MODEL NO:AWD-L / AIW-L / AIW-LS

Recharge You have purchased a quality precision weighing instrument that


1. When to recharge requires handling with care. Read entire contents of this Operation
After turning on the scale or in usage, the figures keep shinning, Manual prior to operating your new instrument.
which means the power of battery is not full. It is time to recharge. If
it needs to use, it may continue about ten hours. Disclaimer Notice
When the message “Lo-bAT” appears, which indicate the power Calibrate your instrument using reference weights of the appropriate
of battery is exhausted. Please turn off the power switch tolerance (class).
immediately and begin to recharge the battery to prolong its life. An instrument can be no more accurate than the standard to which it
2. Recharge Notes has been compared.
It will take 6-8 hours to full the battery. For assistance in the selection of reference weights, please contact
the factory.
Maintenance Caution: Changes or modifications not expressly approved by the
Please take good care of the scale for it’s a precision manufacturer could void the user’s authority to operate this
instrument. equipment.
Place an object on the pan softly in less to destroy load cell Introduction
for the power of moment hit is too heavy. Thank you for choosing one of our instruments. Your instrument is
In order to keep the scale in formal; please recharge the designed and manufactured to the most rigorous standards in order
battery in time. to give you years of service. First, check the contents of the shipping
carton. You should find the following:

*Manual *Instrument *Power cable


Next, follow the instructions for installing your instrument.
Now you are ready to begin using your instrument. To take
advantage of its many features, carefully read your operating
manual. It contains step-by-step procedures, examples, and other
vital information.

Warning: Use of this product in a manner not specified by the


manufacturer may impair any safety protection provided by the
equipment.

-8- -1-
Specification F4-Calibration
1. AWD-L F4-CAL
Capacity 3Kg 6Kg 15Kg 30Kg When it displays F4-CAL
Division-A 0.5g 1g 2g 5g
Press the key P
Division-B 0.2g 0.5g 1g 2g
Division-C 0.1g 0.2g 0.5g 1g It will display CAL 1
Mini. Cap. 10g/4g/2g 20g/10g/4g 40g/20g/10g 100g/40g/20g
Press the key P
Pan size 285 x 240 mm
Power AC: 110V/220V; DC:4V 4Ah It will display Put 1

2.AIW-L Put any STANDARD WEIGHT on the pan. After a little while,then press the
Capacity 60Kg 150Kg 300Kg key
P
Division-A 10g 20g 100g Use the key Pto put into correct weight value.and then use the
Division-B 5g 10g 50g key to confirm.
P
Division-C 2g 5g ----
It will display value of the standard weight.
The scale will finish calibration.
Mini. Cap. 200g/100g/400g 400g/200g/100g 2kg/1kg
Pan size 400 x 500 mm (size can be changed as customers’ request) FC-Protocols
Power AC: 110V/220V; DC:4V 4Ah
0:Yaohua A7 protocol 5:Word or excel protocol
Preparation
1: Yaohua A1 protocol 6:CAS protocol
This product is intended for indoor use.
Select a suitable work area. 2: Yaohua A12 protocol 7: Enter protocol
Work area should be relatively free from drafts and vibrations. 3: Excel Protocol 8: BCD code protocol
Work surface should be level and rigid.
Do not locate near magnetic materials or equipment/instruments 4: Big display protocol 9: NLP50 protocol
which use magnets in their design.
Avoid areas which have variations in room temperatures or have
excessive room temperatures. Room temperatures above 40°C
or 0°C could affect instrument operation and accuracy.

-2- -7-
Programme
Function keys
Turn on the scale, and press the key   .The display
will shows F1-dIU
Captures a new center of zero.
Press key P for entering setting and confirm setting.
Press key F for setting and goes to next step. Reduce gross weight on pan as tare weight.
The programme sequence as follows:
Display Option Parameter
Units convert key
F1-dIU Division 1 2 5 10 20 50
Used for counting function.
F2-PNt Decimal point 0.0 0.00 0.000 0.0000

F3-CAP Capacity If cap is 30kg, 30.000 P print key


F4-CAL Calibration only use for distributor
F Funtion key
0.5 1.0 1.5 2.0
F5-trA Zero-tracking
2.5 3.0 3.5 4.0
0 when ok will alarm
F6-ALr Alarm setting 1 when hi and low will alarm
Operation
2 not alarm Getting Started
0 back light always on Turn the instrument on by pressing the power switch on.
F7-bLy Backlight switch 1 no back light The display will down count from 999999… while the unit is being
2 when it is weighing will on
2400 4800 9600* 19200 updated.
F8-bUd Baud rate (9600,for connect to printer and PC) Allow the unit to warm-up for 5 minutes.
0 not print or not transmit
Taring (zeroing)
1 press key P to print or transmit All models have taring (zeroing) capabilities up to their total
F9-odE Print or transmit mode
2 automatic print or automatic transmit weight capacity.
3 continuous print or continuous transmit
FA-tPE Protocol 0-9, have all protocols in the market 1. Place sample container on pan and then press the
key, and will point to NET on the display.
Fb-dnY Animal weighing set 0 weak 1 hard 3 lock except back to 0
2. Now place sample in its container.
0 off 3. When the scale is stable, the display shows the weight of the
FC-COL Alarm 3-color setting
1 on
sample.
C. Function key
Hold on to press the key F

-6- -3-
1 oFF Only normal weighing function and counting function Counting Function
Count the desired amount sample pieces (10, 20, 50, 100 total
2 dnY Have animal weighing function pieces) and place on the pan.
3 hold Peak hold weighing function
Press the to display total numbers of 10, 20, 50, 100.The
4 Alar Alarm function
number will cycle, press the again to select sample size
5 LP50 Can choose print forms from LP50 printer
C-1 oFF, only have normal weighing function once it is displayed.
C-2 dnY, animal weighing function
In this mode, press the key F, to switch on this function.

C-3 hold, peak hold function


In this mode, press the key F to switch on this function, will have a
hold show on the display. After weighing, the figure will be shown on
the display, till press the key F to return back to next weighing.

C-4 alarm,
In this mode, can choose hi-low rang.
Press the key F, to input low range, then press the key P to confirm;
Press the key F to input hi range, then press the key P to confirm.
***
If you want, OK to give alarm, F6-Alr choose 0
If you want hi-low to give alarm, F6-Alr choose 1
If you no need give any alarm, F6-Alr choose 2
***
If you want 3 colors to work during alarm function,
FC-COL choose 1
If you don’t need 3 colors to work during alarm fucntion,
FC-COL choose 0
C-5 NLP50
In this mode, press the key F to choose needed print forms from
NLP printer.