PDF Service Manual Discovery PDF Compress
PDF Service Manual Discovery PDF Compress
PDF Service Manual Discovery PDF Compress
DISCOVERY
QDR® Series
FAN BEAM X-RA
X-RAY
Y BONE DENSITOMETER
TECHNICAL MANUAL
December 2010
Notice
The information contained in this manual is confidential and proprietary to Hologic, Inc.
This information is provided only to authorized representatives of Hologic's customers
cu stomers
solely for the purpose of facilitating the use of Hologic's products. No information
contained herein may be disclosed to any unauthorized person for any purpose whatsoever
without the prior written consent of Hologic, Inc.
The procedures described in this document are intended for use by direct employees of
Hologic, Inc., or authorized Hologic Equipment Resellers and their trained field
engineers. Any unauthorized or untrained persons performing these procedures may affect
the warranty of the Discovery QDR Series.
Exercise proper caution when servicing the system. There are dangerous and potentially
lethal voltages accessible within the Discovery system. To avoid exposure to shock
hazards, the Main Circuit Breaker should be switched off, and the power cord removed,
before working inside any part of the system.
The Discovery QDR Series Fan Beam X-ray Bone Densitometer produces ionizing
radiation in the form of X-rays. It may be dangerous
d angerous to the patient, operator or field
engineer unless safe exposure factors and operating instructions are observed. To avoid
unsafe exposure, do not attempt to service the equipment unless you are a Hologic, Inc.,
certified field engineers. Exercise proper caution when servicing the system. A dosimeter
(film badge) should always be worn while on site. Dose and scatter measurements must be
taken after each service call to ensure that the parameters are still within specifications.
Hologic, Inc., has made all effort to ensure that the information in this manual is accurate
and complete. Hologic, Inc., shall not, however, be liable for any technical or editorial
errors or omissions contained herein, or for incidental, special or consequential dangers in
connection with the furnishing or use of this manual.
manua l. The information in this manual is
subject to change without notice.
Discovery and the Hologic logo are registered trademarks of Hologic, Inc. All other
products and company names used in this manual are trademarks and registered
trademarks of other manufactures.
Printed in U.S.A.
Copyright© 2003-2010 by Hologic, Inc., All rights reserved
ii
Discovery QDR Series Technical Manual
Chapter 1 -
INTRODUCTION
1.1 System Overview .............................................
...............................................................................................
.................................................. 1-1
1.1.1 X-Ray Scanning Principles
Principles .......................................................................
........................... ............................................ 1-2
1.2 Functional Overview .......................................................................................... 1-4
1.3 Product Specifications ................................................................
.....................................................................................
..................... 1-10
1.3.1 Exam Mode(s) Performance ................................................................... 1-12
1.3.2 Duty Cycle: ...........................................................................................
................................... ..........................................................
.. 1-13
1.3.3 Leakage Technique Factors .................................................................... 1-13
1.3.4 Minimum Beam Filtration ..........................................................
......................................................................
............ 1-13
1.3.5 Measured Half Value Layer (HVL) At Different Operating Potentials . 1-13
1.3.6 Line Voltage and Maximum Line Current ............................................. 1-13
1.3.7 Technique Factors for Maximum Line Current ......................................
................................... ... 1-13
1.3.8 Maximum Deviation ...............................................................................
............................. .................................................. 1-14
1.3.9 Measurement Criteria for Technique Factors ......................................... 1-14
Chapter 2 -
FUNCTIONAL DESCRIPTION
2.1 Computer ........................................................................
.. .........................................................................................................
................................... 2-1
2.2 PCI Communications Controller Boar Board d .............................................................
................................ ............................. 2-1
2.2.1 Interface Connections ...............................................................................
................................................................... ............ 2-1
2.3 Distribution Board .....................................................
.............................................................................................
........................................ 2-3
2.3.1 Power ..............................................................................................
......................... ...............................................................................
.......... 2-3
2.3.2 Interface Connections ...............................................................................
................................................................... ............ 2-4
2.4 Motor Controller Board .....................................................................................
................................................................. .................... 2-5
2.4.1 Power ..............................................................................................
......................... ...............................................................................
.......... 2-5
2.4.2 Interface Connections ...............................................................................
................................................................... ............ 2-6
2.5 TZ Drive Board (A and SL Only) ...................................................................... 2-6
Chapter 3 -
INSTALLATION
3.1 Pre-Installation
Pre-Installation Requirements ............................................................................ 3-1
3.1.1 Required Tools .......................................................................................... 3-1
3.1.2 Required Documentation ..........................................................................
.................................................................... ...... 3-1
3.1.3 Room
Room and Doorway Size ..........................................................................
............... ........................................................... 3-2
3.1.4 Arrange For Help ..............................................................................
.. ....................................................................................
........ 3-6
3.2 Uncrate and Move to Destination ...................................................................... 3-6
iv Table of Contents
Discovery QDR Series Technical Manual
3.4.3
3.4.4 Adjust Belt
Calibrate Tension
Motors ................................................................................
............................................................ ....................
....................................................................
.....................................................................................
................. 3-32
3-33
3.4.5 X-Ray Beam Alignment ...............................................................
........................................................................
......... 3-47
3.4.6 Calibrate Aperture (Cannot be run on Ci and Wi models) ..................... 3-53
3.4.7 Check Laser Positioning Offset ..............................................................
............................................... ............... 3-54
3.5 A/D Gain Control
Control Adjustment
Adjustment .........................................................................
........................ ................................................. 3-54
3.5.1 Check and Verify
Verify the A/D Gain ............................................................. 3-54
3.5.2 Adjust
Adjust the A/D Gain ...............................................................................
............................................................... ................ 3-55
3.5.3 Perform Detector Flattening ............................................................
..................................................................
...... 3-55
3.5.4 Perform Lateral Alignment Test (A and SL) .......................................... 3-56
3.5.5 Check HVPS/S (Tank)
(Tank) For Radiation Leakage ...................................... 3-58
3.5.6 Calibrate For Area, BMD And BMC ...................................................... 3-60
3.5.7 Scan
3.5.8 Scan Thickness
Thickness Measurement
Measurement & & Calibration
Calibration (A (C and
and W) SL)........................
....................... 3-61
3-61
3.5.9 Calibration of Area and BMC, for Array Scan Modes Modes ........................... 3-62
3.5.10 RECALYZE and Add Array AP Scans to the QC Database ................ 3-63
3.5.11 Install Software Options .......................................................................
............................................................ ........... 3-64
3.5.12 Test Scan Modes ..........................................
...................................................................................
......................................... 3-64
3.5.13 Finish
Finish Assembling Unit ........................................................................
............................................................... ......... 3-66
3.5.14 Measure X-Ray Dose To Patient .......................................................... 3-66
3.5.15 Measure X-Ray Scatter From Phantom ................................................
......................... ....................... 3-67
3.5.16 Perform QC ........................................................................
...........................................................................................
................... 3-67
3.5.17 Run Reproducibility Test ......................................................................
.................................................... .................. 3-67
3.5.18 Table Top Radiographic Uniformi Uniformity ty (A, W and Wi) ........................... 3-68
3.5.19 Instruments using
using Body Composition
Composition Analysis (BCA) ........................ 3-69
3.5.20 Instruments using BMD Whole Body Analysis ................................... 3-70
3.6 The Radiation Measurement
Measurement Report ................................................................
................................. ............................... 3-70
Chapter 4 -
ALIGNMENT AND CALIBRATION
4.1 Check Table Alignment ................................................................................
................ .....................................................................
..... 4-1
4.2 Align the Table ..................................................................................................
.................................................................. ................................ 4-1
4.2.1 Table Edge to T-Rail (“A” Dimension) Adjustment ................................ 4-1
4.2.2 Front to Back T-Rail and Table Edge/Rail Gap Adjustment .................... 4-3
4.3 C-Arm Parallelism Adjustment .........................................................
........................................................................
............... 4-3
4.4 X-Ray Beam Alignment ...................................................................................
.............................................................. ..................... 4-4
4.5 Aperture Calibration .........................................................................................
.................................................................... ..................... 4-9
Table of Contents v
Discovery QDR Series Technical Manual
4.14.4 RECALYZE and Add Array AP Scans to the QC Database ................ 4-38
Chapter 5 -
REMOVE AND REPLACE PROCEDURES
5.1 Recommended Tools .........................................................................................
............................................................. ............................ 5-1
5.2 Electronics
Electronics Tray FRUS
FRUS ......................................................................................
................. ..................................................................... 5-1
5.2.1 Electronics Tray Printed Circuit Boards ................................................... 5-2
5.2.2 C-Arm Y Motor or Gearcase .................................................................... 5-3
5.2.3 C-Arm
C-Arm Y Encoder ..............................................................................
.. ...................................................................................
....... 5-4
5.3 Table Y FRUs (A, W and Wi
Wi Only) .................................................................. 5-4
5.3.1 Control Panel (All
(All models)
models) ..............................................................
.......................................................................
......... 5-5
5.3.2 PCBs
5.3.3 Table Under
Y Belt Right-Side of the Table ........................................................
................................. .......................
..............................................................................................
........................................................ ...................................... 5-5
5-6
5.3.4 Table
Table Y Motor or Gearcase ...................................................................... 5-7
5.3.5 Table Y Encoder ...........................................................................
.......................................................................................
............ 5-7
5.4 Table X FRUS ...............................................................
...................................................................................................
.................................... 5-8
5.4.1 Table X Motor Controller PCB ................................................................ 5-9
5.4.2 Table X Belt ............................................................................................
........................................................ .................................... 5-10
5.4.3 Table
Table X Motor or Gearcase .................................................................... 5-10
5.4.4 Table X Encoder ...........................................................................
.....................................................................................
.......... 5-11
5.5 Table Z FRUs (A and SL only) ......................................................................
..................................................................... . 5-12
5.5.1 Pedestal ...................................................................................................
........................................................ ........................................... 5-12
5.5.2 The Linear Rotary String (Encoder) ....................................................... 5-14
5.6 Lower C-Arm FRUS ........................................................................................
...................................................................... .................. 5-16
vi Table of Contents
Discovery QDR Series Technical Manual
Chapter 6 -
FAULT ISOLATION
Chapter 7 -
PREVENTIVE MAINTENANCE
7.1 Customer Preventive Maintenance .................................................................... 7-1
7.2 Field Service Preventive Maintenance .............................................................. 7-1
7.2.1 Guide Rail and Bearing Maintenance .......................................................
...................... ................................. 7-3
Chapter 8 -
PCB SUMMARY INFORMATION
Chapter 9 -
SOFTWARE TOOLS
9.1 X-Ray Survey ....................................................................................................
...................................................................... .............................. 9-1
9.1.1 Invoking X-Ray Survey ............................................................................
................................................................ ............ 9-1
9.1.2 The Main Display .....................................................................................
.................................................................. ................... 9-2
9.1.3 The Display Controls ................................................................................
.................................. .............................................. 9-3
9.2 SQDRIVER .........................................................................
......... ..............................................................................................
.............................. 9-9
9.3 SQVERIFY .............................................................................................
......................................................................................................
......... 9-10
9 -10
9.4 SQKEYPAD .........................................................
....................................................................................................
........................................... 9-10
Appendix A -
ERROR MESSAGES
the
sizescanning arm over
of scan desired, the
the region of
operator interest.
initiates theAfter
scan entering patient
with a single data and selecting the type and
keystroke.
The operator is not required to select technique factors as tube current and voltage are pre-
selected and fixed. Since testing is performed by fan beam method, rather than by flooding the
area as in conventional radiography, the scanning time is a function of the dimensions of the area
to be measured, the desired resolution, and the desired precision.
BMC results are expressed in grams of calcium hydroxyapatite and
an d BMD is reported in grams/
cm2 of the same compound.
In most cases, no additional shielding is necessary for patient,
pa tient, operator, or room. The Discovery
system can be placed in any convenient non-shielded examination room. Contact your state
regulatory agency for details about additional shielding requirements, if any
an y.
1-1
Discovery QDR Series Technical Manual
1-2
Discovery QDR Series Technical Manual
2. Esti
Estima
mate
te k as
as an
an aver
average
age valu
valuee of:
of:
k = [Ltissue - Lair] / [Htissue - Hair]
where Ltissue indicates a low-energy measurement with tissue-equivalent material
interposed by the filter drum, and L air, Htissue and Hair are similarly defined.
Note: The subscript
subscript "air" designates the filter drum segment that is empty (i.e., contains
neither bone- nor tissue-equivalent material).
3. Using this value of k, calculate
calculate Q for each
each point
point scanned
scanned using the formul
formulaa given
given
above (Q = L - kH). This array of Q values constitutes a "Q scan". Displays the Q
scan.
4. Compile
Compile a histogr
histogram
am of the Q values.
values. Because
Because a large
large portion
portion of the scan
scan con-
tains soft tissue only, this histogram will have a large peak.
p eak. Choose a threshold
value just above this peak, and apply that value to discriminate, point by point in
the Q scan, between "bone" points (whose Q is above the threshold) and "non-
bone" points (whose Q is below the threshold).
1-3
Discovery QDR Series Technical Manual
1-4
Discovery QDR Series Technical Manual
1-5
Discovery QDR Series Technical Manual
AC
AC from
AC Power
DIN Rail
Strip
(Scanner)
AC
Printer Monitor
Computer
AC
Video
Phone
Modem
Keyboard Jack
Network
Network I/F Card
Jack
Mouse PCI Discovery
Communications Scanner
Controller
1-6
Discovery QDR Series Technical Manual
To/From AR
Motor
Communications JP10 AR Stepper
Controller Motor Driver AR
Encdr
TY
Motor
TY Stepper
JP5
Signal Motor Driver TY
Encdr
Distribution
Board
+/-15V JP6 TX
From DIN Motor
TX Stepper
Rail JP7
Motor Driver TX
and LVPS +24V
JP8 Encdr
AY
AY Stepper Motor
JP11
Motor Driver
AY
Encdr
TZ
Motors
TZ Stepper
Motor Driver TZ
AC
Encdrs
1-7
Discovery QDR Series Technical Manual
motors, turn power to the instrument on or off, and an indicator showing when X-rays are
being generated.
The C-Arm Interface Board controls the C-Arm Subsystem. Commands to the subsystem
arrive from the Distribution Board. The commands processed by the C-Arm Interface
Board are sent to the Positioning Laser, the X-Ray Controller (XRC), the Aperture Motor,
the Detector assembly, or the Filter Drum Assembly. The commands sent to the
Positioning Laserwhich
produce X-rays, are simple
poweron/off
level commands. Commands
of pulse to produce, andsent
the to the mode
pulse XRC tell it when
to be used. to
These commands will subsequently control the way the X-Ray Source
S ource (Tank) is driven by
the XRC. Instructions to the Aperture Motor are for selecting the aperture to be used for
the scan. The commands to the Filter Drum Assembly turn the Filter Drum on or off and
synchronize the filter drum with the AC line waveform. The last set of command signals is
sent to the Detector Assembly to synchronize the acquisition of data with the production
of the X-rays from the source (Tank).
Detector Assembly
Laser
Assembly
Detector Board
(Detectors, Muxes,
and A/D Converter)
X-Ray X-Ray
Source
Controller
(Tank)
From
DIN
Rail
1-8
Discovery QDR Series Technical Manual
Interface Board and the Distribution Board. Mathematical algorithms applied to this data
produce the scans seen by the operator and to calculate the BMD, BMC, and area of the
bone matter scanned.
Block Description
Computer Controls an
and cco
ommands al
all Di
Discovery ha
hardware mo
modules.
PCI Communications Controls the flow of commands to and from the Scanner modules via the
Controller Board communications bus.
Distri
Distribut
bution
ion Boar
Board
d Pro
Provid
vides
es the
the interc
interconn
onnect
ection
ionss betwee
between
n the Di
Disco
scover
very
y Operat
Operator'
or'ss Con
Consol
solee and th
thee
Scanner and distributes DC voltages.
Cont
Contro
roll P
Pan
anel
el Prov
Provid
ides
es swit
switch
ches
es fo
forr man
manua
uall
lly
y mov
movin
ing
g the
the C-Ar
C-Arm
m aand
nd Pati
Patien
entt Tab
Table
le.. Als
Also
o
provides Emergency Stop and Instrument On/Off switches.
TX Stepper Motor Controls the motion of the Patient Table
Table in and out motor and monitors table
Driver position information from the encoder
encoder..
TY Stepper Motor Controls the motion of the Patient Table
Table left to right motor and monitors table
Driver position information from the encoder (A and W only).
TZ Stepper Motor Controls the motion of the Patient Table left and right pedestal motor and monitors
Driver table position information from the string encoders (A and SL only).
AR Stepper Motor Controls the rotational motion of the C-Arm and monitors C-arm position
Driver information from the encoder (A and SL only).
AY Stepper Motor Controls the motion of the C-Arm left and right motor and monitors C-Arm
Driver position information from the encoder
encoder..
C-Arm
C-Arm Inte
Interfa
rface
ce Con
Contro
trols
ls the Apertu
Aperture
re and Fil
Filter
ter Drum
Drum moto
motors,
rs, gen
genera
erates
tes timin
timing
g aand
nd con
contro
troll sign
signals
als
for the X-Ray Controller and the Data Acquisition System, monitors the tape
switches, and provides power to the Positioning Laser.
X-
X-Ra
Ray
y Cont
Contro
roll
ller
er Cont
Contro
rols
ls tthe
he oper
operat
atio
ion
n of th
thee X-ra
X-ray
y Sour
Source
ce..
X-
X-Ra
Ray
y Sou
Sourc
rcee Uni
Unitt Ge
Gene
nera
rate
tess tthe
he X-
X-ra
ray
ybbea
eam.
m.
De
Dete
tect
ctor
or Boa
Board
rd Co
Conv
nver
erts
ts the
the X-ra
X-rays
ys iint
nto
o elec
electr
tric
ical
al sig
signa
nals
ls.. In
Inte
tegr
grat
ates
es the
the si
sign
gnal
alss fr
from
om the
the So
Soli
lid
d
State Detectors and converts them to a digital value in the Analog-to-Digital
converter circuitry.
Cont
Contro
roll Pan
Panel
el Prov
Provid
ides
es th
thee ope
opera
rato
torr wit
with
h a mea
means
ns to repo
reposi
siti
tion
on the
the tab
table
le,, C
C-A
-Arm
rm,, and
and to turn
turn o
on
n
the laser.
Positi
Positioni
oning
ng Lase
Laserr Provid
Provides
es a laser
laser ccros
rossha
shair
ir b
beam
eam to assi
assist
st in
in po
posit
sition
ioning
ing the pat
patien
ientt on the Pat
Patien
ientt
Table.
1-9
Discovery QDR Series Technical Manual
1-10
Discovery QDR Series Technical Manual
1-11
Discovery QDR Series Technical Manual
1-12
Discovery QDR Series Technical Manual
1-13
Discovery QDR Series Technical Manual
1-14
Section 2
FUNCTIONAL DESCRIPTION
This section provides a detailed functional description along with interconnection diagrams and
descriptions of the Hologic Discovery. Refer to Section 1 for a block diagram and a brief
functional description of each block.
2.1 Computer
The Discovery Scanner is interfaced to a PCI Communications Controller Board/ISA Bus
computer which controls table and C-arm movement, X-ray
X -ray generation, performs all necessary
calculations, and manages both the patient and QC database information.
The computer is a Pentium III-based (or higher) PC that comes equipped with floppy disk drive,
hard disk drive, keyboard, mouse, 17” color monitor, and CDROM R/W for archiving scans and
backups of the system database. For details pertaining to the computer and its associated
components, please refer to the documentation shipped with each unit.
2-1
Discovery QDR Series Technical Manual
2-2
Discovery QDR Series Technical Manual
PCI
Communication DISTRIBUTION
Command BOARD
Board
KP1085_004-0201
this board
Control is a Interface
Panel separate section ofathe
Board is Detector
separate Board.
board In the
located A and
under the SL versions, the
table.
The Distribution Board has provision for three jumpers that can be installed to override
the EMERGENCY signal lines when troubleshooting.
2.3.1 Power
The Distribution Board receives +24 and +/-15VDC from the Multi-voltage DC Switching
Power Module. The +24VDC is applied through four individual circuit breakers to the
Table X, Table
Table Y,
Y, Arm R, and Arm Y Motor Controller Boards.
Boards. The +24 and +/-15VDC
are passed to the C-Arm Interface Board. The +24 VDC is also reduced to +7 and +5VDC
by regulators to power op-amps and analog switches located on this board. The +7VDC is
passed to the Control Panel Controller Board. The +5VDC powers the digital section of
2-3
Discovery QDR Series Technical Manual
the Distribution Board. The +24VDC power supply is not closely regulated and its outputs
may range from +24V to +35V under normal conditions.
Note: +7VDC may measure anywhere from
from +6.25VDC to +7.25VDC. This is true
everywhere +7VDC is shown in this manual.
Limits for +/- 15VDC
Six green LEDs indicate the status of the +28 (on QDR-4500s, +24 on Discovery), +24,
+15, -15, +7 and +5VDC
+5VD C (ON indicates the respective voltage is present). Five red LEDs
indicate the status of the five circuit breakers applying voltage
vo ltage to the motor drivers/
controller.. ON indicates the circuit breaker has been
controller b een tripped by an over-current condition.
OPERATOR’S
CONSOLE
COMPUTER
COMMUNICATIONS
CONTROLLER
BOARD
INSTRUMENT
ELECTRONICS
TRAY
2-4
Discovery QDR Series Technical Manual
2.4.1 Power
Power input to the Motor Controller Board is +24VDC. This voltage provides the motor
drive power and is converted down
do wn to +5VDC for use by logic ccircuits
ircuits on the board. The
+5VDC is also converted to -5VDC. Two green LEDs provide a visual indication of the
power present on the Motor Controller Board.
2-5
Discovery QDR Series Technical Manual
2-6
Discovery QDR Series Technical Manual
board is located in the Electronics Tray in the rear of the bottom of the Scanner base
assembly.
The TZ Drive Board communicates with the PCI Board, via the Distribution Board, to
drive the pedestal motors under computer control. Manual repositioning of the pedestal
may be required in case of an emergency. In this case, manual control is provided through
the Table switches of the Scanner’
Scanner ’s Operator Control Panel.
The TZ Drive board monitors the position of both pedestals using signals received from an
associated Position Encoder connected to the respective pedestal.
The TZ drive motors are designed to run at a 5% duty cycle. If the motors overheat, the
built-in thermal cutouts may trip and cause the motors to stop functioning. If this occurs,
you must wait about 20 minutes before functionality is restored.
Table 2-4.TZ Drive Service Switches
Switch Function
Modee (Normal/
Mod (Normal/Ser
Servic
vice)
e) Determ
Determine
iness whether
whether the
the TZ Drive
Drive is in
in Normal
Normal or Servi
Service
ce
operation.
Di
Dire
recti
ction
on (U
(Up/
p/Dow
Down)
n) When
When the
the TZ
TZ driv
drivee iiss in
in Ser
Servi
vice
ce mode
mode,, dete
determ
rmin
ines
es th
thee
direction of pedestal movement (not active in normal
mode).
Left When the TZ drive is in Service mode, moves the left
pedestal in the direction specified by the Direction switch
(not active in normal mode).
Right When the TZ drive is in Service mode, moves the right
pedestal in the direction specified by the Direction switch
(not active in normal mode).
Reset Resets the board after manual operation. The TZ Drive
board must be reset after any manual operation.
2.5.2 Power
Power input to the TZ Drive Board is +24VDC from the Distribution Board and 240VAC
240VAC
from the DIN Rail. The +24VDC powers circuitry located on this board and is reduced to
2-7
Discovery QDR Series Technical Manual
+5VDC. The +5VDC is converted to -5VDC and +/-3VDC. The +/-5VDC powers logic
circuitry on this board, while the +/- 3VDC provides the reference voltage for the position
sensors. Two
Two green LEDs provide visual indication of the +24 and +5VDC status (ON
indicates the respective voltage is present).
