Atmos c31
Atmos c31
Atmos c31
ATMOS® C 31
ENT Unit
Operating Instructions
506.7350.B
506.7510.B
2015-08 Index: 34
0124
ATMOS
MedizinTechnik GmbH & Co. KG
ATMOS will supply a service manual containing detailed
circuit descriptions and schematics as well as information
on adjustment and servicing to service organizations au-
thorized by ATMOS.
Kudwig-Kegel-Straße 16 Reprints, also in extracts, only with written permission by
ATMOS.
79853 Lenzkirch
Subject to alterations. Errors excepted.
Deutschland
Abbreviations / symbols in these operating instructions:
Telefon: + (49) 7653 689-0
Indicating a list
• Subdivision of a list/activity.
Fax:
+ (49) 7653 689-190 The recommended sequence must be followed in each
case!
+ (49) 7653 689-493 (Service Center)
Indicating particularly important advice!
e-mail: [email protected] Describing the effect of an activity.
Internet: http://www.atmosmed.de
Application organ:
mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
ENT unit: Short term use on the patient (up to 30 days)
Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60 minutes)
Application site:
Application sites are clinics and practices for ENT doctors and
phoniatrists. The examination and/or therapy with the ENT unit
may only be executed by medically trained persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses.
The ear irrigation should not be applied to an infected auditory
canal or a perforated eardrum.
I ON (power) as to
IEC 417/5007 and DIN 30600/16
Water temperature for ear rinsing
too high
O OFF (power) as to
IEC 417/5007and DIN 30600/16
Correct water temperature for ear rinsing
Suction system
Button for decreasing lamp intensity
Compressed-air system
Dangerous voltage as to
IEC 417/5036, DIN 30600/131
Cleaning in flow procedure (hose rinsing)
Protective earth (ground) as to
IEC 417/5019, DIN 30600/1545
Cold-light system
Fuse as to
IEC 417/5016, DIN 30600/0186
Microscope connection (cold light)
Type B equipment
Short-time operation
Do not lean against it!
Footswitch
The ATMOS® C 31 is produced according to IEC 601 / The ENT unit requires clean water (drinking-water quality)
EN 60601 and listed in the following classes: for the operation. In case the clean water cannot be pro-
• VDE Class of protection 1 vided by the water supply, a pre-filter has to be installed.
• Class IIa (EEC 93/42). The relevant country specific regulations for the installation
The unit may be connected only to a socket outlet with have to be considered !
earthing contact installed according to the rules of the trade. The control panel must be well visible to and in reach of
The unit should not be positioned directly next to a wall, the operator.
because of the ventilation openings on the rear side! Do not place used contaminated instruments on the ENT
When a microscope is mounted: Screw the unit into the unit!
ground or position it near a wall. If the microscope is in an The ambient conditions specified in section 9.0 "Technical
unfavourable position it could tilt the unit backwards. specifications" must be strictly observed!
Attention! Mirror and endoscope heaters may generate The suction system of the ATMOS® C 31 is only to be used
temperatures above 40°C! for the suction of fluids in medical ranges. Never remove
Attention with the cold-light source! explosive, inflammable or corrosive gases or fluids.
Because of the high energy of the light there is a large Switch off main switch after finishing work in practice and
amount of heat emission at the point of the optical system. close water supply, if present.
Avoid too small a distance between the tissue and the field
of light emission of the light guide resp. of the endoscope, The ATMOS® C 31 may be operated only in rooms used
as this may cause coagulation of the patient's tissue. When for medical purposes, but not in areas subject to explosion
using the endoscope avoid the direct contact between area hazards and in oxygen rich environments.
of light emission and the tissue.
Attention, Fire Hazard! The ATMOS® C 31 fully complies with the electromagnetic
Do never place the area of light emission from the light immunity requirements of standard IEC 601-1-2 / EN
guide or from the endoscope onto heat-absorbing surfaces 60601-1-2 "Electromagnetic compatibility - Medical
(dark pieces of cloth, etc.), because this will cause Electrical Equipment".
unacceptable high heating or even ignition of the material.
