin LPC213X Applied in
socket P0.14 repairing
2 GND ECG module
floating ground
J8 USB port 1 VBUS 5V VBUS signal
2 USB-
3 USB+ JTAG — — TRST
signal
4 TCK JTAG — — TCK
signal

3.11.3 Plug-in Diagram and Components


The blue broken line means electrical insulation, the right part is power supply input and
communication ports, and the left part is electrocardiasignal processing circuit.

3-42
System Principle

Lead cable socket J2

J8 USB port
(2109 exclusive)

J4 USB port
(2108 exclusive)

ARM BOOT socket


J7

Footswitch socket
J5 (2105 exclusive)

Main system socket J3


DC_IN socket J1 JTAG debugging
(2105 exclusive)
socket J6
At the back of board

Fig 3-18 ECG Plug-in Diagram and Components Layout

3.12 USB-to-IDE Port Board


The structure of USB-to-IDE port board is shown as Table 3-19. The kernel is GL811S, which
supports protocols USB2.0 and ATA/ATAPI-6 1.0, it will convert IDE port to USB port.
The chip is connected to 12MHz crystal X1, the power supply voltage of the chip is 3.3V, and it
is powered by the VBUS of USB. LDO (U2) on the board will convert 5V voltage into 3.3V.

3-43
System Principle

Crystal

Electrostatic
IDE USB
GL811S prevention,
socket port
filter

Power
5V supply VBUS
module

Fig 3-19 Block Diagram of USB-to-IDE Module

3.12.1 Socket Definition

Table 3-42 Socket Definition of USB-to-IDE

Socket Name Pin definition Remarks


number
on PCB Pin No. Pin name Specification

J1 IDE port 1 RESET IDE device reset


signal
3~18 DD0~DD15 IDE device data
cable(16 bit)
2,19,22 GND Earth grounding
,24,26,
28,30,4
0
20 NC Free
21 DMARQ DMARQ signal
23 DIOW DIOW signal
25 DIOR DIOR signal
27 IORDY IORDY signal
29 DMACK DMACK signal
31 INTRQ INTRQ signal
32 OBSOLETE Pull-down 5V
33 DA1 DA1 signal
34 CBLID/PDIA NC
G
35 DA0 DA0 signal
36 DA2 DA2 signal
37 CS0 CS0 signal
38 CS1 CS1 signal
39 DASP Communication
indicator light

3-44
System Principle

J2 USB port 1,2 VBUS 5V


3 DM USB signal DM
4 DP USB signal DP
5,6 GND System grounding

3.13 Power Supply System

3.13.1 Basic Functions of Power Supply System


The power supply system, which powers the whole system, consists of power supply connecting
board, isolating transformer, adapter, the power supply master board and auxiliary board. The
layout of each part is shown as Table 3-20. In Fig 8-7, the input power supply is mains supply,
the different connections of primary and secondary windings are changed by the
voltage-controlled switch, so the AC input voltage can be changed in different countries. There
are four AC power outputs on the power supply connecting board, of which, three are for the
external devices and one for adapter. The adapter output is connected to the power supply
master board and auxiliary board via system master board, the DC outputs provided to the main
machine by the two power supply boards are listed as follows:

Table 3-43 DC Output of Power Supply System

No. Output Board Remarks


1 +12V Master Controlled by power_on signal
2 5Vstb Master Keep output
3 5Vstb_CPU Master Controlled by 5Vstb_CPU_EN signal
4 +5V Master Controlled by power_on signal
5 +3.3V Master Controlled by power_on signal
6 THV Master Controlled by power_on signal
7 +2.5V Auxiliary Controlled by power_on signal
8 +1.2V Auxiliary Controlled by power_on signal
9 -5V Auxiliary Controlled by power_on signal
10 -12V Auxiliary Controlled by power_on signal

3.13.2 Basic Principle of Power Supply System


There are four AC power outputs on the power supply connection board, of which, three are for
the external devices and one for adapter, the +12V output is input to the DC bus of the power
supply master board and the auxiliary board.

The system master board and power supply master board are connected through a 64-pin
socket, signal is defined as Table A-8. In Table 3-20, the inputs of related signals on the system
master board and power supply to the power supply master board is finished through the 64-pin
socket. The power supply master board and auxiliary board output +12V, +5V, 5Vstb, 5Vstb-cpu,

3-45
System Principle

+3.3V, THV; the two boards are connected through 26-pin socket (signal definition is shown as
Table 3-44). The Vbus+, +5v, EDC_Power signals that input to the power supply master board
are transformed into -12V, -5V, +2.5V, and +1.2V outputs by the auxiliary board, the four outputs
are returned to the power supply master board through 26-pin socket, and then be output to
each electro-terminal on the system master board through 64-pin socket.

Isola-
Primary winding
ting tap
Mains input

transfor Power
mer
Secondary
winding tap
supply
connec-
ting
board
Peripherals output
AC output
(3 channels)

Power
supply
adapter
Edc_power Edc_power Edc_power

Power_on Power_on
CW_mode
5Vstb_cpu_en
THV_range
TEMP_D+
TEMP_D- start

+12V VBUS+

+5V +5V

+5Vstb
Power Power
System
supply supply
master +5Vstb_cpu
master auxiliary
board
+3.3V board board

THV

+2.5V +2.5V

+1.5V +1.5V

-5V -5V

-12V -12V

Fig 3-20 Block Diagram of Power Supply System

3-46
System Principle

Table 3-44 Definition of Signals between the Power Supply Master Board and Auxiliary Board

No. Name No. Name


1 GND 2 -5.3V
3 Scan_status 4 -12V
5 Power_on 6 +2.5V
7 B_Battery_NTC 8 +2.5V
9 A_Battery_NTC 10 GND
11 GND 12 GND
13 B_Battery+ 14 1.5V_feedback
15 A_Battery+ 16 +1.5V
17 Edc_power 18 +1.5V
19 GND 20 GND
21 GND 22 GND
23 Vbus+ 24 +5VL
25 GND 26 Start

3.13.2.1 5vstb Circuit


The Edc_power output from the adapter is the input of 5vstb circuit, 5vstb is gained after linear
regular voltage chip MIC5202-5.0YMS.
MIC5202-5.0YMS integrated over-current and over-temperature protection functions internally.

3.13.2.2 +12V Circuit


Voltage rise-and-fall IC LTC3780EG#PBF of Linear Company is applied in +12V circuit, whether
the input voltage is higher or lower than 12V, the output voltage will keep at 12V.
Both +12V over-current and over-voltage protection will not be locked, the over-current point is
6.5A~9.5A, and the output voltage is resumed to normal after over-current disappeared.

3.13.2.3 +5V and +3.3V Circuit


Both +5V and +3.3V are provided by LTM4600 of Linear Company, the chip is of voltage drop
synchronous, integrates two MOSs and power inductor.
LTM4600 integrated over-current protection function internally.

3.13.2.4 THV Circuit


THV consists of a step-down circuit and step-up circuit, with the output voltage of the step-down
circuit is 5~10V (the voltage is controlled by THV_range, 0V is corresponding to 5V of THV,
and 4V to 10V of THV), and output voltage of the step-up circuit is 20~140V (the voltage is
controlled by THV_range, 0V is corresponding to 20V of THV, and 4V to 140V of THV), the
step-down circuit keeps working all the time; the step-up circuit is controlled by cw_mode, it
works only when the cw_mode is of low level.
The model of step-down circuit control chip u10 is TPS54350, it integrates a MOS.
The model of step-up circuit IC U2 is the commonly used TL594CDR2G, there is PMOS U6

3-47
System Principle

between the input of step-up circuit and +12V.


There is no over-voltage protection in step-down circuit, while such function exists in the step-up
circuit (20V~140V), the over-voltage is about 170V, the voltage keeps at about 170V without
been locked.
There is over-current protection in both step-up and step-down circuits, over-current appears in
any circuit will stop the work of both circuits and lock the circuit.

3.13.2.5 +2.5V and +1.5V Circuit


Both +2.5V and +1.5V are provided by the commonly used synchronous rectifying voltage drop
IC EL7566, it is a chip of high reliability with few malfunction appeared in the products.
EL7566 integrated over-current protection function internally, it will not be locked in case of
over-current, and it will resume to normal status after over-current disappeared.

3.13.2.6 -5V and -12V Circuit


Both -5V and -12V are provided by the commonly used IC Max1847 that applied to generate
negative voltage. Both circuits have over-current and over-voltage protection that both outputs
will be stopped in case of over-voltage and over-current in any circuit, and the circuits will be
locked.

3.14 System Power Supply Distribution


The power supply distribution is shown as follows:

Table 3-45 Power Supply Distribution

No. Power supply Loads


1 +5VSTB Power supply management CPLD, crystal, 244 chip and so on.
CPU module, audio operational amplifier, transmitting board, CW
2 +12V board, keyboard board, CD-ROM driver, fan
3 +5VSTB_CPU CPU module
4 -5V High voltage insulation, CW panel
5 -12V CW panel
CW panel, high voltage isolating, beamformer FPGA, hard disk, key
board, CD-ROM-driver, audio encode and decode chip, electrostatic
prevention chip, parallel port drive, gain and remote control high
6 +5V voltage control DA, probe module, video collection
7 +2.5V DDR, FPGA
8 +1.5V FPGA
FPGA, channel selection control CPLD, ADC, transmitting board,
VCA, audio encode and decode chip, 2245 chip, clock drive chip,
crystal, system reset chip, FPGA configure FLASH, video encoder,
9 +3.3V video amplifier, SSRAM, USB HUB
10 PHV Transmitting board, CW board

3-48
System Principle

3.15 LCD

3.15.1 Inverter

3.15.1.1 Block Diagram and Principle Diagram

Voltage
feedback

Voltage-up
Switch power PWM control MOSFET
transfer
Switch-
AC ing
power
input

Output to the
light
On/Off

Luminance
control Current
feedback

Fig 3-21 Block Diagram of LCD Inverter

Fig 3-22 Principle Diagram of LCD Inverter Power Supply

3-49
System Principle

Fig 3-23 Principle Diagram of LCD Inverter

3.15.1.2 Working Principle


Both power supply and inverter are integrated on the power supply board, the power supply,
which provides 12V and 5V power to the inverter and AD board, adopts power switch circuit with
wide voltage range. The inverter works in full bridge mode that 4 MOSFETs are connected into
a bridge, the work status (close/ open) is driven by the main IC, and AC voltage is imposed on
the primary winding of the transformer. The secondary square wave of the transformer, after LC
resonance, is changed into sine wave to drive the CCFL gives out light. The light is controlled by
the switch signal (the signal will control the enable pin of the main IC), the brightness of the light
is controlled by the brightness signal, which will control the duty cycle of square wave of main
IC.
The power supply board consists of open circuit, over-voltage and over-current protection circuit.
For example, when open circuit happened to any of the light, the open circuit protection circuit
will act; when the current is too high, the feedback signal from the light to main IC will limit the
duty ratio of the square wave to decrease the current for circuit protection; when the output
voltage is too high, the feedback signal from the light to main IC will turn off PWM drive.

3-50
System Principle

3.15.1.3 Sockets Definition

CN4
CN5 CN1

CN100
AC power input
CN2
CN3

Fig 3-24 Sockets Location

Table 3-46 Socket of between Power Supply and Inverter

CN1 Pin definition


Pin1 ADJ(Backlight Adjust)
Pin2 EN(Backlight ON/Off)
Pin3 GND
Pin4 GND
Pin5 12V
Pin6 12V
Pin7 PS(Power Saving)

Lights drive output sockets CN2, CN3, CN4, and CN5 are defined as follows:

Table 3-47 Output Socket Definition of Lights Drive

Pin Symbol Description


1 Vout -H High-voltage terminal
2 Vout -L Low voltage terminal

AC power input socket CN100 (socket model: JST, B03P-VL600V20A) are defined as follows:

3-51
System Principle

Table 3-48 AC Power Input Socket

Pin Symbol
1 L
2 N
3 GND

3.15.2 AD Controller Board

3.15.2.1 Working Principle and Block Diagram

LVDS
Panel
Power
Module&
Inverter

Cry
Scaler EEPROM
stal
MCU
Cry IIC
stal

VGA Input KeyPad

Fig 3-25 Principle Diagram of AD Controller Board

After VGA signal is input to RTD2620, MCU will carry out functions including signal recognition,
zoom, color process and transform signal into LVDS signal and then display it to LCD. U6
(MTV412) is MCU, functions of MCU include: controlling on RTD2620 and external related
operation, such as operation on the keys and IIC. U1, U3, U4 carry out DC-DC voltage
converting, thus providing normal voltages to each circuit.

3.15.2.2 Main ICs and the Functions

Table 3-49 Functions of IC

Item Model Function description

U1 LM2596-33 DC-DC converting, 12V to 3.3V

U3 AMS1117 2.5 3.3V-to-2.5V converting

U4 AIC1804-18PM 3.3V-to-1.8V converting

3-52
System Principle

U5 24C16 EEPROM

U6 MTV412 MCU

U7 RTD2620 Signal processing

U8 K4D263238I GDDR

U9 4435 Panel Power On/Off

Y1/Y2 K24.000 Crystal

3.15.2.3 Socket Definition

Fig 3-26.Sockets Location

Table 3-50 LVDS Sockets

J1 Pin definition J1 Pin definition


Pin1 GND Pin2 GND
Pin3 RXOIN3+ Pin4 RXOIN3-
Pin5 GND Pin6 RXOCKIN+
Pin7 RXOCKIN- Pin8 GND
Pin9 RXOIN2+ Pin10 RXOIN2-
Pin11 GND Pin12 RXOIN1+
Pin13 RXOIN1- Pin14 GND

3-53
System Principle

Pin15 RXOIN0+ Pin16 RXOIN0-


Pin17 GND Pin18 GND
Pin19 PANEL_VCC Pin20 PANEL_VCC

Table 3-51 Input Socket of Speaker

J2 Pin definition
Pin1 L-SDBX+
Pin2 L-SDBX-
Pin3 R-SDBX+
Pin4 R-SDBX-

Table 3-52 Signal Input Socket

J3 Pin definition
Pin1 VSYNC Pin2 GND
Pin3 HSYNC Pin4 GND
Pin5 RED Pin6 BLUE
Pin7 GREEN Pin8 GND
Pin9 GND Pin10 GND

Table 3-53 Output Socket of Speaker

J4 Pin definition
Pin1 L-SDBX+
Pin2 L-SDBX-
Pin3 R-SDBX+
Pin4 R-SDBX-

Table 3-54 Socket of Key Board

J5 Pin definition
Pin1 R_LED
Pin2 G_LED
Pin3 AD0
Pin4 AD1
Pin5 GND
Pin6 GND
Pin7 12V
Pin8 5V

3-54
System Principle

Table 3-55 Control Socket of Power Supply and Converter

J6 Pin definition
Pin1 ADJ(Backlight Adjust)
Pin2 EN(Backlight ON/OFF)
Pin3 GND
Pin4 GND
Pin5 12V
Pin6 12V
Pin7 PS(Power Saving)

3-55
System Structure and Assembly / Disassembly

4 System Structure and Assembly


/ Disassembly

4.1 Overall Exploded View

Fig 4-1 Overall Exploded View

4-1
System Structure and Assembly / Disassembly

No. Material
Name No. Material number Name
number
2109-30-76182 Main unit left 15 Monitor module
1 2109-30-76336
cover
2109-30-76194 DVD-R/W 16 2109-30-76189 Up/down air spring
2
module module
2109-20-76214 Isolating 17 2109-20-76078 Rotating axis barrel
3
transformer
2109-30-76181 Main unit rear 18 2109-30-76178 Front cover of main
4
cover unit module
2109-30-76199 Main unit box 19 2109-30-76183 Main unit right
5
module cover
6 2109-30-76188 IO module 20 2109-30-76193 ECG module
2109-30-76187 Power input 21 2109-30-76258 Probe board
7
module
2109-21-76337 Power adapter 22 2109-30-76179 Main unit frontal
8 and connecting cover
cable
2109-30-76177 Main unit top 23 2109-20-76091 Cover of printer
9
cover compartment
2109-20-76059 Cast aluminium 24 2109-30-76180 Footswitch
10
-51 keyboard base
2109-30-76169 Support arm 25 2109-30-76185 Main unit housing
11
base module
2109-20-76071 Cover of cast 26 2109-20-76145 Dustproof net
12
aluminium base bracket
2109-30-76261 Control panel 27 2109-30-76184 System soft switch
13
module
14 2109-30-76176 Support arm

4-2
System Structure and Assembly / Disassembly

4.2 Exploded View of Each Module

4.2.1 Exploded View of LCD Display and Support Arm

8 11

10 12

13
6

9
5

4
3
2

14
1

Fig 4-2 Exploded View of LCD Display and Support Arm

No. Material
Name No. Material number Name
number
Cover of lower Cover of upper
1 2109-20-76057 8 2109-20-76058-51
support arm support arm
Upper arm
2 2111-20-73313 turning auxiliary 9 2111-20-73285-51 Decorative cover
pressing block
Dish-shaped
rd
3 M6T-020002--- spring, the 3 10 2111-20-73325 Damp axis
series D=50
Upper arm
4 2111-20-73296 turning auxiliary 11 2111-20-73288 Monitor rear cover
washer
Monitor
Lower support
5 2109-30-76170 12 2109-20-76334 components-install
arm
board module
Upper support
6 2109-30-76174 13 2111-30-73334 Frontal cover
arm
Bottom
Monitor cable lights(included in
7 2111-20-73289 14 2111-30-73335
cover the frontal cover
module)

4-3
System Structure and Assembly / Disassembly

4.2.2 Exploded View of Monitor Components-install Board

Fig 4-3 Exploded View of Monitor Components-install Board

4-4
System Structure and Assembly / Disassembly

4.2.3 Exploded View of Main Unit Box Module

Fig 4-4 Exploded View of Main Unit Box Module

4-5
System Structure and Assembly / Disassembly

Material
No. Name No. Material number Name
number
Power and
Power supply main
1 2109-20-76326 signal cable of 16 2109-30-76310
board
SATA
HDD bracket Power supply
2 2109-20-76305 17 2108-20-65745
separating board
Hard disk 40G
Power supply
3 0000-10-11267 5400rpm, 2.5 18 2109-30-76312
auxiliary board
inch SATA port
Bottom cover of
4 2109-20-76150 Main box 19 2108-20-65744
power supply box
IO connecting Shielding cover of
5 2109-30-76201 20 2109-20-76158
board port board
Probe
6 2109-30-76211 connecting 21 2109-30-76224 IO port board
board
Board guide-pin Transmitting board
7 2109-20-76155 22 2109-20-76149
socket shielding cover
Shielding frame
of probe
8 2109-20-76154 23 2109-30-76456 Transmitting board
connecting
board
Shielding cover CW board shielding
9 2109-20-76472 24 2108-20-65750
of CPU board cover
10 0000-10-11188 CPU board 25 2108-30-65899 CW Doppler board
CPU heat
11 2109-20-76329 26 2109-30-76294 Master board
conducting mat
Lower shielding
12 2109-20-76473 CPU radiator 27 2109-20-76061 cover of master
board
CPU fan and
13 2109-20-76238 connecting the 28 2109-20-76153 Air inlet fan frame
cable
Cover of main Air inlet fan and the
14 2109-20-76151 29 2109-20-76231
unit box connecting cable
Top cover of
15 2108-20-65743 power supply
box

4-6
System Structure and Assembly / Disassembly

4.2.4 Exploded View of Probe Board Module

Fig 4-5 Exploded View of Probe Board Module

Material Name Name


No. No. Material number
number
Probe board Probe aluminium
1 2109-20-76161 5 2105-20-40131
shielding cover shielding cover
Guide-pin
2 2109-20-76160 6 2109-30-76311 Probe board
installation bar
Aluminium
Probe control
3 2105-20-40132 shielding frame 7 2109-30-76309
board
of probe port
Probe
Probe board
4 2105-20-40133 installation 8 2109-20-76162
shielding cover
bracket

4.3 Overall Disassemble and Parts Replace

4.3.1 Disassemble the Monitor Module


1) Lock the upper support arm by the up/ down spanner. (Refer to 2.2.3 )
2) Turn the monitor to the level position when toggle the deflector rod left side. (Refer to
2.2.2 )

3) Remove the two M4X12 screws which are used to secure the monitor cable, and then
remove the monitor cable cover aslant.

