WP21B WP21E Operation Manual PDF
WP21B WP21E Operation Manual PDF
WP21B WP21E Operation Manual PDF
Statement
This manual will help you understand the operation and maintenance
of the product better. It is reminded that the product shall be used
strictly complying with this manual. User’s operation failing to comply
with this manual may result in malfunction or accident for which Genrui
Biotech Inc. (hereinafter called Genrui) cannot be held liable.
The user shall understand that nothing in this manual grants him,
expressly or implicitly, any right or license to use any of the intellectual
properties of Genrui.
Genrui reserves the right to modify and update this manual without
I
prior notice.
Warning
This system can only be operated by test professionals, doctors, lab
workers or personnel for system maintenance and troubleshooting,
who have been trained by Genrui.
Make sure to use the instrument under the provisions of this Operation
Manual. Otherwise, the instrument may not work normally, the test
result is unreliable and it may damage the instrument parts and
endanger people’s safety.
II
Table of Contents
Chapter 1 Brief Introduction .............................................................. 1
2.1 Safety..................................................................................... 7
III
3.4 Packing list .......................................................................... 12
IV
6.1 QC review ............................................................................ 35
V
7.3 Preventive maintenance and inspection for safety purpose 64
Appendix ........................................................................................... 72
VI
WP21 Series Operation Manual Chapter 1 Brief Introduction
Basic performance
Deviation of
Within ± 3nm
filter accuracy
1
WP21 Series Operation Manual Chapter 1 Brief Introduction
Sample
≤ 1%
carryover rate
Intra-batch
clinical
ALT ≤ 5%, TP ≤ 2.5%, UREA ≤ 3.5%
precision
(CV%)
Basic parameters
Power ≤ 180VA
2
WP21 Series Operation Manual Chapter 1 Brief Introduction
Note:
The following symbols listed in the table below are applicable to this
manual.
Symbol Meaning
Biohazard!
The background color of this symbol is yellow, the
symbol itself and the outline is black.
3
WP21 Series Operation Manual Chapter 1 Brief Introduction
Symbol Meaning
The following screen printing and labels listed in the table below are
applicable to this manual.
Label Meaning
Production date
~ Alternating current
Power ON
4
WP21 Series Operation Manual Chapter 1 Brief Introduction
Label Meaning
Power OFF
Figures in this manual are only for illustration or examples, not for other
purposes. Subject to the real object while in operation, modification
may be made without prior notice.
3) Man-made damage;
5
WP21 Series Operation Manual Chapter 1 Brief Introduction
1.5 Training
In order to help users properly use the analyzer and give full play to its
performance, Genrui holds periodical free training for users throughout
the world, please visit Genrui website or contact the local offices to get
the training schedule.
Please carefully read this manual and the related reagents instructions,
if not operating in accordance with the provisions of this instrument
manual, it may damage the provided protection or greatly affect the
test results.
6
WP21 Series Operation Manual Chapter 2 Safety and Precautions
The following are the warning symbols related to the analyzer, if ignore
these symbols, the users may run the risk of death or serious injury.
The following warning symbols are not sorted by their importance. All
the symbols have the equivalent importance.
2.1 Safety
Electric Shock
High Temperature
Strong Light
7
WP21 Series Operation Manual Chapter 2 Safety and Precautions
Biohazard
2.2 Precautions
Scope of Application
Operators
8
WP21 Series Operation Manual Chapter 2 Safety and Precautions
Application Environment
Electromagnetic Interference
Liquid Leakage
Water Quality
9
WP21 Series Operation Manual Chapter 2 Safety and Precautions
Poor Grounding
Device Connection
Analysis Parameters
10
WP21 Series Operation Manual Chapter 3 Installation
Chapter 3 Installation
3.1 Instrument models
WP21B: Involving seven wavelengths, i.e. 340, 380, 405, 505, 546,
578 and 620nm; a type of semi-auto chemistry analyzer with touch
screen
WP21E: Involving six wavelengths, i.e. 340, 405, 505, 546, 578 and
620nm; a type of semi-auto chemistry analyzer without touch screen
Once you find any of the above issues, please do not open the
package and do not power on the instrument. Otherwise, it may lead to
electric shock. Please notify the local distributor immediately.
