Dr. Life DTS Tourniquet

Download as pdf or txt
Download as pdf or txt
You are on page 1of 26

Product Standard No :

Product Reg No :

User's Manual

Add : 689-31 Keumjung-dong, Kunpo City, Kyunggi-do


Tel : 031-459-7211
Fax : 031-459-7215
E-mail: [email protected]

Important for safe use: The User's Manual is for safeguarding your injury or property
loss. Please carefully read it to use the system properly.
Table
1. Basic Information
1.1 Introduction 4
1.2 Definition 4
1.3 Index 4

2. Clinical Information
2.1 Applications Scope and Effects 5
2.2 Non-application Cases 5
2.3 Warnings 5
2.4 Cautions 5
2.5 Use and Storage 6
2.6 Trouble Signs 7

3. Description
3.1 Outline 7
3.2 Functions 8
3.3 Composition of the System 11
3.4 Upper Side 14
3.5 Front Side 14
3.6 Rear Side 14
3.7 Bottom Side 15
3.8 Technical Data 15
3.9 Product Label 15

4. Product Installation
4.1 Packaging 16
4.2 Inspection 16

5. Use of Product
5.1 Preparation 17
5.2 Install and Use 17
5.3 Use Stoppage 18
5.4 Recommendation 18

6. Troubleshooting
6.1 In-advance Checkups 18
6.2 Other Problems 19
7. Storage
7.1 Storage 19
7.2 Cleanup 19

8. Appendix
8.1 Purchase for Product and Its Parts/Accessories 20
8.2 Exchange of Parts and Additional Purchase 20
8.3 Warranty Period and Returnables 20
Appendix A. Warranty Certificate 21
1. Basic Information
1.1 Introduction

Thank you for your purchasing our DTS-2000W product. It is designed to inflate the limb, after cuff
application, to reduce or block blood circulation for surgical use, which must be in accordance with
your surgical instruction.

This electronic pneumatic tourniquet is used to apply controlled amount of pressure to the circumstance
of extremities through cuffs to stop bleeding in an operative field, having two channels each of which
has two cuffs. Pressure scope applicable for patients ranges from 50mmHg to 600mmHg, depending on
bleeding condition and patient's status. The Double Cuff(with two bladders) is used in case of
operation in need of many hours to prevent bleeding area from damage, either enabling you to change
the monitoring area by pressing the button of 'CUFF CHANGE'.

Its battery, when fully charged, lasts for min. 12 hours. of continual use, not causing any problem
in cases of out-of-blue electricity failure in an operational field, with easy-to-use and cleanability
features of LCD and touch-sensor switches.

The User's Manual, inclusive of many important information on proper use, maintenance and storage,
must require your full understanding prior to clinical applications in surgical settings to let there
be any tourniquet-related problem.

1.2 Definitions
1.2.1 Preventive signals are as follows;

Warning :The symbol may cause life threatening situation when not complied strictly.

Caution : The symbol may cause the user damaged or property loss when not complied
strictly.
Stop : The symbol represents strict compliance by the user.

No Disassembly : The instrument shall be strictly prohibited from disassembly in any case.

1.3 Index
1) bold letters : main information
2) PRESSURE : pressure value(mmHg) for SET process
3) INFLATE : process infusing air to cuff
4) DEFLATE : process discharging air from cuff
5) mmHg : unit of pressure value
6) cuff : air pocket

2. Clinical Information
2.1 Usage Application
This electronic pneumatic tourniquet, DTS-2000W, is used to apply controlled amount of pressure to
the circumstance of extremities(upper and lower) through cuffs to stop bleeding in an operative field,
having a total of four outputs in two channels. The pressurized gas is exerted to the extremity exposed
for surgical operation in order to create a bloodless operative field. The instrument is categorized
into an operative device, requiring your doctor's instruction, based on well-established
understanding of how to use.

2.2 Non-application Cases


※ The instrument must be used according to a surgeon's instruction.

