MA H250-M40 Krone Flowmeter
MA H250-M40 Krone Flowmeter
MA H250-M40 Krone Flowmeter
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
Copyright 2022 by
KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
1 Safety instructions 6
2 Device description 13
3 Installation 21
4 Electrical connections 26
5 Start-up 40
6 Operation 41
7 Service 64
8 Technical data 78
INFORMATION!
This device is a Group 1, Class A device as specified within CISPR11. It is intended for use in
industrial environment. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated disturbances.
INFORMATION!
Electromagnetic fields with field strengths greater than 10 V/m at the place of installation can
affect the measurement accuracy.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
The variable area flowmeters are suitable for measuring clean gases, vapours and liquids.
Intended use
• The product may not contain any ferromagnetic particles or solids. It may be necessary to
install magnetic filters or mechanical filters.
• The product must be sufficiently liquid and free of deposits.
• Avoid pressure surges and pulsing flows.
• Open valves slowly. Do not use solenoid valves.
• For accurate flow measurement, the application data should be consistent with the sizing
data and calibration of the variable area flowmeter.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
CAUTION!
Do not use any abrasive media containing solid particles or highly viscous media.
1.3 Conformity
The manufacturer certifies successful testing of the product by applying the conformity mark on
the device.
This device fulfills the statutory requirements of the relevant directives and
regulations.
For further information on directives, regulations, standards and certifications, please refer to
the declaration of conformity which you can download from the manufacturer's website.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
CAUTION!
For devices used in SIL applications, additional safety notes apply.
For detailed information refer to the "Safety manual".
Example: PED/G1/III/H
G Gases and vapours
1 Fluid group 1
III Category III
H Conformity assessment method according to Module H
Table 1-2: Example of PED key
The PED key identification can be found on the nameplate of the device (for details refer to
Nameplate on page 19).
INFORMATION!
The stated pressures (PS) and temperatures (TS) only apply as refers to the pressure resistance
of the sensor body. As regards the functionality of the entire device, further restrictions of the
maximum temperature may need to be observed (e.g. ATEX approval). Devices rated below
category I due to their size, do not receive the conformity mark in the scope of the PED. These
devices are subject to applicable sound engineering practice (SEP).
Residual risk
A risk analysis in accordance with the pressure equipment directive has been carried out for the
devices. The residual risk is described as follows:
• The devices are designed according to the valid and applicable rules and standards for static
operation and their pressure resistance is calculated for the declared maximum pressure
and temperature (no calculation for cyclical change).
• Responsibility for the use of the measuring devices with regard to corrosion resistance of the
used materials against the measured fluid lies solely with the operator.
• Avoid abrasion.
• Avoid pulsation and cavitation.
• Protect devices from vibration and high-frequency oscillation.
• Draining (backflow) may be delayed due to the float in the measuring tube.
• Implement appropriate measures to counteract external fire hazards
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.5.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer cannot accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of icons as shown below.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
i RESULT
This symbol refers to all important consequences of the previous actions.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
2. H250/RR/M40
• Additional LCD, measured value and/or flow counter
• 2 configurable binary outputs, limit value or pulse output
• 1 binary input, Start / Stop / Reset flow counter
• 2-wire current output 4…20 mA, HART® communication
• Intrinsically safe (Ex i) or in explosion-proof enclosure (Ex d)
Optional versions:
• H250 with indicator M40 as high temperature version HT
• H250H for use in horizontal pipelines
• H250U for use in vertical fall pipes
• H250F with hygienic measuring tube design for Food & Pharma
• H250C with PTFE / TFM liner for aggressive media
Indicator options
• M40 - Aluminium, two-layer powder coating (epoxy / polyester)
• M40R - Stainless steel without coating
For more details refer to the supplementary instructions "H250 M40 Foundation Fieldbus" or
"H250 M40 Profibus PA".
Use of damping
• Generally when CIV and DIV floats are used for gas measurement.
• For devices with a nominal size of DN100 / 4".
• For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
The magnets on the eddy current brake surround the pointer vane without touching it, damping
its movement. The result is a much steadier pointer position, without distorting the measured
value. The eddy current brake can be retrofitted during operation without recalibration.
Note the maximum tightening torque (0.12 Nm) for the turnbuckle!
2.3 Nameplate
INFORMATION!
Check on the device nameplates, that the device is supplied according to your order.
* positions which are not needed are omitted (no blank positions)
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry, dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the measuring device in the original packaging.
• The permissible storage temperatures for standard devices are: -40...+80°C / -40...+176°F
3.3 Transport
If possible, transport the device to the installation point in the original packaging.
• The variable area flowmeter H250 must be installed vertically (within 5° of the vertical).
H250Hs are installed horizontally (within 5° of the horizontal).
H250U devices are installed upright with the flow direction from top to bottom (within 5° of
the vertical).
• A straight unimpeded inlet run of ≥ 5 DN upstream of the device and a straight outlet run of
≥ 3 DN downstream of the device are recommended. Especially for devices equal or bigger
than DN50 / 2" in order to minimise wear by turbulent float movements.
• Screws, bolts and gaskets are to be provided by the customer and must be selected in
accordance with the pressure rating of the connection or the operating pressure.