4-7
System Structure and Assembly / Disassembly

1. Screw M4X12 2. Monitor Cable Cover 3. Monitor Module


4. Support Arm Module 5.Deflector Rod

Fig 4-6 Disassemble the Monitor Module (1)

4) Remove the 1 M4X12 screw securing the ground wire holder, take out the cable plug
and remove the 6 M4X12 screws that are used to secure the monitor.

1. Signal Cable 2. Audio Cable 3. Power Supply 4.M4X12 Screw


5. Ground Wire Holder 6. Screw M4X12 7. Monitor Module 8. Support Arm

Fig 4-7 Disassemble the Monitor Module (2)

4-8
System Structure and Assembly / Disassembly

5)Hold the lateral sides of the monitor, raise up the monitor to separate it with the support
arm hooker when the monitor is 20° from the vertical direction, and then remove the monitor.

Fig 4-8 Disassemble the Monitor Module (3)

4.3.2 Disassemble the Bottom Lights of the Monitor


1) Turn the monitor to one side of the main unit.
2) Remove the 2 M3X8 screws securing the light cover, and then remove the cover.
3) Take the lights from the light hanger gently.

4-9
System Structure and Assembly / Disassembly

1. Bottom Light 2. Transparent Cover 3. M3X8 Screw

Fig 4-9 Disassemble the Bottom Light of Monitor

4.3.3 Replace the Speaker on the Monitor


1) Disassemble the monitor module. (Please refer to 4.3.1 )
2) Remove the 5 M4X12 screws on the rear cover of the monitor, and then remove the
cover.

1.Frontal Cover of the Monitor 2. Rear Cover of the Monitor 3.M4X12 Screw

Fig 4-10 Replace the Speaker (1)

4-10
System Structure and Assembly / Disassembly

3) Pull out the speaker connecting cable plug and cut off the tie which is used to secure the
cable, remove the self-driving screws which are used to secure the speaker (4 screws at
each side), and then remove the speaker.

1. Self-driving Screw PT3X8 2 Speaker 3. Cable Tie


4. Monitor Front Cover Module 5.Speaker Connecting Cable Plug

Fig 4-11 Replace the Speaker (2)

4.3.4 Replace the OSD Board of Monitor


1) Disassemble the rear cover of the monitor. (Please refer to procedure 1, and 2 as
described in 4.3.3 ).
2) Pull out the connecting cable plug on the OSD board.
3) Remove the 2 PT3X8 self-driving screws which are used to secure the OSD board, and
then remove the OSD board.

1. Connecting Cable Plug 2. OSD Board 3. PT3X8 Screw 4.Connnecting Cable Plug 5. Monitor

Fig 4-12 Replace the OSD Board

4-11
System Structure and Assembly / Disassembly

4.3.5 Replace the Bottom Light Board of the Monitor


1) Disassemble the rear cover of the monitor. (Please refer to procedure 1, and 2 as
described in4.3.3 ).
2) Pull out the connecting cable plug on the bottom light module.
3) Remove the 3 self-driving screws PT3X8 which are used to secure the light module, and
then take away the module.

1. Bottom Light Module 2. Connecting Cable Plug 3. Screw PT3X8 4. Monitor Front Cover

Fig 4-13 Replace the Bottom Light Board of Monitor (1)

4) Remove the 2 M3X8 sunk screws securing the transparent cover, and remove the
cover.
5) Take off the two light tubes.
6) Remove the 3 self-driving screws PT3X8 which are used to secure the monitor light
board, and then remove the board.

4-12
System Structure and Assembly / Disassembly

1. ScrewPT3X8 2. Light Board 3. Light Secure Base


4. Light Tube 5. Cover 6. Screw M3X8

Fig 4-14 Replace the Bottom Light Board of Monitor (2)

4.3.6 Replace the Power Supply Inverter Board of the LCD


1)Remove the rear cover of the LCD.(Please refer to procedure 1, and 2 as described in
4.3.3 ).
2)Pull out the connecting cable plug of OSD board and connecting cable plug of speaker
output from the left side of LCD components-install board module.
3)Remove the 10 M3X8 screws which are used to secure the PCB shielding cover, and
remove the cover.

4-13
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. PCB Shielding Cover 3. Connecting Cable Plug of OSD Board
4. Connecting Cable Plug of Speaker Output

Fig 4-15 Replace the Power Supply Inverter Board of the Monitor (1)

4)Pull out the 4 LCD connecting cable plugs at the left side of LCD power supply converter,
and pull out the connecting cable plug at the right side between the LCD power supply board
and the driving board.
5)Remove the 5 M3X8 screws which are used to secure the LCD power supply inverter, and
then remove the power supply inverter.

1. LCD Power Supply Cable Plug 2. M3X8 Screw 3. LCD Power Supply inverter Board
4. Cable connecting LCD Power Supply Board with Drive Board 5. LCD Control Board

Fig 4-16 Replace the Power Supply Inverter Board of LCD (2)

4-14
System Structure and Assembly / Disassembly

4.3.7 Replace the LCD Screen and LCD Control Board


1) Disassemble the rear cover of the monitor, pull out the LCD OSD board connecting
cable plug and the speaker output connecting cable plug. (Please refer to procedure 1,
and 2 as described in 4.3.3 ).
2) Remove the 11 M3X8 screws securing the monitor components-install board, and
remove the board.

1. M3X8 Screw 2. Monitor Components-install Board 3. Front Cover Module

Fig 4-17 Replace the LCD Screen and LCD Control Board (1)

3)Remove the 10 M3X8 screws which are used to secure the PCB shielding cover, and then
remove the cover.

1. M3X8 Screw 2. PCB Shielding Cover 3. LCD Secure Frame

Fig 4-18 Replace the LCD Screen and LCD Control Board (2)

4-15
System Structure and Assembly / Disassembly

4)Pull out the 4 LCD connecting cable plugs at the left side of LCD power supply inverter,
and pull out the connecting cable plug on the LCD control board between the LCD power
supply board and drive board, pull out the LCD LVDS connecting cable plug.
5)Cut off the 2 ties which are used to secure the connecting cable, remove the 5 M3X8
screws which are used to secure LCD power supply inverter and the 5 M3X8 screws which
are used to secure LCD control board, and then take away the inverter board and LCD
control board.

1. Power Supply Converter Board 2. LCD Connecting Cable Board 3. M3X8 Screw
4. Connecting Cable between LCD Power Supply Board and Drive Board
5. M3X8 Screw 6.LCD LVDS Connecting Cable

Fig 4-19 Replace the LCD Screen and LCD Control Board (3)

6)Remove the 4 M3X8 screws which are used to secure LCD screen, take away the LCD
screen, and then finish the replacement of LCD screen and LCD control board.

4-16
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. LCD Shielding Cover 3. LCD Secure Frame

Fig 4-20 Replace the LCD Screen and LCD Control Board (4)

4.3.8 Replace the Air Spring of Support Arm


1) Stamp down the 4 caster brakes to lock the machine.
2) Take away the monitor module. (Refer to 4.3.1 )
3) Take the cover of upper support arm from the support arm module.

4-17
System Structure and Assembly / Disassembly

1. Upper Support Arm Cover 2. Upper Support Arm Module 3. Caster Brake

Fig 4-21 Replace the Air Spring of Support Arm (1)

4) Lock the upper support arm by the up/down spanner of upper support arm. (Refer to
2.2.3 ).
5) Push down the upper support arm by left hand, screw out the securing pins and remove
them. And then leave the arm carefully.

4-18
System Structure and Assembly / Disassembly

1. M5 Pin 2. Elastic Washer M5 3. Support Arm Module

Fig 4-22 Replace the Air Spring of Support Arm (2)

NOTE: Please drop some fixing glue onto the tail of the pin when install the pin.

6) Remove the 3 M4X8 screws which are used to secure the air spring, screw out the air
spring and connection base at the position of lifting ear of the air spring (use the clamp
to remove the cementing agent at the lifting ear).

1. M4X8 Screw 2. Air Spring and Connection Base 3. Lifting Ear

Fig 4-23 Replace the Air Spring of Support Arm (3)

7)Dispatch the shield ring by the straight screwdriver, take out the air spring pin from the

4-19
System Structure and Assembly / Disassembly

right side and then pull out the air spring.

1. Shield Ring Ø4 2. Air Spring Connection Base 3. Pin 4. Air Spring

Fig 4-24 Replace the Air Spring of Support Arm (4)

NOTE: After the support arm (including the monitor) is installed, please check if
the monitor can stay at any height, if can’t, then: a. If it keeps rising, first
adjust the screw anti-clockwise to be still by the straight screwdriver, and
then check again, if that situation continuous, replace the air spring; b. If
it keeps lowering down, first adjust the screw clockwise to be still by the
straight screwdriver, and then check again, if that situation continuous,
replace the air spring.

1. Adjusting Screw

Fig 4-25 Replace the Air Spring of Support Arm (5)

4.3.9 Disassemble the Support Arm


1) Disassemble the monitor and the upper support arm cover. (Please refer to 4.3.1 and
4.3.5 ).
2) Remove the 2 M4X12 screws which are used to secure the lower support arm cover,
slightly move the cover laterally and take it off.

4-20
System Structure and Assembly / Disassembly

1. Support Arm Module 2. Lower Support Arm Cover 3. M4X12 Screw

Fig 4-26 Disassemble the Support Arm Module (1)

3) Cut off the tie on the monitor signal cable, and pull out the cable from the joint of the
support arm.
4) Take off the plastic cover from the support arm using the straight screwdriver, and then
take off the 3 internal hexagonal screws M5X16 as well as the washer, at last separate
the support arm module from the base module.

1. Signal Cable 2. Cable Tie 3. Joint of the Support Arm


4. Support Module 5. Decorative Cover 6. M5X16 Screw and Washer

Fig 4-27 Disassemble the Support Arm Module (2)

4.3.10 Disassemble the Control Panel


1) Turn the keyboard module to the right/left side of the main machine, remove the 9
M4X12 screws which are used to secure the keyboard module at both sides of the

4-21
System Structure and Assembly / Disassembly

casting aluminium base.

1. M4X12 Screw 2. Casting Aluminium Base

Fig 4-28 Disassemble the Control Panel Module (1) (Bottom View)

2) Slightly pull out the control panel module forward in the frontal handle direction, lift the
frontal part of the panel and draw out the external cable plug connected to the control
panel module.
3) Remove the 1 M3X8 screw that jointed with the base module, and separate the control
panel module from the base.

1. Control Panel Module 2. M3X8 Screw 3. Grounding Wire


4. Casting Aluminium Base 5 M4X12 Screws 6. Cable Plug

Fig 4-29 Disassemble the Control Panel Module (2)

4-22
System Structure and Assembly / Disassembly

4.3.11 Replace the Encoder Connecting Board

4.3.11.1 Replace the Encoder 6-to-1 Connecting Board


1) Disassemble the control panel module. (Refer to 4.3.7 ).
2) Draw out the 6 knobs on the connecting board.

1. Encoder Knob 2. Control Panel Module

Fig 4-30 Replace the Encoder 6-to-1 Connecting Board (1)

3) Pull out the connecting cable plug on the control panel, remove the 6 M3X8 screws
which are used to secure the encoder 6-to-1 connecting board, then replace the board.

1. Control Panel 2. M3X8 Screw 3. Encoder 6-to-1 Connecting Board 4. Connecting

Fig 4-31 Replace the Encoder 6-to-1 Connecting Board (2)

4-23
System Structure and Assembly / Disassembly

4.3.11.2 Replace the Single-encoder Connecting Board


1) Disassemble the control panel module.(Refer to 4.3.7 )
2) Pull out the knobs on the encoder connecting boards. (1 knob on 1 board).

1. Encoder Knob 2. Control Panel Module

Fig 4-32 Replace the Single-encoder Connecting Board (1)

3) Pull out the connecting cable plug of the single encoder connecting board from the
control panel, remove the 2 M3X8 screws and replace the connecting board.

1. M3X8 Screw 2. Connecting Board of Single Encoder


3. Connecting Cable Plug 4. Control Panel

Fig 4-33 Replace the Single-encoder Connecting Board (2)

4.3.12 Replace the Trackball


1) Disassemble the control panel module. (Refer to 4.3.7 )

4-24
System Structure and Assembly / Disassembly

2) Pull out the connecting cable plug connected to the trackball, remove the 2 M3X8
screws which are used to secure the trackball, and take away the trackball.

1. Connecting Cable Plug 2. M3X8 Screw 3. Trackball 4. Control Panel

Fig 4-34 Replace the Trackball

4.3.13 Disassemble the Control Panel Buckled Board


1) Disassemble the control panel module. (Refer to 4.3.7 )
2) Loosen the board clip on the left and right sides of the buckled board, and the board will
automatically spring up, and then take the board out from the socket.

1. Buckled Board 2. Board Clamp 3. Control Panel

Fig 4-35 Replace the Control Panel Buckled Board

4.3.14 Replace the Control Panel


1) Disassemble the control panel module.(Refer to 4.3.7 )

4-25
System Structure and Assembly / Disassembly

2) Pull out the 8 TGC knobs and 9 encoder knobs upward from the control panel. (Refer to
4.3.8 )

1. TGC Knob 2. Encoder Knob 3. Control Panel Module

Fig 4-36 Replace the Control Panel (1)

3) Take off the encoder 6-to-1 connecting board, each single encoder connecting board
and the control panel buckled board.(Refer to 4.3.8 and 4.3.10 )
4) Pull out the connecting cable plug of the buzzer from the control panel.
5) Remove the 26 PT3X10 screws on the control panel and the 1 M3X8 screw which is
used to secure the grounding wire.
6) Take out the control panel and the keys from the keyboard cover, and then separate
them to finish the replacement.

4-26
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. Encoder 6-to-1 Connecting Board 3. Single Encoder Connecting Board
4. M3X8 Screw 5. Control Panel Buckled Board 6. PT3X10 Self-driving Screw
7. Control Panel 8. Silica Gel Key 9. Keyboard Cover

Fig 4-37 Replace the Control Panel (2)

4.3.15 Disassemble the Casting Aluminium Base Cover


1) Disassemble the monitor module, support arm module and control panel module. (Refer
to 4.3.1 , 4.3.6 and 4.3.7 )
2) Take off the 1 transducer gel holder and the 4 probe holders, and then remove the 4
M4X8 screws which are used to secure the casting aluminium base cover from the
frontal surface.

1. Transducer Gel Holder 2. Probe Holder 3. Control Panel Module


4. M4X12 Screw 5. Casting Aluminium Base Cover

Fig 4-38 Disassemble the Casting Aluminium Base Cover

3) Remove the 15 M4X12 screws which are used to secure the casting alumilinium cover
at the back of the cover, and then remove the cover.

4-27
System Structure and Assembly / Disassembly

1. Casting Aluminium Base Cover 2. Casting Aluminium Base 3. M4X12 Screw

Fig 4-39 Disassemble the Casting Aluminium Cover (2)

4.3.16 Disassemble the Up/ Down Handle


1) Disassemble the casting aluminium base cover. (Refer to 4.3.12 )
2) Remove the 1 M4X12 screw which is used to secure the handle, take off the large flat
washer, and then take off the handle.

1. M4X12 Screw 2. Large Flat Washer 3. Up/Down Handle 4. Turning Axis Socket

Fig 4-40 Disassemble the Up/ Handle

4.3.17 Disassemble the Turning Handle Module


1) Disassemble the casting aluminium base cover. (Refer to 4.3.12 Replace the
Trackball)
2) Remove the 2 M4X8 screws securing the fixing block of balancing pole, pull the dragline
to the direction of frontal machine to separate the head of dragline from the locating hole,
then remove the fixing block and pull out the turning handle.

4-28
System Structure and Assembly / Disassembly

1. Screw M4X8 2. Fixing Block of Balancing Pole 3. Dragline 4. Turning Handle

Fig 4-41 Disassemble the Rotating Handle

4.3.18 Disassemble the Turning Block Cover


1) Disassemble the cast aluminium base cover.(Refer to 4.3.12 )
2) Remove the 4 M4X8 screws which are used to secure the cover, after taking off the
turning block cover, the turning block as well as the spring can be dissembled.

1. M4X8 Screw 2. Turning Block Wearing Piece 3. Turning Block 4. Spring 5. Dragline

Fig 4-42 Disassemble the Turning Block Cover

4.3.19 Disassemble the Frontal Cover of the Main Unit


1) Remove the 3 M4×8 screws which are used to secure the base of the main unit rack
below the footswitch board.

2) Pull out the footswitch board module horizontally from the machine.

4-29
System Structure and Assembly / Disassembly

1. Main Unit Rack Base 2. Footswitch Board Module 3. M4X8 Screw

Fig 4-43 Disassemble the Frontal Cover of the Main Unit (1)

3) Pull out the plugs (three at each side) that are filled in the left and right lateral sides,
and then remove the screws (three at each side) which are installed on the two sides of
the rack.
4) Remove the 4 M4X8 screws which are used to secure the rack below the main unit
cover, remove the cover downward slightly and take off the frontal cover when the
cover is separated from the slot at the top side.

1. Plug 2. M4X8 Screw 3. Frontal Shell of Main Unit


4. M4X8 Screw 5. Main Unit Rack

Fig 4-44 Disassemble the Frontal Cover of the Main Unit (2)

4-30
System Structure and Assembly / Disassembly

4.3.20 Disassemble the Rear Cover of the Main Unit


1) Remove the 3 M4X8 screws which are used to secure the rack at the back of the rear
cover.
2) Pull out the 7 plugs filled in the bolt hole at the rear cover of the main unit from the back
side of the machine, and remove the 7 M4X8 screws which are used to secure the rack.

1. M4X8 Screw 2. Plug 3. M4X8 Screw

4. Back Cover of the Main Unit 5. Main Unit Rack

Fig 4-45 Disassemble the Back Cover of the Main Unit

4.3.21 Disassemble the IO Module


1) Disassemble the rear cover of the main unit. (Please refer to 4.3.17 )
2) Remove the 4 M4X8 screws which are used to secure the IO module, then the IO
module can be turned around within 90°outward.
3) To disassemble the IO module, first pull out the connecting cable, then slightly lift the
module and separate it from the rack shaft.

