Humacount 5D: - User Manual
Humacount 5D: - User Manual
Humacount 5D: - User Manual
| User Manual
SYSTEM VERSION
COPYRIGHT
Copyright 2019, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or
distributed by means of electronic systems, without prior permission of HUMAN in
writing. Since all precautionary measures were taken into account in producing these operating
instructions, the manufacturer accepts no responsibility for any errors or omissions. This
includes any liability for damage that could arise from possible incorrect operation based on this
information. Subject to changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 7
1.1 INTRODUCTION 7
1.2 USER WARRANTY 7
1.3 INTENDED USE OF THE INSTRUMENT 7
1.4 GENERAL SAFETY WARNINGS 8
1.5 DISPOSAL MANAGEMENT CONCEPT 8
1.6 BIOHAZARD WARNING 9
1.7 INSTRUMENT DISINFECTION 9
1.8 WHO SHOULD READ THIS MANUAL OF HUMACOUNT 5D 9
1.9 SPECIAL SYMBOLS USED IN THIS MANUAL AND ON THE ANALYZER 10
2 SYSTEM DESCRIPTION 13
2.1 INSTALLATION 13
2.2 INSTALLATION REQUIREMENTS 13
2.3 UNPACKING 14
2.4 INTRODUCTION 15
2.5 WHO SHOULD READ THIS MANUAL 15
2.6 STRUCTURE OF THE ANALYZER 15
2.6.1 Main unit 16
2.6.2 Touch screen 17
2.6.3 Aspirate key 17
2.6.4 Power/Status indicator 18
2.6.5 Power switch 18
2.6.6 USB interface 19
2.6.7 Network interface 19
2.6.8 External Equipment (Optional) 19
4 SAMPLE ANALYSIS 49
4.1 INTRODUCTION 49
4.2 INTERFACE INTRODUCTION 49
4.3 FUNCTIONS OF THE BUTTONS. 50
4.4 ENTERING SAMPLE INFORMATION 51
4.5 RUNNING SAMPLES 55
4.6 DEALING WITH THE ANALYSIS RESULTS 57
4.6.1 Automatic saving of analysis results 57
4.6.2 Parameter Flags 57
4.6.3 Flags of Abnormal Blood Cell Differential or Morphology 57
4.7 FUNCTIONS OF THE BUTTONS 60
4.7.1 Previous/Next 60
4.7.2 Mode & ID 60
4.7.3 Validate/Cancel Validation 60
4.7.4 Print 60
4.7.5 Patient Information 60
4.7.6 Customized Parameters 64
4.7.7 Microscopic Exam. Parameters 65
4.7.8 Communication 66
4.7.9 Edit Result 66
5 RESULT REVIEW 69
5.1 INTRODUCTION 69
5.2 INTERFACE INTRODUCTION 69
5.3 SAMPLE LIST 70
5.4 FUNCTIONS OF THE BUTTONS 70
5.4.1 Print 70
5.4.2 Delete 70
5.4.3 Export 71
5.4.4 Edit Result 74
5.4.5 Patient Info. 74
5.4.6 Query 79
5.4.7 Results 80
5.4.8 Microscopic Exam. Parameters 81
CONTENTS
6 DAILY OPERATIONS 87
6.1 PRE-OPERATION PREPARATION 87
6.2 STARTUP 88
6.3 DAILY QUALITY CONTROL 89
6.4 SAMPLE COLLECTION AND HANDLING 90
6.4.1 Sample Analysis 90
6.4.2 Interfering substances 90
7 ADVANCED OPERATION 91
7.1 QUALITY CONTROL 91
7.1.1 L-J QC overview 91
7.1.2 QC Settings 91
7.1.3 Quality Control Analysis 96
7.1.4 QC Result Review 102
7.2 X-B QUALITY CONTROL 111
7.2.1 QC Principle 111
7.2.2 QC Settings 112
7.2.3 Quality Control Analysis 115
7.2.4 QC Result Review 115
7.3 SETUP 122
7.3.1 Introduction 122
7.3.2 Interface Introduction 123
7.3.3 System Settings 124
7.3.4 Parameter Settings 128
7.3.5 User management 145
7.3.6 Print Settings 149
7.3.7 Auxiliary Settings 155
7.3.8 Connecting the Analyzer System 157
7.4 RANGES OF PARAMETERS. 165
7.4.1 Reference/Normal ranges 165
7.4.2 Analytical measurement ranges 166
8 CALIBRATION 167
8.1 INTRODUCTION 167
8.2 WHEN TO CALIBRATE 167
8.3 HOW TO CALIBRATE 168
8.3.1 Manual Calibration 168
8.3.2 Auto Calibration Using Calibrators 170
8.3.3 Auto Calibration Using Fresh Blood Samples 173
8.4 VERIFYING CALIBRATION COEFFICIENTS 175
10 SERVICE 181
10.1 INTRODUCTION 181
10.1.1 Cleaning 184
10.1.2 Maintenance 185
10.2 COMPREHENSIVE DEVICE MAINTENANCE 189
10.2.1 Fluidics Initialization 190
10.2.2 Auto Clean 195
10.2.3 Auto Prompt for Cleanser Soak 196
10.2.4 Auto Sleep 196
10.3 SELF-TEST 197
10.3.1 Syringe and Sampling Mechanism 197
10.3.2 Pressure and Vacuum 198
10.3.3 Valve & Pump 199
10.3.4 Others 199
10.4 SYSTEM STATUS 200
10.4.1 Temperature 201
10.4.2 Voltage and Current 201
10.4.3 Disk Information 202
10.5 LOG 202
10.5.1 All Logs 203
10.5.2 Parameter Revision Logs 203
10.5.3 Fault Logs 204
10.5.4 Other Logs 205
CONTENTS
11 TROUBLESHOOTING 211
11.1 INTRODUCTION 211
11.2 DEALING WITH ERROR MESSAGES 211
11.3 ERROR MESSAGE REFERENCE 212
12 APPENDIX 223
Safety Instructions 7
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
Figure 1
Biological Hazard Symbol
!
Follow the instruction in the manual related to the
Note symbol. The symbol highlight the important infor-
mation in operating procedures that calls for special
attention.
Puncture Warning:
The sampling probe is sharp and may contain biohaz-
ardous materials. Special care should be taken when
working with it.
Laser Warning:
This sign serves as a reminder of laser radiation.
Caution
Biohazard
Network interface
When you
It means...
see...
Protective grounding
Lot No.
Expiry date
Serial No.
Date of manufacture
Manufacturer
Storage temperature
Avoid sunlight
Keep dry
No rolling
No stacking
Recyclable materials
2 SYSTEM DESCRIPTION
2.1 Installation
The analyzer should only be installed by Human or its authorized agents. You
need to provide the appropriate environment and space. When the analyzer
needs to be relocated, please contact Human or your local agents. When you
receive the analyzer, please notify Human or your local agent immediately.
WARNING
-- Connect only to a properly grounded outlet.
-- Before turning on the analyzer, make sure the input voltage
meets the requirements.
CAUTION
-- Using a patch board may introduce electrical interference and
generate incorrect analysis results. Please place the analyzer
near the electrical outlet to avoid using the patch board.
-- Please use the original electrical wires shipped with the ana-
lyzer. Using other electrical wires may damage the analyzer or
generate incorrect analysis results.
2.3 Unpacking
Please unpack the analyzer by taking the following steps:
1. Open the outer packing box; take out the accessory pack; take out the ana-
lyzer together with the protective and cushioning materials.
2. Remove the foam and the protective PE bag.
3. Open the right door (open the linear-shaped cam lock on the right door with
a slotted screwdriver).
4. Remove the binder clips, which are used for fixating two conveyor belts. To
avoid the possible collision resulting from the slippage caused by shaking
and slanting during transportation, the central position of those two belts is
fixated with binder clips before they are shipped from the factory. The binder
clips must be removed during unpacking.
5. Remove the binder clips, which are used for fixating sampling assembly. To
avoid damage during the transportation, the sampling assembly of the ana-
lyzer is fixated with clamps. Do remove the clamps before using the analyzer.
2.4 Introduction
The HumaCount 5D is a quantitative, automated hematology analyzer and
5-part differential counter used in clinical laboratories. This section describes
in details the intended use, measurement parameters, structure, user interface
and compatible reagents of the analyzer.
WARNING
-- Please check the firmness of all the doors, covers and boards
before running the analyzer.
-- The analyzer is heavy, so moving by one person alone may
cause injury. It is advisable for two people to move it together
when the transportation is necessary, and make sure you fol-
low the instructions and use the proper tools.
-- Connect only to a properly grounded outlet.
-- To avoid electrical shocks, disconnect the power supply before
opening the cover.
-- To prevent fire, use the fuses with specified model number
and working current. The sampling probe is sharp and may
contain biohazardous materials. Special care should be taken
when working with it. This sign warns of laser radiation. Do
not look directly at the laser beams or see through the optical
instrument.
-- The sampling probe is sharp and may contain biohazardous
materials. Special care should be taken when working with it.
Figure 2
Front of the analyzer
1 Touch screen 1
2 Power/Status indicator
3 Sample probe
4 Aspirate key
Figure 3
Back of the analyzer
3
1 Power switch
2 AC input
3 Cooling fan
4 Waste outlet
5 Diluent inlet
6 Waste level detection con-
nector
4 2
7 Diluent presence detection
5 1
connector
8 Ground studs
6 7 8
Figure 4
Side views of the analyzer
2
1 Network interface
1
2 USB interface
3 3 Right side door buckle
4 Left side small door buckle
4
Table 2
Instrument Status Indicator Status Remarks
Shutdown Off The main unit has been shut
Main Unit Status Indicators
down.
Stopped running with Red light on Stopped running with the
error conditions occurrence of errors
Running with error Red light flickering Running with the occur-
conditions rence of errors
Time sequence Yellow light on Initialization or sleep status
deactivated irrelevant to running
Running Green light flickering Execution of the sequence
actions is in process.
Ready Green light on Execution of the sequence
actions is allowed.
While the analyzer is running, if the status indicator turns dim or off, please
contact Human or Human’s agent for maintenance.
CAUTION
-- To avoid damage, do not power on/off the analyzer repetitively
within a short time.
-- A power switch is located in the bottom back of the analyzer.
It turns on or shuts down the analyzer. Please use shut down
button of the software to turn off the analyzer.
2.7 Accessories
Waste Container 1
3.1.1 INTRODUCTION
The measurement methods used in this analyzer are: the electrical Impedance
method for determining the RBC and PLT related parameters; the colorimetric
method for determining the HGB; laser-based flow cytometry for determining
the WBC related parameters. During each analysis cycle, the sample is aspirated,
diluted and mixed before the determination for each parameter is performed.
3.1.2 ASPIRATION
The analyzer supports Whole Blood mode (including Venous Whole Blood and
Capillary Whole Blood) and Predilute mode. In Whole Blood mode, the analyzer
will aspirate quantitative whole blood sample. In Predilute mode, the analyzer
will aspirate the prediluted sample (with the dilution ratio of 1:25) which is a
mixture of 20 μl of whole blood/capillary blood sample and 480 μl of diluent
the diluted sample thus prepared is then delivered to the analyzer for sampling
and aspiration.
3.1.3 DILUTION
After being aspirated into the analyzer, the sample is divided into two parts. Af-
ter the reaction with reagents in parallel dilution procedures, each part forms
the sample for red blood cell/platelet, white blood cell count/hemoglobin
! Note: CBC mode, namely
complete blood cell count, is
intended for counting only, not
measurement and white blood cell differential measurement. To meet different for white blood cell classification.
needs, the analyzer offers two working modes (Whole Blood and Predilute), and CBC+DIFF mode is intended for
two measurement modes (CBC and CBC+DIFF). Taking CBC+DIFF mode as an both counting and white blood
example, this section introduces the dilution procedures of the test sample in cell classification.
Whole Blood mode and Predilute mode separately. (The dilution procedure in
CBC mode is not introduced here since it’s the same as that in CBC+DIFF mode.)
22
▼
Whole-blood CBC+DIFF Mode
the samples
▼ ▼
Sampling Sampling
HC5D-Diluent
▼
HC5D-Diluent
▼
HC5D-Diff Lyse
▼
▼ ▼
Prepare WBC diff Dilute the sample
1 samples with certain
dilution ratios ▼
Take the sample that
has been diluted once
HC5D CBC Lyse from the WBC bath
▼
HC5D-Diluent
▼
▼ ▼
Prepare WBC samples Prepare RBC samples
3 and HGB samples with and PLT samples with 2
certain dilution ratios certain dilution ratios
20 ul of blood sample
▼ ▼
Sampling Sampling
Lyse Diluent
▼ ▼
Prepare WBC diff Dilute the sample
1 samples with certain
dilution ratios ▼
Take the sample that
has been diluted once
HC5D CBC from the WBC bath
▼
Lyse
HC5D-
▼
Diluent
▼ ▼
Prepare WBC samples Prepare RBC samples
3 and HGB samples with and PLT samples with 2
certain dilution ratios certain dilution ratios
3.1.5 FLUSHING
After each analysis cycle, each component of the analyzer is flushed.
Figure 7
WBC Measurement
The intensity of scattered light reflects the blood cell size and intracellular densi-
ty. The low-angle scattered light reflects cell size, while the high-angle scattered
light reflects intracellular density (nucleus size and density). The optical detec-
tor receives this scattered light and converts it into electrical pulses. Pulse data
thus collected can be used to draw a 3-dimensional distribution (scattergram)
as shown in Figure 8.
Figure 8
DIFF channel scattergram
BAS
Conduct dual channel detection to the white blood cells (WBCs). Use three-an-
gle laser scattering and flow cytometry for the count and classification of vari-
ous kinds of WBCs in dual channels. The abbreviations for the low angle scatter
channel is: LS. MS means Medium angel scatter channel and HS stands for High
angle scatter channel. By analysing the DIFF channel scattergram, the analyzer
presents the Lym%, Mon%, Eos% and Neu%. The independent WBC/Baso chan-
nel shall use a specific kind of hemolytic agent that can extract the Baso cell
specificity, so as to reserve the complete information of Baso cells.