The 240VAC
240VAC power is connected
c onnected through control
co ntrol relays to the pedestal
pe destal motors.
2-8
Discovery QDR Series Technical Manual
2-9
Discovery QDR Series Technical Manual
2.6.1 Power
The Control Panel Controller Board receives
rece ives +7VDC from the Distribution Board. The +7
is reduced to +5VDC to power the circuitry on this board and is applied to the Operator
Control Panel to power the LEDs. A green LED, on this board, provides a visual
indication of the +5VDC power (ON indicates the voltage
v oltage is present).
2-10
Discovery QDR Series Technical Manual
To/From
Distribution Board
ARD+, ARD- L0_PWR* - L7_PWR*
ATD+, ATD- TZ_PWR*
SYSRESET+, SYSRESET- XRAY_LIGHT_PWR*
EMERGENCY+, EMERGENCY- SW0 - SW2
To/From
TILT_A
C-Arm
TILT_B
Tilt Switch
(A/SL)
2-11
Discovery QDR Series Technical Manual
Pin Pin
L0_PWR* Turns on the ENABLE switch LED. JP6-11 JP1-11
L1_PWR* Turns on the HOME switch LED. JP6-12 JP1-12
L2_PWR* Turns on the LOAD switch LED. JP6-13 JP1-13
L3_PWR* Turns on the TABLE switch IN/OUT LED. JP6-14 JP1-14
L4_PWR* Not used. JP6-15 JP1-15
L5_PWR* Turns on the C-ARM switch RIGHT/LEFT JP6-16 JP1-16
LED.
L6_PWR* Not used. JP6-17 JP1-17
L7_PWR* Turns on the Laser LED. JP6-18 JP1-18
XRAY_LIGHT_PWR* Turns on the X_RAY LED JP6-19 JP1-19
SW0 Control signal to determine the state of the JP6-4 JP1-4
C-ARM and LASER switches.
SW1 Control signal to determine the state of the JP6-5 JP1-5
TABLE IN/OUT switches.
2-12
Discovery QDR Series Technical Manual
the DAS. It also provides buffering for various signal lines. The C-Arm Interface Board is
located near the rear of the lower C-Arm’s
C-Arm’s horizontal shelf.
Note: On the Discovery A model, the Data
Data Acquisition System
System (DAS) consis
consists
ts of the
the
Analog to Digital Converter, Integrator/Multiplexor, and Solid State Detector
Boards. On the Discovery SL, C, and W models, a single-board, 128-Channel
Detector Assembly is used.
The C-Arm Interface provides circuitry to:
• Control
Control and
and monitor
monitor operation
operation of of the
the X-ray
X-ray Control
Controller
ler Board.
Board. Four LEDs (two red,
one green, and one
o ne yellow) on this board, provide a visual indication of the status
of the X-Ray Controller and the X-ray control circuitry of this board.
• Genera
Generatete timin
timingg referen
references
ces to
to the DAS and and PCI Commu
Communicnicati
ation
on Comman
Command d Board.
Board.
• Cont
Contro
roll powe
powerr to the
the Pos
Posititio
ioni
ning
ng Las
Laser
er..
• Move the Aperture
Aperture stepper motor and monitomonitorr its
its mechanically
mechanically linked position-
position-
sensing device.
• Move
Move the steppe
stepperr motor
motor of thethe Referen
Reference ce Drum
Drum device
device and monit
monitor
or the
the encoded
encoded
signals returned from each drum of the device. Two green LEDs, on this board,
provide a visual indication of the Reference Drum operation.
• Gene
Generarate
te +7V
+7VDCDC forfor the
the DAS
DAS sys
syste
tem
m
2.7.2 Power
The C-Arm Interface Board receives +24 and +/-15VDC from the Distribution Board. The
+24VDC powers the Aperture and Reference Drum stepper motor driver circuitrycircuitry.. The +/-
15V is passed through this board to the DAS. The +24VDC is also reduced to +5 and
+7VDC. The +5VDC powers the digital
d igital section of this board and the laser while the
+7VDC is applied to the DAS.
Three green LEDs, on this board, provide a visual indication of the +24 and +5VDC
power (ON indicates the respective voltage is present). The third LED monitors the
+28VDC used on the QDR-4500 models.
mode ls. On the Discovery, this voltage is actually
+24VDC.
2-13
Discovery QDR Series Technical Manual
CONTINUITY 1
ATD_CARM+, ATD_CARM-
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
ARD_CARM+, ARD_CARM-
SRD+, SRD-
SRCLK+, SRCLK-
SRFRM+, SRFRM-
EMERGENCY_CARM, EMERGENCY_CPANEL
EMERGENCY_CPANEL
LINESYNC+, LINESYNC-
INTEGRATE+, INTEGRATE-
C-ARM
DISTRIBUTION SYSRST_CARM+, SYSRST_CARM-
INTERFACE
BOARD EMERGENCY+, EMERGENCY-
BOARD
XRAY_LIGHT+, XRAY_LIGHT-
+15V
15V_RET
-15V
24V
24V_RET
28V
28V_RET
CONTINUITY 2
2-14
Discovery QDR Series Technical Manual
Signal Description 1 2
Dist C-ARM
Pin Pin
SRCLK+ Synchronizes data through the C-Arm Interface JP1-21 JP1-21
SRCLK- Board from the DAS. JP1-22 JP1-22
SRFRM+ Synchronous channel data frame through JP1-24 JP1-24
SRFRM- Distribution Board to PCI Communication JP1-25 JP1-25
Command Board from the DAS.
XR_ZEROX_CC+ AC line zero-crossing signal used for system wide JP1-30 JP1-30
XR_ZEROX_CC- synchronization. Generated by the C-Arm JP1-31 JP1-31
Interface Board.
INTEGR_CC+ Synchronous signal for Detector Integrate period. JP1-33 JP1-33
INTERG_CC- Generated by the C-Arm Interface Board. JP1-34 JP1-34
SYSRST_CARM+ Resets the C-Arm Interface Board. JP1-39 JP1-39
SYSRST_CARM- JP1-40 JP1-40
EMERGENCY_CARM+ Removes power from the Scanner motor drivers JP1-42 JP1-42
HW_EMGNCY_RET- and the X-ray system JP1-43 JP1-43
XRAY_LIGHT+ Applies power to the AUX X-RAY
X-RAY light outlet on JP1-48 JP1-48
XRAR_LIGHT- the Power Console. JP1-49 JP1-49
2-15
Discovery QDR Series Technical Manual
LASER
ASSEMBLY
ANALOG TO
INTEGRATOR SILICON
DIGITAL
MULTIPLEXOR DETECTORS
CONVERTER
TO/FROM
DISTRIBUTION APERTURE
C-ARM MOTOR AND
INTERFACE
SENSOR
DRUM
MOTOR AND
ENCODER PICKUP
X_RAY
X-RAY
CONTROLLER SOURCE
UNIT
2.8.1 Interface
Figure 2-8 shows the interface connections between the C-Arm Interface Board and the X-
Ray Controller Assembly.
Assembly. Table 2-11 describes the interface signals and identifies the
connectors and their pin assignments. Note that the AC input power comes directly from
the DIN Rail Power Module and connects
c onnects to the X-Ray Controller Assembly.
2-16
Discovery QDR Series Technical Manual
XR_RLY_ON+, XR_RLY_ON-
XR_FREQ+, XR_FREQ-
HIGH VO LT
LTAGE
AGE
XR_BEAMON+, XR_BEAMON-
TRANSFORMER
XR_ISET+, XR_ISET- PRIMARY
XR_KV1+, XR_KV1-
C-ARM XR_KV0+, XR_KV0-
INTERFACE
ACLINE+, ACLINE- X-RAY X-RAY
BOARD
LIGHTON+, LIGHTON- SOURCE
IBEAM+, IBEAM- CONTROLLER
ASSEMBLY UNIT
KVP+, KVP-
XRFAULT+, XRFAULT- X_RAY TU BE
X_RAY
FILAMENT
TRANSFORMER
From DIN Rail 240 VAC PRIMARY
Power Module
2-17
Discovery QDR Series Technical Manual
2-18
Discovery QDR Series Technical Manual
capacitors. The current load in the H-bridge is sensed and adjusted. The output v
voltage
oltage
from the H-Bridge is up to 200 volts peak.
The Filament circuit is a regulator. It receives current feedback from the tank indicating
the tube current. 10 volts
vo lts at the feedback input is equivalent to 1
100 milliamps of current in
the tube. The 10 volts are reduced to 5 and then fed to an A/D circuit to be converted to a
digital value for use by the control circuits. The value into the A/D is sampled towards the
end
to anofEPROM.
the pulseIfwhere it is stable.
the current
c urrent Afterthe
is correct, conversion, theEPROM
output of the digital value is fed
will be as anAnything
0 volts. address
other than 0 volts (“+” = high current, “-“= low current) will be added and accumulated in
two stages forming an error integrator. This digital error number is converted back to an
analog value and then used as the threshold of a comparator. The output of the comparator,
through additional circuitry,
circuitry, adjusts the duty cycle of the modulated filament voltage.
Refer to Table 2-11 for pin assignments.
rectifier
tube andcircuit, andcurrent.
the beam the sensing circuits that monitor the high voltage applied to the X-ray
Table 2-12 describes the interconnections between the XX-Ray
-Ray Controller Assembly and
the X-Ray Source unit. The table also identifies the X-Ray Controller Assembly board
connector and pin assignments for each interconnection signal.
Table 2-12. X-Ray Controller Assembly/X-Ray Source Unit Interface
Signal Description XRC Board Controller Pin Tank
Terminal
HV_XFMR_PRI+ High Voltage Transformer H-Br
H-Briidg
dgee (J
(JP1)
P1) JP1-
1-1,
1, JP1
P1-2
-2 TB1-3
HV_XFMR_PRI- Primary JP1-3, JP1-4 TB1-4
FIL_XFMR_A Fila
Filame
ment
nt Tra
Trans
nsfo
form
rmer
er Pri
Prima
mary
ry I/
I/O
O & Logi
Logicc JP2-7 TB1-6*
FIL_XFMR_B (JP6) JP2-8 TB1-9
IF+ Beam Current Sense "+" I/O & Logic JP2-4 TB1-10
(JP6)
IF- Beam Current Sense "-" I/O & Logic JP2-5 TB1-7
(Chassis) (JP6)
TP1 Anode pulse monitor I/O & Logic JP2-2 TB1-12
(JP6)
TP2 Cathode pulse monitor I/O & Logic JP2-1 TB1-11
(JP6)
2-19
Discovery QDR Series Technical Manual
crystal attached. X-rays striking the cadmium tungstate crystals are converted
c onverted into photons
of visible light. The diodes sense this light and convert the light into a current, which is
amplified in a current to voltage converter. The converter output is applied to an integrator
through analog switches. The analog switches all operate in parallel and are turned on
during the X-ray pulse. These switches are turned off during the integrator hold time to
prevent the introduction of integration noise into the data.
da ta. The turning on and off of these
switches is controlled by the INTEGRATE signal under control of the software. The
signal from the detectors is then integrated with the final voltage obtained
o btained held in a
sample/hold circuit. This voltage is sampled, in photodiode order, by the multiplexor
circuitry under control of the software. The output voltage is then sent to the analog-to-
digital conversion circuitry where it is changed to a digital value usable by the software.
The Detector Assembly is located in the upper
uppe r end of the C-arm directly above the X-ray
source. There is a lead radiation shield between the detector board and the cover to stop
any X-rays that may get through the detectors.
2-20
Discovery QDR Series Technical Manual
2.10.4 Power
The 128-Channel Detector Assembly receives +/-15V and +5V from the C-Arm Interface
Board. The +/-15V is passed through this board to the Detector subsection. It is also
regulated to +/- 12V by series regulators to power operational
o perational amplifiers and analog
switches located on this board. The +/- 15V is also used to generate +/- 5V to power the
analog-to-digital converter circuit. The +7V is reduced to +5V to power the digital section
of this board.
STFRM+, STFRM-
SRD+, SRD-
SRCLK+, SRCLK-
SRFRM+, SRFRM-
SYSRST_DAS+, SYSRST_DAS-
C-ARM INTERFACE INTEGRATE+, INTEGRATE- 128-Channel
BOARD XR_ZEROX_DAS+, XR_ZEROX_DAS- Detector
Assembly
+15V
15V_RET
-15V
+7V
2-21
Discovery QDR Series Technical Manual
STD+
STD- Synchronous data to the
Converter Analog/Digital
board. JP10-3
JP10-4 P1-12
P1-13
STCLK+ Synchronizes data to the Analog/Digital JP10-6 P1-15
STCLK- Converter board. JP10-7 P1-16
STFRM+ Synchronizes DSP Communications. JP10-9 P1-18
STFRM- JP10-10 P1-19
SRD+ Synchronous Data from the Analog/Digital JP10-12 P1-3
SRD- board. JP10-13 P1-4
SRCLK+ Synchronizes data from the Analog/Digital JP10-15 P1-6
SRCLK- Converter board. JP10-16 P1-7
SRFRM+ Synchronizes DS
DSP Co
Communications. JP10-18 P1-9
SRFRM- JP10-19 P1-10
SYSRST_DAS+ Re
Rese
sets
ts the An
Anaalo
log/
g/Di
Digi
gittal Co
Con
nve
vert
rter
er bo
boaard. JP10
JP10-2
-21
1 P1-21
SYSRST_DAS- JP10-22 P1-22
INTEGRATE+ Data integration signal. Generated by the C- JP10-24 P1-24
INTEGRATE- Arm Interface Board. JP10-25 P1-25
XR_ZEROX_DAS+ AC line zero-crossing signal used for system JP10-27 P1-27
XR_ZEROX_DAS- wide synchronization. Generated by the C-Arm JP10-28 P1-28
Interface Board.
+15V JP10-32 P1-32
JP10-33 P1-33
Notes:
Notes: 1.C-Arm
1.C-Arm = C-Arm
C-Arm Interf
Interface
ace Board
Board
2.DA=128-Channel Detector Assembly
2-22
Discovery QDR Series Technical Manual
2.11.2
2.11.2 Power
Pow er
The Solid State Detector board receives +/-15V from the Integrator/Multiplexor board.
Voltage regulators, located on this board, convert this voltage to +/-12V to power the
amplifiers and switches.
CHANNEL0 - CHANNEL215
SILICON
INTEGRATOR/MULTIPLEXOR DETECTOR
INTEGA - INTEGF
BOARD BOARDS
+15V
15V_RET
-15V
2-23
Discovery QDR Series Technical Manual
Notes:1.Refer to schematic drawing 140-0048-SD for connector and pin assignments (cannot scope these
signals).
2.Refer to schematic drawing 140-0050-SD for connector and pin assignments (cannot scope these
signals).
3.Note that the CHANNEL hardware numbers are reversed from the software numbers. e.g.
Hardware
CHANNEL0 = software CHANNEL215.
2-24
Discovery QDR Series Technical Manual
2.12.1 Power
The Integrator/Multiplexor board receives +/-15V and +5V from the Analog to Digital
board. The +/-15V is passed through this board to the Solid State Detector boards. Voltage
Voltage
regulators, located on this board, convert this voltage to +/-12V to power circuitry
contained on this board. Analog and digital returns are kept separate.
+15V
15V_RTN
-15V
+5 V
G ND
2-25
Discovery QDR Series Technical Manual
2-26
Discovery QDR Series Technical Manual
2.13.1 Power
The ADC board receives +/-15V and + 7V from the C-arm Interface board. The +/-15V is
passed through this board to the Integrator/multiplexor board. It is also reduced to +/-12V
by series regulators to power op-amplifiers and analog switches located on this board. The
+/-15V also generates +/-5V to power the analog-to-digital
ana log-to-digital converter circuit. The +7V is
reduced to +5V to power the digital section of this board. The +5V is also passed on to the
Integrator/Multiplexor board.
2-27
Discovery QDR Series Technical Manual
STD+, STD-
STCLK+, STCLK-
STFRM+, STFRM-
SRD+, SRD-
SRCLK+, SRCLK-
SRFRM+, SRFRM-
SYSRST_DAS+, SYSRST_DAS-
INTEGRATE+, INTEGRATE- ANALOG/DIGITAL
C-ARM INTERFACE
BOARD XR_ZEROX_DAS+, XR_ZEROX_DAS- CONVERTOR
BOARD
+15V
15V_RET
-15V
+7V
2-28
Discovery QDR Series Technical Manual
2-29
2 29
Section 3
INSTALLATION
*Max. Patient Weight for systems built prior to February 2006 is 350 lb (159Kg)
Note: with
Twothe
System Backup
scanner disks
and the areshipped
other shippedin with
the each
cratesystem.
with theOne
PC.will
Markbe the
included
one
shipped with the scanner and note
no te that if the Serial Numbers on the babackup
ckup disks
do not agree, the disk shipped with the scanner is the correct disk.
Installation Inventory Check List
(Quantities are one (1) each unless otherwise specified)
[ ] Main Discovery Assembly [ ] Voltage Selection Kit (100/120/230)
[ ] Printer Paper [ ] C-Arm Assembly
[ ] Discovery User's Guide [ ] Tabletop Pad
[ ] Computer, Mouse, and Keyboard [ ] Hip Positioner
[ ] Laser or Business Inkjet Printer [ ] Knee (Block) Elevation Pad
[ ] Spine Phantom [ ] Foot Restraint
[ ] VGA or Flat Panel Monitor [ ] CDROM R/W
[ ] Upper C-Arm Assembly [ ] 3.5" Floppy Disks
[ ] Discovery Console [ ] Miscellaneous Hardware Box
[ ] Communication Cables [ ] System Backup Disks (2)
[ ] Pedestal Covers (Left) [ ] C-Arm Cap and Shoulder Covers
Stainless Steel
Cable Guard
Bearing Block
Front Endcap
3-9. Table
Figure 3-9. Table Locking “L” Bracket Location
3-11
Before the Discovery can be tilted to vertical, you must slide the table to the left and clamp
it so it clears the floor when the unit is tilted
tilted to vertical. To accomplish this, do the
following:
1. Remo
Remove
ve the
the ttabl
ablet
etop
op pad.
pad.
2. Remove thethe three Phillip
Phillipss screws
screws holding
holding the tabletop.
tabletop. Two
Two of the screws
screws are
are at
the left corners and one at the right back corner looking down on the tabletop.
3. Slide the
the tabletop
tabletop to the right
right until
until it clears the table
table base panel.
panel.
4.. Refe
4 Referr
rrin
ing
g to Figure 3-10,
3-10, adjust the Table End Stop as pictured.
Bracket
Important: The screws that hold the auxiliary caster bars must be re-tightened
(Figure 3-12)
3-12) to prevent the unit from dropping when moved.
Ramps
2. Remove the
the caster bars
bars and cross
cross brace.
brace. Position
Position the unit
unit in its
its final location
location in
the room.
3. Leve
Levell the un
unit
it..
4. Remove
Remove the
the 2 carriag
carriage-t
e-to-b
o-base
ase shipp
shipping
ing bracke
brackets
ts (Figure
(Figure 3-19).
3-19).
5. Discovery A, W & Wi - Remove the table end stop (Figure
(Figure 3-10).
3-10).
6. Remove the
the table
table locking
locking bracket
bracket located
located on the
the left rear side
side of the scanner
(Figure 3-9).
3-9).
3-16
1
1.. Refer to Figure 3-20.
3-20. Reposition the belt tensioning mechanism. The belt ten-
sioning mechanism is turned around to facilitate shipping and moving the assem-
bly (the shipping bracket is not used after the mechanism is repositioned).
2. Remove
Remove the
the back C-arm
C-arm cover
cover (2
(2 Phillip
Phillipss screws
screws).
).
3. Remove the
the two 1/2"
1/2" bolts that lock
lock the upper
upper C-arm in
in place (left
(left in place
place during
shipment).
4. Remove the
the screws
screws for the
the C-arm shoulder
shoulder cover (left
(left in place
place during
during shipment).
shipment).
5. Remove the
the two C-arm
C-arm retaini
retaining
ng brackets
brackets (to
(to allo
allow
w upper C-arm
C-arm to
to be set in
in
place).
6. Remove
Remove the
the cover
cover from
from the
the upper
upper C-arm.
C-arm.
7. Install
Install the upper
upper C-arm
C-arm onto
onto the lower
lower C-arm
C-arm (tilt
(tilt the front
front of the upper
upper C-arm up
to slide it in place).
8. Instal
Installl the
the retai
retainin
ning
g bracke
brackets.
ts.
3-17
Shipping
Bracket
2. Confirm
Confirm that you have the
the proper
proper kit for
for the site’
site’ss AC voltage
voltage (100/120
(100/120//
230VAC).
3-18
3. Measure
Measure the voltage
voltage between
between ground
ground and neutral
neutral at the wall
wall outlet
outlet that
that will
will be
used for the system. If it is more than TBD mV call Hologic Technical Support
before continuing.
4. Remove foot
foot end cover
cover on the
the Electroni
Electronics
cs Tray
Tray (on the
the left side facing
facing the
the
machine) and the cover of the Torroid Assembly
Assembly..
5. Remove the
the AC input cable from
from the kit,
kit, install
install the strain
strain relief,
relief, and
and route the
the
cable though bottom of the Torroid Assembly.
6. Screw the
the strain
strain relief
relief into
into the bushing
bushing at the
the bottom
bottom of the Torroi
Torroid
d Assembly
Assembly and
fix the cable to the exposed stud using the cable clamp and NyLock nut provided
in the kit.
Grn/Yel
TRANSFORMER
Grn/Yel
TRANSFORMER
Grn/Yel
TRANSFORMER
8. Install
Install the circuit
circuit
roid Assembly breaker
breaker
using from
from the
the screws voltage
voltageinkit
provided kithe
t into
kitthe
the hole in the
the rear of the
the Tor-
Tor-
9. Attach the
the two wires
wires (brown
(brown and blue)
blue) from TB1
TB1 to the Line
Line terminals
terminals (top
(top ter-
ter-
minals) on the circuit breaker.
10. Attach the two wires (brown
(brown and blue) from TB2 to the Load terminals (bottom
(bottom
terminals) on the circuit breaker.
11. Use the jumpers provided to
to configure the terminal block as shown in the tables
below.
Wire From To
Blue Circuit Breaker - Load Pin 1 – TB2
C z Brown Circuit Breaker – Load Pin 7 – TB2
A H
V 0
6 Blue Jumper Pin 1 – TB2 Pin 4 – TB2
0 /
2 0
1 5 Brown Jumper Pin 3 – TB2 Pin 7 – TB2
Wire From To
C z Blue Circuit Breaker - Load Pin 1 – TB2
A H
V 0 Brown Circuit Breaker – Load Pin 6 – TB2
0 6
/
3 0
2 5 Brown Jumper Pin 3 – TB2 Pin 4 – TB2
3-20
Wire From To
Blue Circuit Breaker - Load Pin 1 – TB2
C z Brown Circuit Breaker – Load Pin 5 – TB2
A H
V 0 Blue Jumper Pin 1 – TB2 Pin 4 – TB2
0 6
/
0 0
1 5
Brown Jumper Pin 2 – TB2 Pin 5 – TB2
Figure 3-21
3-21.
Assembly . These
Cove
Coverr. diagrams can also be found on tthe
he inside of the T
Torroid
orroid
SAFETY PRECAUTIONS
There are a number of safety precautions that MUST be observed when servicing the
Discovery systems.
