Multimetro
Multimetro
Multimetro
GDM-834X Series
USER MANUAL
GW INSTEK PART NO. 82DM-83420E01
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
OPERATION .................................................................. 29
Basic Measurement Overview ................... 31
AC/DC Voltage Measurement ................... 33
AC/DC Current Measurement ................... 39
Resistance Measurement .......................... 43
Diode Test ................................................ 45
Capacitance Measurement ........................ 46
Continuity Test .......................................... 48
Frequency/Period Measurement ............... 51
Temperature Measurement ....................... 53
Dual Measurement Overview .................... 57
Advanced Measurement Overview ............ 62
dBm/dB/W Measurement ......................... 63
Max/Min Measurement ............................. 66
Relative Measurement .............................. 67
Hold Measurement ................................... 69
Compare Measurement ............................. 70
Math Measurement ................................... 72
SYSTEM/DISPLAY CONFIGURATION.............................. 76
View Serial Number .................................. 77
View Version Number ............................... 77
Brightness Settings ................................... 78
Input Resistance Settings ......................... 79
3
GDM-834X Series User Manual
4
Table of Contents
SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you
must follow during operation and storage. Read the following
before any operation to ensure your safety and to keep the
instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
5
GDM-834X Series User Manual
Safety Guidelines
General Make sure that the voltage input level does not
Guideline exceed DC1000V/AC750V.
Make sure the current input level does not
CAUTION exceed 12A.
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that can
lead to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at the source of a
low-voltage installation or at building
installations (Note below).
Do not disassemble the instrument unless you
are qualified as service personnel.
Make sure that the COM terminal to earth is
limited to 500Vpk.
(Note) EN 61010-1:2010 specifies the measurement categories and
their requirements as follows. The GDM-834X falls under category II
600V.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
6
SAFETY INSTRUCTIONS
7
GDM-834X Series User Manual
8
SAFETY INSTRUCTIONS
9
GDM-834X Series User Manual
GETTING STARTED
This chapter describes the GDM-8342 and GDM-8341
multimeters in a nutshell, including accessories, and package
contents, their main features and front / rear panel introduction.
Characteristics ................................................................ 11
Accessories ................................................................ 12
Package Contents....................................................... 13
Appearance ..................................................................... 14
GDM-8342/8341 Front Panel .................................... 14
Display Overview ....................................................... 20
Rear Panel .................................................................. 22
10
GETTING STARTED
Characteristics
The GDM-8342 and GDM-8341 are portable, dual-display digital
multimeters suitable for a wide range of applications, such as
production testing, research, and field verification.
Interface Voltage/Resistance/Diode/Capacitance/
Temperature* input
Current input
USB device port as standard for remote control
USB host* for data logging
Optional GPIB* (factory install)
Calibration port (for service operators only)
Excel Add-In for easy-to-use remote control,
data logging and for saving/recalling setups
11
GDM-834X Series User Manual
Accessories
12
GETTING STARTED
Package Contents
Check the contents before using the instrument.
13
GDM-834X Series User Manual
Appearance
MAX
MAX
COM
1000V
0.5A 750V
POWER
DCI
DC + AC
ACI dB dBm TEMP LOCAL
RANGE
FUSE
T0.5A
MAX
500Vpk COM MAX
12A
Terminal
DCV ACV W / / Hz/P 2ND 250V
Auto
REL# MATH COMP INT/EXT USB STO Enter
SHIFT
REL MX/MN HOLD TRIG MENU
EXIT
14
GETTING STARTED
15
GDM-834X Series User Manual
16
GETTING STARTED
17
GDM-834X Series User Manual
18
GETTING STARTED
SHIFT/EXIT SHIFT
When used as a SHIFT key, it is
EXIT used to access the secondary
functions associated with the
measurement keys.
19
GDM-834X Series User Manual
Display Overview
Primary measurement units
Primary measurement Secondary Display
function icons Secondary
GDM-8342 50000 Counts Dual Measurement Multimeter
measurement units
AC DC 2W 4W AUTO F M S HOLD REL MIN MAX RMT nμF
RATIO UNCAL MEM USBSTO nμF C F S W
C F S W M k W Hz
M k W Hz m m V A
m m V A AC DC 2W 4W AUTOMATH LIMIT ERR SHIFT EXT dBm
Secondary
% dB m 2ND COMP STO RCL FILT RS232 LAN USB GPIB REAR measurement
function icons
20
GETTING STARTED
21
GDM-834X Series User Manual
Rear Panel
LINE RATING
Power Cord
GPIB 50 / 60 Hz
15VA MAX. 120
FUSE LINE
100
220
100VAC
0.125AT
0.063AT
120VAC
220VAC
240VAC
240
Socket
DISCONNECT POWER CORD AND
TEST LEADS BEFORE REPLACING FUSE
CALIBRATION
USE ONLY
WARNING
SER.NO. LABEL
TO AVOID SHOCK,
REMOVE INPUTS BEFORE OPENING.
GPIB
GPIB Port
The GPIB port can be
used for remote control.
This is a factory installed
option for the GDM-8342
only.
Power Cord
120 Accepts the power cord.
Socket
AC 100/120/220/240V
100
220
22
GETTING STARTED
Fuse Socket
120 Holds the main fuse:
100
220
240 100/120 VAC: 0.125AT
220/240 VAC: 0.063AT
USB Device
Type B USB port. This port is used
for remote control.
23
GDM-834X Series User Manual
Set Up
From the base of the handle, gently pull the handle out sideways
and then rotate it to one of the following positions.
Horizontal
position
Tilt stand
position
Carry position
24
GETTING STARTED
Power Up
25
GDM-834X Series User Manual
DCV
26
GETTING STARTED
RANGE
Auto
Enter
SHIFT
EXIT
RANGE
USB STO
MENU
27
GDM-834X Series User Manual
RANGE
Auto
Enter
RANGE
28
OPERATION
OPERATION
Basic Measurement Overview ......................................... 31
Refresh Rate ............................................................... 31
Reading Indicator ....................................................... 31
Automatic/Manual Triggering ................................... 32
29
GDM-834X Series User Manual
30
OPERATION
Refresh Rate
31
GDM-834X Series User Manual
Reading Indicator
Automatic/Manual Triggering
32
OPERATION
33
GDM-834X Series User Manual
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
34
OPERATION
35
GDM-834X Series User Manual
For example:
A C D
+5V
DC 5V Range 0V
-5V
B E
36
OPERATION
PK-PK
PK-PK
PK-PK
PK-PK
PK-PK
PK-PK
37
GDM-834X Series User Manual
Spike >9.0
38
OPERATION
39
GDM-834X Series User Manual
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
Max 12A
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
Max 0.5A
40
OPERATION
41
GDM-834X Series User Manual
For example:
A C D
+5A
DC 5A Range 0V
-5A
B E
42
OPERATION
Resistance Measurement
k W
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
43
GDM-834X Series User Manual
44
OPERATION
Diode Test
The diode test checks the forward bias characteristics of a diode by
running a constant forward bias current of approximately 0.83mA
through the DUT.
