GA1000 Series User Manual v1
GA1000 Series User Manual v1
GA1000 Series User Manual v1
Announcement
The content in this manual could be changed without prior notice.
NANJING GLARUN-ATTEN TECHNOLOGY CO., LTD. LTD provides no warranty of any kind to
this manual and assumes no liability or responsibility for any mistakes possibly contained in the
manual or any occasional or subsequent loss caused by using this manual.
Copyright
© 2012 NANJING GLARUN-ATTEN TECHNOLOGY CO., LTD. LTD All Rights Reserved.
Trademark Logo
is the registered trademark of NANJING GLARUN-ATTEN
TECHNOLOGY CO. LTD.
Declaration
ÿ The products of the company are under protection of the People's Republic
of China Patent that has been approved or is pending.
ÿ The company reserves the right to change the specifications and prices.
Warning
The warning announcement points out the operation or condition that
may endanger the operators.
notice
The notice announcement points out the operation or condition that
may cause damage to the product or other properties.
Terms on the product. The following terms may appear on the product:
Danger It represents that harms may be caused to you at once if you perform
the operation.
Warning It represents that latent harms may be caused to you if you perform the
operation.
Notice It represents the damage possibly caused to the product or other properties
if you perform the operation.
Characters on the product. The following characters may appear on the productÿ
notice
protective Measuring chassis
please read
the manual ground terminal ground terminal
Machine Translated by Google
Machine Translated by Google
GA1000 series of digital storage oscilloscope is small in size and flexible to operate. It adopts a
TFT LCD and a pop-up menu for display, and can remarkably improve the working efficiency by its
ease of use.
GA1000 series of oscilloscope has excellent performance, various functions and competitive cost
to performance ratio.
Characteristicsÿ
ÿ The oscilloscope has a totally new ultrathin appearance design, and is small in
size and more portable
ÿ A 7-inch widescreen color TFT LCD displays clear, crisp and more stable waveform
display
ÿ Pass/Fail function
Machine Translated by Google
ÿ The pop-up menu display mode realizes more flexible and more natural for
users' operations
ÿ USB Host: support storage of USB flash disc and upgrading of USB flash disc
system software
ÿ Certificate of approval
ÿ USB cable
Content Summary
The manual introduces related information about the operation of this series of digital
oscilloscope and comprises the following chapters and sections:
ÿ Chapter “Introduction” introduces the front panel, the user interface, the function
check and the probe compensation of the oscilloscope.
ÿ Chapter “System Prompt and Fault Recovery” introduces the system prompts
and lists some simple faults and solutions so that the users are capable to
rectify simple faults.
ÿ Chapter “Service and Support” introduces the warranty and technical support of this
series of product.
catalog
Chapter 1 Introduction ……………………………………………………………………………………………. …………one
1.3 Probe……………………………………………………………………………………… 6
2.2 Connector…………………………………………………………………………………….…………………11
…………………………………………………………………………………………………….20
2.7.3 Coupling……………………………………………………………………………………………………42
2.9.1 XY mode……………………………………………………………..…………………………50
2.10.3 Measurement………………………………………………………………………………………55
2.12.3 Self-correction……………………………………………………………….71
analysis …………………………………………………………………………………………… 89
Introduction one
The content below simply describes and introduces the front panel and the back part of this
series of digital oscilloscope so that you can know this series of digital oscilloscope well within
the shortest time.
This series of digital oscilloscope has a front panel that is simple but clear in function, and is
convenient for a user to finish basic operations. Knobs and functional keys are arranged on the
panel. Five gray keys which have been arranged in a row on the left side of a display screen are
option keys. The current menu can be provided with different options. The other keys are function
keys that can be set in different functional menus or directly obtain specific function application.
The knobs can be used for quickly regulating the corresponding setups of the oscilloscope.
Attention: in this specifications, ÿÿrepresents the keys and knobs of the oscilloscope,
and '''' represents the menu options displayed in a program interface.
Function Keys
Print Screen
Universal Knob
Menu Controls
Auto Setup
Advanced Trigger
ControlsHorizontal Controls
Vertical Controls
one
Machine Translated by Google
ÿ ÿÿ ÿ ÿ
ÿ
ÿ
ÿÿÿ ? ??? ?
no. Details
ÿ Trigger status
Armed: The oscilloscope is acquiring pre-triggering data. All triggers are ignored
in this state.
Ready: The oscilloscope has acquired all the pre-triggering data and is ready to
accept a trigger.
Trig'd: The oscilloscope has caught a single trigger and acquires the data after
trigger
Stop: The oscilloscope has stopped acquiring waveform data.
scan mode.
2
Machine Translated by Google
The horizontal [POSITION] knob is used for regulating the horizontal trigger
position
ÿ The identifier displays a zero electrical level standard point of the channel
waveform
3
Machine Translated by Google
12 3
4
Machine Translated by Google
2. Set a switch on a probe of the oscilloscope as ×1 and connect the probe with a connector
CH1BNC of the oscilloscope. Connect a hook-shaped head of the probe to a probe
compensation signal connector marked with ''1KHz'', and clamp a grounding hook
marked with “GND” by a grounding clamp, as shown in the figure
below:
3ÿPress [AUTO]. Within few seconds, CH1 displays a square wave of which the
frequency is 1kHz and the peak-to-peak voltage value is 3V.
4. Connect the probe with channel 2, and CH2 displays the same waveform after [AUTO]
is pressed.
5
Machine Translated by Google
1.3 Probe
1.3.1 Probe safety
Check and obey the rated values of the probe parameters before using it.
A protective device surrounding the probe main body can prevent fingers from electric
shock.
Probe protective
device
Connect the probe to the oscilloscope and ground the ground terminal before any
measurement.
Attention:
• Keep fingers behind the protective curve on the probe main body to prevent electric shock
when using the probe.
• Do not contact the metal part on the top of the probe when the probe is connected to a voltage
source.
• The signal measured by the oscilloscope uses ''ground''as reference voltage, and the ground
terminal should be grounded correctly to prevent short circuit.
Manually set the probe options, press down the vertical menu key, and then select the
“PROBE” option (such as [CH1] ÿ“PROBE”).
6
Machine Translated by Google
• When the “ATTENUATION” switch is set as ×1, the probe limits the bandwidth of the
oscilloscope within 0-10 MHz (different probes have different specifications).
Ensure that the switch is set to be ×10when to use the full bandwidth of the
oscilloscope.
one. Set the probe coefficient to 10X in channel menu, set the switch on the probe to ×10,
and connect the probe of the oscilloscope with channel 1.
2. Connect the end part of the probe to the probe compensation connector “1KHz”,
clamping the connector “GND” by the ground clamp, turn on the channel displayer,
and then press [AUTO] to display the waveform.
7
Machine Translated by Google
4ÿ If necessary, rotate the adjustable capacitor on the probe handle to realize suitable
compensated.
8
Machine Translated by Google
Function Introduction
and Operation
DIGITAL STORAGE OSCILLOSCOPE
2
This chapter introduces the functional keys and
operations of the front panel of the series of
oscilloscope in detail.