11
WP21 Series Operation Manual Chapter 3 Installation
If the outer package is intact, please open the packaging box and
check the items in the box at the presence of the distributor or Genrui
personnel:
Temperature 10℃-30℃
Atmospheric
70.0kPa-106.0kPa
pressure
12
WP21 Series Operation Manual Chapter 3 Installation
analyzer
2) If the articles in the box are correct, take out the instrument
carefully and put it on the stable and flat workbench.
3) Ground the grounding rod at the rear panel of this analyzer via an
earth wire specially designed for medical instruments (grounding
impedance < 0.1Ω).
4) Take out the power cable, plug one end into the power jack at the
rear panel and insert the other end into the grounded three-wire
socket.
13
WP21 Series Operation Manual Chapter 3 Installation
14
WP21 Series Operation Manual Chapter 4 Structure
Chapter 4 Structure
4.1 Front view of the instrument
Built-in thermal Display screen
printer
Sliding door at
the housing cover
Cuvettes
Keyboard
Figure 1
Figure 2
15
WP21 Series Operation Manual Chapter 4 Structure
Figure 3
16
WP21 Series Operation Manual Chapter 4 Structure
During routine work, operators mainly use Items 1 and 3; Item 1 can
conduct the sample testing and Item 3 can help you view or print a
report.
17
WP21 Series Operation Manual Chapter 5 Routine Operation
18
WP21 Series Operation Manual Chapter 5 Routine Operation
1) Startup: Turn on the power supply and enter the main interface
after the system self-checking. Preheat the analyzer for more
than 15 minutes.
Precautions:
1) This analyzer should keep away from moisture and sunlight and
also should be maintained periodically.
19
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-1
Enter the desired item number (e.g. ―1 ALT‖) to enter the measurement
interface of this item, as illustrated in Figure 5-2:
Figure 5-2
20
WP21 Series Operation Manual Chapter 5 Routine Operation
After the light-source lamp becomes stable, the system will open the
measurement interface, or you can press ―Skip‖ to skip the preheating
step and enter the measurement interface, as illustrated in Figure 5-3:
Figure 5-3
First, by pressing the number key ―1‖, the system will automatically
switch to the corresponding filter and test the original value of the
measurement channel without illumination (dark-current value), as
illustrated in Figure 5-4:
Figure 5-4
21
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-5
Figure 5-6
22
WP21 Series Operation Manual Chapter 5 Routine Operation
After testing the dark current, the system will automatically enter the
―Aspirate Cleaning Solution‖ interface, as illustrated in Figure 5-7:
Figure 5-7
23
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-8
Place distilled water onto the analyzer and press the number key ―1‖ to
confirm. Then, the instrument will begin to aspirate distilled water and
measure its ADC value, the screen is displayed as Figure 5-9:
Figure 5-9
24
WP21 Series Operation Manual Chapter 5 Routine Operation
Every time after the filter wheel rotates, the system will prompt you to
aspirate distilled water for calibration of absorbance zero-value. If the
difference between the original value and zero value of distilled water
is less than 1000, the system will display ―Abnormal Light Path‖. At this
time, please check the light path. If the difference between the original
value and zero value of distilled water is more than 1000 but less than
30,000, the system will display ―Abnormal ADC‖. At this time, please
check the ADC amplification factor and whether there are air bubbles
in the liquid lines.
When a test item is measured at the first time and its blank test has
been defined, the system will automatically enter the blank test
interface after completion of absorbance testing, and the words ―ASP
Blank‖ continuously flashes. The screen is displayed as Figure 5-10:
Figure 5-10
After placing the blank solution onto the analyzer, press the number
key ―1‖, the instrument will begin to aspirate the blank solution. After
the test, the absorbance value of the blank solution will display on the
screen and prompts you to remove the blank solution. The screen is
displayed as Figure 5-11:
25
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-11
Figure 5-12
After placing the standard solution onto the analyzer, press the number
key ―1‖, the instrument will begin to aspirate the standard solution. After
the test, the absorbance value of the standard solution will display on
the screen and prompts you to remove the standard solution. The
screen is displayed as Figure 5-13:
26
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-13
If the analyzer has tested the blank and standard solutions for this item,
then after measurement of the sample solution, the system will directly
enter into the testing interface for next sample, as illustrated in Figure
5-14. If this item is tested for the first time and its blank and standard
have been defined, then the blank and standard testing should be
performed before sample measurement.