Switch the light off when you do not require the light over The ATMOS ® C 31 may not be operated with units
a prolonged period of time. not complying with the requirements of standard EN
60601-1 "Medical Electrical Equipment" and EN 60601-
Care is to be paid in respect to light sources when working 1-2 "Electromagnetic compatibility (Medical Electrical
with endoscopes. Harmful heat rays are eliminated by Equipment)".
special built-in filters: but the intensity of the light is very
high. Do not look directly into the light outlets! In case ATMOS is not liable for personal injury and damage to
of possible light failure remove the endoscope from the property if
working area. • no original ATMOS parts are being used,
• the advice for use in these operating instructions is not
The ATMOS® C 31 may be used in supervised operation
being observed,
by qualified personnel only which has been authorised
by ATMOS and which has been trained for operating the • assembly, new settings, alterations, extensions and
appliance (IEC 601-1/EN 60601-1). repairs have been carried out by personnel not authorised
by ATMOS.
The mains voltage specified on the type plate must
correspond with the data of the power supply system. This operation manual corresponds with the construction
of the unit and with the current status of safety-related
Make sure prior to every application of the equipment that standards at the time of printing. Proprietary rights are
it is technically safe and in proper condition. Damaged existing for all described circuits, processes, names,
leads and hoses must be replaced immediately! software programs and units.
Display instruments and valves must be checked for correct Please pay also attention to the safety information in
function in regular intervals! following chapters.
Inspection of compressed air and vacuum display by Please note:
service technician every 2 years! A medical insulating transformer with earth leakage
Correct configuration in assembly of country-specific monitor or any similar safety system acc. to EN 60601-1
connections: is required, if several devices are connected over one
green/yellow: protective conductor (PE) common power supply. The transformer must correspond to
blue: neutral conductor (N) the power consumption of all the devices to be connected.
black resp. brown: phase (L)
When connecting several devices on one grounding
Units with water consuming systems may only be switched receptacle, the allowed strain and leakage current have
on when the water supply is guaranteed! to be observed!
This product is not re-sterilisable. Repeated reuse of Never leave the patient unattended at the treatment unit.
components which are marked with a 2 is forbidden.
In case of repeated reuse these components lose their It is not allowed to use flammable substances with the
function and there is a high infection risk. device.
2 3 4 5 6
18
17
16 7
15
8
14 13 10
12 11
®
Fig. 2. ATMOS C 31 Service compartment
Connection nipple with hose to the vacuum pump (green coding)
Secretion canister
Hose rinsing container (optional)
Secretion hose (black coding)
Hose for automatic secretion canister
evacuation (optional) (transparent coding)
Connecting electrodes for automatic
secretion canister evacuation (optional)
Attachment for secretion hose
Lid (version with bacterial filter plate)
10
Connect waste water hose with unit (, Fig. 5) and the
G3/4" connection fitting (insert pertaining seals).
Fig. 5. Connection for wase water hose
Inlet connector for non-heating apparatus
3.7 Starting-up the ATMOS® C 31
Electrical connection
• Join the inlet connector for non-heating apparatus(,
fig.5) with the earth socket outlet using the mains
cable supplied.
11
Fig. 8. CPC-Coupling
Connecting the hose for ear rinsing (optional)
The hose for the warm water system is screwed on the
CPC coupling.
12
13
Purpose:
• Suction of fluids and secretions; collecting the secre-
tion in the secretion canister.
Automatic activation of the suction mode by taking out the
suction attachment.
Control of suction rate by means of rotary knob for vacuum
control ( , fig. 11).
Indication of set vacuum at vacuumgauge (, fig. 11) (keep
suction attachment closed by hand).
Secretion is collected in 1.25 l canister with mechanical
overflow safety and water-repellent bacterial filter.
Prevents ingress of secretion into the pump.
14
Fig. 12b Disposable secretion canister systems Please observe the valid hygiene and waste
! disposal measures.
Only the recommended ATMOS disposable
secretion canister with integrated filter may
! be used!
15
Headlamp
Microscope
Light guide 1
Light guide 2
16
Fig. 15. Mirorr quick heater
Key
Heating coil
17
19
The tank for the rinsing fluid of the hose rinsing is located
in the service compartment at the front of the unit. For
optimum cleaning, add the ATMOS special cleanser
(Art.No. 080.0005.0) to the rinsing liquid. The tank may,
optionally, be automatically filled up with fresh water. Water
supply is then controlled by a level switch. A safety switch
in the tank holder next to the tank prevents the filling if the
tank is missing.
