Pen Type Multimeter User Manual
Pen Type Multimeter User Manual
Pen Type Multimeter User Manual
Pen Type
Multimeter
User Manual
Overview
This digital multimeter is designed and produced in accordance with the safety
requirements of the international electrical safety standard IEC-61010 for electronic
measuring instruments and handheld digital multimeters, and meets the 600V CAT IV
and 1000V CAT of IEC61010 III requirements.
Before using this instrument, please read the instruction manual carefully and pay
attention to the relevant safe work guidelines. If the meter is used near equipment
with relatively large electromagnetic interference, the meter's readings will become
unstable and may experience large errors. Do not use if the meter or test lead is
damaged. When working around bare conductors or buses, you must be extremely
careful. Do not use this meter near explosive gas, steam or dust. The input value
being measured must not exceed the input limit value specified in each range to
prevent damage to the meter. When the meter is connected to the line under test, do
not touch the unused input terminal. When the measured voltage exceeds 60VDC or
30VAC, please be careful to prevent electric shock. Before changing the range, make
sure that the meter is not connected to the circuit under test. Do not measure
resistance or continuity on a live circuit.
General operation
1.1 Button Function
1.2 Battery
This multimeter uses 2 x AAA 1.5V batteries for power supply. The batteries must be
correctly installed in the battery box of the meter. When the battery indicator
appears, replace the battery immediately.
Energy Saving
The meter will automatically enter power save mode after approx. 15 minutes of not
being used. You can reactivate the meter by pressing any button.
Battery Replacement
If the symbol appears on the screen, the batteries must be replaced. Failure to do
so will cause incorrect measurements. The meter requires 2 x AAA batteries.
Note: If the test probe on the meter is too short, the red insulating
shield can be removed by sliding it off in the direction of the arrow
shown in Fig.1
Fig.1
2.2 Measuring Resistance.
Note: When the meter measures large capacitance, it takes a certain time to stabilize
the reading.
1. Cut off the power supply and discharge all high voltage capacitors in the circuit
under test.
2. Turn the rotary switch to the NCV position.
3. Insert the black test pen into the COM input socket.
4. Connect the black test probe tip and the tip of the meter to the circuit under
test (in series).
5. Connect the power to the circuit, and then read the displayed reading.
6. If the display only shows "OL", it means that the input exceeds the selected
range. The maximum allowable current is 200mA.
GENERAL SPECIFICATIONS
ELECTRICAL SPECIFICATIONS
DC VOLTAGE
Range Accuracy Resolution
4V 1mV
400V 100mV
600V ± (1%+3d) 1V
AC VOLTAGE
Range Accuracy Resolution
4V ± (1%+3d) 1mV
40V 10mV
400V 100mV
600V ± (1.2%+5d) 1V
AC CURRENT
Range Accuracy Resolution
CAPACITANCE
Range Accuracy Resolution
RESISTANCE
Range Accuracy Resolution
400Ω 0.1Ω
4KΩ 1Ω
400KΩ 100Ω
4MΩ 1KΩ