3400 Operator Manual
3400 Operator Manual
3400 Operator Manual
Operator Manual
KSS - September 2005 - Edition 3
Model 3400 - Table of Contents 1 of 110
Table of Contents
1 Introduction
1.1 Operation Conventions .......................................................................... 10
1.2 Accessories............................................................................................ 11
2 Initialization
2.1 Electrical Connections ........................................................................... 13
2.2 Battery Charging .................................................................................... 15
2.2.1 Using the 3400....................................................................... 15
2.2.2 Using the External Battery Charger ....................................... 16
2.3 Loading Printer Paper ............................................................................ 16
2.4 Getting Started....................................................................................... 17
2.5 Screen Orientation ................................................................................. 19
2.5.1 Overview................................................................................ 19
2.5.2 Sliding Menu .......................................................................... 20
2.5.3 Battery Status ........................................................................ 20
2.6 Quick Start: Taking a Count................................................................... 21
2.7 Main Screen Sliding Menu ..................................................................... 23
3 Configuration
3.1 Setup Overview...................................................................................... 25
3.2 Global Settings....................................................................................... 27
3.2.1 Basic Tab Overview............................................................... 28
3.2.2 Sounds Tab ........................................................................... 30
3.2.3 Users Tab Overview .............................................................. 32
3.2.4 Options and Accessories Tab................................................ 32
3.3 Sample Screen ...................................................................................... 34
3.3.1 General Tab Overview........................................................... 35
3.3.2 Timing Tab Overview............................................................. 40
3.3.3 Environmental Units Tab ....................................................... 42
3.4 Alarm Setup ........................................................................................... 43
3.4.1 Count Tab Overview .............................................................. 45
3.4.2 Environmental Tab Overview................................................. 46
3.5 Group Settings ....................................................................................... 48
KSS - September 2005 - Edition 3
4 Operating Procedures
5 Password Protection
5.1 Password Overview ............................................................................... 71
5.2 Administrator Login ................................................................................ 71
5.3 Enabling Password Protection ............................................................... 72
5.4 Changing a Password............................................................................ 72
5.5 Replacing a Forgotten Password........................................................... 74
Appendix E: Certifications
E.1 Overview ................................................................................................ 97
Manual Overview
WARNING
A warning is used to indicate a condition which, if not met, could cause serious personal injury
and/or death. Do not move beyond a warning until all conditions have been met.
CAUTION:
A caution is used to indicate a condition which, if not met, could cause damage to the
equipment. Do not move beyond a caution until all conditions have been met.
Note:
A note is used to indicate important information or instructions that should be considered
before operating the equipment.
WARNING
KSS - September 2005 - Edition 3
Attempts by untrained personnel to disassemble, alter, modify or adjust the electronics may
result in personal injury and damage to the Model 3400.
WARNING
Use of controls or adjustments, or performance of procedures other than those specified
herein may result in hazardous radiation exposure.
CAUTION:
Observe standard electrostatic discharge (ESD) procedures when servicing or installing
equipment to prevent damage.
Warranty
Hach Ultra Analytics warrants this instrument to be free of defects in materials and
workmanship for a period of two (2) years from the shipping date. If any instrument covered
under this warranty proves defective during this period, Hach Ultra Analytics will, at its option,
either repair the defective product without charge for parts and labor, or provide an equivalent
replacement in exchange for the defective product.
To obtain service under this warranty, the customer must notify the nearest Hach Ultra
Analytics service support center on or before the expiration of the warranty period and follow
their instructions for return of the defective instrument. The customer is responsible for all costs
associated with packaging and transporting the defective unit to the service support center, and
must prepay all shipping charges. Hach Ultra Analytics will pay for return shipping if the
shipment is to a location within the same country as the service support center.