4-31
System Structure and Assembly / Disassembly

1. Connecting Cable 2. I/O Module 3. M4X8 Screw

Fig 4-46 Disassemble the IO Module

4.3.22 Disassemble the Power Supply Input Module


1) Disassemble the rear cover of the main unit. (Please refer to 4.3.17 )
2) Remove the 6 M4X8 screws which are used to secure the module.

1. Main Unit Rack 2. M4X8 Screw 3. Power Supply Input Module

Fig 4-47 Disassemble the Power Supply Input Module (1)

3) Turn the module 90°to insert its two lugs into the rectangular hole at the lower part of
the rack.
4) If the module has to be disassembled, first remove the 5 M4X8 screws securing the
rack base, and then pull out the cables.

4-32
System Structure and Assembly / Disassembly

1. M4X8 Screw 2. Connecting Cable 3. Power Supply Input Module

Fig 4-48 Disassemble the Power Supply Input Module (2)

4.3.23 Disassemble the DVD-R/W Module


1) Disassemble the rear cover of the main unit. (Please refer to 4.3.17 )
2) Pull out the DVD-R/W cable and cable behind the USB-to-IDE port board, remove the 2
M4X8 screws which are used to secure the DVD-R/W module.
3) Pull out the module from the back of the machine, separate the frontal edge fold from
the main unit rack.

1. DVD-R/W Module 2.M4X8 Screw 3. Main Unit Rack


4. Elastic Sheet of the Main Unit Rack 5. Edge fold of DVD-R/W

Fig 4-49 Disassemble the DVD-R/W Module

4) When assemble the module, put the edge fold of the module totally on the elastic sheet
of main unit, and close the bottom guarding edge to the main unit rack, meanwhile, try to
press down the DVD button and make some changes if necessary.

4-33
System Structure and Assembly / Disassembly

1. Elastic Sheet of Main Unit Rack 2. DVD-R/W Module


3. DVD-R/W Guarding Edge 4. DVD-R/W Front Edge Fold

Fig 4-50 Assemble the DVD-R/W Module

4.3.24 Disassemble the DVD and USB-to-IDE Port Board


1) Disassemble the DVD-R/W module. (Please refer to 4.3.20 )
2) Remove the DVD-R/W M3X8 screws at the two sides (two screws at each side), and
take out the DVD-R/W module outwardly. (Note: Properly connect the connector of the
DVD-R/W with the connector on the port board when assemble the module, take care
when insert the module.)

1. DVD Rack 2. M3X8 Screw 3. DVD-R/W


Fig 4-51 Disassemble the DVD-R/W

3) Remove the 2 M3X8 screws that connecting the USB-to-IDE port board with the DVD
rack, take off the port board and DVD crooked-angle part.

1. M3X8 Screw 2. USB-to-IDE Port Board 3. DVD Rack

Fig 4-52 Disassemble the USB-to-IDE Port Board (1)

4-34
System Structure and Assembly / Disassembly

4) Remove the 2 M3X8 screws which are used to secure the DVD patching board
crooked-angle part, disassemble and replace the USB-to-IDE port board.

1. DVD Crooked-angle Part 2. Screw M3X8 3. USB-to-IDE Port Board

Fig 4-53 Disassemble the USB-to-IDE Port Board (2)

4.3.25 Disassemble the Power Supply Adapter


1) Remove the 4 M4X8 screws on the IO module, and turn it within a certain angle ( can be
turned within 90°). (Please refer to 4.3.18 )
2) Remove the 6 M4X8 screws on the power supply input module, and turn it 90°to insert
its two lugs into the rectangular hole at the lower part of the rack.(Please refer to
procedure 1, 2, 3 as described in 4.3.19 )
3) Remove the 2 M4X12 screws which are used to secure the power supply adapter rack,
slightly lift the rack and then take off the rack and power supply adapter, at last, pull out
the connecting cable to complete the disassembly.(Note: Pull out the connecting cable
plug between the adapter and the main unit box.)

1. Main Unit Rack 2. M4X12 Screw 3. Power Supply Adapter


4. Connecting Cable 5. Adapter Rack
Fig 4-54 Disassemble the Power Supply Adapter

4-35
System Structure and Assembly / Disassembly

4.3.26 Disassemble the IO Frontal Board and ECG Board


1) Disassemble the frontal cover of the main unit. (Please refer to 4.3.16 )
2) Pull out the cable connecting to the ECG module, remove the 2 M4 X8 screws securing
the ECG module, and then take out the ECG module.

1. Main Unit Rack 2. M4X8 Screw 3. ECG Module

Fig 4-55 Disassemble the ECG Module

3) Remove the 4 M3 X8 screws on the enclosure of ECG module, take off the internal
cables and disassemble the ECG base module and ECG installation cover module.

1. ECG Base Module 2. M3X8 Screw 3. ECG Installation Cover Module

Fig 4-56 Disassemble the IO Frontal Board and ECG Board (1)

4) Pull out the internal cables, remove the 4 M3X8 screws securing the IO frontal board,
and then take off the I/O frontal Board.

4-36
System Structure and Assembly / Disassembly

1. ECG Frontal cover 2. I/O Frontal Board 3. M3X8 Screw

Fig 4-57 Disassemble the I/O Frontal Board and ECG Board (2)

5) Remove the 4 M3X8 screws securing the ECG board, and then take off the ECG. (When
configured with ECG module.)

1. ECG Base 2. M3X8 Screw 3. ECG Board


Fig 4-58 Disassemble the I/O Frontal Board and ECG Board (3)

4.3.27 Disassemble the Probe Board


1) Disassemble the frontal cover of the main unit. (Please refer to 4.3.16 )
2) Remove the 8 M4x8 screws which are used to secure the probe board module, hold the
2 handles and stably pull out the probe board module.

1. Main Unit Rack 2. M4X8 Screw 3. Probe board Module

Fig 4-59 Disassemble the Probe Board Module (1)

4-37
System Structure and Assembly / Disassembly

3) Remove the 10 M4X8 screws which are used to secure the probe board shielding cover
and then take off the cover.

1. Probe Board Shielding Cover 2. Shielding Plate of Probe Board 3. M4X8 Screw

Fig 4-60 Disassemble the Probe Board (2)

4) Remove the 6 M3X8 screws which are used to secure the probe control board, vertically
pull out the board, and then slightly separate the 3 connectors from the board.

1.M3X8 Screw 2. Probe Control Board

Fig 4-61 Disassemble the Probe Board (3)

5) Remove the 19 M3X8 screws which are used to secure the shielding cover of the probe
board and aluminium shielding frame, then remove the 7 M3X8 screws and 2 M3X7
bolts securing the probe board and the probe board shielding cover, and at last take
away the 4 gaskets, then the probe board shielding cover can be taken off.

4-38
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. Shielding Cover of the Probe Board 3. Probe Board

4. M3X8 Screw 5. M3X7 Stud Screw 6. D3 Gasket

Fig 4-62 Disassemble the Probe Board (4)

6) Remove the M3X8 sunk screws on the aluminium shielding frames and shielding plates
of the transducer socket (6 for each), then remove the 3 shielding frames of the
transducer. (Do not move the pseudo transducer at the leftmost).

1. Aluminium Shielding Plate 2. Aluminium Shielding Frame 3. M3X8 Sunk Screw

Fig 4-63 Disassemble the Probe Board (5)

7) Remove the M3×8 sunk screws securing the probe installing bracket and the aluminium
probe shielding cover (there are 2 sunk screws on each part), then remove the 2
M2.5X8 panhead screws securing the probe port and probe installing bracket, and then

4-39
System Structure and Assembly / Disassembly

take off the 6 installing brackets.

1. Probe Port 2. Probe Installing Bracket

3. M3X8 Sunk Screw 4. M2.5X8 Panhead Screw

Fig 4-64 Disassemble the Probe Board (6)

8) Remove the 12 M3X8 screws securing the transducer board and aluminium shielding
plate, then remove the 4 M3X7+8-6 stud screws and 8 gaskets, and then separate the
shielding plate from the transducer board.

1. Aluminium Shielding Plate 2. Transducer Board 3. M3X8 Screw


4. M3X7 Stud Screw 5. D3 Gasket

Fig 4-65 Disassemble the Probe Board (7)

4.3.28 Replace the Air Spring


1) Trample down the brake of the caster to lock the caster, raise the keyboard and the
monitor to the top.
2) Disassembly the control panel module and frontal cover of the main unit. (Please refer to
4.3.7 and 4.3.16 )

4-40
System Structure and Assembly / Disassembly

3) Pull out the lower cross-pin of the air spring from the guide holder.

1. Main Unit Rack 2. Lower Cross-pin of the Air Spring

Fig 4-66 Replace the Air Spring (1)

4) Draw the spanner of the retainer cross-pin outwardly, turn it 90°clockwise, and the
spanner will be locked after releasing. (There are lock and unlock symbols marked on
the retainer).
(Note: Remove the ECG module first if it is an obstacle for the operation of the spanner).

1. Spanner of the Retainer Cross-pin

Fig 4-67 Replace the Air Spring (2)

5) Hold the keyboard handle by both hands, press down the up/down handle on the
keyboard handle and raise the keyboard. When you hearing “Cluck”, it means the
cross-pin is inserted into the hole of the back-up block, the keyboard is vertically locked
with the main unit.
6) Remove the 6 M5X16 inner hexagon screws, the flat washer, and elastic washer by the
inner hexagon spanner, and then take out the air spring module from the machine.

4-41
System Structure and Assembly / Disassembly

1. M5X16 Inner Hexagon Screw 2. Elastic Washer 3. Flat Washer 4. Air Spring Module

Fig 4-68 Replace the Air Spring (3)

7) Disassemble the retaining ring from the upper cross-pin of air spring by the cross
screwdriver, and then take off the rotary gear.

1. Retaining ring 2. Rotary Gear 3. Upper Cross-pin 4. Air Spring

Fig 4-69 Replace the Air Spring (4)

8) Screw out the upper connecting link and lower connecting link of the air spring, and then
finish the disassembly of the air spring.

4-42
System Structure and Assembly / Disassembly

1. Upper Connecting Link 2. Air Spring 3. Lower Connecting Link

Fig 4-70 Replace the Air Spring (5)

4.3.29 Disassemble the Isolating Transformer


1) Disassemble the probe board module. (Please refer to procedure 1 and 2 as described
in section 4.3.24 ).
2) Remove the 4 M4X8 screws securing the IO module, turn the module a certain angle.
(within 90°) (Please refer to procedure 1 and 2 as described in section 4.3.18 ).
3) Remove the 6 M4X8 screws securing the power supply input module, turn the module
90°, insert the two supporting lugs into the rectangular hole. (Please refer to procedure
1, 2, 3 as described in section 4.3.19 ).
4) Remove the M4X8 screws securing the grounding wire of the transformer, disassemble
the flat washer, elastic washer and the 4 M5X10 inner hexagon screws securing the
isolating transformer, then take off the transformer.

4-43
System Structure and Assembly / Disassembly

1. M5X10 Inner Hexagon Screw 2. Elastic Washer 3. Flat Washer 4. Isolating Transformer

Fig 4-71 Disassemble the Isolating Transformer

4.3.30 Replace the HDD


1) Disassemble the IO module. (Please refer to section 4.3.18 ).
2) Remove the 2 M4X8 screws securing the HDD bracket.

1. HD Module 2. M4X8 Screw

Fig 4-72 Replace the HDD (1)

4-44
System Structure and Assembly / Disassembly

3) Take out the HDD bracket and pull out the HDD plug.

1. HDD 2. HDD Plug 3. HDD Bracket

Fig 4-73 Replace the HDD (2)

4) Take off the HDD and its bracket, remove the 4 M3X4 panhead screws securing the
HDD, and then finish the replacement.

1. HDD 2. M3X4 Screw 3. HDD Bracket

Fig 4-74 Replace the HDD (3)

4.3.31 Disassemble the Main Unit Box Module


1) Disassemble the ECG and IO modules. (Please refer to section 4.3.18 and procedure
1, 2 as described in section 4.3.23 ).
2) Remove the 6 M4X8 screws securing the power supply input module, turn the module
90°, insert the two supporting lugs into the rectangular hole. (Please refer to procedure
1, 2, 3 as described in section 4.3.19 ).
3) Remove the 4 M4X8 screws securing the main unit box module from the front part of the
machine.

4-45
System Structure and Assembly / Disassembly

1. Main Unit Box Module 2. M4X8 Screw


Fig 4-75 Disassemble the Main Unit Box (1)

4) Pull out the cable plugs on the IO port board, take off the IO module, and remove the 4
M4X8 screws securing the main unit box module from the back part of the machine.

1. Main Unit Box Module 2. M4X8 Screw (used to fix the cable)
3. M4X8 Screw (To fix the main unit box)

Fig 4-76 Disassemble the Main Unit Box (2)

5) Hold the handle of the main unit box, slowly pull out the main unit box module, pull out
the plugs of network cable and power supply adapter (when the plugs can be seen on

4-46
System Structure and Assembly / Disassembly

the top part of the main unit box), and then finish the disassembly.

1. Main Unit Box Module 2. Handle of the Main Unit Box


3. Plug of the Network Cable 4. Plug of the Power Supply

Fig 4-77 Disassemble the Main Unit Box (3)

6) When assemble the main unit box module, make sure to align the guide-pin socket with
the guide-pin, and align the plug of probe connecting board with the socket on the probe
board.

1. Transducer Board Module 2. Guide Pin 3. Guide-pin Socket 4. Main Unit Box module

Fig 4-78 Disassemble the Main Unit Box Module

4.3.32 Replace the IO Port Board


1) Remove the 4 M4X8 screws securing the IO module, and turn the module a certain
angle (within 90°).(Please refer to procedure 1 and 2 as described in section 4.3.18 ).

4-47
System Structure and Assembly / Disassembly

2) Pull out all the connecting cables on the IO port board, and remove the 4 M3X8 screws
securing the IO port board.
If it is not convenient to remove the screws above, you can first remove the 6 M4X8 screws
securing the power supply input module, turn the module 90°, insert the two supporting
lugs into the rectangular hole. (Please refer to procedure 1, 2, 3 as described in section
4.3.19 ).

1. IO Port Board 2. M3X8 Screw

Fig 4-79 Replace the IO Port Board (1)

3) Pull the two spanners to separate the socket of IO port board from the socket of IO
connecting board, and then take out the IO port board upward.

1. Spanner 2. IO Port Board Gap 3. Limit Bridge

Fig 4-80 Replace the IO Port Board (2)

4) Remove the 4 M3X8 screws securing the IO port board shielding cover and then finish
the replacement.

4-48
System Structure and Assembly / Disassembly

1. Shielding Cover 2. I/O Port Board 3. M3X8. Screw

Fig 4-81 Replace the IO Port Board (3)

4.3.33 Replace the IO Connecting Board


1) Disassemble the main unit box module. (Please refer to section 4.3.28 ).
2) Disassemble the IO port board. (Please refer to procedure 2, 3 as described in section
4.3.29 ).
3) Remove the 11 M3X8 screws securing the main unit box cover, and then take off the
cover.

4-49
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. Main Unit Box Cover 3. IO Connecting Board


4. Cables between the Master Board and IO Connecting Board

Fig 4-82 Replace the IO Connecting Board (1)

4) Remove the 6 M3X8 screws securing the IO connecting board, pull out the IO
connecting board outward, and draw out the cable plugs on the IO connecting board,
then finish the replacement.

1. IO Connecting Board 2. M3X8 Screw 3. Cable Plug

Fig 4-83 Replace the IO Connecting Board (1)

4-50
System Structure and Assembly / Disassembly

4.3.34 Disassemble the Transducer Connecting Board


1) Disassemble the main unit box. (Please refer to section 4.3.28 ).
2) Remove the 6 M3X8 screws securing the probe connecting board shielding cover, and
then take off the cover.

1. Probe Connecting Board 2. Shielding Cover 3. M3X8 Screw

Fig 4-84 Disassemble the Probe Connecting Board (1)

3) Remove the 3 M3X8 screws, pull the 2 spanners to separate the board from the sockets
and then pull out the probe connecting board.

1. M3X8 Screw 2. Probe Connecting Board 3. Spanner

Fig 4-85 Disassemble the Probe Connecting Board (2)

4) Remove the 4 M3X8 screws securing the guide-pin socket, and then replace the
connecting board.

4-51
System Structure and Assembly / Disassembly

1. Guide-pin Socket 2. Probe Connecting Board 3. M3X8 Screw

Fig 4-86 Disassemble the Probe Connecting Board (3)

4.3.35 Replace the Power Supply Main Board and Auxiliary

Board
1) Remove the main unit box cover. (Please refer to the procedure 1, 2, 3 as described in
section 4.3.30 ).
2) Remove the 4 M3X6 panhead screws securing the power supply module, draw out the
cables on the power supply module and take out the module.

1. M3X8 Screw 2. Power Supply Module 3. Main Unit Box

Fig 4-87 Replace the Main Board and Auxiliary Board of Power Supply (1)

3) Remove the 2 M3X6 panhead screws securing the power supply box, take off the cover;
remove the 4 M3X6 screws securing the power supply main board; take out the main
board, the power supply box clapboard and power supply auxiliary board assembly.

4-52
System Structure and Assembly / Disassembly

1. Upper Cover of Power Supply Box 2. M3X6 Screw

3. M3X6 Screw 4 Power Supply Main Board

Fig 4-88 Replace the Power Supply of Main Board and Auxiliary Board (2)

4) Raise the power supply main board to separate it from the connecting socket of the
auxiliary board, and then take off the power supply main board.

1. Power Supply Main Board 3. Clapboard 3. Connecting Socket

Fig 4-89 Replace the Main Board and Auxiliary Board of Power Supply (3)

5) Remove the 4 M3X6 panhead screws securing the power supply auxiliary board, and
then take off the auxiliary board.

4-53
System Structure and Assembly / Disassembly

1. M3X6 Screw 2. Power Supply Main Board 3. Clapboard

Fig 4-90 Replace the Main Board and Auxiliary Board of Power Supply (4)

4.3.36 Replace the Transmitting Board


1) Disassemble the cover of main unit box. (Please refer to procedure 1, 2, 3 as described
in section 4.3.30 ).
2) Remove the 3 M2.5X8 panhead screws securing the transmitting shielding cover and
lower shielding cover of master board, then remove the 1 M3X8 screw securing the
transmitting shielding cover and main unit box, take off the 2 board pressing blocks,
remove the 2 M3X6 screws securing the transmitting board. Pull out the board upward,
and then remove the board. (If the double glue on the block became effectless, please
replace the block or the glue.)

4-54
System Structure and Assembly / Disassembly

1. M2.5X8 Screw 2. Shielding Cover of the Transmitting Board


3. M3X6 Screw 4. Transmitting Board

Fig 4-91 Replace the Transmitting Board

4.3.37 Replace the CW Board


1) Disassemble the cover of main unit box. (Please refer to procedure 1, 2, 3 as described
in section 4.3.30 ).
2) Remove the 4 M3X8 screws securing the shielding cover of CW board and the 2 screws
securing the CW board, then replace the board.

