User Manual Autoclave Ha-Bav150
User Manual Autoclave Ha-Bav150
User Manual Autoclave Ha-Bav150
Computer control
USER’S MANUAL
Alarm:
1. Please turn on the water inlet valve before using the unit, to ensure the correct using of the
vacuum pump.
2. While turning on the power, the VACUUM PROTECTION lamp is on, which means the
wrong wire connection of the electrical wires.
1. Connect the power, turn on the water resource valve, then turn on the power control switch,
the full screen would show 8, after 10 seconds, the screen would show the jacket temperature
and sterilizing room temperature, the time is 0.
2. Load the sterilizing objects, and lock tightly of the sterilizing room. The screen would show
the door lamp and low water indicating lamp is on.
3. SET:
I. Press SET key, the sterilizing screen would show P01, and set the sterilizing temperature
ranged from 126~134.
II.. Press SET key, the sterilizing screen would show P02, the vacuum times( 3 times), if the
vacuum times is zero, then press the SET key to set the exchange time(10 mins).
III. Press the SET key, the sterilizing screen would show PO3, set the sterilizing time
(20~40mins).
IV. Press the SET key, the sterilizing screen would show PO4, set the drying time (0~99mins).
If no need, then the drying time would set “0”.
VI. Press the SET key, the screen would show the initial figures, and at the meanwhile the
parameter is saved.
1
4. Press the confirm key to start. (Then after the above set, the operator could directly start the
operation from the first step). While the low water indicating lamp and the water inlet lamp is
off, the door lamp and the high water indicating lamp will be on.
5.1 Vacuum procedure: By the sterilizing is finished, it starts to proceed the drying. After the
end of the procedure, the door can not be open will the inner chamber pressure is back to 0. And
after open the door, press Print key to obtain the sterilizing record.
5.2 Non Vacuum procedure: By the sterilizing is finished, it would automatically exhaus
t the steam automatically, it MUST wait the inner chamber pressure is back to 0 and t
he chamber temperature is below 105°C, then proceed the drying. After the drying is fi
nished the unit would beep to alarm, then press the Print key to obtain the sterilizing r
ecord.
6. please read the user’s manual carefully before using the machine.
Attention:
1. Only the gasket temperature is below 105°C, then the unit can be continue using.
2. Should release the water every day after using, and shut of the drain valve every on
e day.
3. Adjust the time: Press the SET key, turn on the power switch, and press the c
onfirm key to adjust the Year, Month, Day and Hour.
Alarm: To ensure the safety and correctly using of the unit, please read the users’
manual carefully, and each operation step should follow the manual said. It would cause
damage caused by un-proper using.
2
USERS’ MANUAL
1. Synopsis
WS series horizontal pressure steam sterilizer is the equipment that disinfects and sterilizes the
objects with saturated steam, its structure is horizontal and heating way is: electric heating and
direct steam heating. It is composed of sterilization room, jacket, shell, door, steam generator,
pipe system, control system, indicating system and power-supply system.
2.Application Scope
The product is applicable to disinfect and sterilize medical devices, dressings, glass wares and
hydroponics medium etc in the steam method.
4. Basic Parameters
Basic Parameters of Sterilizer (Table 1)
code Name HA-BAV150
4 Heat evenness ≤1
5 Temp.range 40~134℃
3
7 Dry time range 1min~99min
13
10 11 12
Fig. 1
4
1. Jacket gauge 2. Chamber gauge 3. Shell 4. LED indicatin 5. Printer 6. Door
7. Power switch 8.control system 9. Circuit breaker 10. Water inlet 11.water
releasing
12. Air exhausting 13.Emergency valve
Fig. 2
Nixie Tube
The upper end of the display panel is installing three nixie tubes to display data.
Display contents in correct order are:jacket temperature, sterilization room temperature and
time.
Indicator Light
To be used for displaying the input and output and of working process. From left to right in
correct order are as below:
Door------door indicator light; light on means the door has been closed.
High water----the water level of the steam generator is at the high level.
Low water----the water level of the steam generator is at the low level.