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
45
GDM-834X Series User Manual
Capacitance Measurement
The capacitance measurement function checks the capacitance of a
component.
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
46
OPERATION
47
GDM-834X Series User Manual
Continuity Test
The continuity test checks that the resistance in the DUT is low
enough to be considered continuous (of a conductive nature).
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
48
OPERATION
49
GDM-834X Series User Manual
50
OPERATION
Frequency/Period Measurement
The GDM-8342/8341 can be used to measure the frequency or
period of a signal.
k Hz
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
51
GDM-834X Series User Manual
k Hz
Note The 2nd key is only used to toggle the view of the
second display between the menu function (FREQ
or PERIOD) and the voltage range. The voltage
range can actually be set without switching to the
2ND display.
52
OPERATION
Temperature Measurement
The GDM-8342 can measure temperature using a thermocouple. To
measure temperature, the DMM accepts a thermocouple input and
calculates the temperature from the voltage fluctuation. The
thermocouple type and reference junction temperature are also
considered. Temperature measurement is only supported on the
GDM-8342.
Display Temp.
Measurement units Sensor type
S
C
53
GDM-834X Series User Manual
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
2. Go to TEMP on level 1.
3. Go to UNIT on level 2.
54
OPERATION
2. Go to TEMP on level 1.
3. Go to SENSOR on level 2.
55
GDM-834X Series User Manual
2. Go to TEMP on level 1.
3. Go to SIM on level 2.
56
OPERATION
57
GDM-834X Series User Manual
DC AUTO S
m A
V AC AUTO
2ND
58
OPERATION
Do not hold the 2ND key. This will turn the dual
Note measurement mode off.
Turn Off 2nd To turn Off the 2nd measurement, press and
Measurement hold the 2nd key for more than 1 second.
59
GDM-834X Series User Manual
INPUT
V W
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
12A
T0.5A
250V
60
OPERATION
INPUT
V W
Load
CAT 600V
MAX
MAX 1000V
0.5A 750V
MAX
500Vpk COM MAX
FUSE
T0.5A
12A 0~12A
250V
0~0.5A
61
GDM-834X Series User Manual
Basic Measurement
Advanced ACV/ ACI/
Meas. DCV DCI Ω Hz/P TEMP* DIODE CAP
dB ● X X X X X X
dBm ● X X X X X X
Max/Min ● ● ● ● ● X ●
Relative ● ● ● ● ● X ●
Hold ● ● ● ● ● X X
Compare ● ● ● ● ● X ●
Math ● ● ● ● ● X X
62
OPERATION
dBm/dB/W Measurement
dBm/dB Calculation
W= Vreading2/Ref
Where:
Vreading= Input Voltage, ACV or DCV;
Rref= Reference resistance simulating an output
load;
dBmref= Reference dBm value
Measuring dBm/W
63
GDM-834X Series User Manual
DC S
W
dB m
View the result in When the reference resistance is less than 50Ω,
Watts it is possible to calculate the power (in watts). If
the reference resistance is equal to or greater
than 50Ω, then this step can be ignored.
DC S
W
W
dB m
64
OPERATION
Measure dB
dB is defined as [dBm−dBmref]. When the dB measurement is
activated, the DMM calculates the dBm using the reading at the
first moment and stores it as dBmref.
Display dB Voltage
measurement reading
DC S
dB
View the dBm To view the dBm reference value, press the
Reference Value 2ND key.
65
GDM-834X Series User Manual
Max/Min Measurement
Maximum and Minimum measurement function stores the highest
(maximum) or lowest (minimum) reading and shows it on the 1st
display when the 2ND key is pressed.
AC AUTO S MAX
View Max/Min Press the 2ND key to view the Max or Min
Value value.
AC AUTO S MAX
66
OPERATION
Relative Measurement
Relative measurement stores a value, typically the data at that
instant, as the reference. The measurement following the reference
is displayed as the delta between the reference. The reference value
will be cleared upon exit.
Display Relative
Range
value
AC S REL
View Relative Press the 2ND key to view the relative reference
Reference Value value at full scale.
Display Relative
reference value
67
GDM-834X Series User Manual
68
OPERATION
Hold Measurement
The Hold Measurement function retains the current measurement
data and updates it only when it exceeds the set threshold (as a
percentage of the retained value).
AC S HOLD
Set the Hold Use the Up and Down arrow keys to select a
Threshold hold threshold level, as a percentage.
69
GDM-834X Series User Manual
Compare Measurement
Compare measurement checks to see if the measurement data stays
between a specified upper (high) and lower (low) limit.
70
OPERATION
V
COMP
71
GDM-834X Series User Manual
Math Measurement
Measure MX+B
72
OPERATION
decimal point.
DC AUTO S
m V MATH
Measure 1/X
73
GDM-834X Series User Manual
V MATH
Measure Percentage
74
OPERATION
V MATH
75
GDM-834X Series User Manual
SYSTEM/DISPLAY
CONFIGURATION
View Serial Number ........................................................ 77
76
SYSTEM/DISPLAY CONFIGURATION
2. Go to SYSTEM on level 1.
3. Go to S/N on level 2.
Display
2. Go to SYSTEM on level 1.
3. Go to VER on level 2.
Display
77
GDM-834X Series User Manual
Brightness Settings
The display has 5 settable brightness levels.
2. Go to SYSTEM on level 1.
3. Go to LIGHT on level 2.
78
SYSTEM/DISPLAY CONFIGURATION
2. Go to MEAS on level 1.
3. Go to INPUT R on level 2.
79
GDM-834X Series User Manual
2. Go to MEAS on level 1.
3. Go to INJACK on level 2.
80
SYSTEM/DISPLAY CONFIGURATION
Compatibility Settings
2. Go to SYSTEM on level 1.
3. Go to LANG on level 2.
81
GDM-834X Series User Manual
2. Go to SYSTEM on level 1.
3. Go to FACTORY on level 2.
82
USB STORE
USB STORE
The GDM-8342 is able to save/log measurement results to a
USB stick.
Please note that this function is not available for the GDM-8341,
however similar functionality is possible on a PC via remote
control using the Excel Add-In, GDM-834x Excel Addins. See
the GDM-834X Series Excel Add-In manual for details.