This warranty shall not apply to any defect, failure, or damage caused by improper use or
maintenance or by inadequate maintenance or care. This warranty shall not apply to damage
resulting from attempts by personnel other than Hach Ultra Analytics representatives, or
factory-authorized and trained personnel, to install, repair or service the instrument; to damage
resulting from improper use or connection to incompatible equipment; or to instruments that
have been modified or integrated with other products when the effect of such modification or
integration materially increases the time or difficulty of servicing the instrument.
THIS WARRANTY IS GIVEN BY HACH ULTRA ANALYTICS WITH RESPECT TO THIS
INSTRUMENT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. HACH
ULTRA ANALYTICS AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR NON-CONTRACTUAL PURPOSE.
HACH ULTRA ANALYTICS’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE
PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER
FOR BREACH OF THIS WARRANTY. HACH ULTRA ANALYTICS AND ITS VENDORS WILL
NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES EVEN IF HACH ULTRA ANALYTICS OR ITS VENDORS HAS BEEN GIVEN
ADVANCED NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Revision History
• Edition 1, August 2005. Initial release.
KSS - September 2005 - Edition 3
Acknowledgements
• Hytrel is a trademark of E. I. Du Pont De Nemours & Company Corporation.
• Long Life Laser is a trademark or registered trademark of Hach Ultra Analytics.
• Windows CE is a trademark of Microsoft Corporation.
KSS - September 2005 - Edition 3
1 Introduction
This manual describes use of the 3400 Series Particle Counters, described in Table 1-1. Refer
to “Performance Specifications” on page 87 for a complete list of specifications.
Table 1-1 : 3400 Series Particle Counter Model Numbers
Flow Rate
Model Number Minimum Particle Size Channel (µm)
L/min cubic ft/min
3413 28.3 1 0.3
3415 28.3 1 0.5
3423 50 1.77 0.3
3425 50 1.77 0.5
intake valve
touch screen
power button
battery indicator
KSS - September 2005 - Edition 3
Hach Ultra Analytics particle counters use a laser-diode light source and collection optics for
particle detection. Particles scatter light from the laser diode. The collection optics focus the
light onto a photo diode that converts the bursts of light into electrical pulses. The pulse height
is proportional to the particle size. Pulses are counted and their amplitude is measured for
particle sizing. Results are displayed as particle counts in the specified size range (differential
count mode) or as total particle counts (cumulative count mode). A microprocessor controls all
instrument functions. Count data is displayed as cumulative counts or differential counts.
The particle counter is thoroughly inspected and tested at the factory and is ready for use upon
receipt. When received, inspect the shipping carton for damage. If the carton is damaged,
notify the carrier and save the carton for carrier inspection. Inspect the counter for broken parts,
scratches, dents, or other damage.
• A pop-up box is a warning or note to make sure actions are taken as intended. Touch
OK to continue or Cancel to stop.
1.2 Accessories
Several accessories are available to tailor the counter’s functioning. These accessories can be
ordered from a local Hach Ultra Analytics representative or from the factory by calling either a
local representative or the factory at 800.866.7889 or +1 541.472.6500 during the week from
8:00 a.m. to 5:00 p.m. PT.
Spare batteries Optional1 Spare batteries allow maximum flexibility with portable use.
2 Initialization
From the left side of the 3400, shown in Figure 2-2, connect:
• DC connector for the power supply
• Air velocity probe connector
• RH/Temperature probe connector
All probes are plug-and-play and should be recognized by the 3400 as soon as they are
connected.
printer USB client USB host handle touchscreen power button battery status
Note:
Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400;
contact Customer Service for additional support at 800.866.7889 or 541.472.6500.
4) Connect the power supply to external power. The unit will automatically turn on when
charged batteries and/or AC power is connected.
5) Once the unit is on, charge the batteries until they are fully charged. The battery status
light indicates the level of power in the battery as described in Table 2-1.
Note:
The batteries cannot overcharge.