Water inlet----water inlet indicating light, means the water inlet the generator.
Pump protection----pump protection light, means the water supply is stopped in the vacuum
pump. (if this lamp is on together with the power lamp, it means the wrong phase protection)
Negative pressure---the negative pressure indicating lamp, it will light on by proceeding the
vacuum.
Heating------Heating indicator light, light on means electric heating pipe is in the holmic
heating state.
Sterilization------Sterilization indicator light, light on means sterilization cycle has been in the
5
sterilization state.
Drying------Drying indicator light, light on means sterilization cycle has been in the
sterilization state.
Starting------operation indicator light; light on means the operation stroke has been started, it is
in sterilization cycle.
Water Inlet------Water inlet indicator light; light on means it is inletting water inside
evaporator.
Slow Exhaust------Slow exhaust indicator light, light on means slow exhaust valve is opened to
exhaust steam, which is mainly used for drainage of sterilization room.
Fast Exhaust------Fast exhaust indicator light, light on means fast exhaust valve is opened to
exhaust steam, which is mainly used for freezing air replacement and exhaust steam.
Full Exhaust-----Full exhaust indicator light, light on means the sterilization room and
evaporator are in the exhausting steam in the fault state.
Water break------Water break indicator light, light on means evaporator is in low water level
state.
End ------the sterilizing cycle end then this lamp will be on.
Print------Print indicator light, light on means the printer is printing the sterilization record
which has been saved. The sterilization record contains sterilization parameter and time and
date before the sterilization is finished. During the period of indicator light on, press printer key
again is in vain.
6
▲ ----- Increase key
It is used to set the increase of sterilization parameter value.
▼ ------- Decrease key
It is used to set the decrease of sterilization parameter value.
Starting------Starting key
It is used to start the working stroke and set the color system time in control system.
Exhaust Steam------Exhaust steam key
It is a press-key used to reset the control system when it occurs abnormal situation during
working process. After the system is reset, the sterilization room will exhaust steam.
Print ------Print key
It is used to print the parameters of sterilization record process in working.
7
g) It is installed with high water level device, after pressing starting button, the inlet indicator
light is on, the water supply will enter into water tank automatically, and it will heat up
immediately after reaching high water level, meanwhile the indicator light is off.
h) It has automatic control functions:
◆If the door of sterilization room is not closed tightly and it can not sterilize, prompting the
door indicator light is off, now we can conduct drying operation.
◆If the door is opened by error in the starting process of sterilization room, the electromagnetic
valve which enters into sterilization room will be closed automatically, and the door will be
mechanical self-locked ≥0.027MPa.
◆If starting without selecting to set sterilization time, the buzzer will be alarmed to prompt.
◆It is installed with low water level protective device, when the water level is lower than the
requirement and lack of water in the entire process, it can cut off the heating power supply
automatically and meanwhile, water break indicator light is on, heating and sterilization
indicating is off and the buzzer is alarmed and the steam is exhausted automatically.
◆It is installed with overload leakage protective device, when it occurs electrical leakage or
over-current, it can cut off the power supply automatically and the indicator light of electric
appliance control panel is off.
◆It is installed with temperature controlling failure protective device, the fine-open spring
safety valve is fixed on the top of sterilizer, which can open automatically and release
over-pressure steam, when the steam pressure in the jacket exceeds the maximum permissible
range and will close automatically when the steam pressure reduces to normal. It plays a very
important role in ensuring safe application; the operation pressure range of safety valve is
0.24~0.25MPa.See Table 2.
i) It is installed with manual emergency switch valve, when the equipment is switched off
suddenly and need to open the door to fetch the objects, it is allowed to open the manual ball
valve on the back of the equipment manually, releasing the pressure of sterilizer and open the
door to fetch objects after pressure gauge returns to zero.
j) It is installed with manual maintenance switch valve, when the evaporator of the equipment
needs cleaning and draining, open the manual valve in the left bottom side of the equipment to
exhaust the waste water from pressure box.
k) Installed with electromagnetic valve:
◆The inlet valve; press start key, the program will be closed after it is inlet water to high water
level automatically.