4-55
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. Shielding Cover of CW Board 3. M3X6 Screw 4. CW Board

Fig 4-92 Replace the CW Board

4.3.38 Replace the CPU Board


1) Disassemble the cover of main unit box. (Please refer to procedure 1, 2, 3 as described
in section 4.3.30 ).
2) Pull out the plug of CPU fan, remove the 4 M3X35 panhead screws securing the fan,
take off the fan as well as the connecting cable; then remove the 4 screws securing the
CPU module, and then replace the module.

4-56
System Structure and Assembly / Disassembly

1. M3X35 Screw 2. Fan and Connecting Cable 3. M3X8 Screw 4. CPU Module

Fig 4-93 Replace the CPU Board (1)

3) Remove the 2 M3X8 screws and 2 M2.5X16 screws securing the CPU radiator and
shielding cover, take off the cover; remove the 2 M2.5X8 screws securing the CPU on
the radiator, and then replace the CPU board. (Heat outlet washer has to be replaced.)
Pay attention to the direction of the fan during the assembly, and take care of the memory
bank on the board when replace the CPU.

1. M2.5X16 Screw 2. M3X8 Screw 3. CPU Shielding Cover


4. M2.5X8 Screw 5. CPU Board 6. CPU Radiator 7. CPU Heat Outlet
Washer
Fig 4-94 Replace the CPU Board (2)

4-57
System Structure and Assembly / Disassembly

4.3.38.1 Disassemble the Memory Bank


Loosen the two clips outward, the memory will automatically spring up, then replace the
memory.

1. Memory Bank

Fig 4-95 Disassemble the Memory Bank

4.3.39 Replace the Master Board


1) Disassemble the main unit box cover, the IO connecting board, probe connecting board,
power supply board, transmitting board, CW board and CPU module. (Please refer to
the procedure 1, 2, 3 as described in section 4.3.30 , procedure 2, 3 in section 4.3.31 ,
procedure 2 in section 4.3.32 , procedure 2 in section 4.3.33 , procedure 2 in section
4.3.34 and procedure 2 in section 4.3.35 ).
2) Remove the 3 M3X8 screws securing the master board, pull out the cable plug of the fan
and HDD, then finish the replacement of the master board.

4-58
System Structure and Assembly / Disassembly

1. M3X8 Screw 2. Master Board

Fig 4-96 Replace the Master Board

4.3.40 Replace the Air Inlet Fan


1) Remove the main unit box cover. (Please refer to procedure 1, 2, 3 as described in
section 4.3.30 ).
2) Pull out the cable plug of the fan, remove the 4 M3X8 screws securing the fan module,
turn the fan slightly inward and then take out the fan module.

1. M3X8 Screw 2. Fan Module 3. Plug

Fig 4-97 Replace the Air inlet Fan (1)

4-59
System Structure and Assembly / Disassembly

3) Cut off the tie securing the magnetic ring, remove the 8 screws securing the fan covers,
take off the fan after removing the cover.

1. Screw 2. Fan Cover 3. Fan Bracket 4. Magnetic Ring 5. Fan and the Cables

Fig 4-98 Replace the Air inlet Fan (2)

4.3.41 Disassemble the Dust Net


1) Grab the lower side of the dust net frame, then pull out the net.

1. Main Unit 2. Dust Net

Fig 4-99 Disassemble the Dust Net

4.3.42 Replace the Caster


1) Lock the casters that needn’t to be replaced. (Front or rear)
2) Place a jack (with maximal load: over 100kg) in the mounting area of casters, and lift the
casters at the same time, thus the system will not topple to one side.

3) Place a piece of wood at the middle of front/ rear bottom. (150 mm ´ 150 mm ´ 220 mm,
W´L´H).

4-60
System Structure and Assembly / Disassembly

Fig 4-100 Replace the Caster (1)

4) Remove the 4 M8X16 screws securing the caster by the inner hexagon spanner, take
out the flat washers, and then disassemble the caster for replacement.

1. Caster 2. Flat Washer 3. M8X16 Inner Hexagon Screw

Fig 4-101 Replace the Caster (2)

4.4 Installation of Peripherals

4.4.1 Installation of W/B Video Printer


1) Remove the cover of the printer placing compartment, connect the power cable, control
cable and signal cable that reserved in the compartment to the printer.

4-61
System Structure and Assembly / Disassembly

1. Printer Placing Compartment 2. W/B Printer 3. Cover

Fig 4-102 Installation of W/B Printer (1)

2) Place the printer into the compartment and grab it with the frontal cover of the main unit.
(To avoid slippage)

Fig 4-103 Installation of W/B Printer (2)

3) Take away the cover of the compartment.

4-62
Maintenance Requirements

5 Maintenance Requirements

5.1 Tools Used in Maintenance

5.1.1 Tools, Measurement Devices, Consumables


Tools and Measurement Devices
Tool/Device Qty. Description
Resin or stainless steel container 1 Can hold two transducers
Plastic bag 1 About 2m´2m (0.3mm
thickness)
Vacuum dust-collector 1
Air blower (600W ~16000 mm/min) 1 Makita 4014NV or like devices
Cleaner
Brush 1 About a toothbrush size
Air blowout device and blower brush 1 For fan cleaning

Consumables
Consumable Qty. Description
Aluminum foil 1
Physiological saline (0.85 - 0.95%) To fill the whole container.
Glass cleaner 1 To clean the display
Neutral cleanser
Soft cloth

NOTE: 1 Do not use soluble solvent (such as paint thinner or gasoline) or


abrasive cleanser to clean the system. Otherwise, the system may
be damaged.
2 Do not use hydrocarbon detergent or detergent used on OA
equipments to clean the screen. Otherwise, the display
performance may degrade.

5-1
Maintenance Requirements

5.1.2 Standard Configuration of Tools and Consumables


Tools
Tool Model Manufacturer Specification/Standard
Screwdrivers (all models and sizes) Not specified
Blunt cross screwdriver (big) Not specified
Wire cutting pliers Not specified
Sharp nose pliers Not specified
Diagonal cutting pliers Not specified
Electric iron and its holder Not specified
De-soldering gun Not specified
Wrench suite Not specified
Adjustable wrench Not specified
Tweezers Not specified
Hammer Not specified
Metric general wrench suite Not specified
Cable (all models) Not specified
Knife Not specified

Consumables

Consumable Model Manufacturer Specification/Standard


Insulating tape Not specified
Hishilite tube (heat shrink tube), Not specified
insulating tube (all models)
Cable (all models) Not specified
Screws, nuts, gaskets (all models) Not specified
Solder stick
File folders (all models)
Alcohol for cleaning and disinfection

5.2 Maintenance Personnel


To ensure the system performance and safety, only Mindray engineers or engineers
authorized by Mindray can perform maintenance.

5-2
Checking

6 Checking

6.1 System Status Checking

6.1.1 Checking Plan


Check plan within one year after installation:

Check times
Model 6 months after 12 months after Remarks
installation installation

DC-3 series - 1

6.1.2 Confirmation before Checking

6.1.2.1 Checking System Status


Check the following items or the log together with the customer to confirm the system status.
1) Any abnormality when the system is running;
2) Occasional abnormality;
3) Other items the customer requires to check.

6.1.2.2 Checking Operation Environment


Check the ambient temperature and humidity. The measurements related to safety features
are sensitive to humidity, and measurement position and circuit as well.

If the insulation feature of the system degrades as the service time increases or due to the
system malfunctions, the deflection range of measurement result may increase as the
humidity increases.

6.1.2.3 Checking System Setup


Check the system’s preset, contrast and brightness of display, optional parts, transducers,
etc.

6.1.2.4 Checking the System Version


Press [Setup] key on the control panel to enter the menu of Setup, select [About] to enter the
system information page for system version checking.

6-1
Checking

6.1.2.5 Checking System Running Status


Check if all menus and dialog boxes can pop up normally.

Check if measurements can be performed in each exam mode.

Check if exam modes can be switched normally.

Check if images modes can be switched normally.

Check if transducers can be switched normally.

6.2 Checking Functionality

6.2.1 Flow of Functionality Checking


Time and date displaying

½
¯
Check initial operations

½
¯
Check general operations

½
¯
Check keys on the control panel

½
¯
Check the trackball

½
¯
Check the transducers

½
¯
Check the fans

6.2.2 Details of Functionality Check


1) Time and date displaying

l Confirm the time and date displayed are correct.

l If they are incorrect, correct them.

6-2
Checking

2) Checking the initial operations

l Check the functions of keys under each mode.

l Check the images produced by each transducer and print them.

l Check the functions of peripherals.

3) Checking general operations

l Check if the [Freeze] key can work normally.

l Check if the total gain can be adjusted properly.

l Check if the CW, Color, PW, M and B modes can be switched.

4) Checking keys on the control panel

l Check all keys on the control panel from left to right and from up to down to see if
they can work properly.

l If a video printer is connected to the system, check if the [Print] key can work
normally.

5) Checking the trackball

l Press the [Freeze] key to enter the image freeze status. Press the [Measure] key
to enter the measurement status.

l Measure the vertical distance and horizontal distance to see if the trackball can
work normally.

6) Checking the peripherals

l Check if the CD-ROM-drive can jet normally.

l Check if the breaker can work normally.

l Check if the transducers can be pulled out properly. Check if the transducers can
work normally after been connected to the main unit.

7) Checking the fans

l Check if the system fan (fan in the main unit) can work normally after the main
unit is powered on.

l Check if the fans make any abnormal noise when they are working.

6-3
Checking

6.3 Safety Checking

6.3.1 Electrical Safety Checking


When performing the following checks, be sure to use the instruments specified by the
manufacturers. For example, the instruments applied for electrical safety checking are the
601pro series instruments of BIO-TEK.

· Flows of checking

1.Measure the impedance of protection earth conductor


½
¯
2. Measure earth leakage current
½
¯
3. Measure enclosure leakage current
½
¯
4. Measure patient leakage current
½
¯
5. Measure patient leakage current (110 % mains
voltage on the applied part)
½
¯
6. Measure patient auxiliary current

· Performing steps:

CAUTION: 1 Before the test, read the manual of 601pro instrument


carefully. During the test, be sure to use the accessories of
601pro instrument.
2 Before the test, follow the steps below to set the test
standard, number of leads, type and degree of protection
against electric shock.
3 If “Inv” appears in “[ ]” on the screen, it means the test is
invalid. At this time, please check if the settings are correct.

To select the standard: MAIN MENUàSYSTEM SETUPàTEST STANDARDàIEC601-1;


To set the type and degree of protection against electric shock:

6-4
Checking

MAIN MENUàCLASS/TYPEàCALSS I/TYPE BF;


To set DC enable: MAIN MENUàSYSTEM SETUPàDC;
To set number of leads: Press [View Present Settings] key à EDIT LEADS à press the
numeric key [5].
1) To correct the conductor impedance: Press the shortcut key 3 on the instrument and then
select “25A” current. Perform the “Correcting leads” procedures described in the
operation manual of 601pro.

2) To measure earth leakage current: Press the shortcut key 4 on the instrument. In each
status specified in the Table I below, press the soft key 4 to select AP-Earth and No
AP-Earth to measure the earth leakage current.

Table I Test Status of Earth Leakage Current

Status DUT OUTLET


Polarity Earth L2
Normal Nor Pol No Earth L2
Normal Rev Pol No Earth L2
Single failure Nor Pol No Earth No L2
Single failure Rev Pol No Earth No L2
3) To measure enclosure leakage current: Press the shortcut key 5 on the instrument. In
each status specified in the Table II below, press the soft key 4 to select AP-Earth and No
AP-Earth to measure the enclosure leakage current.

NOTE: There must be physiological saline between the transducer and


the aluminum foil.

Table II Testing Status of Enclosure/Patient Leakage Current

Status DUT OUTLET


Polarity Earth L2
Normal Nor Pol Earth L2
Normal Rev Pol Earth L2
Single failure Nor Pol No Earth L2
Single failure Rev Pol No Earth L2
Single failure Nor Pol Earth No L2
Single failure Rev Pol Earth No L2

4) To measure patient leakage current: Press the shortcut key 6 on the instrument.
In each status specified in the Table II above, press the soft key 4 to select

6-5
Checking

RA-Earth (RA is connected to the transducer) and V1-V6-Earth (V1/V2/V3 are


connected to ECG electrode ) to measure the patient leakage current.

5) To measure patient leakage current (110%mains voltage on the applied part):


Press the shortcut key 7 on the instrument. In each status specified in the Table III
below, press the soft key 4 to select RA-Earth (RA is connected to the transducer)
and V1-V6-Earth (V1/V2/V3 are connected to ECG electrode ) to measure the
patient leakage current.

NOTE: 1 Be careful when perform the test. Voltage on the applied part
should be 10% higher than the mains voltage.
2 There must be physiological saline between the transducer and the
aluminum foil.

Table III Testing Status of Patient Leakage Current (110% mains voltage on the applied
part)

Status DUT OUTLET


Polarity Earth L2
Single failure Nor Pol Earth L2
Single failure Rev Pol Earth L2

NOTE: In these conditions, polarity of the applied part is changed by the


instrument automatically.

6) To measure patient auxiliary current: Press the shortcut key 8 on the instrument.
In each status specified in the Table II above, press the soft key 4 to select
RA-ALL (to measure the patient auxiliary current).

6-6
Checking

Type Check item Procedure Standard


limited value

Test when the system is not working:

(1) Do not connect the device to be tested


(described as DUT in the following
contents) to the auxiliary power socket on
the front board of testing instrument.
Connect the red cable (accessory of Less than or
601pro) to the red input end on the front equal to 0.1
board of 601pro and connect the black W.
cable to the green input end. Then connect
the other end of red cable to the screw of
Impedance of DUT keyboard holder, and connect the
protection black cable to the equipotential terminal
earth (A2-A1).
conductor
(2) Connect DUT to the auxiliary power
socket on the front panel of 601pro,
and connect the red cable to the screw
Electrical of probe bracket of DUT. Less than or
Safety equal to 0.2
(3) The measure position is from the screw
W.
of DUT keyboard holder to the earth pin
(A2-A3) of the DUT 3-way power plug.

See Fig A.

Earth leakage Test when the system is working: Normal


current Connect the power plug of DUT to auxiliary status:
power socket of 601pro (B1). Less than or
equal to 500
Pack the transducer with metal foil and then uA.
put the transducer into the container filled Single-failure
with physiological saline. Connect the metal status:
foil to the RA input end of 601pro, and Less than or
connect ECG lead to the other input equal to
terminals (V1/V2/V3) (B2, B3, B4, B5). 1000 uA.

Please see Fig B.

6-7
Checking

Type Check Procedure Standard Limited


item Value

Test when the system is working:

(1) Connect the screw of keyboard holder to the


RED input terminal (A2) of the instrument by
the red cable, and connect the DUT power
input plug to the auxiliary power socket (B1)
Normal status:
of 601 PRO.
Less than or
equal to 100 uA
(2) Pack the transducer with metal foil and then
Enclosure put the transducer into the container filled
Single-failure
leakage with physiological saline. Connect the metal
status:
current foil to the RA input end of 601pro, and
connect ECG lead to the other input Less than or
terminals. (V1/V2/V3) (B2, B3, B4, B5). equal to 500 uA

(3) Test the enclosure leakage current according


to the requirements stipulated in procedure
3.

Electri Please see Fig B.


-city
Safety Test when the system is working: Normal status:

(1) Pack the transducer with metal foil and then AC less than or
put the transducer into the container filled equal to 100 uA;
with physiological saline. Connect the DC less than or
aluminium foil to the RA input terminal (B2) equal to 10 uA.
of the instrument with the cable provided by
the instrument, and also plug the 3-way Single-failure
power plug of DUT into the auxiliary power status:
Patient socket (B1) of the instrument.
leakage AC less than or
current (2) Connect ECG electrode to (V1/V2/V3) (B3, equal to 500 uA;
B4, B5) input terminals of the instrument by DC less than or
the input clip. equal to 50 uA.

(3) Test the patient leakage current according to


the requirements stipulated in procedure 4.

Please see Fig B.

6-8
Checking

Type Check Procedure Standard Limited


item Value

Patient Test when the system is working:


leakage
(1) Pack the transducer with metal foil and then
current
put the transducer into the container filled
(110 %
with physiological saline. Connect the
mains aluminium foil to the RA input terminal (B2) of
voltage on the instrument with the cable provided by the
the applied instrument, and also plug power input plug of
part) DUT into the auxiliary power socket (B1) of
Single-failure
601 PRO.
status:
Less than or
(2) Connect ECG electrode to (V1/V2/V3) (B3, equal to 5000 uA.
Electri B4, B5) input terminals of the instrument by
-city the input clip.
Safety

(3) Do the test according to procedure 5 after


the instrument is connected.

Please see Fig B.

6-9
Checking

Type Check Procedure Standard Limited


item Value

Test when the system is working:


Patient
auxiliary (1) Pack the transducer with metal foil and then
current put the transducer into the container filled
with physiological saline. Connect the
aluminium foil to the RA input terminal (B2) of
the instrument with the cable provided by the
instrument.
Normal status:
(2) Connect ECG electrode to (V1/V2/V3) (B3, AC less than or
B4, B5) input terminals of the instrument by equal to 100 uA;
the input clip. DC less than or
equal to 10 uA.
(3) Plug the power input plug of DUT into the
auxiliary power socket (B1) of the instrument. Single-failure
status:
(4) Do the test according to procedure 6.
AC less than or
(5) Connect one ECG electrode to the RA input equal to 500 uA;
terminal of the instrument, and connect the DC less than or
other two to V1/V2, and the aluminium foil to equal to 50 uA.
V3. Plug the 3-way power plug of DUT into
the auxiliary power socket of instrument.
Make sure to connect each ECG electrode to
RA by turns, and connect the other applied
parts according to the above method.

Please see Fig B.

6-10
Checking

6-11
Checking

6.3.2 Mechanical Safety Checking


Evaluation
Perform the evaluation by eye-measuring and checks.

· Checking flow

Check the caster

½
¯
Check the connection of the caster

½
¯
Check the handle

½
¯
Check the rotating part and strutting piece of
keyboard

½
¯
Check the structure of up/ down system

½
¯
Check the monitor and structure of support arm

½
¯
Check the status of the peripheral instruments
½
¯
Check the cable connection

½
¯
Check the transducer appearance

½
¯
Check other mechanical structures

If the check result is “fail”, the system is in abnormal status. Stop the system and adopt proper

6-12
Checking

measures.

Type Check Procedure Tool

Mechanical 1) Check by sight if the caster has any crack.


safety
Caster
2) Operate the caster, make sure the caster can
be locked and loosened normally.

1) Check by sight if the caster is skew, and make


sure no damage to the screw and no screw is
missing.
Connection
of the caster
2) Check with the spanner to make sure that there Inner
is no looseness between the caster and the base hexagon
connection screw. spanner

1) Check by sight if the handle has any crack.


Handle
2) Pull the handle to make sure it is not loose.