ÿ [SET TO 50%]: set the trigger electric level as midpoint of the signal amplitude.
ÿ [CURSORS]: display the “CURSOR” menu. The [UNIVERSAL] knob can be used for
regulating the position of the cursor when the “CURSOR” menu is displayed and
9
Machine Translated by Google
suitable waveform.
10
Machine Translated by Google
2.2 Connector
ÿ EXT TRIG: be used as an input connector of an external trigger source. Use [TRIG
MENU] to select “EXT” or “EXT/5” trigger source, and the trigger signal source
can be used for triggering in the third channel while acquiring data in two
channels.
ÿ This product is grounded by a protective ground wire of a power cord. For avoiding electric shock, please
ensure that the product is reliably grounded before connecting the input end or output end of the
product.
ÿ The ground wire of the probe is connected to the ground only. Please do not connect the
ground wire to high voltage.
11th
Machine Translated by Google
default setup
The [DEFAULT SETUP] key represents the default setup function, most of the options
and control setups of the factory are recalled by pressing them, some setups are not
changed, and the following setups are not reset:
• Language options
• Calibration data
12
Machine Translated by Google
Universal knob
13
Machine Translated by Google
Volt/div knob
As seen in the figure above, each channel has individual vertical menu key and knob for
regulating the vertical gear and offset. Moreover, the waveform display of the
input signal.
Coupling
Reject DC component of the input signal and an AC
HUNGRY
Bandwidth RUN Limit the bandwidth to 20MHz, and reduce the noise.
limit STOP
14
Machine Translated by Google
volts/div
fine tuning Fine tuning changes the resolution as small step in
1000X
this key to skip to the digital filter menu (seen in table 2-5).
digital filter
back Page2/2 Press the key to get back to page 1 of the menu.
Filter Upper Set the frequency upper limit using the [UNIVERSAL] knob
limit
Filter lower Set the frequency lower limit using the [UNIVERSAL]
15
Machine Translated by Google
limit knob
ÿ If the channel adopts a DC coupling mode, you can quickly measure the DC component of the
signal by observing the difference between the waveform and the signal ground.
ÿ If the channel adopts an AC coupling mode, the DC component in the signal is filtered. By
this mode, the AC component of the signal is displayed at a higher sensitivity.
ÿ If the channel adopts a GND coupling mode, cut off the input signal. Inside the channel, the
channel input is connected with a zero volt reference electric level.
By using CH1 as an example, the measured signal is a sine signal with DC offset:
component and the AC component contained in the measured signal can pass
AC identification DC identification
Figure 2-6 Set AC Figure 2-7 Set DC
and the AC component contained in the measured signal are rejected, as shown in
figures 2-8.
16
Machine Translated by Google
GND identification
Using Channel CH1 as an example, the measured signal is a pulse signal with
high-frequency oscillation:
ÿ Press [CH1]ÿ“Bandwidth limit”ÿ“OFF”, and set the bandwidth limit as OFF state.
Vertical gear regulation comprises two modes, including coarse tuning and fine tuning,
and the vertical gear range is 2mV/div ~ 5V/div when the probe is set as 1X .
17
Machine Translated by Google
To be matched with the probe attenuation coefficient, the probe attenuation proportion
coefficient should be accordingly regulated in the channel operation menu. If the probe
attenuation coefficient is 10:1, the proportion of the input channel of the oscilloscope
should be set as 10X so as to avoid the errors of the displayed gear information and
the measured data.
18
Machine Translated by Google
Probe coefficient
19
Machine Translated by Google
1. The knob is used for regulating vertical offset of the waveform of the channel in
where the knob is. The resolution is changed according to the vertical gear.
[Volt/div] knob
1. The knob can be used for regulating the vertical gear of the channel in where the
knob is so as to amplify or attenuate the signal of the channel waveform. the gear
information of the channel is displayed in the lower part of the screen.
2. The knob can be pressed to switch between “Coarse tuning” and “Fine tuning”, and
the sensitivity of the vertical gear is determined by a 1-2-5 stepping mode in coarse
tuning Fine tuning represents further regulation at the current gear so as to display
the waveform needed.
20
Machine Translated by Google
21
Machine Translated by Google
waveform operation
Result identification
Use FFT (Fast Fourier Transform) to transform a time domain (YT) signal into a
system.
ÿ Analyze vibration.
CH1
Signal source Select CH1or CH2 as signal source
CH2
Hanning
Hamming
windows Select window type FFT
Rectang
blackman
1×
10×
22
Machine Translated by Google
FFT Window
As the oscilloscope is used for FFT transformation on waveform recording with finite
length, the FFT algorithm is based on periodic signal. When the waveform period
number within the finite length is an integer, YT waveform has the same amplitude at
the start point and the end point, and no interruption is caused to the waveform. thigh
when the period number is not an integer, the waveform has different amplitude at the
start point and the end point, so that high-frequency transient interruption will be
caused in the joint. In a frequency domain, this effect is named as leakage. Therefore,
function, and forcibly make the values at the start point and the end point be 0.
Determine to use what kind of window according to the item and source signal
characteristics to be measured.
23
Machine Translated by Google
better frequency
resolution and worse
Sine, period and narrow-band random
Hanning amplitude resolution in
noise
comparison with
rectangular window
Frequency of windows
Transient or short pulse where the
Hamming is a little better
Hamming signal electric levels are considerably
than that of window
different there-before and there-after
Hanning
ÿ Access the signal to CH1 or CH2, and press [AUTO] to display a YT waveform.
waveform does not surpass the screen and the screen displays a plurality of
period.
ÿ According to Nyquist law, revolve [s/div] to make the sampling rate at least two
ÿ Press [MATH].
24
Machine Translated by Google
ÿ Regulate spectrum amplitude gear and offset at page 2 of the FFT function menu
so as to better observe FFT operation result.
Measure two items of the FFT spectrum: amplitude and frequency. the cursor
measurement takes 2.11.2 for reference.
Use the horizontal cursor to measure the amplitude and use the vertical cursor to
1. Measure FFT spectrum amplitude, and use CH2 as an example, the operation is as
follow:
5) revolve [s/div] to regulate the sampling rate (larger than double of the input
frequency).
25
Machine Translated by Google
10) Press “CurA”, and revolve the [UNIVERSAL] knob to move cursor A to the lowest
11) Press “CurB”, and revolve the [UNIVERSAL] knob to move cursor B to the highest
point of the FFT waveform.
12) The value of Delta V on the screen is the amplitude of the FFT waveform, as shown
in figures 2-20.
4) Press “CurA”, and revolve the [UNIVERSAL] knob to move cursor A to the highest
position of the FFT spectrum.
5) The value of CurA on the screen is the frequency of the highest point of the FFT
spectrum, and the frequency should be equal to the frequency of the input signal,
as shown in figure 2-21.
26
Machine Translated by Google
ÿ A signal with DC component or offset will cause error or offset of an FFT waveform
component. An AC coupling mode can be selected to reduce DC components.