HIGH VOLTAGE: Voltage levels that can ca n injure or be fatal are present through the
Discovery systems. The line voltage (100, 120, 230 volts) is supplied to the Torroid
Torroid
Assembly and the scanner.
scanner. The X-ray source unit contains 140kV as well as other AC
and DC voltages. The pedestal motors use 230 volts regardless of the line voltage. UsUsee
caution when checking, calibrating, and troubleshooting. Always trip the main breaker
when replacing components.
X-RADIATION: Service personnel are required to wear a dosimeter.
dosimeter. Do not leave the
system unattended in X-RAY SURVEY (SURVEY mode).
ESD PRECAUTIONS:
must take precautions whenTo handling
prevent damage due to Remove
components. ESD (Electrostatic
any chargeDischarge), you
from your body
by wearing an approved and properly grounded wristwrist strap. Keep PCBs in their ESD
protective bag until you are ready to install them. Treat defective PCBs as new to
prevent any additional damage.
Note: Enter this information into the Radiation Measurement Report (see example on
page 3-70).
3-70).
3. Turn
Turn on the computer
computer and
and confirm
confirm that all options
options shipped
shipped with the
the instrument
instrument or
or
listed on the sales order have been installed.
3. Click the Start button on the desktop and then select Search For Files or Fold-
3.
ers...
4. In the
the Searc
Search
h Result
Resultss window
window,, select
select All files and folders, type service in the
All or part of the file name: edit box, and then click on the Search button.
3-24
5. The
The soft
softwa
ware
re now
now sear
searche for SERVICE. A list in the right panel will show you
chess for
the files found. In the list, click the “service” item shown in C:\QDR\Utilities
and drag it to the desktop. Close the Search window and then double-click the
Service icon on the desktop. The software will now be restarted in service mode.
3.3.5.1 Table
Table Alignment Procedure
After you have taken the measurements in the Checking Table Alignment section, use the
procedures below to align the table. Note that if both the “A” dimension and the bracket-
to-guide rail gap are out of specification, you should recheck the measurements after
performing the first adjustment.
3.3.5.2
To changeTable
Tthe
able“A”Edge to T-Rail
dimension, do the(“A” Dimension) Adjus
following: Adjustment
tment
1. Remove
Remove the
the outer
outer and inner
inner covers
covers from
from the righ
rightt pedest
pedestal.
al.
2. Loosen the four
four bolts
bolts fastening
fastening the top of
of the pedesta
pedestall and the
the four bolts fastening
fastening
the bottom of the pedestal (see Figure 3-25).
3-25).
3. Adjust
Adjust the
the table
table so the
the “A” dimen
dimensio
sion
n is within
within the
the specifi
specificat
cation.
ion.
4. Tight
Tighten
en the
the bolts
bolts and
and check
check the
the table
table ali
alignm
gnment
ent agai
again.
n.
If the alignment is within the specification, replace the pedestal covers. If you still note a
change in the “A” dimension, continue with the following steps.
3-25
+/- .06”
63.125”
FRONT & REAR
+/- .25”
“A” “A”
Studs and
Note: Even if the brackets are secured to the frames, they may not be securely
fastened to the pedestal. If not, remove the table to access the screws that
3-26
head
to Allen
make screws.
sure While
that these the table
screws is off,asitwell.
are tight is a good idea to remove the pedestal
6. Move the
the C-arm
C-arm until
until it is parallel
parallel to the tank using
using the
the digital
digital level
level to
to measure
measure
the angles.
3.4.1 Check
WARNING:
ARNING: X-r
Tube
X-rays
ays aare
kVatedPeak
re gener
generated during
Potential
during tthis
his proc
procedu
edure.
re. Keep
Keep ha
hands
nds,, head
head and o
othe
therr bod
body
y par
parts
ts out
out
of beam.
Proper operation of the Discovery requires that the X-ray tube generate X-ray pulses of
80kVp, 100kVp, 120kVp, and 140kVp, all ± 10%. The peak potential check must be
performed at installation time and whenever the X-ray source or X-ray controller is
repaired or replaced. Because it would be very dangerous to directly monitor the kVp
potentials, there is a 10,000
1 0,000 to 1 voltage divider circuit inside the High Voltage
Voltage Power
Supply/Source (HVPS/S). By monitoring this divided voltage, one can determine the peak
potentials being impressed on the X-ray Tube. The monitoring can be done on the terminal
strip on top of the X-ray Source.
WARNING:
ARNING: Alt
Althou
houghgh th
thee test
test vol
voltag
tagee is low
low, there
there are
are ele
elevat
vated
ed vo
volta
ltages
ges n
near
ear aand
nd ar
aroun
ound
d the ttest
est
points.
1. Remove
Remove the
the tank
tank cover
cover (2(2 Philli
Phillips
ps screw
screws)
s)..
Set up the scope as
a s follows:
Channel 1:2V/div (0.2V/div if using x10 probes)
Channel 2:2V/div (0.2V/div if using x10 probes)
Time base:5.0ms/div
Trigger on line (positive slope)
Set Vertical Mode to Add
Invert Channel 2.
Connect Channel 1 to TB1-PIN 12 (on the tank)
Connect Channel 2 to TB1-PIN 11 (on the tank)
AW-00336_001-0318
AW-00336_001-0319
1. If the
the tank cover is
is not off,
off, remove
remove it (2 Phillips
Phillips screws).
screws).
2. Set up the
the scope
scope as fol
follo
lows
ws::
Channel 1:2V/div (0.2V/div if using x10 probes)
Channel 2:2V/div (0.2V/div if using x10 probes)
Time base:5.0ms/div
Trigger on Line (positive slope)
Set Vertical Mode to Add
Invert Channel 2.
Connect Channel 1 to TB1-PIN 10 (on the tank)
Connect Channel 2 to TB1-PIN 7 (on the tank)
3. Ground both channels
channels and
and move the
the trace to the bottom
bottom of the screen.
screen. Remove
Remove
the grounds and go to DC Coupling.
4. Instrument
Instrument power
power,, computer
computer power,
power, and the X-ray
X-ray enable key should
should all
all be on.
5. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities|Service Utilities|X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
6. Change X-ray Mode to1 and Aperture to 7. (Note: On Ci and Wi systems, the
6.
Aperture parameter is set to a default value of -1 by the system software. Do not
change this value on Ci and Wi systems.)
7 . Click X-Rays (F2) to turn X-Rays on.
7.
8. Observe
Observe the oscillo
oscilloscope.
scope. You should
should see a trace
trace similar
similar to Figure 3-19,
3-19, approxi-
mately 4ms pulses with a peak amplitude of 3V (ignore the overshoot and mea-
sure current on the back, after
a fter it settles out). This corresponds to 3ma ± 35% tube
current.
9 . Click X-Rays (F2) to turn X-Rays off..
9.
Note: Enter this information into the Radiation Measurement Report (see
example on page 3-70).
3-70).
Changee X-ray Mode to 3.
10.. Chang
10
3-30
AW-00336_001-0320
AW-00336_001-0321
2. Turn
Turn off the
the Discovery
Discovery computer
computer power
power and main
main circuit
circuit breaker on
on the rear of
of
the foot end pedestal.
3. Remove
Remove the
the Elect
Electron
ronics
ics Tray covers
covers..
4. Loosen the two
two mounting
mounting bolts holding
holding the belt tension
tension block.
5. Adjust the tension
tension nut
nut so that
that the spring
spring is
is compressed
compressed to 7/8 inch.
inch.
6. Tighten
Tighten the two
two mountin
mounting
g bolts
bolts holding
holding the tension
tension block.
3-32
Each motor (except TZ) requires the corresponding protocol calibration file in the
PROTOCOL sub-directory (e.g., for MOTOR$AY
MOTOR$AY, the calibration protocol is
MOTOR_AY.PRO).
To perform the calibration
ca libration procedure:
1. From
From the Dis
Discove
covery
ry Main
Main Menu
Menu screen,
screen, select Utilities|Service Utili-
select
ties|SQDRIVER (you must be in Service Mode).
2. At the CARM$$$$> prompt, type MOTOR$XX<Enter>, where XX equals TZ,
2.
AY, TY, TX, or AR depending on which motor you are calibrating.
3. Measure
Measure the distance
distance moved
moved using
using the bottom
bottom edge
edge of the
the top pedestal
pedestal cover
cover and
the floor. The system displays:
3-33
3.4.4.2 MOTOR$AY
To perform the calibration
ca libration procedure:
1. Select
Select the
the AY
AY motor
motor device
device driv
driver
er by typing: MOTOR$AY<Enter>
typing:
2. At the MOTOR$AY> prompt in SQDRIVER, type: CALIBRATE<Enter>
2.
The program prompts:
Press <Enter> when the AY motor reaches the LEFT mechanical limit.
Press <ESC> to stop calibration.
3. The program
program moves the AY AY motor to
to the left.
left. When AY
AY hits the
the left mechanic
mechanical
al
stop the first time press <Esc>.
4. Check the
the position
position value.
value. It must
must be 3750±
3750±5.
5. Adjust
Adjust if necessa
necessary
ry..
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 3750±5 (i.e., in the range 3745-3755) and then tighten the
coupling setscrew.
setscrew.
3-35
5. Repeat the
the calibration
calibration procedure
procedure above
above but now,
now, when AY
AY hits the
the left mechani-
mechani-
cal stop, press <Enter>.
The program then starts AY
AY moving to the right and prompts:
Press <Enter> when the AY motor reaches the RIGHT mechanical
limit.Press <ESC> to stop calibration.
6. When
When AY
AY hits
hits the right
right mechan
mechanica
icall stop, press <Enter>.
stop, press
The program then switches to graphics mode and draws the Encoder Vs
Distance calibration grid. It steps the AY
AY motor to the left in one inch
increments until the motor hits the left mechanical stop, and then steps
the AY
AY motor to the right in one inch
in ch increments until the motor hits the
right mechanical stop. The program plots the encoder data during the
scan (in raw encoder units,
un its, 0-4095).
Note: Press <Esc> anytime during the scan to abort
a bort the calibration procedure.
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fit parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.6 and 86.5) should be within 0.3 of eeach
ach other.
The program displays the positive and negative limits as horizontal dashed lines.
3-36
6) 363 (EncoderOffset).
origin of motion. The The value
origin Asubtracted
of AYY motion isfrom the encoder
the extreme rightreadback to set
mechanical the so
stop,
this value should be the same as
a s the first field below.
below.
3-37
7,8) 363,3719 (NegLimit,PosLimit). The encoder readings for the negative (right) and
positive (left) mechanical stops.
The pos_limit_position and the neg_limit_position are the limits of motion, in microns,
in the positive and negative direction.
Note: The PosLimitOffset and NegLimitOf
NegLimitOffset
fset entries in the [AyMotor] section of the
SQDRIVER.INI
If filenot
these entries are determine
present, the motion
or are zero,limits relative
the motion to the
limits aremechanical
set to the stops.
mechanical limits, as in the above example).
Note: The last calibration scan data is
is saved in the file
file MOTOR_AY
MOTOR_AY.DAT
.DAT. You
You can
reanalyze the data–e.g., after editing SQDRIVER.INI by typing the command
CALIBRATE @MOTOR_AY.DAT at the MOTOR$AY> prompt in SQDRIVER.
3. The program
program moves
moves the TY
TY motor
motor to the
the left.
left. When TY
TY hits the left
left mechanical
mechanical
stop the first time press <Esc>.
4. Check the
the position
position value.
value. It must
must be 3750±
3750±5.
5. Adjust
Adjust if necessa
necessary
ry..
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 3750±5 (i.e., in the range 3745-3755) and then tighten the
coupling setscrew.
setscrew.
5. Repeat the
the calibration
calibration procedure
procedure above
above but now,
now, when TY hits
hits the left
left mechani-
mechani-
cal stop, press <Enter>.
The program then starts TY moving to the right and prompts:
Press <Enter> when the TY motor reaches the RIGHT mechanical limit.
Press <ESC> to stop calibration.
When TY hits the right mechanical stop, press <Enter>. The program then
switches to graphics mode and draws the Encoder Vs Distance calibration grid. It
steps the TY motor to the left in one inch increments until the motor hits the left
mechanical stop, and then steps the TY motor to the right in one inch increments
until the motor hits the right mechanical stop. The program plots the encoder
data during the scan (in raw encoder units, 0-4095).
3-38
Note: Press <Esc> anytime during the scan to abort the calibration procedure.
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fit parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.5 and 85.6)
85 .6) should be within 0.3 of ea
each
ch other.
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fit parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.5 and 85.6)
85 .6) should be within 0.3 of ea
each
ch other.
The program displays the positive and negative limits as horizontal dashed lines.
Note: The
The PosLimitOffset and NegLimitOff
NegLimitOffsetset entries in the [TyMotor] section of
the SQDRIVER.INI file determine the motion limits relative to the
mechanical stops. If these entries are not present, or are zero, the motion
limits are set to the mechanical limits.
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
3-39
neg_limit_position=0
Update Driver INI-File [Y/N] ?
Note:The last calibration scan data is saved in the file MOTOR_TY.DAT. You
can reanalyze the data–e.g., after editing SQDRIVER.INI –by typing the
command CALIBRATE @MOTOR_T @MOTOR_TY Y.DAT at the MOTOR$TY>
prompt in SQDRIVER.
3.4.4.4 MOTOR$TX
1. Select
Select the
the TX motor
motor devic
devicee driver
driver by typing:: MOTOR$TX<Enter>
by typing
2. At the
the MOTOR$
MOTOR$TX>
TX> promp
promptt in SQDRI
SQDRIVER type: CALIBRATE<Enter>
VER,, type:
This message displays on all systems. On only A and SL systems, you must use a
digital level to set the C-arm to 0º +0º. On other systems, ignore the message.
3.
3. Press <Enter> and the program prompts:
Press <Enter> when the TX motor reaches the OUTER mechanical limit.
Press <ESC> to stop calibration.
4. The program
program moves
moves the table to
to the front.
front. When
When TX hits
hits the
the outer mechanical
mechanical
stop for the first time press <Esc>.
5. Check the
the position
position value.
value. It must
must be 1000±
1000±5.
5. Adjust
Adjust if necessa
necessary
ry..
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 1000±5 (i.e., in the range 995-1005) and then tighten the
coupling setscrew.
setscrew.
6. Repeat the
the calibratio
calibration
n procedure
procedure above but
but now,
now, when TX hits
hits the
the outer
outer mechan-
ical stop, press <Enter>.
3-40
The program switches to graphics mode and draws the Encoder Vs Distance cal-
ibration grid. It steps the TX motor in by 1" increments
incre ments until the inner mechani-
cal stop is hit. The program plots the encoder
encode r data during the scan (in raw
encoder units, 0-4095).
Press <Esc> at anytime during the scan to terminate the calibration procedure.
neg_limit_position=0
Update Driver INI-File [Y/N] ?
3-41
The program moves the TY and AY AY motors to their center positions, then moves
the TZ motor to its topmost position and prompts:
Use the Table
Table IN / OUT switch to move the C-Arm to 0 degrees.
Press <Enter> when the C-Arm is positioned.
Press <ESC> to stop calibration.
4. Remove
Remove the C-arm
C-arm Tank cover
cover.
5. Place the
the digital
digital level on the top of the tank assembly (not on the top of the C-
tank assembly
arm)) and use the Table IN/OUT switch on the operator panel to move the C-arm
arm
until it is 0º ±0.0º
6.. Remo
6 Remove
ve the
the leve
levell.
7. Press <Enter>.
7.
The program prompts:
Press <ESC> to stop calibration.
3-42
The program moves the TZ table to its top most position and moves the TX table
inwards until it almost touches the C-arm. It rotates the C-arm by 2 degrees to
obtain an initial estimate of the encoder calibration and then prompts:
Press <Enter> when the AR motor reaches the AP mechanical
mec hanical limit.
Press <ESC> to stop calibration.
The program rotates the C-arm counter clockwise (i.e., the tank assembly
a ssembly moves
away from the front of the machine). When the C-arm hits the AP mechanical
limit the first time, press <ESC>.
8. Check the
the position
position value.
value. It must
must be 250±5
250±5.. Adjust
Adjust if necessar
necessary
y.
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 250±5 (i.e., in the range 245-255) and then tighten the
coupling setscrew.
setscrew.
9. Before
Before repeating
repeating the
the calibration
calibration procedur
procedure,
e, rotate
rotate the C-arm
C-arm back
back to approxi-
approxi-
mately 0º by:
Typing the command: MOVE_REL 1470<Enter>.
Place a level on top of the X-ray tank assembly (not the top of the C-arm) and
use the Table IN/OUT switch on the operator panel to move the C-arm until it is
0º ±0.1º. Remove the level, wait until the rotation completes, and then repeat the
calibration procedure above but now, when the C-arm hits the AP mechanical
stop,
3-43
Presss <Enter>.
12.. Pres
12
The program prompts:
Press <ESC> to stop calibration.
and moves the C-arm back to the approximate 0° position. The program then
prompts:
Use the Table
Table IN / OUT switch to move the C-Arm to 0 degrees.
Press <Enter> when the C-Arm is positioned.
Press <ESC> to stop calibration.
Place a level on top of the X-ray tank assembly (not the top of the C-arm) and
use the Table IN/OUT switch on the operator panel to move the C-arm until it is
0º ±0.1º. Remove the level and then
Presss <ENTER>.
13. Pres
The program prompts:
and begins rotating the C-arm clockwise (i.e., the tank assembly moves toward
the front of the machine). When the C-arm hits the LA
LATERAL
TERAL mechanical limit,
Presss <ENTER>.
14. Pres
The program prompts:
15. Place the digital level on top of the X-ray tank assembly
assembly and use the Table
Table IN/
OUT switch on the operator panel
pa nel to move the C-arm until it is at 83 °±0.1º (do
not make this measurement with the cosmetic covering
cove ring on the tank assembly).
and moves the C-arm and the X-table back to their initial 0 ° positions.
The program then switches to graphics mode and draws the Encoder Vs. Angle
calibration
reaches thegrid. It steps the
83° position andAR motor
then stepsclockwise in 1 ° counter
the AR motor increments until the
clockwise inmotor
1°
increments until the motor return to approximately 0 °. The program plots the
encoder data during the scan (in raw encoder units, 0-4095).
Note:Press <ESC> anytime during the scan to abort the calibration procedure.
When the calibration scan completes, the program computes the linear fits to the
positive and negative rotation. The linear fit parameters are displayed at the top
left and top right of the plot in the form Y = Intercept + Slope
S lope * X. The two slope
values should be within 0.1 of eacheac h other.
The program displays the positive and negative limits as horizontal dashed lines
Note: The PosLimitOffset and NegLimitOff
NegLimitOffset
set entries in the [ArMotor] section of
the SQDRIVER.INI file determine the motion limits relative to the mechanical
stops. If these entries are not present, or are zero, the motion limits are set to the
mechanical limits.
neg_limit_position=-1395
Update Driver INI-File [Y/N] ?
Alignment Pins
Holding Clamps
A C
Alignment Pin
Holding Clamp
Fixture
B, D
View from Front
Step A. Insert the left side of the Alignment Fixture into the left side of the
Detector Opening so that the vertical edges of the four Alignment
Pins are secure.
Step B. Compress the Holding Clamps.
Step C. Raise the right side of the Alignment Fixture into the Detector
Opening.
Step D. Release the Holding Clamps.
Note: If the Alignment Fixture is inserted with the Alignment Pins on the
right, the procedure works equally well.
1. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities|Service Utilities|X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
2.. Set Aperture to 10 (set Aperture to -1 for Ci and Wi models), Hi gain to 2, Lo
2
gain to 1 and X-ray Mode to 3.
3-48
3. Place the
the Aperture
Aperture Alignment
Alignment Pin through
through the
the hole in
in the Aperture
Aperture Plate.
Plate. It
It
should drop straight through and into the alignment hole in the Filter Drum
Assembly base plate and be perpendicular
perp endicular to the Aperture Plate. If it is not per-
form steps 4 through 7. Otherwise, proceed to step 8.
4.. Hit the <Esc> key to return to the Discovery Main screen.
4
5. From
From the Dis
Discove
covery
ry Main
Main Menu scre
screen,
en, selec
selectt Utilities|Service Utili-
ties|SQDRIVER.
6. At the CARM$$$$> prompt, type move_aper_rel +100 (to move the Aper-
6.
ture Plate right) or move_aper_rel –100 (to the move Aperture Plate left) until
the Alignment Pin is both vertical and perpendicular to the Aperture Plate. (Place
the block on the base plate next to the pin and compare the pin to the vertical sur-
face of the block to see if the pin is perpendicular).
7. Type exit <Enter> to return to the Discovery Main Screen.
7.
8. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities|Service Utilities|X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
9. Do no
nott cha
chang he Aperture setting, the Hi gain setting (set to 2), the Lo gain
ngee tthe
setting (set to 1) or the X-ray mode setting (set to 3).
For Display, click Graph/
10.. For
10
11.
11. Cl
Click X-ray (F2) to turn on x-rays.
ick
3-49
The correct display should show a flat graph with an amplitude of aapproximately
pproximately 5.5 volts
with the alignment fixture installed (approximately 6.5 volts with alignment fixture
removed. volts. If thebeam alignment is not correct, perform
pe rform the following procedure.
Figure 3-24
12. Set the machine in
in the Center Table
Table position.
3-50
Adjusting
Screws and
Jam Nuts
FRONT
1.. Move the front Filter Drum alignment screw until the X-ray signal reaches max-
1
imum voltage.
Note: The last direction
direction turned should
should be clockwise
clockwise to eliminate
eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
2.. Move the rear Filter Drum alignment screw until the X-ray signal reaches maxi-
2
mum voltage.
Note: The last direction
direction turned should
should be clockwise
clockwise to eliminate
eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
3-51
3. Finger tighten
tighten the jam
jam nuts on both Filter
Filter Drum alignm
alignment
ent screws,
screws, then
then gently
gently
snug them with a small wrench (use ofo f excessive force will change the bea
beamm
alignment). After the jam nuts are snug, re-check the graph for proper beam
alignment.
Note: The X-rays
X-rays should still
still show
show peak amplitude.
amplitude.
4. Turn
Turn off
off the X-rays
X-rays using
using the
the <F2> key.
5. Remove
Remove the
the alignme
alignment
nt test
test fixtu
fixture
re from
from the C-arm
C-arm..
3. Start
Start the QDR/AP
QDR/APEX
EX soft
softwar
waree in Servic
Servicee mode.
mode.
4. On the
the QDR
QDR/A
/APEX
PEX main
main scr
screen
een,, select Utilities|Service Utilities|X-Ray
select
Survey.
5. Set Pulse to 1, press <ALT><P>.
6. Set
a. Hi Gains to 2.
b. Lo Gains to 1.
c. X-ray Mo to 3.
Mode to
d. o 7.
Aperture tto
5. Press <Enter>.
6. Displa
Display
y the X-Ra
X-Ray
y Survey
Survey bar graph,
graph, press
press <Ctr
<Ctrl><
l><Pag
Pagee Down>.
Down>.
7. Obse
Observ
rvee Hi Air
Air, pre
press
ss <Alt
<Alt><
><S>
S>..
8. Turn
urn on
on X-ra
X-rays
ys,, p
pre
ress
ss <F1>
<F1>..
9. Verify that the average
average X-ray
X-ray signal
signal level is 6.25V
6.25V and that all
all detectors
detectors are
within the 4.5V to 8.5V range. If these conditions are met, the procedure is com-
plete. Do not adjust the A/D gain.
If these conditions are not met, continue to “Adjust the A/D Gain” on page 3-55
3-55..
This procedure flattens the X-ray beam for each scan mode and finds the metal edge of the
table.
Note: All covers,
covers, table
table mat, etc.,
etc., normally
normally in the
the X-ray path,
path, MUST be
be on the Scanner
Scanner
before running Beam Flattening.
1. Restart
Restart the Discover
Discovery
y software
software in service
service mode (if
(if not already). Press the Center
already). Press
Table button and turn on the laser.