8
◆Drainage valve; electrify and press starting key, the programme will be opened automatically
at the temperature ≤108℃ when it is exhausting air, heating and sterilizing automatically, at the
temperature is ≥108℃, it will be closed automatically.
◆Exhaust steam valve; electrify and press starting key, the programme will be closed after it
finishes, exhausting air and it inlets water to high water level. When sterilization is completed
and press emergency stop key, the programme exhausts steam and open automatically.
◆Sterilization valve; electrify and press starting key, the program inlets steam automatically;
when water inlet reaches to the high water level, the door is closed tightly, vice versa, the door
is opened.
◆vacuum valve: it will start to work automatically.
Fig.3
7.5 After installing the equipment, inspect the equipment completely once again, after
confirmation without fault, it can start trial run. Refer to operation method, open the manual
water inlet valve, observe the state of water source pressure gauge and examine whether the
pressure is in the required pressure range. Switch on power supply and operate trial run
referring to operation method.
7.6 It has been inspected by rising temperature before the products leave factory, but after it is
installed and ready, the user shall perform trail run according to user’ manual; if there is any
problem, please inquire the reasons and solve it, it can’t be put into operation until it runs
normally.
8. Application Method
8.1 Operation Method of the Door
9
Because the device is a pressure vessel, it is important to operate the door exactly, please read
the following items carefully before operating the door.
A. Close the door
Before closing the door, make sure there are no cracks, damage and dirt on the sealing material
of the door and make sure there is no damage and dirt on the contact face between the main
body and the sealing material of the door damage and dirt.
First turn the handle counterclockwise to limit position when closing the door, so as to make all
the radiation bars of the door in contraction position. Then get the door close to the main body,
turn the handle clockwise, at the beginning, turn a bit, after the door indicator light gives out
signal, continue to turn the handle for 1/4-1/2 cycle. Strictly prohibit rotating the door
counterclockwise after the door is closed.
B. Open the door
The door can be opened only when the inner drum pressure is equal to the ambient atmospheric
pressure.
Note: The temperature of the door body is higher when sterilizing; prohibited to touch to
avoid being burned.
10
◆Castings of injection syringe influences drying effect
◆Unused products will be polluted during application period after germicidal treatment.
◆It is easy to mix for using used products and unused products after sterilization
◆Cover of castings is easy to attach water drops.
(3) When the unit is usd fro the retort pouch, the factory suggestion is important for its
performance whether it has heat resisting property and air permeability, please contact wi the
engineer of our factory.
(4) Please do not put the moisture sterilization articles, or which would affect the drying
function..
8.3 Parameter Setting Method
(1)The equipment program is optional and set freely, it is supposed as follows:
◆Replacement time: 0~60 minutes.
◆Sterilization time: 0~99 minutes.
◆Drying time: 0~99 minutes.
◆Sterilization temperature: 105~134℃ (it is generally 115℃~134℃ in practical operation).
(2)Programming Method is shown as follows:
Instruction of Parameter Definition
◆Sterilization temperature: the lowest temperature of the unit can be reach to start the
sterilizing.
◆Vacuum times: The vacuum times and the time for the steam inlet the chamber to remove the
cold air.
◆Sterilization time: Time for timing in sterilization process;
◆Drying time: Time for drying stroke timing (excluding the exhaust steam time in drying
stroke);
Examples:
Example for Setting Parameter
1. Turn on the power supply, it displays 8888 8888 8888 on display panel nixie tube, all the
indicator lights are on, after control system initialization is finished, it displays interlayer
temperature, sterilization room temperature and number 0 in turn.
2. Press setting key, it displays temperature numeral value needed to be set on the nixie tube,
displaying sterilization temperature, the displayed value is the numeral value that is set last time.
The numeral value flashes by 1S frequency. Meanwhile the sterilization indicator light is on,
meaning it is setting sterilization temperature.