Remove the basin from its support and hold it under the
ear to be rinsed.
20
Purpose:
• a) Medicaments can be applied to the nasopharynx.
21
After the switch ( , Fig. 26) has been turned on, the
headlamp connection is ready for operation.
The connector of the headlamp, e.g. CLAR 55 Art.
No. 502.0161.0, is plugged into the jack ( , Fig. 25)
(connection cable is included in standard delivery of this
option).
When connecting nystagmus binoculars (cable available
as accessory, Art.No. 502.0141.6), the voltage is reduced
automatically (4V) even if you work with low voltage.
22
4.3.11 Endoscope heating (option)
Endoscope heating is switched on when the main switch
is activated.
Endoscopes are warmed to approx. 40°C.
The door of the waste bin (, fig. 1, page 7) is fitted with
a "kick box" locking device. The waste bin automatically
opens a little by lightly touching the door with the hand
Fig. 29. Tonue patches and swab dispenser or foot.
23
Fig. 31. Tulip-shaped ear rinsing bowl 4.3.17 Hygiene filter for the ear irrigation
system (option)
Connect the disposable hygiene filter to the irrigation
handle. Exchange it daily, as specified in the hygiene plan.
Insert the filter in the direction of the water flow between
the handle and the nozzle.
! Maximum load: 5 kg !
24
5.1 General information on cleaning and 5.1.3 Secretion canister (without autom.
disinfection secretion canister evacuation),
bacterial filter and suction hose
Set main switch to OFF prior to cleaning and
disinfection! At the end of every working-day, following parts must be
cleaned and disinfected:
The described measures for cleaning do not re
place the respective precautions for operating the Secretion canister with lid system and bacterial filter:
unit ! • Detach all hose connections carefully on the lid system
and take secretion canister out carefully to prevent spills and
For disinfection, you may use all surface and instrument contamination of the area. Dispose of secretions
disinfectants listed on page 26. properly.
• Grip lid system firmly, open lid of filter housing by turning
Always observe the concentration specifications and in anti-clockwise direction.
instructions by the respective manufacturer ! • For systems with bacterial filter plates or integrated
bacterial filter:
Do not use Remove filter and dispose of it.
• Disinfectants containing concentrated organic or an- • With external bacterial filter:
organic acids or bases, since these may cause To open the bacterial filter cover please turn it. Remove
corrosion damages; filter and dispose of it.
• Disinfectants containing chloramides, phenol deri- • After cleaning a new filter must be inserted (with
vatives or anionic tensides, since these may cause bacterial filter plate: smooth side down).
stress cracks in the material used for the housing of See chapter 6.1.
the unit. • The bacterial filter is a disposable and must be disposed
of. The bacterial filter must be exchanged at least once
a day.
25
000.0219.0 506.5121.0
! The sprayer tube must be exchanged after each patient.
000.0577.0
When fixing the twin tube nozzle again, make sure
that the mark (0,X or milling area) on the nozzle shows
Fig. 32. upwards !
506.5120.0
055.0028.0
505.0280.0
005.0026.0
Sprayer tubes are available as spare parts (see fig. 32 /
fig. 33).
506.5227.0
(055.0028.0)
(505.0280.0)
(505.0258.0)
(005.0026.0)
000.0577.0
(055.0021.0)
5.1.6 Instrument trays
Fig. 33.
506.5225.0 Before disinfection, thoroughly rinse the trays under running
water. You may, of course, add a cleanser (deter-gent) or
surface disinfectant.
• Thoroughly rinse the trays under running water.
26
27
Other surfaces
If using aldehyde-containing or amine-containing disinfectants at the same object, this may result in discolourations.
28
29
30
open
31
Fig. 35. Hose to the secretion canister
Release lever
Tube cassette
Drainage hose
Connecting nipple to the rear wall
32
Grid and sleeve might get very hot. Allow them to cool
down before changing the heating coil !
Fig. 36. Mirror quick-heater
Grid
Fig. 37. Removing the heat
33
Fix the light module again with the two locking screws.
Fig.38/c.
34
35
Fig. 40. Fuse support
Before contacting the ATMOS Service, please check your unit acc. to the following charts.