83
GDM-834X Series User Manual
CSV Format
84
USB STORE
Filename Format
85
GDM-834X Series User Manual
Operator Mode
86
USB STORE
2. Go to USBSTO on level 1.
3. Go to MODE on level 2.
87
GDM-834X Series User Manual
2. Go to USBSTO on level 1.
3. Go to RECORD on level 2.
88
USB STORE
89
GDM-834X Series User Manual
2ND
90
USB STORE
91
GDM-834X Series User Manual
92
USB STORE
Timer
93
GDM-834X Series User Manual
2. Go to USBSTO on level 1.
3. Go to TIMER on level 2.
94
USB STORE
Date
2. Go to USBSTO on level 1.
3. Go to DATE on level 2.
95
GDM-834X Series User Manual
Save to USB
Procedure 1. Insert a USB stick into the USB Host port on the
front panel.
96
USB STORE
DC S
USBSTO
Procedure 1. Insert a USB stick into the USB Host port on the
front panel.
97
GDM-834X Series User Manual
DC S
USBSTO
98
USB STORE
99
GDM-834X Series User Manual
GW000-02.CSV
OR
100
REMOTE CONTROL
REMOTE CONTROL
This chapter describes basic configuration of IEEE488.2 based
remote control. For a command list, refer to the Command
Overview chapter on page 106.
101
GDM-834X Series User Manual
USB Interface
The USB device port on the rear panel is used for remote control.
The USB port on the DMM will appear as a virtual COM port to a
connected PC. Any terminal program that can communicate via a
serial port can be used for remote control. Before the DMM can be
used for remote control the USB driver included on the User
Manual CD, must first be installed.
Stop bit 1
2. Press MENU.
3. Go to I/O on level 1.
4. Go to USB on level 2.
102
REMOTE CONTROL
GPIB Interface
In addition to the USB port, an optional GPIB port (GDM-8342
only) on the rear panel can be used for remote control.
2. Press MENU.
3. Go to I/O on level 1.
4. Go to GPIB on level 2.
103
GDM-834X Series User Manual
No loop or parallel
connections
104
REMOTE CONTROL
AUTO RMT
k Hz
105
GDM-834X Series User Manual
COMMAND OVERVIEW
The Command overview chapter lists all programming commands
in functional order as well as alphabetical order. The command
syntax section shows you the basic syntax rules you have to apply
when using commands.
Command Syntax
CONFigure
CONFigure:VOLTage:DC :VOLTage
106
COMMAND OVERVIEW
Command types
Example CONFigure:VOLTage:DC
Example CONFigure:RANGe?
107
GDM-834X Series User Manual
Short form
CONF:DIOD
conf:diod
[SENSe:]UNIT?
Both [SENSe:]UNIT? and UNIT? are both valid
forms.
1 2 3
108
COMMAND OVERVIEW
Example conf:volt:dc 1
This will set the measurement
item to DC Voltage and the range
to 5V. There is no 1V range so the
DMM selects the next available
range, 5V.
109
GDM-834X Series User Manual
Return CR+ LF
Message
110
COMMAND OVERVIEW
Command List
Configure Commands (Display 1)
CONFigure:VOLTage:DC..................................... 116
CONFigure:VOLTage:AC ..................................... 116
CONFigure:VOLTage:DCAC ................................ 116
CONFigure:CURRent:DC ..................................... 116
CONFigure:CURRent:AC ..................................... 117
CONFigure:CURRent:DCAC ................................ 117
CONFigure:RESistance ........................................ 117
CONFigure:FREQuency ....................................... 117
CONFigure:PERiod .............................................. 118
CONFigure:CONTinuity ...................................... 118
CONFigure:DIODe .............................................. 118
CONFigure:TEMPerature:TCOuple ..................... 118
CONFigure:CAPacitance ..................................... 118
CONFigure:FUNCtion? ........................................ 118
CONFigure:RANGe? ............................................ 119
CONFigure:AUTO ................................................ 119
CONFigure:AUTO? .............................................. 119
111
GDM-834X Series User Manual
Measure Commands
MEASure:VOLTage:DC? ...................................... 123
MEASure:VOLTage:AC? ....................................... 123
MEASure:VOLTage:DCAC? ................................. 123
MEASure:CURRent:DC? ...................................... 123
MEASure:CURRent:AC? ....................................... 124
MEASure:CURRent:DCAC?.................................. 124
MEASure:RESistance? ......................................... 124
MEASure:FREQuency? ........................................ 124
MEASure:PERiod? ................................................ 124
MEASure:CONTinuity? ........................................ 125
MEASure:DIODe? ................................................ 125
MEASure:TEMPerature:TCOuple? ...................... 125
MEASure2:VOLTage:DC? .................................... 125
MEASure2:VOLTage:AC? ..................................... 125
MEASure2:CURRent:DC? .................................... 126
MEASure2:CURRent:AC? ..................................... 126
MEASure2:RESistance? ....................................... 126
MEASure2:FREQuency? ...................................... 126
MEASure2:PERiod? .............................................. 126
Sense Commands
[SENSe:]TEMPerature:TCOuple:TYPE ................ 127
[SENSe:]TEMPerature:TCOuple:TYPE? ............... 127
[SENSe:]TEMPerature:RJUNction:SIMulated ..... 127
[SENSe:]TEMPerature:RJUNction:SIMulated? .... 127
[SENSe:]DETector:RATE ...................................... 127
[SENSe:]DETector:RATE? .................................... 127
[SENSe:]FREQuency:INPutjack ........................... 128
[SENSe:]FREQuency:INPutjack? ......................... 128
[SENSe:]PERiod:INPutjack .................................. 128
[SENSe:]PERiod:INPutjack? ................................. 128
[SENSe:]CONTinuity:THReshold ........................ 128
[SENSe:]CONTinuity:THReshold? ....................... 128
[SENSe:]UNIT ...................................................... 129
[SENSe:]UNIT? ..................................................... 129
[SENSe:]FUNCtion[1/2] ....................................... 129
[SENSe:]FUNCtion[1/2]? ..................................... 129
112
COMMAND OVERVIEW
Calculate Commands
CALCulate:FUNCtion ........................................... 130
CALCulate:FUNCtion? ......................................... 130
CALCulate:STATe ................................................. 130
CALCulate:STATe? ............................................... 130
CALCulate:MINimum? ........................................ 130
CALCulate:MAXimum? ........................................ 130
CALCulate:HOLD:REFerence .............................. 131
CALCulate:HOLD:REFerence? ............................. 131
CALCulate:REL:REFerence ................................... 131
CALCulate:REL:REFerence? ................................. 131
CALCulate:LIMit:LOWer ...................................... 131
CALCulate:LIMit:LOWer? .................................... 131
CALCulate:LIMit:UPPer ....................................... 131
CALCulate:LIMit:UPPer? ...................................... 132
CALCulate:DB:REFerence .................................... 132
CALCulate:DB:REFerence? .................................. 132
CALCulate:DBM:REFerence ................................. 132
CALCulate:DBM:REFerence? ............................... 132
CALCulate:MATH:MMFactor .............................. 132
CALCulate:MATH:MMFactor?............................. 132
CALCulate:MATH:MBFactor ............................... 133
CALCulate:MATH:MBFactor? .............................. 133
CALCulate:MATH:PERCent ................................. 133