Batteries will charge when the AC power cord is connected. An external battery charger is also
available for maximum efficiency; refer to “Technical Data” on page 87 for ordering information
and contact a local Hach Ultra Analytics agent or Hach Ultra Analytics at 800.866.7889 or
541.472.6500 to order.
• Batteries charged in the 3400 will take approximately 10 hours to charge completely.
• Batteries charged in an external charger will take approximately 7 hours to charge
completely.
Battery status is indicated by the state of the LED on the front of the unit, as described in
Table 2-1.
CAUTION:
The printer should not be operated without paper as damage may occur to the print head. To
load paper in the printer, follow the instructions in “Loading Printer Paper” on page 16. If the
particle counter must be operated without paper in the printer, be sure to set the Print Mode to
None as described in “Printing Automatically” on page 60.
CAUTION:
Paper used in this printer is temperature-sensitive on one side and must go into the printer as
explained in the instructions. Do not substitute other types of paper.
WARNING
Do not attempt to sample reactive gases (such as hydrogen or oxygen) with the particle
counter. Reactive gases create an explosion hazard in the counter. Contact a local Hach Ultra
Analytics representative or the factory at 800.866.7889 or +1 541.472.6500 for more
information on product implementation.
CAUTION:
Always remove the protective cap from the inlet tube before performing counts or damage to
KSS - September 2005 - Edition 3
2) Remove protective cap from the inlet tube (on top of counter).
WARNING
Exposure to infrared radiation may cause eye injury. Do not look into counter inlet tube while
counter is on.
3) After shipping or storage, the counter batteries may need to be charged. Until the
batteries are completely charged, run the 3400 on AC power by connecting the AC-to-
DC power supply to the DC connector on the left side of the unit and plugging the unit
into AC power.
CAUTION:
Only use the HUA-furnished power supply with the 3400. Substituting a power supply may
damage the unit and will void the product warranty.
4) The counter should turn on once AC power is connected. After a splash screen, the
Main screen displays, as shown in Figure 2-4.
Temperature
Relative humidity
Air velocity
Location Name
Buffer record displayed
Location ID Show next record
Last Location Number Show previous record
Next Location Number
Date
Time
Sliding menu tab Run icon
Battery status
Batteries will charge when the AC power cord is connected. An external battery charger is also
KSS - September 2005 - Edition 3
available for maximum efficiency; refer to “Technical Data” on page 87 for ordering information
and contact a local Hach Ultra Analytics agent or Hach Ultra Analytics at 800.866.7889 or
541.472.6500 to order.
Touch the Battery Status icon to invoke the Battery Diagnostics screen, shown in Figure 6-3
on page 76.
Note:
Alarms are indicated by a red color change and flashing text at the bottom of the touchscreen,
as shown in Figure 2-10. An audible sound may sound if one has been set as described in
“Sounds Tab” on page 30; touch the screen to silence the alarm.
KSS - September 2005 - Edition 3
2) The count will stop automatically. To manually stop counts, press the Stop icon. When
the count is complete, the total number of particles counted during the count time
displays as well as the values for any other parameter measured (relative humidity,
temperature, air velocity).
Fig 2-11 : Stop
Icon
• Counter setup
• Alarm settings
• System diagnostics
Refer to “Setup Overview” on page 25 to adjust the settings such as alarm limits, units of
KSS - September 2005 - Edition 3
3 Configuration
Setup icon
3) When each icon is pressed, it invokes a screen that allows configuration of a certain
aspect of the 3400.
This section describes each subsequent screen in detail; Table 3-1 briefly describes the
functions accessed by each icon.
Only printing information. These screens contain information useful for system diagnostics, including cali-
bration expiration information. These screens are for factory use only. Refer to “Factory Settings Screen”
on page 79.
Factory
KSS - September 2005 - Edition 3
Fig 3-7 : System 3) The Global Settings screen displays. Select the Basic tab, shown in Figure 3-8.
Icon
7) When the correct time is entered, touch the Enter button on the screen.