Failure with
the Unit
Main switch is activated, no voltage – no voltage to power plug • check fuse, connect other units (lamp)
at the unit, no function, pilot lamps to the socket, if necessary
do not light up – fuse is defective • replace 10A fuse on the rear of the
unit
– power plug or cable is defective • check power plug or cable and
replace, if necessary
Mirror is not warmed up with the – no voltage • check voltage by connecting other con-
mirror quick heater suming devices
– heating coil is defective • replace heating coil (section 6.3)
Mirror preheater remains cold – heating plate body in the mirror • have heating plate body replaced by a
preheater plate is defective service technician
– no voltage • check voltage by connecting other con-
suming devices
36
Failure with
the Unit
Weak suction or no suction at all – suction hose is clogged • rinse the suction hose with water
(to do so hose can also be removed)
– float of overflow safety closes the • check filling level in the secretion
suction opening canister
– lid of secretion canister is not
closed tightly • check that lid of the secretion canister
– bacterial filter is clogged is closed properly
– hose connections are leaking • replace bacterial filter
– connecting hoses in the secretions • seal up or replace connections
canister compartment are broken • check hose connections, remove
– secretion penetrated the suction kinks
pump • have pump cleaned by service
technician
No suction, but vacuumgauge indica- – suction hose is clogged • rinse the suction hose with water
tes -0.7 bar (to do so hose can also be removed)
– float of overflow safety closes the • check filling level in the secretion
suction opening canister
– bacterial filter is clogged
– connecting hoses in the secretion • replace bacterial filter
canister compartment are broken • check hose connections, remove
– overflow safety is closed kinks
• hang up suction attachment, open
regulating valve and tap on the secretion
canister; the float balls of the overflow
safety must drop back into the resting
position
No suction and suction motor does – photoelectric barrier / electronics • clean photoelectric barrier
not start defective or photoelectric barrier • have photoelectric barrier checked by
is contaminated a service technician
No suction; suction motor fails to – compressed-air handle and • position handles corresponding to
start although compressed-air suction hose in handle support switching function
pump starts transposed
37
Failure with
the Unit
No light at one light exit – halogen bulb burnt out • replace halogen bulb (chapter 6.4)
Failure with
the Unit
Compressor fails to start although – compressed-air handle and • position handles corresponding to
suction motor starts suction hose in handle support switching function
transposed
Pressure of compressed air is – hose connections are leaking • check hose connections
too low < 2.3 bar • have hose connections inside the unit
checked by a service technician
Compressed air does not switch off – photoelectric barrier is • clean photoelectric barrier
contaminated or defective • have photoelectric barrier replaced
by a service technician
Sprayer does not work – sprayer is clogged • clean sprayer (section 5.1.5)
– twin tube nozzle wrongly added • see section 5.1.5
38
Failure with
the Unit
Failure with
the Unit
Failure with
the Unit
Secretion is no longer sucked off – electrode is contaminated • switch off unit and clean electrode
– tube pump defective • replace tube cassette
Failure with
the Unit
Rinsing container is overflowing – float-type switch is blocked • clean float tube (re-establishment of
– float-type switch is defective free movement of float)
• inform service technician
39
Failure with
the Unit
Rinsing basin does not properly – sieve in rinsing bowl is clogged • clean or replace the sieve
empty – waste water hoses are clogged • have waste water hoses checked by
a service technician or plumber
Basin rinsing does not function at – rinsing head is clogged (calcified) • decalcify rinsing head
all or only insufficiently – rinsing bowl is leaking • replace rinsing bowl
Endoscope supports are no longer – overtemperature protection switch • Have the switch / control / sensor
heated was activated exchanged by your service technician.
– control is defective
If, nevertheless, the errors cannot be removed inform the ATMOS service staff. Do not start any attempts to repair the unit
yourself !
40
41
8.2.1 Suction
Accessories: Article-No.