• The inner diameter of the flange deviates from the standard dimensions. Flange seal
standard DIN 2690 or ASME B16.21 can be applied.
• Align the gaskets. Tighten nuts with the tightening torques of the appropriate pressure
rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, refer to chapter
"Tightening torques".
• Control devices are to be positioned downstream of the measuring device.
• Shutoff devices are preferably to be positioned upstream of the measuring device.
• Before connecting, blow or flush out the pipes leading to the device.
• Piping for gas flow need to be dried before the device is installed.
• Use connectors suitable for the particular device version.
• Align the piping centrically with the process connections on the measuring device so they are
free of stresses.
• If necessary, the piping has to be supported to reduce the vibrations transmitted to the
measuring device.
• Do not lay signal cables directly next to cables for the power supply.
• When mounting several measuring devices next to each other or next to ferromagnetic
system components or solenoid valves, a minimum lateral distance must be ensured.
• The device must not be heated by radiated heat (e.g. exposure to the sun) to an electronics
housing surface temperature above the maximum permissible ambient temperature. If it is
necessary to prevent damage from heat sources, a heat protection (e.g. sun shade) has to be
installed.
Minimum distances
When mounting several measuring devices next to each other or next to ferromagnetic system
components or solenoid valves, a minimum lateral distance a > 300 mm / 11.8" must be ensured.
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for
horizontal installation are to be installed in the pipeline so that the indicator is located on the
side of the measuring tube. The maximum product and ambient temperatures indicated as well
as the measuring accuracy are based on lateral installation of the indicator.
CAUTION!
The heat insulation ① may only reach to the rear of the housing ② . The area around the cable
entries ③ must be freely accessible.
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations.
Any work done on the electrical components of the measuring device may only be carried out by
properly trained specialists.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
The connection terminals have a pluggable design and can be removed to connect the cables.
The built-in limit switch types are shown on the nameplate of the indicator.
Terminal number 1 2 3 4 5 6
Connection 2-wire NAMUR - + - +
Connection 3-wire + - + -
Limit setting
CAUTION!
If the maximum torque (0.2 Nm) is exceeded, the locking screw can be torn off during tightening!
Power supply
INFORMATION!
The supply voltage has to be between 14 VDC and 32 VDC. This is based on the total resistance of
the measuring loop. To calculate this, the resistance of each component in the measuring loop
(not including the device) must be added up.
The required supply voltage can be calculated using the following formula:
Uext. = RL * 22 mA + 14 V
with
Uext. = minimum supply voltage
RL = total measuring loop resistance
INFORMATION!
The power supply has to be able to supply a minimum of 22 mA.
When HART® communication is carried out with the ESK4A, the analogue measured data
transmission (4...20 mA) is not impaired in any way.
Exception for multidrop operation. In multidrop operation a maximum of 15 devices with HART®
function can be operated in parallel, whereby their current outputs are switched to inactive
(I approx. 4.5 mA per device).
RL = (Uext. - 14 V) / 22 mA
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary, e.g if noise levels higher than the NE 21
specification are anticipated.
Configuration
The ESK4A can be configured via HART® communication. DD (Device Description) for AMS and
PDM as well as a DTM (Device Type Manager) for PACTwareTM are available for the
configuration. They can be downloaded free of charge from the website of the manufacturer.
The current flow rate can be transmitted using the integrated HART® communication. A flow
counter can be configured. Two limit values can be monitored. The limit values are assigned
either to flow values or to the counter overflow.
The binary inputs/outputs are electrically isolated from each other and from the ESK4A current
output.
INFORMATION!
The binary inputs/outputs can only be operated if the power supply is applied to ESK4A terminal
11+ and 12-. The binary inputs/outputs are inactive by default and must thus be activated prior to
first use (for details refer to ESK4-T menu on page 49).
Binary output B1 B2
Terminal no. 1 2 3 4 5 6
Connection NAMUR + - + -
Connection transistor output + BOC + BOC
Table 4-3: Terminal assignment of the binary output
Value range applies only when connected to a switching amplifier with the following reference
values:
• Open-circuit voltage U0 = 8.2 VDC
• Internal resistance Ri = 1 kΩ
To ensure the value ranges, a load RL between 250 Ω and 1 kΩ is recommended for the passive transistor
output with a nominal voltage of 24 VDC.
If other loads are used, caution is advised as the range of values of the signal voltages then no longer
corresponds to the range of values for the inputs of process control systems and controls (IEC 60946).
CAUTION!
The upper limit of the signal current must not be exceeded as this may damage the transistor
output.
= + I+
I
-
= + B+
- Boc
The pulse output B2 (B1) is a passive "Open Collector" output which is electrically isolated from
the current output and output B1 (B2). It can be operated as a low-resistance output (refer to
previous figure) or as a NAMUR output (refer to figure "Connection switching output").
The pulse width ton can be configured from 50...500 ms in the indicator menu.
Binary input B3
Terminal no. 7 8
Connection + -
Table 4-7: Value range for NAMUR
This binary input can be activated in the menu of the indicator and can be configured to ACTIVE
HI or ACTIVE LO.
If the input is set as ACTIVE LO, an interruption causes the counter to be reset.
For further information on configuration of each function refer to ESK4-T menu on page 49.