Conduct precise and reliable WBC/Baso cell counting combined with three-an-
gle laser scattering and flow cytometry. Monocytes are after the lyse process
bigger than the lymphocytes; therefore you see them on the LS axis shown on
top of the lymphocytes. or the EOS are more complex than the Lym and there-
fore you see them on the HS-axis further right. The size of ALY is bigger than
LYM but smaller than MON. So ALY area is measured between LYM and MON.
The size of LIC is much bigger than NEU, so it will have much stronger LS signal.
It will always cluster in the area which deviated from the NEU cluster area.
obtained with the working principle of laser-based flow cytometry. The unit of
the number of cells is 109/l.
• White Blood Cell count
WBC count is the number of leukocytes measured directly by counting the leucocytes passing through
the flow chamber
• Number of Basophils (Bas#)
Bas# is the number of Basophils measured directly by counting the basophils passing through the flow
chamber
3.1.8 HGB
The HGB is calculated using the following equation and expressed in g/l.
Blank Photocurrent
HGB (g/l) = Constant x Ln ( Sample Photocurrent )
Figure 9
Constant current source
Electrical Impedance method
Count bath
Front
bath
Aperture
Each pulse is amplified and compared to the internal reference voltage channel,
which only accepts the pulses of a certain amplitude. If the pulse generated is
above the RBC/PLT lower threshold value, it is counted as a RBC/PLT. The analyz-
er presents the RBC/PLT histogram, where the x-coordinate represents the cell
volume (fl) and the y-coordinate represents the number of the cells.
3.1.9.2 RBC
-- Red Blood Cell count
RBC (1012/l) is the number of erythrocytes measured directly by counting the
erythrocytes passing through the aperture.
-- Mean Corpuscular Volume (MCV)
Based on the RBC histogram, this analyzer calculates the MCV and expresses
the result in fl.
-- Hematocrit (HCT), Mean Corpuscular Hemoglobin (MCH), Mean Corpuscu-
lar Hemoglobin Concentration (MCHC)
This analyzer calculates the HCT (%), MCH (pg) and MCHC (g/l) as follows,
where the RBC is expressed in 1012/l, MCV in fl and HGB in g/l.
RBC x MCV
HCT =
10
HGB
MCH =
RBC
HGB
MCHC = x 100
HCT
3.1.9.3 PLT
-- Platelet count
PLT is measured directly by counting the platelets passing through the ap-
erture.
-- Mean Platelet Volume (MPV, fl)
Based on the PLT histogram, this analyzer calculates the MPV.
-- Platelet Distribution Width (PDW) or PDWcv
The PDW is the distribution width of the platelets. It is determined at the
20% level of peak height of the histogram
-- Plateletcrit (PCT)
This analyzer calculates the PCT as follows and expresses it in %, where the
PLT is expressed in 109/l and the MPV in fl.
PLT x MPV
PCT =
10000
30
Table 3
Type Parameter Name Abbreviation CBC CBC+DIFF
WBC White Blood Cell count WBC * *
(15 items) Percentage of Neutrophils Neu% / *
Percentage of Lymphocytes Lym% / *
Percentage of Monocytes Mon% / *
Percentage of Eosinophils Eos% / *
Percentage of Basophils Bas% / *
Number of Neutrophils Neu# / *
Number of Lymphocytes Lym# / *
Number of Monocytes Mon# / *
Number of Eosinophils Eos# / *
Number of Basophils Bas# / *
Percentage of Abnormal ALY% (RUO) / *
Lymphocytes
Percentage of Large LIC% (RUO) / *
Immature Cells
Number of Abnormal ALY# (RUO) / *
Lymphocytes
Number of Large LIC# (RUO) / *
Immature Cells
RBC Red Blood Cell count RBC * *
(8 items) Hemoglobin Concentration HGB * *
Mean Corpuscular Volume MCV * *
Mean Corpuscular Hemoglobin MCH * *
Mean Corpuscular MCHC * *
Hemoglobin Concentration
Red Blood Cell Distribution RDW-CV * *
Width - Coefficient of Variation
Red Blood Cell Distribution RDW-SD * *
Width - Standard Deviation
Hematocrit HCT * *
! -- “*” means the parameter is provided in the mode. “/” means the pa-
rameter is not provided.
-- ALY%, LIC%, ALY#, P-LCC, P-CCR and LIC# are parameters for research
use only (RUO), not for diagnostic use.
The interface can be divided into several areas as follows according to their func-
tions:
-- Menu navigation area
On the top of the screen is the menu navigation area. Once a menu button is
pressed, the system goes immediately to the corresponding screen.
-- Menu content display area
It displays the selected screen and the corresponding function buttons.
-- Error message area
Upon the occurrence of a system failure, the corresponding error message
will appear in this area. When there is more than one failure, the error mes-
sage for the latest failure will appear in this area. Click in this area, you can
deal with the failures in the pop-up dialogue box of troubleshooting help.
For more information, see chapter 11 Troubleshooting.
-- Status display area
On the lower left of the screen is the status display area where the connec-
tion status between the analyzer and the LIS system and printer status are
displayed from left to right. The icons change with the status of the main
unit, as shown in Table 4.
Table 4
Status Icon Remarks
LIS/HIS Grey icon The computer is not connected to the LIS/
Status Icon Description
status HIS.
Black icon The computer is connected to the LIS/HIS.
Print status Grey icon The external printer is not connected to the
analyzer yet.
Colour icon The external printer is connected to the
analyzer.
Ingredients:
Sodium Chloride, Sodium Sulfate, Buffering agent; Antifungal and antibac-
terial agents.
Applicable Analyzer:
Applicable to the HumaCount 5D hematology analyzers.
How to Use:
Restore HC5D-Diluent to the use temperature, unpack the reagent, insert
the corresponding catheter to the diluent barrel according to the colour cor-
respondence between the reagent bottle cap and the analyzer bottle cap
assembly joint, screw the bottle cap assembly, replace the reagent according
to the operator's manual of the analyzer, and then conduct the blank deter-
mination to make sure that the measured value is within the blank counting
range required in the operator's manual of the analyzer before sample test-
ing. Refer to the operator's manual of the analyzer for further information.
34
Specifications:
The product should be clear liquid without particles, sediment and flocs.
Background result: WBC≤0.2x109/l
RBC≤0.2x1012/l
HGB≤1g/l
PLT≤10x109l
Precautions:
• Do not inhale. In case of inhaling, immediately go see a doctor. Avoid skin
or eye contact. In case of skin contact, flush the affected area with plenty
of water immediately. In case of eye contact, flush the affected area with
plenty of water immediately and go see a doctor.
• Use HC5D-Diluent before the expiration date.
• If product is frozen (either partially or completely), thaw completely,
warm to room temperature and mix thoroughly by gentle inversion. Ver-
ify background results before analysing patient samples.
• Dispose of reagents, waste and consumables according to government
regulations.
• Test result may be unreliable if:
Reagent is expired or invalid
Reagent is polluted with dust from the air
Sample is not processed cleanly
Reagent is mixed or used with third party reagents.
Ingredients:
Surface active agent; Buffering agent; Antifungal and antibacterial agents.
Applicable Analyzer:
Applicable to the HumaCount 5D hematology analyzers.
How to Use:
Restore HC5D CBC Lyse to the use temperature, unpack the reagent, insert
the corresponding catheter to the diluent barrel according to the colour cor-
respondence between the reagent bottle cap and the analyzer bottle cap as-
sembly joint, screw the bottle cap assembly, replace the reagent according
to the operator's manual of the analyzer, and then conduct the blank deter-
mination to make sure that the measured value is within the blank counting
range required in the operator's manual of the analyzer before sample test-
ing. Refer to the operator's manual of the analyzer for further information.
Specifications:
Ingredients:
Surface active agent; Buffering agent; Antifungal and antibacterial agents.
Applicable Analyzer:
Applicable to the HumaCount 5D hematology analyzers.
How to Use:
Restore HC5D Diff Lyse to the use temperature, unpack the reagent, insert
the corresponding catheter to the diluent barrel according to the colour cor-
respondence between the reagent bottle cap and the analyzer bottle cap as-
sembly joint, screw the bottle cap assembly, replace the reagent according
to the operator's manual of the analyzer, and then conduct the blank deter-
mination to make sure that the measured value is within the blank counting
range required in the operator's manual of the analyzer before sample test-
ing. Refer to the operator's manual of the analyzer for further information.
Specifications:
The product should be clear liquid without particles, precipitate or floccule.
Background result: WBC≤0.2x109/l
Precautions:
• Do not inhale. In case of inhaling, immediately go see a doctor. Avoid skin
or eye contact. In case of skin contact, flush the affected area with plenty
of water immediately. In case of eye contact, flush the affected area with
plenty of water immediately and go see a doctor.
• Use HC5D Diff Lyse before the expiration date.
• If product is frozen (either partially or completely), thaw completely,
warm to room temperature and mix thoroughly by gentle inversion. Ver-
ify background results before analysing patient samples.
• Dispose of reagents, waste and consumables according to government
regulations.
• Test result may be unreliable if:
Reagent is expired or invalid
Reagent is polluted with dust from the air
Sample is not processed cleanly
Reagent is mixed or used with third party reagents.
Ingredients:
Surface active agent, Sodium Hydroxide, Sodium Hypochlorite
Applicable Analyzer:
Applicable to the HumaCount 5D hematology analyzers.
38
How to Use:
Follow the instructions on the analyzer operating screen and set the un-
capped cleanser to the sample probe when required. Then press the aspirate
key to start cleanser maintenance. Refer to the operator's manual for further
information.
Specifications:
The product should be clear liquid without particles, sediment and flocs.
Background result: WBC≤0.2x109/l
RBC≤0.2x1012/l
HGB≤1g/l
PLT≤10x109/l
Precautions:
• Do not inhale. In case of inhaling, immediately go see a doctor. Avoid skin
or eye contact. In case of skin contact, flush the affected area with plenty
of water immediately. In case of eye contact, flush the affected area with
plenty of water immediately and go see a doctor.
• Use HC5D Clean before the expiration date.
• If product is frozen (either partially or completely), thaw completely,
warm to room temperature and mix thoroughly by gentle inversion. Ver-
ify background results before analysing patient samples.
• Dispose of reagents, waste and consumables according to government
regulations.
• Test result may be unreliable if:
Reagent is expired or invalid
Reagent is polluted with dust from the air
Sample is not processed cleanly
Reagent is mixed or used with third party reagents.
-- Performance characteristics
Assigned values of control and calibrator are presented as a Mean and
Range. The Mean is derived from replicate testing on instruments operated
and maintained according to the manufacturer’s instructions. The Range is
an estimate of variation between laboratories and also takes into account
inherent imprecision of the method and expected biological variability of
the control material. Assay values on a new lot of control should be con-
firmed before the new lot is put into routine use. Test the new lot when the
instrument is in good working order and quality control results on the old
lot are acceptable. The laboratory’s recovered mean should be within the
assay range. For greater control sensitivity each laboratory should establish
its own mean and acceptable range and periodically reevaluate the mean.
The laboratory range may include values outside of the assay range. The
user may establish assay values not listed on the Assay Sheet, if the control
is suitable for the method.
Hazard statements
The following hazard statements are valid for the respective REF numbers.
[REF] 16450/20 HC5D CBC Lyse
H411 Toxic to aquatic life with long lasting effects.
[REF] 16450/60 HC5D Clean
H314 Causes severe skin burns and eye damage.
Precautionary statements
The following precautionary statements apply to the following REF numbers
[REF] 16450/10 HC5D Diluent
[REF] 16450/20 HC5D CBC Lyse
[REF] 16450/30 HC5D Diff Lyse
[REF] 16450/40 HC5D Control
CAUTION
Do not turn on the analyzer immediately after its shutdown. Wait
at least 10 seconds before power-on to avoid damage to the ma-
chine.
Figure 11
2. Click Yes.
The system starts to execute the shutdown sequence and a message box
pops up showing the procedures for cleanser maintenance.
3. Follow the instructions and set the cleanser under the sample probe and
press the aspirate key on the analyzer or click Aspirate to run the cleanser
aspiration. Upon the completion of cleanser maintenance, you’ll be prompt-
ed that the cleanser maintenance is completed.
4. Place the [O/I] switch at the back of the main unit in the [O] position.
5. After shutdown, empty the waste in the waste container, and dispose it.
WARNING
Be sure to dispose reagents, waste, samples, consumables, etc. ac-
cording to local legislations and regulations.
44
WARNING
Do not touch the patients’ blood sample directly.
! Note:
-- Do not re-use such disposable products as collection tubes, test tubes,
capillary tubes, etc.
-- Prepare the samples as per the procedures recommended by the rea-
gent manufacturer.
-- Be sure to use clean K2EDTA vacutainer blood collection tubes with
anticoagulant, fused silica glass/plastic test tubes, centrifugal tubes
and borosilicate glass capillary tubes.
-- Be sure to use the Human-specified disposable products including
vacutainer blood collection tube, vacutainer blood collection tubes
with anticoagulant and capillary tubes etc.
-- For the whole blood samples to be used for WBC classification or PLT
count, store them at room temperature and run them within 8 hours
after collection.
-- If you do not need the PLT, MCV and WBC differential results, you can
store the samples in a refrigerator (2°C - 8°C) for 24 hours. You need to
warm the keep samples at room temperature for at least 30 minutes
before running them.
-- Be sure to mix any sample that has been prepared for a while before
running it.
CAUTION
Never put pressure on the skintissue next to the capillary collec-
tion tissue to avoid dilution of blood by tissue liquids.
46
Figure 12
Menu Screen
3. Take a clean centrifugal tube, uncap it and present it to the sample probe
in a manner as shown in the following picture in which the probe tip is ver-
tically in contact with the bottom of the tube so as to avoid bubbles, liquid
attached to the inner wall or spatter.
4. Press the aspirate key and add the HC5D diluent (480 μl at a time) automati-
cally by the analyzer. After the diluent is added and you hear a beep, you can
remove the centrifugal tube.
5. If more portions of diluent are needed, repeat steps 3~4.
6. Add 20 μl of blood to the diluent, close the tube cap and shake the tube to
mix the sample.
7. After the prediluted sample is prepared, click Cancel to exit dispensing the
diluent.
-- Ideally you use the HUMAN capillary (REF: 16070/30) and put it together
with the blood inside the centrifuged tube. Close the tube and mix careful-
ly until all blood is mashed out the capillary and dissolved homogeneously.