ÿ DBVrms vertical scale can be used for displaying the FFT waveform within a large dynamic
range, and this scale displays the vertical amplitude at a logarithm mode.
ÿ Nyquist Function: Reconstruction of the waveform needs to adopt a sampling rate that is
double of the input frequency.
During actual measurement, the waveform can be compared with the reference
Condition that detailed circuit working point reference waveforms are provided.
Press [REF] to display the reference waveform menu, as shown in figure 2-22:
27
Machine Translated by Google
Operation procedure:
3. Revolve the vertical [POSITION] and knob [Volt/div] to regulate the vertical position and
the gear of the reference waveform.
6. Select “REF A ON” or “REF B ON” to recall the reference waveform, as shown in figure
2-23.
ÿ The horizontal position and gear of the waveform cannot be regulated at a reference
waveform state.
28
Machine Translated by Google
s/div
[HORI MENU]
Press this key to display the horizontal menu “HORI MENU”, and window expansion
can be implemented at this menu. Besides, horizontal displacement can be set by the
horizontal [POSITION] knob. M represents a main time base, and Z represents an
expanded timebase.
Option Description
main time base Horizontal time scale system of the oscilloscope
Define one window region by two cursors, and use the horizontal
window setup
[POSITION] and the knob [s/div] to regulate the window
29
Machine Translated by Google
region
Expand the window setup region to the whole screen to increase
window expansion the resolution relative to the main time base, so that the image
details can be conveniently checked.
The horizontal knob [s/div] changes the horizontal scale (time base), and the
[POSITION] knob changes the horizontal position (triggered displacement) triggered in
the internal memory. The center of the screen in the horizontal direction is a time
reference point of the waveform. Change of the horizontal scale will cause expansion
or contraction of the waveform relative to the center of the screen, while the change of
the horizontal position is relative to the position of a trigger point.
1. Regulate the horizontal position (trigger the position relative to the center of the
display screen) of the waveform (including MATH). The resolution of this control
knob is changed according to the time base.
2. Use the press down function of this knob to make the horizontal displacement return
to zero, namely back to the central position of the screen.
[S/div] knob
1. The knob is used for changing the horizontal time scale so as to conveniently
observe the most suitable waveform.
2. The knob is used for regulating the main time base. When the window expansion
mode is adopted, the knob is used for changing the expansion time base so as to
change the window width.
When the time base is set to be 100ms/div or more slowly and the trigger mode is set
to “Auto”, the oscilloscope enters the scan mode. At this mode, waveform display is
renewed from left to right. At the mode, no waveform trigger or horizontal position
control exists. The channel coupling should be set as direct current when a
30
Machine Translated by Google
3. Revolve [s/div ](to regulate the size of the window) and the horizontal [POSITION] (to
regulate the position of the window) to select the window of the waveform to be
observed, as shown in figure 2-26. The expansion time base cannot be slower than
the main time base.
Press the “Window” button after the window is set. At the moment, the waveform in the
selected window is expanded to the full screen to display. Figure 2-27 shows the
expansion result after the window is set.
31
Machine Translated by Google
the right point of the signal, which is very important to signal display. trigger control
can be used for stabilizing repeated waveform and acquiring single-pulse waveform.
After the trigger is correctly set, the oscilloscope can transform an unstable display
As shown in the figure below, one knob and three keys are in the trigger control region
(TRIGGER).
[TRIG MENU]
[LEVEL] knob
Set a signal voltage corresponding to the trigger point to the trigger electric level for
sampling. Press the knob to set the trigger electric level as zero electric level.
[SET TO 50ÿ]
Use the key to quickly stabilize the waveform. The oscilloscope could automatically
trigger the electric level as a center electric level of the signal. The key is very useful
and can be used for quickly setting the trigger electric level.
32
Machine Translated by Google
[FORCE]
No matter whether the oscilloscope detects the trigger, the key can be pressed to
perform force trigger once so as to finish acquisition of the current waveform. the key
Pre-trigger/delay trigger
The key is used for triggering the sampled data before/after the event.
The trigger position is generally set in the horizontal center of the screen, so that the
pre-trigger and delay information can be observed. The horizontal [POSITION] knob
check more pre-trigger information or delay trigger information. For instance, if burs
generated by the circuit are caught, the causes of generation of the burrs may be found
Attention: pre-trigger and delay trigger are invalid at a slow scan state.
The “Signal source” option can be used for selecting a signal as a trigger source for the
used for “Edge” trigger). The EXT/5 setup option is used for expanding the external
2.7.2 Types
This series of oscilloscopes provide five trigger types: Edge, Pulse, Video, Slope and
alternation.
1. Edge trigger: the most basic as well as the most common trigger type,
33
Machine Translated by Google
CH2 displayed
Does not display the trigger signal; the Ext option uses
the signal connected to the EXT TRIG front panel BNC
EXT
and allows a trigger level range of -12V
Signal to +12V
source
Same as Ext option, but attenuates the signal by a factor
EXT/5 of five, and allows a trigger level range of +6V to -6V. This
extends the trigger level range.
34
Machine Translated by Google
Operation steps:
3) Press “Signal source” to select “CH1”, “CH2”, “EXT”, “EXT/5” or “AC Line” according to
4) Press “Slope” to select “Rise edge”, “Fall edge” or “Rise and fall edge”.
trigger condition.
Normal: Refresh the waveform when the waveform satisfies the trigger
condition, otherwise, do not refresh the waveform and wait for
Single: Acquire the waveform once when the waveform satisfies the trigger
35
Machine Translated by Google
b. Press “Coupling” option key to select “DC”, “AC”, “HF rejection” or “LF rejection”.
CH1
EXT/5
36
Machine Translated by Google
Trigger mode Normal applicable to application of most of the “Pulse width” trigger.
single
Operation description:
3) Press the “Condition” option key to select “ ” (positive pulse width <),
“
” (positive pulse width >), “ ” (positive pulse width =),
“
” (negative pulse width <), “ ” (negative pulse width >) or
Press “Next Page 1/2” to skip to page 2 of the pulse trigger menu, and set the
37
Machine Translated by Google
trigger mode and the trigger coupling like the setup of the edge trigger.
Video trigger: Perform field or row video trigger on a standard video signal.
CH2
Signal source See the signal source in 2.7.1.
EXT
EXT/5
standard
Trigger under negative synchronous pulse
polarity
inverse phase Trigger under positive synchronous pulse
all rows
Specified row
Synchronization Select suitable video for synchronization
odd field
even field
Figure 2-31 Page 1 of video trigger menu Figure 2-32 Page 2 of video trigger menu
38
Machine Translated by Google
Operation description
Setup of the signal source of the video trigger is like that of the edge trigger.
4) Press “Synchronization” to select “All row”, “Specified row”, “Odd field” or “Even
field”. Revolve the [UNIVERSAL] knob to set the number of the specified rows if
Slope trigger
39
Machine Translated by Google
Type Slope
CH1
Signal CH2
source EXT See the signal source in 2.7.1.
