NCC-61 With RET ENG - User Manual V2020 - 01 - 21 - 011900 (046-090)
NCC-61 With RET ENG - User Manual V2020 - 01 - 21 - 011900 (046-090)
NCC-61 With RET ENG - User Manual V2020 - 01 - 21 - 011900 (046-090)
(2) Eight inches of space behind the analyzer must be left for air flow.
(3) There should be 50 cm of space above to either side of the analyzer for
service access.
(4) Sufficient space is required beneath for placing reagents, waste
containers.
WARNING
46
Chapter 4 Installation
(7) Away from communication equipment which may interfere the analyzer by
producing high frequency electric wave.
WARNING
47
Chapter 4 Installation
(2) 250ml of sodium hypochlorite solution (12% chlorine) to handle the waste
biological risk.
WARNING
48
Chapter 4 Installation
NOTE
⮚ After installation, all tubes should be in a nature relaxed state and without
distortion.
Take out the lyse tube with red faucet from the accessories box, and inset
it to the LYSE connector on the back panel. Place the other end of the tube
into the lyse container and twist the cap tightly.
2. DILUENT Tubing Installation
Take out the diluent tube with blue faucet from the accessories box, and
inset it to the DILUENT connector on the back panel. Place the other end of
the tube into the diluent container and twist the cap tightly.
3. DETERGENT Tubing Installation
Take out the detergent tube with green faucet from the accessories box,
and inset it to the DETERGENT connector on the back panel. Place the other
end of the tube into the detergent container and twist the cap tightly.
4. SHEATH Tubing Installation
Take out the sheath tube with yellow faucet from the accessories box,
and inset it to the SHEATH connector on the back panel. Place the other end
of the tube into the sheath container and twist the cap tightly.
5. WASTE Tubing Installation
49
Chapter 4 Installation
Take out the waste tube with faucet from the accessories box, and inset it
to the connector on the back panel. Inset BNC plug to the SENSOR connector
on the left panel. Tightly twist the tube’s cap clockwise onto the waste
container. Place the waster container on the level at least 50cm lower than the
analyzer.
8. Install the ink cartridges and paper according to the instructions. Ensure the
printer is adjusted to the correct receiver size.
9. Connect the power cord to a grounded outlet and turn the power on.
50
Chapter 4 Installation
NOTE
51
Chapter 5 Principles of Operation
5. Principles of Operation
5.1 Overview
NCC-61 uses electrical impedance method (also known as Coulter
principle) to detect the amount and volume distribution of RBC and PLT. The
colorimetric method is for determining the content of HGB. The Multi-angle
laser scattered method is for the five part differential of WBC. Three
separated channels are used for getting the blood cells counting results
respectively.
(1)
WBC and five part differential data of sheath flow regulator are detected by
laser.
(2) HGB is detected by colorimetric assay in WBC/HGB transducer.
(3) The data of RBC and PLT is detected by electrical impedance analysis in
RBC transducer.
The analyzer aspirates, dilutes and mixes the samples and then
detects parameters in each counting process.
Diluent sampling: ≤ 20 µL
The whole blood sample is aspirated into the analyzer by the precision
52
Chapter 5 Principles of Operation
the results of WBC count/HGB test, WBC/PLT count and WBC five part
differential.
53
Chapter 5 Principles of Operation
54
Chapter 5 Principles of Operation
55
Chapter 5 Principles of Operation
56
Chapter 5 Principles of Operation
The gray area is the ghost cells. It reflects that RBC dissolve into pieces
on the scatter plot; green is for lymphocyte group; pink is for monocyte group;
blue is for neutrophil; white is for basophil group; red is for eosinophil group.
Get the value of WOC and WIC simultaneously by laser and electrical
impedance methods
57
Chapter 5 Principles of Operation
Lym% = Lym#/WBC
● Monocyte Number (Mon#)
● Monocyte Percent
Mon% = Mon# /WBC
● Neutrophil Number (Neu#)
● Neutrophil Percent
Neu%=Neu#/WBC
● Eosinophil Number (Eos#)
● Eosinophil Percent
Eos%=Eos#/WBC
● Basophil Number( Baso#)
● Basophil Percent
Baso%=Baso#/WBC
58
Chapter 5 Principles of Operation
59
Chapter 5 Principles of Operation
Figure 5-4 Volumetric Metering
The volumetric metering unit controls the sample size passing through
the pore during counting to obtain the exact counting results in quantitative
samples. The volumetric metering unit includes metering tube and two
photoelectric sensors.