Holder for ear speculae / Politzer olives ........................................................................................................508.0545.0
Politzer olives, Teflon, universal size .............................................................................................................000.0241.0
Politzer olives, Teflon, for children .................................................................................................................000.0241.1
Sprayer attachment, straight ..........................................................................................................................505.0280.0
Twin tube with nozzle .....................................................................................................................................000.0219.0
Sprayer complete, for oily medicaments ........................................................................................................506.5120.0
Sprayer complete, straight .............................................................................................................................506.5225.0
42
Fibre optic cable, length: 1.8 m, 90° Storz angular connection .....................................................................508.0664.0
Headlamp CLAR 73, fibre optic cable 2.3 m, Storz, with angular connection................................................502.0515.0
Headlamp, light model, fibre optic cable, straight connection ........................................................................502.0515.5
Headlamp acc. to Binner, Wolf, fibre optic cable ...........................................................................................502.0516.0
Headlamp CLAR 55, warm light.....................................................................................................................502.0161.0
Connecting cable for headlamp CLAR 55......................................................................................................502.0142.6
Nystagmus binoculars acc. to Dr. Blessing ....................................................................................................502.0189.0
Connecting cable for headlamp CLAR 55......................................................................................................502.0141.6
Teflon element for endoscopes, diameter 2.8 mm - 4 mm .............................................................................508.0777.5
Tank for flexible endoscopes (ATMOS, Olympus) .........................................................................................508.0790.0
Tank for flexible endoscopes (Storz) ..............................................................................................................508.0792.0
Adaptor sleeve for Storz connection ..............................................................................................................507.0940.5
Adaptor sleeve for Olympus connection ........................................................................................................507.0940.6
Adaptor sleeve for Pentax connection ...........................................................................................................507.0940.7
Adaptor sleeve for Wolf connection ...............................................................................................................507.0940.8
Instrument tray set, melamine, consisting of 2 large and 2 small trays .........................................................506.7031.0
Instrument tray set, aluminium-anodized, consisting of 2 large and 2 small trays,
holder for ear speculae / Politzer olives .........................................................................................................506.7032.0
Instrument tray set, stainless steel, consisting of 2 large and 2 small trays,
holder for ear speculae / Politzer olives .........................................................................................................506.7033.0
Ear rinsing bowl .............................................................................................................................................505.0353.0
Plastic quiver for disinfecting solution, to be mounted to cable holder ..........................................................506.7015.0
Instrument intermediate-deposit with paper in large aluminium resp. stainless steel tray ............................. 508.0533.0
Connecting cable for potential balance (5 m) ................................................................................................008.0596.0
Instrument tray, aluminium-anodized, 184 x 142 mm ....................................................................................508.0058.0
Instrument tray, aluminium-anodized, 284 x 184 mm ....................................................................................505.0516.0
Instrument tray, stainless steel, 180 x 140 mm ..............................................................................................508.0058.2
Instrument tray, stainless steel, 280 x 180 mm ..............................................................................................505.0516.2
Instrument tray, melamine, 190 x 150 mm .....................................................................................................000.0746.0
Instrument tray, melamine, 300 x 190 mm .....................................................................................................000.0747.0
Serrated instrument holder, big ......................................................................................................................508.0566.0
Serrated instrument holder, small ..................................................................................................................508.0567.0
Holder for ear speculae / Politzer olives, for aluminium resp. stainless steel tray .........................................508.0545.0
43
44
Voltage 230 V~ ± 10 %
50/60 Hz
Special voltage (optional): 110 V~ ± 10 % ; 127 V~ ± 10 %
50/60 Hz
Current input max. 5 A (230 V~)
max. 10 A (110 / 127 V~)
Supply Fixed connection, power cable: 5 m
Power consumption max. 1150 VA
Air flow rate of pump 40 l pump: 40 l/min ± 10 % (free flow), vacuum: 90 % of ambient pressure;
55 l pump: 55 l/min ± 10 % (free flow), vacuum: 97 % of ambient pressure;
secretion canister 1,25 l TPX
Cold-light system 4-channels-cold light; processor-controlled, halogen pin socket lamp 24V/150 W,
durability: 300 h
Ear rinsing system Water temperature adjustment, 37°C ± 2 ° C, filling quantity 4.5 l,
flow approx. 450 ml/min
Optional compressed-air system 21 l/min ± 10 % (free flow), pressure 2200 hPa
Operating time Continuous operation
Fuses T 6,3 A (f. 230 V~, 50/60 Hz), T 12 A (f. 110 V~, 127 V~, 50/60 Hz)
Protective earth conductor resistance max. 0,1 Ω
Earth leakage current max. 0,5 mA
Enclosure leakage current max. 0,1 mA
Patient leakage current max. 0,1 mA
Ambient conditions
Transport/storage -10...+50°C; 30...95 % humidity without condensing
air pressure 500...1060 hPa
Operation +10...+35°C; 30...95 % humidity without condensing
air pressure 700...1060 hPa
Dimensions HxWxD Column
with 1 Base Cabinet 93 x 47 x 55 cm
with 2 Base Cabinets 93 x 128 x 60 cm
Weight (basic unit) Column max. 60 kg (with all options)
with 1 Base Cabinet max. 70 kg
with 2 Base Cabinets max. 80 kg
Period tests Repeat test of the electrical safety every 24 months.