Figure 5-14
27
WP21 Series Operation Manual Chapter 5 Routine Operation
After placing the sample solution onto the analyzer, press the number
key ―1‖, the instrument will begin to aspirate the sample solution. After
the test, the absorbance value of the sample solution will display on the
screen and prompts you to remove the sample solution. The test
results are displayed as shown in Figure 5-15, meanwhile, the system
will automatically measure the next patient sample under the same test
item. If you want to reduce the cross-contamination between samples,
place distilled water onto the analyzer, press the number key ―0‖, the
instrument begin to aspirate the distilled water, immediately remove it
after hearing the ―di-di‖ sound, the instrument will clean the liquid path
and drain the distilled water within the cuvette.
Figure 5-15
If the item is tested by the kinetic method, the screen will dynamically
display the reaction curve during measurement of the standard and
sample, as illustrated in Figure 5-16:
28
WP21 Series Operation Manual Chapter 5 Routine Operation
Figure 5-16
Figure 5-17
Press any key on the standby interface to return to the last interface
before standby.
29
WP21 Series Operation Manual Chapter 5 Routine Operation
5.3.7 Setup
In the ―Measure‖ menu, choose an item, after heating, you can set this
parameter. Press ―Set‖ or the number key ―2‖ on the keypad, the
screen is displayed as Figure 5-18:
Figure 5-18
In this interface, you can set the printing, edit a test item, sample or QC
and switch between different modes.
30
WP21 Series Operation Manual Chapter 5 Routine Operation
5.3.7.4 QC editing
Press the number key ―4‖ on the ―Set‖ interface to display a screen like
Figure 5-19:
Figure 5-19
In the ―Edit QC‖ interface, you can perform the following operations:
a) Item
Operation: The test item has been selected when entering the
measurement, so it cannot be changed in this ―Edit QC‖ interface.
b) LN
Operation: Press the number key ―1‖ to set the LN. The LN
number can be freely revised in the range of 1~99999999999999,
but repetition is not allowable. If you do not change it, the default
value is 0, which means no increase will occur.
c) Target value
31
WP21 Series Operation Manual Chapter 5 Routine Operation
Press the number key ―5‖ and display a screen like Figure 5-20:
Figure 5-20
5.3.8 Search
Press the number key ―3‖ to search the ―QC Review‖ or ―Result
Review‖ menu. See details in 6.1 or 6.2.3.
2) Press the number key ―1‖ on the ―Item Choice‖ window to choose
the ―ALT‖ item. Then, the system begins to preheat and will enter the
measurement interface after the temperature is stable. Enter the
number key ―1‖ to select ―Measure‖. When the system is ready, the
screen will display ―ASP Rinse‖. Place the cleaning solution onto the
analyzer and press ―1 Sample‖ to start the washing operation. After
completion, the screen displays ―ASP DI Water‖. Place the distilled
water onto the instrument and press ―1 Sample‖ to measure the
absorbance.
4) After aspirating the sample solution, the analyzer will display the
reaction curve on its screen during testing, as illustrated in Figure 5-21:
Figure 5-21
33
WP21 Series Operation Manual Chapter 5 Routine Operation
2) Press the number key ―5‖ on the ―Item Choice‖ window to choose
the ―TP‖ item. Then, the system begins to preheat and will enter the
measurement interface after the temperature is stable. Enter the
number key ―1‖ to select ―Measure‖. When the system is ready, the
screen will display ―ASP Rinse‖. Place the cleaning solution onto the
analyzer and press the key ―1‖ to start the washing operation.
4) At this time, the screen displays ―ASP Standard Solution‖ (if the
screen displays ―ASP Sample‖, please press the number key ―2‖ to
open the ―ASP Standard Solution‖ window). Place the standard
solution onto the instrument and press ―1‖ to perform measurement.
