Ba01271pen 0517

Download as pdf or txt
Download as pdf or txt
You are on page 1of 48

BA01271P/00/EN/05.

17 Products Solutions Services


71353696

Operating Instructions
Cerabar PMC11, PMC21, PMP11,
PMP21, PMP23
Process pressure measurement
Pressure transducer for safe measurement and
monitoring of absolute and gauge pressure
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Order code: XXXXX-XXXXXX


1. Ser. no.: XXXXXXXXXXXX
Ext. ord. cd.: XXX.XXXX.XX

Serial number

2. www.endress.com/deviceviewer Endress+Hauser
Operations App

3.

A0023555

• Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
• To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser distributor will supply you with current information and updates to
these Instructions.

2 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Table of contents

Table of contents

1 Document information . . . . . . . . . . . . . . 4 9 Maintenance . . . . . . . . . . . . . . . . . . . . . . 26


1.1 Document function . . . . . . . . . . . . . . . . . . . . . 4 9.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . 27
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Documentation . . . . . . . . . . . . . . . . . . . . . . . . 5 10 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
1.4 Terms and abbreviations . . . . . . . . . . . . . . . . . 7
1.5 Turn down calculation . . . . . . . . . . . . . . . . . . . 8 10.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . 28
10.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

2 Basic safety instructions . . . . . . . . . . . . 9


11 Accessories . . . . . . . . . . . . . . . . . . . . . . . 29
2.1 Requirements concerning the staff . . . . . . . . . . 9
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . 9 11.1 Weld-in adapter . . . . . . . . . . . . . . . . . . . . . . 29
2.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . 10 11.2 Process adapter M24 . . . . . . . . . . . . . . . . . . . 29
2.4 Operational safety . . . . . . . . . . . . . . . . . . . . . 10 11.3 Plug-on display PHX20 . . . . . . . . . . . . . . . . . 30
2.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . 10 11.4 M12 plug connectors . . . . . . . . . . . . . . . . . . . 30

3 Product description . . . . . . . . . . . . . . . . 11 12 Technical data . . . . . . . . . . . . . . . . . . . . 31


3.1 Product design . . . . . . . . . . . . . . . . . . . . . . . . 11 12.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.2 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 12.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
12.3 Performance characteristics of ceramic
process isolating diaphragm . . . . . . . . . . . . . . 37
4 Incoming acceptance and product 12.4 Performance characteristics of metallic
identification . . . . . . . . . . . . . . . . . . . . . 13 process isolating diaphragm . . . . . . . . . . . . . . 39
12.5 Environment . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . 13
12.6 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
4.2 Product identification . . . . . . . . . . . . . . . . . . . 14
4.3 Storage and transport . . . . . . . . . . . . . . . . . . 14
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5 Installation . . . . . . . . . . . . . . . . . . . . . . . 16
5.1 Mounting dimensions . . . . . . . . . . . . . . . . . . 16
5.2 Installation conditions . . . . . . . . . . . . . . . . . . 16
5.3 Influence of the installation position . . . . . . . . 16
5.4 Mounting location . . . . . . . . . . . . . . . . . . . . . 17
5.5 Mounting the profiled seal for the universal
process adapter . . . . . . . . . . . . . . . . . . . . . . . 18
5.6 Mounting instructions for oxygen
applications . . . . . . . . . . . . . . . . . . . . . . . . . . 18
5.7 Post-installation check . . . . . . . . . . . . . . . . . . 19

6 Electrical connection . . . . . . . . . . . . . . 20
6.1 Connecting the measuring unit . . . . . . . . . . . . 20
6.2 Connection conditions . . . . . . . . . . . . . . . . . . 21
6.3 Connection data . . . . . . . . . . . . . . . . . . . . . . . 21
6.4 Post-connection check . . . . . . . . . . . . . . . . . . 22

7 Operation options . . . . . . . . . . . . . . . . . 23
7.1 Plug-on display PHX20 (optional) . . . . . . . . . . 23

8 Diagnostics and troubleshooting . . . 26


8.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 26
8.2 Response of output to errors . . . . . . . . . . . . . . 26
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Endress+Hauser 3
Document information Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

1 Document information

1.1 Document function


These Operating Instructions contain all the information that is required in various phases
of the life cycle of the device: from product identification, incoming acceptance and
storage, to mounting, connection, operation and commissioning through to
troubleshooting, maintenance and disposal.

1.2 Symbols used

1.2.1 Safety symbols

Symbol Meaning

DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.

WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.

CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.

NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.

1.2.2 Electrical symbols

Symbol Meaning Symbol Meaning

Protective ground connection Ground connection


A terminal which must be connected A grounded terminal which, as far as
to ground prior to establishing any the operator is concerned, is
other connections. grounded via a grounding system.

1.2.3 Tool symbols

Symbol Meaning

Open-ended wrench
A0011222

1.2.4 Symbols for certain types of information

Symbol Meaning

Permitted
Procedures, processes or actions that are permitted.

Forbidden
Procedures, processes or actions that are forbidden.

Tip
Indicates additional information.

4 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Document information

Symbol Meaning

Reference to documentation

A Reference to page

Reference to graphic

1. , 2. , 3. … Series of steps

Result of a step

Visual inspection

1.2.5 Symbols in graphics

Symbol Meaning

1, 2, 3 ... Item numbers

1. , 2. , 3. … Series of steps

A, B, C, ... Views

1.3 Documentation
The document types listed are available:
In the Download Area of the Endress+Hauser Internet site: www.endress.com →
Download

1.3.1 Technical Information (TI): planning aid for your device


PMC11: TI01133P
PMP11: TI01133P
PMC21: TI01133P
PMP21: TI01133P
PMP23: TI01203P
The document contains all the technical data on the device and provides an overview of
the accessories and other products that can be ordered for the device.

1.3.2 Brief Operating Instructions (KA): getting the 1st measured


value quickly
KA01164P:
The Brief Operating Instructions contain all the essential information from incoming
acceptance to initial commissioning.

Endress+Hauser 5
Document information Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

1.3.3 Safety Instructions (XA)


Safety Instructions (XA) are supplied with the device depending on the approval. These
instructions are an integral part of the Operating Instructions.

Device Directive Documentation Option 1)

PMP21 ATEX II 1/2G Ex ia IIC T4 Ga/Gb XA01271P BA


PMP23

PMC21 ATEX II 2G Ex ia IIC T4 Gb XA01271P BB

PMC21 ATEX II 3G Ex ec IIC T4 Gc XA01533P BC


PMP21

PMC21 FM IS Cl. I, Div.1 Gr. A-D T4 XA01321P FA


PMP21
PMP23

PMC21 CSA C/US IS Cl. I Div. 1 Gr. A-D XA01322P CB


PMP21
PMP23

PMC21 EAC Ex ia IIC T4 Ga/Gb XA01540P GA


PMP21
PMP23

PMC21 IEC Ex ia IIC T4 Ga/Gb XA01271P IA


PMP21
PMP23

PMC21 NEPSI Ex ia IIC T4 XA01363P NA


PMP21
PMP23

PMC21 TIIS Ex ia IIC T4 In preparation TA


PMP21
PMP23

1) Product Configurator, order code for "Approval"

The nameplate provides information on the Safety Instructions (XA) that are relevant
for the device.

