PQ Test Protocol MPA Sphere 20231211 064846

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

OVP - PQ Test Protocol

Company: Novartis Pharmaceuticals


Operator: Admin
Instrument Type: MPA Sphere
Optics Configuration: Sphere Background with: NIR, Quartz, RT-Pbs [External]
Accessory: None
Instrument Serial Number: 2006068
Instrument Firmware Version: 1.470 Oct 5 2006
OPUS/DB Version: OPUS 8.5(SP1) Build: 8, 7, 10 / DB: 8,7,10,139
Overall Test Result FAILED
Test expires: Test failed
Test Date/Time: 12/11/2023, 6:48:46 AM (GMT-8)
Test Spectra Path: C:\Users\Public\Documents\Bruker\OPUS_8.7.10\Validation\Data
\20231211\064846
Date of last PQ Reference
Measurement: 12/11/2023
Comment:

Signal to Noise Test

Minimum S/N(area 1): 500 Measured S/N: 2969

100% Line Test

Maximum 100% Line Deviation: 0.5 Measured 100% Line Deviation: 0.52

Interferogram Peak Test

Minimum Amplitude[%] : 70 Measured Amplitude[%]: 100.0

Energy Test

Maximum allowed Value: 30 Measured Value: 0.3

Wavenumber Accuracy Test - BRM 2065 (37.0 Deg. C, Peak is T. corr.)

Sample Material: BRM 2065


Specified Peak: 10245.50 cm-1 Maximum Deviation: 0.60 cm-1
Measured Peak: 10245.80 cm-1
Corrected Peak: 10245.53 cm-1 Measured Deviation: 0.03 cm-1

Photometric Reproducibility Test - Glass Filter A

Maximum Deviation[%]: 0.8 Measured Deviation[%]: 0.04

Overall Test Result = FAILED

Date and Signature Date and Signature

________________
Page 1 of 5 Page Signature
OVP - PQ Test Protocol
Signal to Noise Test, Spectra
100% Line Test, Spectra
Energy Test, Spectra
101.0
99.5 100.0 100.5
Transmittance [%]
99.0

9000 8000 7000 6000 5000


Wavenumber cm-1
1 PQ_MPA_Sphere_Energy_1.0
2 PQ_MPA_Sphere_Energy_2.0
3 PQ_MPA_Sphere_Energy_3.0
4 PQ_MPA_Sphere_Energy_4.0
5 PQ_MPA_Sphere_Energy_5.0

1 1f36c60d-377e-4152-8afa-80fdd5772ef6
2 4627c3a0-a7f2-46ce-85be-b3d71a9bcea7
3 db032420-eefd-4ac7-974a-84c98faf579d
4 9ad49670-06bc-4c90-a3b6-3625dafb2561
5 ffb2dced-08fe-46c8-991c-f07c0e821f3a

Energy Test, Measurement Parameters


Number of Scans: 10 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Open Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 128
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 2 of 5 Page Signature
OVP - PQ Test Protocol
Interferogram Peak Test, Spectrum
0.4
0.2
0.0
-0.2
-0.4

14000 12000 10000 8000 6000 4000 2000 0


Points
PQ_MPA_Sphere_Energy_1.0

1f36c60d-377e-4152-8afa-80fdd5772ef6

Interferogram Peak Test, Measurement Parameters


Number of Scans: 10 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Open Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 128
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 3 of 5 Page Signature
OVP - PQ Test Protocol
Wavenumber Accuracy Test BRM 2065 (37.0 Deg. C, Peak is T. corr.), Spectrum

10245.80
0.8
Absorbance Units
0.4 0.6
0.2

10300 10250 10200 10150


Wavenumber cm-1
PQ_MPA_Sphere_Frequency.1

c4485a57-bcde-4a7d-a1e8-9617e5933b2a

Wavenumber Accuracy Test BRM 2065 , Measurement Parameters


Number of Scans: 32 Preamplifier Gain: Ref
Resolution [cm-1]: 4 Zero filling: 8
Aperture: Filter BRM2065 Phase Correction: Power Spectrum
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 64
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 4 of 5 Page Signature
OVP - PQ Test Protocol
Photometric Reproducibility Test - Glass Filter A, Spectrum
25
Transmittance [%]
10 15 520

10000 9000 8000 7000 6000 5000


Wavenumber cm-1
PQ_MPA_Sphere_GFA.0

391458b8-3cb7-47b3-b248-5b447a2601de

Photometric Reproducibility Test - Glass Filter A, Measurement Parameters


Number of Scans: 32 Preamplifier Gain: Ref
Resolution [cm-1]: 8 Zero filling: 2
Aperture: Filter NG9 Phase Correction: Mertz
Acquisition Mode: Double Sided,Forward-Backward Velocity: 10 KHz
Apodization Function: Blackman-Harris 3-Term Phase Resolution: 64
Low Pass Filter: 10 KHz High Pass Filter: Open

________________
Page 5 of 5 Page Signature

You might also like