8) To return to the Counter Setup screen, press the Return icon, shown in Figure 3-11,
on the Basic tab.
9) To return to the Main screen, press the Return icon.
Setting Sounds
1) From the Main screen, select the Setup icon from the sliding menu.
2) From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-13.
3) The Global Settings screen displays. Select the Sounds tab, shown in Figure 3-12.
4) To establish sounds, select the desired sound from the dropdown menus for the four
Fig 3-13 : System
sound fields:
Icon
• User Feedback
• Stop Error
• Warning
• Alarm Limit
Setting Volume
1) From the Main screen, select the Setup icon from the sliding menu.
2) From the Counter Setup screen shown in Figure 3-2 on page 25, select the System
icon, shown in Figure 3-13.
3) The Global Settings screen displays. Select the Sounds tab, shown in Figure 3-12.
4) Set the volume for the 3400 by dragging the Volume slider to the desired level.
Fig 3-15 : System
Icon • Drag the slider to the right to increase volume.
• Drag the slider to the left to decrease volume.
On the Options and Accessories tab, set the parameters listed in Table 3-4.
5) When a new language is selected, the 3400 will prompt a restart, as shown in
Figure 3-20.
6) Press OK to continue and restart the 3400; press X to cancel and restart the 3400
manually later. To return to the Counter Setup screen, press the Return icon. To return
to the Main screen, press the Return icon.
7) Once the 3400 restarts, all screens should appear in the selected language.
On the General tab, set the parameters listed in Table 3-5. The General tab is shown in
Figure 3-23.
2) If the Location ID field was touched, a numeric entry screen appears, as shown in
Figure 3-25.
Location ID
KSS - September 2005 - Edition 3
Location name
5) When the count cycle value is entered, touch the Enter button on the screen.
6) To return to the Counter Setup screen, press the Return icon, shown in Figure 3-29,
on the General tab.
7) To return to the Main screen, press the Return icon.
On the Timing tab, set the parameters listed in Table 3-6. The Timing tab is shown in
Figure 3-34.
KSS - September 2005 - Edition 3
On the Environmental Units tab, set the parameters listed in Table 3-7. The Environmental
Units tab is shown in Figure 3-37. All parameters are set using dropdown menus.
Flow Rate).
Icon
4) A dropdown menu appears, listing the choices available.
5) Highlight a choice to select it.
6) To return to the Counter Setup screen, press the Return icon, shown in Figure 3-39,
on the Environmental Units tab.
7) To return to the Main screen, press the Return icon.
Alarms icon
2) On the Alarm Settings screen, shown in Figure 3-41, alarm functions are organized
into two tabs:
• Count, discussed on page 45
• Environmental, discussed on page 46
Note:
The Alarm Settings screen may also be accessed from the Alarms icon, shown in
Figure 3-42, on the Counter Setup screen (refer to Figure 3-2 on page 25).
4) Press Edit.
5) A numeric entry screen appears, as shown in Figure 3-25 on page 37.
6) Enter the number by touching the screen.
Note:
The maximum limit for any particle size is 9,999,999.
KSS - September 2005 - Edition 3
7) When the correct limit is entered, touch the Enter button on the screen.
8) To return to the Counter Setup screen, press the Return icon, shown in Figure 3-45,
on the Count tab.
9) To return to the Main screen, press the Return icon.
Red screen
Alarm! message
1) From the sliding menu on the Main screen, touch the Alarms icon, shown in
Figure 3-49, to invoke the Alarm Settings screen.
2) Touch the Environmental Alarm tab to invoke the screen shown in Figure 3-48. The
Environmental Alarm tab displays both high and low values for each of the possible
Fig 3-49 : Alarms environmental measurements.
Icon
• The High value indicates the maximum acceptable range of the measurement.
If this alarm is enabled, the alarm will activate when the measurement is higher
than this value.