EXT/5
positive slope>
positive slope<
positive slope= Conditions of signal slope relative to set
Conditions
negative slope> slope (set by time)
negative slope<
negative slope=
Revolve the [UNIVERSAL] knob to set the
Figure 2-33 Page 1 of slope trigger menu Figure 2-34 Page 2 of slope trigger menu
Type Slope
upper boundary
vertical lower boundary Select boundary, and regulate the size of the
windows upper and lower window by the [LEVEL] knob.
boundaries
40
Machine Translated by Google
auto
single
Operation description:
2. Press [AUTO].
7. Press “Time setup”, and revolve the [UNIVERSAL] knob to regulate the slope time.
10. Revolve the trigger electric level [LEVEL] knob until the waveform is stably
triggered.
alternating trigger
When alternating trigger is selected, the trigger signal comes from two channels. the
mode is used for simultaneously observing two unrelated signals. Two different triggers
types (edge, pulse, video and slope) can be selected for the signals from the two
channels. During setup, the trigger types of the two channels and the trigger electric
level information are respectively displayed in the right lower corner of the screen. the
setups of the four trigger types are seen in 2.7.2.
41
Machine Translated by Google
Operation description:
As shown in figure 2-35, the following steps can be carried out for simultaneously
2. Press [AUTO].
2.7.3 Coupling
The “Coupling” option is used to determine which part of signal passes through the trigger
circuit. It is conductive to stably display the waveform. Press [TRIG MENU] to
use trigger coupling. Select the “Coupling” option in the “Setup” menu after selecting
the trigger type, including DC coupling, AC coupling, HF rejection and LF rejection. the
functions of all the options are specifically shown in table 2-14.
42
Machine Translated by Google
The trigger hold-off function is used for stably displaying complex waveform. the
hold-off time represents a time interval between twice trigger detected by the
oscilloscope. During the hold-off period, the oscilloscope performs no trigger. ace
shown in figure 2-36 below, regulate the hold-off time according to one pulse
sequence so that the oscilloscope only triggers at the first pulse in the row.
43
Machine Translated by Google
Acquisition Peak value For detecting burrs and reducing “Fake wave
mode detection phenomenon”.
Average time
( 4, 16, 32, 64 Select average time.
, 128 , 256 )
Interpolation cine Enable sine interpolation
type Linear Enable linear interpolation
real time Set the sampling mode as real-time sampling
Sampling mode
sampling
Sampling rate Display the sampling rate of the system
ÿ Sampling: The oscilloscope samples the signal at uniform time intervals to form a waveform.
44
Machine Translated by Google
Figure 2-38 Sampling mode Figure 2-39 Peak value detection mode
ÿ Peak value detection: The oscilloscope finds out the maximal value and the
minimal value of the input signal in each sampling interval and uses these values
to display the waveform.
Advantage: spike pulses that may be missed can be acquired and displayed and
signal confusion can be avoided at the mode.
Shortcoming: Loud noise is displayed at the mode.
ÿ Average value: The oscilloscope acquires a plurality of waveforms and displays the
final waveforms after averaging the waveforms.
Advantage: Random or unrelated noises in the displayed signal can be reduced at
the mode. The signal shown in figure 2-40 has loud noises, while the signal shown
in figure 2-41 adopts the average mode, so the noises are greatly reduced. What
calls for attention is that: the higher the average time is, the better the waveform
quality is, but the slower the refreshing speeds of the waveform is.
45
Machine Translated by Google
ÿ Real-time sampling: The real-time sampling mode realizes suffusion of the storage
space during each sampling. The real-time sampling rate is at
most 1GSa/s.
ÿ [RUN/STOP]: Press the key to start continuous data acquisition or stop acquisition.
ÿ [SINGLE]: Press the key after the oscilloscope detects the trigger and stop after wave
acquisition is finished once.
Figure 2-42
46
Machine Translated by Google
At a small time base gear (50ns/div or smaller), there are few original sampling points
in one period and the displayed waveform has bad quality, so an interpolation function
is adopted to increase “sampling point” and rebuild the waveform, so that the
measurement precision is improved. Sine interpolation is suitable for sine, while linear
interpolation is suitable for triangular wave, square wave and similar. As shown in
figure 2-43 below, after sine interpolation, the displayed waveform is good, while figure
2-44 shows the waveform after linear interpolation of a sine signal, absolutely, the
waveform is bad, and the interpolation function is not selected rightly.
47
Machine Translated by Google
Type
point
No interpolation link line is displayed between the sampling
points. Display sample points directly
OFF
1 second
Infinite
grid
Set the grid brightness by the [UNIVERSAL] knob.
brightness
Figure 2-45 Page 1 of the display menu Figure 2-46 Page 2 of the display menu
48
Machine Translated by Google
5s
menu display 10s Set the duration time of the menu display.
20s
Infinite
Operation description:
Press [DISPLAY] to skip to the “DISPLAY” menu, and press “Type” to select “Vector” or
"Point".
Set continue:
Press “Continue” to select “OFF”, “1s”, “2s”, “5s” or “Infinite”. By utilizing the option,
Figure 2-47 Utilize the continue option to observe amplitude modulated wave
49
Machine Translated by Google
Press “Wave brightness”, and revolve the [UNIVERSAL] knob to regulate the display
Press “Grid brightness”, and revolve the [UNIVERSAL] knob to regulate the display
Press “ Next Page1/2” to skip to page 2 of the display menu. Press “Format” to select
“YT” or “XY”.
Press “Screen” to select “Normal” or “Inverse” to set the color of the screen.
Press the “Menu display” option key to select “2s”, “5s”, “10s”, “20s” or “Infinite” to
set the maintained display time length of the menu on the screen.
2.9.1 XY mode
Use the XY format to analyze phase difference. At the format, the voltage of channel 1
determines coordinate X (horizontal) of a point, while the voltage of channel 2
determines coordinate Y (vertical) of the point. The oscilloscope uses a non-triggered
sampling mode to display the data as a spot. Figure 2-48 shows the YT mode, and it
shows that signals of the two channels have the same amplitude and frequency and
the phase difference is 90 degrees. After switching to XY mode, the waveform is
shown in figure 2-49 below.
50
Machine Translated by Google
ÿ The oscilloscope can acquire the waveform according to a normal YT mode at any
sampling rate and can check the corresponding waveform at XY mode.
ÿ The [Volt/div] and the vertical [POSITION] of channel 1 control the setup of the
horizontal scale and position.
ÿ The [Volt/div] and the vertical [POSITION] of channel 2 control the setup of the
vertical scale and position.
ÿ Revolve the [s/div] knob to regulate the sampling rate so as to observe the
waveform better.
ÿ Cursor
ÿ Trigger control
At the vector display mode, the oscilloscope connects the sampling points in a digital interpolation
manner, and the interpolation manner includes linear interpolation and sine interpolation. The sine
interpolation manner is suitable for the real-time sampling mode and is available at a time base of
50ns or faster at the real-time sampling mode.
51
Machine Translated by Google
By using the method, estimation can be made quickly and intuitively. for instance,
waveform amplitude can be observed, and a probable measurement result is judged
according to the vertical scale. The method realizes simple measurement by
multiplying the vertical scale number of the signal with the vertical gear Volt/div.