6 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Document information

1.4 Terms and abbreviations

p
0
LRL LRV URV URL MWP OPL
A0029505

Item Term/ Explanation


abbreviation

1 OPL The OPL (over pressure limit = sensor overload limit) for the measuring device
depends on the lowest-rated element, with regard to pressure, of the selected
components, i.e. the process connection has to be taken into consideration in
addition to the measuring cell. Also observe pressure-temperature dependency. For
the relevant standards and additional notes, see the "Pressure specifications" section
→  44 .
The OPL may only be applied for a limited period of time.

2 MWP The MWP (maximum working pressure) for the sensors depends on the lowest-
rated element, with regard to pressure, of the selected components, i.e. the process
connection has to be taken into consideration in addition to the measuring cell.
Also observe pressure-temperature dependency. For the relevant standards and
additional notes, see the "Pressure specifications" section →  44 .
The MWP may be applied at the device for an unlimited period.
The MWP can also be found on the nameplate.

3 Maximum sensor Span between LRL and URL


measuring range This sensor measuring range is equivalent to the maximum calibratable/adjustable
span.

4 Calibrated/ Span between LRV and URV


adjusted span Factory setting: 0 to URL
Other calibrated spans can be ordered as customized spans.

p - Pressure

- LRL Lower range limit

- URL Upper range limit

- LRV Lower range value

- URV Upper range value

- TD (turn down) Turn down


The turn down is preset at the factory and cannot be changed.
Example - see the following section.

Endress+Hauser 7
Document information Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

1.5 Turn down calculation

1=2 3
LRL LRV URV URL

A0029545

1 Calibrated/adjusted span
2 Zero point-based span
3 URL sensor

Example

• Sensor:10 bar (150 psi) • Calibrated/adjusted span: 0 to 5 bar (0 to 75 psi)


• Upper range value (URL) = 10 bar (150 psi) • Lower range value (LRV) = 0 bar (0 psi)
• Upper range value (URV) = 5 bar (75 psi)
Turn down (TD):

URL
TD =
|URV - LRV|

10 bar (150 psi)


TD = = 2
|5 bar (75 psi) - 0 bar (0 psi)|

In this example, the TD is 2:1.


This span is based on the zero point.

8 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Basic safety instructions

2 Basic safety instructions

2.1 Requirements concerning the staff


The personnel for installation, commissioning, diagnostics and maintenance must fulfill
the following requirements:
‣ Trained, qualified specialists: must have a relevant qualification for this specific
function and task
‣ Are authorized by the plant owner/operator
‣ Are familiar with federal/national regulations
‣ Before beginning work, the specialist staff must have read and understood the
instructions in the Operating Instructions and supplementary documentation as well as
in the certificates (depending on the application)
‣ Following instructions and basic conditions
The operating personnel must fulfill the following requirements:
‣ Being instructed and authorized according to the requirements of the task by the
facility's owner-operator
‣ Following the instructions in these Operating Instructions

2.2 Designated use

2.2.1 Application and media


The Cerabar is used to measure absolute and gauge pressure in gases, vapors and liquids.
The process-wetted materials of the measuring device must have an adequate level of
resistance to the media.
The measuring device may be used for the following measurements (process variables)
• in compliance with the limit values specified under "Technical data"
• in compliance with the conditions that are listed in additional documentation such as the
XA and this manual.

Measured process variable


• PMC11: gauge pressure
• PMP11: gauge pressure
• PMC21: gauge pressure or absolute pressure
• PMP21: gauge pressure or absolute pressure
• PMP23: gauge pressure or absolute pressure

Calculated process variable


Pressure

2.2.2 Incorrect use


The manufacturer is not liable for damage caused by improper or non-designated use.
Verification for borderline cases:
‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance
in verifying the corrosion resistance of process-wetted materials, but does not accept
any warranty or liability.

2.2.3 Residual risks


When in operation, the housing may reach a temperature close to the process
temperature.

Endress+Hauser 9
Basic safety instructions Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Danger of burns from contact with surfaces!


‣ For elevated process temperatures, ensure protection against contact to prevent burns.

2.3 Workplace safety


For work on and with the device:
‣ Wear the required personal protective equipment according to federal/national
regulations.
‣ Switch off the supply voltage before connecting the device.

2.4 Operational safety


Risk of injury!
‣ Operate the device in proper technical condition and fail-safe condition only.
‣ The operator is responsible for interference-free operation of the device.
Conversions to the device
Unauthorized modifications to the device are not permitted and can lead to unforeseeable
dangers.
‣ If, despite this, modifications are required, consult with Endress+Hauser.
Hazardous area
To eliminate the risk of danger to persons or the facility when the device is used in the
approval-related area (e.g. explosion protection, pressure equipment safety):
‣ Check the nameplate to verify if the device ordered can be put to its intended use in the
approval-related area.
‣ Observe the specifications in the separate supplementary documentation, such as the
XA or SD, which is an integral part of these Instructions.

2.5 Product safety


This measuring device is designed in accordance with good engineering practice to meet
state-of-the-art safety requirements, has been tested, and left the factory in a condition in
which it is safe to operate.
It meets general safety standards and legal requirements. It also complies with the EU
directives listed in the device-specific EU Declaration of Conformity. Endress+Hauser
confirms this by affixing the CE mark to the device.

10 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Product description

3 Product description

3.1 Product design


Overview Position Description

A B C-1 C-2 A Valve plug

B Cable

C- 1 M12 plug
Housing cap made of plastic

C- 2 M12 plug
A0027231 A0027232 A0021987 A0027289 For Ex ec and IP69: metal housing cap
Metal housing cap can also be ordered as an option.

D D Housing
E Process connection (sample illustration)

A0027226

A0027215

A0027227

3.2 Function

3.2.1 Calculating the pressure

Devices with ceramic process isolating diaphragm (Ceraphire®)


The ceramic sensor is an oil-free sensor, i.e. the process pressure acts directly on the
robust ceramic process isolating diaphragm and causes it to deflect. A pressure-dependent

Endress+Hauser 11
Product description Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

change in capacitance is measured at the electrodes of the ceramic substrate and the
process isolating diaphragm. The measuring range is determined by the thickness of the
ceramic process isolating diaphragm.

Devices with metallic process isolating diaphragm


The process pressure deflects the metal process isolating diaphragm of the sensor and a fill
fluid transfers the pressure to a Wheatstone bridge (semiconductor technology). The
pressure-dependent change in the bridge output voltage is measured and evaluated.

12 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Incoming acceptance and product identification

4 Incoming acceptance and product


identification

4.1 Incoming acceptance

DELIVERY NOTE

A0028673

1=2

A0016870

Is the order code on the delivery note (1) identical to the order code on
the product sticker (2)?

A0022100

A0028673

A0022103

Are the goods undamaged?

A0028673
DELIVERY NOTE

A0022105

Do the data on the nameplate correspond to the order specifications and


the delivery note?

Endress+Hauser 13
Incoming acceptance and product identification Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

A0028673

A0022106

Is the documentation available?


If required (see nameplate): Are the safety instructions (XA) present?

If one of these conditions does not apply, please contact your


Endress+Hauser sales office.

4.2 Product identification


The following options are available for identification of the measuring device:
• Nameplate specifications
• Order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer
(www.endress.com/deviceviewer): All information about the measuring device is
displayed.
For an overview of the technical documentation provided, enter the serial number from
the nameplates in the W@M Device Viewer (www.endress.com/deviceviewer)

4.2.1 Nameplate

1
2 Cerabar
Made in Germany,
D-79689 Maulburg
3 Ord. cd.:
4 Ser. no.:
5 Ext. ord. cd.:

Date:
TAG:

A0024456

1 Manufacturer's address
2 Device name
3 Order number
4 Serial number
5 Extended order number

4.3 Storage and transport

4.3.1 Storage conditions


Use original packaging.
Store the measuring device in clean and dry conditions and protect from damage caused by
shocks (EN 837-2).