Rotating 1) Pull the keyboard to make sure it is not loose.


part and
strutting 2) Turn the keyboard within 45°both left and right
piece of to see if it can move smoothly and flexibly, without
keyboard any abnormity such as giving out friction noise.

3) Rotate the main body by moving the handle,


leave the handle at the left end, right end and at
the middle position, and check if the upper part of
the main body can stay still without any turning.

1) Posit the main body to the lowest position and


then loosen the up/ down handle, then raise the
cast aluminium base to check if the upper main
body can keep still.

2) Posit the main body to the highest position and


then loosen the handle, then press down the cast
Structure of
aluminium base cover to check if the upper main
the up/down
body can keep still.
system

3) When the main body is at the highest position,


hold still the up/down handle, the body shown not
fall down

4) The machine can be smoothly moved without


any abnormity such as giving out friction noise.

6-13
Checking

1) Check by sight if any inclination happened to


the monitor

2) Manually operate the monitor to make sure the


monitor can act normally when turned left/ right,
lifted/ lowered, and no abnormal noise exists.

Structure of 3) Manually turn the monitor left/ right, make sure


the monitor there is no obvious looseness.
and support
arm 4) Lower the monitor to posit the support arm at
level and slope-down, check if the monitor can
keep still at the two positions when no outside
force is imposed on it.

5) Take apart the upper cover of the support arm,


check by sight if any cable is scratched or clipped
out that the core can be seen.

Status of the Check manually if any instrument is loose.


peripheral
instruments

1) Check by sight to make sure that there is no Phillips


damage happened to the cable both inside and screwdriver
outside the instrument.
Cable
connection
2) Check manually to make sure that there is no
looseness and falling off happened to the cable
inside the instrument

Probe Check by sight to make sure that there is no


appearance crack, peel, loose and damage happened to the
transducer

Other Check to make sure that there is no part is crack,


mechanical and no conducting part is exposing to the outside
structures

6-14
Checking

6.4 Image Checking

6.4.1 Checking the B/W Image Phantom Data and Image

Recording

6.4.1.1 System Setups


The user-defined setups are adopted for all the setups which aren’t mentioned in this manual.
For any setups changed due to special reason, they shall be recorded as the additional
information.

6.4.1.2 Image Recording and Archiving


Print the images and archive them with data recorded.

6.4.1.3 Checking Flow


Check the lateral and axial resolution
¯
Check the penetration
¯
Check the spot features
¯
Record and check the images
Perform the checks above on all transducers used by customers.

6.4.2 Checking Phantom Data

6.4.2.1 Lateral / Axial Resolution


1) Put some gel on the phantom and then scan the phantom with a transducer.

2) After obtaining an optimal image, freeze the image and record it.
Condition: system preset parameters.

6.4.2.2 Penetration
1) Put some gel on the phantom and then scan the phantom with a transducer.

2) Adjust the gain to make the soft tissue spot displayed at the deepest position.

3) Measure the depths of noise and of the soft tissue boundary, and record the images for
the measurement.

6.4.2.3 Spot Features


Evaluate the change of image quality after the system is used for a long time. The evaluation
items include the gain, and the periodic record of the images described above.

6-15
Checking

6.4.2.4 Recording images


Archive the images printed out according to the descriptions above.

6.4.3 Final Operation Checking and Image Archiving

6.4.3.1 Operation Checking


Check image operations with each transducer in each mode.

6.4.3.2 Image Printing and Archiving


Print images and archive them with data recorded.

6.4.3.3 Checking Flow


Functionality checking

½
¯
Image operations checking

½
¯
Image archiving

Perform all the checks above on each transducer.

6.4.3.4 Checking Details


1) Functionality check
Check changes of the exam mode, measurements and related information.

2) Image operation check


Perform image operations with each transducer in each mode.

3) Image archiving
Print images obtained through the steps above and archive them.

6.4.4 Color flow image check


The service engineers should be trained by Mindray’s professionals, and the method is to use
a transducer to examine and capture a color carotid flow image. The color image should be
evaluated depending on whether the flow is full, whether there is an artifact and overflow, and
whether the flow is flat.

6-16
System Maintenance

7 System Maintenance

7.1 System Cleaning

7.1.1 Cleaning Flow


Disconnect the power cable from the power
socket

½
¯
Disconnect the accessories

½
¯
Disassemble the enclosure of main unit

½
¯
Disassemble the box of main unit

½
¯
Clean the interior of main unit and the fans

½
¯
Clean the covers and control panel

½
¯
Clean the screen

½
¯
Clean the peripherals and accessories

½
¯
Clean the cable and the power cord

7-1
System Maintenance

7.1.2 Details for Cleaning


1) Clean the interior of main unit and fans

Disconnect the power cord from the power socket.

Disassemble accessories such as transducers and printers.

Disassemble the enclosure of main unit, and the relative shielding covers to check the
PCBA and power supply part. If the PCBA have no abnormality, do not disassemble
them.

Cover the main unit with a big plastic bag.

Put the mouth of the vacuum dust-collector into the plastic bag and turn on the
dust-collector to collect the dust.

Open a small hole on the top of the plastic bag to insert the mouth of the air blower. Then
turn on the blower to remove the dust.

Take away the plastic bag carefully, avoiding dust rising.

Use the dust collector to remove all remaining dust.

2) Clean the enclosure and control panel

Use neutral cleanser to remove the dust on the enclosure of main unit and on the control
panel. If it is difficult to clean the control panel, disassemble the keys first and then use
neutral cleanser to clean it.

3) Clean the display

Make sure the display is not skew and the fixing mechanism is secured.

4) Transducers
Remove the dust on the transducer. Check if it has any abnormality such as crack. Then
connect it to the main unit. Use a soft brush to clean the transducer holder.

5) Video printer

Clean the thermal head according to the user manual of the printer.

6) Cable / cord
Use neutral cleanser to clean the cable and power cord.

7) Cleaning of the trackball


a)Disassemble
Press the bulges on the clamping ring by both the hands and turn the ring about 45°
clockwise until it lifts. Take out the ring and the rotary ball. Be careful not to drop the
ball. Shown as follows:

7-2
System Maintenance

Rotary ball

Clamping ring Top cover

b)Cleaning
Clean the two long shafts, the bearing and the rotary ball with soft dry cloth or paper.

Bearing Long shaft

Long shaft

c)Installing
Put the rotary ball back in the trackball and then align the clamping ring click with the
top cover notch. Press the bulges on the ring with both hands and turn the ring about
45°counterclockwise until the ring clicks. As the bulges are flush with the top cover,
the ring is secured.

7-3
System Maintenance

Clamping ring click

Top cover
notch

7.2 Software Maintenance

NOTE: Do not turn off the power supply in the process of upgrading,
otherwise, it will lead to upgrading failure and can’t be restarted.

7.2.1 Confirming the System Starting


There are three courses from starting the system to enter the ultrasound interface, they are
BIOS guiding, WINDOWS guiding and ultrasound software guiding.
BIOS guiding: from the appearing of boot-strap interface 1 to the disappearing of the blue
progress bar that located at the bottom of the interface 1.
WINDOWS guiding: from the company logo to appear the boot-strap interface 2, till the blue
progress bar that located at the bottom of the interface 2 disappears.
Ultrasound software guiding: from the appearing of boot-strap interface 3 to the disappearing
of the blue progress bar that located at the bottom of the interface 3. And again, this part can
be divided into the following parts:
Step Phenomenon Task
1 Appear To initialize the system clock and soft interruption
boot-strap module.
interface 3, but
no progress bar
2 Appear the To initialize the preset server, need to preset the data.
progress bar, but
not any progress
3 1/7 progress bar To set the region, language and font.

7-4
System Maintenance

4 2/7 progress bar Can’t be seen.


5 3/7 progress bar To create the operation interface (can’t be seen), to
initialize the icons and the function bank.
6 4/7 progress bar To initialize the keyboard.
7 5/7 progress bar To initialize the ultrasound system, USB manager, video
printer, VCR, CD-RW, file manager, iVision player and
so on.
8 6/7 progress bar To initialize the application software, including the
measurement, patient management, report printing,
system monitor, running log, fore/ end image
parameters, POD data is needed.
9 7/7 progress bar End, switch to the ultrasound system operation
interface.

7.2.2 Preset
Press [Setup] key, pops up the preset menu.

7.2.3 Viewing System Information


In the Setup menu, select “System”, and the system information dialog box appears.

7-5
System Maintenance

NOTE Be sure to confirm the system information before and after the software
maintenance.

7.2.4 Preset Data Management


1)Press [Setup] key to enter into the preset status.
2)Move the cursor onto “Manage Settings” and press [Set].

7-6
System Maintenance

3)Select Export (Import or Restore factory) as required.

A B

D
C D E

Mark Function
A To export the selected presets, and save them in different files.
B To import the preset parameters according to the selection, or to
restore it to the original factory default.
C To export all the preset parameters, and save them in a file.
D To import all the preset parameters at one time.
E To restore all preset parameters to the default value.

7.2.5 Maintenance Status


Copy the files to be used into a USB flash drive. Then insert the USB flash drive into the USB
port on the main unit.
1. Enter into the setup status; (Please refer to 7.2.2 )
2. Select “Maintenance”.

7-7
System Maintenance

3. Press [Set], the following dialog box pops up.

4. Input the correct password, select “OK”, and press [Set] to enter to “Maintenance” menu.

7.2.6 Software Upgrading

NOTE 1 The process is slow when upgrade FP2, please wait.

2 Do not turn off the power supply in the process of upgrading.

7-8
System Maintenance

7.2.6.1 Upgrading
This upgrading is done through an upgrading package, the content to be upgraded is the
content changed in the last version prior to the upgrading package.
1. Enter into maintenance status (Please refer to 7.2.5 );
2. Select “Upgrade”, press [Set] key;

3. Select the correct file in the Load File dialog box, and click “OK”, the system will start the
upgrading, and the progress will be displayed on the bottom of the screen.

4. After the upgrading is complete, the successful prompt appears. Restart the system
according to prompts.

7-9
System Maintenance

NOTE: Select the proper upgrading program (package) whether or not ECG
module is configured.

7.2.6.2 Upgrading a Single Item


1. Enter into maintenance status (Please refer to 7.2.5 )
2. Press Ctrl+Shift+8, the items for upgrading appear one by one.

3. Select the item to be upgraded, and press [Set], select the right file in the Load File dialog
box, and click “OK” then the system begins to upgrade the file and the progress will be
displayed at the bottom of the screen.
4. After the upgrading is completed, the system will give out related information and prompt
the user to restart the system.

7-10
System Maintenance

NOTE: Please shutdown the system and turn off the power supply before
restarting the system after completing all the upgrading.

7.2.7 Model Configuration


1. Enter into maintenance status. (Please refer to 7.2.5 )
2. Choose the “Config” menu, and press [Set].

3. Enter the right password in the popped up dialog box.

4. The “Config” dialog box pops up.

5. Move the cursor to the “Load Config Script”, press Set, and the following dialogue box pops
up.

7-11
System Maintenance

6. Select correct files, and click “OK”. After the operation is complete, the successful prompt
appears. Restart the system according to prompts.

7.2.8 Install the Software of Optional Devices


1. Enter the preset status. (Please refer to 7.2.2 )
2. Choose the “System” menu;
3. In the system setup dialog box, select “Option”, as shown in the figure below:

7-12
System Maintenance

4. In the optional list, select the desire software function, and click “Install”;
5. Select the corresponding file with suffix “key” in the dialogue box, and then click “OK”.
The files and the devices are of one-to-one correspondence, install the software.
6. After successful installation, return to the system preset interface, the corresponding
optional modules should be shown as installed. Click “OK” to close the system preset dialog
box.

7-13
System Maintenance

7.3 System Self-diagnosis

7.3.1 Self-diagnosis Interface

Item

Select

Area

Diagram

Area

Status

7.3.1.1 Option Area of Self-diagnosis Items


The testing items can be set in this area, each item is classified into the corresponding upper
item, if the check box is selected (that is ), it means this testing item is put into the testing
sequence, otherwise, it will not. When the upper item is selected, all the corresponding default
lower items are also selected, if the upper item is cancelled, all the corresponding default
lower items will be cancelled.
Function of [OutDentAll]: click this button, the closed testing items will be opened;
Function of [InDentAll]: click this button, the opened testing items will be closed;
Function of [SelectAll]: click this button to select all the testing items;
Function of [SelectNone]: click this button to cancel all the selected testing items.

7.3.1.2 Display and Control Area of the Testing Information


This area displays the testing results of all items.
Messages: to display the information of all testing items, of which,
Index means the testing sequence number;
TestName means the testing name;
TestResult means the testing result;
Remark displays remark information, e.g. if the test failed, the detailed error information can

7-14
System Maintenance

be checked by right-drawing the level scroll bar.


OnFail drop-down list box: two options, they are Continue and Stop, when select Continue,
the test will be continued even though a test item fails during testing, while Stop means the
test will stop once a test item fails during testing.
Loops check-box: this item determines whether to recycle the test, if it is selected, the test will
be repeated according to the number entered in the right box.
Start button: click this button to perform the tests one by one, and at the same time, the button
will change into Stop button; click Stop, the test will be stopped, and the button changes to
Start, waiting for the next test.
NextFail button: there will have many fails after the tests are finished, click this button to
browse the details one by one. Click this button for once, the current failed test will be
displayed in the first row of the table, when click the button again, the next failed test will be
displayed in the first row of the table.
ClearHistory button: click this button to clear all the tested results listed in the Messages, and
the flow diagram will be returned to the original status.
Quit button: click this button to exit the self-diagnosis and return to the operation interface.

7.3.1.3 Status Bar


Some information as well as the testing progress are displayed in the status bar.
The version and issued date of the system self-diagnosis software will be displayed at the left
side of the status bar when the program is running.
The information of versions is displayed in the following forms:

Table 7-1 Version Information of System Self-checking Software

Model Application Information


DC-3 Production DC-3 SelfTest Software for Manufacuture Version:xx; Release
test Date:YYYYMMDDXX
DC-3 User and DC-3 SelfTest Software for Maintainese Version:xx; Release
user Date:YYYYMMDDXX
service
During the test, the progress bar is displayed at the right side of the status bar to indicate the
testing progress.

7.3.1.4 Structure Diagram Area


This area shows the structure of the whole hardware system, each module will be colorized
according to the test result during the testing. Red means communication failure of the
module or between the related module and the system; while green means the module is
working normally.

7-15
System Maintenance

7.3.2 Description of Testing Items


Differences exist in the testing items of different user groups.

Table 7-2 Differences in the Test Items of Each Version

DC-3
Testing Item
Manufacturer Maintainece
Register Read MainBoard ID Read
DSP Logic Version Read
MultiFunction Logic Version Read
Interconnect Test The Connection of BF and DSP(System Control
Bus)
The Connection of BF and DSP(RF Data Port)
The Connection of CPU and DSP(PCI)
The Connection of CPU and MultiFunction (PCI)
Memory Test DataBuffer SSRAM0 Test
DataBuffer SSRAM1 Test
Frame Correlation SSRAM Test
RealTime Data Upload DDR1 Test
Scan Parameter Setting DDR Test
MuliFunction DDR Test
Power Circuit Test Battery A Test
Battery B Test
Scan Mode and PHV Control Test
System Monitor Test Voltage Monitor Test
FrontEnd Circuit Test ADC Digital Port SelfTest
CW Circuit Port Test
Adjust Module Connunication Test
Prober Port Test
Prober Expension Module Port Test
KeyBoard Communication Test
Port Test
Serial Port Test
Video Test
S-Video Test
Audio Test
ECG Module Port Test
Note: : means this testing item exists in the system self-diagnosis software of this version.
: means this testing item doesn’t exist in the system self-diagnosis software of this
version.

7-16
System Maintenance

7.3.3 Details of the Testing Items

7.3.3.1 Register Read


1)MainBoard ID Read
Testing function:
Read boards and cards IDs of the main board.
Details description:
The boards and cards IDs of the main board will be displayed in the Remark column of
Messages list box in the format of:
MainBoard Version:x.x.
2)DSP Logic Version Read
Testing function:
Read the DSP logic version.
Details description:
DSP logic version will be displayed in the Remark column of Messages list box in the format
of:
The Release Date is YY/MM/DD.
The Compile Version of BF Logic is :xx;
3)MultiFunction Logic Version Read
Testing function:
Read the multifunction FPGA logic version.
Details description:
Multifunction FPGA logic version will be displayed in the Remark column of Messages list box
in the format of:
The Version of:
MultiFunction FPGA is: xx.
The date of MultiFunction Modified is: YY/MM/DD.

7.3.3.2 InterConnect Test


1)The connection of BF and DSP (System Control Bus)
Testing function:
Test the connection of system control bus between BF and DSP.
Details description:
Test if the system control bus between BF and DSP is working normally.
2)The connection of BF and DSP (RF Data Port)
Testing function:

7-17
System Maintenance

Test the connection of RF data port between BF and DSP.


Details description:
Test if the RF data port between BF and DSP is working normally.
3)The Connection of CPU and DSP (PCI)
Testing function:
Test the connection of CPU module and DSP (PCI).
Details description:
Test if the PCI channel between the CPU module and DSP is working normally.
4)The Connection of CPU and MultiFunction (PCI)
Testing function:
Test the (PCI) connection of CPU and MultiFunction FPGA.
Details description:
Test if the PCI channel between the CPU module and MultiFunction FPGA is working
normally.

7.3.3.3 Memory Test


1)DataBuffer SSRAM0 Test
Testing function:
Test the databuffer SSRAM0.
Details description:
Overall test on SSRAM U36, which including two parts:
Test if the connections between FPGA U310 and SSRAM U36 (including the data bus,
address bus and control signal cable) are normal.
Test if the SSRAM U36 itself is normal.
2)DataBuffer SSRAM1 Test
Testing function:
Test the data-buffer SSRAM1.
Details description:
Overall test on SSRAM U313, which including two parts:
Test if the connections between FPGA U310 and SSRAM U313 (including the data bus,
address bus and control signal cable) are normal.
Test if the SSRAM U313 itself is normal.
3)Frame Correlation SSRAM Test
Testing function:
Test frame correlation SSRAM.
Details description:

7-18
System Maintenance

Overall test on SSRAM U314, which including two parts:


Test if the connections between FPGA U310 and SSRAM U314 (including the data bus,
address bus and control signal cable) are normal.
Test if the SSRAM U314 itself is normal.
4)RealTime Data Upload DDR1 Test
Testing function:
Real-time data upload DDR1 test.
Details description:
Overall test on DDRAM U312, which including two parts:
Test if the connections between FPGA U310 and DDRAM U312 (including the data bus,
address bus and control signal cable) are normal.
Test if the DDRAM U312 itself is normal.
5)Scan Parameter Setting DDR Test
Testing function:
Scan parameter setting DDR test.
Details description:
Overall test on DDRAM U311, which including two parts:
Test if the connections between FPGA U310 and DDRAM U311 (including the data bus,
address bus and control signal cable) are normal.
Test if the DDRAM U311 itself is normal.
6)MultiFunction DDR Test
Testing function:
Multi-function DDR test.
Details description:
Overall test on DDRAM U28, which including two parts:
Test if the connections between FPGA U30 and DDRAM U28 (including the data bus, address
bus and control signal cable) are normal.
Test if the DDRAM U28 itself is normal.