Leave capillary inside.
-- You can also dispense 480 μl of diluent by pipette into the tube.
-- The prediluted sample prepared after single blood collection can be counted
twice.
-- Be sure to keep dust from the prepared diluent.
-- Be sure to run the prediluted samples within 30 minutes after the mixing.
-- Be sure to mix any sample that has been prepared for a while before running it.
-- Be sure to evaluate predilute stability based on your laboratory’s sample
population and sample collection techniques or methods.
48
-- The centrifugal tube shall be placed vertically upward, not tilted or upside
down. Otherwise, the inner wall of the tube would be stained with excessive
sample, resulting in waste. Moreover, it may cause unevenly mixed sample
and unreliable analysis results.
Figure 14
4 SAMPLE ANALYSIS
4.1 Introduction
Sample analysis is the most important function of the auto hematology analyz-
er. You can get the blood cell count, HGB concentration and the 5-part classifica-
tion counting results of the white blood cells by performing the sample analysis.
The summary of sample analysis procedures are as follows:
1 Figure 15
Sample analysis interface
2
1 Function buttons
2 Patient information area
3 Analysis results area
3 4 Information of the next
sample
4
50
Related descriptions:
-- Function buttons
You can perform operations such as setting the mode for the samples,
pre-entering information, reviewing previous/next records and printing.
Click and view all function buttons.
! You can also enter sample/patient information after the sample analysis
is completed. For details, please refer to 8 Result Review. Detailed steps
are shown below:
1. Click the Edit Patient button in the function button area. The interface as
shown in Figure 16 will pop up on the screen.
Figure 16
Pre-entering Patient Information
!
Sampling Time Date and time when the Click the date control for the Note: The system automati-
sample is collected. settings. cally displays the current time
-- The input sequence of the as sampling time. The sampling
controls is the same with the time can be no later than the cur-
date format on the top right
rent system time.
corner of the dialogue box.
For example, if the data for-
mat is yyyy/MM/dd HH:mm,
you should input the data in
the sequence of year, month,
date, hour, and minute.
-- Click or to select
the date or click the textbox
to enter them directly.
-- Click to clear the current
data and re-enter the infor-
mation.
Submitter Personnel submitting the Select from the dropdown list or
sample. input directly.
54
Delivery Time Date and time when the Click the date control for the
! Note: The system automati-
cally displays the current time
as sample delivery time. The de-
sample is delivered. settings.
-- The input sequence of the
controls is the same with the
livery time can be no later than
date format on the top right
the current system time and can-
corner of the dialogue box.
not be earlier than the sampling For example, if the data for-
time. mat is yyyy/MM/dd HH:mm,
you should input the data in
the sequence of year, month,
date, hour, and minute.
-- Click or to select
the date or click the textbox
to enter them directly.
-- Click to clear the current
data and re-enter the infor-
mation.
! -- The tube (or centrifugal tube) shall be placed vertically upward, not
tilted or upside down. Otherwise, the inner wall of the tube may be
stained with excessive sample, resulting in waste. Moreover, it may
cause unevenly mixed sample and unreliable analysis results.
-- During aspiration, the tip of the probe should be kept at a certain dis-
tance from the bottom of the sample container, otherwise the accura-
cy of aspiration volume will be affected.
-- Keep the tip of the probe from contacting with the wall of the test
tube to avoid blood splashing.
-- Proper reference range shall be selected on the Setup interface before
analysis. Otherwise, the results may be flagged erroneously.
-- The default system setting for counting mode is Venous Whole Blood
(VWB)-CBC+DIFF.
-- When the analyzer is running the samples, you can switch to Review
interface to perform operations including browsing and exporting,
etc., and you can also switch to other interfaces. But all the functions
related to the fluidics sequence are not available. Take the following
steps to perform sample analysis.
Figure 17
Mode & ID Settings
6. Select the blood sample mode Venous Whole Blood (VWB), Capillary Whole
Blood (CWB) or Predilute (PD) of the sample.
7. Select the measurement mode CBC or CBC+DIFF according to the actual test
case, and enter the Sample ID.
Table 5
Parameter It means Operation
CBC Complete Blood Count with no differ- Selected from the
Sample Analysis
ential count for white blood cells. The radio box.
Parameter Descriptions
counting results comprise 13 parame-
ters, 3 histograms (including WBC, RBC
through the keyboard (including CBC+DIFF Complete Blood Count plus differential Selected from the
special characters) are allowed for count for white blood cells. The count- radio box.
the Sample ID. Chinese and other ing results comprise 23 measurement
languages (such as Japanese, Ko- parameters, 4 RUO parameters, one DIFF
rean, etc) are not supported. scattergram, one BASO scattergram, and
three histograms (including WBC, RBC
! The length of the entries rang-
es from 1 to 25 and the entries
shall not be empty.
and PLT).
11. When you hear a beep, remove the sample tube. The analyzer will automat-
ically run the sample and the analysis status icon and analyzer indicator is
flickering in green. When the analysis is complete, the analyzer indicator re-
turns to constantly-on green.
12. Repeat steps 1~9 to run the remaining samples.
Table 6
Flag Type Flag information
WBC Abnormal Leucocytosis
Flags of abnormal blood cell
Leucopenia
differential or morphology
Neutrophilia
Neutropenia
Lymphocytosis
Lymphopenia
Monocytosis
Eosinophilia
Basophilia
Suspicious WBC abnormal
Abnor. WBC scattergram
Abnor. WBC histogram
Left Shift?
Immature Cell?
RBC Lyse Resistant?
Abn./Atypical Lym?
Abnormal WBC Channel
Abnormal DIFF Channel
RBC/HGB Abnormal Erythrocytosis
Anisocytosis
Macrocytosis
Microcytosis
Anemia
Hypochromia
Suspicious Abnor. RBC Distr.
Dimorphologic
Iron Deficiency?
HGB Abnor./Interfere?
RBC Clump?
Abnormal RBC Channel
Abnormal HGB Channel
PLT Abnormal Thrombocytosis
Thrombopenia
Suspicious Abnor. PLT Distr.
PLT Clump?
The system shows flags for abnormal or suspicious items in different samples
and measurement modes in accordance with the impact of the abnormal or sus-
picious WBC, RBC or PLT items on the results of the parameters. The correlation
is shown in the following table:
-- “” indicates that flags will be displayed in the mode.”” indicates that flags
will not be displayed in the mode.
-- When the PLT value is less than 100×109/l, a manual count by the micro-
scope is recommended.
60
4.7.1 PREVIOUS/NEXT
Click Previous, and the screen will display the sample analysis results prior to
the current one.
Click Next, and the screen will display the sample analysis results after the cur-
rent one.
4.7.4 PRINT
You can click Print to print the report of the sample result.
1. Click Edit Patient to enter the patient information setting interface as shown
in Figure 18.
Figure 18
Patient Information
Sampling Date and time when Click the date control for the settings.
Time the sample is collect- -- The input sequence of the controls
ed. is the same with the date format on
the top right corner of the dialogue
box. For example, if the data format
is yyyy/MM/dd HH:mm, you should
input the data in the sequence of
year, month, date, hour, and min-
ute.
-- Click or to select a date
and time or enter the information
in the textbox directly.
-- Click to clear the current data
and re-enter the information.
Note:
The sampling time can be no later than
the current system time.
Submitter Personnel submitting Select from the dropdown list or input
the sample. directly.
Mode & ID Counting mode of You do not need to enter it and it will
the selected sample. be displayed automatically.
The format is blood
sample mode-meas-
urement mode.
Delivery Date and time when Click the date control for the settings.
Time the sample is deliv- -- The input sequence of the controls
ered. is the same with the date format
on the top right corner of the dia-
logue box. For example, if the data
format is yyyy/MM/dd HH:mm,
you should input the data in the se-
quence of year, month, date, hour,
and minute.
-- Click or to select a date
and time or enter the information
in the textbox directly.
-- Click to clear the current data
and re-enter the information.
Note:
The delivery time can be no later than
the current system time and cannot be
earlier than the sampling time.
Operator Personnel running the You do not need to enter it and it will
sample. be displayed automatically.
64
Run Time Time when the sam- You do not need to enter it and it will
ple is run. be displayed automatically.
Approver Personnel validating This parameter will be automatically
the sample. displayed after the sample is validated.
Report Time The date and time when This parameter will be automatically
the report is printed displayed after the report is printed.
for the first time.
Diagnosis Suspected diagnosis Input in the textbox directly.
information.
Remarks Clarifications or notes. Input in the textbox directly.
3. Click Apply or OK to save the settings.
Figure 19
Customized Parameters
2. Click the cell corresponding to its Value column of the parameter, and en-
ter the value. If the unit and reference range of parameters have been set
in the Setup > Parameter >Custom Para. interface, the corresponding unit
and range (lower limit~upper limit) will be displayed in this tab. When both
the value and range of parameters are numbers, and the number is out of
the reference range, the relevant mark or will be displayed in the Flag
column. Please refer to chapter 4.7.6 Customized Parameters for customized
parameters settings.
Figure 20
Adding a New Microscopic
Exam. Parameter
Refer to Table 9 for parameter description and operation methods regarding the
microscopic examination.
4.7.8 COMMUNICATION
You can transmit the current sample data (except the background sample) to
the LIS/HIS system in the Sample Analysis interface.
You can edit the parameter result of the selected sample as per the following
steps.
The Edit Result dialogue box will pop up on the screen as shown in Figure 21.
Figure 21
Editing Parameter Result
! The result of the parameter that you modified manually will be flagged
with an M. If any parameter result is then changed due to the one that
you modified manually, it will be flagged with an m.
! --
--
1.
Validated samples are not allowed to be deleted.
The common user has no access to delete the sample records.
Click to unfold all function buttons.
2. Click Delete, and then click Yes in the pop-up dialogue box to delete
the sample:
Figure 22
Delete Sample Records
68
5 RESULT REVIEW
5.1 Introduction
Upon the completion of each sample analysis, the analyzer will automatically
save the sample information, result data, flag messages, histograms and scat-
tergrams to the Review Database. In the Review Interface, you can browse the
saved sample information, result data, flag messages, histograms and scatter-
grams, and can search, compare or export the saved sample information.
1 Figure 23
Review
1 Function buttons
2 Result list
3 Direction button
2
4 Current page/Total pages
5 Sequence Number/ Total
number of results
Interface Description:
-- Result list: You can browse detailed sample records.
-- Function buttons: You can perform the operations such as comparing or
searching the sample results, deleting and viewing the Run Charts, export-
ing and printing reports.
70
-- Direction button: If you click different direction buttons, the list will move
toward the corresponding directions.
• From left to right, it indicates in sequence: The first column, moving to
the left page, moving to the right page, and the last column.
• From top to bottom, it indicates in sequence: The first page, the previous
page, the next page, and the last page.
5.4.1 PRINT
Click Print to print the result report of the selected sample.
5.4.2 DELETE
-- Validated samples are not allowed to be deleted.
-- The common user has no access to delete the sample records.
1. Select one or several sample records to be deleted.
2. Click Delete.
A prompt box will pop up on the screen as shown below.
Figure 24
Delete Sample Records
5.4.3 EXPORT
The operator can export the sample data to the USB flash disk for backup. There
are two ways of exporting the sample data: Exporting selected records and ex-
porting records of specified dates.
-- Export Selected Records
• Insert a USB flash disk in the USB interface on the analyzer.
• Select records to be backed up, and click Export.
As shown in the following figure, the export range of the system is
Selected Records by default.
Figure 25
Export Selected Records
• Select the data export path in the popup dialogue box, enter the backup
file name, and click Save.
• The file will be exported to the root directory of the USB flash disk
(/udisk/sda1) and named in the format of SampleInfo_yyyyMMdd_
hhmmss.csv. Among which, yyyyMMdd_hhmmss means data export
year, month, date, hour, minute, and second.
Figure 26
-- Click Save.
The system pops up a dialogue box as shown below to indicate that the data
export is successful.
Figure 27
Figure 28
Export Records of the
Specified Dates
Figure 29
74
1. Select a row of record from the result list and click the Edit Result button.
The Edit Result dialogue box will pop up on the screen as shown in Figure 30.
Figure 30
Editing Parameter Result
Figure 31
Patient Info
!
Note: The bed No. is required to
be filled only for inpatients.
Sampling Time Date and time when Click the date control for the set-
the sample is collect- tings.
ed. -- The input sequence of the con-
trols is the same with the date
format on the top right corner of
the dialogue box. For example,
if the data format is yyyy/MM/
dd, you should input the data in
the sequence of year, month, and
date.
-- Click or to select a
date and time or enter the infor-
mation in the textbox directly.
-- Click to clear the current data
and re-enter the information.
Delivery Time Date and time when Click the date control for the set-
the sample is deliv- tings.
ered. -- The input sequence of the con-
trols is the same with the date
format on the top right corner of
the dialogue box. For example,
if the data format is yyyy/MM/
dd, you should input the data in
the sequence of year, month, and
date.
-- Click or to select a
date and time or enter the infor-
mation in the textbox directly.
-- Click to clear the current data
and re-enter the information.
! Note:
The delivery time can be no later
than the current system time and
cannot be earlier than the sampling
time.
Operator Personnel running the You do not need to enter it and it will
sample. be displayed automatically.
Run Time Time when the sam- You do not need to enter it and it will
ple is run. be displayed automatically.
Approver Personnel validating This parameter will be automatically
the sample. displayed after the sample is validat-
ed.
Report Time The date and time This parameter will be automatically
when the report is displayed after the report is printed.
printed for the first
time.
Diagnosis Suspected diagnosis Input in the textbox directly.
information.
Remarks Clarifications or notes. Input in the textbox directly.
5.4.6 QUERY
You can view the test results of a patient within a certain test date range by
entering the query conditions. The procedures are shown as below:
1. Click the Query button to enter the multi-conditional query dialogue box as
shown below.
Figure 32
Query Conditions
! -- Auto select checked by default indicates that the query result is being
selected (with a blue background colour). If it’s unchecked, the query
result will remain on a white background colour.
-- Click All Samples to close the current window, display all the samples
again and restore all the filter conditions to the default values.
