T16 User Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9
At a glance
Powered by AI
The document outlines proper usage and safety instructions for an indicator device.

The device should be properly grounded and static protection measures should be taken. It should not be used in hazardous environments or exposed to extreme conditions. Proper authorization is required before opening or modifying the device.

The device has modes for normal weighing, peak detection to measure maximum weights, and averaging weights for unstable platforms. It can also perform check weighing functions to check if weights are within set limits.

XK3118T16

indicator
User Manual

May 2011 version

l Please read this Manual before operation

l Please well keep it for future reference

Keli Sensing Technology (Ningbo) Co., Ltd.


Using attention of the indicator
▲ The load cell and indicator should be reliably connected. The load cell shilded wire should be
grounded.
▲ When the indicator is connected to power supply,could not plug all the connection wire,avoid static
electric damage the indicator or load cell.
▲ Load cell and indicator are static sensitive devices, so static-proof measures should be taken.
▲ In the season of thunder, credible thunder-proof measures should be taken, in order to avoid devices’
damage and operators’ physical safety.
▲ Do not use in the areas of flammable or explosive gases, and pressure vessel.
▲ Load cell and indicator shall be far away from strong electric field, magnetic field, seriously corrosive
materials, and flammable & explosive materials.
▲ Do not use strong solvent (such as:benzene、nitryl oils) to clean the shell。
▲ Do not fill in any liquids or electric granule into the system, so as to avoid device damage or electric
shock to the operators.
▲ It is not allowed to unlock the lead seal without any authorization of technical department. Calibrated
after open lead seal.
☆ Battery is consumable, not including in the guarantee range.
☆ To extend battery life, be sure to fully charged before use.
☆ If the indicator is not working for a long time, the battery should be chaged every two months,each time
should be charged 24 hours.
☆ When transiting or installing, make sure it is carefully managed. Great Vibration, strike or knock should
be avoid to prevent the battery short-connected inside or damaged.
◆ The indicator should be placed flatly and not be in the straight sunshine for clear indication and
long-term service life.
◆ The indicator should not be placed in a vibrating area full of dust, and not in a sloppy environment.
◆ Before plug the connection wire between indicator and external device, you must first cut off the power
supply of indicator.
◆ Regulations about external connection in the manual should be strictly observed. It is not allowed
change the connection at random.
◆ It is not allowed to open the indicator, or we do not guarantee to repair.In the indicator there is high pressure
strong electric,non-professional people do not repair the indicator,avoid to make more big damage to the
indicator ,or accident happen.
◆ If not man-made trouble happens within one year after the date of sale under regular service condition,
we guarantee to keep it in good repair free of charge for life. Please send the troubled indicator and
guarantee card (the serial number should be coincident) to the special repair shop or the supplier.
◆ If it is over gurantee time or man-made trouble ,and the indicator is broken,manufacturer will charge to
repair the indicator.
Cause of function improvement,this manaul would be different with the real product,please contact
with manufacturer for the latest version.
Dear customer,thanks for choosing Keli products.
Before using the indicator,please read the user manual carefully.

Keli Sensing Technology (Ningbo) Co., Ltd.