7.3.3.4 Power Circuit Test


1)Battery A Test, Battery B Test
This testing item does not included in Model DC-3, only in M5.
2)Scan Mode and PHV Control Test
Testing function:
Mainly test if the PHV control is normal under all kinds of scanning modes.
Details description:

7-19
System Maintenance

The circuits concerned including the connecting circuit between Beamformer (U302) and DA
chip (U297), U279, connecting circuit between DA output of U297 to the power supply module,
as well as PHV power circuit.

NOTE: It may take about 3 minutes to finish this test, please wait.

7.3.3.5 System Monitor Test


Testing function:
The voltage detection test is to test if power detection circuit of the system main board and
power circuit of each group can work normally.
Details description:
To test if the connecting circuit between PC module and power monitor IC U9, U9, and the
12V, +5V, 2.5V, 1.5V, -5V and 3.3V voltages on the boards are normally working.

7.3.3.6 FrontEnd Circuit Test


1)ADC Digital Port SelfTest
Testing function:
To test if the frontend circuit ADC digital port is normally working.
Details description:
The concerned circuits include the connecting circuit between BeamFomer (U302) and
frontend AD (U298~U301) and AD converter itself.
2)ADC DC Bias SelfTest
Testing function:
Frontend ADC DC biasing test, to judge if the noise level is normal by gathering the system
noise through the frontend circuit ADC.
Details description:
The concerned circuits include the connecting circuit between BeamFomer (U302) and
AD(U298~U301) as well as frontend AD itself.

7.3.3.7 Port Test


1)CW Circuit Interface Test
Testing function:
It is CW circuit interface test.
Details description:
The main function is to test if connection between CW module and the system, as well as CW
module itself are normally working.
2)AD Controller Board Communication Test

7-20
System Maintenance

Testing function:
It is AD controller board communication test.
Details description:
It is to test if SM bus between PC module and AD controller board is normal, and if AD
controller board is normal.
3)Probe Port Test
Testing function:
To test the ID communication channel between the probe board and the probe is normally
working.
Details description:
Plug the probe 7L4A into the A socket of the probe board before testing the DC-3 series
instruments.

DC-3 probe is equal to probe 7L4A


4)KeyBoard Communication Test
Testing function:
It is to test the keyboard board assembly (key, encoder, trackball, TGC and key backlight),
boards and cards ID read, keyboard FPGA version read and keyboard operating program
version read.
Details description:
The keyboard testing interface is shown as follows:

7-21
System Maintenance

Click [Start] to start the testing, each control on the simulating keyboard is corresponding to
the one control on the actual keyboard.
Press down the key on the actual keyboard, the corresponding key on the simulating
keyboard will flash and change into green, flashing means the key is being tested, while
green means the key past the test.
Scroll the trackball, the cursor on the interface will be shifted accordingly.

Encoder rotating function test: on the simulating keyboard represents the encoder on
the actual keyboard, when the rotating function of the encoder is being tested, the ball
corresponding to the encoder will flash and change into green. Flashing means the encoder is
being tested, while green means the encoder past the tested. Meanwhile, with the rotating of

the encoder, the green ball around will rotate in the same direction with the encoder.

Encoder key test: press down the key of the encoder, the ball corresponding to the encoder
will flash and change into green. Flashing means the encoder being tested, while green

means the encoder past the test. Meanwhile, the central key of will change with the

changing of the encoder.

TGC test: the 8 TGCs on the keyboard are corresponding to controls on the

simulating keyboard respectively. Slide the TGC on the keyboard, the TGC on the simulating

7-22
System Maintenance

keyboard will slide accordingly.


Keyboard back lights test: when the keyboard is initialized, all the two-color lights will change
into green; when being tested, the color will change into orange under the control of the
running program.
Zoom, Depth LED lights test: press down Zoom/Depth encoder, the Zoom and Depth lights
will be turned on alternately under the control of the running program.
B C D LED lights test: when rotate the gain encoder, the B, C, D LED lights will be turned on
cyclically under the control of the related programs.
After all parts are tested, the keyboard will be shown as below. (Note: when test the Alt key,
you have to press down Ctrl + Alt).

If all parts of keyboard are in normal conditions, press down [KeyBoard is OK and Quit] button,
otherwise, press down [KeyBoard is Bad and Quit] button.
5)Serial Port Test
Testing function:
Serial port test, the main function is to test if the serial ports are normally working.
Details description:
Connect a tester with the serial port before the testing.
6)Audio Test
Testing function:
The main function is to test if the two buzzers and the related circuits are normally working.

7-23
System Maintenance

Details description:
During the test, the left side of machine will play out “Ding~~”, and pops up the following
dialog box to ask if you can hear “Ding~~” playing out of the left side machine. If you didn’t
hear it, please click “Retry” to play it again, and click “No, I Can’t”, if you still can’t hear it after
trying many times. And if you heard it, please click “Yes, I Can”.

After that, the program will control the right side of the machine to play out “Ding~~”, and pop
up the same dialog box, please do the same operation as described above.
7)ECG Module Port Test
Test function:
The main function is to test if the ECG module is normally working.
Details description:
After the ECG module is connected to the machine, it will communicate the system
self-checking module during the self-checking process, and the ECG module can be judged if
it is normally working according to the feedback information.

NOTE: If no ECG module is configured, this item is a Fail item during the
self-checking.

7-24
Troubleshooting

8 Troubleshooting

8.1 Malfunction

8.1.1 Examine and Repair Flow of Blank Screen


N

Blank Y
Blank screen?
screen

Replace master
N Can the machine be Replace CPU
board
started?

Machine Y
can’t Y
be started

VGA output to the


monitor Blank screen? N Over

Replace IO
Any display? N
port board

If the LCD signal Re-connect or


N
power connected replace the cable
well?

Replace monitor
Y Blank screen?
module

Over

Fig 8-1 Examine and Repair Flow of Blank Screen

8-1
Troubleshooting

8.1.2 Examine and Repair Flow When No Image Displayed in

the Image Area

Over No image

Replace the master Any probe


N
board connected

Y
Y

Noise-free
Remove all probes
Image?

Connect the probe,


No image
replace power N Over
displayed?
supply module

Follow the
flows to
check blank
screen

Fig 8-2 Examine and Repair Flow When No Image Displayed in the Image Area

8-2
Troubleshooting

8.1.3 Examine and Repair Flow When Black Area Exists in

the Image
Black area
Over in the
image

Replace the master Any probe


N
board connected

Y
Y

Black area? Remove all probes

Replace the probe Black area? N Over

Y N

Black area
Replace probe
exists with all N Black area?
control board
probes

Replace Y
Y
transmitting board

Replace probe
board

Replace the master


Y Black area?
board

Over

Fig 8-3 Examine and Repair Flow When Black Area Exists in the Image

8.1.4 Abnormal Image in the Image Area


Problems concerned in this section including B, C, PW images abnormity, to exclude these
malfunctions, first you shoud make sure there is no power supply malfunction according to Fig
8-8, and then change the master board.

8-3
Troubleshooting

8.1.5 CW Image Abnormity


To exclude this problem, there are three steps to be carried out:
Replace the power supply module to judge if it is normal, for the examine and repair flow,
please refer to Fig 8-8.
Replace the CW board to judge if it is normal, for the malfunction collection, please refer to
General Malfunction of CW Board.
If CW board has no problem, you have to change the master board.

8.1.6 Examine and Repair Flow When the System Can’t

Start-up

Machine
can’t be
started

Any output in
N Replace adapter
adapter?

Properly connect Soft switch properly


N Y Can’t be started?
the soft switch connected?

Y N

Replace IO
Can't be started? Y Over
port board

N
Over N Can’t be started? Y Replace CPU

Replace master
Y Can’t be started?
board

Fig 8-4 Examine and Repair Flow When the System Can’t Start-up

8-4
Troubleshooting

8.2 Modules and Boards Malfunction

8.2.1 LCD Display

8.2.1.1 No Displaying in the LCD and the Indicator Lights are OFF

Check if LCD control


board is properly
connected with OSD
board
N
Y Reconnect the two
boards

Check if power supply


(12V/ 5V) to J6 on LCD
control board is normal N
Check if power supply
Y (12V/ 5V) to CN1 is normal

N Check if power cable of


Replace LCD
control board and Y CN100 is properly
LCD screen connected

Check if CN1 is
Y
properly
connected
Replace the
N
power supply
Y
board
Reconnect the
cable

Note: LCD control board should match with the screen, and them should be replaced together if required.

8-5
Troubleshooting

8.2.1.2 LCD Malfunction, While the Indicator Lights Can be ON

8.2.1.3 General Malfunctions of the AD Controller Board


No. Possible
Failure Measure
Cause
The image is located Try Auto Adjust under the image without
in the wrong area, or boundary.
1 lateral and vertical
strips appear on the
image.
The image color is Try Auto Color under the black/ white
2
abnormal. image.
Check the socket of the back-light board,
There is signal input,
check if LVDS is connected well with
3 but nothing displayed
PANEL, and if LVDS is properly connected
on the screen.
with J1 socket of AD board.
The displaying image Check if AD board is securely connected
4 is abnormally with the master board, and LVDS is
fluttering. properly connected.
Check if the display protection program or
Appears remnant timer is acting when the same image is
5
image. displayed for a long time.
The remnant image is caused by the LCD

8-6
Troubleshooting

itself, please avoid displaying the same


image for a long time.
Check if socket of J5 key board is
connected well;
The indicator light is
Check if J6 socket power cable is
6 OFF, no response to
connected well;
the switches
Check the 12V, 5V power supply of J6
socket is normal.
If the output of main unit is normal, but the
display indicator light is orange and no
The color of the
display on the screen, please check if J3
7 indicator lights are
signal input port is connected well and if
orange
there are voltages input to HSYNC,
VSYNC synchronization signals.

8.2.2 General Malfunctions of the Control Panel Unit

8.2.2.1 General Malfunctions of the Control Panel Buckled Board


No. Failure Possible Cause Measure
1 Indicator light No 3.3V power supply to the Check the 3.3V power supply
D4 is OFF control panel. output on the control panel.
2 Indicator light USB is not properly Check if USB cable is damaged
D1 is OFF enumerated. or dropped down.
FPGA configure file hasn’t Burn the FPGA configure file.
been burned.
Indicator light Check if the kernel voltage of Replace U20.
3 D3 is OFF FPGA is normal, test the
sometimes. voltage drop of C27, the
standard voltage is 1.2V.
FLASH malfunction Replace U3.

8.2.2.2 General Malfunctions of the Control Panel


No. Failure Possible Cause Measure
1 Indicator light No 12V power supply to the Check the 12V power supply
D6 is OFF. control panel. circuit of the control panel.
2 Indicator light No 3.3V power supply to the Check the 3.3V power supply
D175 is OFF. control panel. circuit of the control panel.
3 Indicator light No 5V power supply to the Check the 5V power supply circuit
D176 is OFF. control panel. of the control panel.
4 When Trackball connecting cable Check if the trackball connecting
operating the malfunction. cable is properly connected.

8-7
Troubleshooting

trackball, it has Trackball damaged. Check if the trackball is


no response. dust-covered or damaged.

8.2.3 General Malfunctions of the Probe Board


No. Failure Possible Cause Measure
Check if the two boards of the Properly buckle the probe board
probe board module have and probe control board.
been properly buckled.
The probe expander can’t be Check if fuse F1 of the probe
powered (no sound of control board is conducted, if it is
switching the relay can be damaged, please replace the
heard when powered on) fuse.
Test if the VCC (5V), VDD Check if there is short circuit, if no
(3.3V) voltages on the probe such problem can be found, take
control board are normal. the following methods:
The probe Replace U1 if 3.3V voltage is
can‘t be abnormal;
recognized by If the 5V voltage is over-low, the
1
the probe port main unit may be damaged (the
of the main possibility is little).
unit.
Check the connection of each Connect the boards properly. If
boards, pay more attention to the socket is damaged, please
check if the socket connecting replace the board.
the probe connecting board
and probe board is loose or
been damaged.(make sure
that the communication cable
is connected well).
CPLD U3 of the probe control Return the probe control board for
board or the driver’s U3, U4 testing and repairing.
are damaged.
U9 (A socket), U12 (B socket) Replace the corresponding unit.
or U13(C socket) of the probe
Probe A, B, C control board are damaged.
sockets can’t
The resistors corresponding to Replace the corresponding
2 be recognized
AID, BID or CID signal are resistor.
by the main
shortened or opened.
unit.
Note: Carry out testing on the testing jig to determine the
malfunction before returning the module for repairing.
When the Check if the two boards of the Properly connect the boards, or
probes of A, B, probe board module are replace the corresponding socket.
3
C sockets are properly connected, and if
working, about socket is damaged or the leg

8-8
Troubleshooting

1/8 or 1/4 of deviated.


fixed area has The parts in the relay control Replace the damaged parts.
no image circuit of the corresponding
displayed. probe are damaged.
If all other probes have the same
problem, may be the board of the
Check if other probes have the main unit is damaged.
same problem. If only one probe has this
problem, may be the probe is
damaged.
If the three sockets have this
problem, the possible causes are:
1. The probe board sockets
P1~P3 have been damaged;
When the 2. The probe module is not
probes of A, B, properly connected with the
C socket are probe connecting board (the
working, there possibility is little);
4
are one or 3. If the two passes are in
several vertical series, it may caused by the
blank passes Check if the probes of sockets
damaged corresponding
in the images. A, B, C have the same
relays.
problem.
If the problem only exists in one of
the socket, the possible causes
are:
1. The two boards are not
buckled well;
2. The “board-to-board” socket
is damaged;
The probe port of the probe board
module is damaged (the
possibility is little).

8.2.4 General Malfunctions of CW Doppler Board


No. Failure Possible Cause Measure
Make sure the board is properly Fix the connections, make
The 3.3V or 5V connected to the master board, and sure there are power supply
power supply there is power supply to the CW existing in J1, J2 power pins,
1 indicator light is board from the master board. otherwise, please replace
OFF after the master board.
powered on. The L41,L40 are effectless. Replace the effectless
inductor.
CW output Test the frequency of the output Replace U30, U31 if error
2
signal signal of U30, and U31; if they are exists in output signals of

8-9
Troubleshooting

normal, then test the output of U12, U30, U31, if error still exists,
and U14 to check if the output signal replace U13; if U30, U31 and
is normal and if there are distortion U13 are normal but error
in the signal. exists in output signals of
U12, U14, first check if the
peripheral resistors and
frequency
power supply of U12, U14
error.
are OK, if they are normal,
then replace U12, U14.
The pre-demodulation low-pass filter Check if there is damage to
is damaged. the filtering peripheral units,
and replace the damaged
units.
2.5V voltage input/output error in U2, Replace U6, U2 or U4.
and U4.
U1,U5 is damaged, which can be Replace U1 or U5.
concluded from the same input but
Amplitude
different output.
inconsistency
3 of the CW U13 is damaged, which can be Replace U13.
output signal I, concluded from the output
Q. inconsistency.
There are amplifier or peripheral Test and compare the testing
devices be damaged in the points, and then replace the
amplification stage or filtering unit of damaged units.
I and Q.
U13 is damaged. Replace U13
The pre-demodulation low-pass filter Check if there is damage to
is damaged. the filtering peripheral units,
and replace the damage
units.
High harmonic
Output signal saturation distortion in Check if any peripheral unit
wave appears
4 the amplification stage of I, Q. in the amplification stage is
in the CW
damaged, if any, please
output signal.
replace it.
I, Q wall filter output distortion. Check if there is damage to
the filtering peripheral units,
and replace the damage
units.
5 No CW output Electronic switch is damaged. Replace U11.
signal The low-pass filter is damaged. Check if any amplifier in the
amplification stage is
damaged, if any, replace it.
Demodulation IC U13 is damaged. Replace U13.
Amplifier damaged in the Replace the damaged units.
amplification stage and wall filter of I,

8-10
Troubleshooting

Q.
Damage to ADC IC U6 or signal end Replace U6, U2 or U4.
to difference circuit.

8.2.5 General Malfunctions of USB-to-IDE Port Board


No. Failure Possible Cause Measure
After DVD RW Check if USB connecting cable can Replace USB connecting
is connected, work normally. cable.
1 Windows Malfunction of USB communication. Refer to USB
doesn’t ask to communication
test the device. troubleshooting.
No power on Malfunction in the 5V voltage testing Replace USB connecting
the board point. cable or inductor L1.
(refer to Malfunction in the 3.3V voltage Replace U2.
2
testing method testing point.
of board power
supply).
USB communication malfunction on USB-to-IDE port board
USB signal input circuit is shown as Fig 8-5. DP and DM are USB signal differential input pair,
they are connected to ESD protection IC IP4220 through J2 (PIN5 2mm). The electrostatic
level of this IC is 8000V. L1 is the common-mode inductance specified for USB.

Fig 8-5 USB Port

To judge the USB communication malfunction, you can test if L1 is working normally, if not
(the conducting resistor is less than 1Ohm), it means malfunction exists in L1. And if L1 is
normal, please replace IP4220CZ6 to judge the malfunction. At last, test should be done to
U3 and the peripheral circuit if the former method can’t determine the malfunction.

Power supply test of USB-to-IDE port board


The board is powered by VBUS, which will be converted into 3.3V through LDOSPX1117, the
circuit is shown as follows:

8-11
Troubleshooting

Fig 8-6 USB-to-IDE Power Supply Circuit

5V testing point: the voltage of L2, the standard value is 5V±5%;


3.3V testing point: the voltage-drop of C10, the standard value is 3.3V±5%.

8.2.6 PC System Related Malfunctions


No. Failure Possible Cause Measure
If 5VSTB has no output, the
malfunction may exist in the
power supply module.
PC can’t be 5VSTB and 12V have no voltage
1 If just the 12V has no output,
started. output for PC board.
the malfunction may exist in
the power supply
management module.
2 The system Management sequential Check if CPLD is damaged.
gets reset malfunction in PC power supply.
repeatedly after
powered on.
3 System startup System access speed is slow Remove the flash memory
is very low. because when flash memory is and use it after accessing the
applied on the system, the system operation system.
will be restarted from the flash
memory.
The battery has no power. Replace for a new battery.
System clock is
4 RTC clock of CPU can’t work Replace CPU.
abnormal.
normally.
5 Malfunction in FPGA is damaged. Replace the corresponding
PCI peripheral FPGA.
units
(multi-function
FPGA, DSP
FPGA).
6 Malfunction in Multi-function FPGA can’t work Check if multi-function FPGA

8-12
Troubleshooting

normally. is damaged.
testing of Multi-function FPGA is normal, but If the testing of voltage,
system voltage, ADT7462 can’t work normally. temperature and fan rotating
temperature, as speed is abnormal while the
well as fan multi-function FPGA is
rotating speed. normal, then the malfunction
must exist in ADT7462.
Multi-function FPGA can’t work Check if multi-function FPGA
normally. is damaged.
Malfunction in Multi-function FPGA is normal, but If Video/SVideo output is
7 Video & the video converting IC 25874 abnormal while multi-function
S-Video output can’t work normally. FPGA is normal, then the
malfunction must exist in
25874.
Check if the bus current to the If the USB port can’t be used
USB port can USB is more than 1A. because of surge, just to turn
8
not be applied. off the system and then
restart it.