CALCulate:MATH:PERCent? ................................ 133
CALCulate:NULL:OFFSet..................................... 133
CALCulate:NULL:OFFSet? ................................... 133
Trigger Commands
READ?................................................................... 134
VAL1?.................................................................... 134
VAL2?.................................................................... 134
TRIGger:SOURce ................................................. 134
TRIGger:SOURce? ................................................ 134
TRIGger:AUTO ..................................................... 135
TRIGger:AUTO? ................................................... 135
SAMPle:COUNt ................................................... 135
SAMPle:COUNt? .................................................. 135
TRIGger:COUNt ................................................... 135
TRIGger:COUNt? ................................................. 135
113
GDM-834X Series User Manual
System Commands
SYSTem:BEEPer:STATe ....................................... 136
SYSTem:BEEPer:STATe? ...................................... 136
SYSTem:BEEPer:ERRor ........................................ 136
SYSTem:BEEPer:ERRor? ...................................... 136
SYSTem:ERRor? ................................................... 136
SYSTem:VERSion? ............................................... 136
SYSTem:DISPlay .................................................. 137
SYSTem:DISPlay? ................................................ 137
SYSTem:SERial? ................................................... 137
SYSTem:SCPi:MODE ........................................... 137
SYSTem:SCPi:MODE? ......................................... 137
INPut:IMPedance:AUTO ..................................... 137
INPut:IMPedance:AUTO? ................................... 137
Status Commands
STATus:QUEStionable:ENABle ........................... 138
STATus:QUEStionable:ENABle? ......................... 138
STATus:QUEStionable:EVENt? ........................... 138
STATus:PRESet .................................................... 138
Interface Commands
SYSTem:LOCal ..................................................... 138
SYSTem:REMote .................................................. 138
SYSTem:RWLock ................................................. 138
Common Commands
*CLS ..................................................................... 139
*ESE? .................................................................... 139
*ESE ..................................................................... 139
*ESR? ................................................................... 139
*IDN? ................................................................... 139
*OPC? .................................................................. 139
*OPC .................................................................... 140
*PSC? ................................................................... 140
*PSC ..................................................................... 140
*RST ..................................................................... 140
*SRE? ................................................................... 140
*SRE ..................................................................... 140
114
COMMAND OVERVIEW
*STB?.................................................................... 140
*TRG..................................................................... 140
115
GDM-834X Series User Manual
CONFigure Commands
CONFigure:VOLTage:DC
Sets measurement to DC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:VOLT:DC 5
Sets the voltage range to 5 volts.
CONFigure:VOLTage:AC
Sets measurement to AC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:VOLT:AC
Sets the AC range to auto range.
CONFigure:VOLTage:DCAC
Sets measurement to DC+AC Voltage on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:VOLT:DCAC
Sets the DC+AC voltage range to auto range.
CONFigure:CURRent:DC
Sets measurement to DC Current on the first display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:DC 50e-3
Sets the DC current range to 50mA.
116
COMMAND OVERVIEW
CONFigure:CURRent:AC
Sets measurement to AC Current on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:AC 50e-2
Sets the measurement mode to ACI with a 500mA range.
CONFigure:CURRent:DCAC
Sets measurement to DC+AC Current on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CURR:DCAC 50e-2
Sets the measurement mode to DC+AC Current with a 500mA
range.
CONFigure:RESistance
Sets measurement to 2W Resistance on the first display and
specifies range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:RES 50e3
Sets the range to 50kΩ.
CONFigure:FREQuency
Sets measurement to Frequency on the first display and specifies
the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:FREQ MAX
Sets the frequency measurement range to max.
117
GDM-834X Series User Manual
CONFigure:PERiod
Sets measurement to Period on the first display and specifies the
range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:PER
Sets the DMM to period measurement using the previous range.
CONFigure:CONTinuity
Sets measurement to Continuity on the first display.
Parameter: None
CONFigure:DIODe
Sets measurement to Diode on the first display.
Parameter: None
CONFigure:TEMPerature:TCOuple
Sets measurement to Temperature thermocouple (T-CUP) on the
first display.
Parameter: [None] | [Type(J | K | T)]
Example: CONF:TEMP:TCO J
Sets the measurement mode to TCO with a type J sensor.
CONFigure:CAPacitance
Sets measurement to Capacitance on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF:CAP 5E-5
Sets the measurement mode to Capacitance with a 50μF Range.
CONFigure:FUNCtion?
Returns the current function on the first display.
Return parameter: VOLT, VOLT:AC,VOLT:DCAC, CURR,
CURR:AC,CURR:DCAC, RES, FREQ, PER, TEMP, DIOD, CONT,
CAP
118
COMMAND OVERVIEW
CONFigure:RANGe?
Returns the current range on the first display.
Return Parameter:
DCV: 0 .5(500mV), 5(5V), 50(50V), 500(500V), 1000(1000V)
ACV: 0.5(500mV), 5(5V), 50(50V), 500(500V), 750(750V)
ACI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
DCI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
RES: 50E+1(500Ω ) 50E+2(5kΩ ), 50E+3(50kΩ ), 50E+4 (500kΩ ),
50E+5(5MΩ ), 50E+6(50MΩ )
CAP: 5E-9(5nF), 5E-8(50nF), 5E-7(500nF), 5E-6(5μF), 5E-5(50μF)
CONFigure:AUTO
Sets Auto-Range on or off on the first display.
Parameter: ON | OFF
Example: CONF:AUTO ON
CONFigure:AUTO?
Returns the Auto-Range status of the function on the 1st display.
Return Parameter: 0|1, 1=Auto range, 0=Manual range
CONFigure2:VOLTage:DC
Sets measurement to DC Voltage on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:VOLT:DC 5
Sets the voltage range to 5 volts.
119
GDM-834X Series User Manual
CONFigure2:VOLTage:AC
Sets measurement to AC Voltage on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:VOLT:AC
Sets the measurement mode to AC voltage.
CONFigure2:CURRent:DC
Sets measurement to DC Current on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:CURR:DC 50e-3
Sets the DC current range to 50mA on the second display.
CONFigure2:CURRent:AC
Sets measurement to AC Current on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:CURR:AC 50e-2
Sets the measurement mode to ACI with a 500mA range on the
second display.
CONFigure2:RESistance
Sets measurement to 2W Resistance on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:RES 50e3
Sets the range to 50kΩ on the second display.
