Anesthesia Machine
Anesthesia Machine
Anesthesia Machine
-1-
Anesthesia Machine with Ventilator AJ-2105
(2 Vaporizers, 3 Gas)
AJ-2105 Anesthesia Machine, with its ergonomic design, integrates a concise and
smooth structure and high mobility. The microcomputer-controlled Ventilator adopts
exported sensors and other components. Its scientific and proper design as well as
rigorous production process is what demonstrates the inner quality of Anjue products:
mechanical ventilation is made secure and visible by the folded respiratory bellows,
and the security of air supply is guaranteed by pneumatic transmission and
measurement system. With multiple modern technologies applied, the Anesthesia
Machine is both outstanding in performance and versatile in functions.
The Ventilator adopts microcomputer control with high-definition 8.4” TFT color
LCD display, integrating the functions of respiratory control and displaying the
monitoring parameters. It is versatile in function and excellent in performance.
The high-precision Vaporizer is designed with a stable concentration output,
automatic compensation for pressure, temperature and flow rate. And it is
available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and
-2-
Sevoflurane, ensuring a successful anesthesia operation.
High-definition five-tube flowmeter (oxygen, nitrous oxide, and air), designed with
stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the
Customer.
Modularized design and upgrading. Upgrading is available as required by the
Customer.
TECHNICAL PARAMETERS
Part of Anesthesia Machine
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter O2 : 0.1~1.0L/min 1.1~10L/min
N2O: 0.1~1.0L/min 1.1~10L/min
Air : 0.1~10L/min
■O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Part of Ventilator
■Display mode High-definition 8.4” TFT color LCD screen display
■Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL,
SIGH, IMV, VCV, PCV
■Ventilation parameters Tidal volume 20~1500 ml
Rate 1~99 bpm
SIMV rate 2~20 bpm
I:E 4:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
-3-
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
(70-150 adjustable)
Inspiratory platform 0~50%
■Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate,
SIMV rate, total respiratory rate, I/E, peak
pressure of airway, average pressure,
pressure - time waveform, flow rate - time
waveform, PEEP, inspiratory trigger pressure,
Inspiratory platform, autonomous inspiratory
rate, lung compliance, airway resistance,
flow-volume loop, pressure-volume loop
-4-
Anesthesia Machine with Ventilator AJ-2103
(2 Vaporizers, 2 Gas)
AJ-2103 Anesthesia Machine, with its ergonomic design, integrates a concise and
smooth structure and high mobility. The microcomputer-controlled Ventilator adopts
exported sensors and other components. Its scientific and proper design as well as
rigorous production process is what demonstrates the inner quality of Anjue products:
mechanical ventilation is made secure and visible by the folded respiratory bellows,
and the security of air supply is guaranteed by pneumatic transmission and
measurement system. With multiple modern technologies applied, the Anesthesia
Machine is both outstanding in performance and versatile in functions.
The Ventilator adopts microcomputer control with high-definition 5.7” LCD display,
integrating the functions of respiratory control and displaying the monitoring
-5-
parameters. It is versatile in function and excellent in performance.
The high-precision Vaporizer is designed with a stable concentration output,
automatic compensation for pressure, temperature and flow rate. And it is
available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and
Sevoflurane, ensuring a successful anesthesia operation.
High-definition four-tube flowmeter, designed with stoppers for oxygen and
nitrous oxide to control the output oxygen concentration.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the
Customer.
Modularized design and upgrading. Upgrading is available as required by the
Customer.
TECHNICAL PARAMETERS
Part of Anesthesia Machine
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter O2: 0.1~1.0L/min 1.1~10L/min
N2O: 0.1~1.0L/min 1.1~10L/min
■O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Part of Ventilator
■Display mode High-definition 5.7” LCD screen display
■Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
■Ventilation parameters Tidal volume 50~1500 ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
-6-
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
■Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate,
SIMV rate, total respiratory rate, I/E, peak
pressure of airway, pressure – time waveform,
flow rate - time waveform, PEEP, inspiratory
trigger pressure
■Monitoring of oxygen concentration (21%~100%)
■Security Alarm System
Oxygen Concentration alarm upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
■ Power AC 220V±10% 50Hz±1Hz
(UPS, Storage battery)
■(Mechanical Arms)
■(End-expiratory CO2 monitoring)
*Items within brackets are optional.
-7-
Anesthesia Machine with Ventilator AJ-2102
-8-
TECHNICAL PARAMETERS
Part of Anesthesia Machine
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
■Flow meter O2: 0.1~10L/min N2O: 0.1~10L/min
■O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off.
Part of Ventilator
■Display mode High-definition 5.7” LCD screen display
■Ventilation mode IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH
■Ventilation parameters Tidal volume 50~1500ml
Rate 2~99bmp
SIMV rate 2~20bpm
I:E 2:1~1:8
Inspiratory trigger pressure -1.0~2.0 kPa
PEEP 0~2.0 kPa
Pressure range 1.0~6.0 kPa
SIGH 1.5 times the inspiratory time
■Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate,
SIMV rate, total respiratory rate, I/E, peak
pressure of airway, pressure - time waveform,
flow rate - time waveform, PEEP, inspiratory
trigger pressure
■Monitoring of oxygen concentration (21%~100%)
■Security Alarm System
Oxygen Concentration alarm upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
-9-
low limit setting range 0.4~2.0 kPa
Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input for 15 seconds
Power alarm
■ Power AC 220V±10% 50Hz±1Hz
UPS, Storage battery)
*Items within brackets are optional.
- 10 -
Anesthesia Machine with Ventilator AJ-2101A
TECHNICAL PARAMETERS
Part of Anesthesia
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2, N 2O: 0.27~0.55MPa
■Flow meter O2: 0.1~1.0L/min 1.1~10L/min
■Flow rate of rapid oxygen supply 25~75L/min
- 11 -
■Low oxygen pressure alarm There will be sound alarm when the oxygen
pressure < 0.2MPa
■Vaporizer It has the function of automatic compensation
based on pressure, temperature, and flow rate.
The regulation range of evaporator concentration
is 0~5 vol%.
Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for
application as required by the customer.
Part of Ventilator
■Display mode High-definition 4.8” LCD screen display
■Ventilation mode IPPV, MANUAL (Manual monitoring of tidal
volume, ventilation volume, respiratory rate)
■Ventilation parameters Tidal volume 50~1500ml
Rate 4~80 bpm
I:E 2:1~1:4
Pressure range 1.0~6.0 kPa
■Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, I/E,
peak pressure of airway, inspiratory trigger
pressure, pressure - time waveform
■Security Alarm System
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
TV alarm upper limit setting range 0.5~2.0L
low limit setting range 0.02~1.0L
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input in 15s
Power alarm
■Power AC 220V±10% 50Hz±1Hz
- 12 -
Anesthesia Machine AJ-2100
TECHNICAL PARAMETERS
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2, N 2O: 0.27~0.55MPa
■Flow meter O2: 0.1~1.0L/min 1.1~10L/min
■Flow rate of rapid oxygen supply 25~75L/min
■Vaporizer Among Halothane, Enflurane, Isoflurane and
Sevoflurane, one can be chosen for application
as required by the customer.
■Respiratory circuit working mode: all-close, semi-close, semi-open
releasing pressure: 1~4 kPa
■Respiratory bellows Bellows for adults, bellows for children
tidal volume range:0~1500 ml
- 13 -