3. Click Query.
The system will display all the query results which meet the conditions.
5.4.7 RESULTS
In the Review interface, you can click Results to browse the selected sample re-
sults, parameter results and flag messages. The procedures are shown as below:
1. Select a result to review in results interface.
2. Click to unfold all function buttons.
3. Click Results to enter the results interface of the selected sample. In the Re-
sults interface, you can view sample information such as parameter results,
graph results and flag messages. In addition, you can also print the analysis
report.
Figure 33
Figure 34
Adding a New Microscopic
Exam. Parameter
Table 12
Parameter It means Operation
Sample Type Type of sample for Click the Sample Type dropdown
Microscopic Exam. Parameters
microscopic examination. list box and select the type of sam-
-- Venous blood ple for microscopic examination.
-- Capillary
-- Cord blood
-- Blood
Exam. Time Time of microscopic Click the date control for the
examination. settings.
-- The input sequence of the con-
trols is the same with the date
format on the top right corner
of the dialogue box. For exam-
ple, if the data format is yyyy/
MM/dd, you should input the
data in the sequence of year,
month, and date.
-- Click or to select
a date and time or enter the
information in the textbox di-
rectly.
-- Click to clear the current
data and re-enter the informa-
tion.
! Note:
The Microscopic exam. time can
be no later than the current sys-
tem time.
Microscopic Description of cells Enter the morphology information
Description morphology. for cells into the multi-line text-
box.
Figure 35
4. Click the cell corresponding to its Value column of the parameter, and
enter the value. If the unit and reference range of parameters have been
set in the Setup > Parameter > Custom Para. interface, the correspond-
ing unit and range (lower limit~upper limit) will be displayed in this tab.
When both the value and range of parameters are numbers, and the number
is out of the reference range, the relevant mark or will be displayed
in the Flag column. Please refer to chapter 4.7.6 Customized Parameters for
customized parameters settings.
84
-- Click OK.
The screen will show the parameter result run chart of the selected sample.
See Figure 37.
Figure 37
Run Chart
Figure 38
Viewing the Run Chart of Samples
on the Current Page
• Click OK.
The screen will show the parameter result run chart of the selected sample.
• Click the date edit box, set a date range in the pop-up dialogue box, then
click OK.
Figure 40
86
The input sequence of the controls is the same with the date format on
the top right corner of the dialogue box. For example, if the data for-
mat is yyyy/MM/dd, you should input the data in the sequence of year,
month, and date.
Click or to select a date and time or enter the information in
the textbox directly.
Click Cancel to keep the selected run date.
• Click OK. The screen will show the parameter result run chart of the se-
lected sample.
6 DAILY OPERATIONS
Preparation before
the operation
Startup
Daily Quality
Control
Sample Collection
and Handling
Sample Analysis
Shutdown
6.2 Startup
This section introduces the operations related to the startup of the analyzer.
Note:
Figure 41
Login
The initial user name and password of administrator are admin, which was
set by service engineer.
1 to 12 digits of numeric characters can be entered for the user name and
the password. No Chinese characters are allowed.
! Note:
-- The background test is designed for detecting particle interference
and electrical interference. For the background reference range of
each parameter, please see A.4.2 Normal Background.
-- The sample ID for the background test is background.
-- If the background results exceed the Ref. Range for the first time dur-
ing fluidics initialization, the analyzer will run the background test one
more time.
-- Running a test when there is a Background abnormal, you would ob-
tain an unreliable testing result.
-- If any error is detected during initialization (e.g. the background re-
sults exceed the Ref. Range), the analyzer will activate the alarm. For
details, see chapter 11 Troubleshooting.
-- To lock or switch a user, click on the menu screen and click Yes on the
pop-up dialogue box. The system will return to the login dialogue box.
Enter the user name and password, click , then you can log in again or
log in the software interface with another user identity.
Table 13
Parameter Interference
WBC > 5 NRBCs/100 WBCs, PLT clumps/ large PLTs
RBC WBC Count >75.0 x 103/µl
MCV WBC Count >75.0 x 103/µl
PLT PLT clumps/ large PLTs
Hemoglobin WBC count > 75.0 x 103 /ul, Lipids > 280 g/dl
Differential > NRBCs/ 100 WBCs, PLT clumps/ large PLTs
7 ADVANCED OPERATION
Analysis of the results with statistical methods allows the inference that sample
results are reliable. HUMAN recommends running the QC program on a daily
basis with low, normal and high level controls. A new lot of controls should be
analyzed in parallel with the current lot prior to their Exp. dates. This may be
accomplished by running the new lot of controls twice a day for five days using
any empty QC file.
-- You should only use the HUMAN-specified controls and reagents. Store and
use the controls and reagents by following the instructions for use of the
controls and reagents.
-- Controls beyond their Exp. date shall not be used. Controls (similar to stand-
ard blood samples) must be well mixed before use.
-- General users only have the access for browsing and executing the QC
analysis.
7.1.2 QC SETTINGS
Before running a new batch of controls, you need to assign a QC file to each batch
of controls. You can complete the QC settings by setting QC information in the
QC files. Only users with administrator-level access can edit the L-J settings.
92
Figure 42
L-J Quality Control
Figure 43
Entering QC Information
(Target values)
You can also select the QC file of which data has been set and then click Copy,
and edit the content based on the original data. Set related information of the
controls with reference to Table 14.
! Note:
The lot No. can not be empty
and up to 16 digits can be en-
tered. You can enter characters,
numbers, letters and special
characters, but no Chinese char-
acters are allowed.
Level Level of the controls, includ- Select from the dropdown list.
ing 3 levels, i.e. High, Normal
and Low.
Exp. Date Exp. date of the controls. The default Exp. Date is the cur-
rent system date and needs to be
changed to the actual Exp. date
of the controls.
QC Mode QC mode of the controls, in- Select from the dropdown list.
cluding Whole Blood-DIFF
and Predilute-DIFF.
94
! Note:
You can click Set Limits to set
the display form of the limits or
the calculation method of the
limits among the preset values.
-- By SD: the limits displays in
form of absolute value. Click
2SD or 3SD to select either
double or triple standard devi-
ation to be the limits.
-- By CV: the limits displays in
form of percentage. Click 2CV
or 3CV to select either double
or triple coefficient of varia-
tion to be the limits.
In use Set if you want to specify the It’s unchecked by default. Set the
QC sample ID in the selected parameter according to the actu-
file so that you can run the QC al situation.
sample in the interface other
than the QC interface.
-- If it’s checked, you can run
the sample with the corre-
sponding sample ID in any
interface and the system
will run the QC analysis for
this sample.
-- If it’s not checked, you can
only run the QC sample in
the QC interface.
Existing/ The existing data and total QC Read only.
Total results in the current QC file.
Up to 500 QC results can be
saved for each QC file.
4. According to the target list of the corresponding lot No., enter the target val-
ue and limits into the textboxes of the parameters to be included in the QC
run.
5. Click the Save button to save all the settings of the QC.
Figure 44
4. Click Yes.
All selected QC files together with their QC results will be completely deleted.
96
! Note:
-- You should only use the Human-specified controls and reagents. Store
and use the controls and reagents as instructed by instructions for use
of the controls and reagents. Using other controls may lead to incor-
rect QC results.
-- Before being used for analysis mix gently the controls that have been
settled for a while according to the instruction of controls.
-- Be sure to use the Human-specified disposable products including
vacutainer blood collection tube, vacutainer blood collection tubes
with anticoagulant and capillary tubes etc.
-- If the blood-sample mode is Predilute, then a reminder of predilute
counting will pop up if the user presses the aspirate key to perform the
counting. To close the prompt, please refer to Auxiliary Settings.
WARNING
-- The sample probe (needle) is sharp and potentially biohazard-
ous. Exercise caution to avoid contact with the probe when
working around it.
-- The sample may spill from the unclosed collection tubes and
cause biohazard. Exercise caution to the unclosed collection
tubes.
-- Collection tubes broken may cause personal injury and/or bio-
hazard. Be sure to place the collection tubes in the right adapter
before running, otherwise, the collection tubes may be broken
and cause biohazard.
-- Keep your clothes, hairs and hands away from the moving parts
to avoid injury.
-- Reagents can be irritating to the eyes, skin, and mucosa. Wear
proper personal protective equipment (e.g. gloves, lab uniforms,
etc.) and follow laboratory safety procedures when handling
them in the laboratory.
-- If the reagent accidentally comes in contact with your skin,
wash it off immediately with plenty of water and see a doctor
if necessary. Do the same if you accidentally get any of the rea-
gent in your eyes.
After completing the QC settings, users can perform the QC analysis in the QC
Analysis interface.
Figure 45
QC Analysis
3. Select the QC file No. to be run. The screen will display the corresponding
information and QC parameters.
4. Be sure that the level of the control to be run is the same with the current QC
file, and the control to be run is not expired.
5. Prepare the controls according to control instructions. If the QC mode is Pre-
dilute-DIFF, Predilute the controls with reference to chapter 6.4 Sample Coll-
ection and Handling and get diluted QC samples.
! Note:
-- Be sure to evaluate predilute stability based on your laboratory’s sam-
ple population and sample collection techniques or methods.
Figure 46
Mixing the Controls
7. In the ready for counting state (namely, the indicator light of the main unit
is green), place the controls under the sample probe where the probe can
aspirate the well-mixed controls.
8. Press the aspirate key and start running the controls.
9. Upon the completion of the aspiration, you’ll hear a beep and you can re-
move the controls. When the running of QC analysis is complete, the QC
results will be displayed in the current screen (as shown in Figure 47) and
saved in the QC file automatically.
Figure 47
Analysis Results
10. Perform the above procedures to continue running the controls if necessary.
! Note:
-- If the QC file is outdated, its valid period will be displayed in red.
-- “” or “” alarm symbol will be displayed next to the results with
deviations exceeding the set limits.
100
! Note:
-- You should only use the HUMAN-specified controls and reagents.
Store and use the controls and reagents as instructed by instructions
for use of the controls and reagents. Using other controls may lead to
incorrect QC results.
-- Before being used for analysis shake well the controls that have been
settled for a while.
-- Be sure to use the HUMAN-specified disposable products including
vacutainer blood collection tube, vacutainer blood collection tubes
with anticoagulant and capillary tubes etc.
-- If the blood-sample mode is Predilute, then a reminder of predilute
counting will pop up if the user presses the aspirate key to perform the
counting. To close the prompt, please refer to Auxiliary Settings.
After completing the QC settings, you can place the controls among the daily
samples and perform analysis together in the Sample Analysis interface. After
the analysis is completed, the system will store the results to the QC file with
the corresponding ID.
Specific steps for performing QC analysis in the Sample Analysis interface are
as follows:
1. Prepare the controls according to the set control mode and control instruc-
tions.
2. In pre-dilution mode: Predilute the controls with reference to chapter 3.1.13
Sample Collection and Handling and get diluted QC samples if the QC mode
is Predilute.
3. Enter the set QC Sample ID in the Sample ID edit box (other options can be
ignored). Refer to chapter 7.1.2.1 Entering QC Information for the setting of
the QC Sample ID.
4. Well mix the prepared controls.
5. In the ready for counting state of the analyzer (namely, the indicator light of
the main unit is green), place the controls under the sample probe where the
probe can aspirate the well-mixed controls.
6. Press the aspirate key and start running the controls.
7. Upon the completion of the aspiration, you’ll hear a beep and you can re-
move the controls. When the running of the controls is complete, the QC
results will be saved in the QC file automatically.
8. Perform the above procedures to continue running the controls if necessary.
! Note:
-- It is not recommended to exceed the limits higher than shown on the
target value sheet of the control.
-- Be sure to evaluate predilute stability based on your laboratory’s sam-
ple population and sample collection techniques or methods.
Figure 48
102
Figure 49
Graph Interface
3. Select the QC file No. you want to review. The screen will display the corre-
sponding information and the graph. See Figure 50.
Figure 50
QC Graph
4. Click the buttons at the right side of the QC graph, then you can browse QC
graphs of different parameters; click the buttons at the bottom of the QC
graph, then you can browse all QC results.
Figure 51
3 2 1
4 7 6
8
104
Interface Description:
1. The Mean, SD and CV% of all the QC results of each parameter in the current
graph.
2. The saving date and time of the QC points located on the grey line.
3. The operator who run the QC analysis and obtained the QC points located
on the grey line.
4. The QC results of the parameters that correspond to the QC points located
on the grey line.
5. The QC points in each graph are displayed from left to right according to the
sequence from the earliest to the latest. The QC points are connected by a
line to illustrate the distribution trend.
6. The QC point corresponds to each QC result. Only the selected QC point dis-
plays its value under the parameter. The black QC point indicates the value is
within the limit; the red QC point indicates the value is out of the limit.
7. When you clicking a QC point in the graph, the QC points of other parame-
ters saved together with this one will be marked by a grey line.
8. The relative position of the QC point located on the grey line and the total QC
points saved currently.
! Note:
-- The outliers are excluded from the calculation of Mean, SD and CV%.
-- See below how to define a value as outlier.
Delete
The administrator can delete the QC results by the following steps:
-- Delete a single QC result
• Move the grey line to the desired QC result, and click Delete.
• Select Current Data in the pop-up dialogue box as shown in Figure 52.
Figure 52
Deleting Current QC Data
(QC Graph)
• Click OK.
-- Deleting all the QC results in the current QC file. Click Delete, select All Data
in the pop-up dialogue box, then click OK. See Figure 53.
Figure 53
Deleting all QC Data (QC Graph)
Figure 54
Deleting all QC Data (QC Graph)
2. You can select the reason from the given ones or manually enter the reasons
(up to 200 characters) into the textbox after selecting Others.
3. Click OK to save the reasons for the outliers and exit.
! Note:
If you enter the reason for the group of QC points whose results are ac-
tually within the limits, then their corresponding QC data both in the QC
Graph and QC Table will be displayed in red. And the data will return in
black if you cancel the reason and then save the changes.
106
Print
You can have the QC data of the current page or all QC data in the QC file printed
by clicking the Print button.
7.1.4.2 Table
All the samples, controls, calibrators, reagents, wastes and areas in contact with
them are potentially biohazardous. Wear proper personal protective equipment
(e.g. gloves, lab uniforms, etc.) and follow laboratory safety procedures when
handling them and the relevant areas in the laboratory.