1) Warning : Do not apply to patients infected in their extremities, or those with


VTE(Venous Thromboembolism).

2) Warning : Do not apply to the area in the state of tumor or trauma.

3) Warning : Do not apply to those whose extremities are inserted by artificial joint,
viscose, metal or silicon. It may be detrimental for possible inflammation.

4) Warning : Do not apply to those with heart disease, hypertension, malignant tumor,
purulent disease, high fever, general prostration and so on. It may worsen the symptom.

5) Warning : Do not apply to those who were post-operated until healed completely,
especially for such operations on viscera, circulatory system, nerves and cerebral area
being prohibited from use for one year after operation. The use may cause detrimental effect
against post-operational recovery.

2.3 Warnings

1) Warning : Do not use the instrument until fully understandable for how to connect and
use.

2) Warning : Do not manipulate the instrument whose gas tubing is directed to nose, mouth
and ear. It may give rise to physical injury.

2.4 Cautions
1) Stop : The product is usable for voltage range of AC 100V - 240V~ 50/60Hz. The use
setting apart from the allowed scope may cause fire or electric shock.

2) Stop : Do not plug in or plug out with wet hands, which may cause electric shock.
3) Stop : Do not damage electric cords or forcibly twist or fabricate them. The damaged
tubing may cause fire or electric shock.

4) Caution : Do not use any longer when there occurs strange odor from the system during
manipulation. In this case immediately plug off the instrument and call our service center.
Further use may cause fire or electric shock. (Service Center ☎ 031-424-0091)

5) Caution : When not in use, when situated in thunder or lightning or when in abrupt
interruption of electricity, plug it off from the power source.

2.5 Use and Storage


1) Stop : Use in ambient room temperature ranging from 0℃ to 40℃. The use deviating the
scope may cause defect to the system and fire.
2) Stop : Do not use or store in the vicinity of heating resources or direct sunlight. It
may cause defect, deformation or fire.

3) Warning : Do not use in a moist place(such as bathing room, sauna, etc.), which may
cause fire or electric shock.

4) Warning : Mount the instrument to the stand firmly in a plain surface. Leaving it to the
unstable state may cause falling, defect or malfunction.
6) Stop : Do not apply anti-blooding cuff to the extremities wearing accessories or wrist
watch, which may cause physical injury or damage to cuffs.

7) Stop : Do not infuse pressurized gas to cuffs not applied to extremities, except for
testing. It may cause defect or deterioration of durability.
8) Stop : Do not place cuffs neighboring to flammable origins like a heater or cigarette.
It may cause damage to cuffs or fire.
9) Stop : Surgical cuffs must not be placed in the vicinity of sharp materials such as
needles, scissors. It may cause damage to air chamber of cuffs or malfunctions.

10) Stop : Do not allow the instrument stained with such substances as oil, benzene,
alcohol, benzine or chemical. It may worsen durability in the system.

11) Stop : Let hoses not bended or folded, which may cause malfunction.
12) Stop : Do not infuse gas to the cuff using the method not allowed by this Manual. It
may cause cuff damage or physical injury.
13) Stop : Do not allow the instrument stained with such substances as oil, benzene,
alcohol, benzine or chemical when you clean the system. Use dry clothes.

14) No Disassembly : The system must be never dismantled by unauthorized person. It may
cause fire or electric shock.

15) Stop : Do not let clips or staples including liquid substance be inserted into the
instrument, which may be highly detrimental causing short circuit or heat generation. If
inserted already, call your nearest distributor for servicing.
2.6 Trouble Signs

2.6.1 When the following troubles occur, immediately take a proper countermeasure.
1) When there occurs warning beep sound at the same time showing error message in LCD window,
immediately check its possible reasons. In the event those situation happens repeatedly, call
the shop where you purchased it.
2) In the event that air is not discharged properly, stop the operation immediately.
3) Whenever you receive even the minimal electricity while touching the instrument, immediately
plug off the cord from the source and call your shop of purchase.