As shown in Figure 5-5, empty the metering tube before testing. The
liquid level of metering tube declines slowly as the sample passing through
the pore. When the liquid level passes through the start detector, one
electrical signal generates, and the analyzer starts counting. When the liquid
level reaches the stop detector, it also generates an electrical signal, then the
counting finishes. If there were bubbles or other abnormal stream in the flow
system, "bubble" or "clog" alarm pops up. Please refer to Chapter 11
Troubleshooting.
60
Chapter 5 Principles of Operation
The analyzer gets the number of red blood cell (RBC) by measuring the
corresponding electrical pulse numbers of RBC directly. The unit is 10^12/L.
RBC = n ×10^12 / L
● MCV
● HCT
The
hematocrit (HCT) is the ratio of red blood cells to plasma. It is
expressed as a percentage of the whole blood volume. The HCT is calculated
from the RBC count and the MCV as follows.
● MCH
● MCHC
61
Chapter 5 Principles of Operation
● RDW-CV
The RDW-CV is derived from the RBC histogram and being expressed in
percent.
● RDW-SD
The RDW-SD is the width of 20% peak value of red blood cell distribution
histogram .The unit is fL.
PLT = n ×10^9/ L
● MPV
The mean platelet volume (MPV) is derived from the PLT histogram after
the PLT count has been determined. The unit is fL.
● PDW
62
Chapter 5 Principles of Operation
● PCT
The PLT is calculated as follows. The unit of PLT is 10^9/L. The unit of
MPV is fL
63
Chapter 5 Principles of Operation
Illuminated with polarized light, the stained dotted or net substances will
strengthen scatted light on wide-angle direction:
64
Chapter 5 Principles of Operation
65
Chapter 5 Principles of Operation
5.7.4 RETIC_ABS
5.7.5 IRF
IRF has more RNA than mature reticulocytes and absorb more stain. So
their wide-angle scattering light signal is larger. IRF is classified as
reticulocyte population which exceeds preconcerted scattering threshold, as
the purple part in Figure 5-8.
Phoenix NCC-61 User Manual
66
Chapter 5 Principles of Operation
2) Monitor bone marrow toxic insults from drugs (for example, AZT)
4) Classify anemia
67
Chapter 6 Settings
6. Settings
6.1 Overview
Initialization setting of NCC-61 has been done before delivery. Setting of
the interface at the first boot is default. To meet the different needs, some
parameters can be reset.
6.2 Settings
Click “Setup” to enter setting interface, see Figure 6-1.
68
Chapter 6 Settings
Soak sheath flow regulator: click this button to clean inner wall of sheath
flow regulator.
6.4 X-B QC
Click “X-B QC” to enter QC interface. Please refer to Chapter 7 for
details.
6.5 X-R QC
6.6 X QC
6.7 Limit
70
Chapter 6 Settings
6.8 Time
Click “Time” to set it.
71
Chapter 6 Settings
6.9 Parameter
Choose unit of WBC, RBC, PLT and HGB/MCHC and modify the reaction
time of RRBC. Click “Default” to revert RRBC reaction time to factory settings.
Click “OK” to save modified settings.
6.10 Print
Printer type: USB port printer (A5), USB port printer (A4)
72
Chapter 6 Settings
Auto print: open/close auto print. If it’s open, test result is auto printed
after counting. If it’s closed, it needs to manual print.
Printer title: input hospital name here, hospital name displays in printed
report title. Click “OK” to save the modified settings.
6.11 Transmit
Click “Transmit” to enter the interface as shown in Figure 6-8.
Set the local IP, server IP, local mask, local gateway and port number as
connecting with LIS system. The native mask and the local gateway can be
selected by default, the others shall be reset.
Select either “On” or “Off” auto transmit as connecting with LIS system. “Trans
Histo” and “Trans Scatter” can be selected.
73
Chapter 6 Settings
6.12 Maint.Set
74
Chapter 6 Settings
settings as blank test is necessary in each boot. The analyzer does not
perform it if it is “Off”.
Auto sleep: the analyzer automatically enters the dormant state without
any operation for an interval of time. Users can adjust dormancy length
according to the necessary.
