Fluid Loss of Clay Component of Geosynthetic Clay Liners: Standard Test Method For
Fluid Loss of Clay Component of Geosynthetic Clay Liners: Standard Test Method For
Fluid Loss of Clay Component of Geosynthetic Clay Liners: Standard Test Method For
ε1 NOTE—Designation was changed to dual, units statement in 1.4 and units, where applicable, were corrected editorially
in January 2016.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
D5891/D5891M − 02 (2016)´1
5.6 ASTM Calibration Immersion Thermometer, 0 to 105 6 5.13 Filter Paper, 90-mm, very dense, hardened with
0.5°C (see Specification E1). smooth lint-free surface, must be used.7 These papers have
high wet strength permitting application of high pressure
5.7 Mixer—11 000 6 300 rpm under load with single
during filtration. They also have good resistance to alkalies and
sine-wave impeller approximately 25 mm [1.0 in.] in diameter acids.
(mounted flash side up).5 The impeller shall be replaced when
it weighs a minimum of 5.1 g, from an original weight of about 6. Reagents
5.5 g. New blades will be weighed prior to installation in order
6.1 Purity of Reagents—Unless otherwise indicated, refer-
to ensure conformance to manufacturing criteria. Mixer speed
ences to water shall be understood to mean reagent water
under sample loading shall be determined and documented
conforming to Specification D1193, Type I, II, or III. Such
once every 90 days unless the manufacturer has documented
water is best prepared by distillation or the passage of tap water
objective evidence to extend calibration time.
through an ion-exchange resin.
NOTE 1—Sterling Multimixer Model 9B with 9B29X impeller blades or 6.2 Specification D1193 for reagent water, Type I, II, or III.
equivalent may be obtained from the suppliers given in Footnote 6.
5.8 Mixing Container—Approximate dimensions are 7. Hazards
180 mm [7 in.] deep, 97-mm [313⁄16-in.] inner diameter at top, 7.1 Safety Precautions—Establish appropriate safety and
and 70-mm [23⁄4-in.] inner diameter at bottom.6 health practices for high-pressure equipment prior to use.
NOTE 2—Mixing containers or equivalent may be obtained from the
8. Sampling and Selection
suppliers given in Footnote 5.
8.1 Conduct the sampling in accordance with Test Method
5.9 Timers, 30 min, two interval, mechanical or electrical,
E725.
precision 60.1 min.
5.10 Spatula, flat blade, to dislodge clay mineral clumps 9. Procedure
adhering to the mixing container walls. 9.1 Grind the clay mineral sample to greater than 100 %
5.11 Covered or Sealed Container, of 400- to 600-mL passing a 100 mesh U.S. Standard Sieve, and a minimum of
capacity. 65 % passing a 200-mesh U.S. Standard Sieve with a mortar
and pestle or laboratory hammer mill as required.
5.12 Ambient Temperature/Low-Pressure Filter Press, con- 9.2 Weigh 22.50 6 0.01 g of the whole composite of finely
forming to API RP 131, Section 3.2. This filter press consists ground clay mineral with “as received” moisture, typically 5 to
mainly of a cylindrical cell having an inside diameter of 10 %, onto a weighing paper. If bentonite is removed from a
76.2 mm [3 in.] and a height of at least 64.0 mm [2.5 in.]. This GCL product, the bentonite would be dried to less than 10 %
chamber is made of materials resistant to strongly alkaline moisture prior to weighing.
solutions, and is so fitted that a pressure medium can be
conveniently admitted into and bled from the top. Arrangement 9.3 Measure 350 6 5 mL of reagent water with the 500-mL
is also such that a sheet of 90-mm filter paper can be placed in graduated cylinder and added to the mixing cup. Place the cup
the bottom of the chamber just above a suitable support. The on the mixer, and add the clay mineral slowly over approxi-
filtration area is 4580 6 60 mm2 [7.1 6 0.1 in2]. Below the mately 30 s.
support is a drain tube for discharging the filtrate into a 9.4 After stirring for 5 6 0.5 min, remove the container
graduated cylinder. Sealing is accomplished with gaskets, and from mixer, and scrape its sides with the spatula to dislodge
the entire assembly supported by a stand. A mini-press or any clay clinging to the container wall. Ensure that all of the
half-area press does not directly correlate with the results dislodged clay mineral clinging to the spatula is incorporated
obtained when using the above described standard-sized press. into the suspension.