Storage temperature range


–40 to +85 °C (–40 to +185 °F)

14 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Incoming acceptance and product identification

4.3.2 Transporting the product to the measuring point


LWARNING
Incorrect transport!
Housing and diaphragm may become damaged, and there is a risk of injury!
‣ Transport the measuring device to the measuring point in its original packaging or by
the process connection.

Endress+Hauser 15
Installation Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

5 Installation

5.1 Mounting dimensions


For dimensions, see the "Mechanical construction" section in the Technical Information.

5.2 Installation conditions


• No moisture may enter the housing when installing or operating the device, or when
establishing the electrical connection.
• For metal M12 plugs: Do not remove the protection cap (only for IP69 and Ex ec version)
of the M12 plug terminal until shortly before the electrical connection.
• Do not clean or touch process isolating diaphragms with hard and/or pointed objects.
• Do not remove the protector on the process isolating diaphragm until just before
installation.
• Always firmly tighten the cable entry.
• Point the cable and connector downwards where possible to prevent moisture from
entering (e.g. rain or condensation water).
• Protect the housing from impact
• The following instruction applies for devices with a gauge pressure sensor and M12 plug
or valve plug:

NOTICE
If a heated device is cooled during the cleaning process (e.g. by cold water), a vacuum
develops for a short time, whereby moisture can penetrate the sensor through the
pressure compensation element (1).
Device could be destroyed!
‣ In this case, mount the device with the pressure compensation element (1) pointing
diagonally downwards where possible or to the side.

1 1 1
A0022252

5.3 Influence of the installation position


Any orientation is possible. However, the orientation may cause a zero point shift i.e. the
measured value does not show zero when the vessel is empty or partially full.

A B C

A0024708

16 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Installation

Type Process isolating Process isolating diaphragm Process isolating diaphragm


diaphragm axis is pointing upwards (B) pointing downwards (C)
horizontal (A)

PMP11 Calibration position, no Up to +4 mbar (+0.058 psi) Up to –4 mbar (–0.058 psi)


PMP21 effect
PMP23

PMC11, PMC21 Calibration position, no Up to Up to


< 1 bar (15 psi) effect +0.3 mbar (+0.0044 psi) –0.3 mbar (–0.0044 psi)

PMC11, PMC21 Calibration position, no Up to +3 mbar (+0.0435 psi) Up to –3 mbar (–0.0435 psi)
≥1 bar (15 psi) effect

5.4 Mounting location

5.4.1 Pressure measurement

Pressure measurement in gases


Mount the device with shutoff device above the tapping point so that any condensate can
flow into the process.

A0021904

1 Device
2 Shutoff device

Pressure measurement in vapors


For pressure measurement in vapors, use a siphon. The siphon reduces the temperature to
almost ambient temperature. Mount the device with a shutoff device at the same height as
the tapping point.
Advantage:
only minor/negligible heat effects on the device.
Note the max. permitted ambient temperature of the transmitter!

3
A0024395

1 Device
2 Shutoff device
3 Siphon

Endress+Hauser 17
Installation Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Pressure measurement in liquids


Mount the device with a shutoff device at the same height as the tapping point.

2 1
A0024399

1 Device
2 Shutoff device

5.4.2 Level measurement


• Always install the device below the lowest measuring point.
• Do not install the device at the following positions:
– In the filling curtain
– In the tank outlet
– In the suction area of a pump
– Or at a point in the tank which could be affected by pressure pulses from the agitator.

A0024405

5.5 Mounting the profiled seal for the universal process


adapter
For details on mounting, see KA00096F/00/A3.

5.6 Mounting instructions for oxygen applications


Oxygen and other gases can react explosively to oils, grease and plastics, such that, among
other things, the following precautions must be taken:
• All components of the system, such as measuring devices, must be cleaned in accordance
with the BAM requirements.
• Dependent on the materials used, a certain maximum temperature and a maximum
pressure for oxygen applications must not be exceeded.
• The following table lists devices (devices only, not accessories or enclosed accessories),
which are suitable for gaseous oxygen applications.

1)
Device pmax for oxygen applications Tmax for oxygen applications Option

PMC21 40 bar (600 psi) –10 to +60 °C (+14 to +140 °F) HB

1) Product Configurator, order code for "Service"

18 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Installation

5.7 Post-installation check


 Is the device undamaged (visual inspection)?

Does the device conform to the measuring point specifications?


For example:
 • Process temperature
• Process pressure
• Ambient temperature range
• Measuring range

 Are the measuring point identification and labeling correct (visual inspection)?

 Is the device adequately protected from precipitation and direct sunlight?

 Are the securing screws tightened securely?

 Does the pressure compensation element point downwards at an angle or to the side?

 To prevent the penetration of moisture: are the connecting cables/plugs pointing downwards?

Endress+Hauser 19
Electrical connection Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

6 Electrical connection

6.1 Connecting the measuring unit

6.1.1 Terminal assignment


LWARNING
Risk of injury from the uncontrolled activation of processes!
‣ Switch off the supply voltage before connecting the device.
‣ Make sure that downstream processes are not started unintentionally.
LWARNING
Supply voltage might be connected!
Risk of explosion!
‣ Make sure that no supply voltage is applied when connecting.
‣ Switch off the supply voltage before connecting the device.
LWARNING
Limitation of electrical safety due to incorrect connection!
‣ In accordance with IEC/EN61010 a separate circuit breaker must be provided for the
device .
‣ The device must be operated with a 500 mA fine-wire fuse (slow-blow).
‣ When using the measuring device in hazardous areas, installation must comply with
the corresponding national standards and regulations and the Safety Instructions or
Installation or Control Drawings.
‣ All explosion protection data are given in separate documentation which is available
upon request. The Ex documentation is supplied as standard with all devices approved
for use in explosion hazardous areas.
‣ Protective circuits against reverse polarity are integrated.
Connect the device in the following order:
1. Check whether the supply voltage matches the supply voltage indicated on the
nameplate.
2. Connect device in accordance with the following diagram.
Switch on supply voltage.
For devices with a cable connection: do not close reference air hose (see (a) in the
following drawings)! Protect reference air hose against penetration by water/condensate.

4 to 20 mA output
Device M12 plug Valve plug Cable

PMC11 0.5A 0.5A 1 0.5A


PMP11 L+ 1 L+ L+
PMC21 2 1
3 2
PMP21 3 4 L–
PMP23 L– 3
2
A0023487 L–
(a)
A0022823 A0023783

1 brown = L+
2 blue = L-
3 green/yellow = ground connection
(a) reference air hose

20 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Electrical connection

0 to 10 V output
Device M12 plug Valve plug Cable

PMC11 0.5A 1 -
PMP11 L+ 3
2 1
4
3
2 L–
L–
A0017576
L+
0.5A
A0022822

6.1.2 Supply voltage


LWARNING
Supply voltage might be connected!
Risk of explosion!
‣ When using the measuring device in hazardous areas, installation must comply with
the corresponding national standards and regulations as well as the Safety Instructions.
‣ All explosion protection data are given in separate documentation which is available
upon request. The Ex documentation is supplied as standard with all devices approved
for use in explosion hazardous areas.