8.2.7 Power Supply Malfunction


Most of the devices on the power supply connecting board are passive one, so they are not
easy to be damaged, but it is possible that the fuse may be damaged due to peripheral device
over-current or short circuit.
When malfunction exists in the voltage selection switch of the power supply connecting board,
there are two possible causes:
1. Connection of 120Vac: when input 230Vac, the isolating transformer may be saturated,
which will lead to the temperature fuse in the transformer be burnt out or cause the breaker
tripped.
2. Connection of 230Vac: when input 120Vac, the transformer copper loss will get severe,
which will lead to greater heat emitting. This will not be obvious when the overall load is
insignificant.
The following figure shows the internal winding and the different connections of the isolating
transformer.

8-13
Troubleshooting

Periphe
rals
90~140Vac Primary windings are

Main
unit
connected in parallel, which
increased the load capability

198~264Vac Secondary windings are


connected in series, which
increased the voltage-adapt

Periphe
capability

Main

rals
unit
Fig 8-7 Connections of the Transformer Winding

When the malfunction exists in the adapter of the power supply system or the DC power, the
malfunction can be determined according to the flows in Fig 8-8.

8-14
Troubleshooting

M a k e s u re th e b o a rd s
a re p ro p e rly c o n n e c te d

C o n n e c t A C a d a p te r a n d
AC pow er

M a lfu n c tio n in T H V , o r
N o rm a l s h o rt -c irc u it / o v e r-c u rre n t A b n o rm a l If T H V is n o rm a l?
in th e lo a d

If e d c _ p o w e r o n
M a lfu n c tio n in a d a p te r o r
p o w e r s u p p ly N o rm a l
s h o rt c irc u it o n p o w e r A b n o rm a l
m a s te r b o a rd is
s u p p ly m a s te r b o a rd
n o rm a l?

R e p a ir + 2 .5 V a n d 1 .5 V
If + 2 .5 V a n d 1 .5 V
u n its o n th e p o w e r A b n o rm a l
a re n o rm a l?
N o rm a l s u p p ly a u x ilia ry b o a rd

R e p a ir p o w e r s u p p ly N o rm a l
If 5 V s tb is
m a s te r b o a rd U 9 re la te d A b n o rm a l
n o rm a l?
u n it

R e p a ir -5 . 3 V a n d -1 2 V
If -5 .3 V a n d
u n its o n th e p o w e r A b n o rm a l
N o rm a l -1 2 V a re n o rm a l?
s u p p ly a u x ilia ry b o a rd

W hen N o rm a l
R e p a ir p o w e r s u p p ly
5 V s tb _ c p u _ e n is
m a s te r b o a rd 5 v s tb _ c p u A b n o rm a l
0 , if 5 V s tb _ c p u is
re la te d u n it
n o rm a l?

O ver

N o rm a l

R e p a ir th e p o w e r s u p p ly W h en P ow er_ o n
m a s te r b o a rd + 1 2 V A b n o rm a l is 0 , if + 1 2 V is
re la te d u n it n o rm a l?

N o rm a l

R e p a ir th e p o w e r s u p p ly
m a s te r b o a rd + 5 V A b n o rm a l If + 5 V is n o rm a l ?
re la te d u n it

N o rm a l

R e p a ir th e p o w e r s u p p ly
If + 3 .3 V is
m a s te r b o a rd + 3 .3 V A b n o rm a l
n o rm a l?
re la te d u n it

N o rm a l

R e p a ir th e p o w e r s u p p ly
If o u tp u t o f U 1 (-
m a s te r b o a rd U 1 re la te d A b n o rm a l
1 2 V ) is n o rm a l?
u n it

Fig 8-8 Examine and Repair Flow of DC-DC Power Supply

8-15
Troubleshooting

8.2.7.1 Power Supply Connecting Board


When malfunction exists in the power supply connecting board, you should first check if the
setting of the two voltage selection switches of the isolating transformer are correct and then
observe if the fuses corresponding to the 3 peripheral devices’ outputs are damaged.

8.2.7.2 Isolating Transformer


The malfunction in the isolating transformer may be caused by the burnt-out of the internal
temperature fuse, disconnect the transformer external wiring and then test if the two primary
windings are opened.

8.2.7.3 Malfunctions in the Power Supply Master Board and Auxiliary


Board

8.2.7.4 Troubleshooting of 5vstb Circuit


If the malfunction exists in 5vstb, first test if the input voltage of PIN1 of U9 is normal; if it is
normal, then it can be concluded that U9 is damaged or there are over-current or short circuit
happened at the load end.

8.2.7.5 Troubleshooting of +12v Circuit


Before the test, please make sure that the power_on signal is of low level, and the input
voltage VBUS+ is normal.
Examine and repair flow:
Test the voltage of PIN8 of U3, which should be higher than 4V;
Test the voltage of PIN19 of U3, which should be higher than 6V;
Check if there are drive pulse at the G polarity of MOS Q11, Q18, Q24 and Q25 by the
oscilloscope. The normal status is that there should have 2 MOS have drive pulse at the G
polarity.
Check if there are any damage to the 4 MOS Q11, Q18, Q24, Q25, and the 2 diodes D5, and
D7.

8.2.7.6 Troubleshooting of +5v and +3.3v Circuit


Before the test, please make sure that the +12V input as well as the input voltage VBUS+ are
normal.
Examine and repair flow:
Test the level of PIN23, which should be high;
Check by sight if R6 (+3.3v unit) and R12 (+5v unit) are properly jointed.
Check by sight if the joint of IC is deviated, which will lead to short circuit between the pins at
the bottom of IC.

8-16
Troubleshooting

8.2.7.7 Troubleshooting of THV Circuit


Before the test, please make sure that the +12V output as well as the U1 negative voltage
output are normal.
Low voltage section and high voltage section are independent from each other. First, check if
the THV is at circuit protection status because of over-current and short circuit, and this can
be done by the ocp signal (the ocp signal is of high level in the case of over-current).
When the malfunction exists in the low voltage section, please check if U10 (TPS54350), U7,
D1, R142 and R143 are in good conditions.
When the malfunction exists in the high voltage section, please check if U6 (TL594), U2, Q5,
D9, R110 and R111 are in good conditions.
Troubleshooting of U10 and the peripheral circuits:
Test the voltage of PIN7, which should be higher than 0.5V;
Test the voltage of PIN12, which should be higher than 8V;
Test the voltage of PIN12, which should be 8V higher than that of PIN15;
Test the voltage of PIN14 and PIN15, which should have pulse voltage wave.
Troubleshooting of U6 and the peripheral circuits:
Before the test, please make sure that the cw_mode is of low voltage, that means the
instrument is working at high voltage section.
Test the voltage of PIN8 and PIN11, which should be higher than 12V;
Test the voltage of PIN14, which should be higher than +5V;
Test the voltage of PIN5, which should have oscillatory wave;
Test the voltage of PIN15, which should be higher than 1.6V;
Test PIN9 and PIN10, which should have drive pulse.

8.2.7.8 Troubleshooting of +2.5V and +1.5V circuit units


Before power on, please check by sight if EL7566 and the peripheral units are properly jointed,
and if any pins of EL7566 are of short circuit and dry joint.
Examine and repair flow:
Test the voltage of PIN19~PIN21, which should be +5V;
Test the voltage of PIN22, which should be +5V;
Test PIN27, which should have oscillatory wave output;
Test PIN8~PIN13, which should have pulse voltage wave.

8.2.7.9 Troubleshooting of -5V and -12V Circuits


Make sure the power supply VBUS+ to the -5V and -12v circuits are at the normal range.
Examine and repair flow after powered on:
Test the voltage of PIN15 (power supply to the power supply chip), which should be the same

8-17
Troubleshooting

with VBUS.
Test the voltage of PIN5 (the reference voltage pin), which should be 1.25V.
Test the voltage of PIN8 (over-voltage/ over-current protection control pin), which should be
higher than 0.5V.
Test PIN14 (drive pin), which should output drive pulse.

8.3 Software Troubleshooting

8.3.1 Troubleshooting in the Case of HDD Damaged


If the hard disk is damaged due to some special causes, please make a new one follow to the
following steps:
1. Use the Nero tool, select the menu [Rewriter| R/W CD-ROM image file…”], and then select
the image file (2108-30-66180), rewrite a hard disk initialized CD-ROM.
2. Set the BIOS of ultrasound system.
2.1) Start the ultrasound system, press F2 on the control panel, enter the BIOS password to
begin the BIOS setting.
2.2) Set 【CMOS Restore Condition】 as Never in the 【Exit】 page; then click Save
Changes--Save CMOS To Flash; set 【Boot Order】as Boot from USB CDROM in the 【Boot】
page (Select USB CDROM by the up arrow or down arrow, press the ‘Shift” key while
pressing, at each pressing of “+”, the USB CDROM will move up a step, we should put the
item to the top).
3. Make a hard disk
3.1)Start the power, put the hard disk into the CD-ROM driver, it will prompt “Press any key to
boot CD-ROM…” on the screen, press any key to automatically enter into the Windows XP
pre-installation environment.
3.2)The control window will appear on the screen after successfully started, the version
information will be displayed, and will prompt “Press any key to continue …”, press any key to
continue the hard disk initialization.
3.3)The initialization is automatically processed, and “Over” means the whole process is
completed, and it will display “Press any key to continue …” again, press any key to restart
the system, then a hard disk is made.
4. Change the BIOS setting after the hard disk is repaired. Enter into BIOS, cancel the
function to start from USB-CDROM by setting the 【Boot Order】 on 【Boot】 page, (select the
USB CDROM by up arrow or down arrow, press the ‘Shift” key while pressing, at each
pressing of “-”, the USB CDROM will move down a step, we should put the item below the
HDD), set 【CMOS Restore Condition】 as Always on 【Exit】 page, then click Save

8-18
Troubleshooting

Changes--Save CMOS To Flash; at last, click【Exit Saving Changes】 on【Exit】 page to exit.

All the data on the hard disk will be deleted after formatting!
WARNING

Note: To make a new hard disk, you have to re-install the XPE system (see XP Operation
System Troubleshooting), and run the user service ultrasound software restore program (see
Ultrasound Software System Troubleshooting); and then configure the machine according to
the model, at last, configure the optional functions according to the function CD-ROM owned
by the user. (see 7.2.7 and 7.2.8 ).

8.3.2 XP Operation System Troubleshooting


1. Use Nero tool, select the menu [Rewriter| Rewrite CD-ROM image file”], add the XPE setup
program to rewrite a CD-ROM with the XPE setup program (2108-30-66108).
2. Set the BIOS of the ultrasound system
2.1) Start the ultrasound system, press F2 on the control panel, enter the BIOS password to
begin the BIOS setting.
2.2) Set 【CMOS Restore Condition】 as Never on 【Exit】 page, click Save Changes--Save
CMOS To Flash; set 【Boot Order】 as Boot from USB CDROM on 【Boot】 page (select USB
CDROM by the up arrow or down arrow, press “Shift” key while pressing, at each pressing of
“+”, the USB CDROM will move up a step, we should move the item to the top), at last, select
【Exit Saving Changes】 on 【Exit】 page to exit.
3. Setup XP operation system
3.1) Put the XPE setup image CD-ROM into the CD-ROM driver, it will prompt “Press any key
to boot CD-ROM…” on the screen, press any key to automatically enter into the Windows XP
pre-installation environment.
3.2) The control window will appear on the screen after been successfully booted, the version
information will be displayed, and will prompt “Press any key to continue …”, press any key to
continue the Windows© XPE installation.
3.3) “Over” means the whole process is completed, and it will display “Press any key to
continue …” again, press any key to reboot the system.

NOTE: 1 The Windows setup is automatically processed, never do


anything during the setup to avoid problem.
2 XPE will re-setup the drive program after XP is rebooted, it
may ask to reboot again, setup will be completed after
rebooting.

8-19
Troubleshooting

4. Change the BIOS setting after the hard disk is repaired. Enter into BIOS, cancel the
function to boot from USB-CDROM by setting the 【Boot Order】 on 【Boot】 page, (select the
USB CDROM by up arrow or down arrow, press the ‘Shift” key while pressing, at each
pressing of “-”, the USB CDROM will move down a step, we should put the item below the
HDD), set 【CMOS Restore Condition】 as Always on 【Exit】 page, then click Save
Changes--Save CMOS To Flash; at last, click【Exit Saving Changes】 on【Exit】 page to exit.

1 When the XPE system is restored, the ultrasound


WARNING:
software as well as all the related data will be deleted
(including the user preset files and backup information
in patient data bank), but the patient information and
logs in D disk will be reserved, and all files in the E disk
will be reserved, too.
2 If the system is restored, you have to run the user
service ultrasound software restore program (see
Ultrasound Software System Troubleshooting); and then
configure the machine according to the model, at last,
configure the optional functions according to the
function CD-ROM owned by the user. (see 7.2.7 and
7.2.8 ).

8.3.3 Ultrasound Software System Troubleshooting

8.3.3.1 System Malfunction Classification and Simple Troubleshooting


There are two situations if the ultrasound software is damaged:
Situation 1: System can not normally enter into the ultrasound system interface, if this
happened, first, you have to restore the XPE system (see XP Operation System
Troubleshooting), and then to restore the ultrasound system software;
Situation 2: System can enter into the ultrasound operation system, but some of the functions
can not be normally operated. There are two methods to deal with this problem:
2.1) Upgrade the software according to the software maintenance methods that introduced in
Chapter 7;
2.2) If no response to the first method, then restore the ultrasound system software according
to the software maintenance methods that introduced in Chapter 7.

8.3.3.2 Ultrasound System Software Restore


Preparation: use Nero tool, select the menu [Rewriter| Rewrite CD-ROM image file”], select
the ultrasound system restore software (2109-30-76419) to rewrite an ultrasound system

8-20
Troubleshooting

restore software.
1)The ultrasound system software must be setup in XP interface, if the XPE system is setup
just for a short time, the system can be restored by the ultrasound system restore software; if
the machine is still running under the ultrasound system, some simple operations are required
before restoring the system by the ultrasound system restore software.
1.1) To restore the ultrasound system software in the case that the XPE system is setup just
for a short time: put the prepared ultrasound system restore software CD-ROM to the
CD-ROM drive, the system software setup will be automatically started. When the interface
prompts “System need reboot!Please input Enter key to quit the Program!”, press [Enter] on
the control panel and the system will return to the Windows desk. Take out the CD-ROM and
press the system soft switch to turn off the main machine and then switch off the breaker.
After the system is booted, it will enter into the ultrasound system.
1.2)To restore the ultrasound system software in the case that the machine is still running
under the ultrasound system:
Press “Ctrl + Shift + =” on the control panel, and enter the correct password in the popped out
dialogue box, open the online debugger, and enter “shellapp off” in the debugging column,
press [Enter], turn off the machine and then reboot it. After the machine is booted, it will
directly enter into the Windows desk; delete the M5, and PatientBak file folders in C disk, and
the PADIENTDATA_2108 folder in D disk; put the prepared ultrasound system restore
CD-ROM into the CD-ROM drive, the setup will be automatically started, when it prompts
“System need reboot!Please input Enter key to quit the Program!”, press [Enter] and the
system will return to Windows desk. Take out the CD-ROM and press the system soft switch
to turn off the main machine and then switch off the breaker. After the system is started, it will
enter into the ultrasound system.;
2)If the ultrasound system is restored, the machine has to be configured, also, the optional
functions should be configured according to the function CD-ROM owned by the user. (see
7.2.7 and 7.2.8 )
After the ultrasound software system can be applied on the
WARNING:
machine, the original ultrasound software, the related
information (including the user preset files and backup
information in patient data bank ), as well as the patient
information in the D disk will be deleted when restore the
system of the machine, so remember to backup all the
information before deletion.

8-21
Troubleshooting

8.3.4 Patient Databank Troubleshooting


All information will be backed up when the main databank is working, when any error
appeared during the working, the backed up information can be used to restore the databank
automatically without any manual operation.
1) Press “Ctrl + Shift + =” on the control panel, and enter the correct password in the popped
out dialog box, open the online debugger, and enter “shellapp on” in the debugging
column, press [Enter], turn off the machine and then reboot it.
2) After the machine is rebooted, it will enter into Windows desk, manually delete
D:\PATIENT_2108 and C:\PATINET.BAK, and run the doppler.exe file under directory of
C:\M5\TargetData\exe to start the ultrasound software.
3) Press “Ctrl + Shift + =” on the control panel, and enter the correct password in the popped
out dialog box, open the online debugger, and enter “shellapp on” in the debugging
column, press [Enter], turn off the machine and then restart it.
Manual databank deletion is non-reverse, after deletion, all the
WARNING:
saved patient information will be lost forever.

NOTE: If the patient databank loaded to the external mediums is


damaged, there is no way to restore it.