4) Although cuffs were connected properly but when warning signs arise, immediately change your
system to 'OFF' condition.
5) During the use, when there produces abnormal sound in motor operation or when the LCD window
is fluctuating, immediately stop the use of it any longer and call your distributor shop of
purchase.

3. Description
3.1 Outline
The device, being attached to the upper or lower extremities, is electronic pneumatic tourniquet
designed to reduce blooding in a surgical operation. It is composed of 8 single cuffs and 6 double
cuffs being required to select the most appropriate one taking into consideration the operational
duration, patient conditions and where to apply. Wrong selection of cuff may retard the effective
anti-blooding.

The product is designed to freely administer pressure level inside the cuff using digital pressure
sensor, controlling motor pump and solenoid valve, while consistently maintaining a preset pressure
level by monitoring digital pressure sensor even after the set value is reached.

The feature making the desired pressure level maintained all the time during the use, not preventing
from over excessive pressure imposed to cuffs, demonstrates it automatically makes it possible to
the pressure level not changeable during use through self-diagnosing calibration function.

The battery lasting for max. 12 hours with no power source prevents from any accidental impact to
the operation due to sudden electrical interruption.
3.2 Functions
3.2.1 Nomenclature

1 Channel 1 Pressure Control Button 9 Channel 1 Exhaust Button


2 Channel 1 Time Control Button 10 Channel 1 Cuff Conversion Button
3 Channel 1 Pressurization Button 11 Storage Button
4 Channel 1 Exhaust Control Button 12 Warning Sound Elimination Button
5 Channel 1 Cuff Conversion Button 13 LCD Display Window
6 Channel 2 Pressure Control Button 14 Channel 1 Air Outlet
7 Channel 2 Time Control Button 15 Channel 2 Air Outlet
8 Channel 2 Pressurization Button

All switches in the pad are applied by touch sensors, enjoyable for smooth touching in operation with
no need of forcing onto them.

①CH 1 Pressure Control Button is the one inputting pressure value applicable for CH1, which
is described as following Fig. 1;
By moving ▲, ▼ buttons up/down you input your desired
value between 50mmHg~600mmHg depending on patient's
condition and specific demands of the limb. The value below
which SET appears means that the number is the preset
pressure value. If not, the value is the present actual pressure
value existing inside the cuff.

Fig. 1 : method for setting CH1 pressure value

③ CH1 Time Control Button is the one to set time value applicable to CH1 according to Fig. 2 as
follows;

By moving ▲, ▼ buttons up/down you input your desired


time value ranging from 5 min to 180 min(3 hrs) subject to
patient and operation conditions. The value over which SET
appears means that the number is the preset time value. If
not, the value represents the actual elapsed time for
anti-blooding.

Fig. 2 : method for setting CH1 time value


③④⑤ CH1 operational buttons are the ones to make CH1 to INFLATE, CUFF CHANGE and DEFLATE,
where you must press and hold buttons for up to 2 seconds. The operational description is made in
Fig. 3 as follows;

Fig 3: operational buttons for CH1

CH1 Control Panel Operational Signals

INFLATE DEFLATE OK!


- INFLATE Button: Pressing and holding the button for 2 sec is made to pressurize gas into the cuff for
preventing blooding, which flickers in the bottom of LCD display to show the recent status of operation.
When pressurization is met to the present value, the display shows "OK!" signal that means the
preparation of surgical occupation.
- CUFF CHANGE Button: It is used for changing the anti-blooding area to prevent possible damage to
skin, nerve and muscular system in surgical operations exceeding more than an hour. It can work only
under the conditions that "OK!" appears in the operational display or when the applied cuff is in the state
of malfunction. To make it operative, press and hold the button for up to 2 seconds. It functions to
infuse gas to non-pressurized cuff, where the operational status and cuff operating status are the same
to the case of INFLATE button. Once the preset pressure value is reached, the old pressurized gas is
consecutively discharged, being replaced by new pressurized gas to the cuff. In this case the presently
pressurized cuff is shown in the bottom of LCD display.
- DEFLATE Button: It is used to exhaust the infused gas after finishing surgical operation, which is
necessary to press for up to 2 seconds. When depressurization within the cuff is satisfactorily made, the
system returns the state for the next use.