Figure 55
3. Select the QC file No. you want to review. The screen will display the corre-
sponding information and the table.
4. Click the buttons at the bottom of the table to browse the QC data of de-
sired parameters; click the buttons on the right of the table to browse the
QC results.
Editing
Choose a row in the QC table and click Edit, then you can edit the selected QC
data. The edited data will be marked with an E. See Figure 56.
Figure 56
Editing QC Results
Restoring
Click Restore to cancel the editing of the QC results. After the data is restored,
the E mark will disappear.
Delete
With the administrator-level access, users can delete the selected QC data, QC
data on the current page and all QC data.
-- Delete a selected QC result
• Click the column containing the desired QC result, and then click Delete.
• Select Current Data in the pop-up dialogue box as shown in Figure 57.
Figure 57
Deleting Current QC Data
(QC Graph)
• Click OK.
108
Figure 58
• Click OK.
-- Delete all QC results
! Note:
Please be careful to perform this operation as it will delete all QC data of
the selected QC file and cannot be reverted.
-- Click Delete.
-- Select All Data in the pop-up dialogue box.
-- Click OK.
The interface pops up a dialogue box as shown below.
Figure 59
Print
You can print all the QC data or the data within the specified date range of the
selected QC file.
Detailed steps are shown below:
1. Select a QC file No. to be printed.
2. Click Print.
The interface pops up a dialogue box as shown below.
Figure 60
Deleting all QC Data (QC Graph)
Export
If you wish to export the information and the result of the current QC file, do as
follows:
Figure 61
Deleting all QC Data (QC Graph)
4. Select an export path for the data and enter the file name. The file will be ex-
ported to the root directory of the USB flash disk (/udisk/sda1) and named in
the format of SampleInfo_yyyyMMdd_hhmmss.csv. Among which, yyyyM-
Mdd_hhmmss means data export year, month, date, hour, minute, and sec-
ond.
5. Click Save. When the export is finished, a message box as shown below will
pop up.
Figure 62
7.2.1 QC Principle
The X-B analysis is a weighted moving average analysis that uses values ob-
tained from patient samples. It uses the 3 red cell indices, MCV, MCH and MCHC
to indicate the hematology instrument performance.
This is a QC method without controls, and reflects only an indicator for a correct
analyzer performance. HUMAN highly recommends to use control materials.
Both methods reflect the analysis performance of the analyzer from different
perspective. Thus, one method should not be replaced with the other. It is rec-
ommended the X-B analysis be activated when the sample volume of your labo-
ratory is greater then 100 samples per day.
Effective use of X-B requires randomization of samples and a normal cross sec-
tion of patients to prevent skewing of indices. A reference range is established
by the given reference values as well as lower and upper limits for the purpose
of observing the variation of QC results within the reference range. The analyzer
performs X-B QC for three parameters, MCV, MCH, and MCHC.
Twenty to two hundred samples can be grouped together for X-B numerical
analysis. The samples are derived from the results of normal analyzer counting,
with no distinction of whole-blood or predilute mode. The analyzer can save
maximum 500 X-B QC results. When the saved QC results have reached the
maximum number, the newest result will overwrite the oldest.
112
7.2.2 QC SETTINGS
! Note:
-- Only users with administrator-level access can edit the X-B settings.
-- Perform the QC Settings before running the controls. You can com-
plete the QC settings by entering the QC information.
Figure 63
! Note:
Once the Samples/Group is changed, the number of valid sample results
will be recalculated. For example, if 20 valid samples are needed for the
X-B QC calculation, when you change the value of Samples/Group after
10 group of valid sample results have been acquired, these 10 group of
results will be discarded, and only valid sample results generated after-
wards will be used in the QC calculation.
5. Click the Open button of X-B to open the X-B quality control. The samples
results will be included to calculate the X-B.
6. Enter the targets and limits for the QC parameters.
! Note:
-- All the targets and limits for the QC parameters must be entered.
-- When first use, the default setting will provide the Initial values for
the targets and limits of the three QC parameters.
-- If the QC data have existed in the QC file, you are not allowed to edit
the target and limits.
-- You can set the display form of the limits or the calculation method of
the limits among the preset values see chapter 7.2.2.2
7. Set the valid upper and lower limits for the QC parameter in Sample Validity
Setting field. Setting sample validity is to set the valid range of four QC pa-
rameters, RBC, MCV, MCH and MCHC. To be incorporated into X-B QC calcu-
lation, the sample results should satisfy the validity ranges of all these four
parameters.
! Note:
Once the Samples/Group is changed, the number of valid sample results
will be recalculated. For example, if 20 valid samples are needed for the
X-B QC calculation, when you change the value of Samples/Group after
10 group of valid sample results have been acquired, these 10 group of
results will be discarded, and only valid sample results generated after-
wards will be used in the QC calculation.
8. Click the Save button to save all the settings of the QC. If the entered value
exceeds the acceptable range or the upper limit is lower than the lower limit,
a reminder message will pop up and you will be prompted to re-enter the
correct data and save the entry again.
114
Figure 64
! Note:
-- If QC data are existing in the QC file, you are not allowed to restore the
parameters.
-- Clicking Restores Defaults can only store the default settings of
Target, Limits and Sample Validity Setting, while Samples/Group, X-B
QC switch and limit settings cannot be restored.
7.2.4.1 Graph
Access the X-B QC Graph interface by taking the following steps:
1. Click QC to access the QC interface.
2. Select X-B from the dropdown list of the QC Type.
3. Click Graph. The X-B QC Graph interface will be displayed, see Figure 65.
116
Figure 65
QC Graph
4. You can also drag the scroll bar down to the graph horizontally to browse all
the QC results.
Figure 66
4
2
Delete
The administrator can delete the QC results by the following steps:
-- Delete a single QC result
• Move the green line to the desired QC result, and click Delete.
• Select Current Data in the pop-up dialogue box as shown in Figure 67.
Figure 67
Deleting Current QC Data
(QC Graph)
• Click OK.
118
-- Deleting all the QC results in the current QC file. Click Delete, select All Data
in the pop-up dialogue box, then click OK. See Figure 68.
Figure 68
Deleting all QC Data
(QC Graph)
Print
Click the Print button to print the QC graph.
7.2.4.2 Table
Access the X-B QC Table interface by taking the following steps:
1. Click QC to access the QC interface.
2. Select X-B from the dropdown list of the QC Type.
3. Click QC Table. The X-B QC table interface will be displayed. See Figure 69.
Figure 69
QC Table
Figure 70
3 5 4
Delete
With the administrator-level access, users can delete the selected QC data, QC
data on the current page and all QC data.
-- Delete a selected QC result
• Click the column containing the desired QC result, and then click Delete.
• Select Current Data in the pop-up dialogue box as shown in Figure 71.
Figure 71
Deleting Current QC Data
(QC Graph)
• Click OK.
120
Figure 72
Deleting all QC Data (QC Graph)
• Click OK.
-- Delete all QC results.
! Note:
-- Please be careful to perform this operation as it will delete all QC data
of the selected QC file and cannot be reverted.
• Click Delete.
• Select All Data in the pop-up dialogue box.
• Click OK. The interface pops up a dialogue box as shown below.
• Click Yes to delete all the QC results in the current QC file.
Figure 73
Print
You can print all the QC data or the data within the specified date range of the
selected QC file. Detailed steps are shown below:
1. Select a QC file No. to be printed.
1. Click Print.
The interface pops up a dialogue box as shown below.
Figure 74
Export
If you wish to export the information and the result of the current QC file, do as
follows:
1. Insert a USB flash disk in the USB interface on the analyzer.
2. Click Export.
3. A dialogue box will pop up as shown below.
Figure 75
122
4. Select an export path for the data and enter the file name. The file will be ex-
ported to the root directory of the USB flash disk (/udisk/sda1) and named in
the format of SampleInfo_yyyyMMdd_hhmmss.csv. Among which, yyyyM-
Mdd_hhmmss means data export year, month, date, hour, minute, and
second.
5. Click Save. When the export is finished, a message box as shown below will
pop up.
Figure 76
Export successfully
7.3 Setup
7.3.1 INTRODUCTION
This chapter introduces the daily operations from the startup to the shutdown
of the analyzer.
A flow chart indicating the common daily operation process is presented in
chapter 6 daily operations.
The analyzer has been initialized before delivery. The interfaces upon the initial
startup of the analyzer are system settings by default. Some parameters of the
analyzer can be reset to meet various demands in practical applications. The
analyzer divides the operators into two access levels, common user and admin-
istrator. Note that an administrator can access all the functions accessible to a
common user. This chapter introduces how to customize your analyzer as an
administrator.
Figure 77
Setup
The administrator is allowed to set the following functions in the Setup inter-
face:
-- System settings
-- Parameter settings
-- Meterage settings
-- LIS communication
-- User management
-- Print settings
-- Auxiliary settings
124
Figure 78
2. Click the Date and Time dropdown list and set the current date and time of
the system in the popup dialogue box.
Figure 79
Related descriptions:
-- The input sequence of the controls is the same with the date format on
the top right corner of the dialogue box. For example, if the data format is
yyyy/MM/dd HH:mm:ss, you should input the data in the sequence of year,
month, date, hour, minute, and second.
-- Click or to select a date and time or enter the information in the
textbox directly.
-- Click Cancel to close the window and keep the current date and time set-
tings.
Figure 80
5. Click Apply. The system message will pop up, indicating the successful set-
ting. See Figure 81.
Figure 81
Successful Setting
of the Date Format
The date and time at the bottom right corner will be displayed in the newly set
format as shown in
Figure 82
Figure 83
As shown in Figure 83, You can set to turn the soft keyboard on or off.
-- Soft Keyboard
• On (default)
You can enter content using the soft keyboard popped up on the screen.
Functions and applications for the keys are shown in Figure 84.
6
Figure 84
1 Toggling between upper
and lower case
2 Toggling between number
and symbol input
1
3 Hiding the soft keyboard
4 Hiding the soft keyboard
5 Line feed/Enter key
6 Delete key 2 3 4 5
• Off
You need to use an externally connected USB keyboard for entering con-
tent.
Figure 85
Setting Lab Information
! Note:
Only the administrator has the access for setting the lab information.
General users are only allowed to browse such information. Refer to the
table below for the detailed instructions of parameter setting
Figure 86
Auto Maintenance
Auto Sleep
In the Wait textbox, the administrators can set the waiting time for enter-
ing the sleep state after the main unit is halted. The range is between 15 and
120 minutes and the default value is 60 minutes.
Figure 87
Shortcut Code
You can set the shortcut code for the following items: Department, Submitter,
Patient Type, Gender, Area, Bed No. and Sample Type.
Figure 88
Adding a New Item
130
Figure 89
Information of the
Newly Added Department
Figure 90
Editing Item/Shortcut Code
2. Modify the Name, Shortcut Code and Remarks in each textbox according to
the actual demand.
• Newly added department name must be entered and it can not be the
same as existing ones.
• The shortcut code is not necessary to be entered, but once set, every code
must be unique.
3. Click OK to save the information.
Figure 91
Deleting a Department
Figure 92
Shortcut Code
Figure 93
2. Click the parameter, of which the unit is to be set, from the parameter list
(such as WBC).
3. Select a new parameter unit from the Unit Options list.
Figure 94
Figure 95
Ref. Range
Figure 96
Copying a Ref. Group
You can edit the new reference group. Save and close the screen, and then the
copied reference group will be shown in the reference group list.
Figure 97
Copying a Ref. Group
! Note: The reference group name entered is not allowed to be empty nor
the same as the existing ones.
Figure 98
Adding a New Ref. Group
2. Complete the entries for each parameter with reference to the parameter
description in Table 16.
136
! Note:
The reference group name en-
tered is not allowed to be empty
nor the same as the existing ones.
Lower Limit of Lower limit of age of the Enter an integer value in the text-
Age reference group. box and select the age unit (year,
month, week, day, or hour) from
the drop list on the right.
! Note:
The Lower Limit of Age must
be smaller than the Upper Limit
of Age.
Upper Limit of Upper limit of age of Enter an integer value in the text-
Age the reference group. box and select the age unit (year,
month, week, day, or hour) from
the drop list on the right.
! Note:
The Lower Limit of Age must
be smaller than the Upper Limit
of Age.
Gender Gender of the reference Select Man, Woman, Not defined
group. from the dropdown list. The de-
fault setting is empty.
Lower Limit (of Lower limit of param- Click the Lower Limit cell which
parameter) eters of the reference corresponds to the parameter and
group. If the test result enter a new value.
is lower than this value,
it would be regarded as
clinically abnormal. ! Note:
The Lower Limit must be small-
er than the Upper Limit.
Upper Limit Upper limit of param- Click the Lower Limit cell which
(of parameter) eters of the reference corresponds to the parameter and
group If the test result enter a new value.
is higher than this value,
it would be regarded as
clinically abnormal.
! Note:
The Lower Limit must be small-
er than the Upper Limit.
Figure 99
Adding a New Ref. Group
2. Refer to Table 16 for the description of the parameters to finish the editing.
! Note:
-- For the built-in reference group, you can modify the upper limit and
lower limit of the parameters, but not its name, the upper limit and
lower limit of age as well as gender.
-- Click Set as default to restore the setting of the selected reference
group to the default value.
-- Non-built-in reference group (which is added by user) cannot restore
defaults.
Figure 100
Setting Default Ref. Group
Automatically match the Customized Ref. Match the reference group Table 17
customized reference group Group
Rules for Matching the Reference
according to age and gender
Group
Unchecked N/A Built-in reference group
Checked None Built-in reference group
Checked Created Preferentially match the
customized reference group
! Note:
When the customized ref. groups are used to match the reference group,
the matching will be performed from top down according to the custom-
ized ref. groups displayed in the screen.
! Note:
The operations of adding, editing, deleting and adjusting the list order do
not affect the sample record in which the microscopic examination re-
sults have been entered and saved. Such operations are only valid for the
record in which the microscopic examination results have not been saved,
and the samples analyzed after the setting operations.
Figure 101
Microscopic Exam. Settings
140
Figure 102
Adding a New Microscopic
Exam. Parameter
2. Input the parameter name and its coding system in the corresponding text-
boxes.