ADD:NO199,CHANGXING ROAD,JIANGBEI DISTRICT,NINGBO,CHINA
Service line:400-887-4165 800-857-4165
FAX: 0574-87562298
POST CODE:315033
WEBSITE:http://www.kelichina.com
The company reserves the right to modify the instruction.
1.0 General
XK3118T16 indicator , adopts high-strength interference single-chip microprocessor, fixed with
high-precision △-∑ A/D conversion technique. It is used in platform scale、bench scale, ect.
Main functions: accumulating、up and down limit ,peak keeping,average value,RS232 ,scoreboard.
Technical parameters
u A/D Conversion Mode: Adopt Δ-Σtechnology,20bitAD code,10times/second;
u The smallest input voltage of Maximum number of load cell intervals(n): ≈1.5uV/e;
u Bridge voltage of load cell:DC 5V,could connect 1~6 pcs of 350Ω load cell;
u Signal input range: -16mV~18mV;
u Load cell connection mode: Four wire mode;
u Division: 1/2/5/10/20/50 optional;
u Display: 6 bits LED,12 pcs of status indicator,3 pcs of power indicator,1 pc
of AC indicator;
u Scoreboard port: current loop/RS232 output,Baud rate:600bps;
u Serial communication port:
Transmission: RS232 ;
Baud rate: 1200/2400/4800/9600 optional;
u Power supply: AC 110~220V,50Hz~60Hz;
u Power supply: Build-in battery 6V/4Ah;
u Working Temperature: 0oC~40oC, Humidity 10%~85%(RH) ,no condensation;
u Storing Temperature:: -10oC~40oC,Humidity 10%~95%(RH) ,no condensation;
u Weight : about 1.5kg;
u Dimension(L*W*H) : about 236*150*150mm。
2.0 INSTALLATION
2.1 Diagram of indicator
XK3118T16 front board
XK3118T16 back side

2.2 Load cell and indicator connection(Left side) and serial port communication wire connection(Right side)
5 pin aviation plug DB 9 pin Instruction
Instruction Current loop of
PC RS232
scoreboard
Scoreboard
3 Indicator
① Excitation E+ ⑧ current loop ②
4 2 sending(TXD)
output(+)
5 1 Scoreboard Indicator
② Excitation E- ⑨ current loop ③ receiving( RX
output(-) D)

③ Shilded ⑤ GND

④ Signal S+ ②、⑤ could


connect with
⑤ Signal S-
scoreboard(RS232)
▲!Connection between load cell and indicator must be reliable; Connection or disconnection are not
allowed when the indicator is on, which may damage the indicator or load cells.
▲!Static protection must be properly adopted as the load cell and indicator are all static sensitive
equipments. Welding or other strong electricity operation should be strictly forbidden. During
thunderstorm season, proper lightening protection should be taken care of to protect the load cells
and indicators from damaging by lightening and to ensure the personal safety and the safely running
of the weighing and related equipments.
3.0 OPERATION
3.1 KEY INSTRUCTION

Fn 1: Fn2: Fn3:
KEY NAME Under weighing status,press Under weighing,press Go into parameter
lightly. long time setting
Turn off Turn on and off of the indicator NO NO
F1 NO NO NO
C Exit “NET” displaying mode NO Exit parameter setting
Look at cumulative
Total Total,refer 3.5 Modify the parameter
sum,refer 3.5
Peak keeping,average
M NO NO
value,normal weighing exchange
Flash digital move
Tare Tare operation,refer 3.4 NO
right
Unit change,according to
Go into parameter
Function parameter setting’s function to Exit parameter setting
setting,refer 3.6
realize,refere 3.6
Confirm the
Zero Zero setting,refer 3.3 NO
parameter setted

Note: below use “↙” to replace.


3.2 Power on and auto-zero
Indicator power supply is controlled by on-off key. After power-on, the indicator will first self-check
automatically. Even if the platform departs zero, It will be zero as long as it is within the zero range. The
indicator will show zero and the light of “Zero” will shine. If it departs over zero range, it will show current
reading.
When the indicator turn on,press on-off key(about two seconds),the indicator is power-off.
The back side of the indicator,there is a switch ,it is for AC power supply.

3.2 MANNUALLY ZERO


Indicator will back to zero when pressing “ZERO” key if the gross weight is within the range of manually zero
and stable. Manually zero is not valid under the “NET” displaying mode;
3.3 TARE
TARE Operation can be carried out if both gross weight and net weight are above zero and stable. The
indicator will display “0” after pressing the “TARE” key. The tared weight is current gross weight. And the
indicator will enter “NET” displaying mode and the “NET” indicator lights will be on;
The indicator will exit “NET” displaying mode after pressing “TARE” when the gross weight is “0” .Under “NET”
displaying mode,press “C" will exit “NET” displaying mode.