ESK4-FF / ESK4-PA
• 2-wire, bus-supplied
• Polarity protected
• Bus voltage 9...32 VDC
• Nominal current 16 mA
① Pin number HAN® 7D K1/K2 NAMUR R1/R2 Reed ESK4A Terminal number
contacts contacts
NAMUR Reed
1 ② NAMUR MIN (-) ② Reed MIN - 1 1
2 ② NAMUR MIN (+) ② Reed MIN - 2 3
3 ③ NAMUR MAX (-) ③ Reed MAX - 4 4
4 ③ NAMUR MAX (+) ③ Reed MAX - 5 6
5 - - ④ 4...20mA (+) 11
6 - - ④ 4...20mA (-) 12
7 - - -
8 - - -
Table 4-9: Terminal assignment - Harting plug to terminal number
INFORMATION!
No Harting connection for module ESK4-FF/PA is provided and is only available on request.
DANGER!
The grounding cable should not transmit any interference voltages.
Do not use this grounding cable to ground any other electrical devices.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class
must be ensured again.
Start
After the device is switched on, the display shows
• "INITIALISING"
• Firmware version IO module
The device first performs a self-test. Here, all of the parameters preset for the customer are
analysed and checked for plausibility. The device then switches to measuring mode and indicates
the current measured value.
Operation
INFORMATION!
The device is low-maintenance
Comply with the application limits with regard to temperature of the medium and ambient
temperature.
CAUTION!
When activating the loop check mode, ensure that no alarms are unintentionally triggered in the
downstream system components.
• Press and hold the microswitch ① for more than 6 seconds to activate the loop check mode ③ .
The current output jumps to constant 4 mA.
• Change the current output from constant 4 mA to constant 20 mA as often as you like by briefly pressing
it (less than 6 seconds) to check the function of the measuring circuit.
• Exit loop check mode by holding down the microswitch (longer than 6 seconds). The current output
jumps back to measuring mode ② .
INFORMATION!
If the microswitch has not been pressed for longer than 60 seconds, the ESK4A automatically
returns to measuring mode ② .
CAUTION!
The switching point of the magnetic sensors is directly at the height of the corresponding circle
(see figure). Only touch the circle vertically and from the front using the bar magnet. Touching it
from the side may cause an error in the measurement as the float position is recorded by the
magnetic field sensors.
The mechanical keys and keys for the bar magnet have the same functionality. In this
documentation the keys are represented as symbols to describe the operating functions:
Key Symbol
① Enter ↲
② up ↑
③ right →
Table 6-1: ESK4-T operation keys
If you are already at the lowest level (function level), the → key takes you to the change mode,
where you can set data and values.
If you are at the first level (main menu), you can use the ↲ key to exit the menu mode and return
to the measuring mode.
If settings were changed, the query whether they should be saved will appear. Confirm this query
with the → key or cancel it with the ↲ key.
If an operation lock is set, the code ( → → → ↲ ↲ ↲ ↑ ↑ ↑ ) must be entered. The code can be
set in menu 3.15. The code shown here is set at the factory but not activated. If no key is activated
within 5 seconds or an incorrect code is entered, a warning message is displayed and the
indication returns to measuring mode.
Exit operation
Operation is exited by pressing the ↲ key several times.
Save Yes → Changes are accepted. An update is carried out and the display returns to
measuring mode.
Save No ↲ Changes are discarded and the display returns to measuring mode.
Table 6-4: Change the settings in the menu
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal
plausibility check.
If inplausible entries were made, a warning message is shown. If this warning is confirmed with
the ↲ key, the display returns to the respective menu item without saving the relevant change.
A new entry can now be made.
Display Display
Example: 7.2 1x → Fct. 3.11.1
m3/h MEAS’D VALUE
1x → Fct. 1 1x → 10.0000
OPERATION m3/h
2x ↑ Fct. 3 4x ↑ 10000
INSTALLATION l/h
1x → Fct 3.1 Confirm with →
LANGUAGE deny ↲
10x ↑ Fct 3.11 3x ↲ 7200
F.S.& UNIT l/h
Table 6-5: Example: Changing the flow unit from m3/h to l/h
In addition to the predefined units shown here, a user-defined unit can be activated in Menu 3.12
by entering a conversion factor and a unit designation as free text.