• The Parameter Name can not be empty and up to 32 characters can be
entered.
• The Code System is the code ID of the parameter. It is used for LIS trans-
mission only when the parameter is transmitted to the LIS. You may not
input the value if it is not needed. Up to20 characters can be entered.
3. Click OK. The name of the new parameter will be displayed in the microscop-
ic exam. parameter list.
Figure 103
Editing a Microscopic Exam.
Parameter
Figure 104
Deleting a Microscopic Exam.
Parameter
Figure 105
Setting RUO Parameters
142
! Note:
-- Display RUO Parameters
It’s checked by default, which means the information regarding the
RUO parameters will be displayed in the counting results. If it’s un-
checked, the RUO parameters, the * mark and the declaration will not
be displayed in the counting results.
• Display “*” mark
It’s checked by default, which means the “*” mark will be dis-
played in the counting results; If it’s unchecked, the “*” mark and
the declaration will not be displayed.
• Display declaration
It’s checked by default, which means the declaration will be dis-
played in the counting results; if it’s unchecked, the declaration
will not be displayed.
-- Print RUO parameters
It’s checked by default, which means the RUO parameters will be
printed in the report. If it’s unchecked, the RUO parameters, the “*”
mark and the declaration will not be printed in the report.
• Print “*” mark
It’s checked by default, which means the “*” mark will be printed
in the report. If it’s unchecked, the “*” mark and the declaration
will not be printed in the report.
• Print declaration
It’s checked by default, which means the declaration will be
printed in the report. If it’s unchecked, the declaration will not be
printed in the report.
-- Editing Declaration
The default declaration is: “*” means “research use only, not for
diagnostic use”. You can modify the declaration in the textbox as per
the actual demand. Up to 50 characters can be entered, including all
characters, numbers, letters and other special characters (except “/”
and “\”) on the keyboard.
! Note:
Any change made to the display settings or printing of the RUO param-
eters, the “*” mark and the declaration will be applied to all the RUO pa-
rameters (before and after the change).
Figure 106
Customized Parameter Settings
Figure 107
Adding a Customized Parameter
2. Click the textboxes of Parameter Name and Unit respectively, and enter the
name and unit of the customized parameter.
3. Click corresponding cells of the Upper Limit and Lower Limit of the reference
group, and input values. You can also customize the reference group accord-
ing to the actual situation. For details, see chapter 7.3.4.11 Ref. Range.
4. Click OK. The added parameter will be displayed in the customized param-
eter list.
2. Click the textboxes of Parameter Name and Unit respectively, and modify
the name and unit of the customized parameter.
3. Click corresponding cells of the Upper Limit and Lower Limit of the reference
group, and modify the values. You can also customize the reference group
according to the actual situation. For details, see chapter 7.3.4.11 Ref. Range.
4. Click Save.
Figure 109
User management
Figure 110
! Note:
User Group includes Common User and Administrator. Users are assigned
different access levels according to the user group they belong to. Click
OK after the setting is complete. The information of the new user will be
shown in the user list.
Figure 111
! Note:
The administrator cannot delete his/her own information.
Figure 112
After it is set successfully, the default user name will be displayed in the login
box next time and you only needs to enter the corresponding password. See
Figure 113.
Figure 113
Login after Setting the Default
Use
148
Figure 114
Changing Password
! Note:
You can only change his/her own password and cannot change the pass-
word of other users.
Figure 115
Resetting Password
! Note:
The administrator is allowed to reset the password of all administrators
and general users; general users do not have the access to reset the pass-
word.
Figure 116
Printer Settings
You can set the printer and driver of the system in the Printer selection. See
Figure 117.
Figure 117
Printer Settings
-- Printer Driver
The system automatically detects the printer driver by default.
-- Printer
Select a printer to be used from the dropdown list. If the dropdown list is
blank, it indicates that no printer has been installed for the operating sys-
tem. In this case, install a printer, and then perform the relevant settings and
printing operations.
-- Printer Resolution
Select a proper resolution from the dropdown list. The higher the resolution
of the printer, the better the print quality.
150
Report Settings
You can set relevant parameters of the report in the Report Settings combo box.
See Figure 118
Figure 118
Report Print Setting
-- Report Title
Enter the title of the report in the Report Title textbox. The default setting is
Hematology Analysis Report.
-- Copies
You can enter the number of copies to be printed for a report in the Copies
textbox according to the actual demand. Click to increase the number of
copies and click to decrease the number of copies or enter the number of
copies in the edit box directly. Range of the copies is between 1 and 100 and
the default value is 1.
Format Settings
Report type and template of prints can be set in the Format Settings combo box.
See Figure 119.
Figure 119
Format Settings
! Note:
Before importing/exporting template, insert a USB flash disk in the USB
interface on the analyzer.
-- Exporting template
Select the template to be exported from the dropdown list of
Template and click Export. Select the export path in the popup dia-
logue box, and click Save.
-- Importing template
Click Import and select the required template in the pop-up dialogue
box, then click Open.
-- Deleting template
Select the template to be deleted from the dropdown list of the
Template.
! Note:
Only customized templates can be deleted, the built-in templates can not
be deleted.
152
Auto Settings
-- Autoprint
The default setting is Off, which means the report should be printed manu-
ally after the results are obtained.
If it is set to On, the system will automatically print the report of the
sample as per the current report template once the counting results are
obtained.
-- If Print after validation is checked, the autoprint function becomes invalid.
-- Auto print is not applicable for the background results.
-- Auto print after validation
It’s unchecked by default, which means the system can print the report au-
tomatically without validation. If it’s checked, the report will be printed au-
tomatically after it’s been validated instead of being printed right after the
results are obtained each time.
The parameter is valid only when the Autoprint is set to On.
-- Auto validate when printing
It’s unchecked by default, which means the report will not be automatically
validated by the system at the time of printing.
If it’s checked, the report will be automatically validated and printed by the
system at the time of printing.
Printing Options
-- Print Flag
It’s checked by default, which means the flag information will be printed in
the report. If it’s not checked, it will not be printed.
-- Print Ref. Range
It’s checked by default, which means the reference range of the parameter
will be shown in the printed report; If it’s unchecked, the results alone, rather
than reference range, will be shown in the printed report and the reference
range will not.
When Auto Increment is selected as the Sample ID entry method, you can add
a prefix to a certain batch or samples for identification. Enter the prefix length
ranging from 0 to 24 (e.g. 2) of the sample ID in the Prefix Length textbox. The
prefix length will be applied to all sample IDs after the setting is saved.
Predilute
Set if you wish to see a pop-up dialogue box when you perform the Predilute
counting.
-- Ask for confirmation (default setting): In the Predilute mode, when you press
the aspirate key to start the analysis, a dialogue box will pop up to remind
you that the ongoing analysis is for Predilute counting.
-- Do not ask for confirmation: The dialogue box for confirming the Predilute
counting will not pop up.
Other
-- Show Result Edited Flags
It’s unchecked by default, which means the edited results are marked with
an M at the end, while the corresponding results with manual modifications
are marked with an m at the end. M or m is displayed between the result
data and the parameter unit by default.
If unchecked, the edited result will not be marked with an M or m.
-- Automatically generate the delivery date
It is checked by default, which means you do not need to manually enter the
Delivery Time when you modify patient information after running a sample.
The operating date will be displayed in the date textbox.
! Note:
When Autoprint is On, a page remains to be printed with one report.
-- Update blank test time before be printed
It’s unchecked by default, which means the blank test time will not be
processed by the system.
If it’s checked, the Delivery Time will be automatically updated as the
Run Time by the system at the time of printing.
-- Print as black and white
It’s unchecked by default, which means the report will be printed ac-
cording to the default settings of the printer.
If it’s checked, the report will be printed as black and white.
Figure 120
Auxiliary Settings
The administrator is allowed to set the following functions in the Auxiliary Set-
tings interface:
-- Sample Numbering Rules
-- Startup sample IP and mode
-- Predilute
-- Other
-- Prefix Length
When Auto Increment is selected as the Sample ID entry method, you can
add a prefix to a certain batch of samples for identification. Enter the prefix
length ranging from 0 to 24 (e.g. 2) of the sample ID in the Prefix Length text-
box. The prefix length will applied to all sample IDs after the setting is saved.
! Note:
When Autoprint is On, a page remains to be printed with one report. If
the Effective tomorrow is checked, the modification of the next sample ID
and mode after startup will become effective on the next day.
-- Continue using the sample ID and mode before the last shutdown. If
checked, the system will by default add 1 to the last sample ID analyz-
ed before shutdown as the next sample ID after startup.
Predilute
Set if you wish to see a pop-up dialogue box when you perform the Predilute
counting.
-- Ask for confirmation (default setting): In the Predilute mode, when you press
the aspirate key to start the analysis, a dialogue box will pop up to remind
you that the ongoing analysis is for Predilute counting.
-- Do not ask for confirmation: The dialogue box for confirming the Predilute
counting will not pop up.
Other
-- Show Result Edited Flags
It’s unchecked by default, which means the edited results are marked with
an M at the end, while the corresponding results with manual modifications
are marked with an m at the end. M or m is displayed between the result
data and the parameter unit by default. If unchecked, the edited result will
not be marked with an M or m.
Figure 121
Connecting the electrical devices
158
WARNING
-- Be sure to dispose reagents, waste, samples, consumables, etc.
according to local legislations and regulations.
-- Reagents can be irritating to the eyes, skin, and mucosa. Wear
proper personal protective equipment (e.g. gloves, lab uni-
forms, etc.) and follow laboratory safety procedures when han-
dling them in the laboratory.
-- If the reagent accidentally comes in contact with your skin,
wash it off immediately with plenty of water and see a doctor
if necessary. Do the same if you accidentally get any of the re-
agent in your eyes.
CAUTION
-- Please make sure the length of the diluent pipe and the waste
pipe should be no longer than 1500 mm; the length of the lyse
pipe and the cleanser pipe should be no longer than 850 mm.
-- Tighten the panel connector of the fluidic line so that the over-
all fluidic line is closed to prevent leakage and seepage caused
by siphonage, etc. Refer to Figure 122 for the connection of the
reagents placed outside the analyzer.
Figure 122
Connecting reagents placed
outside the analyzer
Refer to Figure 123 for the connection of the reagent placed inside the analyzer
HumaCount 5D.
Figure 123
Connecting reagents placed
inside the analyzer HumaCount
5D (left door opened)
HC5D HC5D
7.3.8.3 Installing the Diluent Float Sensor and Replacing the Reagents
Please install the diluent float sensor and replace the diluent as per the ap-
proaches stated in this section.
Figure 124
Installing the Diluent Float
Sensor
Figure 125
Installing the Waste Float Sensor
The waste container can be replaced according to the steps mentioned above.
The replaced waste shall be properly disposed to avoid contamination.
WARNING
-- Be sure to dispose reagents, waste, samples, consumables, etc.
according to local legislations and regulations.
Please contact the Human customer engineer to get Description of LIS Commu-
nication Protocol for Human Hematology Analyzers to complete the support of
the LIS workstation to the LIS communication protocol.
Figure 126
Host Network Settings
4. Set the IP address and other network information of the analyzer according
to the actual situation.
• If the network is accessed through a router on the site, please select Ob-
tain an IP address automatically and Obtain DNS server address auto-
matically.
• If the network is accessed through a network switch, or the analyzer is
directly connected to the LIS on the site, please select Use the follow-
ing address, so as to manually set the IP address and subnet mask of the
analyzer. The IP addresses of the analyzer and LIS must be in the same
network segment. Furthermore, their subnet masks shall be the same,
while other parameters can maintain null. Click OK to save the settings
and close the dialogue box.
Figure 127
Communication Settings
3. Input the IP address and port of LIS workstation in Network Settings area.
Find the IP address and port of LIS in the network setup interface in the LIS
workstation; if IP address can’t be found, try the method below:
• Enter the operating system of LIS workstation.
• Press combination key [Windows+R] to open the Run window.
• Input cmd, and then click OK.
• Input the ipconfig command into the cmd.exe window popped out. The
interface shows similar content as follows:
164
Figure 128
The IPv4 address in the red box is the IP address of LIS workstation.
-- The IP address 192.168.8.44 of the LIS workstation shown as above is used as
an example, real IP should be in the same network segment with LIS server.
4. Click OK to save the settings.
5. Check if the connection is successful.
The LIS icon in the upper right side on the analyzer screen turns from grey to
black , which indicates auto hematology analyzer software is connected to LIS
successfully. If the icon stays grey, the connection fails. Please check if the IP
address and port of LIS is correct and reconnect as the steps above; if the prob-
lem still exists, please contact the hospital network administrator or HUMAN
customer service engineer to handle it.
Table 19
Range Display Range Linearity Range
Parameter Low High Low High Unit
Display ranges and Linearity
WBC 0 999 0 300 109/l
ranges of all parameters of Hu-
NEU# 0 999 0 300 109/l
maCount 5D. LYM# 0 999 0 300 109/l
MON# 0 999 0 300 109/l
EOS# 0 999 0 300 109/l
BAS# 0 999 0 300 109/l
ALY# 0 999 0 300 109/l
LIC# 0 999 0 300 109/l
NEU% 0 99.9 NA NA %
LYM% 0 99.9 NA NA %
MON% 0 99.9 NA NA %
EOS% 0 99.9 NA NA %
BAS% 0 99.9 NA NA %
ALY% 0 99.9 NA NA %
LIC% 0 99.9 NA NA %
RBC 0 18 0 8.5 1012/l
HGB 0 300 0 250 g/l
MCHC 0 9999 NA NA g/l
MCH 0 999 NA NA pg
HCT 0 80 0 67 %
MCV 0 250 NA NA fl
RDW-SD 0 999 NA NA fl
RDW-CV 0 99.9 NA NA %
PLT 0 5000 0 3000 109/l
MPV 0 99.9 NA NA fl
PDW 0 999 NA NA NA
PCT 0 0.999 NA NA %
P-LCR 0 99.9 NA NA 109/l
P-LCC 0 5000 NA NA %
8 CALIBRATION
8.1 Introduction
Calibration is a procedure to standardize the analyzer by determining its devi-
ation, if any, from calibration references and to apply any necessary correction
factors. To get accurate blood analysis results, perform calibration of the analyz-
er following the procedures given in this chapter when it’s needed.