2. If you are
are using an Anthropomo
Anthropomorphic
rphic Spine
Spine Phantom
Phantom (resem
(resembles
bles vertebra
vertebrall bod-
ies, Part Number 030-1967), place the phantom on end (vertical) with the spinal
3-55
processes pointing towards the head end of the table. The laser crosshair should
be 1.5 inch in from the left end and centered. (Some phantoms will have a target
hole, if not, use a ruler).
If you are using a DXA
DX A Quality Control Phantom (aluminum representation of
vertebral bodies that does
d oes not resemble real bone, Part Number ASY-01564),
ASY-01564),
place the phantom on end with the flattening target facing up. Orient the pha
phantom
ntom
so the metal edge is to the right (head end) and tissue to the left (foot end) as
viewed from the front of the table. Align the phantom flattening target to the laser
crosshair.
3. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities/Service Utilities/X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
4.. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to
4
7.
5. For Display, click Graph.
5.
6 . Click X-Rays (F2) to turn X-Rays on.
6.
7. Check that
that phantom
phantom covers
covers the whole
whole beam. This
This is critical.
critical. If
If phantom
phantom does not
cover the whole beam, move it until it does. Keep the phantom as straight as pos-
sible.
8. Turn
urn off
off X-r
X-ray
ayss wit
with the <F2> key and press <Esc> to exit X-Ray Survey.
h the Surv ey.
Reboot the PC, logon as a s Field Service, and restart the software in Service Mode
to return to Discovery Main Menu screen.
9.. Select Utilities|Service Utilities|Detector Flattening. Click the Continue
9
button. This procedure can takes about 10 minutes on a C or W model and 55
minutes on an A or SL.
10. When flattening is complete, run
run a QC scan and cancel adding it to the QC plot.
11. Use the Scan File Plot utility to check X-ray beam flatness of the QC scan
scan per-
formed in Step 10.
4. Align the
the phantom
phantom so that the
the laser crossh
crosshair
air is on the
the phantom
phantom laster
laster target.
target.
Using the laser as a guide, adjust the phantom to be parallel to the table edges
(just like when performing a QC scan).
5. On the APEX/
APEX/QDR
QDR main
main scree
screen,
n, select
select Perfo
Perform
rm Exam.
Exam.
6. Create
Create a ttest
est patien
patientt biogr
biography
aphy..
7. Select
Select OK
OK at pati
patient
ent Conf
Confirm
irmati
ation
on Screen
Screen..
8. On the Select
Select Scan Type
Type screen
screen select AP/Late
AP/Lateral
ral Pair.
Pair. Uncheck Use
Use Default
Default
Scan Mode and click Next.
9. Select
Select Arra
Array
y and
and clic
click
k Next
Next..
10. On the Select Lateral Scan Mode screen
screen select Array and click Next.
11. Verify the alignment of the
the spine phantom to the laser.
laser.
12. Change Scan Length to 6.0 inches and click Start
Start Scan to perform tthe
he AP scan.
13. When scan has finished,
finished, select Spine as the analysis method, and click Next.
14. Analyze the scan, and click Start
Start Position. At the bottom of screen verify that
that the
centerline angle is + 2 degrees. If OK, go to the next step. If not OK, restart the
procedure (go back to Step 1), being careful to align
a lign the phantom properly to the
laser.
15. If you are using an Anthropomorphic Spine Phantom (resembles
(resembles vertebral bod-
ies, Part Number 030-1967), skip this step.
If you are using a DXA
DX A Quality Control Phantom (aluminum representations of
vertebral bodies, Part Number ASY-01564),
ASY-01564), place 2 sheets of paper along the left
and right edges of the phantom as shown in Figure 3-27.
Figure 3-27.
3-57
Next, rotate the phantom “bone” segments towards the front of the scanner as
shown in Figure 3-28. Be careful to keep the phantom left and right edges paral-
lel to the sheets of paper.
Figure 3-28.
16. Click close, and then hold the Enable Lateral switch on the control panel to
to move
the system to the lateral position.
17. Click Start
Start Scan to perform
perform the lateral
lateral scan.
18. When the scan has finished, press
press and hold the Enable Lateral sswitch
witch to return
the system to AP position.
19. Analyze the lateral scan (refer
(refer to User Guide for instructions if necessary).
necessary). When
done, click Close.
20. At the Exit Analysis screen select Report. Print a report of the procedure just per-
formed, and include the report in the paperwork being returned to Hologic.
3.5.5.1 Principles
Radiation leakage from the HVPS/S results from a defect in the lead liner of the tank or its
cover. If there is a hole in the lead liner, the resulting leakage will produce a pencil
pe ncil beam
of X-rays. If there is a crack in the lead liner, the resulting leakage will produce a fan
beam. In either case, the initial survey must be taken close to and almost touching the tank
with the survey meter probe. If a leak is found, the meter indicates a spike in the reading.
If this occurs, you must then move the probe one-meter
one -meter (approx. 40”) along the beam from
that spot to determine if the leakage is within acceptable levels.
The Victoreen 450P has a response time of:
Range Time
0-4mR/h 3.3 seconds
0-40mR/h 4.5 seconds
1. Turn
Turn on X-r
X-rays
ays by pres
pressin
sing the <F1> key.
g the
2. Slowly
Slowly move the the meter probe (V
(Victoreen
ictoreen Model
Model 450P oror equivalent)
equivalent) about all
all
accessible surfaces (see table above for meter response times) to detect any leaks.
3. If a leak
leak is detected
detected (a
(a spike in
in the reading)
reading),, measure
measure a point
point at a one
one meter
meter
(approx. 40”) distance from the leak and insure that the reading is 10µGy/h
(1.0mrad/h) or less.
1. Position
Position the
the survey
survey meter one meter
meter (approx.
(approx. 40”)
40”) from the tank
tank and move
move it all
around the outside of the instrument.
2. The
The rrea
eadi
ding
ng must
must be 10µGy/h (1.0mrad/h) or less. Record the highest reading in
the service report.
3. Press the <F2> key to turn off X-rays, remove the tool, and
3. a nd replace the tank
cover.
Note: Enter this information into the Radiation Measurement Report (see example on
page 3-70).
3-70).
tests. Also, because the tabletop pad will have a slight affect on the test results, all
scans must be performed with it in place.
WARNING:
ARNING: X-r
X-rays
ays aare
re produ
produced
ced d
duri
uring
ng most
most of the
these
se tests
tests.. Keep h
hand
ands,
s, head
head an
and
d oth
other
er bo
body
dy parts
parts
out of the X-ray beam path. The tester should also be wearing an approved radiation
dosimetry badge.
Note: form
Be sure the spine
of "Spine phantom
Phantom has been
#nnn" entered
(where nnn isinthe
thenumber
patientof
database in the
the phantom)
so that the QC plot program can identify it properly.
properly.
3-61
8. Click
Click Star
Startt Scan
Scan to perf
perform
orm Latera
Laterall compo
component
nent..
9. Press
Press and
and hold
hold the
the Enable
Enable swit
switch
ch to retur
return
n to AP
AP positi
position.
on.
10. Analyze
Analyze scan (refer
(refer to user guide
guide for Lateral
Lateral analysis)
analysis) and close
close the screen.
screen.
11. Click exit
exit to close the
the Exit Analys
Analysis
is screen.
screen.
12. Restart
Restart QDR for Windows
Windows software
software in service
service mode (if not already).
already). From
From the
QDR Main window, select Utilities, Service Utilities, Field Service Calibration.
13. Select Highs/Lows
Highs/Lows for Array
Array Spine
Spine Phantom
Phantom
14. Select the AP scan
scan just
just performed
performed and
and click next
15. Click continue
continue on the Calculat
Calculated
ed highs and lows
lows page.
16. Select Highs/Lows
Highs/Lows for Array
Array Lateral
Lateral Phantom.
Phantom.
17. Click continue
continue on the Calculat
Calculated
ed highs and lows
lows page.
18. Click Temp
Temp Write
Write to write
write temporary results
results and click continue
continue to exit the write
dialog box.
1. From
From the
the Discove
Discovery
ry Main
Main Menu
Menu screen,
screen, select
select Utilities|Service Utili-
ties|Auto Analyze.
3-63
2. Select
Select Anal
Analyze
yzed
d sc
scans
ans,, click RECALYZE, and select the 20 scans
click sca ns just acquired
and click the Next button.
3. The scans
scans will
will now be recalc
recalcula
ulated
ted to the
the new cali
calibrat
bration
ion..
4. Highli
Highlight
ght the
the scans
scans from
from the
the Exclu
Excluded
ded list
list and clic
clickk the Include Scans button
and then the Finish button. (Selection is easier if you click Scan Date bar above
the list. This will order the scans by date and
a nd time).From the Discovery Main
Menu screen, select QC|QC Data Management|Select Scans .
5. Select the QC|QC Data Management|Plot option from the Discovery Main
Menu Screen to generate an AREA, BMC, and BMD plot for each array mode.
Verify that the QC plots include the factory scans and the scans just performed.
6. Ensure
Ensure that
that all
all scans
scans fall
fall between
between the
the two,
two, dotted
dotted limit
limit lines
lines..
7. Print the BMD, BMC, Area HiAir and LoAir
LoAir plots.
plots. Include
Include the printed
printed plots
plots
with the other paperwork being returned to Hologic.
Note: For more detail
detailss on performing
performing the
the QC setup and producing
producing QC plots,
plots, refer
refer to the
the
Discovery User's Guide.
functionalduring
problems in all modes of operation.
each scan mode. Verify
Verify that there are no elec
electrical
trical or mechanical
[ ] Perform an AP Spi
Spine
ne Array,
Array, Fast Array,
Array, and Express (Ci and Wi
Wi only) scan.
[ ] Perform a Left Hip Array,
Array, Fast Array,
Array, and Express (Ci and Wi
Wi only) scan.
[ ] Perform a Whole Body scan
scan (A, W,
W, and Wi
Wi only).
[ ] Perform a Forearm scan.
[ ] Perform the IVA
IVA Scan QC Procedure described below.
below.
• at the
the tim
timee of
of inst
instal
alla
lati
tion
on
• after the repair
repair of any
any parts
parts associ
associated
ated with the generati
generation
on or detection
detection of x-rays
3-64
Each service engineer has been provided with an 11-step aluminum phantom (Step Wedge
Penetrometer, P/N 099-0716) for checking image contrast
c ontrast and a line-pair phantom (X-ray
Test Pattern, P/N 099-0715) for verifying scan resolution.
Checking the quality of the IVA
IVA scans requires three scans:
• single-ener
erg
gy AP
• si
sing
ngle
le-e
-ene
nerrgy later
ateral
al
• dual-energy AP
AP
To perform an IVA scan QC, do the following:
1. Press the Center Table button to center the table.
1.
2.. Press the Laser button to turn the laser ON. If the laser goes off
2 off during the align-
ment process just press the Laser button again to turn the laser ON.
3. Place the x-ray
x-ray Step
Step Wedge
Wedge Phantom
Phantom (P/N
(P/N 099-0716)
099-0716) on the table
table approxima
approximately
tely
an inch to the right of the crosshair.
crosshair. Center the laser line
line approximately through
the center of the Step Wedge.
4. The Step Wedge
Wedge Phantom
Phantom should
should be placed with
with its
its thickest
thickest end to
to the left (foot
(foot
end) and its thinnest end on the right with its edges parallel to the table.
5. Skip steps 6 through 8 for Models Ci and Wi.
6. The X-ray
X-ray Test
Test Pattern
Pattern Phantom
Phantom (P/N
(P/N 099-0715)
099-0715) should
should be aligned
aligned approxi-
approxi-
mately an inch to the right of the Step Wedge
Wedge Phantom.
7. The x-ray test pattern
pattern should
should be placed
placed on the
the table with
with the lines
lines the Test
Test Pattern
Pattern
Phantom perpendicular to the motion of the C-arm during sscanning.
canning. The laser
line should pass through
through the center of the x-ray Test
Test Pattern Phantom. The 1.8
and 0.6 test patterns should be farthest from the Step Wedge
Wedge Phantom. The 0.6
test pattern should be nearest to the front of the table.
8. Depre the Laser button on the C-Arm to turn OFF the laser.
press th
9. Click the Perform Exam button at the Discovery Main Screen.
9.
10. Select the IVA Test as the patient. If this name is not present, then click the New
Select the
patient button. Enter IVA Test as the new patient last name. Leave the first
name and initials blank and then click the OK button.
11.
11. Click
Click the name IVA Test from the patient names. Click the OK button.
the name
2. In Patient Confirmation Screen, click the OK button.
12.
1
the Select Scan Type screen, select IVA Imaging.
13.. At the
13
Uncheck the Use default scan mode check box and then click the Next>> but-
14. Uncheck
ton.
Select SE AP Image then click the Next>> button.
15.. Select
15
16.. Select
16 Select None then click the Next>> button.
17. From the Scan Parameters screen set the scan length to 10.0 inches.
From the
3-65
18. Cl
Clic
ick the Start Scan button to run a single-energy AP scan.
k the
19. From the Hologic Analyze screen, perform the next 3 steps.
From the
Note: Intensity may be adjusted to visualize scan.
20. Count the 12 steps associated with the Step Wedge
Count the Wedge Phantom.
The number of steps visible on the Step Phantom should be 11 and 1 more for the
table.
21. Skip this step on models Ci and Wi. By clicking the + sign to increase the mag-
nification and using the scroll bar to examine the x-ray test pattern image on the
screen, you should be able to see three lines for the 0.8 test pattern image .
22. When completed click the Cancel button.
completed,, click
A radiation dose measurement must be performed at installation time, and should be done
routinely whenever the machine is serviced. This test requires the use of a Victoreen
Model 450P or equivalent measuring device set to measure
mea sure dose over time (mR/Hr).
Note: The dose measurements with the machine fully assembled and the
measurements must be done with
table mat in place.
place.
Note: Enter this information into the Radiation Measurement Report (see
example on page 3-70).
3-70).
page 3-70).
3-70).
3.5.16 Perform QC
Once the machine has been
bee n fully assembled and calibrated, at least one QC scan should be
performed.
• Perfor
Perform
m the dail
daily
y QC proced
procedure
ure and
and enter
enter the
the scan
scan into
into the QC
QC databa
database.
se.
insure that the coefficient of variation for absolute radiation is less than 4%. The plot
screen does not readily display the coefficient of variation for absolute radiation.
ra diation.
You must use the formula:
coefficient of variation for absolute radiation = standard deviation (SD) x.0028
x.0 028
Therefore:
If… Then…
the Standard Deviation (SD) for both High Air the coefficient of variation for absolute radiation is
(HiA) and Low Air (LoA) is less than or equal to less than 4%.
14,
To do this:
1. If necessary
necessary,, restart
restart the
the Discovery
Discovery softwa
software
re in Service
Service Mode.
Mode.
2. Perform
Perform 10 arr
array
ay spin
spinee scans
scans (using Utilities|Service Utilities|Autoscan
(using
from the Discovery Main Menu screen).
3. Analyze
Analyze the fir
first
st scan
scan manual
manually
ly..
4. Include
Include the analyzed
analyzed Scans
Scans into
into the QC database.
database. On the
the APEX main
main screen,
screen,
select QC>QC data Management>Select Scans. In the Excluded Scans
pane, select (highlight) the ten scans just analyzed and click Include Scans. Click
OK when done.
5. Put the results
results in the QC database
database (if
(if you have not already
already done so).
so).
6. From the Discove
Discovery
ry Main
Main Menu,
Menu, sele ct QC|QC Data Management|Plot. Use
select
a start and end date that will only select the scans that you have just completed.
7. Select
Select QC
QC Paramet
Parameter
er to Plot,
Plot, HiA,
HiA, and
and Plot.
Plot.
8. Plot
Plot the
the data
data for the 10 scans
scans in step
step 1.
9. Assu
Assure
re that
that the
the SD is
is 3 or less
less..
Make sure that only the pad is on the table. Clear the table of anything else.
3-68
3. Perf
Perfor
ormm one
one Whol Body air scan (a scan with nothing on
Wholee Body o n the table). Do NOT
interrupt the scan for any reason. When the scan completes, select Exit Exam.
Do not analyze the scan that was just performed.
4. From the Discove
Discovery
ry Main
Main Menu
Menu screen, select Utilities|Service Utili-
screen, select
ties|Table
ties|Table Top Radiographic Uniformity.
5. Select
Select the
the Whol
Wholee Body
Body scan
scan and clic
click
k the
the Next> button.
6. Table Top
Top Radiographi
Radiographicc Uniformit
Uniformityy results
results tabs are now displayed.
displayed. Select
Select the
the
Low Air Tab.
11. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly
aligned.
If the Global Stats S.D. is greater than or equal to 2.0 , then the machine is not
aligned properly. Check the following:
• If the
the image
image appear
appearss "stre
"streaky"
aky" or
or "banded
"banded",", check
check for
for loose
loose wires
wires or other
other
debris between the table and the tank.
• Check
Check thethe Aper
Apertu
ture
re Cal
Calibibra
rati
tion
on (se
(seee page
page 3-53).
3-53).
• Check
heck AC Li Line vo
voltage.
ge.
• Che
heck
ck +/-
+/- 15V
15V Powe
Powerr Sup
Suppl
ply
y
• Chec
Check k Dete
Detectctor
or Fla
Flatt
tten
enin
ingg on pag
pagee 3-55.
3-55.
• Re- daiily QC (see the Discovery User's Guide).
e-rrun da Guide).
Print a copy of the test results and inlcude it with the other paperwork being
be ing
returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the
global S.D. is less than 2.0, you are done. If the global S.D. if equal or greater
than 2.0, print a copy of the test results and include it with the other pa
paperwork
perwork
being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed
to the table itself. If so, continue with this procedure.
proce dure. Make sure that the machine
has body composition loaded. If it does,
do es, then follow the appropriate procedure
below.
Section 4
ALIGNMENT AND CALIBRATION
+/- .06”
63.125”
FRONT & REAR
+/- .25”
“A” “A”
Studs and
Adjustment Nuts
T-Rail
Upper Bracket
the bolts.
4. Install
Install the
the pedestal
pedestal covers that were removed
removed in Step
Step 1.
4. Rotate
Rotate the C-Arm
C-Arm until
until the
the center of
of gravity
gravity for
for the upper
upper C-Arm
C-Arm is over the
the
lower C-Arm (approximately 60 degrees rotated).
4-3
5. Loosen
Loosen eight
eight 1/4"
1/4" bolts
bolts (4 on
on each
each side)
side)..
6. Move the C-arm
C-arm until
until it is parall
parallel
el to the tank
tank using the
the digital
digital level to measure
measure
the angles.
Insert the alignment fixture (see the following three figures) into the detector opening.
Alignment Pins
Holding Clamps
A C
Alignment Pin
Holding Clamp
Fixture
B, D
View from Front
Step A. Insert the left side of the Alignment Fixture into the left side of the
Detector Opening so that the vertical edges of the four Alignment
Pins are secure.
Step B. Compress the Holding Clamps.
Step C. Raise the right side of the Alignment Fixture into the Detector
Opening.
Step D. Release the Holding Clamps.
Note: If the Alignment Fixture is inserted with the Alignment Pins on the
right, the procedure works equally well.
4-5
4.. Hit the <Esc> key to return to the Discovery Main screen.
4
5. From
From the Dis
Discove
covery
ry Main
Main Menu scre
screen, selectt Utilities|Service Utili-
en, selec
ties|SQDRIVER.
6. At the CARM$$$$> prompt, type move_aper_rel +100 (to move the Aper-
6.
ture Plate right) or move_aper_rel –100 (to the move Aperture Plate left) until
the Alignment Pin is both vertical and perpendicular to the Aperture Plate. (Place
the block on the base plate next to the pin and compare the pin to the vertical sur-
face of the block to see if the pin is perpendicular).
7. Type exit <Enter> to return to the Discovery Main Screen.
7.
8. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities|Service Utilities|X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
9. Do no
nott cha
chang hiss Aperture setting (set to 10), the Hi gain setting (set to 2), the
ngee tthi
Lo gain setting (set to 1) or the X-ray mode setting (set to 3).
For Display, click Graph/
10.. For
10
11.
11. Cl
Click X-ray (F2) to turn on x-rays.
ick
The correct
with display
alignment should
fixture show(approximately
installed a flat graph with
6.5anvolts
amplitude of aapproximately
pproximately
with alignment 5.5 volts
fixture removed).
If the beam alignment is not correct, perform the following beam alignment procedure.
4-6
To perform beam
b eam alignment:
1. Set the
the machin
machinee in the
the Center
Center Tabl
Tablee positi
position.
on.
Adjusting
Screws and
Jam Nuts
8.. Move the front Filter Drum alignment screw until the X-ray signal reaches max-
8
imum voltage.
Note: The last direction
direction turned should
should be clockwise
clockwise to eliminate
eliminate backlash when
the jam nuts are tightened. The plot should show peak amplitude.
9.. Move the rear Filter Drum alignment screw until the X-ray signal reaches maxi-
9
mum voltage.
4-7
X -A
-A l i n m e
enn t N ut
ut 4 x
A l i n m en
e n t S c re
re w
14. Turn
Turn on the X-rays
X-rays with
with the <F1> key.
For a Discovery A only, adjust the array X-alignment screw in one direction until
15. For
the trace drops off (the signal on the end detectors of the array will drop off).
Then, count the turns while moving it in the other direction until the trace falls
off on the detector on the other end of the array. Set the adjustment in the middle
by turning the screw back half the number of turns counted.
For Discovery C, Ci, W,
16.. For
16 W, Wi, and SL models, detectors 44 through 171 will be
turned on. You
You will have to exit this routine, remove the alignment pin, run the
Aperture Calibration procedure (Aperture Calibration cannot be run on Ci and
Wi), and then return to X-Ray Survey. When you enter X-Ray Survey do not
change this Aperture setting (set to 7), the Hi gain setting (set to 2), the Lo gain
setting (set to 1) or the X-ray mode setting (set to 3).
For Display, click Graph.
17.. For
17
18.. Cl
18 Click X-ray (F2) to turn x-rays on.
ick
4-8
19. Follow the procedure in in step 3 while checking detectors number 44 and 171 to
check for the fall off of x-rays.
20. Tighten
Tighten the four array
array X-alignment
X-alignment nuts.
nuts.
21.. Cl
21 Click X-ray (F2) to turn x-rays off.
ick
Use the following table to determine which calibration procedures you need
nee d to perform on
a given Discovery model.
Perform the calibration procedures if
indicated (*) in order from left to
right.
Model TZ AY TY TX AR
A * * * * *
SL * * N/A * *
W N/A * * * N/A
Wi N/A * * * N/A
C N/A * N/A * N/A
Ci N/A * N/A * N.A
Each motor (except TZ) requires the corresponding protocol calibration file in the
PROTOCOL sub-directory (e.g., for MOTOR$AY
MOTOR$AY, the calibration protocol is
MOTOR_AY.PRO).
To perform the calibration
ca libration procedure:
1. From
From the Dis
Discove
covery
ry Main
Main Menu
Menu screen,
screen, select Utilities|Service Utili-
select
ties|SQDRIVER (you must be in Service Mode).
2. At the CARM$$$$> prompt, type MOTOR$XX<Enter>, where XX equals TZ,
2.
AY, TY, TX, or AR depending on which motor you are calibrating.
3. Measure
Measure the distance
distance moved
moved using
using the bottom
bottom edge
edge of the
the top pedestal
pedestal cover
cover and
the floor. The system displays:
Total Distance Moved By Pedestal [20.0 cm]?
4-10
4.6.0.2 MOTOR$AY
If you have to replace the encoder, before beginning this procedure, loosen the set screws
that secure the encoder coupling
co upling to the pulley shaft.