• The Low value indicates the minimum acceptable range of the measurement. If
this alarm is enabled, the alarm will activate when the measurement drops
below this value.
3) Select the High measurement limit to modify by touching the existing limit.
4) A numeric entry screen appears, similar to the one shown in Figure 3-25.
5) Enter the number by touching the screen.
6) When the correct limit is entered, touch the Enter button on the screen.
KSS - September 2005 - Edition 3
7) Select the Low measurement limit to modify by touching the existing limit.
8) A numeric entry screen appears, similar to the one shown in Figure 3-25.
9) Enter the number by touching the screen.
10) When the correct limit is entered, touch the Enter button on the screen.
11) Enable the alarm by touching the Enabled checkbox next to the limit.
12) To return to the Counter Setup screen, press the Return icon, shown in Figure 3-50,
on the Environmental Alarm tab.
13) Press the Return icon to return to the Main screen.
14) When alarms are triggered, the screen will flash red, the out of range value will be
Fig 3-50 : Return
highlighted on the Main screen, and an Alarm! will appear at the bottom of the screen,
Icon
as shown in Figure 3-51. If an audible alarm has been set as described in “Sounds Tab”
on page 30, a sound will be heard. Touch the screen to silence the alarm.
Red screen
Alarm! message
Refer to “Group Settings” on page 93 for forms that may be copied to keep track of group
setting parameters.
Note:
The Default setting may be used to return the counter to its factory configuration. The default
setting may not be deleted or modified.
1) To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-52.
There are four functions from the Group Setting Management screen:
• Saving current settings as a group, discussed on page 50
• Loading a group setting, discussed on page 51
• Deleting a group setting, discussed on page 52
• Printing the last run from a group setting, discussed on page 52
1) To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 on page 25 and select the Setup icon, shown in Figure 3-55.
2) From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon.
Fig 3-55 : Setup
3) The Group Setting Management screen displays as shown in Figure 3-54.
KSS - September 2005 - Edition 3
Icon
4) In the Defined Groups list, highlight the <New> item by touching it. A keyboard entry
appears, as shown in Figure 3-56.
5) Enter a new group name by touching the keyboard, then press Enter to return to the
Group Setting Management screen.
6) Press Save to save the 3400’s current settings as the new group’s settings.
7) To return to the Counter Setup screen, press the Return icon shown in Figure 3-57.
Fig 3-57 : Return
Icon 8) To return to the Main screen, press the Return icon.
1) To access Group Setting Management, from the Main screen, invoke the sliding
menu shown in Figure 3-1 and select the Setup icon, shown in Figure 3-59.
2) From the Counter Setup screen shown in Figure 3-2 on page 25, select the Group
icon, shown in Figure 3-61.
Fig 3-59 : Setup
3) The Group Setting Management screen displays as shown in Figure 3-54.
Icon
4) Highlight a group then press Delete to delete the group.
5) To return to the Counter Setup screen, press the Return icon on the Group Setting
Management screen.
6) To return to the Main screen, press the Return icon.
4 Operating Procedures
4.1 Logging In
If password protection has been enabled, login to use the 3400. If password protection is not
enabled, no login is required.
Instrument access can be shown by the Lock icon on the Main screen, as described in
Table 4-1.
Indicates that the instrument is unlocked. When password protection is enabled, touching
this icon invokes a login screen.
Indicates that the instrument is locked. When password protection is enabled, touching this
icon logs the previous user out.
Note:
For information on enabling password protection, see “Enabling Password Protection” on
page 72.
1) There are two methods to login, depending on whether the 3400 has first been turned
on or if another user is already logged in:
• To log in after the Model 3400 is initialized, touch the screen. When password
protection is enabled, users must log in to proceed.
• To login to the unit with another user already logged in, touch the Lock icon to
log out the previous user, then touch the Lock icon again to login.
2) A login screen appears like the one shown in Figure 4-1. Touch the Name field to enter
the user name and a keyboard entry screen appears.