Electronic version Device Supply voltage

4 to 20 mA output PMC11 10 to 30 V DC
PMP11
PMC21
PMP21
PMP23

0 to 10 V output PMC11 12 to 30 V DC
PMP11

6.1.3 Current consumption and alarm signal

Number of wires Device Normal operation Alarm signal 1)

2 PMC11 ≤ 26 mA > 21 mA
PMP11
PMC21
PMP21
PMP23

3 PMC11 < 12 mA 11 V
PMP11

1) For MAX alarm (factory setting)

6.2 Connection conditions

6.2.1 Cable specification


For valve plug: < 1.5 mm2 (16 AWG) and Ø3.5 to 6.5 mm (0.14 to 0.26 in)

6.3 Connection data

Endress+Hauser 21
Electrical connection Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

6.3.1 Load (for 4 to 20 mA devices )


In order to guarantee sufficient terminal voltage in two-wire devices, a maximum load
resistance RL (including line resistance) must not be exceeded, depending on the supply
voltage UB of the supply unit.

R L max
[W ]
1068

614 2 R L max £ U B - 6.5V


22mA

10 20 30 UB
[V]
1

A0029452

1 Power supply 10 to 30 V DC
2 RLmax maximum load resistance
UB Supply voltage

6.3.2 Load resistance (for 0 to 10 V devices)


The load resistance must be ≥ 5 [kΩ].

6.4 Post-connection check


 Is the device or cable undamaged (visual check)?

 Do the cables comply with the requirements ?

 Do the cables have adequate strain relief?

 Are all the cable glands installed, firmly tightened and leak-tight?

 Does the supply voltage match the specifications on the nameplate?

 Is the terminal assignment correct ?

 If required: Has protective ground connection been established ?

22 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Operation options

7 Operation options

7.1 Plug-on display PHX20 (optional)


No display or other operation facility is required to operate the device. However, devices
with a valve plug can be fitted with the optional local display PHX20.
A 1-line liquid crystal display (LCD) is used. The local display shows measured values, fault
messages and information messages. The device display can be turned in 90° steps.
Depending on the orientation of the device, it is therefore easy to read the measured
values.

7.1.1 Storage conditions


• Use original packaging.
• Storage temperature range: –30 to +80 °C (–22 to +176 °F)

7.1.2 Installation

1 2 3 4

A0022208

1. Place seals between the sensor/plug-on display and the plug-on display/connector.
2. Fit the plug-on display (2) between the connector (3) and connector socket (1) of the
sensor.
3. Replace the securing screw (4) with the longer screw included in the delivery.
4. An adhesive label (included in the delivery) with information on the technical unit
can be fixed beneath the LED display.

Mounting dimensions
42 (1.65)

42 (1.65) 48 (1.89)

A0022210

7.1.3 Technical data

Display: 4-digit, red LED display

Digit height: 7.62 mm; programmable decimal point setting

Display range: -1999 to 9999

Accuracy: 0.2% of the span ±1 digit

Endress+Hauser 23
Operation options Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Electrical connection: To transmitter with 4 to 20 mA output and elbow plug DIN 43 650,
reverse polarity protection

Display power supply: Not required, powered automatically from the power loop

Voltage drop: ≤ 5 V (corresponds to load: max. 250 Ω)

Rate of conversion: 3 measurements per second

Damping: 0.3 to 20 s (configurable)

Data backup: Non-volatile EEPROM

Error message: • HI: overrange


• LO: below range

Programming: Via 2 keys, menu-guided, display range scaling, decimal point, damping,
error message

Degree of protection: IP 65

Influence of temperature on the 0.1% / 10 K


display:

Electromagnetic compatibility Interference emission as per EN 50081, interference immunity as per EN


(EMC): 50082

Permitted current load: Max. 60 mA

Ambient temperature: 0 to +60 °C (+32 to +140 °F)

Housing material: Pa6 GF30 plastic, blue


Front screen made from red PMMA

Order number: 52022914

7.1.4 Electrical connection

Pin assignment
LWARNING
Is the supply voltage switched off?
Risk of electric shock!
‣ Switch off the supply voltage before connecting the device.
• PIN 1: L+ (supply voltage UB)
• PIN 2: L- (0 V)
• PIN 3: Not assigned

Supply voltage
The supply voltage (mostly 24 V DC) must be greater than the sum of the voltage drop Us
at the sensor, the voltage drop of 5 V at the display and other voltage loss Ua (such as
additional evaluation and power loss).
The following therefore applies: Ub = Us + 5 V + Ua

Post-connection check

 Is the device or cable undamaged (visual check)?

 Are all cable glands installed, securely tightened and leak-tight?

 If supply voltage is present, is the device ready for operation and do values appear on the display module?

24 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Operation options

7.1.5 Commissioning
NOTICE
Risk of injury from the uncontrolled activation of processes!
‣ Ensure that no uncontrolled processes are activated in the system.

Configuring the menu items


To configure the menu items, loosen the 4 Phillips screws (1) on the display and remove
the cover.

1 1

A B
1 1

A0022209

A Move down and select the items


B Move up and select the items
A+B Select the item to configure or acknowledge the setting

Setting the decimal point

Press the B key until "dP" is displayed. d P

Press A+B to set the decimal point: - - -. -

Press B or A for up or down: - -. - -

Press A+B to quit the setting function and to go to the "dP" menu item. d P

Out of range setting


Message if signal is below 4 mA or above 20 mA:
• "HI" message = over range
• "LO" message = below range

Press the B key until "HILO" is displayed. H I L O

Press A+B for the setting function (message inactive): o F F

Press B or A for up or down (message active): o n

Press A+B to quit the setting function and to go to the "HILO" menu item. H I L O

Please note: If the "HILO" message is not active, the error "Er06" is displayed if the display
range (-1999 to +9999) is exceeded.

Changing to the measuring mode


Depending on the menu item selected, press key A or B between 1 and 8 times.

Endress+Hauser 25
Diagnostics and troubleshooting Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

8 Diagnostics and troubleshooting

8.1 Troubleshooting
If an illegal configuration exists in the device, the device switches to error mode.

General errors
Error Possible cause Solution

Device is not responding. Supply voltage does not match that specified on Apply correct voltage.
the nameplate.

Supply voltage has incorrect polarity. Reverse polarity of supply voltage.

Connecting cables are not in contact with the Check the contacting of the cables
terminals. and correct if necessary.

Output current Signal line is not wired correctly. Check wiring.


≤ 3.6 mA

8.2 Response of output to errors


The response of the output to error is regulated in accordance with NAMUR NE43.
Factory setting MAX alarm: >21 mA

8.2.1 alarm current

Device Description Option

PMC21 Adjusted min. alarm current IA 1)


PMP21
PMP23

1) Product Configurator order code for "Service"

8.3 Disposal
When disposing, separate and recycle the device components based on the materials.

9 Maintenance
No special maintenance work is required.
Keep the pressure compensation element (1) free from contamination.

A0022141

26 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Maintenance

9.1 Exterior cleaning


Please note the following points when cleaning the device:
• The cleaning agents used should not corrode the surface and the seals.
• Mechanical damage to the process isolating diaphragm, e.g. due to sharp objects, must
be avoided.
• Observe the degree of protection of the device. See the nameplate if necessary →  14.

Endress+Hauser 27
Repair Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

10 Repair

10.1 General notes

10.1.1 Repair concept


Repairs are not possible.