To perform the calibration
ca libration procedure:
1. Select
Select the
the AY
AY motor
motor device
device driv
driver
er by typing: MOTOR$AY<Enter>
typing:
2. At the MOTOR$AY> prompt in SQDRIVER, type: CALIBRATE<Enter>
2.
The program prompts:
Press <Enter> when the AY motor reaches the LEFT mechanical limit.
Press <ESC> to stop calibration.
3. The program
program moves the AY AY motor to
to the left.
left. When AY
AY hits the
the left mechanic
mechanical
al
stop the first time press <Esc>.
4. Check the
the position
position value.
value. It must
must be 3750±
3750±5.
5. Adjust
Adjust if necessa
necessary
ry..
4-12
6. When
When AY
AY hits
hits the right
right mechan
mechanica
icall stop, press <Enter>.
stop, press
The program then switches to graphics mode and draws the Encoder Vs
Distance calibration grid. It steps the AY
AY motor to the left in one inch
increments until the motor hits the left mechanical stop, and then steps
the AY
AY motor to the right in one inch
in ch increments until the motor hits the
right mechanical stop. The program plots the encoder data during the
scan (in raw encoder units,
un its, 0-4095).
Note: Press <Esc> anytime during the scan to abort the calibration procedure.
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fit parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.6 and 86.5)
86 .5) should be within 0.3 of ea
each
ch other.
4-13
The program displays the positive and negative limits as horizontal dashed lines.
Note: The PosLimitOffset and NegLimitOf
NegLimitOffset
fset entries in the [AyMotor] section
of the SQDRIVER.INI file determine the motion limits relative to the
mechanical stops. If these entries are not present, or are zero, the motion
limits are set to the mechanical limits).
change s the plot title to PRESS ANY KEY TO CONTINUE.
The program then changes
7. Press the <Enter> key and the program prompts.
motor_direction=1
calibrate_position=1,2288,41187,209,61339,363,363,3750
pos_limit_position=984946
neg_limit_position=0
Update Driver INI-File [Y/N] ?
4-14
8. If the position
position limits
limits are within
within specifi
specification
cations, Type Y<Enter> to accept the cal-
s, Type
ibration values.
7,8) 363,3719 (NegLimit,PosLimit). The encoder readings for the negative (right) and
positive (left) mechanical stops.
The pos_limit_position and the neg_limit_position are the limits of motion, in microns,
in the positive and negative direction.
Note: The PosLimitOffset and NegLimitOf
NegLimitOffset
fset entries in the [AyMotor] section of the
SQDRIVER.INI file determine the motion limits relative to the mechanical stops.
If these entries are not present, or are zero, the motion limits are set to the
mechanical limits, as in the above example).
Note: The last calibration scan data is
is saved in the file
file MOTOR_AY
MOTOR_AY.DAT
.DAT. You
You can
reanalyze the data–e.g., after editing SQDRIVER.INI by typing the command
CALIBRATE @MOTOR_AY.DAT at the MOTOR$AY> prompt in SQDRIVER.
4-15
3. The program
program moves
moves the TY
TY motor
motor to the
the left.
left. When TY
TY hits the left
left mechanical
mechanical
stop the first time press <Esc>.
4. Check the
the position
position value.
value. It must
must be 3750±
3750±5.
5. Adjust
Adjust if necessa
necessary
ry..
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 3750±5 (i.e., in the range 3745-3755) and then tighten the
coupling setscrew.
setscrew.
5. Repeat the
the calibration
calibration procedure
procedure above
above but now,
now, when TY hits
hits the left
left mechani-
mechani-
cal stop, press <Enter>.
The program then starts TY moving to the right and prompts:
Press <Enter> when the TY motor reaches the RIGHT mechanical limit.
Press <ESC> to stop calibration.
When TY hits the right mechanical stop, press <Enter>. The program then
switches to graphics mode and draws the Encoder Vs Distance calibration grid. It
steps the TY motor to the left in one inch increments until the motor hits the left
mechanical stop, and then steps the TY motor to the right in one inch increments
until the motor hits the right mechanical stop. The program plots the encoder
data during the scan (in raw encoder units, 0-4095).
Note: Press <Esc> anytime during the scan to abort the calibration procedure.
4-16
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fit parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.5 and 85.6)
85 .6) should be within 0.3 of ea
each
ch other.
When the calibration scan completes, the program computes the linear fit for
both the positive and negative motion. The linear fir parameters are displayed at
the top left and top right of the plot in the form Y = Intercept + Slope * X. The
two slopes (e.g., 86.5 and 85.6)
85 .6) should be within 0.3 of ea
each
ch other.
The program displays the positive and negative limits as horizontal dashed lines.
Note: The PosLimitOffset and NegLimitOf
NegLimitOffset
fset entries in the [TyMotor] section
of the SQDRIVER.INI file determine the motion limits relative to the
mechanical stops. If these entries are not present, or are
a re zero, the motion
limits are set to the mechanical limits.
6. The prog
program
ram then
then changes
changes the plot
plot title to PRESS ANY KEY TO CONTINUE .
title to
Press the <Enter> key and the program prompts.
motor_direction=0
calibrate_position=1,2287,27446,154,45219,238,238,3742
pos_limit_position=1028879
neg_limit_position=0
Update Driver INI-File [Y/N] ?
4-17
Note: The last calibration scan data is saved in the file MOTOR_TY.DA
MOTOR_TY.DAT T. You
You
can reanalyze the data–e.g., after editing SQDRIVER.INI –by typing the
command CALIBRATE @MOTOR_T @MOTOR_TY Y.DAT at the MOTOR$TY>
prompt in SQDRIVER.
4.6.0.4 MOTOR$TX
If you have replaced the encoder, before beginning this procedure, loosen the set screws
that secure the encoder coupling to the pulley shaft and make sure that the tank assembly
is not under the operator console. For normal installations, this is unnecessary.
1. Select
Select the
the TX motor
motor devic
devicee driver
driver by typing:: MOTOR$TX<Enter>
by typing
2. At the
the MOTOR$
MOTOR$TX>
TX> promp
promptt in SQDRI
SQDRIVER type: CALIBRATE<Enter>
VER,, type:
The program prompts:.
Press <ESC> to stop calibration.
4. The program
program moves
moves the TX
TX motor
motor to the
the left.
left. When TX
TX hits the left
left mechanical
mechanical
stop the first time press <Esc>.
5. Check the
the position
position value.
value. It must
must be 1000±
1000±5.
5. Adjust
Adjust if necessa
necessary
ry..
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 1000±5 (i.e., in the range 995-1005) and then tighten the
coupling setscrew.
setscrew.
6. Repeat the
the calibration
calibration procedure
procedure above
above but now,
now, when AY
AY hits the
the left mechani-
mechani-
cal stop, press <Enter>.
The program switches to graphics mode and draws the Encoder Vs Distance cal-
ibration grid. It steps the TX motor in by 1" increments
incre ments until the inner mechani-
cal stop is
encoder hit. The
units, program plots the encoder
0-4095). encode r data during the scan (in raw
7. Press <Esc> at anytime during the scan to terminate the calibration procedure.
4-18
The program then changes the plot title to PRESS ANY KEY TO CONTINUE.
8. Press the <Enter> key and the program prompts:
motor_direction=1
calibrate_position=1,2287,27446,43,12563,1026,1026,3096
pos_limit_position=604777
neg_limit_position=0
Update Driver INI-File [Y/N] ?
4-19
Note: The last calibration scan data is saved in the file MOT
MOTOR_TX.DA
OR_TX.DAT T. You
You
can reanalyze the data–e.g., after editing SQDRIVER.INI –by typing the
command CALIBRATE @MOTOR_TX.DAT at the MOTOR$TX>
prompt in SQDRIVER.
The program moves the TY and AY AY motors to their center positions, then moves
the TZ motor to its topmost position and prompts:
Use the Table
Table IN / OUT switch to move the C-Arm to 0 degrees.
Press <Enter> when the C-Arm is positioned.
Press <ESC> to stop calibration.
4. Remove
Remove the C-arm
C-arm Tank cover
cover.
5. Place the
the digital
digital level on the top of the tank assembly (not on the top of the C-
tank assembly
arm)) and use the Table IN/OUT switch on the operator panel to move the C-arm
arm
until it is 0º ±0.0º
6.. Remo
6 Remove
ve the
the leve
levell.
7. Press <Enter>.
7.
The program prompts:
Press <ESC> to stop calibration.
4-20
The program moves the TZ table to its top most position and moves the TX table
inwards until it almost touches the C-arm. It rotates the C-arm by 2 degrees to
obtain an initial estimate of the encoder calibration and then prompts:
Press <Enter> when the AR motor reaches the AP mechanical limit.
Press <ESC> to stop calibration.
8. The program
program rotates
rotates the
the C-arm counter
counter clockwise
clockwise (i.e.,
(i.e., the tank
tank assembly
assembly moves
moves
away from the front of the machine). When the C-arm hits the AP mechanical
limit the first time, press <ESC>.Check the position value. It must be 250±5.
Adjust if necessary.
If out of range, loosen the coupling
c oupling setscrew and manually rotate the encoder
until the readback is 250±5 (i.e., in the range 245-255) and then tighten the cou-
pling setscrew.
9. Before
Before repeating
repeating the
the calibration
calibration procedur
procedure,
e, rotate
rotate the C-arm
C-arm back
back to approxi-
approxi-
mately 0º by:
Typing the command: MOVE_REL 1470<Enter>.
10. Repeat the calibration procedure above but now,
now, when A
AY
Y hits the left mechani-
cal stop, press <Enter>.
Press <ESC> to stop calibration. Press <Enter> when the AR motor reaches the AP
mechanical limit.
and moves
mately 83°.the C-arm
It then and the
changes theX-table
prompttogether
to: until the C-arm is at approxi-
4-23
and moves the C-arm and the X-table back to their initial 0 ° positions.
The program then switches to graphics mode and draws the Encoder Vs. Angle
calibration grid. It steps the AR motor clockwise in 1 ° increments until the motor
reaches the 83° position and then steps the AR motor counter clockwise in 1 °
increments until the motor return to approximately 0 °. The program plots the
encoder data during the scan (in raw encoder units, 0-4095).
Note: Press <ESC> anytime during the scan to abort the calibration procedure.
ress
When the calibration scan completes, the program computes the linear fits to the
positive and negative rotation. The linear fit parameters are displayed at the top
left and top right of the plot in the form Y = Intercept + Slope
S lope * X. The two slope
values should be within 0.1 of eacheac h other.
The program displays the positive and negative limits as horizontal dashed lines
Note: The PosLimitOffset and NegLimitOff
NegLimitOffset
set entries in the [ArMotor] section
of the SQDRIVER.INI file determine the motion limits relative to the
mechanical stops. If these entries are not present, or are zero, the motion
limits are set to the mechanical limits.
neg_limit_position=-1395
Update Driver INI-File [Y/N] ?
For information about calibration parameters, see “Calibration Parameters” on
page 4-25.
4-25.
ype Y<Enter> to accept the calibration values or N<Enter> to reject them.
18.. Type
18
19. Exit SQDRIVER
SQDRIVER by typing exit
exit and clicking Retu
Return.
rn.
4-26
• On system
systemss without
without a separate
separate A/D Converter
Converter board,
board, the adjust
adjustment
ment is the
the only
only poten-
poten-
tiometer on the X-Ray Detector assembly.
Caution:Any adjustment of this potentiometer affects the QC highs and lows. You must
run detector flattening after adjusting the A/D gain. DO NOT ADJUST THE A/
D GAIN UNLESS ABSOLUTELY
ABSOLUTELY NECESSARY AS DEFINED BY “Check
and Verify
Verify the A/D Gain” on page 4-27.
4-27.
4.8.1 Check and Verify the A/D Gain
1. Instal
Installl all syst
system
em covers
covers that
that are
are normal
normally
ly in the
the X-ray
X-ray beam
beam..
2. In
Inst
stal
alll the
the tab
table
le pad
pad ont
onto
o the
the tabl
table.
e.
3. Start
Start the QDR/AP
QDR/APEX
EX soft
softwar
waree in Servic
Servicee mode.
mode.
4. On the
the QDR
QDR/A
/APEX
PEX main
main scr
screen
een,, select Utilities|Service Utilities|X-Ray
select
Survey.
5. Set Pulse to 1, press <ALT><P>.
6. Set
a. Hi Gains to 2.
b. Lo Gains to 1.
c. X-ray Mo to 3.
Mode to
d. o 7.
Aperture tto
5. Press <Enter>.
6. Displa
Display
y the X-Ra
X-Ray
y Survey
Survey bar
bar graph,
graph, pres
presss <Ctrl>
<Ctrl><Pag
<Pagee Down>.
Down>.
7. Obse
Observ
rvee Hi Air
Air, pre
press
ss <Alt
<Alt><
><S>
S>..
8. Turn
urn on
on X-ra
X-rays
ys,, p
pre
ress
ss <F1>
<F1>..
9. Verify that the average
average X-ray
X-ray signal
signal level is 6.25V
6.25V and that all
all detectors
detectors are
within the 4.5V to 8.5V range. If these conditions are met, the procedure is com-
plete. Do not adjust the A/D gain.
If these conditions are not met, continue to “Adjust the A/D Gain” on page 4-27
4-27..
Segment
Signal
Jigger Brass
Signal
500μs
Brass
Jigger Signal
Segment
Signal
500μs
This procedure flattens the X-ray beam for each scan mode and finds the metal edge of the
table.
Note: All covers,
covers, table
table mat, etc.,
etc., normally
normally in the
the X-ray path,
path, MUST be
be on the Scanner
Scanner
before running Beam Flattening.
1. Restart
Restart the Discov
Discovery
ery software
software in service
service mode (if
(if not already).
already). Press the Center
Press the
Table button and turn on the laser.
2. If you are
are using an Anthropomo
Anthropomorphic
rphic Spine
Spine Phantom
Phantom (resem
(resembles
bles vertebra
vertebrall bod-
ies, Part Number 030-1967), place the phantom on end (vertical) with the spinal
processes pointing towards the head end of the table. The laser crosshair should
be 1.5ifinch
hole, not,in from
use the left end and centered. (Some phantoms will have a target
a ruler).
If you are using a DXA Quality Control Phantom (aluminum representation of
vertebral bodies, Part Number ASY-01564),
ASY-01564), place the phantom on its end
e nd with
the flattening target facing up. Orient the phantom so the metal edge is to the
right (head end) and tissue to the left (foot end) as viewed from the front of the
table. Align the phantom
ph antom flattening target to the la
laser
ser crosshair.
crosshair.
3. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities/Service Utilities/X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
4.. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to
4
7.
4-29
7. Check that
that phantom
phantom covers
covers the whole
whole beam. This
This is critical.
critical. If
If phantom
phantom does not
cover the whole beam, move it until it does. Keep the phantom as straight as pos-
sible.
8. Turn
urn off
off X-r
X-ray
ayss wit
with the <F2> key and press <Esc> to exit X-Ray Survey.
h the Surv ey.
Reboot the PC, logon asa s Field Service, and restart the software in Service Mode
to return to Discovery Main Menu screen.
9.. Select Utilities|Service Utilities|Detector Flattening. Click the Continue
9
button. This procedure can takes about 10 minutes on a C or W model and 50
minutes on an A or SL.
10. When flattening is complete, run
run a QC scan and cancel adding it to the QC plot.
11. Use the Scan File Plot utility to check X-ray beam flatness of the QC scan
scan per-
formed in Step 10.
4-30
Figure 4-10.
Next, rotate the phantom “bone” segments towards the front of the scanner as
shown in Figure 4-11. Be careful to keep the phantom left and right edges parallel
to the sheets of paper.
4-31
Figure 4-11.
16. Click close, and then hold the Enable Lateral switch on the control panel to
to move
the system to the lateral position.
17. Click Start
Start Scan to perform
perform the lateral
lateral scan.
18. When the scan has finished, press and hold the Enable Lateral sswitch
witch to return
the system to AP position.
20. At the Exit Analysis screen select Report. Print a report of the procedure just per-
formed, and include the report in the paperwork being returned to Hologic.
4.12 Table
Table Top Radiograph
Rad iographic
ic Uni
Uniformity
formity
This procedure verifies proper X-ray beam alignment (for Discovery A and W):
1. Restart
Restart the Discover
Discovery
y software
software in
in service
service mode (if
(if not already).
already).
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit
un it in
the Patient ID field.
Make sure that only the pad is on the table. Clear the table o
off anything else.
4-32
3. Perfo
Perform
rm one
one Who
Whole Body air scan (a scan with nothing on
le Body o n the table). Do NOT
interrupt the scan for any reason. Select Exit Exam when complete. Do not ana-
lyze the scan you just performed.
4. From the Disc
Discover
overy
y Main
Main Menu
Menu screen,
screen, select Utilities|Service Utili-
select
ties|Table
ties|Table Top Radiographic Uniformity.
5. Select
Select the Whole
Whole Body
Body scan
scan and
and click
click the Next> button.
6. Table Top
Top Radiographic
Radiographic Uniformit
Uniformity
y results
results tabs are now display
displayed.
ed. Select the
the
Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly
aligned. Print a copy of the test results and include it with the other paperwork
being returned to Hologic.
If the Global Stats S.D. is greater than or equal to 2.0 , then the machine is not aligned
properly.. Check the following:
properly
• If the
the image
image appears
appears "streaky"
"streaky" or "banded",
"banded", check for
for loose
loose wires or other
other debris
debris
between the table and the tank.
• Check the “Aperture Calibration ” on page 4-9 4-9..
• Chec
Check k AC Line
Line vo
volt
ltag
age.
e.
• Chec
Check k +/
+/- 15V
15V Power
ower Supp
Supplly
• Check “Detector Flattening ” on page 4-29.
4-29.
• Re-r
Re-run
un da (see the Discovery User's Guide).
daiily QC (s Guide).
After realignment, repeat the Table Top
Top Radiographic Uniformity test. If the global S.D. is
less than 2.0, print a copy of the test results and include it with the other paperwork being
returned to Hologic.
If the global S.D. if equal or greater than 2.0, the non-uniformity may be attributed to the
table itself. If so, continue with this procedure. Make sure that the machine has body
composition loaded. If it does, then follow the appropriate procedure in 4.12.1 or 4.12.2.
4.12.2.
4-33
3. Compare
Compare values displayed
displayed with
with actual values print
printed
ed on the phantom.
phantom. If there
there is a
discrepancy, check to make sure you have the phantom that shipped with the unit.
Check with Hologic Technical Support if needed.
4-34
4. "Spine
Be surePhantom
the
the spine
spine#nnnn
#phantom has been
nnnn"" (where nnnnentered
entere
is thed number
in the patient
pat
ofient
thedatabas
databasee inso
phantom) thethat
form
form
theof
QC plot program can
c an identify it properly.
5. From
From the
the Discov
Discovery
ery Main
Main Menu
Menu scree
screen, selectt Perform Exam, select the Spine
n, selec
Phantom, and click the OK button at Patient Information dialog box.
6.. Select AP Lumbar Spine, uncheck default scan check box, and click the
6
Next> button.
7.. Select Array, click the Next> button, set the scan length to 6 inches, and start
7
the scan.
8. Anal
Analyz
yzee the
the scan
scan and
and clic
click the Close button to exit the Analysis screen.
k the
9. Start
Start the Field Service
Service Calibratio
Calibration
n program selecting Utilities|Service Utili-
program by selecting
ties|Field Service Calibration from the Discovery Main Menu screen.
10. Select the Spine Highs/Lows
Highs/Lows for Array Spine phantom option
option to automatically
calculate the new thickness indicators for array spine.
11. Select the Spine Phantom scanned in step 6 from the list of Analyzed Scans and
click the Next button. Click the Continue button on the Successfully Calcu-
lated message box.
12. Click on the Temp
Temp Write
Write button and click the Continue
Continue button on the Success-
fully Wrote message box. Exit Field Service Calibration.
4-35
Note: over.
Do not move the arm or table, otherwise the test will have to be start
started
ed
4. Change the
the scan length
length to
to 6 inches,
inches, and click
click start
start scan
scan to beg
begin
in the AP
AP compo-
compo-
nent.
5. When scan
scan has finished
finished,, select
select Spine
Spine as the analysis
analysis method, click Next.
method, and click
6. Analyze
Analyze the scan, and
and click start
start positi
position.
on. Check at bottom
bottom of
of screen that
that the cen-
cen-
terline angle is + or - 2 Degrees.
Degree s.
7. Click close, and then
then hold the enable
enable button
button to
to lateral
lateral positi
position.
on.
8. Click
Click Start
Start Scan
Scan to
to perform
perform Late
Lateral
ral comp
componen
onent.
t.
9. Press and hold
hold the Enable switch
switch to return
return to AP position.
position.
10. Analyze scan (refer to user guide for Lateral analysis)
analysis) and close the screen.
11. Click exit to
to close the Exit Analysis
Analysis screen.
screen.
12. Restart QDR for Windows software
software in service mode (if(if not already). From the
the
QDR Main window, select Utilities, Service Utilities, Field Service Calibration.
13. Select Highs/Lo
Highs/Lows
ws for Array Spine
Spine Phantom
14. Select the
the AP scan just performed and click Next.
performed
15. Click continue
continue on the Calculated
Calculated highs
highs and lows page.
page.
16. Select Highs/Lo
Highs/Lows
ws for Array Lateral
Lateral Phantom.
Phantom.
17. Click continue
continue on the Calculated
Calculated highs
highs and lows page.
page.
4-36
7. ties|Auto
From
From the Dis
DAnalyze
iscove
covery
ry Main
Ma
. in Menu
Menu screen,
screen, sele
select
ct Utilities|Service Utili-
8. Click on unanalyzed
unanalyzed Scans and select
select the 19
19 scans, click on Compare radio but-
scans, click
ton and then click the next button.
9. Now,
Now, sselect
elect the scan
scan analyzed
analyzed in
in step
step 6 and
and click the next button.
click the
10. The 19 scans will
will now be compared
compared with the analyzed
analyzed scan.
11. From the Discover
Discovery
y Main Menu screen, select Utilities|Service Utili-
screen, select
ties|Field Service Calibration.
Click the ACF/BCF button. Select the 20 scans, click next, and then click Con-
12. Click
tinue at the Successfully Calculated dialog box.
4-37
After the machine has been calibrated for thickness measurement in the array mode, it
must be calibrated for the array AREA and BMC measurements.
1. From
From the Dis
Discove
covery
ry Main
Main Menu
Menu screen,
screen, select Utilities|Service Utili-
select
ties|Auto Analyze.
2. Select
Select Anal
Analyze
yzed
d scans
scans,, click RECALYZE, and select the 20 scans
click sca ns just acquired
and click the Next button.
3. . The scans
scans will
will now
now be recalcul
recalculated
ated to
to the new
new calibrati
calibration.
on.
4. Highlight
Highlight the
the scans
scans from
from the Excluded list and click the Include Scans button
Excluded list
and then the Finish button. (Selection is easier if you click Scan Date bar above
the list. This will order the scans by date and
a nd time).From the Discovery Main
Menu screen, select QC|QC Data Management|Select Scans .
5. From
From the
the Discove
Discovery ry Main
Main Menu, selectt the QC|QC Data Management|Plot
Menu, selec
to create plots for BMD, BMC, AREA, HiAir, and LoAir plot for each array
mode. Verify that the QC plots inlcude QC scans from at least the last two years.
6. Ensure that all
all scans
scans fall
fall between
between the two,
two, dotted
dotted limit
limit lines.
lines.
7. Print the BMD,
BMD, BMC,
BMC, Area, HiAir,
HiAir, and LoAir
LoAir plots.
plots. Include
Include them with the other
other
paperwork being returned to Hologic.
Note: For more detail
detailss on performing
performing the
the QC setup and producing
producing QC plots,
plots, refer
refer to the
the
Discovery User's Guide.