8-22
Definition of Commonly Used Sockets and Functions in

System Maintenance

Figure A-1 Layout of Each Module and Socket on the Master Board
Ports of probe module

CW board
Receiving unit

T ra n sm ittin g
b o a rd

Signal processing unit


Master
board

Power supply module


Extended port

CPU module
Dual-USB

SVIDEO
Network
Adapter
port

port

port

1
Definition of Commonly Used Sockets and Functions in System Maintenance

Table 8-1-1 Sockets between the Master Board and Probe Board

CON1 CON2
PIN PIN PIN
PI NAM PI NAM PI NAM PI PIN PI PIN PI PIN
N E N E N E N NAME N NAME N NAME
1 GND 2 PE1 3 GND 1 PE69 2 PE70 3 PE71
4 PE2 5 GND 6 PE3 4 PE72 5 PE73 6 PE74
7 GND 8 PE4 9 PE5 7 PE75 8 GND 9 PE76
10 PE6 11 PE7 12 GND 10 GND 11 PE77 12 PE78
13 PE8 14 GND 15 PE9 13 PE79 14 PE80 15 GND
PE1 PE1
16 GND 17 0 18 1 16 PE81 17 GND 18 PE82
PE1 PE1
19 2 20 3 21 GND 19 GND 20 PE83 21 PE84
PE1 PE1
22 4 23 GND 24 5 22 PE85 23 PE86 24 GND
PE1 PE1
25 GND 26 6 27 7 25 PE87 26 GND 27 PE88
PE1 PE1
28 8 29 9 30 GND 28 GND 29 PE89 30 PE90
PE2 PE2
31 0 32 GND 33 1 31 PE91 32 PE92 33 GND
PE2 PE2
34 GND 35 2 36 3 34 PE93 35 GND 36 PE94
PE2 PE2
37 4 38 5 39 GND 37 GND 38 PE95 39 PE96
PE2 PE2
40 6 41 GND 42 7 40 PE97 41 PE98 42 GND
PE2 PE2
43 GND 44 8 45 9 43 PE99 44 GND 45 PE100
PE3 PE3 PE10
46 0 47 1 48 GND 46 PE101 47 2 48 PE103
PE3 PE3 PE10
49 2 50 GND 51 3 49 PE104 50 5 51 GND
PE3 PE3
52 GND 53 4 54 5 52 PE106 53 GND 54 PE107
PE3 PE3 PE10
55 6 56 7 57 GND 55 GND 56 8 57 PE109
PE3 PE3 PE11
58 8 59 GND 60 9 58 PE110 59 1 60 PE112
PE4 PE4
61 GND 62 0 63 1 61 PE113 62 GND 63 PE114
PE4 PE4 PE11
64 2 65 3 66 GND 64 PE115 65 6 66 PE117
PE4 PE4 PE11
67 4 68 GND 69 5 67 PE118 68 9 69 GND

2
Definition of Commonly Used Sockets and Functions in System Maintenance

PE4 PE4
70 GND 71 6 72 7 70 PE120 71 GND 72 PE121
PE4 PE4 PE12
73 8 74 9 75 GND 73 GND 74 2 75 PE123
PE5 PE5 PE12
76 0 77 GND 78 1 76 PE124 77 5 78 PE126
PE5 PE5
79 GND 80 2 81 3 79 PE127 80 GND 81 PE128
PE5 PE5 SPI_C
82 4 83 5 84 GND 82 SPI_CLK 83 S 84 SPI_DIN
PE5 PE5 SPI_DOU FLAS
85 6 86 GND 87 7 85 T 86 H_WP 87 GND
PE5 PE5 FLASH_P PROBE_
88 GND 89 8 90 9 88 OWER 89 GND 90 ID6
PE6 PE6 PROB PROBEI
91 0 92 1 93 GND 91 GND 92 E_ID7 93 D_CS0
RELA
PE6 PE6 PROBEID Y_EN RELAY_
94 2 95 GND 96 3 94 _CS1 95 0 96 EN1
PROBE_
PE6 PE6 EXP_PRE PRESEN
97 GND 98 4 99 5 97 SENT 98 GND 99 T
10 PE6 10 PE6 10 10 10 10
0 6 1 7 2 GND 0 GND 1 GND 2 GND
10 10 10 PE6 10 10 10
3 GND 4 GND 5 8 3 VCC 4 VCC 5 VCC

Table A-1 Definition of Input Sockets on Transmitting Board

CON1

PIN SIG PIN SIG PIN SIG PIN SIG

1 TPU58 2 TPU73 61 TPU82 62 TPU90

3 GND 4 GND 63 GND 64 GND

5 TPU50 6 TPU75 65 TPU127 66 TPU128

7 GND 8 GND 67 GND 68 GND

9 TPU79 10 TPU77 69 TPU89 70 TPU92

11 GND 12 GND 71 GND 72 GND

13 TPU71 14 TPU83 73 TPU91 74 TPU123

15 GND 16 GND 75 GND 76 GND

17 TPU49 18 TPU85 77 TPU121 78 TPU122

19 GND 20 GND 79 GND 80 GND

3
Definition of Commonly Used Sockets and Functions in System Maintenance

21 TPU57 22 TPU87 81 TPU124 82 TPU126

23 GND 24 GND 83 GND 84 GND

25 TPU41 26 TPU76 85 TPU119 86 TPU118

27 GND 28 GND 87 GND 88 GND

29 TPU56 30 TPU74 89 TPU116 90 TPU120

31 GND 32 GND 91 GND 92 GND

33 TPU72 34 TPU78 93 TPU112 94 TPU113

35 GND 36 GND 95 GND 96 GND

37 TPU54 38 TPU88 97 TPU105 98 TPU103

39 GND 40 GND 99 GND 100 GND

41 TPU68 42 TPU86 101 TPU114 102 TPU110

43 GND 44 GND 103 GND 104 GND

45 TPU80 46 TPU84 105 TPU104 106 TPU107

47 GND 48 GND 107 GND 108 GND

49 TPU66 50 TPU81 109 TPU108 110 TPU106

51 GND 52 GND 111 GND 112 GND

53 TPU93 54 TPU96 113 TPU101 114 TPU97

55 GND 56 GND 115 TPU100 116 TPU99

57 TPU95 58 TPU94 117 GND 118 GND

59 GND 60 GND 119 TPU98 120 TPU102

CON2

PIN SIG PIN SIG PIN SIG PIN SIG

1 TPU5 2 TPU34 61 TPU16 62 TPU61

3 GND 4 GND 63 GND 64 GND

5 TPU2 6 TPU42 65 TPU40 66 TPU45

7 TPU3 8 TPU46 67 TPU18 68 TPU37

9 GND 10 GND 69 GND 70 GND

11 TPU1 12 TPU25 71 TPU20 72 TPU29

13 TPU4 14 TPU48 73 TPU22 74 TPU70

15 GND 16 GND 75 GND 76 GND

17 TPU8 18 TPU23 77 TPU24 78 TPU69

19 TPU10 20 TPU62 79 TPU26 80 TPU52

21 GND 22 GND 81 GND 82 GND

23 TPU12 24 TPU21 83 TPU59 84 TPU60

4
Definition of Commonly Used Sockets and Functions in System Maintenance

25 TPU14 26 TPU64 85 TPU44 86 TPU51

27 GND 28 GND 87 GND 88 GND

29 TPU11 30 TPU19 89 TPU43 90 TPU125

31 TPU9 32 TPU27 91 TPU36 92 TPU115

33 GND 34 GND 93 GND 94 GND

35 TPU7 36 TPU33 95 TPU35 96 TPU109

37 TPU17 38 TPU65 97 TPU117 98 TPU111

39 GND 40 GND 99 GND 100 GND

41 TPU15 42 TPU55 101 VDD 102 VDD

43 TPU13 44 TPU63 103 GND 104 GND

45 GND 46 GND 105 GND 106 GND

47 TPU6 48 TPU47 107 GND 108 GND

49 TPU28 50 TPU39 109 12V 110 12V

51 GND 52 GND 111 GND 112 GND

53 TPU30 54 TPU31 113 GND 114 GND

55 TPU32 56 TPU67 115 116

57 GND 58 GND 117 PHV 118 PHV

59 TPU38 60 TPU53 119 PHV 120 PHV

Table A-2 Definition of Output Sockets on Transmitting Board

CON1

PIN SIG PIN SIG PIN SIG PIN SIG

1 POUT1 2 POUT2 61 POUT33 62 POUT34

3 GND 4 GND 63 GND 64 GND

5 POUT3 6 POUT4 65 POUT35 66 POUT36

7 POUT5 8 POUT6 67 GND 68 GND

9 GND 10 GND 69 POUT37 70 POUT38

11 POUT7 12 POUT8 71 GND 72 GND

13 POUT9 14 POUT10 73 POUT39 74 POUT40

15 GND 16 GND 75 GND 76 GND

17 POUT11 18 POUT12 77 POUT41 78 POUT42

19 GND 20 GND 79 GND 80 GND

21 POUT13 22 POUT14 81 POUT43 82 POUT44

5
Definition of Commonly Used Sockets and Functions in System Maintenance

23 GND 24 GND 83 GND 84 GND

25 POUT15 26 POUT16 85 POUT45 86 POUT46

27 GND 28 GND 87 GND 88 GND

29 POUT17 30 POUT18 89 POUT47 90 POUT48

31 GND 32 GND 91 GND 92 GND

33 POUT19 34 POUT20 93 POUT49 94 POUT50

35 GND 36 GND 95 GND 96 GND

37 POUT21 38 POUT22 97 POUT51 98 POUT52

39 GND 40 GND 99 GND 100 GND

41 POUT23 42 POUT24 101 POUT53 102 POUT54

43 GND 44 GND 103 GND 104 GND

45 POUT25 46 POUT26 105 POUT55 106 POUT56

47 GND 48 GND 107 GND 108 GND

49 POUT27 50 POUT28 109 POUT57 110 POUT58

51 GND 52 GND 111 GND 112 GND

53 POUT29 54 POUT30 113 POUT59 114 POUT60

55 GND 56 GND 115 POUT61 116 POUT62

57 POUT31 58 POUT32 117 GND 118 GND

59 GND 60 GND 119 POUT63 120 POUT64

CON2

PIN SIG PIN SIG PIN SIG PIN SIG

1 POUT65 2 POUT66 61 POUT97 62 POUT98

3 GND 4 GND 63 GND 64 GND

5 POUT67 6 POUT68 65 POUT99 66 POUT100

7 POUT69 8 POUT70 67 GND 68 GND

9 GND 10 GND 69 POUT101 70 POUT102

11 POUT71 12 POUT72 71 GND 72 GND

13 POUT73 14 POUT74 73 POUT103 74 POUT104

15 GND 16 GND 75 GND 76 GND

17 POUT75 18 POUT76 77 POUT105 78 POUT106

19 GND 20 GND 79 GND 80 GND

21 POUT77 22 POUT78 81 POUT107 82 POUT108

23 GND 24 GND 83 GND 84 GND

25 POUT79 26 POUT80 85 POUT109 86 POUT110

6
Definition of Commonly Used Sockets and Functions in System Maintenance

27 GND 28 GND 87 GND 88 GND

29 POUT81 30 POUT82 89 POUT111 90 POUT112

31 GND 32 GND 91 GND 92 GND

33 POUT83 34 POUT84 93 POUT113 94 POUT114

35 GND 36 GND 95 GND 96 GND

37 POUT85 38 POUT86 97 POUT115 98 POUT116

39 GND 40 GND 99 GND 100 GND

41 POUT87 42 POUT88 101 POUT117 102 POUT118

43 GND 44 GND 103 GND 104 GND

45 POUT89 46 POUT90 105 POUT119 106 POUT120

47 GND 48 GND 107 GND 108 GND

49 POUT91 50 POUT92 109 POUT121 110 POUT122

51 GND 52 GND 111 GND 112 GND

53 POUT93 54 POUT94 113 POUT123 114 POUT124

55 GND 56 GND 115 POUT125 116 POUT126

57 POUT95 58 POUT96 117 GND 118 GND

59 GND 60 GND 119 POUT127 120 POUT128

Table A-3 Pins Definition of CW CON1

SIGNA SIGNA
PIN L PIN L
1 GND 2 GND
3 CW0 4 CW1
5 GND 6 GND
7 CW2 8 CW3
9 GND 10 GND
11 CW4 12 CW5
13 GND 14 GND
15 CW6 16 CW7
17 GND 18 GND
19 CW8 20 CW9
21 GND 22 GND
23 GND 24 GND
25 12V 26 12V
27 GND 28 GND
29 -12V 30 -12V
31 GND 32 GND

7
Definition of Commonly Used Sockets and Functions in System Maintenance

33 5V 34 5V
35 GND 36 GND
37 -5V 38 -5V
39 GND 40 GND

Table A-4 Pins Definition of CW CON2

PIN
NUM Signal
name Note
1 AD_NRST AD reset signal( Alreadyposited to the right position, can be reserved)
2 RESERVED Reserved, not be used temporarily
3 RESERVED Reserved, not be used temporarily
4 RESERVED Reserved, not be used temporarily
5 GND
6 GND
7 LOCLK_P Quadrature demodulation local oscillation input (If it is difference, it is positive end)
8 RESERVED Reserved, not be used temporarily
9 LOCLK_N Quadrature demodulation local oscillation input (If it is difference, it is negative end)
10 RESERVED Reserved, not be used temporarily
11 GND
12 GND
13 RESERVED Reserved, not be used temporarily
14 RESERVED Reserved, not be used temporarily
15 RESERVED Reserved, not be used temporarily
16 AD_HPFD AD internal high-pass filter enable control
(Already posited to normal open, can be reserved)
17 GND
18 GND
19 AD audio serial port left/ right output clock
AD_LRCK (Audio serial port left/ right (or word) clock)
20 AD audio serial port output data
AD_DATA (Audio serial port left and right channel PCM data)
21 AD audio serial port output bit clock
AD_BCK (Audio serial port bit clock)
22 SW_CTRL Analog switch channel selection control signal
23 GND
24 GND
25 TPU Input of pencil probe drive signal
26 RESERVED Reserved, not be used temporarily
27 GND
28 GND
29 RESERVED Reserved, not be used temporarily
30 RESERVED Reserved, not be used temporarily

8
Definition of Commonly Used Sockets and Functions in System Maintenance

31 RESERVED Reserved, not be used temporarily


32 GND
33 GND
34 GND
35 D3V3
36 D3V3
37 GND
38 GND
39 HV Remote control, high voltage
40 HV Remote control, high voltage

Table A-5 Signal Definition between the Master Board and 4D Board

Pin Signal Pin Signal


1 +12V 2 +12V
3 +12V 4 +12V
5 +12V 6 +12V
7 GND 8 GND
9 GND 10 USB-2
11 USB_2_3_OC_N 12 USB+2
13 Gnd 14 Gnd
15 +5V 16 +5V
17 +5V 18 +5V
19 GND 20 GND
21 GND 22 GND
23 +3.3V 24 +3.3V
25 GND 26 GND
27 STEP 28 DIR
29 SLEEP 30 Reserved
31 Reserved 32 Reserved
33 GND 34 GND
35 SPI_CLK ( Reserve SPI Ad 36 SPI_DIN/Rxd
port)/Txd
37 SPI_DOUT 38 SPI_/CS
39 GND 40 GND

Table A-6 Pins Definition between the Master Board and AD Control Board

PIN NUM NET NAME PIN NUM NET NAME


1 GND 2 GND
3 12V 4 12V
5 12V 6 12V
7 12V 8 12V

9
Definition of Commonly Used Sockets and Functions in System Maintenance

9 12V 10 12V
11 GND 12 GND
13 GND 14 GND
15 GND 16 GND
17 GND 18 GND
19 5V 20 5V
21 5V 22 5V
23 3.3V 24 3.3V
25 3.3V 26 3.3V
27 3.3V 28 3.3V
29 3.3V 30 3.3V
31 GND 32 GND
33 GND 34 VGA_SDA
35 GND 36 GND
37 GND 38 VGA_SCL
39 GND 40 GND
41 GND 42 VGA_VSYNC
43 GND 44 GND
45 GND 46 VGA_HSYNC
47 GND 48 GND
49 GND 50 VGA_BLUE
51 GND 52 GND
53 GND 54 VGA_GREEN
55 GND 56 GND
57 GND 58 VGA_RED
59 GND 60 GND

Table A-7 Pins Definition between the Master Board and Power Supply Module

PIN NUM Name PIN NUM Name


1 EDC_Power 2 EDC_Power
3 EDC_Power 4 EDC_Power
5 GND 6 GND
7 GND 8 GND
9 A_Battery+ 10 A_Battery+
11 A_Battery- 12 A_Battery-
13 B_Battery+ 14 B_Battery+
15 B_Battery- 16 B_Battery-
17 Gnd 18 Gnd
19 A_Battery_NTC 20 B_Battery_NTC
21 EDC_Status 22 Power_ON
23 Gnd 24 Scan_Status

10
Definition of Commonly Used Sockets and Functions in System Maintenance

25 Temperature_D+ 26 CW_Mode
27 Temperature_D- 28 +5VStb_CPU_En
29 Gnd 30 Gnd
31 +5VStb 32 +5VStb_CPU
33 +12V 34 +12V
35 Gnd 36 Gnd
37 GND 38 GND
39 -12V 40 -5V
41 Gnd 42 Gnd
43 +5V 44 +5V
45 Gnd 46 Gnd
47 +2.5V 48 +2.5V
49 Gnd 50 Gnd
51 Gnd 52 NC
53 +1.5V 54 +1.5V
55 Gnd 56 Gnd
57 3.3V 58 3.3V
59 Gnd 60 Gnd
61 Gnd 62 Gnd
63 THV 64 THV_Range

Table A-8 Signal Definition between the Master Board and Power Supply Module

No. Name Direction Description Notes


1 +12V In EDC in (external adapter input) 10A
2 +11.1V In Battery in (internal battery input) 9200mA
3 EDC_Status Out To indicate whether the main unit is 5V TTL level
connected with the mains power
supply that high level means they are
connected and normal power supply,
while low level means battery power
supply
4 Scan_Status In Never charge the battery during 5V TTL level
scanning, and low level means
scanning is being performed.
5 CW_mode In Control the output range of remote 5V TTL level
control high voltage, when it is low
level, the output voltage adjustable
range of THV is 20~140V, when it is
high level, the output voltage
adjustable range of THV is 5~10V.
6 Power_ON IN Control the power supply output, 5V TTL level
output in the power supply board
should be normal in the case of low
level.
+5Vstb_CPU_EN IN +5Vstb_CPU output control, normal 5V TTL level
output in +5Vstb_CPU in the case of
low level.

11
Definition of Commonly Used Sockets and Functions in System Maintenance

7 THV_Range IN Control the output range of remote


control high voltage
8 Temperature_D+ Out Control the ambient temperature of
power supply module.
9 Temperature_D- out Control the ambient temperature of
power supply module.
10 A_Battery_NTC IN Control the temperature of battery A.
11 B_Battery_NTC IN Control the temperature of battery B.
12 +12V Out
13 THV Out Remote control high voltage.
14 +5V Out
15 +5VStb Out
16 +5VStb_CPU Out
17 +3.3V Out
18 +2.5A Out
19 +1.5V Out
20 -5V Out
21 -12V Out Exclusive for CW board.
22 NC / No connection.

Table A-9 Signal Definition between Mater Board and Extended Socket

PIN Signal Name PIN Signal Name PIN Signal Name PIN Signal Name
A1 USB+4 B1 USB-4 C1 USB+5 D1 USB-5
A2 VBUS_SYS45 B2 GND C2 VBUS_SYS45 D2 GND
A3 GND B3 VGA_HS C3 VGA_RED D3 VGA_GREEN
A4 UART_RX0 B4 VGA_VS C4 GND D4 GND
A5 UART_TX0 B5 GND C5 VGA_BLUE D5 D0
A6 GND B6 AUDIO_L C6 GND D6 D1
A7 RM_PRINT B7 AUDIO_R C7 D2 D7 GND
A8 RM_BUSY B8 GND C8 D3 D8 AUTOFD_N
A9 GND B9 D7 C9 D4 D9 BUSY
A10 MIC_L B10 PE C10 GND D10 GND
A11 MIC_R B11 STROBE_N C11 D5 D11 ACK_N
A12 VIDEO B12 INIT_N C12 D6 D12 SELIN_N
\ \ B13 SEL \ \ D13 ERROR_N

Table A-10 Signal Definition between Master Board and Keyboard Board

No. Name Direction No. Name Direction


1 +3.3V 2 +3.3V
3 +3.3V 4 +3.3V
5 Gnd 6 Gnd
7 Gnd 8 Gnd
9 +5V 10 +5V

12
Definition of Commonly Used Sockets and Functions in System Maintenance

11 +5V 12 +5V
13 Gnd 14 Gnd
15 Gnd 16 Gnd
17 Gnd 18 Gnd
19 Work_Status_O 20 Work_Status_G
21 Battery_Status_O 22 Battery_Status_G
23 Gnd 24 EDC_Status_G
25 USB_DN 26 LCD_SW
27 USB_DP 28 Power_BTN
29 Gnd 30 Gnd

13
P/N:H-2109-20-76227(V1.0)