⑥ CH2 Pressure Control Button is the one inputting pressure value applicable for CH2. functional
use is the same to that of CH1 Pressure Control Button.

⑦ CH2 Time Control Button is the one to set time value applicable to CH2. Its functional use is
the same to that of CH1 Time Control Button.

⑧⑨⑩ CH2 Operational Buttons are the ones to make CH2 to INFLATE, CUFF CHANGE and DEFLATE,
where you must press and hold buttons for up to 2 seconds. Their functional uses are the same to
those of CH1 Operational Button.

⑪ Set Value Storage Button To store the set values, press and hold this button for 2 seconds.

⑫ Warn Sound Removal Button It eliminates the warning sounds caused by excessive lead time in
operation or system errors.

⑬LCD Display Window It shown the status of system operation.

⑭ CH1 Air Outlet is to connect cuff connectors(hose) to supply cuffs with pressure gas.

⑮ CH2 Air Outlet is to connect cuff connectors(hose) to supply cuffs with pressure gas.

3.2.2 Error Messages


1) NO CUFF! : It is shown when the desired level of pressure is not detected even after some lapse
of time from the system operation, which requires the visual inspection on whether cuffs are well
connected with hoses. When the same error message is repeatedly shown under good connecting of cuffs
with hoses, contact your product seller.

2) LEAK AIR! : It is displayed when pressure level is not reached to the preset value, though a certain
level of pressure is detectable, which requires the visual inspection on whether cuffs are well
connected with hoses or if there occurs leakage. When the same error message is repeatedly shown under
good connecting of cuffs with hoses, contact your product seller.

3) LOW BATTERY! : It is displayed when battery power is in the low state while the system is not
connected to AC power, requiring immediate connecting the system to the power source. If not, it may
cause shorter lifetime of the system.

4) TIME OVER! : It is flickered in LCD, together with the warning sound, when elapsing the preset
time after anti-blooding function, requiring proper measure following by instruction of a surgeon.

3.2.3 Other Functions


1) To save battery power, the instrument automatically enters the sleep mode after 3 minutes of
non-operation, and activates again as soon as you press any button in the panel for one time. As
to the low power state of battery, the instrument is changed to the sleep mode after 20 seconds of
non-operation while generating warning message for every two minutes. In this case immediately
connect it to the power source for batter recharging. Complete discharging may cause shorter lifetime
to battery.

2) When the battery is in short of power voltage, the yellow-colored battery symbol in LED flickers
continuously while generating the 1st warn sound. And when it is outreached to the point of 2~3 minutes
before its complete discharging is made, the system produces the 2nd warn sound. Therefore, when
heard of battery warning sound, the system MUST be plugged in to AC power source in order not to
bring about operational stoppage causing a detrimental impact to your surgical operative field.
Complete discharging in a battery may impede recharging successfully, so keep in mind that the battery
MUST not be discharged in any case including storage.

3) The instrument(DTS-2000W) automatically calibrates the pressure value through digital pressure
sensor when connected to the power for the first time, which has highest accuracy in representation
of values all the time.

3.2.4 Functional Flow


- System's Operational Flow(Incl. Air Flow)
3.3 Composition of the System
3.3.1 Main Components
1) DTS-2000W Main Body

Rated Power Source : AC 100 - 240V~ 50/60Hz

Rated Consumption Power : 50W

Output Channels : 4 channels

Allowable Pressure Scope in Channel : 50 ~ 600mmHg

Allowable Tolerance : each value input ± 10mmHg


2) Single Bladder Cuff

① material: cotton(65%) + polyester(35%)


texturization
② SIZE
- Child single cuff: 400mm X 70mm (Basic)