10.2 Return
The measuring device must be returned if the wrong device has been ordered or delivered.
As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged
to follow certain procedures when handling any returned products that have been in
contact with medium. To ensure swift, safe and professional device returns, please read
the return procedures and conditions on the Endress+Hauser website at
www.services.endress.com/return-material

28 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Accessories

11 Accessories

11.1 Weld-in adapter


Various weld-in adapters are available for installation in vessels or pipes.

Device Description Option 1) Order number

PMP23 Weld-in adapter M24, d=65, 316L PM 71041381

PMP23 Weld-in adapter M24, d=65, 316L 3.1 EN10204-3.1 PN 71041383


material, inspection certificate

PMP21 Weld-in adapter G½, 316L QA 52002643

PMP21 Weld-in adapter G½, 316L 3.1 EN10204-3.1 material, QB 52010172


inspection certificate

PMP21 Weld-in tool adapter G½, brass QC 52005082

PMP23 Weld-in adapter G1, 316L, conical metal joint QE 52005087

PMP23 Weld-in adapter G1, 316L, 3.1, conical metal joint, QF 52010171
EN10204-3.1 material, inspection certificate

PMP23 Weld-in tool adapter G1, brass QG 52005272

PMP23 Weld-in adapter G1, 316L, silicone O-ring seal QJ 52001051

PMP23 Weld-in adapter G1, 316L, 3.1, silicone O-ring seal, QK 52011896
EN10204-3.1 material, inspection certificate

PMP23 Weld-in adapter Uni D65, 316L QL 214880-0002

PMP23 Weld-in adapter Uni D65, 316L 3.1 EN10204-3.1 QM 52010174


material, inspection certificate

PMP23 Weld-in tool adapter Uni D65/D85, brass QN 71114210

PMP23 Weld-in adapter Uni D85, 316L QP 52006262

PMP23 Weld-in adapter Uni D85, 316L 3.1 EN10204-3.1 QR 52010173


material, inspection certificate

1) Product Configurator, order code for "Enclosed accessories"

If installed horizontally and weld-in adapters with a leakage hole are used, ensure that the
leakage hole is pointing down. This allows leaks to be detected as quickly as possible.

11.2 Process adapter M24


The following process adapters can be ordered for the process connections with order
option X2J and X3J:

Device Description Order number Order number with inspection certificate 3.1 EN10204

PMP23 Varivent F DN32 PN40 52023996 52024003

PMP23 Varivent N DN50 PN40 52023997 52024004

PMP23 DIN11851 DN40 52023999 52024006

PMP23 DIN11851 DN50 52023998 52024005

PMP23 SMS 1½" 52026997 52026999

PMP23 Clamp 1½" 52023994 52024001

PMP23 Clamp 2" 52023995 52024002

Endress+Hauser 29
Accessories Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

11.3 Plug-on display PHX20


→  23

11.4 M12 plug connectors


Connector Degree of protection Material Option 1) Order
number

M12 IP67 • Union nut: Cu Sn/Ni R1 52006263


(self-terminated connection • Body: PBT
at M12 plug) • Seal: NBR
(0.8)
ø20

53 (2.09)
A0024475

M12 90 degrees IP67 • Union nut: GD Zn/Ni RZ 52010285


with 5m (16 ft) cable • Body: PUR
• Cable: PVC
(1.1)
28

³40 (1.57)
A0024476

M12 90 degrees IP67 • Union nut: GD Zn/Ni RM 71114212


(self-terminated connection • Body: PBT
at M12 plug) • Seal: NBR

28
(1.1)
(1.38)
35

20 (0.79)
A0024478

M12 90 degrees IP69 2) • Union nut: 316L RW 52024216


with 5m (16 ft) cable (1.4435)
(terminated at one end) • Body and cable:
PVC and PUR
(1.1)
28

³40 (1.57)
A0024477

1) Product Configurator, order code for "Enclosed accessories"


2) Designation of the IP protection class according to DIN EN 60529. Previous designation "IP69K" according
to DIN 40050 Part 9 is no longer valid (standard withdrawn on November 1, 2012). The tests required by
both standards are identical.

30 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12 Technical data

12.1 Input

12.1.1 Measured variable

Measured process variable


• PMC11: gauge pressure
• PMP11: gauge pressure
• PMC21: gauge pressure or absolute pressure
• PMP21: gauge pressure or absolute pressure
• PMP23: gauge pressure or absolute pressure

Calculated process variable


Pressure

12.1.2 Measuring range

Ceramic process isolating diaphragm

Sensor Device Maximum Lowest MWP OPL Factory settings 2) Option 3)


Sensor measuring range calibratable
span 1)
lower (LRL) upper (URL)

[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]

Devices for gauge pressure measurement


4)
100 mbar (1.5 psi) PMC21 –0.1 (–1.5) +0.1 (+1.5) 0.02 (0.3) 2.7 (40.5) 4 (60) 0 to 100 mbar (0 to 1.5 psi) 1C
5)
250 mbar (4 psi) PMC21 –0.25 (–4) +0.25 (+4) 0.05 (1) 3.3 (49.5) 5 (75) 0 to 250 mbar (0 to 4 psi) 1E
6)
400 mbar (6 psi) PMC11 –0.4 (–6) +0.4 (+6) 0.08 (1.2) 5.3 (79.5) 8 (120) 0 to 400 mbar (0 to 6 psi) 1F
PMC21
6)
1 bar (15 psi) PMC11 –1 (–15) +1 (+15) 0.2 (3) 6.7 (100.5) 10 (150) 0 to 1 bar (0 to 15 psi) 1H
PMC21
6)
2 bar (30 psi) PMC11 –1 (–15) +2 (+30) 0.4 (0.6) 12 (180) 18 (270) 0 to 2 bar (0 to 30 psi) 1K
PMC21
6)
4 bar (60 psi) PMC11 –1 (–15) +4 (+60) 0.8 (1.2) 16.7 (250.5) 25 (375) 0 to 4 bar (0 to 60 psi) 1M
PMC21
6)
6 bar (90 psi) PMC11 –1 (–15) +6 (+90) 2.4 (36) 26.7 (400.5) 40 (600) 0 to 6 bar (0 to 90 psi) 1N
PMC21
6)
10 bar (150 psi) PMC11 –1 (–15) +10 (+150) 2 (30) 26.7 (400.5) 40 (600) 0 to 10 bar (0 to 150 psi) 1P
PMC21
6)
16 bar (240 psi) PMC11 –1 (–15) +16 (+240) 6.4 (96) 40 (600) 60 (900) 0 to 16 bar (0 to 240 psi) 1Q
PMC21
6)
25 bar (375 psi) PMC11 –1 (–15) +25 (+375) 10 (150) 40 (600) 60 (900) 0 to 25 bar (0 to 375 psi) 1R
PMC21
6)
40 bar (600 psi) PMC11 –1 (–15) +40 (+600) 8 (120) 40 (600) 60 (900) 0 to 40 bar (0 to 600 psi) 1S
PMC21

Endress+Hauser 31
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Sensor Device Maximum Lowest MWP OPL Factory settings 2) Option 3)


Sensor measuring range calibratable
span 1)
lower (LRL) upper (URL)

[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]

