Da550012 Ba 2000

Download as pdf or txt
Download as pdf or txt
You are on page 1of 6

Gas Analysis

Oxygen Flue Gas Analyzer BA 2000

Some combustion processes, e.g. process heaters, steam


Fast response time
boilers or heating furnaces the air required to achieve optimal
system efficiency can easily fluctuate. And the economic
Tool-less filter change
optimum of the process fluctuates within a relatively narrow
range. Both elevated NOx or SOx emission due to an excess
Easy handling
supply of air (excess O2) as well as energy loss within the
system due to incomplete combustion (lack of O2) require Flue gas temperatures up to 1600 °C (2912 °F)
measuring the oxygen level in the flue gas of the combustion
process. Sampling near the combustion chamber is therefore Durable ZrO2 measuring cell
just as vital as using a rapid response sensor to allow for
responding to changes in the combustion gas temperature Display includes O2
and/or other variables in the combustion process promptly.
The BA 2000 was developed specifically for this application. 4-20 mA output signal

Ambient temperature -20 to +70 °C (-4 to 158 °F)

No reference gas required

No test gas required

No gas conditioning required

Calibration with instrument air

DA550012 Buhler Technologies LLC, 1030 West Hamlin Road, Rochester Hills, MI 48309
Phone: 248.652.1546, Fax: 248.652.1598
02/2020
e-mail: [email protected]
page 1 / 6 Internet: www.buhlertech.com
BA 2000
Description
The injector built into the filter housing constantly supplies the ZrO2 sensor with fresh process gas. The self-regulating probe
part is heated to 180 °C (356 °F) to prevent condensation. The ZrO2 sensor in the BA 2000 provides accurate, extremely fast mea-
surements.
No reference gas needed for operation. The sensor's 1-point calibration uses instrument air also needed to operate the injector. If
necessary, 2-point calibration may also be performed. The test gas additionally needed for calibration in this case should ideally
correspond with the O2 concentration of the sample gas.
A filter built into the probe protects the measuring cell from dust exposure. The filter can be changed in seconds without tools
by simply turning the handle 90°.
The filter elements are available in ceramic, sintered or star-pleated stainless steel.
Together with the filters listed, the BA 2000 can be used for gases with a dust load of up to approx. 2 g/m³.
The BA 2000 will provide all the information required for safe operation. The controller features a display with key pad for enter-
ing commands, alarm output, calibration function and 4 - 20 mA output signal.
t90 times depending on volume flow and pressure

t 90 [sec]
t 90 [sec]
t 90 [sec]

Instrument air pressure [bar]


Instrument air volume flow [L/h] Sample gas volume flow [L/h] at a process pressure of 1 bar (14.5 psi) absolute

Oxygen measurement principle with ZrO2 cells


The NERNST equation serves as the basis for determining the oxygen concentration in gases by zirconium dioxide measuring
cell.

U Cell voltage in V
R Universal gas constant, R = 8.31447 J/(molּK)
po2 , air T Measuring temperature in K
(l) U = RT In
4F po2 , sample gas F Faraday-constant, F = 96485.34 C/mol
po2 , air Partial pressure of oxygen at the reference electrode in dry air in Pa
po2 , Sample gas Partial pressure of oxygen at the reference electrode in dry air in Pa

The conductivity of the oxide ions of zirconium oxide increases exponentially with the temperature and reaches adequate val-
ues above 600 °C (1112 °F).
Provided the total pressures of the gases are about equal on both electrodes (in this case volume concentrations can be used in
place of partial pressures), after adding the numeric values for the constants in equation (I) results in the following equation for
the oxygen concentration.
φ o2 Oxygen concentration in sample gas in Vol.-%
U
(-46,42 . ) U Potential difference in mV
(lI) φ o2 = 20,9 . e T
T Measuring temperature in K
20,9 Oxygen concentration in dry air in Vol.-%

The BA 2000 uses a potentiometric cell. The reference and the sample gas electrode are located in two different gas chambers
with different oxygen partial pressure. The two chambers are separated by the gas-tight ZrO2 tube. The electrodes generate
e.m.f. (electromotive force) proportional to the partial pressure difference of the oxygen. The NERNST equation applies.

2 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020
BA 2000
Technical Data

BA 2000 Technical Data


Sampling tube length: 0.5...2 m (1.6...6.6 ft)
Voltage: 115 or 230 V, 50/60 Hz
Probe heat output: 400 W
Measuring range: 0.1 to 21 Vol.-% O2
Output signal: 4-20 mA = 0-21 Vol.-% O2 (scalable 0-2.5/0-5/0-10/0-15)
Accuracy: relative error < 5 %
Sensor T90 time: < 15 sec
Alarm Sensor: Upper and lower limit of nominal value for heating (fixed)
Upper and lower limit of O2 concentration (adjustable)
Probe alarm: Insufficient temperature
Ambient temperature: -20 … +70 °C (-4 … 158 °F)
Process temperature: up to 1600 °C (2912 °F), depending on sampling tube
Probe operating temperature: max. 200 °C (392 °F)
Probe material: 1.4571
Test gas 1-point calibration: Instrument air 20.9 Vol.-% O2
Test gases 2-point calibration: Instrument air 20.9 Vol.-% O2 and
test gas 0.1 to 15 Vol.-% O2

Ordering instructions

Item no. Description


55200099 BA 2000, 230 V 50/60Hz
55201099 BA 2000-MF, 230 V 50/60Hz
55202099 BA 2000-SE, 230 V 50/60Hz
55200098 BA 2000, 115 V 50/60Hz
55201098 BA 2000-MF, 115 V 50/60Hz
55202098 BA 2000-SE, 115 V 50/60Hz
55200098I BA 2000I, 115 V 50/60Hz, US sized
55201098I BA 2000I-MF, 115 V 50/60Hz, US sized
55202098I BA 2000I-SE, 115 V 50/60Hz, US sized
55200099I BA 2000I, 230 V 50/60Hz, US sized
55201099I BA 2000I-MF, 230 V 50/60Hz, US sized
55202099I BA 2000I-SE, 230 V 50/60Hz, US sized
MF = separate sample gas recovery
SE = separated electronics up to approx. 15 m (49.2 ft)

Adapter flanges

Item no. Description


55200001 Adapter flange DN65 PN6 to Servomex
55200002 Adapter flange DN65 PN6 to Thermox
55200001I Adapter flange DN3-150 to Servomex
55200002I Adapter flange DN3-150 to Thermox

DA550012 ◦ 02/2020 We reserve the right to amend specification. Buhler Technologies LLC 3
BA 2000
Drawings

4 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020
BA 2000

DA550012 ◦ 02/2020 We reserve the right to amend specification. Buhler Technologies LLC 5
BA 2000

6 Buhler Technologies LLC We reserve the right to amend specification. DA550012 ◦ 02/2020

You might also like