3.5 Accumulating、Accumulating display and accumulating deletion


When net weight is over min capacity( 5 division values)and stable, press“Σ”to add present net weights.
The light of “Total”will shine, and show total value. 1 seconds later, it will show accumulated times【n ***】,another
1 second later, it will exit from accumulation status. Next operation should be done when net weight is less than min
weighing capacity;
Press“Σ” for a long time in the status of weighing, and stop until the buzzer rings. At this time,“Fn” and “Total” will
shines and you can see total accumulating value. Press “↙”to check the accumulating time, then press“↙”to exit;
Press“Σ” for a long time to accumulating display, press key “→”the indicator will show【CLRAr-】 ,which means to
delete the accumulating values or not. Press“↙”to delete. Press“ C”to exit, but not delete.

Operation as below:

Keep Pressing “TOTAL” Display Total Value Press “↙” to see the adding times Press “↙” to exit

Press “→” to clear total value Press “↙” to confirm clear

Press “C” not clear

Note:During kg and lb unit switching,the indicator automatically remove cumulative data and cumulative times.
3.6 OTHER PARAMETER SETTING AND FUNCTION
Parameter setting mode can be entered if keep pressing “FUNCTION” under normal weighing status until
buzzer alarms. Detailed operations are as follows:

Steps Operations Display Note


Keep Prssing
“Function” Setting:
“FUNCTION” to Enter
1 【Fn **】 【 Lb】
:One Key Switch (kg/lb)
“↑”to Switch
【 --】:No Function
“↙”to Confirm
Power Save Setting:
【oFF】 :Power save mode off
【 oN】:Open power save mode. The power save mode will

“↑”to Switch be entered 3 minutes after weight stable. The indicator will
2 【PS **】
“↙”to Confirm only display date circultly in last digit and the display light
is the lowest.
【 onP 】: Enhanced power save mode which will
automatically turn off the indicator after 2 minutes power
save mode.
“↑”to Switch
3 【br****】 Baud Rate Setting:1200~9600bps Optional
“↙”to Confirm
“↑”to Switch Communicate Mode Setting:1~7 Optional, Detailed format
4 【Co *】
“↙”to Confirm followed
High Setting:

Press “→” the flash digit will move towards right


5 High Setting 【H*****】
Press “↑” to increase the number of flash digit
Press “↙”to confirm and enter next step such as 2000
Low Setting:

Press “→” the flash digit will move towards right

6 Low Setting 【L*****】 Press “↑” to increase the number of flash digit

Press “↙”to confirm and enter next step such as 1000

If weight is higher than High setting the “HI” light will be on


If weight is lower than Low setting the “LO” light will be on
If the weight is between High and Low setting then the “OK” light
will be on
Note:1 、 If low setting value is higher than high setting
value,weight function would be abnormal.2、If kg and lb unit
changing,high and low setting need be redone.

Set brightness:press key“↑”,and choose from“0~7”,the brightness


7 Brightness setting 【dIS***】
change from weak to strong.
Average stable range Set average stable range,according to the actual situation 1d、3d、
8 【Stb***】
setting 6d、9d、12d、15d.

★ If complete parameter settings, just press "C" key to quickly exit;During parameter setting, if more
than 3 minutes not operate, it will automatic return to weighing mode.
Kg lb unit switching:

When allow kg and lb Units switching, press key "FN" to switch unit kg and lb.

Peak Features:
In some cases, need to collect some of the peaks, and the function is designed for this situation.
When you press the "M" key, the instrument will be in general mode, peak mode, the average mode. When the
peak indicator light,the instrument go into peak detection status.When the weight is less than 5d,peak hold for
8 seconds,then automatically cleared.When the new peak value is bigger than old one,the max value update
automatically.Or the max value would not update.Press key "C",could quickly clear the peak value,re-test the
peak value.

Under peak hold status, the upper and lower check weighing function, cumulative, tare is invalid.

Average Features:
In some cases, need to be dynamic weighing, the weighing platform is not stable at this time. Instrument will
automatically collect the weight and calculate based on the average.If it is in the range of average,the result
would be locked.If it is not in the range,the instrument will unlock the value.When the zero indicator light,the
instrument automatically remove lock value.
When the peak and average instruction light off,the instrument is under normal weighing state.