Devices for absolute pressure measurement

100 mbar (1.5 psi) 6) PMC21 0 +0.1 (+1.5) 0.1 (1.5) 2.7 (40.5) 4 (60) 0 to 100 mbar (0 to 1.5 psi) 2C
6)
250 mbar (4 psi) PMC21 0 +0.25 (+4) 0.25 (4) 3.3 (49.5) 5 (75) 0 to 250 mbar (0 to 4 psi) 2E
6)
400 mbar (6 psi) PMC21 0 +0.4 (+6) 0.4 (6) 5.3 (79.5) 8 (120) 0 to 400 mbar (0 to 6 psi) 2F
6)
1 bar (15 psi) PMC21 0 +1 (+15) 0.4 (6) 6.7 (100.5) 10 (150) 0 to 1 bar (0 to 15 psi) 2H
6)
2 bar (30 psi) PMC21 0 +2 (+30) 0.4 (0.6) 12 (180) 18 (270) 0 to 2 bar (0 to 30 psi) 2K
6)
4 bar (60 psi) PMC21 0 +4 (+60) 0.8 (1.2) 16.7 (250.5) 25 (375) 0 to 4 bar (0 to 60 psi) 2M
6)
10 bar (150 psi) PMC21 0 +10 (+150) 2 (30) 26.7 (400.5) 40 (600) 0 to 10 bar (0 to 150 psi) 2P
6)
40 bar (600 psi) PMC21 0 +40 (+600) 8 (120) 40 (600) 60 (900) 0 to 40 bar (0 to 600 psi) 2S

1) Highest turn down that can be set at the factory: 5:1. The turn down is preset and cannot be changed.
2) Other measuring ranges (e.g. –1 to +5 bar (–15 to 75 psi)) can be ordered with customer-specific settings (see the Product Configurator, order
code for "Calibration; Unit" option "J"). It is possible to invert the output signal (LRV = 20 mA; URV = 4 mA). Prerequisite: URV < LRV
3) Product Configurator, order code for "Sensor range"
4) Vacuum resistance: 0.7 bar (10.5 psi) abs
5) Vacuum resistance: 0.5 bar (7.5 psi) abs
6) Vacuum resistance: 0 bar (0 psi) abs

Maximum turn down which can be ordered for absolute pressure and gauge pressure
sensors
Devices for gauge pressure measurement
• 6 bar (90 psi), 16 bar (240 psi), 25 bar (375 psi): TD 1:1 to TD 2.5:1
• All other measuring ranges: TD 1:1 to TD 5:1
Devices for absolute pressure measurement
• 100 mbar (1.5 psi), 250 mbar (4 psi), 400 mbar (6 psi): TD 1:1
• 1 bar (15 psi): TD 1:1 to TD 2.5:1
• All other measuring ranges: TD 1:1 to TD 5:1

32 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

Metal process isolating diaphragm

Sensor Device Maximum Lowest MWP OPL Factory settings 2) Option 3)


Sensor measuring range calibratable
span 1)
lower (LRL) upper (URL)

[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]

Devices for gauge pressure measurement


4)
400 mbar (6 psi) PMP11 –0.4 (–6) +0.4 (+6) 0.4 (6) 1 (15) 1.6 (24) 0 to 400 mbar (0 to 6 psi) 1F
PMP21
PMP23

1 bar (15 psi) 4) PMP11 –1 (–15) +1 (+15) 0.4 (6) 2.7 (40.5) 4 (60) 0 to 1 bar (0 to 15 psi) 1H
PMP21
PMP23

2 bar (30 psi) 4) PMP11 –1 (–15) +2 (+30) 0.4 (6) 6.7 (100.5) 10 (150) 0 to 2 bar (0 to 30 psi) 1K
PMP21
PMP23

4 bar (60 psi) 4) PMP11 –1 (–15) +4 (+60) 0.8 (12) 10.7 (160.5) 16 (240) 0 to 4 bar (0 to 60 psi) 1M
PMP21
PMP23

6 bar (90 psi) 4) PMP11 –1 (–15) +6 (+90) 2.4 (36) 16 (240) 24 (360) 0 to 6 bar (0 to 90 psi) 1N
PMP21
PMP23

10 bar (150 psi) 4) PMP11 –1 (–15) +10 (+150) 2 (30) 25 (375) 40 (600) 0 to 10 bar (0 to 150 psi) 1P
PMP21
PMP23

16 bar (240 psi) 4) PMP11 –1 (–15) +16 (+240) 5 (75) 25 (375) 64 (960) 0 to 16 bar (0 to 240 psi) 1Q
PMP21
PMP23

25 bar (375 psi) 4) PMP11 –1 (–15) +25 (+375) 5 (75) 25 (375) 100 (1500) 0 to 25 bar (0 to 375 psi) 1R
PMP21
PMP23

40 bar (600 psi) 4) PMP11 –1 (–15) +40 (+600) 8 (120) 100 (1500) 160 (2400) 0 to 40 bar (0 to 600 psi) 1S
PMP21
PMP23

100 bar (1 500 psi) 4) PMP21 –1 (–15) +100 (+1500) 20 (300) 100 (1500) 160 (2400) 0 to 100 bar (0 to 1 500 psi) 1U
4)
400 bar (6 000 psi) PMP21 –1 (–15) +400 (+6000) 80 (1200) 400 (6000) 600 (9000) 0 to 400 bar (0 to 6 000 psi) 1W

Endress+Hauser 33
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Sensor Device Maximum Lowest MWP OPL Factory settings 2) Option 3)


Sensor measuring range calibratable
span 1)
lower (LRL) upper (URL)

[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]

Devices for absolute pressure measurement

400 mbar (6 psi) 4) PMP21 0 (0) 0.4 (+6) 0.4 (6) 1 (15) 1.6 (24) 0 to 400 mbar (0 to 6 psi) 2F
PMP23

1 bar (15 psi) 4) PMP21 0 (0) 1 (+15) 0.4 (6) 2.7 (40.5) 4 (60) 0 to 1 bar (0 to 15 psi) 2H
PMP23

2 bar (30 psi) 4) PMP21 0 (0) 2 (+30) 0.4 (6) 6.7 (100.5) 10 (150) 0 to 2 bar (0 to 30 psi) 2K
PMP23

4 bar (60 psi) 4) PMP21 0 (0) 4 (+60) 0.8 (12) 10.7 (160.5) 16 (240) 0 to 4 bar (0 to 60 psi) 2M
PMP23

10 bar (150 psi) 4) PMP21 0 (0) 10 (+150) 2 (30) 25 (375) 40 (600) 0 to 10 bar (0 to 150 psi) 2P
PMP23

40 bar (600 psi) 4) PMP21 0 (0) +40 (+600) 8 (120) 100 (1500) 160 (2400) 0 to 40 bar (0 to 600 psi) 2S
PMP23

100 bar (1 500 psi) 4) PMP21 0 (0) +100 (+1500) 20 (300) 100 (1500) 160 (2400) 0 to 100 bar (0 to 1 500 psi) 2U
4)
400 bar (6 000 psi) PMP21 0 (0) +400 (+6000) 80 (1200) 400 (6000) 600 (9000) 0 to 400 bar (0 to 6 000 psi) 2W

1) Highest turn down that can be set at the factory: 5:1. The turn down is preset and cannot be changed.
2) Other measuring ranges (e.g. –1 to +5 bar (–15 to 75 psi)) can be ordered with customer-specific settings (see the Product Configurator, order
code for "Calibration; Unit" option "J"). It is possible to invert the output signal (LRV = 20 mA; URV = 4 mA). Prerequisite: URV < LRV
3) Product Configurator, order code for "Sensor range"
4) Vacuum resistance: 0.01 bar (0.145 psi) abs

Maximum turn down which can be ordered for absolute pressure and gauge pressure
sensors