- Arm single cuff: 520mm X 75mm (Basic)

- Thigh single cuff: 800mm X 90mm (Basic)

3) Dual Bladder Cuff

① MATERIAL: cotton(65%) + polyester(35%)


textrication
② SIZE
- Arm - L: 565mm X 100mm (Basic)
- Thigh - M: 800mm X 150mm (Basic)

4) Connecting Tubing

① material : PVC
② SIZE : 2M
③ use : the one end is connected to a jack of
the system and the opposite plug connected to cuffs
for use.
5) Stand

① SIZE : 1020mm X 540mm


② weight : 8.5Kg
③ use
- Mount Stand Fixing Bracket to the system.
- Assemble the bracket and stand to connect each other firmly
using joint screws to use.

3.3.2 connection
※ The instrument must be mounted in a specialty fixture or fixed in the safe place such as the
top of a table.
1) mounting of main body
① example of mounting to a specialty fixture
- Mount the main body paced to a fixture(bracket) to fix them each other as shown
hereunder.

- The set is again connected with a stand firmly using joint screws to ensure they are
tightly combined.
- The stand must be ensured to be in the plain and safe place.

2) Connection of System with Cuffs


① Single Cuff
- The connecting hose as shown in the Fig. is pushed to a channel of the main body until
click sound is heard.
- The opposite plug of the said hose is connected to the cuff to use in a manner that
ensures there is no leakage. Improper connecting causes malfunction to the system.

- After making sure the connection, power on the main body and input desired pressure and
time values for operation.

② Double Cuff
- The connecting hose as shown in the Fig. is pushed to a channel of the main body until
click sound is heard. Match the color between horses and cuffs while connecting.
- The opposite plug of the said hose is connected to the cuff to use in a manner that
ensures there is no leakage. Improper connecting causes malfunction to the system.

- After making sure the connection, power on the main body and input your desired
pressure and time values for operation.

3.3.3 Fits and Cautions


1) The use of the instrument must be proceeded in compliance with your doctor's instruction.
2) Make sure the condition of cuff before its placement. The cuff torn out or damaged with
possibility of leakage may cause insufficient anti-blooding effect.
3) Do not operate the system until the cuff is completely fit to a patient. When inflated cuff
to some extent is applied to the patient, it may cause a physical burden as well as
possibility of damage to cuff itself. It further weaken the close tightening to the patient's
skin, which lowers desired anti-blooding level.
4) The choice of different types of cuffs must be individualized depending on primarily proper
fit to a patient and surgical procedures. Wrong selection of a cuff may give rise to
unexpected result far beyond the desired effectuation.
5) When fitting to the upper level of arms or thigh, the thickest part in a cuff must be
applied for the fit.
6) The pressure level imposed to a cuff must be minimized as possible as you can as a means of
preventing from possible impact to blood vessels, nerver, muscular system and skin.
7) In general, it shall be recommended that the fix time through a tourniquet be elapsed not
exceeding 60 min for the upper extremity and 90 min for the lower extremity. When required for
further application, make the tightening strength loose for 15 min tentatively to reduce
inflammatory response.
8) In case of double cuff, be careful of its application. When operation starts, the red-color
cuff is first operated in case of CH1, contrary to the blue-color cuff being operated first in
case of CH2, while the gray-color cuff is operated only by pressing "Cuff Change" button.
9) When cuffs are fit to the patient, continuously monitor the body temperature, blood pressure
and pain.
10) In case that a cuff of 2CH is additionally applicable, take heed of wrong cuff not to be
inflated or discharged; CH1 for blue color and CH2 for red color.
11) For next use, the cuffs used should be disinfected properly.