The cursor measurement includes two modes: Manual mode and Tracking mode.
1. Manual mode:
Horizontal cursors or vertical cursors appear in pair to measure time or voltage, and
the distance between the cursors can be manually regulated. The signal source should
be set as a waveform to be measured before the cursors are used.
2. Tracking mode:
A horizontal cursor is intersected with a vertical cursor to form a cross cursor. the
cross cursor is automatically located on the waveform, and the horizontal position of
the cross cursor on the waveform is regulated by selecting “Cur A” or “Cur B” and
rotating the [UNIVERSAL] knob. The coordinates of the cursor point will be displayed
on the screen of the oscilloscope.
52
Machine Translated by Google
parameters.
CH1 Select the input channel of the measured signal.
CH2
REF A
REF B
The manual cursor measurement mode is used for measuring the coordinate values
and increments of one pair of horizontal or vertical cursors. Ensure the signal source to
ÿ Voltage cursor: The voltage cursor appears on the display screen as a horizontal
ÿ Time cursor: The time cursor appears on the display screen as a vertical line, and
ÿ Cursor movement: Select the cursors first and use the [UNIVERSAL] knob to move
4. Press “Signal source” to select CH1, CH2, MATH, REF A or REF B according to the
5. Select “Cur A” and revolve the [UNIVERSAL] knob to regulate the position of
53
Machine Translated by Google
cursor A.
6. Select “Cur B” and revolve the [UNIVERSAL] knob to regulate the position of
cursor B.
7. Display the measured values on the left upper corner of the screen.
Figure 2-50 Cursor manual mode Figure 2-51 Cursor tracking mode
No cursors
At cursor tracking measurement mode, the cross cursors are displayed on the
measured waveform, the cursors are automatically located on the waveform by moving
the horizontal position between the cursors, and simultaneously the horizontal and
vertical coordinates of the current located point, and the horizontal and vertical
increments between the two cursors are displayed. The horizontal coordinate is
displayed as a time value, and the vertical coordinate is displayed as a voltage value,
54
Machine Translated by Google
3. Press “Cursor A”, and select the input channel CH1 or CH2 of the tracked signal.
4. Press “Cursor B”, and select the input channel CH1 or CH2 of the tracked signal.
5. Select “Cur A”, and rotate the [UNIVERSAL] knob to horizontally move cursor A.
6. Select “Cur B”, and rotate the [UNIVERSAL] knob to horizontally move cursor B.
7. Display the measured values on the left upper corner of the screen:
AÿT : Position (namely the time based on the horizontal central position) of cursor A
in the horizontal direction.
AÿV : Position (namely the voltage based on the grounded point of the channel) of
cursor A in the vertical direction.
BÿT : Position (namely the time based on the horizontal central position) of cursor B
in the horizontal direction.
BÿV : Position (namely the voltage based on the grounded point of the channel) of
cursor B in the vertical direction.
Delta T : Horizontal distance (namely the time value between the two cursors) of
cursor A and cursor B.
Delta V : Vertical distance (namely the voltage value between the two cursors) of
cursor A and cursor B.
2.10.3 Measurement
measure
55
Machine Translated by Google
measurement and delay measurement; and there are 32 measurement minor items in
total. At most five items can be displayed once. Press [MEASURE] to skip to the
measurement menu to display the measurement result at first, as shown in figure 2-53,
and press any option key to change the measurement type, as shown in figure 2-54.
Option Description
voltage Press the key to skip to the voltage measurement menu.
measurement
time Press the key to skip to the time measurement menu.
measurement
delay Press the key to skip to the delay measurement menu.
measurement
All measurement Press the key to skip to the all measurement menu.
back Press the key to get back to the measurement result.
56
Machine Translated by Google
menu 1
57
Machine Translated by Google
time parameters.
measurement stop
Stop all measurement on the time type
parameters.
58
Machine Translated by Google
Amplitude Voltage between the top value and the bottom value of the
waveform
Arithmetic mean value of the waveform in the first
Periodic average value
period
average value Calculate the arithmetic mean voltage in the whole recording.
periodic mean Namely an effective value. Calculate the actual mean square
root value of the waveform in the first complete period.
square root
mean square root Actual mean square root voltage of the whole waveform
FOVERShoot Ratio of the difference of the minimal value and the bottom
value of the waveform to the amplitude after falling
rise time Measure the time between 10% and 90% of the
fall time Measure the time between 90% and 10% of the first fall
positive duty ratio Ratio of the positive pulse width to the period
negative duty ratio Ratio of the negative pulse width to the period
represent the
59
Machine Translated by Google
FRR Time between the first rise edges of signal source 1 and
signal source 2
FRF Time between the first rise edge of signal source 1 and
the first fall edge of signal source 2
FFR Time between the first fall edge of signal source 1 and
the first rise edge of signal source 2
FFF Time between the first fall edges of signal source 1 and
signal source 2
LRR Time between the first rise edge of signal source 1 and
the last rise edge of signal source 2
LRF Time between the first rise edge of signal source 1 and
the last fall edge of signal source 2
LFR Time between the first fall edge of signal source 1 and
the last rise edge of signal source 2
LFF Time between the first fall edge of signal source 1 and
the last fall edge of signal source 2
2. Press the first option key to skip to the measurement function menu 1 in a
4. Press “Signal source”, and select CH1 or CH2 according to the signal input channel.
5. Press “Measurement type” or revolve the [UNIVERSAL] knob to select the voltage
parameters to be measured.
The corresponding icons and parameter values will be displayed in the menu
60
Machine Translated by Google
6. When get back to the measurement result region in the homepage, the selected
parameters and the corresponding values will be displayed in the first option.
Other option measurement types can be changed by the same method, and the
homepage can display five parameters every time.
The voltage parameters can be measured by using the all measurement function by the
following steps:
1. Press [MEASURE] to skip to the “MEASURE” menu.
2. Press any option key in the homepage to skip to the measurement function menu 1.
5. Press “Voltage test” to select “Run”. At the moment, all the voltage parameter
values will be simultaneously displayed on the screen. (As shown in figure 2-56)
61
Machine Translated by Google
Type Description
image save is only saved in a USB flash disk and opened by computer
software.
the oscilloscope;
The saved image is the whole screen image (including waveform, menu and the like);
save/recall setup
ÿ Save types
The save types include setup save, waveform save, image save, CSV and factory
setups, wherein the setups and waveforms can be saved and then recalled, the images
and CSV are saved only, and the factory setups are recalled only.
62
Machine Translated by Google
As shown in figure 2-58, the setups are saved in equipment (the equipment is the
oscilloscope), and 20 groups of data (NO.1-NO.20) can be saved. The save steps are as
follow:
5. Press “Recall” to recall the setup data from the save position.
“Data is read successfully”: The setups are successfully recalled and executed; “Blank
63
Machine Translated by Google
1. Press [SAVE/RECALL] and select the “Setup save” type, as shown in figure 2-59.
2. Plug the USB flash disc, and prompt “USB save equipment is connected
successfully” after the connection is normal.