Pressure can be applied with any nonhazardous fluid medium, 9.5 Replace the container on the mixer, and continue to stir
either gas or liquid. Presses are equipped with pressure for a cumulative total stirring time of 20 6 0.1 min. The
regulators and can be obtained with portable pressure container may need to be removed from the mixer and the sides
cylinders, midget pressure cartridges, or means of utilizing scraped to dislodge any clay clinging to container walls after
hydraulic pressure. another 5 or 10 min of stirring.
NOTE 3—Ambient temperature/low-pressure filter press conforming to 9.6 Age the clay mineral suspension for a minimum of 16 h
API RP 131, Section 3.2, or equivalent, may be obtained from the in a sealed or covered container at ambient temperature.
suppliers given in Footnote 6. Record the initial temperature, final temperature, and actual
hydration aging time.
9.7 After aging the clay mineral suspension, shake vigor-
5
For example, Sterling Multimixer Model 9B with 9B29X impeller blades ously to break its gel strength, and then pour the suspension
available from Fann Instrument Co., P.O. Box 4350, Houston, TX 77210, has been
found suitable for this purpose.
6
For example, Hamilton Beach Mixer Cup No. M110-D, or equivalent, has been
7
found suitable for this purpose. Mixing containers supplied by Fann Instrument Co., For example, Whatman No. 50, S & S No. 576, or equivalent, have been found
P.O. Box 4350, Houston, TX 77210. suitable for this purpose.
2
D5891/D5891M − 02 (2016)´1
into the mixer container. Stir the suspension on the mixer for 11.1.4 Any modifications to the test method or unusual
5 6 0.5 min to completely disperse the clay mineral slurry. observations which may effect the test results,
9.8 Assemble the dry filter cell with filter paper and gaskets, 11.1.5 Calculated fluid loss as millilitres to the nearest
and immediately after remixing the clay mineral slurry, pour it 0.1 mL, and
into the filter cell and complete assembly of the filter cell. Place 11.1.6 Temperature of the slurry at the start and completion
the filter cell in the filter frame and close the relief valve. Place of the test to the nearest 0.5°C.
a 10 mL graduated cylinder under the filter cell drain tube. 12. Precision and Bias
9.9 Set one timer for 7.5 6 0.1 min and the second timer for 12.1 Interlaboratory Test Program—An interlaboratory
30 6 0.1 min. Start both timers and adjust pressure on the fluid study of the test method was run in 1999. The design of the
loss cell to 100 6 2 psi. Starting the timers and adding 100 psi experiment, similar to that of Practice E691. Seven different
pressure should be completed in less than 15 s. Supply pressure clay mineral samples were distributed to seven laboratories.
by compressed air, nitrogen, helium, or carbon dioxide. Three sets of test results were generated for each sample by
9.10 At 7.5 6 0.1 min on the first timer, remove the each of the laboratories.
graduated cylinder and any adhering liquid on the drain tube, 12.2 Test Results—The precision information is given in
and discard. Immediately place a clean dry 10-mL graduated Table 1. The average fluid loss values ranged from 9 to 22 for
cylinder under the drain tube, and collect the fluid for 22.5 6 the seven clay mineral samples tested. However, since the
0.1 min to the end of the second timer. This corrects the fluid statistics were not related to the magnitude of the test result, the
loss value for any initial unpredictable spurt loss from the fluid precision values have been presented in terms of coefficients of
loss cell. Remove the graduated cylinder after the second time varaiation, CV %.
interval and record the volume of fluid collected.
12.3 Bias—The procedure in Test Method D5891/D5891M
10. Calculation for measuring the fluid loss value of the clay mineral compo-
10.1 Calculate the fluid loss in millilitres using Eq 1: nent of geosynthetic clay liners has no bias because the values
of swell index can be defined only in terms of this test method.
Fluid loss volume 5 2 ~ mL filtrate volume for last 22.5 min. interval! mL
(1)
TABLE 1 Test Results
11. Report Statistic ILS Range
Within laboratory repeatability limit, CV %r 1.8 to 4.7 %
11.1 Report the following information: Between laboratory reproducibility limit, CV %R 6 to 18 %
11.1.1 Source of clay mineral, including sample identifica- 95 % confidence limit 5 to 13 %
tion or lot number, Within laboratory repeatability, 2.8 CV %r
95 % confidence limit 11.8 to 51 %
11.1.2 Method of sampling used, Between laboratory reproducibility, 2.8 CV %R
11.1.3 ASTM Test Method number used to perform the test,
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/