Sample HARDWARE.INI
HARDWARE.INI (Calibration) File
HARDWARE.INI Variables Set by… Descriptions
[Version] .INI File Section Name
HardwareModel=14 Factory Discovery model (C=20, W=21, A=23, SL=22).
[Laser]
LaserYOffset=-38100 Factory Laser Offset Value
[TxMotor]
PosLimitPosition=409575 Factory Positive limit in microns (= 16 1/8 inches)
SpineInitPos=208000 Factory TX center position in microns (= 8.18 inches)
MaxSpinePos=233400 Factory + Limit for spine placement
MinSpinePos=182600 Factory - Limit for spine placement
PatientUnloadPos=0 Factory Table position for unloading patients
APLatDiff=376200 Factory AP/Lateral Di
Difference
MaxTrackOffset=25400 Factory
4-38
[TyMotor]
SpineInitPos=503175 Factory TX center position in microns (= 8.18 inches)
FlatPos=503175 Factory TX position for Detector Flattening
LocalMotionPos=503175 Factory Y Center table position using Control Panel
ScanPosTolerance=25400 Factory +/- Tolerance for Sc
Scan Y P
Po
osition (1
(1 in
in)
[TzMotor]
SpineInitPos=500 Factory TZ Start Position (AP)
LatInitPos=98425 Factory TZ Start Position (LAT)
LocalMotionPos=500 Factory TZ Center Positioning Control Panel
PatientUnloadPos=500 Factory Patient Unload Position
ScanPosTolerance=3175 Factory +/- Tolerance for TZ
[AyMotor]
SpineInitPos=460000 Factory C-Arm scan start position
FlatPos=510000 Factory Detector flattening start position
FlatMoveDist=25400 Factory Detector flattening move distance
LocalMotionHomePos=460000 Factory Center C-Arm position
YTiltOffset=-127000 Factory
TBarStartPos=152400 Factory Tissue BAR Start Position
[ArMotor]
SpineInitPos=500 Factory Spine Initial Position
LateralInitPos=4980 Factory Lateral initial Position
ScanPosTolerance=60 Factory +/- Tolerance for Scanning
LocalMotionMax=60 Factory Not Max Talerance for Center
LocalMotionMin=-60 Factory Min Center Position Tolerance
[Gain] Collimator & gain selection
SpineHiGain=3,3 Factory
SpineLoGain=3,3 Factory
SpineHiGain=6,3 Factory
SpineLoGain=6,2 Factory
SpineHiGain=7,2 Factory
SpineLoGain=7,1 Factory
SpineHiGain=8,1 Factory
4-39
SpineLoGain=8,0 Factory
SpineHiGain=10,2 Factory
SpineLoGain=10,1 Factory
DecubHiGain=3,3 Factory
DecubLoGain=3,3 Factory
HipHiGain=6,3 Factory
HipLoGain=6,2 Factory
HipHiGain=7,2 Factory
HipLoGain=7,1 Factory
IVAHiGain=6,1 Factory
IVALoGain=6,1 Factory
[Flattening]
DailyFlatFile=FLT721NA.DCL Automatic
[Aperture] Size Parameters: Number, Length, Width
Size=0, 2.510, 0.04 Factory
Size=1, 2.510, 0.04 Factory
Size=2, 2.510, 0.04 Factory
Size=3, 2.510, 0.04 Factory
Size=4, 2.510, 0.04 Factory
Size=5, 2.510, 0.04 Factory
Size=6, 2.400, 0.04 Factory
Size=7, 2.400, 0.02 Factory
Size=8, 2.510, 0.04 Factory
Size=9, 2.510, 0.04 Factory
Size=10, 2.510, 0.04 Factory
Size=11, 2.510, 0.04 Factory
Size=12, 2.510, 0.04 Factory
Size=13, 2.510, 0.04 Factory
Tolerance=10 +/- Tolerance to center of aperture
Slot=0, 4439 Factory Number and distance to center of aperture
Slot=1, 4139 Factory
Slot=2, 3679 Factory
Slot=3, 3464 Factory
4-40
Block=10,1559 Factory
Block=11,1309 Factory
Block=12,1059 Factory
4-41
Block=13,809 Factory
Block=14,559 Factory
Block=15,309 Factory
[HeatUnits]
Q4_HA
Q4_HAT
1.445 T = 0.617 0.980 Fa
Fact
cto
ory Si
Sim
themStep
il
ilar
ar to Q4,
Q4, exc
excep
phantom eptt th
with thextra
emmea
eas
sur
urem
emen
ents
absorber ts are
are tak
take
material en on
on
placed
on top of it.
T4 = 2.628 Factory Thickness (height in inches) of the block phantom.
T4_HAT = 7.00 Factory Thickness (height in inches) of the block phantom
with extra absorber.
DELTA0 = 0.493 Factory Thickness of tissue segment in the filter wheel.
T0_N = 6.85 Fact. or F.S. Overall height of the phantom (in inches) shipped
with the Discovery. (Varies with each phantom.)
HIA_N = 1305.39 Automatic Hi Air,
LOA_N = 1622.56 Automatic Low Air,
4-42
4-43
QC_HILO_FACT_MAX =
QC_HILO_FACT_MAX Fa
Fact
cto
ory Lo
Lon
ng-
g-te
term
rm dr
driift chec
check
k war
warning
ning messa
essag
ge lim
limit.
it.
200.0
QC_HILO_FACT_MIN =
QC_HILO_FACT_MIN Fa
Fact
cto
ory Lo
Lon
ng-
g-tter
erm
m dr
drif
iftt chec
check
k erro
rorr mess
messaage lim
limit.
it.
100.0
UPDATED = 06/20/94 Au
Auto
toma
mati
ticc Da
Date
te and
and ttim
imee the
the ARR
ARRC.T
C.TXT
XT file
file wa
wass las
lastt upd
updat
ated
ed..
12:00:00
UPDA
UP DATE
TED_
D_N
N= Auto
Automa
mati
ticc Da
Date
te La
Last
st Up
Upda
date
ted
d - No
Norm
rmal
al
UPDA
UPDATE
TED_
D_NL
NL = Auto
Automa
mati
ticc Da
Date
te La
Last
st Up
Upda
date
ted
d - Late
Latera
rall
Section 5
REMOVE AND REPLACE
PROCEDURES
This section describes how to remove and replace the Field Replaceable Units (FRUs) in the
Discovery.. To safely perform a FRU removal or replacement, take care to follow the procedure
Discovery
precisely as written.
Note: Whenever
Whenever a component
component in the x-ray
x-ray path is
is replaced,
replaced, you must
must recalibrate
recalibrate an
and
d rerun QC.
QC.
Front
Din Rail
Encoder Assy.
Power Supply
Multi-Voltage
Multi-Voltage DC
Distribution Board
Motor Control Board
C-Arm Y Belt
Stepper Motor
C-Arm Y Belt
To remove and replace the C-arm Y-Belt,
Y-Belt, refer to Figure 5-1 and follow the procedure
below:
5-2
1. Move
Move the
the C-arm
C-arm to the
the cente
centerr of the table.
table.
2. Shut down
down the computer
computer and
and then turn
turn off the main
main circuit
circuit breaker
breaker on the foot
foot end
pedestal.
3. At the encoder
encoder end
end of the belt, loosen
loosen the
the 2 bolts
bolts holding
holding the tension
tension block,
remove the tensioning nut, and remove and replace the belt.
4. Instal
Installl the tens
tension
ion spri
spring
ng and tens
tension
ioning
ing nut.
nut.
5. Tighten
Tighten the
the tension
tension nut so that
that the spring
spring compres
compresses
ses to 7/8
7/8 inch. The
The bracket cut-
cut-
out can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut).
6. Tighten
Tighten the
the two mounting
mounting bolts
bolts holding
holding the tension
tension block.
7. Restore
Restore power,
power, boot the compute
computerr and login as Service.
Service.
8. Perfo
Perform
rm the
the proc
proced
edure
ure in “MOTOR$AY”
“MOTOR$AY” on page 3-35.
3-35 .
5-3
Control Panel
A/SL - 9 button
C/W - 7 button
KP1085_004-0502
5-2. T
Figure 5-2. Table
able Y FRUs (C and W Models only)
Table Y Encoder
Table Y Belt
5-6
10. Perform
Perform the procedu
procedure
re in “MOTOR$TY (A, W and Wi models only)” on page 3-
38.
38.
Note: When starting
starting this procedure, make sure the encoder coupling setscrew is
not tightened.
12. Perform
Perform the procedu
procedure
re in “MOTOR$TY (A, W and Wi models only)” on page 3-
38.
38.
Note: When startin
starting
g this procedure,
procedure, make
make sure the
the encoder
encoder coupling setscrew is not
coupling setscrew
tightened.
3. Pull the
the tabletop
tabletop to the left
left far enough
enough to remove
remove the right
right table
table rail end cover
cover,,
and remove the cover
cove r. Mark this as the right cover so that it is not confused later
with the left cover (they are not interchangeable).
4. Pull the
the tabletop
tabletop to the right
right far enough
enough to remove
remove the left
left table
table rail end cover
cover,,
and remove the cover
cove r. Mark this as the left cover so that it is not confused later
with the right cover (they are not interchangeable).
5. Unplug the encoder
encoder from
from the Motor
Motor Control
Controller
ler Board
Board and pull
pull out the cable.
cable.
6. Remove the couplin
coupling
g holding
holding the encoder
encoder to the
the belt
belt drive.
drive.
7. Remove
Remove the
the encoder
encoder from
from the
the bracket
bracket assem
assembly
bly..
8. Replace the
the encoder
encoder on the bracket
bracket,, install
install the
the coupling
coupling but do
do not tighten
tighten the
the
screws.
9. Restore
Restore power,
power, boot the compute
computerr and login as Service.
Service.
10. Perform the MOTOR$TY calibration
calibration procedure on page 3-38.
3-38.
Note: When starting this procedure, make sure the encoder
e ncoder coupling setscrew is not
tightened.
5.4 Table X FRUS
This section describes how to remove and replace the FRUs associated with Table X
motion of the Discovery (see Figure 5-4).
5-4).
5-8
Table X Encoder
Table X Belt
Front Endcap
Table X Motor
5-10
5-11
11. Restore
Restore power,
power, boot the computer
computer and login as Service.
12. Perform
Perform the procedur
proceduree in “MOTOR$TX” on page 3-40.
3-40.
Note: When starting
starting the MOTOR$TX
MOTOR$TX calibration
calibration procedure,
procedure, make sure the
encoder coupling setscrew is not tightened.
Linear Encoder
(old style)
Pedestal Power Plug
String Encoder
(new style, right side,
left side not shown)
5.5.1 Pedestal
To remove and replace either pedestal, refer to Figure 5-6 and follow the procedure below.
The procedure requires two pieces of 2 x 4 lumber, approximately 3 feet long, and two
pieces of foam padding.
5-12
1. Remove
Remove the
the Elect
Electron
ronics
ics Tray covers
covers..
2. Remove 6 flat head
head Phillips
Phillips screws
screws from the upper
upper pedestal
pedestal cover
cover and remove
remove the
the
cover.
3. Remove 5 flat head Phillip
Phillipss scre
screws
ws from
from the lower pedest
pedestal
al cover and remove
remove the
cover.
4. Move the
the C-arm towards
towards the pedestal
pedestal to be replaced,
replaced, but
but ensure
ensure that
that the service
service
switches on the TZ Motor Controller board are accessible.
a ccessible.
5. Place two
two 2 x 4s on
on end, on top
top of the
the tank. The
The 2 x 4s should
should be
be oriented
oriented front to
back, on the tank, to support the table. Place foam padding between the 2 x 4s
and the table to protect the tape safety switch.
6. Push
Push in
in the
the Emerg
Emergency
ency Stop
Stop swit
switch.
ch.
7. On the TZ Motor Controller
Controller board, set
set the Normal/
Normal/Servi
Service
ce switch
switch to Service
Service and
and
set the Direction switch to Down.
8. Press the
the left and
and right switches
switches together
together until
until the table
table just rests
rests on the
the 2 x 4s.
9. Remove the
the top bolts
bolts on
on the pedestal
pedestal (9/16”)
(9/16”).. The cosmetic
cosmetic bracket
bracket will come
off.
Note: If removing the left pedestal, the left T-rail
T-rail will become loose. Be careful
not to drop this T-rail.
10. Press the appropriate pedestal switch
switch on the TZ Motor Controller board to lower
the pedestal completely.
completely.
11. Remove the pedestal power plug, the lower four pedestal bolts,
bolts, and remove the
the
pedestal.
5-13
12. Remove the end plates from the old pedestal and install on the new pedestal
being careful to maintain their orientation (6mm Allen screws). Apply a small
amount of Loctite when installing the screws.
13. Replace the pedestal and install the pedestal lower bolts
bolts,, but leave them loose for
now.
14. Replace
Replace the power
power plug.
plug.
15. On the TZ Motor Controller
Controller board, set the Direction
Direction switch to Up.
16. Press the appropriate pedestal switch
switch on the TZ Motor Controller board to raise
raise
the pedestal until it just touches the T-rail.
17. Install the cosmetic bracket
bracket and upper pedestal bolts leaving tthe
he bolts loose for
now.
18. Using the switches on the TZ Motor Controller board, raise both pedestals and
remove the 2 x 4s and foam.
19. Measure from the inside of one T-rail
T-rail to the inside of the other
other.. It must be 65
inches at both the front and back. If it is not, move the pedestal until the measure-
ments are correct.
20. Tighten
Tighten the lower and upper pedestal
pedestal bolts.
bolts.
21. On the TZ Motor Controller
Controller board, set the Direction
Direction switch to Down.
22. Using the switches on the TZ Motor Controller board, lower both pedestals all
the way down.
23. On the TZ Motor Controller
Controller board, set the Service switch
switch to Normal and press
Reset (large black button).
MOTOR$TZ calibration by selecting Utili-
24. Login as Service and perform the MOTOR$TZ
ties|Service Utilities|SQDRIVER from the Discovery Main Menu screen
(you must be in Service Mode). At the CARM$$$> prompt, type MOTOR$TZ
<Enter>
Refer to the MOTOR$TZ calibration procedure on page 3-33 for details.
25. After the calibration process completes, replace all covers.
Refer to Figure 5-5 and Figure 5-6 and follow the procedure below:
5-14
1. Remove
Remove the electr
electroni
onics
cs tray
tray covers.
covers.
2. Push in the
the Emergency
Emergency Stop
Stop Switch
Switch then press
press and hold
hold the Table
Table Up Switch
Switch to
move the table all the way up.
WARNING
ARNING:: Shut d
down
own th
thee computer
computer and
and then
then turn off
off th
thee main circu
circuit
it br
breaker
eaker on the foot
end pedestal.
WARNING
ARNING:: Do no
nott allow
allow the string
string to snap back
back into
into th
thee enc
encoder
oder after being exte
extended.
nded. This
can permanently damage the unit.
8. Route the
the new encoder
encoder cable
cable through
through the path
path of the
the old encoder
encoder and
and plug it
it into
into
the TZ Drive board.
9. Install
Install the
the rotary
rotary encoder at the bottom
bottom using the
the screw
screw and two
two washers
washers pro-
vided.
Note: The bracket
bracket must sit
sit on the top surface
surface of the base
base frame before
before
tightening.
10. Replace the linear encoder top screw by extending the string and fastening to the
the
top using screw and washers provided.
11. Turn on the instrument
instrument power,
power, boot the computer and login as Service.
12. Perform
Perform the MOTOR$TZ
MOTOR$TZ calibration selecting Utilities|Service Utili-
calibration by selecting
ties|SQDRIVER from the Discovery Main Menu screen (you must be in Ser-
vice Mode). At the CARM$$$> prompt, type MOTOR$TZ <Enter>
Refer to “MOTOR$TZ (Discovery A and SL only)” on page 3-33 for details.
13. When calibration and adjustment
adjustment is done,
done, replace the pedestal covers.
5.5.2.2 Adjustment
Use this section only when the pedestals
pede stals are not within the 20 counts specified.
1. Typ
ypee Y<En
Y<Entter
er>.
>.
A screen similar to the following appears.
DeviceState E_OK
PedestalMode OPERATIONAL
5-15
Table Status OK
Left Pedestal AT_TARGET
Right Pedestal AT_TARGET
Position 199649
PosLimitPosition 199930
NegLimitPosition 0
LowerLeft 384
UpperLeft 3235
LowerRight 383 Adjust the Right String
UpperRight 3234 Encoder center hub until
PositionAverage 8 EncoderPosition
DriverVersion 8.02 equals UpperLeft.
DeviceVersion 2.30
EncoderPosition 3230
EncoderNegLimit 383
EncoderPosLimit 3234
A384;3235;383;3234
5-16
C-Arm Interface Bd
Tank Assembly
L-Bracket (2x)
9.
9. Perform “Area, BMD, and BMC Calibration” on page 4-34
4-34..
10. Log out of Service
Service and login
login as User.
User.
11. Perfo
Perform
rm “T
“Test
est Scan Modes” on page
pa ge 3-64.
3-64.
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
3.. Set Aperture to 7.
3
4. Shut down
down the computer
computer and
and then turn
turn off the main
main circuit
circuit breaker
breaker on the foot
foot end
pedestal.
5. Remove the
the lower
lower C-arm
C-arm cover by removing
removing 2 Phillip
Phillipss screws
screws and slidin
sliding
g it out
from the front. Also, remove the bottom cover.
6. Remove
Remove the
the 4 Philli
Phillips
ps screw
screwss holding
holding the
the XRC
XRC (see
(see Figure 5-7).
5-7).
7. Pull the
the XRC forwar
forwardd far enough
enough for
for access to the Filter
Filter Drum (i
(itt is not neces-
sary to remove the XRC cables.
8. Remove the rear
rear tank
tank cover
cover (covers
(covers the
the C-arm
C-arm Interface
Interface board).
board).
9. Unplug the
the 2 Filter
Filter Drum cables from the C-arm
C-arm Interface
Interface board.
board.
10. Remove 3 hex head screws (3/32
(3/32 inch Allen screws).
screws).
5-18
1. Using the
the Control
Control Panel,
Panel, move
move the table
table in as far as it
it will
will go for
for easier
easier access.
access.
2. Start
Start the
the X-Ray
X-Ray Surve
Survey
y Utilit
Utility
y by selec
selecting Utilities|Service Utilities|X-Ray
ting
Survey (see “X-Ray Survey” on page 9-1 for information on the utility).
3.. Set Aperture to 7.
3
4. Shut down
down the computer
computer and
and then turn
turn off the main
main circuit
circuit breaker
breaker on the foot
foot end
pedestal.
5. Remove the
the lower
lower C-arm
C-arm cover by removing
removing 2 Phillip
Phillipss screws
screws and slidin
sliding
g it out
from the front.
6. Remove
Remove the XRC (see
(see proced
procedure
ure above).
above).
7. Remove the Filte
Filterr Drum
Drum Assembl
Assembly
y (see
(see procedure
procedure above).
8. For A and SL models
models only:
only: Lock the
the C-arm
C-arm in place
place by securing
securing the
the shipping
shipping
brackets on each side of the arm.
5-19
21. Starting
Starting with
with the
the “Check Tube kV Peak Potential” on page 3-28,
3-28, perform all the
procedures in that section in order, with the exception of the Calibrate Motors
and Check Laser Positioning Offset procedures.
Gas Spring
C-Arm R Encoder
Motor Controller
Board
C-Arm R Motor
C-Arm R Belt
C-Arm R
Encoder Belt
5-10. C-Arm R FRUs (Inside View)
Figure 5-10.
5-23
7. Remove the
the belt from the front
front belt
belt clamp (four Phillips
Phillips screws)
screws)..
8. Install
Install the new
new belt in reverse
reverse order
order,, front belt
belt clamp
clamp first,
first, then over
over the pulleys,
pulleys,
and then to the rear belt clamp.
9. Tight
ighten
en bot
both
h belt
belt clam
clamps
ps..
5-24
12. Replace the cables and replace the belt on the motor pulleys. Be ssure
ure to restore
the cable shield and ground strap terminations.
13. Replace the belt bracket and loosen the belt tension block (two 1/4" bolt
bolts).
s).
14. Adjust the tension nut so that the spring is
is compressed to 7/8" from
from the inside of
1. Using Emer
Emergency
gency Motion,
Motion, rotate
rotate the
the C-Arm
C-Arm to the Lateral
Lateral position.
position.
2. Remove the
the retaining
retaining clip
clip from
from the top
top of the
the gas spring
spring and
and remove
remove the spring
spring
from the stud.
3. Remove
Remove the
the nut that
that holds
holds the lower
lower stud
stud in place.
place.
4. Remove
Remove the
the stud
stud and spri
spring
ng toget
together
her..
5. Prior to installin
installing
g the new spring,
spring, lubric
lubricate
ate the ball
ball studs
studs at both ends
ends with
with white
white
lithium grease.
6. Put the
the lower
lower stud
stud on the
the new spring
spring and attach
attach the retaining
retaining clip.
7. Install
Install the top of the
the spring
spring to the top stud and
and replace
replace the retainin
retaining
g clip.
080-1085_004-0513
4. Remove the cables and the screws on the existing ADC Board (140-0087) and
remove the board.
Note: The 180-0185 and 180-0189 ribbon cables on the ADC Board will not be
reused. Also, the four 6-32x.25 screws (200-0356) securing the ADC
board will not be reused.
180-0292 010-1653
laser cable
180-0189
180-0185
140-0087
180-0419
200-0356 (4x)
5-12. Detector Assembly 010-1653
Figure 5-12.
5-27
ASY-00271
180-0292
laser cable
370-0056
180-0419
5-28
Laser Assembly
Laser
KP1085_004-0514
Counterweights
Convertor Bd
Convertor
Analog to Digital
7. Restore
Restore power,
power, boot the compute
computerr and login as Service.
Service.
8.
8. Perform “A/D Gain Control Adjustment” on page 4-264-26..
9. Replace
Replace any covers
covers that
that were
were removed
removed..
10. Perfo
Perform
rm “Detector Flattening ” on page 4-29.
4-29.
5-30
5.8.1 Aperture
To remove and replaceStepper Motor
the Aperture Stepper Motor (320-0041), refer to Figure 5-16 and
follow the procedure below:
Motor Shaft Bracket
Aperture Stepper Motor
Motor Mount
Aperture
Posit
osition
ion Belt
Tension Nuts
PCB Bracket
Clamping Nut
5.9.1 Drum
To remove Encoder
and replace PCB
the Drum Encoder
Encod er PCB (140-0089), refer to Figure 5-16,
5-16, Figure 5-
17 and Figure 5-18 and follow the procedure below:
Motor Mount
2. assembly
Remove the
assembly,t,he
andAperture
Aper ture Assem
Assembly
2 Phillips
P hillips blythat
screws by secure
removing
removing
the 4aperture
Phillips
Phillipsmotor
screws
screws that
thatt,secure
moun
mount, setocure the
the
the drum
end plates.
3. Unpl
Unplug
ug the
the PCB
PCB ca
cabl
ble.
e.
4. Remove and
and replace
replace the PCB
PCB (ensure
(ensure that
that the encoder
encoder wheels
wheels are
are not touching
touching
the sensors.
5. Attach Ch1 probe
probe to SEGMENT
SEGMENT test
test point
point (TP3)
(TP3) on the
the C-Arm
C-Arm Interface
Interface board.
board.
6. Attach Ch2 probe
probe to BRASS
BRASS test
test point
point (TP2)
(TP2) on the C-Arm Interface
Interface board.
board.
7. Conn
Connect
ect pr
prob
obee tto
o gro
groun
und.
d.
8. Go to X-Ray
X-Ray Survey
Survey and turn
turn on the
the Filter
Filter Drum
Drum motor
motor..