Function Setting
1.1.1 OUTPUT B1 INACTIVE
1.2.1 OUTPUT B2 INACTIVE
1.3.1 PULSE WIDTH 50ms
1.3.2 PULSE/UNIT 1 Pulse/Unit
1.4 DISPLAY MEAS’D VALUE
1.4.2 ROTATION 0°
1.5 TIME CONST 1.0s
1.6.1 COUNTER NO
1.6.2 ERROR NO
1.6.3 RE-INIT IO NO
3.1 LANGUAGE ENGLISH
3.2 FUNCTION B1 INACTIVE
3.3 CONTACT B1 NORM.OPEN
3.4 FUNCTION B2 INACTIVE
3.5 CONTACT B2 NORM.OPEN
3.6 FUNCTION B3 INACTIVE
3.7 MULTIDROP POLLING ADD: 00
3.8 4mA CALIBR. 4.000mA
3.9 20mA CALIBR. 20.000mA
3.10 ALARM CURR. ALARM HIGH
3.11 F.S.& UNIT Application specific
3.11.2 COUNTER Application specific
3.12 USER DEFINED User defined unit / factor
3.13 L.FL.CUTOFF 4% ON; 6% OFF
3.14 DESCRIPTOR Free text
3.15 ENTRY CODE OFF
3.16 DEFAULT VAL. NO
5.1 DIS/ENABLE -
5.1.1 BLOCKED FL. OFF
5.1.2 EXT. MAG. F. OFF
5.1.3 INT. MAG. F. OFF
5.1.4 PUL. FLOW OFF
5.2 AUTO CAL. -
5.2.1 BLOCKED FL. OFF
5.2.2 EXT. MAG. F. OFF
5.2.3 INT. MAG. F. OFF
5.2.4 PUL. FLOW OFF
5.3 MAP EVENT -
5.3.1 BLOCKED FL. Failure
Function Setting
5.3.2 MAG. FIELD Maintenance required
5.3.3 PUL. FLOW Out of specification
5.4 BLOCKED FL. -
5.4.1 PERIOD 1800 s
5.4.2 MIN. FLOW 50 %
5.4.3 THRESHOLD 0.075 %
5.5 EXT. MAG. F. -
5.5.1 HOLD TIME 4.0 s
5.5.2 MIN MAG 26.0 %
5.5.3 MAX MAG 74.0 %
5.6 INT. MAG. F -
5.6.1 HOLD TIME 4.0 s
5.6.2 MIN MAG 70.5 %
5.6.3 MAX MAG 90.0 %
5.7 PUL. FLOW -
5.7.1 HOLD TIME 4.0 s
5.7.2 PERIOD 0.2 s
5.7.3 THRESHOLD 30.0 %
Table 6-10: Factory settings
1.1.1
INACTIVE -
MEAS.VAL. B1 Switching point of flow value
Value range: 0.0…full scale range
The switching point is entered in flow units.
If the current flow value exceeds this pre-determined switching point, output B1 changes
its binary state ③ .
A configuration takes place in Fct. 1.3.1 PULSE WIDTH, 1.3.2 PULSE/UNIT and 3.11.2
COUNTER.
1.1.2 HYST. B1 Hysteresis for the switching point of the flow value
Value range: 0.0…switching point
If the current flow value exceeds the pre-determined switching point from Fct. 1.1.1,
output B1 changes its binary state ③ .
In order for output B1 to change its binary state back to the initial setting, the switching
point made smaller by the hysteresis must be undershot ④ .
Example:
In 1.1.1, a switching point of 200 L/h is set.
The possible value range for the hysteresis is then 0.0…200 L/h.
With a hysteresis value of 0, the switching point has no hysteresis ( ③ = ④ ).
If a hysteresis value of 20 L/h is entered, output B1 changes its binary state to the initial
setting, if it drops below 180 L/h ④ .
1.2.1
INACTIVE -
MEAS.VAL. B2 Refer to MEAS.VAL. B1
In Fct. 1.2.2 a hysteresis can be specified additionally.
CNT. VAL. B2 Refer to CNT. VAL. B1
PULSE WIDTH B2 Refer to PULSE WIDTH B1
A configuration takes place in Fct. 1.3.1 PULSE WIDTH, 1.3.2 PULSE/UNIT and 3.11.2
COUNTER.
1.2.2 HYST. B2 Refer to HYST. B1
1.3 PULSE OUTP. -
1.3.1 PULSE WIDTH
50ms Ti = 50 ms; fmax = 10 Hz
max. pulses/h = 36000
100ms Ti = 100 ms; fmax = 5 Hz
max. pulses/h = 18000
200ms Ti = 200 ms; fmax = 2.5 Hz
max. pulses/h = 9000
500ms Ti = 500 ms; fmax = 1 Hz
max. pulses/h = 3600
1.3.2 PULSE/UNIT 0.001...1000
Pulse per volume or mass unit of counter (setting in Fct. 3.11.2), which can be output via
one of the binary outputs. The maximum frequency for the pulse output (refer to Fct 1.3.1)
cannot be exceeded, even at maximum flow rate (full scale).
Example:
Final value Qmax = 1200 L/h; volume unit of counter = Litre; pulse width = 100ms;
If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in one hour at maximum
flow rate.
Number of max. permissible pulses:
1.4 DISPLAY Different measured values can be selected for permanent or alternating display.
The reading on the display can be rotated.
1.4.1
MEAS’D VALUE Permanent display of flow in flow units
COUNTER Permanent display of totalizer
MV/COUNTER Alternating display of flow value in flow units and totalizer
MV&COUNTER Simultaneous display of flow value and totalizer
PERCENT Permanent display of flow value in percent
1.4.2 ROTATION
0° The reading is not rotated.
90° The reading on the display is rotated by 90°.
180° The reading on the display is rotated by 180°.
270° The reading on the display is rotated by 270°.
1.5 TIME CONST 0.0...20.0s
Given in seconds
The output variables (value of the current loop and indicated flow value) follow the current
process by the value set here (in seconds) with a time delay.
Note:
If the current flow is polled via HART® communication, the transferred measured value is
also played back with a delay.