120
COMMAND OVERVIEW
CONFigure2:FREQuency
Sets measurement to Frequency on the second display and
specifies the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:FREQ MAX
Sets the frequency measurement range to max on the second
display.
CONFigure2:PERiod
Sets measurement to Period on the second display and specifies
the range.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: CONF2:PER
Sets the DMM to period measurement using the previous range.
CONFigure2:OFF
Turns the second display function off.
Parameter: None.
CONFigure2:FUNCtion?
Returns the current function on the second display.
Return parameter: VOLT, VOLT:AC, CURR, CURR:AC, RES,
FREQ, PER, NON
121
GDM-834X Series User Manual
CONFigure2:RANGe?
Returns the range of the current function on the second display.
Return parameter:
DCV: 0 .5(500mV), 5(5V), 50(50V), 500(500V), 1000(1000V)
ACV: 0.5(500mV), 5(5V), 50(50V), 500(500V), 750(750V)
ACI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
DCI: 0.0005(500μA), 0.005 (5mA), 0.05(50mA), 0.5(500mA), 5(5A),
10(10A)
RES: 50E+1(500Ω ) 50E+2(5kΩ ), 50E+3(50kΩ ), 50E+4 (500kΩ ),
50E+5(5MΩ ), 50E+6(50MΩ )
CONFigure2:AUTO
Sets Auto-Range on or off on the 2nd display.
Parameter: ON | OFF
Example: CONF2:AUTO ON
CONFigure2:AUTO?
Returns the Auto-Range status of the function on the 2nd
display.
Return Parameter: 0|1, 1=Auto range, 0=Manual range
122
COMMAND OVERVIEW
Measure Commands
MEASure:VOLTage:DC?
Returns the DC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DC?
>+0.488E-4
Returns the DC voltage measurement as 0.0488 mV.
MEASure:VOLTage:AC?
Returns the AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:AC?
>+0.511E-3
Returns the AC voltage measurement as 0.511 mV.
MEASure:VOLTage:DCAC?
Returns the DC+AC voltage measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:VOLT:DCAC?
>+0.326E-3
Returns the DC+AC voltage measurement as 0.326 mV.
MEASure:CURRent:DC?
Returns the DC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:DC?
>+0.234E-4
Returns the DC current measurement as 0.0234 mA.
123
GDM-834X Series User Manual
MEASure:CURRent:AC?
Returns the AC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:AC?
> +0.387E-2
Returns the AC current measurement as 3.87mA.
MEASure:CURRent:DCAC?
Returns the DC+AC current measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:CURR:DCAC?
>+0.123E-4
Returns the DC+AC current measurement as 0.0123 mA.
MEASure:RESistance?
Returns the 2W resistance measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:RES?
> +1.1937E+3
Returns the 2W measurement as 1.1937kΩ .
MEASure:FREQuency?
Returns the frequency measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:FREQ?
> +2.3708E+2
Returns the frequency (237.08Hz).
MEASure:PERiod?
Returns the period measurement on the first display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS:PER? MAX
Returns the period at the maximum range.
124
COMMAND OVERVIEW
MEASure:CONTinuity?
Returns the continuity measurement on the first display.
Example: MEAS:CONT?
Returns the continuity.
MEASure:DIODe?
Returns the diode measurement on the first display.
Example: MEAS:DIOD?
Returns the diode measurement.
MEASure:TEMPerature:TCOuple?
Returns the temperature for the selected thermocouple type on
the first display.
Parameter:[NONE] | J | K | T
Example: MEAS:TEMP:TCO? J
> +2.50E+1
Returns the temperature.
MEASure2:VOLTage:DC?
Returns the DC voltage measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:VOLT:DC?
>+0.488E-4
Returns the DC voltage measurement as 0.0488 mV.
MEASure2:VOLTage:AC?
Returns the AC voltage measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:VOLT:AC?
>+0.511E-3
Returns the AC voltage measurement as 0.511 mV.
125
GDM-834X Series User Manual
MEASure2:CURRent:DC?
Returns the DC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:CURR:DC?
>+0.234E-4
Returns the DC current measurement as 0.0234 mA.
MEASure2:CURRent:AC?
Returns the AC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:CURR:AC?
> +0.387E-2
Returns the AC current measurement.
MEASure2:RESistance?
Returns the 2W resistance measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:RES?
> +1.1912E+3
Returns the 2W measurement.
MEASure2:FREQuency?
Returns the frequency measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:FREQ?
> +2.3712E+2
Returns the frequency (237.12Hz).
MEASure2:PERiod?
Returns the period measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)]
Example: MEAS2:PER? MAX
Returns the period at the maximum range.
126
COMMAND OVERVIEW
SENSe Commands
[SENSe:]TEMPerature:TCOuple:TYPE
Sets thermocouple type.
Parameter: Type(J | K | T)
Example: SENS:TEMP:TCO:TYPE J
Sets the thermocouple to type J.
[SENSe:]TEMPerature:TCOuple:TYPE?
Returns the thermocouple type.
Return parameter: J, K, T
[SENSe:]TEMPerature:RJUNction:SIMulated
Set temperature simulation value.
Parameter: <NRf>(0.00 ~ 50.00)
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25˚C.
[SENSe:]TEMPerature:RJUNction:SIMulated?
Returns temperature simulation value.
Return parameter: <NR1> (+0000~+5000) ,where +0000=0.00˚C,
+5000=50.00˚C
[SENSe:]DETector:RATE
Sets the detection rate (sample rate)
Parameter: RATE(S | M | F)
Example: SENS:DET:RATE S
Sets the rate to slow (S).
[SENSe:]DETector:RATE?
Returns the sample rate.
Return parameter: SLOW, MID, FAST
127
GDM-834X Series User Manual
[SENSe:]FREQuency:INPutjack
Assigns an input terminal for the frequency function.
Parameter: (0|1|2) 0=volt, 1=500mA, 2=10A
Example: SENS:FREQ:INP 0
Sets the input jack to the Volt input terminal.
[SENSe:]FREQuency:INPutjack?
Returns the assigned input terminal used for the frequency
function.
Return Parameter: VOLT, 500mA, 10A
[SENSe:]PERiod:INPutjack
Assigns an input terminal for the period function.
Parameter: (0|1|2) 0=volt, 1=500mA, 2=10A
Example: SENS:PER:INP 0
Sets the input jack to the Volt input terminal.
[SENSe:]PERiod:INPutjack?
Returns the assigned input terminal used for the period function.
Return Parameter: VOLT, 500mA, 10A
[SENSe:]CONTinuity:THReshold
Sets the continuity threshold in ohms.