Check weighing Features:


In some cases, need to re-check function, when the weight is less or equal than lower limit setting value, the
lower indicator "LO" light, when the weigh is between the upper and lower limit setting value, qualified light
"OK" light, when the weight is greater or equal to the upper limit setting value, the maximum light "HI" light.
Note: upper and lower limits setting value
COMMUNICATION MODE FORMAT:

Number Note
Serial of each
frame
Reversely send the Net Weight date. For example if the net weight is 23.45kg,
1 8 ASCII code 54.3200 will be sent. And if the net weight is –23.45kg, ASCII code
54.320- will be sent.
2 8 Reversely send Gross Weight date. The format is same as serial 1
Positively send the Net Weight date with unit. For example if the netweightis
3 14 23.45kg, ASCII code =0023.45(kg)will be sent.
End with Hex number OD,OA
4 14 Positively send the Net Weight date with unit. The format is same as serial 3
Order Response Mode:Order mode 02 “Order” 03(Hex)
There are 5 pieces order,ASCII code 'A'~'E'. Take gross weight 23.45kg,
netweight13.45kg and tare 10.00kg for example
'A':Read gross weight, indicator back:GW:0023.45(kg)
Not
5 'B':Read net weight, indicator back:NW:0013.45(kg)
confirm
'C':Read tare, indicator back:TW:0010.00(kg)
'D':Manually zero, indicator back:'D'
'E':Tare operation, indicator back:'E'
All order back add 02 at the beginning and 03 at the end(Hex)
Net and total weight can be automatically output in Total operation and the date
6
can be printed if connected with serial printer
7 If choose RS232 of Keli scoreboard,then current loop of scoreboard is invalide.

3.7 Battery Maintenance and Caution


The indicator uses 6V/4A battery. If the indicator is fully charged,no power save mode,could continue work 30
hours.If under power save mode,could use 10 hours more.When connect with AC,and open the back side of the
indicator,indicator will charge the battery automatically. The AC pilot lamp get lighted.
Battery is consumable, not including in the guarantee range.
★ To extend battery life, be sure to fully charged before use.
If the indicator is not working for a long time, the battery should be chaged every two months,each time
should be charged 24 hours.
★ When transiting or installing, make sure it is carefully managed. Great Vibration, strike or knock should
be avoid to prevent the battery short-connected inside or damaged.

3.8 Low Battery Alarm and Auto Turn Off


When use battery only for working,if voltage is more than 6.1V,three pilot lamps all get lighted.If voltage is between
5.8~6.1V,two pilot lamps get lighted.If voltage is between 5.5~5.8V,one pilot lamp get lighted.If the voltage is close to
5.5V,the indicator will turn off automatically.
4.0 ERROR NOTICE
Display Note
Err 01 Exceed the Zero Range,or setting Zero is not stable
Err 02 Not Meet the Requirement of Total
Err 03 Weight Overloaded,or load cell connection is not good

Err 04 Weight not stable during Calibration


Load Calibration Error. Too low load or Calibration code too small or AD
Err 05
reverse
Err 06 not meet requirement of tare,scale is not stable or over load.
Err 09 Data read verify Error, Data Memory Damaged
Err 10 Boot verify Error, SCM damaged

Appendix Packing list


NO Name Model QTY Remark
1 indicator XK3118T16 1pc

2 User manual XK3118T16user manual 1pc

3 debugging manual XK3118T16debugging manual 1 pc

4 Eligiable certificate XK3118T16Eligiable certificate 1 pc

5 Plug of load cell wire 5-pin 16M Aviation plug 1pc

6 Plug of the indicator XK3118T16 plug 1pc

7 Screw 1 M4×8 1 set

8 Screw 2 M6×25 4pcs

9 Hex nut 1 M4 1pc

10 Hex nut 2 M6 4pcs

11 Power fuse 0.5A,Φ5mm×20mm 1pc

Keli Sensing Technology (Ningbo) Co., Ltd..