Calculate the factor after the testing is completed.
34
WP21 Series Operation Manual Chapter 6 Software Operation
Press the number key ―2‖ on the main menu and display a screen like
Figure 6-1:
Figure 6-1
a) To input a QC number
b) To search results
Press the number key ―2‖ to view the QC results. The screen is
35
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-2
c) Statistics
Press the number key ―3‖ on the ―QC Review‖ interface to display a
screen like Figure 6-3:
Figure 6-3
36
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-4
Press the number key ―1‖ on the ―Result‖ interface to display a screen
like Figure 6-5:
37
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-5
You can edit the information about patient samples on the sample
editing interface, and the specific operations are described as follows:
6.2.1.1 No.
If you need to measure another test item for the same patient, change
the current ―No.‖ to the original ―No.‖ of this patient.
6.2.1.2 ID
6.2.1.3 Sex
6.2.1.4 Age
38
WP21 Series Operation Manual Chapter 6 Software Operation
Press the number key ―2‖ on the ―Result Review‖ interface to add
results. The screens are displayed as Figures 6-6.1 and 6-6.2:
Figure 6-6.1
Figure 6-6.2
39
WP21 Series Operation Manual Chapter 6 Software Operation
1) No.
If you need to measure another test item for the same patient,
change the current ―No.‖ to the original ―No.‖ of this patient.
2) ID
3) Sex
4) Age
Operation: Press ―Set‖ or the number key ―1‖ to set the value.
5) Blood group
6) Item
Operation: Press ―Set‖ or the number key ―1‖ to set the item
7) Decimal digits
Options: 0, 1, 2, 3, 4, 5
8) Results
40
WP21 Series Operation Manual Chapter 6 Software Operation
9) Unit
Options: μ/L, μ/mL, μmol/L, mmol/L, g/L, mg/L, g/dL, g/mL, X10
12/dL, ABS
If you enter a patient number that does not exist, the system will
display ―This No. is not present‖. Then, the system will automatically
return to the ―Add Results‖ interface.
Press ―3‖ on the ―Result‖ interface to display a screen like Figure 6-7:
Figure 6-7
2) Entering the ID
41
WP21 Series Operation Manual Chapter 6 Software Operation
3) Printing
Operation: Press the number key ―1‖ to print all the results of the
currently displayed ―No.‖.
4) Searching
Operation: Press the number key ―2‖ to view all the results of the
currently displayed sample number. The screen is displayed as
Figure 6-8:
Figure 6-8
Press the number key ―1‖ to print all the current results; press ―Up‖
or ―Down‖ to view the last or next result.
Operation: Press the number key ―2‖ to print all the results of the
day's tests.
6) Exit
42
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-9
Press the up and down cursor keys to jump to the desired page.
Press the number key ―9‖ and display a screen like Figure 6-10:
Figure 6-10
You can set up the name of a new item and its parameters.
43
WP21 Series Operation Manual Chapter 6 Software Operation
44
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-11
1) Method
45
WP21 Series Operation Manual Chapter 6 Software Operation
2) Wavelength
Options: 1 or 2
3) Temperature
4) Aspiration volume
5) Decimal digits
Options: 0, 1, 2, 3, 4,
6) Unit
7) Reagent blank
Options: Yes, No
46
WP21 Series Operation Manual Chapter 6 Software Operation
8) Wavelength
Options: 340nm, 380nm, 405 nm, 505 nm, 546 nm, 578 nm,
620 nm
9) Sampling points
11) Correction
12) Calibration
Operation: Choose ―>‖ or ―<‖ using left and right cursor keys
and press the number key ―1‖ to set the value.
47
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-12
Figure 6-13
48
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-14
Press ―2‖ to cancel the changes and return to the ―Program‖ interface.
Directly press ―Menu‖ to exit the ―Program‖ interface. At this time, the
changes you made will not be saved.
49
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-15
Press the number key ―1‖ on the ―Setup‖ interface to display a screen
like Figure 6-16:
Figure 6-16
Press ―1‖ to select or cancel this option. When the option is checked,
the instrument will automatically print the result report after measuring
a patient’s sample.