Parameter: <NRf> (0 ~ 1000)
Example: SENS:CONT:THR 500
Sets the continuity threshold to 500
[SENSe:]CONTinuity:THReshold?
Returns the continuity threshold.
128
COMMAND OVERVIEW
[SENSe:]UNIT
Sets the temperature unit.
Parameter: C|F
Example: SENS:UNIT C
Sets the temperature unit to ˚C.
[SENSe:]UNIT?
Returns the temperature unit.
[SENSe:]FUNCtion[1/2]
Sets the function for the first or second display.
Parameter:
(display1):"VOLT[:DC]", "VOLT:AC", “VOLT:DCAC”,
"CURR[:DC]", "CURR:AC", “CURR:DCAC”, "RES", "FREQ",
"PER", "TEMP:TCO", "DIOD", "CONT", “CAP”
(display2): "VOLT[:DC]", "VOLT:AC", "CURR[:DC]",
"CURR:AC", "RES", "FREQ", "PER", "NON"
Example: SENS:FUNC1 "VOLT:DC"
Sets the 1st display to the DCV function.
[SENSe:]FUNCtion[1/2]?
Returns the function displayed on the first or second display.
Return parameter:
(display 1): VOLT, VOLT:AC,VOLT:DCAC, CURR,
CURR:AC,CURR:DCAC, RES, FREQ, PER, TEMP:TCO, DIOD,
CONT, CAP
(display 2): VOLT, VOLT:AC, CURR, CURR:AC, RES, FREQ,
PER, NON
129
GDM-834X Series User Manual
CALCulate Commands
CALCulate:FUNCtion
Sets the Advanced function.
Parameter: OFF | MIN | MAX | HOLD | REL | COMP | DB |
DBM | MXB | INV | REF
Example: CALC:FUNC REL
Sets the Advanced function to REL (relative)
CALCulate:FUNCtion?
Returns the current Advanced function.
CALCulate:STATe
Turns the Advanced function on/off.
Parameter: ON|OFF
Example: CALC:STAT OFF
Turns the Advanced function off.
CALCulate:STATe?
Returns the status of the Advanced function.
Return Parameter: 0 | 1, 1=ON, 0=OFF
CALCulate:MINimum?
Returns the minimum value from the Max/Min measurement.
CALCulate:MAXimum?
Returns the maximum value from the Max/Min measurement.
130
COMMAND OVERVIEW
CALCulate:HOLD:REFerence
Sets the percentage threshold for the Hold function.
Parameter: <NRf> (0.01, 0.1, 1, 10)
Example: CALC:HOLD:REF 10
Sets the hold percentage to 10%.
CALCulate:HOLD:REFerence?
Returns the percentage threshold from the Hold function.
CALCulate:REL:REFerence
Sets the reference value for the relative function.
Parameter: <NRf> | MIN | MAX
Example: CALC:REL:REF MAX
Sets the reference value to the maximum allowed.
CALCulate:REL:REFerence?
Returns the reference value from the relative function.
CALCulate:LIMit:LOWer
Sets the lower limit of the compare function.
Para meter: <NRf> | MIN | MAX
Example: CALC:LIM:LOW 1.0
Sets the lower limit to 1.0
CALCulate:LIMit:LOWer?
Returns the lower limit of the compare function.
CALCulate:LIMit:UPPer
Sets the upper limit of the compare function.
Para meter: <NRf> | MIN | MAX
Example: CALC:LIM:UPP 1.0
Sets the upper limit to 1.0
131
GDM-834X Series User Manual
CALCulate:LIMit:UPPer?
Returns the upper limit of the compare function.
CALCulate:DB:REFerence
Sets the reference value for the dB function.
Parameter: <NRf> | MIN | MAX
Example: CALC:DB:REF MAX
Sets the reference voltage for dB measurements to the maximum
allowed.
CALCulate:DB:REFerence?
Returns the reference voltage from the dB function.
CALCulate:DBM:REFerence
Sets the resistance value for the dBm function.
Parameter: <NRf> | MIN | MAX
Example: CALC:DBM:REF MAX
Sets the resistance value for dBm measurements to the maximum
allowed.
CALCulate:DBM:REFerence?
Returns the resistance value from the dBm function.
CALCulate:MATH:MMFactor
Sets the scale factor M for math measurements.
Parameter: <NRf> | MIN | MAX
Example: CALC:MATH:MMF MIN
Sets the scale factor M to the minimum allowed value.
CALCulate:MATH:MMFactor?
Returns the scale factor M used in the math measurement.
132
COMMAND OVERVIEW
CALCulate:MATH:MBFactor
Sets the offset factor B for math measurements.
Parameter: <NRf> | MIN | MAX
Example: CALC:MATH:MBF MIN
Sets the offset factor B to the minimum allowed value.
CALCulate:MATH:MBFactor?
Returns the offset factor B used in the math measurement.
CALCulate:MATH:PERCent
Sets the reference value for the Percent function.
Parameter: <NRf> | MIN | MAX
Example: CALC:MATH:PERC MAX
Sets the reference value for the Percent function to the maximum.
CALCulate:MATH:PERCent?
Returns the reference value setting for the Percent function.
CALCulate:NULL:OFFSet
Sets the reference value for the relative function. This command
is analogous to the CALCulate:REL:REFerence command.
Parameter: <NRf> | MIN | MAX
Example: CALC:NULL:OFFS MAX
Sets the reference value to the maximum allowed.
CALCulate:NULL:OFFSet?
Returns the reference value from the relative function. This
query is analogous to the CALCulate:REL:REFerence? query.
133
GDM-834X Series User Manual
TRIGger Commands
READ?
Returns 1st and 2nd display value.
VAL1?
Returns the 1st display reading
Example: SAMP:COUN 100
VAL1?
>+0.333E-4,V DC
>+0.389E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 1st display.
VAL2?
Returns the 2nd display reading.
Example: SAMP:COUN 100
VAL2?
>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 2nd display.
TRIGger:SOURce
Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.
TRIGger:SOURce?
Returns current trigger source.
134
COMMAND OVERVIEW
TRIGger:AUTO
Turns Trigger Auto mode on/off.
Parameters: ON | OFF
Example: TRIG:AUTO OFF
Turns the Trigger Auto mode off.
TRIGger:AUTO?
Returns the Trigger Auto mode.
Return parameter: 0|1, 0=OFF, 1=ON
SAMPle:COUNt
Sets the number of samples.
Parameter: <NR1>(1 ~ 9999) | MIN | MAX
Example: SAMP:COUN 10
Sets the number of samples to 10.
SAMPle:COUNt?
Returns the number of samples.
Parameter: None | MIN | MAX
TRIGger:COUNt
Sets the number of trigger counts.
Parameter: <NR1>(1 ~ 9999) | MIN | MAX
Example: TRIG:COUN 10
Sets the number of trigger counts to 10.