9. As the filter
filter drum
drum motor turns on
on and rotates
rotates look on the
the C-Arm
C-Arm Interface
Interface board.
board.
a. The yellow
yellow “Index”
“Index” LED blinks
blinks (on then
then off)
off) as the
the Index mark passes
passes the
optical interrupter.
interrupter.
b. Verify the
the green “Top
“Top of Drum”
Drum” and “Brass
“Brass on Top”
Top” LED (D8)
(D8) blink
blink as the
drum spins.
c. The green
green “AC Lock”
Lock” LED
LED (D7) lights
lights (steady
(steady,, not blinki
blinking)
ng) after
after a few
few rev-
olutions of the drum.
10. Measure from the rising edge of the
the BRASS to the rising edge of the SEGMENT
(it does not matter which edge leads).
lead s). The time measured must be less than 500
µs (500 µs = 0.5 ms). See Figure 5-19 and Figure 5-20.
5-20.
5-35
Segment
Signal
Jigger Brass
Signal
500μs
Jigger Brass
Signal
Segment
Signal
500μs
5-36
3. Loosen the
the 2 drum motor mount
mount screws,
screws, the idler
idler screw
screw and the outer pulley
pulley set
screw.. Remove the belts.
screw
4. Install
Install the Filter
Filter Drum Alignm
Alignment
ent Pin (099-01
(099-0110),
10), small
small end first,
first, through
through the
the
slotted holes and into the small hole at the base plate. If the pin is installed prop-
erly, the drum will not rotate.
5. Install
Install the 150
150 tooth
tooth belt on
on the back pulley (ensure
(ensure the belt
belt is positi
positioned
oned under
under
the idler).
6. Instal
Installl the 130
130 tooth
tooth belt
belt on
on the front
front pulley
pulley..
7. Tighten
Tighten the
the 2 drum motor mount screws
screws (the
(the motor
motor mount
mount is spring
spring loaded
loaded to
seek proper tension).
8. Tighten
Tighten the idler
idler screw
screw (the
(the idler is
is spring
spring loaded to
to seek proper
proper tension).
tension).
9. Tight
Tighten
en the
the outer
outer pull
pulley
ey set
set screw
screw..
10. Remove the Filter
Filter Drum
Drum Alignment
Alignment Pin.
11. Pull the
the Emergency
Emergency Stop
Stop button.
button.
5.9.3 Stepper
Stepper Motor Assembly
To remove and replace the Stepper Motor Assembly,
Assembly, refer to Figure 5-18 and follow the
procedure below:
1. Shut down
down the computer
computer and
and then turn
turn off the main
main circuit
circuit breaker
breaker on the foot
foot end
pedestal.
2. Remove
Remove the Apertur
Aperturee Asse
Assembl
mbly
y.
3. Remove
Remove the
the drum
drum belts
belts (see
(see proc
procedur
eduree above).
above).
4. Remove
Remove the stepper
stepper motor
motor pulley
pulley..
5. Remove and replace
replace the
the motor
motor (4 flathead
flathead Phillip
Phillipss screws).
screws).
6. Replace the
the pulley
pulley,, use Loctite
Loctite 222
222 (540-0100)
(540-0100) on
on the set
set screw
screw.
7. Refer to the
the Drum Belts procedure
procedure above
above and replace
replace the belts.
belts.
Idler Arm
Bearing
250-0046
Endplate
Alignment Hole
Side Plate
Lead Shield
Endplate
5-38
6. Remove
Remove both
both drum
drum bel
beltt pull
pulley
eys.
s.
7. Remove
Remove the drum
drum from
from the endplat
endplates.
es.
8. Remove one
one drum endcap
endcap (4 Phillips
Phillips screws
screws)) and remove
remove the
the inner drum.
drum.
9. Repl
Replac
acee the
the bear
bearin
ings
gs..
10. Reassemble the drum and replace (and tighten) the screws in the endcap.
Note: When replacing
replacing the
the end cap ensure that
that the flats
flats on each side of the drum
drum
shaft are facing the same way (this happens where the two side-by-side
holes line up).
11. Replace the 2 endplates. The endplate with the cutout (for the PCB) goes on the
drum end away from the alignment hole (see Figure 5-21).
5-21).
12. Replace the
the lead shields and side
side plates (6 screws).
screws).
13. Align the 2 endplates onto the holes
holes on the recessed area on the base plate. Orient
the front endplate towards the 2 slotted holes. Orient the rear endplate on the
opposite set of holes and position so that the endplate side with cutout (for PCB)
is facing the "C" cutout of the base plate. Replace the 4 flat washers, split locks,
and 4-40 x 3/8 screws that secure the endplates.
14. Insert the Drum Spacing Fixture (099-0296) between the endcap and the front
endplate. Push the filter drum against the fixture to center it between the two
blocks. Do not remove the fixture at this time.
15. Install the spacers
spacers and pulleys at the front endplate (see Figure 5-18).
5-18). Ensure that
the pulley set screws line up with the flat, and use Loctite 222 on set screws.
Press the filter drum against the fixture, and the pulley against the endplate, while
tightening the set
se t screw.
Note: Leave the outer pulley set screw loose until belts are installed.
16. Remove the
the Drum Spacing
Spacing Fixture.
Fixture.
17. Replace
Replace the inner encoder wheel
wheel (see
(see Figure 5-17).
5-17). Press (squeeze together) the
drum and encoder wheel against the endplate and tighten the set screw (use Loc-
tite 222).
18. Replace
Replace the outer encoder wheel.
wheel.
19. Replace the drum encoder PCB. Ensure that the encoder wheels do not contact
the sensors.
20. Replace the stepper motor assembly (2 screws on the m
motor
otor mount, and 2 screws
on the base plate). Leave the motor mount screws loose for now.
21. Replace the drum belts
belts (see the
the Drum Belts procedure).
22. Tighten outer pulley set screw after belts
belts are installed.
installed.
23. Refer to the Aperture Assembly
Assembly procedure and install tthe
he aperture assembly.
assembly.
5-39
Bearing Endcap
250-0045
Bearing Bearing
250-0045 250-0045
Bearing
250-0045
Endcap
5-40
5.11
5.11 FRU Lists
The following tables provide the information necessary to identify and order the correct
FRU.
Please note:
• The table
tabless are liste
listed
d by Figure,
Figure, and
and then by
by Cable,
Cable, Miscel
Miscellan
laneous
eous,, Mobile,
Mobile, and
and
then Special Tools.
• The "Used
"Used On"
On" column
column indica
indicates
tes whic
whichh Discove
Discovery
ry model(
model(s)
s) the
the Part Numb
Numberer is
used; "All" indicates all Discovery models.
5-41
5-43
5-45
Section 6
FAULT ISOLATION
This section provides information to help identify the source of a problem in the Discovery
system. The three general categories are:
Problem... Refer to the section titled... On page...
Dead system, or power problem Power Problems 6-1
Scanner motion problem Motion Problems 6-2
Computer display problem Display Problems 6-5
6-1
Also...
Problem... Possible cause... Corrective action...
Laser does not Safety feature if system thinks C-arm
C -arm Restart in Service mode
turn on is at head of table From Service Utilities, run SQDRIVER
At CARM$$$$> prompt, type
LASER_SAFTY
LASER_SAFT Y 0 <Enter>
Exit from SQDRIVER
Reboot the computer
Start by identifying the bad axis (if it is not obvious). Ask the operator for symptoms and
check the error log. See Figure 6-1 for possible motion directions.
6-2
AY
AR
TZ
TX
TY
6-3
CAUTION: Running any motor subsystem to its limit (in either direction) does not
damage the motor. However, if the motor encoder is not properly
calibrated, it may be damaged.
6-5
Also...
Run the hardware checker see “SQVERIFY” on page 9-10
6-6
6.7.3 Noise
The term "noise" is used here to describe any flaw, or irregularity in the display (dots,
specks, uneven lines, etc.) or similar problem. The following suggestions apply to a
Discovery system that exhibits noise in the display:
Check... Refer to...
Tube kV peak potential see “Check Tube kV Peak Potential” on page 3-28
Tube current see “Check Tube Current ” on page 3-30
X-ray beam alignment see “X-Ray Beam Alignment ” on page 4-4
Signal strength and noise see “X-Ray Survey” on page 9-1
Aperture position and aperture belt see “Aperture Calibration ” on page 4-9
Filter Drum is turning Figure 5-16
Filter Drum belt Figure 5-21
Green LED’s on C-Arm Interface board Figure 5-7,
5-7, Section 8
Aperture plate assembly, first Figure 5-17
precollimator,, ssecond
precollimator econd precollimator and
collimator for specs of lead and other
deformities.
Also...
Run the hardware checker see “SQVERIFY” on page 9-10
6.7.4 No Display
The term "no display" is used here to describe:
• no scan display
• co
comp
mple
lettely
ely whi
whitte scr
scree
eenn
• co
comp
mple
lete
telly dar
darkk scr
scree
een
n
• “blo
“blotc
tchy
hy”” or
or ““gr
grai
ainy
ny”” dis
displ
play
ay
This type of display problem is usually related to the detector subsystem or the X-ray
subsystem.
The following suggestions apply to a Discovery system that exhibits no scan display:
Check... Refer to...
Tube kV peak potential see “Check Tube kV Peak Potential” on page 3-28
Tube current see “Check Tube Current ” on page 3-30
X-ray production see “Field Service Preventive Maintenance” on page 7-1
Signal strength and noise see “X-Ray Survey” on page 9-1
Filter Drum is turning Figure 5-6
5-6,, Figure 5-18 through Figure 5-23
Filter Drum belt Figure 5-18,
5-18, Figure 5-23
Green LED’s on C-Arm Interface board Figure 5-6,
5-6, Section 8
6-7
No display may indicate a bad Detector see “Data Acquisition System (C, W, and SL) ” on
Assembly page 2-20
Ch
Chec
eck
k ap
aper
erttur
uree po
posi
siti
tion
on an
and
d ape
pert
rtur
uree bel
eltt see “Aperture Calibration ” on page 4-9
Also...
The bone and/or tissue segments in the filter drum may be Figure 5-18
defective or out of spec. If so, replace and restart system testing. through
Figure 5-24
6.11.1 No X-Rays
If... Check... Refer to...
X-Ray (24VDC
Distribution LED)
Board on lit
is not X-ray EnableBoard
Distribution LED should
on Control
be onPanel and
6-9
block and pin line up, the alignment brackets need to be repositioned. (Check
captive screws and jam nuts.)
2. If you align
align the X-ray
X-ray beam to both
both drop-off
drop-off points
points but the range
range is narrow
narrow,, (on
aperture 10 you should have 3-8 turns of the hex driver) the collimator cup and/or
7. and
Check for
for leadAssembly.
Aperture fragment
fragments
Assembly. s in the
the X-ray
X-ray beam. Inspec
Inspectt the Collimat
Collimator
or Cup, disk,
6-10
8. If a failure
failure occurs
occurs during
during a whole body
body,, recheck the
the TX motor
motor cal files.
files. The X-
rays may be hitting the table edge.
e dge. Make sure the drives are op
operating
erating normally.
normally.
Examine the Detector Flattening scans for any unusual indications using Utili-
ties|Scan File Plot and selecting the WBAIRQC scan. If an error message is
displayed, go to “Detector Flattening ” on page 4-29.
4-29.
If... Suggestions
(Software is inhibiting laser when table head is near the laser)
Move the C-arm to the center of the table and retry.
Do the following:
1. Check black
black and
and yellow
yellow wires
wires from
from the C-arm
C-arm Interf
Interface
ace PCB
PCB to the
the
Laser does
not turn on underside of the Detector Assembly.
Assembly.
2. Shut off
off instrument
instrument power (Emerg
(Emergency
ency Stop
Stop Switch
Switch on Control
Control
Panel).
3. Discon
Disconnect
nect laser
laser power
power connect
connector
or..
4. Connect a DVM to the
the plug coming
coming from the Detector
Detector Assembly
Assembly..
5. Tur
urn
n las
laser
er po
powe
werr on.
on.
6. Check
Check co
conne
nnect
ctor
or for
for +5V
+5VDC
DC..
Check C-arm Interface Board, Distribution Board, all cables.
Check computer.
At installation only, type “laser safety=0” in SQDRIVER and reboot.
Check AY
AY motor calibration file. If data is corrupt or uncalibrated, laser may
not turn on, despite position of C-arm.
6-11
Laser Check laser block assembly mirrors for breaks, cracks, or misalignment
crosshair
Check that shutter is open
beam is
defective Check that aperture is unobstructed
Loosen
10)
10) the securing screw to adjust the laser to perpendicular
p erpendicular (see Figure 5-
Front
6.15.3 Tightening
Tightening the Lexan Cup Screws
Tighten the Lexan Cup seals in the following order using the torque settings in the above
table:
7
1 4
5 6
3 2
1 4
3 2
2 4
1 3
6-5. Transformer Screws
Figure 6-5.
3. Repeat this
this pattern
pattern in
in a clockwise
clockwise direction
direction and tighten
tighten all remaining
remaining screws.
screws.
6-14
replaceable units or
loosen or remove anyrequired
screwsadjustments
on the tank. inside the tank, so there is no reason to
5
1 L
3 A
C
I
T
I
R
FRONT C
T
S
O
M
6 2 4
Section 7
PREVENTIVE MAINTENANCE
This section lists the procedures that should be performed by trained service personnel at least
once per year, and preferably at six-month intervals.
NOTICE
The Discovery system meets applicable FDA radiation performance standards through
its useful and expected life provided no components or parts are removed from the
system and no unauthorized adjustment
ad justment or unauthorized replacement of certified
components is perform
7-1
[ ] Print copies of the baseline phantom scan, recent phantom scan, and the results of scan
averaging for the 10 spine scans in the previous step. Compare these printouts with
those from the last PM for possible problems, and send the printouts to Hologic to be
kept with the service records for this system.
[ ] From the Discovery Main Menu, select QC|QC Data Management|Plot. Setup the
Discovery Main
QC Plot to include QC scans from at
a t least the last wo years. Display and print a copy
of the BMD, BMC, Area, HiAir, and LoAir plots to be included in the paperwork
being returned to Hologic.
[ ] During a scan, verify that
that pressing the red Emergency switch immediately stops
Emergency STOP switch
all C-Arm motion, table motion, and X-ray production. X-ray production should be
monitored by a Victoreen 450P or equivalent.
[ ]T
Test
est all tape
tape switches. The tape switches are located along the length
length of the front and
back of the table, and mounted on the frame under the table. Verify
Verify that pressing a tape
switch immediately stops all C-Arm motion, table motion, and X-ray production. X-
ray production should be monitored by a Victoreen 450P or equivalent.
Note: The front (table) tape switch is disabled until the C-arm is at 0 °, so this switch
should be checked with the C-arm at a position greater than 30 °.
[ ] Replace (if necessary) and adjust the motor drive belts as follows:
Note: Detailed instructions for drive belt replacement and adjustments can be found
in the REMOVE AND REPLACE section of this manual.
[ ] For each drive belt (C-Arm Y,
Y, C-Arm R, Table
Table X, and Table
Table Y) perform the following:
• Loosen
Loosen the
the 2 mount
mounting
ing bolts
bolts hold
holding
ing the tensio
tension
n block.
block.
• Tighten
Tighten the
the tension
tension nut so that
that the spring
spring is compressed
compressed to 7/8 inch.
inch. The bracket
cutout can be used as a measuring guide (the inside of the washer should be flush
with the bracket cut). For the AR tension spring, adjust to 7/8" from the inside of
one washer to the inside of the other.
• Tighten
Tighten the two mounting
mounting bolts holding
holding each
each drive
drive belt’
belt’ss ttensio
ension
n block.
block.
Note: It is not necessary to perform the MOTOR calibration procedures after
tensioning the motor drive belts.
[ ] Ensure that all cable connections are tight.
tight.
[ ] Run CHECKDISK on the C drive. This utility
utility program can be started by clicking on
start|Run…, then typing CHKDSK, and then clicking the OK button. DO NOT run
Scandisk on any Archive media.
[ ] Run Dis
Disk
k Defragmenter toto and compact and re-order the files on the computer's hard
c licking on Start|Run…, then typing
disk. This utility program can be started by clicking
DEFRAG, and then clicking the OK button.
7-2
[ ] Run a computer virus checker (any major brand that is current for latest
latest virus types).
[ ] Clean the fan filters, paying special attention to the computer fan filter.
filter.
[ ] Clean all exterior metal surfaces and wipe off the rails.
[ ] Clean the monitor screen, keyboard, and mouse ball rollers.
[ ] Clean the inkjet printer rollers and ink cartridge electrical contacts with isopropyl
alcohol.
Motion
Motio n
Guide Bearing
Grease Fitting
Seal
Guide Rail
2. Gr
Greas
easee the
the guid
guidee bear
bearin
ings
gs..
Note: The bearings must be greased every 100km (62mi) of travel, depending
on usage. This may range between
be tween one and two years. The following table
lists more information about the grease to be used.
Generic Brand Names Manufacturer
EP-2 Lithium Beacon 325 ESSO
soap-based grease Alvania Grease RA Shell
Mobilux Grease No. 2 Mobil
Isoflex Super LDS 18 Kluber
Consult Hologic Field Service for more information on obtaining the proper tools,
material, and procedure.
7-4
Section 8
PCB SUMMARY
SUMM ARY INFORMA
INFORMATION
TION
Distribution Board
LED’s1 Signal Voltage Source Jumpers Refer to...
D1 On +7VDC Int Voltage Reg. JP1 (panel) Out
8-2
C-Arm Interface
LED’s Signal Voltage Source Jumpers Refer to...
D1 On +28VDC Ext Signal Dist. JP7 DRUM Out
D3 On +24VDC Ext Signal Dist. JP8 MAIN Out
D6 On +5VDC Int. Voltage W1, NORMAL(1-2) In
Reg. W2 TEST (2-3)
- - +7VDC Int. Voltage W3 SYS-RESET In
Reg.
- - +15VDC Ext Signal Dist. W4 TEST- Out
MODE
- - -15VDC Ext Signal Dist. W5 Safety In
switches
Drum JP15 Redundant for
D7 On LOCKED W5 may be Out if W5 is Figure 5-6
In
D8 Flash- DRUM AT
ing TOP
X-Ray
D9 Off X-RAY
CPU
ERROR
D10 Off X-RAY
ENABLE
D11 On X-RAY
CPU OK
D12 Off X-RAY
FAULT
8-4
Section 9
SOFTWARE TOOLS
The Discovery system software includes software tools to troubleshoot the system. They are:
• X-Ra
X-Ray
y Sur
Surve
vey
y Too
Tooll Dis
Disk
k
• SQDRIVER
• SQVERIFY
• Emer
Emerge
genc
ncy
y Moti
Motion
on (SQ
(SQKE
KEYP
YPAD
AD))
9.1.1
1. Invokingas theX-Ray
Log on as Fiel
Field SerSurvey
d Servic
vicee and start
start the
the QDR
QDR Softwa
Software
re in Serv
Service
ice mode
mode..
2. Click Utility > Service Utilities > X-Ray Survey and the X-Ray Survey utility screen
will appear (see Figure 9-1).
9-1).
Main Display
AW-00335_001-0902
Display Controls Close utility
9-1
1. The Main
Main Display
Display shows either
either a bar-grap
bar-graph
h or numeric
numeric fields
fields for each detector
detector
depending on which Display button is selected.
2. The Display
Display Controls
Controls that
that affect
affect the
the data
data shown in the
the main display
display.. Not all con-
con-
trols are used by the field engineer.
3. The Close button which exits the utility.
AW-00335_001-0902
9-2
AW-00335_001-0903
AW-00335_001-0904
9-4. X-Ray
Figure 9-4. Data as a Raw Data
AW-00335_001-0905
The
and itdefault
shouldfor this button
a lways
always remainis OFF.
OFF. This
The Field
dealsEngineer willthe
with testing notsystem
use theinFilter Wheel
different button
mode.
Note: The next set of controls have the label Filter Wheel Off Phase which indicates the
state of the Filter Wheel. These controls (HI/LO, BONE/AIR/TISSUE, 0/1/2/3 and 0/1)
indicate where the Filter Wheel will stop when it is not moving.
9-4
Label Information
This area also provides information on the Hi Voltage Setting (either ON or OFF),
Frequency, Filter Errors and HiLo Errors.
AW-00335_001-0906
9-5
9.1.3.3.4 Hi Gain.
This is a spinner control with a range of 0 to 11. Changing Hi Gain changes the displayed
detector values for the Hi portions of the current segment. The associated actual gain
values are shown in the following table.
10 4 4 16
11 4 8 32
9.1.3.3.5 Lo Gain.
This is a spinner control with a range of 0 to 11. Changing Lo Gain changes the displayed
detector values for the Lo portions of the current segment. The associated actual gain
values are shown in the above table.
9-6
9.1.3.3.7 Aperture.
This is a spinner control with a range of 0 to 13. This controls the slit through which the
X-rays pass to limit the dimensions of the X-ray beam.
bea m. This control allows the selection of
a specific aperture in the system and can be used to test aperture movement.
AW-00335_001-0907
9.1.3.4.1 Display
This has two buttons: GRAPH and TABLE. This controls what is seen on the Main
Display. In GRAPH the Main Display shows as a bar graph (see Figure 9-3) 9-3) with scale
markings on the left side according to how Scale is set (see below). It shows the lo (brown
tinted) and the hi (green tinted) voltages. In TABLE the Main Display shows the data
from the detectors as numeric data.
9.1.3.4.2 Scale
This has three buttons: RAW, LOG and VOLTS. The Default is VOLTS. The Field
Engineer will not use RAW or LOG and the button should always remain the default
(VOLTS). This control sets up the scale markings on the Main Display when the Display
is set to GRAPH and changes the units for the numeric value when the TABLE
TABLE is
displayed.
9.1.3.4.3 Segment
This has three buttons: BONE, TISSUE and AIR. The
Th e default is AIR. This controls the
data segment that are displayed in the GRASPH. The Field Engineer will not use either
BONE or TISSUE and the button should always remain the default (AIR).
9.1.3.4.4 Zoom
By default, In and Zoom
the GARAPH Out.
is displayed at minimum resolution When either function
reaches it maximum the button will be grayed out. Zoom In increases the resolution,
meaning that less of the total data range is displayed on the graph but differences in ba
barr
9-7
heights are magnified. The horizontal scale isn't affected and all detectors are displayed
regardless of the zoom value. Zoom Out will reverse what Zoom In has done.
AW-00335_001-0908
9.1.3.5.2 Reset
The Reset button re-starts the process of computing the means and standard deviations for
each filter-segment. It resets the number of samples to zero. It also resets the bits in the
Status Bits display.
AW-00335_001-0909
9-9.Moving
Figure 9-9. the Detector
includes (values
samples the current
used(most recent) the
to calculate value,
meanmean
andvalue, standard
standard deviation
deviation) and of
for each number
the sixof
filter-segments.
filter-segments. The pop-up window can be dragged wherever the user wants and remains
until Show Details is actively released.
9-10.
Figure 9-10. Show Details Window
9.2 SQDRIVER
The SQDRIVER program is accessed from the Discovery Main Menu by selecting
Utilities|Service Utilities|SQDRIVER. This program is used to calibrate the motors
and can be used as an effect tool for troubleshooting motor movement problems.
9-9
9.3 SQVERIFY
The SQVERIFY program is accessed from the Discovery Main Menu by selecting
Utilities|Service Utilities|SQVERIFY. This program is a diagnostic program, which
can be used to
SQCHECK perform
User a series
Manual of low-level
(080-0707) tests of
for detailed system components.
explanations of its use. Refer to the
9.4 SQKEYPAD
The SQKEYPAD
SQKEYPAD program is accessed
acce ssed from the Discovery Main Menu by selecting
Utilities|Emergency Motion. This program provides the Field Service Engineer with a
method for moving all motors even in conditions where a detected software error
condition would normally prevent any movement of the motors. This program can be used
as a quick means of checking most motor control functions.