6.13 Version
Click “Version” to pop up version dialog. See Figure 6-10.
75
Chapter 6 Settings
6.14 User
Click “Add” to pop up “Add user” dialog to edit new user’s name,
password and group. “Group” is divided into “Ordinary user” and
“Administrator”, which are given different permissions. The administrator's
permissions are higher than the Ordinary user’s. The administrator can
operate all the functions, while the general user can not delete data, use the
export function or calibrate the analyzer. See Figure 6-12.
6.15 Service
77
Chapter 6 Settings
6.16 Reagent
The activation date, reagent amount, valid period of lyse, sheath, diluents
and detergent. For example, click “Change” of diluent to change diluent, see
the popup dialog in Figure 6-15.
Take out the diluent activation card from the diluent container and click
“Activate”. 15 seconds countdown starts. Put the IC card onto card reader and
hear a "tick" sound, which means successful card read. Successful activation
displays in dialog box. The activation date is the current date after activating.
The valid period is three months. The reagent balance is the current
maximum reagent dose. The remaining amount subtracts the amount
consumed by the analyzer during operation. The activation method of other
reagents is the same as diluent’s.
78
Chapter 6 Settings
79
Chapter 6 Settings
6.18Display
80
Chapter 6 Settings
figure6-17 Display
CAUTION
81
Chapter7 Daily Operation
7. Daily Operation
7.1 Overview
This chapter describes the whole procedures of daily operation from
startup to shutoff, and explains the process of different modes of sample
analysis in detail.
CAUTION
82
Chapter7 Daily Operation
7.2 Preparations
Check the analyzer as the following steps before startup.
Ensure diluent, lyse, detergent and sheath meet the test requirements.
Ensure the tubing of reagents and waste connected well and without
bending.
Ensure the power is on and the cable has been connected with the
analyzer and the computer properly.
WARNING
7.3 Startup
Turn on the power switch on the left panel, then the status indicator on
the front panel turns orange. The analyzer automatically checks the operation
Phoenix NCC-61 User Manual
83
Chapter7 Daily Operation
Figure 7-2.
The analyzer enters test interface after entering password and user’s
name. See Figure 7-3.
84
Chapter7 Daily Operation
After startup, blank test should be done before sample test. Operator can
set
to run it automatically after startup, see Chapter 6 Settings for details. The
acceptable range of blank test is listed in Table 7-1.
WBC ≤0.20x10∧9 /L
RBC ≤0.02x10∧12 /L
HGB ≤1g /L
PLT ≤10.0x10∧9 /L
If the blank result is out of this range, please repeat the above procedures
until it is in this range. If the results are still out of this range after repeating
five times, please refer to Section 11.4.2 of Chapter 11 Troubleshooting.
85
Chapter7 Daily Operation
WARNING
CAUTION
86
Chapter7 Daily Operation
collection.
1. Heparin
2. Sodium citrate
4. EDTA
In the salt of EDTA, use EDTA K2(United States and Japan)and EDTA
K3(United States and Europe),sometimes NA2EDTA. And EDTA K2,
EDTA K3 which recommend by ISCH in1993 are most widely used in the
blood test of the world. But other EDTA salts can also be used. EDTA
could lead to Pseudo-thrombocytopenia through Platelet aggregation.
(Incidence is about 1/800)
5. Fluoride
87
Chapter7 Daily Operation
2. Take a clean test tube under the aspiration probe, press “Drain” button
on front panel. The system automatically drains 500 μL diluent from
aspiration probe. It is recommended to put the test tube close to the
aspiration probe, so as to avoid bubbles or spillage.
3. Please quickly inject 20uL peripheral blood into the test tube filled with
diluents and mix it well.
CAUTION
⮚ Avoid the collected diluent mixing with dust; otherwise it may cause
analytical error.
⮚ Peripheral blood and diluent after full reaction, should be placed for 3
minutes, and then only after blending again that can do the analyze.
⮚ Ensure that the sample has been analyzed within 30 minutes after dilution,
otherwise the analysis results are not reliable.
88
Chapter7 Daily Operation
sampling method and the technical level to evaluate the stability of the
results under the diluent mode.
89
Chapter7 Daily Operation
ID: only numbers can be input here. If there’s no SN input, the analyzer
Phoenix NCC-61 User Manual
90