KSS - September 2005 - Edition 3
6) Press Enter.
7) If login is successful, the Main screen should appear, similar to the one shown in
Figure 4-6. If login is unsuccessful, a Login Failed message appears and the
password should be re-entered.
CAUTION:
Always remove the protective cap from the inlet tube before performing counts or damage to
the counter will occur.
1) If still in place, remove protective cap from the inlet tube (on top of counter).
WARNING
Exposure to infrared radiation may cause eye injury. Do not look into counter inlet tube while
counter is on.
Table 4-2 describes the function of the buttons on the Main screen.
3) To start counting particles, press the Run icon, shown in Figure 4-4, on the
touchscreen.
Particle counting begins.
Fig 4-4 : Run Icon • The Run icon becomes a Stop icon, shown in Figure 4-5, until the count is
completed.
• The display indicates the counts and any other parameters measured.
Note:
Count or environmental alarms will be indicated by a red color and flashing text at the bottom
of the touchscreen, as well as an audible alarm if one was set as described in “Sounds Tab” on
Fig 4-5 : Stop
Icon page 30.
4) The count stops automatically. To manually stop counts, press the Stop icon. When the
count is complete, the total number of particles counted during the count time displays
as well as the values for any other parameter measured (relative humidity, temperature,
air velocity).
Refer to “Configuration” on page 25 to adjust the settings, such as alarm limits, units of
measurement, and time and date.
2) A numeric entry screen appears. Enter the buffer record number by touching the
screen.
3) When the value is entered, touch the Enter button on the screen.
4) The buffer record information appears on the Main screen.
CAUTION:
When the buffer is full, the buffer must be cleared before further sampling can occur. No
sampling can occur when the buffer is full.
3) A pop-up box will ask for confirmation to clear the buffer, as shown in Figure 4-10.
CAUTION:
Once the buffer is cleared, the records are erased and cannot be retrieved.
CAUTION:
The printer should not be operated without paper as damage may occur to the print head. To
load paper in the printer, follow the instructions in “Loading Printer Paper” on page 16. If the
particle counter must be operated without paper in the printer, be sure to set the Print Mode to
None as described in “Printing Automatically” on page 60.
CAUTION:
Paper used in this printer is temperature-sensitive on one side and must go into the printer as
explained in the instructions.
Table 4-3 describes the function of the fields and buttons on the Print Center screen.
Average Prints an average of all count cycles in each particle size range
Average
Print icon
3) Touch the Print Mode dropdown menu to make a selection. Highlight one of the
following choices, shown in Figure 4-14:
• Cycles: prints results of the first count cycle and multiples of the count cycle programmed.
• None: no counts print automatically
• Alarms: prints results only when a count alarm limit has been exceeded.
• All: prints results after completion of each count cycle.
Note:
Printing or exporting data from the buffer does not clear the buffer.
4) To return to the Main Screen, touch the Return icon. A sample printout is shown in
Figure 4-16
.
Note:
Printing or exporting data from the buffer does not clear the buffer.
4) The data prints. Example printouts are shown in Figure 4-18 and Figure 4-19.
Note:
Printing or exporting data from the buffer does not clear the buffer.
5) To return to the Main Screen, touch the Return icon. KSS - September 2005 - Edition 3
1) Position the counter at the first sample location; turn power on.
2) Assign a location number to the first location using the procedure described in “Taking
Counts” on page 55.
3) Set the number of cycles to be taken at each location using the procedure described in
“Changing the Count Cycles” on page 38.
4) Set the sample time of each cycle using the procedure described in “Setting Sample,
Hold, or Delay Times” on page 41 and enter a hold time.
5) Set mode to Counts/cubic meter as described in “Changing the Count Mode” on
page 40.
6) Set mode to Automatic or Beep (not Manual) as described in “Changing the Run
Mode” on page 39.
7) Press Run to start the first count cycle. The counter stops when samples are complete.