Device Range 400 mbar (6 psi) 1 bar (15 psi) 2 bar (30 psi)
6 bar (90 psi) 4 bar (60 psi)
16 bar (240 psi) 10 bar (150 psi)
25 to 400 bar (375 to 6 000 psi)
25 to 40 bar (375 to 600 psi)

PMP11 0.5% TD 1:1 TD 1:1 to TD 2.5:1 TD 1:1 to TD 5:1

PMP21 0.3% TD 1:1 TD 1:1 to TD 2.5:1 TD 1:1 to TD 5:1

PMP23 0.3% TD 1:1 TD 1:1 to TD 2.5:1 TD 1:1 to TD 5:1

34 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12.2 Output

12.2.1 Output signal

Description Option 1)

4 to 20 mA (2-wire) 1

PMC11: 0 to 10 V output (3-wire) 2


PMP11: 0 to 10 V output (3-wire)

1) Product Configurator, order code for "Output"

12.2.2 Signal range 4 to 20 mA


3.8 mA to 20.5 mA

12.2.3 Load (for 4 to 20 mA devices )


In order to guarantee sufficient terminal voltage in two-wire devices, a maximum load
resistance RL (including line resistance) must not be exceeded, depending on the supply
voltage UB of the supply unit.

R L max
[W ]
1068

614 2 R L max £ U B - 6.5V


22mA

10 20 30 UB
[V]
1

A0029452

1 Power supply 10 to 30 V DC
2 RLmax maximum load resistance
UB Supply voltage

12.2.4 Load resistance (for 0 to 10 V devices)


The load resistance must be ≥ 5 [kΩ].

12.2.5 Signal on alarm 4 to 20 mA


The response of the output to error is regulated in accordance with NAMUR NE43.
Factory setting MAX alarm: >21 mA

alarm current

Device Description Option

PMC21 Adjusted min. alarm current IA 1)


PMP21
PMP23

1) Product Configurator order code for "Service"

Endress+Hauser 35
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

12.2.6 Dead time, time constant


Presentation of the dead time and the time constant:

I
100 %
90 %

63 %

t1 t2 t
t3

A0019786

12.2.7 Dynamic behavior

Analog electronics

Dead time (t1) [ms] Time constant (T63), t2 [ms] Time constant (T90), t3 [ms]

6 ms 10 ms 15 ms

36 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12.3 Performance characteristics of ceramic process


isolating diaphragm

12.3.1 Reference operating conditions


• As per IEC 60770
• Ambient temperature TA = constant, in the range of:+21 to +33 °C (+70 to +91 °F)
• Humidity φ = constant, in the range of 5 to 80 % rH
• Ambient pressure pA = constant, in the range of:860 to 1 060 mbar (12.47 to 15.37 psi)
• Position of measuring cell = constant, in range: horizontal ±1° (see also "Influence of the
installation position" section →  16)
• Zero based span
• Material of process isolating diaphragm: Al2O3 (aluminum-oxide ceramic, Ceraphire®)
• Supply voltage: 24 V DC ±3 V DC
• Load: 320 Ω (at 4 to 20 mA output)

12.3.2 Measuring uncertainty for small absolute pressure measuring


ranges
The smallest extended uncertainty of measurement that can delivered by our
standards is:
• in range 1 to 30 mbar (0.0145 to 0.435 psi): 0.4 % of reading
• in range < 1 mbar (0.0145 psi): 1 % of reading.

12.3.3 Influence of the installation position


→  16

12.3.4 Resolution
Current output: min. 1.6 μA

12.3.5 Reference accuracy


The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the
pressure hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11]
in accordance with the limit point method as per [DIN EN 60770].

Device % of the calibrated span to the maximum turn down

Reference accuracy Non-linearity 1) Non-repeatability


2)
PMC11 ±0.5 ±0.1 ±0.1

PMC21 ±0.3 ±0.1 ±0.1

1) The non-linearity for the 40 bar (600 psi) sensor can be up to ± 0.15% of the calibrated span up to the
maximum turn down.
2) For devices with 0 to 10 V output, a non-linearity of up to max. 0.3 V can occur for signal values below
0.03 V.

Endress+Hauser 37
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Overview of the turn down ranges →  32

Measuring ranges Turn down Device % of URL

100 mbar (1.5 psi)to 1:1 to TD 5:1 PMC11 ±0.5


40 bar (600 psi)
PMC21 ±0.3 1)

1) For the 100 mbar (1.5 psi) and 250 mbar (4 psi) measuring ranges, the following applies: In the event of
heat effects on the initial reference conditions, an additional deviation of max. 0.3 mbar (4.5 psi) from the
zero point or the output span is possible.

12.3.6 Thermal change of the zero output and the output span

Measuring cell –20 to +85 °C (–4 to +185 °F) –40 to –20 °C (–40 to –4 °F)
+85 to +100 °C (+185 to +212 °F)

% of URL for TD 1:1

<1 bar (15 psi) <1 <1.2

≥ 1 bar (15 psi) <0.8 <1

12.3.7 Long-term stability

1 year 5 years 8 years

% of URL

±0.2 ±0.4 ±0.45

12.3.8 Switch-on time


≤2 s (For small measuring ranges, pay attention to the thermal compensation effects.)

38 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12.4 Performance characteristics of metallic process


isolating diaphragm

12.4.1 Reference operating conditions


• As per IEC 60770
• Ambient temperature TA = constant, in the range of:+21 to +33 °C (+70 to +91 °F)
• Humidity φ = constant, in the range of 5 to 80 % rH
• Ambient pressure pA = constant, in the range of:860 to 1 060 mbar (12.47 to 15.37 psi)
• Position of measuring cell = constant, in range: horizontal ±1° (see also "Influence of the
installation position" section →  16)
• Zero based span
• Process isolating diaphragm material: AISI 316L (1.4435)
• Filling oil: NSF-H1 synthetic oil in accordance with FDA 21 CFR 178.3570
• Supply voltage: 24 V DC ±3 V DC
• Load: 320 Ω (at 4 to 20 mA output)

12.4.2 Measuring uncertainty for small absolute pressure measuring


ranges
The smallest extended uncertainty of measurement that can delivered by our
standards is:
• in range 1 to 30 mbar (0.0145 to 0.435 psi): 0.4 % of reading
• in range < 1 mbar (0.0145 psi): 1 % of reading.

12.4.3 Influence of the installation position


→  16

12.4.4 Resolution
Current output: min. 1.6 μA

12.4.5 Reference accuracy


The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the
pressure hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11]
in accordance with the limit point method as per [DIN EN 60770].

Device % of the calibrated span to the maximum turn down

Reference accuracy Non-linearity Non-repeatability


1)
PMP11 ±0.5 ±0.1 ±0.1

PMP21 ±0.3 ±0.1 ±0.1

PMP23 ±0.3 ±0.1 ±0.1

1) For devices with 0 to 10 V output, a non-linearity of up to max. 0.3 V can occur for signal values below
0.015 V.

