XK3118T1 Technical Manual


1. Main Technical Parameter:


AD Switch Mode: ∑—∆,10times per second

Load Cell Excitation: DC +5V,can connect with 1-6pcs 350Ω strain gauge load cell.

Signal Input Range: -16mV~18mV;

Input Sensitivity: ≥1.5uV/e

:Adopts 6 wire type(Long wire auto compensation)


Load Cell Connection:

: AC 110~240V,50~60Hz;inbuilt 6V/4AH maintenance-free


Indicator Power Supply:

lead-acid battery

: 0~40 。
Working Temperature:

II. :
Connection with Load Cell:

Assignment

① -Excitation

② -Feed Back

⑥ +Excitation

⑦ +Feed Back

⑧ -Signal

⑨ +Signal

⑤ Shield Wire

Short connect PIN 1 and 2, PIN 6 and 7 when connected a load cell with 4 wire cable;

!Connection between load cell and indicator must be reliable, shield wire must be
▲!

connected to ground reliably. Connection or disconnection are not allowed when the
indicator is on, which may damage the indicator or load cells.

1
Keli Electirc
▲ ! Static protection must be properly adopted as the load cell and indicator are all

static sensitive equipments. Welding or other strong electricity operation should be


strictly forbidden. During thunderstorm season, proper lightening protection should be
tanken care of to protect the load cells and indicators from damaging by lightening and
to ensure the personal safety and the safely running of the weighing and related
equipments.

III. Calibration

Turn on the indicator, after self check it will enter normal weiging display. Warm up for 15~30
minutes and open the calibration cover in the back side of the indicator. The calibration status

Steps Operation Display Note


【 CAL 】 Means calibration status has enterred. Press ”↙” for next step
1 “↑”to select division 【E 01】 Divisions Setting:1、2、5、10、20、50
“↑”to select tadix point :Indicator directly shows the Radix Point
Radix Point Setting:
2 “↙” to confirm 【dC 0】

Full Capacity Setting Full Capacity Setting:


Press ”→” the flash digit will move towards right
3 【F030.00】
Press ”↑” to increase the number of flash digit
Press “ ” to confirm and enter next step such as 3000
Press for confirm after Zero Calibration:Make indicator under noload status, press “ ” after
4 date stable 【noLoAd】
date stable to confirm Zero
Input Loading Weight
【AdLoAd 】 Linearity Calibration:Load weight, the closer to the full capacity the
After display for better. Operation is same as full capacity setting.
2 seconds, auto
5 Press “↙” to confirm and automatically exit calibration status 5 seconds
switch to load
input after the stable indicator light on
【003.000】 Such as:3000

Back to Working Status 【 3.000】 After calibration, the current weight will be displayed
will be enterred after pressing the calibration butto. Following are the details:

IV. Setting for Other Calibraion Parameter

After enter the calibration status and display 【CAL】, press “↑”and the indicator will display

【Zero】 and then press “↙” to enter setting for other calibration parameter. Following are the

details:
Steps Operation Display Note
Means other calibration parameter setting has enterred.
【Zero】
Press ”↙” for next step
“↑”to switch
1 【Zot *.*】 Zero Trace:0~4d
“↙”to confirm
“↑”to switch
2 【nt **】 :
Manual Set Zero:
“↙”to confirm

2
Keli Electirc
0,2,4,10,20,100 % of full capacity


Auto Set Zero:

0,2,4,10,20,100 % of full capacity


“↑”to switch
3 【At **】
“↙”to confirm Select【--】for turn off zero saving function, i.e. the last

turn off zero that manually set will be the turn on zero
when the indiator turn on next time and the auto zero
will not be carried out again.

Filter:

“↑”to switch
4 【FL ***】 【S t b】:Stable
“↙”to confirm

【SEn】
:Sensitive