8) Turn the counter off to avoid contamination between samples and move counter to the
next location.
9) Assign a location number to the second location, then press Run to start the first count
cycle.
10) Repeat step 8 and step 9 for the remaining locations.
11) From the sliding menu on the Main screen, touch the Print icon shown in Figure 4-22
to invoke the Print Center screen.
12) Press the icon of the report to print:
• ISO 14644-1
Fig 4-22 : Print
Icon • FS 209E
• BS 5295
13) The report prints. Example printouts are shown in Figure 4-23 and Figure 4-24.
KSS - September 2005 - Edition 3
LOCATION ###
Fig 4-24 : FS 209E Printout
2) From the sliding menu on the Main screen, touch the Export icon shown in Figure
4-25 to invoke the Export Sample Data screen.
Export icon
4) From the Export Sample Data, touch to select either a Comma Separated File or a
Tab Separated File.
5) To change the default file name, touch the File name: field to invoke a keyboard entry
screen. When finished entering the file name, press Enter to return to the Export
Sample Data screen.
6) Touch Export. A status bar appears at the bottom to indicate how much of the
download has been completed, as shown in Figure 4-27.
7) A pop-up box appears to indicate when the export is complete; see Figure 4-28. Press
OK to close this pop-up.
• Wait at least 60 seconds for the 3400 to recognize the USB drive.
• Wait until the USB status light has stopped flashing before removing the
memory device.
The Status value represents several elements of unit status; environmental, count, and
concentration alarms can be indicated with one value. The definitions of these values appear
in Table 4-5.
To determine the alarm conditions present, take the Status value and subtract the largest
possible value listed in Table 4-5. Continue to subtract the largest possible value until the value
equals zero (0).
For example, a Status value of 16512 represents a concentration alam, as shown in Table 4-4.
5 32 System Alarm
6 64 Count Alarm
7 128 Concentration Alarm
8 256 Channel 1 Count Alarm
9 512 Channel 2 Count Alarm
5 Password Protection
2) Touch the Name field to enter the user name and a keyboard entry screen appears.
3) Enter Admin by touching the screen.
4) When the name is entered, touch the Enter button on the screen.
5) Touch the Password field to invoke the keyboard screen.
6) Enter the default administrator password, 123456.
7) Press Enter.
CAUTION:
To ensure system security, change the default administrator password following the procedure
described in “Changing a Password” on page 72.
4) Touch the checkbox next to User Logon Required to enable password protection.
5) To save changes and return to the Counter Setup screen, press the Return icon.
6) To return to the Main screen, select the Return icon.
To change a password:
1) From the login screen shown in Figure 5-2, touch the Change Password button to
invoke the Change Password screen shown in Figure 5-6.
WARNING
If a hazardous chemical comes in contact with the touchscreen, clean the touchscreen
immediately to prevent human contact with the chemical.
CAUTION:
Do not execute the following procedure without receiving upgrade files and explicit direction
from Hach Ultra Analytics.
1) Download a self-extracting .ZIP software update file from Hach Ultra Analytics.
Extract the files to a compatible USB flash drive1. Do not move or delete any files.
KSS - September 2005 - Edition 3
2)
3) Remove the batteries and AC power from the Model 3400.
4) Plug the USB flash drive into the USB port.
5) Apply AC power to the Model 3400.
1. Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400; contact
Customer Service for additional support at 800.866.7889 or 541.472.6500.
6) Upon successful completion of the update process, the instrument will restart.
7) To verify the success of the update process, go to the Diagnostics screen as
described in “Diagnostics Screen Overview” and verify the version number located on
the lower left-hand portion of the screen matches the version on the upgrade.
8) Once a successful update is confirmed, delete the files from the USB drive.
Diagnostics icon
The Diagnostics screen, shown in Figure 6-5, provides information that may be helpful in
troubleshooting connections or calibration data.
