Application of XRD, XRF
Application of XRD, XRF
Application of XRD, XRF
X-RAY
ANALYTICAL
SECURITY MEDICAL
XRF XRD
WDX EDX
AIRPORT CARGO RADIOLOGY
Atom returns to the ground state by transferring an electron from higher energy
state to the vacancy position created in the lower energy state and emits the
excess energy in the form of x-rays. This process is called x-ray fluorescence.
The process of detecting and analyzing the emitted x-rays is called “X- ray
Fluorescence Analysis.”
For example, if a hole is created in the K-shell and is filled by the electron
from L-shell then the difference in the energies of the two levels will be
emitted as x-rays.
X-RAYS
PRINCIPAL OF XRF
BRAGG’S PRINCIPLE: n 2d.sin
where d is interplannar distance of planes, is wavelength of the emitted line
is scattering angle and n is integer
MOSELEY PRINCIPLE: C( Z )
where ν is frequency of the emitted line, Z is the atomic number (number of protons)
C & σ are constants
Based on the characterstic atomic structure of each element,
chractestic X-ray emitted of the atom from excited state to
ground state.
If bragg’s condition is satisfy within atomic structure,
coherent characterstic wave length or energy of emitted X-
rays emitted and identifed.
Atomic number Z is measured based on the detected wave
length or energy by using moseley law. contd
QULITATIVE/QUANTITAIVE ANALYSIS OF XRF
X-ray tube serves as x-ray source and excites the elements in the
sample to emit fluorescent radiation. The emitted fluorescent
radiation from the sample is collected and intensity is registered by
detection system
The difference between these two types lies in their detection system
WDXRF
A wavelength dispersive detection system physically separates
the X-Rays according to their wavelengths.
DETECTOR
SAMPLE
X-RAY
TUBE
• A detector is used to convert X-ray energy into voltage signals; this information is sent to a
pulse processor, which measures the energy of the signals and passes them onto an
analyzer.
• They are sorted by energy with help of multi-channel analyzer (energy is characteristic for
each element) and frequency of appearance (characteristic for concentration) and sent to
data display and analysis.
• The most common detector now is Si(Li) detector cooled to cryogenic temperatures with
liquid nitrogen.
COMPARISON OF EDXRF AND WDXRF SPECTROMETERS
Instrument calibration
Metals, Slags
Glass
INSTRUMENT AND LAB PREPARATION
Constituent Maximum
difference The maximum permissible values for the
between
duplicates (%±)
SiO2 0.16
standard deviation are listed.
Al2O3 0.20
The two most critical elements are CaO and
Fe2O3 0.10
CaO 0.20 SiO2.
MgO 0.16
SO3 0.10 If either of these elements exceeds the values
Na2O 0.03
K2O 0.03 shown the sample preparation procedure is
TiO2 0.02
P2O5 0.03 unacceptable and the cause of deviation must
Mn2O3 0.03
be sought and rectified.
• INSTRUMENT CALIBRATION
CALIBRATION PROCEDURE
Constituent Typical Concentration Ranges for
portland cement
After success in the First step: SiO2 20 – 23
preparation of at least 7 NIST Portland Al2O3 3–6
cement reference materials can be Fe2O3 2–4
undertaken CaO 61 – 67
MgO 0.5 – 4
• NIST 1880a, 1881a, 1882, 1883, SO3 2.0 – 4.5
1884a, 1885a, 1886a, 1886b, 2689, Na2O 0.1 – 0.5
2691, 8486, 633, 634 K2O 0.1 – 1.3
Intensities of the elements are measured TiO2 0.2 – 0.3
P2O5 0.1 – 0.25
Calibration curves are derived through Mn2O3 0.05 – 0.3
correlation with the known concentrations
The prepared reference materials used for
calibration are analysed as unknowns.
CONCENTRATION
These analytical results constitute the first
set of test data.
CALIBRATION PLOT
INTENSITY
MAXIMUM PERMISSIBLE VARIATIONS IN RESULTS FOR
RAPID ANALYTICAL METHODS
Constituent Instrument Errors %(±) Maximum difference Standard deviation (%±)
between duplicates
(%±) IS 12803
SiO2 0.05 0.2 0.2
Al2O3 0.05 0.2 0.1
Fe2O3 0.05 0.1 0.1
CaO 0.05 0.3 0.5
MgO 0.05 0.2 0.2
SO3 0.05 0.1 0.1
Na2O 0.005 0.05 0.05
K2O 0.005 0.05 0.05
TiO2 0.005 0.03 0.05
P2O5 0.005 0.03 0.03
Mn2O3 0.005 0.03 0.03
• One day later: A new set of the same reference materials used for calibration is
prepared and now analysed as Unknowns. These analytical results constitute the
second set of test.
• Precision of the method is calculated by the difference between the two sets of
analyses obtained on separate days. Differences between the results may not exceed the
maximum differences between duplicates shown here
XRF technique is very sensitive , therefore choosing the sample
and preparing it for the analysis are very important. Only a few
grams of sample is analysed in this technique, so it should be chosen
in such a way that the sample should represent the entire bulk
material.
Samples prepared in this way has several advantages over the samples
prepared by pressed pellet method
Contamination from the flux and decrease in the line intensity due to dilution
are the main draw backs of this technique
TYPICAL SAMPLES FOR X-RAY FLUORESCENCE
PRESSED PELLETS
FUSION BEADS
SOURCES OF ERROR RELATED TO SAMPLE
Sample preparation method
Mineralogical effects
Sample selection
10 μm
COARSE PARTICLES
~ 50 μm
When the particles are coarser, only a few particles will be analysed and the
sample may not represent the entire bulk material.
10 μm
FINE PARTICLES
~ 10 μm
When the particles are finer, then sufficient number of particles will be analysed
and the sample well represents the entire bulk material.