Overview of the turn down ranges →  34

Endress+Hauser 39
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

12.4.6 Thermal change of the zero output and the output span
PMP11, PMP21

Measuring cell –20 to +85 °C (–4 to +185 °F) –20 to –40 °C (–4 to –40 °F)
+85 to +100 °C (+185 to +212 °F)

% of the calibrated span for TD 1:1

<1 bar (15 psi) <1 <1.2

≥ 1 bar (15 psi) <0.8 <1

PMP23

Measuring cell –10 to +85 °C (+14 to +185 °F) +85 to +100 °C (+185 to +212 °F)

% of the calibrated span for TD 1:1

<1 bar (15 psi) <1 <1.2

≥ 1 bar (15 psi) <0.8 <1

12.4.7 Long-term stability

1 year 5 years 8 years

% of URL

±0.2 ±0.4 ±0.45

12.4.8 Switch-on time


≤2 s

40 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12.5 Environment

12.5.1 Ambient temperature range

Device Ambient temperature range 1)

PMC11 –40 to +70 °C (–40 to +158 °F)


PMP11

PMC21 –40 to +85 °C (–40 to +185 °F)


PMP21
PMP23

PMC21 Devices for hazardous areas: –40 to +70 °C (–40 to +158 °F)
PMP21
PMP23

1) Exception: the following cable is designed for an operating temperature range of


–25 to +70 °C (–13 to +158 °F): Product Configurator order code for "Accessory enclosed", option "RZ".

12.5.2 Storage temperature range


–40 to +85 °C (–40 to +185 °F)

12.5.3 Climate class

Device Climate class Note

PMP23 Class 4K4H Air temperature: –20 to +55 °C (–4 to +131 °F),
relative humidity: 4 to 100 %
satisfied according to DIN EN 60721-3-4 (condensation is possible)

PMC11 Class 3K5 Air temperature: –5 to +45 °C (+23 to +113 °F),


PMP11 relative humidity: 4 to 95 %
PMC21 satisfied according to IEC 721-3-3 (condensation not possible)
PMP21

12.5.4 Degree of protection

Device Connection Climate class Option 1)

PMP21 Cable5 m (16 ft) IP66/68 2) NEMA type 4X/6P enclosure A


PMP21
PMP23

PMP21 Cable10 m (33 ft) IP66/68 2) NEMA type 4X/6P enclosure B


PMP21
PMP23

PMP21 Cable25 m (82 ft) IP66/68 2) NEMA type 4X/6P enclosure C


PMP21
PMP23

PMC11 M12 plug IP65 NEMA type 4X enclosure L


PMP11

PMC21 M12 plug IP65/67 NEMA type 4X enclosure M


PMP21
PMP23

PMP23 M12 plug made of metal IP66/69 3) NEMA type 4X enclosure N

Endress+Hauser 41
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

Device Connection Climate class Option 1)

PMC11 Valve plug ISO4400 M16 IP65 NEMA type 4X enclosure U


PMP11
PMC21
PMP21
PMP23

PMC11 Valve plug ISO4400 NPT ½ IP65 NEMA type 4X enclosure V


PMP11
PMC21
PMP21
PMP23

1) Product Configurator, order code for "Electrical connection"


2) IP 68 (1.83m H2O for 24 h)
3) Designation of the IP protection class according to DIN EN 60529. Previous designation "IP69K" according
to DIN 40050 Part 9 is no longer valid (standard withdrawn on November 1, 2012). The tests required by
both standards are identical.

12.5.5 Vibration resistance

Test standard Vibration resistance

IEC 60068-2-64:2008 Guaranteed for 5 to 2000Hz: 0.05g2/Hz

12.5.6 Electromagnetic compatibility


• Interference emission as per EN 61326-1 equipment B
• Interference immunity as per EN 61326-1 (industrial environment)
• NAMUR recommendation EMC (NE21)
• Maximum deviation: 1.5% for TD 1:1
For more details, refer to the Declaration of Conformity.

42 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Technical data

12.6 Process

12.6.1 Process temperature range for devices with ceramic process


isolating diaphragm

Device Process temperature range

PMC11 –25 to +85 °C (–13 to +185 °F)

PMC21 –25 to +100 °C (–13 to +212 °F)

PMC21 for oxygen applications –10 to +60 °C (+14 to +140 °F)

• For saturated steam applications, use a device with a metal process isolating diaphragm,
or provide a siphon for temperature isolation when installing.
• Pay attention to the process temperature range of the seal. See also the following table.

Seal Notes Process temperature range Option

FKM - –20 to +100 °C (–4 to +212 °F) A 1)

FKM Cleaned for oxygen service –10 to +60 °C (+14 to +140 °F) A 1) and HB 2)

EPDM 70 - –25 to +100 °C (–13 to +212 °F) J 1)

1) Product Configurator, order code for "Seal"


2) Product Configurator, order code for "Service"

Applications with changes in temperature


Frequent extreme changes in temperatures can temporarily cause measuring errors.
Temperature compensation takes place after a few minutes. Internal temperature
compensation is faster the smaller the change in temperature and the longer the time
interval.
For further information please contact your local Endress+Hauser Sales Center.

12.6.2 Process temperature range for devices with metallic process


isolating diaphragm

Device Process temperature range

PMP11 –25 to +85 °C (–13 to +185 °F)

PMP21 –40 to +100 °C (–40 to +212 °F)

PMP23 –10 to +100 °C (+14 to +212 °F)

PMP23 At +135°C (+275 °F) for a maximum of one hour (device in operation but not within
Sterilization in place (SIP) measuring specification)

Applications with changes in temperature


Frequent extreme changes in temperatures can temporarily cause measuring errors.
Internal temperature compensation is faster the smaller the change in temperature and
the longer the time interval.
For further information please contact your local Endress+Hauser Sales Center.

Endress+Hauser 43
Technical data Cerabar PMC11, PMC21, PMP11, PMP21, PMP23

12.6.3 Pressure specifications


LWARNING
The maximum pressure for the measuring device depends on the lowest-rated
element with regard to pressure.
‣ For pressure specifications, see the "Measuring range" section and "Mechanical
construction" section in the Technical Information.
‣ The Pressure Equipment Directive (2014/68/EU) uses the abbreviation "PS". The
abbreviation "PS" corresponds to the MWP (maximum working pressure) of the
measuring device.
‣ MWP (maximum working pressure): The MWP (maximum working pressure) is
specified on the nameplate. This value is based on a reference temperature of
+20 °C (+68 °F) and may be applied to the device for an unlimited period of time.
Observe the temperature dependency of the MWP.
‣ OPL (over pressure limit): The test pressure corresponds to the over pressure limit of
the sensor and may only be applied temporarily to ensure that the measurement is
within the specifications and no permanent damage develops. In the case of sensor
range and process connections where the over pressure limit (OPL) of the process
connection is smaller than the nominal value of the sensor, the device is set at the
factory, at the very maximum, to the OPL value of the process connection. If you want
to use the entire sensor range, select a process connection with a higher OPL value.
‣ Oxygen applications: In oxygen applications, the values for "pmax and Tmax for oxygen
applications may not be exceeded .
‣ Devices with ceramic process isolating diaphragm: avoid steam hammering! Steam
hammering can cause zero point drifts. Recommendation: Residue (water droplets or
condensation) may remain on the process isolating diaphragm following CIP cleaning
and can result in local steam hammering the next time steam cleaning takes place. In
practice, drying the process isolating diaphragm (e.g. by blowing) has proved to prevent
steam hammering.

44 Endress+Hauser
Cerabar PMC11, PMC21, PMP11, PMP21, PMP23 Index

Index
A
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

C
CE mark (declaration of conformity) . . . . . . . . . . . . . 10
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

D
Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . 10
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

E
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

F
Field of application
Residual risks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

O
Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

P
Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

R
Repair concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

S
Safety instructions
Basic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . . 6
Staff
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

T
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

U
Use of the measuring device
see Designated use
Use of the measuring devices