Press ―2‖ to select or cancel this option. When the option is checked, if
you use the kinetics method to measure samples, the instrument will
automatically print the reaction curve after measuring a patient’s
50
WP21 Series Operation Manual Chapter 6 Software Operation
sample.
Press ―3‖ to select or cancel this option. When the option is checked,
absorbance values will be added into the result report after measuring
a patient’s sample.
In the system setup interface, press ―2‖ to open the ―Set Time‖ window.
The screen is displayed as Figure 6-17:
Figure 6-17
Press the left and right direction keys or directly click the left and right
keys on the screen to select the date and time needing to be changed.
After entering the desired values, press ―Enter‖ to save the changes or
press ―Exit‖ to cancel the changes.
In the system setup interface, press ―3‖ to open the ―COM‖ window.
The screen is displayed as Figure 6-18:
51
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-18
You can set the parameters for COM communication. Press left and
right keys to select the desired parameters.
You can select ―On‖ or ―Off‖ for COM status. If you select ―On‖, the
parameters will be displayed in the COM parameter-setup interface.
There are five options for baud rate, i.e. 9600, 19200, 38400, 57600
and 115200.
After determining the baud rate, you can select the check parameters;
for parity check, there are three options, including ―Odd‖, ―Even‖ and
―None‖.
In the system setup interface, press ―4‖ to open the ―Contrast‖ window.
The screen is displayed as Figure 6-19:
52
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-19
Adjust the backlight as needed by pressing the up, down, left and right
keys on the keyboard.
In the system setup interface, press ―5‖ to open the ―Cuvette‖ window.
The screen is displayed as Figure 6-20:
Figure 6-20
In this interface, you can switch between ―Flow Cell‖ and ―Cuvette‖, the
system default is ―Flow Cell‖. If you need to use cuvette to test samples,
press ―5‖ to switch to manual measurement mode.
53
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-21
In the ―Maintain‖ interface, press ―1‖ to open the ―TEMP ADJ‖ window.
The screen is displayed as Figure 6-22:
Figure 6-22
54
WP21 Series Operation Manual Chapter 6 Software Operation
Press the up, down, left and right keys on the keyboard to set the
parameters.
For ―Set TEMP‖, you have two options: 30℃ and 37℃.
In the ―Maintain‖ interface, press ―2‖ to open the ―Set ASP‖ window.
The screen is displayed as Figure 6-23:
Figure 6-23
By using this function, you can arbitrarily set the ―Clean VOL‖ and
―Sample VOL‖ in the range of 0~9999μL. The set cleaning volume
must ensure that the pipelines are clean after washing. The default
value of ―Clean VOL‖ is 1200μL and that of ―Sample VOL‖ is 1000μL,
which can be adjusted according to the actual situation.
55
WP21 Series Operation Manual Chapter 6 Software Operation
In the ―Maintain‖ interface, press ―3‖ to open the ―ASP ADJ‖ window.
The screen is displayed as Figure 6-24:
Figure 6-24
Enter the desired values in the fields of ―Set VOL‖ and ―Real VOL‖ and
get the aspiration volume determined in the ―Set VOL‖ field. Press
―Sample‖ to start aspiration and observe whether the aspirated volume
is accurate. If the sample after aspiration is exhausted, increase the
value in ―Real VOL‖; if the sample after aspiration remains a large
amount, decrease the value in ―Real VOL‖.
In the ―Maintain‖ interface, press ―4‖ to open the ―ABS‖ window. The
screen is displayed as Figure 6-25:
56
WP21 Series Operation Manual Chapter 6 Software Operation
57
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-25
In the ―Maintain‖ interface, press ―5‖ to open the ―ADC ADJ‖ window.
The screen is displayed as Figure 6-26:
58
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-26
Place distilled water onto the instrument and click ―Sample‖ or press
―1‖ to start aspiration.
In the ―Maintain‖ interface, press ―6‖ to open the ―Restore‖ window. The
screen is displayed as Figure 6-27:
59
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-27
1) To delete results
Press ―1‖ to delete all the patient results stored in the instrument.
2) To delete QC
3) To initialize items
Press ―3‖ to initialize the parameter settings of all the built-in items in
the instrument.