1.6 RESET Local resetting of totalizer and acknowledgement of warnings.
There is always a safety prompt ("YES / NO"), to avoid accidental resetting.
Note:
External resetting of the totalizer can be installed with the binary input B3.
1.6.1 COUNTER Confirming with "YES" resets the value of the totalizer to 0.0.
1.6.2 ERROR Confirming with "YES" acknowledges all existing warnings.
Note:
Confirming with "YES" acknowledges existing errors and warnings.
1.6.3 RE-INIT IO As a general rule, data is transferred from the counter module to the ESK4 and vice versa
when the device starts. For safety, another data transmission can be performed by
selecting this menu item and confirming with "YES".
Table 6-12: Menu description - 1 OPERATION
Note:
The test is not available in HART® multi-drop mode (refer to Fct. 3.7)
Attention:
During the test, the value of the current loop does not follow the current process.
2.1.1 NORMAL OP The value of the current loop follows the current process.
2.1.2 4.0mA The value of the current loop no longer follows the current process.
It is set to the selected current value.
2.1.3 5.6mA
2.1.4 7.2mA
2.1.5 8.8mA
2.1.6 10.4mA
2.1.7 12.0mA
2.1.8 13.6mA
2.1.9 15.2mA
2.1.10 16.8mA
2.1.11 18.4mA
2.1.12 20.0mA
2.1.13 21.6mA
2.2 ALARM CURR. <3.6 / >21 mA
Testing the alarm current HIGH/LOW according to the settings in Fct. 3.10.
2.3 OUTPUT B1 Testing the binary switching output B1 by changing its binary status.
Attention:
During the test, the binary status does not correspond to the current process.
2.3.1 NORMAL OP The binary status of the switching output corresponds to the current process.
2.3.2 OPEN The binary status of the switching output no longer corresponds to the current process.
The selected status is tested.
2.3.3 CLOSED
2.4 OUTPUT B2 Refer to 2.2 OUTPUT B1
2.4.1 NORMAL OP
2.4.2 OPEN
2.4.3 CLOSED
2.5 INPUT B3 The current binary status of the binary input B3 is displayed. External switching of the
voltage applied to input B3 activates a change of the displayed binary status.
Note:
A reaction to the switching of the voltage applied to input B3 only occurs if the function B3
is activated (refer to Fct. 3.6).
3 INSTALLATION
Selection / Input Explanation
3.1 LANGUAGE Language selection for the displayed menu texts.
3.1.1 ENGLISH The menu texts are displayed in the selected language.
3.1.2 DEUTSCH
3.1.3 FRANCAIS
3.1.4 ITALIANO
3.1.5 ESPANOL
3.1.6 CESKY
3.1.7 POLSKI
3.1.8 NEDERLANDS
3.1.9 DANSK
3.2 FUNCTION B1 -
INACTIVE The binary switching output B1 has no function.
SWITCH POINT The binary switching point B1 acts as a limit switch depending on the current flow value.
Note:
The configuration in Fct. 1.3.1 and Fct. 1.3.2 applies to both pulse outputs. If both B1 and B2
are configured as pulse outputs, both binary outputs behave in exactly the same way.
If an ampere meter detects in the 4...20 mA loop a deviation from the desired value of
"4.000mA", the measured value must be entered. The correction value is accepted
following the query to save and confirm with yes.
Note:
This function is not available in HART® multi-drop mode.
Attention:
During calibration, the value of the current loop does not follow the current process.
If an ampere meter detects in the 4...20 mA loop a deviation from the desired value of
"20.000mA", the measured value must be entered. The correction value is accepted
following the query to save and confirm with yes.
Note:
This function is not available in HART® multi-drop mode.
ALARM LOW Error indication via the current loop is activated (failure signal "low" acc. to NE43).
Note:
1) This function is not available in HART® multi-drop mode.
2) This function is supported from ER 2.2.x.
3.11 F.S. & UNIT By changing the unit, the respective final value is scaled accordingly. Depending on
calibration, either units for volume or mass flow measurement are available for selection.
3.11.1 MEAS'D VALUE For units for volume flow or mass flow, refer to Overview of the units ESK4-T on page 45.
3.11.2 COUNTER For units for volume counter or mass counter and pulse output, refer to Overview of the
units ESK4-T on page 45.
3.12 USER DEFINED User-defined unit with conversion to calibrated unit.
3.13 L.FL.CUTOFF L.FL.CUTOFF stands for low flow cutoff.
To ensure a stable zero point of the current output, the current output can be set to a
stable value of "4.00mA" ③ in a selectable range.
3.13.1
CONTROL OFF LFC function is not active.
CONTROL ON L.FL.CUTOFF function is active.
3.13.2 ON VALUE Switch-on value ① :
Value range 1…19% (of full scale range)
The flow is greater than the switch-on value. The current output corresponds to this.
If the flow rate falls, the current output follows until the switch-on value ① .
If the flow value continues to fall, the current output is switched to "4.00mA" ③ .
Note:
The switch-on value to be set must be smaller than the previously selected switch-off
value.
3.13.3 OFF VALUE Switch-off value ② :
Value range 2…20% (of full scale range)
The flow rate is 0. The current output is "4.00mA" ③ .