! Note:
-- Calibration procedures can only be performed by users with the ad-
ministrator-level access. The login users with the access level of gen-
eral users can not perform the calibration procedures but only browse
the calibration coefficients.
-- You should only use the Human-specified calibrators and reagents.
Store and use the calibrator and reagents following the instructions
for use of the calibrations and reagents.
-- The analyzer identifies a sample as a calibration sample only if the
analysis is started from the Cal interface.
-- The calculation of repeatability is included in the calibration proce-
dure.
! Note:
-- All of the measured parameters must be calibrated before readings of
this analyzer can be used as valid analysis results.
-- For laboratories conducting routine tests, the calibration should be ap-
plied at least once every six months.
168
1. Check and make sure enough reagents have been prepared for the calibra-
tion. You need to start over the calibration if the reagents run out during the
process.
2. Do the background check. If the analyzer alarms are activated for abnormal
background results, see 11 Troubleshooting for solutions. Run the median
controls in whole blood-CBC+DIFF mode consecutively for 11 times, take and
view repeatability of the counting results from the 2nd run through the 11th
run in the Review interface and make sure they are within the range speci-
fied in A.4.4 Repeatability.
3. Run the corresponding diluent for 3 times immediately after running the
high-level controls for 3 times and calculate the carryover by the following
formulae:
Figure 129
Manual Calibration
The calibration coefficients of whole blood mode and predilute mode are dis-
played on the manual interface.
! Note:
-- The login users with the access level of general users can not perform
the calibration procedures but only browse the calibration coefficients
on the current screen. To perform the calibration, please log out and
then log in as users with administrator-level access.
3. Check the calibration coefficient and calculate the new coefficient using the
following equation.
For example, the WBC reference value of a calibrator is 8.3, and the current
calibration coefficient of the whole blood mode is 99.00%. Run the calibrator
in whole blood mode for 11 consecutive times and calculate the WBC results of
the 2nd to 11th runs (n=10):The obtained CV is 1.1% and the Mean is 8.22, which
meet the requirements. The new calibration coefficient is obtained:
99.00% x 8.3
New calibration factor = = 99.96%
8.22
The calculated calibration coefficients shall be between 75%~125%. In case
of an invalid calibration coefficient, try to find out the reason (e.g. calibration
material not thoroughly mixed, incorrect operation, etc.).Then recalibrate the
analyzer and recalculate the calibration coefficients.
4. Enter the new calibration coefficients into the factor cell of the parameter
that requires calibration. The entered calibration coefficients shall be be-
tween 75.0%~125.0% (calculation results rounded to two decimal places).
5. Click Save.
-- If the new calibration coefficient is valid and different from the original val-
ue, the following dialogue box will pop up.
170
Figure 130
Calibration set successfully
On the screen, the calibration coefficient is refreshed to be the new one and the
calibration date is refreshed to be the current system date.
-- If the new calibration coefficients are invalid, the message box will pop up.
Click OK to close the message box and enter a valid factor.
Figure 131
Invalid Coefficients
! Note:
-- Only Human-specified calibrators shall be used. Human will not be re-
sponsible for any erroneous result caused by using other calibrators.
-- See the instructions for use of the calibrators for the lot No., Exp. Date
and the target. Complete the calibration with calibrators as per the
following procedure:
Figure 132
Auto Calibration Using
Calibrators
3. Enter the lot No. of the calibrator into the Lot No. box.
4. Click the Exp. Date box, and then edit the Exp. Date.
! Note:
-- The Exp. Date can be no earlier than the current system date.
-- The entered Exp. Date should be either the Exp. Date printed on the la-
beling or the open-container expiration date, whichever is earlier. The
open-container expiration date is calculated as follows: The date on
which the container is opened + the open-container stability in days.
5. Input the target values of the parameters in the corresponding cell of the
Target.
6. Prepare the calibrators following their instructions for use and place the cal-
ibrators under the sampling probe.
7. Press the aspirate key to start the calibration counting. After every calibra-
tion run, the progress bar will close automatically and the analyzer will have
different responses according to different analysis results.
-- If the results are valid and within the linearity range, they will be displayed
directly.
172
! Note:
-- After the valid calibration result is obtained, the parameters with cor-
responding checkboxes ticked off will be involved in the calculation of
the calibration coefficients by default.
-- If you switch to other interfaces before the new calibration coefficients
are obtained, the system will discard the current calibration data and
keep the original calibration coefficients.
8. To get 10 valid counting results, repeat steps 6~7 ten times. The analyzer will,
by default, calculate the Mean, CV% and the new calibration coefficients
based on all the ticked-off calibration data according to the formulae.
9. You can select a few groups of data for the calculation of the calibration co-
efficients which can be obtained unless at least 5 groups of ticked-off data
are included. Each time when you tick off or uncheck the checkboxes, the
calibration coefficients will be refreshed and displayed in time. When the
amount of the valid calibration data in the list reaches 10, a message box of
Calibrator calibration done! will pop up. Click OK to close the message box.
! Note:
The out-of-range CV% does not influence the display of the calibration
coefficients.
Figure 133
Auto Calibration Using
Calibrators
174
! Note:
The reference instrument must be a properly running standard analyzer
so as to ensure the accuracy of the reference values.
5. Enter the reference values for the parameters to be calibrated in the corre-
sponding Target textbox.
6. Place the blood sample under the sampling probe, press the aspirate key
on the analyzer to run the samples. The system will calculate the values for
WBC, RBC, HGB, MCV and PLT of the sample. If used in VWB mode, use 20 µl
of calibrator material. If used in Pre-diluted mode, first dispense diluent by
the analyzer into a bullet tube, than add 20 µl of calibrator material into the
bullet tube -prefilled with diluent.
7. Repeat step 6 for 10 times and calculate the counting results for sample No.
1 in the 10 runs. The system will calculate the Mean, CV and Calibration co-
efficient for each parameter of the sample. If the obtained calibration coef-
ficient for any sample is not within the valid range or CV% or any calibrated
parameters does not meet the repeatability, a dialogue box indicating in-
valid new calibration coefficient will pop up when you are selecting other
blood samples. Click Yes to clear the calibration data of the sample. Redo the
calibration or redo after running another sample meeting all criteria.
8. Refer to steps 6~7 and perform the counting operations for the remaining
four blood samples. The system will calculate the Mean, CV and Calibration
Coefficient for each parameter of the remaining 4 blood samples.
9. Click Calculate. The system will calculate the average of the calibration coef-
ficients, namely, the mean calibration coefficient (%), as the new calibration
coefficient based on the five blood samples. You can also check at least three
accurate calibration coefficients and the system will re-calculate the mean
calibration coefficient (%).
! Note:
The mean calibration coefficient is invalid if its absolute value of deviation
from the original calibration coefficient is greater than or equal to 5%.
9 REAGENT MANAGEMENT
Once the new reagent is connected to the analyzer, you can set the reagent con-
figurations, including validity period, residue volume and reagent barcode on
the Reagent Management interface. Upon the completion of reagent configura-
tion, you can perform the procedures for reagent replacement.
! Note:
-- Reagents can be irritating to the eyes, skin, and mucosa. Wear prop-
er personal protective equipment (e.g. gloves, lab uniforms, etc.) and
follow laboratory safety procedures when handling them in the labo-
ratory.
-- If the reagent accidentally comes in contact with your skin, wash it off
immediately with plenty of water and see a doctor if necessary. Do the
same if you accidentally get any of the reagent in your eyes.
Figure 134
178
Table 20
Parameter NOTE
Current Model Current model of the analyzer.
Parameter Description for
-- Open system
Reagent Management
-- Closed system
Reagent setting procedures for different analyzer mod-
els vary, please refer to chapter 9.2 Setting Reagent In-
formation.
Reagent Name Name of the reagent.
Exp. Date Exp. Date of the unopened reagent will be shown upon
the completion of the reagent settings.
Any reagent, regardless of its container being opened or
not, should not be used beyond this date.
Open-container The date on which the reagent container is opened. The
Date default open-container date is the date on which the re-
agent settings are completed.
Period after The validity period (days) after the reagent container is
opening (PAO) opened. It will be shown upon the completion of the re-
agent settings.
Open-container Expiration date of the opened reagent, and it will be
Exp. Date shown upon the completion of the reagent settings.
Residue Volume The current residue volume of the reagent, and it will
be shown in ml upon the completion of the reagent set-
tings. The unit is ml.
Figure 136
RF Card Verification
2. Put the RF card attached to reagent packing on the RF card reader in front of
the analyzer. The beeping of the card reader and a pop-up dialogue box as
shown in Figure 137 indicate the successful reagent settings.
Figure 137
Successful Reagent Settings
! Note:
-- The RF card is intended for single use only.
-- If RF card verification fails, please follow the system prompts and use
a valid RF card for re-reading.
3. Click OK.
4. Click Close to exit.
! Note:
Once the reagent settings are successfully completed, the system prompt
at the top right corner of the screen will show that the reagent has not
been replaced. To remove this error, click the error message and then click
Remove Error in the pop-up dialogue box. The analyzer will complete the
replacement of the reagent and remove the error.
180
After the replacement is completed, a message box as shown below will pop up
on the screen.
Figure 138
! Note:
When you have changed the reagents, run a background check to see if
the results meet the requirement. If RF card verification fails, please fol-
low the system prompts and use a valid RF card for re-reading.
10 SERVICE
10.1 Introduction
CAUTION
-- Performing unauthorized maintenance procedures can dam-
age your analyzer. Do not perform any maintenance proce-
dures that are not described in this chapter.
-- In case of problems not specified in this manual, contact Hu-
man customer service department or your local agent for as-
sistance.
-- Only Human-supplied parts can be used for maintenance. For
any question, contact Human customer service department or
your local agent.
-- Exercise caution to avoid contact with the sharp sample probe
when performing maintenance.
WARNING
-- Connect only to a properly grounded outlet.
-- Before turning on the analyzer, make sure the input voltage
meets the requirements.
! Note:
-- After long-distance transportation, the reagent must be allowed to
settle for more than one day before use.
-- When you have changed the diluents, cleansers or lyses, run a back-
ground check to see if the results meet the requirement.
182
Figure 139
Service
Figure 140
Reagent Replacement
4. Click the name of the reagent that needs to be replaced, such as Replace All
Reagents. After the replacement is completed, the following message box
will pop up.
Figure 141
! Note:
-- After long-distance transportation, the reagent must be allowed to
settle for more than one day before use.
-- When you have changed the diluents, cleansers or lyses, run a back-
ground check to see if the results meet the requirement.
184
10.1.1 CLEANING
Clean corresponding parts according to the actual situation:
-- WBC bath
You should clean the WBC bath when:
-- The background of the scattergram has abnormal excessive cells
-- the background of WBC- and/or HGB-specific parameters exceeds the refer-
ence range
-- RBC bath
When the background of RBC- and (or) PLT-specific parameters exceeds the ref-
erence range, you should clean the RBC bath.
-- Flow chamber
When the background of the scattergram has abnormal excessive cells, or bad
differential of WBC, you should clean the flow chamber.
-- Sample probe
When the sample probe is dirty, you should clean the sample probe.
The cleaning procedures are as follows:
1. Click the Service icon in the menu page to access the Service interface.
Figure 142
Service
Figure 143
Cleaning
3. Click the icon of the part that needs to be cleaned, such as Clean Sample
Probe. When the system cleaning is complete, the message box will pop up
to show that the cleaning is done.
Figure 144
10.1.2 MAINTENANCE
Maintenance of the analyzer includes: unclogging, cleanser soak, cleanser soak
for WBC channel and cleanser soak for RBC channel.
10.1.2.1 Unclogging
If clogging is found, or it is suspected that the counting results are not accurate
due to aperture clogging, you can perform the unclogging operations.
The unclogging procedures are shown as follows:
186
1. Click the Service icon in the menu page to access the Service interface.
Figure 145
Service
Figure 146
Maintenance
The system will start clogging, and a message box will pop up. After the unclog-
ging is completed, a message box will pop up to show that the clogging is done.
Figure 147
Figure 148
Service
Figure 149
Figure 150
Cleanser Soak
Figure 151
Cleanser Soak
5. Present the cleanser to the sample probe as per the prompt, and press the
aspirate key or click the Aspirate button. Cleanser soaking... and the soaking
time will appear as shown below. See Figure 152.
Figure 152
Cleanser Maintenance Done
6. Click Close.
7. Perform the above procedures to perform the cleanser soak again if neces-
sary.
Figure 153
Service
Figure 154
Figure 155
4. Click OK.
Figure 156
Figure 157
4. Click OK.
192
Figure 159
4. Click Yes.
• A dialogue box will pop up as shown below.
Figure 160
5. Remove all reagent pickup tube assemblies according to the prompt, and
then click OK to start emptying the fluidic system. After the emptying is
complete, a message box will pop up.
6. Place the [O/I] switch at the left side of the main unit in the [O] position.
7. After shutdown, empty the waste in the waste container, and dispose it.
WARNING
-- Be sure to dispose reagents, waste, samples, consumables, etc.
according to local legislations and regulations.
Figure 161
Figure 162
4. Click Yes.
The interface pops up a dialogue box as shown below.
194
Figure 163
5. Remove all reagent pickup tube assemblies according to the prompt, and
then click OK to start emptying the fluidic system. After the emptying is
complete, a message box will pop up.
Figure 164
6. Place all reagent pickup tube assemblies into the distilled water, and then
click OK to start priming.
! Note:
-- Be sure to use distilled water in order to ensure the normal use of the
device in the future. In addition, the beaker holding the distilled water
needs to be cleaned thoroughly.
-- The diluent pipe and lyse pipes should be stored separately in two
beakers. System performs the filling operation. After the filling is com-
pleted, the following dialogue box will pop up.
Figure 165
7. Take out the diluent and lyse pipes from the distilled water as per the prompt,
then click OK.
A dialogue box will pop up to prompt you to power off the device.
! Note:
Prepare to ship done. Please power off the analyzer!
1. Place the [O/I] switch at the left side of the main unit in the [O] position.
2. After shutdown, empty the waste in the waste container, and dispose it.
WARNING
-- Be sure to dispose reagents, waste, samples, consumables, etc.
according to local legislations and regulations.