The data is saved in root directory of the USB flash disc, and the file name is defaulted
as GASxxxx.SET, such as GAS0001.SET.
1. Press [SAVE/RECALL], and select the “Setup save” type, as shown in figure 2-60.
2. Plug the USB flash disc, and prompt “USB save equipment is connected
successfully” after the connection is normal.
4. Press “Recall” to skip to a USB flash disc file interface, as shown in figure 2-60.
6. Press “Recall” to recall the setup data from the save position.
“Data is read successfully”: The setups are successfully recalled and executed; “Blank
save unit”: File is corrupted.
64
Machine Translated by Google
Press “Back” to the waveform display interface if the current USB flash disc has no
setup files.
ÿ The save/recall waveform acquisition step is the same as the save/recall setup.
The oscilloscope enters “STOP” state after the saved waveform is recalled and
displayed.
ÿ Image save
As shown in figure 2-61, waveform images can be saved in the USB flash disc but
cannot be recalled by the oscilloscope. The images are in standard BMP format and
The images are saved in the root directory of the USB flash disc, and the file name is
[PRINT] print button: It is a shortcut key and can immediately save the screen image in
“Save”: It can be used for saving the screen image in a special menu (see the menu bar
65
Machine Translated by Google
ÿ CSV save
As shown in figure 2-62, the following steps can be carried out to save the CSV file in
the USB flash disk:
3. Plug the USB flash disc, and prompt “USB save equipment is connected
The data is saved in the root directory of the USB flash disc, and the file name is
defaulted as GASxxxx.CSV, such as GAS0001.CSV.
66
Machine Translated by Google
When the save type is set as factory setup, press the “Recall” key to restore the
67
Machine Translated by Google
68
Machine Translated by Google
Figure 2-65 Page 3 of the auxiliary menu Figure 2-66 Page 4 of the auxiliary menu
69
Machine Translated by Google
stop
Next Page 4/4 Press the option key to skip to page 1.
Select “System state” in the [UTILITY] menu to display the system state. As shown in
figure 2-67, the system state includes software and hardware versions, local model
This series of oscilloscope is provided with a plurality of languages that can be selected by users.
To select a display language, press [UTILITY], press “Language type” in page 1 of the “UTILITY” menu,
and switch the language menu for display. Figure 2-68 shows an English menu.
70
Machine Translated by Google
For self correction, please disconnect all the probes or wires from the input connector, then
press [UTILITY], press “ Self correction” in page 2 of the “UTILITY” menu, and run
the self correction program according to a screen prompt.
2.12.4 Self-test
Press [UTILITY], and press “Self test” in page 2 of the “UTILITY” menu. The self-test
includes screen test, keyboard test and lighting test.
Operation steps:
1. Screen test:
Select “Screen test” to skip to a screen test interface, as shown in figure 2-70. at the
moment, the screen displays a prompt message of “Press SINGLE key to continue,
Press RUN/STOP key to exit”, namely start the test when prompting to press [SINGLE].
Different colors are displayed on the screen. Observe whether the screen has serious
color cast or other display errors.
71
Machine Translated by Google
2. Keyboard test
Select “Keyboard test” to skip to a keyboard test interface, as shown in figure 2-71. A
rectangular region in the interface represents the key at the corresponding position on
the panel; a lathy rectangle represents the knob at the corresponding position of the
panel; and a square represents the press-down function of the corresponding knob.
Test all the keys and knobs, and observe whether the keys and knobs respond
correctly.
ÿ Tested keys are in green, and the knob region is in red, such as “-16” in the figure,
counterclockwise rotation represents “-”, clockwise rotation represents “+”, and the number
represents the number of revolution.
ÿ A prompt message of “ Press RUN/STOP key three times to exit” is displayed at the bottom
of the screen to remind the user to exit the test method.
3. Lightening test
Select “Lightening test” to skip to a lighting test interface, as shown in figure 2-72.
At the moment, the screen displays a prompt message of “Press SINGLE key to
72
Machine Translated by Google
continue, Press RUN/STOP key to exit”. After continuously pressing the [SINGLE] key,
the corresponding region on the screen will be in green when the key is lightened. RUN
and STOP share one key, so the key is in green at RUN state and in red at STOP state.
This series of oscilloscope can upgrade the software by the USB flash disc, which
needs about 5 minutes.
1. Plug the USB flash disc in which a firmware program is saved in USB Host interface
3. Press the “Next” option key to skip to page 3 of the auxiliary menu.
Reboot the machine after finishing upgrading, and the software version is upgraded.
ÿ Upgrading should be performed again by rebooting the machine if power is off or upgrading
is failed during upgrading.
2.12.6 Pass/fail
“Pass/fail” is used for judging whether the input signal is in a built rule range and
73
Machine Translated by Google
outputting the past or failed waveform so as to detect the change condition of the
signal.
run pass/fail
Operation
ÿ stop pass/fail
Figure 2-73 Page 1 of pass/fail menu Figure 2-74 Page 2 of pass/fail menu
74
Machine Translated by Google
Figure 2-75 Page 1 of rule setup menu Figure 2-76 Page 2 of rule setup menu
75
Machine Translated by Google
5) Press “Signal source selection” to select the signal input channel. In figure 2-73
CH2 is selected.
7) Press “Rule setup” to skip to page 2 of the rule setup menu, as shown in figure 2-75.
8) Press “Horizontal regulation” and “Vertical regulation”, and use the [UNIVERSAL]
knob to regulate the horizontal tolerance range and the vertical tolerance range.
9) Press “Build rule” to build a rule template, or press the “Recall” key in the next page
of the menu to recall the saved rules.
10) Get back to skip to page 2 of the pass/fail menu, and set the “Output” option as
pass.
11) Get back to page 1 of the pass/fail menu, and press “Operation” to start.
As shown in figure 2-77, after the rules are built completely, PASS starts counting if
the signal of CH2 satisfies the rules during start, and Pass would stop counting
while Fail would start counting when the signal surpasses the set template (the
amplitude of the signal in the figure is smaller).
Pass/Fail output
The pass/fail function can be used for outputting a negative pulse train through a
Pass/Fail BNC interface on a rear panel of the oscilloscope.
76
Machine Translated by Google
The waveform recording function can be used for recording waveforms input by
channel 1 and channel 2. The user can set the time interval of the frames within
1ms-999s. At most 1000 frames of waveforms can be recorded. The waveforms can be
replayed after being recorded.
Waveform recording: Record the waveforms at a specified time interval until reaching
the set end frame number.
As shown in figure 2-78, the operation steps of waveform recording are as follows:
77
Machine Translated by Google
5. Press the “Signal source” key to select the signal channel to be recorded.
6. Select the “Time interval” option, and use the [UNIVERSAL] knob to regulate the time
interval of frame to frame in waveform recording.
7. Select the “End frame” option, and use the [UNIVERSAL] knob to regulate the
maximal frame number in the waveform recording.