Recommended: inspection according to the manufacturers specifications.
Protection class (EN 60601-1) I
Degree of protection Type B equipment
45
46
After each
procedure
staff who are familiar
Recommendations with reprocessing.
C D S
Monthly
Weekly
(Please fill in the
Reusable parts Cleaning Disinfection Sterilisation responsible person
Daily
-> use a water-based
overhead marker)
Secretion canister
Housing for external
X X2)4)5) Manual or automatic cleaning and disinfection X
bacterial filter
/ Bacterial filter cover X X 2)4)5)
Manual or automatic cleaning and disinfection X
Bacterial filter plate with
/ blue marking / Disposable. Exchange daily or when blocked X
bacterial filter
Sprayer head X X2)4)5) Multiple rinsing of the sprayer head with water X
Weekly exchange of the hose or when changing the
Hose at sprayer head X X X
medication
Cleaning and disinfection (manual or automatic)
Sprayer bottle X X 2)4)5)6)
X
weekly or when changing the medication
Exchange after each application;
Politzer olive X X2)4)5)6) X
afterwards cleaning and disinfection
Exchange after each application;
Politzer connection X X2)4)5)6) X
afterwards cleaning and disinfection
47
After each
procedure
staff who are familiar
Recommendations with reprocessing.
C D S
Monthly
Weekly
(Please fill in the
Reusable parts Cleaning Disinfection Sterilisation responsable person
Daily
-> use a water-based
overhead marker)
Endoscope management
Plastic quiver X X2)4)5) Cleaning with a brush; disinfect afterwards X
Instrument management
Immerse instruments into solution immediately after use,
fully wetting is required, air must be removed from any
cavities, after the contact time instruments must be rinsed
ENT instruments X X2)4)5) X X
with water, has to be dried and sterilised afterwards. Ple-
ase also observe the ATMOS operating
instructions for ENT instruments.
Instrument bowl with co- Cleaning with a brush;
X X2)4)5) X
ver disinfect afterwards (manual or automatic)
Visualisation
ATMOS® Cam 21 / 31 X X3) Wipe cleaning and wipe disinfection X
ATMOS® Strobo 21 LED X X3) Wipe cleaning and wipe disinfection X
Flexible scope X X1),7),8) X1) Immediate pre-cleaning after the procedure X
Surfaces
Housing X X3) Wipe cleaning and wipe disinfection X
Roller shutter X X3) Wipe cleaning and wipe disinfection X
Drawers X X3) Wipe cleaning and wipe disinfection X
Recommended disinfectants 4)
Instruments - manual disinfection: 7)
Endoscopes - manual disinfection: The above stated hygiene requirements are based on the regulations
according to the Medical Devices Act, the Medical Devices Operator
• Korsolex® basic (Bode Chemie) • Helipur® H plus N (BBraun) Ordinance, §18 IfSG and the recommendations of the Robert Koch
3)
Surface disinfection for coated surfaces: • Korsolex® plus (Bode Chemie) • Helix® Ultra (BBraun) Institute.
• Green & Clean SK (ATMOS) • Korsolex® extra (Bode Chemie) • Korsolex® Basic (Bode Chemie) Definition of the required reprocessing steps result from the recom-
• Dismozon® plus (Bode Chemie) neodisher® Septo MED (Dr. Weigert) • neodisher® MediClean forte (Dr. Weigert) mendations of the Robert Koch Institute: „Requirements for the repro-
cessing of medical products“. The medical products were categorised
• Kohrsolin® FF (Bode Chemie) • neodisher® Septo 3000 (Dr. Weigert) • Sekusept® aktiv (Ecolab) in the risk groups uncritical, semicritical and critical. The reprocessing
• Perform® (Schülke & Mayr) • Sekusept® aktiv (Ecolab) measures mentioned in this cleaning
Endoscopes - automatic disinfection:
8)
• Terralin® Protect (Schülke & Mayr) Gigasept® Instru AF (Schülke & Mayr) and disinfection plan are a recommendation of ATMOS Medizin-
• Korsolex® Basic (Bode Chemie)
• Technik. Any additional reprocessing measures are at the operator‘s
• Gigazyme® (Schülke & Mayr) •neodisher® MediClean forte (Dr. Weigert) discretion.