! Note:
Prepare to ship done. Please power off the analyzer!
196
! Note:
-- At the Self-test or Status interface, the analyzer does not ask for con-
firmation to perform the cleanser soak.
-- If the analyzer is running or has problems when the conditions of auto
prompt for cleanser soak is satisfied, the analyzer will prompt again
after the current operation is completed or the problems are resolved.
-- After cleanser soak is completed, the accumulative count values will
be cleared automatically.
-- Cleanser soak is an important step in comprehensive device mainte-
nance. It is recommended not to stop soaking halfway.
10.3 Self-test
This feature is to test if some important components of the device can function
properly or not, including syringe and sampling assembly self-test, pressure and
vacuum self-test, valve self-test and other self-test.
Note:
! If the testing result is abnormal, you should try again for several times;
if the abnormalities persist, please contact Human customer service
department or your local agent.
Figure 166
Syringe
3. Click the part that needs to be tested, e.g. Sample Syringe, and wait for the
self-inspection results. After the self-test is completed, a dialogue box will
pop up to show the self-test results.
Figure 167
Syringe Self-test Results
Figure 168
Pressure and Vacuum Self-in-
spection
The system will perform the corresponding self-test operations. After the self-
test is completed, a dialogue box will pop up to show the self-test results.
Figure 169
Figure 170
Valve Self-test
3. Click the desired valve number (e.g. 1), then confirm whether it works prop-
erly by the sound of its opening and closing.
10.3.4 OTHERS
You can perform the self-test for RBC aperture voltage.
RBC aperture voltage
The self-test procedure of RBC aperture voltage is shown as below:
1. Click the Service icon in the menu page to access the Service interface.
2. Click Other Self-test in the Self-test selection.
The interface as shown in Figure 171 will pop up on the screen.
Figure 171
Pressure and Vacuum
Self-inspection
200
3. Click RBC Aperture Voltage to start self-test. The system will perform the
corresponding self-test operations. After the self-inspection is completed, a
dialogue box will pop up to show the self-inspection results.
Figure 172
RBC Aperture Voltage Self-test
Results
RF Card Reader
You can carry out a self-test on its built-in RF card reader. The operation proce-
dures are as shown below.
1. Click the Service icon in the menu page to access the Service interface.
2. Click Other Self-test in the Self-test selection.
3. Click the icon of RF Reader to start self-test.
Figure 173
Syringe Self-test Results
10.4.1 TEMPERATURE
1. Click the Service icon in the menu page to access the Service interface.
2. Click Temperature in the Status selection.
• The interface as shown in Figure 174 will pop up on the screen.
Figure 174
View Temperature Status
User can view the current temperature information of the analyzer, including
the temperature of preheating bath temperature, ambient temperature and the
temperature of the optical system. If the results of the temperature testing ex-
ceed the normal range, they will be highlighted by the red background.
Figure 175
Voltage and Current
You can view the voltage and current information of the analyzer. The voltage
or current value that exceeds the normal range will be displayed in a red back-
ground.
202
Figure 176
Disk Information
10.5 Log
In the Log interface, you can view the records of Set Paras, Other Logs, Fault Logs
and All Logs.
-- If a new record is added when the log is full, the newest record will overwrite
the oldest one automatically.
-- The administrator can view both his/her own operation logs and the general
users’ operation logs, while the general users can only review their own op-
eration logs.
-- The log can keep records of up to 5 years.
Figure 177
All Logs
3. Select the dates in the two date textboxes, and then you can view the all
logs within the date range, including operation time, log information and
the operator.
Figure 178
Parameter Revision Logs
3. Select the dates in the two date textboxes, and then you can view the pa-
rameter revision logs within the date range, including the revision date and
time, revision summary and the operator.
Figure 179
Fault Logs
3. Select the dates in the two date textboxes, and then you can view the fault
logs within the date range, including date and time when the faults occur,
fault description and the operator.
You can view other logs besides parameter revision logs and fault logs.
Figure 180
Other Logs
3. Select the dates in the two date textboxes to view the logs within the date
range, including operation date and time, operation records and the opera-
tor.
Figure 181
Data Cleanup
3. Click the End time combo box, set the date range of the data to be cleaned
up in the pop-up dialogue box.
Figure 182
-- The input sequence of the controls is the same with the date format on
the top right corner of the dialogue box. For example, if the data format is
yyyy/MM/dd, you should input the data in the sequence of year, month, and
day.
-- Click or to select a date.
-- Click Cancel to cancel the process. The data will not be cleaned up.
-- For example, If the End time is set to 2016/03/31, the data generated from
system installation
-- Date to 31 March 2016 will be cleared.
4. Click OK to save the settings and close the dialogue box.
5. Select the data to be cleaned up.
-- Counting results
-- L-J QC results
-- X-B QC results
-- Log files
-- Scattergram
Figure 184
Version Information
208
Figure 185
Figure 186
6. Click OK to exit.
11 TROUBLESHOOTING
11.1 Introduction
This chapter contains information that is helpful in locating and resolving prob-
lems that may occur during the operation of your analyzer.
This chapter is not a complete service manual and is limited to problems that
are readily diagnosed and/or corrected by the user of the analyzer. If the recom-
mended solution fails to solve the problem, contact Human customer service
department or your local agent.
Figure 187
Error Messages
You can refer to the following steps to deal with the error messages.
1. Click the error message area.
2. Touch the screen to disable the beep.
3. Click Remove Error. Normally, the system will automatically remove the er-
rors. For errors which cannot be removed automatically, you can take appro-
priate actions by following the error help information or 11.3 Error Message
Reference.
212
Table 21
Problem Name Troubleshooting Information
-12V power is not 1. Please power off the analyzer directly and restart later.
Error Message Reference
working properly. 2. If the error still exists, contact our customer service
department.
Optical 1. Close the optical assembly cover.
assembly cover is 2. Click the Remove Error button to remove this error.
open. 3. If the error still exists, contact our customer service
department.
The CC source 1. Please power off the analyzer directly and restart later.
voltage is 2. If the error still exists, contact our customer service
abnormal. department.
Abnormal laser 1. Please power off the analyzer directly and restart later.
current. 2. If the error still exists, contact our customer service
department.
Startup failure. 1. Click the Remove Error button to remove this error.
2. If the error still exists, contact our customer service
department.
Startup 1. Click the Remove Error button to remove this error.
initialization is not 2. If the error still exists, contact our customer service
executed. department.
The right-side door 1. Close the right side door.
is open. 2. Click the Remove Error button to remove this error.
3. If the error still exists, contact our customer service
department.
+12V power is not 1. Please power off the analyzer directly and restart later.
working properly. 2. If the error still exists, contact our customer service
department.
HC5D-Diluent 1. Check if the HC5D-Diluent expires. If so, replace it with
expiration. a new container of HC5D-Diluent.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
HC5D CBC Lyse 1. Check if the HC5D CBC Lyse expires. If so, replace it
expiration with a new container of HC5D CBC Lyse.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
HC5D Diff Lyse 1. Check if the HC5D Diff Lyse expires. If so, replace it
expiration with a new container of HC5D Diff Lyse.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
Preheating bath 1. Click the Remove Error button to remove this error.
temperature out of 2. If the error still exists, contact our customer service
working range. department.
Abnormal HGB Please contact our customer service department.
background volt-
age.
Abnormal RBC ap- 1. Click the Remove Error button to remove this error.
erture voltage. 2. If the error still exists, contact our customer service
department.
Abnormal back- 1. Check whether the diluent is contaminated.
ground. 2. If not, click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
Failed to read 1. Click the Remove Error button to remove this error.
sample syringe 2. If the error still exists, contact our customer service
parameter. department.
Failed to configure 1. Click the Remove Error button to remove this error.
sample syringe 2. If the error still exists, contact our customer service
parameter. department.
Sample syringe 1. Click the Remove Error button to remove this error.
timeout 2. If the error still exists, contact our customer service
department.
Sample syringe is 1. Click the Remove Error button to remove this error.
busy. 2. If the error still exists, contact our customer service
department.
Vertical motor in- 1. Click the Remove Error button to remove this error.
struction parame- 2. If the error still exists, contact our customer service
ter error. department.
214
Failed to read ver- 1. Click the Remove Error button to remove this error.
tical motor param- 2. If the error still exists, contact our customer service
eter. department.
Vertical motor 1. Click the Remove Error button to remove this error.
timeout 2. If the error still exists, contact our customer service
department.
Failed to read the 1. Click the Remove Error button to remove this error.
remaining steps of 2. If the error still exists, contact our customer service
vertical motor. department.
The vertical motor 1. Click the Remove Error button to remove this error.
is busy. 2. If the error still exists, contact our customer service
department.
Failed to read 1. Make sure the temperature sensor is correctly installed.
preheating bath 2. Click the Remove Error button to remove the error.
temperature. 3. If the error still exists, contact our customer service
department.
Failed to read opti- 1. Make sure the temperature sensor is correctly installed.
cal system temper- 2. Click the Remove Error button to remove the error.
ature. 3. If the error still exists, contact our customer service
department.
Failed to read am- 1. Make sure the temperature sensor is correctly installed.
bient temperature. 2. Click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
Waste is full. 1. Empty the waste container or install a new waste con-
tainer.
2. Click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
The setting tem- 1. Click the Remove Error button to remove this error.
perature of optical 2. If the error still exists, contact our customer service
system out of department.
range.
Optical system 1. Click the Remove Error button to remove this error.
temperature out of 2. If the error still exists, contact our customer service
working range. department.
Flow cell clog. 1. Click the Remove Error button to remove this error.
2. If the error still exists, contact our customer service
department.
Failed to read 1. Click the Remove Error button to remove this error.
horizontal motor 2. If the error still exists, contact our customer service
parameter. department.
Failed to configure1. Click the Remove Error button to remove this error.
Horizontal motor 2. If the error still exists, contact our customer service
parameter. department.
Horizontal motor 1. Click the Remove Error button to remove this error.
timeout 2. If the error still exists, contact our customer service
department.
The optocoupler of 1. Click the Remove Error button to remove this error.
the horizontal mo- 2. If the error still exists, contact our customer service
tor is not working department.
properly.
The horizontal 1. Click the Remove Error button to remove this error.
motor is busy. 2. If the error still exists, contact our customer service
department.
No HC5D-Diluent 1. Check whether the HC5D-Diluent container is empty.
If so, install a new container of HC5D-Diluent.
2. Click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
HC5D CBC Lyse 1. Check whether the HC5D CBC Lyse is running out or
running out or air there are air bubbles in the inlet tubing of HC5D CBC
bubbles in inlet Lyse. If it is running out, install a new container of HC5D
tubing. CBC Lyse; If there is still plenty of HC5D CBC Lyse or there
are bubbles, perform step 2.
2. Click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
HC5D Diff Lyse 1. Check whether the HC5D Diff Lyse is running out or
running out or air there are air bubbles in the inlet tubing of HC5D Diff Lyse.
bubbles in inlet If it is running out, install a new container of HC5D Diff
tubing. Lyse; If there is still plenty of HC5D Diff Lyse or there are
bubbles, perform step 2.
2. Click the Remove Error button to remove the error.
3. If the error still exists, contact our customer service
department.
HC5D-Diluent not 1. Click the Remove Error button to remove this error.
replaced. 2. If the error still exists, contact our customer service
department.
HC5D CBC Lyse not 1. Click the Remove Error button to remove this error.
replaced. 2. If the error still exists, contact our customer service
department.
HC5D Diff Lyse not 1. Click the Remove Error button to remove this error.
replaced. 2. If the error still exists, contact our customer service
department.
216
DIFF probe clog- 1. Click the Remove Error button to remove this error.
ging 2. If the error still exists, contact our customer service
department.
Abnormal 12V driv- 1. Please power off the analyzer directly and restart later.
ing power supply. 2. If the error still exists, contact our customer service
department.
Abnormal 24V driv- 1. Please power off the analyzer directly and restart later.
ing power supply. 2. If the error still exists, contact our customer service
department.
Insufficient 1. Check whether the HC5D-Diluent container is empty.
HC5D-Diluent. If so, install a new container of HC5D-Diluent.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
Insufficient HC5D 1. Check whether the HC5D CBC Lyse container is empty.
CBC Lyse. If so, install a new container of HC5D CBC Lyse.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
Insufficient HC5D 1. Check whether the HC5D Diff Lyse container is empty.
Diff Lyse. If so, install a new container of HC5D Diff Lyse.
2. Click the Remove Error button, the Reagent Manage-
ment screen will be displayed.
3. Set the reagent information by referring to chapter 9
Reagent Management.
4. If the error still exists, contact our customer service
department.
Appendix A Specifications
A.1 Classification
According to the CE classification, the Auto Hematology Analyzer belongs to in
vitro diagnostic medical devices, rather than those covered by Annex II and de-
vices for performance evaluation.
A.2 Reagents
A.3 Parameters
A.4.5 Repeatability
These repeatability requirements apply only to the situation in which a qualified
sample has been run for 11 times and the results of the 2nd to 11th runs are used
to calculate the repeatabilities.
A.4.6 Carryover
WARNING
Accessory equipment connected to the analogue and digital in-
terfaces must comply with the relevant Safety and EMC stand-
ards (e.g., IEC 60950 Safety of Information Technology Equipment
Standard and CISPR 22 EMC of Information Technology Equip-
ment Standard (CLASS B)).Anyone who connects additional equip-
ment to the signal input or output ports and configures an IVD
system is responsible for ensuring that the system works properly
and complies with the safety and EMC requirements. If you have
any problem, consult the technical services department of your
local agent.
-- Analyzer
-- Touch screen: 10.4 inches embedded touch screen with a resolution of 800×600
-- One LAN interface
-- USB interfaces
-- Power
-- Voltage: A.C 100V~240V
-- Input power: ≤200VA
-- Frequency: 50/60 Hz
-- Keyboard (Optional, USB)
! Note:
Be sure to use and store the analyzer in the specified environment.
Figure 188
Height
Depth
Width
Table 32
Analyzer Dimensions and Weight
Width (mm) 364
Height 498
Depth (mm) 431
Weight (kg) 28
12 APPENDIX
Appendix B Terms and Abbreviations
CWB Capillary Whole Blood
PD Predilute
VWB Venous Whole Blood