If the flow rate increases, the current output remains at "4.00mA" until it has reached the
switch-off value ② .
Note:
The switch-off value to be set must be greater than the previously selected switch-on
value.
5 NE107 DIAG.
Selection / Input Explanation
5.1 DIS/ENABLE The different application diagnostic functions can be switched on/off separately.
5.1.1 BLOCKED FL. Detection of blocked float.
OFF Not active.
ON Active
5.1.2 EXT. MAG. F. Detection of external magnetic field interference (e.g. caused by high DC current).
In order to detect all kind of magnetic field interference it is recommended to switch on
5.1.3 INT MAG-F., too.
OFF Not active
ON Active.
5.1.3 INT. MAG. F. Detection of internal magnetic field interference (e.g. missing float, upside down float).
OFF Not active.
ON Active.
5.1.4 PUL. FLOW Detection of pulsating flow/oscillating float.
OFF Not active.
ON Active.
5.2 AUTO CAL. The factory setting of parameters for the application diagnostics can be optimised by
executing the auto calibration of the respective function. During auto calibration the NE107
application diagnostic is inactive and the status "function check" is activated. The auto
calibration is executed until the function is switched off again. When switching off the new
determined parameters are stored in the non-volatile memory and are used for the
diagnostic functions. Power off/on during auto calibration interrupts the auto calibration
and previous parameters are used.
5.2.1 BLOCKED FL. Auto calibration to determine the threshold parameter for blocked float.
OFF Not active.
ON Active. The auto calibration is executed until switched to "OFF" again. During auto
calibration, the flow should be as constant as possible.
5.2.2 EXT. MAG. F. Auto calibration to determine the magnetic field limits for external field detection.
During auto calibration, the complete flow range from 0 to 100% should be applied.
OFF Not active.
ON Active. The auto calibration is executed until switched to "OFF" again.
5.2.3 INT. MAG. F. Auto calibration to determine the magnetic field limits for internal field detection.
During auto calibration, the complete flow range from 0 to 100% should be applied.
OFF Not active.
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also
be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to
the measuring tube and the indicator.
We advise that inspections are carried out at least once per year.
The device must be removed from the pipeline before cleaning.
CAUTION!
Pressurised pipes have to be depressurised before removing the device.
Empty the pipelines as completely as possible.
In the case of devices used for measuring aggressive or hazardous media, appropriate safety
precautions must be taken with regard to residual liquids in the measuring unit.
New gaskets have to be used when re-installing the device in the pipeline.
Avoid electrostatic charges when cleaning the surfaces (e.g. sight window).
• Float damping
Indicator M40:
• Limit switch module K1 / K2
• Current output 4..20 mA ESK4A
• Counter module with LCD and I/O ESK4-T
• Fieldbus interface ESK4-PA / FF
CAUTION!
Without a recalibration an additional measuring error is to be expected.
For the setting of the limits refer to Connection of limit switches K1/K2 on page 27.
The contact module connecting terminals feature a pluggable design and can be removed in
order to connect the cables.
WARNING!
Do not damage the indicating element!
WARNING!
Do not damage the indicating element!
• The ESK4A plug-in tongues ① are plugged in under the two bolts ② on the base plate.
• Slight pressure is used to press the ESK4A onto the spring pins ③ until it stops, firmly
attaching the ESK4A.
• Device identification
• Device address
• Serial number
• Measuring point identifier
• Digital measured value query in flow units, % and mA
• Test / setting functions
• Calibration 4.00 and 20.00 mA
• Set current output to any desired value
More detailed information can be found in the service manual included with each retrofit or
replacement set.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
INFORMATION!
Other spare parts on request.
INFORMATION!
For more precise information, please contact your local sales office.
WARNING!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
• Due to statutory regulations on environmental protection and safeguarding the health and
safety of the personnel, the manufacturer may only handle, test and repair returned devices
that have been in contact with products without risk to personnel and environment.
• This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
WARNING!
If the device has been operated with toxic, caustic, radioactive, flammable or water-endangering
products, you are kindly requested:
• to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from
such dangerous substances,
• to enclose a certificate with the device confirming that it is safe to handle and stating the
product used.
Company: Address:
Department: Name:
Fax number:
Manufacturer order number or serial number:
We hereby confirm that there is no risk to persons or the environment caused by any residual media contained in this
device when it is returned.
Date: Signature:
Stamp:
7.6 Disposal
LEGAL NOTICE!
Disposal must be carried out in accordance with legislation applicable in your country.
INFORMATION!
• Wear personal protective equipment
• Make sure to use a stable workstation / bench to disassemble
The variable area flowmeter consists of a mechanical, metallic measuring unit ① and an
indicator housing attached to it ② , which may contain electronic components.
The metallic measuring unit can be sent to metal recycling after the indicator housing has been
removed. Depending on the version, the measuring unit may weigh between 2...50 kg /
4.4...110.2 lb.
CAUTION!
Check and ensure that the measuring units, including all cavities, are free of hazardous
substances by rinsing or neutralising if required. This should be done especially in the case of
devices that have been operated with toxic, corrosive, radioactive, flammable or water-
contaminating products to protect both the environment and the health and safety of staff.
The indicator housing may contain optional electronic components which can be seen when the
cover is opened. They may be removed individually from the indicator housing.