Other surfaces: • Gigasept FF neu (Schülke & Mayr) • Gigasept® FF neu (Schülke & Mayr) All the recommended disinfectants which are stated herein are listed
• Dismozon® plus (Bode Chemie) • Endozime® AW Plus (Ruhof)
disinfectants (VAH/RKI) and have been tested on their suitability of
use on the ATMOS® C 21 / C 31. ATMOS MedizinTechnik cannot be
• Kohrsolin® FF (Bode Chemie) 5)
Instruments - automatic disinfection: • ADAPTACLEAN™ (ASP) hold liable for any damage caused by wrong concentration of the dis-
• Mikrobac® forte (Bode Chemie) • Dismoclean® 24 Vario (Bode Chemie) infectants or by the application of any other disinfectants.
• Perform® (Schülke & Mayr) • Dismoclean® 28 alka med (Bode Chemie) Important information Patients with suspicion of a clinical disease or who developed a
transmissible spongiform encephalopathy (CJK, vCJK, etc.) have to
Terralin® Protect (Schülke & Mayr) • Dismoclean® twin basic/twin zyme (Bode Chemie) be treated at facilities which are able to provide for the necessary
• Surface disinfection FD 312 (Dürr Dental) Wipe cleaning and wipe disinfection: All surfaces
• neodisher® FA (Dr. Weigert) preventive measures against infection. The reprocessing of the reu-
• Quick disinfection B 30 (Orochemie) have to be wiped with a clean (disposable) wipe sable instruments and material may only be performed at facilities
• neodisher® MediClean forte (Dr. Weigert) which have an externally certified QM Management acc. to DIN EN
which is damped with disinfectant solution; the
• Thermosept® alka clean forte (Schülke & Mayr) ISO 13485.
entire surface has to be wiped thoroughly and may
• Thermosept® RKN-zym (Schülke & Mayr) The Medical Devices Act, IfSG, the RKI directives, BGR 250 and
not be dried afterwards. TRBA 250 always have to be considered.
1)
Please observe the manufacturer‘s operating
Please see the manufacturer‘s instructions for con- Wrong concentration of disinfectants instructions. ATMOS MedizinTechnik GmbH & Co. KG
2)
Alternative to manual cleaning und disinfection: Ludwig-Kegel-Str. 16 79853 Lenzkirch/Germany
centration, contact time, temperature and the com- may lead to damage!
Wash-Disinfector 78°C / 172°F Phone +49 7653 689-0 Fax +49 7653 689-190
patibility of materials.
[email protected] www.atmosmed.de
48 6)
Material dimensionally stable at 134°C
2015-08 Index: 07
■ Portable and mobile HF communication facilities can influence medical electrical equipment.
■ The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
The device may not be used directly next to other devices or piled up with other devices.
If operation next to or piled with other devices is necessary, please watch the device to check its intended
operation in this arrangement.
The ATMOS® C 31 is intended for use in the electromagnetic environment specified below. The customer or user of
the ATMOS® C 31 should ensure that it is used in such an environment.
49
< 5 % UT < 5 % UT
(>95 % Dip of the UT) (>95 % Einbruch der UT)
for 5 s für 5 s
NOTE UT is the mains alternating current prior to application of the test levels.
50
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines might not be applicable in any case. The propagation of electromagnetic sizes is
influenced by absorptions and reflections of buildings, objects and people.
a The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radio
equipment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly.
To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to
be considered. If the measured field strength at the location where the ATMOS® C 31
is used exceeds the above compliance level, the ATMOS® C 31 is to be observed to
verify the intended use. If abnormal performance characteristics are noted, additional measures might
be necessary, e. g. a changed arrangement or another location for the device.
b Within the frequency range of 150 kHz to 80 MHz the field strength is to be below 3 V/m.
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2 These guidelines are not applicable in any case. The propagation of electromagnetic sizes is
influenced by absorptions and reflections of buildings, objects and people.
51
www.atmosmed.com