Press key,the nixie tube will stop flashing when the set sterilization temperature valve
increases by the way of plus 1. The upper limit value of the set sterilization temperature is
134℃, and go on to press key, the value is kept at 134℃.
Press key, the nixie tube will stop flashing when the set sterilization temperature valve
decreases by the way of minus 1. There is no lower limit value for setting sterilization
temperature, it decreases to 115℃, return to 134℃ and decreases circularly.
Press key for a long time, the nixie tube will stop flashing when the set sterilization
temperature valve increases quickly and continuously by the way of plus 1.
Press key for a long time, the nixie tube will stop flashing when the set sterilization
temperature valve decreases quickly and continuously by the way of minus 1.
If not to press key or key for 10S, the nixie tube will recover to flash. Now the set
11
sterilization temperature is not saved.
3. Press setting key again, the nixie tube which displays the value 0 displays the replacement
time value needed to be set, the displayed value is the one set last time. The value is flashing by
1S frequency (if the previous sterilization parameter is still in setting state and not flashing, the
value will be flashing 10S later). Meanwhile the sterilization indicator light is off, which means
it is setting replacement time.
Press key, the nixie tube will stop flashing when the set replacement time value increases by
the way of plus 1. Set the upper limit value of replacement time for 60min, go on to press
key, the value is kept at 60min.
Press key, the nixie tube will stop flashing when the set replacement time value decreases by
the way of minus 1. There is no lower limit value for setting replacement time, it decreases to 0,
return to 60 min and decreases circularly.
Press key for a long time, the nixie tube will stop flashing when the set replacement time
increases quickly and continuously by the way of plus 1.
Press key for a long time, the nixie tube will stop flashing when the set replacement time
value decreases quickly and continuously by the way of minus 1.
If not to press key or key for 10S, the nixie tube will recover to flash. Now the set
replacement time is not saved.
4. Press setting key for a third time, the nixie tube which displays the value 0 displays the
sterilization time value needed to be set, and the displayed value is the one set last time. The
value is flashing by 1S frequency (if the previous sterilization parameter is still in setting state
and not flashing, the value will be flashing 10S later). Meanwhile the sterilization indicator
light is on, which means setting the sterilization time.
Press key, the nixie tube will stop flashing when the set sterilization time value increase by
the way of plus 1. The upper limit value of setting sterilization time is 99min, go on to
presskey; the value is kept at 99min.
Press key; the nixie tube will stop flashing when the set sterilization time value decreases by
the way of minus 1. There is no lower limit value for setting sterilization time, it decreases to 0,
then return to 99min, it decreases circularly.
Press key for a long time, the nixie tube will stop flashing when the set sterilization time
value increases quickly and continuously by the way of plus 1.
Press key for a long time, the nixie tube will stop flashing when the set sterilization time
value decreases quickly and continuously by the way of minus 1.
If not to presskey or key for 10S, the nixie tube will recover to flash. Now the set
sterilization time is not saved.
5. Press setting key for a fourth time, the nixie tube which displays the value 0 displays the
drying time value needs to be set, the displayed value is the one set last time. The value is
flashing by 1S frequency (if the previous sterilization parameter is still in setting state and not
flashing, the value will be flashing 10S later). Meanwhile the drying indicator light is on, which
means setting the drying time.
Press key, the nixie tube will stop flashing when the set drying time value increases by the
way of plus 1. The upper limit value of setting drying time is 99min, go on to press key, the
value is kept at 99min.
Press key, the nixie tube will stop flashing when the set drying time value decreases by the
12
way of minus 1. There is lower limit value for the set drying time, it decreases to 0, then return
to 99min, it decreases circularly.
Press key for a long time, the nixie tube will stop flashing when the set drying time value
increases quickly and continuously by the way of plus 1.
Press key for a long time, the nixie tube will stop flashing when the set drying time value
decreases quickly and continuously by the way of minus 1.
If to press key or key for 10S, the nixie tube will recover to flash. Now the set drying time
is not saved.
6. Press setting key for a fifth time, nixie recovers the display when turning on the power
supply and meanwhile save the set parameters.