TRIGger:COUNt?
Returns the number of trigger counts.
Parameter: None | MIN | MAX
135
GDM-834X Series User Manual
SYSTem:BEEPer:STATe
Selects the beeper mode; no beep, beep on fail and beep on pass.
Parameter: <NR1>(0 | 1 | 2) 0=no beep, 2=fail, 1=pass
Example: SYST:BEEP:STAT 0
Turns the beeper off.
SYSTem:BEEPer:STATe?
Returns the beeper mode.
Return parameter: Beep on Pass | Beep on Fail | No Beep
SYSTem:BEEPer:ERRor
Sets the beeper to sound on an SCPI error.
Parameter: ON | OFF
Example: SYST:BEEP:ERR ON
Allows the beeper to sound when an SCPI error occurs.
SYSTem:BEEPer:ERRor?
Returns the beeper error mode.
Return parameter: 0|1, 0=OFF, 1=ON
SYSTem:ERRor?
Returns the current system error, if any.
SYSTem:VERSion?
Returns system version.
Return Parameter: X.XX.
136
COMMAND OVERVIEW
SYSTem:DISPlay
Turns the Display on/off.
Parameter: ON | OFF
Example: SYST:DISP ON
Turns the display on.
SYSTem:DISPlay?
Returns the status of the display
Return parameter: 0|1, 0=OFF, 1=ON
SYSTem:SERial?
Returns the serial number (eight characters/numbers)
SYSTem:SCPi:MODE
Sets the SCPI mode.
Parameter: NORM| COMP
(NORM=Normal, COMP= Compatible to GDM8246)
Example: SYST:SCP:MODE NORM
Sets the SCPI mode to normal.
SYSTem:SCPi:MODE?
Returns the SCPI mode.
Return parameter: NORMAL | COMPATIBLE
INPut:IMPedance:AUTO
Sets the input impedance for DCV mode.
Parameter: ON(10G)|OFF(10M)
Example: INP:IMP:AUTO ON
Turns the Automatic input impedance on.
INPut:IMPedance:AUTO?
Returns the input impedance mode.
Return parameter: <Boolean>(0|1) (0=OFF(10M), 1=ON(10G))
137
GDM-834X Series User Manual
STATus:QUEStionable:ENABle
Set bits in the Questionable Data Enable register.
STATus:QUEStionable:ENABle?
Returns the contents of the Questionable Data Enable register.
STATus:QUEStionable:EVENt?
Returns the contents of the Questionable Data Event register.
STATus:PRESet
Clears the Questionable Data Enable register.
Example: STAT:PRES
Interface Commands
SYSTem:LOCal
Enables local control (front panel control) and disables remote
control.
SYSTem:REMote
Enables remote control and disables local control (front panel
control)
SYSTem:RWLock
Enables remote control and disables local control (front panel
control). This command is analogous to the SYSTem:REMote
command.
138
COMMAND OVERVIEW
*CLS
Clears the Event Status register (Output Queue, Operation Event
Status, Questionable Event Status, Standard Event Status)
*ESE?
Returns the ESER (Event Status Enable Register) contents.
Example: *ESE?
>130
Returns 130. ESER=10000010
*ESE
Sets the ESER contents.
Parameter: <NR1> (0~255)
Example: *ESE 65
Sets the ESER to 01000001
*ESR?
Returns SESR (Standard Event Status Register) contents.
Example: *ESR?
>198
Returns 198. SESR=11000110
*IDN?
Returns the manufacturer, model No., serial number and system
version number.
Example: *IDN?
>GWInstek,GDM8342,00000000,1.0
*OPC?
“1” is placed in the output queue when all the pending
operations are completed.
139
GDM-834X Series User Manual
*OPC
Sets operation complete bit (bit0) in SERS (Standard Event Status
Register) when all pending operations are completed.
*PSC?
Returns power On clear status.
Return parameter: <Boolean>(0|1) 0= don’t clear, 1=clear
*PSC
Clears power On status.
Parameter: <Boolean>(0|1) 0=don’t clear, 1= clear
*RST
Recalls default panel setup.
*SRE?
Returns the SRER (Service Request Enable Register) contents.
*SRE
Sets SRER contents.
Parameter: <NR1>(0~255)
Example: *SRE 7
Sets the SRER to 00000111.
*STB?
Returns the SBR (Status Byte Register) contents.
Example:*STB?
>81
Returns the contents of the SBR as 01010001.
*TRG
Manually triggers the DMM.
140
COMMAND OVERVIEW
For the following command sets, please refer to the status system
diagram on page 148.
STAT: QUES:EVEN?
STAT: QUES: ENAB
STAT: QUES: ENAB?
*ESR?
*ESE
*ESE?
*STB?
*SRE
*SRE?
141
GDM-834X Series User Manual
FAQ
The DMM performance doesn’t match the specifications.
There are a number of reasons why the measured value may not
match the expected values.
1. Ensure that all connections are connected securely and have a
good contact at all times. Poor contacts could result in erroneous
measurements.
2. Ensure that the appropriate input resistance has been set in the
System menu. For 500mv and 5V ranges, the input resistance can be
set to either 10MΩ or 10GΩ.
3. When measuring AC voltage or current, the RMS of the voltage
peak is measured, not the voltage peak. See page 37 for details.
4. The measurement rate settings can have an effect on the accuracy
of the measurement. Slow measurements are more accurate, while
the fast rate is not as accurate.
5. Ensure that an appropriate range setting is used. If a too-large
range is used, the resolution or the measurement may be affected.
142
APPENDIX
APPENDIX
System Menu Tree
GDM-8342 only
115200 OFF
57600 ON 1
38400 :
19200 30
9600
GDM-8342 only
TYPE J 23.00 C
TYPE K F
TYPE T
GDM-8342 only
143
GDM-834X Series User Manual
TEMP Menu
SENSOR: TYPE J
SIM: 23.00
UNIT: C
I/O Menu
USB: BAUD: 115200
GPIB: OFF
USBSTO Menu
MODE: SIMPLE
RECORD:NORMAL
TIMER: 00:00:00
DATE: 13.03.01
144
APPENDIX
Steps 1. Turn the DMM off and take out the power cord.
145
GDM-834X Series User Manual
UT
IN P
VW
V
600 MA
X
T 0V
CA 1 0 00 V
75
X
M AA
0 .5
X
M AA
X M 12
M A pk C O
V
500
E
FUS5A
T0. V
25 0
2 1
146
APPENDIX
UT
IN P
VW
V
600 MA
X
T 0V
CA 1 0 00 V
75
X
M AA
0 .5
X
M A pk C O
M 12
X
M AA 3
V
500
E
FUS5A
T0. V
250
147
GDM-834X Series User Manual
Status system
The diagram below is a description of the status system
For the following command sets, please refer to the diagram above.