MINERALOGY EFFECT
Mineralogical effect is defined as the difference of absorption of the
fluorescent radiation in the particles of minerals of different chemical
compositions.
E.g Calibration with OPC samples and analyse of PSC or 5% limestone addition
OPC or PPC samples
INTERFERENCE ON XRF INTENSITY
In the case of interference on XRF intensity of one
element due to others in the sample, the equation
Y=Mx+C takes the form of Y=Mx+C+terms involving
interfernces.
Interfering elements are
Si by Mg and Al
Al by Mg and S
Fe by Ca and Si
Ca by K
Mg by Ca
MATRIX EFFECTS
Intesity of a line not only depends on the concentration of the element from which it
is originating, but also on the presence of the other elements in the matrix
Absorption
Enhancement
Characteristic radiation of one element excites other element in the matrix and
enhances its signal
CALIBRATION VALIDATION
• Calibration plot is validated by refernce
samples analysis (NIST).
• Validation of unknown sample results by
conventional chemical analysis method.
• Drift correction of the calibration curve, If
check sample analysis is not within the limits
of repeatability.
sample CaO SiO2 Al2O3 Fe2O3 MgO SO3 Na2O K2O LOI
638 62.09 21.48 4.45 3.55 3.83 2.34 0.13 0.59 0.95
638* 62.05 21.52 4.48 3.43 3.70 2.10 0.15 0.54 0.95
calibration
310
62.46 20.88 5.28 3.72 2.29 2.08 0.287 0.60 0.88
310’
conventional
62.20 21.09 5.36 3.78 2.33 2.11 0.29 0.61 0.89
ACCURACY
Constituent Maximum difference of
• Accuracy of the method is the average of
duplicates from SRM
measured by calculating the Certificate value (%)
average value of the two separate SiO2 0.2
determinations for each element of Al2O3 0.2
each standard. Fe2O3 0.1
The difference between this CaO 0.3
Operator
Regression technique
Standards accuracy
Standards stability
Standards composition
Standards range
Calculation philosophy
SAMPLING
APPLICATIONS OF XRF IN CEMENT INDUSTRY
Quarry guidance
ACCURATE
AMENABLE TO AUTOMATION
• This technique is not very suitable to find the low z (< 11) elements
CASE STUDY
QUALITY CONTROL SAMPLING POINTS
MINES COMPOISTION
• Specific compliance of the avaliable materials and to ensure
that the major components will meet raw mix requirments
• Quality of the limestone deposited in the mines may not be
uniform and diffrent grades/types of limestone available at
diffrent locations.
• Blending of diffrent grades of limestones is controlled by
XRF in desired proportions and transpotrted to the stockpile.
LOI
40.23 36.15 32.42
CaO 46.78 43.23 38.6
SiO2 6.74 15.12 21.69
Al2O3 0.66 1.23 2.45
Fe2O3 0.55 0.71 1.08
MgO 5.08 3.27 2.94
Na2O 0.06 0.07 0.24
K2O 0.15 0.36 0.69
Cl- 0.004 0.004 0.006
STOCKPILE COMPOSITION
• Large quantity preseved/stored material for correct composition with
unifrom quality
• Pile quality is monitered by analysing the chemical analysis of
diffrent locations of samples and if quality is not upto desired level,
corrective action such as adding required quality of material in
desired quantity.
• General practice of pile quality parameter is Lime Saturation Factor
(LSF) and quick control/moniter technique is XRF analysis.
RAW MATERIALS
Sample CaO SiO2 Fe2O3 MgO Na2O K2O Cl-
LOI (%) Al2O3 (%)
Code (%) (%) (%) (%) (%) (%) (%)
LST 35.16 41.68 12.92 1.16 0.68 3.56 0.08 0.32 35.16
Material Name
Raw Mix1
Limestone
95.4
Mill Scale 0.6
Bauxite 4.0
• The raw materials with desired weight proportions are conveyed to grinding mill and
percentages of these materials are designed based on the quality of the materials and quantity
of such materials is controlled/monitered by XRF
• Corrective actions are designed for quanity increase or decresed or addition of
sweetner/additive based on the quality of the materials.
RAW MEAL
• The quality of raw meal is an
key parameter and it is
monitered and ensured by XRF
• If the quality of the material is
not meeting the required levels,
corrective action will be taken.
RM-1
35.87 13.22 41.69 3.56 2.68 2.48 99.80 2.56 0.93
RM-2
CL-1 21.56 63.3 5.45 4.17 4.57 0.14 0.13 0.49 0.37
CL-2 21.43 63.23 5.72 3.94 4.71 0.14 0.16 0.52 0.38
SiO2 (%) CaO MgO Fe2O3 Al2O3 SO3 Na2O K2O TiO2
(%) (%) (%) (%) (%) (%) (%) (%)
OPC-1
21.02 61.66 5.21 3.98 4.40 2.02 0.13 0.47 0.35
OPC-2
20.90 61.59 5.46 3.76 4.53 2.02 0.16 0.50 0.36
SUMMARY
• The defination of quality is the same product each time within
defined limits.
• Quality depends on how fast the measurment is done to enalble a
process/quality manager to assess whether the quality of the product
is maintained.
• Quality of the materials can be cntrolled by getting the analytical
results within reasonable time period and feedback given to
respective departement in order to take neccesary steps accordingly.
• Quality control and process optimization for cement maufacture
mainely depends on XRF
• Monitoring and controling the quality of materials from starting
stage to final product.
• XRF allows greater understanding of the quality of the material
within 30 mins and therefore greater control of the entire process
focusing on the ultimate quality of product of reduced costs.
• XRF plays a key role in on-line quality control of the final desired
product.
THANK YOU