In the ―Maintain‖ interface, press ―7‖ to open the ―Soak‖ window. The
screen is displayed as Figure 6-28:
60
WP21 Series Operation Manual Chapter 6 Software Operation
Figure 6-28
Figure 6-29
61
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
2) Daily maintenance
3) Weekly maintenance
Wash the lines with distilled water several times and then aspirate the
cleaning solution once (main ingredient: sodium hypochlorite, diluted to
5%); after that, maintain the cleaning solution in the lines for ten
minutes and finally wash the lines with distilled water for three times.
Fill distilled water into the lines after washing. Neutral or alkaline
cleaning solutions are applicable to this instrument, but acidic cleaning
solution is prohibited.
Washing: Dip a piece of wet towel with cleaning solution to wipe the
analyzer tabletop, until all the stains are removed. It is recommended
to perform this operation once a day.
62
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
Note: Make sure to turn off the power switch and pull out the plug
before cleaning the analyzer housing. Please use clean water, neutral
cleansers or corrosion-free detergents for washing; do not use organic
solvents or hot water to avoid instrument damage due to deformity.
5) Waste disposal
Note: Make sure to turn off the power switch and pull out the plug
before cleaning the analyzer housing. Please use clean water,
neutral cleansers or corrosion-free detergents for washing. Do
not use organic solvents or hot water to avoid instrument damage
due to deformity.
63
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
4) When the expected service life of this instrument ends (about eight
years), it is recommended to stop using it or resume using after overall
troubleshooting, maintenance and care by Genrui.
5) This analyzer can only be used by personnel who are trained and
authorized by Genrui or its representative; otherwise, the protection
provided by the instrument may be impaired or the test results may be
significantly affected.
64
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
Parts: None
65
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
Fault
Cause Analysis Troubleshooting
Description
66
WP21 Series Operation Manual Chapter 7 Daily Use and Maintenance
Fault
Cause Analysis Troubleshooting
Description
B The cuvette is
unclean
C The light-source
lamp is ageing
If the instrument fault is still present, or there are other faults, please
contact our customer service division.
67
WP21 Series Operation Manual Chapter 8 Shipment and Storage
8.2 Storage
The packaged analyzer should be stored in a well-ventilated room that
has the environmental temperature of -10℃ to +55℃ and relative
humidity of 93% and is free from corrosive gases.
Note: The illustrations are for reference only, subject to pictures on the
outer package.
69
WP21 Series Operation Manual Chapter 9 Commonly-used Consumables
and Ordering Information
Ordering Information
Thermal
1 printing 80mm × 35mm
paper
Light-source
2 6V/10W, PHILIP, 6605, G4
lamp
Aspiration
4 φ1.5mm × φ0.8mm
tube
Connection
5 φ3mm × φ1.0mm
tube
7 Filter 340nm
8 Filter 380nm
70
WP21 Series Operation Manual Chapter 9 Commonly-used Consumables
and Ordering Information
9 Filter 405nm
10 Filter 505nm
11 Filter 546nm
12 Filter 578nm
13 Filter 620nm
14 Filter 700nm
Ordering method
71
WP21 Series Operation Manual Appendix
Appendix
Appendix 1 Replace printing paper
1. Open the printer cover and remove the roller (first take out the
terminal without gears and then the other terminal with gears;
please handle them gently), and take out the residual printing
paper.
2. Put the paper roll into the paper trough. Please note that there are
differences between the sides of thermal printing paper. Please
make sure the thermal side faces downward. Place the printing
paper below the roller, as illustrated in the figure below. Please
ensure that the movement direction of the printing paper should be
vertical with the roller orientation, thereby avoiding inability of
normal printing due to paper jam.
Non-thermal
Paper roller
side upward
3. Install the roller: First make the terminal with gears aligned and
then gently press down the other terminal.
4. Press the ―Feed‖ key, observe whether the paper moves smoothly
and the printer roller is vertical. Adjust it as needed. When a
segment of the printing paper moves out of the printer (passing the
printer cover), press the ―Feed‖ key again to stop it.