Time Clock Setting
The control system consists of real-time clock, the clock mainly provides the time parameter
which is in accordance with the local time for printer, after the arrival of equipment, users shall
adjust time parameter according to the practical situation, the detailed adjusting method is as
follows。
Press starting key and turn on power supply switch, then a voice of “Di”from buzzer can be
heard. It displays 8888 8888 8888 on nixie tube of the panel from left to right in correct order,
all indicator lights are on, after 10S, the nixie tube displays time, they are: year (e.g. 08 year),
month (e.g. 04 month), day (e.g.20days), hour(e.g. 12 hours), minute(e.g.00 minute),
second(e.g.00 second) in turn.
1) Press starting key for the first time, the value of year is flashing, presses key or key to
adjust the value of year. If not to press key or key, the value will be flashing 10S later.
2) Press starting key for the second time, the value of month is flashing, press key or key
to adjust the value of month. If not to press key or key, the value will be flashing 10S later.
3) Press starting key for the third time, the value of day is flashing, press key or key to
adjust the value of day. If not to press key or key, the value will be flashing 10S later.
4) Press starting key for the fourth time, the value of hour is flashing, press key or key to
adjust the value of hour. If not to press key or key, the value will be flashing 10S later.
5) Press starting key for the fifth time, the value of minute is flashing, press key or key to
adjust the value of minute. If not to press key or key, the value will be flashing 10S later.
6) Press starting key for the sixth time, the value of second is flashing, press key or key to
adjust the value of second. If not to press key or key, the value will be flashing 10S later.
After adjusting parameters, the parameters will be saved automatically.
Turn off the power supply and the clock parameter adjustment is finished.
Adjust the clock parameters every six months to make sure the time exact.
8.4 Instruction of Sterilization Sequence
1)Preparation Stroke
In preparation procedure, it mainly finishes: load the sterilized articles, turn on the water source
switch, check whether the water source pressure complies with the requirement and close the
door of sterilization room.
Electrify and plug the attaching plug in power supply socket, turn on the breaker switch and the
power supply switch of control system. It displays 8888 8888 8888 on the nixie tube of display
panel, all indicator lights are on, after the control system initiation, the nixie displays jacket
temperature and sterilization room temperature 0, the slow exhaust steam light and fast
13
exhaust steam light of indicator light are on, then set working parameters (if necessary). The
setting range can refer to sterilization reference table for various kinds of articles (see Table 3)
and comply with different sterilization articles to set sterilization temperature/sterilization time.
14
Curve Diagram for Setting Working Procedure (Taking parameters for example)
15
The replacement time is decreasing.
When the replacement time reduces to 0, the fast exhaust steam valve will be closed and the
inner chamber will continue to heat up. The fast exhaust steam valve indicator light is off.
Start pulsating movement of fast exhaust steam valve to drainage before not reaching
sterilization temperature.
The nixie tube displays jacket temperature, sterilization temperature and sterilization time in
correct order.
Enter sterilization stroke when reaching the set sterilization temperature.
3) Sterilization Stroke
After entering sterilization stroke and reaching the sterilization temperature, it will start to
sterilize and timing. During the stroke, stop timing when the temperature is lower than the set
temperature. Close the inlet valve when the temperature is 1℃ higher than sterilization
temperature, inspect the delay exhaust steam automatically when the temperature is 3℃ higher
than sterilization temperature.
The nixie tube displays jacket temperature, sterilization temperature and sterilization time
decreasing in correct order.
The pulsation of slow exhaust valve is opened to drainage.
When the sterilization time is over, it will enter drying stroke.
4) Drying Stroke
Enter the drying stroke and remove the steam out of inner chamber, when the temperature in the
sterilization room is lower than 105℃, the buzzer will alarm to remind the users to open the
door for drying. Pay attention to the pressure of sterilization room and open the door for drying.
Now the nixie tube displays jacket temperature, sterilization temperature and drying time
decreasing in correct order.
When the drying time reduces to 0, the entire sterilization period is finished (Note, if to set the
drying time to be 0, The exhausting steam and directly buzzing of the buzzer will be transferred
to finish).