3.4 Upper Side


3.5 Front Side

3.6 Rear Side


3.8 Technical Data

※ The product has the following specifications;

1) Product Specifications
Div Specification Remark

Rated Voltage AC 100 - 240V~ 50/60Hz

Rated Consumption Power 50W in case of complete


Stand-by Consumption Power below 25W battery discharging

Rated Voltage 25W in case of complete


Stand-by Consumption Power below 1W battery recharging

Power Noise Level below 2 KV

Operation Noise Level below 60dB

Max Pressure 600 mmHg ± 10mmHg

Wt(Main Body) 5.5 Kg

Size 225(W) X 185(D) X 240(H)mm

3.9 Product Label


3.9.1 Product

3.9.2 Symbols
※DTS-2000W has information on the following symbols in its back panel side.

Caution : Refer to Related BF-type Clinical


Documents. Instrument

European Accreditation Mark Manufacturer Mark

Service Provider Mark Manufacturing Date

LOT NO SERIAL NO

3.9.3 DTS-2000W LABEL


※ DTS-2000W has a product-descriptive label attached in its rear side of the main body.
Product Informational Sticker

4. Product Installation

4.1 Packaging
DTS-2000W has product box and parts box.
4.1.1 Components of the Product
DTS-2000W is packaged in order as follows in its main box.
1) PART BOX
① connecting hose(blue) X 1
② connecting hose(red) X 1
③ connecting hose(gray) X 2
④ power supply cord X 1
2) User's Manual 1 ea.
3) Main Body : DTS-2000W
4.2 Product Inspection

※ Caution : Do not use sharp cutting tools such as knife and scissors in an attempt to open a
package because cuffs are positioned on the top of the main body.
4.2.1 Inspection
1) Make sure the box includes the following components properly.
① main body DTS-2000W 1 SET
② connecting hose(blue) X 1
③ connecting hose(red) X 1
④ connecting hose(gray) X 2
⑤ User's Manual 1 ea.

2) Make further sure the followings are fit properly.


① product model no.(DTS-2000W)
② connecting hoses composed of 1 blue-color, 1 red-color and 2 gray-color hoses
③ whether sockets of connecting hoses are properly fit to outlet jacks in the system
④ whether sockets of connecting hoses are fit properly to jacks in cuffs
⑤ the power supply cord applicable for your local country.
⑥ whether the rated voltage shown in the rear side attaching to the main body is fit for your
power source

4.2.2 In Case of Troubles in Product


※ In the event that there lacks in required components in a box or there finds malfunction(s)
in the Product, please do not hesitate to call the nearest distributor or the location where
you purchased the product.
4.2.3 Special Remark
※ The case of defect/damage to cuffs, connecting hoses and/or a main body due to your
uncarefulness during opening the packs shall not be applicable for our free warranty
policies.
* Caution : When not in use, it is recommended the Product after disconnecting from the power
source be repacked in a box or placed in a safe area free from dust or moisture.

5. Use of Product
5.1 Preparation
The product must be handled under superintendence by your surgeon. For fit to patients, take into
consideration the fit location, type of cuffs to be used and value of pressurization in a right manner.
If used improperly, it may cause detrimental effect to a patient under operation.

5.2 Use and Install


1) To begin with, mount the main body in a fixture to be refixed to a stand. And then place the stand
in the plain area and connect the instrument with a power supply cord.
2) Connect the main body with cuffs selected and connecting hoses according to the procedure as follows;
- cuffs selected → connecting hoses → air outlets in a main body
3) Set pressure value and time best fit for conditions in the target patient and surgical operation.
4) Apply the select cuff to extremity in a patient.
5) Press "INFLATE" button and hold it for 2 seconds to start out pressurization into the cuff.
6) The representation of "OK!" in LCD window verifies the required pressurization level is reached,
where verification of blood flow in the applied extremity is made for determining to commence the
operation.
7) As for use of double cuffs, press and hold “CUFF CHANGE" button for 2 seconds, if required, to
change pressurized fit location to a different location.
8) To extinguish the inflated gas pressure existing in cuffs, press and hold “DEFLATE" button for
2 seconds to discharge it completely.