72
WP21 Series Operation Manual Appendix
5. Pull out the printing paper from the paper outlet at the printer cover
and close the cover.
Note: The “Feed” key is available on the main interface, all the
screens waiting for keyboard entering, the “Print/Delete Result
Reports” window in the “Result Report” interface and the QC
report interface. This function is disabled on all the other
interfaces to protect the printer.
1. Hold the connector of the pump tube with a hand, and then
remove the liquid inlet pipe and the waste liquid pipe.
5. Install the new tube as per the reverse steps of removing it. Let the
tube slide along the gap of the tube rack into the fixation trough.
73
WP21 Series Operation Manual Appendix
Peristaltic Housing
pump
tube
Waste
liquid
pipe
Bottom
2. Press ―6‖ on the main interface to open the ―Maintain‖ window and
then press ―5‖ to enter the ―ADC ADJ‖ interface.
74
WP21 Series Operation Manual Appendix
1 1 8 8
2 2 9 9
3 3 10 10
4 4 11 11
5 5 …… ……
6 6 …… ……
7 7 …… ……
75
WP21 Series Operation Manual Appendix
1. All the data sent is encoded with Code ASCII, e.g. using COM
tools to switch to the character format.
Normal mode
QC mode
Version number: 1
Instrument type: 1
No.: 00173
Item number: 1
76
WP21 Series Operation Manual Appendix
Unit: μ/mL
Range: 1.20--2.31
Item Item
Abbreviation Name Abbreviation Name
No. No.
Alanine Serum
1 ALT 24 Mg
transaminase magnesium
Alkaline
2 ALP 25 Cu Serum copper
phosphatase
Glutamic
3 GOT oxaloacetic 26 Zn Serum zinc
transaminase
γ-glutamyl Serum
4 γ-GT 27 P
transpeptidase phosphorus
Total iron-binding
7 TBIL Total bilirubin 30 TIBC
capacity
77
WP21 Series Operation Manual Appendix
Item Item
Abbreviation Name Abbreviation Name
No. No.
Albumin/
12 UA Uric acid 35 A/G
globulin ratio
Carbon-dioxide
13 CO2 36 UDBIL Indirect bilirubin
combining power
Immuno-
14 NH3 Blood ammonia 37 IGG
globulin G
Immuno-
15 GLU Blood glucose 38 IGM
globulin M
High density
17 HDLC 40 CHE Cholinesterase
lipoprotein
Low density
18 LDLC 41 CK Creatine kinase
lipoprotein
α-hydroxybutyric
21 AMS Amylase 44 α -HBDH
dehydrogenase
Monoamine
23 Ca Serum calcium 46 MAO
oxidase
Each test item should be tested with the corresponding reagent kit.
78
WP21 Series Operation Manual Appendix
The test combinations built-in this instrument and their respective test
items are as follows:
79
WP21 Series Operation Manual Appendix
Host shell ○ ○ ○ ○ ○ ○
Labels ○ ○ ○ ○ ○ ○
Closure assembly ○ ○ ○ ○ ○ ○
Accessories
Maintenance tools ○ ○ ○ ○ ○ ○
80
WP21 Series Operation Manual Appendix
Hazardous substances
Parts name
Pb Hg Cd Cr(VI) PBB PBDE
○: means the content of the hazardous substance in all homogeneous materials of the
part is in the limited requirement according to the standard of SJ/T 11363-2006.
×: means the content of the hazardous substance in at least one of the homogeneous
materials of the part is beyond the limited requirement according to the standard of SJ/T
11363-2006.
(1): some parts of the circuit board used lead solder during processing.
Notice: the product marked with ―×‖ is because there has no other technologies or parts
to be replaced at present stage, under normal use conditions, leak and mutation will not
occur in 5 years, and it will not cause environment pollution or harm to people and
property.
81
P01. 91. 300077- 04
Add: 6F, Shanshui BuildingB, Nanshan Yungu Innovation Industrial Park, 1183 Liuxian Blvd,
Nanshan District, 518055, Shenzhen, P.R. China
TEL: +86 755-2683 5560 FAX: +86 755-2667 8789
E- mail: [email protected] WEB: www.genrui-bio.com