INFORMATION!
Silicone is used when potting the PCBs in the polyamide housings. It is thus not easy to separate
the components by hand.
① For the indicator, the flow-dependent height of the float in the measuring unit is transmitted
by means of a magnetic coupling and displayed on a scale.
② For a built-in signal converter (ESK4A), the flow-dependent height of the float in the
measuring unit is detected by the S1 and S2 magnetic field sensors and electronically
processed.
HT High-temperature version
ESK4A 2-wire current output 4…20 mA with HART® 7
ESK4-T ESK4A + ESK4-IO (LCD, binary status outputs, digital counter and pulse output)
ESK4-FF FOUNDATION FIELDBUS interface
ESK4-PA PROFIBUS PA interface
Table 8-1: Used abbreviations
Measuring system
Application range Flow measurement of liquids, gases and vapours
Function / Measuring principle Float measuring principle
Measured value
Primary measured value Float position
Secondary measured value Operating volume flow, standard volume flow or mass flow
Measurement accuracy
Max. permissible error 1.5% of full scale range
H250 /RR /HC /F
1.6% of measured value according to VDI/VDE 3513-2 (qG = 50%)
Max. permissible error 2.0% of full scale range
H250/C (ceramic, PTFE), H250H, H250U,
H250 (100 : 1), H250 Low Flow 2.5% of measured value according to VDI/VDE 3513-2 (qG = 50%)
Measurement repeatability
H250 /RR /HC /F 0.25%
H250H, H250U, H250 (100 : 1) 0.5%
Operating conditions
Temperature
Max. operating temperature TS -196...+300°C / -321...+572°F
Depending on the version (refer to nameplate)
Higher temperatures up to +400°C /+752°F are on request.
Mechanical indicators For detailed information on product and ambient temperatures refer to
Temperatures for mechanical indicators without auxiliary power on page 82.
Devices with electrical components For detailed information on product and ambient temperatures refer to
Temperatures for devices with electrical components on page 83.
Pressure
Max. operating pressure PS, max. test Depending on the version (refer to nameplate)
pressure PT
Min. required operating pressure 2 times greater than pressure loss (refer to measuring ranges)
Pressure tightness (vacuum) H250/C For detailed information refer to Pressure tightness (vacuum) H250/C on
page 85.
Pressure/temperature for the optional heating of the measuring cone
DN15...DN50 PS = 40 barg / 580 psig, TS = 300°C / 572°F
DN80...DN100 PS = 25 barg / 363 psig, TS = 300°C / 572°F
Ingress protection
M40, M40R IP66/68 according to EN 60529, NEMA 4/4X/6 according to NEMA 250
M40R IP69K according to DIN 40050-9
Float damping during gas measurement recommended
DN15...25 / 1/2...1" Operating pressure <0.3 barg / 4.4 psig
DN50...100 / 2...4" Operating pressure <0.2 barg / 2.9 psig
Installation conditions
Inlet section ≥ 5 x DN
Outlet section ≥ 3 x DN
Note: Straight inlet and outlet sections are not required for accuracy
reasons, but may reduce instable flow indication caused by fluctuating flow
profiles and may increase lifetime because of reduced mechanical wear.
It is recommended to consider the inlet / outlet sections especially for
devices ≥ DN50 / 2".
Protect devices from vibration and high- Use in accordance with IEC 61298-3 in control room or field with medium
frequency oscillation vibration level.
Materials
H250/RR
Flange & measuring tube Stainless steel 1.4401 / 1.4404, 316 / 316L (dual certification)
Float & float guide 1.4404 / 316L
Retaining ring 1.4571 / 316 Ti
H250/HC
Flange Hastelloy® C276 / 2.4819, Hastelloy® C4 / 2.4610, solid or plated
Measuring tube, float & float guide Hastelloy® C276 / 2.4819, Hastelloy® C4 / 2.4610
Retaining ring Hastelloy® C2000 / 2.4675
H250/F - Food
Flange & measuring tube Stainless steel 1.4404 / 316L
Float & float guide Stainless steel 1.4404 / 316L
H250/C - Ceramic/PTFE (DN100 / 4" only PTFE)
Flange & measuring tube Stainless steel 1.4571 with TFM/PTFE
TFM/PTFE liner (electrically non-conductive), conductive PTFE on request
Float PTFE or Al2O3 with FFKM gasket
Float guide Al2O3 and PTFE
Ring orifice Al2O3
Indicators
M40 Aluminium, two-layer powder coating (epoxy / polyester)
M40R Stainless steel without coating 1.4408 / CF8M
Offshore Wet coating on request
Cable glands / blanking plugs
Standard Polyamide
Optional Nickel-plated brass or stainless steel
Other options on request
Special materials e.g. SMO 254/6Mo/1.4547, Titanium Grade 2, Hastelloy® C276 / 2.4819,
Hastelloy® C4 / 2.4610, Monel® / 2.4360, Inconel® / 2.4856,...
Float damping PEEK (only for gas) or ceramic
Gaskets Standard for devices with female thread as insert: O-ring FPM / FKM
Options Surface passivation of all metallic, wetted parts e.g. SilcoNert®2000 or
Dursan®, material certificates, NACE MR0175 / MR0103, non-destructive
material tests, pressure/leakage tests, final cleaning
Process connections
Options Flanges, clamp connections, screw connections and thread connections.