KSS - September 2005 - Edition 3
If any diagnostics have failed, the word FAIL appears in red, as shown in Figure 6-6.
Fig 6-7 :
Diagnostics Icon
2) Touch the Print icon.
3) The printout appears similar to the one shown in Figure 6-9.
The Factory Settings screen shows the unit’s identifying information, including model, serial
number, nominal flow, and calibration information. This information is view-only and may not be
edited.
To display the Factory Settings screen:
1) From the sliding menu on the Main screen, touch the Login icon, shown in Figure 6-10
to invoke a Login screen.
Fig 6-10 : Main Screen Sliding Menu, Setup and Login Icons
5) Touch the Factory icon to display the Factory Settings screen, shown in
Figure 6-12.
The Model 3400 must be calibrated every year; the calibration expiration date appears on the
Counter tab. For information on returning the 3400 to the factory for recalibration, follow all
procedures listed in “Return Procedures” on page 85 or contact Hach Ultra Analytics at
800.866.7889 or 541.472.6500.
KSS - September 2005 - Edition 3
From the Test tab, print the factory settings for records, use in troubleshooting, or to assist
Technical Support.
2) The factory settings print similarly to the example shown in Figure 6-16.
SENSOR CALIBRATION
Fig 6-16 : Factory Setup Printout
KSS - September 2005 - Edition 3
WARNING
The following actions must be performed when returning any unit for any reason to prevent
personal injury and/or damage to the unit.
• When the sensor is not in use, or is being stored, verify that the flow cell is clean. Non-
volatile compounds can coat the flow cell windows, making cleaning very difficult.
1 Channel selections can be selected at time of order in a range from 0.3 µm to 25.0 µm; however, 0.3 µm and 25.0 µm cannot be configured
together.
2 Battery life is estimated with the following conditions: Two fully-charged batteries placed in a 50 L/min unit sampling for 20 minutes (1 m3 sample),
print record, a 5 minute hold time (simulating move to new location), then repeating this cycle. The backlight time-out set to two minutes.
1. Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with
the 3400; contact Customer Service for additional support at 800.866.7889 or
541.472.6500.
Item Quantity
Spare Lithium Ion Smart Batteries 2
External battery charger 2
2-roll pack thermal printer paper 2 (4 rolls total) KSS - September 2005 - Edition 3
1. Hach Ultra Analytics recommends using Hach Ultra-furnished USB flash drives with the 3400;
contact Customer Service for additional support at 800.866.7889 or 541.472.6500.
KSS - September 2005 - Edition 3
Main Screen
Counter Setup
General
Timing
Environmental Units
Count
Environmental
Counter
Calibration
Test
Alarm Settings
Count
Environmental
System Diagnostics
Battery Diagnostics
Print Center
Print Cycles
This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.
Group Settings Record Form
Sample Setup
Temperature:
KSS - September 2005 - Edition 3
Air Velocity:
Flow Rate:
This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.
Group Settings Record Form
Alarm Settings
Print Center
Print mode:
Cycles:
KSS - September 2005 - Edition 3
This page of the 3400 Series Particle Counter Operator Manual, PN 701241, may be photocopied and distributed internally by customers for
use in conjunction with a 3400 Series Airborne Particle Counter. Not for proprietary use. All rights reserved, Hach Ultra Analytics.
Model 3400 - Certifications 97 of 110
Appendix E: Certifications
E.1 Overview
This appendix contains the Certificate of Conformity and the WEEE statement.
KSS - September 2005 - Edition 3
F.1 Overview
The Status value represents several elements of unit status; environmental, count, and
concentration alarms can be indicated with one value. The definitions of these values appear
in Table F-1.
To determine the alarm conditions present, take the Status value and subtract the largest
possible value listed in Table F-1. Continue to subtract the largest possible value until the value
equals zero (0).
For more information, refer to “Exported Data Status Bit Masks” on page 69.
Annex
User notes
KSS - September 2005 - Edition 3