78
Machine Translated by Google
during replay, the current frame number is displayed on the screen; after replay is
stopped, the [UNIVERSAL] knob is used to observe all frames of waveforms between
the start frame and the end frame.
The current recorded waveforms can be replayed by carrying out the following
steps: 1. Press [UTILITY] to skip to the [UTILITY] menu.
4. Press “Time interval” to set the time interval of the replayed waveform frame to
frame
6. Press “Start frame”, and revolve the [UNIVERSAL] knob to regulate the start frame
number of the replayed waveforms.
7. Press “End frame”, and use the [UNIVERSAL] knob to regulate the end frame number
of the replayed waveforms.
8. Press the “Back Page 2/ 2” option key to get back to page 1 of the waveform replay
function menu.
The interface setup is used for setting RS-232 Baud rate, and the Baud rate can be set
as 300, 2400, 4800, 9600, 19200 or 38400.
79
Machine Translated by Google
[HELP] is a help function key, and the user can enter or exit the help state by pressing
the key. The user can recall the corresponding help information by pressing the keys
when entering the help state.
The submenus in each main menu have the corresponding help information. Attention:
in order to check the help information of the options in the next page of the submenu,
please exit the help state at first, switch to the next page of the menu, then enter the
help state again, and then press the option key to check the corresponding help
information. Figure 2-81 shows the help information of [CH1].
80
Machine Translated by Google
Application examples
DIGITAL STORAGE OSCILLOSCOPE
3
This chapter mainly introduces several application
your
ÿ Simple measurement
ÿ Cursor measurement
ÿ Application of XY function
signal analysis
ÿ Screen capture
Machine Translated by Google
Machine Translated by Google
The oscilloscope can be used for automatically measuring most of the display signals,
and the frequency and peak-to-peak values of the signal are measured according to
the following steps:
81
Machine Translated by Google
The corresponding icons and measured values will be displayed in the third option.
ÿ Display of the measurement results on the screen is changed with the change of the
measured signal.
ÿ Please regulate [Volt/div] or [s/div] if the reading of the “value” is displayed as ****.
82
Machine Translated by Google
5. Press the “CurA” option key, and revolve the [UNIVERSAL] knob to place the cursor A
at an edge of the spike pulse.
6. Press the “CurB”option key, and revolve the [UNIVERSAL] knob to place the cursor B
at an edge that is clearest to the spike pulse.
As shown in figure 3-1ÿtime increment Delta T and frequency increment 1/Delta T (measured
the width of spike pulse) will be displayed in the left upper corner of the display screen.
Carry out the following steps to measure the amplitude of the spike pulse:
83
Machine Translated by Google
5. Press the “CurA”option key, and revolve the [UNIVERSAL] knob to place the cursor A
at the highest wave peak of the spike pulse.
6. Press the “CurB”option key, and revolve the [UNIVERSAL] knob to place the cursor B
at the lowest point of the spike pulse.
ÿ Voltage at cursor A
ÿ Voltage at cursor B
84
Machine Translated by Google
first, otherwise, observe the signal at first at an automatic trigger mode or at a normal
trigger mode so as to determine a suitable trigger electric level and a trigger edge.
1. Set the attenuation coefficient as 10× in the probe option, and set the switch on the
probe as × 10.
(2) Respectively set the type as “EDGE TRIGGER”, set the signal source as [CH1], set the
slope as “RISE EDGE”; set the trigger mode as “ONCE”, and set the coupling as
“DIRECT CURRENT”.
(3) Regulate the horizontal time base and the vertical gear to a suitable range.
(5) Press [RUN/STOP] to wait appearance of a signal satisfying the trigger condition,
acquire the signal once and display the signal on the screen.
By this function, occasional events can be easily caught, such as sudden burrs with large
amplitude; the trigger electric level is set to be an electric level that is a little larger than a
normal signal; the key [RUN/STOP] is pressed to wait, and the machine automatically carries
out triggering and records the waveform within a period before and after triggering when burs
appear. The [POSITION] panel is revolved to observe the waveform before appearance of the
burrs.
85
Machine Translated by Google
displayed on the oscilloscope, and the signal may contain much information that
the circuit, and the following steps can be carried out for analyzing the noise better:
2. Press the “ACQUIRE MODE” key or revolve the [UNIVERSAL] knob to select “PEAK
VALUE DETECTION”.
Carry out the following steps to reduce random noise in the display screen:
2. Press the “ACQUIRE MODE” option key or revolve the [UNIVERSAL] knob to select
“AVERAGE VALUE”.
3. Press the “AVERATE TIME ” option key to check the influence caused to the waveform by
changing the frequency of operating average operation, wherein the average operation can
reduce the random noise and make the detailed information of the signal be easily checked.
86
Machine Translated by Google
4. Connect the probe of channel 1 to an input point of the network, and connect the
probe of channel 2 to an output point of the network.
5. Press [AUTO].
6. Revolve [Volt/div] to make that the signal amplitudes displayed on the two channels
are roughly the same.
The phase difference angle theta is equal to +/- arcsin (A/B) or theta is equal to +/-
arcsin (C/D) according to the formula sin theta=A/B or C/D, wherein theta represents
the phase difference angle between the channels, and A, B, C and D are defined in the
figure above.
87
Machine Translated by Google
The two signals have the same amplitude if the frequency or phase difference of the
two measured signals is integer multiples of pi/4. The Lissajous pattern of the special
signals is shown in figure 3-4, wherein m:n is a frequency ratio, K represents a phase
difference coefficient, and theta=K x pi/4. Figure 2-49 is a measured Lissajous pattern
when the two signals have the same frequency and amplitude and the phase difference
is pi/2.
88
Machine Translated by Google
It is a difference signal, and the waveform can be displayed better by using the
arithmetical function of the oscilloscope.
The following steps can be carried out by activating the difference signal connected to
channel 1 and channel 2:
4. Press [AUTO].
7. Select “CH1-CH2”, and represent the difference of the waveforms of channel 1 and
channel 2 by the arithmetical operation results.
8. Press the fourth and fifth function options to regulate the vertical scale and position
of the arithmetical operation waveform.
ÿ Two probes should be compensated at first, and the differences of the probe
compensation may cause errors of the measurement results.
89
Machine Translated by Google
2. Select “PRINT BUTTON” as “STORED IMAGE”. Set the [PRINT] shortcut key as
4. Regulate the vertical gear [Volt/div] and the time base gear [s/div] to display the best
waveform
5. Plug the USB flash disc, and notice that “USB MEMORY IS CONNECTED
SUCCESSFULLY” if the connection is normal.
6. Press [PRINT] to carry out storage, and notice that “DATA IS STORED
SUCCESSFULLY”.
90
Machine Translated by Google
ÿ Horizontal position reaches the limit: It prompts that the horizontal movement
position has already reached the limit and cannot be regulated any more.
ÿ Voltage gear reaches the limit: It prompts that the vertical voltage gear has already
reached the lowest gear 2mV/div or the highest gear 5mV/div and cannot be regulated
any more.
ÿ Voltage displacement reaches the limit: The system will pop up this prompt message
if channel offset is regulated to be too high or too low by revolving the knob
[VERTICAL POSITION].