For detailed information refer to Process connections on page 85.
[°C] [°F]
ESK4A, ESK4-FF, ESK4-PA 1 -40...+70 -40...+158
Limit switches SJ3,5-SN / I7S23,5-N / Reed SPST -40...+70 -40...+158
Limit switches SC3,5-N0 / SJ3,5-S1N / SB3,5-E2 -25...+70 -13...+158
Table 8-4: Ambient temperature in °C and °F
1 Decreasing display contrast outside of the temperature range of 0...+60°C / +32...+140°F.
INFORMATION!
The device must not be heated by radiated heat (e.g. exposure to the sun) to an electronics
housing surface temperature above the maximum permissible ambient temperature.
A sun shade is available as option.
• Reference point is the connection of the equipotential bonding conductor of the M40 indicator.
• The temperatures at the reference point are to be determined in the most unfavourable
operating situation.
• The measuring unit should be insulated properly.
Nominal size Float Liner [mbara] [psia] [mbara] [psia] [mbara] [psia]
DN15...100 PTFE PTFE 100 1.45 - - - -
DN15...80 Ceramic PTFE 100 1.45 250 3.63 - -
DN15...80 Ceramic TFM / Ceramic 100 1.45 100 1.45 100 1.45
Table 8-8: Pressure tightness (vacuum) H250/C
INFORMATION!
Higher pressure ratings and other connections are available on request.
INFORMATION!
For more details refer to the supplementary instructions "H250 M40 Foundation Fieldbus".
ESK4-PA Profibus PA
Physical Layer IEC 61158-2 and FISCO model
Communication standard Profibus PA profile 3.02
PNO ID 4531 HEX
Power supply Bus supply: 9...32 VDC, intrinsically safe max. 30 VDC
Nominal current 16 mA
Error current 23 mA
Boot-up current after 10 ms < Nominal current
Table 8-13: ESK4-PA
INFORMATION!
For more details refer to the supplementary instructions "H250 M40 Profibus PA".
ESK4-T with LCD, binary inputs and outputs and digital counter
Binary output
Two binary outputs Galvanically isolated, passive
Mode Switching output NAMUR or transistor (OC)
Configurable as Switch contact or pulse output Normally closed / normally open or
max. 10 pulses/s
NAMUR switching output
Power supply 8.2 VDC
Signal current > 3 mA switching value not reached < 1 mA switching value reached
Switching output transistor (Open Collector)
Power supply Nominal 24 VDC, maximum 30 VDC
Pmax 500 mW
Continuous current Max. 100 mA
No-load current I0 ≤ 2 mA
Pulse output
Ton Configurable from 50...500 ms
Toff Depending on flow rate
Pulse weight Configurable in flow units e.g. 5 pulses/m3
Binary input
Input Galvanically isolated
Mode Reset counters or Start / Stop
Configurable as active HI / active LO
H signal 16...30 VDC
Internal resistance Ri Typical 20 kΩ
Ton (active) ≥ 500 ms
LC display
Technology Passive graphic LCD
Display Measurement with units and/or counter reading with units.
Counter reading max. 11-digit with power failure proof storage.
Binary flags for limit value status. 0...100% bargraph for measurement.
NE 107 diagnostic status symbols. Plain text menu for configuration.
Configuration Local plain text menu navigation via microswitch or bar magnet or using
DD/DTM software
Table 8-14: ESK4-T with LCD, binary inputs and outputs and digital counter
8.2.6 Approvals
a b d h
EN ASME c① e② Øf g j②
[mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 1/2" 94 3.70 114 4.49 20 0.80 97 3.82 197 7.76
DN25 1" 94 3.70 127 5.00 32 1.28 109 4.27 209 8.23
DN50 2" 107 4.22 141 5.55 65 2.57 125 4.90 222 8.74
DN80 3" 107 4.22 157 6.18 89 3.51 143 5.61 238 9.37
DN100 4" 107 4.22 167 6.57 114 4.50 150 5.91 248 9.76
Table 8-16: Dimensions in mm and inch
① Without cable gland
② Ex d, Ex t, Ex ec: + 10 mm / 0.39"
ISO 228 / ASME B1.20.1 ISO 228 / ASME B1.20.1 H250/F ① Clamp H250/F screw connection
Female thread screwed Female thread welded connection DIN 11851
Weight
H250 Heating
Measuring span: 10 : 1
Flow values: Values = 100% Water: +20°C / +68°F Air: +20°C / +68°F, 1.013 bara / 14.7 psia
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate.
Other flow ranges on request. Conversion of other media or operating data is performed using
the calculation method in accordance with VDI/VDE directive 3513.
Float ► TIV CIV DIV TIV TIV DIV TIV TIV CIV DIV
Alu. Alu.
Float ► TIV CIV DIV TIV TIV DIV TIV TIV CIV DIV
Alu. Alu.
Liner / Float ► PTFE Ceram. PTFE Ceram. PTFE Ceram. PTFE Ceram.
Liner / Float ► PTFE Ceram. PTFE Ceram. PTFE Ceram. PTFE Ceram.