9-10
Section A
ERROR MESSAGES
For any system error there may be one, or more, messages. Messages shown in the Err
Error
or Message
Message
List section below may be the first message, or a subsequent message.
All messages, for each system error, are saved in the ERRLOG file.
NOTE:
Depending on the software running on the unit additional error messages may be available that are not described in
this appendix.
A-1
Error Message List
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Address in the response message did not match the address in the sent message. [E_ADDRESS_MISMA
[E_ADDRESS_MISMATCH_I]
TCH_I]
Hardware d0 is found to be below zero, which is not possible if there are X-rays reaching the
or detector.
Software
Action: This message should be reported to Hologic
Hologic Tec
Technical
hnical Support. Include as
much information as possible about the scan.
Arm and table are in danger of colliding! [E_COLLISION_IMMINENT_I] Comment: The C/C has calculated that the C-arm and table are within 1/4".
Action: If the C-arm and table are really within this distance, the problem is with the
application (or operator) that moved them so close. If they are not, then either the
calibration information for the system is wrong or the mechanical stops for the table
and C-arm have moved. One or more of the motors may have stopped short of the
actual mechanical stop during calibration. Verify
Verify that the positive limit positions for
each motor correspond to the published specs for the model. To recover from this,
select the Utility option from the main menu and then select the emergency Motion
option from the sub-menu. By holding down the Enable button on the operator control
panel, you can use the keypad to move the table and C-arm away from each other.
A-3
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Bad checksum in message. [E_BAD_BLOCK_I] Comment: The checksum calculated by the C/C for a message did not match the one in
the message packet.
Action: Either the driver is incorrectly configured, one or more devices are jumpered to
the debug mode, there is a hardware problem on the c ommunication link (cable or
connectors) or one of more of the DSP chips in the C/C or DAS board is not properly
seated. Verify that the DEVICE=C:\MENU\MAPIDEV.SYS line in CONFIG.SYS has
the options /OVERLAP=0 and /HANDSHAKE=1. Check each stepper motor board
and verify that its address is correctly set. Verify that the TZ motor board does not
have the debug jumper installed. Verify
Verify that the communications cable is installed and
properly secured on the back of the C/C board. If the problem persists, it may be
because one (or both) of the DSP chips on the C/C or DAS board is not properly seated
in the PGA socket. Try "massaging" the DSP chip on the C/C board and the one on the
DAS board. If the problem changes (gets worse or better) when you do so, replace the
appropriate board. If the problem
pr oblem persists, place all devices in emulation mode (e.g., by
turning off instrument power and booting in the workstation
w orkstation mode). Disconnect all
devices except the C-arm from the communications link. Turn off C-arm emulation. If
there are no problems, proceed to re-connect each device and take it out of emulation
until the offending device is found. Replace the appropriate board.
BoundaryLine Internal Software Error Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Cannot generate bad detector map, too many bad detectors Action: One or more detectors is bad. See instructions for running hardware checking
program.
Cannot obtain results of scan ________ of __________
Software Comment: For some reason, the system was unable to obtain the results of a scan that
was selected for normals plotting.
Can't open a window Software Action: Ensure that only Hologic software is installed on the computer. If so, report
this message to Hologic Technical
Technical Support. Include as much information as possible
about what you were doing at the time.
A-4
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Corrupted Scan Data Detected Software This message should be reported to Hologic Technical Support. In
Include as much
information as possible about what you were doing at the time.
Could not find file extension in Software This message should be reported to Hologic Technical Support. In
Include as much
information as possible about what you were doing at the time.
Couldn't write the new record Hardware Comment: The system was unable to write a reference curve that was entered or edited.
or
Software Action: Check that the hard drive is not full. That is, check the message "Room for __
scans" in the status window and ensure that it does not say zero. If it is full, then
archive and delete one or more scans and try again.
If the hard drive is not full, then you have either a hard drive failure or a corrupt
reference curve database.
db_File error __ ..... Software Any message that begins this way should be reported to Hologic Technical Support.
or Include as much information as possible about what you were doing at the time.
db_VISTA error __ .....
MAP
APII Comm
Commeent:
nt: The
The MAP
MAPI drdriver
ver w
was
as alrea
lready
dy op
opeene
ned
d by an
anot
othe
herr ap
appl
pliicat
atio
ion
n whe
when
n the
the
driver attempted to go ON_LINE (i.e., the driver requires exclusive access to the
MAPI driver).
A-5
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
STATE Comments:
1. An application attem
attempted
pted to load or execute a state machine
machine when one was already
running, or an application attempted to execute a state machine while one or more
devices were still executing a command, or an attempt to abort the curr
currently
ently
MOTOR_
MOTOR_xx
xx Commen
Comment: t: An applic
applicati
ation
on atte
attempt
mpted
ed to modify
modify ste
steppe
pperr mo
motortor parame
parameter
terss wh
while
ile the
motor was moving (acceleration distance/type, first/fina
first/finall rates, position average, etc.),
or an application issued a MOVE_ABS while a MOVE_REL was in progress (or (o r vice
versa).
C-Arm Comments:
1. An application attem
attempted
pted to control the laser positioning light when it was under
local control or an application attempted to turn on the laser while the C-arm and
table were within the laser safety distance of each other.
Action: Be sure the table and C-arm are properly positioned before starting a scan.
If the problem persists, report this error to Hologic Technical
Technical Support.
2. An application attem
attempted
pted to issue another move command to the apertu
aperture
re shuttle
while it was still moving.
Action: The aperture may not be calibrated and an application attempted to move it
to a calibrated position.
position. The aperture shuttle on the C-arm may be jajammed
mmed and the
driver is still attempting to position it in response to a previous MOVE_ABS
command. Visual
Visually
ly inspect the aperture shuttle and use the MOVE_REL command
to verify that the shuttle is moving. If it is not, fix the mechanical or ele
electrical
ctrical
problem with the aperture shuttle and then recalibrate the aperture.
A-6
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Moto
Motor_T
r_TZ
Z Comm
Commen ent:
t: The
The T
TZ
Z sy
syst
stem
em is in se
serv
rvic
icee m
mode
ode;; the
the pede
pedest
stal
alss aare
re mo
movi
ving
ng and
and aan
n
application attempts
attempts to move them in the opposite direction; or the TZ system is
performing a calibration.
Action: Verify
Verify that the TZ board
bo ard service mode switch is in the off position.
po sition.
Device has not been configured [E_NOT_CONFIGURED_I] Comment: Invalid motor calibration data has been sent to the driver (e.g., one or more
of the motor calibration parameters–number of steps, step distance, number of ticks or
tick distance–is less than or zero).
Action: Verify
Verify that the system has been properly calibrated and the normal
nor mal Hologic
system startup procedure has been followed (i.e., reboot the machine and if the
problem persists, recalibrate the motor).
Action: None.
ST
STA
ATE Comm
Commen
ent:
t: An ap
appl
plic
icat
atio
ion
n att
attem
empt
pted
ed to exec
execut
utee a st
stat
atee mac
machi
hine
ne bu
butt tthe
he st
stat
atee mac
machi
hine
ne is
not ready to execute because one or
o r more of the necessary data structures has not been
downloaded (program or task table).
A-7
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
C-Ar
C-Arm
m Comm
Commen
entt: A
Att
ttem
empt
pt to turn
turn on X-
X-ra
rays
ys wh
when
en ei
eith
ther
er th
thee filt
filter
er drum
drum is not
not ye
yett lo
lock
cked
ed (if
(if it
is on) or the initial 30-second X-ray delay after startup has not elapsed. Reported as the
device state if the C-arm is reset. Reported as command error for any IS_xxx command
executed from an application (as opposed to executing from a state machine).
Action: Wait
Wait until the filter drum synchronizes to the A/C line and until the high
voltage relay in the X-ray controller is turned on (30 seconds after the system starts).
Moto
Motor_x
r_xx
x Repo
Reportrted
ed as co
comm
mman
and
d err
error
or for
for aany
ny IS
IS_xx
_xxx
x com
comma
mand
nd exec
execut
uted
ed from
from an appl
applic
icat
atio
ion
n (as
(as
opposed to executing from a state machine). Reported as device state if the motor is
reset.
Action: Determine why the motor was reset (e.g., power line brown out).
Moto
Motor_T
r_TZ
Z Repor
Reporte
ted
d as co
comm
mman
andd err
error
or ffor
or aany
ny m
move
ove comm
comman
and
d unti
untill va
vali
lid
d ccal
alib
ibra
rati
tion
on inf
infor
orma
mati
tion
on
has been downloaded. Reported as command error for any IS_xxx comm command
and executed
from an application (as opposed to executing from a state machine).
Action: Verify
Verify that the TZ driver has been calibrated and that the correct Hologic
system startup procedure has been followed (i.e., reboot the system and if the problem
persists, recalibrate the TZ motor)
Device time-out waiting for requested action [E_TIME_OUT_I] Comment: For everything except the state machine, indicates that a message was sent
to the device and a response was not received within 150ms.
Action: Follow the procedure under "BAD CHECKSUM IN MESSAGE." for isolating
the failed device and replace or repair it.
Comment: For the state machine, indicates that one of the self-limiting commands has
timed out or that the state machine continued executing into the next cycle.
A-8
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Driver is off line [E_OFF_LINE_I] Comment: Returned by any driver whenever a command is written to the driver and the
system is OFF_LINE.
Action: Verify
Verify that no other software, TSR or operating system is running while you do
a scan (e.g., do not run a scan with while connected to a network or attempt to run a
scan under Windows). This may also be due to a communications error between the AT AT
and the C-arm. Follow the procedure under "BAD CHECKSUM IN MESSAGE. MESSAGE."" to
diagnose I/O errors.
Emergency stop active [E_EMERGENCY_STOP_I] Comment: The system emergency stop is active, either because the operator has
pressed the emergency stop button, one of the emergency stop strips has been touched,
or because the C/C has calculated that the C-arm and the table are within 1/10" of each
other.
Action: If the emergency stop button on the operator panel is down (on),
(on ), pull it up.
Verify that nothing is touching the emergency stop strips on the table. If the C-arm and
table are within 1/10", the problem is with the application (or operator) that moved
them so close. If they are not, see the discussion about
about
E_COLLISION_IMMINENT_I. If both E_COLLISION_IMMINE
E_COLLISION_IMMINENT_I NT_I and
E_EMERGENCY_STOP_I are active, the only way to recover is to manually push the
table and C-arm away from each other and then follow the recovery procedure under
E_COLLISION_IMMINENT_I.
End of file Action: This could occur at the start of a lateral scan if the currently selected scan is not
an analyzed companion AP scan.
A-9
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Error closing optical file ________________ Hardware There is a problem writing an optical file. Either the medium is full, or there is a
hardware malfunction.
Error flushing optical file _______________
Action: Check for disk full. Go to DOS, type DIR D: and check for free space.
Error writing optical record ____________ If the disk is not full, call Hologic Technical
Technical Support.
Error copying file _________ Hardware There was an error copying the named file.
Action: Verify
Verify that there was space on the target
target drive. Check the "Room for .. "
message in the status window.
window. Run appropriate diagnostics.
Hardware There is a problem creating a new file (which may be on your hard drive, your diskette,
or or your optical). Either the medium is full, or there is a ha
hardware
rdware error.
error.
Software
Action: First determine which disk drive is involved. The filename should start with a
drive letter,
letter, followed by a colon. Drive A: is diskette, C: is hard drive, and D: is the
optical.
Error initializing graphics module. Action: This can be caused by not enough available memory or an incorrect graphics
adapter. Verify that only Hologic software is installed on the system and that the
graphics adapter has not been modified.
Error in region of interest structure Software Action: This message should be reported to Hologic Technical Support. I nc
nclude as
much information as possible about what you were doing at the time.
Error opening the flattening input file Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Error reading flattening configuration record
Error reading flattening factor record Action: Most likely cause is no flattening done for given scan mode. Re-run BIGFLAT.
BIGFLAT.
A-10
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Error reading flattening records Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Error reading Optical drive: ____________ Hardware There was a disk problem while reading a file.
Error reading
reading file _______ : _
__________
_________ Action: Run appropriate diagnostics.
Error reading reference attenuation record Action: Flattening may not have been done for this scan mode. Re-run BIGFLAT
BIGFLAT.
Failure at file ________ line ______ Software These messages should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
file extension doesn't start with 'P' in ...
flattening records are not initialized Action: Disk error or missing software. Check operation of hard disk, and check
consistency of system software with CHEKPART.
Garbled Optical File Software The optical file that you are attempting to restore scans from seems to be clobbered.
GetDKernel Internal Software Error Software These messages should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
GetKernel Internal Software Error
A-11
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Histogram Overflow in datahist Software These messages should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Histogram Smoothing Error
Internal Buffer Size Exceeded Software This message should be reported to Hologic Technical Support. In
Include as much
information as possible about what you were doing at the time.
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Internal driver error [E_INTERNAL_ERROR_I] Comment: If returned in response to an ON_LINE command, indicates that the MAPI
Comment: If reported as the device status for the C-arm, it indicates an unrecognized
response from the C-arm micro.
Comment: If reported as the abort code for the state machine, it indicates that the
internal driver event queue was full when the state machine attempted to issue the
system-wide abort code for state machine termination.
Internal Error:....
Error:...... Software Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Internal Software Error: ......
Internal stack error [E_STACK_ERROR_I] Comment: Returned by the state machine driver to indicate that the local stack for a
task has overflowed or underflowed.
Invalid ... Software In general, any message that begins with the word "Invalid" should be reported to
Hologic Technical
Technical Support. Include as much information as possible about what you
were doing at the time. See the following "Invalid..." errors for more information.
A-13
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Invalid flattening data [E_INVALID_FLAT_DATA_S] Comment: The most likely cause is that the signal for one or more of
o f the detectors is
higher than allowed.
Action: One or more detectors is bad. See instructions for running hardware checking
program.
Invalid identification file. Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
A-14
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Stat
Statee Comm
Commenent:
t: Retu
Return
rned
ed as the
the dev
devic
icee ssta
tate
te if the C-ar
C-arm
m mic
micro
ropro
proce
cess
ssor
or re
retu
turn
rned
ed an
unrecognized response to the fast query.
Comment: For the state machine driver, this error indicates that either the protocol
attempted to execute a command that is not supported; that an instruction that requires
a register reference did not provide one; or, more commonly
commonly,, an attempt was made to
access the external data buffers and they were either not available or were still being
using by the application.
Action: If you are running any program other than standard Hologic software, remove
it and retry the scan (e.g.,
( e.g., do not run a scan while connected to a network or attempt to
run a scan under Windows). Otherwise, report this error to Hologic T Technical
echnical Support.
Mot
Motor_
or_xx
xx Comment
Comment:: Fo
Forr a sstep
tepper
per motor
motor,, tthe
he appl
applica
icati
tion
on atte
attempt
mpted
ed to m
move
ove a moto
motorr in the
opposite direction in which it is currently moving (i.e., a motor's motion can be
extended in the same direction as the current motion but it cannot reverse motion).
Invalid parameter(s) [E_INVALID_PARAM_I] Comment: The value of the parameter associated with a device command was out of
range.
Invalid serial number in ARRC.TXT. Action: This is not an error if the machine is not yet calibrated.
Invalid serial number in environ.bat file. Action: Machine serial number is not valid. Call Hologic Technical Support.
Invalid values in bad detector map Action: See instructions for running hardware checking program.
A-15
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
I/O Error .... Hardware There was an error reading or writing a disk drive. The message should identify the
or drive and say something about the nature of the error.
err or.
Software
Action: If this is an "out of space" error, verify that there is room on the target drive.
Check the "Room for ..." message in the status window.
window. Otherwise, run appropriate
diagnostics on the disk and controller; if the problem involves diskette (Drive A:), try a
different diskette.
Limit Exceeded: ...... Software Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Action: When moving the system using the operator control panel, wait until all
motion stops before starting an application that requires motor movement (e.g., the
scan program).
Machine serial number is not valid. Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type in configuration files is inconsistent. Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type is not valid. Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Memory allocation error. Action: Verify that only Hologic software is installed on the system.
Message packet canceled [E_CANCELED_I] Comment: The message packet was canceled
canceled before the response was received. This is
not an error but merely records the current state of a message packet.
packet. The driver
normally cancels all outstanding message packets when it goes off line.
Action: None.
A-16
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Action: Because the driver must communicate with the C-arm every A/C line cycle,
and because all other devices share the communications link, any device that causes an
I/O error (e.g., a timeout) generally causes this error in the C-arm. The first thing to do
is to determine if the problem is really with the C-arm or with another device in the
system. Place all devices in emulati
emulation
on mode (e.g., by turning off instrument power and
booting up in the workstation
workstation mode). Then, first take tthe
he C-arm out of emulation.
emulation. If it
reports no errors, then proceed to take each one ofo f the other devices out of emulation
mode until the one in
in error is found. Also see the dis
discussion
cussion under "BAD
CHECKSUM IN MESSAGE." for diagnosing I/O errors.
Message packet sent and waiting for response [E_SENT_I] Comment: The message has been sent but the response has not yet been received.
Reported as the C-arm device status if the fast query response for a cycle is not
received before the next cycle.
Missing tissue bar initialization file Software The file that contains the calibration information for the tissue bar is missing or has not
been installed.
Action: Install the tissue bar initialization software that comes with the tissue bar.
A-17
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Motion reversal is primarily due to the stepper motor cutting out because of a thermal
overload and usually occurs on the AY motor. In most cases, a mechanical problem
(loose belt, loose encoder coupling, etc.) causes this problem. In some cases, a bad
encoder and in fewer cases, a bad drive board causes this problem.
Action: Wait
Wait a few minutes for the motor to cool down and re-try the scan.
Check the drive belt tension. Check the encoder coupling setscrew.
setscrew. If the problem is
not corrected, replace the drive board. If these steps fail, replace the encoder motor.
The procedure for these operations can be found in the Remove and ReplaceReplace section of
this manual. If the problem persists, contact Hologic Technic
Technicalal Support.
Motor is at a limit [E_LIMIT_POSITION_I] Comment: An application tried to move the motor beyond the limit position (i.e., the
motor need not necessarily be at a limit).
Action: Be sure that the table and C-arm are at the correct starting position before
starting a scan. Verify that the range of motion (positive limit position) of the motors
(AR, AY, TX, TY and TZ) are correct for the QDR model. Also see the discussion
under E_COLLISION_IMMINENT_I.
Action: To
To proceed, select emergency Motion on the Utility menu. Then push the table
away from the arm. To restart the system, select retry by pressing the <ENTER> key.
No A/C line interrupts! [E_WORKSTATION_BOOT_I] Action: Check that instrument power is on. When the problem is corrected, press
<Enter> to continue. If you want to leave instrument power off, press <ESC> to boot in
the workstation mode. In this mode, you can not do any scans even if you turn
instrument power back on. This massage can occur if the Emergency Stop sswitch
witch is
pushed in.
Refer to the “NO A/C Line Interrupts” message at start-up on page 6-9.
6-9 .
A-18
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Hardware The QDR is equipped with motion detectors which detect actual motion in the X, Y
and C-arm rotation motors. This message occurs when one of these motors was
commanded to move and no motion was detected.
Action: Check for loose belt couplers, tighten set screws A/R.
Check for coupler hitting bearing block, move coupler A/R.
Check motion encoder, replace if defective.
Check Motor & Drive board, replace if defective.
Check stepper translator, replace if defective.
MOTOR$
MOTOR$xx
xx Whenev
Whenever
er a mo
motor
tor iiss mov
moving,
ing, the driv
driver
er ch
check
eckss its p
posi
ositi
tion
on every
every three
three sseco
econds
nds an
and
d
stops stepping if the incremental positions at these check intervals do not match the
motion parameters (final rate) to within approximately 50%.
Action: Determine why the motor is not moving (check motor coupling, drive belt, and
encoder shaft coupling).
MOTOR$TZ
MOTOR$TZ The devic
devicee st
state
ate reported
reported wh
wheneve
eneverr the TZ firmw
firmware
are repor
reports
ts a pedestal
pedestal ttimeou
imeoutt during
during
a move command.
No data for attenuation curves Software This message should be reported to Hologic Technical Support. In
Include as much
information as possible about what you were doing at the time.
Not enough memory Software Ensure that only Hologic software is installed on the computer. If so, this message
should be reported to Hologic Tec
Technical
hnical Support. Include as much information as
possible about what you were doing at the time.
Old Whole Body Context Type Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Out of memory Software Ensure that only Hologic software is installed on the computer. If so, this message
should be reported to Hologic Tec
Technical
hnical Support. Include as much information as
possible about what you were doing at the time.
A-19
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
Overflow calculating flattening or daily factors Action: Most likely cause is saturation of one or more detectors during the flattening
procedure.
Parameter size is invalid [E_INVALID_SIZE_I] Comment: The device state reported by the C-arm if the response to a fast query is not
a minimum length to encompass all necessary information.
Patient File Record __ Too Small Hardware These messages indicate problems performing I/O operations (usually to the hard
or drive, but possibly to diskette or optical).
Patient File Record __ Too Large Software
Action: Ensure that there is adequate space on the hard drive (Check the "Room for ..."
message in the status
status window.
window. You should always leave
leave room for 1 or 2 scans). There
may be a disk problem; appropriate
appropriate diagnostics should be run. If you are unable to find
a hardware problem, the message should be reported to Hologic Tec Technical
hnical Support.
Include as much information as possible about what you were doing at the time.
PGLINE: ... Software Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
A-20
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
ROI Limit Error Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Scan File Record __ Too Small (__) Hardware These messages indicate problems performing I/O operations (usually to the hard
drive, but possibly to diskette or optical).
Scan File Record __ Too Large (__)
Action: Ensure that there is adequate space on the hard drive (Check the "Room for ..."
Scan has 0 points or lines message in the status
status window.
window. You should always leave
leave room for 1 or 2 scans). There
may be a disk problem; appropriate
appropriate diagnostics should be run. If you are unable to find
a hardware problem, the message should be reported to Hologic Tec Technical
hnical Support.
Include as much information as possible about what you were doing at the time.
Software Error ....... Software Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Starting phase Out of Range Software This message should be reported to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Test failed - absolute max or min limit exceeded Action: One or more detectors is bad. See instructions for running hardware checking
program.
The analyze data file either did not exist or did not contain any valid entries
Software The software was not correctly installed (No ANALYZE.DAT file). This might also be
due to a hardware problem with the hard disk drive.
A-21
HARDWARE
ERROR MESSAGE /SOFTWARE LOCUS COMMENT/ACTION
ERROR
User When restoring files from optical ("Optical" on main menu), you selected an archive
or that contains no scans.
Software
Action: Report this problem to Hologic Technic
Technical
al Support.
Software The reference curve database appears to be totally empty. This condition should never
occur in normal operations.
Action: The Hologic software may not have been installed correctly.
Hardware The X-ray Controller assembly generates an interrupt every 1/120 of a second (60HZ
line) or every 1/100 of a second (50Hz line). The PC clock generates 18 interrupts a
second (regardless of line frequency). If two clock ticks occur wit
with
h NO Motor & Drive
board interrupts between them, then this message is produced.
Action: Report this error to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.
Refer to the “NO A/C Line Interrupts” message at start-up on page 6-9.
6-9 .
Truncated Patient Record ... Hardware This message indicate problems performing I/O operations (usually to the hard drive,
or but possibly to diskette or optical).
Software
Action: Ensure that there is adequate space on the hard drive (Check the "Room for ..."
message in the status
status window.
window. You should always leave
leave room for 1 or 2 scans). There
may be a disk problem; appropriate
appropriate diagnostics should be run. If you are unable to find
a hardware problem, the message should be reported to Hologic Tec Technical
hnical Support.
Include as much information as possible about what you were doing at the time.