5) Finish Stroke
When the buzzer buzzes, take out the sterilization articles.
Note: As for sterilized fluid, pay attention to not opening the door directly, but turn the door in
place, open the door to 5 mm and take out the articles, until the liquid temperature of the inner
chamber reduces below 50 ℃.
Warning After entering working state, please do not try to open the door.
9 Attentions in Use
9.1 Precautions prior to use
A. Horizontal pressure steam sterilizer is an equipment, which takes advantage of saturated
steam to sterilize the dressing packages and various medical appliances. Please pay attention to
the following:
(1)Sterilization articles of this device are the ones with the characteristics of high temperature
resistance and high humidity resistance.
(3)Sterilization articles of this device shall be washed prior to use. Otherwise, the dirt
attached will affect the sterilization effect.
(4)Under normal circumstances, it is different from the drying degree of the sterilization
16
articles.
B. Security
Since the device belongs to Class I pressure vessel, please confirm the responsible person in
order to ensure the proper and safe way to use this device.
C. Horizontal pressure steam sterilizer in normal use will produce sound pressure ≤85 dB,
please be relaxed to use.
9.2 Daily use and maintenance
Warning Since the device is Class I pressure vessel, in order to use properly and safely every
day, please pay attention to the following while repairing and maintaining.
(1)Please refer to “Operation of the door" when opening the door, and check whether the
sealing materials of the door have been cracked and damaged.
Please refer to "Repair and Maintenance" for sealing material of door.
(2)Please be careful not to be burnt when adding or taking out the sterilization articles.
(3)Please confirm the indications of the pressure gauges during the use of this device. Be sure
to cut off power supply and check the control system when the pressure exceeds 0.25 MPa.
(4)Please check whether the inner chamber and exhaust steam port has sundries before use
every day. If there are sundries piled up on the filter net, it will cause incomplete or bad
sterilization.
(5)Please be sure to maintain the performance of device, please self-check at regular time.
17
◆Prepare No.4 high-temperature grease (ZBE36009—88) ZN6—4
◆Gauze glove
◆Wipes
◆Necessary tools
Order of filling up grease lubricant is as follows:
◆Disassemble the ball-end part of fixed radiation bar.
◆Clean up the aging grease lubricant in screw thread and the ball-end part of fixed radiation
bar.
◆Re-install the disassembled parts in original order.
◆Rotate the door several times repeatedly to have it smeared evenly.
Don’t make the door rotate in the open position in order to avoid the radiation bar extending to
damage the external appearance of the equipment.
10.4 Maintenance for Other Components
Safety valve: pull up the draw rod of safety valve every other month, exhaust steam several
times repeatedly to avoid failure.
Pressure gauge: inspect it regularly, check every year.
Breaker: inspect it regularly, if there is any fault, please replace it in time.
10.5 The replaced fuse specification shall comply with the regulation in the user’ manual and it
is the same with the original fuse specification.
10.6 Make sure the grounding connection of sterilizer and electric outlet is in good state.
10.7 Common Fault and Troubleshooting
FAULT No. Reason
E**3 reserve
10.8 During the sterilization, if it buzzer alarming appears and the nixie tube displays error
message, it means there is fault in working process, the details are shown as follows:
Error 1
It means the door is not closed tightly or the door switch is off. Check the door switch or the
door.
If to press the starting key without closing the door, it will display the above error message.
If to display the error message, turn on fast exhaust steam valve and slow exhaust steam valve
and turn off electric heating pipe.
Error 2
18
It means the water level reaches to the low level inside the evaporator, check the low water
level switch or whether it needs to add water.
If to display the error message, turn on fast exhaust steam valve, slow exhaust steam valve and
inlet valve and turn off electric heating pipe.
At the moment, if it needs to add water, please do not open the door and cut off the power
supply till the pressure in sterilization room and evaporator returns to 0, or turn off the power
supply to wait the evaporator to be cool gradually then feed the water.
11 Sign Instruction
Grounding sign
19
20