ÿ Time base gear reaches the limit: The system will prompt this message if the horizontal
time base is still regulated after being regulated to the fastest or
slowest gear.
ÿ Hold-off time reaches the limit: The system will prompt this message when the hold-
off time is regulated to a smallest or largest value by revolving the knob [UNIVERSAL].
ÿ Function unavailable: The system cannot set some functions at some special
modes.
ÿ No signal: This message is prompted when the signal does not satisfy the automatic
setup condition during automatic setup.
ÿ Setup reaches the limit: Regulation cannot be continued if the pulse width
reaches the minimal value 20.0ns or the maximal value 10.0s when the knob
[UNIVERSAL] is used to regulate the pulse width.
ÿ Blank storage unit: This message appears when the option key [CALLOUT] is pressed
if no waveform or setup is stored at the storage position in waveform
storage or setup storage.
ÿ USB storage device is connected successfully: This message will appear when
the USB flash disc is plugged in a USB Host interface of the oscilloscope.
ÿ USB storage device is disconnected: This message will appear when the USB flash
disc is unplugged.
91
Machine Translated by Google
ÿ Data is stored successfully: It means that the setup data, waveform data and
image data are already stored in the oscilloscope or the USB flash disc.
ÿ Data is read successfully: It means that the setup data or the waveform data is
successfully called out from the oscilloscope or the USB flash disc.
ÿ USB storage device is unconnected: This message will be prompted if the USB
flash disc is not plugged for storage when the option “store to” is set as “USB
flash disc” or the option “PRINT BUTTON” is set as “stored image ” in the menu
“SAVE/RECALL”.
ÿ Waveform recording is finished: The system will prompt this message when
waveform recording is ended.
92
Machine Translated by Google
(4) Please contact Gratten if the product still cannot be used properly.
2. Treat the fault according to the following steps if the waveform of the signal does
not appear in the picture after the signal is acquired:
(1) Check whether the probe is normally connected to the signal connecting wire.
(2) Check whether the signal connecting wire is normally connected to the BNC.
(3) Check whether the probe is normally connected with a to-be-measured signal.
3. Check whether the channel attenuation coefficient accords with the probe
attenuation ratio actually used if the measured voltage amplitude is 10 times larger
or 10 times smaller than the actual value.
(1) Check whether the signal source option of the trigger panel accords with the
signal channel actually used.
(2) Check the trigger type: general signals apply “EDGE TRIGGER” mode, video
signals apply “VIDEO TRIGGER” mode, and the waveform is only stably
displayed when suitable trigger mode is applied.
Please check whether the trigger mode of the trigger panel is at the “NORMAL” or
“SINGLE” position and whether the trigger electric level surpasses the waveform
range, if so, center the trigger electric level, or set the trigger mode as “AUTO”
93
Machine Translated by Google
6. If the display speed is lower after starting the average sampling mode or setting a
long afterglow time.
It is normal.
(2) The horizontal time base gear may be too low, and the display can be improved
by increasing the horizontal time base to increase the horizontal resolution.
(3) The display type may be “VECTOR”, and the linked line between the sampling
points may cause ladder-like display of the waveform. This problem can be
solved by setup the display type as “POINT” display mode.
94
Machine Translated by Google
Please contact the nearest sales and service agency of Gratten if needing services or
requesting a complete copy of the warranty bill.
95
Machine Translated by Google
Machine Translated by Google
appendix
DIGITAL STORAGE OSCILLOSCOPE
6
This chapter introduces the technical specifications
and default settings of the GA1000 series of
oscilloscope.
Machine Translated by Google
Machine Translated by Google
are set as ×10 and this series of digital oscilloscope. To check whether the
oscilloscope satisfies the technical specifications, the oscilloscope should satisfy the
ÿ The oscilloscope should run for at least 30min within a specified operation
temperature range.
operation range is 5 DEG C or higher, and this operation can be carried out by
the “AUXILIARY FUNCTION” menu.
All the specifications are ensured to satisfy the requirement except that marked with
“TYPICAL” sign.
Technical specifications
Input
Maximum input
400V (DC+AC peak value, 1Mÿ input impedance)
voltage
96
Machine Translated by Google
vertical system
channel amount 2
300MHz(only to GA1302CAL)
bandwidth 200MHz(only to GA1202CAL)
100MHz(only to GA1102CAL)
AC plus accuracy 2mV/div ÿ±4%, the rest gears ÿ±3%
DC measurement
±[DC measurement accuracy x reading+ (1% x vertical
accuracy displacement reading) +0.2div]
horizontal system
(only to GA1302CAL)
horizontal
100div
displacement range
97
Machine Translated by Google
XY mode phase
±3 Degrees
difference
trigger system
CH1, CH2:±10div
Trigger electric level
EXT: ±1.5V
range
EXT/5: ±7.5V
EXT/5: ÿ0.75V
measurement system
automatic
Maximal valve, minimal value, peak-to-peak value,
98
Machine Translated by Google
Reading resolution
6 bits
ratio
General specifications
display system
99
Machine Translated by Google
background intensity
300 Cd/m2
(typical)
waveform display
14×8 grids
range
power supply
AC power supply
45Hz to 440Hz
frequency range
environment
Operating: 10 ºC to +40 ºC
temperature
Non operating: -20 ºC to +60 ºC
cooling Forced cooling of fans
Mechanical specifications
To you 339mm×149mm×111mm (width x height x depth)
one hundred
Machine Translated by Google
volts/div 200mV
Operation CH1+CH2
FFT operation
horizontal
1×
amplification
position 0.0s
Type OFF
Signal CH1
CURSOR
voltage ±2G
time ±2G
101
Machine Translated by Google
average value
Amplitude
period
rise time
type vector
go on OFF
Format YT
Set No.1
REFA/REFB REFA
REF
REFA OFF
REFB OFF
102
Machine Translated by Google
Type edge
Slope rise
TRIGGER (EDGE)
Trigger mode auto
Type pulse
TRIGGER Condition
Type Video
TRIGGER polarity
standard NTSC
Condition
103
Machine Translated by Google
Type alternation
104
Machine Translated by Google
Do not let the liquid crystal displayer be from direct irradiation of sunshine
when storing or placing it.
Attention: Do not place the instrument or probe in fog, liquid or solvent for avoiding
damage of the instrument or probe.
Cleaning:
Usually check the instrument and the probe according to operation condition, and
clean the external surface of the instrument according to the following steps:
1. Use a soft rag to wipe surface dust outside the instrument and the probe, and do not
scratch the transparent plastic protective screen when cleaning the liquid crystal
displayer.
2. Use a wet soft cloth to clean the instrument, and please cut off the power supply. for
more complete cleaning, please use a hydro-solvent of 75% of isopropanol.
ÿ To avoid damage of the surface of the instrument of probe, please do not use any corrosive
reagent or chemical cleaning reagent.
ÿ Before use, after the power on, please confirm that the instrument is completely dry so as to
prevent short circuit or even personal injury caused by moisture.
105