Da550012 Ba 2000
Da550012 Ba 2000
Da550012 Ba 2000
DA550012 Buhler Technologies LLC, 1030 West Hamlin Road, Rochester Hills, MI 48309
Phone: 248.652.1546, Fax: 248.652.1598
02/2020
e-mail: [email protected]
page 1 / 6 Internet: www.buhlertech.com
BA 2000
Description
The injector built into the filter housing constantly supplies the ZrO2 sensor with fresh process gas. The self-regulating probe
part is heated to 180 °C (356 °F) to prevent condensation. The ZrO2 sensor in the BA 2000 provides accurate, extremely fast mea-
surements.
No reference gas needed for operation. The sensor's 1-point calibration uses instrument air also needed to operate the injector. If
necessary, 2-point calibration may also be performed. The test gas additionally needed for calibration in this case should ideally
correspond with the O2 concentration of the sample gas.
A filter built into the probe protects the measuring cell from dust exposure. The filter can be changed in seconds without tools
by simply turning the handle 90°.
The filter elements are available in ceramic, sintered or star-pleated stainless steel.
Together with the filters listed, the BA 2000 can be used for gases with a dust load of up to approx. 2 g/m³.
The BA 2000 will provide all the information required for safe operation. The controller features a display with key pad for enter-
ing commands, alarm output, calibration function and 4 - 20 mA output signal.
t90 times depending on volume flow and pressure
t 90 [sec]
t 90 [sec]
t 90 [sec]
U Cell voltage in V
R Universal gas constant, R = 8.31447 J/(molּK)
po2 , air T Measuring temperature in K
(l) U = RT In
4F po2 , sample gas F Faraday-constant, F = 96485.34 C/mol
po2 , air Partial pressure of oxygen at the reference electrode in dry air in Pa
po2 , Sample gas Partial pressure of oxygen at the reference electrode in dry air in Pa
The conductivity of the oxide ions of zirconium oxide increases exponentially with the temperature and reaches adequate val-
ues above 600 °C (1112 °F).
Provided the total pressures of the gases are about equal on both electrodes (in this case volume concentrations can be used in
place of partial pressures), after adding the numeric values for the constants in equation (I) results in the following equation for
the oxygen concentration.
φ o2 Oxygen concentration in sample gas in Vol.-%
U
(-46,42 . ) U Potential difference in mV
(lI) φ o2 = 20,9 . e T
T Measuring temperature in K
20,9 Oxygen concentration in dry air in Vol.-%
The BA 2000 uses a potentiometric cell. The reference and the sample gas electrode are located in two different gas chambers
with different oxygen partial pressure. The two chambers are separated by the gas-tight ZrO2 tube. The electrodes generate
e.m.f. (electromotive force) proportional to the partial pressure difference of the oxygen. The NERNST equation applies.
2 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020
BA 2000
Technical Data
Ordering instructions
Adapter flanges
DA550012 ◦ 02/2020 We reserve the right to amend specification. Buhler Technologies LLC 3
BA 2000
Drawings
4 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020
BA 2000
DA550012 ◦ 02/2020 We reserve the right to amend specification. Buhler Technologies LLC 5
BA 2000
6 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020