5.3 Stop of Use


※ In the following cases, make sure to let your doctor know the problem and take a measure according
to his/her instructions.
1) When cuff(s) is inflated, but the value in the system represents low pressure value,
2) In case that the system generates an error message,
3) When buttons in a panel are not operated properly, and
4) When the LCD display is unstably fluctuating

5.4 Recommendation
※ The instrument is for clinical applications in a surgical operation. It is of outstanding importance
that sufficient comprehension of how to use be accomplished prior to practical applications according
to instruction by a person in charge of a surgical operative field. Feel free to contact us at local
agent if you have any inquiries or a matter inconvenient.

6. Troubleshooting
6.1 In-advance Checkups
NO problem cause troubleshooting
connection to power
Plug in correctly to a power supply socket.
supply source
Power is not ON to anrated power Check if your power supply is in the range of AC 100
1 - 240V~ 50/60Hz.
instrument
Check if FUSE is in the state of shortcut.
FUSE shortcut
(FUSE is located inside AC INLET.)

Power is ON, but Plug off and then plug in again to check up the
2 power supply
operation is not made operation is made properly.

Check if the system and its stand is closely


Loud noise is tightened.
installation
3 generated during
environment
operation
Make sure not to put something onto the system.

connection of hoses Connect the system with hoses properly.


Air is not discharged
4
from hoses hoses bent Connect hoses after straightening the bent area.

6.2 Other Problems


NO problem cause troubleshooting

defect in air hoses


Air is infused not
1 defect in air hose Call your product seller for inquiry on repairing.
strongly
socket
fault in internal
parts
Power is ON, but
2 fault in internal Call your product seller for inquiry on repairing.
operation is not
parts (except fault of FUSE)
made

※ Be advised that faults caused by your careless, improper use or handling are not covered by
our Warranty Policy.

7. Storage

7.1 Storage
1) Avoid the area susceptible for direct sunlight.
2) Avoid the moist/wet area such as bath room or lavatory.
3) Avoid storage in dusty area and keep the product packed in a box when not in use for long
time.

7.2 Cleanup
1) The system must be cleaned with dry clothes--It is prohibited to use detergent or other
chemicals because it may cause discoloring or defect to the system.
2) When cleaning cuffs, wash them with clothes as much as min 90% of moisture is removed, and
then clean them again with dry clothes to get rid of any moisture.
3) Be careful for foreign materials not to be inserted into cuffs and/or hoses.
4) In case of foreign materials inserted into the system, do not disassemble it but call for our
local agent.

8.3 Warranty Period and Returnables

1) In case of product return not caused by the product malfunctions, the expense for sending
back shall be borne to the sender.
2) We provide customer-oriented product warranty policy expect for the cases as follows, unless
otherwise stipulated separately;
① defects caused by a strong impact to the product
② in the event the user intentionally refabricate the inside of the product or repair it
③ in the event the product is used in the forbidden areas not stipulated herein
④ in other cases of manipulation, installation and storage with methods not stipulated herein
⑤ Cuffs are not warranted because they are disposable items.
※ Free-of-Charge Warranty Period
1) Up to 1 Year as for Instrument
2) Up to 6 Months as for Cuffs

Warranty Certificate
Much appreciated on using our device. We, Daesung Maref are doing our best to
improve the quality of our products. We will do our best effort to be a good companion
of users.

* In case of followings, we can not repair our device gratuitously even


though warranty period:
1. Disorder happened by strong impact.
2. In case user repair or reproduce internal part arbitrarily.
3. In case of using the device in prohibited place.
4. In case of against our <HOW TO USE>.
5. Cuff is articles of consumption.

DESCRIPTION Automatic Tourniquet System

MODEL NAME DTS-2000W

SERIAL NO.

WARRANTY CONTROLLER: 1 YEAR, CUFF: 6MONTHS

BUYING PLACE

BUYING DATE
689-31, GUMJUNG-DONG, GUNPO-SHI, 435-862, KOREA,
TEL : +82-31-459-7211, FAX: +82-31-459-7215, E-mail :[email protected], Website: www.dsmaref.com