1. Main Technical Parameter:
:
AD Switch Mode: ∑—∆,10times per second
Load Cell Excitation: DC +5V,can connect with 1-6pcs 350Ω strain gauge load cell.
lead-acid battery
: 0~40 。
Working Temperature:
II. :
Connection with Load Cell:
Assignment
① -Excitation
② -Feed Back
⑥ +Excitation
⑦ +Feed Back
⑧ -Signal
⑨ +Signal
⑤ Shield Wire
Short connect PIN 1 and 2, PIN 6 and 7 when connected a load cell with 4 wire cable;
!Connection between load cell and indicator must be reliable, shield wire must be
▲!
connected to ground reliably. Connection or disconnection are not allowed when the
indicator is on, which may damage the indicator or load cells.
1
Keli Electirc
▲ ! Static protection must be properly adopted as the load cell and indicator are all
III. Calibration
Turn on the indicator, after self check it will enter normal weiging display. Warm up for 15~30
minutes and open the calibration cover in the back side of the indicator. The calibration status
Back to Working Status 【 3.000】 After calibration, the current weight will be displayed
will be enterred after pressing the calibration butto. Following are the details:
After enter the calibration status and display 【CAL】, press “↑”and the indicator will display
【Zero】 and then press “↙” to enter setting for other calibration parameter. Following are the
details:
Steps Operation Display Note
Means other calibration parameter setting has enterred.
【Zero】
Press ”↙” for next step
“↑”to switch
1 【Zot *.*】 Zero Trace:0~4d
“↙”to confirm
“↑”to switch
2 【nt **】 :
Manual Set Zero:
“↙”to confirm
2
Keli Electirc
0,2,4,10,20,100 % of full capacity
:
Auto Set Zero:
turn off zero that manually set will be the turn on zero
when the indiator turn on next time and the auto zero
will not be carried out again.
Filter:
“↑”to switch
4 【FL ***】 【S t b】:Stable
“↙”to confirm
【SEn】
:Sensitive