STAT: QUES: EVEN?
STAT: QUES: ENAB
STAT: QUES: ENAB?
*ESR?
*ESE
*ESE?
*STB?
*SRE
*SRE?
148
APPENDIX
Specifications
The specifications apply when the DMM is warmed up for at least
30 minutes and operates in slow rate.
Below are the basic conditions required to operate the DMM within
specifications:
Calibration: Yearly
Operating Temperature Specification: 18~28˚C (64.4~82.4˚F)
Relative Humidity: 80% (Non condensing)
Accuracy: ± (% of Reading + Digits)
AC measurements are based on a 50% duty cycle.
The power supply cable must be grounded to ensure accuracy.
All specifications are applicable to the main (1st) display only.
General Specifications
Specification Conditions:
Temperature: 23˚C±5˚C
Humidity: <80%RH, 75%RH for resistance measurement readings greater than
10MΩ.
Operating Environment: (0~50˚C)
Temperature Range: 0~35˚C, Relative Humidity: <80%RH;
>35˚C, Relative Humidity: <70%RH
Indoor use only
Altitude: 2000 meters
Pollution degree 2
Storage Conditions (-10~70˚C)
Temperature Range: 0~35˚C, Relative Humidity: <90%RH;
>35˚C, Relative Humidity: <80%RH
General:
Power Consumption: Max 15VA
Dimensions: 265 mm (W) X 107 mm (H) X 302 mm (D)
Weight: Approximately 2.9 kg
149
GDM-834X Series User Manual
DC Voltage
Accuracy Input
Range Resolution Full Scale (1 year 23˚C ±5˚C) Resistance
10MΩ or
500mV 10μV 510.00
>10GΩ
10MΩ or
5V 100μV 5.1000
0.02%+4 >10GΩ
50V 1mV 51.000 11.1MΩ
500V 10mV 510.00 10.1MΩ
1000V 100mV 1020.0 10MΩ
* When the input value exceeds the full scale of the selected range, the display will
show -OL- (over load) on the display.
* The specifications are guaranteed to an input voltage of 1000V. A beeping alarm
will go off when the input voltage is higher than 1000V.
* Input protection of 1000V peak on all ranges.
DC Current
Accuracy Shunt Burden
Range Resolution Full Scale (1 year 23˚C ±5˚C) Resistance Voltage
500μA 10nA 510.00 0.05%+5 100Ω 0.06V max
5mA 100nA 5.1000 0.05%+4 100Ω 0.6V max
50mA 1μA 51.000 0.05%+4 1Ω 0.14V max
500mA 10μA 510.00 0.10%+4 1Ω 1.4V max
5A 100μA 5.1000 0.25%+5 10mΩ 0.5V max
10 A 1mA 12.000 0.25%+5 10mΩ 0.8V max
* 500μA~500mA range has a 3.6V voltage limit protection and 0.5A fuse
protection. And 10A range has a 12A fuse protection.
* When the input value exceeds the full scale of the selected range, the display will
show -OL- (over load) on the display.
* The specifications are guaranteed to an input of 10A. A beeping alarm will go off
when the input value is higher than 10A.
150
APPENDIX
151
GDM-834X Series User Manual
Resistance
Accuracy
(1 year 23˚C
Resistance Resolution Full Scale Test Current ±5˚C)[2]
500Ω 10mΩ 510.00 0.83mA 0.1%+5 [1]
5kΩ 100mΩ 5.1000 0.83mA 0.1%+3 [1]
50kΩ 1Ω 51.000 83μA 0.1%+3
500kΩ 10Ω 510.00 8.3μA 0.1%+3
5MΩ 100Ω 5.1000 830nA 0.1%+3
50MΩ 1KΩ 51.000 560nA//10MΩ 0.3%+3
[1] Using the REL function. If you don't use the REL function then increase the
error by 0.2Ω.
[2] When measuring resistances greater than 500kΩ, please use shielded test leads
to eliminate the noise interference that may be induced by standard test leads.
* Open circuit voltage approximates 6V max on 500~5MΩ range, approximates
5.5V max on 50MΩ range.
* Input protection of 500V peak on all ranges.
Diode
Accuracy
Range Resolution Full Scale Test Current (1 year 23˚C ±5˚C)
5V 100μV 5.1000 0.83mA 0.05%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.
Continuity
Accuracy
Range Resolution Full Scale Test Current (1 year 23˚C ±5˚C)
5000.0Ω 100mΩ 5100.0 0.83mA 0.1%+5
* Input protection of 500V peak. *Open circuit voltage approximates 6V.
152
APPENDIX
Capacitance
Accuracy
Range Resolution Full Scale Test Current (1 year 23˚C ±5˚C) [1]
5nF: 0.5~1nF [2] 2.0%+20
0.001nF 5.100 8.3μA
5nF: 1~5nF [2] 2.0%+10
50nF: 5~10nF [2] 2.0%+30
0.01nF 51.00 8.3μA
50nF: 10~50nF [2] 2.0%+10
500nF 0.1nF 510.0 83μA
5μF 1nF 5.100 0.56mA 2.0%+4
50μF 10nF 51.00 0.83mA
[1] For the 5nF ~ 50μF range, make sure that the input is greater than 10% of the
range.
[2] Need to use the REL function.
* Input protection of 500V peak on all ranges.
Frequency
Measurement Range Accuracy (1 year 23˚C ±5˚C)
10Hz ~ 500Hz 0.01%+5
500Hz ~ 500kHz 0.01%+3
* AC + DC measurements do not allow frequency measurements.
* Input protection of 1000V peak on all ranges.
153
GDM-834X Series User Manual
Temperature Specifications
Measurement Accuracy (1 year
Sensor Type Range Resolution 23˚C ±5˚C)
J
Thermocouple K -200 ~ +300°C 0.1°C 2 °C
T
* Note: The temperature specifications do not include sensor error.
* Note: This feature is not supported on the GDM-8341.
154
APPENDIX
Dimensions
GDM-8342/GDM-8341
MAX
MAX 1000V
0.5A 750V
107
RANGE MAX
POWER DC + AC
LOCAL 500Vpk COM MAX
DCI ACI dB dBm TEMP FUSE
12A
T0.5A
DCV ACV W / / Hz/P 2ND 250V
Auto
REL# MATH COMP INT/EXT USB STO Enter
SHIFT
REL MX/MN HOLD TRIG MENU
EXIT
RANGE
231 302
265
155
GDM-834X Series User Manual
Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No. 7-1, Jhongsing Rd, Tucheng Dist., New Taipei City 236, Taiwan
156
INDEX
157
GDM-834X Series User Manual
158